Está en la página 1de 63

AD-5000

Weigh Scale Indicator

Installation/Calibration/Operation
Version 4.2a 4/99
Program S12.x

A&D Weighing
1555 Mc Candless Dr.
Milpitas, CA. 95053
408-263-5333
Fax: 408-263-0119
Table of Contents
Introduction .........................................................................................................1
What’s In Each Section ............................................................................................................. 2
WARRANTY INFORMATION.................................................................................................... 2

General Description ............................................................................................3


AD-5000 Display and Keyboard Diagram ................................................................................ 3
The AD-5000 Display ................................................................................................................ 4
The AD-5000 Keyboard............................................................................................................. 4
Numeric Keypad ................................................................................................................. 4
Function Keys ..................................................................................................................... 4
Control Keys........................................................................................................................ 4

Installation and Setup.........................................................................................5


Unpacking the AD-5000 ............................................................................................................ 5
Item check list ..................................................................................................................... 5
Installation Guide....................................................................................................................... 5
Connect AD-5000 to load cell ............................................................................................. 5
Connect AD-5000 to the printer (optional) ......................................................................... 5
Connect AD-5000 to host device (optional) ....................................................................... 6
Apply AC Power to the AD-5000 ........................................................................................ 6
Setup Guide............................................................................................................................... 7
Initialize to Factory Defaults................................................................................................ 7
Setup & "Quick" Calibration Guide...................................................................................... 7
Calibrate the Weight Indicator............................................................................................. 7

Using the AD-5000 ..............................................................................................8


Keyboard Layout ....................................................................................................................... 8
Operation of Function Keys ....................................................................................................... 8
Zero..................................................................................................................................... 8
Tare..................................................................................................................................... 9
Gross/Net ............................................................................................................................ 9
Print..................................................................................................................................... 9
F1 and F2............................................................................................................................ 9
Memory Register Entry ....................................................................................................... 9
Execute a Scale Basic Function ......................................................................................... 9
Display Time & Date .......................................................................................................... 9

Built In Application Programs..........................................................................10


Fill to Setpoint.......................................................................................................................... 11
Checkweigh: Under/Between/Over ......................................................................................... 12
Weigh-in / Weigh-out ............................................................................................................... 13
Axle Weigh – Short Scale........................................................................................................ 15
Axle Weigh – Long Scale, Unattended.................................................................................... 16
ID, Tare, and Total................................................................................................................... 17
Peak Hold Function ................................................................................................................. 18
Set Operation Parameters ................................................................................19
Access Operation Parameters and Parameter Functions....................................................... 19
Operation Parameters Reference Tables................................................................................ 19
Scale Parameters .................................................................................................................... 23
Parameter 1. Load Cell mv/V (3 mv/v)........................................................................ 23
Parameter 2. Digital Filter (9)........................................................................................ 23
Parameter 3. Display Update Rate (1)........................................................................... 23
Parameter 4. Motion Detection Delay (6) ..................................................................... 24
Parameter 5. Motion Detection Band (6) ...................................................................... 24
Parameter 6. Display Blanking During Motion (No Blanking) ....................................... 24
Parameter 7. Zero Tracking Delay (0) .......................................................................... 24
Parameter 8. Zero tracking band (12)........................................................................... 24
Parameter 9. Push Button Zero Percent (100) ............................................................. 24
Parameter 10. Primary Units Type (1 [lb]) .................................................................... 25
Parameter 11. Decimal Point Position (0 [no decimal point]) ....................................... 25
Parameter 12. Count-by (1) .......................................................................................... 25
Parameter 13. Alternate Units Type (2 [kg]) ................................................................. 25
Parameter 14. Alternate Decimal Point (1) ................................................................... 25
Parameter 15. Alternate Count-by (5)........................................................................... 26
Parameter 16. Alt Units Conversion Factor (45,360).................................................... 26
Parameter 17. Full Scale Graduations (10,000) ........................................................... 26
Parameter 18. Overload Graduations (10,200) ........................................................... 26
Parameter 19. Deadload Offset ...................................................................................... 26
Parameter 20. Disable Power-On Zero (1 [disabled]) .................................................... 26
Parameter 21. Deadload Factor ..................................................................................... 27
Parameter 22. Weight Conversion Factor ..................................................................... 27
Serial I/O Ports ........................................................................................................................ 27
Parameter 23 Serial Port 1 Mode (4 [8 data bits, no parity]) ........................................ 27
Parameter 24 Serial Port 1 Baud Rate (1 [9600 baud])................................................ 28
Parameter 25 Serial Port 2 Mode (4 [8 data bits, no parity] ) ....................................... 28
Parameter 26 Serial Port 2 Baud Rate (1 [9600 baud])................................................ 28
Parameter 27 TX2 Control / Station ID (255 [single station]) ..................................... 28
Parameter 28 TX2 Format (2 Condec format) ............................................................ 29
Printer Parameters .................................................................................................................. 30
Parameter 29 Printer Port Select (2 [Serial Port 1]) ..................................................... 30
Parameter 30 Auto Print LF after CR (1 [yes]).............................................................. 30
Parameter 31 End of Print Character (12 [form feed]).................................................. 30
Parameters 32, 33 34, 84, 85, 86 and 87: Print Labels ................................................. 30
Parameters 35, 36, 37, and 38: Print Codes ................................................................. 31
Parameter 39 Enable TTL Remote Control (1 [enabled])............................................. 31
Display Intensity, Battery, Watch Dog Timer........................................................................... 31
Parameter 40 Display Intensity (10) .............................................................................. 31
Parameter 41 Battery Backup (0 [no]) ........................................................................... 31
Parameter 42 Enable Watch Dog Timer (0 [no]) ........................................................... 31
Parameters 43-50 Fixed Registers.................................................................................. 31
Parameters 51-58 Keyboard Event Functions ................................................................. 32

Parameter Functions ........................................................................................33


Parameter Functions ............................................................................................................... 33
Access Parameter Functions .................................................................................................. 33
Function 59 Display Calibration Audit Number. ..................................................................... 34
Function 60 Calibrate Scale - Deadload First ........................................................................ 34
Function 61 Calibrate Scale - Span First............................................................................... 35
Function 62 Calibrate Deadload Only .................................................................................... 35
Function 63 Adjust Gain Calibration....................................................................................... 35
Function 64 Configure Passwords ......................................................................................... 36
Function 65 Configure Print Formats ..................................................................................... 37
Function 66 Set Time and Date.............................................................................................. 41
Function 67 Display Operation Parameters ........................................................................... 41
Function 68 Print Operation Parameters................................................................................ 41
Function 69 Diagnostic Tests ................................................................................................. 41
Function 70 Initialize Operation Parameters to Factory Defaults........................................... 42
Function 71 Configure Event Monitor..................................................................................... 42
Function 72 Scale Basic Function Editor................................................................................ 43
Function 73 Initialize Function Memory.................................................................................. 43
Function 74 Configure Timers ................................................................................................ 43
Parameter 75 Enable Peak Detect........................................................................................ 44
Parameter 76 Debug mode for Event monitors and Scale Basic.......................................... 44
Function 78 Initialize ID memory ............................................................................................ 44
Parameter 79 Enable Multi-point Linearization ...................................................................... 44
Function 80 Calibrate - Multi-point Linearization.................................................................... 44

Diagnostic Tests ...............................................................................................46


Table of Diagnostic Tests........................................................................................................ 46
Diagnostic Test 1: Serial Com Port 1 - Display Input Data .................................................... 46
Diagnostic Test 2: Serial Com Port 1 – Display Errors .......................................................... 46
Diagnostic Test 3: Serial Com Port 1 - Transmit Data ........................................................... 47
Diagnostic Test 4: Serial Com Port 2 - Display Input Data .................................................... 47
Diagnostic Test 5: Serial Com Port 2 – Display Errors .......................................................... 47
Diagnostic Test 6: Serial Com Port 2 - Transmit Data ........................................................... 47
Diagnostic Test 7: Test parallel output ................................................................................... 47
Diagnostic Test 8: Test A/D.................................................................................................... 48
Diagnostic Test 9: Perform internal memory test ................................................................... 48
Diagnostic Test 10: Lamp Test............................................................................................... 48
Diagnostic Test 11: Print the EAROM Configuration Table ................................................... 48
Diagnostic Tests 12 & 13: Loop Back Tests .......................................................................... 48

Troubleshooting................................................................................................49
Problems And Probable Causes ............................................................................................. 49
Error Messages ....................................................................................................................... 50

Hardware and Wiring ........................................................................................51


Load Cell Connector - TB3...................................................................................................... 51
Serial Port 1 Connector - TB 1 ................................................................................................ 52
Serial Port 2 Connector - TB 2 ................................................................................................ 52
TTL I/O Port Connector - TB 4 ............................................................................................... 53
P.C. Board Diagram - Connectors, Jumpers, and Switches ................................................... 54
Sample Cable Drawings .......................................................................................................... 55
Load Cell ........................................................................................................................... 55
IDS 150A/152/160/320...................................................................................................... 55
IDS DT210/IDS 550/IDS 710/PC (Continuous Output Port)............................................. 55
Hewlett Packard Smart Wand HBCR-8xxx series ............................................................ 56
Parallel (Centronix) Printer................................................................................................ 57
Remote Switch Inputs ....................................................................................................... 57

AD-5000 Specifications ....................................................................................58

ASCII CHART .....................................................................................................59


A&D5000 Users Manual version 4.2.a Introduction

Introduction
Congratulations on your purchase and welcome to the AD-5000 User's Manual. This
manual describes the installation, calibration, setup, and operation of the AD-5000 weigh
scale indicator. The AD-5000 is a microprocessor based weight indicator with rugged
design and state-of-the-art technology. It is designed for use in industrial environments
to perform the following functions:
• Analog to Digital conversion (A/D) - converts the analog signal from a strain gauge
"load cell" to a numeric weight.
• Signal Processing - filters electrical and mechanical noise from the load cell signal
using electronic circuits and digital filtering.
• Basic Weighing - push button zero, tare functions (push button and keyboard tare),
display of gross or net weight and scale status, and transmission of weight data to a
printer and a computer.
• Additional Functions - trip level event monitor, programmable "Scale Basic”
functions, TTL level ‘setpoint’ inputs and outputs, and remote control via TTL inputs
or RS232 serial input.

The AD-5000's stainless-steel enclosure is designed for industrial environments and can
be used in wash down applications.
The AD-5000 has two bi-directional serial communication ports that are used for printer
output, bar code input, computer communications, and remote control. The serial ports
support RS232, RS422, RS485, and 20ma Current Loop communication.
The AD-5000 has one parallel TTL I/O port for output to a printer or for setpoint
input/output. The TTL inputs can be used for remote control (ZERO, TARE,
GROSS/NET, and PRINT).
Options available include, analog output, solid state relay board, counting program.

1
A&D5000 Users Manual version 4.2.a Introduction

What’s In Each Section


This manual is organized and divided into separate sections, which describe the
installation, calibration, setup, operation, and testing of the AD-5000. Some Sections,
such as Installation, Setup and Calibration are intended for trained scale technicians
familiar with the installation of computer equipment. Other sections present information
applicable to a user, installer, or system integrator.
The importance of certain information in various sections depends on what features of the
AD-5000 are to be utilized. It is recommended that each installer and user reviews all
sections of the manual and determines what information is useful to his personal needs.
Installation and Setup allow a user to quickly get the AD-5000 in a working manner for
simple use and testing.
Set Operation Parameters and Parameter Functions concentrate upon additional
versatile scale setup parameters and interfacing to peripheral devices such as printers,
computer, PLCs, remote displays/scoreboards, etc.
Diagnostic Tests and Troubleshooting supply fast and simple diagnostic and
troubleshooting instructions for virtually any abnormal situation.
Hardware and Wiring Diagrams provide all external-wiring instructions.

Scale Basic / Event Monitor Programming


The AD-5000 Indicator offers "Scale Basic" programming capabilities which allow you
to customize the operation of the AD-5000 to your application. Scale Basic is designed
to be similar to the common BASIC language used in computer programming.
Event Monitors work with Scale Basic functions to provide event driven programming
capabilities, such as activating setpoint relays. Events trigger actions; Event Monitors
activate Scale Basic functions.
A Scale Basic & Event Monitor Programming Manual may be obtained through your
A&D Dealer

2
A&D5000 Users Manual version 4.2.f General Description

General Description
This section gives you a general description on the operation of the AD-5000's keyboard
and display. The keyboard is used to initiate weighing functions, for data entry,
maintenance, setup, and testing. The numeric display is used for weight and error display
and for data entry. The LED status lights indicate the status o the weigh scale.
Additional information about the keyboard and display is provided in section: Using the
AD-5000.

AD-5000 Display and Keyboard Diagram

3
A&D5000 Users Manual version 4.2.f General Description

The AD-5000 Display


The AD-5000 DISPLAY consists of a six character NUMERIC display and six status
LED’s The NUMERIC display is normally used to display the weight on the scale. The
LED’s display the scale status. When lighted they indicate the following conditions:
GROSS- The number displayed is the gross weight on the scale.
NET - The number displayed is the net weight on the scale.
MOTION- The weight on the scale is changing (not stable).
ZERO- The scale is at "Center of Zero".
lb- The weight displayed is in increments of pounds.
kg- The weight displayed is in increments of kilograms.

The AD-5000 Keyboard


The keyboard is used to initiate weighing functions and for data input and setup of the
AD-5000. It is divided into 3 sections: Numeric Keypad, Function Keys, and Control
keys as shown on the previous page.

Numeric Keypad
The numeric keypad is used for numeric data entry. Alphanumeric entry is performed
using a three-digit code for every alphanumeric character. Some system parameters may
require alphanumeric entry. Instructions for alphanumeric entry are provided where
needed.

Function Keys
The function keys ZERO, TARE, GROSS/NET, UNITS and PRINT are used to perform
the functions of zeroing the scale, taring the scale, switching the weight display between
gross and net display, switching the weight display between LB and KG display, and
printing the weight. The function keys F1 and F2 are programmable using Scale Basic
functions.

Control Keys
The control keys ENTER, and CLEAR are used to control the data entry operations for
ID’s, setpoints and setup.

4
A&D5000 Users Manual version 4.0.a Installation and Setup

Installation and Setup


This section provides information about unpacking, installing, and setup of the AD-5000.
It also directs the installer to the appropriate sections of the manual for hardware and
setup installation.

Unpacking the AD-5000


Installation begins with unpacking the AD-5000. Observe any instructions or cautions,
which appear in the shipping container. Check the items in the shipping container against
the following item check list or the packing list. A&D will not accept responsibility for
shortages against the packing list unless notified within 30 days. All equipment and
accessories are inspected and tested before shipment. Inspect the equipment carefully. If
shipping damage is evident, notify the carrier immediately. Take photographs if
necessary. You are responsible for final claim and negotiations with the carrier. Save the
packing and shipping container in case you need to store, return, or transport the AD-
5000 for any reason.

Item check list


• AD-5000 - Weight Scale Indicator
• Power cord
• This User's Manual
• Capacity labels
• A sheet that describes the capacity label placement
NOTE: The container may include additional items that have been purchased separately.

Installation Guide

Connect AD-5000 to load cell


Connect the load cell to the Load Cell Terminal Block located inside AD-5000’s
enclosure. See the HARDWARE INSTALLATION AND WIRING section for pin
assignments and a sample cable drawing.

Connect AD-5000 to the printer (optional)


The AD-5000 can be connected to a printer using Com Port 1 (TB1) for RS232 or
Current Loop interfaces or by using the TTL I/O Port (TB4) for Centronix compatible
interfaces. See Hardware and Wiring Diagrams for pin assignments and sample cable
drawings. See Set Operation Parameters, and Parameter Functions for other
communication setup parameters.

5
A&D5000 Users Manual version 4.0.a Installation and Setup

Connect AD-5000 to host device (optional)


The Serial Communications Port 2 (TB2) provides a continuous output of scale data.
This is used for connection to a host device such as a computer or to a remote display
(scoreboard). Continuous output transmission can be enabled or disabled and the output
format can be customized to interface with the host device. See Hardware and Wiring
Diagrams for pin assignments and sample cable drawings. See Set Operation
Parameters, and Parameter Functions for other communication setup parameters.

Apply AC Power to the AD-5000


After the load cell and optional equipment has been connected, apply power through the
power cord.
Note: The socket outlet should be installed near the equipment and shall be easily
accessible.

On power-up, the AD-5000 performs the following tests:


1. A/D Conversion Test - the AD-5000 verifies that A/D conversions are taking place.
2. Keyboard Scan - the keyboard is scanned to verify that no keys are engaged.
3. Memory Test - the AD-5000 displays "rTEST" while it performs a memory test on
internal memory. If a problem is encountered during the test, an error message is
displayed.
4. Lamp Test - all segments on the numeric display and all LED's are turned on for 5
seconds.
5. Conversion Display - the AD-5000 displays "Ad xxx", where "xxx" is the conversion
count. This test takes approximately 10 seconds.
6. Stability Check - the AD-5000 displays "C xxxx” where "xxxx" is the raw data
difference in scale readings. The AD-5000 waits until the data difference is within a
stable range.
To exit the above power on test functions, press the CLEAR key during any test. This is
possible only if the AD-5000 has not been locked.
The AD-5000 is now ready for initialization, calibration, and configuration. Continue
with Setup Guide.

6
A&D5000 Users Manual version 4.0.a Installation and Setup

Setup Guide
Setup the AD-5000 in the following order:

Initialize to Factory Defaults


The AD-5000 is initialized at the factory. It must be re-initialized if the program
memory (EPROM) or the parameter memory (EAROM) is replaced. The Initialize
function clears the memory and sets the configuration parameters to their default values.
Use parameter function 70 described in Parameter Functions to initialize the AD-5000.
NOTE: Calibration and setup must be performed after the AD-5000 is initialized.

Setup & "Quick" Calibration Guide


The AD-5000 should be configured BEFORE calibration. Set Operation Parameters
details all of the scale parameters that are available. To quickly get the AD-5000 up and
running, the following Quick Calibration Guide lists only the mandatory parameters that
must be defined. Enter the following scale parameters using the selection outlined in Set
Operation Parameters and Parameter Functions.

Parameter (section) Purpose (Factory Setting)


1 (Set Operation Parameters) Load Cell mv/V (3)
11 (Set Operation Parameters ) Decimal Point (none)
12 (Set Operation Parameters ) Count-by (1)
17 (Set Operation Parameters ) Full Scale Graduations (10,000)
18 (Set Operation Parameters ) Overload Graduations (10,200)
60 (Parameter Functions) Calibrate Scale

Calibrate the Weight Indicator


Use Parameter Function 60 to calibrate the weight indicator.

7
A&D5000 Users Manual version 4.0.a Built In Application Programs

Using the AD-5000


The AD-5000 has seven function keys located at the bottom of the keyboard, a numeric
keypad, and two control keys (ENTER, and CLEAR). This section of the AD-5000
manual describes what the keys are used for and how to use them.

Keyboard Layout

Operation of Function Keys


The function keys are located on the bottom of the AD-5000’s front panel.

Zero
Press the ZERO key to zero the scale. The following conditions must be met for the
ZERO key to function:
• The gross weight is within the configured zero acquisition range: 1% to 100% of full
scale.
• The scale is stable (no motion).

8
A&D5000 Users Manual version 4.0.a Built In Application Programs

Tare
AUTO TARE:
Press the TARE key. The AD-5000 reads the weight on the scale and stores it in the
tare register. The AD-5000 switches to NET mode and displays the net weight.
KEYBOARD TARE:
Use the numeric keys to enter a tare weight. Press the TARE key. The AD-5000
stores the entered weight in the tare register and switches to NET display mode.
CLEAR TARE:
Press the numeric ZERO key. Press the TARE key. The tare weight is cleared to
zero and the AD-5000 switches to the GROSS mode.

Gross/Net
Press the Gross/Net key to alternate between Gross and Net display modes.

Print
Press the PRINT key to print weight data. Page format 1 is printed if the AD-5000 is in
Gross display mode, page format 2 is printed in the Net display mode. See section
Parameter Functions for setup of print formats.

F1 and F2
The F1and F2 keys can be configured to execute any scale basic function (parameters 57
and 58). They are set at the factory to execute the Fill to Setpoint functions.
When the indicator is in data entry mode, the F1 key can be used for entering a minus
sign, the F2 key can be used to enter a decimal point.

Memory Register Entry


Use the numeric keypad to enter the memory register number (1 to 16). Press the
ENTER key. The AD-5000 displays "reg xx" for 1 second and then displays the current
memory register value. Press the CLEAR key if no change is to be made, or enter a new
value and then press the ENTER key. Memory registers are used by the Event monitor to
calculate setpoint values. NOTE: memory registers are set to zero at power up unless the
battery/Time & Date option is installed.

Execute a Scale Basic Function


The AD-5000 provides up to 16 user programmable functions. User programs 1 thru 16
can be executed by entering 21 to 36 respectively. See the "Scale Basic / Event Monitor"
manual for more information.

Display Time & Date


Enter 19 to display the time, press Enter to display the date, press Enter again to return to
Gross/Net display.

9
A&D5000 Users Manual version 4.0.a Built In Application Programs

Built In Application Programs


The following built in programs are available:
1. Fill to Setpoint.
2. Checkweigh: Under/Between/Over.
3. Weigh-in / Weigh-out.
4. Axle Weigh – Short Scale.
5. Axle Weigh – Long Scale.
6. ID Tare and Total.
7. Peak Hold.

Use parameter function 73, “Initialize Function Memory”, to activate a built in program.
Key in the number of the built in program you wish to run, then the enter key to initialize
and store program. The initialize function erases function memory and setpoint monitor
memory, and initializes all parameters that have a password level of 1 or 2. Use
parameter function 73 before configuring parameters with password level of 1 or 2.
Example: CFG 73 = 1 for fill to set point program.
CFG 73 = 4 for axle weigh-short scale.
See Set operation parameters ( page 20 ) for system initializing before starting built in
applications.

10
A&D5000 Users Manual version 4.0.a Built In Application Programs

Fill to Setpoint # 1
Press the F1 key to enter a setpoint. Place an empty container on the scale. Press the F2
key to begin filling. TTL output 1 turns on until the weight on the scale is equal to or
greater than the setpoint.
Installation : Connect TTL output 1 to relay or lamp actuator.
TB4 - Pin 1 TTL output 1
TB4 - Pin 16 Ground (GND)

Setup Parameters:
(see chapter 5: Set Operation Parameters for setup details)
Parameter 43 Minimum scale weight to begin filling. Verifies that a container is on the
scale. Set to 0 if there is no minimum. Default value is 0.
Parameter 44 Minimum fill amount. Set to some small value to insure that a fill amount
has been entered. Default value is 0.
Parameter 45 Maximum fill amount. Maximum fill weight that will ever be used.
Default value is 5,000.
Parameter 46 Scale empty weight. Maximum weight to begin fill. Verifies that an empty
container is on the scale.
Operators Functions

Enter setpoint
Press the F1 key to enter the setpoint. Enter the setpoint weight, then press the ENTER
key.

Start fill operation


Place the container on the scale (if required). Press the F2 key to begin fill operation.

Stop fill operations


Press the Clear key to stop fill operation.

Print Results
Press the Print key to print the Gross, Tare, Net results of the fill operation.

Error Codes
Error 30 = scale not empty. (Over parameter 46)
Error 31 = minimum weight error: container is not on scale. ( Under parameter 43)
Error 32 = tare weight error: tare function failed, scale not stable (motion)
Error 33 = setpoint less than minimum setpoint amount. (Less than parameter 44)
Error 34 = setpoint more than maximum setpoint amount. (Over parameter 45)

11
A&D5000 Users Manual version 4.0.a Built In Application Programs

Checkweigh: Under/Between/Over #2
Monitor the scale for activity. When the scale goes above the ‘empty’ weight, wait for
stable weight, then output an indication of Under, Between (OK), or Over. The cycle
resets when the scale goes below the empty weight.

Installation
Connect TTL outputs to relay or lamp actuators.
Under weight indication TTL 1 TB4 - Pin 1
Between weight indication TTL 2 TB4 - Pin 2
Over weight indication TTL 3 TB4 - Pin 3
Ground GND TB4 - Pin 16

Configure ‘empty’ scale weight. (Parameter 43) default value is 100. This is the
minimum weight the scale must pass before starting check weight. This is also the
reset weight the scale must drop below before resetting.

Operators Functions

Enter setpoint weights


Press the F1 key. The display prompts “SP 1. Use the numeric keyboard to enter
setpoint 1, then press the ENTER key. The display prompts “SP 2”. Use the numeric
keyboard to enter setpoint 2, then press the ENTER key. NOTE: use the F2 key to enter
a decimal point if needed. To keep previous setpoints ENTER after setpoint displays.
If over weight TTL 3 will go high (5 v). Readjust weight and F2 to recalculate, or remove
Container to auto reset.
If under weight TTL 1 will go high (5v). Readjust weight and F2 to recalculate, or
remove container to auto reset.
NOTE: The scale must drop below empty scale weight (parameter 43) before auto
resetting Or the container must be recalulated before continuing.

Recalculate
Press the F2 key to re-check the weight on the scale.

12
A&D5000 Users Manual version 4.0.a Built In Application Programs

Weigh-in / Weigh-out # 3
This program uses the ID memory for weigh-in / weigh-out transactions. An ID record
is created using the Weigh-in function. The Weigh-out function recalls the ID, calculates
Gross/Tare/Net of the transaction, prints the transaction, and adds the net weight to the
ID’s totals register. The ID data is saved for further weigh-in/weigh-out transactions, or
weigh-out only transactions. A sequential transaction number is maintained for printing
purposes.

Weigh-In
Press the F1 key. The display prompts "id". Enter up to 8 digits for the weigh-in ID and
then press the ENTER key. The scale weight is stored in the ID's tare register and print
page 3 is sent to the printer. If the ID has been weighed in but not weighed-out then "Err
20" is displayed. Press the CLEAR key to exit without changing the ID data, press the
ENTER key to enter the new ID data.

Weigh-Out
Press the F2 key. The display prompts "id". Enter up to 8 digits for the weigh-out ID
and then press the ENTER key. The indicator recalls the ID from memory, calculates
the gross, tare and net weights, totals the net weight and sends the weigh-out data (page
2) to the printer. If an ID is entered that is not found in memory the displays "Err 21".
Press the CLEAR key to exit the weigh-out function.
The print formats label “ GROSS” for weigh-in gross transactions, “(IN) TARE” for
weigh-in tare transactions, and “MEM TARE” to designate Memory tare for ID’s
recalled 2 or more times. Once an id has been weighed in and weighed out that id can be
weighed out multiple times without weighing back in every time. The tare weight Is
recalled from memory and used in the gross, tare, net calulations.

Print ID’s, Display ID’s, Clear Totals, Erase ID’s


The ID maintenance functions are accessed by entering ID 0. Press the F1 (or F2) key,
press the numeric 0 key, and then press the ENTER key. The display prompts "rePort".
Press the UNITS (↓) key or the PRINT (↑) key to scroll through the ID file maintenance
menu.
REPORT: Press the ENTER key when the display prompts "rePort". The display then
prompts "id". Enter an ID number to be printed or enter the numeric 0 to print all ID's.
The ID #, TARE weight and NET TOTAL will be sent to serial port #1.
DISPLAY ID’s: Press the ENTER key when the display prompts “diSPLA”. The
display then prompts “id”. Enter an ID number to display the data in its registers, or
enter 0 to scan ID memory. The unit will display the ID#, current TARE weight and the
NET TOTAL of that ID.
CLEAR TOTALS: Press the ENTER key when "CLEAr" is displayed. The display
prompts "id". Enter an ID number to clear or enter the numeric 0 to clear all ID totals.

13
A&D5000 Users Manual version 4.0.a Built In Application Programs

ERASE ID'S: Press the ENTER key when the display prompts "ErASE". The display
then prompts "id". Enter an ID to be erased or enter the numeric 0 to erase all ID's.
REBUILD ID MEMORY: Press the ENTER key when the display prompts “rebuiLd”.
The display then prompts “id”. Enter the numeric 0 then press ENTER to rebuild ID
memory.
COUNT UNUSED ID’S: Press the ENTER key when the display prompts “Count”.
The display prompts “FrExxx” where xxx is the number of unused ID records in
memory.

Set Beginning Sequence Number


Key in the number 9 then enter. The display prompts "reg 9" for 1 second and then
displays the current value. Press the CLEAR key if no change is to be made, or enter a
new value and then press the ENTER key. The next sequence number printed will be the
entered value +1.

Options
• Print key: Set parameter 52 = 4 (execute user function 4) to enable the print key to
print the weight on the scale. The print data is sent to Serial Communications Port 1.
If units other than lb are used, change print label 85 to the correct units. (this option
is not available for printers that use the parallel output port).
• Units key: Set parameter 53 = 127 (execute Units function) if weighing of dual units
is required.

Notes
• The ID file maintenance functions are password level 1 protected. If a password has
been configured (function 64) the operator will be prompted for the password before
the file maintenance menu is displayed.
• ID memory is cleared to zero at power up unless parameter 41 (battery backup) is set
to 1.

14
A&D5000 Users Manual version 4.0.a Built In Application Programs

Axle Weigh – Short Scale # 4


Automatically weigh multiple axles (or containers), print each axle weight, and print axle
total.

Installation
Connect the Red light / Green light relays.
Green light = TTL out 1 TB4 pin 1
Red light = TTL out 2 TB4 pin 2
Optional TTL out 3 TB4 pin 3

Setup Parameters
Configure the following parameters:

Parameter 43, Trip level weight. Determines when an axle is on the scale. The scale must
exceed this weight before the axle weighing begins. Default value is 400.

Parameter 74, 1 Red light delay timer. This provides a delay time before the Red
light output is turned on.

Parameter 74, 2 Green light delay timer. The amount of delay time before the
green light is turned on.

Parameter 74, 3 Finish timer. Amount of time to wait before automatically


activating the finish function.

Enter CFG 74. The AD-5000 will prompt ( tr ). Key in the timer number you plan to set
and enter. The AD-5000 will prompt L for length. Key in the delay time for that timer,
then
enter. The delay times are in tenths of a second. 1 = one tenth of a second. The red
light is # 1, green is # 2 and the finish timer is # 3.

Operators Functions
The unit is capable of operating unattended, with no operator intervention. The indicator
determines when an axle is on the scale by the increase (or decrease) of weight on the
scale.
A finish timer (parameter 74-3) determines when to complete a weigh cycle (print total
and wait for empty scale).
Press the F1 key to manually finish a weigh cycle before finish timer has timed out.
Press the F2 key to abort weigh cycle.
Optional TTL 3 is providing approx (5v) while the scale is busy, and will only drop low
after the scale has finished and dropped below the trip weight.

15
A&D5000 Users Manual version 4.0.a Built In Application Programs

Axle Weigh – Long Scale, Unattended # 5


Automatically weigh multiple axles (or containers), print each axle weight, and print axle
total.

Installation
Connect the Red light / Green light relays.
Green light = TTL out 1 TB4 pin 1
Red light = TTL out 2 TB4 pin 2
Optional TTLout 3 TB4 pin 3

Setup Parameters
Configure the following parameters:

Parameter 43 Trip level weight. Determines when an axle is on the scale. The scale must
exceed this weight before axle weighing begins.

Parameter 74, 1 Red light delay timer. This provides a delay time before the red light
output is turned on.

Parameter 74, 2 Green light delay timer. The amount of delay time before the green
light is turned on.

Parameter 74, 3 Finish timer. Amount of time to wait before automatically activating
the finish function.

Enter CFG 74. The AD-5000 will prompt ( tr ). Key in the timer number you plan to set
and enter. The AD-5000 will prompt L for length. Key in the delay time for that
timer,then
enter. The delay times are in tenths of a second. 1 = one tenth of a second. The red
light is # 1, green is # 2 and the finish timer is # 3.

Operators Functions
The unit is capable of operating unattended, with no operator intervention. The indicator
determines when an axle is on the scale by the increase of weight on the scale.
A finish timer (parameter 74-3) determines when to complete a weigh cycle (print total
and wait for empty scale).
Press the F1 key to manually finish a weigh cycle before finish timer has timed out.
Press the F2 key to abort weigh cycle.
Optional TTL 3 is providing approx (5v) while the scale is busy, and will only drop low
after the scale has finished and dropped below the trip weight.

16
A&D5000 Users Manual version 4.0.a Built In Application Programs

ID, Tare, and Total # 6


This program uses the ID memory to store tare weights and totals.

Enter ID Tare Data


Press the F1 key. The display prompts "id". Enter up to 8 digits for ID and then press
the ENTER key. The display prompts “tArE”. Key in the tare weight and enter.

Recall ID and Tare


Press the F2 key. The display prompts "id". Enter up to 8 digits for the weigh-out ID
and then press the ENTER key. The indicator recalls the ID from memory and copies the
ID tare to the tare register.

Print and Total ID


To Recall an ID press the F2 key and the AD-5000 will prompt “id”. Key in the ID# and
then ENTER. The NET indicator light will light up on keypad, then press the Print key.
The SEQUENCE number, ID number, Gross, Tare, and Net are printed. The Net weight
is added to the ID’s totals register.

Print ID’s, Display ID’s, Clear Totals, Erase ID’s


The ID maintenance functions are accessed by entering ID 0. Press the F1 (or F2) key,
press the numeric 0 key, and then press the ENTER key. The display prompts "rePort".
Press the UNITS (↓) key or the PRINT (↑) key to scroll through the ID file maintenance
menu.
REPORT: Press the ENTER key when the display prompts "rePort". The display then
prompts "id". Enter an ID number to be printed or enter the numeric 0 to print all ID's.
The ID#, TARE, and NET totals will be sent to serial port # 1.
DISPLAY ID’s: Press the ENTER key when the display prompts “diSPLA”. The
display then prompts “id”. Enter an ID number to display the data in its registers, or
enter 0 to scan ID memory. The unit will display the ID#, TARE weight, NET TOTAL.
CLEAR TOTALS: Press the ENTER key when "CLEAr" is displayed. The display
prompts "id". Enter an ID number to clear or enter the numeric 0 to clear all ID totals.
ERASE ID'S: Press the ENTER key when the display prompts "ErASE". The display
then prompts "id". Enter an ID to be erased or enter the numeric 0 to erase all ID's.
REBUILD ID MEMORY: Press the ENTER key when the display prompts “rebuiLd”.
The display then prompts “id”. Enter the numeric 0 then press ENTER to rebuild ID
memory.
COUNT UNUSED ID’S: Press the ENTER key when the display prompts “Count”. The
display prompts “FrExxx” where xxx is the number of unused ID records in memory.

17
A&D5000 Users Manual version 4.0.a Built In Application Programs

Set Beginning Sequence Number


Use the numeric keypad to enter number 9. The display prompts "reg 9" for 1 second and
then displays the current value. Press the CLEAR key if no change is to be made, or
enter a new value and then press the ENTER key. The next sequence number printed
will be the entered value +1.

Notes
• The ID file maintenance functions are password level 1 protected. If a password has
been configured (function 64) the operator will be prompted for the password before
the file maintenance menu is displayed.
• ID memory is cleared to zero at power up unless parameter 41 (battery back up) is set
to 1.

Peak Hold Function # 7


The peak hold register maintains the peak weight detected by the weight indicator.

Enable Peak Hold


Access configuration parameter 75. Enter a 1 to enable peak hold, enter a 0 to disable
peak hold.

Clear Peak Hold Register


Press the CLEAR key to reset the peak hold register to 0.
Use the Scale Basic built in function Clear Peak [143] to reset the peak hold register.

Display Peak Weight


Use the GROSS/NET key to scroll through the display options as follows:
Initial mode: Display gross weight. Gross LED on.
First GROSS/NET press: Display net weight. Net LED on.
Next GROSS/NET press: Display gross peak weight. Gross LED flashing.
Next GROSS/NET press: Display net peak weight. Net LED flashing.
Next GROSS/NET press: Display gross weight. Gross LED on.
Etc.

Print Peak Weight


Use function 65 (Configure Print Formats) to format the print output using:
Print Item 19: print Gross Peak weight.
Print Item 20: print Net Peak weight.

18
A&D5000 Users Manual version 4.0.a Set Operation Parameters

Set Operation Parameters


This section describes the operation parameters of the AD-5000 and how their settings
affect the operation of the weight indicator.

Access Operation Parameters and Parameter Functions


The parameters are accessed by holding the CLEAR key down and then pressing the
ENTER key. The numeric display will prompt “CFG xx” where xx is the currently
selected parameter. Enter the parameter number to be modified and then press the
ENTER key. NOTE: the UNITS key (↓) and the PRINT key (↑) can be used to scroll
the CFG parameter number. Press the UNITS key to increment the CFG number, press
the PRINT key to decrement the CFG number.
To exit configuration and store changes, press the CLEAR key when “CFG xx” is being
displayed.
To exit configuration and NOT store changes, enter 999 when “CFG xx” is being
displayed. The display prompts “Abort”. Press the ENTER key to abort (exit without
saving changes) or press the CLEAR key if you do not want to abort.

Operation Parameters Reference Tables


The reference tables provide an overview of the operation parameters. The sections
following the reference tables describe the setup parameters in detail and the procedures
for modifying them. Use the ‘Field Settings’ column of the reference tables to document
any changes to the setup parameters.
Reference Table column heading descriptions:
• No.: Use this number to access the parameter to be modified. For example to modify
the Digital Filter enter CFG 2.
• Parameter Name: The name of the parameter.
• Default: The value set at the factory. The Parameter Function “Initialize Operation
Parameters to Factory Defaults” sets the parameter to this value.
• Max Value: The maximum value for numeric parameters, the maximum number of
characters for alphanumeric parameters.
• Pas: The AD-5000 has 4 password levels. Level 1 requires entering password-1
before accessing the parameter. Level 2 requires password-2. Level 3 requires
password-2 and the Lock Switch to be in the unlocked position. Level 4 inhibits Scale
Basic functions from being uploaded to a PC.
• Field Setting: Record any changes to the parameters in the Field setting column.

19
A&D5000 Users Manual version 4.0.a Set Operation Parameters

Scale Parameters
No Parameter Name Default Max Value Pas Field Setting
1 Loadcell mv/v 3 4 3
2 Digital filter 9 18 3
3 Display Update Rate 1 200 3
4 Motion delay time 6 250 3
5 Motion band 6 250 3
6 Motion blanks display 0 (no) 1 3
7 Zero tracking delay 0 250 3
8 Zero tracking band 12 250 3
9 Push button zero percent 100 100 3
10 Primary units type 1 (LB) 6(oz) 3
11 Decimal point position 0 4 3
12 Count-by 1 100 3
13 Alternate units type 2 (kg) 6(oz) 3
14 Alt decimal point 1 4 3
15 Alternate count-by 5 100 3
16 Alt conversion factor 45,360 3
17 Full Scale display digits 10,000 120,000 3
18 Over Weight DD 10,200 120,200 3
19 Deadload offset 3
20 Disable Power-On Zero 1 1 3
21 Deadload Factor 3
22 Weight conversion fact. 3
I/O Port Parameters
No Parameter Name Default Max Value Pas Field Setting
23 Serial port 1 mode 4(8N1) 6 2
24 Serial port 1 baud rate 1(9600) 6 2
25 Serial port 2 mode 4(8N1) 6 2
26 Serial port 2 baud rate 1(9600) 6 2
27 TX2 control / Station Id 255 255 1
28 TX2 Data Format 1 (AND) 4 2
29 Printer port select 2 (Com1) 2 2
30 Autoprint LF after CR 1 (yes) 1 2
31 End Of Print character 12 (ff.) 255 2
32 print label 32 16 char 2
33 print label 33 16 char 2
34 print label 34 16 char 2
84 print label 84 30 char 2
85 print label 85 30 char 2
86 print label 86 30 char 2
87 print label 87 30 char 2

20
A&D5000 Users Manual version 4.0.a Set Operation Parameters

88 print label 88 30 char 2


89 print label 89 30 char 2
90 print label 90 30 char 2
35 print code 1 3 codes 2
36 print code 2 3 codes 2
37 print code 3 3 codes 2
38 print code 4 3 codes 2
39 TTL remote control 1 (yes) 1 2
Display Intensity, Battery, Watchdog Timer
No Parameter Name Default Max Value Pas Field Setting
40 display intensity 10 15 2
41 battery backup 0 (no) 1 2
42 Enable watch dog timer 1 (yes) 1 1
Fixed Setpoints
No Parameter Name Default Max Value Pas Field Setting
43 fixed setpoint 43 0 1
44 fixed setpoint 44 0 1
45 fixed setpoint 45 0 1
46 fixed setpoint 46 0 1
47 fixed setpoint 47 0 1
48 fixed setpoint 48 0 1
49 fixed setpoint 49 0 1
50 fixed setpoint 50 0 1
Keyboard Events
No Parameter Name Default Max Value Pas Field Setting
51 Power on start 0 255 2
52 PRINT key 136 255 2
53 UNITS key 127 255 2
54 Gross/Net key 128 255 2
55 TARE key 129 255 2
56 ZERO key 130 255 2
57 F1 key 138 255 2
58 F2 key 139 255 2

21
A&D5000 Users Manual version 4.0.a Set Operation Parameters

Parameter Functions
Fn. No Function Name Password Level
59 Display Calibration Audit Number 0
60 Calibrate Scale - Deadload first 3
61 Calibrate Scale - span first 3
62 Calibrate Deadload Only 3
63 Adjust Gain Calibration 3
64 Configure Passwords 2
65 Configure Print Formats 2
66 Set Time and Date 2
67 Display Operation Parameters 0
68 Print Operation Parameters 0
69 Diagnostic Tests 0
70 Initialize Operation Parameters to factory defaults. 3
71 Configure Event Monitor 2
72 Function Editor 2
73 Initialize Function Memory 2
74 Set Debug mode for Event monitors and Scale basic fns. 2
75 Enable Peak Detect (optional) 2
76 Set debug mode for event monitors and scale basic fns. 2
78 Erase Id Memory 1
79 Enable Multi-point linearization (default = disabled [0]) 3
80 Calibrate using Multi-point linearization 3
81 Analog output span (default = 10,000) 1
82 Analog output offset (default = 0) 1
83 Analog output register (default = 0 [disabled]) 1

22
A&D5000 Users Manual version 4.0.a Set Operation Parameters

Scale Parameters
The scale parameters configure the AD-5000 for the weigh platform or load cell that it is
connected to. The preceding table titled “SCALE PARAMETERS” lists the scale
parameters and their factory settings. Use the “Field setting” column to record any
changes made to the factory default values.
Many of the following parameters are entered in terms of “graduations”. The full-scale
graduations can be found by dividing the full-scale weight by the count-by weight. For
example, if the scale is configured for 500.00kg x 0.05kg, the full-scale graduations is
500.00/0.05 = 10,000 graduations to full scale.
The numbers in parenthesis are the default factory settings.

Parameter 1. Load Cell mv/V (3 mv/v)


The AD-5000 can be used with load cell outputs of 1 mv/V, 2 mv/V, or 3 mv/V. Enter a
1, 2, or 3 for the load cell type used in your application.
Setting A/D gain
0 200
1 mV/v 100
2 mV/v 100
3 mV/v 50
4 25
Parameter 2. Digital Filter (9)
The digital filter removes scale vibrations and other noise signals from the weight
display. The filter range is between 0 and 18. A setting of 0 provides no filtering; a
setting of 18 provides maximum filtering. The higher the filters number the longer the
settle time of the weight display. For example:
Filter Setting Settle Time
3 0.4 sec
6 0.8 sec
9 1.75 sec
12 3.5 sec
15 7.0 sec
Use the smallest filter setting that provides satisfactory display stability.

Parameter 3. Display Update Rate (1)


Enter the time between display updates (in tenths of a second). An update rate of 1
causes a 0. 1-second delay between display updates, a rate of 200 causes a 20 second
delay between updates. NOTE: motion detection is calculated using the weight values of
the display rate. If the update rate is 50, the motion calculation is based on weight
readings sampled at 0.5sec intervals.

23
A&D5000 Users Manual version 4.0.a Set Operation Parameters

Parameter 4. Motion Detection Delay (6)


Motion detection delay is used by the motion detection function to determine if the scale
is in motion (not stable). Enter the amount of time (in tenths of a second) that the scale
must be stable for motion status to be false (scale status = stable).
Example: 0 = disables motion detection.
2 = 0.2 second delay before motion = false (stable = true).
15 = 1.5 second delay.
30 = 3.0 second delay.
Parameter 5. Motion Detection Band (6)
The motion band is used by the MOTION DETECT function to determine if the scale is
in motion (not stable). If the amount of change between display updates (parameter 2) is
greater than the motion band, a MOTION status is generated (unstable condition). The
motion band is entered in units of ¼ dd, i.e.. For a motion band of:
½ display division enter 2 ½ x 4 = 2.
2 display divisions enter 8 2 x 4 = 8.
4.5 display divisions enter 18 4.5 x 4 = 18.

Parameter 6. Display Blanking During Motion (No Blanking)


When display blanking is enabled, the display is blanked (turned off) when the scale is in
motion (not stable). Enter a 1 (yes) to enable display blanking, enter a 0 (no) to disable
display blanking.

Parameter 7. Zero Tracking Delay (0)


The automatic zero tracking function removes small deviations from zero. The AD-5000
adjusts the zero setting by approximately 1/30th of a display digit per delay interval. The
delay interval is the number effective A/D conversions between adjustments.

Parameter 8. Zero tracking band (12)


The zero tracking band is the maximum displacement from zero wherein zero tracking is
active. In order for zero tracking to be active, the following conditions must be met:
• Scale weight is within Zero Track Band.
• Scale is stable (no motion)
• Zero Track Band is greater than 0 (zero).
The zero tracking band is entered in units of ¼ dd, i.e.. For a band of:
½ display division enter 2 ½ x 4 = 2.
2 display divisions enter 8 2 x 4 = 8.
4.5 display divisions enter 18 4.5 x 4 = 18
Parameter 9. Push Button Zero Percent (100)
Enter the percent of full-scale capacity that can be acquired as the zero point with the
ZERO key. 1 = 1%, 2 = 2%, ... 100 = 100%.

24
A&D5000 Users Manual version 4.0.a Set Operation Parameters

Parameter 10. Primary Units Type (1 [LB])


The scale is calibrated in the primary units. The UNITS key is used to toggle the weight
display between primary units and alternate units. Enter the number that corresponds
with the primary units:
0 = UNDEFINED, no units printed
1 = LB
2 = kg
3 = ton
4 = tne
5 = gram
6 = oz
7 = use print label 32 for units
8 = use print code 1 for units
NOTE: when parameter 10 is changed parameters 13 and 16 must be programmed.

Parameter 11. Decimal Point Position (0 [no decimal point])


Enter the decimal point position for the primary units display. For example, if the scale is
calibrated x 0.05 the decimal position = 2.

Parameter 12. Count-by (1)


Enter the primary units count-by. For example, if the scale is configured for:
X 0.05 count-by = 5
X 0.2 count-by = 2
X1 count-by = 1
X5 count-by = 5
X 50 count-by = 50
Note: the entry for 0.05 and 5 is the same. The decimal point position (parameter 11)
differentiates between the count-by 0.05 and count-by 5.

Parameter 13. Alternate Units Type (2 [kg])


Alternate units are the alternate display and print unit of measure that can be used in
addition to the primary units. The UNITS switch on the AD-5000’s keyboard toggles
the weight display between primary units and secondary units.
0 = UNDEFINED, no units printed
1 = LB
2 = kg
3 = ton
4 = tne
5 = gram
6 = oz
7 = use print label 32 for units
8 = use print code 1 for units
Parameter 14. Alternate Decimal Point (1)
Enter the decimal point position for the alternate units display.

25
A&D5000 Users Manual version 4.0.a Set Operation Parameters

Parameter 15. Alternate Count-by (5)


Enter the alternate units count-by.

Parameter 16. Alt Units Conversion Factor (45,360)


The AD-5000 calculates the alternate weight by multiplying the primary weight by the
conversion factor and then dividing by 100,000. The following table lists some common
conversions.
Alternate Units Conversion Factor Table
Conversion Conversion Factor Conv. Fact. x 100,000
LB to kg 0.45359 kg/lb 45,359
kg to LB 2.20462 lb/kg 220,462
LB to Ton 0.0005 Ton/lb 50
kg to tonne 0.001 tonne/kg 100
gram to oz 0.352739 oz/gram 352,739
oz to gram 28.3495 gram/oz 2,834,950
lb to lb 1 100,000
kb to kg
etc.
Enter the conversion factor x 100,000 for the configured alternate units.

Parameter 17. Full Scale Graduations (10,000)


Enter the number of graduations at full scale. The full-scale graduations are calculated
using the count-by, decimal position, and full-scale weight.
Full Scale Graduations = Full Scale Weight / count-by.
For example, if Full Scale Weight = 600.0 and count-by = 0.2 then Full Scale
Graduations = 600.0 / 0.2 = 3000.

Parameter 18. Overload Graduations (10,200)


Enter the number of graduations at which the scale is in overload status. The AD-5000
displays Err oL on the numeric display when the scale weight exceeds overload. For
example, if the Full Scale Graduations is 10,000 grads and the scale is rated for 2%
overload, then Overload Graduations = 10,000 + (10,000 x 0.02) = 10,200.

Parameter 19. Deadload Offset


This parameter is automatically calculated by the configure function (60 & 80).

Parameter 20. Disable Power-On Zero (1 [disabled])


Enter a 1 to disable the power-on zero function. Enter a 0 to enable power-on zero. The
power-on zero function attempts to acquire zero when the power on tests has completed.
If the weight on the scale is above the zero range, the display prompts “Err 0”. Press
the ENTER key to view the scale weight. Remove the weight from the scale to enter
weighing mode.

26
A&D5000 Users Manual version 4.0.a Set Operation Parameters

Parameter 21. Deadload Factor


The Deadload factor is automatically calculated by the calibration function. This
parameter is not accessible from the keyboard.

Parameter 22. Weight Conversion Factor


The weight conversion factor is automatically calculated by the calibration function.
This parameter is not accessible from the keyboard.

Serial I/O Ports


The AD-5000 is equipped with two bi-directional serial communication ports that can
communicate to an optional printer, a computer, bar code reader, a score board or other
serial communication devices. The baud rate and format can be set independently for
each port. Serial port 1 is normally used to output weight data and ID reports to a
printer. Serial port 2 is normally used for continuous output of weight data. Continuous
transmission can be enabled or disabled and the output format can be customized using
Parameters 27 and 28.
Communication between serial devices requires that both devices have the same baud
rate (transmission speed), data bit length, and parity. The mode parameter selects the
data bit length and parity; the baud parameter selects the baud rate.

Parameter 23 Serial Port 1 Mode (4 [8 data bits, no parity])


Enter the mode number from the table below to select the mode parameter in the mode
description column.
Mode Number Mode Description
1 7 data bits, no parity
2 7 data bits, even parity
3 7 data bits, odd parity
4 8 data bits, no parity
5 8 data bits, even parity
6 8 data bits, odd parity

27
A&D5000 Users Manual version 4.0.a Set Operation Parameters

Parameter 24 Serial Port 1 Baud Rate (1 [9600 baud])


Enter a baud number from the table below to select a baud rate from the description
column.
Baud Number Baud Description
1 9600 Baud
2 4800 Baud
3 2400 Baud
4 1200 Baud
5 600 Baud
6 300 Baud

Parameter 25 Serial Port 2 Mode (4 [8 data bits, no parity] )


Enter the mode number from the table below to select the mode parameter in the mode
description column.
Mode Number Mode Description
1 7 data bits, no parity
2 7 data bits, even parity
3 7 data bits, odd parity
4 8 data bits, no parity
5 8 data bits, even parity
16 8 data bits, odd parity

Parameter 26 Serial Port 2 Baud Rate (1 [9600 baud])


Enter a baud number from the table below to select a baud rate from the description
column.
Baud Number Baud Description
1 9600 Baud
2 4800 Baud
3 2400 Baud
4 1200 Baud
5 600 Baud
6 300 Baud
Parameter 27 TX2 Control / Station ID (255 [single station])
Serial Port 2 can be used for continuous transmission of scale data or as a network buss
or for single station remote control.
• Enter a 0 to enable continuous transmission of TX 2 data. The data sent is configured
in parameter 28 (TX2 Format).
• Enter a station ID (1 to 254) for multi-drop network configuration
• Enter 255 for single station remote control

28
A&D5000 Users Manual version 4.0.a Set Operation Parameters

Parameter 28 TX2 Format (2 Condec format)


Serial Port 2 can be used for continuous transmission of scale data. The TX2 Format
parameter determines what is sent. There are five selections:

1. -AND: <status>,<mode>,<polarity>,<weight>,<units>,<terminate>
Status: OL (overload), ST (stable), US (unstable, in motion)
Mode: GS (gross mode), NT (net mode)
Polarity: + (positive weight), - (negative weight)
Weight: 7 characters data, NOT zero blanked, with decimal point
Units: lb, kg, t (ton), tn (tonne), g<sp> (gram), oz
Terminate: lf, cr (line feed, carriage return)

2. -CONDEC: <stx><polarity><weight><units><mode>>status><terminate>
Polarity: space (positive weight), - (negative weight)
Weight: 7 characters, zero blanked, with decimal point
Units: L (lb), K (kg), T (ton), G (gram), O (oz)
Mode: G (gross mode), N (net mode)
Status: <space> (valid weight), M (motion), O (over wt),
I (invalid)
Terminate: lf, cr (line feed, carriage return)

3. -WEIGHTRONIX: <mode1><mode2><polarity><weight><space><units><term>
Mode1: space (gross mode), 0 (zero if net mode)
Mode2: G (gross mode), N (net mode)
Polarity: + (positive weight), - (negative weight)
Weight: 6 characters, NOT zero blanked, with decimal point
Units: lb, kg, oz, g <space>(gram), t<space> (ton), tn (tonne)
Terminate: lf, cr (line feed, carriage return)

4. - OHAS: <polarity><weight><space><units><2 spaces><mode><status><term>


Polarity: space (positive weight), - (negative weight)
Weight: 7 digits, zero blanked, with decimal point
Units: lb, kg, oz, g <space>(gram), t<space> (ton), tn (tonne)
Mode: G (gross mode), N (net mode)
Status: <space> (valid weight), M (motion), O (over wt),
I (invalid)
Terminate: lf, cr (line feed, carriage return)

5. Print label 85: transmits the contents of parameter 85 (print label 85)
Data entered into a print label can include register data:
[253] [reg no] buffer register data, minimum size (zero blanked no spaces)
[254] [reg no] buffer data, 7 character field, zero blanked, right justified.

29
A&D5000 Users Manual version 4.0.a Set Operation Parameters

Printer Parameters
The AD-5000 provides a universal printer interface for various printers. The printer
interface allows you to select and customize the data output for serial or parallel printers.
The serial interface option uses Serial Port 1 (TB1) for connection to the printer. The
parallel interface uses the TTL I/O Port (TB4) for connection to the printer. Refer to the
“HARDWARE INSTALLATION AND WIRING “ section for information on connector
pin designations. NOTE: The parallel interface may not be available if the AD-5000 is
configured to use setpoint outputs.

Parameter 29 Printer Port Select (2 [Serial Port 1])


Enter a number from the following list to select a printer option:
1 Parallel Printer interface.
2 Serial Printer interface.
Parameter 30 Auto Print LF after CR (1 [yes])
Some printers require a line feed character (LF) to advance the paper and some printers
advance the paper with only a carriage return character (CR).
Enter a 1 to automatically add a line feed after a carriage return.
Enter a 0 to NOT add a line feed after a carriage return.

Parameter 31 End of Print Character (12 [form feed])


The end of print character is sent to the printer at the end of a print transmission. If your
printer requires a special character to release the paper or to form feed the paper, enter the
code here. The ASCII chart in the appendix lists the ASCII codes and their numeric
values. Enter a 0 (zero) if no end of print character is required.

Parameters 32, 33 34, 84, 85, 86, 87, 88, 89 and 90: Print Labels
The AD-5000 provide ten programmable print labels. They are available for printing
company name, address, station ID, or other messages. Labels 32, 33, and 34 are up to
16 characters long; labels 84-90 are up to 30 characters long. Use the ASCII chart in the
appendix to look up the numeric values of the ASCII codes. Data entered into a print
label can include register data:
[252] [item no] buffer page format item (see Function 65 ‘Print Item List’)
[253] [reg no] buffer register data, minimum size (zero blanked no spaces)
[254] [reg no] buffer register data, 7 character field, right justified
Select parameter 84 to begin Print Label 84 entry. The display prompts “C0 xxx” where
xxx is the current character 0 entry. Enter:
84, 111, 116, 97, 108, 32, 254, 1, 32, 116, 111, 110, 32, 252, 12, 13, 0
T o t a l <reg1> t o n <date> <cr>
Remember to enter a 0 as the last entry of the print label.

30
A&D5000 Users Manual version 4.0.a Set Operation Parameters

Parameters 35, 36, 37, and 38: Print Codes


The AD-5000 provides four programmable print codes. The print codes are used to send
special setup parameters to the printer, such as select print size or select print font. Each
print code is up to four characters long. Consult your printer manual for the print codes
that you may want to use. Enter the codes in decimal format (an ASCII carriage return =
13). Select parameter 35 to begin print code 1 entry. The AD-5000 prompts “C0 XXX”
where xxx is the current character 0 print code value. Enter data for character 0. The
AD-5000 prompts “C1 XXX”. Continue the code entry. Enter a 0 to end print code
entry.

Parameter 39 Enable TTL Remote Control (1 [enabled])


Use parameter 39 to enable remote control using TTL inputs 1 thru 4 (TB4 pins 10-13).
Enter a 1 to use TTL inputs 1-4 to activate scale basic functions 1 - 4. For example,
grounding TB4 pin 10 (TTL input 1) activates scale basic function 1.
Enter a 0 to disable TTL remote control.

Display Intensity, Battery, Watch Dog Timer

Parameter 40 Display Intensity (10)


Use parameter 40 to set the brightness of the AD-5000’s numeric display and LED
indicators. The brightness can be set from 1 (dimmest) to 15 (brightest).

Parameter 41 Battery backup ( Saves memory data (0 [no])


Parameter 41 is used to preserve the time and date, memory registers, and ID memory
when power is lost. Set parameter 41 to ON (1) this will prevent the AD-5000 from
clearing the memory when power is restored, set parameter 41 to OFF (0) to erase
memory when power is restored.
Parameter 42 Enable Watch Dog Timer (0 [no])
The watch dog timer resets the AD-5000 when the weight display is not updated within
the time-out time (0.2 sec). Set the Enable ON (1) to reset after time-out. Set the Enable
OFF (0) to turn off the watch dog timer.

Parameters 43-50 Fixed Registers


The Fixed Registers are used by the Trip-Level Event Monitor to determine when a trip-
level is reached. See the “Scale Basic / Event Monitor” manual for more information.
Parameters 43 thru 50 select Fixed Setpoints 1 thru 8. Select the parameter number to be
changed and then enter the trip-level weight.

31
A&D5000 Users Manual version 4.0.a Set Operation Parameters

Parameters 51-58 Keyboard Event Functions


The keyboard event functions determine what actions take place when a key is pressed.
The Power on start function activates a function when power is applied to the indicator.

Number Parameter Name Default Value Password


51 Power on start function 0 (no function executed) 2
52 PRINT key function 136 (Print function) 2
53 UNITS key function 127 (Units toggle function) 2
54 Gross/Net key function 128 (Gross/Net toggle) 2
55 TARE key function 129 (Tare function) 2
56 ZERO key function 130 (Tare function) 2
57 Function Key 1 (F1) 0 (no function executed) 2
58 Function Key 2 (F2) 0 (no function executed) 2

32
A&D5000 Users Manual version 4.0.a Parameter Functions

Parameter Functions
The parameter functions are used to set parameters that are too complex for a single
numeric entry. The following table lists the parameter functions:

Parameter Functions
Fn. No Function Name Password Level
59 Display Calibration Audit Number 0
60 Calibrate Scale - Deadload first 3
61 Calibrate Scale - span first 3
62 Calibrate Deadload Only 3
63 Adjust Gain Calibration 3
64 Configure Passwords 2
65 Configure Print Formats 2
66 Set Time and Date 2
67 Display Operation Parameters 0
68 Print Operation Parameters 0
69 Diagnostic Tests 0
70 Initialize Operation Parameters to factory defaults. 3
71 Configure Event Monitor 2
72 Function Editor 2
73 Initialize Function Memory 2
74 Set Debug mode for Event monitors and Scale basic fns. 2
75 Enable Peak Detect (optional) 2
76 Set debug mode for event monitors and scale basic fns. 2
78 Erase Id Memory 1
79 Enable Multi-point linearization (default = disabled [0]) 3
80 Calibrate using Multi-point linearization 3
81 Analog output span (default = 10,000) 1
82 Analog output span (default = 0) 1
83 Analog output register (default = 0 [disabled]) 1

The Password Level is used to prevent unauthorized access to parameter functions. See
Parameter Function 64, “Configure Passwords”, for more information on passwords.

Access Parameter Functions


Hold the CLEAR key down and press the ENTER key. The numeric display prompts
“CFG xx” where xx is the currently selected parameter. Enter a parameter function
number and then press the ENTER key. NOTE: use the UNITS key (↓) and the PRINT
key (↑) to scroll the CFG parameter number. Press the UNITS key to increment the CFG
number, press the PRINT key to decrement the CFG number.

33
A&D5000 Users Manual version 4.0.a Parameter Functions

To exit configuration and store changes, press the CLEAR key when “CFG xx” is being
displayed.
To exit configuration and NOT store changes, enter 999 when “CFG xx” is being
displayed. The display prompts “Abort”. Press the ENTER key to abort (exit without
saving changes) or press the CLEAR key if you do not want to abort.

Example: Begin Calibration of the Scale-


The AD-5000 is in the display weight mode. Hold the CLEAR key down and press the
ENTER key to enter the configure mode. The AD-5000 prompts “CFG xx” where xx is
the last selected parameter. Enter 60 and then press the ENTER key. The AD-5000
prompts “dEAdld”. Begin the calibration process or press the CLEAR key to exit.

Function 59 Display Calibration Audit Number.


Use function 59 to view the audit and configuration numbers, the software ID number,
and the status of the internal function lock.
The AD-5000 provides a calibration audit number and a configuration number. They are
used to indicate changes to calibration data and configuration data and can be used in
place of mechanical seals by some regulatory agencies. The AD-5000 audit number
increments each time a calibration parameter is changed (Deadload or weight conversion
factor). The configuration number increments each time a password level 3 parameter is
changed. If physical seals are used, then place the internal function lock in the locked
position before sealing. Calibration and password level 3 parameters can not be accessed
when the function lock is in the locked position.
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
2. Enter CFG 59. The AD-5000 prompts “A xxxx” where xxxx is the audit number.
3. Press the ENTER key. The AD-5000 prompts “C xxxx” where xxxx is the current
calibration number.
4. Press the ENTER key. The AD-5000 prompts “bixxxx” where bixxxx is the ID
number of the software installed in the AD-5000.
5. Press the ENTER key. The AD-5000 prompt “un-LoC” if the function lock is in the
unlocked position or “Loc Ed” if the function is in the locked position.

Function 60 Calibrate Scale - Deadload First


Use function 60 to calibrate the scale by reading the deadload first and then reading the
span. Set the Scale Operation Parameters before using function 60. Best results are
obtained when the AD-5000 and load cells have reached thermal equilibrium (power
applied for approximately 30 to 120 minutes depending on load cell size). NOTE:
function 60 automatically adjusts parameter 19 (deadload offset).
Hold the CLEAR key down and press the ENTER key to enter the configure mode.

34
A&D5000 Users Manual version 4.0.a Parameter Functions

1. Enter CFG 60. The AD-5000 prompts “dEAdld”. Remove all weight from the scale.
Press the ENTER key. The AD-5000 reads the scale’s deadload weight. NOTE:
press the CONTINUE key (Units) to skip deadload read.
2. The display alternates between “CALib” and the weight on the scale. Place the
calibration weight on the scale. Enter the calibration weight using the numeric
keypad. Press the ENTER key. The AD-5000 reads the calibration weight. If an
error is made in data entry, press the PRINT (↑) key to re-enter the calibration
weight.
3. The AD-5000 prompts “CFG 60”. Press the CLEAR key to exit the configure mode.

Function 61 Calibrate Scale - Span First


Use function 61 to calibrate the scale by reading the span first and then reading the
deadload. Set the Scale Operation Parameters before using function 61. Best results are
obtained when the AD-5000 and load cells have reached thermal equilibrium (power
applied for approximately 30 to 120 minutes depending on load cell size).
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
2. Enter CFG 61. The AD-5000 display alternates between “CALib” and the weight on
the scale. Place the calibration weight on the scale. Enter the calibration weight and
press the ENTER key. The AD-5000 reads the calibration weight.
3. The AD-5000 prompts “dEAdld”. Remove all weight from the scale. Press the
ENTER key. The AD-5000 reads the scale’s deadload weight.
4. The AD-5000 prompts “CFG 61”. Press the CLEAR key to exit the configure mode.

Function 62 Calibrate Deadload Only


Use function 62 to re-calibrate the deadload weight without affecting the span calibration.
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
2. Enter CFG 62. The AD-5000 prompts “dEAdld”. Remove all weight from the scale.
Press the ENTER key. The AD-5000 reads the scale’s deadload weight.
The AD-5000 prompts “CFG 62”. Press the CLEAR key to exit the configure mode.

Function 63 Adjust Gain Calibration


Use function 63 to make small adjustments to the span calibration factor. Zero the scale
and then place a known weight on the scale.
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
2. Enter CFG 63. The AD-5000 displays the weight on the scale along with a center of
digit indication in the least significant digit position of the display (the minus sign (-)
indicates the center of digit.

35
A&D5000 Users Manual version 4.0.a Parameter Functions

3. Press the ‘1’ key to adjust by 1, press the ‘2’ key to adjust by 10, the ‘3’ key to adjust
by 100, or the 4 key to adjust by 1000. Use the UNITS key (↓) and the PRINT key
(↑) to increase or decrease the displayed weight until the displayed weight is the same
as the weight on the scale. The center of digit indicator (-) should be displayed for
maximum accuracy.
4. Press the CLEAR key to exit Adjust Gain procedure.

Adjust Gain – 5 Point Linearization


If five point linearization is enabled, the gain adjust procedure begins with the display
alternating between the calibration number (example: Cal 1) and the calibration weight.
• Press the ENTER key to adjust the gain at the displayed calibration point.
• Press the UNITS key (↓) to adjust the next calibration point
• Press the PRINT key (↑) adjust the previous calibration point.
• Press the ZERO key to clear a calibration point. (Must re-calibrate to reuse).

Function 64 Configure Passwords


Use function 64 to enter password data. The AD-5000 provides two security passwords
that can be activated and entered to protect configuration data. Password 1 is used to
protect I/O port configuration, setpoint, and event monitor data. Password 2 is used to
protect calibration and scale configuration data. Security level 3 provides extra
protection for password level 2 data. Security level 3 requires the entry of password 2
and the Function Lock inside the AD-5000 to be in the unlocked position. Password 3
will inhibit viewing, changing, and uploading of Scale Basic functions to a PC when
password is active but will allow downloading a new Scale Basic program from the PC to
the indicator. All passwords will be disabled once the new program has been downloaded
into the indicator and should be reactivated.
The Operation Parameter Reference Table lists the configuration parameters and the
password/security level required to access them. The AD-5000 prompts “PASS 1” when
a parameter is selected that requires password level 1. If the entered password is
incorrect, the AD-5000 displays “Error”. Press the ENTER key to try again. Press the
CLEAR to exit from password entry.
IMPORTANT NOTE: Write down the passwords and put them in a safe place. If you
loose and forget a password, there is no way to access the protected functions.
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
2. Enter CFG 64. The AD-5000 prompts “P1 xxxx” where xxxx is the current password
number. Enter a new password number or press the ENTER key to use the displayed
number. Enter a 0 (zero) for the password number to disable the password.
3. The AD-5000 prompts “P2 xxxx”. Enter password 2 data.
4. The AD-5000 prompts “P3 xxxx”. Enter password 3 data.

36
A&D5000 Users Manual version 4.0.a Parameter Functions

Function 65 Configure Print Formats


Use function 65 to modify print formats to meet your print requirements. The print
formats are grouped into PAGES that are used for generating print data onto a form.
Each page is organized into rows and columns as follows:
Line No. Column No. Item No.

Line No. = Line number on the page to print the item on.
Column No. = Column position where item print begins.
Item No. = Item to print.
Print format design requires you to become familiar with the PAGE LIST and the PRINT
ITEM LIST. The PAGE LIST describes how each PAGE is used. The PRINT ITEM
LIST describes what ITEMS can be included in each page. Use the print format
worksheets to design new print formats. The PRINT FORMAT DESIGN EXAMPLE
illustrates how a print format is designed. A list of the default print formats is found in
DEFAULT PRINT FORMATS. The print formats (PAGES) are initialized to factory
settings when the Initialize System Parameters function is used. (Parameter Function 73).

Modify or Design a Print Format


1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
Enter CFG number 65.
2. The AD-5000 prompts “PAGE”. Enter the Page number that you want to modify.
3. The AD-5000 prompts “Px E1” where x is the page number that you entered and E1
indicates that the current entry being modified is entry 1. Press the ENTER key.
4. The AD-5000 prompts “L x” where x is the line number on the page that entry 1 is
to print on. Enter a new line number or press ENTER to use the displayed number.
5. The AD-5000 prompts “C x” where x is the column number on the page that entry 1
is to print on. Enter a new column number or press ENTER to use the displayed
number.
6. The AD-5000 prompts “i x” where x is the item number that is to be printed. (See
Print Item List for a description of items that can be printed). Enter a new item
number or press the ENTER key to use the displayed number.
7. The AD-5000 prompts “Px E2”. Use the procedure above to modify entry 2. Use
the UNITS key (↓) to scroll forward in the page table, use the PRINT key (↑) to
scroll backwards.
8. Enter a zero (0) for the line number (L 0) after the last entry that is to be printed.
Press the CLEAR key to exit from the current page. NOTE: If the 16th entry is used,
then a zero entry for the line number is not necessary.

37
A&D5000 Users Manual version 4.0.a Parameter Functions

Page List
FORMAT DESCRIPTION
PAGE 1 Prints when the AD-5000 is in the GROSS weight
mode.
PAGE 2 Prints when the AD-5000 is in the NET weight mode.
PAGE 3 Used by Scale Basic functions when needed.
PAGE 4 Used by Scale Basic functions when needed.

Print Item List


Item # Description
2-4 Print Label 1, 2, and 3 (see chap. 5: Set I/O Port, parameters 32-34)
84-87 Print Labels 4, 5, 6, and 7. (parameters 84-87)
5 Gross Weight (primary units)
6 Tare Weight (primary units)
7 Net Weight (primary units)
8 Gross Weight labeled “WEIGHT” (primary units)
9 Scale Status (used by continuous transmit)
10 Time and Date
11 Time
12 Date
15 Gross Weight (alternate units)
16 Tare Weight (alternate units)
17 Net Weight (alternate units)
18 The Displayed weight (in the displayed units and mode)
21 Print Code 1 (see section: Set I/O Port Parameters, 35-38)
22 Print Code 2
23 Print Code 3
24 Print Code 4
25 Gross Weight (display units)
26 Tare Weight (display units)
27 Net Weight (display units)
28 Gross Weight (display units, no ‘Gross’ label)
29 Tare Weight (display units, no ‘Tare’ label)
30 Net Weight (display units, no ‘Net’ label)
32 Space holder, does nothing.
33 Form Feed (ASCII 12)
34 Carriage Return (ASCII 13)
35 Line Feed (ASCII 10)
40 ID number
41 ID total (ID register 1)
42-46 ID registers 2 through 6
51-66 Memory registers 1 through 16
84-87 Print Labels 84 to 87 (see parameters 84 to 87)

38
A&D5000 Users Manual version 4.0.a Parameter Functions

Print Format Design Example


EXAMPLE PRINT:

1111111111222222222233333333334
COLUMN 1234567890123456789012345678901234567890
LINE
1
2
3 GROSS 54785 LB
4 TARE 12451 LB
5 NET 42243 LB
6
7 TIME 03:45 PM DATE 06 AUG 98

EXAMPLE FORMAT DESIGN WORKSHEET:


Page:1 Item Description Line No. Column Item No.
E1 Gross Weight 3 4 5
E2 Tare Weight 4 4 6
E3 Net Weight 5 4 7
E4 Time and Date 7 1 10
E5 End of Print 0 0 0

NOTE: A zero in the Line No column terminates printing. Use zero only after all
required information has been entered. The maximum number of entries is 16. If the 16th
entry is used, a line number 0 is not required.

Default Print Formats


The print formats (PAGES) are initialized to the following factory settings when the AD-
5000 is initialized.
Page 1 Item Description Line No. Column Item No.
E1 The Displayed Weight 1 1 18
E2 End of Print 0 0 0

Page 2 Item Description Line No. Column Item No.


E1 Gross Wt. (primary units) 3 5 5
E2 Tare Weight (primary units) 4 5 6
E3 Net Weight (primary units) 5 5 7
E4 End of Print 0 0 0

39
A&D5000 Users Manual version 4.0.a Parameter Functions

Format Design Worksheets

Page Item Description Line No. Column Item No.


E1
E2
E3
E4
E5
E6
E7
E8
E9
E10
E11
E12
E13
E14
E15
E16

Page Item Description Line No. Column Item No.


E1
E2
E3
E4
E5
E6
E7
E8
E9
E10
E11
E12
E13
E14
E15
E16

40
A&D5000 Users Manual version 4.0.a Parameter Functions

Function 66 Set Time and Date


Use function 66 to set the time and date. NOTE: Battery Backup parameter 41 should be
set to (1) on so if the unit looses power the clock data will be saved. The time and date
can be displayed by entering 19 when the meter is in the idle mode.
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
2. Enter CFG 66. The AD-5000 prompts “t”. Enter the time using five digits. The last
digit should be 0 for AM, 1 for PM, or 2 for 24 hour time.
3. The AD-5000 prompts “D”. Enter the date using six digits in the format MMDDYY.

Function 67 Display Operation Parameters


Use function 67 to view operation parameters. This function allows viewing the
parameters when the password is not available.
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
2. Enter CFG 67. The AD-5000 prompts “PAR xx” where xx is the currently selected
parameter number. Press the ENTER key to view the parameter data or enter a
parameter number to view.
3. Press the CLEAR key to exit function 67.

Function 68 Print Operation Parameters


1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
2. Enter CFG 68. The AD-5000 prompts “Print”. Press the ENTER key to print the
operation parameters or press the CLEAR key to exit without printing.

Function 69 Diagnostic Tests


1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
2. Enter CFG 69. The AD-5000 prompts “diA xx” where xx is the currently selected
test number. Enter a test number or press the ENTER key to use the displayed
number.

41
A&D5000 Users Manual version 4.0.a Parameter Functions

3. The AD-5000 begins the selected test. Press the CLEAR key when the test is
complete.
Test # Hardware Tested Description
1 Serial Com. Port 1 Display input data.
2 Serial Com. Port 1 Display input error count.
3 Serial Com. Port 1 Transmit data.
4 Serial Com. Port 2 Display input data.
5 Serial Com. Port 2 Display input error count.
6 Serial Com. Port 2 Transmit data.
7 Parallel Port Transmit data.
8 A/D converter Display raw conversion data.
9 Ram Memory Test memory for errors.
10 Display Turn on all LED’s and display
segments.
11 EAROM config Print the EAROM configuration data.
12 Serial Com. Port 1 Transmit/Receive loop-back test.
13 Serial Com. Port 2 Transmit/Receive loop-back test.
See section: Diagnostic Tests for more information on each test.

Function 70 Initialize Operation Parameters to Factory Defaults


Use function 70 to reset the AD-5000’s operation parameters to their factory default
values. This function is used to reset the AD-5000’s earom when a new EPROM
program is installed. Do NOT use this function unless you are ready to begin installation
from the beginning.
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
2. Enter CFG 70. The AD-5000 prompts “rESEt”. Press the ENTER key to reset
operation parameters or press the CLEAR key to exit without changing the operation
parameters.
3. NOTE: enter 10809 when the display prompts “rESEt” to reset the calibration factors
to the factory defaults.

Function 71 Configure Event Monitor


Use function 71 to enter setpoint data into the Event Monitor. See the “Scale Basic /
Event Monitor” manual” for more information on the event monitor parameters.
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
2. Enter CFG 71. The AD-5000 prompts “Eno”. Enter the Event number to be
modified.

42
A&D5000 Users Manual version 4.0.a Parameter Functions

3. The AD-5000 prompts “P0 xx” where P0 is parameter 0 and xx is the current value
of parameter 0. Enter new data for P0 or press the ENTER key to use the displayed
data.
4. Use the UNITS key (↓) or the PRINT key (↑) to scroll through the parameter
numbers. Press the CLEAR key to exit from the current Event number. Press the
CLEAR key again to exit from event monitor data entry.

Function 72 Scale Basic Function Editor


Use function 72 to create or modify Scale Basic functions. See the “Scale Basic / Event
Monitor” manual for more information on Scale Basic functions.
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
2. Enter CFG 72. The AD-5000 prompts “Fn”. Enter the function number to be
modified.
3. The AD-5000 prompts “0- xxx” where 0- is entry zero, and xxx is the current value
for entry zero. Enter new data for 0- or press the ENTER key to use the displayed
data.
4. Use the UNITS key (↓) or the PRINT key (↑) to scroll through the entry numbers.
Press the CLEAR key to exit from the Function Editor.

Function 73 Initialize Function Memory


The initialize function erases function memory and setpoint monitor memory, and
initializes all parameters with password level 1 or 2.
Function 73 prompts “rs Fn” . Press the Enter key to initialize only or enter a number
below to activate a built in program.
1. Fill to Setpoint.
2. Checkweigh: Over/Between/Under
3. Weigh-in / Weigh-out
4. Axle Weigh – Short Scale
5. Axle Weigh – Long Scale
6. ID, Tare, Total
7. Peak Hold

Function 74 Configure Timers


The display prompts “tr” when function 74 is activated. Enter the timer number to be
modified (1 to 5) then press the ENTER key. The first parameter (L) is the time length in
tenths of a second. The second parameter (Fn) is the function number to be executed
when the timer times out. Timer # 5 auto reloads after it counts down to zero. This
provides more accurate timing of repetitive events (such as pulse outputs, speed
calculations and Interval timing, etc.) See timer events in the SCALE BASIC
programming guide.

43
A&D5000 Users Manual version 4.0.a Parameter Functions

Parameter 75 Enable Peak Detect


Enter a 1 to enable peak detect, enter a 0 to disable peak detect.

Parameter 76 Debug mode for Event monitors and Scale Basic


The Debug mode is used to ‘debug’ event monitors and scale basic functions. Enter a 1
to activate debug mode, enter a 0 to disable debug mode. When debug mode is enabled
Scale Basic functions ‘single step’. The subroutine stack level and the Function number
are displayed when entering a function. Example: “0 F12”. Press the ENTER key to
begin execution of the function. The instruction address and instruction code is
displayed. Example: “21 255”. Continue to press the ENTER key to step thru the
program. NOTE: the first step displayed for all scale basic functions is a GOTO (244).

Function 78 Initialize ID memory


Use function 78 to initialize id memory after it has been corrupted. The ID memory
becomes corrupted when power is lost (when no battery is installed) or when the battery
is allowed to discharge below 2 volts.

Parameter 79 Enable Multi-point Linearization


Enter a to 1 to enable multi-point linearization, enter a to 0 to disable multi-point
linearization. The AD-5000 uses the calibration factors set in function 80 below when
multi-point linearization is enabled.

Function 80 Calibrate - Multi-point Linearization


Use function 80 to calibrate the scale using up to 5 calibration points to 'linearize' the
load cell. Set the Scale Operation Parameters before using function 80. Best results are
obtained when the AD-5000 and load cells have reached thermal equilibrium (power
applied for approximately 30 to 120 minutes depending on load cell size). NOTE:
function 80 automatically adjusts parameter 19 (deadload offset).
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.
2. Enter CFG 80. The AD-5000 prompts "dEAdld". Remove all weight from the scale.
Press the ENTER key to obtain the deadload weight. Press the Units key (↓) to
continue without reading a new deadload weight.
3. The AD-5000 ‘s display alternates between "CAL 1" and the weight on the scale.
Place the first calibration weight on the scale. Enter the calibration weight using the
numeric keypad. Press the ENTER key. The AD-5000 reads the calibration weight.
4. The AD-5000 ‘s display alternates between "CAL 2" and the weight on the scale.
Repeat step 3. The AD-5000 will accept up to five calibration points. Press the
CLEAR key to complete calibration if less than five calibration points are used.
5. The AD-5000 prompts "CFG 80". Press the CLEAR key to exit the configure mode

44
A&D5000 Users Manual version 4.0.a Parameter Functions

• Parameter 79 must be set to 1 to enable the AD-5000 to use the multi-point


linearization calibration parameters.
• Use the UNITS key (↓) to continue to the next calibration point if the current point
has already been calibrated.

45
A&D5000 Users Manual version 4.0.a Diagnostic Tests

Diagnostic Tests
The DIAGNOSTIC TESTS are accessed using configuration function 69. Hold the
CLEAR key down and press the ENTER key. The display prompts “CFG xx” where xx
is the currently selected parameter. Enter 69 and press the ENTER key to access the
DIAGNOSTIC TESTS.
The display prompts “diA xx” where xx is the last selected diagnostic function. Select a
test number from the table below. Enter the test number and press the ENTER key.

Table of Diagnostic Tests


Test No. Hardware Tested Description
1 Serial Com. Port 1 Display input data.
2 Serial Com. Port 1 Display input error count.
3 Serial Com. Port 1 Transmit data.
4 Serial Com. Port 2 Display input data.
5 Serial Com. Port 2 Display input error count.
6 Serial Com. Port 2 Transmit data.
7 Parallel Port Transmit data.
8 A/D converter Display raw conversion data.
9 Ram Memory Test memory for errors.
10 Display Turn on all LED’s and display segments.
11 EAROM Print the EAROM configuration data.
12 Serial Com. Port 1 Transmit/Receive loop-back test.
13 Serial Com. Port 2 Transmit/Receive loop-back test.

Diagnostic Test 1: Serial Com Port 1 - Display Input Data


This test displays serial data as it is received by serial communications port 1. The
numeric display has limited alpha display capability, however numeric and some alpha
characters are legible. If all data being received is unintelligible, the baud rate is
probably incorrect. If nothing is displayed on the display, then check the following:
1. RS232/Current switch on the main circuit board.
2. Cabling between the communicating units.
3. Make sure the sending unit is actually sending data.
Press the CLEAR key to exit diagnostic test 1.
Diagnostic Test 2: Serial Com Port 1 – Display Errors
This test displays the number of framing and parity errors that are detected in the input
data stream. The error count is set to zero when entering this test. New errors are
displayed as “PxxFxx” where Pxx is the number of parity errors and Fxx is the number of
framing errors. If the error count exceeds 99, the display will remain at 99.
Press the CLEAR key to exit Diagnostic test 2.

46
A&D5000 Users Manual version 4.0.a Diagnostic Tests

Diagnostic Test 3: Serial Com Port 1 - Transmit Data


This test transmits data out of serial communications port 1. The display prompts: “OUT
0”. Press the ENTER key to send data. Serial Port 1 transmits “1234567890ABCDEF”
and prompts “OUT 1” to indicate that 1 transmission has occurred. Press the ENTER key
again or press the CLEAR key to exit diagnostic test 3.
NOTE: If the error message “Er CtS” appears, the data can not be transmitted because the
handshake signal, clear-to-send (TB1 pin 7), is low or not connected. If clear-to-send is
not used jumper it to data-terminal-ready (TB1 pin 4)

Diagnostic Test 4: Serial Com Port 2 - Display Input Data


This test displays serial data as it is received by serial communications port 2. The
numeric display has limited alpha display capability, however numeric and some alpha
characters are legible. If all data being received is unintelligible, the baud rate is
probably incorrect. If nothing is displayed on the display, then check the following:
1. RS232/RS485 jumper on the main circuit board.
2. Cabling between the communicating units.
3. Make sure the sending unit is actually sending data.
Press the CLEAR key to exit diagnostic test 4.

Diagnostic Test 5: Serial Com Port 2 – Display Errors


This test displays the number of framing and parity errors that are detected in the input
data stream. The error count is set to zero when entering this test. New errors are
displayed as “PxxFxx” where Pxx is the number of parity errors and Fxx is the number of
framing errors. If the error count exceeds 99, the display will remain at 99.
Press the CLEAR key to exit Diagnostic test 5.

Diagnostic Test 6: Serial Com Port 2 - Transmit Data


This test transmits data out of serial communications port 2. The display prompts: “OUT
0”. Press the ENTER key to send data. Serial Port 2 transmits “1234567890ABCDEF”
and prompts “OUT 1” to indicate that 1 transmission has occurred. Press the ENTER key
again or press the CLEAR key to exit diagnostic test 6.

Diagnostic Test 7: Test parallel output


This test transmits data out the parallel TTL I/O port. The display prompts: “dAtA”.
Use the numeric keyboard to enter data. Press the ENTER key to transmit the data. If
the data is successfully sent, “ 1= xxx” is displayed, where xxx is the first three
characters of the data sent and 1= is the number of times that the data has been sent.
Press the ENTER key to send the data again. Press the CLEAR key for the “dAtA”
prompt. Enter new data to transmit or press the CLEAR key to exit diagnostic test 3.
If the data is not successfully sent, the display will remain blank. If data is not sent
within 15 seconds, “Err 13” is displayed. Press the CLEAR key to exit from the error

47
A&D5000 Users Manual version 4.0.a Diagnostic Tests

display. The data will not transmit if the BUSY signal (TB4 pin 14) is active (TTL high).
Check the cable connections and the sending device.

Diagnostic Test 8: Test A/D


This test displays the (filtered) raw data being read from the A/D converter. Use this test
to determine optimum deadload and gain settings. If the number displayed at deadload is
less than 1000 or greater than 250,000 then use the deadload offset (parameter 19) to add
or subtract deadload from the signal input. NOTE: deadload offset is automatically set
using the calibration function 60. If the full-scale number displayed is less than 380,000
then use the “Loadcell mv/V” parameter to increase the gain. If the full-scale signal is
greater than 1,000,000 then use the “Loadcell mv/V” parameter to decrease the gain.
If no data is being displayed, the A/D converter is “locked up”. Turn power off and then
on and re-activate the test. Check load cell wiring for reversed signal or excitation
signals.
Press the PRINT key to scroll the AD test function from:
1. Display raw data (digital filter on).
2. Display and transmit via serial port 2, raw data (digital filter off).
3. Display A/D drift factor. Press the ENTER key to clear the drift factor to 0.
Press the CLEAR key to exit AD test.

Diagnostic Test 9: Perform internal memory test


This test writes data to memory and then reads it back to test for failed memory locations.
The existing data in memory is not affected by this test. The display prompts “r TEST”
while the test is in progress. The display prompts “PAS 32” if the test is successful. If an
error is detected the “Exxxx” is displayed, where xxxx is the location of the error.

Diagnostic Test 10: Lamp Test


This test turns on all the LED displays and numeric display segments. Replace any LED
or display that does not illuminate. Press the CLEAR key to exit diagnostic test 10.

Diagnostic Test 11: Print the EAROM Configuration Table


This function sends the contents to the EAROM configuration memory to the printer.
Use this function to document your setup of the when installation is complete.

Diagnostic Tests 12 & 13: Loop Back Tests


These tests are used by the factory to test the serial ports transmit and receive functions.
The transmit and receive lines must be jumpered together and the clear-to-send must be
jumpered to data-terminal-ready for this test.

48
A&D5000 Users Manual version 4.0.a Troubleshooting

Troubleshooting
The following table describes the probable causes to some problems you may encounter.
Most problems can be resolved by using the information provided in the table.

Problems and Probable Causes

Problem Probable Causes


Printer does not print * Cable connection problem or invalid pin connections.
(serial RS232 ) * Baud rate and format on serial port 1 does not match printer.
* Printer is not ready or it is busy. Check CTS signal TB1 pin 7.
* Transmitter/receiver IC damaged ( U6 MC145406).
Continuous output does * “TX CONTINUOUS” disabled (configuration Parameter 28)
not update * Cable connection problem or invalid pin connections.
* Baud rate and format on serial port 2 does not match receiver.
* Transmitter/receiver IC damaged (U6 MC145406).
Input port 1 not working * Baud rate and format on serial port 1 does not match host.
*Cable connection problem or invalid pin connections.
*Internal switch SW1 (CL/RS232) not set properly for input
type.
*RS232 transmitter/receiver IC damaged (U6 MC145406).
*Current loop receiver IC damaged (U3 4N36).
Input port 2 not working * Baud rate and format on serial port 2 does not match host.
* Cable connection problem or invalid pin connections.
* Internal jumper JP3 (RS232/422/485) not set properly.
* RS232 transmitter/receiver IC damaged (U6 MC145406).
* RS422/485 transmitter/receiver IC damaged (RS422{U14
75176} RS485{U10 75176}).
“Er CtS” error message. * Cable connection problem. Check CTS signal TB1 pin 5. This
signal must be high in order to transmit.
* Printer is not ready or it is busy.
* Transmitter/receiver IC damaged (U6 MC145406).
“Err 13” error message * This error occurs when the parallel output has been selected
(Printer Error) for printing and the printer is busy, a printer fault has occurred,
or there is a problem with the cable.
Invalid weight display * Invalid calibration parameters. * Load cell or load cell input
or weight fluctuating cable problems.

49
A&D5000 Users Manual version 4.0.a Troubleshooting

Error Messages

Error Description
Err 0 Power on zero error. See parameter 20.
Err 1 Keyboard error. Occurs during power up while a key on the keyboard is
being pressed or one of the remote switch inputs is activated.
Err 2 Restart Trap. The microprocessor accessed nonexistent memory location.
Usual cause is A/C power “glitch” or static electric discharge.
Err 3 Watchdog Time-out. The weight display has not been updated within the
watchdog time-out period. Usual cause is A/C power glitch or static electric
discharge.
Err 4 Battery Error. Battery voltage level was below 2 V when the AD-5000 was
powered off.
Err 5 Earom memory error. A checksum error has occurred when reading the
configuration earom memory. Check calibration parameters and rewrite the
earom memory. If error continues, replace earom (U13)
Err 5.1 Earom Time-out, unable to write to the earom memory.
Err 6 Ram memory error. A ram memory error has occurred. Replace U8 if error
continues.
Err 7 The A/D converter is “locked up”. Check load cell wiring.
Err 8 Negative dead-load error.
Err 9 Count-by error. The entered value is outside the scale count-by set during
calibration.
Err 10 Not used
Err 11 A/D converter not working.
Err 12 Negative deadload. Check Loadcell wiring. Enter a deadload offset in
configuration parameter 19.
Err 13 Printers busy error. The busy signal (TB4 pin 14) is active (TTL hi).
Err 14 Page Format line length error. The page format output exceeded the
maximum line length.
Err 15 The function selected is locked. Switch 2 on the AD-5000’s circuit board
locks and unlocks configuration parameters.
Err 16 Scale Basic Stack Overflow
Err 18 Referenced a Function Instruction that does not exist.
Err 20 Double weigh in.
Err 21 ID not found.
Err 22 Scale Basic prompt error.
Err 23 ID memory full.
Err 24 Link list error
Err OL Calibration error or scale is over-loaded (above capacity).
undEr Calibration error or scale is under-loaded ( below readable weight )

50
A&D5000 Users Manual version 4.0.a A&D5000 Specifications

Hardware and Wiring


This section describes installation and wiring information for the interface ports. There
are two bi-directional serial ports, one parallel port with TTL inputs, outputs, and one
load cell input port. The serial ports are used to interface to a printer and to a host device
or PC for continuous output. The parallel port is used to interface to parallel printers and
for remote switch input and relay control using the TTL I/O. The load cell input is used
to interface to the scale platform.

Load Cell Connector - TB3


NOTE: JP1 and JP2 on the circuit board can be used to disable remote sensing.

Load Cell - TB3


Pin # Signal Name
1 + Signal
2 - Signal
3 + Excitation
4 + Sense
5 - Sense
6 - Excitation
7 Analog ground

Note: Terminate load cell cable shield to chassis ground on the exterior
of unit, NOT to terminal block pin # 7.

Place jumpers as shown to DISABLE remote sense. Move jumpers to the


top to enable remote sense.

Remote Sense Jumpers - Remote sense DISABLED

JP1 JP2

Load Cell Terminal Block TB3

51
A&D5000 Users Manual version 4.0.a A&D5000 Specifications

Serial Port 1 Connector – COM 1


For RS232, input, set switch S1 to RS 232 position, and for current loop input, set the
switch to CL position.

Serial Port 1 –COM 1


Pin # Signal Name
1 +5V
2 RS232 TXD (transmit data)
3 RS232 RXD (receive data))
4 DTR (220 ohm to +5V)
5 Ground (GND)
6 Current Loop Transmit
7 RS232 CTS (busy)
8 RS232 RTS (request to send)
9 Current Loop Input -
10 Current Loop Input +
NOTE: Jumper pin 7 to pin 4 if CTS is not used. Com port 1 will not transmit if CTS is
low.

Serial Port 2 Connector – COM 2


For RS232 input jumper JP3 in the RS232 position. For RS422/RS485 input jumper JP3
In the RS485 position.

Serial Port 2 – COM 2


Pin # Signal Name
1 +5V
2 RS232 TXD (transmit data)
3 RS232 RXD (receive data)
4 DTR (220 ohm to +5V)
5 Ground (GND)
6 Current Loop Transmit
7 RS485 Transmit -
8 RS485 Transmit +
9 RS485 Receive -
10 RS485 Receive +

52
A&D5000 Users Manual version 4.0.a A&D5000 Specifications

TTL I/O Port Connector - TB 4


TTL outputs 1 thru 8 will drive 24ma at TTL low level, and 8ma at TTL hi level. Output
9 is an open collector output with 4.7k ohm resistor to +5V.

TTL I/O - TB 4
Pin # Signal Name
1 TTL Output 1 (Printer Data 1)
2 TTL Output 2 (Printer Data 2)
3 TTL Output 3 (Printer Data 3)
4 TTL Output 4 (Printer Data 4)
5 TTL Output 5 (Printer Data 5)
6 TTL Output 6 (Printer Data 6)
7 TTL Output 7 (Printer Data 7)
8 TTL Output 8 (Printer Data 8)
9 TTL Output 9 (Printer Strobe -)
10 TTL Input 1 (Scale Basic Fn 1)
11 TTL Input 2 (Scale Basic Fn 2)
12 TTL Input 3 (Scale Basic Fn 3)
13 TTL Input 4 (Scale Basic Fn 4)
14 TTL Input 5 (Printer Busy +)
15 TTL Input 6 (Printer Error )
16 Ground (GND)

53
test connector
Power Supply Connector
A&D5000 Users Manual version 4.0.a Serial Port 1
A&D5000 Specifications
Serial Port 2
1 2 34 5 67 8 9 1 2 34 5 67 8 9

Keyboard Connector

REMOTE SENSE

Load Cell Cal lock


1 2 3 4 5 6 7
TTL I/O

Analog Out test and relay connector

P.C. Board Diagram - Connectors, Jumpers, and Switches

Remote Sense DISABLED


JP2 JP1 Serial Port 2
RS485 input

Remote Sense ENABLED Serial Port 2


JP2 JP1
RS232 input

54
A&D5000 Users Manual version 4.0.a A&D5000 Specifications

Sample Cable Drawings

Load Cell

LOAD CELL TB 3

1 Signal +
2 Signal -
3 Excitation +
4 Sense +
5 Sense -
6 Excitation -
7 Shield

Note: connect sheild to the enclosure connector hood

IDS 150A/152/160/320

Serial Com Port 1 IDS150A/152/160/320


(RS232)
TB1
2 TXD
RXD
3
5 GND
GND
7
7 CTS RTS 4

NOTE: jumper CTS (pin 7) to pin 4 if CTS is not used.


Com port 1 will not transmit if CTS is TTL low.
DB 25 PIN MALE

Serial Com Port 1 IDS150A/152/160/320


TB1 (20ma Current Loop)
4 +5R CL+ 8
CL1 TX CL- 22
6
7
DB 25 PIN MALE
1

55
A&D5000 Users Manual version 4.0.a A&D5000 Specifications

IDS DT210/ IDS 550/ IDS 710/PC (Continuous Output Port)

Serial Com Port 2 DT210 / IDS550 / IDS710


TB2 (20ma Current Loop)
+5R CL+
4 8
CL1 TX CL-
6 22

DB 25 PIN MALE

Serial Com Port 2 DT210 / IDS550 / IDS710


TB2
(RS232)
TXD RXD
2 3
GND GND
5 7

DB 25 PIN MALE

Hewlett Packard Smart Wand HBCR-8xxx series

Serial Com Port 1 Bar Code Reader HBCR-8xxx


(RS232)
TB1
RXD TXD
3 2
GND GND
5 7
+5V Power
1 9
NOTE: set SW1 to RS232 position.

DB 9 PIN Female

56
A&D5000 Users Manual version 4.0.a A&D5000 Specifications

Parallel (Centronix) Printer

TTL I/O Port PRINTER


FAULT - 32
15
BUSY + 11
14
1 Data 1 2
2 Data 2 3
3 Data 3 4
4 Data 4 5
5 Data 5 6
6 Data 6 7
7 Data 7 8
Data 8 9
8
STROBE - 1
9
Ground (GND) 16
16

TB 4 36 PIN CENTRONIX

Remote Switch Inputs

TTL I/O Port TB 4 Scale Basic Function 1

10 Scale Basic Function 2


11
12 Scale Basic Function 3

13
16 Scale Basic Function 4

GND

57
A&D5000 Users Manual version 4.0.a A&D5000 Specifications

AD-5000 Specifications
Display 1 inch x 6 digit numeric, 6 LED status indicators.
Keyboard Sealed keyboard with tactile feel.

I/O Ports Load Cell Interface Port.


Serial port 1 - RS232 or Cur. Loop
Serial port 2 - RS232 or RS485
9 TTL outputs
8 TTL inputs

Display Graduations 120,000 graduations (industrial)


10,000 hb44.
Display Increments 1, x2, x5, x10, x20, x50, x100
Decimal Point 0 to 4 decimal places
Internal Graduations 1,024,000

Conversion Rate 60/sec @60hz line frequency


Signal Sensitivity 0.1 uV/graduation (min)
Signal Range 0.5 mV/V - 3.9 mV/V

Load Cell Excitation 10 VDC


Load Cell Current 240ma (8, 350 ohm Load Cells)

Auto Zero Tracking 0 to 60 grads, @1/4 grad increments


Auto Zero Delay 0 to 25 sec, @0.1 sec increments
Motion Detect 0 to 60 grads, @1/4 grad increments
Motion Delay 0 to 25 sec, @0.1 sec increments
Digital Filter 19 settings ( 0 to 18 )
Digital Calibration Front Panel - zero & span stored in EAROM

Watchdog Timer Fault tolerant operation (212ms)


Power Input 115/220 VAC, 50/60 Hz. Only the 115VAC
version has been tested to comply with the
Underwriters Laboratories Standards for safety.
Operating Temperature -10 to 40 degrees centigrade

Standard Features Time and Date Clock


Network Interface
Fill to set point, Check weigh, Weigh In/Out, Axle weigh
long and short scales, ID Tare and Total, Peak Hold.

Options Counting program


Analog output 0- 24, 0-20, 4-20 (mA) and 0 – 5 volt DC
12 to 28 VDC operation
220 VAC operation

58
A&D5000 Users Manual version 4.0.a ASCII Chart

ASCII CHART

ASCII DEC ASCII DEC ASCII DEC ASCII DEC


Null 00 SP 32 @ 64 ` 96
SOH 01 ! 33 A 65 a 97
STX 02 " 34 B 66 b 98
ETX 03 # 35 C 67 c 99
EOT 04 $ 36 D 68 d 100
ENQ 05 % 37 E 69 e 101
ACK 06 & 38 F 70 f 102
Bell 07 ' 39 G 71 g 103
BS 08 ( 40 H 72 h 104
HT 09 ) 41 I 73 i 105
LF 10 * 42 J 74 j 106
VT 11 + 43 K 75 k 107
FF 12 , 44 L 76 l 108
CR 13 - 45 M 77 m 109
SO 14 . 46 N 78 n 110
SI 15 / 47 O 79 o 111
DLE 16 0 48 P 80 p 112
XON 17 1 49 Q 81 q 113
TAPE 18 2 50 R 82 r 114
XOFF 19 3 51 S 83 s 115
DC4 20 4 52 T 84 t 116
NAK 21 5 53 U 85 u 117
SYN 22 6 54 V 86 v 118
ETB 23 7 55 W 87 w 119
CAN 24 8 56 X 88 x 120
EM 25 9 57 Y 89 y 121
SUB 26 : 58 Z 90 z 122
ESC 27 ; 59 [ 91 { 123
FS 28 > 60 \ 92 | 124
GS 29 = 61 ] 93 } 125
RS 30 > 62 ^ 94 ~ 126
US 31 ? 63 <- 95 DEL 127

59

También podría gustarte