cambios de diseo y alcance de distribucin y como consecuencia del desarrollo tcnico continuo. Humphrey, HFA, Matrix, FDT, Guided Progression Analysis, GPA, HFA-NET Pro, SITA, SITA-SWAP, STATPAC, EasyConnect,
Visual Field Index, VFI, Cirrus, GDx y FORUM son marcas comerciales o marcas comerciales registradas de Carl Zeiss Meditec, Inc. en los Estados Unidos u otros pases. Todas las dems referencias a marcas son marcas
Cirugia oftalmolgica: Shinjuku Ku 509 Coldhams lane
Es posible que los contenidos de este folleto no coincidan con el estado de aprobacin actual del producto en su pas. Si desea obtener ms informacin, pngase en contacto con nuestro representante local. Sujeto a
Telfono: +49 800 470 50 30 Corea del Sur Tokyo 160-0003 Cambridge
iol.order@meditec.zeiss.com Carl Zeiss Co. Ltd. 22 Honchio-Cho CAMBS
Seoul 121-828 Japn CB1 3JS,
Argentina Mapo-gu Instrumentos oftalmolgicos: Reino Unido
Carl Zeiss Argentina S.A. 141-1, Sangsu-dong Telfono: +81 3 33 55 0331 Telfono: +44 1707 871200
Calle Nahuel Huapi 4015 / 25 2F, BR Elitel Bldg. medsales@zeiss.co.jp info@zeiss.co.uk
C1430 BCO Buenos Aires Corea del Sur Instrumentos quirrgicos:
Argentina Telfono: +82 2 3140 2600 Telfono: +81 3 33 55 0341 Singapur
Telfono: +54 11 45 45 66 61 korea@zeiss.co.kr cmskoho@zeiss.co.jp Carl Zeiss Ptd. Ltd.
comerciales o marcas comerciales registradas de sus respectivas compaas en los Estados Unidos o en otros pases. 2012 Carl Zeiss Meditec, Inc. Todos los derechos reservados.
bruzzi@zeiss.com.ar 50 Kaki Bukit Place
Espaa Malasia Singapore 415926
Australia Carl Zeiss Meditec Iberia S.A.U. Carl Zeiss Sdn Bhd. Singapur
Carl Zeiss Pty Ltd Ronda de Poniente, 15 Lot2, Jalan 243/51 A Telfono: +65 6741 9600
Tenancy Office 4, Level 1 Tres Cantos 46100 Petaling Jaya info@zeiss.com.sg
40-52 Talavera Road 28760 Madrid Selangor Darul Ehsan
North Ryde NSW 2113 Espaa Malasia Sudfrica
Australia Telfono: +34 91 203 37 00 Telfono: +60 3 7877 50 58 Carl Zeiss (Pty.) Ltd.
Telfono: +61 2 9020 1333 info@zeiss.es malaysia@zeiss.com.sg 363 Oak Avenue
med@zeiss.com Ferndale
Francia Mxico Randburg 2194
Austria Carl Zeiss Meditec France SAS Carl Zeiss de Mxico S.A. de C.V. Sudfrica
Carl Zeiss GmbH 60, route de Sartrouville Avenida Miguel ngel de Quevedo Telfono: +27 11 886 9510
Laxenburger Str. 2 78230 Le Pecq 496 info@zeiss.co.za
1100 Viena Francia 04010 Ciudad de Mxico
Austria Telfono: +33 1 34 80 21 00 Mxico Suecia
Telfono: +43 1 79 51 80 med@zeiss.fr Telfono: +52 55 59 99 0200 Carl Zeiss AB
austria@zeiss.org cz-mexico@zeiss.org Tegeluddsvaegen 76
Hong Kong 10254 Estocolmo
Blgica Carl Zeiss Far East Co. Ltd. Nueva Zelanda Suecia
Carl Zeiss NV-SA Units 11-12. 25/F Carl Zeiss (N.Z.) Ltd. Telfono: +46 84 59 25 00
Ikaroslaan 49 Tower 2, Ever Gain Plaza 15B Paramount Drive info@zeiss.se
1930 Zaventem No. 88 Container Port Road P.O. Box 121 - 1001
Blgica Kwai Chung Henderson, Auckland 0650 Suiza
Telfono: + 32 2 719 39 11 Hong Kong Nueva Zelanda Carl Zeiss AG
Carl Zeiss Meditec, Inc. Carl Zeiss Meditec AG Carl Zeiss Meditec AG
5160 Hacienda Drive Goeschwitzer Strasse 5152 Goeschwitzer Strasse 5152
Dublin, CA 94568 07745 Jena 07745 Jena
Estados Unidos de Amrica Alemania Alemania
www.meditec.zeiss.com/matrix www.meditec.zeiss.com/matrix www.meditec.zeiss.com/forum
www.meditec.zeiss.com/contacts www.meditec.zeiss.com/contacts www.meditec.zeiss.com/contacts
HFA II-i
Caractersticas FDT Matrix 800
720i 740i 745i 750i
Control de fijacin
Monitor de punto ciego Heijl-Krakau
Control ocular por vdeo Bibliografa seleccionada
1 Bengtsson B, Heijl A. A visual field index for calculation 17 Kusaba K, Kawanami M, Ban Y. [Utilidad de la perimetra de
Seguimiento de la mirada
of glaucoma rate of progression. Am J Ophthalmol. Feb la tecnologa de duplicacin de frecuencia en la deteccin del
Seguimiento de la cabeza 2008;145(2):343-53. glaucoma en el programa de revisin mdica]. [Artculo en
Control del vrtice 2 Leung CK, Cheung CY, Weinreb RN, et al. Evaluation of japons] Nippon Ganka Gakkai Zasshi. Sep 2004;108(9):554-9.
Funcin de control ocular por vdeo remoto retinal nerve fiber layer progression in glaucoma: a study on 18 Tatemichi M, Nakano T, Tanaka K, et al. Performance of
optical coherence tomography guided progression analysis. glaucoma mass screening with only a visual field test using
Interfaz del operador InvestOphthalmol Vis Sci. Jan 2010;51(1):217-22. Epub frequency-doubling technology perimetry. Am J Ophthalmol.
Pantalla LCD LCD CRT tctil 2009 Aug 13. Oct2002;134(4):529-37. Erratum in: Sep 2003;136(3):592.
Teclado 3 Albanis CV, Quinones RA. Use of Matrix Frequency Doubling 19 Jonas JB, Xu L, Wang YX, et al. The Beijing Eye Study. Acta
Technology (FDT) to Assess Visual Field Status Following Unreliable Ophthalmol. May 2009;87(3):247-61.
Estmulo Standard Automated Perimetry (SAP). Invest Ophthalmol Vis Sci. 20 Thomas R, Naveen S, Nirmalan P, Parikh R. Detection of Ocular
Duplicacin de frecuencia Apr 2008;49:1078. Disease by a Vision Center Technician & The Role of Frequency
Blanco sobre blanco 4 Racette L, Medeiros FA, Zangwill LM, et al. Diagnostic accuracy Doubling Technology Perimetry in this Setting. Br J Ophthalmol.
of the Matrix 24-2 and original N-30 frequency doubling Aug 2009 (18): [Epub ahead of print].
Rojo o azul sobre blanco
technology tests compared with standard automated perimetery. 21 Wang Y, Xu L, Jonas JB. Prevalence and causes of visual field loss
Azul sobre amarillo (SWAP) InvestOphthalmol Vis Sci. 2008;49:954-960. as determined by frequency doubling perimetry in urban and rural
Caractersticas generales de las pruebas 5 Sample PA, Medeiros FA, Racette L, et al. Identifying adult Chinese. Am J Ophthalmol. Jun 2006;141(6):1078-1086.
glaucomatous vision loss with visual-function-specific perimetry in 22 Heijl A, Leske MC, Bengtsson B, Hussein M. Measuring visual field
Tamaos del estmulo 10 2, 5, 10 Goldmann Goldmann Goldmann Goldmann
the diagnostic innovations in glaucoma study. Invest Ophthalmol progression in the Early Manifest Glaucoma Trial. Acta Ophthalmol
III I-V I-V I-V
Vis Sci. 2006;47:3381-3389. Scand. Jun 2003;81(3):286-293.
Pruebas del umbral foveal 6 Sample PA, Bosworth CF, Blumenthal EZ, Girkin C, Weinreb 23 Bengtsson B, Heijl A. Normal intersubject threshold variability and
Medicin automtica de la pupila RN. Visual function-specific perimetry for indirect comparison normal limits of the SITA SWAP and full threshold SWAP perimetric
of different ganglion cell populations in glaucoma. Invest programs. Invest Ophthalmol Vis Sci. Nov 2003;44(11):5029-34.
Almacenamiento de pruebas
Ophthalmol Vis Sci. 2000;41:1783-1790. 24 Bengtsson B. A new rapid threshold algorithm for short-
Definido por el usuario 7 Johnson CA, Samuels SJ. Screening for glaucomatous visual field wavelength automated perimetry. Invest Ophthalmol Vis Sci.
Funciones del software loss with frequency-doubling perimetry. Invest Ophthalmol Vis Sci. Mar2003; 44(3):1388-94.
Feb 1997;38(2):413-425. 25 Casas-Llera P, Rebolleda G, Muoz-Negrete FJ, Arnalich-Montiel
STATPAC 2: anlisis de campo nico
8 Parikh R, Naik M, Mathai A, et al. Role of frequency doubling F, Prez-Lpez M, Fernndez-Buenaga R. Visual field index rate
Prueba de hemicampo del glaucoma (PHG) technology perimetry in screening of diabetic retinopathy. and event-based glaucoma progression analysis: comparison in a
Visual Field Index (VFI) Indian J Ophthalmol. Mar 2006;54(1):17-22. glaucoma population. Br J Ophthalmol. Dec 2009; 93(12):1576-9.
9 Realini T, Lai MQ, Barber L. Impact of diabetes on glaucoma Epub Jun 16 2009.
Guided Progression Analysis (GPA)
screening using frequency-doubling perimetry. Ophthalmology. 26 A partir de 3 dioptras.
Resumen de campo consecutivo
Nov 2004;111(11):2133-6. 27 Medeiros FA, Sample PA, Zangwill LM, et al. A Statistical Approach
Conexin en red 10 Thomas D, Thomas R, Muliyil JP, et al. Role of frequency doubling to the Evaluation of Covariate Effects on the Receiver Operating
Conectividad FORUM perimetry in detecting neuro-ophthalmic visual field defects. Characteristic Curves of Diagnostic Tests in Glaucoma. Invest
Am. J Ophthalmol. June 2001;131(6):734-41. Ophthalmol Vis Sci. Jun 2006;47:2520-2527.
Conectividad DICOM
11 Wall M, Neahring RK, Woodward KR. Sensitivity and specificity 28 Giuffre I. Frequency Doubling Technology vs Standard Automated
Impresora of frequency doubling perimetry in neuro-ophthalmic disorders: Perimetry in Ocular Hypertensive Patients. Open Ophthalmol J.
Impresora trmica Soporte general y nativo para PCL 3, PCL 5 e impresora postscript a comparison with conventional automated perimetry. Jan2009;3:6-9.
para impresoras locales, compartidas y en red InvestOphthalmol Vis Sci. Apr 2002;43(4):1277-83. 29 Patel A, Wollstein G, Ishikawa H, Schuman J. Comparison of Visual
12 Detry-Morel M, Zeyen T, Kestelyn P, Collignon J, Goethals M. Field Defects Using Matrix Perimetry and Standard Achromatic
Almacenamiento, recuperacin y Belgian Glaucoma Society. Screening for glaucoma in a general Perimetry. Ophthalmology. Mar 2007;114(3):480-487.
anlisisdedatos population with the non-mydriatic fundus camera and the 30 White AJ, Sun H, Swanson WH, Lee BB. An examination of
Disco duro 40 GB 40 GB 40 GB 40 GB frequency doubling perimeter. Eur J Ophthalmol. Sep-Oct physiological mechanisms underlying the frequency-doubling
USB 2004;14(5):387-93. illusion. Invest Ophthalmol Vis Sci. Nov 2002;43(11):3590-9.
13 Quigley HA. Identification of glaucoma-related visual field 31 Sakata LM, DeLeon-Ortega J, Arthur SN, et al. Detecting
Unidad de CD-R/W
abnormality with the screening protocol of frequency doubling Visual Function Abnormalities Using the Swedish Interactive
Dimensiones technology. Am J Ophthalmol. June 1998;125(6):819-29. Threshold Algorithm and Matrix Perimetry in Eyes with
Altura 43 cm 43 cm 60 cm 14 Robin TA, Mller A, Rait J, et al. Performance of community-based Glaucomatous Appearance of the Optic Disc. Arch Ophthalmol.
glaucoma screening using Frequency Doubling Technology and Mar2007;125:340-345.
Ancho 25 cm 31 cm 58 cm
Heidelberg Retinal Tomography. Ophthalmic Epidemiol. June 32 Tafreshi A, Sample P, Liebmann J, et al. Visual Function Specific
Profundidad 48 cm 85 cm 51 cm 2005;12(3):167-78. Perimetry to Identify Glaucomatous Visual Field Loss Using
Peso 8,6 kg 17,4 kg 40 kg 15 Iwase A, Tomidokoro A, Araie M, et al. Performance of frequency- Three Different Definitions of Visual Field Abnormality. Invest
doubling technology perimetry in a population-based prevalence Ophthalmol Vis Sci. Mar 2009;50(3),1234-1240.
Requisitos elctricos
survey of glaucoma: the Tajimi study. Ophthalmology. Jan 33 A partir de 3 dioptras se precisan lentes de prueba para Matrix y
100-120V, 50/60Hz 100-240V 100-120V, 50/60Hz 2007;114(1):27-32. Epub 2006 Oct 27. a partir de 7 dioptras para FDT.
230V, 50/60Hz ~ 50/60Hz 16 Saito M, Yamashiro H, Matsumoto H, Shirato S. [Utilidad de
200VA mx. la tecnologa de duplicacin de frecuencia como prueba de
//PERIMETRA Estndares
deteccin del glaucoma] [Artculo en japons] Nippon Ganka
Gakkai Zasshi. Jan 2001;105(1):20-3.
CREADO POR CARL ZEISS Cumple con las normativas UL, CSA y CE
11 12
Analizadores de campo visual Humphrey
C
onectividad para la gestin de datos: los
informes de los pacientes y otros datos pueden
transferirse electrnicamente al sistema de gestin
de datos de oftalmologa FORUM, a los registros
mdicos electrnicos (RME) y a otras soluciones de
gestin de datos de ZEISS.
S
eguimiento de la mirada: realiza un seguimiento
nico del ngulo de fijacin de la mirada e imprime
el error de la mirada para ayudar al mdico a juzgar
la calidad de la prueba.
3
//PRECISIN
CREADO POR CARL ZEISS
El software Guided Progression Analysis (GPA) del HFA distingue con precisin la progresin estadsticamente
significativa de la prdida del campo visual de la variabilidad aleatoria. El anlisis se basa en conocimientos empricos
detallados de la variabilidad encontrada en varias etapas de la prdida de campo visual glaucomatosa a partir de los
datos adquiridos en numerosos ensayos multicntricos de todo el mundo.
Tasa
3 del VFI del
anlisis de progresin
2 Anlisis de tendencias del
historial completo del campo
visual del paciente.25
Grfico
4 del VFI
Un anlisis de regresin
3 delos valores del VFI y
4
previsin de 3 a 5 aos.
5
Barra
5 del VFI
Una descripcin grfica de la
visin til restante del paciente
segn el valor del VFI actual
junto con una previsin de 3 a
5 aos de la lnea de regresin
del VFI si la tendencia actual se
mantiene.
6
Resumen
6 del
campo visual actual
Informe completo del campo
visual actual, que incluye
VFI, DM (desviacin media),
7
DEP (desviacin estndar del
modelo), mapa del anlisis de
progresin y alerta de GPA.
Alerta
7 de GPA
Un mensaje que indica si se
ha detectado un deterioro
estadsticamente significativo
en las siguientes exploraciones.
5
//CONFIANZA
CREADO POR CARL ZEISS
Utilizar un campmetro nunca ha sido tan fcil como con Humphrey Matrix.
Es la solucin ideal para que los consultorios concurridos puedan diagnosticar y realizar
rpidamente pruebas de umbral. Adems de simplificar las pruebas de campo visual,
numerososestudios demuestran que la perimetra de duplicacin de frecuencia puede
detectarunaprdida de campo visual que resultara imposible de detectar con otros mtodos.3-6
6
Resumen de campo consecutivo
7
//FIABILIDAD
CREADO POR CARL ZEISS
Humphrey FDT
Tecnologa de duplicacin de frecuencia para la
deteccin de la prdida de campo visual eficiente.
Clnicamente demostrado
Varios estudios7-21 han revelado que el Humphrey FDT detecta la prdida de campo visual causada
por distintas enfermedades oculares, incluido el glaucoma. As, el FDT es ideal para los mdicos
que deseen identificar a los pacientes que necesiten asistencia oftalmolgica adicional. El FDT ha
sido clnicamente validado en ms de 170 publicaciones referenciadas.
El HFA II-i y Humphrey Matrix 800 proporcionan una amplia gama de soluciones de
conectividad a redes de consultorios, registros mdicos electrnicos y FORUM, el
completo sistema de gestin de datos de Carl Zeiss Meditec.
9
Especificaciones
tcnicas
HFA II-i
Especificaciones FDT Matrix
720i 740i 745i 750i
Especificaciones de la prueba
Intervalo temporal mximo (grados) 30 30 89 89 89 89
Duracin del estmulo 300 ms 300 ms 200 ms 200 ms 200 ms 200 ms
Distancia de control del campo visual Infinito Infinito 30 cm 30 cm 30 cm 30 cm
Iluminacin de fondo 100 cd/m2 100 cd/m2 31.5 ASB 31.5 ASB 31.5 ASB 31.5 ASB
Biblioteca de pruebas de umbral
N-30
C-20
24-2, 30-2, 10-2, mcula
60-4, escaln nasal
10
HFA II-i
Caractersticas FDT Matrix 800
720i 740i 745i 750i
Control de fijacin
Monitor de punto ciego Heijl-Krakau
Control ocular por vdeo Bibliografa seleccionada
1 Bengtsson B, Heijl A. A visual field index for calculation 17 Kusaba K, Kawanami M, Ban Y. [Utilidad de la perimetra de
Seguimiento de la mirada
of glaucoma rate of progression. Am J Ophthalmol. Feb la tecnologa de duplicacin de frecuencia en la deteccin del
Seguimiento de la cabeza 2008;145(2):343-53. glaucoma en el programa de revisin mdica]. [Artculo en
Control del vrtice 2 Leung CK, Cheung CY, Weinreb RN, et al. Evaluation of japons] Nippon Ganka Gakkai Zasshi. Sep 2004;108(9):554-9.
Funcin de control ocular por vdeo remoto retinal nerve fiber layer progression in glaucoma: a study on 18 Tatemichi M, Nakano T, Tanaka K, et al. Performance of
optical coherence tomography guided progression analysis. glaucoma mass screening with only a visual field test using
Interfaz del operador InvestOphthalmol Vis Sci. Jan 2010;51(1):217-22. Epub frequency-doubling technology perimetry. Am J Ophthalmol.
Pantalla LCD LCD CRT tctil 2009 Aug 13. Oct2002;134(4):529-37. Erratum in: Sep 2003;136(3):592.
Teclado 3 Albanis CV, Quinones RA. Use of Matrix Frequency Doubling 19 Jonas JB, Xu L, Wang YX, et al. The Beijing Eye Study. Acta
Technology (FDT) to Assess Visual Field Status Following Unreliable Ophthalmol. May 2009;87(3):247-61.
Estmulo Standard Automated Perimetry (SAP). Invest Ophthalmol Vis Sci. 20 Thomas R, Naveen S, Nirmalan P, Parikh R. Detection of Ocular
Duplicacin de frecuencia Apr 2008;49:1078. Disease by a Vision Center Technician & The Role of Frequency
Blanco sobre blanco 4 Racette L, Medeiros FA, Zangwill LM, et al. Diagnostic accuracy Doubling Technology Perimetry in this Setting. Br J Ophthalmol.
of the Matrix 24-2 and original N-30 frequency doubling Aug 2009 (18): [Epub ahead of print].
Rojo o azul sobre blanco
technology tests compared with standard automated perimetery. 21 Wang Y, Xu L, Jonas JB. Prevalence and causes of visual field loss
Azul sobre amarillo (SWAP) InvestOphthalmol Vis Sci. 2008;49:954-960. as determined by frequency doubling perimetry in urban and rural
Caractersticas generales de las pruebas 5 Sample PA, Medeiros FA, Racette L, et al. Identifying adult Chinese. Am J Ophthalmol. Jun 2006;141(6):1078-1086.
glaucomatous vision loss with visual-function-specific perimetry in 22 Heijl A, Leske MC, Bengtsson B, Hussein M. Measuring visual field
Tamaos del estmulo 10 2, 5, 10 Goldmann Goldmann Goldmann Goldmann
the diagnostic innovations in glaucoma study. Invest Ophthalmol progression in the Early Manifest Glaucoma Trial. Acta Ophthalmol
III I-V I-V I-V
Vis Sci. 2006;47:3381-3389. Scand. Jun 2003;81(3):286-293.
Pruebas del umbral foveal 6 Sample PA, Bosworth CF, Blumenthal EZ, Girkin C, Weinreb 23 Bengtsson B, Heijl A. Normal intersubject threshold variability and
Medicin automtica de la pupila RN. Visual function-specific perimetry for indirect comparison normal limits of the SITA SWAP and full threshold SWAP perimetric
of different ganglion cell populations in glaucoma. Invest programs. Invest Ophthalmol Vis Sci. Nov 2003;44(11):5029-34.
Almacenamiento de pruebas
Ophthalmol Vis Sci. 2000;41:1783-1790. 24 Bengtsson B. A new rapid threshold algorithm for short-
Definido por el usuario 7 Johnson CA, Samuels SJ. Screening for glaucomatous visual field wavelength automated perimetry. Invest Ophthalmol Vis Sci.
Funciones del software loss with frequency-doubling perimetry. Invest Ophthalmol Vis Sci. Mar2003; 44(3):1388-94.
Feb 1997;38(2):413-425. 25 Casas-Llera P, Rebolleda G, Muoz-Negrete FJ, Arnalich-Montiel
STATPAC 2: anlisis de campo nico
8 Parikh R, Naik M, Mathai A, et al. Role of frequency doubling F, Prez-Lpez M, Fernndez-Buenaga R. Visual field index rate
Prueba de hemicampo del glaucoma (PHG) technology perimetry in screening of diabetic retinopathy. and event-based glaucoma progression analysis: comparison in a
Visual Field Index (VFI) Indian J Ophthalmol. Mar 2006;54(1):17-22. glaucoma population. Br J Ophthalmol. Dec 2009; 93(12):1576-9.
9 Realini T, Lai MQ, Barber L. Impact of diabetes on glaucoma Epub Jun 16 2009.
Guided Progression Analysis (GPA)
screening using frequency-doubling perimetry. Ophthalmology. 26 A partir de 3 dioptras.
Resumen de campo consecutivo
Nov 2004;111(11):2133-6. 27 Medeiros FA, Sample PA, Zangwill LM, et al. A Statistical Approach
Conexin en red 10 Thomas D, Thomas R, Muliyil JP, et al. Role of frequency doubling to the Evaluation of Covariate Effects on the Receiver Operating
Conectividad FORUM perimetry in detecting neuro-ophthalmic visual field defects. Characteristic Curves of Diagnostic Tests in Glaucoma. Invest
Am. J Ophthalmol. June 2001;131(6):734-41. Ophthalmol Vis Sci. Jun 2006;47:2520-2527.
Conectividad DICOM
11 Wall M, Neahring RK, Woodward KR. Sensitivity and specificity 28 Giuffre I. Frequency Doubling Technology vs Standard Automated
Impresora of frequency doubling perimetry in neuro-ophthalmic disorders: Perimetry in Ocular Hypertensive Patients. Open Ophthalmol J.
Impresora trmica Soporte general y nativo para PCL 3, PCL 5 e impresora postscript a comparison with conventional automated perimetry. Jan2009;3:6-9.
para impresoras locales, compartidas y en red InvestOphthalmol Vis Sci. Apr 2002;43(4):1277-83. 29 Patel A, Wollstein G, Ishikawa H, Schuman J. Comparison of Visual
12 Detry-Morel M, Zeyen T, Kestelyn P, Collignon J, Goethals M. Field Defects Using Matrix Perimetry and Standard Achromatic
Almacenamiento, recuperacin y Belgian Glaucoma Society. Screening for glaucoma in a general Perimetry. Ophthalmology. Mar 2007;114(3):480-487.
anlisisdedatos population with the non-mydriatic fundus camera and the 30 White AJ, Sun H, Swanson WH, Lee BB. An examination of
Disco duro 40 GB 40 GB 40 GB 40 GB frequency doubling perimeter. Eur J Ophthalmol. Sep-Oct physiological mechanisms underlying the frequency-doubling
USB 2004;14(5):387-93. illusion. Invest Ophthalmol Vis Sci. Nov 2002;43(11):3590-9.
13 Quigley HA. Identification of glaucoma-related visual field 31 Sakata LM, DeLeon-Ortega J, Arthur SN, et al. Detecting
Unidad de CD-R/W
abnormality with the screening protocol of frequency doubling Visual Function Abnormalities Using the Swedish Interactive
Dimensiones technology. Am J Ophthalmol. June 1998;125(6):819-29. Threshold Algorithm and Matrix Perimetry in Eyes with
Altura 43 cm 43 cm 60 cm 14 Robin TA, Mller A, Rait J, et al. Performance of community-based Glaucomatous Appearance of the Optic Disc. Arch Ophthalmol.
glaucoma screening using Frequency Doubling Technology and Mar2007;125:340-345.
Ancho 25 cm 31 cm 58 cm
Heidelberg Retinal Tomography. Ophthalmic Epidemiol. June 32 Tafreshi A, Sample P, Liebmann J, et al. Visual Function Specific
Profundidad 48 cm 85 cm 51 cm 2005;12(3):167-78. Perimetry to Identify Glaucomatous Visual Field Loss Using
Peso 8,6 kg 17,4 kg 40 kg 15 Iwase A, Tomidokoro A, Araie M, et al. Performance of frequency- Three Different Definitions of Visual Field Abnormality. Invest
doubling technology perimetry in a population-based prevalence Ophthalmol Vis Sci. Mar 2009;50(3),1234-1240.
Requisitos elctricos
survey of glaucoma: the Tajimi study. Ophthalmology. Jan 33 A partir de 3 dioptras se precisan lentes de prueba para Matrix y
100-120V, 50/60Hz 100-240V 100-120V, 50/60Hz 2007;114(1):27-32. Epub 2006 Oct 27. a partir de 7 dioptras para FDT.
230V, 50/60Hz ~ 50/60Hz 16 Saito M, Yamashiro H, Matsumoto H, Shirato S. [Utilidad de
200VA mx. la tecnologa de duplicacin de frecuencia como prueba de
//PERIMETRA Estndares
deteccin del glaucoma] [Artculo en japons] Nippon Ganka
Gakkai Zasshi. Jan 2001;105(1):20-3.
CREADO POR CARL ZEISS Cumple con las normativas UL, CSA y CE
11 12
HFA II-i
Caractersticas FDT Matrix 800
720i 740i 745i 750i
Control de fijacin
Monitor de punto ciego Heijl-Krakau
Control ocular por vdeo Bibliografa seleccionada
1 Bengtsson B, Heijl A. A visual field index for calculation 17 Kusaba K, Kawanami M, Ban Y. [Utilidad de la perimetra de
Seguimiento de la mirada
of glaucoma rate of progression. Am J Ophthalmol. Feb la tecnologa de duplicacin de frecuencia en la deteccin del
Seguimiento de la cabeza 2008;145(2):343-53. glaucoma en el programa de revisin mdica]. [Artculo en
Control del vrtice 2 Leung CK, Cheung CY, Weinreb RN, et al. Evaluation of japons] Nippon Ganka Gakkai Zasshi. Sep 2004;108(9):554-9.
Funcin de control ocular por vdeo remoto retinal nerve fiber layer progression in glaucoma: a study on 18 Tatemichi M, Nakano T, Tanaka K, et al. Performance of
optical coherence tomography guided progression analysis. glaucoma mass screening with only a visual field test using
Interfaz del operador InvestOphthalmol Vis Sci. Jan 2010;51(1):217-22. Epub frequency-doubling technology perimetry. Am J Ophthalmol.
Pantalla LCD LCD CRT tctil 2009 Aug 13. Oct2002;134(4):529-37. Erratum in: Sep 2003;136(3):592.
Teclado 3 Albanis CV, Quinones RA. Use of Matrix Frequency Doubling 19 Jonas JB, Xu L, Wang YX, et al. The Beijing Eye Study. Acta
Technology (FDT) to Assess Visual Field Status Following Unreliable Ophthalmol. May 2009;87(3):247-61.
Estmulo Standard Automated Perimetry (SAP). Invest Ophthalmol Vis Sci. 20 Thomas R, Naveen S, Nirmalan P, Parikh R. Detection of Ocular
Duplicacin de frecuencia Apr 2008;49:1078. Disease by a Vision Center Technician & The Role of Frequency
Blanco sobre blanco 4 Racette L, Medeiros FA, Zangwill LM, et al. Diagnostic accuracy Doubling Technology Perimetry in this Setting. Br J Ophthalmol.
of the Matrix 24-2 and original N-30 frequency doubling Aug 2009 (18): [Epub ahead of print].
Rojo o azul sobre blanco
technology tests compared with standard automated perimetery. 21 Wang Y, Xu L, Jonas JB. Prevalence and causes of visual field loss
Azul sobre amarillo (SWAP) InvestOphthalmol Vis Sci. 2008;49:954-960. as determined by frequency doubling perimetry in urban and rural
Caractersticas generales de las pruebas 5 Sample PA, Medeiros FA, Racette L, et al. Identifying adult Chinese. Am J Ophthalmol. Jun 2006;141(6):1078-1086.
glaucomatous vision loss with visual-function-specific perimetry in 22 Heijl A, Leske MC, Bengtsson B, Hussein M. Measuring visual field
Tamaos del estmulo 10 2, 5, 10 Goldmann Goldmann Goldmann Goldmann
the diagnostic innovations in glaucoma study. Invest Ophthalmol progression in the Early Manifest Glaucoma Trial. Acta Ophthalmol
III I-V I-V I-V
Vis Sci. 2006;47:3381-3389. Scand. Jun 2003;81(3):286-293.
Pruebas del umbral foveal 6 Sample PA, Bosworth CF, Blumenthal EZ, Girkin C, Weinreb 23 Bengtsson B, Heijl A. Normal intersubject threshold variability and
Medicin automtica de la pupila RN. Visual function-specific perimetry for indirect comparison normal limits of the SITA SWAP and full threshold SWAP perimetric
of different ganglion cell populations in glaucoma. Invest programs. Invest Ophthalmol Vis Sci. Nov 2003;44(11):5029-34.
Almacenamiento de pruebas
Ophthalmol Vis Sci. 2000;41:1783-1790. 24 Bengtsson B. A new rapid threshold algorithm for short-
Definido por el usuario 7 Johnson CA, Samuels SJ. Screening for glaucomatous visual field wavelength automated perimetry. Invest Ophthalmol Vis Sci.
Funciones del software loss with frequency-doubling perimetry. Invest Ophthalmol Vis Sci. Mar2003; 44(3):1388-94.
Feb 1997;38(2):413-425. 25 Casas-Llera P, Rebolleda G, Muoz-Negrete FJ, Arnalich-Montiel
STATPAC 2: anlisis de campo nico
8 Parikh R, Naik M, Mathai A, et al. Role of frequency doubling F, Prez-Lpez M, Fernndez-Buenaga R. Visual field index rate
Prueba de hemicampo del glaucoma (PHG) technology perimetry in screening of diabetic retinopathy. and event-based glaucoma progression analysis: comparison in a
Visual Field Index (VFI) Indian J Ophthalmol. Mar 2006;54(1):17-22. glaucoma population. Br J Ophthalmol. Dec 2009; 93(12):1576-9.
9 Realini T, Lai MQ, Barber L. Impact of diabetes on glaucoma Epub Jun 16 2009.
Guided Progression Analysis (GPA)
screening using frequency-doubling perimetry. Ophthalmology. 26 A partir de 3 dioptras.
Resumen de campo consecutivo
Nov 2004;111(11):2133-6. 27 Medeiros FA, Sample PA, Zangwill LM, et al. A Statistical Approach
Conexin en red 10 Thomas D, Thomas R, Muliyil JP, et al. Role of frequency doubling to the Evaluation of Covariate Effects on the Receiver Operating
Conectividad FORUM perimetry in detecting neuro-ophthalmic visual field defects. Characteristic Curves of Diagnostic Tests in Glaucoma. Invest
Am. J Ophthalmol. June 2001;131(6):734-41. Ophthalmol Vis Sci. Jun 2006;47:2520-2527.
Conectividad DICOM
11 Wall M, Neahring RK, Woodward KR. Sensitivity and specificity 28 Giuffre I. Frequency Doubling Technology vs Standard Automated
Impresora of frequency doubling perimetry in neuro-ophthalmic disorders: Perimetry in Ocular Hypertensive Patients. Open Ophthalmol J.
Impresora trmica Soporte general y nativo para PCL 3, PCL 5 e impresora postscript a comparison with conventional automated perimetry. Jan2009;3:6-9.
para impresoras locales, compartidas y en red InvestOphthalmol Vis Sci. Apr 2002;43(4):1277-83. 29 Patel A, Wollstein G, Ishikawa H, Schuman J. Comparison of Visual
12 Detry-Morel M, Zeyen T, Kestelyn P, Collignon J, Goethals M. Field Defects Using Matrix Perimetry and Standard Achromatic
Almacenamiento, recuperacin y Belgian Glaucoma Society. Screening for glaucoma in a general Perimetry. Ophthalmology. Mar 2007;114(3):480-487.
anlisisdedatos population with the non-mydriatic fundus camera and the 30 White AJ, Sun H, Swanson WH, Lee BB. An examination of
Disco duro 40 GB 40 GB 40 GB 40 GB frequency doubling perimeter. Eur J Ophthalmol. Sep-Oct physiological mechanisms underlying the frequency-doubling
USB 2004;14(5):387-93. illusion. Invest Ophthalmol Vis Sci. Nov 2002;43(11):3590-9.
13 Quigley HA. Identification of glaucoma-related visual field 31 Sakata LM, DeLeon-Ortega J, Arthur SN, et al. Detecting
Unidad de CD-R/W
abnormality with the screening protocol of frequency doubling Visual Function Abnormalities Using the Swedish Interactive
Dimensiones technology. Am J Ophthalmol. June 1998;125(6):819-29. Threshold Algorithm and Matrix Perimetry in Eyes with
Altura 43 cm 43 cm 60 cm 14 Robin TA, Mller A, Rait J, et al. Performance of community-based Glaucomatous Appearance of the Optic Disc. Arch Ophthalmol.
glaucoma screening using Frequency Doubling Technology and Mar2007;125:340-345.
Ancho 25 cm 31 cm 58 cm
Heidelberg Retinal Tomography. Ophthalmic Epidemiol. June 32 Tafreshi A, Sample P, Liebmann J, et al. Visual Function Specific
Profundidad 48 cm 85 cm 51 cm 2005;12(3):167-78. Perimetry to Identify Glaucomatous Visual Field Loss Using
Peso 8,6 kg 17,4 kg 40 kg 15 Iwase A, Tomidokoro A, Araie M, et al. Performance of frequency- Three Different Definitions of Visual Field Abnormality. Invest
doubling technology perimetry in a population-based prevalence Ophthalmol Vis Sci. Mar 2009;50(3),1234-1240.
Requisitos elctricos
survey of glaucoma: the Tajimi study. Ophthalmology. Jan 33 A partir de 3 dioptras se precisan lentes de prueba para Matrix y
100-120V, 50/60Hz 100-240V 100-120V, 50/60Hz 2007;114(1):27-32. Epub 2006 Oct 27. a partir de 7 dioptras para FDT.
230V, 50/60Hz ~ 50/60Hz 16 Saito M, Yamashiro H, Matsumoto H, Shirato S. [Utilidad de
200VA mx. la tecnologa de duplicacin de frecuencia como prueba de
//PERIMETRA Estndares
deteccin del glaucoma] [Artculo en japons] Nippon Ganka
Gakkai Zasshi. Jan 2001;105(1):20-3.
CREADO POR CARL ZEISS Cumple con las normativas UL, CSA y CE
11 12
Puntos de contacto locales: China Italia Repblica Checa
Carl Zeiss Shanghai Co. Ltd. Carl Zeiss S.p.A. Carl Zeiss spol. s.r.o.
Alemania 1/f., Ke Yuan Building Viale delle Industrie 20 Radlick 14/3201
Carl Zeiss Meditec 11 Ri Yin Nan Road 20020 Arese (Miln) 150 00 Praga 5
Vertriebsgesellschaft mbH Waigaoqiao Free Trade Zone Italia Repblica Checa
Carl-Zeiss-Strasse 22 2005 Yang Gao Bei Road Telfono: +39 02 93773 1 Telfono: +420 233 101 221
73447 Oberkochen Shanghai 200131 infomed@zeiss.it zeiss@zeiss.cz
Alemania China
Telfono: +49 7364 20 6000 Telfono: +86 21 5048 17 17 Japn Reino Unido
vertrieb@meditec.zeiss.com sro@zeiss.com.cn Carl Zeiss Meditec Japan Co. Ltd. Carl Zeiss Ltd.
cambios de diseo y alcance de distribucin y como consecuencia del desarrollo tcnico continuo. Humphrey, HFA, Matrix, FDT, Guided Progression Analysis, GPA, HFA-NET Pro, SITA, SITA-SWAP, STATPAC, EasyConnect,
Visual Field Index, VFI, Cirrus, GDx y FORUM son marcas comerciales o marcas comerciales registradas de Carl Zeiss Meditec, Inc. en los Estados Unidos u otros pases. Todas las dems referencias a marcas son marcas
Cirugia oftalmolgica: Shinjuku Ku 509 Coldhams lane
Es posible que los contenidos de este folleto no coincidan con el estado de aprobacin actual del producto en su pas. Si desea obtener ms informacin, pngase en contacto con nuestro representante local. Sujeto a
Telfono: +49 800 470 50 30 Corea del Sur Tokyo 160-0003 Cambridge
iol.order@meditec.zeiss.com Carl Zeiss Co. Ltd. 22 Honchio-Cho CAMBS
Seoul 121-828 Japn CB1 3JS,
Argentina Mapo-gu Instrumentos oftalmolgicos: Reino Unido
Carl Zeiss Argentina S.A. 141-1, Sangsu-dong Telfono: +81 3 33 55 0331 Telfono: +44 1707 871200
Calle Nahuel Huapi 4015 / 25 2F, BR Elitel Bldg. medsales@zeiss.co.jp info@zeiss.co.uk
C1430 BCO Buenos Aires Corea del Sur Instrumentos quirrgicos:
Argentina Telfono: +82 2 3140 2600 Telfono: +81 3 33 55 0341 Singapur
Telfono: +54 11 45 45 66 61 korea@zeiss.co.kr cmskoho@zeiss.co.jp Carl Zeiss Ptd. Ltd.
comerciales o marcas comerciales registradas de sus respectivas compaas en los Estados Unidos o en otros pases. 2012 Carl Zeiss Meditec, Inc. Todos los derechos reservados.
bruzzi@zeiss.com.ar 50 Kaki Bukit Place
Espaa Malasia Singapore 415926
Australia Carl Zeiss Meditec Iberia S.A.U. Carl Zeiss Sdn Bhd. Singapur
Carl Zeiss Pty Ltd Ronda de Poniente, 15 Lot2, Jalan 243/51 A Telfono: +65 6741 9600
Tenancy Office 4, Level 1 Tres Cantos 46100 Petaling Jaya info@zeiss.com.sg
40-52 Talavera Road 28760 Madrid Selangor Darul Ehsan
North Ryde NSW 2113 Espaa Malasia Sudfrica
Australia Telfono: +34 91 203 37 00 Telfono: +60 3 7877 50 58 Carl Zeiss (Pty.) Ltd.
Telfono: +61 2 9020 1333 info@zeiss.es malaysia@zeiss.com.sg 363 Oak Avenue
med@zeiss.com Ferndale
Francia Mxico Randburg 2194
Austria Carl Zeiss Meditec France SAS Carl Zeiss de Mxico S.A. de C.V. Sudfrica
Carl Zeiss GmbH 60, route de Sartrouville Avenida Miguel ngel de Quevedo Telfono: +27 11 886 9510
Laxenburger Str. 2 78230 Le Pecq 496 info@zeiss.co.za
1100 Viena Francia 04010 Ciudad de Mxico
Austria Telfono: +33 1 34 80 21 00 Mxico Suecia
Telfono: +43 1 79 51 80 med@zeiss.fr Telfono: +52 55 59 99 0200 Carl Zeiss AB
austria@zeiss.org cz-mexico@zeiss.org Tegeluddsvaegen 76
Hong Kong 10254 Estocolmo
Blgica Carl Zeiss Far East Co. Ltd. Nueva Zelanda Suecia
Carl Zeiss NV-SA Units 11-12. 25/F Carl Zeiss (N.Z.) Ltd. Telfono: +46 84 59 25 00
Ikaroslaan 49 Tower 2, Ever Gain Plaza 15B Paramount Drive info@zeiss.se
1930 Zaventem No. 88 Container Port Road P.O. Box 121 - 1001
Blgica Kwai Chung Henderson, Auckland 0650 Suiza
Telfono: + 32 2 719 39 11 Hong Kong Nueva Zelanda Carl Zeiss AG
Carl Zeiss Meditec, Inc. Carl Zeiss Meditec AG Carl Zeiss Meditec AG
5160 Hacienda Drive Goeschwitzer Strasse 5152 Goeschwitzer Strasse 5152
Dublin, CA 94568 07745 Jena 07745 Jena
Estados Unidos de Amrica Alemania Alemania
www.meditec.zeiss.com/matrix www.meditec.zeiss.com/matrix www.meditec.zeiss.com/forum
www.meditec.zeiss.com/contacts www.meditec.zeiss.com/contacts www.meditec.zeiss.com/contacts
Puntos de contacto locales: China Italia Repblica Checa
Carl Zeiss Shanghai Co. Ltd. Carl Zeiss S.p.A. Carl Zeiss spol. s.r.o.
Alemania 1/f., Ke Yuan Building Viale delle Industrie 20 Radlick 14/3201
Carl Zeiss Meditec 11 Ri Yin Nan Road 20020 Arese (Miln) 150 00 Praga 5
Vertriebsgesellschaft mbH Waigaoqiao Free Trade Zone Italia Repblica Checa
Carl-Zeiss-Strasse 22 2005 Yang Gao Bei Road Telfono: +39 02 93773 1 Telfono: +420 233 101 221
73447 Oberkochen Shanghai 200131 infomed@zeiss.it zeiss@zeiss.cz
Alemania China
Telfono: +49 7364 20 6000 Telfono: +86 21 5048 17 17 Japn Reino Unido
vertrieb@meditec.zeiss.com sro@zeiss.com.cn Carl Zeiss Meditec Japan Co. Ltd. Carl Zeiss Ltd.
cambios de diseo y alcance de distribucin y como consecuencia del desarrollo tcnico continuo. Humphrey, HFA, Matrix, FDT, Guided Progression Analysis, GPA, HFA-NET Pro, SITA, SITA-SWAP, STATPAC, EasyConnect,
Visual Field Index, VFI, Cirrus, GDx y FORUM son marcas comerciales o marcas comerciales registradas de Carl Zeiss Meditec, Inc. en los Estados Unidos u otros pases. Todas las dems referencias a marcas son marcas
Cirugia oftalmolgica: Shinjuku Ku 509 Coldhams lane
Es posible que los contenidos de este folleto no coincidan con el estado de aprobacin actual del producto en su pas. Si desea obtener ms informacin, pngase en contacto con nuestro representante local. Sujeto a
Telfono: +49 800 470 50 30 Corea del Sur Tokyo 160-0003 Cambridge
iol.order@meditec.zeiss.com Carl Zeiss Co. Ltd. 22 Honchio-Cho CAMBS
Seoul 121-828 Japn CB1 3JS,
Argentina Mapo-gu Instrumentos oftalmolgicos: Reino Unido
Carl Zeiss Argentina S.A. 141-1, Sangsu-dong Telfono: +81 3 33 55 0331 Telfono: +44 1707 871200
Calle Nahuel Huapi 4015 / 25 2F, BR Elitel Bldg. medsales@zeiss.co.jp info@zeiss.co.uk
C1430 BCO Buenos Aires Corea del Sur Instrumentos quirrgicos:
Argentina Telfono: +82 2 3140 2600 Telfono: +81 3 33 55 0341 Singapur
Telfono: +54 11 45 45 66 61 korea@zeiss.co.kr cmskoho@zeiss.co.jp Carl Zeiss Ptd. Ltd.
comerciales o marcas comerciales registradas de sus respectivas compaas en los Estados Unidos o en otros pases. 2012 Carl Zeiss Meditec, Inc. Todos los derechos reservados.
bruzzi@zeiss.com.ar 50 Kaki Bukit Place
Espaa Malasia Singapore 415926
Australia Carl Zeiss Meditec Iberia S.A.U. Carl Zeiss Sdn Bhd. Singapur
Carl Zeiss Pty Ltd Ronda de Poniente, 15 Lot2, Jalan 243/51 A Telfono: +65 6741 9600
Tenancy Office 4, Level 1 Tres Cantos 46100 Petaling Jaya info@zeiss.com.sg
40-52 Talavera Road 28760 Madrid Selangor Darul Ehsan
North Ryde NSW 2113 Espaa Malasia Sudfrica
Australia Telfono: +34 91 203 37 00 Telfono: +60 3 7877 50 58 Carl Zeiss (Pty.) Ltd.
Telfono: +61 2 9020 1333 info@zeiss.es malaysia@zeiss.com.sg 363 Oak Avenue
med@zeiss.com Ferndale
Francia Mxico Randburg 2194
Austria Carl Zeiss Meditec France SAS Carl Zeiss de Mxico S.A. de C.V. Sudfrica
Carl Zeiss GmbH 60, route de Sartrouville Avenida Miguel ngel de Quevedo Telfono: +27 11 886 9510
Laxenburger Str. 2 78230 Le Pecq 496 info@zeiss.co.za
1100 Viena Francia 04010 Ciudad de Mxico
Austria Telfono: +33 1 34 80 21 00 Mxico Suecia
Telfono: +43 1 79 51 80 med@zeiss.fr Telfono: +52 55 59 99 0200 Carl Zeiss AB
austria@zeiss.org cz-mexico@zeiss.org Tegeluddsvaegen 76
Hong Kong 10254 Estocolmo
Blgica Carl Zeiss Far East Co. Ltd. Nueva Zelanda Suecia
Carl Zeiss NV-SA Units 11-12. 25/F Carl Zeiss (N.Z.) Ltd. Telfono: +46 84 59 25 00
Ikaroslaan 49 Tower 2, Ever Gain Plaza 15B Paramount Drive info@zeiss.se
1930 Zaventem No. 88 Container Port Road P.O. Box 121 - 1001
Blgica Kwai Chung Henderson, Auckland 0650 Suiza
Telfono: + 32 2 719 39 11 Hong Kong Nueva Zelanda Carl Zeiss AG
Carl Zeiss Meditec, Inc. Carl Zeiss Meditec AG Carl Zeiss Meditec AG
5160 Hacienda Drive Goeschwitzer Strasse 5152 Goeschwitzer Strasse 5152
Dublin, CA 94568 07745 Jena 07745 Jena
Estados Unidos de Amrica Alemania Alemania
www.meditec.zeiss.com/matrix www.meditec.zeiss.com/matrix www.meditec.zeiss.com/forum
www.meditec.zeiss.com/contacts www.meditec.zeiss.com/contacts www.meditec.zeiss.com/contacts