CONFIGURACIÓN
CONFIGURACION
• Consiste en el ingreso de datos de transductor,
zapata, material, etc
• Es importante para que el equipo diseñe las
leyes focales y los parametros
correspondientes
GENERACION DE UN GRUPO
Vi / Re =Ta
DETALLES DE LA CONFIGURACION
By Nick Bublitz
Olympus NDT
Calibrations-Required items
• Choose a reference
screen height to equalize
sensitivity to (80% FSH)
and a tolerance
Set Section
• Choose whether to
calibrate the whole
angular sweep in one
step or break it up into
separate calibrations- if
your calibration block
has adequate spacing
between SDHs you can
usually do the whole
range.
Set Gate A on Echo A
Set the gate to encompass the
SDH reflection. Adjust the
start/width/threshold of the
gate.
By Nick Bublitz
Olympus NDT
Calibrations-Required items
• Choose a reference
screen height to equalize
sensitivity to (80% FSH)
and a tolerance
Getting Started
• Choose a SDH and
couple your probe above
it
• Use a guide system to
keep the probe from
skewing side to side.
Set Section
• Choose whether to
calibrate the whole
angular sweep in one
step or break it up into
separate calibrations- if
your calibration block
has adequate spacing
between SDHs you can
usually do the whole
range.
Set Gate A on Echo A
Set the gate to encompass the
SDH reflection. Adjust the
start/width/threshold of the gate.