Manual Oasciloscopio OD 512 Promax PDF
Manual Oasciloscopio OD 512 Promax PDF
OSCILOSCOPIO DE 20 MHz
20 MHz OSCILLOSCOPE
OSCILLOSCOPE DE 20 MHz
0 MI0843
MANUAL DE INSTRUCCIONES OD-512
SAFETY NOTES
Read the instruction manual before using the equipment, mainly "SAFETY
RULES" paragraph.
Warning and Caution statements may appear in this manual to avoid injury hazard
or damage to this product or other property.
SUMARIO
CONTENTS
SOMMAIRE
INDICE
1. GENERALIDADES ................................................................................................ 1
1.1 DESCRIPCIÓN ...................................................................................................... 1
1.2 ESPECIFICACIONES ............................................................................................... 2
2. PRESCRIPCIONES DE SEGURIDAD ................................................................... 5
2.1 GENERALES ......................................................................................................... 5
2.2 PRECAUCIONES ESPECÍFICAS................................................................................. 7
2.3 EJEMPLOS DESCRIPTIVOS DE LAS CATEGORÍAS DE SOBRETENSIÓN ............................ 7
3. INSTALACIÓN ...................................................................................................... 9
3.1 ALIMENTACIÓN ..................................................................................................... 9
3.2 ORIENTACIÓN DEL ASA ........................................................................................ 10
4. INSTRUCCIONES DE MANEJO .......................................................................... 11
4.1 DESCRIPCIÓN DE MANDOS Y ELEMENTOS ............................................................... 11
4.2 PUESTA EN MARCHA ........................................................................................... 16
4.2.1 Operaciones preliminares ........................................................................ 16
4.2.2 Ajuste de la rotación de la traza ............................................................... 17
4.2.3 Ajuste de las sondas ................................................................................ 17
4.2.4 Ajuste del nivel de DC.............................................................................. 17
4.3 FORMA DE UTILIZACIÓN ....................................................................................... 18
4.3.1 Operación con una sola traza .................................................................. 18
4.3.2 Operación con doble traza ....................................................................... 19
4.3.3 Medida aditiva y substractiva ................................................................... 19
4.3.4 Opciones de disparo ................................................................................ 20
4.3.5 Funcionamiento en modo X-Y .................................................................. 24
5. APLICACIONES .................................................................................................. 25
5.1 APLICACIONES USANDO UN SOLO CANAL................................................................ 25
5.1.1 Mediciones de amplitud ........................................................................... 25
5.1.2 Mediciones de intervalos de tiempo ......................................................... 27
5.1.3 Medición del tiempo de subida ................................................................. 30
5.1.4 Medición de la profundidad de modulación en amplitud ............................ 32
5.1.5 Medición de frecuencia ............................................................................ 32
5.2 APLICACIONES USANDO LOS DOS CANALES ............................................................ 33
5.2.1 Comparación de niveles........................................................................... 33
5.2.2 Reparación de aparatos estéreo .............................................................. 33
5.2.3 Reparación de TV .................................................................................... 33
5.2.4 Análisis de vídeo compuesto.................................................................... 33
5.2.5 Mediciones de diferencia de fase ............................................................. 34
6. MANTENIMIENTO ............................................................................................... 39
6.1 INSTRUCCIONES DE ENVÍO ................................................................................... 39
6.2 SUSTITUCIÓN DEL FUSIBLE DE RED........................................................................ 39
6.3 FUSIBLES NO SUSTITUIBLES POR EL USUARIO ......................................................... 40
6.4 RECOMENDACIONES DE LIMPIEZA ......................................................................... 40
MANUAL DE INSTRUCCIONES OD-512
MANUAL DE INSTRUCCIONES OD-512
OSCILOSCOPIO DE 20 MHz
OD-512
1. GENERALIDADES
1.1 Descripción
3) Disparo alternado
Incluso con la observación de 2 formas de onda diferentes, cada una de las formas
de onda puede disparase de forma estable.
5) Salida CH1
La salida CH1 en el panel posterior del equipo puede ser aplicada a otros
instrumentos de medida como por ejemplo un frecuencímetro.
6) Entrada eje Z
La capacidad de modulación de intensidad del haz permite añadir marcadores de
frecuencia o de tiempo. Traza blanco con señal positiva, compatible TTL.
06/2003 Página 1
MANUAL DE INSTRUCCIONES OD-512
1.2 Especificaciones
Sensibilidad de 5 mV a 5 V/DIV, 10 pasos en secuencia 1-2-5
Precisión ≤3% ( x 5 MAG: ≤5%)
Control variable de
Hasta 1/2.5 o inferior del valor indicado en el panel
sensibilidad
DC a 20 MHz ( x 5MAG: DC a 7 MHz)
Ancho de banda Acoplo AC: Frecuencia límite inferior 10 Hz. (Con referencia a 100 kHz, 8 DIV.
Respuesta en frecuencia a -3 dB).
Tiempo de subida Aprox. 17.5 ns (x 5 MAG: Aprox. 50 ns)
Impedancia de entrada Aprox. 1 MΩ // Aprox. 25 pF
Características señal Sobredisparo: ≤ 5% (en la escala 10 mV/DIV)
cuadrada Otras distorsiones y otras escalas: 5% añadido al valor anterior
Ajuste de DC Ajustable por el usuario
<±0.1 DIV de la variación de amplitud cuando se desplaza verticalmente una señal
Linealidad
de 2 DIV desde el centro de la retícula
CH1: sólo canal CH1
EJE CH2: sólo canal CH2
VERTICAL Modos de operación DUAL: se muestran los canales CH1 y CH2. Modo ALT (alternado) o CHOP
(recortado) seleccionable en todas las velocidades de barrido.
ADD: CH1 + CH2 suma algebraica
Frecuencia de
repetición de recorte Aprox. 250 kHz
(Chopping)
Acoplo de entrada AC, GND, DC
Tensión máxima de
300 V DC + pico AC (AC: frecuencia 1 kHz o inferior)
entrada
Relación de rechazo 50:1 o mejor para una señal senoidal de 50 kHz. (Para sensibilidades de CH1 y
en modo común CH2 iguales).
Aislamiento entre
>1000:1 a 50 kHz
canales
>30:1 a 20 MHz
(escala 5 mV/DIV)
Superior a 20 mV/DIV sobre una carga de 50 Ω, ancho de banda mínimo de 50 Hz
Señal salida CH1
a 5 MHz.
CH2 INV BAL Punto de variación equilibrado: ≤ 1 DIV (Referida al centro de la retícula).
CH1, CH2, LINE, EXT (CH1 y CH2 solo pueden seleccionarse en los modos
Fuente de disparo vertical DUAL o ADD). En el modo ALT, interruptor TRIG. ALT pulsado, se alterna
la fuente de disparo.
Acoplo AC: de 20 Hz a ancho de banda completo
Pendiente +/-
20 Hz ∼ 2 MHz: 0.5 DIV, TRIG-ALT: 2 DIV, EXT: 200 mV
Sensibilidad 2 ∼ 20 MHz: 1.5 DIV, TRIG-ALT:3 DIV, EXT: 800 mV
TV: Impulso de Sincronismo mayor que 1 DIV (EXT: 1V)
AUTO: Barrido en el modo libre cuando no se aplica ninguna señal a la entrada de
disparo. (Aplicable para señales repetitivas de frecuencia 25 Hz o
superior).
NORM: Se produce barrido únicamente cuando se recibe una señal de disparo.
DISPARO No es posible ver la traza sin ningún disparo. Este modo debe utilizarse
cuando la frecuencia de la señal es 25 Hz o inferior.
Modos de disparo
TV-V: Este modo se utiliza para visualizar la imagen vertical completa de una
señal de televisión.
TV-H: Este modo se utiliza para visualizar la imagen horizontal completa de
una señal de televisión.
(Ambos, TV-V y TV-H, sincronizan únicamente para señales de sincronismo
negativas).
Entrada señal disparo
Aprox.: 1 MΩ // aprox. 25 pF
exterior (TRIG IN)
Impedancia de entrada
300 V (DC+AC pico), AC: Frecuencia inferior a 1 kHz
Máx. tensión de entrada
Página 2 06/2003
MANUAL DE INSTRUCCIONES OD-512
06/2003 Página 3
MANUAL DE INSTRUCCIONES OD-512
Página 4 06/2003
MANUAL DE INSTRUCCIONES OD-512
2. PRESCRIPCIONES DE SEGURIDAD
2.1 Generales
* Al emplear cualquiera de los siguientes accesorios debe hacerse solo con los tipos
especificados a fin de preservar la seguridad.
Cable de red
Sustitución del fusible de red, que deberá ser del tipo y valor indicados.
06/2003 Página 5
MANUAL DE INSTRUCCIONES OD-512
CORRIENTE CONTINUA
CORRIENTE ALTERNA
ALTERNA Y CONTINUA
TERMINAL DE TIERRA
TERMINAL DE PROTECCIÓN
TERMINAL A CARCASA
EQUIPOTENCIALIDAD
MARCHA
PARO
DOBLE AISLAMIENTO
(Protección CLASE II)
PRECAUCIÓN
(Riesgo de choque eléctrico)
FUSIBLE
Página 6 06/2003
MANUAL DE INSTRUCCIONES OD-512
06/2003 Página 7
MANUAL DE INSTRUCCIONES OD-512
Página 8 06/2003
MANUAL DE INSTRUCCIONES OD-512
3. INSTALACIÓN
3.1 Alimentación
El osciloscopio OD-512 está preparado para ser alimentado con tensiones de red
de 115 ó 230 V AC 50-60 Hz. La tensión de red puede seleccionarse desde la propia
base de red.
3.- Insertar la tapita portafusibles, haciendo coincidir el índice [A] con la indicación de la
tensión de red deseada [B].
PRECAUCIÓN:
EL APARATO VIENE PREPARADO DE FABRICA PARA 230 V.
06/2003 Página 9
MANUAL DE INSTRUCCIONES OD-512
Para girar el asa, sujete con ambas manos las bases del asa que están en
contacto con el chasis del osciloscopio y sepárelas ligeramente del chasis del
oscilocopio, de esta manera podrá girar el asa hasta la posición que desee.
Página 10 06/2003
MANUAL DE INSTRUCCIONES OD-512
4. INSTRUCCIONES DE MANEJO
Panel frontal
[1] Conector CAL. Proporciona una onda cuadrada de amplitud calibrada para
ajustes de la sonda y la calibración del amplificador vertical.
CRT
[2] INTEN. Control de ajuste del brillo de la traza en la pantalla del TRC. Aumenta
girando el mando en el sentido de las agujas del reloj.
06/2003 Página 11
MANUAL DE INSTRUCCIONES OD-512
[8] CH1 (X). Conector de entrada del canal vertical CH1. En el modo de
operación X-Y se convierte en la entrada del eje X.
[20] CH2 (Y). Conector de entrada del canal vertical CH2. En el modo de
operación X-Y se convierte en la entrada del eje Y.
[7] [22] VOLTS/DIV. Selectores de la sensibilidad del eje vertical CH1 y CH2
respectivamente, de 5 mV/DIV a 5 V/DIV en 10 pasos.
[9] [21] VARIABLE. Ajuste fino de la sensibilidad del eje vertical CH1 y CH2
respectivamente, con un factor ≥ 1/2.5 del valor seleccionado por el control
VOLTS/DIV. La sensibilidad de los amplificadores verticales VOLTS/DIV está
calibrada cuando los controles VARIABLE señalan la posición marcada como
CAL. Cuando se extraen estos controles (modo PULL x5 MAG) la sensibilidad
del amplificador se multiplica por 5.
[13] [17] DC BAL. Ajustes del balance DC de los amplificadores de entrada CH1 y CH2
respectivamente. Para más detalles ver el párrafo 4.2.4 Ajuste del nivel de DC.
[14] MODE. Selección del modo de operación del los amplificadores CH1 y CH2.
CH1 El osciloscopio trabaja como un instrumento de un único canal, sólo
se visualiza la señal en CH1.
CH2 El osciloscopio trabaja como un instrumento de un único canal, sólo
se visualiza la señal en CH2.
DUAL El osciloscopio trabaja como un instrumento de doble canal: se
visualizan las señales en CH1 y CH2 (ver la descripción del control
[12])
ADD Se visualiza la suma algebraica de las señales CH1 y CH2. También
se puede visualizar la diferencia de señales, para ello se debe invertir
la polaridad de la señal CH2 mediante el control [16].
Página 12 06/2003
MANUAL DE INSTRUCCIONES OD-512
CHOP Modo recortado. Tecla pulsada. Las trazas de las señales en CH1 y
CH2 se recortan, es decir cada barrido horizontal del haz de
electrones del TRC se recorta en secciones alternando en cada una
de ellas el canal representado. Recomendado para velocidades de
barrido bajas.
DISPARO (TRIGGER)
[24] Conector TRIG IN. Permite aplicar una señal externa de sincronismo a los
circuitos de disparo.
[27] TRIG. ALT. Cuando el interruptor VERT MODE [14] selecciona el estado
DUAL o ADD y el control SOURCE [23] selecciona CH1 o CH2, al pulsar esta
tecla se alternan las señales en CH1 y CH2 como fuente de disparo.
06/2003 Página 13
MANUAL DE INSTRUCCIONES OD-512
[29] TIME/DIV. Selector del tiempo de barrido de 0.2 µs/DIV a 0.5 s/DIV en 20
pasos. La posición X-Y se selecciona en el modo de operación
Página 14 06/2003
MANUAL DE INSTRUCCIONES OD-512
Panel posterior
[34] Z AXIS INPUT. Para aplicar una señal que module la intensidad del TRC. El
brillo de la traza se reduce con una señal positiva y aumenta con una negativa.
[35] Conector CH1 OUTPUT. Proporciona una salida amplificada del canal 1,
adecuada para controlar un frecuencímetro u otros instrumentos.
[38] Pies para apoyar el osciloscopio sobre su panel posterior. También utilizados
para enrollar el cable de red.
06/2003 Página 15
MANUAL DE INSTRUCCIONES OD-512
1. Pulse el interruptor I/O [6]. El piloto POWER [5] deberá iluminarse. Después de 20
segundos, la traza aparecerá en la pantalla. Si la traza no aparece en 60 segundos
aproximadamente, revise el estado de los mandos y elementos de acuerdo con el punto 1.
ATENCIÓN
Se ha utilizado un material resistente al envejecimiento del TRC. De todas
formas, si se deja el TRC con un punto o traza extremadamente brillante y
por un período de tiempo largo, puede dañarse la pantalla. Por tanto, si la
medición requiere un brillo muy intenso asegúrese de bajar el control
INTEN inmediatamente después. También acostúmbrese a bajar el brillo
cuando deje de observar el osciloscopio durante un tiempo.
3. Gire el control CH1 POSITION [11] para desplazar la traza del canal CH1
hasta la línea central de la retícula.
4. Gire el control POSITION [32] para ajustar el borde izquierdo de la traza
con la línea del extremo izquierdo de la retícula .
Página 16 06/2003
MANUAL DE INSTRUCCIONES OD-512
Por medio del control de posición del canal CH1 POSITION [11], situar la
traza de forma que coincida con la línea horizontal central de la retícula.
Conectar una sonda (x10) a la entrada CH1 [8] y conectar la punta de la sonda al
conector CAL [1]. En la pantalla del TRC debe aparecer una onda cuadrada con una
amplitud de cuatro divisiones (4 x 0.5 V/DIV = 2 Vpp).
06/2003 Página 17
MANUAL DE INSTRUCCIONES OD-512
La operación con una sola traza con una única base de tiempo y disparo interno
es el modo operativo más elemental del OD-512. Utilice este modo cuando desee
observar únicamente una señal, sin la molestia de otras trazas en el TRC. Puesto que se
trata de un osciloscopio de dos canales, se puede escoger uno de los dos. El canal CH1
tiene un terminal de salida: utilice la salida CH1 en el panel posterior si desea medir la
frecuencia de la señal con un frecuencímetro mientras observa la forma de la onda. El
canal CH2 tiene un interruptor para invertir la polaridad, aunque ello añade flexibilidad, no
es muy útil para la operación habitual con una sola traza.
El OD-512 se configura para la operación con una sola traza de la siguiente forma:
ATENCIÓN
No aplique una señal mayor de 300 V (DC + pico AC).
Página 18 06/2003
MANUAL DE INSTRUCCIONES OD-512
6. Si la señal que desea observar es de una frecuencia tan alta que incluso en la
posición de 0.2 µs del selector TIME/DIV [29] aparecen demasiados ciclos en la pantalla,
pulse la tecla x10 MAG [31]. Ello aumenta la velocidad de barrido efectiva por un factor
10, así 0.2 µs/DIV pasan a ser 20 ns/DIV, 0.5 µs resultan 50 ns/DIV, etc. Las posiciones
0.2 y 0.5 µs x 10 no están calibradas, mientras que las posiciones 1 µs e inferiores si
están calibradas.
7. Si la señal que desea observar es una señal contínua o posee una frecuencia
muy baja, por lo que el acoplamiento en AC atenúa o distorsiona la señal, lleve el
selector AC/GND/DC [10] o [18] a DC.
Sitúe el selector VERT MODE [14] en la posición DUAL de manera que también
se visualice la señal del canal 2. El funcionamiento es idéntico al descrito en el párrafo
4.3.1 con las siguientes excepciones:
Las operaciones aditiva y substractiva son dos formas de operación con dos
canales en las cuales se combinan las dos señales para visualizar una única traza. En la
operación aditiva la traza resultante representa la suma algebraica de las señales CH1 y
CH2. En la operación substractiva, la traza resultante representa la diferencia algebraica
entre las señales CH1 y CH2.
06/2003 Página 19
MANUAL DE INSTRUCCIONES OD-512
4. Lleve el selector MODE [14] a la posición ADD. De esta forma la única traza
resultante será la suma algebraica de las señales CH1 y CH2. Tanto el control
POSITION [11] como [19] se pueden utilizar para desplazar la traza resultante.
NOTA
Si las señales de entrada se encuentran en fase, la amplitud de la traza
resultante será la suma aritmética de las trazas individuales (por ej., 4.2 div.
+ 1.2 div. = 5.4 div.). Si las señales de entrada tienen un desfase de 180º, la
amplitud será la diferencia (por ej., 4.2 div. – 1.2 div. = 3.0 div.).
Página 20 06/2003
MANUAL DE INSTRUCCIONES OD-512
Las posiciones TV-V y TV-H del selector Trigger MODE insertan un separador
de sincronismo de TV en la cadena de disparo, de forma que una señal de disparo
limpia, tanto vertical como horizontal, pueda extraerse de una señal de vídeo compuesta
(fig. 7a). Para disparar el osciloscopio en la frecuencia de sincronismo vertical (fig. 7b),
lleve el selector Trigger MODE a TV-V. Para disparar el osciloscopio en la frecuencia de
sincronismo horizontal (línea) (fig. 7c), lleve el selector MODE a TV-H. Para obtener los
mejores resultados, la polaridad de sincronismo TV debe ser negativa (fig. 7d), cuando
se utiliza el separador de sincronismo.
06/2003 Página 21
MANUAL DE INSTRUCCIONES OD-512
LINE La señal de la red eléctrica se utiliza como señal de disparo. Este medio es
efectivo cuando la señal a medir posee una frecuencia relacionada con la de la
red eléctrica, especialmente para la medida de niveles de ruido AC bajos de
equipos de audio, circuitos con tiristores, etc.
Página 22 06/2003
MANUAL DE INSTRUCCIONES OD-512
06/2003 Página 23
MANUAL DE INSTRUCCIONES OD-512
1. Gire totalmente el selector TIME/DIV [29] en el sentido de las agujas del reloj
a la posición X-Y.
ATENCIÓN
Reduzca la intensidad de la traza para que el punto sin deflexionar no dañe
el fósforo del TRC.
5. La señal vertical (eje Y) se puede invertir pulsando el interruptor CH2 INV [16].
Página 24 06/2003
MANUAL DE INSTRUCCIONES OD-512
5. APLICACIONES
Tensiones pico a pico. Para medir la tensión pico a pico, proceda de la forma
siguiente.
2. Ajuste el selector TIME/DIV [29] para obtener dos o tres ciclos de forma de
onda y el conmutador VOLTS/DIV [7] o [22] para conseguir en la pantalla la visualización
más amplia.
3. Use el control POSITION apropiado [11] o [19] para situar los picos
negativos de la señal en la línea horizontal de la retícula más cercana por debajo de sus
picos (fig. 9).
4. Use el control POSITION [32] para situar uno de los picos positivos en la
línea vertical central de la retícula. Esta línea tiene marcas de calibración adicionales
espaciadas 0.2 divisiones.
7. Si mide una onda senoidal inferior a 100 Hz, o una onda rectangular por
debajo de los 1000 Hz, sitúe el selector AC/GND/DC en DC.
06/2003 Página 25
MANUAL DE INSTRUCCIONES OD-512
2. Ajuste el selector TIME/DIV [29] para obtener un ciclo completo de onda y gire
el conmutador VOLTS/DIV para conseguir una amplitud de traza de 4 o 6 divisiones (fig.
8).
Página 26 06/2003
MANUAL DE INSTRUCCIONES OD-512
4. Use el control POSITION apropiado [11] o [19] para situar la traza sobre
la línea horizontal central de la retícula. No obstante, si sabe que la tensión de la señal
es totalmente positiva, use la línea inferior de la retícula. Si sabe que la tensión de la
señal es totalmente negativa use la línea superior. Dicha línea recibe el nombre de línea
de referencia de tierra.
NOTA
Los controles Vertical POSITION no deben manipularse de nuevo hasta
que no termine la medición.
ATENCIÓN
Asegúrese de que la forma de onda no se está leyendo con una tensión DC
de alta amplitud antes de variar el conmutador AC/GND/DC.
3. Use el control POSITION [11] o [19] para situar la traza de manera que la
línea horizontal central de la retícula pase a través de los puntos de la forma de onda
entre los cuales quiera hacer la medición.
06/2003 Página 27
MANUAL DE INSTRUCCIONES OD-512
Una técnica alternativa útil para impulsos de amplitud inferior a una división
consiste en tirar del botón x10 MAG [31] para una magnificación x10, y resitúar el
impulso en la pantalla con el control POSITION [32]. En algunas aplicaciones, la
amplitud del impulso también se llama de tiempo. Entonces, la distancia entre los puntos
(B) y (C) se denomina de tiempo. Esta se puede medir de la misma manera que la
amplitud de impulso.
PW (x100) A → B (x100)
Ciclo de trabajo (%) = =
PERIODO A →C
15 ms x 100
Ciclo de trabajo de ejemplo = = 21.4%
70 ms
Página 28 06/2003
MANUAL DE INSTRUCCIONES OD-512
a. División 10 ms
b. División 2 ms
06/2003 Página 29
MANUAL DE INSTRUCCIONES OD-512
2. Ajuste el selector TIME DIV [29] para visualizar unos 2 ciclos de impulso.
Asegúrese de que el control x 10 MAG [31] no se encuentra pulsado.
4. Ajuste el selector VOLTS/DIV del canal CH1 [9] para conseguir que el pico
positivo del impulso exceda el 100 % de la línea de la retícula, y el pico del impulso
negativo exceda la línea 0 %, entonces gire el control VARIABLE [9] en el sentido
contrario a las agujas del reloj hasta que los picos de los impulsos positivo y negativo
descansen exactamente en las líneas 100 % y 0 % de la retícula (fig. 10).
5. Utilice el control POSITION [32] para desplazar la traza de forma que el
borde delantero pase a través de la intersección de la línea 10 % y la vertical central de
la retícula.
Página 30 06/2003
MANUAL DE INSTRUCCIONES OD-512
a. VISUALIZACIÓN BÁSICA
06/2003 Página 31
MANUAL DE INSTRUCCIONES OD-512
A-B
Mod.(%) = x 100
A +B
Página 32 06/2003
MANUAL DE INSTRUCCIONES OD-512
5.2.3 Reparación de TV
06/2003 Página 33
MANUAL DE INSTRUCCIONES OD-512
Método de doble traza. Este método funciona con cualquier tipo de forma de
onda. De hecho, a menudo incluso funciona si se comparan formas de onda diferentes.
Es efectivo para medir diferencias de fase grandes o pequeñas, en cualquier frecuencia
hasta 40 MHz.
NOTA
En las frecuencias altas utilice sondas idénticas y correctamente
compensadas, o longitudes iguales del mismo tipo de cable coaxial para
asegurar tiempos de retardo iguales.
2. Sitúe el selector Trigger SOURCE [23] en el canal que tenga la traza más
limpia y estable. Desplace temporalmente la traza del otro canal de la pantalla mediante
el control POSITION [11] o [19].
4. Use el control Trigger LEVEL [28] para asegurarse que la traza cruza la línea
horizontal central de la retícula al inicio del barrido o muy cerca de éste (fig. 14).
Página 34 06/2003
MANUAL DE INSTRUCCIONES OD-512
ADVERTENCIA
Reduzca la intensidad de la traza para que el punto sin deflexión no dañe
el fósforo del TRC.
3. Conecte una señal al conector CH1 [8], y la otra señal al conector CH2 [20].
06/2003 Página 35
MANUAL DE INSTRUCCIONES OD-512
Página 36 06/2003
MANUAL DE INSTRUCCIONES OD-512
NOTA
La conversión seno a ángulo se puede llevar a cabo utilizando tablas
trigonométricas o una calculadora científica.
06/2003 Página 37
MANUAL DE INSTRUCCIONES OD-512
Página 38 06/2003
MANUAL DE INSTRUCCIONES OD-512
6. MANTENIMIENTO
06/2003 Página 39
MANUAL DE INSTRUCCIONES OD-512
IDENTIFICADOR DE CARACTERÍSTICAS
POSICIÓN
F601 0.25 A - T - 250 V 5 x 20 mm para circuito impreso
PRECAUCIÓN
PRECAUCIÓN
La caja se limpiará con una ligera solución de detergente con agua y aplicada
mediante un paño suave humedecido.
Página 40 06/2003
INSTRUCTION MANUAL OD-512
TABLE OF CONTENTS
1. GENERAL .............................................................................................................. 1
1.1 DESCRIPTION ....................................................................................................... 1
1.2 SPECIFICATIONS ................................................................................................... 2
2. SAFETY RULES...................................................................................................... 5
2.1 GENERAL............................................................................................................. 5
2.2 SPECIFIC PRECAUTIONS ......................................................................................... 7
2.3 DESCRIPTIVE EXAMPLES OF OVER-VOLTAGE CATEGORIES......................................... 7
3. INSTALLATION....................................................................................................... 9
3.1 POWER SUPPLY ................................................................................................... 9
3.2 HANDLE POSITION ............................................................................................... 10
4. INSTRUCTIONS .................................................................................................... 11
4.1 DESCRIPTION OF CONTROLS AND ELEMENTS ........................................................... 11
4.2 START UP .......................................................................................................... 15
4.2.1 Preliminary Operations................................................................................. 15
4.2.2 Trace Rotation Adjustment ........................................................................... 16
4.2.3 Probe Compensation ................................................................................... 16
4.2.4 DC BAL Adjustments ................................................................................... 16
4.3 OPERATION METHOD .................................................................................... 17
4.3.1 Single-Channel Operation ............................................................................ 17
4.3.2 Dual-Channel Operation............................................................................... 18
4.3.3 ADD and SUBTRACTION Measurement ...................................................... 18
4.3.4 Triggering options ........................................................................................ 19
4.3.5 X-Y Operation.............................................................................................. 22
5. MEASUREMENT APPLICATIONS ........................................................................ 25
5.1 ONE-CHANNEL APPLICATIONS............................................................................... 25
5.1.1 Amplitude Measurements............................................................................. 25
5.1.2 Time Interval Measurement .......................................................................... 27
5.1.3 Risetime Measurement ................................................................................ 30
5.1.4 Amplitude modulation index measurement.................................................... 32
5.1.5 Frequency Measurement ............................................................................. 32
5.2 DUAL-CHANNEL APPLICATIONS ............................................................................. 33
5.2.1 Level comparison......................................................................................... 33
5.2.2 Stereo Appliance Repairing.......................................................................... 33
5.2.3 TV Repairing ............................................................................................... 33
5.2.4 Composite video signal analysis ................................................................... 33
5.2.5 Phase Difference Measurement ................................................................... 34
6. MAINTENANCE................................................................................................... 37
6.1 INSTRUCTIONS FOR RETURNING BY MAIL ................................................................. 37
6.2 MAINS FUSE REPLACEMENT .................................................................................. 37
6.3 FUSES THAT CANNOT BE REPLACED BY THE USER .................................................. 38
6.4 CLEANING RECOMMENDATIONS ............................................................................ 38
INSTRUCTION MANUAL OD-512
INSTRUCTION MANUAL OD-512
20 MHz OSCILLOSCOPE
OD-512
1. GENERAL
1.1 Description
The OD-512 is sturdy, easy to operate and exhibits high operational
reliability. This is a portable-type, dual-channel oscilloscope, its bandwidth of DC
is up to 20 MHz, and its maximum sensitivity is 1 mV/DIV. The time base provides
P
a maximum sweep time of 0.2 s/DIV. The sweep speed becomes 100 ns/DIV
after magnifying 10 times. The oscilloscope uses a 6-inch rectangular type
cathode-ray tube with red internal graticule.
3) Alternate triggering:
Even with an observation of two different frequency waveforms, each waveform
can be triggered stably.
4) TV sync triggering:
The oscilloscope has a sync separator circuit for TV-V and TV-H signals triggering.
5) CH1 Output:
CH1 output located on rear panel can be applied to frequency counter or other
instruments.
6) Z-Axis Input:
Beam intensity modulation capability permits time or frequency markers to be
added. Trace blank with positive signal, TTL compatible.
7) X-Y operation:
In this mode of operation CH1 can be applied as a horizontal deflection (X-axis)
while CH2 provides vertical deflection (Y-axis).
06/2003 Page 1
INSTRUCTION MANUAL OD-512
1.2 Specifications
Sensitivity 5 mV to 5 V/DIV, 10 steps in 1-2-5 sequence
Accuracy ≤ 3% (x 5 MAG: ≤5%)
Vernier Vertical
To 1/2.5 or less of panel-indicated value
sensitivity
DC to 20 MHz (x 5 MAG: DC to 7 MHz)
Frequency bandwidth AC coupling: Low limit frequency of 10 Hz
(With reference to 100 kHz, 8 DIV. Frequency response at –3 dB)
Rise time Approx. 17.5 ns (x 5 MAG: Approx. 50 ns)
Input impedance :
Approx. 1 M // Approx. 25 pF
Square Wave Overshoot: ≤ 5% (At 10 mV/DIV range)
Characteristics Other distortions and other ranges: 5% added to the above value
DC Balance Shift Panel adjustable
< 0.1 DIV of amplitude change when waveform of 2 DIV at graticule centre
Linearity
is moved vertically
VERTICAL CH1: CH1 single channel
AXIS CH2: CH2 single channel
Vertical modes DUAL: CH1 and CH2 are displayed. ALT or CHOP selectable at
any sweep rate
ADD: CH1 + CH2 algebraic addition
Chopping Repetition
Approx. 250 kHz
Frequency
Input Coupling AC, GND, DC
Maximum Input Voltage 300 V DC + peak (AC: frequency 1 kHz or lower)
Common Mode Rejection 50:1 or better at 50 kHz sinusoidal wave. (When sensitivities of CH1 and
Ratio CH2 are set equally)
Isolation between
>1000: 1 at 50 kHz
channels
>30: 1 at 20 MHz
(At 5m V/DIV range)
CH1 signal output At least 20 mV/div into a 50 terminal, Bandwidth is 50Hz to 5MHz at least.
CH2 INV BAL. Balanced point variation: ≤1 DIV (Reference at centre graticule)
CH1, CH2, LINE, EXT (CH1 and CH2 can be selected only in the DUAL or
ADD vertical mode).
Triggering source
In ALT mode, if the TRIG. ALT switch is pushed in, it can alternate
triggering of two different source.
Coupling AC: 20 Hz to full bandwidth
Slope +/-
20 Hz to 2 MHz: 0.5 DIV, TRIG-ALT: 2 DIV, EXT: 200 mV
Sensitivity 2 to 20 MHz: 1.5 DIV, TRIG-ALT: 3 DIV, EXT: 800 mV
TV: Sync pulse more than 1 DIV (EXT: 1V)
Triggering modes AUTO: Sweeps run in the free mode when no triggering input signal is
applied. (Applicable for repetitive signals of frequency 25 Hz or over)..
TRIGGERING NORM: When no triggering signal is applied, the trace is in the ready state,
but not displayed.
TV-V: This setting is used when observing the entire vertical picture of
television signal.
TV-H: This setting is used when observing the entire horizontal picture of
television signal.
(Both TV-V and TV-H synchronise only when the synchronising signal is
negative )
EXT Triggering Signal
Input Approx.: 1M ohm // approx. 25 Pf
Input Impedance 300 V (DC+AC peak), AC: Frequency not higher than 1 kHz
Max. Input Voltage
Sweep Time P
0.2 s to 0.5 s/DIV, 20 steps in 1-2-5 sequence
Sweep Time Accuracy 3%
Vernier Sweep Time
≤ 1/2.5 of panel-indicated value
Control
HORIZIONTAL Sweep Magnification 10 times (maximum sweep time 100 ns/DIV)
AXIS x10 MAG Sweep Time
Accuracy
5%, (20 ns to 50 ns are uncalibrated)
Linearity 3%, x10 MAG: 5% (20 ns and 50 ns are uncalibrated)
Position shift caused
Within 2 div. at CRT screen centre
by x 10MAG
Page 2 06/2003
INSTRUCTION MANUAL OD-512
Sensitivity Same as vertical axis (X-axis: CH1 input signal; Y-axis: CH2 input signal)
X-Y MODE Frequency Bandwidth DC to at least 500 kHz
X-Y Phase Difference ≤ 30 at DC to 50 kHz
Sensitivity 5 Vp-p (Positive-going signal decreases intensity)
Frequency Bandwidth DC to 2 MHz
Z AXIS
Input resistance :
Approx. 47 k
Maximum Input Voltage 30 V (DC+AC peak, AC frequency ≤ 1 kHz)
Waveform Positive-going Square wave
Frequency Approx. 1 kHz
CALIBRATION
Duty Ratio Within 48:52
2 Vp-p 2%.
VOLTAGE
Output Voltage
Output Impedance :
Approx. 1 k
Type 6-inch rectangular type, internal graticule
Phosphor P 31
Acceleration Voltage Approx. 2 kV
CRT
Effective Screen Size 8 x 10 DIV (1 DIV = 10 mm (0.39 in)).
Graticule Internal
Trace Rotation Provided
POWER SUPPLY
AC 115 V (97 to 132 V), 230 V (195 to 250 V) selectable,
Mains voltage
50 Hz or 60 Hz
Consumption 35 W
OPERATING ENVIRONMENTAL CONDITIONS
Indoor use
Altitude Up to 2000 m
To satisfy specifications: 10 º to 35 ºC
Temperature range
Maximum operating ranges: 0 º to 40 ºC
Relative humidity 85 % maximum, non condensing
Storage temperature
From –10 º to 70 ºC, 70% maximum
and relative humidity
MECHANICAL SPECIFICATIONS
Dimensions 310 W x 150 H x 455 D mm
Weight Approx. 8 Kg (17,6 lbs)
ACCESSORIES
1 x Mains cord CA-006
06/2003 Page 3
INSTRUCTION MANUAL OD-512
Page 4 06/2003
INSTRUCTION MANUAL OD-512
2. SAFETY RULES
2.1 General
* Use this equipment connected only to systems with their negative of measurement
connected to ground potential.
* This is a Class I equipment, for safety reasons plug it to a supply line with the
corresponding ground terminal.
* When using some of the following accessories use only the specified ones to ensure
safety.
Mains cord.
* The user is only authorised to carry out the following maintenance operations:
Mains fuse replacement, that should fit indicated type and value.
Any other change on the equipment should be carried out by qualified personnel.
06/2003 Page 5
INSTRUCTION MANUAL OD-512
DIRECT CURRENT
ALTERNATING CURRENT
GROUND TERMINAL
PROTECTIVE CONDUCTOR
FRAME TERMINAL
EQUIPOTENTIALITY
ON (Supply)
OFF (Supply)
DOUBLE INSULATION
(Class II protection)
CAUTION
(Risk of electric shock)
FUSE
Page 6 06/2003
INSTRUCTION MANUAL OD-512
Do not use the equipment immediately after moving it to a room with very
different temperature. Wait for a while until the equipment reaches room temperature.
Do not apply voltages that exceed the limits of probes or input connectors.
3.
06/2003 Page 7
INSTRUCTION MANUAL OD-512
Page 8 06/2003
INSTRUCTION MANUAL OD-512
3. INSTALLATION
CAUTION:
FACTORY SETS INSTRUMENT MAINS VOLTAGE TO 230 V
06/2003 Page 9
INSTRUCTION MANUAL OD-512
OD-512 oscilloscope has a movable handle for transport and for positioning the
oscilloscope to obtain the best angle with the display.
To change handle position, take the handle supports that are in contact with the
oscilloscope chassis and detach slightly from the chassis so you can change its
position.
Page 10 06/2003
INSTRUCTION MANUAL OD-512
4. INSTRUCTIONS
[1] CAL. The terminal provides a calibrated amplitude square wave of 2 Vp-p at
1 kHz for probe adjustment and vertical amplifier calibration.
CRT
[2]. INTEN. Control the brightness of the spot or trace. The brightness increases
when turning the control clockwise.
[4] TRACE ROTATION. Semi-fix potentiometer for aligning the horizontal trace in
parallel with graticule lines.
[5] POWER led. It lights when the POWER switch of the oscilloscope is ON.
06/2003 Page 11
INSTRUCTION MANUAL OD-512
[8] CH1 (X). The input terminal of vertical channel CH1. In X-Y operation mode is X-
axis.
[20] CH2 (Y). Input terminal of vertical channel CH2. In operation mode X-Y is Y-axis.
[10] [18] AC/GND/DC. Select connection mode between input terminal and vertical
amplifier for channel CH1 and CH2 respectively.
AC Blocks DC component of the input signal.
GND Connect the vertical amplifier of the CH1-CH2 to ground, establishing a
ground reference.
DC Connect CH1-CH2 amplifier to the input terminal, visualising AC & DC
component of the signal.
[7] [22] VOLTS/DIV. Selects the CH1 / CH2 vertical axis sensitivity from 5 mV/DIV to
5 V/DIV with 10 ranges totally.
[9] [21] VARIABLE. Fine adjustment of sensitivity of CH1 & CH2 vertical axis respectively
with a factor of ≥ 1/2.5 of the indicated value. The sensitivity is calibrated to
specific value in the CAL position of VARIABLE control. When this knob is pulled
out (x5 MAG state), it will multiply 5 by the amplifier sensitivity.
[13] [17] DC BAL. The knobs are used for adjusting the DC balance of the CH1/CH2 input
amplifiers. See paragraph 4.2.4 DC level Adjustment for details.
[11] [19] POSITION. Vertical positioning control of the CH1/CH2 trace on the CRT
display. Clockwise turning shifts trace up and anticlockwise turning shifts trace
down.
Page 12 06/2003
INSTRUCTION MANUAL OD-512
TRIGGERING
[24] TRIG IN terminal. It allows to apply an external triggering signal to the triggering
circuits. Input terminal is used for external triggering signal.
[27] TRIG. ALT. When switch VERT MODE [14] is set in DUAL or ADD state and
switch SOURCE [23] is set in CH1 or CH2, with the engagement of this switch, it
will alternately select CH1 & Ch2 for the internal triggering source signal.
[28] LEVEL. Selects the amplitude of the signal that makes the triggering occur. That
is, the start point of the waveform. When turning it clockwise (+), the triggering
level moves upward on the display waveform. When turning it anticlockwise (-),
the triggering level moves downward on the display waveform.
[29] TIME/DIV. Selects sweep time. Sweep time ranges are available in 20 steps from
0.2 µs/DIV to 0.5 s/DIV. Position X-Y se selected when using the instrument in
the X-Y mode.
[30] SWP. VAR. Vernier and continuously varied control of sweep time between
positions of the selector TIME/DIV [29].
TIME/DIV calibrations are exact only when VARIABLE control is completely
turned clockwise and a “click” is produced.
06/2003 Page 13
INSTRUCTION MANUAL OD-512
[32] POSITION. Horizontal positioning control of the traces on the CRT display.
When turning this control clockwise the trace moves to the right on the CRT.
When turning this control anticlockwise the trace moves to the left on the CRT.
Rear Panel
[34] Z AXIS INPUT. Input terminal for external CRT intensity modulation signal. Trace
brightness is reduced when a positive signal is applied and it is increased when a
negative signal is applied.
[35] CH1 OUTPUT input. Delivers the CH1 signal with a voltage of approximately
20 mV per 1 DIV into a 50 :
termination. Suitable for frequency counters and
other instruments.
[36] Power input connector. It allows to connect or disconnect the mains cord.
[37] Fuse and line voltage selector. It allows to change the operating voltage range.
[38] Studs for laying the oscilloscope on its back to operate it in the upward posture.
Also used to take up the power cord.
Page 14 06/2003
INSTRUCTION MANUAL OD-512
4.2 Start Up
Connect power cord to power input connector [36], then connect it to an AC line
outlet suitable for the instrument previously set AC line voltage, and then continues as
follows:
1. Press the I/O switch [6]. The POWER [5] led will be on. In about 20 seconds, a
trace will appear on the CRT screen. If no trace appears in about 60 seconds, check the
switch and control setting (previous paragraphs).
2. Adjust the trace to an appropriate brightness and image with INTEN [2] and
FOCUS [3] control knob respectively.
WARNING
Ageing-resistance material has been used in the CRT. In any case, the
screen can be damaged if a spot or a trace extremely bright is set on it
for a long time. So if an intense brightness is needed for a measurement
make sure to decrease intensity (INTEN) just after the measurement is
performed. You should get used to decrease brightness when your not
going to observe the oscilloscope for a while too.
3. Align the trace with the horizontal central line of the graticule by adjusting the
CH1 POSITION [11] control knob.
4. Adjust the POSITION [32] control knob to align the left end of the trace with
the graticule left end line.
06/2003 Page 15
INSTRUCTION MANUAL OD-512
Preadjust the instrument until a trace just as indicated in the previous paragraph is
obtained.
Align the trace with the horizontal central line of the graticule by adjusting the CH1
POSITION [11] control knob.
If the trace is not parallel to the central line, adjust the TRACE ROTATION [4]
potentiometer by means of a suitable screwdriver until the trace becomes parallel.
Connect a probe (x10) to the CH1 [8] input terminal and connect the probe tip to the
CAL 2 Vpp [1] terminal. On the CRT screen should be displayed a square waveform with
four divisions amplitude (4 x 0.5 V/DIV = 2 Vpp).
Remove the probe from CAL 2 Vpp [1] terminal. Set switch V-MODE [14] to CH2
position and repeat the process for CH2, each one with its own probe.
The attenuator (ATT) balance of the vertical axis can be made easily.
1. Set the input coupling switches of CH1 [10] and CH2 [18] to GND and set the
TRIG MODE [25] to AUTO. Then position the base line to the centre.
Page 16 06/2003
INSTRUCTION MANUAL OD-512
Single trace and base time operation and internal triggering is the most basic
operating mode of the OD-512. Select this mode when there is only one signal to be
observed, without the inconvenience of other traces on the CRT screen. Due it is a two
channel oscilloscope, any of these two channels can be selected. CH1 is provided with and
output terminal: use the CH1 output terminal on the rear panel to measure signal frequency
using a frequency counter while observing waveform. CH2 have a switch that inverts
polarity, although it adds flexibility, it has no use when operating in single trace mode.
1. Set the following controls of the instrument as shown below. Observe the
selected triggering source (CH1 or CH2 SOURCE) corresponds to the selected channel
(CH1 or CH2 V MODE).
3. Connect the signal to be observed to the appropriate input terminal CH1 [8] or
CH2 [20] and adjust the appropriate VOLTS/DIV selector [7] or [22] so that signal
waveform is completely displayed on the screen.
WARNING
Do not apply signal higher than 300 V (DC + AC peak)
4. Set TIME/DIV [29] so that the appropriate number of signal cycles appears on
screen. For some measurements the appropriate number of cycles is only two or 3 cycles;
for other measurements the appropriate number of cycles is 50-100 (displayed as a band).
If necessary, adjust Trigger LEVEL [28] control for a stable display of the signal.
5. If the signal is so weak that even in the 5 mV position of the VOLTS/DIV [7] or
[22] selector the trace is not high enough for triggering or for the image to be useful, pull out
the button of the VAR PULL x5 MAG [9] or [21] control. This provides a sensitivity of
2 mV/DIV when VOLTS/DIV [7] or [22] is set to 10 mV and of 1 mV/DIV when it is set
to 5 mV. In any case, when this mode is selected, the bandwidth is reduced to 7 MHz, and
the trace noise can be considerable.
06/2003 Page 17
INSTRUCTION MANUAL OD-512
P
6. If the frequency of signal is so high that even in the 0.2 s position of the
TIME/DIV [29] selector appears too many cycles on screen, push the x10 MAG [31] key.
P
This increases the effective sweep speed 10 times, so 0.2 s/DIV position now corres-
P
ponds to 20 ns/DIV, 0.5 s position corresponds to 50 ns/DIV, etc. 0.2 and 0.5 s x 10 P
P
positions are not calibrated, whereas 1 s and lower positions are calibrated.
7. If the frequency of the signal is very low or its zero (DC signal), so the AC
coupling attenuates or distorts the signal, set AC/GND/DC [10] or [18] selector to DC.
Trigger MODE [25] switch should be set to NORM again if the frequency of the
signal is lower than 25 Hz, and possibly readjust the Trigger LEVEL [28] control.
Set VERT MODE [14] selector to the DUAL position so the channel 2 signal is
displayed too. The operation is identical to the one described in the section 4.3.1 with the
following exceptions:
1. Sweep mode can be set to ALT (alternate) or CHOP by means of the ALT
CHOP [12] control. ALT mode is useful when measuring signals with a quite high frequency
(TIME/DIV [29] selector set to 0.2 ms/DIV or faster). CHOP mode is useful for signals with a
quite low frequency (TIME/DIV [29] selector set to 0.5 ms or slower), in this mode the
signals are chopped and are displayed at a frequency of 250 KHz.
2. If both channels displays signals with the same frequency, select by means of
the Trigger SOURCE [23] the channel with the waveform with the sharpest slope. If the
signals do not have the same frequency but they are harmonics, use the channel with the
lower frequency signal for triggering. Bear in mind too that if the channel used for triggering
is disconnected, the image will not be stable.
Add and subtraction operations are two types of dual-channel operation in which
the two signals are combined to display a unique trace. In ADD operation the resultant trace
displays the algebraic sum of the CH1 and CH2 signals. In subtraction operation the
resultant trace displays the algebraic difference of CH1 and CH2 signals.
2. Make sure that VOLTS/DIV [7] and [22] selectors are set to the same position
and that VARIABLE [9] and [21] are set to CAL position. If signal levels are very different,
set both VOLTS/DIV selectors to the position that completely displays on screen the signal
with the highest amplitude.
3. Use the channel with the highest amplitude signal for triggering.
4. Set MODE [14] selector to ADD position. This way the unique resultant trace will
be the algebraic sum of the CH1 and CH2 signals. POSITION [11] or [19] can be used
to positioning the resultant trace.
Page 18 06/2003
INSTRUCTION MANUAL OD-512
NOTE
If the input signals are in phase, the amplitude of the resultant trace will be
the arithmetic sum of the individual traces (for ex. 4.2 div + 1.2 div = 5.4 div).
If the phase difference of the input signals is 180º, the resultant amplitude is
the arithmetic difference (fore ex. 4.2 div – 1.2 div = 3.0 div).
5. If the peak-peak amplitude of the resultant trace is very small, turn both
VOLTS/DIV [7] and [22] selectors to increase the height of the displayed trace. Make sure
that both selectors are set to the same position.
To obtain the subtraction, proceed as before but now push the CH2 INV [16] key.
The resultant trace will be the algebraic difference of the CH1 and CH2 signals. Now, if the
input signals are in phase, the resultant trace will be the arithmetic difference of the
individual traces (for ex. 4.2 div – 1.2 div = 3.0 div). If the input signals have a phase
difference of 180 º, the resultant trace amplitude will be the arithmetic sum of the individual
traces (for ex. 4.2 div + 1.2 div = 5.4 div).
When NORM triggering mode is selected, the CRT beam does not sweep
horizontally the screen until the selected triggering signal crosses the threshold defined by
means of the TRIG LEVEL [28] control. Nevertheless, it is not suitable to use this triggering
mode as default because, if no signal is present or if controls are not properly set, no trace
will be displayed on CRT screen. A lot of time can be spent to find out that this absence of
traces is due to an incorrect setting of the Vertical POSITIONING [11] or [19] or VOLT/DIV
[7] or [22] selector.
Trigger AUTO mode solves this problem provoking the time base to free runs
without a trigger signal. This generates a single horizontal line when no signal is present
and a vertical deflection display without triggering, when exists a vertical signal but
triggering control is misplaced. So, this way, it can be automatically found out what is
wrong. The only problem of the AUTO mode is that signal with a frequency lower than
25 Hz can not certainly trigger the time base, nor complex signals of any frequency,
possibly. So, usually Trigger MODE [25] selector is set to AUTO position, but setting it to
NORM position if any signal (especially if its frequency is lower than 25 Hz) can not
displayed as a stable waveform.
The TV-V and TV-H positions of the Trigger MODE separate de TV synchronism
signal so a clear triggering signal, both vertical and horizontal, can be extracted from the
composite video signal (fig. 7a). Setting the Trigger MODE selector to TV-V permits
selection of vertical sync pulses for sweep triggering. Vertical sync pulses are selected as
trigger to permit viewing of vertical fields and frames of video. Setting the Trigger MODE
selector to TV-H position permits selection of horizontal sync pulses for sweep triggering
when viewing composite video waveforms. Horizontal sync pulses are selected as trigger to
permit viewing of horizontal lines of video. This oscilloscope synchronises with only (-)
negative polarity, that is, the sync pulses are negative and the video is positive (figure 7d).
06/2003 Page 19
INSTRUCTION MANUAL OD-512
The displayed signal itself or a trigger signal, which has a time relationship with the
displayed signal is required to be applied to the trigger circuit to display a stationary signal
on the CRT screen. The SOURCE [23] switch is used for selecting these trigger sources.
CH2: The signal applied to the vertical input terminal is branched off away from
the preamplifier and is fed to the trigger circuit through the VERT MODE
switch. Since the trigger signal is the measured signal itself, a stable
waveform can be readily displayed on the CRT screen. When in the DUAL
or ADD operation, the selected signal through the SOURCE switch is used
as a trigger source signal
LINE: The AC power line frequency is used as the triggering signal This method
is effective when the measured signal has a relationship with the AC line
frequency, especially for measurements of low level AC noise of audio
equipment, thyristor circuits, etc.
Page 20 06/2003
INSTRUCTION MANUAL OD-512
EXT: The sweep is triggered by an external signal applied to the external trigger
input terminal. An external signal which has a periodic relationship with the
measured signal is used. Because the measured signal is not used as the
triggering signal, the waveforms can be displayed more independent than
the measured signal
LEVEL [28] controls allows to adjust the start of the sweep to almost any desired
point on a waveform. The effect of the LEVEL control over the trace is shown in figure 6c.
Marks + and – of this control panel refers to the cross point or zero of the waveform and to
the more positive (+) and negative (-) points than this one. If the trigger slope is very steep,
as in square waves or digital pulses, the will be no apparent change in the displayed trace
until LEVEL control is rotated past the most positive or negative point, in which case the
display will free run (AUTO sweep mode) or disappear completely (NORM sweep mode).
Try to trigger at the mid point of slow-rise waveforms (as sine and triangular waveforms)
since these are usually cleanest spots on such waveforms.
The SLOPE [26] switch selects the slope (polarity) triggering signal (figure 6)). That
is to say when the SLOPE [26] switch is set to (+) position (up), the sweep is developed
from the trigger source waveform as it crosses the threshold level in a positive-going
direction. When it is set to (-) position (down), a sweep trigger is developed from the trigger
source waveform as it crosses the threshold level in negative-going direction. Always select
the steepest and the most stable slope or edge. For example, small changes in the
amplitude of the sawtooth shown in figure 6a will cause jittering (distortion) if the time base
is triggered on the positive (ramp) slope, but have no effect if triggering occurs on the
negative slope (a fast-fall edge). In the example shown in figure 6b, both leading and tailing
edges are very steep trace to jitter, making observation difficult. Triggering from the stable
leading edge (+slope) yields a trace that has only the tailing-edge jitter of the original signal.
If you are ever in doubt, or have an unsatisfactory display, try both slopes to find the best
way.
06/2003 Page 21
INSTRUCTION MANUAL OD-512
Figure 6.- Functions of TRIG LEVEL [28] control and SLOPE [26] switch
TRIG ALT [27] switch is used to select alternate triggering and alternate display
when the DUAL-trace VERT MODE is selected (the switch has effect in the CH1, CH2, or
ADD modes). In the alternate triggering mode (when select dual-trace operation), the
trigger source alternates between CH1 and CH2 with each sweep. This is convenient for
checking amplitudes, wave-shape, or waveform period measurements, and even permits
simultaneously observing two waveforms which are not related to frequency or period.
However, this setting is not suitable for phase or timing comparison measurements. For
such measurements, both traces must be triggered by the same sync signal.
Page 22 06/2003
INSTRUCTION MANUAL OD-512
The internal timebase of the OD-512 are not utilised in X-Y operation; deflection in
both the vertical and horizontal directions is via external signals. Vertical channel 1 serves
as the X-axis (horizontal) signal processor, so horizontal and vertical axis have identical
control facilities.
All of the V MODE, and trigger switches, as well as there associated controls and
connectors, are inoperative in the X-Y mode.
1. Set the TIME/IDV [29] control to the X-Y position (fully clockwise).
CAUTION
Reduce the trace intensity, lest the undeflected spot damage the CRT
phosphor.
2. Apply the vertical signal to the CH2 or Y IN connector, and the horizontal signal
to the CH1 or X IN [8] connector. Once the trace is deflected, restore normal brightness.
3. Adjust X the trace position vertically (Y-axis) with the CH2 Vertical POSITION
control [19]. Adjust the trace position horizontally (X-axis) with the Horizontal POSITION
[32] control. The CH1 Vertical POSITION [11] control has no effect during X-Y operation.
4. Adjust the trace height with the CH2 VOLTS/DIV [22] switch and trace width with
the CH1 VOLTS/DIV [7] switch. The x5MAG [9] and [21] switches on the VARIABLE
controls can be used if greater is necessary.
5. Vertical (Y-axis) signal may be inverted setting CH2 INV [16] switch.
06/2003 Page 23
INSTRUCTION MANUAL OD-512
Page 24 06/2003
INSTRUCTION MANUAL OD-512
5. MEASUREMENT APPLICATIONS
The modern triggered sweep oscilloscope has two major measurement functions.
The first of these is amplitude. The oscilloscope has an advantage over most other forms of
amplitude measurement in that complex as well as simply waveforms can be totally
characterised (i.e. complete voltage information is available)
Oscilloscope voltage measurements generally fall into one of two types: peak to
peak or instantaneous. Peak to peak (p-p) measurement simply notes the total amplitude
between extremes without regard to polarity reference. Instantaneous voltage
measurement indicates the exact voltage from each every point on the waveform to a
ground reference. When making either type of measurements, make sure that the
VARIABLE controls are click-stopped fully clockwise.
1. Set up the oscilloscope for the vertical mode operation proceeding as indicated
in paragraph 4.3.
2. Adjust the TIME/DIV [29] switch for two or three cycles of waveform, and set the
VOLTS/DIV switch [77] or [22] for the largest-possible totally-on-screen display.
3. Use the appropriate vertical POSITION control [11] or [19] to position the
negative signal peaks on the nearest horizontal graticule line below the signal peaks (fig. 9).
4. Use the horizontal POSITION control [32] to position one of the positive peaks
on the central vertical graticule line. This line has additional calibration marks equal to 0.2
major division each.
5. Count the number of divisions from the graticule line touching the negative signal
peak to the intersection of the positive signal peak with the central vertical graticule line.
Multiply this number by the VOLTS/DIV switch setting to get the peak-to peak voltage of the
waveform. For example, if the VOLTS/DIV switch were set to 2 V, the waveform shown in
figure 9 would be 8.0 Vp-p (4.0 DIV x 2 V).
7. If measuring a sine wave below 100 Hz, or a rectangular wave below 1000 Hz,
flip the AG/GND/DC switch to DC.
06/2003 Page 25
INSTRUCTION MANUAL OD-512
Page 26 06/2003
INSTRUCTION MANUAL OD-512
1. Set up the oscilloscope for the vertical mode operation proceeding as indicated
in paragraph 4.3.
2. Adjust the TIME/DIV switch [29] for one complete cycle of waveform, an set the
VOLTS/DIV switch for the trace amplitude of 4 to 6 division (see figure 8).
4. Use the appropriate vertical POSITON control [11] or [19] to set the trace on the
central horizontal graticule line. However, if you know the signal voltage is wholly positive
use the bottom most graticule line. If you know the signal voltage is wholly negative use the
top most graticule line. This line is the so called ground-reference line.
NOTE:
Vertical POSITION control setting can not be changed until the measurement
ends.
5. Flip the AC/GND/DC switch [10] to DC. The polarity of all points above the
ground reference line is positive; all points below the ground-reference line are negative.
NOTE:
Make certain the waveform is not riding on a high-amplitude DC voltage wave
before flipping the AC/GND/DC switch.
6. Use the Horizontal POSITION control [32] to position any point of interest on the central
vertical graticule line. This line has additional calibration marks equal to 0.2 major division
each. The voltage relative to ground at any point selected is equal to the number of division
from that point to the ground reference line multiplied by the VOLTS/DIV setting. In the
example used for figure 10, the voltage for a 0.5 V/DIV scale is 2.5 V (5.0 DIV x 5 V).
Basic Technique. The basic technique for measuring time interval is described in
the following steps. This same technique applies to the specific procedures and variations
that follow.
2. Set TIME/DIV [29] switch so the desired interval appears on the screen as large
as possible. Make sure SWP:VAR control [30] is at CAL position (fully clockwise). If it is
not, any measured value may be incorrect.
3. Use the Vertical POSITION control [11] or [19] to position the trace so the central
horizontal graticule line passes through the points on the waveform between which you
want to make the measurement.
06/2003 Page 27
INSTRUCTION MANUAL OD-512
4. Use Horizontal POSITION control [32] to set the left-most measurement point on
a nearly vertical graticule line.
5. Count the number of horizontal divisions between the step 4 graticule line and
the second measurement point. Measure to tenth of a major division. Note that each minor
division on the central horizontal graticule line is 0.2 major division.
6. To determine the time interval between the two measurement points, multiply the
number of horizontal divisions counted in step 5 by the setting of the TIME/DIV switch. If the
x10 MAG [31] is set to x10 (x10 magnification), be certain to divide the TIME/DIV switch
setting by 10.
Period, pulse width, and duty cycle. The basic technique in the preceding
paragraph can be used to determine pulse parameters such as period, pulse width, duty
cycle etc.
The period of a pulse or any other waveform is the time it takes for one full cycle of
the signal. In figure 9a, the distance between points (A) and (C) represents one cycle; the
time interval of this distance is the period. The time scale for the CRT display of figure 9 is
10 ms/DIV, so the period is 70 milliseconds in this example.
Pulse width is 15 milliseconds. However, 1.5 divisions is a rather small distance for
accurate measurements, so it is advisable to use a faster sweep speed for this particular
measurement. Increasing the sweep to 2 ms/DIV as in figure 9b gives a large display,
allowing more accurate measurement.
An alternative technique useful when pulse width is smaller than a division consists
of pulling x10 MAG button [31] to obtain a magnification x10, and repositioning the pulse on
the screen with the POSITION control [32]. In some applications the pulse width is
called offeime. Then the distance between points (B) and (C) is then called offeime. This
can be measured in the same manner as pulse width.
When pulse width and period are known, duty cycle can be calculated. Duty cycle is
the percentage of the period (or total on and off times) represented by the pulse width (on
time), that the pulse is at its high level.
PW(100) A → B(100)
Duty cycle (%) = =
Period A→C
15 ms x 100
Duty cycle of example = = 21.4%
70 ms
Page 28 06/2003
INSTRUCTION MANUAL OD-512
a. 10 ms division.
b. 2 ms division
06/2003 Page 29
INSTRUCTION MANUAL OD-512
Rise time is the time for the leading edge of a pulse to rise from 10% to 90% of the
total pulse amplitude.
Fall time is the time required for the trailing edge of a pulse to drop from 90% of
total pulse amplitude to 10%. Rise time and fall time, which may be collectively called
transition time, are measured in essentially the same manner.
1. Connect the pulse to be measured to CH1 [8] and set the AC/GND/DC switch
[10] to AC.
2. Adjust the TIME/DIV selector [29] to display about two cycles of the pulse. Make
certain the x 10 MAG [31] control is not pulled.
3. Centre the pulse vertically with the CH1 Vertical POSITION control [11]
4. Adjust the CH1 VOLTS/DIV switch [9] to make the positive pulse peak exceed
the 100% graticule line, and the negative pulse peak exceed the 0% line. Then rotate the
VARIABLE control [9] clockwise until the positive and negative pulse peaks rest exactly on
the 100% and 0% graticule lines (figure 10).
5. Use the horizontal POSITION control [32] to shift the trace so the leading
edge passes through the intersection of the 10% and central vertical line.
6. If the rise time slow compared to the period, no further control manipulations are
necessary. If the rise time is fast (leading edge almost vertical) , set x10 MAG control [31]
to x10 MAG position and reposition the trace as in step 5.
7. Count the number of horizontal divisions between the central vertical line (10%
point) and the intersection of the trace with the 90% line.
8. Multiply the number of divisions counted in step 7 by the setting of the TIME/DIV
switch [29] to find the measured rise time. If x10 magnification is used, divide the TIME/DIV
setting by 10. For example, if the time base setting in figure 10 was 1 s/DIV (1000 ns), the
rise time would be 360 ns (1000 ns / 10 = 100 ns, 100 ns x 3.6 DIV = 360 ns).
9. To measure fall time, simply shift the trace horizontally until a tailing edge passes
through the 10% and centre vertical graticule lines, and repeat steps 7 and 8.
Page 30 06/2003
INSTRUCTION MANUAL OD-512
06/2003 Page 31
INSTRUCTION MANUAL OD-512
A -B
Mod(%) = x100
A+B
Page 32 06/2003
INSTRUCTION MANUAL OD-512
Connect the different elements as shown in the previous figure and set CH1 and
CH2 switches to the same position. Use the Horizontal POSITION control [11] and [19]
to shift the traces until they overlap. Amplitude difference will be due to the amplifier. Adjust
VOLTS/DIV and VARIABLE switches of one channel so the traces overlap completely. if
any difference exists, it is due to distortion. To display distortion only, invert the polarity of
the channel two signal by means of CH2 INV [16] and select ADD mode. The remaining
signal corresponds to the distortion. When no distortion is present a horizontal straight line
would be displayed.
5.2.3 TV Repairing
OD-512 includes a TV synchronism separator circuit that makes easy the horizontal
or vertical synchronism observation, blanking pulses, image components, as well
composite video signal (complete).
06/2003 Page 33
INSTRUCTION MANUAL OD-512
Phase difference or phase angle between two signals can be measured using the
dual-trace feature of the oscilloscope, or by operating the oscilloscope in the X-Y mode.
Dual-trace Method. This method works with any type of waveform. In fact, it will
often work even if different waveforms are being compared. This method is effective in
measuring large or small differences in phase, at any frequency up to 40 MHz.
NOTE:
At high frequencies use identical and correctly-compensated probes, or
equal lengths of the same type of coaxial cable to ensure equal delay times.
2. Position the Trigger SOURCE switch [23] to the channel with the cleanest and
most stable trace. Temporally move the other channel’s trace off the screen by means of its
Vertical POSITION control [11] or [19].
3. Centre the stable trace (trigger source) with its Vertical POSITION control
[11] or [19], and adjust its amplitude to exactly 6 vertical by means of its VOLTS/DIV switch
[7] or [22] and VARIABLE control [9] or [21].
4. Use the Trigger LEVEL control [28] to ensure that the trace crosses the central
horizontal graticule line at or near the beginning of the sweep (figure 14).
5. Use the TIME/DIV switch [29], and the VARIABLE control [30] and the Horizontal
POSITION control [32] to display one cycle of trace over 7.2 divisions. When this is
done, each major horizontal division represents 50º and each minor division represents 10º.
Page 34 06/2003
INSTRUCTION MANUAL OD-512
6.Move the off-screen trace back on the CRT with its Vertical POSITION
control [11] or [19], centring it vertically. Use the associated VOLTS/DIV switch [7] or [22]
and VARIABLE control [9] or [21] to adjust its amplitude to exactly 6 vertical divisions.
8. If the phase difference is less than 50º (one major division), set the x10 MAG
switch to x10 position and use the Horizontal POSITION control [32] (if needed) to
position the measurement area back on screen. With x10 magnification, each major
division is 5º and each minor division is 1º.
Lessajous Pattern Method. This method is used primarily with sine waves.
1. Rotate the TIME/DIV switch [29] fully clockwise to its X-Y position.
WARNING
Reduce the trace intensity least the undeflected spot damage the CRT
phosphor.
2. Make sure that x10 MAG [18] control is engaged so, if it is not, an error of 180º
would be introduced.
3. Connect one signal to CH1 connector [8], and other signal to the CH2
connector [20].
5. Adjust the CH1 VOLTS/DIV switch [7] for the largest-possible on-screen display.
7. Count the number of division subtended by the trace along the central vertical
graticule line (dimension A). You can now shift the trace vertically with CH2
POSITION
control [19] to a major division line for easier counting.
8. The phase difference (θ angle) between the two signals is equal to the arcsine of
dimension A : B (the step 7 number divided by 6). For example, the step 7 value of the
figure 15 pattern is 2.0. Dividing this by 6 yields 0.3334, whose arcsine is 19.5º.
9. The simple formula in figure 15 works for angles less than 90º. For angles over
90º (leftware tilt), add 90º to the angle found in step 7. Figure 15 shows the Lissajous
patterns of various phase angels: use this as guide in determining whether or not to add the
additional 90º.
NOTE:
The sine-to-angle conversion can be accomplished by using trigonometric
tables or a scientific calculator.
06/2003 Page 35
INSTRUCTION MANUAL OD-512
Page 36 06/2003
INSTRUCTION MANUAL OD-512
6. MAINTENANCE
3) Replace the damaged fuse for another one with the same characteristics.
4) Replace the fuse holder lid, aligning mark with the desired mains voltage mark.
06/2003 Page 37
INSTRUCTION MANUAL OD-512
CAUTION
To clean the cover, be sure the unit is turned off.
CAUTION
Do not use scented hydrocarbons or chlorized solvents. Such products may attack
the plastics used in the construction of the cover.
The cover should be cleaned by means of a light solution of detergent and water
applied with a soft cloth.
Page 38 06/2003
MANUEL D’UTILISATION OD-512
SOMMAIRE
1. GÉNÉRALITÉS....................................................................................................... 1
1.1 DESCRIPTION...................................................................................................... 1
1.2 SPÉCIFICATIONS ................................................................................................. 2
2. PRESCRIPTIONS DE SÉCURITÉ............................................................................ 5
2.1 GÉNÉRALES ....................................................................................................... 5
2.2 PRÉCAUTIONS SPÉCIFIQUES ................................................................................. 7
2.3 EXEMPLES DE CATÉGORIES DE SURTENSION .......................................................... 7
3. INSTALLATION....................................................................................................... 9
3.1 POWER SUPPLY .................................................................................................. 9
3.2 ORIENTATION DE L’ANSE..................................................................................... 10
4. MODE D’EMPLOI .................................................................................................. 11
4.1 DESCRIPTION DES COMMANDES ET DES ÉLÉMENTS ................................................ 11
4.2 MISE EN MARCHE .............................................................................................. 16
4.2.1 Opérations préliminaires........................................................................... 16
4.2.2 Réglage de la rotation de la trace.............................................................. 17
4.2.3 Réglage des sondes................................................................................. 17
4.2.4 Réglage du niveau de DC......................................................................... 17
4.3 FORME D’UTILISATION ........................................................................................ 18
4.3.1 Opération avec une seule trace ................................................................ 18
4.3.2 Opération avec double trace ..................................................................... 19
4.3.3 Mesures additive et soustractive ............................................................... 19
4.3.4 Options de déclenchement ....................................................................... 20
4.3.5 Fonctionnement en mode X-Y .................................................................. 24
5. APPLICATIONS .................................................................................................... 25
5.1 APPLICATIONS EN UTILISANT UN SEUL CANAL ......................................................... 25
5.1.1 Mesures d’amplitude ................................................................................ 25
5.1.2 Mesures d’intervalles de temps................................................................. 27
5.1.3 Mesure du temps de montée .................................................................... 30
5.1.4 Mesure de la profondeur de modulation en amplitude ................................ 32
5.1.5 Mesure de fréquence................................................................................ 32
5.2 APPLICATIONS UTILISANT LES DEUX CANAUX ......................................................... 33
5.2.1 Comparaison de niveaux .......................................................................... 33
5.2.2 Réparation d’appareils stéréo ................................................................... 33
5.2.3 Réparation de TV ..................................................................................... 33
5.2.4 Analyse de vidéo composée ..................................................................... 33
5.2.5 Mesures de différence de phase ............................................................... 34
6. ENTRETIEN .......................................................................................................... 39
6.1 INSTRUCTIONS D’ENVOI ...................................................................................... 39
6.2 REMPLACEMENT DU FUSIBLE DE SECTEUR ............................................................ 39
6.3 FUSIBLES NE POUVANT PAS ÊTRE SUBSTITUÉS PAR L’UTILISATEUR............................ 40
6.4 RECOMMANDATIONS DE NETTOYAGE .................................................................... 40
MANUEL D’UTILISATION OD-512
MANUEL D’UTILISATION OD-512
OSCILLOSCOPE DE 20 MHz
OD-512
1. GÉNÉRALITÉS
1.1 Description
L’OD-512 est un instrument portatif, robuste, facile à utiliser et fabriqué sous un
strict contrôle de qualité. Il s’agit d’un oscilloscope de double canal, avec une largeur de
bande de DC à 20 MHz et une sensibilité maximum de 1 mV/DIV. La base de temps
permet une vitesse de balayage maximum de 0.2 µs/DIV. Lorsque l’on multiplie à l’aide de
l’agrandissement (MAG) par 10, la vitesse de balayage est de 100 ns/DIV. L’oscilloscope
possède un écran rectangulaire de 6 pouces avec une grille interne rouge.
Étant donné qu’il possède une vaste gamme de possibilités, il s’agit d’un
équipement spécialement indiqué pour les laboratoires, les écoles et les travaux qui
requièrent une grande souplesse.
3) Déclenchement alterné
Même avec l’observation de 2 formes d’onde différentes, chacune des formes d’onde
peut être déclenchée de manière stable.
5) Sortie CH1
La sortie CH1 dans le panneau postérieur de l’équipement peut être utilisée avec
d’autres instruments de mesure tels qu’un fréquencemètre, par exemple.
6) Entrée axe Z
La capacité de modulation d’intensité du faisceau permet d’ajouter des marqueurs de
fréquence ou de temps. Trace blanche avec signal positif, compatible TTL.
06/2003 Page 1
MANUEL D’UTILISATION OD-512
1.2 Spécifications
Sensibilité de 5 mV à 5 V/DIV, 10 pas en séquence 1-2-5
Précision ≤ 3% (x 5 MAG: ≤5%)
Contrôle variable de
Jusqu’à 1/2.5 ou inférieur à la valeur indiquée sur le panneau
sensibilité
DC à 20 MHz (x 5 MAG: DC à 7 MHz)
Largeur de bande Couplage AC: fréquence limite inférieure à 10 Hz
(avec référence à 100 kHz, 8 DIV. Réponse en fréquence à –3 dB)
Temps de montée Environ 17.5 ns (x 5 MAG: environ 50 ns)
Impédance d’entrée :
Environ 1 M // Environ 25 pF
Caractéristiques signal Surdéclenchement: ≤ 5% (dans l’échelle 10 mV/DIV)
carré Autres distorsions et autres échelles: 5% ajoutés à la valeur précédente
Réglage de DC Réglable par l’utilisateur
<± 0.1 DIV de la variation d’amplitude lorsqu’un signal de 2 DIV se déplace
Linéarité
verticalement depuis le centre de la grille.
CH1: canal CH1 seul
CH2: canal CH2 seul
AXE
DUAL: les canaux CH1 et CH2 sont visibles. Le mode ALT (alterné) ou
VERTICAL Modes d’opération
CHOP (découpé) pouvant être sélectionné dans toutes les
vitesses de balayage.
ADD: CH1 + CH2 somme algébrique
Fréquence de répétition
de découpage Environ 250 kHz
(Chopping)
Couplage d’entrée AC, GND, DC
Tension maximum
300 V DC + pic (AC: fréquence 1 kHz ou inférieure)
d’entrée
Rapport de rejet en 50:1 ou mieux pour un signal sinusoïdal de 50 kHz.
mode commun (Pour des sensibilités de CH1 et CH2 égales)
Isolation entre canaux >1000: 1 à 50 kHz
(échelle 5m V/DIV) >30: 1 à 20 MHz
Supérieur à 20 mV/div sur une charge de 50 Ω,
Signal de sortie CH1
largeur de bande minimum de 50 Hz à 5 MHz.
CH2 INV BAL Point de variation équilibré : ≤1 DIV (par rapport au centre de la grille)
CH1, CH2, LINE, EXT (CH1 et CH2 ne peuvent être sélectionnés que dans
Source de les modes verticaux DUAL ou ADD).
déclenchement Dans le mode ALT, interrupteur TRIG. ALT enfoncé, on alterne la source
de déclenchement.
Couplage AC: de 20 Hz à largeur de bande complète.
Pente +/-
20 Hz à 2 MHz: 0.5 DIV, TRIG-ALT: 2 DIV, EXT: 200 mV
Sensibilité 2 Hz à 20 MHz: 1.5 DIV, TRIG-ALT: 3 DIV, EXT: 800 mV
TV: Impulsion de synchronisme supérieure à 1 DIV (EXT: 1V)
AUTO: Balayage dans le mode libre lorsqu’aucun signal n’est appliqué
déclenchement. (Applicable pour des signaux répétitifs de fréquen
supérieure).
NORM: Il ne se produit un balayage que lorsque l’on reçoit un signal de
DÉCLEN-
déclenchement. Il n’est pas possible de voir la trace sans aucun
CHEMENT déclenchement. Ce mode doit être utilisé lorsque la fréquence du
Modes de
signal est de 25 Hz ou inférieure.
déclenchement
TV-V: Ce mode est utilisé pour voir l’image verticale complète d’un signal
de télévision.
TV-H: Ce mode est utilisé pour voir l’image horizontale complète d’un
signal de télévision.
(Les deux, TV-V et TV-H, synchronisent seulement pour les signaux de
synchronisme négatifs).
Entrée signal (TRIG IN)
déclenchement extérieur Environ: 1M ohm // Environ 25 Pf
Impédance d’entrée 300 V (DC+AC pIc), AC: fréquence inférieure à 1 kHz
Tension maximum
d’entrée
Page 2 06/2003
MANUEL D’UTILISATION OD-512
Temps de balayage P
0.2 s à 0.5 s/DIV, en 20 pas séquence 1-2-5
Précision du temps de
± 3%
balayage
Contrôle du temps de
≤ 1/2.5 de la valeur indiquée sur le panneau
balayage variable
Augmentation du
x 10 (temps de balayage maximum 100 ns/DIV)
AXE balayage
HORIZONTAL Précision du temps de
balayage augmenté ± 5%, (de 20 ns à 50 ns, non calibré)
x 10 MAG
Linéarité ± 3%, x10 MAG: ± 5% (20 ns et 50 ns non calibrés)
Déviation de la
position en mode À l’intérieur de 2 div. dans le centre de l’écran du TRC
augmenté x 10 MAG
La même que l’axe vertical (axe X: signal à l’entrée CH1; axe Y: signal à
Sensibilité
FONCTIONNE- l’entrée CH2)
MENT X-Y Largeur de bande DC à 500 kHz (ou supérieure)
Différence de phase X-Y ≤ 30 de DC à 50 kHz
Sensibilité 5 Vp-p (la brillance de la trace est réduite avec un signal positif)
Largeur de bande DC à 2 MHz
AXE Z Impédance d’entrée :
Environ 47 k
Tension d’entrée 30 V (DC+AC pic, AC fréquence ≤ 1 kHz)
maximum
Forme d’onde Signal carré (positif)
Fréquence Environ 1 kHz
SIGNAL DE Cycle de travail 48:52
CALIBRAGE
Tension de sortie 2 Vp-p ± 2%.
Impédance de sortie :
Environ 1 k
Type Rectangulaire de 6 pouces
TUBE DE Phosphore P 31
RAYONS Tension d’accélération Environ 2 kV
CATHODIQUES Zone utile 8 x 10 DIV (1 DIV = 10 mm (0.39 pouces)).
(TRC) Grille Interne
Rotation de la trace Incorporée
ALIMENTATION
Tension de secteur AC 115 V (97 à 132 V), 230 V (195 à 250 V) pouvant être sélectionnée, 50 Hz ou 60 Hz
Consommation 35 W
CONDITIONS D’AMBIANCE
Utilisation en intérieurs
Altitude Jusqu’à 2000 m
Pour satisfaire les spécifications: de 10 º à 35 ºC
Température
Marges maximums d’opération: de 0 º à 40 ºC
Humidité relative 85 % maximum sans condensation
Température et
humidité relative de De –10 º à 70 ºC, 70% maximum
stockage
CARACTÉRISTIQUES MÉCANIQUES
Dimensions 310 W x 150 H x 455 D mm
Poids Approx. 8 Kg (17,6 lbs)
ACCESSOIRES INCLUS
1 x Câble de secteur coudé CA-006
06/2003 Page 3
MANUEL D’UTILISATION OD-512
Page 4 06/2003
MANUEL D’UTILISATION OD-512
2. PRESCRIPTIONS DE SÉCURITÉ
2.1 Générales
* S’assurer que le dispositif à mesurer a le négatif de mesure connecté à la terre ou
est isolé du secteur.
* Cet appareil est un appareil de Classe I. Pour des raisons de sécurité il doit être
connecté aux lignes d’alimentation avec la prise de terre correspondante.
* Cet équipement peut être utilisé dans des installations de Catégorie de Surtension II
et des ambiances ayant un Degré de Pollution 1 (voir 2.3).
* Lorsque l’on emploiera l’un des accessoires suivants, respecter les spécifications de
sécurité.
Câble de secteur
* Ne jamais oublier les marges spécifiées aussi bien pour l’alimentation que pour la
mesure.
* Respecter dans tous les cas les conditions d’ambiance maximums spécifiées pour
l’appareil.
Les instructions spécifiques pour cette intervention sont fournies dans le chapitre
Entretien.
Tout autre changement dans l’appareil devra être effectué exclusivement par du
personnel spécialisé.
06/2003 Page 5
MANUEL D’UTILISATION OD-512
COURANT CONTINU
COURANT ALTERNATIF
ALTERNATIF ET CONTINU
TERMINAL DE TERRE
TERMINAL DE PROTECTION
TERMINAL À LA CARCASSE
ÉQUIPOTENTIALITÉ
MARCHE
ARRÊT
DOUBLE ISOLATION
(Protection CLASSE II)
PRÉCAUTION
(Risque de choc électrique)
FUSIBLE
Page 6 06/2003
MANUEL D’UTILISATION OD-512
Ne pas utiliser l’appareil immédiatement après l’avoir déplacé entre deux lieux ayant
des températures très différentes. Attendre un certain temps afin qu’il s’adapte à la nouvelle
température.
Ne pas appliquer de tensions excédant les limites des sondes ou des connecteurs
d’entrée.
06/2003 Page 7
MANUEL D’UTILISATION OD-512
Page 8 06/2003
MANUEL D’UTILISATION OD-512
3. INSTALLATION
3.- Replacer le cache porte-fusibles, en faisant coïncider l’indice [A] avec l’indication de la
tension de secteur souhaitée [B].
ATTENTION :
L’APPAREIL EST PRÉPARÉ EN USINE POUR UNE TENSION DE 230 V.
06/2003 Page 9
MANUEL D’UTILISATION OD-512
Pour la tourner, prendre avec les deux mains les bases de l’anse qui sont en
contact avec le châssis de l’oscilloscope et les séparer légèrement de celui-ci. De cette
manière, on pourra faire tourner l’anse jusqu’à la position souhaitée.
Page 10 06/2003
MANUEL D’UTILISATION OD-512
4. MODE D’EMPLOI
Panneau frontal
[1] Connecteur CAL. Il fournit une onde carrée d’amplitude calibrée pour les
réglages de la sonde et le calibrage de l’amplificateur vertical.
CRT
[2]. INTEN. Contrôle de réglage de la brillance de la trace sur l’écran du TRC. En
augmentation lorsque l’on tourne la commande dans le sens des aiguilles d’une
montre.
[4] TRACE ROTATION. Alignement de la trace par rapport aux lignes horizontales
de la grille du TRC.
06/2003 Page 11
MANUEL D’UTILISATION OD-512
[8] CH1 (X). Connecteur d’entrée du canal vertical CH1. Dans le mode d’opération
X-Y, cela devient l’entrée de l’axe X.
[20] CH2 (Y). Connecteur d’entrée du canal vertical CH2. Dans le mode d’opération
X-Y, cela devient l’entrée de l’axe Y.
[9] [21] VARIABLE. Réglage fin de la sensibilité de l’axe vertical CH1 et CH2
respectivement, avec un facteur ≥ 1/2.5 de la valeur sélectionnée par le contrôle
VOLTS/DIV. La sensibilité des amplificateurs verticaux VOLTS/DIV est calibrée
lorsque les contrôles VARIABLE signalent la position marquée comme CAL.
Lorsque l’on retire ces contrôles (mode PULL x 5 MAG), la sensibilité de
l’amplificateur est multipliée par 5.
[13] [17] DC BAL. Réglages de la balance DC des amplificateurs d’entrée CH1 et CH2
respectivement. Pour plus de détails, se reporter au paragraphe 4.2.4 Réglage du
niveau de DC.
Page 12 06/2003
MANUEL D’UTILISATION OD-512
DÉCLENCHEMENT (TRIGGER)
06/2003 Page 13
MANUEL D’UTILISATION OD-512
[27] TRIG. ALT. Lorsque l’interrupteur VERT MODE [14] sélectionne l’état DUAL ou
ADD et que le contrôle SOURCE [23] sélectionne CH1 ou CH2, en poussant
cette touche on fait alterner les signaux en CH1 et CH2 comme source de
déclenchement.
[29] TIME/DIV. Sélecteur du temps de balayage de 0.2 µs/DIV à 0.5 s/DIV en 20 pas.
La position X-Y est sélectionnée dans le mode d’opération.
[30] SWP. VAR. Réglage variable et continu de la vitesse de balayage entre les
positions du sélecteur TIME/DIV [29].
Les calibrages TIME/DIV ne sont exacts que lorsque le contrôle VARIABLE a été
tourné jusqu’à son extrémité dans le sens des aiguilles d’une montre et que l’on a
entendu ‘’click’’.
Page 14 06/2003
MANUEL D’UTILISATION OD-512
Panneau postérieur
[34] Z AXIS INPUT. Pour appliquer un signal qui module l’intensité du TRC. La
brillance de la trace baisse avec un signal positif et augmente avec un signal
négatif.
[35] Connecteur CH1 OUTPUT permettant une sortie amplifiée du canal 1, adéquate
pour contrôler un fréquencemètre ou tout autre instrument.
[38] Pieds pour appuyer l’oscilloscope sur son panneau postérieur. Ils servent aussi à
enrouler le câble de secteur.
06/2003 Page 15
MANUEL D’UTILISATION OD-512
1. Enfoncer l’interrupteur I/O [6]. Le pilote POWER [5] doit s’illuminer. Après 20
secondes, la trace doit apparaître à l’écran. Si la trace n’apparaît pas au bout de 60
secondes approximativement, réviser l’état des commandes et des éléments
conformément au tableau 1.
ATTENTION
Un matériau résistant au vieillissement du TRC a été utilisé. Cependant, si on
laisse le TRC avec un point ou une trace extrêmement brillant pendant une
période de temps très longue, on risque d’endommager l’écran. Par
conséquent, si la mesure requiert une brillance très intense, prendre soin de
baisser le contrôle INTEN immédiatement après celle-ci. Prendre aussi
l’habitude de baisser la brillance lorsque l’on doit travailler avec
l’oscilloscope pendant un certain temps.
Page 16 06/2003
MANUEL D’UTILISATION OD-512
Prérégler l’instrument jusqu’à obtenir une trace, comme indiqué dans le chapitre
précédent.
Connecter une sonde (x 10) à l’entrée CH1 [8] et connecter la pointe de la sonde
au connecteur CAL [1]. Sur l’écran du TRC on doit voir apparaître une onde carrée ayant
une amplitude de quatre divisions (4 x 0.5 V/DIV = 2 Vpp).
Retirer la sonde de CAL 2 Vpp [1]. Mettre le commutateur V-MODE [14] dans la
position CH2 et répéter le processus pour le canal CH2, chacun avec sa propre sonde.
06/2003 Page 17
MANUEL D’UTILISATION OD-512
L’opération avec une seule trace et une base de temps unique, ainsi qu’un
déclenchement interne est le mode opératoire le plus élémentaire de l’OD-512. Utiliser ce
mode lorsque l’on souhaite observer un seul signal, sans être gêné par les autres traces
dans le TRC. Étant donné qu’il s’agit d’un oscilloscope de deux canaux, on peut choisir l’un
ou l’autre. Le canal CH1 a un terminal de sortie. Utiliser la sortie CH1 dans le panneau
postérieur si l’on souhaite mesurer la fréquence du signal avec un fréquencemètre pendant
qu’on observe la forme de l’onde. Le canal CH2 a un interrupteur pour inverser la polarité.
Bien que cela apporte une certaine flexibilité, ce n’est pas très utile pour l’opération
habituelle avec une seule trace.
L’appareil OD-512 peut être configuré pour l’opération avec une seule trace de la
manière suivante:
1. Mettre les contrôles suivants comme indiqué ci-dessous. Vérifier que la source
de déclenchement sélectionnée (CH1 ou CH2 SOURCE) correspond au canal sélectionné
(CH1 ou CH2 V MODE).
ATTENTION
Ne pas appliquer un signal supérieur à 300 V (DC + pic AC).
Page 18 06/2003
MANUEL D’UTILISATION OD-512
5. Si le signal que l’on souhaite observer est si faible que même dans la position de
5 mV du sélecteur VOLTS/DIV [7] ou [22] il ne se produit pas de trace d’une hauteur
suffisante pour déclencher ou d’image utile, faire varier le bouton de contrôle VAR PULL x5
MAG [9] ou [21] correspondant. Cela permet une sensibilité de 2 mV/DIV lorsque le
sélecteur VOLTS/DIV [7] ou [22] se trouve dans la position 10 mV, et de 1 mV/div quand il
se trouve à 5 mV. Dans tous les cas, lorsque l’on choisit ce système, la largeur de bande
passante diminue à 7 MHz, et le bruit de la trace peut être appréciable.
6. Si le signal que l’on souhaite observer est d’une fréquence si haute que même
dans la position de 0.2 µs du sélecteur TIME/DIV [29] apparaissent trop de cycles à l’écran,
enfoncer la touche x10 MAG [31]. Cela augmente la vitesse de balayage effective par un
facteur 10; ainsi 0.2 µs/DIV devient 20 ns/DIV; 0.5 µs devient 50 ns/DIV; etc. Les positions
0.2 et 0.5 µs x 10 ne sont pas calibrées, alors que les positions 1 µs et inférieures le sont.
7. Si le signal que l’on souhaite observer est un signal continu ou d’une fréquence
très basse, et que le couplage en AC atténue ou distorsionne le signal, mettre le sélecteur
AC/GND/DC [10] ou [18] sur DC.
Placer le sélecteur VERT MODE [14] dans la position DUAL de telle manière que
l’on puisse aussi voir le signal du canal 2. Le fonctionnement est identique à celui qui est
décrit dans le paragraphe 4.3.1 aux exceptions suivantes près:
Les opérations additive et soustractive sont deux formes d’opération avec deux
canaux dans lesquelles les deux signaux sont combinés pour voir une unique trace. Dans
l’opération additive, la trace résultante représente la somme algébrique des signaux CH1 et
CH2. Dans l’opération soustractive, la trace résultante représente la différence algébrique
entre les signaux CH1 et CH2.
06/2003 Page 19
MANUEL D’UTILISATION OD-512
2. S’assurer que les deux sélecteurs VOLTS/DIV [7] et [22] se trouvent dans la
même position et que les contrôles VARIABLE [9] et [21] sont bien enclenchés dans la
position CAL. Si les niveaux de signal sont très différents, mettre les deux sélecteurs
VOLTS/DIV dans la position qui permet une vision complète à l’écran du signal de plus
grande amplitude.
4. Mettre le sélecteur MODE [14] dans la position ADD. De cette manière, la seule
trace résultante sera la somme algébrique des signaux CH1 et CH2. Aussi bien le contrôle
POSITION [11] que [19] peuvent être utilisés pour déplacer la trace résultante.
REMARQUE
Si les signaux d’entrée se trouvent en phase, l’amplitude de la trace
résultante sera la somme arithmétique des traces individuelles (par ex. : 4.2
div. + 1.2 div. = 5.4 div.). Si les signaux d’entrée on un déphasage de 180 º,
l’amplitude sera la différence (par ex. : 4.2 div. – 1.2 div. = 3.0 div.).
5. Si l’amplitude pic à pic de la trace résultante est très petite, tourner les deux
sélecteurs VOLTS/DIV [7] et [22] pour augmenter la hauteur visible. S’assurer que les deux
sont dans la même position.
Page 20 06/2003
MANUEL D’UTILISATION OD-512
06/2003 Page 21
MANUEL D’UTILISATION OD-512
Page 22 06/2003
MANUEL D’UTILISATION OD-512
06/2003 Page 23
MANUEL D’UTILISATION OD-512
La base de temps interne de l’appareil OD-512 n’est pas utilisée dans l’opération
X-Y; la déflexion, aussi bien dans la direction verticale que dans la direction horizontale, se
fait au travers de signaux externes. Le canal vertical 1 sert de traitement de signal de l’axe
X (horizontal); de cette manière, les axes vertical et horizontal ont des facilités de contrôle
identiques.
1. Tourner totalement le sélecteur TIME/DIV [29] dans le sens des aiguilles d’une
montre jusqu’à la position X-Y.
ATTENTION
Réduire l’intensité de la trace pour que le point sans déflexion n’endommage
pas le phosphore du TRC.
5. Le signal vertical (axe Y) peut être inversé en enfonçant l’interrupteur CH2 INV
[16].
Page 24 06/2003
MANUEL D’UTILISATION OD-512
5. APPLICATIONS
La mesure de tension à l’aide de l’oscilloscope est limitée à deux types: mesure pic
à pic et mesure instantanée. La première indique l’amplitude totale entre les extrêmes sans
tenir compte de la polarité de référence. La mesure de tension instantanée indique la
tension exacte depuis chaque point de la forme d’onde à une référence de terre. Lorsque
l’on fait les deux types de mesure, vérifier que les contrôles VARIABLE se trouvent bien
dans leur position extrême dans le sens des aiguilles d’une montre.
Tensions pic à pic. Pour mesurer la tension pic à pic, procéder de la manière
suivante.
2. Régler le sélecteur TIME/DIV [29] pour obtenir deux ou trois cycles de forme
d’onde et le commutateur VOLTS/DIV [7] ou [22] pour obtenir à l’écran la vision la plus
large.
3. Utiliser le contrôle POSITION approprié [11] ou [19] pour placer les pics
négatifs du signal sur la ligne horizontale de la grille la plus proche en-dessous de ces pics
(fig. 9).
4. Utiliser le contrôle POSITION [32] pour placer l’un des pics positifs sur la
ligne verticale centrale de la grille. Cette ligne a des marques de calibrage supplémentaires
espacées de 0.2 divisions.
7. Si l’on mesure une onde sinusoïdale inférieure à 100 Hz, ou une onde
rectangulaire en-dessous de 1000 Hz, situer le sélecteur AC/GND/DC en DC.
06/2003 Page 25
MANUEL D’UTILISATION OD-512
Figure 7.- Mesure de tension pic à pic. Distance verticale, ligne de référence
Page 26 06/2003
MANUEL D’UTILISATION OD-512
4. Utiliser le contrôle POSITION [11] ou [19] approprié pour placer la trace sur
la ligne horizontale centrale de la grille. Toutefois, si l’on sait que la tension du signal est
totalement positive, utiliser la ligne inférieure de la grille. Si l’on sait que la tension du signal
est totalement négative, utiliser la ligne supérieure. Cette ligne reçoit le nom de ligne de
référence de terre.
REMARQUE
Les contrôles Vertical POSITION ne doivent pas être manipulés à nouveau
avant la fin de la mesure.
5. Mettre le sélecteur AC/GND/DC [10] dans la position DC. La polarité de tous les
points situés au-dessus de la ligne de référence de terre est positive; tous les points situés
en-dessous de la ligne de référence de terre sont négatifs.
ATTENTION
Vérifier que l’on ne lit pas la forme d’onde avec une tension DC de haute
amplitude avant de faire varier le commutateur AC/GND/DC.
6. Utiliser le contrôle POSITION [32] pour placer tout point intéressant dans la
ligne verticale centrale de la grille. Cette ligne dispose de marques de calibrage
supplémentaires égales à la division par 0.2 de chacune. La tension relative à la terre en
tout point sélectionné est égale au nombre de divisions depuis ce point jusqu’à la ligne de
référence de terre multiplié par la position de VOLTS/DIV. Dans l’exemple utilisé dans la
figure 10, la tension pour l’échelle 0.5 V/DIV est de 2.5 V (5 DIV x 0.5 V/DIV).
2. Placer le commutateur TIME/DIV [29] de telle manière que l’intervalle que l’on
souhaite mesurer occupe la plus grande partie de l’écran possible. S’assurer que le
contrôle SWP. VAR. [30] se trouve bien dans la position extrême dans le sens des aiguilles
d’une montre (CAL). Si ce n’était pas le cas, aucune mesure d’intervalle de temps dans ces
conditions ne serait précise.
06/2003 Page 27
MANUEL D’UTILISATION OD-512
6. Pour déterminer l’intervalle de temps entre les deux points de mesure, multiplier
le nombre de divisions horizontales au point 5 par la position du commutateur TIME/DIV
[29]. Si le bouton x10 MAG [31] n’est pas enfoncé (agrandissement x10), prendre soin de
diviser la position du commutateur TIME/DIV par 10.
La période d’une impulsion ou de toute autre forme d’onde est le temps mis pour
réaliser un cycle complet du signal. Sur la figure 9a, la distance entre les points (A) et (C)
représente un cycle. L’échelle de temps pour la visualisation du TRC de la figure 9.a est de
10 ms/DIV; par la suite dans cet exemple, la période est de 70 ms.
L’amplitude d’impulsion est la distance entre les points (A) et (B). Dans notre
exemple, il est souhaitable qu’il soit de 1.5 divisions, ainsi l’amplitude d’impulsion est de
15 ms. Toutefois, pour des mesures de précision, 1.5 divisions est une distance plutôt
petite, il est donc conseillé d’utiliser une vitesse de balayage plus rapide pour ce type de
mesures. En augmentant la vitesse de balayage à 2 ms/DIV, comme le montre la figure 9b,
on obtient une visualisation plus étendue, ce qui permet une mesure plus précise.
Une technique alternative utile pour des impulsions d’amplitude inférieure à une
division consiste à marquer avec le bouton x10 MAG [31] pour un agrandissement x10, et
replacer l’impulsion sur l’écran avec le contrôle POSITION [32]. Dans certaines
applications, l’amplitude d’impulsion s’appelle aussi temps. Dans ce cas, la distance entre
les points (B) et (C) s’appelle aussi temps; et on peut la mesurer de la même manière que
l’amplitude d’impulsion.
PW(100) A → B(100)
Cycle de travail (%) = =
Period A→C
15 ms x 100
Exemple de cycle de travail = = 21.4%
70 ms
Page 28 06/2003
MANUEL D’UTILISATION OD-512
a. Division 10 ms
b. Division 2 ms
06/2003 Page 29
MANUEL D’UTILISATION OD-512
Le temps de montée est le temps nécessaire pour que le bord antérieur d’une
impulsion monte de 10 % à 90 % de l’amplitude totale de l’impulsion.
Le temps de descente est le temps nécessaire pour que le bord postérieur d’une
impulsion descende de 90 % de l’amplitude totale de l’impulsion à 10 %. Le temps de
montée et celui de descente, qui peuvent être appelés ensemble temps de transition, se
mesurent pour l’essentiel de la même manière.
2. Régler le sélecteur TIME DIV [29] pour voir environ 2 cycles d’impulsion.
S’assurer que le contrôle x10 MAG [31] n’est pas enfoncé.
4. Régler le sélecteur VOLTS/DIV du canal CH1 [9] pour obtenir que le pic positif
de l’impulsion dépasse 100 % de la ligne de la grille, et le pic de l’impulsion négatif dépasse
la ligne 0 %. Tourner ensuite le contrôle VARIABLE [9] dans le sens contraire des aiguilles
d’une montre jusqu’à ce que les pics des impulsions positif et négatif reposent exactement
sur les lignes 100 % et 0 % de la grille (fig. 10).
6. Si le temps de montée est lent par rapport à la période, il n’est pas nécessaire de
manipuler les contrôles. Si le temps de montée est bref (bord antérieur presque vertical)
enfoncer la touche x10 MAG [31] et replacer la trace comme dans le point 5 (fig. 10).
Page 30 06/2003
MANUEL D’UTILISATION OD-512
a. VISUALISATION DE BASE
06/2003 Page 31
MANUEL D’UTILISATION OD-512
A -B
Mod(%) = x100
A+B
Lorsque l’on a besoin de réaliser des mesures de fréquence avec une grande
précision, il est recommandé d’utiliser un fréquencemètre. On peut connecter celui-ci à la
sortie CH1 OUTPUT [35]. Toutefois, si l’on ne dispose pas de fréquencemètre, ou que la
modulation et/ou le bruit l’inhibent, l’oscilloscope permet aussi de mesurer les fréquences.
Page 32 06/2003
MANUEL D’UTILISATION OD-512
Après avoir effectué les connexions comme il est indiqué sur la figure, placer de la
même manière les commandes de CH1 et CH2, et déplacer les traces à l’aide des
commandes POSITION [11] et [19] pour essayer de les faire coïncider. La différence
d’amplitude verticale correspond à l’amplification de l’amplificateur. En agissant sur l’un des
atténuateurs (VOLTS/DIV et VARIABLE) on tentera de faire en sorte que les deux formes
d’onde coïncident complètement. S’il existe des différences, celles-ci correspondent à la
distorsion. Pour montrer la distorsion seule, inverser la polarité du signal du canal 2 en
activant la fonction CH2 INV [16] et sélectionner le mode ADD. On peut tenter de réduire
au minimum le signal différent à l’aide de l’une des commandes VARIABLE. Ce qui reste
correspond à la distorsion. En l’absence de celle-ci, on doit voir une ligne droite horizontale.
Les appareils stéréo ont deux amplificateurs égaux. L’observation simultanée des
deux facilite la localisation des pannes.
5.2.3 Réparation de TV
La forme d’onde la plus importante dans le service de TV est le signal qui est
composé d’impulsions de synchronisme, de pieds d’effacement, du composant d’image,
etc. La figure 13 montre des signaux de vidéo composée synchronisés avec les impulsions
de synchronisme horizontales et avec les impulsions d’effacement verticales.
06/2003 Page 33
MANUEL D’UTILISATION OD-512
La différence de phase ou angle de phase entre deux signaux peut être mesurée
en utilisant la caractéristique de double trace de l’oscilloscope, ou en faisant fonctionner
l’oscilloscope en mode X-Y.
Méthode de double trace. Cette méthode fonctionne avec n’importe quel type de
forme d’onde. De fait, elle fonctionne souvent même si l’on compare des formes d’onde
différentes. Elle est efficace pour mesurer des différences de phase grandes ou petites, à
n’importe quelle fréquence jusqu’à 40 MHz.
Pour mesurer des différences de phase avec la méthode de double trace, procéder
de la manière suivante:
REMARQUE
Dans les fréquences élevées, utiliser des sondes identiques et correctement
compensées, ou des longueurs égales du même type de câble coaxial pour
assurer des temps de retardement égaux.
2. Mettre le sélecteur Trigger SOURCE [23] dans le canal qui a la trace la plus
propre et la plus stable. Déplacer temporairement la trace de l’autre canal de l’écran à
l’aide du contrôle POSITION [11] ou [19].
4. Utiliser le contrôle Trigger LEVEL [28] pour s’assurer que la trace croise la ligne
horizontale centrale de la grille au début du balayage ou très près de celui-ci (fig. 14).
Page 34 06/2003
MANUEL D’UTILISATION OD-512
1. Tourner le sélecteur TIME/DIV [29] dans le sens des aiguilles d’une montre
jusqu’à la position X-Y.
AVERTISSEMENT
Réduire l’intensité de la trace afin que le point sans déflexion n’endommage
pas le phosphore du TRC.
2. S’assurer que le bouton x10 MAG [18] est enfoncé. Dans le cas contraire, une
erreur de phase de 180 º s’introduira.
5. Régler le sélecteur CH1 VOLTS/DIV [7] pour faire en sorte de voir l’écran le plus
largement possible.
06/2003 Page 35
MANUEL D’UTILISATION OD-512
Page 36 06/2003
MANUEL D’UTILISATION OD-512
8. La différence de phase (angle θ) entre les deux signaux est égale au sinus de
l’arc de dimension A : B (le chiffre correspondant au point 7 divisé par 6). Par exemple, la
valeur du point 7 de la figure 15 est 2.0. En le divisant par 6, on obtient 0.3334, dont le
sinus est 19.5 º.
9. La formule simple de la figure 15 est valide pour des angles plus petits que 90 º.
Pour des angles plus grands (inclinaison vers la gauche), ajouter 90 º à l’angle du point 7.
La figure 15 montre la figure de Lissajous de divers angles de phase: l’utiliser comme
référence pour déterminer si l’on ajoute 90 º ou non.
REMARQUE
La conversion sinus d’un angle peut être effectuée en utilisant une table
trigonométrique ou une calculette scientifique.
06/2003 Page 37
MANUEL D’UTILISATION OD-512
Page 38 06/2003
MANUEL D’UTILISATION OD-512
6. ENTRETIEN
06/2003 Page 39
MANUEL D’UTILISATION OD-512
PRÉCAUTION
Pour nettoyer la boîte, s’assurer que l’appareil est bien déconnecté.
PRÉCAUTION
Ne pas utiliser pour le nettoyage d’hydrocarbures aromatiques ni de dissolvants
chlorés. Ces produits peuvent en effet attaquer les matériaux employés pour la
fabrication de la boîte.
La boîte sera nettoyée avec une solution de détergent et d’eau appliquée à l’aide
d’un tissu légèrement humidifié.
Page 40 06/2003