Está en la página 1de 2

XN CAL™

XN-SERIES (XN-10, XN-20) ASSAY


FSR USE ONLY This Product Must Be Run In Calibrator Mode

Lot Number 01392101 Exp. Date 2020-06-21

Effective Date: XN-10 and XN-20 Targets and Limits


20-May-20 WBC WBC-D * WBC-P* RBC HGB PLT** RETIC% RBC-O* MPV RET-He RDW-SD RDW-CV
Target 6.73 6.92 6.75 4.38 12.0 248 2.09 4.36 9.9 25.7 45.8 16.0
Lower Limit 6.43 6.52 6.35 4.29 11.7 237 1.84 4.22 9.5 24.3 43.8 15.0
Upper Limit 7.03 7.32 7.15 4.47 12.3 259 2.34 4.50 10.3 27.1 47.8 17.0

XN-10 and XN-20 DCP Targets and Limits


Effective Date: Use Between: 20-May-20 02-Jun-20
RET- RET- RET- PLT-F- PLT-F-
20-May-20 WNR-X* WNR-Y* WNR-Z* WDF-X* WDF-Y* WDF-Z* WPC-X* WPC-Y* WPC-Z*
RBC-X* RBC-Y* RBC-Z* RBC-X* RBC-Y*
PLT-F-Z* HCT MCV MCHC

Target 180.1 168.7 114.2 169.9 104.4 119.0 210.7 199.3 129.6 23.2 161.4 26.0 83.7 232.2 38.3 35.0 80.1 34.3
Lower Limit 177.1 165.7 111.2 166.9 99.4 116.0 207.7 197.3 127.6 22.2 158.4 24.0 81.7 230.2 36.3 34.0 78.1 33.7
Upper Limit 183.1 171.7 117.2 172.9 109.4 122.0 213.7 201.3 131.6 24.2 164.4 28.0 85.7 234.2 40.3 36.0 82.1 34.9

Effective Date: Use Between: 03-Jun-20 12-Jun-20


RET- RET- RET- PLT-F- PLT-F-
03-Jun-20 WNR-X* WNR-Y* WNR-Z* WDF-X* WDF-Y* WDF-Z* WPC-X* WPC-Y* WPC-Z*
RBC-X* RBC-Y* RBC-Z* RBC-X* RBC-Y*
PLT-F-Z* HCT MCV MCHC

Target 179.4 168.7 114.2 169.9 102.2 119.1 210.7 199.1 130.4 23.1 161.2 25.3 83.1 232.2 38.3 35.2 80.5 34.1
Lower Limit 176.4 165.7 111.2 166.9 97.2 116.1 207.7 197.1 128.4 22.1 158.2 23.3 81.1 230.2 36.3 34.2 78.5 33.5
Upper Limit 182.4 171.7 117.2 172.9 107.2 122.1 213.7 201.1 132.4 24.1 164.2 27.3 85.1 234.2 40.3 36.2 82.5 34.7

Effective Date: Use Between: 13-Jun-20 21-Jun-20


RET- RET- RET- PLT-F- PLT-F-
13-Jun-20 WNR-X* WNR-Y* WNR-Z* WDF-X* WDF-Y* WDF-Z* WPC-X* WPC-Y* WPC-Z*
RBC-X* RBC-Y* RBC-Z* RBC-X* RBC-Y*
PLT-F-Z* HCT MCV MCHC

Target 178.6 168.7 114.2 169.9 101.1 119.3 210.7 198.5 131.0 22.9 161.1 24.7 82.6 232.2 38.3 35.3 80.7 34.1
Lower Limit 175.6 165.7 111.2 166.9 96.1 116.3 207.7 196.5 129.0 21.9 158.1 22.7 80.6 230.2 36.3 34.3 78.7 33.5
Upper Limit 181.6 171.7 117.2 172.9 106.1 122.3 213.7 200.5 133.0 23.9 164.1 26.7 84.6 234.2 40.3 36.3 82.7 34.7
*Detector Control Parameters (DCP) are non-reportable parameters for internal use only used by Service Personnel.
**For XN-V customers calibrate PLT-O using PLT targets and limits.

Distributed by: 350603

Sysmex America, Inc. Rev 6. 17-Apr-2019

Lincolnshire, IL May-20

XN CAL™
XN-L ASSAY
FSR USE ONLY This Product Must Be Run In Calibrator Mode

Lot Number 01392101 Exp. Date 2020-06-21

Effective Date: XN-L Targets and Limits


20-May-20 WBC-C WBC-D * RBC HGB PLT RETIC% RBC-O* MPV RET-He RDW-SD RDW-CV
Target 7.42 7.03 4.40 12.1 251 2.26 4.28 9.8 23.6 46.3 16.1
Lower Limit 7.12 6.63 4.31 11.8 240 2.03 4.14 9.4 22.8 44.3 15.1
Upper Limit 7.72 7.43 4.49 12.4 262 2.49 4.42 10.2 24.4 48.3 17.1

XN-L DCP Targets and Limits


Effective Date: Use Between: 20-May-20 02-Jun-20
RET- RET- RET-
20-May-20 WDF-X* WDF-Y* WDF-Z*
RBC-X* RBC-Y* RBC-Z*
HCT MCV MCHC

Target 161.9 101.3 110.9 21.4 159.3 22.5 35.1 79.8 34.4
Lower Limit 158.9 96.3 107.9 20.4 156.3 20.5 34.1 77.8 33.8
Upper Limit 164.9 106.3 113.9 22.4 162.3 24.5 36.1 81.8 35.0

Effective Date: Use Between: 03-Jun-20 12-Jun-20


RET- RET- RET-
03-Jun-20 WDF-X* WDF-Y* WDF-Z*
RBC-X* RBC-Y* RBC-Z*
HCT MCV MCHC

Target 161.9 99.2 111.0 21.3 159.1 22.0 35.3 80.2 34.3
Lower Limit 158.9 94.2 108.0 20.3 156.1 20.0 34.3 78.2 33.7
Upper Limit 164.9 104.2 114.0 22.3 162.1 24.0 36.3 82.2 34.9

Effective Date: Use Between: 13-Jun-20 21-Jun-20


RET- RET- RET-
13-Jun-20 WDF-X* WDF-Y* WDF-Z*
RBC-X* RBC-Y* RBC-Z*
HCT MCV MCHC

Target 161.9 98.2 111.2 21.1 158.9 21.8 35.4 80.4 34.2
Lower Limit 158.9 93.2 108.2 20.1 155.9 19.8 34.4 78.4 33.6
Upper Limit 164.9 103.2 114.2 22.1 161.9 23.8 36.4 82.4 34.8
*Detector Control Parameters (DCP) are non-reportable parameters for internal use only used by Service Personnel.

Distributed by: 350603

Sysmex America, Inc. Rev 6. 17-Apr-2019

Lincolnshire, IL May-20

2
XN CAL™
XN-SERIES (XN-11, XN-21) ASSAY
FSR USE ONLY This Product Must Be Run In Calibrator Mode

Lot Number 01392101 Exp. Date 2020-06-21

Effective Date: XN-11 and XN-21 Targets and Limits


20-May-20 WBC WBC-D * WBC-P* RBC HGB PLT RETIC% RBC-O* MPV RET-He RDW-SD RDW-CV
Target 6.71 6.94 6.93 4.43 12.13 253 1.86 4.28 11.1 25.4 44.8 16.6
Lower Limit 6.41 6.54 6.53 4.34 11.83 242 1.61 4.14 10.7 24.0 42.8 15.6
Upper Limit 7.01 7.34 7.33 4.52 12.43 264 2.11 4.42 11.5 26.8 46.8 17.6

XN-11 and XN-21 DCP Targets and Limits


Effective Date: Use Between: 20-May-20 02-Jun-20
RET- RET- RET- PLT-F- PLT-F-
20-May-20 WNR-X* WNR-Y* WNR-Z* WDF-X* WDF-Y* WDF-Z* WPC-X* WPC-Y* WPC-Z*
RBC-X* RBC-Y* RBC-Z* RBC-X* RBC-Y*
PLT-F-Z* HCT MCV MCHC

Target 180.8 168.4 114.4 170.2 106.1 119.5 211.5 198.0 129.1 23.4 161.3 25.5 84.1 232.6 38.8 34.1 76.9 35.6
Lower Limit 177.8 165.4 111.4 167.2 101.1 116.5 208.5 196.0 127.1 22.4 158.3 23.5 82.1 230.6 36.8 33.1 74.9 35.0
Upper Limit 183.8 171.4 117.4 173.2 111.1 122.5 214.5 200.0 131.1 24.4 164.3 27.5 86.1 234.6 40.8 35.1 78.9 36.2

Effective Date: Use Between: 03-Jun-20 12-Jun-20


RET- RET- RET- PLT-F- PLT-F-
03-Jun-20 WNR-X* WNR-Y* WNR-Z* WDF-X* WDF-Y* WDF-Z* WPC-X* WPC-Y* WPC-Z*
RBC-X* RBC-Y* RBC-Z* RBC-X* RBC-Y*
PLT-F-Z* HCT MCV MCHC

Target 180.1 168.4 114.4 170.2 103.9 119.6 211.5 197.8 129.8 23.3 161.2 24.9 83.5 232.6 38.8 34.2 77.3 35.4
Lower Limit 177.1 165.4 111.4 167.2 98.9 116.6 208.5 195.8 127.8 22.3 158.2 22.9 81.5 230.6 36.8 33.2 75.3 34.8
Upper Limit 183.1 171.4 117.4 173.2 108.9 122.6 214.5 199.8 131.8 24.3 164.2 26.9 85.5 234.6 40.8 35.2 79.3 36.0

Effective Date: Use Between: 13-Jun-20 21-Jun-20


RET- RET- RET- PLT-F- PLT-F-
13-Jun-20 WNR-X* WNR-Y* WNR-Z* WDF-X* WDF-Y* WDF-Z* WPC-X* WPC-Y* WPC-Z*
RBC-X* RBC-Y* RBC-Z* RBC-X* RBC-Y*
PLT-F-Z* HCT MCV MCHC

Target 179.4 168.4 114.4 170.2 102.9 119.8 211.5 197.2 130.5 23.1 161.0 24.3 83.0 232.6 38.8 34.3 77.5 35.3
Lower Limit 176.4 165.4 111.4 167.2 97.9 116.8 208.5 195.2 128.5 22.1 158.0 22.3 81.0 230.6 36.8 33.3 75.5 34.7
Upper Limit 182.4 171.4 117.4 173.2 107.9 122.8 214.5 199.2 132.5 24.1 164.0 26.3 85.0 234.6 40.8 35.3 79.5 35.9
*Detector Control Parameters (DCP) are non-reportable parameters for internal use only used by Service Personnel.

Distributed by: 350603

Sysmex America, Inc. Rev 6. 17-Apr-2019

Lincolnshire, IL May-20