REPORTE DE CALIBRACION CROMATOGRAFOS DANIEL

FECHA:
DESCRIPCION: CROMATOGRAFO DE LINEA GC700XA C9+ P/N: 2-1-770XA-037
FUNCION: CROMATOGRAFO DE LINEA S/N: 9014066
FRECUENCIA CALIBRACION: MENSUAL TAG: AUT -0105
PROXIMA CALIBRACION: 9/31/2016 TEMPERATURA AMBIENTE 20 Deg. C

GAS PATRON Corrida # 1 Corrida # 2 Corrida # 3 % MOLAR


COMPONENTES
% Mol % Mol % Mol % Mol Reproducibilidad
PROPANO 1.0000 1.0000 1.0003 1.0003 0.00
i-BUTANO 0.3000 0.2999 0.3000 0.3001 0.00
n-BUTANO 0.3030 0.3032 0.3028 0.3031 0.00
NEOPENTANO 0.0000 0.0000 0.0000 0.0000 0.00
i-PENTANO 0.1010 0.1009 0.1013 0.1015 0.00
n-PENTANO 0.1010 0.1005 0.1006 0.1014 0.00
NITROGENO 2.4600 2.4597 2.4595 2.4596 0.00
METANO 89.6375 89.6388 89.6385 89.6367 0.00
DIOXIDO DE CARBONO 0.9990 0.9989 0.9989 0.9990 0.00
ETANO 4.9900 4.9899 4.9896 4.9896 0.00
n-NONAN0 0.0099 0.0099 0.0099 0.0099 0.00
n-HEXANO 0.0594 0.0593 0.0593 0.0593 0.00
n-HEPTANO 0.0199 0.0199 0.0199 0.0199 0.00
n-OCTANO 0.0193 0.0193 0.0193 0.0197 0.00
% TOTAL NORMALIZADO 100.000 100.000 100.000 100.000
% TOTAL NO NORMALIZADO 100.000 100.016 100.026 100.037

STANDAR NORMA ASTM D 1945-03

Cont. Molar Reprod. Repet.


% Mol % Mol % Mol
De 0.0 a 0.1 0.02 0.01
De 0.1 a 1.0 0.07 0.04
De 1.0 a 5.0 0.10 0.07
De 5.0 a 10.0 0.12 0.08
Mayor a 10.0 0.15 0.10
% TOTAL NO NORMALIZADO 100 ± 1 %

OBSERVACIONES :

Los valores obtenidos cumplen con la Norma A.S.T.M. D 1945-03

ESPECIFICACION TECNICA GAS PATRON


GAS No. CILINDRO No.CERTIFICADO CERTIFICADO EMITIDO POR FECHA DE EXPIRA
HELIO
PATRON 75910AW 13312-59459 APPLIED GAS 21-Jan-18

PERSONAL EJECUCION PERSONAL FISCALIZACION

PAMELA STUMVOLL DANIEL RIOS OMAR CAMACHO IVER ARISPE MARIELA JAURE
INCA SRL TECNA TECNA YPFB YPFB
CION CROMATOGRAFOS DANIEL
31-Aug-16
2-1-770XA-037
9014066
AUT -0105
20 Deg. C

% MOLAR
Repetibilidad
0.00
0.00
0.00
0.00
0.00
0.00
0.00
0.00
0.00
0.00
0.00
0.00
0.00
0.00

ORMA ASTM D 1945-03

ON TECNICA GAS PATRON


FECHA DE EXPIRACION

21-Jan-18

PERSONAL FISCALIZACION

MARIELA JAUREGUI
YPFB