Documentos de Académico
Documentos de Profesional
Documentos de Cultura
Ventilacion Mecanica en Pediatria
Ventilacion Mecanica en Pediatria
Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
RETIRADA DE LA VENTILACIN,
COMPLICACIONES Y OTROS TIPOS
DE VENTILACIN
Sociedad Espaola de Cuidados Intensivos Peditricos
patient outcome. Thus, to ensure maximum success, patient readiness for weaning and extubation should be
evaluated through the following steps: the patient must
fulfill pre-established clinical and ventilatory support
criteria for extubation, the patient should be observed
during a breathing trial on minimal or no ventilatory
support, and variables used to predict weaning success
should indicate a favorable outcome. Breathing trials are
usually conducted while the patient breathes spontaneously through a T-tube system or through the ventilator circuit on minimal ventilatory support. Neither of
these methods has proved superior to the other. The best
prognostic indicator of weaning outcome is clinical assessment of respiratory effort. Once mechanical ventilation is discontinued, it may be necessary to treat post-extubation complications or even to resume ventilatory
support.
Key words:
Mechanical ventilation. Weaning. Extubation. T tube.
Support pressure. Post-extubation croup.
Palabras clave:
Ventilacin mecnica. Retirada de la asistencia respiratoria. Extubacin. Tubo en T. Presin de soporte. Estridor por extubacin.
CONCEPTOS GENERALES
La retirada de la VM define el proceso que permite el
paso de la VM a la ventilacin espontnea. Como todo
proceso, se distinguen en l varias fases: fase de partida o
de soporte ventilatorio total; fase de transicin o de soporte ventilatorio parcial y fase final o extubacin. No
todos los pacientes deben pasar por todas las fases del
proceso, ya que en determinados casos es posible pasar
de la fase de soporte total casi directamente a la extuba-
00
155
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
cin (p. ej., en los pacientes que reciben soporte ventilatorio durante unas horas en el postoperatorio de una ciruga mayor no complicada). Como norma general, ser
posible aplicar una retirada rpida de la VM en pacientes
con una patologa aguda ya resuelta y que han recibido
un soporte ventilatorio de corta duracin. Por otro lado,
pacientes con patologa aguda muy grave, patologa aguda cronificada y/o soporte ventilatorio muy prolongado o complejo precisarn una retirada gradual o lenta de
la VM1. As, los principales determinantes de la duracin
de la retirada del soporte ventilatorio son el tipo de enfermedad que motiv la VM y la duracin de la propia
VM. En general, la rapidez del descenso de la asistencia
respiratoria debe ser inversamente proporcional a la duracin de la VM.
156
00
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
cos, benzodiacepinas, etc.) en los requerimientos de frmacos vasoactivos (teniendo en cuenta que estos frmacos sern suspendidos al retirar la VM).
Nivel de conciencia. Por lo que respecta al nivel de conciencia que se considera adecuado, es preciso hacer algunas matizaciones. As, idealmente, el paciente debera estar despierto y colaborador previo a la extubacin, pero en
muchas ocasiones, y sobre todo con pacientes peditricos,
es imposible mantener al nio intubado, despierto y colaborador. Con frecuencia se plantea, sobre todo en lactantes, una situacin de todo o nada, es decir, o se seda profundamente al paciente o se procede a la extubacin sin
apenas poder evaluar su estado de conciencia. En esta situacin es fundamental haber valorado bien el resto de requisitos para la extubacin, ya que si todos se valoran positivamente, por lo general es posible extubar con xito al
lactante que lucha con el respirador. Cabe hacer mencin
especial a los lactantes sometidos a VM postoperatoria que
han recibido dosis significativas de fentanilo durante la
anestesia, los cuales, con frecuencia, intercalan en las primeras 3 a 6 h de postoperatorio intervalos breves de desadaptacin con momentos sedacin profunda; la extubacin diligente o precoz de estos nios antes de que
eliminen suficientemente la anestesia puede generar alguna situacin comprometida ya que, una vez desaparece el
estmulo del tubo endotraqueal, suelen quedar profundamente sedados, realizando algunos de ellos pausas de apnea (por obstruccin de la va area superior) que pueden requerir reintubacin o uso de antdotos (naloxona,
flumazenilo). Un grupo de pacientes que debe considerarse aparte lo constituyen los pacientes con lesiones del sistema nervioso central (SNC). El paradigma de los mismos
son los pacientes con traumatismo craneoenceflico (TCE)
grave. As como el criterio fundamental (desde el punto de
vista neurolgico) para intubar a estos pacientes es un estado de conciencia con una puntuacin en la escala de
Glasgow < 8, no es necesario (y muchas veces no es posible) esperar a que recuperen un Glasgow > 8 para proceder a la extubacin. As, los pacientes con un TCE grave
en quienes quedan importantes secuelas recuperan los
mecanismos necesarios para mantener una ventilacin espontnea (reflejo tusgeno, nauseoso, estmulo respiratorio), mucho antes de que se produzca una recuperacin
neurolgica en trminos de estado de conciencia y conexin ambiental.
00
Valoracin preextubacin
La valoracin preextubacin consiste en un ltimo anlisis de la situacin del paciente con la finalidad de proceder o no a la extubacin. Se trata de un momento tan
importante como la indicacin de la VM, ya que de su correcta o incorrecta realizacin se derivan consecuencias
de gran trascendencia para el paciente. As, una valoracin falsamente negativa que no identifique a un paciente como apto para ser extubado, cuando en realidad s lo
es, expone al paciente a una prolongacin innecesaria de
la VM. Por otro lado, una valoracin falsamente positiva
que no detecte aquellos pacientes en los que es posible
anticipar un fracaso de la retirada de la VM, pone al paciente en una situacin de estrs fsico y psicolgico, adems de exponerlo a una situacin de riesgo vital como
puede ser una reintubacin urgente.
A pesar de la relevancia que tiene la correcta o incorrecta indicacin de la retirada de la VM, este es un
apartado del soporte ventilatorio en el que se ha profundizado poco y en el que escasean datos objetivos extrapolables a la mayora de pacientes crticos, sobre todo a
los nios. A pesar de ello, en los ltimos aos han aparecido varios trabajos que han contribuido a arrojar algo
de luz sobre el tema2-4.
A la hora de evaluar a un paciente es preciso decidir a
qu tipo de prueba o evaluacin va a someterse al paciente y escoger qu parmetros se van a valorar para considerar la prueba positiva o negativa. Respecto al primer
punto existen fundamentalmente dos opciones: a) prueba de respiracin espontnea con tubo en T y b) prueba
con soporte ventilatorio mnimo (CPAP con o sin presin
de soporte).
157
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
Prueba de tubo en T
La prueba de tubo en T es una prueba sencilla que
consiste bsicamente en desconectar al paciente del respirador y dejarlo respirar espontneamente a travs del
tubo endotraqueal. Con la finalidad de garantizar una
FiO2 suficiente y una humidificacin adecuada del gas
inspirado, el tubo endotraqueal se conecta mediante una
pieza en forma de T a una tubuladura por la que circula un flujo de aire y oxgeno humidificados. En el momento de iniciar la prueba, el mdico debe decidir la
FiO2 inspirada (que ser como mnimo la misma que reciba estando conectado al respirador) y el flujo que
debe circular por el circuito. Este ltimo punto tiene una
importancia no siempre reconocida, ya que un flujo inapropiadamente bajo puede propiciar que el paciente respire aire atmosfrico (con FiO2 del 21 %) y que, por lo
tanto, la FiO2 inspirada real del paciente no sea la pautada; adems, un flujo insuficiente en el circuito no lava
el aire espirado y favorece la reinhalacin de aire espirado con la consiguiente hipercapnia. La administracin
de flujos altos en el circuito del tubo en T no tiene ninguna repercusin sobre el paciente, por lo que siempre
debe ajustarse el flujo al alza.
158
00
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
los datos basales como los cambios que se producen durante la prueba preextubacin).
Extubacin: tcnica
Una vez tomada la decisin de extubar al paciente debemos, en primer lugar, preparar el material necesario
para la reintubacin: mascarilla facial, bolsa autohinchable, sondas de aspiracin, cnula de Guedel, tubos endotraqueales, laringoscopio y medicacin (sedoanalgesia,
relajantes musculares). Una vez preparado el material se
realiza una ltima aspiracin de las secreciones traqueales, se comprueba que se ha deshinchado el baln del
neumotaponamiento y se procede a la extubacin, que
puede realizarse a presin negativa o a presin positiva.
La extubacin a presin negativa consiste en retirar el
tubo endotraqueal mientras se aspiran las secreciones traqueales con una sonda introducida a travs del tubo con
la finalidad de dejar la va area limpia de secreciones en
el momento de que el paciente inicie la respiracin espontnea. La extubacin a presin positiva consiste en retirar el tubo endotraqueal al mismo tiempo que se realiza una insuflacin con bolsa autohinchable a travs de
ste, con la finalidad de que al retirar el tubo el paciente
tenga los pulmones expandidos y pueda toser con fuerza para eliminar las secreciones respiratorias.
COMPLICACIONES POSTEXTUBACIN
La principal complicacin de la extubacin es que fracase4. En un porcentaje no insignificante de los casos,
fracasa a pesar de haber seguido una metodologa cuidadosa en la retirada de la VM y en la valoracin preextubacin. Son muchas las causas por las que puede fallar
la retirada de la VM y se enumeran en la tabla 4. La mayora pueden ser intuidas o sospechadas durante la valoracin previa a la extubacin; otras, en cambio, como la
aparicin de obstruccin alta de la va area, bien sea por
laringitis inflamatoria o por lesiones establecidas como
estenosis o granulomas no se manifiestan hasta el momento en que se retira el tubo endotraqueal. As, habr
que anticipar la posible aparicin de estridor postextubacin mediante la valoracin de los siguientes factores de
riesgo: antecedentes de intubacin complicada, reintubaciones previas, tubo endotraqueal grande, desplazamientos repetidos del tubo, presin del neumotaponamiento
excesivamente alta, presencia de infeccin laringotraqueal. En ocasiones, se emplea el tratamiento con corticoides sistmicos para prevenir la aparicin del estridor
postextubacin, pero su utilidad no se ha demostrado. El
tratamiento del estridor postextubacin incluye: adrenalina inhalada (0,5 mg/kg, mximo 5 mg), budesonida inhalada (2 mg), metilprednisolona intravenosa (2 mg/kg),
inhalacin de mezclas de helio y oxgeno y, finalmente,
reintubacin si es preciso. Respecto al uso de mezclas de
helio y oxgeno conviene sealar, que si bien pueden resultar muy tiles para rescatar a algunos pacientes de la
00
reintubacin, en ocasiones pueden hacer que sta se retrase hasta un punto en que la obstruccin de la va area
sea tal que la reintubacin sea tcnicamente muy difcil
o imposible. Por tanto, debe quedar claro que la administracin de mezclas de helio y oxgeno es una medida
paliativa, destinada a ganar tiempo mientras otros tratamientos (adrenalina inhalada, budesonida inhalada, etc.)
controlan la inflamacin larngea.
BIBLIOGRAFA
1. Tobin MJ, Alex CG. Discontinuation of mechanical ventilation. En: Tobin MJ, editor. Mechanical ventilation. New York:
McGraw-Hill, 1994; p. 1177-206.
2. Randolph AG, Wypij D, Venkataraman ST, Gedeit RG, Meert
KL, Luckett PM, et al. Effect of mechanical ventilation weaning
protocols on respiratory outcomes in infants and children.
JAMA 2002;288:2561-8.
3. Farias JA, Retta A, Ala I, Olazarri F, Esteban A, Golubicki A, et
al. A comparison of two methods to perform a breathing trial
before extubation in pediatric intensive care patients. Intensive
Care Med 2001;27:1649-54.
4. Epstein SK. Decision to extubate. Intensive Care Med 2002;28:
535-46.
159
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
Palabras clave:
Ventilacin mecnica. Complicaciones de la ventilacin
mecnica. Lesin inducida por ventilacin. Barotrauma.
Volutrauma. Estridor postextubacin. Neumona nosocomial. Nios.
COMPLICATIONS OF MECHANICAL
VENTILATION
Mechanical ventilation can produce multiple complications. The most important acute complications are mechanical problems (respirator failure, problems with the
connections and circuit, incorrect parameters or alarms),
problems in the airway (disconnection, extubation, malpositioning of the endotracheal tube, leaks, nose erosions,
obstruction of the endotracheal tube due to secretions or
kinking, mainstem bronchus intubation, bronchospasm,
postextubation croup), pulmonary complications (ventilator-induced lung injury with barotrauma, volutrauma and
biotrauma), hemodynamic complications, nosocomial infections (tracheobronchitis, pneumonia, otitis, sinusitis),
failure of adjustment of the respirator to the patient, and
nutritional complications. The most important chronic
problems are subglottal stenosis, chronic pulmonary injury, and psychological alterations.
Key words:
Mechanical ventilation. Mechanical ventilation complications. Ventilator-induced lung injury. Barotrauma. Volutrauma. Postextubation croup. Nosocomial pneumonia.
Children.
INTRODUCCIN
La utilizacin de la VM en el paciente crtico supone un
riesgo de complicaciones, con el agravante de la mayor
dificultad de la VM en el nio, tanto por las caractersticas
propias de la edad peditrica, como de los respiradores
utilizados.
Las caractersticas del nio que dificultan la VM son la
inmadurez pulmonar en el neonato/prematuro, la mayor
frecuencia respiratoria, la respiracin irregular, menor volumen corriente utilizado, menor esfuerzo respiratorio
por parte del nio, la dificultad de la monitorizacin y la
falta de colaboracin del paciente peditrico. Por otro
lado, los respiradores utilizados tienen una menor precisin en volmenes bajos (con lo que aumenta el riesgo
de hipo/hiperventilacin), su sensibilidad es menor ante
pequeos esfuerzos y la adaptacin en cada respiracin
es ms difcil. Es preciso un adecuado conocimiento tanto de las caractersticas especiales del nio, como de las
limitaciones de los respiradores, para conseguir los mximos beneficios de la VM, minimizando los riesgos de
complicaciones.
Las complicaciones de la VM pueden clasificarse en
agudas y crnicas (o secuelas).
COMPLICACIONES AGUDAS
Fallo de dispositivos externos y de programacin
Se produce por prdida de la fuente elctrica al respirador, fallo en la fuente de gases (mala colocacin de
conexiones, etc.), fallo del respirador o error de programacin (alarmas mal colocadas, fugas, tanto por desconexin, como alrededor del tubo endotraqueal, mala
programacin de la sensibilidad, que provoca un exce-
160
00
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
Problemas bronquiales
La presencia del tubo endotraqueal en la va area produce un aumento de la produccin de secreciones que
no pueden ser eliminadas de forma espontnea por el paciente, por lo que requiere aspiraciones peridicas. Las
aspiraciones son las responsables fundamentales del desarrollo de broncospasmo y sangrado de la va respiratoria. Para prevenir el broncospasmo es necesario que las
aspiraciones no sean ni muy intensas ni prolongadas, e
introducir la sonda de aspiracin slo hasta el final del
tubo endotraqueal. No hay evidencias que la administracin previa de broncodilatadores prevenga la aparicin
del broncospasmo secundario a la aspiracin. Si aparece
sangrado por dao en la va area el lavado con suero salino y adrenalina diluida al 1/10.000 puede ayudar a controlarlo.
Desconexin o extubacin
Se produce por sedacin insuficiente, falta de vigilancia
o mala fijacin del tubo endotraqueal. Puede provocar insuficiencia o parada respiratoria, adems de lesin de la
va area (sobre todo en el caso de extubacin con el
baln de neumotaponamiento hinchado)1-3.
Malposicin del tubo endotraqueal
La ms importante es la intubacin en bronquio derecho, por la anatoma propia del rbol bronquial. Puede
provocar hipoxemia e hipercapnia, atelectasia del lbulo
superior derecho y del hemitrax izquierdo, as como
neumotrax o hiperinsuflacin de hemitrax derecho.
Debe sospecharse en un paciente que presenta aumento
de las necesidades ventilatorias, hipoxemia e hipercapnia,
con aumento de presiones en el caso de ventilacin por
volumen y disminucin del VC en ventilacin por presin; en la exploracin existe una hipoventilacin de hemitrax izquierdo.
Fuga de aire alrededor del tubo
Generalmente se debe a la utilizacin de un tubo endotraqueal pequeo o en posicin muy alta, a presiones
altas de la va area en pacientes ventilados sin baln o a
rotura del baln o de su vlvula. La existencia de una
pequea fuga espiratoria es normal en los nios ventilados con tubos sin baln; sin embargo, si la fuga es importante y dificulta la ventilacin es necesario cambiar el
tubo endotraqueal por uno mayor o con baln.
Lesiones en el ala de la nariz y en la zona de fijacin
del tubo endotraqueal
Los nios intubados por la nariz pueden presentar lesiones por decbito en el ala de la nariz, que pueden llegar a necrosis con prdida irreversible del tejido y erosin
en las zonas de fijacin del esparadrapo. Para prevenirlas es recomendable realizar una proteccin de las zonas
de fijacin con hidrocoloides y vigilar de forma peridica la piel y el ala de la nariz.
00
Estridor postextubacin
Los factores de riesgo fundamentales son la dificultad
para la intubacin, la reintubacin o cambios repetidos de
tubo endotraqueal, la utilizacin de presin alta en el baln de neumotaponamiento y la infeccin laringotraqueal
asociada. Se manifiesta con sintomatologa de estridor inspiratorio y dificultad respiratoria tras la retirada de tubo
endotraqueal. El efecto preventivo de la metilprednisolona intravenosa a 2 mg/kg en el momento previo a la extubacin es muy discutido, y no existen pruebas claras sobre su utilidad. Para su tratamiento debe administrarse
adrenalina inhalada 0,5 ml/kg (mximo 10 ml). En caso
de no mejora pueden asociarse la budesonida inhalada
0,5-1 mg, metilprednisolona 2 mg/kg/IV, y/o heliox. En
ocasiones puede ser necesaria la utilizacin de CPAP nasal, VM no invasiva o la reintubacin.
Problemas pulmonares
Lesin inducida por ventilacin mecnica
Mecanismo. La VM no es un proceso fisiolgico, por lo
que puede producirse lesin pulmonar por barotrauma
(presin) o volutrauma (volumen). La utilizacin de presiones pico y meseta elevadas son los factores de riesgo
161
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
ms importantes para la aparicin de barotrauma, con lesin de va area, sobre todo por presiones pico excesivas, y de lesin alveolar por elevadas presiones meseta. La
presin pico inicia la lesin en la va area, mientras que
el volumen excesivo distiende el alvolo, mantenindose
y progresando la lesin. En los ltimos aos se ha demostrado tambin el papel del estiramiento por colapso y
apertura de los alvolos en cada ciclo respiratorio que
produce dao de las fibras elsticas de la pared alveolar y
prdida de surfactante. Tambin se ha sugerido que la VM
al actuar sobre la va area y las clulas alveolares, estimula la liberacin de mediadores inflamatorios que actuaran aumentando la lesin pulmonar (biotrauma). Adems de los efectos mecnicos y biolgico el oxgeno a
concentraciones superiores a 0,6 produce lesin en la va
area y el alvolo pulmonar, y otros factores como la sobreinfeccin pulmonar contribuyen al desarrollo y el mantenimiento de la lesin pulmonar4,5.
Manifestaciones clnicas. La lesin inducida por VM puede producir:
1. Alteraciones del parnquima pulmonar: edema pulmonar, enfisema intersticial, atrapamiento areo, hemorragia pulmonar.
2. Escape areo (neumotrax, neumomediastino, neumopericardio, neumoperitoneo). Los factores de riesgo
ms importantes son: enfermedad pulmonar grave, lucha con el respirador por sedacin insuficiente, utilizacin de presiones o volmenes elevados, ventilacin manual, aspiracin enrgica de secreciones. En caso de
neumotrax pequeo con escasa repercusin clnica el
tratamiento consistir en drenaje con aguja en la lnea
media clavicular del segundo espacio intercostal. Si el
neumotrax es grande, debe realizarse drenaje con catter en la lnea axilar media del quinto espacio intercostal,
y conexin a aspiracin continua con presin negativa a
10, 20 cmH2O y sello de agua.
Prevencin. La mejor prevencin de las complicaciones de la VM es buscar una correcta ventilacin del paciente, mediante la utilizacin de volmenes y presiones mnimas necesarias, la utilizacin de modalidades
de VM de soporte, y realizacin de aspiracin frecuente de secreciones de forma suave. Algunos estudios experimentales han demostrado que en los casos de ventilacin con volmenes y/o picos de presin elevados
la aplicacin de PEEP disminuye el dao inducido por
ventilacin al prevenir el colapso y el estiramiento pulmonar.
Aspiracin pulmonar
Est favorecida por las caractersticas del paciente conectado a VM (posicin horizontal, disminucin del nivel de conciencia por coma o medicacin, y relajacin) y
162
agravado por la utilizacin de tubo endotraqueal sin baln en la edad peditrica y la no colocacin de sonda
nasogstrica en el paciente intubado, o bien administracin de alimentacin gstrica. Para prevenir la aspiracin
pulmonar es recomendable mantener al paciente en posicin semiincorporada 30-45, administracin de alimentacin transpilrica, disminuir al mximo la utilizacin de
relajantes musculares optimizando la sedoanalgesia, colocacin de sonda nasogstrica abierta a bolsa, utilizacin
de baln en los pacientes en los que no est contraindicado y aspiracin frecuente suave supragltica (evitando
el reflejo nauseoso).
Problemas hemodinmicos
La VM aumenta la presin intratorcica (sobre todo
cuando se asocia PEEP elevada), por lo que dificulta el
retorno venoso, aumenta la sobrecarga al ventrculo derecho y disminuye el gasto cardaco sistmico, reduciendo la perfusin de otros rganos. Por ello, es necesario realizar una monitorizacin hemodinmica al
menos de la frecuencia cardaca (FC), la PA y la PVC,
especialmente cuando el paciente precise VM con presin media en la va area elevada, compensando el
efecto negativo de la VM con expansin de volumen
y/o frmacos inotrpicos6.
Problemas infecciosos
La VM facilita un importante riesgo de infecciones. La
incidencia de infecciones respiratorias en pacientes sometidos a VM oscila entre el 6 y el 26 %. El mayor riesgo
sucede entre los 5 y 15 das de VM, con una incidencia
creciente a medida que sta se prolonga. Las infecciones
respiratorias ms frecuentes en el nio sometido a VM
son la neumona asociada a VM, la traqueobronquitis, la
sinusitis y la otitis.
00
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
00
163
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
Problemas nutricionales
La VM dificulta la nutricin del paciente crtico, ya que
aumenta la presin intraabdominal y con frecuencia produce distensin gstrica. Adems, la situacin crtica y
los frmacos sedantes disminuyen la motilidad gstrica,
dificultan la tolerancia digestiva y aumentan el riesgo de
aspiracin. Por otra parte, la malnutricin disminuye la
fuerza muscular respiratoria, prolongando la VM. El nio
con VM debe recibir nutricin precozmente, con un
aporte calrico suficiente para evitar la fatiga respiratoria pero sin llegar a la sobrecarga calrica, que aumenta
las necesidades de ventilacin. La va y tcnica de administracin de la nutricin (oral, gstrica intermitente, gstrica continua, transpilrica o nutricin parenteral) depende del estado general del paciente, el estado de
conciencia, la integridad del aparato digestivo y la administracin de sedantes y relajantes musculares, aunque
siempre debe intentarse el mtodo de nutricin ms fi-
164
Lesin pulmonar
La utilizacin prolongada de VM favorece las complicaciones pulmonares a largo plazo por los mismos mecanismos que produce la lesin aguda y la administracin
de oxgeno a concentraciones superiores a 0,6 produce
toxicidad pulmonar y retinopata. Las complicaciones crnicas o secuelas ms importantes son la displasia broncopulmonar (sobre todo en grandes prematuros que han
precisado VM prolongada con altas concentraciones de
oxgeno), la fibrosis pulmonar y la dependencia crnica
de oxigenoterapia.
00
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
EFECTOS PSICOLGICOS
El nio sometido a VM puede sufrir importantes alteraciones psicolgicas ya que el nio est incmodo, se le
somete de forma peridica a maniobras molestas y dolorosas como la aspiracin endotraqueal o el cambio de fijacin del tubo endotraqueal, se le impide la movilidad y frecuentemente est sujeto de pies y manos y no puede
comunicarse verbalmente ni con sus padres ni con el personal sanitario. Por ello, es necesario que se ajuste la sedacin para hacer la ventilacin lo ms cmoda posible, y
facilitar en los nios que estn conscientes la presencia y
comunicacin gestual con los padres y personal sanitario.
BIBLIOGRAFA
1. Velasco Jabalquinto M, Ulloa Santamara E, Lpez-Herce Cid J.
Ventilacion mecnica. En: Lpez-Herce Cid J, Calvo Rey C,
Lorente M, Jaimovich D, Baltodano A, editors. Manual de cuidados intensivos peditricos. Madrid: Publimed, 2001; p. 620-43.
2. Ruza F, Gonzlez Garrido M. Modalidades de ventilacin artificial en el nio. En: Ruza F, editor. Tratado de Cuidados Intensivos Peditricos. Madrid: Norma, 1994; p. 381-99.
Ventilacin no invasiva
M. Pons dena y F.J. Cambra Lasaosa
Unidad de Cuidados Intensivos Peditricos. Unidad Integrada de Pediatra. Hospital Sant Joan de Du-Clnic.
Universidad de Barcelona. Espaa.
00
165
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
Palabras clave:
Ventilacin mecnica no invasiva. BiPAP. Nios. Presin
de soporte.
NONINVASIVE VENTILATION
Noninvasive ventilation (NIV), i.e. without tracheal intubation, has been reintroduced for the treatment of respiratory failure to reduce the complications of mechanical
ventilation. Nowadays, NIV with positive pressure is the
preferred method, applied through a mask held in place
by a harness. Several masks can be used (nasal, bucconasal, facial) and a variety of means can be used to keep
them in place. Many respirators can be selected, ranging
from those traditionally used in the intensive care unit
(ICU) to specific NIV respirators and conventional ICU respirators with specific software for NIV. Many respiratory
modalities can be used according to the respirator (biphasic positive airway pressure [BIPAP], proportional assist
Key words:
Non invasive ventilation. BiPAP. Children. Support pressure.
INTRODUCCIN
La necesidad de minimizar la agresividad y las complicaciones de las medidas teraputicas en los pacientes con insuficiencia respiratoria aguda (IRA) o crnica agudizada
(IRCA) ha favorecido la reutilizacin de la VNI, es decir, sin
intubacin traqueal. La VNI se utiliz hace aproximadamente 40 aos con los pacientes que desarrollaban IRA en
el curso evolutivo de la poliomielitis, entonces se realizaba
ejerciendo una presin negativa sobre la caja torcica con
los llamados pulmones de acero. Actualmente, la tcnica
ms extendida es la VNIP, que se realiza mediante una mascarilla (nasal o facial) sujetada con arneses. La aplicacin de
esta tcnica ventilatoria en adultos tiene casi 20 aos de
evolucin, pero su introduccin en el paciente peditrico se
inici a mediados de 1990 en Estados Unidos1-3.
MATERIAL NECESARIO
Figura 1. VNI con respirador convencional. Insuficiencia
respiratoria hipercpnica en el postoperatorio
de escoliosis.
166
Mascarillas
Existen mascarillas nasales, nasobucales y de cara completa. En adultos se est empezando a utilizar la interfase
tipo casco (Helmet) (figs. 1 y 2).
La mascarilla nasobucal es ms adecuada para situaciones agudas en que el paciente no pueda respirar slo
por la nariz, pero es ms claustrofbica, con mayor riesgo
de aspiracin en caso de vmito y menos cmoda para
su utilizacin a largo plazo. La mascarilla facial completa
minimiza la posibilidad de fugas, con los mismos inconvenientes que la mascarilla nasobucal. La mascarilla
nasal es ms cmoda, pero precisa que el paciente mantenga la boca cerrada, pues en caso contrario la compensacin de fugas har intolerable la mascarilla. Existen
tambin interfases que no son propiamente mascarillas
como la interfase tipo Adams, que se acopla en las fosas
nasales y se apoya sobre el vrtice craneal (fig. 3) evitando producir apoyo sobre el puente nasal o el Helmet diseado para adultos (fig. 4).
00
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
Sistemas de sujecin
La mascarilla puede sujetarse mediante gorros o cinchas. El gorro es ms fcil y rpido de colocar, pero la
transpiracin que permite es menor y resulta incmodo
en verano. Las cinchas precisan de un mayor entrenamiento para el ajuste adecuado, pero son menos calurosas.
Respirador
Respiradores de VNI y convencionales
Respiradores de ventilacin no invasiva (compensan
fugas, fciles de usar y programar, transportables, permiten uso domiciliario, permiten pocas modalidades, no todos tienen oxgeno incorporado, slo algunos permiten
monitorizacin). Existen modelos ciclados por presin
(BiPAP Respironics) y por volumen (Breas, Onyx).
Respiradores convencionales (no compensan fugas, no
se sincronizan bien con el paciente, no suelen tener posibilidad de uso domiciliario, permiten ms modalidades y
posibilidad de monitorizacin).
Administracin de oxgeno. Aerosoles
y humidificacin
El oxgeno se administra en aquellos respiradores que
no lleven toma incorporada, en la parte proximal de la
tubuladura mediante una pieza en T. En algunas mascarillas (Respironics) puede administrarse conectando
la lnea de oxgeno en unos orificios de la mascarilla
pero se producen turbulencias. La humidificacin puede realizarse con tubuladuras especiales con trampa de
agua que permiten intercalar un humidificador. En cuanto a los aerosoles pueden administrarse intercalando
dispositivos de aerosolizacin en los circuitos de las tubuladuras con vlvula espiratoria (respiradores volumtricos).
CPAP
Se genera un nivel de presin positiva en la va area
mediante un flujo continuo, siendo la respiracin del paciente espontnea. Su utilidad est casi limitada a pacientes con apneas (prematuros, lactantes con bronquiolitis) y
en el edema pulmonar cardiognico. Tambin se utiliza
00
167
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
BiPAP
Modalidad de presin producida mediante una turbina
que administra dos niveles de presin (IPAP durante la
inspiracin y EPAP durante la espiracin) y que permite
la sincronizacin con la respiracin espontnea del paciente mediante un trigger de flujo muy sensible, as
como la compensacin de las fugas alrededor de la mascarilla. Es la modalidad de uso ms generalizado para
todo tipo de pacientes y situaciones clnicas.
PARMETROS A FIJAR
A continuacin se detallan las caractersticas bsicas de
la programacin de los dos modelos de respirador ms
utilizados para VNI, as como la programacin ms adecuada para el uso de ventiladores convencionales.
168
00
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
Parmetros iniciales
1. Sensibilidad al mximo, a ser posible con disparo de
flujo (evitando el autociclado).
2. Ajustar la frecuencia respiratoria:
a) Al mximo: frecuencia parecida a la del paciente y
sincronizar.
b) Al mnimo: permitir el ritmo respiratorio del paciente.
3. En caso de fugas elevadas, puede aadirse un flujo
de oxgeno al asa espiratoria para engaar a la alarma de
bajo volumen minuto y apnea.
Algunos respiradores convencionales de ltima generacin, disponen de una modalidad especfica de VNI (Evita 2 Dura, Evita4, Galileo, etc.)
MONITORIZACIN Y VIGILANCIA
VENTILACIN NO INVASIVA
EN EL PACIENTE AGUDO
1. Frecuencia respiratoria.
2. VC administrado.
3. Volumen de fugas.
4. Score de dificultad respiratoria.
5. Presencia de distensin abdominal.
6. Pulsioximetra.
7. Control gasomtrico (CO2).
8. Respuesta clnica: la respuesta clnica suele producirse en la primera hora, reducindose la taquipnea y el
tiraje o bien, en caso contrario, precipitndose la intubacin por el incremento de trabajo respiratorio inducido
por la desadaptacin a la VNI.
00
Los parmetros que se deben vigilar y monitorizar durante la VNI son los siguientes:
VENTAJAS
Las ventajas de la VNI derivan sobre todo de la eliminacin de las complicaciones asociadas a la ventilacin
mecnica convencional (barotrauma, neumona asociada
al ventilador, lesiones laringotraqueales, traquetis necrosante, polineuropata asociada al uso de paralizantes
musculares, etc.). As pues, supone un nuevo paradigma
que rompe el axioma por el cual todo paciente con insuficiencia respiratoria grave precisa intubacin traqueal y
conexin a un respirador.
En su uso domiciliario, en pacientes neuromusculares,
se ha demostrado una mejora en la calidad de vida. Con
ello mejora el descanso nocturno, disminuye la tasa de
complicaciones respiratorias, y el nmero de ingresos hospitalarios4-7. Posiblemente en pocos aos se podr demostrar tambin una prolongacin de la vida media de
este grupo de pacientes.
INCONVENIENTES Y COMPLICACIONES
La ventilacin no invasiva no est exenta de inconvenientes8. Dentro de los ms frecuentes es el mayor requerimiento de atencin y trabajo sobre el paciente para
169
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
170
los circuitos de fuga controlada, la eliminacin del aire espirado, lo cual podra causar hipercapnia por rebreathing.
INDICACIONES
Las indicaciones de la VNI en pediatra se observan tanto en la fase aguda en el mbito hospitalario como de forma continua en el domicilio.
00
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
00
CONTRAINDICACIONES
La tabla 3 recoge las contraindicaciones de la VNI. La
alteracin de conciencia profunda con ausencia del reflejo de tos e incapacidad de eliminacin de secreciones,
la inestabilidad hemodinmica, la insuficiencia respiratoria grave con neumotrax y las complicaciones locales
son las contraindicaciones ms importantes9,10.
BIBLIOGRAFA
1. Padman R, Lawless ST, Von Nessen S. Use of BIPAP by nasal
mask in the treatment of respiratory insufficiency in pediatric
patients: Preliminary investigation. Pediatric Pulmonol 1994;17:
119-23.
2. Padman R, Lawless ST, Kettrick RG. Noninvasive ventilation via
bilevel positive airway pressure support in pediatric practice.
Crit Care Med 1998;26:169-73.
3. Fortenberry JD, Del Toro J, Jefferson LS, Evey L, Haase D.
Management of pediatric acute hypoxemic respiratory insufficiency with bilevel positive pressure (BiPAP) nasal mask ventilation. Chest 1995;108:1059-64.
4. Netzer N, Werner P, Korinthenberg R, Matthys H. Nasal BIPAP
respiration with controlled respiratory mode in neuromuscular
diseases and severe kyphoscoliosis. Pneumologie 1995;49
(Suppl 1):161-4.
5. Daz-Lobato S, Dorgham A, Garca Tejero MT, Redondo MA,
Echevarri C, Martinez J, et al. Home mechanical ventilation by
nasal mask in a patient with Duchennes disease. Rev Neurol
1996;24:562-3.
6. Nirajan V, Bach JR. Noninvasive management of pediatric neuromuscular ventilatory failure. Crit Care Med 1998;26:2061-5.
7. Guilleminault C, Philip P, Robinson A. Sleep and neuromuscular disease: Bilevel positive airway pressure by nasal mask as a
treatment for sleep disordered breathing in patients with neuromuscular disease. J Neurol Neurosurg Psychiatry 1998;65:
225-32.
8. Hill NS. Complications of noninvasive ventilation. Respir Care.
2000;45:480-1.
9. Garca Teresa MA, Martn Barba C. Ventilacin no invasiva. En:
Lpez-Herce J, Calvo C, Lorente M, Jaimovich D, editores.
171
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
11. Fortenberry JD. Noninvasive ventilation in children with respiratory failure. Crit Care Med 1998;26:2095-96.
Palabras clave:
Ventilacin de alta frecuencia. Nios. Estrategia ventilatoria protectora pulmonar. Ventilacin mecnica. Insu-
HIGH-FREQUENCY VENTILATION
In the era of lung-protective ventilation strategies, high
frequency oscillatory ventilation (HFOV) has attracted renewed interest and its use has dramatically increased in
neonatal and pediatric intensive care units. HFOV is able to
reduce ventilator-induced lung injury by limiting the incidence of volutrauma, atelectrauma, barotrauma and biotrauma. During HFOV, adequate oxygenation and ventilation is achieved by using low tidal volumes and small
pressure swings at supraphysiologic frequencies. Unlike
other high-frequency ventilation modes, HFOV has an active expiration phase. HFOV constitutes a safe and successful ventilation mode for managing pediatric patients with
respiratory insufficiency refractory to optimized conventional mechanical ventilation and provides better results
when initiated early. However, the elective use of HFOV requires further studies to identify its benefits over conventional modes of mechanical ventilation and to support its
routine use as a first line therapy. In the present article, the
Respiratory Working Group of the Spanish Society of Pediatric Critical Care reviews the main issues in the pediatric
application of HFOV. In addition, a general practical protocol and specific management strategies, as well as the monitoring, patient care and other special features of the use
of HFOV in the pediatric setting, are discussed.
Key words:
172
00
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
INTRODUCCIN
La ventilacin de alta frecuencia oscilatoria (VAFO)
constituye una modalidad ventilatoria que puede definirse por su carcter protector pulmonar y la aplicacin sistemtica de una estrategia basada en el reclutamiento alveolar y el establecimiento de un volumen pulmonar
ptimo1-4.
Las evidencias existentes en el campo experimental
muestran cmo la VAFO, aplicada correctamente, es capaz de proteger eficazmente el pulmn y disminuir la incidencia de barotrauma, volutrauma, atelectrauma y biotrauma, con un efecto protector del pulmn equivalente
al de la ventilacin lquida parcial y superior al de la VMC
aplicada incluso con una estrategia de reclutamiento-proteccin1-7. En el mbito peditrico, la VAFO constituye
una tcnica de rescate pulmonar segura y eficaz, con mejores resultados cuanto ms precozmente se inicia su uso,
capaz de disminuir la mortalidad y la incidencia de enfermedad pulmonar crnica. En cuanto a la aplicacin
electiva de VAFO frente a la VMC, en pacientes peditricos con patologa respiratoria que precisa ventilacin
asistida, no existen en la actualidad evidencias que demuestren las ventajas o desventajas de una tcnica frente
a la otra1-4,8-14.
CONCEPTOS
La VAFO, a diferencia de otras modalidades de ventilacin, trata de abrir el pulmn y mantenerlo abierto, utilizando para ello volmenes corrientes muy pequeos
(1-2 ml/kg en comparacin con los 8-15 ml/kg utilizados en
VMC) a frecuencias suprafisiolgicas (240-900 resp./min
[4-15 Hz] frente a las 15-30 resp./min utilizadas en VMC)1-4.
El modo exacto por el que se produce el intercambio
gaseoso durante la VAFO no est completamente aclarado, aunque se sabe que intervienen diferentes mecanismos como la ventilacin alveolar directa, la difusin
molecular, el mezclado cardiognico, la dispersin aumentada de Taylor, el perfil de velocidad asimtrica o el
efecto Pendelluft entre otros.
DESCRIPCIN DE LA TCNICA
El aparato de VAFO requiere tres elementos fundamentales: una fuente de flujo de gas fresco continuo, un mecanismo oscilador capaz de generar oscilaciones de la
columna de gas en el circuito y en la va area a altas frecuencias (en general un diafragma o un pistn), y un
puerto espiratorio con un sistema de filtro low-pass que
mantenga la Paw a lo largo de todo el ciclo respiratorio1-4.
El aparato de VAFO genera as ondas oscilatorias de presin que son de alta amplitud a nivel proximal, pero que
se atenan por las fuerzas resistivas del circuito y las
fuerzas conductivas de la va area, de modo que a nivel alveolar slo llegan ondas de presin de baja amplitud (menores del 10 % de la amplitud producida por el
aparato y medida a nivel proximal). En consecuencia, las
00
presiones pico, tanto inspiratorias como espiratorias, estn marcadamente disminuidas en el alvolo. Adems de
esta menor oscilacin de presin, el VC utilizado es ms
pequeo (menor que el espacio muerto anatmico y fisiolgico) y el volumen teleespiratorio mayor que los
empleados en VMC, y ambos son mantenidos de forma
prcticamente constante (menor oscilacin de volumen).
El volumen minuto contina siendo eficaz gracias a la
utilizacin de frecuencias respiratorias muy altas y la
existencia de un sistema de espiracin activa que, adems de permitir una eliminacin de carbnico ms eficaz
y controlable, disminuye el riesgo de atrapamiento de
aire.
Tipo de oscilatoria
Tasa de flujo
(bias-flow)
Diafragma
Diafragma
Pistn
0-40 l/min
0-60 l/min
0-25 l/min
Paw
3-45 cmH2O
3-55 cmH2O
0-25 cmH2O
Delta-P
(mximo proximal)
Frecuencia
3-15 Hz
3-15 Hz
5-15 Hz
Porcentaje inspiratorio
30-50 %
30-50 %
30-40-50 %
35 kg
Sin lmite
20 kg
Sin lmite
Sin lmite
10-12 kg
173
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
3. Compatibilidad con otras tcnicas. La VAFO es compatible con la utilizacin simultnea de otras modalidades
teraputicas respiratorias como el xido ntrico (NO), el
helio, la ventilacin en prono, la insuflacin traqueal o la
ventilacin lquida parcial12.
INDICACIONES
1. Indicacin electiva o rescate. No existen evidencias
de que la aplicacin de la VAFO de manera electiva tenga
ventajas sobre la VMC. Su papel actual es de rescate, ante
el fracaso de la ventilacin mecnica convencional optimizada. El criterio de inclusin empleado en nuestra unidad
de cuidados intensivos peditricos (UCIP) es el siguiente:
insuficiencia respiratoria grave de cualquier etiologa (infecciosa, inhalatoria, etc.) con un ndice de oxigenacin
(IO) mayor de 13, en dos muestras de gases arteriales en
un intervalo de 6 h, a pesar de la utilizacin de VMC optimizada (ndice de oxigenacin = 100 FiO2 presin media va area/PaO2)1-4,8-14.
2. En la prctica, la mayora de los pacientes que precisan VAFO padecen sndrome de dificultad respiratoria
aguda (SDRA) grave o infeccin pulmonar por virus respiratorio sincitial (VRS).
3. Escape areo. Los pacientes con sndromes de escape
areo grosero (neumotrax, neumoperitoneo, neumomediastino, etc.) responden particularmente bien a la VAFO, lo
cual permite la retirada precoz de los drenajes por la rpida
resolucin del escape; por este motivo, en muchos centros
aplicamos la VAFO de manera electiva en estos casos.
CONTRAINDICACIONES
1. Existencia de resistencias en la va area muy aumentadas.
2. Presin intracraneal elevada.
3. Presin arterial media baja.
4. Dependencia de flujo sanguneo pulmonar pasivo.
Estos criterios de exclusin son relativos y deben ser
aplicados de forma individualizada en cada caso, valorndose la relacin riesgo/beneficio, al no existir evidencias de que la VAFO sea ms perjudicial que cualquier
otra modalidad ventilatoria en dichas circunstancias. Son
condiciones que, tericamente, dificultan la aplicacin de
la tcnica, o disminuyen la probabilidad de respuesta positiva a la misma, pero que no constituyen contraindicaciones absolutas1-4,8-14.
1. Aumento de las resistencias en la va area. Tericamente aumenta el riesgo de atrapamiento areo y favorece el mecanismo de hiperinsuflacin dinmica, aunque la aplicacin de una estrategia adecuada (frecuencias
ms bajas, amplitudes menores) puede minimizar los riesgos potenciales de la VAFO.
2. Presin arterial media baja. La necesidad de una
presin arterial media adecuada previa a la instauracin
174
00
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
3. Una vez conectado el circuito y previamente a la conexin al paciente, deben realizarse las calibraciones obligatorias: del circuito del paciente y de comprobacin de
funcionamiento del ventilador (figuran en la carcasa del
aparato y en el manual del operador). Si no se pasan correctamente las calibraciones, debe buscarse el origen del
problema, que en la mayora de los casos se debe al inadecuado montaje del circuito y/o la existencia de fugas.
Delta-P
Paw
% T. insp.
Regulador
de flujo
Alarmas
Regulador
de presin
(Paw)
Frecuencia
NO Flow
NO
Conexin al
paciente
Sensor NOx
Flujo
Del NO 1.000
Figura 1. Cuadro de mandos del ventilador Sensormedics 3100A, esquema del circuito, y conexin de xido ntrico (NO).
00
175
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
176
a) FiO2 < 40 %.
b) Paw < 15 cmH2O.
c) Delta-P < 40 cmH2O.
Modo de retirada
1. Se puede cambiar a modalidad ventilatoria convencional, recomendndose:
00
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
00
Intervencin
Aumentar delta-P
Aumentar delta-P y aumentar
FiO2***
Aumentar delta-P y disminuir Paw
Disminuir delta-P
Disminuir delta-P y aumentar FiO2
Disminuir delta P y disminuir Paw
Aumentar Delta-P y valorar
aumentar Paw
Aumentar Delta-P y aumentar
Paw***
Aumentar Delta P y disminuir
FiO2
Valorar aumentar Paw
Aumentar Paw
Disminuir FiO2
Disminuir delta P
(y valorar aumentar Paw)
Disminuir delta P y aumentar Paw
Disminuir delta P y disminuir FiO2
177
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
Auscultacin
Las caractersticas de la auscultacin durante la VAFO
son distintas a la VMC, pero no por ello intiles.
1. No es posible identificar los ruidos pulmonares normales durante la VAFO.
2. Ruido de la oscilacin del diafragma. Se debe prestar atencin al tono e intensidad del mismo, y especialmente a su simetra en ambos campos pulmonares. Si
deja de ser simtrico, debe valorarse en primer lugar la
necesidad de aspiracin de secreciones. Si la asimetra
persiste tras una aspiracin adecuada, ser necesario un
control radiolgico para descartar una migracin accidental del tubo endotraqueal e intubacin bronquial selectiva, una atelectasia o un neumotrax.
3. Auscultacin cardaca. No es factible durante la
oscilacin. Si es necesario realizarla, se apagar durante
el menor tiempo posible el oscilador (dejarn de oscilar el
diafragma y el paciente), sin desconectar al paciente
del aparato. De este modo, puede auscultarse el corazn,
evitando el desreclutamiento y manteniendo el paciente mientras tanto en modo CPAP.
4. Ruido emitido por el aparato: El cambio en las caractersticas del ruido emitido por el aparato de VAFO
puede indicar un malfuncionamiento del mismo, sobre
todo la necesidad de recambio del diafragma.
178
Monitorizacin respiratoria
2. Pulsioximetra. Constituye la gua para ajustar la
FiO2. Por lo general, su medicin no se interfiere por
la VAFO, aunque la fiabilidad vara en funcin del tipo de
pulsioxmetro utilizado y el tamao del paciente: en los
nios ms pequeos, el temblor que la VAFO produce
puede originar interferencias constantes que no siempre
son identificables en la curva del pulso y que dan lugar a
lecturas errneas.
2. PaO2 y PaCO2. A travs de controles gasomtricos,
que deben ser al menos horarios durante la estabilizacin
inicial del paciente (primeras 6 h); la periodicidad posterior depender de la evolucin del paciente y se har
siempre que se modifiquen los parmetros del aparato.
Alternativamente puede utilizarse la monitorizacin transcutnea de O2 y CO2, o mejor la intraarterial invasiva continua Paratrend (Philips Medical Systems), que adems
de permitir una evaluacin continua en tiempo real de los
parmetros gasomtricos, agiliza, rentabiliza y aumenta el
rango de seguridad de las maniobras de reclutamiento,
as como el ajuste inmediato de parmetros y la deteccin
precoz de complicaciones. No es posible la realizacin de
capnografa durante la VAFO.
3. Radiografa de trax. La primera radiografa de trax debe realizarse una hora despus de la colocacin del
paciente en VAFO, para determinar si el volumen pulmonar es correcto (nivel diafragmtico en 8.-9. costilla),
especialmente cuando an no se est familiarizado con la
tcnica. Ms tarde, debe repetirse cada 6 h hasta la estabilizacin del paciente. La pauta posterior ser segn indicacin mdica, y siempre que se tengan dudas de si
existe sobre o infradistensin pulmonar, o se sospeche la
presencia de alguna complicacin. La radiografa sirve
para determinar si el volumen pulmonar durante la VAFO
es adecuado, por lo que debe realizarse siempre con el
oscilador funcionando, y poniendo especial cuidado en
evitar la desconexin accidental del ventilador durante el
procedimiento.
Monitorizacin hemodinmica
Es esencial monitorizar la frecuencia cardaca, PA media, PVC, tiempo de relleno capilar y, segn el paciente,
gasto cardaco, saturacin venosa mixta y presin de enclavamiento.
00
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
Humidificacin y calentamiento
Se puede utilizar cualquier sistema convencional de humidificacin y calentamiento, intercalado en el circuito
del paciente. Las calibraciones iniciales deben realizarse
con el humidificador ya colocado. El nico cuidado especial aadido es el vaciado frecuente y parcial del depsito de agua, dejando un sello de lquido que evite el cambio en las presiones del sistema.
EFECTOS SECUNDARIOS
A pesar de su papel protector del pulmn, la VAFO no
elimina el riesgo del dao pulmonar agudo y, adems,
tiene una morbimortalidad asociada dependiente de la
experiencia del equipo que realiza la tcnica y/o de la aplicacin inadecuada de sta. La mortalidad global de los
pacientes peditricos ventilados con VAFO oscila entre el
33 y el 45 %, siendo mayor del 85 % la mortalidad de los
pacientes que no responden a la tcnica18,19.
Patologa combinada
1. La principal complicacin es la aparicin de nuevos
sndromes de escape areo (20-25 % de los casos).
2. Repercusin hemodinmica. Algunos trabajos que sugieren que la VAFO puede disminuir el retorno venoso de
la circulacin sistmica al comprometer la circulacin pulmonar cuando el alvolo es sobredistendido. Sin embargo,
esta disminucin del llenado ventricular izquierdo puede
prevenirse mediante un manejo adecuado de los lquidos y
soporte inotrpico; ms an, existen algunos datos que
apuntan que las interferencias de la VAFO sobre la funcin
miocrdica y el gasto cardaco, son similares o menores que
en la VMC con presiones medias y/o PEEP equivalentes.
00
En los casos de patologa combinada, la estrategia aplicada debe individualizarse, dependiendo de la situacin
del paciente y el predominio de cada enfermedad. Idealmente, en el caso de coexistir escape areo y SDRA, trataremos de solucionar en primer lugar la fuga (hiperoxia
e hipercapnia permisivas), para posteriormente, realizar
la estrategia de SDRA (estrategia general de reclutamiento, con prioridad en descenso de FiO2).
179
Documento descargado de http://www.doyma.es el 28/11/2006. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
monitor de NO en el ms distal (v. fig. 1); o alternativamente, la fuente de NO se pone en T a nivel del humificador, y el sensor en alguno de los tres puertos del circuito (en general, en el ms proximal al ventilador).
Ventilacin en prono
Tiene las mismas indicaciones, limitaciones y complicaciones que durante la VMC. Se debe prestar especial atencin en evitar la desconexin accidental durante el cambio
postural, as como a la variacin en el patrn de vibracin
del paciente (mayor amortiguacin y extensin diferente).
BIBLIOGRAFA
1. Martinn-Torres F, Prez Gil J, Ginesta V, Martinn Snchez JM,
lvarez J. Ventilacin de alta frecuencia oscilatoria: pasado,
presente y futuro. Act Anest Reanim 2001;11:136-49.
2. Martinn-Torres F, Rodrguez Nez A, Martinn Snchez JM.
Ventilacin de alta frecuencia oscilatoria en pacientes peditricos. En: Lpez-Herce, editor. Manual de Cuidados Intensivos
Peditricos. Publimed: Madrid, 2001; p. 644-650.
3. Arnold JH. High-frequency ventilation in the pediatric intensive
care unit. Pediatr Crit Care Med 2000;1:93-9.
4. Priebe GP, Arnold JH. High-frequency oscillatory ventilation in
pediatric patients. Respir Care Clin North Am 2001;7:633-45.
5. Hudson L. Progress in understanding ventilator induced lung
injury. JAMA 1999;282:77-8.
Abreviaturas
CPAP:
FiO2:
PA:
PaCO2:
PACO2:
PaO2:
PAO2:
Paw:
PCP:
PCO2:
PEEP:
PH2O:
180
00