ADVIA 2120i System:

Analytical Technology

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved. Answers for life.
ADVIA 2120i System — analytical technology overview

Cytochemistry, nuclear density, and light scatter for white cell identification

Dual-angle laser light scatter measures size and refractive index for
red cell differentiation

Nucleic acid staining for reticulocyte counting

Dual-angle laser light scatter for platelet differentiation by size and refractive index

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 2 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Interpreting graphics

White cell cytograms


Erythrograms
Size
Platelets
Retics

Refractive index or density


Staining intensity
This presentation is for educational purposes only; cytogram patterns do not represent the full range of information needed for diagnosis.

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 3 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
White cell technology

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 4 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
ADVIA 2120i System — white cell technology

• Dual WBC count – internal QC


• Cytochemical classification according to:
• peroxidase activity
• nuclear density
• size and granularity
• Myeloperoxidase:
excellent stability and differentiation
• Expanded linearity of 400 x 109/L
• Accurate flagging and lower review rates

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 5 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Quality of results – WBC

Two methods

Each method measures 2 cell characteristics

Method 1 — Size and perox staining

Method 2 — Size and nuclear density

Unique Siemens Solution

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 6 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
ADVIA 2120i — Peroxidase cytochemistry

Eosinophil RBC

Neutrophil Heat/SDS
formaldehyde
(Dil 1)

Monocyte

4-Chloro-1
Basophil Naphthol (Dil 2)
H2O2 (Dil 3)

Lymphocyte

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 7 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
WBC analysis

Size
Absorption

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 8 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Quality of results — Baso channel — WBC

High angle detector


measures nuclear
density / complexity

Laser
Zero angle

Low angle detector


measures size

Primary WBC and baso count

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 9 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Basophil / nuclear lobularity channel

1. Noise
2. Blast cell nuclei
3. Mononuclear WBCs

Cell Size
4. Basophils
5. Baso Suspect
6. Saturation
7. Polymorphonuclear WBCs
Nuclear Shape and Density

Accurate WBC counts

No lyse-resistant red cells

No cellular interference

Reduced manual reviews

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 10 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Basophil/Nuclear lobularity channel

Eosinophil RBC

Neutrophil

Heat
Monocyte Phthalic Acid
Surfactant

Basophil

Lymphocyte

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 11 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
WBC analysis

Size
Complexity

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 12 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Quality of results — Baso channel — WBC

High angle detector


measures nuclear
density / complexity

Laser
Zero angle

Low angle detector


measures size

Primary WBC and baso count

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 13 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Basophil / nuclear lobularity channel

1. Noise
2. Blast cell nuclei
3. Mononuclear WBCs

Cell Size
4. Basophils
5. Baso Suspect
6. Saturation
7. Polymorphonuclear WBCs
Nuclear Shape and Density

Accurate WBC counts

No lyse-resistant red cells

No cellular interference

Reduced manual reviews

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 14 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
White cell cytograms

Normal

M1 AML

M3 AML

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 15 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Quality of results — WBC

Dual WBC methodologies

Fewer repeats

Lower review rate

Faster TAT

Diagnostic scattergrams

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 16 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
PANDA
Peroxidase Activity and Nuclear Density Analysis

Courtesy of G. d’Onofrio, MD and G. Zini, MD

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 17 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Peroxidase patterns

P2 P2
P0 P1 P3

P5
P0 P2
P1P3P4 P6

Courtesy of G. d’Onofrio and G. Zini

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 18 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Nuclear density patterns — blasts present?

• D0
Normal rounded MNC cluster shape and no
blast flag
• D1
The MNC cluster is shifted down and left to D0
blast area (usually with flag)

D1 D1

Courtesy of G. d’Onofrio and G. Zini

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 19 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Preclassification grid

P0 P1 P2 P3 P4 P5 P6
D0 CLL severe partial CML CML AIDS
PLL MPO-def. MPO-def. MDS
HCL CMML
ALL (L1-L3)
total MPO-def.

D1 ALL (L1-L3) M1 M2 M2 M2 M3v M3


NHL M5a M4 M4 M3v at. CML
M0 M2 M5a CML-BC M4
M5a M4 M5b
M6 M1
M7

Courtesy of G. d’Onofrio and G. Zini

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 20 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Peripheral blood

P2

P2-D1

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 21 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Preclassification grid

P0 P1 P2 P3 P4 P5 P6
D0 CLL severe partial CML CML AIDS
PLL MPO-def. MPO-def. MDS
HCL CMML
ALL (L1-L3)
total MPO-def.

D1 ALL (L1-L3) M1 M2 M2 M2 M3v M3


NHL M5a M4 M4 M3v at. CML
M0 M2 M5a CML-BC M4
M5a M4 M5b
M6 M1
M7

Courtesy of G. d’Onofrio and G. Zini

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 22 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Peripheral blood

P2

M5/AML

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 23 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
ADVIA 2120i System — Body fluid software

Cerebrospinal fluid

Pleural fluid

Peritoneal fluid

Dialysate

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 24 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Body fluid assay selection

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 25 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
ADVIA 2120i System CSF Assay Flow Chart

Combine equal volume 5 mins


CSF + reagent

Patient CSF Prepared


CSF – 300mL reagent sample
Run
Patient CSF result: sample
RBC/WBC count
and WBC Diff
TAT
5-10 minutes

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 26 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Cerebrospinal fluid result

ADVIA® 2120i System Manual


CSF RBC 16 1
CSF WBC 3838 3089
CSF %MN 31 22
CSF %PMN 69 78
CSF %Neut 69 78
CSF %Lymph 3 3
CSF %Mono 28 19
CSF %EOS* 0 0

Example of bacterial meningitis


* Research use only

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 27 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Cerebrospinal fluid result

ADVIA 2120i System Manual


CSF RBC 2 0
CSF WBC 8 6
CSF %MN 83 81
CSF %PMN 17 19
CSF %Neut 17 19
CSF %Lymph 21 30
CSF %Mono 62 51
CSF %EOS* 0 0

Example of ALL
* Research use only

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 28 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
CSF Assay

• Reduces labor, specialized staffing needs, and turnaround time


• Provides quality control, consistency of results, and 24 hour
testing availability
• White cell linearity to zero
• Increases prepared sample stability to enable sample transportation
and batch testing

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 29 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Red cell and reticulocyte technology

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 30 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Quality of results — RBC

RBC accuracy

Measures size and hgb concentration

Quantitative morphology results

Unique Siemens Solution

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 31 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Red cell technology

%Macro 120fl

60fl

%Micro

Hgb.conc. 28g/dl 41g/dl

%Hypo %Hyper

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 32 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Red blood cells — clinical utility

% hypo, micro etc.

Aids thalassemia diagnosis


120fl

Detects spherocytosis

60fl
B12 / folate vs. liver / chemo

Monitoring of EPO therapy


Hgb.conc. 28g/dl 41g/dl
Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.
Page 33 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Quality of results — RBC

Lipemic samples

HGB 13.0 Plasma replacement


HGB = 12.1
MCHC 37.0

CHCM 34.7 Unique to ADVIA®


Direct HGB 12.2

Eliminates plasma replacement

Not effected by lipemia

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 34 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Differential diagnosis:
Iron deficiency and b-thalassemia

MCV = Normal MCV = 61.1 fL MCV = 66 fL


M/H ratio = 0.63 M/H ratio = 2.05

Diagnosis = Diagnosis =
Iron Deficiency b-Thalassemia

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 35 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Typical erythrograms

Normal Iron deficiency Iron treatment Hemoglobin S+C

Megaloblastic Cryoglobulins Hemolytic Uremia Spherocytosis

Elliptocytosis Sickle Cell Anemia a Thalassemia β Thal Major

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 36 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Reticulocyte method

• RBC and platelets are isovolumetrically sphered


• Oxazine 750 differentially stains the reticulocytes
based on their RNA content

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 37 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Reticulocyte parameters

• Reticulocytes slightly larger than mature


red cells (fL)
• Reticulocyte hemoglobin concentration
proportionately less (g/dL)
• Normal reticulocyte hemoglobin content
similar to mature red cells (pg)

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 38 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Functionality of %Hypo and CHr

% hypochromic cells
Provides history of hemoglobin concentration
over previous 8-12 weeks

CHr
Provides snapshot of current hemoglobin
content in developing red cells

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 39 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
%Hypo/Hyper and CHr assays

• Provides screening capabilities for iron deficiency, especially


pediatrics
• Allows for better treatment management of EPO and iron
• Only method validated for monitoring hemodialysis patients in
NICE and European Guidelines

Beginning After 4 days After 2 weeks After 1 month

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 40 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Cost effective

sTfR % Hypo
100

80
Er-ZPP T Sat

60
Sensitivity

40
Ferritin

20

0
0 20 40 60 80 100
Specificity
ref: Tessitori et al 1997

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 41 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Conclusions (Dr. Iain Macdougall,
Consultant Nephrologist, Kings College, UK)

• EPO has revolutionized the management of renal anemia over the last decade.
The use of EPO for other clinical situations is becoming more widespread,
especially oncology
• It has, however, cast light on the concept of functional iron deficiency
• Serum ferritin and TSAT had major limitations in the detection of this condition
• Red cell and reticulocyte assessments using ADVIA 120, 2120, 2120i
technology provide very useful diagnostic information in this context

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 42 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
ADVIA 2120i System — platelet technology

Large Platelets

RBC Ghosts
Platelets

Debris

RBC Fragments

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 43 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Impedance method

Platelet? Platelet?

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 44 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
ADVIA 2120i optical method

Platelet?
Platelet
Platelet?
Red Blood Cell X
Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.
Page 45 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Quality of results – PLT

Two cell characteristics

Size and RI (color)

Includes large platelets

Excludes micro RBCs, ghosts, fragments

Unique Siemens Solution

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 46 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Platelet method

• Provides two-dimensional analysis:


Large Platelets Volume and refractive Index
• Linearity – platelets 0 to 3500 x 109/L
RBC Ghosts
Platelets
• Size range - 1 to 60 fL

Debris • Large platelets included


RBC Fragments
• RBC fragments excluded
• Refractive Index - 1.35 to 1.40
• Indication of platelet activation

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 47 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Platelets

Cat sample

RBC: 8.2 x 1012/L


MCV: 44.8 fL
PLT: 447 x 109/L
MPV: 17.1 fL

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 48 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Evaluation of ADVIA 120 System platelet method

“ADVIA 120 very close to immuno method”


Advances in Routine Laboratory Platelet Counting
S.J.Machin, P.Harrison, D. Grant and C.Briggs.
British Journal of Haematology. 2000,108, 228-235

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 49 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Platelet evaluation results

Impedance 1
Lipemic
200 Impedance 2

Impedance 3

160 ADVIA 120

120

80

40
ITP
0
0 20 40 60 80 100
Immunophenotypic count (x 109/L)

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 50 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
ADVIA 2120i System — nucleated red cell method

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 51 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
NRBC analysis algorithm

Three independent NRBC counts are determined for each sample:

• Two counts are based on the unstained region of the perox histogram
• The 3rd count is a combination of baso and perox counts (barox count)
• MN + baso from baso lobularity (= ba count)
• Neuts + eos from perox (= rox count)

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 52 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
WBC BAROX

An ‘uncontaminated’ BAROX WBC is created by merging “selected” clusters


of the BAso and peROX channels

WBC BA = MN + B

M N

L E Baso

MN PMN NRBC
NRBC
WBC ROX = N + E
Noise

WBC BAROX = (MN + B) + (N + E )


# NRBC = WBCB – BAROX WBC
Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.
Page 53 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Histogram analysis of perox unstained region

Red Raw histogram data of unstained


region of peroxidase cytogram

Blue Log-normal fit of noise region

Green NRBC events region

Yellow Gaussian fit of lymphocyte


region events

Amber Gaussian fit of NRBC events

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 54 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
ADVIA 2120i System — run screen

Areas where
NRBCs are
located

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 55 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Summary NRBC method

The ADVIA 2120i System NRBC enumeration method is:

• Accurate and specific


• Automatically provides corrected WBC counts and differentials
• Does not rely on an NRBC flag
• Does not impact throughput
• Includes all stages of NRBC
• Every sample at no extra cost

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 56 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Multispecies capability

• Random access of over 50 species


• Species-specific WBC differentials
• Species-specific normal ranges
• Superior differentiation between
platelets and RBCs
• Customizable species-specific flagging
• Playback feature for sample reanalysis

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 57 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Versatile species selection

• Companion animals
• Farm animals
• Racing stables
• Laboratory animals
• Industry standard for clinical trials

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 58 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Benefits of the ADVIA 2120i System

QUALITY results from ONE sample


• Dual WBC methods • Reticulocyte hemoglobin content (CHr)
• Peroxidase differential method • Effective renal patient EPO management
• Dual HGB methods (measured • FDA-approved CSF cell counts
and calculated) and differentials
• Quantitative RBC differential • Clinically useful scattergrams
• 2-D optical platelet count • Practical automation solutions with autoslide,
middleware,
• Reduced film review rates
and multidisciplinary options

Overall cost savings and confidence in results!

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 59 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Why Siemens Hematology?

• Results right the first time


• Low film review rates
• Rapid turn around times
• Clinically useful parameters
• Confidence in results
• Cost effective quality

Practical Automation: Improved Productivity

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 60 A91DX-HHS-130985-GC1-4A00 HHS/Hematology
Questions?

Unrestricted © Siemens Healthcare Diagnostics Inc. 2013 All rights reserved.


Page 61 A91DX-HHS-130985-GC1-4A00 HHS/Hematology