Phased Array
Amplitude Analysis
The majority of both conventional and phased array weld inspections can be
considered Amplitude based inspections as opposed to time of flight
applications like TOFD, corrosion mapping, and composite inspections.
Using side drilled holes (SDH) or notches of different size at different depths a
reference sensitivity can be calibrated for comparison to detected flaws during
the inspection. Typically to 80%.
The smaller the notch or SDH, the more gain required resulting in a more
sensitive inspection and greater potential for a reject.
Amplitude analysis is dependent on the data having been acquired in
accordance with the calibration and inspection rules specified in the
referencing code or procedure.
The MX2 uses an 8 bit amplitude resolution (0-100%) for the digitized A-scan
and a 10 bit resolution (0-200%) for the C-scan gate data and gate readings.
Modifying the gate position in analysis mode will convert 10 bit data to 8 bit
data where only 0-100% amplitude is available. In other words, when you
change the gate position all data between 101-200% is converted to 100%
amplitude.
In the example below readings have been selected for both gate A% and
gate B%.
The gate B reading is 33.7%.
The reading for gate A of 86.5% also indicates with a red background that
at least one focal law in the group is saturated.
Gate A in the A-scan display
is normally red, but when it
flashes red\white as pictured
it indicates that at least one
element in the current focal
law is saturated.
The AdBCurve reading calculates the difference between the peak amplitude
signal in gate A and the sizing curve level. (UT>Advanced>Reference
Amplitude>Typically 80%)
This reading is available for both gate A (AdBCurve) and gate B (BdBCurve).
In the below example of inadequate penetration the reference sensitivity was
calibrated to 80% amplitude using a notch in a piping calibration block.
The AdBCurve reading indicates that the 99.8% amplitude signal in gate A is 1.9
dB over the reference sensitivity of 80%. A negative number would indicate below
the reference curve.
The AdBr reading calculates the difference between the peak amplitude signal in
gate A and the reference. (UT>Advanced>Reference Amplitude>On)
This reading is available for both gate A (AdBr) and gate B (BdBr).
In the below example of inadequate penetration the reference sensitivity calibrated
for 80% at 18.8 dB.
The AdBr reading indicates that the 144% amplitude signal in gate A is 4.2 dB over
the reference sensitivity of 80%.
26 dB (AdBr) 18.8 dB (Ref sensitivity) 3.0 dB (Scan sensitivity) = 4.2 dB.
The AdBA reading calculates the difference between the peak amplitude signal in
gate A and the gate threshold level. (Gate\Alarm>Gate>Position>Threshold>40%)
This reading is available for both gate A (AdBA) and gate B (BdBB).
In the below example slag is detected in gate A at 77.9% amplitude.
The AdBA reading indicates that the 77.9% amplitude signal in gate A is 17.8 dB
over the gate A threshold of 10%.
The %(U(r)) reading displays the amplitude of the signal at the position of the
reference cursor on the UT axis
This reading is available for both the red reference cursor %(U(r)) and the green
measure cursor %(U(m)) on the UT axis.
In the below example porosity is detected in gate A. The signal amplitude at the
position of the reference cursor on the UT axis is 71.4% amplitude.
Envelope
In the amplitude C-scan below, the data point circled represents the maximum
signal detected in gate A by the 55.5 degree focal law at 269mm on the scan axis.
>80%
>80%
In the S-scans below are two indications that will produce the exact same pixel
color in the C-scan. (55.5 degrees at 269mm on the scan axis) One is a geometric
reflector from the weld crown and one is rejectable side wall lack of fusion.
Because both signal 1 and 2 are detected in gate A, it is not possible to
differentiate between them without use of the S-scan or A-scan.
?
OD weld crown geometry (2)
1
2
?
12
1
2
By visualizing the S-scan and A-scan that generated the amplitude C-scan pixel at
269mm it is obvious that the indication is a geometric reflector from the far side
weld crown and need not be evaluated as a flaw.
13
By visualizing the S-scan and A-scan that generated the amplitude C-scan pixel at
173mm it is obvious that the indication is SWLF and must be evaluated as a flaw.
The primary indicator for flaw type characterization is the flaws volumetric location.
14