Blood pressure acts as an important health index for monitoring many health problems.
Medical doctors recommends a regular self-monitoring of blood pressure to make sure that
blood pressure is under normal ranges.
As mobile technology springs to new level, there is a need for easily accessible mobile based
technology to measure blood pressure instead of conventional blood pressure meter that
limit their application for home monitoring.
Much work has been done on B.p measurement of which the most noteworthy
are:
Studies by N. Lutter et al. (2002) express that the blood pressure slightly
depends on blood density and blood viscosity, but essentially relies on the pulse
wave velocity (PWV), the velocity pulse.
Design Methodology
Blood Pressure measurement based on Pulse Transit Time (PTT) approach is chosen for this project.
PTT is a delay time measured from R wave of electrocardiogram (ECG) to Systolic wave peak of
PhotoPlethysmoGramm(PPG).
Since PPG can be acquired from finger, ear, brachial or toes, in this project the PPG is taken from finger
utilize the finger probe of pulse oximeter.
Based on the work of C. C. Y. Poon and Y. T. Zhang ,we calculated our own coefficients by for
establishing relationship between PTT ,SBP and DBP.
In the present study, we investigated the relationship between BP and PTT and by using MIMIC
waveform database to obtain coefficients which could be used as constants for the following equation
given below:
SBP = a*Pwv
DBP = a*Pwv
Where a and b are the coefficients obtained after modification of Bramwell Hills equation, between
SBP and PTT and DBP and PTT.
Electrode
Experimental/Simulated observations
Constants relating SBP,DBP and PTT were tested by the help of MIMIC-II
waveform database.
Only two files from the database were selected as their were irregularities in
data of different types and also much noise was present.
So in-order to test our theory we constructed our own ECG and PPG ,not giving
special emphasis to filtering of the signal as we are concerned only with the
peaks of ECG and PPG waveform.
During our intensive struggle to find the relational constant between PTT and BP, we
came across a almost enlightening concept of Respiratory sinus arrhythmia(RSA).
RSA is the effect of respiration on the R-R interval of ECG and also PPG ,during
inspiration R-R interval shortens and vice-versa.
So we calculated the constants assuming and testing when the person is in relaxed
state i.e. respiratory rate between 20-25 pm.
Using our own hardware and digital BP monitor we calculated approximate constant
for calculating B.P.
We are getting problem in deciding the precise coefficients for the range of respiratory
rate because of absence of real-time B.P. machine to check the efficiency of our
algorithm.