FINAL PROJECT

WIRELESS PATIENT VITAL SIGNS


MONITOR USING ZIGBEE
PROTOCOL ON E-HEALTH
PLATFORM
Presented by:
Philip Tobianto Daely
1105110091
Advisors:
Achmad Rizal ST., MT.

Sugondo Hadiyoso ST., MT.

SCHOOL OF ELECTRICAL ENGINEERING


TELKOM UNIVERSITY
BANDUNG
2015

Background
Objectives
Literature review
Product Design
Testing
Conclusion

Overview

Patient and doctor wont always near each other.

Background

To integrate ECG, HR, RR, SpO2 and body


temperature measurement in one device using
e-Health Platform.
To implement ZigBee protocol in data
transmission from sensors to PC.
To display the data with GUI using real-time
chart and numeric form.

Objective

Vital Signs
ZigBee Protocol
e-Health Sensor Platform

Literature Review

Measurements of basic body functions.


Consist of:

Electrocardiogram
Heart Rate
Respiration Rate
Oxygen Saturation
Body Temperature

Vital Signs

Protocol for WPAN


Based on IEEE 802.15.4
Work in frequency of 2.4 GHz with data rate of
250 kbps.
Support star, mesh and cluster tree network
topology.

ZigBee Protocol

ZigBee Stack Layer

ZigBee Device Types

64-bit addressing

Assigned during manufacturing


Includes a 3-byte OUI from IEEE

16-bit addressing

Received when join a ZigBee network


Not static

ZigBee Addressing

Platform for biometric / medical application


Can monitor many medical sensors, such as:

Pulse oximeter
ECG electrodes
Respiration sensor (thermocouple)
Temperature sensor (thermistor)

e-Health Sensor Platform

Data Acquisition Device (DAD)


Monitoring Device (MD)

Product Design

DAD Block Diagram

DAD Software Flowchart

Device type: End device


PAN ID: 1994
16-bit destination address: 0x0000 (coordinator)

XBee Configuration

MD Block Diagram

Device type: Coordinator


PAN ID: 1994
16-bit destination address: 0xFFFF (broadcast)

Xbee Configuration

GUI

GUI Flowchart

Communication range test


ECG signal detection test
Temperature measurement test
Respiratory rate measurement test

Testing

Tx

Packet

Error

Loss

1 meter

99%

1st floor

6 meter

37

62%

1st floor

15 meter

14

85%

1st floor

20 meter

11

88%

1st floor

5 meter

100%

2nd floor

10 meter

35

63%

3rd floor

No

Range

%Success

End Device
Position

Communication Range Test

ECG Signal Detection Test

Sample (n) Digital Thermometer (C) Tested Device (C) Error (C)
1

35.6

37.5

+1.9

35.6

36.6

+1.0

35.6

37.0

+1.4

35.6

36.9

+1.3

35.6

38.1

+2.5

35.6

37.5

+1.9

35.6

37.4

+1.8

35.6

37.5

+1.9

35.6

37.5

+1.9

10

35.6

36.9

+1.3

Temperature Measurement Test

Sample (n) Manual (breath/minute) Tested Device (breath/minute) Error (breath/minute)


1

19

19

19

17

-2

19

18

-1

19

20

+1

19

19

19

18

-1

19

19

19

19

19

19

-3

10

19

19

-1

Respiratory Rate Measurement Test

Maximum communication range using ZigBee


protocol is 20 meter with success rate of 62%100%
ECG measurement can show a decent signal
that show P, Q, R, S and T component.
Temperature measurement has measurement
error deviation of 1.8C or 5.1% from digital
thermometer result.

Conclusion

Respiratory rate measurement has


measurement error deviation of 1
breath/minute or 5.3 breath/minute from
manual measurement.
From overall results, the designed device has
satisfied the requirement.

Conclusion (cont...)

Questions and Suggestions