Oxygen Monitoring in Acute Medicine


Dave Sainsbury
Staff Anaesthetist, Womens & Childrens Hospital Adelaide, South Australia

Entropy

Life

n n
1/25/2011

2nd law of thermodynamics Entropy is times arrow


Oxygen Monitoring 2

CO2

O2 CnHnOn
1/25/2011 Oxygen Monitoring 3

159 150

149

100

50

I n s p i r e d

100

T ra c h ea

92
A r t e r y

A l v e o l u s

Tissue
40 2

Vein
40

Mitochondria
1/25/2011 Oxygen Monitoring 4

Monitoring Oxygen As a Gas


Inspired Tracheal Alveolar

Oxygen analyser Calculated from FIO2 Calculated: Alveolar air equation End tidal oxygen analyser

1/25/2011

Oxygen Monitoring

Ideal Oxygen Analyser


n n n n n

Rapid response Continuous Robust Low maintenance Do no harm

1/25/2011

Oxygen Monitoring

Oxygen Analysers
n n n

Polarographic sensor Fuel cell Paramagnetic analyzer


Robust Rapid response time Low maintenance

1/25/2011

Oxygen Monitoring

Paramagnetic Oxygen Analyser


Pressure transducer

1/25/2011

Oxygen Monitoring

Oxygram
Inspired 30% 20% 10% 0% Expired

1/25/2011

Oxygen Monitoring

Alveolar Air Equation


n

PAO2 = FIO2 * (Pb- 47) - PaCO2 PAO2 = FIO2 * (Pb- 47) - PaCO2 /R PAO2 = FIO2 * (Pb- 47) - PaCO2 /R(1-FIO2(1-R))

n n

1/25/2011

Oxygen Monitoring

10

159 150

149

100

50

I n s p i r e d

100

T ra c h ea

92
A r t e r y

A l v e o l u s

Tissue
40 2

Vein
40

Mitochondria
1/25/2011 Oxygen Monitoring 11

Oxygen saturation of haemoglobin (%)

100 90 80 70 60 50 40 30 20 10 0 10

20

Vein P50

15

10

20

40

60

80

Partial pressure of oxygen (mmHg)


1/25/2011 Oxygen Monitoring

100

200

300

400

500

600

12

Oxygen content of blood (ml / 100 ml)

Artery

Ideal blood oxygen monitor


n n

Indicate oxygen tension Indicate oxygen supply


Haemoglobin concentration Haemoglobin saturation Blood flow

n n n n n n n
1/25/2011

Rapid response Continuous Safe


Non-invasive

Robust Low maintenance Stable calibration Affordable


Oxygen Monitoring 13

Monitoring blood oxygen

Arterial tension

Blood gas analyser Trans cutaneous Calculated from tension Co-oximeter Pulse oximeter

Arterial saturation

1/25/2011

Oxygen Monitoring

14

Blood gas analyser


n n n n

Polarographic Measures tension Calculates saturation (functional) Problems


invasive intermittent

1/25/2011

Oxygen Monitoring

15

Functional Saturation OxyHb (OxyHb+reducedHb)


7g OxyHb 1g reducedHb 2g metHb

Functional Saturation

= 7/(7+1) = 87%
16

1/25/2011

Oxygen Monitoring

Transcutaneous gas analyser


n n

Polarographic Measures tension


persistent pulmonary hypertension diaphragmatic hernia

Problems
Delicate Heated Repositioning Recalibrating
Oxygen Monitoring 17

1/25/2011

Co oximeters
n n

Measures haemoglobin types by absorption of light To measure 4 types of Haemoglobin measure absorption at 4 different wavelengths OHb reducedHb COHb metHb

1/25/2011

Oxygen Monitoring

18

Fractional Saturation OHb (OHb+reducedHb+COHb +metHb)


7g OxyHb 1g reducedHb 2g MetHb

Fractional Saturation

= 7/(7+1+2) = 70%
19

1/25/2011

Oxygen Monitoring

Oximeters

1875 vonVierordt 1972 Takuo Aoyagi - "Pulse Oximetry" 1980's Nellcor 1988 pulse oximetry mandatory in Australia

1/25/2011

Oxygen Monitoring

20

Systole

Light absorption due to arterial blood

Diastole

Light absorption due to tissues Artery (pulsatile tissue) Non-arterial, (non-pulsatile) tissue

1/25/2011

Oxygen Monitoring

21

10

Red

Infrared

Extinction coefficient

1.0

Oxyhaemoglobin
.1

Deoxyhaemoglobin

.01 600 640 680 720 760 800 840 880 920 960 1000

Light wavelength (NM)


1/25/2011 Oxygen Monitoring 22

Oxygen saturation of haemoglobin (%)

100 90 80 70 60 50 40 30 20 10 0 0.4 1.0 2.0


Oxygen Monitoring

R=

Red absorption Infra-red absorption

3.0

3.4
23

1/25/2011

Oximeter errors
n n

Signal artefact Light absorption artefact

1/25/2011

Oxygen Monitoring

24

Signal artefact
Ambient light Poor perfusion Movement artefact If R = 1 pulse oximeter shows 85%

n n n n

1/25/2011

Oxygen Monitoring

25

Light absorption artefact


n n

carboxyHb metHb

1/25/2011

Oxygen Monitoring

26

10

Red

Infrared

Extinction coefficient

1.0

Oxyhaemoglobin
.1

Deoxyhaemoglobin Carboxyhaemoglobin
720 760 800 840 880 920 960 1000

.01 600 640 680

Light wavelength (NM)


1/25/2011 Oxygen Monitoring 27

10

Red

Infrared

Extinction coefficient

1.0

Methaemoglobin Oxyhaemoglobin

.1

Deoxyhaemoglobin

.01 600 640 680 720 760 800 840 880 920 960 1000

Light wavelength (NM)


1/25/2011 Oxygen Monitoring 28

Light absorption artefact


n

carboxyHb
error showing low functional saturation error showing high fractional saturation

metHb
error toward 85%

1/25/2011

Oxygen Monitoring

29

Light absorption artefact


n

carboxyHb
reads low functional, high fractional

metHb
error toward 85%

n n n n n
1/25/2011

fetal Hb dyes nail polish bilirubin skin pigment


Oxygen Monitoring 30

Monitoring oxygen

Tissue Venous tension Venous saturation

Cytochrome oximeters Blood gas analyser Optical fiber oximeter

1/25/2011

Oxygen Monitoring

31

Monitoring Oxygen
Oxygen supply is critical No single monitor of oxygen supply Oxygen tension versus haemoglobin saturation Functional versus fractional saturation