Está en la página 1de 7

Technical Data Sheet

SMP - PORTABLE FIELD MONITORING SYSTEM


Site surveys, EMF assessments and Validation of places for public use. Certification of telecommunications infrastructure, industrial facilities and machinery. Safety audits of electromagnetic fields. Compliance with European Directive 2004/40/EC.

Description
The SMP Portable Electromagnetic Field Monitoring System can measure electromagnetic field levels generated by any source within the frequency range of the probe attached to the system. For example it can measure fields generated by different sources such as mobile telephone base stations, radio and television transmitters and repeaters, DECT systems, Wi-Fi systems, etc. Those measurements can be made quickly and easily anywhere because the SMP is portable and simple to handle and use. The data stored in the memory can easily be downloaded to a computer via a USB port for later processing in the software tool provided with the unit.

v 1.0

Pallars, 65-71 08018 Barcelona (Spain) | Telephone: (+34) 933 208 055 | Fax: (+34) 933 208 056

www.wavecontrol.com

Technical Data Sheet


The SMP allows measurement of the strength of electromagnetic fields generated by mobile telephone base stations and other telecommunication services, to: Assess, audit and certify telecommunication and industrial facilities. Certify compliance with current applicable law on protection of the general public from electromagnetic fields. EU Council Recommendation 1999/519/EC Certify compliance with current applicable law on protection of workers from electromagnetic fields. European Directive 2004/40/EC

National & international standards


The measurements are performed according to recognized international standards recommendations related to the evaluation of the human exposure to electromagnetic fields: European Union Council Recommendation of 12 July 1999 on the limitation of exposure of the general public to electromagnetic fields (0 Hz to 300 GHz). ICNIRP guidelines for limiting exposure to timevarying electric, magnetic and electromagnetic fields (up to 300 GHz). IEC 62577, IEC 62311, IEC 62369-1. ECC recommendation 02(04) measuring non-ionising electromagnetic radiation (9 kHz 300 GHz). EN 50400, EN 50401, EN 50499, EN 50496, EN 50475, EN 50476, EN 50492, EN 50413, EN 50357. IEEE Std C95.3-2002. Health Canada: Safety Code 6. and

Pallars, 65-71 08018 Barcelona (Spain) | Telephone: (+34) 933 208 055 | Fax: (+34) 933 208 056

www.wavecontrol.com

Technical Data Sheet Main Features


The SMP samples the electromagnetic field strength (V/m) continuously and stores data automatically in internal memory at user-configured intervals with a capacity of more than 62,000 samples (27,300 samples with GPS data).

The SMP is designed for simple and intuitive operation with 2 basic functions: instant and averaged field measurement. storage of field measurements in memory. The SMP integrates a new multimeasurement system that allows instantaneous measurements of the total field level in three (isotropic) axes, Ex, Ey and Ez separately. Displayed measurements can be continuous, or processed into arithmetic average, minimum and maximum values, or international measurement standards. The internal configurable alarm generates an audible signal (warning) when the measured level exceeds the adjustable threshold. The SMP can optionally be equipped with an internal GPS module (factory fit option) so the measured data can be associated with its position and displayed on a GIS (mapping) application. Several field probes are available that can be easily interchanged thanks to its automatic probe recognition system. The SMP contains a rechargeable Li-ion battery allowing continuous field operation for over 10 hours (GPS off).

Pallars, 65-71 08018 Barcelona (Spain) | Telephone: (+34) 933 208 055 | Fax: (+34) 933 208 056

www.wavecontrol.com

Technical Data Sheet


Measurement Functions
Measurement field probes Measurement units Acquisition time Result types Average type Averaging time Alarm Interchangeable; automatic probe detection and recognition system V/m, kV/m, W/cm2, mW/cm2, W/m2 2 samples per second Actual total field value, axis field value (X, Y, Z), maximum and minimum Sliding average (continuous), arithmetic average, normative average 10 s, 30 s, 1 min, 2 min, 5 min, 6 min, 10 min, 15 min, 30 min Audible signal, 2400 Hz; adjustable threshold

Memory
Measurement field probes Capacity 1 s, 5 s, 10 s, 30 s, 1 min, 5 min 62,400 samples (without GPS data) 27,300 samples (with GPS data) Registers Dynamic, up to 100 different memory registers

Interfaces
Probe Data download Firmware update Battery charger Push-pull connector. Automatic detection and recognition Mini USB connector Mini USB connector Coaxial jack connector + LED

Display
Display type Display size Back light Display range Transmissive, monochrome, STN, negative 60 x 78 mm (240 x 320 pixels) White LED 5 digits: 0.0001 to 99999

GPS
General Chip set Position accuracy Update rate Sensitivity Hyper sensitivity with 20 independent channel satellite tracking system SiRF starIII GSC3 1.5 m (CEP50) , 1.8 m (CEP95) 1s - 157 dBm (tracking)

SBAS (Satellite Based Augmentation System) WAAS/EGNOS/MSAS support Geodetic system WGS 84

Measurement Functions
Battery Operation time Temperature range Size (without probe) Weight (without probe) Rechargeable Li-ion battery > 10 hours (GPS enabled) > 14 hours (GPS disabled) -10 C to +45 C 100 x 215 x 40 mm 490 g (including the internal GPS receiver)
www.wavecontrol.com

Pallars, 65-71 08018 Barcelona (Spain) | Telephone: (+34) 933 208 055 | Fax: (+34) 933 208 056

ELEC TRIC FIELD PR O BES

WPF3 - BROAD BAND PROBE Measuring Electric Field from 100 kHz to 3 GHz

Isotropic and RMS measurement FM, TV, mobile telephony, ... High sensitivity starting at 0.3 V/m

Technical Specifications
Sensor type Frequency band Range Isotropic, RMS diode technology 100 kHz - 3 GHz 0.3 - 150 V/m (CW) 0.3 - 25 V/m (RMS) 54 dB 0.3 V/m < 5% 1.8 dB (1 MHz - 3 GHz) 0.5 dB (0.5 V/m - 100 V/m) 1 dB Internal EEPROM memory 24 months (recommended) 0C to +50C +0.5 dB/-1 dB (5C to 50C, related to 25C) 28.5 x 10.5 x 10.5 cm. 95 g.

Dynamic range Sensitivity Resolution Frequency response Linearity Isotropic deviation Calibration Calibration period Operating temperature Temperature response Dimensions Weight

v 2.0
Pallars, 65-71 08018 Barcelona (Spain) | Telephone: (+34) 933 208 055 | Fax: (+34) 933 208 056

www.wavecontrol.com

ELEC TRIC FIELD PR O BES

WPF8 - BROADBAND PROBE Measuring Electric Field from 100 kHz to 8 GHz
Isotropic and RMS measurement New Frequency band from 100kHz to 8 GHz High sensitivity starting at 0.2 V/m

Technical Specifications
Sensor type Frequency band Range Isotropic, RMS diode technology 100 kHz - 8 GHz 0.2 - 120 V/m (CW) 0.2 - 20 V/m (RMS) 52 dB 0.2 V/m < 5% 2 dB (4 MHz - 7.5 GHz) 1 dB (900 MHz - 3 GHz) 0.5 dB (0,5 - 100 V/m) 1.5 dB (100 kHz 8 GHz) Internal EEPROM memory 24 months (recommended) 0C to +50C 0.5 dB (0C to 50C, related to 25C) 28.4 cm x 6 cm 95 g.
v 2.0
Pallars, 65-71 08018 Barcelona (Spain) | Telephone: (+34) 933 208 055 | Fax: (+34) 933 208 056

Dynamic range Sensitivity Resolution Frequency response Linearity Isotropic deviation Calibration Calibration period Operating temperature Temperature response Dimensions Weight

www.wavecontrol.com

ELEC TRIC FIELD PR O BES

WPT - MOBILE TELEPHONY PROBE Measuring 900, 1800 and 2100 bands Electric Field

Isotropic and True RMS measurement The best sensitivity starting at 0.06 V/m High interband attenuation

Technical Specifications
Sensor type Frequency band Range Dynamic range Sensitivity Resolution Interband attenuation Isotropic, True RMS GSM900, GSM1800 and 3G 0.06 61 V/m (RMS) > 60 dB 0.06 V/m 0.1 dB >15 dB between GSM900 GSM1800 >5 dB between GSM1800 3G 1.5 dB (950, 1850 and 2140 MHz) < 0.6 dB (0.15 - 50 V/m) < 0.5 dB Internal EEPROM memory 24 months (recommended)

Frequency response Linearity Isotropic deviation Calibration Calibration period

Operating temperature -30C to +80C Temperature response 0.8 dB (-20C to 60C) Dimensions Weight 28.5 x 10.5 x 10.5 cm. 140 g.

v 2.0
Pallars, 65-71 08018 Barcelona (Spain) | Telephone: (+34) 933 208 055 | Fax: (+34) 933 208 056

www.wavecontrol.com