Certificate of Calibration BO AU aR UNE Everett Service Center Certificate Number: EVL838430 Data Type: As-Left Calibration Date: 28-Oct-2022 Result Summary: In Tolerance 25-Oct-2023 Manufacturer: Fluke Networks Certificate Date: 26-Oct-2022 Model: Nautilus-5011 224°C Serial Number: 19030324 392% Description: DSX-Verification 2G NEXT ARTIFACT Verification Artifact Characterization Rovisi 2 ALCARED SAC Lima, Country: PE Purchase Order: CCS ALVARO JESUS CAYO UREUTA RMA: 32510468 “This calibration is teceable fo the International System of Unis (SD, tough National Maroiogy instilas (NIST, PTE, NRC, NPL, etc], ralometrc technique, cor natwal physical contanis. Tis certeate apples only to the item dented and shall not be reproduced othe! than f ful, wihout the specie writen approve by Fluke Corporation. Calbreton certficaies witout signature are not vald. The calltation has been completed in accordance wih Ficke Electronics Corporation Quality System Document 111.0 Revision 124 andlor Fluke 17025 Quality Manual GSD 111.41 Revision 007 ‘The Data Type found inthis cetcate must be interpreted as: + AS Found Calration data collected before the uns aausted andr repired + AS-Left Calibration data colected attr the unit has ben adjuted and ar repaired * Found-Lett Calixtion data colected without any adjustment and or repair perfec, Measurement resus greater than mis of ero are ineated by: Calibratea OUR «pg cow: evsssan Epp cabee gona © BRE - Deon: 25082003 ERE su: soos BERS Tikes wweubecom ey Fluke Comoation Telephone Facsimile Inerot Revision 220 “OTE eeW eet WAGEDTA —— waa = —— Page 1 of 3 Certificate of Calibration eee Everett Service Center Certificate Number: EVLB36440 Data Type: As-Left Calibration Date; 25-Oct-2022 Result Summary: In Tolerance Calibration Du 25-Oct-2023 Manufacturer: Fluke Networks Certificate Date: 25-Oct-2022 Mode! Nautilus-5011 Temperatu 224°C Serial Number: 19030343 Humidity: 39.2% Description: DSX-Verification Option(s): 2G CDNEXT ARTIFACT Procedur Verification Artitact Characterization 2 Customer: ALCARED SAC. City: LMA Country: PE Purchase Order: CCS ALVARO JESUS CAYO UREUTA RMA: 32510468 Tis calbraton 's aceable to he Intemational Sytem of Units (S), trough National Metrology neltutes (NIST, PTB, NRC, NPL, ete), ramets ochre, ‘or natural physical consiants. This certficate apples only to he Her identi an shall not be reproduced other than in ful what the spect wien approval by Fluke Corporation. Cattration cartficates without signature are not vaid. The ealtraton nas been comploted in accordance with Fluke Electronics Corporation Quality System Document 111.0 Revision 124 andlor Fluke 17025 Quallly Manual GSD 111,41 Revision 07. ‘The Dela Type found in is certeste must be interpreted as: + AS-Found Caltrain data collect before the untis adjusted and or raped, + AS-LeM Calltraton data collected after ne unit hasbeen adjusted and/or repaired, * Found-Lett Calration data collected without any adjustment and / or resi performed, Messurement results greatr than its of error are incest by Calibrated (EHUREN) ggg NGberord PH: came abonama« FEEZ Dvedae §—5-0ct2003 |g RAE sn ssos0a3 | 3 SG E ick Powers ‘we tute com Issued By Fluke Corporation Telephone Facsimile Internet Revision 2.20 -237aN SSW, SWAB USA SST cr wwnaaoe Page 1 of 3 icate of Calibration Everett Service Center 180 80012015 (1010012) Certificate Number: EVL838447 Data Type: As-Left Calibration Date: 25-Oct-2022 Result Summary: In Tolerance Calibration Due: 25-Oct-2023 Manufacturer: Fluke Networks Certificate Date: 25-Oct-2022 Model: Nautilus-6011 Temperature: 22.4°C Serial Number: 19030357 Humidity: 39.2% Description: DSX-Verification Option(; 2G CMDM RL ARTIFACT. Procedure: Verification Artifact Characterization Revision: 2 Customer: ALCARED SAC LIMA, Country: PE Purchase Order: + CCS ALVARO JESUS CAYO UREUTA, RMA: 32510468 ‘This caliretion i Waceable tothe Intemational System of Unis (SD, trough National Meirology nettles (NIST, PTB, NRC, NPL, ole), ralomotia techniques, natural physical constants, This cetest applies only to the ler derived and shall not be reproduced other than in fl hoc the Space writen aphvoval by Fluke Corporation, Cabatcn certcates without signature are not vale. The caltration has been completed in accordance with Fluke Elesttonics Corporation Quaity System Document 111.0 Revision 124 andlor Fluke 17028 Quality Manual QD 111.44 Revision 007 ‘The Daia Type found in this cetcale mus be interpreted a: * AS-Found Calton data colectd before the unt is aeusted an er repaired + AS-Left Calibration data collected alert unit has been acsted and or repaires, * Fund-Lett Calibration data collected without any aojustment and / or repair performed, “Measurement results greater than ints of eror are indeated by Calibrated mCUREN «PET a cus: evaseint Epp cate aeowama ESE Dome 208m BEER sm; “spoaoasr EB RE wa vow fuke.com ey Fluke Corporation Telephone Facsimile tnornet Revision 2.20 “oT om we RADIA aE oo Torn Page 1 of 3 Certificate of Calibration Everett Service Center 180 9001:2016 (1010012) Certificate Number: EVL638442 Data Type: As-Lett Calibration Date eee Result Summary: In Tolerance Calibration Duc 25-Oct-2023 Manufacturer: Fluke Networks Certificate Dat 26-Oct-2022 Model: Nautilus-5011 Temperatur 224°C Serial Number: 19030366 392% Description: DSX-Verification Option(s): 2G TCL ARTIFACT Procedure: Verification Artifact Characterization Revision: 2 Customer: ALCARED SAC city: uMA Country: PE Purchase Order: CCS ALVARO JESUS CAYO UREUTA RMA: 32510468 Ths calbraton is aoeabie othe Inlemational System of Unis (SD, though National Metioay ihstuies (NIST, PTB, NRC, NPL, atc), raiomeir tecniques, ‘or natural physical constant. Thi cetfcate applies erly to the tem dented end shall not be reproduced othe than n ful, without the specs wien approval by Fluke Corporation. Calibration cerfcates witout signature are not vald. The caliation has been compleled in accordance with Fluke Electronics Corporation Quality System Document 111.0 Revision 124 andlor Fluke 17025 Quality Manual QSD 111.41 Revision 007 ‘Tho Data Typo found in thie cetfeate must be interpreted as * As-Found Callbraton data collected before the untis adjusted and / or repaired + As-Left Calibration deta collected aftr he unit has been adjusted and/or repaired. + Found-Left Caltraon data collected without any adustment and / or repair performed. Messurement results greater than its of ror ae neat by" Calibrated eau «gE came: soem | EERE Debi onan «FREE 3n isos «8 BEE Tick powers wacom ‘sued by Fue Corporation Tetapnone Facsim Internet Revision 220 Page 1 of 3 Certificate of Calibration Everett Service Center 18 (1010072) AscLeft Calibration Date: 25-Oct-2022 In Tolerance Calibration Due: 25-Oct-2023, Fluke Networks 25-Oct-2022 Neutilus-5010 24°C SerialNumber: 18210516 30.2% Description: DSX-Verification Option( 2G IIFEXT ARTIFACT Procedure: Verification Artifact Characterization Revi 2 Customer: ALCARED SAC City: LIMA Country: PE Purchase Order: CCS ALVARO JESUS CAYO UREUTA RMA: 32510468 “This calration is Waceabie to the Intemational System of Unis (S), tough National Metoogy instules (NIST, PTB, NRC, NPL, eto), ralomebie techniques, ‘oF naturel physical constants, This certificate apples oly to the item idantted and shad not ba reproduced ether than infu, wihout the specie win approve by Fluke Corporation. Calibration cercates witout signature are not vals. The calbration has deen complelad in accordance with Fluke Electronica Corporation Quality System Document 111.0 Revision 124 andor Fuke 17025 Qualty Manual QSD 111.41 Revision 007. ‘The Data Type found inthis cetficate must be interpreted as AS Found Calibration data coleced before the units asusted and or repaired. + As+Left—Caration data colected ater the unt nas been dusts and o epsited. + FoundLeft Carton data collected without any adjustment and or repair performed, "Measurement rests areatar than its of erar are incest by Calibrated GUORER FEL cus: eves bay cutwe gona = BREE Omoae ont «FE RE om: toaoss, BRE Tid roves vette can fara ute Corporation Telephone Facsimile Internet Revision 220 "7S SSH Eom WA RESTS aa ea Toe cor Page 1 of 3 ae VS Certificate of Calibration 10 900122015 (1010072) Everett Service Center Certificate Number: EVL836447 Data Type: As-Left Calibration Date: 25-Oct-2022 Result Summary: In Tolerance Calibration Du 25-Oct-2023 Manufacturer: Fluke Networks Certificate Date: 25-Oct-2022 Model: Nautilus-5010 Temperatu 224°C Serial Number: 18210488 Humidity: 392% Description: DSX-Verification Option(s): 2G LOOP RESISTANCE ARTIF Procedure: Verification Artifact Characterization Revision: 2 Customer: ALCARED SAC City: Lima Country: PE Purchase Order: CCS ALVARO JESUS CAYO UREUTA RMA: 92510468, This calbraton is teceable fo he Intemational System of Uns (S), trough National Melrougy nstules (NIST, PTB, NRC, NPL, etc), raiomotic tochnques, ‘or natural physica constanis. This vetfcale apples only 1 he lier dented and shall nat be reproduced otyer than in ful, wihodt the specie wren approvsl by Fluke Corporation. Calbraton cortificates without signature are not vad. The calibration Ras Been completed in accordance with Fluke Eloctonics Corporation Quality System Document 111.0 Revision 124 andlor Fluke 17025 Quality Manual QSD 111.41 Revision 007 ‘The Data Type found inthis cmstfcate must be interpreted as: + As-Found Calitration deta cotected before the unitis adjusted and / or repaired + AS-Loft Caloation data colectod afer he uit has been aqjsted and /r repaired. * Found-Lett Caltoration data colecoa witout any aajustment and/or repair performed “Messurement results greater than its of eror ae inacate by cottos BREN EFS Dense: 25002003 f BRE econ _ rs face ae Page 1 of 3 180 9001:2018 (1010/2) Everett Service Center Certificate Number: EVL838448 Data Type: As-Lett Result Summary: In Tolerance Certificate of Calibration Calibration Date: 25-Oct-2022 Calibration Due: 25-Oct-2023 Manufacturer: Fluke Networks Certificate Date: 25-Oct-2022 Model: Nautilus-5010 Temperature: 224°C Serial Number; 18210447 Humidity: 392% Description: DSX-Verification Option( 2G RESISTANCE UNBALANCED Procedure: Verification Artifact Characterization Revision: 2 Customer: ‘ALCARED SAC city: uma Country: PE Purchase Order: CCS ALVARO JESUS CAYO UREUTA RMA: 932510468 This calibration is traceable to he Intemational System of Unis (S), Byough Nalonal Motrology Ineltules (NIST, PTB, NRC, NPL, etc), reloretic techniques, ‘oF natural physical constants. Ts certhicate apples ony tote em ierdied and stall nat be reproduced othe than in fll witha the specie wetton approved by Fluke Corporation. Catoaton certicates witout signature are not vad. The eaibration has been completed in accordance with Fluke Electonics Corporation Quaity System Document 111.0 Revision 124 andor Fuke 17025 Quality Manual QSD 111.41 Revision 007. ‘The Data Type fond in thie catfoate must be interpreted 3: + AS-Found Calton data colecte botore the unite adjusted and or repaired, + AS-Laft Calibration daa collected after the unithas been asjuted and/or ropared. * Found-Lem Calibration data collected without ary adjustment and or repair performed Measurement results grastr than its of ero ae indicat by: calirates Gia) BFE cus: evens Leg cate eonse | BRE mba: sonams «=F EEE on: imo ARE Tl Pome ‘vtincom ery Fluka Corporation Telephone Facsimile internet Revision 220 Se aT —— a aes reas Page 1 of 3 VS Certificate of Calibration 180 9001:2015 (1010012) Everett Service Center Cerificate Number: EVLBSEIaS Data Type: AsLeft Calibration Date: 25-Oct-2022 Result Summary: In Tolerance Calibration Dus 25-Oct-2023 Manufacturer: Fluke Networks: Certificate Date: 25-Oct-2022 Model: Nauti 8010 Tomporature: 224°c ‘Serial Number: 18210478 Humidity: 39.2% Description: DSX-Verification Option(s) ILIFEXT ARTIFACT Procedure: Verification Artifact Characterization Revision: 2 ‘Customer: ALCARED SAC City: UMA Country: PE Purchase Order: CCS ALVARO JESUS CAYO UREUTA RMA: 32510468 ‘This cabratin is raceable tothe Irtematonal Sytem of Uns (SD, trough National Nekology insutes (NIST, PTB, NRC, NPL, etc), relemotis echnigucs, or naturel physical contents, This ceca applies only othe lem Mdenifed and sal ne De reproduced othe tan in ful without the specte wition approval by Fluke Corporation. Caran cerca witheut signa are not vai Tha calteaton has bean completed in accoresnce with Fuke Eleetroniss Corporation Quality Systems Document 111.0 Revision 124 andlor Fluke 17025 Quality Manual QD 11 1.41 Revision 007 ‘The Data Type fund inthis ceticata must be interpreted as: + As~ Found Calibration data collected before the units aciusted and Freire, + As Left Calbration data collected ater the unithas been adjusted and/or repited + Found:Left Callration data collacted without any asjusmaat and ot rea partrmed, ‘Measurement results greater than its of or are inca by ont: onaseus Ey yg - ore Som PERE Owome — dsonams FREE sm ware «8 BRE Wek Powers pecttec as rhs tore repo pe tiene recon 220 arenes RT — Sa — =e Page 1 of 3

También podría gustarte