Está en la página 1de 11

Dentomaxillofacial Radiology (2008) 37, 309–318

’ 2008 The British Institute of Radiology


http://dmfr.birjournals.org

RESEARCH
Image quality vs radiation dose of four cone beam computed
tomography scanners
M Loubele*,1,2, R Jacobs2,3, F Maes1, K Denis4, S White5, W Coudyzer6, I Lambrichts7,
D van Steenberghe2,3 and P Suetens1
1
ESAT-PSI, Katholieke Universiteit Leuven, Belgium; 2Perdiodontology, Katholieke Universiteit Leuven, Belgium; 3Oral Imaging
Center, Katholieke Universiteit Leuven, Belgium; 4EMAP, Xios Hogeschool Limburg, Belgium; 5School of Dentistry, University of
California at Los Angeles, USA; 6Radiology, UZ Leuven, Belgium; 7Morphology Section, Universiteit Hasselt, Belgium

Objectives: To evaluate image quality by examining segmentation accuracy and assess


radiation dose for cone beam CT (CBCT) scanners.
Methods: A skull phantom, scanned by a laser scanner, and a contrast phantom were used
to evaluate segmentation accuracy. The contrast phantom consisted of a polymethyl
methacrylate (PMMA) cylinder with cylindrical inserts of air, bone and PMMA. The
phantoms were scanned on the (1) Accuitomo 3D, (2) MercuRay, (3) NewTom 3G, (4) i-
CAT and (5) Sensation 16. The structures were segmented with an optimal threshold.
Thicknesses of the bone of the mandible and the diameter of the cylinders in the contrast
phantom were measured across lines at corresponding places in the CT image vs a ground
truth. The accuracy was in the 95th percentile of the difference between corresponding
measurements. The correlation between accuracy in skull and contrast phantom was
calculated. The radiation dose was assessed by DPI100,c (dose profile integral 100,c) at the
central hole of a CT dose index (CTDI) phantom.
Results: The results for the DPI100,c were 107 mGy mm for (1), 1569 mGy mm for (2),
446 mGy mm for (3), 249 mGy mm for (4) and 1090 mGy mm for (5). The segmentations in
the contrast phantom were submillimeter accurate in all scanners. The segmentation accuracy
of the mandible was 2.9 mm for (1), 4.2 mm for (2), 3.4 mm for (3), 1.0 mm for (4) and
1.2 mm for (5). The correlation between measurements in the contrast and skull phantom
was below 0.37 mm.
Conclusions: The best radiation dose vs image quality was found for the i-CAT.
Dentomaxillofacial Radiology (2008) 37, 309–318. doi: 10.1259/dmfr/16770531

Keywords: cone beam computed tomography; image quality; radiation dose

Introduction

During the last decade, there has been a trend towards radiation dose for the patient without significant loss of
using three-dimensional (3D) information to assist image quality are being investigated. Examples of these
dentomaxillofacial diagnostics and surgical planning.1 protocols are modified protocols on MSCT scanners4
This could first be realized by the use of conventional or the use of cone beam CT (CBCT).5,6
single and later multislice CT (MSCT).2 Nevertheless, Recently, an impressive number of CBCT scanners
because conventional CT protocols are generally have been introduced in the field of dentomaxillofacial
associated with relatively high radiation dose levels,3 radiology. Unfortunately, hardly any research-based
alternative CT protocols for bone visualization and evidence is available for either optimal image quality or
modelling that would allow the lowering of the effective radiation dose levels. Besides the development and
research of clinical evaluation protocols of these
scanners,7,8 it is essential to develop a technical test
*Correspondence to: Reinhilde Jacobs, Centrum voor Orale Beeldvorming,
Kapucijnvoer 7 blok a bus 7001, 3000 Leuven, Belgium; E-mail:
protocol aiding optimization of the individual scanner
reinhilde.jacobs@med.kleuven.be parameters according to the ALARA (as low as
Received 11 June 2007; revised 7 October 2007; accepted 12 October 2007 reasonably achievable) principle.9 Such optimization
Image quality and radiation dose of CBCT
310 M Loubele et al

procedures can only be approached by a combined


research strategy of radiation dose assessment for
various image quality levels. A technical method for
the evaluation of radiation dose in MSCT is the CT
dose index (CTDI).10 Because of the shortcomings of
these measurements,11 Mori et al12 suggested the use of
the dose profile integral (DPI). The advantage of such a
DPI is efficient data acquisition. The link to the
effective patient dose can then be established by
correlating effective dose levels with those technical
measures.13 For the evaluation of the image quality,
analysis on physical8,14 as well as anthropomorphic
phantoms8 can be used.
This paper is based on a combined research strategy
reported in two of our previous studies.15,16 In the first
study, a method was developed which made it possible
to evaluate the bone segmentation accuracy based on a
ground truth acquired with a laser scanner.15 In the Figure 1 The volume rendering of the laser model of mandible is
shown
second study, the physical image quality was compared
with the radiation dose for four different CBCT
scanners and one MSCT scanner.16 In the present
study, the evaluation of the image quality on an based on the CTDI values as listed by ImPACT, London,
anthropomorphic phantom and a physical phantom is UK (see ImPSCT website http://www.impactscan.
combined with the radiation dose for four CBCT org/ctditables.htm/). After appropriately adapting the
scanners and one MSCT scanner. listed CTDI values to the selected tube current settings of
each evaluated MSCT protocol and correcting for
collimation as described on the ImPACT website, the
DPI was calculated by multiplying the adapted CTDI
Materials and methods value with a length of 60 mm, which was sufficient to
acquire the complete mandible. In this way the radiation
Radiation dose assessment dose was truncated corresponding to the approach for
The radiation dose was measured by means of the DPI100,c, the dose measurements in CBCT.
Equation (1), measured at the central hole of a dedicated
CTDI-phantom. For the i-CAT (Imaging Sciences Image quality assessment
International, Hatfield, PA) and the NewTom 3G
For the evaluation of the image quality of CBCT and
(Quantitative Radiology, Verona, Italy) the DPI100,c was
MSCT, we applied the method previously reported for
measured with an electrometer and an ion chamber with an
observer measurements.17,18 We therefore measured the
active length of 10 cm, calibrated for measurements in CT
thickness of different structures in the CBCT images in
scanners (Radcal Corporation, Monrovia, CA). For the
an automated way. By comparing the measured
measurement of the DPI100,c on the Accuitomo 3D
thickness of the structures in the CT images with the
(Morita, Kyoto, Japan) and the MercuRay CB (Medico
real thickness of the structures, the accuracy of the
Technology Corporation, Kashiwa, Japan), there was no
ion chamber available on site and therefore the DPI100,c measures can be determined. Because it is important to
was measured with strips of thermoluminescent dose- have an upper boundary for the accuracy measure-
meters (TLD) of type TLD 100 (Li:Mg:Ti) (Bicron, Solon, ments, the 95th percentile of the difference between the
OH) placed 1.5 cm from each other. The TLDs for the thickness measurements on the CT image and the
Accuitomo 3D were read out on a fully automated ground truth image can be used. This procedure was
Harshaw 6600 reader (Bicron) in our hospital. The TLDs applied on both a skull phantom and a physical
for the MercuRay were read out with a fully automated phantom (contrast phantom) belonging to the image
Harshaw 8800 Card Reader Workstation (Bicron). The quality kit of the Accuitomo 3D. The skull phantom
DPI100,c was then calculated by approximation of the was used to mimic the diagnostic situation. The test on
integral in (1) using the trapezium rule. the contrast phantom was used to see whether
information about diagnostic quality could be learned
l
based on a physical phantom. In the sections below, the
ð2 complete procedure for the evaluation of the image
DPIl,j ~ d ðzÞdz ½mGy:mm ð1Þ quality will be further elaborated.
{12
Skull phantom
For the technical evaluation of the radiation dose on For the study, the dry skull from a person who donated
the MSCT scanner, the DPI100,c value was calculated his body to research and which was kindly provided by

Dentomaxillofacial Radiology
Image quality and radiation dose of CBCT
M Loubele et al 311

the Department of Morphology, University of Hasselt Registration of point cloud to CT data


(Belgium) was used for the construction of this To assess accuracy, bone thickness measurements on
phantom.15 The skull was cut into four parts: the CT are compared with those of the ground truth. To
mandible, the calvarium, the left zygoma and left part achieve a reliable assessment technique, there should be
of the maxilla, and the right zygoma and part of the absolute agreement regarding the position to be
maxilla. The four different sections were scanned with measured on CT and the laser model. An accurate
an XC50 Cross Scanner (Metris, Leuven, Belgium) with registration of the point cloud to the CT data was
three laser planes, mounted on a Wenzel LH57 3D therefore needed. For this registration, the MIRIT
coordinate measurement machine. With this laser Software19 was used. This software calculates a six-
scanner it is possible to obtain surfaces in the form of parameter rigid transformation T (i.e. a combination of
a point cloud (Figure 1) with an accuracy of 15 mm. a 3D translation and a 3D rotation), mapping every
Because the skull was cut into four sections, it was location in a floating image, A, to a corresponding
possible to acquire the outer and inner bone surface in location in the reference image B by maximizing the
one acquisition, allowing the measurement of bone statistical dependence between intensity values of
thickness along the point cloud. After acquiring the corresponding voxels in both images. A point cloud
laser scan, the skull was placed in a head mould exists of a set of 3D coordinates which represent a
enclosure in the form of a face filled with water for soft- surface and not a volume. This means that a point
tissue simulation. For further analysis of the image cloud does not contain voxels. Therefore, a function
quality, the segmentation accuracy of the mandible will was needed, indicating when a specific point was
be evaluated. located inside or on the bone surface and when it
was located outside the bone surface. This function was
calculated based on the FastRBF Interpolation
Toolbox (FarField Technology Limited, Christchurch,
New Zealand), which computes an implicit function
through the point cloud of the laser data in such a way
that this function evaluates to zero on the surface,
negative inside the surface and positive outside. The
function was evaluated on a cubical grid in which the
complete mandible was situated. This grid had a voxel
size of 0.1 mm. As a result, the discretization of the grid
did not influence the registration accuracy.20

Definition of quality measure


a b For the evaluation of the CT image quality, a measure
needs to be defined. In this paper, the error on the bone
thickness measured on the bone model will be used as the
quality measure, as was done in a previous study.15,16 To
evaluate this error, measurement lines are defined along
the bone surface in the laser model. These measurement
lines are transferred to the CT image based on the
transformation calculated in the previous section. For the
definition of the measurement lines, a cylindrical grid
with the vertical axis through the centre of mass of the
object as z-axis was calculated. This is the central axis of a
cylindrical grid defined by a vertical increment Dz and an
angular increment Dh. In all elements (iDz, jDu) of the
cylindrical grid where bone exists, a measurement line is
defined by two points on the bone surface along a radial
through this element (one point nearest to the z-axis and
one point farthest to the z-axis). With this method, 3589
measurement lines were found. Across these measure-
(mm) ment lines, 1D intensity profiles are calculated by a 3D
c trilinear interpolation of the image intensity at equidi-
Figure 2 Linear measurements are used for image quality assess- stant points between the beginning and the end point of
ment. Measurement lines are defined on (a) the computer model of the each line. The sample distance was 0.1 mm. The
phantom and transferred to (b) the CT image. After proper intersections of each measurement line with the bone
registration between them, (c) the one-dimensional CT intensity
profile along each line is extracted by interpolation and segmented by surface are extracted by thresholding of its 1D intensity
thresholding in order to measure the thickness of the structure of profile by using a global bone threshold. Linear
interest interpolation of the profile values is used to locate

Dentomaxillofacial Radiology
Image quality and radiation dose of CBCT
312 M Loubele et al

candidate intersection points at sub-voxel precision. If cylinders with a diameter of 10 mm consisting of


more than two candidates are found, which is typically aluminum, PMMA, bone equivalent plastic and air
the case when the bone consists of two cortical plates (Figure 2a,b)). Based on the phantom, image-based
surrounding an inner spongiosa, the locations that are measurements of the diameters of the cylinders at
retained are those closest to the reference bone surface various positions were also performed in a similar way
derived from the laser data. to those for the skull phantom. Because the different
parts of this phantom exist as homogeneous materials,
Calculation of threshold value the threshold value was calculated as the average of the
If a human observer measures bone thickness on a CT mean intensity of the bone/aluminium and PMMA.
image, they delineate the bone in their mind by seeking the The evaluation procedure is briefly depicted in
transition between soft-tissue and bone. Such delineation Figure 2. For a more elaborate explanation, refer to
is performed by the computer using a segmentation the previous study.16
algorithm. The easiest segmentation algorithm to use is
the use of a global threshold value. This means that a Experiments
single threshold value is used to segment the whole object Dose measurements and image quality assessment were
everywhere in the image. Some pilot tests were performed performed in one MSCT and four different CBCT
at the start of the study, and the results of this test showed scanners. An overview of the protocols is given in
that the rule applied by Wiemker and Zwartkruis21 Table 1. The MSCT scanner was the Somatom
performed best, therefore this rule was used. They showed Sensation 16 (Siemens, Erlangen, Germany). The
that the transition from a region to segment and their CBCT scanners were the NewTom 3G, the i-CAT,
background corresponds to a local optimum in the the MercuRay and the Accuitomo 3D. When we
cumulative Laplace-weighted histogram. In some cases initiated this study, the different CBCT scanners were
an optimum of the surface, the mean gradient, the volume not available at many sites and hence measurements
or the sphericity histogram can give more information to needed to be performed at different institutions. The
find the ideal threshold value. For the calculation of the measurements with the Somatom Sensation 16 and the
histograms, a cuboid region was indicated on the 3D Accuitomo 3D were performed at the University
images of the CBCT or MSCT images. The materials Hospital Leuven (Leuven, Belgium), the measurements
present in this cuboid region were air, water and bone. To with the NewTom 3G were performed at the UCLA
find the maximum which corresponds to the transition School of Dentistry (Los Angeles, CA), the measure-
between bone and soft tissue on the cumulative histogram ments with the MercuRay CB at were performed at
of the Laplacian, first the intensity values which SmartScan Imaging (Orange, CA) and the measure-
correspond to soft tissue and bone needed to be selected. ments with the i-CAT were performed at Golden State
This intensity region was found by inspection of the X-ray Lab (North Hollywood, CA) and Imaging
histogram of the image intensities. Sciences International (Hatfield, PA). For all CBCT
scanners except for the MercuRay, the protocols suited
Physical phantom for the planning of oral implants were evaluated. Due
For the evaluation of the image quality based on the to practical reasons, we could not evaluate the implant
physical phantom, the results acquired for the segmen- protocol on the MercuRay and therefore the protocol
tation of the contrast phantom in a previous study were with the largest diameter was evaluated. A comparison
used.16 The contrast phantom of the Accuitomo 3D between the radiation dose of the MercuRay and the
was used for the evaluation of the physical per- other CBCT scanners would therefore not be fair. It is,
formance. This phantom is made of polymethyl however, possible to estimate the radiation dose of an
methacrylate (PMMA) with an insert of four smaller MSCT scanner with a similar height as the MercuRay.

Table 1 Overview of the parameters of the evaluated protocols


Tube
Tube potential current Diameter Height Rotation Exposure
Scanner Protocol (kVp) (mAs) (mm) (mm) time (s) time (s) Voxel size (mm) Kernel
NewTom 3G N1 110 22 100 100 36 7.2 0.1860.1861 Standard
NewTom 3G N2 110 22 100 100 36 7.2 0.1860.1860.4 High
NewTom 3G N3 110 22 100 100 36 7.2 0.1860.1860.4 Very high
i-CAT I1 120 11 160 60 10 1.92 0.360.360.3 N/A
i-CAT I2 120 11 160 60 10 1.92 0.460.460.4 N/A
i-CAT I3 120 20 160 60 20 3.67 0.460.460.4 N/A
i-CAT I4 120 40 160 60 40 7.19 0.260.2 0.2 N/A
MercuRay M1 120 150 193 193 10 10 0.3860.3860.38 N/A
Accuitomo 3D A1 80 72 40 30 18 18 0.1360.1360.5 N/A
Sensation 16 S1 120 90 90 60 – – 0.2560.2560.4 H60s
Sensation 16 S2 120 90 90 193 – – 0.2560.2560.4 H60s
N/A, not applicable

Dentomaxillofacial Radiology
Image quality and radiation dose of CBCT
M Loubele et al 313

Therefore a comparison will be made between the The results of the analysis of the segmentation
MercuRay and the Somatom Sensation 16. accuracy can be found in Table 2. The highest accuracy
for the segmentation of the mandible was found for the
Data and statistical analysis i-CAT (protocols I2 and I3), this accuracy was also
For the evaluated protocols on the different scanners, higher than for the Somatom Sensation 16. The lowest
descriptive statistics were used to express the mean and accuracy was found for the MercuRay. The high
standard deviation between the ground truth values accuracy for the i-CAT could be achieved because this
measured on the laser model and the bone thickness image suffered less from the intensity inhomogeneity
measured on the CT images for all measurement lines. and it was therefore possible to calculate a threshold
Positive values indicated an underestimation of the true value based on the complete mandible. The MercuRay
suffered most from the intensity inhomogeneity and
thickness; negative values indicate an overestimation.
there was therefore a lower accuracy for the bone
Also the 95th percentile of the absolute difference
segmentation.
between the ground truth and the measured thickness
was calculated. A similar analysis was performed in the For the segmentation of the cylinder of bone and
aluminium in the contrast phantom, for all protocols an
previous study for the images of the contrast phan-
accuracy better than 1 mm was achieved. The Pearson
tom.16 Finally, the Pearson correlation coefficient
correlation coefficient between the accuracy achieved
between the accuracy achieved on the mandible and
on the mandible and the bone equivalent plastic was
the segmentation of the cylinder in bone and aluminum
0.37 and the correlation coefficient between the
were calculated.
accuracy achieved on the mandible and the aluminium
cylinder was 0.03. This means that based on the
analysis of the physical phantom, no conclusion can
Results be drawn about the segmentation of the mandible.
For the protocols used for oral implant placement,
The setup for the experiment did not fit in the the DPI100,c was the highest on the Somatom Sensation
Accuitomo 3D and therefore a smaller box was used 16. The highest radiation dose for the CBCT scanners
for the experiment on the Accuitomo 3D. Typical was for the NewTom 3G and the lowest for the
histograms, i.e. the histogram of the image intensities Accuitomo 3D. The i-CAT had the highest accuracy
and the cumulative histogram of the Laplacian of the combined with a low radiation dose and had as a
image, are shown in Figures 3–6 for the different CBCT surplus a submillimeter accuracy for the segmentation
scanners. For each image, the region of interest (ROI) is of the mandible. We can therefore say that this scanner
indicated with highlighted intensities. Due to the performed the best on this test.
intensity of the grey values in the NewTom 3G and When the radiation doses of the MercuRay and the
the MercuRay, several ROIs needed to be tested before Somatom Sensation 16 are compared, one can see that
a cumulative histogram of the Laplacian was found the radiation dose levels are similar. Because the
which was suitable for analysis. Such a histogram could radiation dose of the Somatom Sensation 16 is higher
be achieved when only a small region was taken into than the CBCT scanners for placement of oral
account. In the image histograms the intensities which implants, one may therefore expect that if a protocol
were taken into account to find the bone threshold are with a lower height is used on the MercuRay, we will
indicated in a thicker line. also have a radiation dose for the MercuRay that is

a b c
Figure 3 (a) A slice of the NewTom (protocol N2) together with a highlighted rectangle representing the region of interest. (b) Represents the
histogram of the image intensities and (c) represents the cumulative Laplacian histogram. The image intensities which are taken into account are
shown in a thicker line

Dentomaxillofacial Radiology
Image quality and radiation dose of CBCT
314 M Loubele et al

a b c
Figure 4 (a) A slice of the Accuitomo 3D. (b) Represents the histogram of the image intensities and (c) represents the cumulative Laplacian
histogram. The image intensities that are taken into account are shown in a thicker line

similar to Somatom Sensation 16 and thus higher than the X-ray beam at two opposite positions. Because of
the other CBCT scanners the larger distance between the source of the X-ray
beam and the object, and considering the limited
collimation of the X-ray beam, the irradiated area will
Discussion only be small; therefore, one can expect much of the
scattered radiation to be covered when an ion chamber
In this paper, the radiation dose and image quality of with a length of over 10 cm is used. In Figure 7b, a
different CBCT scanners was compared with the similar composition is made, but this time for the i-
radiation dose and image quality of an MSCT scanner. CAT scanner. The figure has been made for a protocol
The radiation dose was evaluated by a technical dose with a height of 13 cm. Because the scanned height is
measurement, the DPI100,c measured in the central hole even longer than the 10 cm of the ion chamber, it is
of a CTDI phantom. Recently, a lot of discussion for obvious that not all scattered radiation is covered by
and against the use of the CTDI or another technical the ion chamber. If we divided the DPI by the slice
dose measurement has been performed.11,12,22,23 We will thickness, for achieving the CTDI, we could obtain a
explain two of the shortcomings of the use of the CTDI. completely wrong dose descriptor. In Figure 7c, we give
The main difference between the CTDI and the DPI is the configuration for the Accuitomo 3D. For this
that the CTDI is the DPI divided by the slice thickness. scanner, a similar reflection can be made as for the
One of the most important shortcomings of the i-CAT, except that the scanned height here is smaller
CTDI is that it is measured over a length of 10 cm, than 10 cm. The use of longer phantoms as proposed
which is too short to include all the scattered radiation by Mori et al12 is not feasible in CBCT scanners for
dose in the protocol of a CBCT scanner. We show this dentomaxillofacial applications.11
in Figure 7. Figure 7a represents the configuration of Another problem for the CTDI measurement is the
the Somatom Sensation 16. The contour of the CTDI way that different CTDI measurements are combined.11
phantom is presented on an axial slice, together with For the CTDI measurements in MSCT scanners, the

a b c
Figure 5 (a) A slice of the MercuRay together with a highlighted rectangle representing the region of interest. (b) Represents the histogram of
the image intensities and (c) represents the cumulative Laplacian histogram. The image intensities which are taken into account are shown in a
thicker line

Dentomaxillofacial Radiology
Image quality and radiation dose of CBCT
M Loubele et al 315

a b c
Figure 6 (a) A slice of the i-CAT (protocol N2) together with a highlighted rectangle representing the region of interest. (b) Represents the
histogram of the image intensities and (c) represents the cumulative Laplacian histogram. The image intensities which are taken into account are
shown in a thicker line

weighted sum of the CTDI at the central position and the measurement of the DPI100 should be given because
the average of four peripheral measurements with this measurement can be obtained with the same
weights of one-third and two-thirds, respectively, is measurement procedure. The aim of future research in
used as global dose estimation. For the dose measure- technical dosimetry will consist of striving for a
ment in this study, only a measurement at the central correlation between the DPI or a new technical
hole was performed and not at the peripheral positions recording method and the effective radiation dose that
because there currently exists no similar rule to combine a patient receives. Also, correction factors will be
these into one dose estimate. This is intuitively calculated and tabulated to compensate for scattered
explained in Figure 7. For the Somatom Sensation 16 radiation, which was not measured by the ion chamber
(Figure 7a), the cross-section of the X-ray beam with of 10 cm, and similar correction factors will be
the phantom is almost rectangular. For the i-CAT tabulated to compensate for the geometry of the
(Figure 7b) and the Accuitomo 3D (Figure 7c), this scanner.
cross-section has a conical shape, which depends on the We assessed the segmentation accuracy of bone
geometry of the scanner and the protocol which was models segmented from CBCT and MSCT images by
used. It is obvious that the contribution of the comparing them with a model acquired from a laser
dosimetry at the central and peripheral holes in the scanner. Our method relies on a correct geometric
phantom does not make the same contribution to the alignment or registration of the CT images with a
radiation dose as for the traditional formula of the volumetric model with a voxel size of 0.1 mm acquired
CTDI. from the laser scanner. The voxel size of this model was
Unfortunately, when we performed a study on the defined as smaller than the voxel sizes of the CT images
technical dose that was reported in the manuals of the and the registration accuracy was therefore not
evaluated CBCT scanners, we noticed that they were determined by the laser model but by the CT images.20
aware of the problem with the CTDI and had therefore The availability of an accurate ground truth made it
adapted the definition of the CTDI based on their own possible to evaluate some rules for the finding of a
opinion. However, we propose that besides the CTDI threshold value for bone. It was found that, certainly

Table 2 Results of the analysis of the image quality and the radiation dose
Skull phantom Contrast phantom
Mandible Bone equivalent plastic Aluminium CTDI phantom
Protocol Prct 95 Mean ¡ Prct 95 Mean ¡ SD Prct 95 Mean ¡ SD DPI100,c
N1 3.5 20.7¡1.5 0.73 20.06¡0.17 0.37 20.06¡0.17 446
N2 3.5 20.72¡1.5 0.59 20.02¡0.13 0.37 20.02¡0.13 446
N3 3.4 20.7¡1.5 0.42 20.07¡0.16 0.4 20.07¡0.16 446
I1 1.4 0.03¡0.7 0.34 0.03¡0.2 0.26 0.05¡0.10 71
I2 1.0 20.03¡0.6 0.6 20.1¡0.19 1.03 0.08¡0.32 71
I3 1.0 20.08¡0.5 0.45 20.04¡0.24 1.07 0.12¡0.29 124
I4 1.1 20.01¡0.6 0.25 20.1¡0.08 0.25 20.19¡0.06 249
M1 4.2 20.1¡2.0 0.42 20.04¡0.38 1.01 20.04¡0.38 1569
A1 2.9 0.5¡1.1 0.3 20.17¡0.09 0.3 20.48¡0.57 107
S1 1.2 0.14¡0.6 0.24 20.07¡0.06 0.22 20.07¡0.06 1090
S2 1.2 0.14¡0.6 0.24 20.07¡0.06 0.22 20.07¡0.06 1677
CTDI, CT dose index; Prct, percentile, SD, standard deviation

Dentomaxillofacial Radiology
Image quality and radiation dose of CBCT
316 M Loubele et al

for the i-CAT scanner and the MSCT scanner, the use contrast phantom is a small phantom, only 5 cm in
of the rule indicated by Wiemker et al21 gave satisfying diameter and 5 cm in height. This means that except for
results. Unfortunately, this rule gave no accurate results the Accuitomo 3D the phantom is positioned comple-
when the images suffered more from intensity inhomo- tely in the field of view and therefore does not suffer
geneity. from the truncated view artefact. Secondly, because the
In a second experiment, the image quality was object is so small, there is less scattered radiation which
evaluated based on a physical phantom. Only a small can influence the image quality. The influence of the
correlation between the results of the segmentation scanner depends on the geometry of the scanner and the
accuracy on the mandible and the segmentation of the spectrum of the X-ray tube.24 How the scatter finally
cylinders in the contrast phantom was found. Several will influence the image quality can be determined by
reasons can be found for this small correlation. The the reconstruction algorithm which is developed by the

c
Figure 7 The configuration of the X-ray beam is presented for three different scanners: (a) Somatom Sensation 16, (b) i-CAT and (c) Accuitomo
3D. Each figure shows two positions of the X-ray beam in the opposite position, together with the contours of a CT dose index (CTDI) phantom.
The dashed-dotted line together with the dotted lines indicates the central and the two peripheral holes of the CTDI phantom

Dentomaxillofacial Radiology
Image quality and radiation dose of CBCT
M Loubele et al 317

CBCT manufacturer.24 Because these different factors larger region was scanned, so this also should be taken
are dependent on each scanner, one cannot predict the into account. The lowest radiation dose was achieved
image quality of bone segmentation based on the for the Accuitomo 3D, but here the smallest region was
segmentation of bone equivalent plastic in the contrast scanned. If, for example, a complete mandible needs to
phantom. This does not mean that physical phantoms be imaged, this would require at least three scans. As a
are useless for the evaluation of image quality of CBCT result, the Accuitomo 3D would require the highest
scanners. These phantoms are very useful for contin- radiation dose. If with the i-CAT a region with a similar
uous quality control of CBCT scanners. A difference, height would be scanned with the parameters of
for example, in contrast and therefore a deterioration of protocol I4, a similar dose to the NewTom 3G would
segmentation accuracy over time can indicate that there be achieved. One can therefore state that the radiation
might be a problem with the CBCT scanner. Also, dose of the i-CAT and the NewTom 3G are similar.
different parameters such as, for example, the However, the i-CAT has better segmentation accuracy
Hounsfield units without influence of scattered radia- for a similar radiation dose and therefore the i-CAT
tion dose can be calculated. performed best in this study.
The benefits of this work over the work of previous
studies17,18 in which the accuracy of linear measure-
ments of CBCT was derived is that the measurements
Conclusion
are performed automatically without much user inter-
action. Therefore, this method does not suffer from
observer and intraobserver variability. In the current A framework for the evaluation of image quality and
study, water was used to simulate soft tissue. Water for radiation dose was presented. The image quality was
the use of soft tissues is not very practical and a solid evaluated by measuring the segmentation accuracy of
water simulating or soft tissue simulating material the mandible in a skull phantom and the segmentation
should therefore be used for the generation of a skull of cylinders in a contrast phantom. The results of the
phantom. However, the use of solid material instead of image quality assessment were accumulated in the 95th
water belongs to a future study. Also in future studies, percentile of the absolute difference between the
different parts of the skull phantom should be measured thickness on the CT scanners and the ground
incorporated for the evaluation of the image quality. truth thickness. The radiation dose was evaluated by a
Besides this, a better definition of the measurement technical measure, the dose profile integral. The highest
lines will also be researched. radiation dose was found for the Somatom Sensation
The Somatom Sensation 16, the MSCT scanner, had 16 and the MercuRay. The lowest was found for the
the highest radiation dose together with the MercuRay. Accuitomo 3D, which also covered the smallest image
Similar results for the MercuRay were found by area. The best segmentation accuracy was found for the
Ludlow et al.25 The NewTom 3G had a higher i-CAT. No correspondence was found between accu-
radiation dose than the other CBCT scanners, but a racy in contrast phantom and skull phantom.

References

1. Jacobs R and van Steenberghe D. Radiographic planning and 10. Dixon RL. A new look of CT dose measurement: beyond CTDI.
assessment of endosseous oral implants. Berlin: Springer Verlag, 1998. Med Phys 2003; 30: 1272–1280.
2. Gahleitner A, Watzek G, Imhof H. Dental CT: imaging 11. Dixon RL. Restructuring CT dosimetry – a realistic strategy for
technique, anatomy, and pathologic conditions of the jaws. Eur the future Requiem for the pencil chamber. Med Phys 2006; 33:
Radiol 2003; 13: 366–376. 3973–3976.
3. Cohnen M, Kemper J, Mobes O, Pazelzik J, Modder U. 12. Mori S, Endo M, Nishizawa K, Tsunoo T, Aoyama T, Fujiwara
Radiation dose in dental radiology. Eur Radiol 2002; 12: 634–637. H, Murase K. Enlarged longitudinal dose profiles in cone-beam
4. Brem MH, Zamani AA, Riva R, Zou KH, Rumboldt Z, Henning CT and the need for modified dosimetry. Med Phys 2005; 32:
FF, et al. Multidetector CT of the paranasal sinus, potential for 1061–1069.
radiation dose reduction. Radiology 2007; 243: 847–852. 13. Leitz W, Axelsson B, Szendrö G. Computed tomography dose
5. Guerrero ME, Jacobs R, Loubele M, Schutyser F, Suetens P, van assessment – a practical approach. Radiat Prot Dosim 1995; 57:
Steenberghe D. State-of-the-art on cone beam CT imaging for 377–380.
preoperative planning of implant placement. Clin Oral Investig 14. Marmulla R, Wörtche R, Mühling J, Hassfeld S. Geometric
2006; 10: 1–7. accuracy of the NewTom 9000 cone beam CT. Dentomaxillofac
6. Scarfe WC, Farman AG, Sukovic P. Clinical applications of Radiol 2005; 34: 28–31.
cone-beam computed tomography in dental practice. J Can Dent 15. Loubele M, Maes F, Vandermeulen D, Denis K, Jacobs R, White
Assoc 2006; 72: 75–80. SC, et al. Assessment of bone segmentation quality of CT
7. Hintze, M Wiese, A Wenzel. Cone beam CT and conventional scanners using laser scanning, Int J Comp Assist Radiol Surg
tomography for the detection of morphological temporomandi- 2006; 1(Suppl): 400–402.
bular joint changes. Dentomaxillofac Radiol 2007; 36: 192–197. 16. Loubele M, Bogaerts R, White SC, Maes F, Bosmans H,
8. Pinsky HM, Dyda S, Pinsky RW, Misch KA, Sarment DP. Sanderink G, et al. Comparative study of image quality and
Accuracy of three-dimensional measurements using cone-beam radiation dose of MSCT and CBCT scanners in dentomaxillo-
CT. Dentomaxillofac Radiol 2006; 35: 410–416. facial radiology. Internal report KUL/ESAT/PSI/0703. KU
9. International Commission on Radiation Protection. 2007 Leuven, ESAT, Leuven, Belgium: February 2007.
Recommendations of the International Commission on Radiological 17. Kobayashi K, Shimoda S, Nakagawa Y, Yamamomto A.
Protection. Annals of the ICRP. Oxford: Elsevier, 2007. Accuracy in measurements of distance using limited cone-beam

Dentomaxillofacial Radiology
Image quality and radiation dose of CBCT
318 M Loubele et al

computerized tomography. Int J Oral Maxillofac Implants 2004; Proceedings of CARS 2001; Berlin, Germany, 2001, pp 611–
19: 228–231. 616.
18. Lascala CA, Panella J, Marques MM. Analysis of the accuracy of 22. Brenner DJ. It is time to retire the computed tomography dose
linear measurements obtained by cone beam computed tomo- index (CTDI) for CT quality and dose optimization. For the
graphy (CBCT-NewTom). Dentomaxillofac Radiol 2004; 33: proposition. Med Phys 2006; 33: 1189–1190.
291–294. 23. McCollough CH. It is time to retire the computed tomography
19. Maes F, Collignon A, Vandermeulen D, Marchal G, Suetens P. dose index (CTDI) for CT quality assurance and dose optimiza-
Multi-modality image registration by maximization of mutual tion. Against the proposition Med Phys 2006; 33: 1190–1191.
information. IEEE TMI 1997; 16: 187–198. 24. Carlsson CA. Imaging modalities in X-ray computerized tomo-
20. West J, Fitzpatrick JM, Wang MY, Dawant BM, Maurer CR Jr, graphy and in selected volume tomography. Phys Med Biol 1999;
Kessler RM, et al. Comparison and evaluation of retrospective 44: R23–R56.
intermodality brain image registration techniques. J Comput 25. Ludlow JB, Davies-LudLow JE, Brooks SL, Howerton WB.
Assist Tomogr 1997; 21: 554–566. Dosimetry of 3 CBCT devices for oral and maxillofacial
21. Wiemker R, Zwartkruis A. Optimal thresholding for 3D radiology: CB MercuRay, NewTom 3G and i-CAT.
segmentation of pulmonary nodules in high resolution CT. Dentomaxillofac Radiol 2006; 35: 219–226.

Dentomaxillofacial Radiology

También podría gustarte