Documentos de Académico
Documentos de Profesional
Documentos de Cultura
Table of Contents
API 2000 TurboIonSpray Ion Source Manual. . . . . . . . . . . . . . . . . . . . . . . . . . 1
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
TurboIonSpray Source Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Optimizing the TurboIonSpray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Changing The Peek Tubing Transfer Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Changing The Stainless Steel Sprayer Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Changing The Grounded Union/Splitter Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Appendix A - Consumable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Appendix B - Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
i
Table of Contents
ii
API 2000 TurboIonSpray Ion Source Manual
Conventions
Within this manual, the following conventions are used:
Introduction
TurboIonSpray is an Atmospheric Pressure Ion Source in which pre-formed ions in
solution are emitted into the gas phase with or without the application of heat. In this way,
quasi-molecular ions can be generated from very labile and high molecular weight
compounds with no thermal degradation.
TurboIonSpray is the technique by which a liquid sample is pumped by a liquid
chromatograph (LC) pump or syringe drive through a sprayer probe, which is maintained
at a high voltage, and is nebulized into the Ion Source creating a mist of highly charged
droplets. Once inside the Ion Source, the droplets evaporate causing the ions to enter the
gas phase by a low energy process called Ion Evaporation.
The use of an orthogonal heated gas extends the rugged and versatile technique of
TurboIonSpray to accept higher flow rates with improved sensitivity. TurboIonSpray will
accept flows from 5 to 1000 µL/min of solvent compositions from 100% aqueous to 100%
organic, such as acetonitrile, without splitting. This allows the use of 1 mm, 2 mm, and 4.6
mm analytical columns with or without splitting.
A heater probe directs a jet of heated dry gas (up to a maximum of 500°C) at the mist
produced by the sprayer. The gas is sprayed across the orifice at an angle of approximately
45° with respect to the Curtain Plate. The liquid spray emerging from the TurboIonSpray
is directed at an angle of about 45° from the opposite direction (or 135°). The
TurboIonSpray effluent and the heated dry gas intersect at an angle of approximately 90°
near the orifice. This interaction helps focus the TurboIonSpray stream and increases the
rate of droplet evaporation resulting in an increased ion signal.
1
API 2000 TurboIonSpray Ion Source Manual
TurboIonSpray Setup
TurboIonSpray is ideally suited for LC/MS/MS quantitative analyses. The sensitivity
increases that are achieved with this technique are both flow rate and analyte dependent.
In the conventional IonSpray source sensitivity decreases with increased flow rate, while
the heated TurboIonSpray process increases ionization efficiency, especially at the higher
flow rates; this results in improved sensitivity. Sensitivity is compound dependent and
compounds of extremely high polarity and low surface activity usually show the greatest
sensitivity increases. The TurboIonSpray technique is mild enough to be used with labile
compounds such as peptides, proteins, and thermally labile pharmaceuticals.
NOTE: The difference between conventional IonSpray and TurboIonSpray is that in the
TurboIonSpray, the ionization of the sample is achieved with additional heater gas
delivered to aid in the desolvation of the spray. Typically the TurboIonSpray is used at
higher flow rates (greater than or equal to 40 µL/min).
Features
The following list outlines the features of the TurboIonSpray ion source:
• Able to function as conventional IonSpray source when the Heater Gas is turned off.
• Able to function with flow rates from 1 µ L/min to 1000 µL/min.
• Able to vaporize 100% aqueous to 100% organic solvents.
Specifications
Temperature Range
• Probe temperature from 0° to 500°C
Liquid Chromatograph
• Interfaces to any liquid chromatograph system
2
API 2000 TurboIonSpray Ion Source Manual
Sprayer Probe
The TurboIonSpray source contains a fixed angle adjustable sprayer probe. The sprayer
probe is adjustable in two directions: towards the orifice from a scale of 0 to 12 (adjusted
by the horizontal adjusting ring mounted on the TurboIonSpray probe), and up and down
towards the orifice from 0 to 12 by the Lateral Adjustment control on the top of the Ion
Source.
3
API 2000 TurboIonSpray Ion Source Manual
Sprayer Probe
Peek Tubing
Grounded Union/Splitter
4
API 2000 TurboIonSpray Ion Source Manual
WARNING! The use of the Peek tubing transfer line connected to the
grounded fitting is mandatory since it prevents any exposed peripherals
connected to the source from floating at high voltage.
WARNING! Peek tubing (or fused silica) with Peek fittings must be used as
the transfer line. Metal tubing or fittings must not be used. The use of metal
may result in the exposure of high voltage to the user.
Refer to the procedure for changing the grounded fitting from a union connection to a
splitter described later in this document.
Heater Probe
The heater probe temperature is set in the applications computer software by modifying
the TEM parameter. The correlation between the parameter value of TEM and the actual
heater temperature is direct (for example TEM 350 is equal to 350°C at the heater). The
heater gas is also set in the applications computer software. The connections for the heater
and gas are made automatically once the source is locked to the interface.
The heater probe temperature is maintained by the Temperature Control Board (TCB)
mounted behind the IonSource panel (below the IonSource housing). The TCB adjusts the
flow of power to the heater element as a function of the difference between the actual
heater temperature and the temperature setting in the applications computer. The probe
temperature is monitored by a thermocouple connected directly to the heater element
which maintains the temperature within ±5 degrees of the applications computer setting.
The operating range for the probe is from approximately 100° to 500°C.
5
API 2000 TurboIonSpray Ion Source Manual
Exhaust Pump
The TurboIonSpray Source requires that the Source Exhaust system is properly connected
and functioning (refer to the figure Exhaust Supply Connection Points under Source
Exhaust Pump in the Optimizing the TurboIonSpray section). A filtered nitrogen or air gas
supply (free from pump oil) is delivered to the Source Exhaust pump at 60 psi pressure at
a flow of at least 4 L per minute. The Source Exhaust pump is used to vent solvent vapors
which develop in the Ion Source plenum.
It is recommended that these vapors be passed through a trap vented to a fume hood or
outside port.
WARNING! Failure to provide proper ventilation of the ion source can result
in hazardous vapors being released into the laboratory environment.
Installation
The TurboIonSpray source connects to the Vacuum Interface housing. Two latches
mounted on the source housing secure the TurboIonSpray against the vacuum interface.
Guide Pins
6
API 2000 TurboIonSpray Ion Source Manual
6. Connect the liquid sample tubing to the grounded fitting which is connected to the
inlet end of the sprayer probe. Ensure that all fittings are properly seated in order to
minimize dead volumes.
NOTE: In order to reduce the band broadening of sample with solvent, thus maximizing
sensitivity, the use of 0.005” ID Peek, or fused silica tubing between the TurboIonSpray
source and the injector is recommended.
WARNING! Ion source may be hot for several minutes after it is removed
from the instrument.
1. Stop all scans and place the instrument in Standby or Overnight Quit status.
2. Turn off the liquid flow entering the ion source.
3. Disconnect the liquid sample tubing from the grounded fitting.
4. Turn the two source latches to the unlocked (up) position.
The sprayer high voltage and heater connectors are automatically disengaged once the
source is unlocked.
5. Slide the source along the guide pins away from the interface to remove it.
7
API 2000 TurboIonSpray Ion Source Manual
CAUTION! When optimizing the Sprayer position make certain not to spray
directly down the orifice. Spraying down the orifice may contaminate the
Vacuum Interface and Vacuum Chamber ion optics and could impact
instrument performance.
ORIFICE
40 µL/min
200 µL/min
1mL/min
8
API 2000 TurboIonSpray Ion Source Manual
The best position is usually a few millimeters off axis to the bottom of the curtain plate
aperture. Multiply charged proteins and peptides introduced at a few microliters per
minute usually require the sprayer to be as close as possible to the Curtain Plate.
NOTE: If the TurboIonSpray voltage is set too high a blue glow can be seen at the tip of
the TurboIonSpray indicating a corona discharge. This will result in decreased sensitivity
and stability of the ion signal.
9
API 2000 TurboIonSpray Ion Source Manual
The table below provides recommended operating conditions for the TurboIonSpray at
three different flow rates. For each flow rate, the Curtain Gas flow rate (from setting 11 to
45 at the applications computer) should be as high as possible. The solvent composition
used for optimization was 1/1 Water/Acetonitrile. These conditions represent a starting
point from which to optimize TurboIonSpray. By an iterative process, the various settings
can be optimized using Flow Injection Analysis to obtain maximum signal-to-noise for the
compound of interest.
10
API 2000 TurboIonSpray Ion Source Manual
In general terms, the higher the Declustering Potential and Focusing Potential voltages,
the greater the energy imparted to the ions entering the analyzing region of the mass
spectrometer. The energy helps to decluster the ions and to reduce the chemical noise in
the spectrum resulting in an increase in sensitivity. Increasing the voltages beyond optimal
conditions can induce fragmentation before the ions enter the mass filters resulting in a
decrease in sensitivity. In some instances this fragmentation can prove a valuable tool
providing additional structural information.
Solvent Composition
Commonly used solvents and modifiers are acetonitrile, methanol, propanol, water, acetic
acid, formic acid, ammonium formate and ammonium acetate. The modifiers such as
TEA, sodium phosphate, TFA and dodecyl sodium sulfate are not commonly used because
they complicate the spectrum with their ion mixtures and cluster combinations. They may
also suppress the strength of the target compound ion signal. The standard concentration
of ammonium formate or ammonium acetate is from 2 to 10 millimole per liter for positive
ions and 2 to 50 millimole per liter for negative ions. The concentration of the organic
acids is 0.01% to 0.5% by volume.
11
API 2000 TurboIonSpray Ion Source Manual
Operating Tips
1. When running the TurboIonSpray, with or without the Heater Probe on, run the
Curtain Gas at as high a flow rate as possible. The Curtain Gas should optimize about
30% higher when TurboIonSpray is in use. This will help to:
a) prevent penetration of the Curtain Gas which can produce a noisy signal,
b) prevent contamination of the orifice, and
c) increase the overall signal to noise ratio
2. The liquid spray from the TurboIonSpray should be directed away from the orifice in
order to:
a) prevent contamination of the orifice,
b) prevent piercing of the Curtain Gas which can create an unstable signal, and
c) prevent electrical shorting due to the presence of the liquid.
3. The higher the liquid flow or the higher the aqueous composition of the solvent, the
higher the temperature and Heater Gas (Gas 2) flow required. However, too high a
temperature can cause thermal degradation of the compound, and result in a high
chemical background noise, while too high a heater gas flow can produce a noisy, or
unstable signal.
12
API 2000 TurboIonSpray Ion Source Manual
WARNING! The use of the Peek tubing transfer line connected to the
grounded fitting is mandatory since it prevents any exposed peripherals
connected to the source from floating at high voltage.
Peek Tubing
3. Obtain a new piece of 0.0025” Peek tubing (it must be at least 30 cm long) and place
Peek connection fittings on each end.
WARNING! Peek tubing (or fused silica) with Peek fittings must be used as
the transfer line. Metal tubing or fittings must not be used. The use of metal
may expose the user to high voltage.
4. Connect the Peek tubing to the probe inlet and grounded union/splitter fitting.
13
API 2000 TurboIonSpray Ion Source Manual
14
API 2000 TurboIonSpray Ion Source Manual
WARNING! The use of the Peek tubing transfer line connected to the
grounded fitting is mandatory since it prevents any exposed peripherals
connected to the source from floating at high voltage.
WARNING! Peek tubing (or fused silica) with Peek fittings must be used as
the transfer line. Metal tubing or fittings must not be used. The use of metal
may result in the exposure of high voltage to the user.
15
Appendix A - Consumable Parts
16
Appendix B - Troubleshooting
Appendix B - Troubleshooting
Introduction
The troubleshooting procedures given here are suitable for use by operators who have
access to an ohm meter and are familiar with its use. These procedures do not require
removal of the instrument covers and do not expose the operator to hazardous voltages.
The heater control circuits are protected by a resetable over current device. This device
may be tripped by excessive heater current or by failure to approach the desired operating
temperature within about 5 minutes. When tripped the device will remove power from the
heater. To reset the protective device turn power off the system for about one minute or
remove the ionizer from the system for about one minute.
Diagnosis of the nature of the fault is done by means of ohm meter readings taken across
the pins of the ionizer 24 pin connector on the ionizer which is shown below.
17
Appendix B - Troubleshooting
18