Está en la página 1de 5

Application Note

Measurements of Dielectric Constant and Loss with the


QuadTech 7000 Series Precision RLC Meters and the
Dielectric Products Type LD-3 Cell
Introduction

This application note describes an integrated system for The complete assembly is shown in the photograph
dielectric measurements on solid and liquid materials below. The shielded terminal of the LD-3 cell fixed
that includes the measurement instrument, the test cell, electrode should be connected to the low terminals of the
and the connecting cables and adapters. This setup meter, both HL and PL, by means of the shielded cables,
allows rapid, precise measurements over a wide a Tee and GR-874 adapter (see photograph). The cable
frequency range. shields provide a ground to the body of the cell. The
unshielded terminal on the barrel of the cell should be
Cables and Connections - The necessary cables are connected to the high terminals of the meter, IH and PH,
supplied with the 7000 Series instrument and the with cables, a Tee and the banana pin adapter. The base
connectors, adapters, and the base plate are supplied with plate keeps the cables from moving while the
Dielectric Products Interface Kit, type LD-3-QT7000, measurements are made which is particularly important
which also includes a connection diagram. at higher frequencies.

QuadTech 7600 Precision RLC Meter Shown with Dielectric Products LD-3 Cell and Mounting Plate
Procedures

Calibration - The 7000 Series instrument should have Computer Programs - Dielectric measurements require
had a full calibration within its 1 year calibration precision calculations and are greatly facilitated by
interval. It should have a new OPEN circuit calibration computer programs. Simple programs are easily written
performed, which, for best accuracy, be made at the test for manually entered data. Programs for automatic data
frequency to be used. To do this, remove the IH and PH transfer from the 7000 Series instruments require
leads from the cell (but keep them connected together) reference to the instruction manuals for these
and move them a few inches from the H cell terminal, instruments. QuadTech has developed windows based
shielded from it by the body of the cell. The OPEN applications software which performs the measurement
calibration should be performed using the procedure in techniques described. This software utilizes the 7000
the Operating Instructions for the 7000 Series instrument Series RLC Meters via RS-232 interface. Contact
used. QuadTech for further information.

A new SHORT circuit zero should also be made. At low Accuracy - The accuracy of the measurements of
frequencies simply plug the H plug (carefully) into the L dielectric properties depends on the procedures used, the
connector and make the SHORT calibration. However, two-fluid method being most accurate, and on the
above 10 kHz, don't move the leads, but rather place a accuracy of the separate capacitance and dissipation
soft metal such as folded aluminum foil between the factor measurements. The latter depend on the test
active electrodes and close the electrodes until contact is conditions as shown in the operation instructions for the
made. Do not contact the guard ring. instrument used but may be substantially better than the
specifications given, particularly if calibrations are made
Procedure (See Measurement Methods) - These at the test frequency to be used and if several
dielectric measurements require a series of two or more measurements are averaged to increase precision.
capacitance and dissipation factor measurements from
which the relative dielectric constant (K or εr) and Note that when extreme accuracy is required, the
dissipation factor (D) of the sample can be calculated. QuadTech 1615 Capacitance Bridge (or 1620
The procedures are outlined on the following pages Capacitance Measuring Assembly, see Appendix C) may
along with their formulas. Further information on these be used, but this bridge requires manual balancing and
measurements is available on the GR Application Note has limited frequency range (50 Hz to 100 kHz).
#11 (provided with the LD-3 cell), ASTM Standard D-
150 and IEC Standard 250.

Definitions
where εο is the permittivity of a vacuum, and ε the
Many different notations are used for dielectric absolute permittivity.
properties, see references above. This application note
will use K, the relative dielectric constant, and D, the
ε o = 0.08854 pF cm
dissipation factor (or tan δ) defined as follows:

K = εr '= ε The capacitance of a parallel-plate air capacitor (two


plates) is
and C = K a ε o Area / spacing

D = tan δ = ε r "/ ε r ' where Ka is the dielectric constant of air

The complex relative permittivity is K a = 1.00053


if the air is dry and at normal atmospheric pressure.
ε r * = ε ε o = ε r ' − jε r "
Measurement Methods, Solids
The Air-Gap Method - This method avoids the error
The Contacting Electrode Method - This method is due to the air layer but requires that the thickness of the
quick and easy, but is the least accurate. The results for sample is known. Its thickness should be measured at
K should be within 10% if the sample is reasonably flat. several points over its area and the measured values
should be averaged to get the thickness h. The
h=h h micrometer used should have the same units as those of
o o
the micrometer on the cell.

H H h h
o
L L
H H

L L
G G
Specimen Air
Cxm and Dxm Ca and Da
G G
First the sample is inserted in the cell and the electrodes Air Specimen and Air
closed with the micrometer until they just touch the Ca and Da Cxa and Dxa
sample. Do not force the electrodes against the
sample. Turn the micrometer with a light finger touch. Set the electrodes to about .02 cm or .01 inch greater
Record the electrometer setting as hm. Set the instrument than the sample thickness, h, and measure the equivalent
to measure parallel capacitance and measure the series capacitance and D, as Ca and Da. Note the
capacitance and dissipation factor of the sample as Cxm micrometer setting as hm and correct this with the
and Dxm. micrometer zero calibration, hmo (see Appendix A), to
get
Open the electrodes and remove the sample. Then close
the electrodes to the same micrometer reading, hm. h o = h m + h mo
Measure C (parallel) and D of empty cell as Ca and Da.
Then insert the sample and measure it as Cxa and Dxa.
Calculate Kx and Dx of the sample from Calculate

K x = (10005)C M = ( h o − h) h o
.
and
D x = D xm − D a  
D x = (D xa − D a )
Ca

 C a − MC xa 
The factor 1.0005 in the formula for Kx corrects for the and
dielectric constant of (dry) air. Subtracting Da from Dxm
 (1 − M )C xa   1.0005 
removes any constant phase error in the instrument. For Kx =   
even better D accuracy, adjust the electrode spacing until  C a − MC xa   1 + D x 2 
the measured capacitance is approximately equal to Cxm
and then measure Da.
The factor (1 + Dx2) converts the series value of Cx to
the equivalent parallel value and is not necessary if Dx is
Note that both Kx and Dx will probably be too low
small. The factor of 1.0005 corrects for the dielectric
because there is always some air between the electrodes
constant of air (if dry). The formula for Dx assumes that
and the sample. This error is smallest for very flat
the true D of air is zero and it makes a correction for a
samples, for thicker samples and for those with low K
constant D error in the instrument.
and D values.
Cf and Df Cxf and Dxf
The Two-Fluid Method - This method is preferred for
specimens whose thickness is difficult to measure and From these measurements calculate
for best accuracy which will be limited by the accuracy
of the C and D measurements. However it requires four h C C (C - C xa )
= 1 − a f xf
C xa C xf (C f - C a )
measurements, two using a second fluid (the first being
air). The dielectric properties of this fluid need not be
ho
known, but it must not react with the specimen and it and
must be stable and safe to use. A silicone fluid such as
C xser C xf C xa (C f − C a )
Dow Corning 200, 1 centistoke viscosity, is most =
generally satisfactory. Ca C a (C xa C f − C xf C a )

The four measurements of series capacitance and D are which is the ratio of the equivalent series capacitance of
outlined in the figure below. Note the spacing is the the sample to Ca. If Dx is close to Df or larger use
same for all measurements and should be just slightly

C a (C xf − C xa )(D xf − D f )
more than the specimen thickness. The accuracy will be
limited mainly by the accuracy of the measurements
D x = D xf +
made. C xa C f − C xf C a
If Dx is very small use
h
o
h
Dx =
(D xa − Da )C xf (C f − C a )
C xa C f − C xf C a
H H which makes a zero D correction.
L L
From the above results calculate
 h   C   1.0005 
K x =    xser   2
 ho   Ca   1 + D x 
G G
Air Specimen and Air As before, the factor of 1.0005 corrects for the dielectric
Ca and Da Cxa and Dxa constant of air (if dry) and the (1 + Dx2) factor converts
Cx to equivalent parallel capacitance.

h h
o

H H

L L

G G
Fluid Specimen and Fluid

Measurement Methods, Liquids


and
Measurements on liquids are simple, the only difficulty is D x = D xm − D a
with handling and cleanup.
Note that the spacing is not critical but should be narrow
enough to make the capacitance large enough to be
Measure equivalent parallel capacitance and D first of air
measured accurately.
(Ca and Da) and then of the liquid to be measured (Cxm
and Dxm), then determine
Other liquid cells are available that are easier to clean and
use a smaller sample (see Appendix B).
C xm
Kx = (1.0005)
Ca
APPENDICES:

(A) Micrometer Zero Calibration - The micrometer zero (C) Other Instruments for Dielectric Measurements
correction may change with used. To check this, measure (QuadTech, Inc.)
the capacitance of the empty cell with a narrow setting,
h1, getting C1, and with a wider setting, h2, getting C2. 1615 Capacitance Bridge (a precision, manual bridge)
The value of the zero offset, hmo, is
h 2 C 2 − h 1C1 1620 Capacitance Measuring Assembly
h mo = (a 1615 with a 1311-A Oscillator and 1232-A Detector)
C1 − C 2
Greater accuracy is obtained by taking measurements at Digibridge Impedance Meters (types 1689, 1689M,
more spacings and averaging the results. 1692 & 1693)

(B) Accessories and Other Test Cells


(Dielectric Products, Co.*)

Interface Cable Set (see above)


High Temperature Cell LD-3T
High temperature Cable Set
Dielectric Cell Top Cover LD-3-TC
Thermocouple & Fitting
Teflon standard TS-100
Liquid Cell 350
Powder and Paste Cell MC-100
Resistivity Cell RF-100

* Dielectrics Products Co. 178 Orchard Street, Watertown, MA 02172 617-924-5688

The information presented here is subject to change and is intended for general information only

QuadTech, Inc.
Tel. 800-253-1230 Website: www.quadtech.com
Printed in U.S.A. PN 035012 July, 1999

También podría gustarte