Está en la página 1de 3

COAGULATION CONTROL N COAGULATION CONTROL N

(CONTROL COAGULACION N) (CONTROL COAGULACION N)


CONTROL NORMAL / NORMAL CONTROL CONTROL NORMAL / NORMAL CONTROL

Quantitative determination of coagulation factors Determinación cuantitativa de factores de coagulación Quantitative determination of coagulation factors Determinación cuantitativa de factores de coagulación
IVD IVD IVD IVD
Store at 2 - 8ºC. Conservar a 2 - 8ºC Store at 2 - 8ºC. Conservar a 2 - 8ºC

PRODUCT CHARACTERISTICS CARACTERISTICAS DEL PRODUCTO PRODUCT CHARACTERISTICS CARACTERISTICAS DEL PRODUCTO
The Control is a lyophilised human plasma, used to evaluate the El Control es un plasma humano liofilizado utilizado para evaluar la The Control is a lyophilised human plasma, used to evaluate El Control es un plasma humano liofilizado utilizado para evaluar la
precision and accuracy of PT, APTT and Fibrinogen precisión y exactitud en la determinación de PT, APTT y Fibrinógeno the precision and accuracy of PT, APTT and Fibrinogen precisión y exactitud en la determinación de PT, APTT y Fibrinógeno
determinations in human plasma. en plasma humano. determinations in human plasma. en plasma humano.
REAGENTS REACTIVOS REAGENTS REACTIVOS
Human plasma collected with <0.4% sodium citrate Plasma humano con citrato sódico <0.4% como Human plasma collected with <0.4% sodium citrate Plasma humano con citrato sódico <0.4% como
anticoagulant, with a NORMAL concentration of coagulation anticoagulante y un nivel de concentración NORMAL de los anticoagulant, with a NORMAL concentration of coagulation anticoagulante y un nivel de concentración NORMAL de los
factors. factores de coagulación. factors. factores de coagulación.
Coagulation factors concentration is indicated below. Su concentración esta indicada en la tabla anexa. Coagulation factors concentration is indicated below. Su concentración esta indicada en la tabla anexa.

PRECAUTIONS PRECAUCIONES PRECAUTIONS PRECAUCIONES


Each unit of source material used in the preparation of this Cada unidad de material usado en la preparación de este reactivo has Each unit of source material used in the preparation of this Cada unidad de material usado en la preparación de este reactivo has
product has been tested by an FDA licensed method and found sido testada por métodos aprobados por la FDA, resultando no product has been tested by an FDA licensed method and sido testada por métodos aprobados por la FDA, resultando no
non-reactive for HBsAg and negative for antibodies to HIV and reactivos a anticuerpos HBsAg, HIV y HCV. Sin embargo, dado que found non-reactive for HBsAg and negative for antibodies to reactivos a anticuerpos HBsAg, HIV y HCV. Sin embargo, dado que
HCV. However, as no known test method can offer complete ningún método puede asegurar completamente que productos HIV and HCV. However, as no known test method can offer ningún método puede asegurar completamente que productos
assurance that products derived from human blood will not derivados de humanos no puedan transmitir enfermedades complete assurance that products derived from human blood derivados de humanos no puedan transmitir enfermedades
transmit infectious diseases, this product must be handled as infecciosas, este producto debe ser manipulado como material will not transmit infectious diseases, this product must be infecciosas, este producto debe ser manipulado como material
potencially infectious biological material. biológico potencialmente infeccioso. handled as potencially infectious biological material. biológico potencialmente infeccioso.
PREPARATION PREPARACION PREPARATION PREPARACION
Reconstitute () with 1.0 mL of distilled water. Swirl gently Reconstituir con 1,0 mL de agua destilada. Mover lentamente en Reconstitute () with 1.0 mL of distilled water. Swirl gently Reconstituir con 1,0 mL de agua destilada. Mover lentamente en
(do not invert vial or mix vigorously) and let stand undisturbed círculos y dejar reposar durante 15 minutos a temperatura ambiente. (do not invert vial or mix vigorously) and let stand undisturbed círculos y dejar reposar durante 15 minutos a temperatura ambiente.
for 15 minutes at room temperature before use. No invertir el frasco ni agitarlo vigorosamente. for 15 minutes at room temperature before use. No invertir el frasco ni agitarlo vigorosamente.
STORAGE AND STABILITY CONSERVACION Y ESTABILIDAD STORAGE AND STABILITY CONSERVACION Y ESTABILIDAD
The control is stable until the expiration date on the label when El control es estable hasta la fecha de caducidad indicada en el The control is stable until the expiration date on the label when El control es estable hasta la fecha de caducidad indicada en el envase
stored tightly closed at 2-8ºC and contaminations are prevented envase cuando se mantiene el vial bien cerrado a 2-8ºC, y se evita stored tightly closed at 2-8ºC and contaminations are cuando se mantiene el vial bien cerrado a 2-8ºC, y se evita la
during their use. Do not use reagents over the expiration date. la contaminación durante su uso. No utilizar reactivos que hayan prevented during their use. Do not use reagents over the contaminación durante su uso. No utilizar reactivos que hayan
After reconstitution, it’s stable for 8 hours tightly closed at 2-8ºC. sobrepasado la fecha de caducidad. expiration date. After reconstitution, it’s stable for 8 hours sobrepasado la fecha de caducidad.
Gently mix contents prior to each use. Después de la reconstitución es estable 8 horas cuando se mantiene tightly closed at 2-8ºC. Gently mix contents prior to each use. Después de la reconstitución es estable 8 horas cuando se mantiene
Erratic values, product color variations, or lack of vacuum in the el vial bien cerrado a 2-8ºC. Mezclar suavemente antes de cada uso. Erratic values, product color variations, or lack of vacuum in el vial bien cerrado a 2-8ºC. Mezclar suavemente antes de cada uso.
vials could indicate product deterioration. However, poor control Valores erráticos, variación del color pueden ser indicativos de la the vials could indicate product deterioration. However, poor Valores erráticos, variación del color pueden ser indicativos de la
performance could also be due to other factors within the test. deterioración del producto. Sin embargo, una escaso control del control performance could also be due to other factors within deterioración del producto. Sin embargo, una escaso control del
método puede ser debido a otros factores internos del test. the test. método puede ser debido a otros factores internos del test.
PROCEDURE
The Control should be run as a sample, at the initiation of PROCEDIMIENTO PROCEDURE PROCEDIMIENTO
testing each day and at least once each shift, or with each El Control debe tratarse como si fuera una muestra; Deben ser The Control should be run as a sample, at the initiation of El Control debe tratarse como si fuera una muestra; Deben ser
group of assays. Controls should also be tested with each analizados al inicio de las pruebas y al menos una vez en cada turno, testing each day and at least once each shift, or with each analizados al inicio de las pruebas y al menos una vez en cada turno,
reagent change or major instrument adjustment. Compare test o con cada grupo de ensayos, cada vez que cambie de reactivo o group of assays. Controls should also be tested with each o con cada grupo de ensayos, cada vez que cambie de reactivo o
results obtained to the expected results for the test method realice un ajuste importante del instrumento. reagent change or major instrument adjustment. Compare realice un ajuste importante del instrumento.
and control plasma. Comparar los resultados obtenidos con los resultados esperados test results obtained to the expected results for the test Comparar los resultados obtenidos con los resultados esperados
según el método y el control. method and control plasma. según el método y el control.

COMPONENT / COMPONENTE METHOD / MÉTODO RANGE / RANGO COMPONENT / COMPONENTE METHOD / METODO RANGE / RANGO
PT (Prothrombin Time / Tiempo de Protombina) 12.2 (10.4 – 14.0) s PT (Prothrombin Time / Tiempo de Protombina) 12.2 (10.4 – 14.0) s
According to Instructions Sheet of According to Instructions Sheet of
APTT Activated Partial Thromboplastin Test corresponding Spinreact reagent/ Según Hoja APTT Activated Partial Thromboplastin Test corresponding Spinreact reagent/ Según
Tiempo de Tromboplatina Parcial Activada
29.9 (25.4 – 34.4) s Tiempo de Tromboplatina Parcial Activada
29.9 (25.4 – 34.4) s
de Instrucciones del reactivo Spinreact Hoja de Instrucciones del reactivo Spinreact
Fibrinogen / Fibrinógeno correspondiente Fibrinogen / Fibrinógeno correspondiente
202 (172 - 232) mg/dL 202 (172- 232) mg/dL

Los resultados reales dependen de muchos factores, entre los cuales Los resultados reales dependen de muchos factores, entre los cuales
Actual results depend on many factors, including lot number, type of Actual results depend on many factors, including lot number, type
se encuentran el número de lote, el tipo de reactivo y el instrumento. se encuentran el número de lote, el tipo de reactivo y el instrumento.
reagent, and instrument. Ranges must be determined in each of reagent, and instrument. Ranges must be determined in each
Los intervalos deben determinarse en cada laboratorio, cada vez que Los intervalos deben determinarse en cada laboratorio, cada vez que
laboratory with each change of lot number, reagent, or instrument. laboratory with each change of lot number, reagent, or instrument.
se cambie de número de lote del control, de reactivo o del instrumento. se cambie de número de lote del control, de reactivo o del instrumento.
This Control data sheet is applicable to sublots. Sequential alphabetical Este Boletín de análisis es aplicable al lote y sublotes. Letras alfabéticas This Control data sheet is applicable to sublots. Sequential alphabetical Este Boletín de análisis es aplicable al lote y sublotes. Letras alfabéticas
letter (e.g. A, B, C etc.) following the lot number. secuenciales (p.e. A, B , C etc.) que siguen al nº de lote. letter (e.g. A, B, C etc.) following the lot number. secuenciales (p.e. A, B , C etc.) que siguen al nº de lote.

1709104 4 x 1 mL 1709104 4 x 1 mL

COIS05 13/06/14 SPINREACT,S.A./S.A.U. Ctra.Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN COIS05 13/06/14 SPINREACT,S.A/S.A.U.. Ctra.Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com
COAGULATION CONTROL N COAGULATION CONTROL N

CONTRÔLE NORMAL / NORMAL CONTROL CONTROL NORMAL / NORMAL CONTROL


Coagulation / Coagulation Coagulation / Coagulación
Quantitative determination of coagulation factors Détermination quantitative de facteurs de coagulation Quantitative determination of coagulation factors Détermination quantitative de facteurs de coagulation
IVD IVD IVD IVD
Store at 2 - 8ºC. Conserver à 2-8ºC Store at 2 - 8ºC. Conserver à 2-8ºC

PRODUCT CHARACTERISTICS CARACTÉRISTIQUES DU PRODUIT PRODUCT CHARACTERISTICS CARACTÉRISTIQUES DU PRODUIT


The Control is a lyophilised human plasma, used to evaluate Le Contrôle est un plasma humain lyophilisé qui est utilisé pour The Control is a lyophilised human plasma, used to evaluate Le Contrôle est un plasma humain lyophilisé qui est utilisé pour
the precision and accuracy of PT, APTT and Fibrinogen évaluer la précision et l’exactitude dans la détermination de PT, the precision and accuracy of PT, APTT and Fibrinogen évaluer la précision et l’exactitude dans la détermination de PT, APTT
determinations in human plasma. APTT et fibrinogène dans le plasma humain. determinations in human plasma. et fibrinogène dans le plasma humain.
REAGENTS RÉACTIFS REAGENTS RÉACTIFS
Human plasma collected with <0.4% sodium citrate Plasma humain avec citrate de sodium <0,4 % comme Human plasma collected with <0.4% sodium citrate Plasma humain avec citrate de sodium <0,4 % comme
anticoagulant, with a NORMAL concentration of anticoagulant et un niveau de concentration NORMAL des anticoagulant, with a NORMAL concentration of anticoagulant et un niveau de concentration NORMAL des
coagulation factors. facteurs de coagulation. coagulation factors. facteurs de coagulation.
Coagulation factors concentration is indicated below. Sa concentration est indiquée dans le tableau en annexe. Coagulation factors concentration is indicated below. Sa concentration est indiquée dans le tableau en annexe.

PRECAUTIONS PRÉCAUTIONS PRECAUTIONS PRÉCAUTIONS


Each unit of source material used in the preparation of this Chaque unité de matériel utilisée dans la préparation de ce réactif a Each unit of source material used in the preparation of this Chaque unité de matériel utilisée dans la préparation de ce réactif a
product has been tested by an FDA licensed method and été testée par des méthodes approuvées par la FDA, ce qui implique product has been tested by an FDA licensed method and été testée par des méthodes approuvées par la FDA, ce qui implique
found non-reactive for HBsAg and negative for antibodies to une non réactivité face aux anticorps HBsAg, HIV et HCV. found non-reactive for HBsAg and negative for antibodies to une non réactivité face aux anticorps HBsAg, HIV et HCV.
HIV and HCV. However, as no known test method can offer Cependant, vu qu’aucune méthode ne peut garantir complètement HIV and HCV. However, as no known test method can offer Cependant, vu qu’aucune méthode ne peut garantir complètement
complete assurance that products derived from human blood que les produits dérivés d’êtres humains ne transmettront pas de complete assurance that products derived from human blood que les produits dérivés d’êtres humains ne transmettront pas de
will not transmit infectious diseases, this product must be maladies infectieuses, ce produit doit être manipulé comme un will not transmit infectious diseases, this product must be maladies infectieuses, ce produit doit être manipulé comme un
handled as potencially infectious biological material. matériel biologique potentiellement infectieux. handled as potencially infectious biological material. matériel biologique potentiellement infectieux.
PREPARATION PRÉPARATION PREPARATION PRÉPARATION
Reconstitute () with 1.0 mL of distilled water. Swirl gently Reconstituer avec 1,0 mL d’eau distillée. Remuer lentement en Reconstitute () with 1.0 mL of distilled water. Swirl gently Reconstituer avec 1,0 mL d’eau distillée. Remuer lentement en
(do not invert vial or mix vigorously) and let stand undisturbed cercles et laisser reposer pendant 15 minutes à température (do not invert vial or mix vigorously) and let stand undisturbed cercles et laisser reposer pendant 15 minutes à température
for 15 minutes at room temperature before use. ambiante. Ne pas retourner le flacon ni l’agiter énergiquement. for 15 minutes at room temperature before use. ambiante. Ne pas retourner le flacon ni l’agiter énergiquement.
STORAGE AND STABILITY CONSERVATION ET STABILITÉ STORAGE AND STABILITY CONSERVATION ET STABILITÉ
The control is stable until the expiration date on the label when Le contrôle est stable jusqu’à la date de péremption indiquée sur The control is stable until the expiration date on the label Le contrôle est stable jusqu’à la date de péremption indiquée sur
stored tightly closed at 2-8ºC and contaminations are l’emballage si le flacon est bien fermé et conservé à 2-8ºC, et que when stored tightly closed at 2-8ºC and contaminations are l’emballage si le flacon est bien fermé et conservé à 2-8ºC, et que l’on
prevented during their use. Do not use reagents over the l’on évite sa contamination pendant l’utilisation. Ne pas utiliser de prevented during their use. Do not use reagents over the évite sa contamination pendant l’utilisation. Ne pas utiliser de réactifs
expiration date. After reconstitution, it’s stable for 8 hours réactifs dont la date de péremption serait dépassée. expiration date. After reconstitution, it’s stable for 8 hours dont la date de péremption serait dépassée.
tightly closed at 2-8ºC. Gently mix contents prior to each use. Après la reconstitution, il est stable 8 heures si le flacon est bien tightly closed at 2-8ºC. Gently mix contents prior to each use. Après la reconstitution, il est stable 8 heures si le flacon est bien
Erratic values, product color variations, or lack of vacuum in fermé et conservé à 2-8ºC. Mélanger doucement avant chaque Erratic values, product color variations, or lack of vacuum in fermé et conservé à 2-8ºC. Mélanger doucement avant chaque
the vials could indicate product deterioration. However, poor utilisation. the vials could indicate product deterioration. However, poor utilisation.
control performance could also be due to other factors within Des valeurs erratiques et une variation de la couleur peuvent control performance could also be due to other factors within Des valeurs erratiques et une variation de la couleur peuvent indiquer
the test. indiquer la détérioration du produit. Toutefois, un faible contrôle de la the test. la détérioration du produit. Toutefois, un faible contrôle de la méthode
méthode peut être dû à d’autres facteurs internes du test. peut être dû à d’autres facteurs internes du test.
PROCEDURE PROCEDURE
The Control should be run as a sample, at the initiation of PROCÉDURE The Control should be run as a sample, at the initiation of PROCÉDURE
testing each day and at least once each shift, or with each Le Contrôle doit être traité à la manière d’un échantillon. Il doit être testing each day and at least once each shift, or with each Le Contrôle doit être traité à la manière d’un échantillon. Il doit être
group of assays. Controls should also be tested with each analysé au début des essais et, au moins, une fois à chaque tour, ou group of assays. Controls should also be tested with each analysé au début des essais et, au moins, une fois à chaque tour, ou
reagent change or major instrument adjustment. Compare test avec chaque groupe d’essais, chaque fois qu’il y a un changement reagent change or major instrument adjustment. Compare avec chaque groupe d’essais, chaque fois qu’il y a un changement de
results obtained to the expected results for the test method de réactif ou qu’un réglage important de l’instrument est effectué. test results obtained to the expected results for the test réactif ou qu’un réglage important de l’instrument est effectué.
and control plasma. Comparer les résultats obtenus avec les résultats attendus en method and control plasma. Comparer les résultats obtenus avec les résultats attendus en
fonction de la méthode et du contrôle. fonction de la méthode et du contrôle.

COMPONENT / COMPOSANT METHOD / MÉTHODE RANGE / PLAGE COMPONENT / COMPOSANT METHOD / MÉTHODE RANGE / PLAGE
PT (Prothrombin Time / Temps de
Prothrombine)
14,0 (11,9 – 16,1) s PT (Prothrombin Time / Temps de Prothrombine) According to Instructions Sheet of 14.0 (11.9 – 16.1) s
According to Instructions Sheet of corresponding
corresponding Spinreact reagent/ Selon fiche
APTT Activated Partial Thromboplastin Test Spinreact reagent/ Selon fiche d’instructions du APTT Activated Partial Thromboplastin Test
Temps de Thromboplastine Partielle Activée réactif Spinreact correspondante
28,6 (24,3 – 32,9) s Temps de Thromboplastine Partielle Activée
d’instructions du réactif Spinreact 28.6 (24.3– 32.9) s
correspondante
Fibrinogen / Fibrinogène 202 (172 - 232) mg/dL Fibrinogen / Fibrinogène 202 (172- 232) mg/dL

Les résultats réels dépendent de nombreux facteurs dont le numéro de Les résultats réels dépendent de nombreux facteurs dont le numéro de
Actual results depend on many factors, including lot number, type of Actual results depend on many factors, including lot number, type
lot, le type de réactif et l’instrument. Les intervalles doivent être lot, le type de réactif et l’instrument. Les intervalles doivent être
reagent, and instrument. Ranges must be determined in each of reagent, and instrument. Ranges must be determined in each
déterminés dans chaque laboratoire, dès que l’on change de numéro déterminés dans chaque laboratoire, dès que l’on change de numéro
laboratory with each change of lot number, reagent, or instrument. laboratory with each change of lot number, reagent, or instrument.
de lot, de réactif ou d’instrument. de lot, de réactif ou d’instrument.
This Control data sheet is applicable to sublots. Sequential alphabetical Ce Bulletin d’analyse est applicable au lot et aux sous-lots. Lettres This Control data sheet is applicable to sublots. Sequential alphabetical Ce Bulletin d’analyse est applicable au lot et aux sous-lots. Lettres
letter (e.g. A, B, C etc.) following the lot number. alphabétiques séquentielles (par ex. A, B, C, etc.) qui suivent le nº de lot. letter (e.g. A, B, C etc.) following the lot number. alphabétiques séquentielles (par ex. A, B, C, etc.) qui suivent le nº de lot.

1709104 4 x 1 mL 1709104 4 x 1 mL
COIS05-F 03/03/16 SPINREACT,S.A./S.A.U. Ctra.Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN COIS05-F 03/03/16 SPINREACT,S.A./S.A.U. Ctra.Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tél. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail : spinreact@spinreact.com Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com
COAGULATION CONTROL N COAGULATION CONTROL N

(CONTROL COAGULAÇÃO P) (CONTROL COAGULAÇÃO P)


CONTROL NORMAL / NORMAL CONTROL CONTROL NORMAL / NORMAL CONTROL

Quantitative determination of coagulation factors Determinação quantitativa de factores de coagulação Quantitative determination of coagulation factors Determinação quantitativa de factores de coagulação
IVD IVD IVD IVD
Store at 2 - 8ºC. Conservar a 2 - 8ºC Store at 2 - 8ºC. Conservar a 2 - 8ºC
PRODUCT CHARACTERISTICS CARACTERÍSTICAS DO PRODUTO PRODUCT CHARACTERISTICS CARACTERÍSTICAS DO PRODUTO
The Control is a lyophilised human plasma, used to evaluate the O Control é um plasma humano liofilizado utilizado para avaliar a The Control is a lyophilised human plasma, used to evaluate the O Control é um plasma humano liofilizado utilizado para avaliar a
precision and accuracy of PT, APTT and Fibrinogen determinations in precisão e exactidão na determinação de PT, APTT e Fibrinogénio no precision and accuracy of PT, APTT and Fibrinogen determinations precisão e exactidão na determinação de PT, APTT e Fibrinogénio no
human plasma. plasma humano. in human plasma. plasma humano.
REAGENTS REAGENTES REAGENTS REAGENTES
Human plasma collected with <0.4% sodium citrate anticoagulant, Plasma humano com citrato de sódio <0.4% como anticoagulante com Human plasma collected with <0.4% sodium citrate Plasma humano com citrato de sódio <0.4% como anticoagulante com
with a NORMAL concentration of coagulation factors. um valor de concentração NORMAL dos factores de coagulação. anticoagulant, with a NORMAL concentration of coagulation um valor de concentração NORMAL dos factores de coagulação.
Coagulation factors concentration is indicated below. A concentração dos factores de coagulação está indicada na tabela factors. A concentração dos factores de coagulação está indicada na tabela
abaixo. Coagulation factors concentration is indicated below. abaixo.
PRECAUTIONS PRECAUÇÕES PRECAUTIONS PRECAUÇÕES
Each unit of source material used in the preparation of this product Cada unidade de material usado na preparação deste produto foi testada Each unit of source material used in the preparation of this product Cada unidade de material usado na preparação deste produto foi testada
has been tested by an FDA licensed method and found non-reactive por um método aprovado pela FDA, com resultado não reactivo a has been tested by an FDA licensed method and found non- por um método aprovado pela FDA, com resultado não reactivo a
for HBsAg and negative for antibodies to HIV and HCV. However, as anticorpos HBsAg, HIV e HCV. No entanto, uma vez que nenhum reactive for HBsAg and negative for antibodies to HIV and HCV. anticorpos HBsAg, HIV e HCV. No entanto, uma vez que nenhum
no known test method can offer complete assurance that products método pode assegurar completamente que produtos derivados de However, as no known test method can offer complete assurance método pode assegurar completamente que produtos derivados de
derived from human blood will not transmit infectious diseases, this sangue humano não possam transmitir doenças infecciosas, este that products derived from human blood will not transmit infectious sangue humano não possam transmitir doenças infecciosas, este
product must be handled as potencially infectious biological material. produto deve ser manipulado como material biológico potencialmente diseases, this product must be handled as potencially infectious produto deve ser manipulado como material biológico potencialmente
infeccioso. biological material. infeccioso.
PREPARATION
Reconstitute () with 1.0 mL of distilled water. Swirl gently (do not PREPARAÇÃO PREPARATION PREPARAÇÃO
invert vial or mix vigorously) and let stand undisturbed for 15 minutes Reconstituir com 1,0 ml de água destilada. Agitar lentamente com Reconstitute () with 1.0 mL of distilled water. Swirl gently (do not Reconstituir com 1,0 ml de água destilada. Agitar lentamente com
at room temperature before use. movimentos circulares e deixar repousar durante 15 minutos à invert vial or mix vigorously) and let stand undisturbed for 15 movimentos circulares e deixar repousar durante 15 minutos à
temperatura ambiente. Não inverter o frasco nem agitar vigorosamente. minutes at room temperature before use. temperatura ambiente. Não inverter o frasco nem agitar vigorosamente.
STORAGE AND STABILITY
The control is stable until the expiration date on the label when stored STORAGE AND STABILITY CONSERVAÇÃO E ESTABILIDADE
CONSERVAÇÃO E ESTABILIDADE The control is stable until the expiration date on the label when O control é estável até ao final do prazo de validade indicada no frasco
tightly closed at 2-8ºC and contaminations are prevented during their O control é estável até ao final do prazo de validade indicada no frasco
use. Do not use reagents over the expiration date. After reconstitution, stored tightly closed at 2-8ºC and contaminations are prevented quando este é mantido bem fechado a2-8ºC, e as contaminações são
quando este é mantido bem fechado a2-8ºC, e as contaminações são during their use. Do not use reagents over the expiration date. After evitadas durante a sua utilização. Não utilizar reagentes com prazo de
it’s stable for 8 hours tightly closed at 2-8ºC. Gently mix contents prior evitadas durante a sua utilização. Não utilizar reagentes com prazo de
to each use. reconstitution, it’s stable for 8 hours tightly closed at 2-8ºC. Gently validade ultrapassado.
validade ultrapassado. mix contents prior to each use. Após a reconstituição, é estável durante 8 horas, se o frasco estiver bem
Após a reconstituição, é estável durante 8 horas, se o frasco estiver bem fechado a 2-8ºC. Agitar suavemente antes de cada utilização.
Erratic values, product color variations, or lack of vacuum in the vials fechado a 2-8ºC. Agitar suavemente antes de cada utilização. Erratic values, product color variations, or lack of vacuum in the
could indicate product deterioration. However, poor control Valores erráticos, variações da cor do produto ou falta de vácuo nos
Valores erráticos, variações da cor do produto ou falta de vácuo nos vials could indicate product deterioration. However, poor control frascos podem ser indicativos da deterioração do produto. No entanto,
performance could also be due to other factors within the test. frascos podem ser indicativos da deterioração do produto. No entanto, performance could also be due to other factors within the test. um fraco controlo do método também pode ser devido a outros factores
um fraco controlo do método também pode ser devido a outros factores internos do teste.
PROCEDURE internos do teste. PROCEDURE
The Control should be run as a sample, at the initiation of The Control should be run as a sample, at the initiation of PROCEDIMENTO
testing each day and at least once each shift, or with each group of PROCEDIMENTO testing each day and at least once each shift, or with each group of O Control deve ser tratado como se fosse uma amostra; Devem ser
assays. Controls should also be tested with each reagent change or O Control deve ser tratado como se fosse uma amostra; Devem ser assays. Controls should also be tested with each reagent change analisados diáriamente antes de iniciar o método e pelo menos uma vez
major instrument adjustment. Compare test results obtained to the analisados diáriamente antes de iniciar o método e pelo menos uma vez or major instrument adjustment. Compare test results obtained to por cada turno ou com cada grupo de ensaios. O Control também deve
expected results for the test method and control plasma. por cada turno ou com cada grupo de ensaios. O Control também deve the expected results for the test method and control plasma. ser testado de cada vez que se muda de reagente ou se realiza um
ser testado de cada vez que se muda de reagente ou se realiza um ajuste importante do instrumento.
ajuste importante do instrumento. Devem ser comparados os resultados obtidos com os resultados
Devem ser comparados os resultados obtidos com os resultados esperados de acordo com o método e o control.
esperados de acordo com o método e o control.

COMPONENT / COMPONENTE METHOD / MÉTODO RANGE / INTERVALO COMPONENT / COMPONENTE METHOD / METODO RANGE / INTERVALO
PT (Prothrombin Time / Tempo de
PT (Prothrombin Time / Tempo de Protombina) 13.1 (11.1 – 15.1) s Protombina)
13.1 (11.1 – 15.1) s
According to Instructions Sheet of According to Instructions Sheet of
APTT Activated Partial Thromboplastin Test corresponding Spinreact reagent/ Conforme a APTT Activated Partial Thromboplastin Test corresponding Spinreact reagent/ Conforme a
Tempo de Tromboplatina Parcial Activada
31.0 (26.3 – 35.7) s Tempo de Tromboplatina Parcial Activada
31.0 (26.3 – 35.7) s
folha de instrução do reagente Spinreact folha de instrução do reagente Spinreact
Fibrinogen / Fibrinogénio correspondente Fibrinogen / Fibrinogénio correspondente
210 (178 - 242) mg/dL 210 (178- 242) mg/dL

Os resultados reais dependem de muitos factores, entre os quais se Os resultados reais dependem de muitos factores, entre os quais se
Actual results depend on many factors, including lot number, type of Actual results depend on many factors, including lot number, type of
encontram o número de lote, o tipo de reagente e o instrumento. Os encontram o número de lote, o tipo de reagente e o instrumento. Os
reagent, and instrument. Ranges must be determined in each laboratory reagent, and instrument. Ranges must be determined in each laboratory
with each change of lot number, reagent, or instrument. intervalos devem ser determinados em cada laboratório, cada vez que with each change of lot number, reagent, or instrument. intervalos devem ser determinados em cada laboratório, cada vez que se
se muda de número de lote do control, de reagente ou de instrumento. muda de número de lote do control, de reagente ou de instrumento.
This Control value sheet is applicable to sublots. Sequential alphabetical Esta Folha de valores é aplicável a lotes e sublotes. Letras alfabéticas This Control value sheet is applicable to sublots. Sequential alphabetical letter Esta Folha de valores é aplicável a lotes e sublotes. Letras alfabéticas
letter (e.g. A, B, C etc.) following the lot number. sequenciais (p.e. A, B , C etc.) que se seguem ao nº de lote. (e.g. A, B, C etc.) following the lot number. sequenciais (p.e. A, B , C etc.) que se seguem ao nº de lote.

1709104 4 x 1 mL 1709104 4 x 1 mL

COIS05-P 13/06/14 SPINREACT,S.A./S.A.U. Ctra.Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN COIS05-P 13/06/14 PINREACT,S.A. /S.A.U.Ctra.Santa Coloma, 7 E-17176 SANT ESTEVE DE BAS (GI) SPAIN
Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com Tel. +34 972 69 08 00 Fax +34 972 69 00 99 e-mail: spinreact@spinreact.com

También podría gustarte