AUSENCIA DE DETERGENTE

EQUIPO: UV-1800 ESP0004.


VIGENCIA OQ: 01/2022.

Software Information
Software Name: UVProbe
Version: 2.31
Mode: Security Mode

Data Information
Filename: \\172.27.100.26\cromatografos\ESP0004\CONTROL DE CALIDAD\ESPECTROFOTOMETRO UV\RESULTADOS AUSENCIAS\2021
\080421 DETERGENTE\DET 080421.pho
Title: DETERMINACIÓN DE TRAZAS DE DETERGENTE ANIONICO AL 14% DODISULFONATO DE SODIO
Analyst: System Administrator
Date/Time: 08/04/2021 15:19:00
Comments: Según Tecnica 2725 Edición 2

Criterio de Aceptación: La concentración de las muestras no pueden ser mayor a 10 µg/mL de dodecilsulfonato de sodio.

Criterio de Aceptación= (1876.4mg/500mL)*(2/100mL)*0.14*(1000µg/mL)= 10.508µg/mL

Balanza: BAL0109 Ultrasonido: ULT0004

Instrument Information
Instrument Name: UV
Instrument Type: UV-1800 Series
Model (S/N): UV 1800 (A11454500305)

Standard Curve
0.512

0.400
Abs.

0.200

-0.047
0.000 5.000 10.508
Conc. (µg/mL)
y = 0.04431 x + 0.00000

Standard Table
Sample ID Type Date Time Conc WL223.0 Comments
1 CC 13251 Std-Repeat 08/04/2021 14:34:04 10.508 0.466 Solución Refeencia
2 CC 13251-2 Std-Repeat 08/04/2021 14:34:05 10.508 0.465 Detergente Anionico 14%
3 CC 13251-3 Std-Repeat 08/04/2021 14:34:05 10.508 0.465 Dodecilsulfonato de Sodio
4 CC 13251-4 Std-Repeat 08/04/2021 14:34:06 10.508 0.466
5 CC 13251-5 Std-Repeat 08/04/2021 14:34:06 10.508 0.466
6 CC 13251-Avg Average 08/04/2021 14:34:06 10.508 0.466 Avg of preceding 5 Samples
7

REALIZADO POR: REVISADO POR:


FECHA: FECHA:
AUSENCIA DE DETERGENTE
EQUIPO: UV-1800 ESP0004.
VIGENCIA OQ: 01/2022.

Sample Table
Sample ID Type Date Time Conc WL223.0 Comments
1 BLANCO Unknown 08/04/2021 14:33:09 0.00585 0.00026 Agua Purificada NA 167249
2 NA 161075 Unknown 08/04/2021 15:11:24 0.00930 0.00041 Iodo plus 10% solución topica
15

Method Parameters
[Wavelengths]
Wavelength Name: WL223.0
Wavelength: 223.00 nm

[Calibration Curve]
Column for Cal. Curve: WL223.0
Cal. Curve Type: Multi Point
Cal. Curve Unit: µg/mL
Selected Wavelength: WL223.0
Calibration Equation: Abs = K1*(Conc) + K0
Zero Interception: Selected

[Measurement Parameters(Standard)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Enabled
Repetitions: 5

[Measurement Parameters(Sample)]
Data Acquired by: Instrument
Delay sample read: Disabled
Repeat: Disabled

[Equations]

[Pass Fail]

[Method Summary]
Title:
Date/Time: 20/10/2020 20:27:51
Comments:
Sample Preparations:

REALIZADO POR: REVISADO POR:


FECHA:
FECHA: