Está en la página 1de 7

Mechanisms, Procedures, and Instruments to Detect Faults and Defects of

Electro-Optical Oceanic Cable Arrays in SSW vs. FW


Colin Sandwith1 and Michael Welch1
1
Applied Physics Laboratory, University of Washington, Seattle, WA 98105 USA

Abstract - All marine cable systems must be immersion tested systems and arrays. The possible causes of faults or defects prior to
prior to final delivery. Selecting and determining the tests and service are listed here and not necessarily repeated herein:
environments is a compromise between risk and consequence of 1. SSW leaks
early system mortality versus the cost and time of the testing. 2. Galvanic cells
Reasons, methods, and instruments for insulation resistance 3. Improper repairs
testing in natural seawater or simulated seawater vs. natural or 4. Thin electrical insulation
fresh water in hydropressure and/or soak tanks are discussed 5. Cracks and/or pinholes
here. Voltmeters, ohmmeters, and meggers detect and locate two 6. O-ring seal leak
general models of electrical faults. Detection, failure analyses, 7. Improper mold or adhesive processing
and a novel method of freezing to determine defect location are 8. Contaminated insulation or jacket
also presented. 9. Unbonds, discontinuous bonding
10. Contaminated overmold, adhesive interfaces
I. INTRODUCTION AND BACKGROUND 11. Design flaw
12. Interface imperfection
We have observed and performed immersion tests using 13. Defective microstructure, nonuniformity
megohmmeters on cable systems and sonar sensor arrays in 14. Material composition variations
simulated seawater (SSW) to detect and locate extremely small 15. Trapped solvent
electrical leak paths, faults, and defects that freshwater (FW) tests 16. Creep, cutting, and contact
cannot. Hypotheses and likely mechanisms for the increased 17. Branching or carbon particle/flake networks
sensitivity when using SSW vs. FW are proposed here. Internal part displacement in connectors, or short-term creep
To reduce the risk of system problems or mortality during and induced electrical or optical problems may also occur and be
after deployment, there is a critical need to conduct hydropressure detected.
as well as preliminary immersion tests, especially for state of the art Many costly lessons have been learned that prove the necessity
cable systems and sonar arrays with splices and connectors. Testing and value of electrical IR, operation, and pressure testing and
is particularly important in new or unproven designs, especially for exercising of cable systems in a medium with identical or more
fractional kilometer-length arrays with multiple inline components challenging chemical and physical characteristics than the service
such as sensors, connectors, splices, varying cable diameters, environment. Experience has shown that testing of a cable system
overmolds, interfaces etc. Testing of the final assembly, which may and a sonar array that was intended for seawater service in fresh
contain repairs or alternate materials that are incorporated during water can lead to false positive results. An array tested in fresh
production, is imperative, especially if the system must work water passed an acceptable limit of about 20 megohms insulation
“perfectly” on the first deployment. resistance; the same test in SSW resulted in considerably lower
insulation resistance—well below the acceptable limits. Detection
II. TEST DESIGN CONSIDERATIONS of such faults and defects prevented an almost certain critical
in-service failure.
Typically, a nominal or minimum electrical insulation resistance When measuring IR with VOMs (volt-ohm-milliammeter
(IR) between the cable conductors and test chamber electrolyte or ohmmeters) and megohmmeters, the conductive test medium needs
wall is determined or set jointly by the designers and customers. We to have the least electrical resistance (highest electrical
find it better to set the limit for IR for the service environment [1, 2]. conductance) and service composition to insure the highest
For example, 20, 40, 60, or preferably 100 megohms of resistance sensitivity to leaks and faults in a reasonable amount of time with a
can be set as the lower limit in SSW. IR readings below the suitable test voltage. Consequently, accurate fault detection and
specified values are expected to be caused by electrical faults. These evaluation of marine cable systems and sonar arrays requires a test
values should be specified taking into consideration circuitry design environment with elevated pressure and voltage factors in a
requirements, experience with cable reliability, quality, life chemically conductive medium with the same composition as that in
expectancy, risk and consequence of infant mortality, molding which the system will be deployed.
processes, manufacturing maturity, possible innate defects, material Although the bulk conductivity and composition of SSW vs. FW
concerns, state of design development and history. Joint, splice, or may not seem significant for IR testing, they are essential for
repair processes and actions need to be stressed by such a test maximum sensitivity and duplicating the aggressive service
because this is the last and final measure to reveal problems before environment during testing. When the possible electrical faults may
acceptance and service. involve microscopic leaks along long, thin, and narrow leak paths,
A reasonable SSW immersion test of electrical IR and diffusion along paths across mold/bond lines, through contaminated
electro-optics system function is twenty-four hours at maximum test dielectrics or charge alignment across imperfect or faulty insulting
pressure after stepping up or cycling up and down to pressure while layers, the highly ionically conductive dielectric medium provided
operating the system and periodically measuring IR of all cable by SSW enables the detection of small faults that might otherwise

Authorized licensed use limited to: M S RAMAIAH INSTITUTE OF TECHNOLOGY. Downloaded on April 24,2010 at 07:12:01 UTC from IEEE Xplore. Restrictions apply.
be undetected. electrode. The galvanic cell is often that produced between the
Early deployment (infant) mortalities in the form of leaks and conductor (often copper alloy), the tank electrolyte (more reactive if
low IR tend to show up in such a test, especially if the test pressure SSW), and the tank wall (often unpainted steel). Examples of leak
exceeds the expected service pressure by a factor of about 1.5 or 2 paths are cracks, pinholes, etc., as previously listed. Any and all
and the megohmmeter voltage exceeds the expected voltages by a leakage contributions/effects are additive and enhanced by the
factor of 5, 100, or more. Such pressure over short times should relatively large surface area of the cable and vessel wall. Fault
produce leaks and creep of plastic insulation or of molded insulation model 1 is best, and at times only, detected by testing in SSW.
over conductors that are likely to occur, causing electrical resistance Fault model 2 is not a leak path via SSW through diffusion or
to decrease to a level that would show early service failures. Cable actual liquid path but by design has a continuous dielectric (typical
systems should be exercised or flexed prior to or during IR electrical insulating material such as jacket or plastic) encasing all
immersion testing to open and expose weak bonds, interfaces, seals, metallic conductors; but the encasement is poor (defective) or
joints, or cracks. Such flexing can open weak and tight cracks to imperfect in terms of continuity, uniformity, thickness, composition,
allow better SSW diffusion or leaking and make the test represent contamination, cohesion, or microstructure. Essentially model 2
realities of deployment and service. faults in SSW might be modeled (thought of) as an imperfect/real
Due to increased costs and potential damage to pressure vessels dielectric whose collective surface, interfaces, and interior that have
by corrosion, there is management and engineering reticence to use SSW ions diffused, forced, or migrated into them. An existing or
SSW in hydropressure, cyclic, or stepped testing. However, the high applied voltage helps diffusion and migration of ions. In certain
to unacceptable cost of service failures often outweighs the people cases the collective jackets may be looked at as a “leaky” dielectric
power and facilities costs of testing. Lessons of corrosion science of a pseudo capacitor. Examples of such model 2 faults would be
can be used to reduce corrosion and contamination of test facilities too thin, poor quality, or damaged insulation jacket, contaminated
and apparatus. Proven methods can reduce corrosion, fouling, and bond lines, poor bond interfaces/lines, or trapped foam or bubble
contamination of pressure test systems [3–11]. paths in insulation, overmolds, or splices. Further, the mechanism of
If natural seawater (NSW) is the expected service environment, conduction by “branching” in plastics or carbon particle/flake
then NSW or SSW should be the test electrolyte. The SSW should networks in rubbers can also be an example of fault cause. Any and
be similar to or comply with ASTM D1141-98(2003) Standard all fault effects tend to be additive and probability of occurrences
Practice for the Preparation of Substitute Ocean Water or be made increases with the immersed length and surface area of the cable.
from proper Cargill sea salt to about 50 mS/cm conductivity or the Interestingly, seawater itself can be characterized as an ionically
conductivity of the service environment [12, 13]. conductive dielectric. Fault model 2 is also best, and at times only,
Not only is the SSW test more sensitive to defects, faults, or detected by testing in SSW.
marine service performance but it is the ONLY test that can give the
highest assurance that the system will work in the expected service IV. TESTING WITH SSW VS. FW
environment, with the required IR behavior. A SSW hydropressure
IR and function test is the best protection against infant and later The conductivity and behavior of SSW vs. FW is not negligible.
mortality. There is no substitute or alternative final acceptance test The differences between hydropressure and/or soak testing in SSW
than the SSW IR hydropressure test to minimize risk. An extended vs. FW seem negligible according to the following argument:
SSW soak IR test following cable exercising and hydrostatic Typical NSW and proper SSW conductivity is about 50 mS/cm.
function testing is the next best step to reduce risk. SSW measured and calculated resistance can be about 0.8 ohms
between the tank/vessel walls and a “typical ohmmeter probe.”
III. TWO FAULT MODELS (This practice is not recommended.) Similarly NW (natural water)
TW (tap water) or FW resistance can be about 80 ohms. Hence, the
There are two definable and detectable kinds, based on test difference between 0.8 ohms and 80 ohms relative to 20 Mohms
response and cause, of marine cable electrical faults, defects, low IR, (megohms) or 100 Mohms limits is negligible.
and erratic electrical readings due to manufacture, material, or This argument, however, does not take into account the dynamic
design problems of interest here. Often, they are detected by and reactive behavior of SSW vs. FW on test surfaces during
different procedures during immersion and hydropressure testing. hydropressure and/or soak testing when using ohmmeters and
Here, the term “fault” refers to an unacceptably low or erratic meggers. Such meters impose a voltage gradient across the
(drifting, unstable, or non-repeating) IR or unacceptable voltage electrolyte that produces dynamic changes in the electrolyte, cable
readings between internal cable conductors, strength members, or interfaces, vessel, or electrodes. Polarization effects are complicated
system metals and the vessel wall or alternate electrode. by the fact that several processes act in concert in the electrolyte and
Fault model 1 is at least one water leak path (leak rate from interfaces. The polarization effects for SSW are much different than
diffusion up to at least a thin but continuous water leak path) those of FW, due to the incredible difference in number,
between one or more cable “metallic” conductors and/or the outside composition, type of ions and neutral/uncharged particles in these
liquid in the test tank. By definition, this fault model can allow at electrolytes. Further, the metal and even nonmetal interfaces can
least a finite amount of DC current to flow from the conductor or change by corrosion and/or charging reactions (tend to be dynamic)
armor through the electrolyte in the leak, bulk electrolyte, and to the depending on which electrolyte is used. Simply stated, SSW
wall or electrode. Essentially model 1 fault in SSW may be viewed behaves like a very reactive and corrosive electrolyte in a resistor
as a weak battery (galvanic cell), which can be in series or parallel and pseudo electrolytic capacitor circuit and at times like the
“or concert” with a “leaky” electrolytic capacitor. The conductor is electrolyte in a galvanic cell in a cable with huge surface areas. [2–5,
one plate, the cable jacket is the normal dielectric, the SSW is an 14, 15]
ionically conductive electrolyte between the jacket and wall or SSW can create galvanic cells. As the classic SW galvanic series

Authorized licensed use limited to: M S RAMAIAH INSTITUTE OF TECHNOLOGY. Downloaded on April 24,2010 at 07:12:01 UTC from IEEE Xplore. Restrictions apply.
clearly shows, the galvanic cells between conductors (engineering measured with typical ohmmeters where all other conditions such as
metals) and pressure vessel metals can develop voltages of up to temperature, area, humidity and composition are constant. However,
about 1.5 volt (1,500 mV). As a FW “galvanic” series would show, conductivity (inverse of resistivity) in electrolytes between reactive
FW cells could exhibit nil to maybe a small fraction of what is electrodes can be thought of as dynamic conditions (multiple valued,
developed in SW cells. drifting and somewhat unrepeatable/irreversible) since they can
SW is very reactive and corrosive compared to FW especially if change due to chemical, electrochemical, and physical factors as a
there is dissolved oxygen and more so under applied voltages. SW result of the typical charging and reactions. Because
corrosion cells, at microscopic leaks with any thin layer of SW, act electrochemistry is at work when voltage is applied across SSW or
like batteries that can be charged and discharged. SW, as a dielectric some FW or interfaces, the measured resistances, voltages, charges,
solution between two plates (metals such as conductors, armor, or and currents tend to be dynamic, much more so in SSW. Hence
conductive/contaminated plastic insulation layers), is capable of standard meters such as YSI Model 63 must be used to measure
developing substantial charges compared to FW, by reactions and conductivity of SW.
ion concentrations across leaks or leaky capacitors, when a voltage Further, because most VOMs ohmmeters apply a voltage (on
field exists or is applied. order of 1 to 3 volts DC or a fixed current) across the probes, any
Composition and conductivity of waters vary. Detection of these corrosion cell, or leaked SW in that test circuit is bound to react and
two fault models depends primarily on the types and degrees of show a dynamic or erroneous reading for pure resistance. On
cable system perfection or faults from manufacturing, material, reversing the polarity of the ohmmeter, the reading will often differ
prior exercising, test pressure, times and instruments, cable jacket by a factor of ½ to 2 or go off scale, read overload, etc. When SSW
IR magnitude, electrolyte leakage, and electrolyte conductivity and is in the test circuit, the ohmmeter reading will often be unsteady
composition. Typical SW is 30 to at least 550 more conductive and drift up or down depending on the polarity and conditions.
than most NW, FW, or TW and tends to be more reactive with any If any corrosion cells exist in the circuit a typical volt meter will
metallic electrodes. Water such as TW or FW from different sources read a few to hundreds of millivolts. Readings will often be
differs greatly in conductivity and is neither predictable nor easily unsteady depending on fault metals, size, dissolved gases, and
reproducible. Pure water is practically nonconductive. Standard composition as well as the polarity, time, and internal resistance of
SSW can be made with consistent, reproducible, and known the voltmeter. Charge can exist from previous ohmmeter
conductivity and composition; hence, a more reliable test solution. measurements or other causes. Discharging can take time (seconds,
SSW ion composition and movement are important. Because real minutes to hours). Corrosion (change in chemistry at the interface,
cable jackets, insulating layers, joints splices, materials, mold lines, polarization, and passivation) can occur and prevent reproducible
O-rings, seals, and repairs are not perfect, ionic diffusion, especially results.
by applied voltage, into and even through such dielectrics or Meggers can be used to detect the dynamics of SW charging and
features, is possible. Further, no matter how perfect a real cable reactions. The charging times, discharging times during readings
outer jacket and surface, there is always a finite leakage current (time constants), drift directions and times and magnitudes can be
through the jacket under these applied voltages. These cable surface informative. Such charging and reactions differ for NSW or SSW vs.
and “bulk” effects are expected to be more dynamic and reactive in for FW.
SSW than in FW. Hence, the magnitude of the differences in Faults can disappear or become undetectable. After removing
megger readings between SSW and FW in cable tests is not cable components from immersion, some microscopic leaks and
expected by these authors to be equal to the measured differences layers of SW or SSW can evaporate naturally or be boiled off by
between resistivities. However, if the cable has an IR on the order of heat or reactions due to current flow. The latter can occur even
gigaohms to infinite by megger reading (ideal cable condition), then during immersion if the volume of electrolyte is microscopic. Once
a difference in IR between tests in SSW and FW at low applied dry by evaporation and perhaps temporarily sealed by salt deposits,
voltages is not expected to differ much, if at all. Hence, if the cable and then testing as high/good IR, it may become difficult to
is real, testing in SSW is critical; but, if the cable is perfect, results reestablish the leak and detect the fault. Re-immersion, pressure,
in SSW vs. FW are equal. and time may help. If such faults must be preserved for failure
VOM and megger readings tend to be dynamic in SSW. analysis by autopsy and testing, then prevent evaporation by wet
Electrolytes do not conduct electricity in the same fashion as metal bagging, avoiding prolonged current flow, and by limiting current
conductors. Metals conduct electricity by motion of “free” electrons flow through the fault while testing post immersion.
due to a voltage difference. Electrolytes such as SSW and
contaminated FW conduct electricity by motion of ions. Ions can V. FAILURE ANALYSES EXPERIENCE AND SSW IR TESTS
exist in an electrolyte by its ionization constants at given
temperatures, pH and concentrations, or by reactions at metal or Fault locations tend to be repetitive. Historical service
conductor interfaces. Positive ions are generated (go into solution) experience and failure analyses have clearly shown that repairs and
at anodes and positive ions such as H+ can be neutralized at reworks are the most likely locations for defects or faults. In service
cathodes by given applied voltages as long as appropriate currents faults occur from tension, compression, impacts, fish, abrasion,
can flow and charges can be developed and balanced. Negative ions cutting, fishing gear, plows, anchors, etc. In production, new or
such as OH– can be generated at cathodes. Typical and classic changed processes, materials, and personnel/operators are likely
corrosion or electro-chemical reactions are also possible at these causes of imperfect or defective cable systems and hence faults.
electrodes [3–11]. Problems tend to be more prevalent at connectors, terminations,
Interestingly, measured resistance at relatively low current flow penetrations, splices, joints, stress concentrations, section changes,
under applied voltage across metals such as typical resistors can be and material changes. Hence, best practices require that while cable
thought of as a relatively static or single-valued quantity when systems are in the acceptance and manufacturing stages, repair

Authorized licensed use limited to: M S RAMAIAH INSTITUTE OF TECHNOLOGY. Downloaded on April 24,2010 at 07:12:01 UTC from IEEE Xplore. Restrictions apply.
procedures and quality control measures at these locations be procedures for ohmmeter and megger measurements.
tracked and carefully documented. These measurement records have 5) Measure resistance with ohmmeter between all safe (to system
proven invaluable even when manufactures and operators are top and meter) combinations of conductors to each other, SSW, and
quality and diligent, as is most often the case. vessel.
Experience has provided the following “lessons learned” 6) Measure IR with megger at appropriate voltage between all safe
concerning manufacturing conditions. Rushed, stressed, behind combinations of conductors to each other, SSW, and vessel.
schedule or less trained/skilled personnel/operators tend to be more Always follow proper and safe (to personnel) megger
likely to make mistakes or create cable quality problems. Best procedures, including discharging procedures.
environments are those where operators are highly trained, 7) Repeat voltmeter mV measurements and procedures.
re-trained often, recognized, and rewarded for good quality products 8) Always wash all residual SSW off of tested cables, arrays or
as well as reporting evidence of defects or needs for review, systems immediately after testing—before any evaporation can
redesign, or repair. Close cooperation and continuous inspections by occur. Flushing with FW by hose, immersion in FW tank, and a
quality control or assurance personnel makes the most reliable repeat FW flushing achieves most cleaning objectives. Once salt
product. Written, reviewed, approved, up-to-date, and followed is deposited by evaporation, removal of deposits becomes
manufacturing procedures, processes, material control, and difficult. Deposits can contaminate other surfaces and increase
inspections are critical to producing the most reliable product. Open, abrasiveness of cable surfaces.
cooperative, competent, and active work/communications between 9) Ensure that before and during any measurement connections are
material control, quality control, and engineering are critical made that all terminals, clips, surfaces, and contacts are clean
ingredients for consistent quality product output and absence of and dry. Avoid wet or humid conditions. Lowest relative
undetected, unreported, or unrepaired defects. humidity is best.
Experience has also provided the following “lessons learned” Detection of model 1 faults. These faults are easiest to detect and
regarding raw material control, electronic component control, can be detected by typical VOM ohmmeter or volt meters during
manufacturing environmental control, and vendor issues. Careful immersion or hydropressure tests. Obvious or blatant leak model 1
and continuous testing, tracking, handling, and storage of raw faults may be detected in FW, though all but the most obvious are
materials and components are critical. All parts and rooms need to easier to detect in SSW. Some faults may not be detected in FW
be monitored continuously and kept clean, at low humidity (< 50% with ohmmeters or voltmeters.
RH), and at temperatures well above (e.g., 10ºF >) the dew point. The typical ohmmeter cannot correctly read the resistance
Specifications for vendors, vendor performance, and consistency directly if there is a galvanic cell in the circuit, because existing
need constant attention, reviews, and periodic updates. Unknown galvanic cells will bias the ohmmeter reading to give a higher or
changes in any ingredients or out-of-specification conditions can lower than actual resistance depending on the polarity of the cell
lead to serious cable system problems. and the ohmmeter applied voltage or set current. By measuring such
Importantly, the most revealing time to detect and observe faults in both polarities, the existence of such a fault through a
quality or the lack thereof, in the form of accidental conditions or seawater path can be detected or proven. The locations of such
imperfections, is during manufacture before the final jackets are faults can be determined by slowly removing, immersing, or wetting
closed up. the cable while making the readings.
Simplified SSW IR Test Steps: Detection of Model 2 faults. These faults may only be detected
1) Prepare SSW to meet the recommended composition and when the electrolyte is SSW. Even when testing in SSW, a megger
conductivity standards [12, 13]. Check conductivity with the may be required to detect such faults. In any case such faults are
YSI model 63 conductivity meter, or equivalent, before, during, more easily detected when tested at maximum allowed voltages
and after cable system test. (such as 50, 100, 500 or higher VDC) and when the cable is
2) Perform and record baseline measurements for all IR variables immersed in a SSW electrolyte as opposed to FW.
before immersion. Prior to or during testing, exercise cable. Higher test pressures in SSW typically increase the likelihood of
3) Immerse as much cable system as possible into SSW vessel. detecting such faults unless, as in some cases, the crack or leak path
Most stringent tests use pressure cycling or at least stepping becomes closed or sealed by higher pressures. Faults may not
during hydropressure or SSW soaking for 24 hours, overnight, develop or be detected until the pressure reaches a certain value or
or several hours. Measure and record initial, during, and end is cycled or causes displacement or creep, which makes contacts or
immersion mV and IR test readings. Periodic system operation closes a conductive circuit within the cable or system. In one case, a
during hydropressure testing is recommended. stiff copper center conductor, which was terminated by cutting with
4) Measure voltage (mV) between all combinations of conductors to dikes at each overmold breakout in a cable array, did not cut into the
each other, SSW, and vessel or alternate immersed electrode. insulation and make contact with other conductors until higher
Read and record voltage polarity (voltmeter terminal to which pressures (near service pressures) were applied. When the pressure
conductor), sign, and actual value (not pass/fail) at first/initial was reduced relaxation occurred, the contact was opened and the
instantaneous reading and that after 15 seconds to the nearest fault was not detected, until the pressure was again increased. Again
mV (sensitivity), drift direction, rate or magnitude, and time to exercising/flexing helps reveal weak imperfections.
stabilize. Make and record the same voltage measurements in Impedance (EIS = electrochemical impedance spectroscopy) as
reverse meter polarity. If results show any significant reversal on well as IR testing of jackets on small-diameter cables has proved
voltage sign or unacceptable magnitude (often a few to several valuable in a previous cable and jacket development program. As
mVs), then track and determine cause. Remedy cause as expected, the impedance of the tested thin cable jackets in these
required. Repeat measurements. Use the same measurement, short test cables dramatically decreases with higher test frequencies
reading, reversal, recording, report, and documentation and to a lesser degree with time of immersion [2]. EIS evaluation of

Authorized licensed use limited to: M S RAMAIAH INSTITUTE OF TECHNOLOGY. Downloaded on April 24,2010 at 07:12:01 UTC from IEEE Xplore. Restrictions apply.
coatings is under development at ASTM and ISO [16]. Tests for
effectiveness, condition, and aging of cable jackets may be applied
or compared with methods to test coatings or paints. The impedance
or capacitance of coatings is used to evaluate their behavior and
aging when exposed to various stressful service environments.

VI. TEST INSTRUMENTS

1) GR-1864 Megohmmeter serial #1864-9700-00; Instruction


Manual Form 1863-0100-00; General Radio 1978, Concord,
Massachusetts, USA
2) High impedance Volt Ohmmeter (modern: typically measure
voltage drop across unknown IR for a set current flow – about 1
microampere); Example: Fluke 87 Series V; Industrial Fig. 2. Autopsied cable with pencil pointing toward the electrical fault and
True-RMS; Instruction Manual PN 2101973, May 2004 galvanic cell at the defective repair.
3) Simpson VOM (older: typically measures IR by leakage current
through unknown IR at known applied voltage)
4) Handheld pH, Conductivity, Salinity and Temperature System;
YSI Model 63
Always refer to and follow instruction manuals to be sure safe
and proper operating techniques are used with all meters.

VII. NOVEL, SIMPLE, AND LEAST DESTRUCTIVE METHOD


TO LOCATE AN INTERNAL SW LEAK

A cable more than several kilometers long (Fig. 1) [17] had a


suspected seawater leak across the internal insulation between a
copper conductor (designed to carry 3 KV) and the outer steel shield
(bronze counter-laid strip design at test time). After some time of
deployment in the ocean, diminishing IR between the center
conductor and the ocean/Earth indicated an electrical problem. At
that time the “suspected” SW leak and “short” was claimed by some
to be caused either by service exposure/handling or fish bite, and it
was claimed that the problems, leaks, shorts, or damage “could not Fig. 3. Closeup of defective and leaking electrical insulation repair
be detected” without extreme (unacceptable) cost and extensive under the shield.
destructive testing.
The SW connection (leaking defect) between the copper
conductor and the outer shield showed clear symptoms of being a
galvanic corrosion cell. It was detected by measuring a few hundred
millivolts in one direction and a negative few hundred millivolts in
the opposite polarity, and by showing about 15 K ohms in one
polarity and 7.5 K ohms with a Simpson VOM in the opposite
polarity between the conductor and shield. The voltage and IR were
both unstable and drifted over the time of measuring by at least
+/– 10%. The ohmmeter reading charged the cell and caused the
typical dynamic mV and IR readings.
First defects or penetrations in the outer jacket were located and
marked along the very long cable by using a special “paint holiday
detector.” The holiday detector was connected, via a commutator on
the take-up reel, between the center copper conductor and a wet
sponge in contact with the outer jacket of the cable during recovery
from the ocean. Several locations where the cable outer jacket had
penetrations through the outer insulating jacket to the outer shield
were detected. The cable was cut in half and the half containing the
SW leak to the copper was determined by voltmeter and ohmmeter
readings.
The SW leak to the center conductor was neither under nor near
any of the detected penetrations in the outer jacket. That
determination was reached by quickly opening and drying the outer
jacket and shield at the few penetrations into the outer jacket over
Fig. 1. Longitudinal cut-away section of submarine cable.
the half cable length that contained the SW galvanic cell and low IR

Authorized licensed use limited to: M S RAMAIAH INSTITUTE OF TECHNOLOGY. Downloaded on April 24,2010 at 07:12:01 UTC from IEEE Xplore. Restrictions apply.
(“short”). It was surmised that SW had leaked through the outer SSW and FW at low applied voltages is not expected to differ
jacket and under the bronze tape shield, to the insulation, and then much if at all.
hosed long distances to some defect in the thick internal insulation. 5) Freezing the NSW or other electrolyte in a corrosion cell
The challenge was to find that defective insulation without reduces or stops corrosion current flow, decreases conductivity
removing the outer jacket over the entire length in the least and reduces or suppresses galvanic voltage; hence, freezing
destructive fashion possible. We successfully tried our novel selected lengths of a cable, while monitoring IR and/or mV,
technique of locating the defect by locally freezing its electrolyte containing a galvanic cell due to leaking can help locate the leak
and observing the changes in readings. The defect was located site. Alternatively, when the electrolyte in a corrosion cell
exactly (within 3/8 inches) in the most suspect 100 foot length by freezes, no typical corrosion reactions can occur; hence, the cell
the novel method of freezing the cable over successive 6 inch voltage approaches zero and the IR becomes that of frozen SW.
lengths until the SW in the leak was frozen and hence reduced the 6) Because water such as TW or FW from different sources varies
corrosion cell voltage to near zero! greatly in conductivity they are neither predictable nor easily
Once located and carefully autopsied, the defective and leaking reproducible. Pure water is practically nonconductive. Standard
insulation repair was obvious (Figs. 2 and 3). It was about 1 inch SSW can be made with consistent, reproducible, and known
long, contained SW and corrosion products. A hair, probably human, conductivity and composition; hence SSW is more reliable as a
could be seen sticking out from the repair (arrow, Fig. 3). The test solution.
manufacturer used records to trace the time of that repair to a period
when management personnel were working the cable production Acknowledgments
line and doing the repair due to an operator absence by vacation or
strike. Technical comments and reviews from Dr. Robert Ruedisueli (U.S.
Navy, NAVSEA, Carderock Division), Dr. Thomas Sanford, Russell
VIII. FUTURE WORK Light, and Jim Carlson (all APL-UW) are greatly appreciated.

1) Statistics are needed to determine the tradeoff involved in IR REFERENCES


SSW soak testing in concert with SSW and/or FW
hydropressure testing to reduce costs and corrosion risks. [1] Sandwith, C.J., E.R. Dunston, and A.A. Sieger, Analysis of an
2) Define effects on tested surfaces and interfaces by concentration, underwater hydrophone recovered from the Barking Sands
activation, and resistance over voltages [2, 8]. Tactical Underwater Range after 21 years of service,”
3) Define effects of double layer capacitance, time of charging Proceedings, Oceans ‘90, Washington, D.C., 1990.
megger, rate of voltage drop by meter read switch and [2] Sandwith, C.J. and R.L. Ruedisueli, Corrosion and aging
voltage/charge relaxation time [2, 8]. tests—via measurements of insulation resistance, impedance,
4) Use AC as well as DC instruments and tests to provide more and electrochemical noise—on jackets of small-diameter,
informative IR, impedance, and EIS diagnostics. Tests for armored, fiber-optic cables with and without simulated
condition and aging of cable jackets may be compared with test biofouling damage, Proceedings, Ocean Community
methods for coatings or paints. The difficulties of transmitting Conference ’98, November 16–18, Baltimore, Maryland
high-frequency AC signal down a long cable must be considered. [CD-ROM], Marine Technology Society, Washington, D.C.
Perform circuit analyses on probable fault scenarios to [3] Sandwith, C.J., and R.K. Kent, Velocity-affected corrosion, in
determine best and expected test results, time constants, and Failure Analysis and Prevention, Vol. 11 of the ASM
other symptoms [10, 15]. Handbook series, edited by W.T. Becker and R.J. Shipley,
788–795, Materials Park, OH, ASM International, 2002.
IX. CONCLUSIONS [4] Shreir, L.L. Corrosion, Newnes-Butterworths, London and
Boston, 1979.
1) Not only is the SSW test more sensitive to defects, faults or [5] ASM Handbook, Ninth Edition, Volume 13: Corrosion, ASM
marine service performance but it is the ONLY test that can give International, 1987.
the highest assurance that the system will work in the expected [6] Sandwith, C.J., and R.L. Ruedisueli, An experiment to reduce
service environment, with the required IR behavior. As with corrosivitiy in submarine sonar domes by nitrogen injections,”
checking each light bulb, a SSW hydropressure IR and function Proceedings, Oceans’95 MTS/IEEE Conference, 1134–1138,
test is the best protection against infant and later mortality. 1995.
There is no substitute or alternative to minimize risk. An [7] Sandwith, C.J., R.L. Ruedisueli, J.P. Papageorge, and B.A. Eng,
extended SSW soak IR testing following hydrostatic function Initial results of monitoring corrosivity in submarine sonar
testing is the next best step to reduce risk. domes: The Dome Water Monitoring Instrumentation System
2) SSW or NSW as solutions for hydropressure and/or soak tests (DWMIS) Experiment, January - August 1993, Proceedings,
provide a more realistic and demanding test of electrical IR, Oceans’96, MTS/IEEE Conference, Brest, France, 1994.
sealing, and leaking defects in electro-optical or mechanical [8] Sandwith, C.J., R.L. Ruedisueli, K.G. Booth, J.P. Papageorge,
cables than does FW. and B.A. Eng, Monitoring corrosion in submarine sonar domes,
3) IR testing in FW can give false passing readings for both model Proceedings, OCEANS ’93 MTS/IEEE Conference, Washington,
1 and 2 faults. D.C., 1993.
4) If a cable system were to exhibit an IR on the order of gigaohms [9] Sandwith, C.J., R.L. Ruedisueli, and K.G. Booth, Seawater
to infinite by megger readings, then a difference in IR between monitoring instrumentation system,” Proceedings, MTS ‘92,
Washington, D.C., 1992.

Authorized licensed use limited to: M S RAMAIAH INSTITUTE OF TECHNOLOGY. Downloaded on April 24,2010 at 07:12:01 UTC from IEEE Xplore. Restrictions apply.
[10] Fontana, M.G., Corrosion Engineering, McGraw-Hill, 1986. [15] Sandwith, C.J., R.L. Ruedisueli, A.L. James, and G.E.
[11] Boyd, W.K., and F.W. Fink, Corrosion of Metals in Marine Gotthardt, Resistance of steel strength wires used in small
Environments, Metals and Ceramics Information Center, fiber-optic cables to seawater corrosion, Proceedings,
Columbus, OH, 1978. Oceans'96 MTS/IEEE Conference, Fort Lauderdale, Florida,
[12] Standard Practice for the Preparation of Substitute Ocean Vol. 1, 511–516, 1996.
Water, Document Number: ASTM D1141-98e1, ASTM [16] Loveday, D., P. Peterson, and B. Rodgers, Evaluation of
International, 2003. organic coatings with electrochemical impedance spectroscopy,
[13] See the Cargill, Inc., website at: Part 2: Application of EIS to coatings, JCT Coatings Tech,
http://www.cargill.com/index.htm October, 2004
[14] Sandwith, C.J., and R.L. Ruedisueli, Corrosion resistance of [17] Sandwith, C.J., W.D. McCormick, J.A. Thornton, D.R. Wise,
certain strength wires used in small fiber-optic cables exposed and R.I. Odom, Demonstration of fiber-strain measurement
to seawater,” Proceedings, Oceans ’97 MTS/IEEE Conference, using a Brillouin optical-fiber time-domain analysis system,
Vol. 2, 957–963, 1997. Proceedings, Ocean Community Conference ’98 [CD-ROM],
MTS, Washington, D.C., 1998.

Authorized licensed use limited to: M S RAMAIAH INSTITUTE OF TECHNOLOGY. Downloaded on April 24,2010 at 07:12:01 UTC from IEEE Xplore. Restrictions apply.

También podría gustarte