Está en la página 1de 172
SERVICE MANUAL ISOLETTE Infant Incubator From Hill-Rom Ait-Shields Products: C400 QT™/C450 @T™ and CE VERSIONS Paris Or Technical Assistance USA (800) 445-3720 Canada (800) 267-2337 Intemational: Contact your distributor. man203 Hill-Rom Air Shields. IsTRUCTIONS REASON FOR CHANGE INSERT THIS SHEET INTO THE MANUAL POWER SUPPLY CALIBRATION. DESCRIBED BELOW. PROCEDURE UPDATE: EQUIPMENT: [colt Infant incubators, Models C400 aT and c¥s0 GT MANUAL TITLE: Service Manual MANUAL CAT. NO: 58 994 50-7 (man20s) ‘ADDENDUM INFORMATION PARAGRAPH 5-9 CALIBRATION. ADD THE FOLLOWING: + ©400.Contoters Only Reeve the Msn Soar rom the Front Paral (CHANGE PARAGRAPH 5.3.2 POWER SUPPLY (REFER TO FIGURES 51 AND 52 FORLOCATION OF ‘TEST POINTS AND ADJUSTMENTS) AS FOLLOWS: +. Connact®ne Corrales the variable ransom. Sothe wanformecto 100, 420, 200.240 nominal ‘1.0 Vac, Check yout Dat Tag forthe proper Ine vaage. Connect fe DVM betwean TPS and TPS (GND) onthe Na board, 2. Reterto paragraph S23, Steps A and, iirwoke th Of-Line Dingess and select DiagnecicTost Number PROCEDURE 1, Ast 1 on the Power Supply Board fo a roading 6400 = 0.200 NOTE: Tre fa coarse adustment on. 2, Cheokethet the Skin ial rads 00 = 0.06 nocessay, reeds AT as ecures ‘TABLE 68 CONTROLLER ASSEMBLY, GROUP 1, PARTS LIST ‘Change Part Nunber of tem 47, Ar Flow Thermisor Assembly, or 68 2:4 841 68 214 26 ‘TABLE 69 CONTROLLER ASSEMBLY, GROUP 2, PARTS LIST Change Part Number of em 17, Air Flew Themistr Assembly, or 68 21488 to 68 214 84 (Gat No, 68 994 £0-7A (ian203) Page ft Pred in USA 1/00