Está en la página 1de 23

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles

for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Designation: D3606 − 17

Standard Test Method for


Determination of Benzene and Toluene in Spark Ignition
Fuels by Gas Chromatography1
This standard is issued under the fixed designation D3606; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1. Scope* priate safety, health, and environmental practices and deter-


1.1 This test method covers the determination of benzene mine the applicability of regulatory limitations prior to use.
and toluene in finished motor and aviation spark ignition fuels 1.8 This international standard was developed in accor-
by gas chromatography. This test method has two procedures: dance with internationally recognized principles on standard-
Procedure A uses capillary column gas chromatography and ization established in the Decision on Principles for the
Procedure B uses packed gas chromatography columns. Pro- Development of International Standards, Guides and Recom-
cedures A and B have separate precisions. mendations issued by the World Trade Organization Technical
Barriers to Trade (TBT) Committee.
1.2 Benzene can be determined between the levels of: (1)
Procedure A 0.1 % and 5.0 % by volume and toluene can be 2. Referenced Documents
determined between the levels 0.4 % and 20 % by volume, and 2.1 ASTM Standards:2
(2) Procedure B benzene can be determined between the levels D4057 Practice for Manual Sampling of Petroleum and
of 0.1 % and 5 % by volume and toluene can be determined Petroleum Products
between the levels of 2 % and 20 % by volume. D6299 Practice for Applying Statistical Quality Assurance
1.3 The precision for this test method was determined using and Control Charting Techniques to Evaluate Analytical
conventional spark ignition fuel (motor ‘gasolines’) as well as Measurement System Performance
those containing blended ethanol and such ethers as MTBE, E288 Specification for Laboratory Glass Volumetric Flasks
ETBE, TAME, and other alcohols such as methanol and E694 Specification for Laboratory Glass Volumetric Appa-
butanols. This method may be used for finished motor and ratus
aviation gasolines that contain ethanol blend up to 20 % and E969 Specification for Glass Volumetric (Transfer) Pipets
butanols blend up to 20 %. E1044 Specification for Glass Serological Pipets (General
1.4 M85 and E85 fuels were not included in the precision Purpose and Kahn)
for Procedure A. M85 and E85 fuels cause interference with E1293 Specification for Glass Measuring Pipets
Procedure B.
1.5 Procedure A uses MIBK as the internal standard. Pro- PROCEDURE A—CAPILLARY WCOT GAS
cedure B uses sec-butanol as the internal standard. The use of CHROMATOGRAPHIC SYSTEMS
Procedure B for fuels containing blended butanols requires that
sec-butanol be below the detection limit in the fuels. 3. Summary of Test Method
3.1 An internal standard, methyl isobutyl ketone (MIBK) is
1.6 The values stated in SI units are to be regarded as
added to the sample which is then introduced into a heated
standard. The values given in parentheses are for information
capillary or programmed temperature vaporization (PTV) in-
only.
jector on a gas chromatograph (GC). The GC is equipped with
1.7 This standard does not purport to address all of the two columns connected in series. The sample passes first
safety concerns, if any, associated with its use. It is the through a column with a nonpolar phase. After toluene has
responsibility of the user of this standard to establish appro- eluted, the flow through the nonpolar column is reversed,
flushing out the components heavier than toluene. The toluene
and lighter components enter a second column which separates
1
This test method is under the jurisdiction of ASTM Committee D02 on
Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of
2
Subcommittee D02.04.0L on Gas Chromatography Methods. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Dec. 1, 2017. Published April 2018. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1977. Last previous edition approved in 2010 as D3606 – 10ε1. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/D3606-17. the ASTM website.

*A Summary of Changes section appears at the end of this standard


Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
1
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17
the aromatic and nonaromatic compounds. The eluted compo- 5.7 Flasks—Volumetric, 25 mL and 100 mL capacity (see
nents are detected by a flame ionization detector (FID). The Specification E288).
detector response is recorded, the peak areas are measured, and 5.8 GC Carrier Gas—Helium or hydrogen 99.999 % pure.
the concentration of each component is calculated with refer- (Warning—Compressed gas under high pressure; hydrogen
ence to the internal standard. flammable.) A hydrogen generator may be used.
NOTE 2—When using hydrogen carrier gas, use precautions such as
4. Significance and Use installation of hydrogen sensors in the gas chromatograph oven. It is
recommended that the hydrogen excess flow, such as from the splitter
4.1 Knowledge of the concentration of benzene may be inlet, should be vented to a safe area such as a ventilated hood or
required for regulatory use, control of gasoline blending, appropriate exhaust approved for such use. Use caution when performing
and/or process optimizations. maintenance by isolating the hydrogen source.
5.9 FID Gas—Air chromatographic grade, dry (Warning—
5. Apparatus and Chemicals Compressed gas under high pressure.) An air generator may be
5.1 Chromatograph—Any gas chromatograph that has the used with purifier.
capability of performing multidimensional chromatography 5.10 FID Gas—Hydrogen, 99.999 % pure (Warning—
that can be operated at approximately the conditions described Compressed gas under high pressure; hydrogen flammable.) A
in this method, and meets all the required chromatographic hydrogen generator may be used.
specifications and method criteria specified in 7.5 and 11.
5.11 FID Make-up Gas, if Required (Refer to Manufactur-
5.1.1 Sample Introduction System—Automated sample in- er’s Requirements)—Nitrogen, 99.999 % pure. (Warning—
jection system such as a liquid autosampler. Compressed gas under high pressure.)
5.1.2 Injector—Capillary split/splitless injector. NOTE 3—Reference the instrument manual for specifics on how to
5.1.3 Detector—At least one flame ionization detector (FID) establish required flow rates for a particular vendor’s instrument.
is required. In a two FID configuration, the optional second 5.12 Chemicals—(Warning—These chemicals are
detector is used to monitor the effluent from the pre-column dangerous, harmful, or fatal if swallowed or inhaled. They are
(refer to configuration A in 7.1.1). flammable to extremely flammable. Vapors can cause flash
5.2 Columns: fires.)
5.2.1 Pre-column—Non-polar, boiling point column, which 5.12.1 Internal Standard—Methyl Isobutyl Ketone (MIBK)
performs the first separation of the hydrocarbons. A dimethyl 99.5 % pure.
polysiloxane phase column with dimensions of 30 m by 5.13 Calibration Standards:
0.25 mm by 0.50 µm has been used successfully. Any column 5.13.1 Benzene—99.5 % minimum purity. (Warning—
with equivalent or better chromatographic efficiency and se- Carcinogen.)
lectivity may be used. 5.13.2 Toluene—99.5 % minimum purity.
5.2.2 Analytical Column—This column separates aromatic 5.13.3 Isooctane (2,2,4–trimethyl pentane)—99.5 % mini-
and non-aromatic compounds. A polyethylene glycol (polar mum purity. This is the solvent used for dilution.
‘wax’) column with the dimensions 60 m by 0.32 mm by 1.0 5.14 Method Chemicals for Resolution Validation:
µm has been used successfully. Any column capable of meeting 5.14.1 Sec-butanol—99.5 % minimum purity.
the resolution requirement in 7.5 may be employed. 5.14.2 Iso-butanol—99.5 % minimum purity.
5.2.3 Restrictor—Uncoated deactivated fused silica. A 5.14.3 n-butanol—99.5 % minimum purity.
100 µm by 42.5 cm restrictor was successfully used when 5.14.4 n-propanol—99.5 % minimum purity.
developing this method. 5.14.5 Ethanol—99.5 % minimum purity (200 proof).
5.3 Chromatography Data System (CDS)—An electronic
device capable of graphical presentation and integration of the 6. Sampling
chromatogram peaks. 6.1 Gasoline—(Warning—Extremely flammable. Vapors
5.4 Microsyringe—Capable of making injections from harmful if inhaled.) Samples to be analyzed by this test method
0.5 µL to 1.0 µL. shall be obtained using the procedures outlined in Practice
D4057.
5.5 Volumetric Pipets, Class A—0.5 mL, 1 mL, 5 mL,
10 mL, 15 mL, and 20 mL capacities (see Specifications E694
7. Apparatus, Configurations, and Method Setup
and E969).
7.1 Configuration—Illustrated and discussed in this test
5.6 Measuring Pipets—1 mL and 2 mL capacities calibrated
method are three configurations used by the laboratories that
in 0.01 mL; 5 mL calibrated in 0.1 mL, for use in dispensing
have participated in the interlaboratory study (ILS). Figs. A1.5
volumes of benzene and toluene not covered by the volumetric
and A1.6 contain graphical representations of these configura-
pipets (see Specifications E1044 and E1293) during prepara-
tions and Table A1.1 provides the method parameters. Valve-
tion of standard samples (see 8.1).
based systems for backflushing may be used and such a
NOTE 1—Other manual or automated volume dispensing equipment
capable of delivering the specified volumes within the stated tolerance configuration is displayed in Fig. A1.7 and was also used in the
limits may be used as an alternative to the requirements stated in 5.5 and ILS. Other configurations may be used provided all of the
5.6. requirements and criteria in this test method are met, such as

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
2
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17
signal to noise (S/N), resolution, calibration, quality control 7.3.2 Configuration B, Single Detector Configuration—To
requirements, and so forth, which are discussed in 7.5 and monitor the flow from the pre-column to establish the back-
Section 11. flush time, the analytical column is removed from the FID and
7.1.1 Configuration A—Has the capability of monitoring the left in the GC oven as displayed in Fig. A1.3. (Warning—Use
effluent from the pre-column via a restrictor. This is a dual caution if using hydrogen as carrier gas; if possible, vent
detector system. One FID is used to monitor the analytical externally of the gas chromatograph’s oven.) The restrictor is
column’s effluent, and the other to monitor the effluent from the installed into the FID.
pre-column. The second FID provides a quicker approach in 7.3.2.1 Follow procedures from 7.3.1.1 through 7.3.1.3.
determining backflush time. Figs. A1.1 and A1.2 represent 7.3.2.2 Remove restrictor from the detector and install the
examples of this configuration. analytical column in the detector for analysis as shown in Fig.
7.1.2 Configuration B—Has the capability of monitoring the A1.4.
effluent from the pre-column via a restrictor; however, only one 7.3.2.3 Follow procedures from 7.3.1.4 to 7.3.1.6.
FID is installed. Figs. A1.3 and A1.4 represent examples of this 7.3.3 Configuration C, No Pre-column Configuration—This
configuration. configuration is not capable of monitoring the chromatography
7.1.3 Configuration C—Does not have the capability to from the pre-column. An example of this configuration is
monitor the effluent from the pre-column. Figs. A1.5 and A1.6 shown in Figs. A1.5 and A1.6 configuration C.
represent an example of this configuration. 7.3.3.1 Inject 0.1 µL to 0.5 µL of standard #1 (8.2) or the
7.2 Conditioning Column—Install the pre-column, analyti- composite mix (8.5) depending on ethanol or butanol blended
cal column, and the restrictor, if the restrictor is being used as gasoline using a backflush time equal to the run-time.
in configurations A and B. Condition the columns per manu- 7.3.3.2 Decrease the backflush time by 1 min intervals until
facturer’s instructions. toluene is no longer present in the chromatogram.
7.3.3.3 Increase the backflush time by 0.02 min until tolu-
7.3 Determine Time to Backflush (BF)—Or backflush time, ene provides a stable area count. The shortest time when the
must be empirically determined for each system and column toluene has a stable response is the backflush time.
set used by the laboratory. Optimization of backflush time is
dependent upon the configuration used. 7.4 Re-establishing Backflush Time—As the column ages
7.3.1 Configuration A, Dual Detector Configuration— with use, retention times will become shorter. If the quality
Follow the instrument setup as shown in Figs. A1.1 and A1.2. control samples fail, backflush time may need to be re-
established on the same column, set especially if a retention
7.3.1.1 Establish analytical parameters set forth in Table
time shift is when a new pre-column, analytical column, and/or
A1.1 or parameters that have met method separation criteria as
restrictor are employed, the backflush time will need to be
outlined in 7.5. Ensure both detector channels are being
re-established.
collected by the CDS.
7.3.1.2 Inject 0.1 µL to 0.5 µL of standard #1 (8.2) or the 7.5 Method Criteria:
butanol composite mix (8.5.2) depending on ethanol or butanol 7.5.1 Verify Resolution for Gasoline Blended With
blended gasoline, respectively. If the gasoline being tested is Ethanol—Using the injection volume and other parameters that
blended with ethanol and not butanol, the composite mix will be used for sample analysis, inject either composite mix
described in 8.5.2 for butanol blend does not need to be described in 8.5.1 or 8.5.2. Benzene shall be sufficiently
prepared. However, if the laboratory wants to prepare this resolved from ethanol and having a resolution (R) value >2.
composite mix, the composite mix described in 8.5.2 may be The internal standard (MIBK) shall have an R value >1.5.
used for establishing backflush time for both blended fuels. Refer to Fig. A1.10 for standard #1 or Fig. A1.11 for an
7.3.1.3 Record the time when toluene returns to baseline example chromatogram displaying resolution for composite
from the pre-column via the restrictor. A chromatogram from mix from 8.5.2.
the pre-column through the restrictor demonstrating this is 7.5.2 Verify Resolution for Gasoline Blended With
shown in Fig. A1.8 for standard #1, and in Fig. A1.9 for the Butanol—Using the injection volume and parameters estab-
composite mix in 8.5.2. lished above and that will be used for sample analysis, inject
7.3.1.4 Next monitor the effluent from the analytical column composite mix described in 8.5.2. The resolution requirements
using the backflush time established in 7.3.1.3. Inject standard are as follows: benzene R > 2; MIBK R > 1.5; toluene R > 0.6.
#1 (8.2) or composite mix for butanol blend (8.5.2). Record the Refer to an example chromatogram found in Fig. A1.11.
area for toluene. 7.5.3 Resolution calculation may be performed by the CDS
7.3.1.5 Reduce the backflush time by 0.02 min. Inject or manually. If using a CDS, ensure the CDS is using the
standard #1 or composite mix. Record the area for toluene. proper resolution calculation, as follows:
Repeat this step until the response for toluene begins to 1.18~ T p2 2 T p1 !
decrease. R p2 5 (1)
W h1⁄2 1W h2⁄2
7.3.1.6 The backflush time for the method will be the time
where:
recorded before the time when toluene decreased. An example
chromatogram from the analytical column of standard #1 is in Rp2 = the resolution of peak 2,
Fig. A1.10 and an example chromatogram of composite mix Tp1 = the retention time of peak 1,
described in 8.5.2 is located in Fig. A1.11. Tp2 = the retention time of peak 2,

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
3
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17

Wh1/2 = the width of peak 1 at half height, and ing the calibration curve 1/y is allowed and will provide
Wh2/2 = the width of peak 2 at half height. enhanced accuracy for lower concentration target samples.
NOTE 5—Some CDS systems plot the area ratio versus the concentra-
7.5.4 Determining Method Run Time or Completion of tion ratio of the benzene and toluene to that of the internal standard.
Chromatography for Butanol Blended Gasoline—Because Consult the system guidelines.
n-butanol may be an isomer in the butanol blend, the runtime 8.5 Composite Mix for Ethanol and Butanol Blends—The
and/or temperature needs to be long enough and temperature composite mix will ensure a necessary separation of ethanol
high enough to ensure n-butanol elutes from the analytical from benzene (R > 2), and that possible contaminants in the
column. Use the composite mix in 8.5.2 for this determination. butanol blend will not co-elute with benzene (R > 2), toluene
Refer to the chromatogram found in Figs. A1.10 and A1.11. (R > 0.6) and MIBK (R > 1.5).
7.5.5 Attaining Detection Limit for Benzene—Using stan- 8.5.1 Composite Mix for Ethanol Blend—Prepare a solution
dard mix #7 (8.1), ensure that at least a signal to noise of 50 to by volume that is 20 % ethanol, 1 % benzene, 4 % MIBK and
1 is attained for benzene. If this is not achieved, reduce the 20 % toluene in iso-octane. This solution may also be used to
instrument’s split flow or increase the injection volume, not to determine the backflush time.
exceed 0.5 µL until this criteria is attained. 8.5.2 Composite Mix for Ethanol and Butanol Blends—
7.5.6 Calibration Requirements—A linear regression, or Prepare a solution by volume that is 20 % ethanol, 1 %
weighted curve, shall meet a criterion correlation coefficient benzene, 4 % sec-butanol, 0.1 % n-propanol, 4 % MIBK, 20 %
equal to or greater than 0.999. toluene, 20 % iso-butanol, 5 % n-butanol in iso-octane. This
solution may also be used to determine the backflush time.
8. Calibration and Standardization
8.1 Standards—Prepare seven standards (Table 1) covering 9. Procedure
the range 0.06 % to 5 % by volume benzene and 0.5 % to 20 % 9.1 Preparation of Sample—Accurately measure 1.0 mL of
by volume toluene. For each standard, measure the volume of MIBK into a 25 mL volumetric flask. Fill to the mark with the
benzene and of toluene listed below into a 100 mL volumetric sample to be analyzed and mix well.
flask. Dilute to volume with isooctane (2,2,4- 9.2 Chromatographic Analysis—Chromatograph the
trimethylpentane), with all components and glassware at am- sample, using the conditions established in Section 7 and as
bient temperature. used for the calibration. The GC column flow must be turned
8.2 Final Standard Preparation—Accurately measure to backflush mode at the time determined in 7.2 so that
1.0 mL of MIBK into a 25 mL-volumetric flask, and fill to the undesirable components do not enter the analytical column.
mark using the first standard (8.1). Continue doing this until all Fig. A1.13 is an example of a Test Method D3606 proficiency
seven (7) blends (concentrations) have been prepared. sample analyzed using configuration A.
NOTE 4—Commercially prepared calibration standards may be used, 9.3 Interpretation of Chromatogram—Identify the benzene,
including those that are pre-mixed with the MIBK internal standard. The
purity of target compounds is given in 5.12. toluene, and the internal standard MIBK peaks from the
retention times of the standards.
8.3 Chromatographic Analysis—Chromatograph each of the NOTE 6—The order of elution will be nonaromatic hydrocarbons,
calibration blends using the conditions established in Section 7 benzene, MIBK and toluene using the dimethyl poly siloxane and polar
using the automated sampling system of the gas chromato- column combination used in the ILS.
graph. 9.4 Measurement of Area—Measure the areas under the
8.4 Calibration—Using a computer data system (CDS), benzene, toluene aromatic peaks and under the MIBK peak.
measure the area of the benzene and toluene aromatic peaks 10. Calculation
and of the internal standard peak. Calculate the area ratio of the
benzene peak area to the MIBK peak area. Plot the concentra- 10.1 Calculate the ratios of the peak areas of benzene and
tion of benzene versus the area ratio. Make the same calcula- toluene to the peak area of MIBK. Determine from the
tion and plot similarly for toluene. Refer to Fig. A1.12 for an calibration curve the liquid volume percent of benzene and
example calibration curve. The calibration shall be linear with toluene corresponding to the calculated peak ratios. A CDS that
a correlation coefficient greater than or equal to 0.999. Weigh- has the capability of processing this calibration and quantitat-
ing results may be used.
10.2 If the results are desired on a mass basis, convert to
TABLE 1 Seven Standards
mass percent as follows:
Benzene Toluene Benzene, mass percent 5 ~ V B ⁄ D ! 3 0.8844 (2)
Std # Volume Volume
mL mL where:
percent percent
1 5 5 20 20
VB = volume percent benzene, and
2 2.5 2.5 15 15 D = relative density of sample at 15.6/15.6 °C (60/60 °F).
3 1.25 1.25 10 10
4 0.67 0.67 5 5 Toluene, mass percent 5 ~ V T ⁄ D ! 3 0.8719 (3)
5 0.33 0.33 2.5 2.5
6 0.12 0.12 1 1 where:
7 0.06 0.06 0.5 0.5
VT = volume percent toluene, and

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
4
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17

D = relative density of sample at 15.6/15.6 °C (60/60 °F). the long run, in the normal and correct operation of the test
method, exceed the values in Table 3 only 1 case in 20.
11. Quality Control NOTE 7—The sample compositions and results of the cooperative study
are filed at ASTM International.
11.1 Test for System Stability and Repeatability—On setting
up the method for the first time and thereafter when major 13.2 Bias—Since there is no accepted reference method
maintenance is performed (change in columns and flows, for suitable for measuring bias for this method, no statement of
example), make six injections of standard #4 (8.2). Calculate bias can be made.
the standard deviation from the six results from the six
injections. Calculate 2.77 × the standard deviation obtained. 14. Keywords
This quantity shall meet repeatability as published 13.1.1. 14.1 aviation gasoline; benzene; gas chromatography; gaso-
Thereafter establish statistical quality control charting (SQC) line; multidimensional gas chromatography; spark ignition
for each batch of analysis analyzed using a production a sample fuels; toluene
similar in composition to samples to be analyzed.
11.2 Reference Material—Analyze one or more reference PROCEDURE B—PACKED GAS
material of known accepted reference value (ARV), such as CHROMATOGRAPHIC COLUMNS
from NIST and/or an independent certified control standard, or
meeting the check standard requirement as specified in Practice 15. Summary of Test Method
D6299 upon setting up the test method, post maintenance 15.1 An internal standard, sec-butanol, is added to the
(calibration, column replacement, cut time adjustments, and so sample. The sample is then introduced into a gas chromato-
forth) and at least quarterly to confirm proper set-up and graph (GC) equipped with two packed GC columns connected
calibration of the gas chromatograph. Test results shall agree in series. The sample passes first through a pre-column packed
with within the 95 % confidence level or reproducibility of the with a nonpolar or equivalent phase, which separates the
benzene or toluene components divided by the square root of 2. lighter components according to boiling point. After toluene
In addition follow any other specification requirements per has eluted, the flow through the nonpolar column is reversed,
regulatory or commercial application(s). Monitor the perfor- flushing out the components heavier than toluene. The lighter
mance of the reference material(s) using SQC charting in components then pass through an analytical column packed
accordance with Practice D6299. with a stationary phase capable of separating benzene and
12. Report toluene from ethanol and other hydrocarbons and oxygenates
such as methanol, MTBE, TAME, ETBE, and butanols. The
12.1 Report the benzene and toluene contents in liquid eluted components are detected by a thermal conductivity
volume percent to the nearest 0.01 %, and indicate that the detector (TCD). The detector response is recorded, the peak
results are from “D3606 Procedure A.” areas are measured and the concentration of each component is
13. Precision and Bias3 calculated with reference to the internal standard.
13.1 The following criteria should be used for judging the 16. Significance and Use
acceptability of results (95 % confidence). The user should
16.1 Knowledge of the concentration of benzene may be
choose the precision statement that reflects the concentration
required for regulatory use, control of gasoline blending,
range of each component under study.
and/or process optimizations.
13.1.1 Repeatability—The difference between successive
test results, obtained by the same operator with the same 17. Apparatus
apparatus under constant operating conditions on identical test
material would, in the long run, in the normal and correct 17.1 Chromatograph—Any multidimensional packed col-
operation of the test method, exceed the values in Table 2 only umn gas chromatographic instrument configuration that has a
in 1 case in 20. backflush system and thermal conductivity detector, and that
13.1.2 Reproducibility—The difference between two, single can meet the specification of this test method, such as column
and independent results, obtained by different operators work- resolution, can be used.
ing in different laboratories on identical test material would, in 17.1.1 Two backflush systems are shown. Fig. A2.1 is a
switching valve system and Fig. A2.2 is a pressure system.
Either one can be used.
3
Supporting data have been filed at ASTM International Headquarters and may
be obtained by requesting Research Report RR:D02-1882. Contact ASTM Customer 17.2 Columns:
Service at service@astm.org.

TABLE 2 Repeatability–Procedure A (WCOT Column) TABLE 3 Reproducibility–Procedure A (WCOT Column)

NOTE 1—X = the volume percent of the component. NOTE 1—X = the volume percent of the component.
Component Range, Volume percent Repeatability Component Range, volume percent Reproducibility
Benzene 0.12 to 5.2 0.03202 (X + 0.2) Benzene 0.12 to 5.2 0.1462 (X + 0.2)
Toluene 0.4 to 20 0.01767 (X + 0.5) Toluene 0.4 to 20 0.04659 (X + 0.5)

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
5
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17
TABLE 4 Repeatability/Reproducibility Example volumetric pipets (see Specifications E1044 and E1293) during
(Benzene)–Procedure A (WCOT Column) preparation of standard samples.
Benzene NOTE 8—Other volume dispensing equipment capable of delivering the
Repeatability Reproducibility
Volume percent specified volumes within the stated tolerance limits may be used as an
0.1 0.010 0.044 alternative to the requirements stated in 17.5 and 17.6.
0.2 0.013 0.058
0.3 0.016 0.073 17.7 Flasks—Class A volumetric (see Specification E288),
0.4 0.019 0.088 25 mL and 100 mL capacity.
0.5 0.022 0.102
0.6 0.026 0.117
0.7 0.029 0.132 18. Materials
0.8 0.032 0.146
0.9 0.035 0.161
18.1 Carrier Gas—Helium, 99.99 % mol pure or better.
1.0 0.038 0.175 (Warning—Compressed gas under high pressure.)
2.0 0.070 0.322
3.0 0.102 0.468 18.2 Solvents:
4.0 0.134 0.614 18.2.1 Methanol—Reagent grade. (Warning—Flammable.
5.0 0.167 0.760 Vapor harmful. Can be fatal or cause blindness if swallowed or
inhaled.)
TABLE 5 Repeatability/Reproducibility Example 18.3 Internal Standard:
(Toluene)–Procedure A (WCOT Column) 18.3.1 sec-butanol—99.5 % mol minimum purity. Free of
Toluene benzene and toluene. (Warning—Flammable. Vapor can be
Repeatability Reproducibility
Volume percent harmful.)
0.4 0.016 0.042
1.0 0.027 0.070
18.4 Calibration Standards:
2.0 0.044 0.116 18.4.1 Benzene—99+ % mol. (Warning—Poison. Carcino-
3.0 0.062 0.163 gen. Harmful or fatal if swallowed. Extremely flammable.
4.0 0.080 0.210
5.0 0.097 0.256 Vapors can cause flash fires.)
6.0 0.115 0.303 18.4.2 Isooctane (2,2,4–trimethyl pentane)—99+ % mol.
7.0 0.133 0.349 (Warning—Extremely flammable. Harmful if inhaled.)
8.0 0.150 0.396
9.0 0.168 0.443 18.4.3 Toluene—(Warning—Flammable. Vapor harmful.)
10.0 0.186 0.489 18.4.4 n-nonane—99+ % mol. (Warning—Flammable. Va-
11.0 0.203 0.536 por harmful.)
12.0 0.221 0.582
13.0 0.239 0.629
14.0 0.256 0.676 19. Sampling
15.0 0.274 0.722
16.0 0.292 0.769 19.1 Spark Ignition Fuel (for example, “Gasoline”)—
17.0 0.309 0.815 (Warning—Extremely flammable. Vapors harmful if inhaled.)
18.0 0.327 0.862
19.0 0.345 0.909
Samples to be analyzed by this test method shall be obtained
20.0 0.362 0.955 using the procedures outlined in Practice D4057.

20. Configuration of Apparatus and Establishment of


Conditions
20.1 Following manufacturer’s guidelines and/or suggested
17.2.1 Columns A (Precolumn) and B (Analytical details in A2.2, A2.3, and A2.4 optimize the chromatographic
Column)—Annex A2.3 gives an example of a commercial system.
column set that has been used successfully for samples 20.1.1 Column System Setup for Valve Backflushing—The
containing ethanol. In addition, Annex A2.4 gives an additional following procedure provides a general guideline for optimiz-
column set originally proposed by the U.S. Environmental ing a system using valve switching as shown in Fig. A2.1.
Protection Agency (EPA) that may be adequate for this Several operational conditions are given in Table A2.1.
analyses. Other packed column sets that give a resolution from 20.1.1.1 Set the valve in the forward flow mode with
ethanol as specified in 20.3 may be used. columns A and B in series (Fig. A2.2), and adjust the primary
17.3 Chromatography Data System (CDS)—An electronic flow control to give the desired flow. Measure the flow rate at
device capable of graphical presentation and integration of the the detector vent, sample side.
chromatogram peaks. 20.1.1.2 Set the valve in the backflush position, measure the
17.4 Microsyringe—5 µL capacity. flow rate at the detector vent, sample side. If the flow has
changed, adjust the secondary or auxiliary flow control to
17.5 Volumetric Pipets, Class A—0.5 mL, 1 mL, 5 mL, obtain the correct flow. Flows should match to within 61 cm3/
10 mL, 15 mL, and 20 mL capacities (see Specifications E694 min.
and E969). 20.1.1.3 Change the valve from forward flow to the back-
17.6 Measuring Pipets—1 mL and 2 mL capacities cali- flush position several times and observe the baseline. There
brated in 0.01 mL; 5 mL calibrated in 0.1 mL, for use in should be no baseline shift or drift after the initial valve kick
dispensing volumes of benzene and toluene not covered by the that results from the pressure surge. If there is a baseline shift,

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
6
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17
increase or decrease the secondary flow control slightly to 21. Calibration and Standardization
balance the baseline. A persistent drift could indicate leaks 21.1 Standard Samples—Prepare seven standard samples
somewhere in the system. (Table 6) covering the range 0.06 % to 5 % by volume benzene
20.2 Determine Time to Backflush—The time to backflush and 0.5 % to 20 % by volume toluene as follows: For each
will vary for each column system and must be determined standard, measure the volume of benzene and of toluene listed
experimentally as follows. Table A2.1 gives examples of below into a 100-mL volumetric flask. Dilute to volume with
backflush times. Prepare a mixture of 5 % by volume isooctane isooctane (2,2,4–trimethyl pentane), with all components and
in n-nonane. With the valve configuration in the forward flow glassware at ambient temperature.
mode, inject 1 µL of the isooctane-n-nonane mixture. Allow the 21.2 Calibration Blends—Accurately measure 1.0 mL of
chromatogram to run until the n-nonane has eluted and the sec-butanol into a 25 mL volumetric flask, and fill to the mark
detector signal has returned to baseline. Measure the time in with the first standard sample (21.1). Continue doing this until
seconds, from the injection until the detector signal returns to all blends have been prepared.
baseline between the isooctane and n-nonane peaks. At this NOTE 9—Commercially prepared calibration standards may be used,
point all of the isooctane, but essentially none of the n-nonane, including those that are pre-mixed with the sec-butanol internal standard.
should have eluted. Repeat the run, including the injection, but 21.3 Chromatographic Analysis—Chromatograph each of
switching the system to the backflush mode at the predeter- the calibration blends using the conditions established in 20.4
mined “time to backflush.” This should result in a chromato- using the following injection technique:
gram of isooctane with little or no n-nonane visible. If
necessary, make additional runs, adjusting the “time to back- 21.4 Injection of Sample:
flush” until this condition of all the isooctane and little or no 21.4.1 Use an automatic liquid sample injection system. If
n-nonane is attained. The “time to backflush” so established, manual injections are to be made, the injection technique in
including the actual valve operations, must be used in all 21.4.2 is necessary so that sharp symmetrical peaks will be
subsequent calibrations and analyses. It is also valid to set obtained.
backflush time after toluene elutes. NOTE 10—ILS precision was based using automated injection systems.
20.3 Resolution of Ethanol/Benzene and Sec-butanol Ben- 21.4.2 Flush the 5 µL microsyringe at least three times with
zene Pairs—Resolution calculation may be performed by the the sample mixture and then fill with about 3 µL of the sample.
CDS or manually. The resolution shall be greater than 1.5 using (Avoid including any air bubbles in the syringe). Slowly eject
calibration standard 4 (Table 6) for benzene. If using a CDS, the sample until 2.0 µL remains in the syringe; wipe the needle
ensure the CDS is using the proper resolution calculation, as with tissue and draw back the plunger to admit 1 µL to 2 µL of
follows: air into the syringe. Insert the needle of the syringe through the
septum cap of the chromatograph and push until the barrel of
1.18~ T p2 2 T p1 !
R p2 5 (4) the syringe is resting against the septum cap; then push the
W h1⁄2 1W h2⁄2
plunger to the hilt and remove the syringe immediately from
where: the chromatograph.
Rp2 = the resolution of peak 2, 21.5 Calibration—Measure the area of both aromatic peaks
Tp1 = the retention time of peak 1, and of the internal standard peak. Calculate the ratio of the
Tp2 = the retention time of peak 2, benzene peak area to the sec-butanol and the ratio of concen-
Wh1/2 = the width of peak 1 at half height, and trations of benzene to that of the sec-butanol. Plot the area
Wh2/2 = the width of peak 2 at half height.
ratios versus and concentration ratios. Make the same calcula-
20.4 To ensure proper resolution at low concentration tion and plot for toluene. The calibrations shall be linear (Fig.
ranges, the ethanol needs to have minimal peak tailing. A2.3 A2.6).
and A2.4 describe the recommended peak symmetry for NOTE 11—Some CDS systems plot the area ratio versus the concentra-
ethanol and the ethanol/benzene resolution required. It is tion ratio of the benzene and toluene to that of the internal standard.
recommended that final system optimization be verified against Consult the system guidelines.
certified reference materials containing benzene at several 22. Procedure
concentration levels of interest and ethanol present at a level in 22.1 Preparation of Sample—Accurately measure 1.0 mL of
the expected samples to confirm the accuracy of the analyses. sec-butanol into a 25 mL volumetric flask. Fill to the mark with
the sample to be tested and mix well.
TABLE 6 Seven Standard Samples 22.2 Chromatographic Analysis—Chromatograph the
Benzene Toluene sample, using the conditions established in Section 20 “time to
Std #
Volume % mL Volume % mL backflush” and the injection technique described in 21.4. The
1 5 5 20 20 valves must be turned to backflush mode at the time deter-
2 2.5 2.5 15 15 mined in 20.2 so that undesirable components do not enter
3 1.25 1.25 10 10
4 0.67 0.67 5 5
Column B.
5 0.33 0.33 2.5 2.5 22.3 Interpretation of Chromatogram—Identify on the chro-
6 0.12 0.12 1 1
7 0.06 0.06 0.5 0.5 matogram the benzene, toluene, and the internal standard
sec-butanol peaks from the retention times of the standards.

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
7
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17
NOTE 12—The order of elution will be nonaromatic hydrocarbons, benzene or toluene components divided by the square root of 2.
ethanol, benzene, sec-butanol, and toluene using the prescribed column In addition follow any other specification requirements per
system in A2.3. Figs. A2.5 and A2.6 are examples of typical chromato-
grams.
regulatory or commercial application(s). Monitor the perfor-
mance of the reference material(s) using SQC charting in
22.4 Measurement of Area—Measure the areas under the accordance with Practice D6299.
aromatic peaks and under the internal standard peak by CDS.
NOTE 13—The precision statement in Section 25 was developed from 26. Precision and Bias
data obtained using electronic integrators or on-line computers. The
precision statement may not apply if other methods of integration or peak 26.1 The following criteria should be used for judging the
area measurement are used. acceptability of results (95 % confidence). The user should
choose the precision statement that reflects the concentration
23. Calculation range of each component under study.
23.1 Calculate the ratios of the peak areas of benzene and 26.1.1 Repeatability—The difference between successive
toluene to the peak area of the internal standard or if using test results, obtained by the same operator with the same
concentration ratios also ratio of benzene and aromatics to that apparatus under constant operating conditions on identical test
of the internal standard. Determine from the appropriate material would, in the long run, in the normal and correct
calibration curve the liquid volume percent of benzene and operation of the test method, exceed the values in Table 7 only
toluene corresponding to the calculated peak ratios. in 1 case in 20.
23.2 If the results are desired on a mass basis, convert to 26.1.2 Reproducibility—The difference between two, single
mass percent as follows: and independent results, obtained by different operators work-
ing in different laboratories on identical test material would, in
Benzene, mass percent 5 ~ V B ⁄ D ! 3 0.8844 (5)
the long run, in the normal and correct operation of the test
where: method, exceed the values in Table 8 only 1 case in 20.
VB = volume percent benzene, and NOTE 14—In order to reflect changes in gasoline composition, the
D = relative density of sample at 15.6/15.6 °C (60/60 °F). precision for this test method was determined in 1994 using both
conventional gasolines as well as gasolines containing oxygenates (ethers
Toluene, mass percent 5 ~ V T ⁄ D ! 3 0.8719 (6) such as methyl tert-butyl ether, ethyl tert-butyl ether, and tert-amyl methyl
ether, and alcohols such as ethanol). This precision should be used when
where: the concentration of benzene (0.1 % to 1.5 % by volume) and toluene
VT = volume percent toluene, and (1.7 % to 9 % by volume) fall within the specified range. The sample
D = relative density of sample at 15.6/15.6 °C (60/60 °F). composition and results of the cooperative study are filed at ASTM
International.4
24. Report NOTE 15—The precision was determined using conventional motor
gasolines without oxygenates and purchased on the open market. This
24.1 Report the benzene and toluene contents in liquid precision should be used when the concentration of benzene exceeds
volume percent to the nearest 0.01 %. Indicate that result is 1.5 % by volume and toluene 9 % by volume. The sample compositions
from “D3606 Procedure B.” and results of the cooperative study are filed at ASTM International.4
26.2 Bias—Since there is no accepted reference method
25. Quality Control suitable for measuring bias for this test method, no statement of
25.1 Test for System Stability and Repeatability—On setting bias can be made.
up the method for the first time and thereafter when major
maintenance (change in columns and flows, for example) is 27. Keywords
performed, make six injections of standard #4 (Table 6). 27.1 aviation gasoline; benzene; ethanol; gas chromatogra-
Calculate the standard deviation from the six results from the phy; gasoline; multidimensional gas chromatography; spark
six injections. Calculate 2.77 × the standard deviation obtained. ignition fuel; toluene
This quantity shall meet repeatability as published in 26.1.1.
Thereafter, establish statistical quality control charting (SQC) 4
Supporting data have been filed at ASTM International Headquarters and may
for each batch of analysis analyzed using a production sample be obtained by requesting Research Report RR:D02-1042. Contact ASTM Customer
similar in composition to samples to be analyzed. Service at service@astm.org.
25.2 Reference Material—Analyze one or more reference
material of known accepted reference value (ARV), such as TABLE 7 Repeatability Benzene/Toluene (Procedure B)
from NIST and/or an independent certified control standard, or NOTE 1—X = the mean volume percent of the component.
meeting the check standard requirement as specified in Practice Range, Volume per-
Component Repeatability See Note
D6299 upon setting up the test method, post maintenance cent
(calibration, column replacement, cut time adjustments, and so Benzene 0.1–1.5 0.03(X) + 0.01 Note 14
forth) and at least quarterly to confirm proper set-up and Benzene >1.5 0.03 Note 15
Toluene 1.7–9 0.03(X) + 0.02 Note 14
calibration of the gas chromatograph. Test results shall agree Toluene >9 0.62 Note 15
with within the 95 % confidence level or reproducibility of the

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
8
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17
TABLE 8 Reproducibility Benzene/Toluene (Procedure B)
NOTE 1—X = the mean volume percent of the component.
Range, Volume
Component Reproducibility See Note
percent
Benzene 0.1–1.5 0.13(X) + 0.05 Note 14
Benzene >1.5 0.28 Note 15
Toluene 1.7–9 0.12(X) + 0.07 Note 14
Toluene >9 1.15 Note 15

TABLE 9 Repeatability/Reproducibilty Example (Benzene)—


Procedure B (Packed Column)
Concentration
r R
Volume percent
0.1 0.01 0.06
0.2 0.02 0.08
0.3 0.02 0.09
0.4 0.02 0.10
0.5 0.03 0.12
0.6 0.03 0.13
0.7 0.03 0.14
0.8 0.03 0.15
0.9 0.04 0.17
1.0 0.04 0.18
1.5 0.06 0.25
2.0 0.03 0.46
3.0 0.03 0.69
4.0 0.03 0.92
5.0 0.03 1.15

TABLE 10 Repeatability/Reproducibilty Example (Toluene)—


Procedure B (Packed Column)
Concentration
r R
Volume percent
0.2 0.21 0.09
0.3 0.21 0.11
0.4 0.21 0.12
0.5 0.22 0.13
0.6 0.22 0.14
0.7 0.22 0.15
0.8 0.22 0.17
0.9 0.23 0.18
1.0 0.23 0.19
1.5 0.25 0.25
2.0 0.26 0.31
3.0 0.29 0.43
4.0 0.32 0.55
5.0 0.35 0.67
7.0 0.41 0.91
9.0 0.47 1.15
10.0 0.62 1.15
14.0 0.62 1.15
17.0 0.62 1.15
20.0 0.62 1.15

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
9
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17
ANNEXES

(Mandatory Information)

A1. PROCEDURE A

A1.1 These configurations and operating conditions are A1.2 See Table A1.1 for Operating Conditions—
approximate and may require optimization for each instrument Configuration A.
and configuration. It is possible that configurations of pressure
switching and/or valve switching may vary among manufac- A1.3 See Figs. A1.1-A1.13 for Example Configurations,
turers. It is important that all critical parameters such as Chromatograms, and Calibration.
resolution, calibration linearity, quality control, and so forth be
met as specified in this test method. Consult manufacturer for
guidance.

TABLE A1.1 Operating Conditions—Configuration A (Example From ILS)


Parameter Configuration A
Pre-Column Dimethyl Polysiloxane (non-polar): 30 m x 0.25 mm x 0.25 µm
Analytical Column CarboWax (Polar): 30 m x 0.25 mm x 0.5 µm
Restrictor 42.5 cm x 0.1 mm (no phase)
Flow rate: Pre-Column Initial: 2.0 mL/min 2.91 min: 1.4 mL/min
Flow rate: Restrictor Initial: 4.0 mL/min 2.91 min: 3.3 2mL/min
Flow rate: Analytical Column Initial: 2.6 mL/min 2.91 min: 2.0 mL/min
Inlet Pressure, kPa (psi) Initial: 151.71 (22.0) 2.91 min: 34.51 (5.0)
Auxillary or Midpoint pressure, kPa (psi) Initial: 89.6 (13.0) 2.91 min: 75.80 (11.0)
Split Ratio 100:1
Oven Parameters Initial Temp: 75 °C hold 8 min Ramp 1: 5 °C/min
Temp 2: 85 °C hold 3 min Ramp 2: 40 °C/min
Temp 3: 140 °C hold 0.4 min
Inlet Temp 200 °C
Detector Temp 200 °C

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
10
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17

FIG. A1.1 Configuration A—Demonstrates Elution of Sample from Pre-column onto Analytical Column Before Backflush Using Dual De-
tector Instrument

FIG. A1.2 Configuration A—Demonstrates How Backflushing is Accomplished After Elution of Toluene

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
11
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17

FIG. A1.3 Configuration B—Demonstrates Elution of Sample from Pre-column onto Analytical Column Before Backflush Using Single
Detector Instrument

FIG. A1.4 Configuration B—Demonstrates How Backflushing is Accomplished After Elution of Toluene

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
12
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17

FIG. A1.5 Configuration C—Demonstrates Elution of Sample from Pre-column onto Analytical Column Before Backflush

FIG. A1.6 Configuration C—Demonstrates How Backflushing is Accomplished After Elution of Toluene

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
13
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17

FIG. A1.7 Example of Valve Configuration

FIG. A1.8 Example of Configuration A Chromatography of Standard #1 Described in 8.2 from Pre-column

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
14
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17

FIG. A1.9 Example of Configuration A Chromatography Through Pre-column for Composite Mix Described in 8.5.2

FIG. A1.10 Example Configuration A Chromatogram of Standard #1 Described in 8.2 Elution from Column B (Analytical Column)

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
15
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17

NOTE 1—sec-Butanol, n-Propanol, iso-Butanol and n-Butanol should only be present in butanol blended fuel.
FIG. A1.11 Example Configuration A Chromatogram of Composite Mix Described in 8.5.2 Through Column B (Analytical Column)

FIG. A1.12 Example Calibration Curve for Benzene Correlation Coefficient r2 = 0.9999

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
16
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17

FIG. A1.13 Configuration A Chromatogram of ASTM Cross Check Sample RFG1312, Ethanol Blend

A2. PROCEDURE B (PACKED COLUMN SYSTEM)

A2.1 These configurations and operating conditions are A2.3 Resolving Benzene From Ethanol Using Packed Col-
approximate and may require optimization for each instrument umn System:
and configuration. It is possible that configurations of pressure A2.3.1 A packed column set used in this test method is
switching and/or valve switching may vary among manufac- described below; it was developed5 to more accurately quantify
turers. It is important that all critical parameters such as
resolution, calibration linearity, quality control, and so forth be
met as specified in this test method. Consult manufacturer for 5
The sole source of supply of Column 2 known to the committee at this time is
guidance. Restek, Inc., 110 Benner Circle, Bellefonte, PA 16823. If you are aware of
alternative suppliers, please provide this information to ASTM International
A2.2 Gas Chromatographic System Configuration and Op- Headquarters. Your comments will receive careful consideration at a meeting of the
timization of Separation—See Figs. A2.1 and A2.2. responsible technical committee, which you may attend.

FIG. A2.1 Procedure B (Packed Column) Example Configuration With Valve Backflush

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
17
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17

FIG. A2.2 Procedure B (Packed Column) Example Configuration With Pressure Backflush

benzene in gasoline containing the oxygenate ethanol. A set also permits the use of N2 (nitrogen) carrier gas with no
requirement for such column or equivalent is that a resolution detrimental analytical effects (see Fig. A2.5); however, the
(RS) values for ethanol/benzene pair and sec-butanol/benzene precision indicated in the test method was developed using
pair of >1.5 be obtained when used in the configuration of B to helium only.
eliminate ethanol interference with benzene and the internal
standard sec-butanol interference with benzene. The analytical A2.3.2 A two column set is employed in the following order
packed column contains a proprietary chromatographic phase. (see Fig. A2.4):
The two column set can be used to quantify benzene in A2.3.2.1 Column 1—6 ft (1.8 m) × 1⁄8 in. OD × 2 mm ID
gasoline with or without ethanol. The robustness of the column (10 % Rtx 1 on 100/120 Silcoport)–Nonpolar Backflush.

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
18
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17

FIG. A2.3 Symmetry Value of Ethanol

FIG. A2.4 Two Column Set

FIG. A2.5 Chromatogram of Spark Ignition Fuel (‘Gasoline’) Using Configuration in Fig. A2.4

A2.3.2.2 Backflush valve. alternative supplier. However, it is imperative that the tubing and the solid
A2.3.2.3 Column 2—15.5 ft (4.76 m) × 1⁄8 in. × 2 mm ID support used to prepare the PDMS packing material be properly deacti-
(BenzoSep proprietary polymer)–Analytical Column.5 vated to prevent excessive tailing of the ethanol into benzene. Should this
occur, accurate quantitation of benzene will be difficult, if not impossible.
A2.3.2.4 TCD–Detector. The symmetry value (S) of ethanol at 10 % peak height should not exceed
NOTE A2.1—It is acceptable to replace the MXT (Restek) Column 1 2.8 when using an alternative Column 1. The symmetry value of ethanol
(which is the nonpolar PDMS backflush column) with a column from an can be calculated either electronically or manually as follows (see Fig.

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
19
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17
A2.5) by drawing a vertical line from the apex of the ethanol peak down method as an alternative to Approach A2.3 where Approach A
to the baseline, dividing the peak into two sections. The peak height of columns may not be readily available.
ethanol is then measured from the apex to the baseline. This value is
multiplied by 10 % and the resulting value is measured and marked above A2.4.2 Three column sections are employed, in the follow-
the baseline inside the peak. A horizontal line is then drawn parallel to the
ing order:
baseline through the 10 % peak height indicating mark. As indicated in
Fig. A2.5, the leading edge of the peak is designated as “A” and the A2.4.2.1 Column 1—5 ft × 1⁄8 in. ID methyl silicone on
trailing edge of the peak is designated as “B.” The distance is measured Chromosorb (10 % OV101 on Chromosorb PAW 80/100).
from the peak leading edge and trailing edge to the vertical line at 10 %
peak height. The symmetry value is then determined as S = B/A. A2.4.2.2 Column 2—5 ft × 1⁄8 in. ID TCEP on Chromosorb
(20 % TCEP on Chromosorb PAW 80/100).
A2.3.2.5 The total length of the two set of columns is 21 ft
(6.57 m). The nonpolar backflush column and the main ana- A2.4.2.3 Column 3—15 ft × 1⁄8 in. ID Carbowax 1540
lytical column are installed as illustrated in Fig. A2.4. A (15 %) on Chromosorb W 60/80.
chromatogram from such system is shown in Fig. A2.5. A2.4.2.4 This combined 20 ft section of columns 2 and 3 is
connected such that the TCEP end of the combined column is
A2.3.3 Examples of Operating Conditions Information:
toward the valve (which places the Carbowax end next to the
A2.3.3.1 Table A2.1 gives examples of operating conditions
detector). See Fig. A2.8 for a valve diagram showing the
using the column set described in A2.3 and configuration in
arrangement of the columns.
Fig. A2.2 (valve system) for Procedure B. Note that backflush
time needs to be established with each column set and A2.4.2.5 The internal standard is 2-butanol (sec-butyl alco-
individual system. Flow rates and oven temperature may need hol or SBA).
to be modified to attain the required resolution for benzene. A2.4.2.6 Gas chromatographic parameters are approxi-
Consult manufacturer of analyzer as necessary. mately as follows:
Column Temperature Isothermal at 135 °C
A2.4 Approach B: Column Head Pressure Approximately 65 psi
Volume Flow Rate Approximately 26.6 cm3/min
A2.4.1 The following second approach to packed columns
selection is based on a suggested EPA Q&A document and is A2.4.2.7 For a sample chromatogram using this
presented here. It may offer guidance to users of the test modification, see Fig. A2.9.

TABLE A2.1 Examples of Operating Conditions


NOTE 1—This backflush time must be determined for each column system.
Parameter Procedure B (Example 1) Procedure B (Example 2) Procedure B (Example 3)
A: Pre-column 6 ft (1.8 m), 1⁄8 in. OD, 2.0 mm ID, 6 ft (1.8 m), 1⁄8 in. OD, 2.0 mm ID, 6 ft (1.8 m), 1⁄8 in. OD, 2.0 mm ID,
nonpolar nonpolar nonpolar
B: Analytical Column 16 ft (4.9 m), 1⁄8 in. OD, 2.0 mm 16 ft (4.9 m), 1⁄8 in. OD, 2.0 mm 16 ft (4.9 m), 1⁄8 in. OD, 2.0 mm
ID, proprietary packing material ID, proprietary packing material ID, proprietary packing material
Flow rate (analytical column) 25.0 mL/min 30.0 mL/min 44.0 mL/min
Aux Flow 25.0 mL/min 30.0 mL/min 44.0 mL/min
Oven Parameters 135 °C hold 12 min 125 °C hold 13 min 135 °C hold 20 min
Injector Temp 200 °C 200 °C 200 °C
Detector Temp (TCD) 200 °C 200 °C 200 °C
Valve Time 0: ON Time 0: ON Time 3 min: OFF
Time 3 min: OFF Time 2 min: OFF Time 19.95 min: ON
Backflush time 3.0 min 2.0 min 3.0 min
Injection Volume 1.0 µL 1.0 µL 1.0 µL

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
20
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17

FIG. A2.6 Example Chromatogram of Calibration Standard using Configuration in Fig. A2.2 and Column Set Described in A2.3

FIG. A2.7 Example Calibration Curve for Benzene Using Configuration in Fig. A2.2 and Column Set Described in A2.3

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
21
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17

FIG. A2.8 Valve Diagram Showing Arrangement of Columns

FIG. A2.9 Example Chromatogram for Improved Resolution of Benzene and Ethanol in Accordance With Fig. A2.8 GC Column Set

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
22
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.
D3606 − 17
SUMMARY OF CHANGES

Subcommittee D02.04 has identified the location of selected changes to this standard since the last issue
(D3606 – 10ε1) that may impact the use of this standard. (Approved Dec. 1, 2017.)

(1) Added Procedure A for use of WCOT capillary columns. (3) New precision for Procedures A.
(2) Modified original D3606 procedure as now Procedure B (4) Modified scope concentration ranges based on ILS test
(packed column). samples for Procedure A.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

Copyright by ASTM Int'l (all rights reserved); Fri Feb 1 14:43:03 EST 2019
23
Downloaded/printed by
INSTITUTO POLITECNICO NACIONAL (INSTITUTO POLITECNICO NACIONAL) pursuant to License Agreement. No further reproductions authorized.

También podría gustarte