Documentos de Académico
Documentos de Profesional
Documentos de Cultura
The tab following this Instruction sheet is the sheet on which you will paste the
generator test data from your Rapidose system. To simplify the analysis, it is
recommended that the template as presented by followed as closely as possible. If
a generator does not have one of the indicated settings and that row is not used,
do not delete the row from the spreadsheet; rather, just clear the cells of that row
of any entries.
The Report sheet is calculated and completed automatically. Except for pasting the
waveform, your intervention should not be required.
Export settings
This spreadsheet makes certain assumptions about the data arrangement, as
discussed above, and the options used for exporting the data. Illustrations are
provided below to help make this spreadsheet as useful as possible.
IMPORTANT: When you import your data be sure to select the following
items in the Rapidose export dialog boxes:
NOTE: If you what other units (Gy, etc), you can of course change the units to match
your preferences provided you change the Data and Report sheets to coincide.
To copy the waveform, click on the "Wave" button, then on the "Screen" button. This
will put a copy of the waveform on the clipboard, which may be then pasted into
the worksheet where indicated.
PARAMETERS
Profile Name Table
Export Date 4/26/2018
Export Time 3:41:39 PM
Export Culture es-ES_tradnl
Index kVp Exp, Time Dose Dose Rate HVL Est, Filt, Filter
- kV s mR mR/s mm mm -
1 0.0 0.072 2.71 37.4300 n.v 4.2 Al
2 45.7 0.072 7.75 107.2000 2.0 4.3 Al
3 64.6 0.073 25.04 343.5000 2.7 4.4 Al
4 70.8 0.073 40.19 549.7000 3.2 4.3 Al
5 71.4 0.071 36.78 521.0000 3.1 4.3 Al
6 71.0 0.063 35.16 557.3000 3.2 4.4 Al
7 71.2 0.063 35.45 560.9000 3.2 4.1 Al
8 56.6 0.431 106.60 247.6000 2.3 4.1 Al
9 57.5 0.612 156.10 255.0000 2.4 4.1 Al
10 58.1 0.512 138.00 269.3000 2.4 4.1 Al
11 57.9 0.243 65.56 270.4000 2.4 4.1 Al
12 57.3 0.093 25.52 275.6000 2.4 4.4 Al
13 70.8 0.073 40.19 549.7000 3.2 4.3 Al
14 71.4 0.071 36.78 521.0000 3.1 4.4 Al
15 70.8 0.073 40.19 549.7000 3.2 4.3 Al
16 71.4 0.071 36.78 521.0000 3.1 4.4 Al
17 70.8 0.073 40.19 549.7000 3.2 4.3 Al
18 71.4 0.071 36.78 521.0000 3.1 4.4 Al
19 70.8 0.073 40.19 549.7000 3.2 4.3 Al
20 71.4 0.071 36.78 521.0000 3.1 4.4 Al
21 70.8 0.073 40.19 549.7000 3.2 4.3 Al
Anode/Filter Tube Failure
- - -
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
A line is left open at row 148 of the Report tab to paste this waveform for the report.
Radiographic Unit Performance Report
Facility Identification System Identification
0 Room 1
Registration ID Radiographic
SANATORIO PEREZ SALUM Unique ID 24
Main Campus Control Manufacturer MADE IN CHINA
cerro leon 1775/Avda. Carlos Dominguez Control Model QIONG HUA
PEDRO JUAN CABALLERO Control S/N YZ-200B
Dr. SABINA PEREZ Control Date 0
(595) 81-481658 Tube Manufacturer MADE IN CHINA
Tube Model DX51-20,40/100
Tube S/N 9112116
Tube Date 9.11
Signature
General Comments:
the unit does not meets regulatory requirements.
X-Ray Machine Performance
mR/mAs
6.0
50 0.3
60 0.8 5.1 5.0
70 1.3 4.0
80 1.9 9.6 3.0
80 2.0 2.0
f(x) = 1.615967897291E-08 x^4.2779117605
80 1.8 15.4 1.0 R² = 0.8135831588
80 1.8
0.0
80 1.8 21.9 50 60 70 80 90 100 110 120 130
kVp
80 5.3
Measured
Set kVp kVp kVp error
50 0.0 -100.0% 40 10%
60 45.7 -23.8% 140 10%
70 64.6 -7.7% 20% 40 -10%
80 71.1 -11.1% 0%
140 -10%
-20%
80 70.8 -11.5% -40%
80 71.4 -10.8% -60%
Measured Measured
Set kVp kVp HVL HVL limit
3.5
50 0 n.v -0.06
3.0
60 45.7 2.0 1.15 2.5
70 64.6 2.7 1.65 2.0
Measured
Set msec msec Error Error (%) 12%
2000 612.2 69.39% 10%
1000 512.4 48.76% 8% 0.0000001 0%
500 510.0 2.00% 6% 10000 0%
250 260.0 4.00% 4%
100 108.0 8.00% 2%
100 70.6 29.40% 0%
100 73.1 26.90% -2% 2000 1000 500 250 100 100 100 100 100
100 70.6 29.40%
100 73.1 26.90% Set Time (msec)
FAIL mA Linearity
L in e a rit y C o e f f ic ie n t
Focal Linearity
mA Spot mR/mAs Coeff.
50 L 1.56
0.12
50 L 2.76 0.28 0 0.1
0.10
50 L 2.62 0.03 1000.00 0.1
0.08
50 L 2.04 0.12
0.06
50 L 8.04 0.59
0.04
200 L 1.92 0.61 0.02
200 L 1.84 0.02 0.00
200 L 2.01 0.04 -0.02
50 50 50 50 200 200 200 200 200
Comment: Radiation output linearity between different stations does not meet standards.