This spreadsheet is provided to assist you in collecting data for quality control testing

of a radiographic x-ray unit.

The tab following this Instruction sheet is the sheet on which you will paste the
generator test data from your Rapidose system. To simplify the analysis, it is
recommended that the template as presented by followed as closely as possible. If
a generator does not have one of the indicated settings and that row is not used,
do not delete the row from the spreadsheet; rather, just clear the cells of that row
of any entries.

The cells on the Data sheet are color coded as follows:


Green cells are for import of data from your Rapidose system.
Yellow cells are where you enter information specific to your testing session
Blue cells are automatically calculated; no user entry is required

The Report sheet is calculated and completed automatically. Except for pasting the
waveform, your intervention should not be required.

Export settings
This spreadsheet makes certain assumptions about the data arrangement, as
discussed above, and the options used for exporting the data. Illustrations are
provided below to help make this spreadsheet as useful as possible.

IMPORTANT: When you import your data be sure to select the following
items in the Rapidose export dialog boxes:
NOTE: If you what other units (Gy, etc), you can of course change the units to match
your preferences provided you change the Data and Report sheets to coincide.

To copy the waveform, click on the "Wave" button, then on the "Screen" button. This
will put a copy of the waveform on the clipboard, which may be then pasted into
the worksheet where indicated.

Please Note ---


These templates are provided with the compliments of Radcal Corporation as samples and
are available for modification by the user. Certain cells are locked and protected but may be
unlocked without a password. (Use Tools/Protection/Unprotect Workbook/[enter])
Radcal assumes no responsibility or liability for any unintentional errors.
It remains the user's responsibility to validate the calculations and to confirm suitability for the particular purpose.
particular purpose.
Place cursor in upper left cell of this box for import
Rapidose by Radcal

PARAMETERS
Profile Name Table
Export Date 4/26/2018
Export Time 3:41:39 PM
Export Culture es-ES_tradnl

Index kVp Exp, Time Dose Dose Rate HVL Est, Filt, Filter
- kV s mR mR/s mm mm -
1 0.0 0.072 2.71 37.4300 n.v 4.2 Al
2 45.7 0.072 7.75 107.2000 2.0 4.3 Al
3 64.6 0.073 25.04 343.5000 2.7 4.4 Al
4 70.8 0.073 40.19 549.7000 3.2 4.3 Al
5 71.4 0.071 36.78 521.0000 3.1 4.3 Al
6 71.0 0.063 35.16 557.3000 3.2 4.4 Al
7 71.2 0.063 35.45 560.9000 3.2 4.1 Al
8 56.6 0.431 106.60 247.6000 2.3 4.1 Al
9 57.5 0.612 156.10 255.0000 2.4 4.1 Al
10 58.1 0.512 138.00 269.3000 2.4 4.1 Al
11 57.9 0.243 65.56 270.4000 2.4 4.1 Al
12 57.3 0.093 25.52 275.6000 2.4 4.4 Al
13 70.8 0.073 40.19 549.7000 3.2 4.3 Al
14 71.4 0.071 36.78 521.0000 3.1 4.4 Al
15 70.8 0.073 40.19 549.7000 3.2 4.3 Al
16 71.4 0.071 36.78 521.0000 3.1 4.4 Al
17 70.8 0.073 40.19 549.7000 3.2 4.3 Al
18 71.4 0.071 36.78 521.0000 3.1 4.4 Al
19 70.8 0.073 40.19 549.7000 3.2 4.3 Al
20 71.4 0.071 36.78 521.0000 3.1 4.4 Al
21 70.8 0.073 40.19 549.7000 3.2 4.3 Al
Anode/Filter Tube Failure
- - -
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
n,a, General False
A line is left open at row 148 of the Report tab to paste this waveform for the report.
Radiographic Unit Performance Report
Facility Identification System Identification
0 Room 1
Registration ID Radiographic
SANATORIO PEREZ SALUM Unique ID 24
Main Campus Control Manufacturer MADE IN CHINA
cerro leon 1775/Avda. Carlos Dominguez Control Model QIONG HUA
PEDRO JUAN CABALLERO Control S/N YZ-200B
Dr. SABINA PEREZ Control Date 0
(595) 81-481658 Tube Manufacturer MADE IN CHINA
Tube Model DX51-20,40/100
Tube S/N 9112116
Tube Date 9.11

Test Date 43452


Tested by 0

Signature

Summary of the results of this performance survey


(N/A = not applicable, N/M = not measured)

Pass Radiation output


Pass Beam quality
FAIL Reproducibility of radiation exposures
Pass kVp accuracy
FAIL Timer accuracy
Pass Timer reproducibility
FAIL mA station radiation output linearity

General Comments:
the unit does not meets regulatory requirements.
X-Ray Machine Performance

Pass Radiation Output


Measurement distance = 30 inches

Set kVp mR/mAs Std. Values

mR/mAs
6.0
50 0.3
60 0.8 5.1 5.0

70 1.3 4.0
80 1.9 9.6 3.0
80 2.0 2.0
f(x) = 1.615967897291E-08 x^4.2779117605
80 1.8 15.4 1.0 R² = 0.8135831588
80 1.8
0.0
80 1.8 21.9 50 60 70 80 90 100 110 120 130
kVp
80 5.3

Comment: Radiation output is typical.

Pass kVp Accuracy


M e a s ure d % E rror

Measured
Set kVp kVp kVp error
50 0.0 -100.0% 40 10%
60 45.7 -23.8% 140 10%
70 64.6 -7.7% 20% 40 -10%
80 71.1 -11.1% 0%
140 -10%
-20%
80 70.8 -11.5% -40%
80 71.4 -10.8% -60%

80 71.0 -11.3% -80%


-100%
80 71.2 -11.0% -120%
80 56.6 -29.3% 40 50 60 70 80 90 100 110 120 130 140
Nominal kVp
The kVp error limit is 10%.

Comment: kVp accuracy meets applicable standards.

Pass Beam Quality


HVL (m m Al)

Measured Measured
Set kVp kVp HVL HVL limit
3.5
50 0 n.v -0.06
3.0
60 45.7 2.0 1.15 2.5
70 64.6 2.7 1.65 2.0

80 71.1 2.9 2.53 1.5


1.0
80 70.8 3.2 2.52 0.5
80 71.4 3.1 2.54 0.0
80 71 3.2 2.53 -0.5 40 60 80 100

80 71.2 3.2 2.54 Nominal kVp


80 56.6 2.3 2.01
Measured HVL must exceed the stated limit.

Comment: X-ray beam filtration is adequate.


FAIL Exposure & Timer Reproducibility
Measured Measured
Exposure mR Exposure msec
#1 36.78 #1 70.6
#2 40.19 #2 73.1
#3 36.78 #3 70.6
#4 40.19 #4 73.1
Average 38.5 Average 71.9
Std. Dev 1.969 Range 2.5
CV 5.12%
CV must be less than 5x the range must be
5%. less than the average.

Comment: Output reproducibility is acceptable. Timer reproducibility is acceptable.

FAIL Timer Accuracy:

Measured
Set msec msec Error Error (%) 12%
2000 612.2 69.39% 10%
1000 512.4 48.76% 8% 0.0000001 0%
500 510.0 2.00% 6% 10000 0%
250 260.0 4.00% 4%
100 108.0 8.00% 2%
100 70.6 29.40% 0%
100 73.1 26.90% -2% 2000 1000 500 250 100 100 100 100 100
100 70.6 29.40%
100 73.1 26.90% Set Time (msec)

Timer error must be less than 10% or the manufacturer specification.

Comment: Timer does not meet accuracy requirements.

FAIL mA Linearity
L in e a rit y C o e f f ic ie n t

Focal Linearity
mA Spot mR/mAs Coeff.
50 L 1.56
0.12
50 L 2.76 0.28 0 0.1
0.10
50 L 2.62 0.03 1000.00 0.1
0.08
50 L 2.04 0.12
0.06
50 L 8.04 0.59
0.04
200 L 1.92 0.61 0.02
200 L 1.84 0.02 0.00
200 L 2.01 0.04 -0.02
50 50 50 50 200 200 200 200 200

200 L 1.84 0.04 Set mA


200 L 2.01 0.04
Linearity coefficient must be less than 0,10.

Comment: Radiation output linearity between different stations does not meet standards.

FAIL Radiation and kV Waveforms


Paste waveform in this
area.

Comment: Waveform is not typical.