Está en la página 1de 8

801163

research-article2018
EJO0010.1177/1120672118801163European Journal of OphthalmologyTojo et al.

EJO European
Journal of
Ophthalmology
Original Research Article

European Journal of Ophthalmology

Comparison of intraocular pressure


1­–8
© The Author(s) 2018
Article reuse guidelines:
fluctuation before and after cataract sagepub.com/journals-permissions
https://doi.org/10.1177/1120672118801163
DOI: 10.1177/1120672118801163

surgeries in normal-tension glaucoma journals.sagepub.com/home/ejo

patients

Naoki Tojo, Mitsuya Otsuka and Atsushi Hayashi

Abstract
Purpose: Cataract surgeries have been shown to reduce intraocular pressure. We used a Sensimed Triggerfish® contact
lens sensor to compare intraocular pressure levels and their fluctuation between before and after cataract surgeries in
patients with normal-tension glaucoma.
Methods: This was a prospective open-label study. Thirteen patients with normal-tension glaucoma were included. All
patients underwent a 1-month washout and discontinued glaucoma medications during this study. In each eye, intraocular
pressure fluctuations over 24 h were measured with the contact lens sensor before and at 3 months after the cataract
surgery. We compared intraocular pressure levels and their fluctuation between before and after cataract surgeries. We
used two approaches to evaluate the amplitude of intraocular pressure fluctuations: dual-harmonic regression analysis,
and measurement of the difference between the maximum and the minimum value.
Results: The mean pre-operative intraocular pressure was 14.7 ± 2.2 mm Hg and mean post-operative intraocular
pressure was 11.4 ± 2.2 mm Hg. Cataract surgery significantly decreased intraocular pressure (p = 0.0005). In both
methods, the post-operative fluctuations in intraocular pressure over 24 h were significantly smaller than their pre-
operative counterparts (dual-harmonic regression analysis: p = 0.0171; difference between the maximum and the
minimum: p = 0.0398).
Conclusion: Cataract surgery decreased both intraocular pressure values and intraocular pressure fluctuations in
normal-tension glaucoma patients.

Keywords
Normal-tension glaucoma, intraocular pressure, contact lens sensor, Triggerfish, fluctuation, cataract surgery

Date received: 17 December 2017; accepted: 28 August 2018

Introduction
Glaucoma is a neurodegenerative disease of the eye and Study,3 20% of NTG subjects showed visual field progres-
the leading cause of blindness in Japan.1 Normal-tension sion despite a reduction in their IOP by more than 30%
glaucoma (NTG) is typically defined as optic disk and/or from baseline. These results suggest that non-IOP-related
visual field glaucomatous changes in an individual who elements may be involved in the progression of NTG.
has never shown an untreated intraocular pressure (IOP)
value above 21 mm Hg.2 High IOP is a well-known signifi-
cant risk factor for the progression of glaucoma.3,4 Department of Ophthalmology, Graduate School of Medicine and
Pharmaceutical Sciences for Education, University of Toyama, Toyama,
In the Tajimi Study, the average IOP for eyes with primary Japan
open-angle glaucoma (POAG) was 15.4 ± 2.8 mm Hg.5
It has been estimated that NTG is present in almost 70% of Corresponding author:
Naoki Tojo, Department of Ophthalmology, Graduate School of
all glaucoma patients in Japan, and the lack of an effective Medicine and Pharmaceutical Sciences for Education, University of
treatment other than lowering the IOP is a significant Toyama, 2630 Sugitani, Toyama 930-0194, Japan.
problem. In the Collaborative Normal Tension Glaucoma Email: tojo-naoki@umin.ac.jp
2 European Journal of Ophthalmology 00(0)

The IOP is known to fluctuate, and several studies have Glaucoma Hemifield Test (GHT) were “outside normal
shown that large IOP fluctuation might be one of the causes limits,” and a cluster of three contiguous points on the pat-
of glaucoma progression.6–11 In support of these findings, tern deviation plot that were depressed at the p < 0.05 level
other studies have reported that large IOP fluctuations con- (occurring in age-matched normal subjects) did not cross
tribute to fast glaucoma progression.6,9,12,13 The determina- the horizontal meridian and were compatible with
tion of IOP fluctuations over the course of a full 24 h may glaucoma.
thus provide better information for clinical decision-making The inclusion criteria for the subjects were (1) the
than discrete measurements of IOP in a single day. case was appropriate for cataract surgery; (2) the best-
A contact lens sensor (CLS) was developed to continu- corrected visual acuity was ⩾0.2 and <0.8; (3) spherical
ously monitor habitual IOP fluctuations over a period of equivalent was >−6 and <+6 diopters; and (4) mean
24 h.14,15 Cataract surgery has been shown to reduce the deviation (MD) was >−20  dB. For patients with
IOP and the number of anti-glaucoma medications in glau- advanced glaucoma, other surgical treatments might be
coma patients.16–18 The purpose of this study was thus to also considered, so we excluded advanced glaucoma
use a CLS to compare the IOP value and IOP fluctuations patients (<−20 dB). When both eyes were indicated,
before and after cataract surgery in patients with NTG. we choose the eye of the severe visual field. The exclu-
sion criteria were (1) primary angle closure glaucoma,
(2) neovascular glaucoma, (3) steroid glaucoma, (4)
Material and methods
history of ocular trauma, (5) retinal disease, (6) post-
Subjects photocoagulation, (7) post-vitrectomy, (8) ocular inflam-
matory disease, (9) prior laser trabeculoplasty or other
This was a prospective, open-label study. Thirteen con- prior glaucoma surgery, (10) IOP of more than 21 mm Hg
secutive patients who were diagnosed with NTG by two after washout of glaucoma medications, and (11) case of
glaucoma specialists (N.T and M.O) at Toyama intraoperative complications.
University Hospital were enrolled. All subjects were
recruited during the period from April 2016 to January
2017. All patients underwent a comprehensive ophthal- Measurement of IOP with GAT
mic examination including refraction, Goldmann goni- All patients underwent a 1-month washout of glaucoma
oscopy, fundus examination, anterior segment optical medications. The untreated IOP was obtained at two read-
coherence tomography (AS-OCT) (CASIA SS-1000; ings on separate occasions prior to the cataract surgeries.
Tomey, Nagoya, Japan), and automated perimetry We confirmed that untreated IOP was less than 21 mm Hg.
(Humphrey Field Analyzer; Carl Zeiss Meditec, Dublin, We measured the post-operative IOP at 3 months after the
CA, USA). We measured IOP with Goldmann applana- cataract surgeries in each patient. Since this study is con-
tion tonometry (GAT), we did not use correction IOP ducted under conditions excluding the effect of glaucoma
value with central corneal thickness (CCT). medications, it is aimed to investigate the effects of pure
Glaucoma medications had already been used in all cataract surgery. We adopted the IOP value at about 10
patients, but additional treatment was needed to lower a.m., just before the start of the 24-h continuous IOP meas-
their IOP due to the progression of their visual field disor- urement with the CLS.
der. The research protocol was approved by the Institutional
Review Board of the University of Toyama, and the proce-
dures used conformed to the tenets of the Declaration of Surgical techniques
Helsinki. After the nature and possible consequences of All patients were operated on by a single surgeon (N.T.)
the study were explained to the patients, written informed using the same surgical protocol under local (peribulbar)
consent was obtained from each. anesthesia. A temporal clear cornel incision was made
(2.8 mm), and conventional phacoemulsification (PEA)
Inclusion and exclusion criteria and intraocular lens implantation were then performed.
PEA was performed with a WHITESTAR SIGNATURE
The diagnosis of NTG was made if all three of the follow- System (Abbott Medical Optics, Santa Ana, CA, USA).
ing criteria were satisfied. (1) Untreated IOP had never
been above 21 mm Hg. We measured untreated IOP at least
two times on separate occasions, and we clinically defined
Post-operative medications
NTG as an untreated IOP value of less than 21 mm Hg. (2) The post-operative treatments consisted of topical steroids
Glaucomatous optic disk neuropathy (a cup/disk ratio of and antibiotics that were reduced over 4 weeks following
⩾0.7, or the presence of notching) was present, accompa- the intervention. Topical non-steroidal anti-inflammatory
nied by corresponding visual field defects. (3) In a thresh- drugs were used for 3 months. All anti-glaucoma medi-
old examination using the Swedish Interactive Threshold cations were discontinued for the duration of the study
Algorithm (SITA) 30-2 program, the scores on the period.
Tojo et al. 3

Figure 1.  Representative results of IOP fluctuation measurement using a contact lens sensor over a 24-h period before and after
cataract surgery.
Gray line: the IOP fluctuation before cataract surgery.
Black line: the IOP fluctuation after cataract surgery. After the surgery, the range of IOP fluctuation was decreased. The two gently sloping lines
show the results of the dual-harmonic regression analysis.

Measurement of IOP fluctuation with the CLS  2π   2π 


IOPt = M + A1 cos  t + φ1  + A2 cos  2 t + φ2 
We measured the IOP fluctuation using the CLS at about  τ   τ 
10 a.m. the day before and the day 3 months after cataract
surgery. Two doctors trained (N.T and M.O) in CLS meas- where A1 is the amplitude of the fundamental cosine fit, A2
urement inserted and removed the CLS in the hospital. is the amplitude of the first harmonic cosine fit, φ1 is the
Monitoring was performed in one eye of each subject; if acrophase of the fundamental cosine fit, φ2 is the acrophase
both eyes had NTG, we inserted the CLS in the eye with of the first harmonic cosine fit, τ is the endogenous circa-
the worse visual field. dian period (set at 24 h because of the entrained condi-
The 24-h IOP was monitored using a Triggerfish® CLS tions), M is the midline estimating statistic of rhythm
(Sensimed, Lausanne, Switzerland). This CLS consists of (MESOR), and τ is time. Unbiased estimates and confi-
a highly oxygen-permeable soft contact lens whose key dence limits of amplitude (half the difference between the
elements are two sensing-resistive strain gauges that can highest and lowest IOP values in a 24-h cycle), MESOR
record circumferential changes in the area of the corneo- (average IOP values in a 24-h cycle), acrophase (time of
scleral junction.14,15 The device is based on a novel the highest IOP value in a 24-h cycle), and bathyphase
approach to IOP monitoring in which changes in corneal (time of the lowest IOP value in a 24-h cycle) were
curvature and circumference are assumed to correspond obtained from modeling each IOP curve.21 The amplitude
to changes in IOP. The unit of measurement used in moni- of the IOP curve was calculated automatically using the
toring the IOP fluctuation with the Triggerfish is not software provided with the Sensimed CLS.
mm Hg but mVeq (millivolt equivalents), which is unique The other method was based on manual calculation of
to the Triggerfish. The median values were monitored for the difference between the maximum value (mVeq) and
30 s every 5 min, providing 288 points over the 24-h the minimum value during the course of the 24 h. This
period. The subjects were instructed to record the time method was used in previous reports.10,11,22 We also evalu-
when they went to bed, and the time when they woke up. ated the IOP fluctuations separately: diurnal IOP and noc-
Subjects were not restricted in their posture during the turnal IOP. The nocturnal/sleep periods were defined based
measurement. on the observation of blink cessation. Eye blink was iden-
We compared IOP fluctuation before and after cataract tified as short and high-amplitude spikes in the CLS sig-
surgery using two methods: dual-harmonic regression anal- nals that were displayed by the software’s zoom function.
ysis, and manual calculation of the difference between the We also verified diurnal/wakefulness periods and noctur-
maximum and minimum values. A sample result from dual- nal/sleep periods during the 24-h assessment based on
harmonic regression analysis is shown in Figure 1. The each subject’s log; subjects were instructed to record the
shapes of these approximate curves were generated by com- time when they returned home, installed their eye drops,
bining two sine curves from raw IOP data over 24 h. This went to bed, and woke up. We also evaluated the relation-
method was used to model the 24-h IOP rhythms19,20 as ship between systemic disease and IOP change.
4 European Journal of Ophthalmology 00(0)

Evaluation of the changes in corneal thickness Table 1.  Ophthalmic data.


and meridians Gender (male) 53.8% (7/13)
Age (years) 70.9 ± 6.2
The CCT and corneal meridians, which may influence
Axial length (mm) 24.0 ± 1.5
IOP measurement, were measured with AS-OCT before CCT (µm) 514 ± 16
and after a 24-h IOP measurement in each eye. All Medications 2.9 ± 1.3
patients were measured with CLS twice (once before and Treated IOP (mm Hg) 12.0 ± 2.0
once after their surgeries), we used the results obtained Baseline IOP (mm Hg) (washout medication) 14.7 ± 2.2
before the surgeries. We measured the differences
between the total corneal meridians and anterior corneal CCT: central corneal thickness; IOP: intraocular pressure.
meridians.
Table 2.  Comparison of the results of IOP and IOP
fluctuations between pre- and post-cataract surgery for
Statistics normal-tension glaucoma.
The Wilcoxon signed-rank test was used to compare the
Pre-operation Post-operation p value
values between before and after IOP and to assess the fluc-
tuation of IOP. Assuming that the standard deviation of the IOP (mm Hg) 14.7 ± 2.2 11.4 ± 2.2 0.0005
IOP daily variation with respect to a paired difference IOP fluctuation (mVeq)
between before and after surgery was 50 mVeq according  Dual-harmonic 174.5 ± 63.9 137.8 ± 48.8 0.0171
to the data to have been collected in a previous study,22 we   24 h 519.6 ± 122.0 433.2 ± 107.0 0.0398
found that a total of eight pairs of values were needed to   Diurnal period 439.0 ± 115.7 389.5 ± 117.0 0.147
detect a meaningful difference of 50 mVeq with respect to   Nocturnal period 265.4 ± 83.6 183.0 ± 54.7 0.0105
the IOP daily variation with 80% power and a two-sided IOP: intraocular pressure.
significance level of 0.05. All statistical analyses were per-
formed using the JMP Pro 11 software program (SAS,
Cary, NC, USA). Statistical significance was defined at the NTG patients (p = 0.0005). Only one patient had an
p < 0.05. IOP at 3 months after surgery that was higher than the pre-
operative IOP.

Results
Measurement of IOP fluctuations
Ophthalmic data We were able to successfully measure the 24-h IOP in
Table 1 shows the baseline data of the subjects, and Table 2 all 13 NTG patients. There were no serious complica-
shows the IOP fluctuation data. The study was performed tions related to wearing the CLS. Figure 1 shows an
on 13 eyes of 13 patients (7 males and 6 females). The example of IOP result with dual-harmonic regression
mean (±standard deviation) age at the time of surgery, analysis. The mean of pre-operative IOP fluctuations was
axial length, CCT, and amount of glaucoma medication 174.5 ± 63.9 mVeq, and the mean post-operative IOP fluc-
were 70.9 ± 6.2 years, 24.0 ± 1.5 mm, 514 ± 16 µm, and tuation was 137.8 ± 48.8 mVeq. The amplitudes of IOP
2.9 ± 1.3 drops, respectively. No patients took oral aceta- fluctuation are shown in Figure 2. As shown in Figure 2,
zolamide. Regarding systemic diseases, there were three the mean amplitude of IOP fluctuations was significantly
patients with hypertension and two patients with diabetes. decreased (p = 0.0171). The peak IOP (the time of peak
All cataract surgeries were performed successfully, and amplitude) occurred during the nocturnal period in all
no intraoperative complications occurred. After the surger- patients. 10 of 13 patients had smaller IOP fluctuations
ies, there were no severe post-operative complications than before the cataract surgery with this method.
such as hypertonia, hypotony, maculopathy, choroidal The results for the second method, that is, the analysis
effusion or hemorrhage, and endophthalmitis. of differences between the maximum and minimum IOP,
are shown in Figure 3. The mean range of IOP fluctuations
over 24 h and the nocturnal period were significantly
Measurement of IOP with GAT smaller than before the surgery (p = 0.0398, p = 0.0105).
The average IOPs before and after the washout of glau- The number of patients who had greater IOP fluctuations
coma medications were 12.0 ± 2.0 and 14.7 ± 2.2 mm Hg, after cataract surgery over 24 h, the diurnal period, and the
respectively. Washout of glaucoma medication increased nocturnal period were 2, 4, and 3, respectively.
the IOP significantly (p = 0.0009). The average IOP at There was no correlation between the presence or
3 months after cataract surgery was 11.4 ± 2.2 mm Hg. absence of systemic disease and post-operative IOP
Cataract surgery was shown to significantly lower IOP in decrease.
Tojo et al. 5

Corneal effects of the CLS


To examine the effects of the CLS on the cornea, we meas-
ured the corneal curvature and CCT with the AS-OCT just
before insertion of the contact lens with the CLS, and
again just after its removal. Table 3 shows the changes of
corneal curvature, thickness, and IOP value. The mean
steeper meridian was significantly changed (p = 0.0332),
and the mean of the flatter meridian was not significantly
changed (p = 0.348). The mean CCT was not significantly
changed (p = 0.172). We measured the corneal curvature
and CCT with the AS-OCT before and 3 months after cata-
ract surgeries. There was no significant change of CCT
and corneal curvature between before and after cataract
surgeries (Table 4).

Discussion
We used a CLS to measure IOP fluctuations in 13 NTG
patients before and after cataract surgery and found that there Figure 2.  Range of IOP fluctuation before and after cataract
were significant decreases in the IOP values and the ampli- surgery (dual-harmonic regression analysis).
tude of 24-h IOP fluctuation at 3 months after surgery. Boxplots of the amplitude of IOP fluctuation with dual-harmonic
regression analysis before and after cataract surgery are shown. Cata-
Previous studies reported that cataract surgery could ract surgery significantly decreased the amplitude of IOP fluctuation
reduce the IOP to within the range of 1.1–4.0 mm Hg.17,23–28 (p = 0.0171).

Figure 3.  Range of IOP fluctuation before and after cataract surgery (the difference between the maximum and minimum).
Boxplots of the range of IOP fluctuation (a) over 24 h, (b) during the diurnal period, and (c) during the nocturnal period before and after
cataract surgery. Cataract surgery significantly decreased the range of IOP fluctuation over 24 h and during the nocturnal periods (p = 0.0398,
p = 0.0105).
6 European Journal of Ophthalmology 00(0)

Table 3.  Change in the CCT and corneal curvature before and after the 24-h measurement with contact lens sensor.

Before measurement After measurement p value


CCT (µm) 514 ± 16 511 ± 19 0.172
Steeper meridian (D) 44.8 ± 2.3 45.3 ± 2.6 0.0332
Flatter meridian (D) 43.7 ± 1.6 43.8 ± 2.1 0.348

CCT: central corneal thickness; D: diopter

Table 4.  Change in the CCT and corneal curvature before and after cataract surgeries.

Pre-operation Post-operation p value


CCT (µm) 514 ± 16 517 ± 17 0.106
Steeper meridian (D) 44.75 ± 2.27 45.05 ± 2.44 0.178
Flatter meridian (D) 43.65 ± 1.57 44.12 ± 1.83 0.223

CCT: central corneal thickness; D: diopter.

However, there were some reports that cataract surgery fluctuation might have possibility to contribute progres-
could not affect for lowering IOP.29,30 In this study, cataract sion glaucoma.
surgery was found to reduce the IOP by 3.3 mm Hg in our There was no significant relationship between system-
NTG patients, which constituted an improvement over the atic disease and IOP lowering or IOP fluctuation change.
previous reports. We considered that this was due to the To state the relationship between systemic diseases and
difference in the number of glaucoma medications. Most of IOP, the number of subjects was too small.
the previous studies reported that cataract surgery reduced All our patients experienced their peak IOP during the
the number of glaucoma medications. In this study, how- night. This result is in accord with the previous reports.11,35
ever, we did not use glaucoma medications. In our study, cataract surgery was found to significantly
There are some reports that IOP fluctuation is one of reduce IOP fluctuations. We suspect that there are two rea-
factors for glaucoma progression and some reports that it sons for the reductions in IOP and IOP fluctuation after
is not so, and it is controversial.31–33 In general, however, cataract surgery. First, the opening of the anterior chamber
because the existing studies on IOP fluctuation can be angle would be expected to reduce these parameters.
divided into the short-term studies (with measurements Second, when patients undergo cataract surgery, the irriga-
throughout a single day) and the long-term studies (with tion and aspiration could washout some materials that
measurements over several years), it has been difficult to obstructed the outflow though the trabecular meshwork.
determine the significance of these variations due to the Our study has several weaknesses. First, it was not pos-
lack of standardization regarding the time between assess- sible to convert the recorded units of mVeq to units of
ments, the methods of measurement, and the definition of mm Hg. The CLS recorded relative IOP from the initial
fluctuation itself. IOP, not the absolute IOP. Moreover, the relationship
Using GAT, Kim et al.34 reported that the 24-h IOP fluc- between the CLS device output and IOP as measured with
tuation in NTG eyes was significantly larger than that in a tonometer is unknown. Therefore, further investigations
healthy eyes. Their method could not adequately account will be needed into the usefulness of CLS for use in meas-
for the variations in IOP throughout the day, because their uring IOP fluctuation. Leonardi et al.15 demonstrated that
methods might overlook the IOP fluctuations at interval the IOP and the output by CLS were directly proportional.
between IOP measurements. However, methods for meas- Some studies have reported that the reproducibility with
uring IOP more frequently must overcome the burdens of CLS was good.21,36–38 In light of the existing data, we con-
hospitalization and sleep disturbance. The new method of sidered that the plotted data with CLS were reliable.
measuring IOP fluctuation with a CLS overcomes these A second limitation is that changes in the corneal curva-
problems and provides a clearer assessment of circadian tures may have affected the values of mVeq during the
patterns of IOP. We also expect that it will lessen the pos- 24-h measurement of IOP fluctuations with CLS. Further
sibility of bias in the measurement of IOP fluctuation, improvement in the CLS will be expected to reduce stress
since the CLS provides objective measurements. Using the on the cornea during the measurement. Cataract surgery
CLS method, some studies have reported that the IOP makes some changes in morphology.39 Measurement IOP
fluctuation in the eyes of patients with NTG was greater with CLS results in the presence of artifacts due to the cor-
than that in healthy eyes.11,35 In this regard, larger IOP neal incision. In our study, there was no significant change
Tojo et al. 7

of corneal curvature between pre-op and post-op, but these Declaration of conflicting interests
small changes might not be negligible. Previous studies The author(s) declared no potential conflicts of interest with
reported that cataract surgery increased IOP with dynamic respect to the research, authorship, and/or publication of this
contour tonometry, and the phaco-induced GAT-measured article.
IOP drop has been questioned.40,41
A third limitation is that the methods for evaluating IOP Funding
fluctuation with CLS have not been standardized. The pre-
The author(s) received no financial support for the research,
vious studies on IOP fluctuation used a variety of slightly
authorship, and/or publication of this article.
different evaluation methods. Some studies evaluated the
range of IOP fluctuations as the difference between the
maximum value (mVeq) and the minimum value during References
the course of the 24 h recording.10,11 This method is simple, 1. Shiose Y, Kitazawa Y, Tsukahara S, et al. Epidemiology of
and the calculations are easily made. However, if the plot- glaucoma in Japan—a nationwide glaucoma survey. Jpn J
ted IOP has some spikes, it might not be suitable for depict- Ophthalmol 1991; 35: 133–155.
ing the overall IOP fluctuations. Mottet et al.21 reported the 2. Schulzer M. Intraocular pressure reduction in normal-tension
method of dual-harmonic regression analysis. Their glaucoma patients. The Normal Tension Glaucoma Study
Group. Ophthalmology 1992; 99: 1468–1470.
method yields a sophisticated curve of IOP fluctuation, but
3. Collaborative normal-tension glaucoma study group. The
it can fail to capture the real IOP peak. Other reports have effectiveness of intraocular pressure reduction in the treat-
evaluated IOP fluctuations using various calculation ment of normal-tension glaucoma. Am J Ophthalmol 1998;
methods.42–44 Given this wide range of approaches, the 126: 498–505.
CLS-based methods for evaluating IOP fluctuation will 4. Heijl A, Leske MC, Bengtsson B, et al. Reduction of
need further study. intraocular pressure and glaucoma progression: results from
The forth limitation is very small number of subjects the Early Manifest Glaucoma Trial. Arch Ophthalmol 2002;
and short term of follow-up period. Only 3 months is 120: 1268–1279.
probably too short in a chronic disease as glaucoma. 5. Iwase A, Suzuki Y, Araie M, et al. The prevalence of pri-
Furthermore, it is desirable to follow-up for long term mary open-angle glaucoma in Japanese: the Tajimi study.
and to measure IOP fluctuation multiple times with CLS. Ophthalmology 2004; 111: 1641–1648.
6. Ma KT, Kim CY, Seong GJ, et al. Intraocular pressure
This fifth limitation is lack a control group to compare
reduction in normal-tension glaucoma patients in South
with NTG patients. It desirable to compare IOP and IOP Korea. Int Ophthalmol 2011; 31: 355–361.
fluctuations between NTG patients and normal subjects 7. Nouri-Mahdavi K, Medeiros FA and Weinreb RN.
before and after cataract surgery. The sixth limitation is to Fluctuation of intraocular pressure as a predictor of visual
ignore the influence of glaucoma medications. The aver- field progression. Arch Ophthalmology 2008; 126: 1168–
age IOP before washout glaucoma medication period 1169.
(12.0 ± 2.0 mm Hg) is quite similar to the average IOP 8. Caprioli J and Coleman AL. Intraocular pressure fluctua-
at 3 months after cataract surgery (11.4 ± 2.2 mm Hg). It tion a risk factor for visual field progression at low intraocu-
desirable to compare IOP when resuming glaucoma medi- lar pressures in the advanced glaucoma intervention study.
cations after cataract surgery. Whether cataract surgery is Ophthalmology 2008; 115: 1123–1129e.3.
better than glaucoma medications treatment is unknown in 9. Nouri-Mahdavi K, Hoffman D, Coleman AL, et al.
Predictive factors for glaucomatous visual field progres-
this study.
sion in the Advanced Glaucoma Intervention Study.
Treatment for glaucoma may be needed not only to Ophthalmology 2004; 111: 1627–1635.
reduce a temporary IOP but also to lessen IOP fluctuation. 10. Tojo N, Hayashi A, Otsuka M, et al. Fluctuations of the
IOP measurement with a CLS might therefore becomes an intraocular pressure in pseudoexfoliation syndrome and
essential test for glaucoma patients. Even though there are normal eyes measured by a contact lens sensor. J Glaucoma
some limitations regarding the use of a CLS, habitual IOP 2016; 25: e463–468.
fluctuation can provide useful information. Our study 11. Tojo N, Abe S, Ishida M, et al. The fluctuation of intraocu-
clearly showed that cataract surgeries decreased both IOP lar pressure measured by a contact lens sensor in normal-
values and IOP fluctuations in our patients with NTG. tension glaucoma patients and nonglaucoma subjects. J
Cataract surgery might thus be one of the treatment strate- Glaucoma 2016; 26: 195–200.
gies for NTG. In addition, the improvement of IOP fluc- 12. Komori S, Ishida K and Yamamoto T. Results of long-term
monitoring of normal-tension glaucoma patients receiving
tuations might be important for the prevention of glaucoma
medical therapy: results of an 18-year follow-up. Graefes
progression. Arch Clin Exp Ophthalmol 2014; 252: 1963–1970.
13. Musch DC, Gillespie BW, Niziol LM, et al. Intraocular
Acknowledgement pressure control and long-term visual field loss in the
The authors thank Professor Hideki Orikasa for his guidance on Collaborative Initial Glaucoma Treatment Study. Ophthal-
statistics. mology 2011; 118: 1766–1773.
8 European Journal of Ophthalmology 00(0)

14. Mansouri K and Shaarawy T. Continuous intraocular pres- 29. Coban-Karatas M, Sizmaz S, Altan-Yaycioglu R, et al. Risk
sure monitoring with a wireless ocular telemetry sensor: factors for intraocular pressure rise following phacoemulsi-
initial clinical experience in patients with open angle glau- fication. Indian J Ophthalmol 2013; 61: 115–118.
coma. Br Ophthalmol 2011; 95: 627–629. 30. Fang EN and Kass MA. Increased intraocular pressure after
15. Leonardi M, Pitchon EM, Bertsch A, et al. Wireless con- cataract surgery. Semin Ophthalmol 1994; 9: 235–242.
tact lens sensor for intraocular pressure monitoring: assess- 31. Altintaş O, Yüksel N, Karabaş VL, et al. Diurnal intraocu-
ment on enucleated pig eyes. Acta Ophthalmol 2009; 87: lar pressure variation in pseudoexfoliation syndrome. Eur J
433–437. Ophthalmol 2004; 14: 495–500.
16. Guan H, Mick A, Porco T, et al. Preoperative factors associ- 32. Bengtsson B, Leske MC, Hyman L, et al. Fluctuation of
ated with IOP reduction after cataract surgery. Optom Vis intraocular pressure and glaucoma progression in the early
Sci 2013; 90: 179–184. manifest glaucoma trial. Ophthalmology 2007; 114: 205–209.
17. Vold S, Ahmed II, Craven ER, et al. Two-year COMPASS 33. Medeiros FA, Weinreb RN, Zangwill LM, et al. Long-term
trial results: supraciliary microstenting with phacoemulsifi- intraocular pressure fluctuations and risk of conversion from
cation in patients with open-angle glaucoma and cataracts. ocular hypertension to glaucoma. Ophthalmology 2008; 115:
Ophthalmology 2016; 123: 2103–2112. 934–940.
18. Chen PP, Lin SC, Junk AK, et al. The effect of phacoemul- 34. Kim YK, Oh WH, Park KH, et al. Circadian blood pressure
sification on intraocular pressure in glaucoma patients: and intraocular pressure patterns in normal tension glau-
a report by the American Academy of Ophthalmology. coma patients with undisturbed sleep. Korean J Ophthalmol
Ophthalmology 2015; 122: 1294–1307. 2010; 24: 23–28.
19. Brown EN and Czeisler CA. The statistical analysis of 35. Agnifili L, Mastropasqua R, Frezzotti P, et al. Circadian
circadian phase and amplitude in constant-routine core- intraocular pressure patterns in healthy subjects, primary
temperature data. J Biol Rhythms 1992; 7: 177–202. open angle and normal tension glaucoma patients with a
20. Gronfier C, Wright KP and Kronauer RE. Entrainment of contact lens sensor. Acta Ophthalmol 2015; 93: e14–21.
the human circadian pacemaker to longer-than-24-h days. 36. Mansouri K, Medeiros FA, Tafreshi A, et al. Continuous
Proc Natl Acad Sci USA 2007; 104: 9081–9086. 24-hour monitoring of intraocular pressure patterns with a
21. Mottet B, Aptel F, Romanet JP, et al. 24-hour intraocular contact lens sensor: safety, tolerability, and reproducibility
pressure rhythm in young healthy subjects evaluated with in patients with glaucoma. Arch Ophthalmol 2012; 130:
continuous monitoring using a contact lens sensor. JAMA 1534–1539.
Ophthalmol 2013; 131: 1507–1516. 37. Mansouri K, Weinreb RN and Liu JH. Efficacy of a contact
22. Tojo N, Oka M, Miyakoshi A, et al. Comparison of fluc- lens sensor for monitoring 24-h intraocular pressure related
tuations of intraocular pressure before and after selective patterns. PLoS ONE 2015; 10: e0125530.
laser trabeculoplasty in normal-tension glaucoma patients. 38. Liu JH, Mansouri K and Weinreb RN. Estimation of 24-hour
J Glaucoma 2014; 23: e138–143. intraocular pressure peak timing and variation using a con-
23. Shoji T, Tanito M, Takahashi H, et al. Phacoviscocanalostomy tact lens sensor. PLoS ONE 2015; 10: e0129529.
versus cataract surgery only in patients with coexisting 39. Bilak S, Simsek A, Capkin M, et al. Biometric and intraocu-
normal-tension glaucoma: midterm outcomes. J Cataract lar pressure change after cataract surgery. Optom Vis Sci
Refract Sur 2007; 33: 1209–1216. 2015; 92: 464–470.
24. Mierzejewski A, Eliks I, Kaluzny B, et al. Cataract phacoe- 40. Heinrich MA, Eppig T, Langenbucher A, et al. Comparison
mulsification and intraocular pressure in glaucoma patients. of Goldmann applanation and dynamic contour tonometry
Klin Oczna 2008; 110: 11–17. before and after cataract surgery. J Cataract Refract Surg
25. Damji KF, Konstas AGP, Liebmann JM, et al. Intraocular 2012; 38: 683–689.
pressure following phacoemulsification in patients with and 41. Plange N, Rennings C, Herr A, et al. Ocular pulse amplitude
without exfoliation syndrome: a 2 year prospective study. before and after cataract surgery. Curr Eye Res 2012; 37:
Br J Ophthalmol 2006; 90: 1014–1018. 115–119.
26. Slabaugh MA, Bojikian KD, Moore DB, et al. The effect 42. Rekas M, Danielewska ME, Byszewska A, et al. Assessing
of phacoemulsification on intraocular pressure in medically efficacy of canaloplasty using continuous 24-hour monitor-
controlled open-angle glaucoma patients. Am J Ophthalmol ing of ocular dimensional changes. Invest Ophthalmol Vis
2014; 157: 26–31. Sci 2016; 57: 2533–2542.
27. Iancu R and Corbu C. Intraocular pressure after phacoemul- 43. Lee JW, Fu L, Chan JC, et al. Twenty-four-hour intraocular
sification in patients with uncontrolled primary open angle pressure related changes following adjuvant selective laser
glaucoma. J Med Life 2014; 7: 11–16. trabeculoplasty for normal tension glaucoma. Medicine
28. Arthur SN, Cantor LB, WuDunn D, et al. Efficacy, safety, 2014; 93: e238.
and survival rates of IOP-lowering effect of phacoemulsi- 44. Tan S, Yu M, Baig N, et al. Circadian intraocular pressure
fication alone or combined with canaloplasty in glaucoma fluctuation and disease progression in primary angle closure
patients. J Glaucoma 2014; 23: 316–320. glaucoma. Invest Ophthalmol Vis Sci 2015(56): 4994–5005.

También podría gustarte