2 THERMAL ANALYSIS

GENERAL PRINCIPLES OF OPERATION

Thermal analysis refers to a variety of techniques in which a property of


a sample is continuously measured as the sample is programmed through
a predetermined temperature profile. Among the most common techni-
ques are thermal gravimetric analysis (TA) and differential scanning
calorimetry @SC).
In TA the mass loss versus increasing temperature of the sample is
recorded. The basic instrumental requirements are simple: a precision
balance, a programmable furnace, and a recorder (Figure 1). Modern
instruments, however, tend to be automated and include software for data
reduction. In addition, provisions are made for surrounding the sample
with an air, nitrogen, or an oxygen atmosphere.
In a DSC experiment the difference in energy input to a sample and
a reference material is measured while the sample and reference are
subjected to a controlled temperature program. DSC requires two cells
equipped with thermocouples in addition to a programmable furnace,
recorder, and gas controller. Automation is even more extensive than in
TA due to the more complicated nature of the instrumentation and
calculations.
A thermal analysis curve is interpreted by relating the measured
property versus temperature data to chemical and physical events
occurring in the sample. It is frequently a qualitative or comparative
technique.
In TA the mass loss can be due to such events as the volatilization
of liquids and the decomposition and evolution of gases from solids. The
onset of volatilization is proportional to the boiling point of the liquid.
The residue remaining at high temperature represents the percent ash
content of the sample. Figure 2 shows the TA spectrum of calcium
oxalate as an example.

17
18 Polymer Characterization

Atmosphere
Control

Furnace Sample Holder

Furnace
Temperature Recorder
Programmer

Temperature
Sensor Recording
Balance

Balance
Control

Figure 1. Typical components of a TA instrument.

Figure 2. Shows the TA spectrum of calcium oxalate.


Thermal Analysis 19

In DSC the measured energy differential corresponds to the heat


content (enthalpy) or the specific heat of the sample. DSC is often used
in conjunction with TA to determine if a reaction is endothermic, such
as melting, vaporization and sublimation, or exothermic, such as
oxidative degradation. It is also used to determine the glass transition
temperature of polymers. Liquids and solids can be analyzed by both
methods of thermal analysis. The sample size is usually limited to lo-
20 mg.
Thermal analysis can be used to characterize the physical and
chemical properties of a system under conditions that simulate real world
applications. It is not simply a sample composition technique.
Much of the data interpretation is empirical in nature and more than
one thermal method may be required to fully understand the chemical
and physical reactions occurring in a sample.
Condensation of volatile reaction products on the sample support
system of a TA can give rise to anomalous weight changes.

THERMAL ANALYSIS OF POLYMERS

A simple example of the relationship between “structure” and


“properties” is the effect of increasing molecular weight of a polymer on
its physical (mechanical) state; a progression from an oily liquid, to a
soft viscoelastic solid, to a hard, glassy elastic solid. Even seemingly
minor rearrangements of atomic structure can have dramatic effects as,
for example, the atactic and syndiotactic stereoisomers of polypropylene--
the first being a viscoelastic amorphous polymer at room temperature
while the second is a strong, fairly rigid plastic with a melting point
above 160°C. At high thermal energies conformational changes via bond
rotations are frequent on the time scale of typical processing operations
and the polymer behaves as a liquid (melt). At lower temperatures the
chains solidifies by either of two mechanisms: by ordered molecular
packing in a crystal lattice, crystdization, or by a gradual freezing out
of long range molecular motions, vitrification. These transformations,
which define the principal rheological regimes of mechanical behavior:
the melt, the rubbery state, and the semicrystalline and glassy amorphous
solids, are accompanied by transitions in thermodynamic properties at the
glass transition temperature, the crystalline melting, and the crystalli-
zation temperatures.
20 Polymer Characterization

Thermal analysis techniques are designed to measure the above


mentioned transitions both by measurements of heat capacity and
mechanical moduhrs (stiffness).

Differential Scanning Calorimetry @SC)

The DSC measures the power (heat energy per unit time) differential
between a small weighed sample of polymer (ea. 10 mg) in a sealed
aluminum pan referenced to an empty pan in order to maintain a zero
temperature differential between them during programmed heating and
cooling temperature scans. The technique is most often used for
characterizing the T,, T,, T,, and heat of fusion of polymers (Figure 3).
The technique can also be used for studying the kinetics of chemical
reactions, e.g., oxidation and decomposition. The conversion of a
measured heat of fusion can be converted to a % crystallinity provided,
of course, the heat of fusion for the 100% crystalline polymer is known.

Thermogravimetric Analysis (TGA)

TGA makes a continuous weighing of a small sample (ca 10 mg) in a


controlled atmosphere (e.g., air or nitrogen) as the temperature is
increased at a programmed linear rate. The thermogram shown in
Figure 4 illustrates weight losses due to desorption of gases (e.g.,
moisture) or decomposition (e.g., HBr loss from halobutyl, CO, from
calcium carbonate filler). TA is a very simple technique for quan-
titatively analyzing for filler content of a polymer compound (e.g.,
carbon black decomposed in air but not nitrogen). While oil can be
readily detected in the thermogram it almost always overlaps with the
temperature range of hydrocarbon polymer degradation. The curves
cannot be reliably deconvoluted since the actual decomposition range of
a polymer in a polymer blend can be affected by the sample morphology.

Thermomechanical Analysis (TMA)

TMA consists of a quartz probe which rests on top of a flat sample (a


few mms square) in a temperature controlled chamber. When setup in
neutral buoyancy (with ‘flat probe’) then as the temperature is increased
the probe rises in direct response to the expansion of the sample yielding
AH/At

I
crystalline melting
oxidation/
glass transition crystallization
(onset)
\ I area =

onset
+
Heating scan
(typically 2OWmin)

c /
exo -
-
Cooling scan

TEMPERATURE II
+
i
l Important characteristics: Ts , T,,, , heat of fusion on heating; Tc on cooling E
2v1’

Figure 3. Illustrates typical polymer DSC thermograms. z


Thermal Analysis 23

thermal expansion coefficient versus temperature scans. Alternatively,


with the ‘penetration probe’ under dead loading a thermal softening
profile is obtained (penetration distance versus temperature). Although
this is a simple and versatile experiment, it gives only a semi-quantitative
indication of mechanical modulus versus temperature. The DMTA,
described below, gives an absolute modulus measurement.

Dynamic Mechanical Thermal Analysis (DMTA)

DMTA is a measurement of the dynamic moduli (in-phase and out-of-


phase) in an oscillatory mechanical deformation experiment during a
programmed temperature scan at controlled frequency. Thermograms
are usually plotted to show elastic modulus, E, and tan 6 versus
temperature (Figure 5). The peak of the tan 6 is a particularly
discriminatory measure of T,, although this is the center of the relaxation
whereas in the DSC experiment the onset temperature of the T,
relaxation is usually reported. In such a case the DSC T, will be lower
than that for DMTA by an amount that varies with the specific polymer.
There is, in addition, a frequency effect which puts the mechanical (ca.
1 Hz) T, about 17°C higher than that for a DSC measurement (ca.
O.OOOlHz)for an assumed activation energy of 400 Id/mole (typical for
polymer TJ. The DMTA has a frequency multiplexing capability which
can be used for calculating activation energies using time-temperature
superposition software.
The temperature range of the DMTA is from -150°C to 300°C and
frequencies from 0.033 to 90 Hz. The sample size for the usual flexural
test mode is 1 mm x 10 mm x 40 mm; slightly less sample is required
in the parallel plate shear mode.
24 Polymer Characterization

cJr1ve Shaft Drive shaft


(selected amDlltude) (selected amplitude)

clamps

(a) Flexure mode (TV) Shear mode