Documentos de Académico
Documentos de Profesional
Documentos de Cultura
To Receiver
Analysis Zone: 140 to 356°F (60 to 180°C) provided that the desired oven temperature is at least
18°F (10°C) above ambient (stable oven temperature ±0.1°F (± 0.05°C) at steady
ambient ±1°F (± 0.5°C) for –4 to 122°F (–20 to 50°C) ambient change)
Contacting Material: Stainless steel or surface-deactivated steel for trace analysis of reactive compounds,
Teflon
Auxiliary Utilities: Instrument air-dry, oil-free, available at 50 PSIG min (3.5 bar) and 2 to 3 scfm (1 to
1.5 sl/sec); carrier gas, zero air, and hydrogen for FID; possible steam tracing or
electrical heating for sample lines
Cycle Time: 2 to 20 min, depending on application and packed or capillary column operation
Special Features: Accepts other inputs (e.g., flow rate and density in the calculation of output, e.g.,
BTU/h); built-in diagnostics with local and remote indication; multicomponent read-
out; stand-alone or networked; communication link (fieldbus or Ethernet directly to
the plant or local LAN and local control system); local and remote operator interface;
single- or multistream analysis
1258
© 2003 by Béla Lipták
8.12 Chromatographs: Gas 1259
FIG. 8.12a
A typical packed column chromatogram showing fully and partially resolved peaks. Retention time of an unretained sample component
(such as air) measured from the time of injection = tM . Retention time of sample components 1 to 8 = tR1 to tR8. The resolution of peaks 5
and 6 = 1.1. The number of theoretical plates calculated using peak 8 = 4158.
where w5 and w6 are the respective peak widths measured at retention times), a precisely regulated temperature environ-
the base of the peak. In this case, the resolution of the two ment must be provided for the column. In some cases, to
peaks is 1.1. Two compounds of equal concentration are minimize the analysis time for samples that contain a wide
baseline resolved at a resolution of 1.5. boiling range of compounds, the temperature of the oven is
The theoretical plate as calculated from a peak (number varied linearly with time. This procedure is called temperature-
8 in Figure 8.12a) using Equation 8.12(2) is found to be 4158 programmed chromatography.
and was obtained using a packed column. Although they employ the same principles, the process
gas chromatograph (PGC) is quite a different instrument than
t 2R 8 its laboratory counterpart. These differences appear as the
N = 16 8.12(2) result of a number of factors, not the least of which include:
w82
• The need for continuous, reliable operation of the unit
and the efficiency of a column of length, L, is • The need for the cycle time of the analysis to be shorter
than the time required for the process control system
N to achieve proper control action
hetp = 8.12(3)
L • The nature of the environment in which the PGC is
placed
The time required for each component to emerge is called • The diverse user attitudes regarding maintenance, and
the retention time, tR, and the magnitude of the detector signal their capability regarding the operation of the unit
(peak height or peak area) is related to the amount of the • The necessity to interface with modern-day computer-
compound present in the injected sample. If these individual controlled systems
retention times and detector responses are made to remain
constant (by adequate control of carrier gas flow rate, column Thus, a premium is placed on simple, reliable design to ensure
temperature, amount of stationary phase, and detector vari- a low mean failure rate, but straightforward, rapid repair when
ables), the process can be automated so that the instrument required, and the strategic placement of appropriate sensors
will provide this information quite precisely and cyclically to aid in diagnosing problems and to provide alarms with
on a repetitive basis. regard to the operation of the unit.
To summarize, the PGC has the following distinguishing
attributes:
BASIC CHROMATOGRAPHIC INSTRUMENTATION
• Located in the plant as close as possible to the sample
The basic instrumentation necessary to accomplish the point to minimize sample transport time
above-described chromatography is extremely simple (see • Dedicated to monitor one or more components in one
Figure 8.12b). or more process streams
It consists of a supply of carrier gas (commonly helium, • Designed for continuous, unattended operation
hydrogen, or nitrogen) with appropriate pressure regulation, • Designed for operation in hazardous environments
a sample introduction device (described later on) to inject a • Designed to withstand exposure to weather, humidity,
fixed volume of sample into the carrier gas stream, a column dust, and corrosive atmospheres
containing the stationary phase, and a detector. Since the • Contains integral hardware and software to allow com-
degree of solubility of the sample components in the station- munication with the process control system as well as
ary phase is temperature dependent (and thus so will be the remote maintenance stations
FIG. 8.12b
Basic elements of a chromatograph.
FIG. 8.12c
Basic elements of a multistream process gas chromatographic system.
Analyzer
FIG. 8.12f
Basic elements of an explosion-proof isothermal air bath oven
heater assembly.
FIG. 8.12h
An exploded view of the “transmitter” illustrates the modular construction of the unit. Rapid field maintenance of the analyzer is met by
replacement of the total oven assembly which includes columns, detectors, valves, actuators, heaters, temperature sensors, and electronic
pressure/flow controller components. In the electronics section, the printed circuit board assemblies are “euro-card” slide rail mounted and
may be removed/replaced without the use of tools. (Courtesy of Rosemont Analytical Inc.)
to form a mating seal with the rotor. The rotor is held against
the stator by a factory-calibrated threaded cap that applies
the correct force. An exploded view of a manually operated
version of such a valve is shown in Figure 8.12j.
The PGC version of this valve is actuated either by a
compressed air operator or by an electrically driven solenoid.
Although these valves meet the stringent demands for capil-
lary chromatography, there are disadvantages in that they
require regular maintenance, since the rotors wear and must
be replaced after about 200,000 cycles. In addition, they
FIG. 8.12i require a much more complicated driver since the simple linear
Ratary valve. motion commonly available must be converted to a rotary one.
The rotor faces are of filled Teflon or similar polymer Sliding Plate The sliding plate valve comes in both linear
that provides chemical resistance as well as sealing and and rotary versions, the former being the most popular by
lubricity. The valve body is made of stainless steel, or other virtue of the simple operator required for lateral rather than
corrosion-resistant alloy, with highly polished conical faces rotary movement of the slider.
FIG. 8.12j
Exploded view of a manually actuated rotary valve. (Courtesy of
Valco Instrument Co. Inc.)
TABLE 8.12m
Characteristics of Columns Used in PGCs
Type mls/min Length, m o.d. mm i.d. mm Flow
Packed
Conventional 0.1–5 3.18 (1/8") 2 10–50
Micropacked 0.02–1 1.59 (1/16") 1 1–10
Capillary 1–100 0.4–0.8 (1/64–1/32) 0.1–0.5 0.5–10
Hardware
FIG. 8.12p
A 10-port rotary valve with an external sample loop configured in Position B for the injection of sample into column 1 and column 2 and,
in Position A, for backflush to vent of the sample components trapped on column 1 while the rest of the sample is separated on column 2
and the sample loop is refilled with fresh sample. (Courtesy of Valco Instrument Co. Inc.)
FIG. 8.12q
Backflush to measure configuration: an 8-port rotary valve is configured so that the backflushed components can be flushed through the
detector and measured as a group. (Courtesy of Valco Instrument Co. Inc.)
FIG. 8.12s
FIG. 8.12r A six-port plate valve configured so that a group of early eluting
Heart-cut configuration: a six-port plate valve is configured so as sample components pass into a storage column which is then isolated
to divert a large portion of a major component in order to simplify from carrier gas flow (the group of components remain stationary
the separation of a small concentration of a sample component on the column except for a samll amount of longitudinal diffusion)
riding on its tail. and stored until the rest of the sample is separated and measured by
the detector. When the valve is again switched, the components on
the storage column are developed and measured by the detector.
Heart-cutting Heart-cutting is one of the most useful column
configurations. It is frequently used for the analysis of trace
to direct the effluent from column 1 into column 2 or through
components that elute immediately after a large concentration
a bypass restrictor to the detector for measurement. In a
of another sample component that interferes with its determina-
manner similar to that used in heart-cutting, two or more
tion (for example, trace amounts of acetylene in ethylene). The
components that are not separated on column 1 are diverted
design consists of two columns, a heart-cut column and an
to column 2, as shown in Figure 8.12s, normal position. In
analysis column, as shown in Figure 8.12r, using a six-port slid-
the other storage position of the valve, carrier gas is not
ing-plate valve. In the normal position of the heart-cut valve, the
available for this leg. Column 2 is thus isolated, and the
effluent of the heart-cut column is diverted to vent. When the
diverted compounds are stored while the rest of the sample
component of interest is about to elute, the heart-cut valve diverts
is developed in column 1 and passed on through the restrictor
it into the analysis column and then returns to its venting posi-
to the detector for measurement. The trapped compounds are
tion. A heart-cut of only the compound of interest and a narrow
then passed to the detector for measurement.
band or “tail” from the major component are thus introduced
into the analysis column. This operation may be repeated for
each component of interest, although it is usually not practical DETECTORS
to measure more than three heart-cut components per analysis.
The trap-and-store configuration consists of two analyt- Of the many detectors that have been proposed for use in PGC
ical columns arranged in series through a column-switching over the years, two emerge as the most popular and have been
valve, as shown in Figure 8.12s. The switching valve is used applied to a far-ranging number of applications. They are the
FIG. 8.12u
Simple constant current circuit for a thermistor bead TCD with a gain
of 5. (From Annino, R. and Villalobos, R., Process Gas Chromatog-
raphy: Fundamentals and Applications, Research Triangle Park, NC:
ISA, copyright 1992 ISA. Used with permission. All rights reserved.)
FIG. 8.12t
Single sensing filament, single reference filament, flow-through TCD
with replaceable elements shown in a classical Wheatstone bridge
circuit. In a properly designed TCD block assembly, the inclusion
of reference filaments imparts greater stability to the signal by
balancing out small temperature variations in the cell block.
Even though the use of a FID involves increased operating air supports the flame. In some designs, the flame points
cost and increased analyzer complexity, its very large downward. This ensures removal of water from the flame
7
dynamic range (10 ), high sensitivity (0.015 coulombs/g C), vicinity together with any solid particles that may be present.
–12
low detection limits (10 gC/sec), and vanishing small dead The presence of either of these substances will contribute to
volume (10 µl) have made it probably one of the most
–2
the background noise of the system and thus reduce minimum
popular GC detectors in use today. With optimized operating detection levels. The response of the detector depends on the
parameters, it is possible to determine as low as 20 pg, or flow rate of all three gases—with the hydrogen flow rate
about 5 ppb, of a sample. being the most critical (the ratio of air to hydrogen is usually
The FID consists of an oxygen-rich hydrogen flame that around 10).
burns organic molecules, producing ionized fragments. These The response of the FID, which varies with the identity
ions are subjected to an electrical field produced by impress- of the component in the flame, depends primarily on the
ing a potential (usually 150 to 300 V) across a jet and the number and type of carbon atoms being oxidized. However,
collecting electrode (the jet is usually made negative, and the the identity and manner in which other atoms are combined
collector positive). with the carbon also influences the response. The concept of
The essential components of a FID are shown schemat- effective carbon number (ECN) has been introduced as a
12
ically in Figure 8.12w. The hydrogen fuel is mixed with the means of relating the various responses. The ECN is the sum
column effluent and then fed to the jet. An outer sheath of of the contributions made by the individual carbon atoms
modified by the functional group contributions (a sample of
these ECNs is given in Table 8.12x).
A number of compounds cannot be detected by the FID.
These include the fixed gases, oxides of nitrogen, H2S, SO2,
COS, CS, CO, CO2, H2O, NH3, and the noble gases. However,
if the carrier gas is doped with an ionizable gas (such as
methane), even these gases can be detected (yielding negative
peaks as they dilute the concentration of ionizing gas during
the elution). Also, in the case of carbon dioxide, monoxide,
or formaldehyde, chemically reacting the gases as they emerge
from the column to form methane makes them FID respon-
sive. Thus, most suppliers of PGCs offer a methanizer, which
is placed between the end of the column and the FID. Hydro-
gen is also teed in at this point. The methanizer consists of
a length of 1/8-in.-o.d. tubing filled with a catalyst of 10%
(wt/wt) nickel on Chromosorb P, over which the reducing
reaction occurs. The reaction is nonselective in that all hydro-
carbons are converted to methane.
The FID is a very inefficient ionizing source, and it is
–14
only its very low noise level (10 A), allowing large ampli-
FIG. 8.12w
fications, that makes this detector so sensitive. As shown in
Schematic illustration of an FID assembly. The capillary column is Figure 8.12y, high-gain, low-noise, electrometers mounted
run up inside the jet to the tip (thus minimizing detector dead as close to the detector as possible are required to exploit the
volume). The collector is positively polarized usually with +300 full range of this detector. In addition, to maintain this low
volts. The flame is supported by an external sheath of air. noise level, highly purified gases should be used.
FIG. 8.12z
Basic flame photometric detector.
FIG. 8.12aa
Schematic of a FPD showing details of a dual flame assembly. The dual flame scheme is used to minimize the quenching of the sulfur
emission by other organic compounds and also to eliminate the flame quenching problem when large concentrations of organics are present.
It also provides the same sulfur response for all compounds since they are all oxidized to SO2 in the lower flame. (Courtesy of Varian
Associates.)
FIG. 8.12cc
Schematic view of a PFPD. (Reprinted with permission from Amirav,
A. and Jing, H., Anal. Chem. 67(18) 3305. Copyright © 1995 Amer-
ican Chemical Society.)
Orifice-Capillary Detector
Miscellaneous Detectors that it does not respond sensitively to many of the compounds
of interest, and it requires much maintenance to maintain
There are a number of other detectors that have been intro- quantitative accuracy.
duced since the birth of GC, but because of issues such as
high maintenance requirements, high customer burden, and Electron Capture Detector The electron capture detector
unreliable operation over extended periods in the field, they (ECD) (see Figure 8.12ff) is yet another type of ionization
have been limited primarily to laboratory applications. detector. The column effluent passes between two electrodes,
However, there are times when the minimum detection limits one of which has been treated with a radioactive source
of the application cannot be satisfied by the basic detectors (tritium or nickel-63; the latter is preferred because of its
normally used in process GC. In those cases, some of these extended detector stability) that emits high-energy electrons.
less-than-robust detectors have been used. These electrons produce large quantities of low-energy ther-
mal electrons in the carrier gas, which are in turn collected
Photoionization Detector The photoionization detector (PID) by the other electrode to produce a steady-state current in the
functions by irradiating the column effluent with high-energy presence of pure carrier gas. Compounds eluting from the
ultraviolet (UV) light generated by a high-voltage discharge column that have an affinity for thermal electrons reduce this
lamp containing a noble gas (e.g., krypton) (Figure 8.12ee). steady-state current, thereby producing the chromatogram.
The ions are collected at a polarized electrode, and the result-
ing current is measured with a FID-type electrometer. This
detector is selective in that only compounds whose ionization
potential is less than the UV radiation will be ionized. Com-
pounds with aromatic ring structures give the highest sensi-
tivities (as much as 10 times the FID), while compounds like
methane and ethane with ionization potentials greater than
12.98 eV give no response with the commonly used PID
tubes (9.5, 10.0, 10.2, 10.9, and 11.7 eV). The PID has a
7
dynamic range of 10 , extending from 2 pg to 30 mg. An
advantage to this detector is that it does not require auxiliary
gases, as does the FID. This advantage is offset by the fact
FIG. 8.12ff
FIG. 8.12ee Schematic of an ECD arranged for constant voltage operation. Also
Principal elements of a PID showing a 200 V supply for the col- shown is the chromatogram produced by the reduction of the stand-
lecting electrodes and a separate 400 V supply for the UV lamp. ing current when compounds of higher electron affinity than the
(From Driscol, J.N. and Spaziani, F.F., “PID Development Gives carrier gas are eluted from the column and pass through the detec-
New Performance Levels,” Research/Development, May 1976.) tor. (Courtesy of Varian Associates.)
The detector is thus highly selective, with halogenated com- of strontium, nickel, or tritium to produce metastable helium
pounds being the most responsive (detection at the picogram atoms by collisions with the high-energy electrons gener-
level). Other groups exhibiting good selectivity include anhy- ated by the source, the newer designs require only helium.
drides, peroxides, conjugated carbonyls, nitriles and nitrates, Two electrodes support a low-current arc through the helium
and sulfur-containing compounds. makeup gas flow. The helium molecules between the elec-
Maintenance is of critical importance with this detector, trodes are elevated from their ground state to form a helium
more so than with any other GC detector (except the helium plasma cloud. As the helium molecules collapse back down
ionization detector, discussed next). It responds exceptionally to the ground state, they give off a high-energy photon that
well to oxygen, necessitating leak-free systems and oxygen- will ionize all compounds having a ionization potential
free carrier gases. Also, response to water vapor can cause lower than 17.7 eV. Thus, the ionization detector will
unstable baselines so that molecular sieve traps (which require respond to volatile inorganics that the FID does not (such
periodic maintenance) are required in the carrier gas lines. as NOx, CO, CO2, O2, N2, H2S, and H2). The ions are
Finally, care must be taken to use this detector only with collected at polarized electrodes to produce a current, as in
columns of very low bleed, as the condensed stationary phase any ionization detector.
in the detector can easily be polymerized (by radiation and One version of this detector utilizes a steady-state helium
electron bombardment) to a hard insoluble deposit, which is plasma cloud, while a more recent design (see Figure 8.12gg)
almost impossible to eliminate and which interferes with the pulses the discharge (pulsed discharge detector (PDD)). Per-
proper functioning of the detector. formance of the discharge ionization detectors is equal to or
Although a pulsed mode of operation has improved the better than that of the older-design ionization detectors that
linear range of the ECD, it has not improved its robustness, require radioactive sources. In its helium photoionization
and all of the above-outlined concerns still apply. mode of operation, it is a universal, nondestructive, high-
sensitivity detector responding to fixed gases in the low parts
Discharge Ionization Detectors In contrast to the older- per billion range. Its response to both inorganic and organic
5
design ionization detectors that required a radioactive source compounds is linear over a wide range (10 ).
FIG. 8.12gg
Schematic of a pulsed discharge detector configured as a PHID and as a PECD. A stable low power, pulsed DC discharge, in helium is
utilized as the ionization source. Eluents from the GC column flowing counter to the flow of helium from the discharge zone are ionized
by photons from the helium discharge and the resulting electrons are focused toward the collector electrode by the two bias electrodes. The
principal mode of ionization is photoionization in the range of 13.5 to 17.7 eV. (Courtesy of Valco Instrument Co. Inc.)
The PDD is extremely versatile. If the carrier gas is doped that are sometimes used. Therefore, it is important that this
with methane at the column exit (see Figure 8.12gg), the PDD regulator have a minimum temperature coefficient (not as
can operate as an electron capture detector with sensitivity important a requirement when the analyzer is mounted in
and response characteristics similar to those of a conventional an analyzer shelter that has a fairly stable temperature envi-
radioactive ECD. Its MDQ for halogenated compounds ronment) and that it can tolerate the input pressure changes
–15 –12
ranges from 10 to 10 g. it will experience from the tank regulator that is likely to be
In addition, when the helium discharge gas is doped with located outside, exposed to the varying temperatures of the
a suitable noble gas such as argon, krypton, or xenon (depend- environment.
ing on the desired final ionizing power), the PDD can function For pressure drops across the column of 75 psi (5.3 bars),
as a specific PID for selective determination of aliphatics, the relative error in peak area will be 1.7 times the relative
aromatics, etc., as discussed in the section devoted to photo- error in the pressure drop. Therefore, to limit the error from
ionization detectors. this source to 0.5%, one must control pressure to within 0.4
The HID is a high-maintenance item. It is the only detec- PSIG (0.03 bars). In addition, assuming one is using a pres-
tor that can measure permanent gases at the 1-ppb level. sure regulator that is referenced to ambient pressure, the
Therefore, leak-free plumbing is an absolute necessity. It error caused by variations in ambient pressure will be 1.4
responds to trace impurities, including water, in the carrier times the relative barometric pressure variation. Thus, errors
gas, thus requiring traps in the carrier gas stream to ensure of 0.2 to 0.5% due to changes in weather are difficult to
16
the removal of these trace impurities. Clearly, this is a dis- eliminate.
advantage for users not willing to devote maintenance time Recently, electronically controlled pressure regulators
to monitor and frequently replace these traps. In addition, the have become available and have been incorporated into
HID is easily fouled by deposits from column bleed, which PGCs. The heart of such a regulator is a solid-state pressure
also contributes to a high noise level. Thus, its lowest detec- transducer that produces an electrical output proportional to
tion limit is reached only with active solid supports such as pressure. This output is used to operate the integral control
molecular sieves. valve and to produce an output signal. In the case of the
transmitter, miniature fluidic thermistors (fluistors) are used
as pressure/flow controllers. Such an addition to the PGC is
a tremendous advantage for maintenance staff since it allows
CARRIER GAS FLOW CONTROL for control of carrier gas flow rate from the remote analyzer
maintenance station (AMS) rather than at the unit.
Controlling carrier gas flow rate is of critical importance if
one is to reproduce peak areas and retention times, and thus
automate the analysis. Two methods, pressure control and
flow control, may be used, but pressure control is by far the PROGRAMMER-CONTROLLER
favored procedure used in process gas chromatographic
instrumentation, for the following reasons: The PC section is an extremely important part of a modern
PGC, containing all the electronics to power the system, the
1. A pressure regulator has an almost immediate response controller, the data reduction hardware and software, and, of
to downstream upsets (such as valve switching). A flow increasing importance, the communications package. Its
controller responds to flow upset only as fast as its design has changed radically over the past few years with the
flow set point allows. advances in electronic technology.
2. A pressure regulator has the ability to provide some The PC unit can in theory be located in the field at the
compensation for variation in retention time caused analyzer (stand-alone) or in a remote area such as the
by changes in temperature. control room or instrument room close by (split architec-
3. A pressure regulator has the ability to supply the ture). The stand-alone design is by far the most popular.
proper carrier gas flow rate to the column even in the Although in this option the analyzer electronics are also
presence of downstream leaks between it and the sam- usually located within the PC section, the primary purpose
ple valve. of the PC is to control all the functions in the analyzer.
These include sample injection and column switching as
All pressure regulators are, to some degree, sensitive to well as various housekeeping tasks, such as modification
temperature and input pressure. Tank regulators will not of the detector signal, auto-zeroing, peak gating, integra-
provide sufficiently stable output pressures under varying tion, conversion of area units to concentration, and data
conditions of temperature and tank pressure. Thus, another transfer to the proper location. The transmitter utilizes what
regulator is mounted in the line close-coupled to the ana- looks like a split-architecture design in that the electronics
lyzer. Ideally, it should be mounted in the analyzer oven are packaged separately from the analyzer, but the two are
compartment to ensure a stable temperature environment, close-coupled to each other in the field to yield a stand-
but few if any can withstand the 398°F (200°C) temperatures alone PGC.
*
These are Instrument Society of America (ISA) definitions of purge sys-
Communication with the plant’s distributed control system
tems. X purge allows a general-purpose instrument to be placed in a Division (DCS) is a must for modern PGCs. However, because of the
1 area, while Y purge allows for a Division 2-rated instrument to be used large number of different DCSs in place—all using commu-
in a Division 1 area. nication systems proprietary to the manufacturer of that
FIG. 8.12hh
I/O characteristics of a typical microprocessor-based PGC.
system—direct interfacing with all networks is all but impos- network is a digital, bidirectional, multidrop, serial-bus com-
sible for the user to ascertain that a gateway is available to munication network used to link isolated field devices. The
interface the PGC to his or her control system. bidirectional specification allows data to be transmitted in
Over the years there has been much discussion and two directions at the same time, multidrop can be interpreted
research into the development of a digital communication as a single bus with many nodes connected to it, and serial
system to replace the traditional 4- to 20-mA analog standard. means that the data are transmitted serially according to RS-
Worldwide adoption of an industry standard is still in the 232 or RS-485 protocol.
future. However, emerging from the various discussions Currently, there are two advanced protocols for all digital
within standards groups is a consensus that the local PGC will communication with field devices for measurement and con-
never be connected directly to the main data highway, but trol of continuous processes. These are the Foundation Field-
rather, communications will take place on a so-called fieldbus. bus (FF) and the Profibus-PA (also in the field is a hybrid
Fieldbus is a generic term that describes a new digital analog/digital protocol called HART, which uses a frequency-
communication network that will be used in industry to based digital signal superimposed on the 4- to 20-mA analog
replace the existing 4- to 20-ma analog signal standard. The signal).
FIG. 8.12ii
Illustration of the various hardware and communication configurations that must be supported by a PGC. (Courtesy of The Foxboro Co.)
Operator Interface
Quantitation
FIG. 8.12ll
A typical multi-PGC interface to the DCS. Some variation will exist among the various vendors. In this configuration only PGC data,
validation, system alarms, and stream sequence control are available to the DCS operator. All control and data communication is available
to the AMS operator. (Courtesy of The Foxboro Co.)
it to the analyzer, and return it either to the process stream volumes that are difficult to purge, thus leading to decreased
at a lower pressure point or to a suitable waste disposal sample system performance.
system. The time necessary to transport the sample to the Recently, there has been some initiative directed toward
analyzer can, in some cases, constitute the largest share of changing the above-discussed conventional design to fit the
17,18
the system dead time or turnaround time. Mounting the ana- specific requirements of the process analyzer. The poten-
lyzer closer to the sampling point is the most direct way of tial of a smart modular miniature process analyzer sampling
decreasing this time. Lacking this alternative, the next best system integral to the analyzer is being explored with the
solution is to increase the flow sample in the line. An added idea of decreasing the installed cost of the analyzer, increas-
plus to using a so-called fast loop is that it provides additional ing reliability of the system, and, at the same time, decreasing
mixing due to turbulent flow, and thus further ensures a cost of manufacture. One problem with sampling systems,
representative sample. however, is that they are application dependent. Thus, the
Sample transport time is a function of sample line length, design must allow for the convenient connection of simple
line diameter, the absolute pressure on the line, and the sam- modules to form the more complex one required for the par-
ple flow rate. Neglecting compressibility, the transport time ticular application. A proposed design (borrowed from the gas
for a gas sample is given as management systems used in the semiconductor industry)
19
shown in Figure 8.12nn uses a substrate assembly that pro-
vides the flow path for a process sample and consists of a
VL(p + 14.7)(530)
t lag = 8.12(4) variety of sample conditioning components. In turn, it is
F(14.7)(t + 460) attached to a manifold system that will also accept various
flow components and provide a flow path between two or
where more substrates.
20,21
V = volume of sample per unit length, ml/cm Thus far, implementation of these ideas has been
L = length of sample line, cm restricted to fairly simple and straightforward applications
F = volume flow rate (ml/min) of sample under standard 22
such as clean, dry, light hydrocarbon streams. The SHS used
conditions (14.7 PSIA, 70 °F) in this case is diagrammed in Figure 8.12oo and consists of
t = temperature of sample, °F four substrate assemblies manifolded together to produce the
p = sample line pressure, PSIG product shown in Figure 8.12pp.
The PGC transmitter described previously is offered as
Additional time must be allowed to purge sample condi- a single- or dual-process stream analyzer (with two sample
tioning equipment and any unswept volumes that are present calibration streams) complete with a modular sample han-
due to poor design. Commonly, systems are designed to dling system for close-coupling to the sample tap.
provide sample flow rates between 0.5 and 15 GPM (2 and Unfortunately, sampling systems usually fail during pro-
50 l/min). cess upset conditions (just when the analysis is most needed
to bring the process back into control). Thus, whenever pos-
Sample Conditioning sible, these systems should be overdesigned to be able to
handle the upset condition. A process data sheet containing
Conditioning of the sample either to allow easier handling the information summarized in Table 8.12qq is required to
or to put it into a form acceptable to the analyzer is always design a proper sample conditioning system. This definition
required to some degree. Conditioning may be as simple as of all the process conditions such as pressure, temperature,
pressure reduction and filtering or as complex as scrubbing phase, particulates, chemical composition, physical properties
and drying. Pressure reduction is the most effective way to (viscosity, density, etc.), and any possible process upsets that
prevent condensation of a condensable gas sample, and main- might occur, such as breakthrough of contaminants and run-
tenance of a minimum pressure is the most effective way to away thermal reactions, ensures a design that will guarantee
prevent a liquid sample from vaporizing. Thus, pressure reg- a correct sample presentation to the analyzer or, alternatively,
ulators for vaporization and pressure reduction are a common a controlled shutdown or process hold with accompanying
element of a conditioning system. In all cases, conditioning alarm signal.
must be accomplished without affecting the sample compo-
sition. Some common sample conditioning elements are Multistream Analysis
given in Table 8.12mm.
The basic technology of sample conditioning has not The PGC must be able to accommodate at least two streams,
changed much over the last 40 years. Currently, a SCS is the process stream and the calibration or standard line. The
constructed using components (listed in Table 8.12mm) that clean calibration sample can bypass the SHS and be plumbed
have evolved from industries such as pneumatic and hydrau- directly into a stream-switching valve that will allow the
lic service. These are mounted in a cabinet (which is usually process stream to be diverted while the standard sample is
secured to the outside wall of the analyzer shelter) and inter- flowing through the analyzer sample loop (Figure 8.12rr).
connected with tubing and fittings. Not being designed spe- The sharing of one PGC among a number of process
cifically to fit process analyzer needs, they have large dead streams is not recommended but, unfortunately, is common
TABLE 8.12mm
Common SHS/SCS Elements
Element Function/Comments
Filter Remove particulates from the sample. Filtering below 10 microns requires
vigilant preventive maintenance unless some sort of self-cleaning filter can be
devised (e.g., cyclone separation with automatic cleanout).
Vaporizing regulator If possible, at least in the case of a PGC, it is best to present the sample as a
vapor. The vaporizing regulator is best located at the sampling point (to provide
short transport times for the sample), but it can be placed at the analyzer.
Condensors/separators Used in applications where condensables are to be removed from the vapor
sample
Coalescers Used to force finely divided liquid droplets to combine into larger droplets so
they can be separated by gravity
Knockout pots Used to collect liquid that has been separated from the sample. Should be
provided with automatic drains.
Aspirators, ejectors This is the preferred pump (by creating a vacuum) if the sample pressure is not
sufficient to drive the sampling system. It is inexpensive and quite reliable
because of the absence of moving parts
Rotameters Consists of a ball or float in a tapered tube. Commonly used to measure flow. It
is quite unreliable as the float tends to stick when the steam becomes dirty. It
is a good indirect indicator of filter integrity.
Pressure regulators Nonbleeding type, using corrosion resistant stainless steel and Teflon. Used to
regulate pressure between sampling point and the analyzer as a means of
controlling sample flow rate. Since the quantity of sample is directly
proportional to the sample in the loop, in most cases a “block and bleed”
configuration is used at the analyzer to equilibrate the sample to atmospheric
pressure (this still does not eliminate inaccuracies due to changes in atmospheric
pressure).
Pressure gauges Installed downstream from regulators as an aid in setting regulator pressure and
as an aid in maintenance checks. Installing in high speed by-pass lines and
downstream of the analyzer eliminates extended turnaround problems that can
occur because of the time required to flush sample from the pressure gauge
bourdon tubes.
Steam or electrical heat To avoid condensation in the sample lines.
tracing
FIG. 8.12nn
Illustration of the architecture of a modular SHS with the various miniature SHS components surface mounted amd interconnected on a
so-called substrate. Individual substrates can then be manifolded together to form the comple sample handling system. (Photograph ©2002
Swagelok Company; text ©2002 R. Annino.)
FIG. 8.12oo
Schematic of a simple SHS circuit for one process stream and a calibration sample. (From Goedert, M. and Guiochon, G., Journal of
Chromatographic Science, 7, 323, 1969.)
FIG. 8.12pp
Hardware implementation of the circuit shown in Figure 8.12oo minus the miniature flow controllers and the pressure indicator (guage).
(From Goedert, M. and Guiochon, G., Journal of Chromatographic Science, 7, 323, 1969.)
practice in the industry as a way of justifying the cost of the continuously, even in the unselected lines, thus ensuring up-
PGC. It is questionable if the savings are real when one con- to-date sample composition when selected. Sample stream 2
siders the increased complexity of the SHS/SCS (one for each is selected in this case and continuously flushes the PGC
stream) and the maintenance problems that are incurred. Add sample loop and all other sample lines.
to this the fact that if the analyzer goes down, the analysis of
more than one stream is affected—a result that may have more Sample Disposal
impact on product cost than purchasing separate analyzers. The
availability of a lower installed cost PGC with its simpler The vent of the PGC is almost totally carrier gas and is
modular SHS may affect a shift in this paradigm in the future. usually vented to the atmosphere, as is the sample loop vent
In any case, stream switching requires a design that pre- line, unless the sample is toxic or otherwise environmentally
vents mixing of the process stream samples. The double- harmful. In such cases, the sample must be vented to a spe-
block-and-bleed-with-bypass purge illustrated in Figure cific waste disposal area. Alternatively, if a low-pressure point
8.12rr is probably the most efficient circuit for accomplishing in the process stream is available, the sample can be returned
this objective. In this circuit, all sample streams are flowing to the process. Since the accuracy of the analysis is directly
TABLE 8.12qq
Minimum Process Sample Information Required for Proper SHS Design
Liquid Vapor
Process pressure and temperature Process pressure and temperature
Bubble point at the process or highest ambient temperature Dew point and lowest ambient temperature
(whichever is highest) and atmospheric pressure
Viscosity Average molecular weight
Specific gravity Specific gravity
Sample composition under normal and upset conditions Sample composition under normal and upset conditions
Sample return point pressure Sample return point pressure
Distance between sampling point and the analyzer Distance between sampling point and the analyzer
Required sample turnaround time (lag time of the SHS/SCS) Required sample turnaround time (lag time of the SHS/SCS)
Maximum pressure for which the PGC sample valve is specified Maximum pressure for which the PGC sample valve is specified
Whether the PGC valve can accept either liquid or vapor sample Whether the PGC valve can accept either liquid or vapor sample
Minimum flow rate required through the PGC sample loop Minimum flow rate required through the PGC sample loop
Corrosiveness of the sample (related to material compatibility) Corrosiveness of the sample (related to material compatibility)
Sample toxicity Sample toxicity
INSTALLATION
TABLE 8.12ss
Typical Utility Requirement
Item Specification
Instrument air 4 scfm per heated zone, 2 scfm purge
Instrument air quality Clean, dry, −40°C dew point, oil free, particles ≤ 5 ppm
microns, ISA grade hydrocarbon free
Carrier and other gases Compressed gases, application dependent, may include helium,
nitrogen, hydrogen, air, etc.
Compressed gas quality Hydrogen, 99.995% ultra pure grade
Burner air, ≤ 1 ppm hydrocarbons and ≤ 5 ppm water
Helium, ≤ 99.995% ultra pure grade
Nitrogen, 99.995% ultra pure grade
AC power 115 VAC, 50/60 Hz 10 amp service (per oven) or
220/230 VAC, 50/60 Hz
TABLE 8.12tt
Breakdown of Installation Costs
Item Traditional PGC PGC Transmitter
Analyzer with basic SHS 45,000 33,000
Analyzer shelter (1/4 pro rata) 35,000 NA
Installation and startup 20,000 5,000
Total* 100,000 38,000
* This is a minimum installation cost as each unit may require specific hardware/software
depending on the specific PGC and the application.