ily Designation: C 1064 - 86 (Reapproved 1993)
Standard Test Method for
Temperature of Freshly Mixed Portland Cement Concrete’
‘This standard is isued under the xed designation C 164; the numb immediatly ellowing the designation indicates the yar of|
crignaladopton on the case of revision the year of last revson
"A urbe in parentheses indicates the yea fast approval A
Sdpecspt eto ()Hdates wt itil cage ine Ue lt tevaion on seapporal
1. Scope
LLL This test method covers the determination of temper-
sture of freshly mixed portland cement concrete
12 The values stated in inch-pound units are to be
rreardel as the standard
1.3 This standard may involve hazardous materials, oper-
ations, and equipment. This standard does not purport to
address all of the safety problems associated with its use. Its,
the responsibility of whoever uses this standard to consult and
ssablish appropriate safety and health practices and deter-
mine the applicability of regulatory limitations prior to use
2. Referenced Documents
2.1 ASTM Standards
C172 Practice for Sampling Freshly Mixed Concrete”
E 1 Specification for ASTM Thermometers?
E77 Test Method for the Inspection and Verification of
Thermometers!
22 NBS Standard:
N.BS. Monograph 150 Liquid-in-Glass Thermometry*
4. Significance and Use
3.1 ‘This test method provides a means for measuring the
temperature of freshly mixed concrete. It may be used to
veiily conformance to a specified requirement for tempera
ture of concrete.
32 Conerete containing aggregate of a nominal max~
imum size greater than 3 in. (75 mm) may require up to 20
min for the transfer of heat from aggregate to mortar. (See
ACI Committee 207.1R Report.)
4 Apparatus
4.1 Container—The container shall be made of
sonasorptive material and large enough to provide at least 3,
in, (75 mm) of concrete in all directions around the sensor of
the temperature measuring device; concrete cover must also
‘be at Ieast three times the nominal maximum size of the
cnarse aggregate,
42 Temperature Measuring Device—The temperature
"This tes method is under the jurisdiction of ASTM Commitee C9 on
‘ee CO9.40 on Fresh Concrete Testing
(Cine edton approved July 17,1986. Published Jay 1986
Namal Book of #S7M Standards, Vai 0402
2a! Book of ASTM Standards, Vol 1403,
‘Avalable fom National Burns of Standard, US. Department of Com
Avalable from Amerian Concrete Isis, Box 19150, Reford Station,
Dea, MI $8219, Other relted documents ako available fom American
(Gace alse we Consntice Repent 305 (Tot Weather Coneeing) ad 306
(ei Weather Conersing)
sa
‘measuring device shall be capable of measuring the temper-
ature of the freshly mixed concrete to +1°F (40.5°C)
‘throughont the entire temperature range likely to be encoun-
tered in the fresh concrete. ASTM liquid-in-glass thermom-
eters having a range from 0 to 120°F (~18 to 49°C) and
conforming to the requirements for ASTM thermometer No.
36°C as prescribed in Specification E 1 are satisfactory. Other
thermometers of the required accuracy, including the metal
immersion type, are acceptable.
4.3 Panial immersion liquid-in-glass thermometers (and.
possibly other types) shall have a permanent mark to which
the device must be immersed without applying a correction
factor.
4.4 Reference Temperature Measuring Devico—The refer-
cence temperature measuring device shall be a liquid-in-glass
thermometer readable to 0.5°F (0.2°C) that has been verified
and calibrated in accordance with Method E 77. The calibra-
tion certificate or report shall be available for inspection,
5. Calibration of Temperature Measuring Device
5.1 Each temperature measuring device used for deter-
mining temperature of fiestily uixed concrete shall be
calibrated annually, or whenever there is a question of
accuracy. This calibration shall be performed by comparing
the readings on the temperature measuring device at two
temperatures at least 30°F (15°C) apart
5.2 Calibration of the temperature measuring devices may
bbe made in oil or other suitable baths having uniform density
if provision is made to:
5.2.1 Maintain the bath temperature constant within
U.5°F (0.2°C) during the period of the test.
5.2.2 Have both the temperature and reference tempera-
‘ure measuring deviees maintained in the bath for a min-
imum of 5 min before reading temperatures.
5.2.3 Continuously circulate the bath liquid to provide a
uniform temperature.
5.24 Slightly tap thermometers containing liquid to avoid
adhesion of the liquid to the glass if the temperature
exposure is being reduced.
5.3 If a limiting temperature is specified, calibrate the
measuring device at a temperature within +5°F (2°C) of the
limiting temperature permitted.
6. Sampling Concrete
6.1 The temperature of freshly mixed concrete may be
‘measured in the transporting equipment provided the sensor
of the temperature measuring device has at least 3 in. (75
‘mm) of concrete cover in all directions around it.@ c 1064
6.2 Temperature of the freshly mixed concrete may be
obtained following conerete placement using the forms as the
container.
6.3 TF the transporting eqnipment or placement forms are
not used as the container, a sample shall be prepared as
follows:
6.3.1 Immediately, prior to sampling the freshly mixed
concrete, dampen (with water) the sample container.
6.3.2 Sample the freshly mixed conerere in accordance
with Method C 172, except that composite samples are not
required if the only’ purpose for obtaining the sample is to
determine temperature.
6.33 Place the freshly mixed concrete into the container.
6.3.4 When concrete contains a nominal maximum size
of aggregate greater than 3 in. (75 mm), it may require 20
min before the temperature is stabilized after mixing,
7. Procedure
7.1. Place the temperature measuring device in the freshly
mixed concrete so that the temperature sensing portion is
submerged a minimum of 3 in. (75 mm). Gently press the
conerete around the temperature measuring device at the
surface of the concrete so that ambient air temperature does
not affect the reading.
7.2 Leave the temperature measuring device in the freshly
mixed concrete for a minimum period of 2 min or until the
temperature reading stabilizes, then read and record the
temperature.
7.3. Complete the temperature measurement of the freshly
mixed concrete within 5 min after obtaining the sample.
8. Report
8.1 Record the measured temperature of the freshly
mixed concrete to the nearest I°F (0.5°C).
9. Precision and Bins
9.1 The precision and bias of this test method have not
been determined. A precision and bias statement will be
included when sufficient test data have been obtained and
analyzed,
‘Te American Society fr Testi ané Materia takes no ooston respecting the val of any pater ight esate in comeaction
wi any tem mentioned inthis standard Usors of this standard are expressly advised that determination ofthe val of any suc
patent rts, andthe isk of iningemert of such rights, are ene ther own responsi
‘Ts standard e sujet ro revision at any tne bythe espensibeechrca commit and must be reviewed oer ve years and
\Ynot revised, ater reaporoved er witharewn. Your commons areivtd ether fo reision os sandard orf ado tana
‘ano spew be soaressea 0 ASTHMoasuares. Your Commas wil recoWe care consteoul a asi of he resort
{echnlealcommtee, which you may ation. I you fa that your carmen hav net rected fa hearing you shout make your
ews knew 1 the ASTM Commitoo on Standards, 1976 Race St, Phisdetphi, PA 19103
522
dip >
1. Scope
LA thi
reducing
develope
the main
that exhi
reduce th
12 Th
for plans
laborator
good enc
provides
‘Annex A
12.7
the time
interiabe
C802, 0
14'T.
ations, «
address
the respe
establish
mine th
2. Refe
24
Corn
Me
C20.
Pr
for
E45
E69
"Th
sand Con
wesc
‘cur
Pan
sam