Está en la página 1de 34

Bubble CPAP System

PRODUCT CATALOGUE
BubbleCPAP
Contents

Section
Delivery Systems 1
Diagrams 1.1
Infant Delivery System (BC151) 1.2
Infant Delivery System (BC161) 1.2
Infant Delivery System (BC171) 1.2
Patient Interface 2
Nasal Tubing Midline 2.1
Nasal Tubing Universal 2.1
Nasal Prongs 2.1
Infant Bonnet Midline 2.1
Headgear Midline 2.1
Chinstrap 2.1
Bubble CPAP System Starter Kits 3
Starter Kit Midline Delivery System 3.1
Starter Kit I/F (Flow Driver compatible) Delivery System 3.1
Nasal Prong Starter Kit 3.2
Infant Bonnet Starter Kit 3.2
Starter Kit Midline F&P 50mm 3.2
Starter Kit Midline F&P 70mm 3.2
Starter Kit Midline F&P 100mm 3.2
Starter Kit Midline Universal 50mm 3.3
Starter Kit Midline Universal 70mm 3.3
Starter Kit Midline Universal 100mm 3.3
Accessories / Devices 4
Single-Use
Gas Supply Line 2.1m 4.1
Gas Supply Line to 2 flow meters 4.1
Midline Pressure Luer Adaptor 4.1
Mounting Options
Humidifier Stand Telescopic 4.2
Mounting Bracket C-Clamp 4.2
Mounting Bracket Pole 15-50mm 4.2
Humidification
Humidifier 4.3
Heater Wire Adaptor MR850 4.3
Heater Wire/Temp Probe Kit 4.3
Probe Temp/Flow Ang 1600mm 4.3
Gas Supply
Bird MBlender 2-16LPM DISS 4.3
Bird MBlender 2-16LPM NIST 4.3
Bird MBlender Pole Mount Bracket 4.3
Sample Setups 5
Pole Mount Setup 5.1
Infant Warmer Setup 5.2
BubbleCPAP
Delivery Systems 1.1

BC161

BC151

10
9
8
7
6

BC171
10
9
8
7
6
BubbleCPAP
Delivery Systems 1.2

Product Code Description

BC151-10 Infant Delivery System


Includes MR290 chamber
Pressure Manifold
CPAP Generator
Single-heated Circuit

BC161-10 Infant Delivery System


Includes MR290 chamber
MR290 Chamber
Pressure Manifold
CPAP Generator
Single-heated Circuit

BC171-10* Infant Delivery System


Includes MR290 chamber
Simplified Pressure Manifold (with attached dry line)
CPAP Generator
Single-heated Circuit
Pressure Damping Filter

* Flow Driver compatible circuit

Pressure Manifold

Simplified Pressure
Manifold
(BC171-10 only)

Single Heated Circuit

10
9
8
7
6
5

Pressure Damping Filter CPAP Generator


(BC171-10 only)
BubbleCPAP
Patient Interface 2.1

Length in mm

Product Code Description Size

Nasal Tubing Midline Nasal Tubing Length


(BC18x series) BC180-05 Nasal tubing 50mm x5 50mm
BC181-05 Nasal tubing 70mm x5 70mm
BC182-05 Nasal tubing 100mm x5 100mm
BC190-05 Nasal tubing Universal 50mm x5 50mm
BC191-05 Nasal tubing Universal 70mm x5 70mm
BC192-05 Nasal tubing Universal 100mm x5 100mm
Nasal Tubing Universal
(BC19x series) Nasal Prongs diameter of nares /
width of septum
Diameter of nares BC3020-10 Nasal prong 020
3 10/bx 3.0mm / 2.0mm
Septum gap BC3520-10 Nasal prong 3520 10/bx 3.5mm / 2.0mm
BC4030-10 Nasal prong 4030 10/bx 4.0mm / 3.0mm
BC4540-10 Nasal prong 4540 10/bx 4.5mm / 4.0mm
BC5040-10 Nasal prong 5040 10/bx 5.0mm / 4.0mm
BC5050-10 Nasal prong 5050 10/bx 5.0mm / 5.0mm
BC5550-10 Nasal prong 5550 10/bx 5.5mm / 5.0mm
BC5560-10 Nasal prong 5560 10/bx 5.5mm / 6.0mm
Nasal Prongs BC6060-10 Nasal prong 6060 10/bx 6.0mm / 6.0mm
BC6070-10 Nasal prong 6070 10/bx 6.0mm / 7.0mm
BC6570-10 Nasal prong 6570 10/bx 6.5mm / 7.0mm

Product Code Description Size

Infant Bonnet head circumference


BC300-05 Infant Bonnet Midline 17-22cm x5 17cm-22cm
BC303-05 Infant Bonnet Midline 22-25cm x5 22cm-25cm
BC306-05 Infant Bonnet Midline 25-29cm x5 25cm-29cm
BC309-05 Infant Bonnet Midline 29-36cm x5 29cm-36cm

Infant Bonnet

Product Code Description Size

Headgear head circumference


BC325-05 Headgear Midline 29-36cm x5 29cm-36cm
BC328-05 Headgear Midline 35-40cm x5 35cm-40cm
BC331-05 Headgear Midline 40-45cm x5 40cm-45cm

Chinstrap face circumference


BC351-10 Chinstrap 20-26cm x10 20cm-26cm
BC353-10 Chinstrap 26-32cm x10 26cm-32cm
Headgear BC355-10 Chinstrap 32-38cm x10 32cm-38cm
BC357-10 Chinstrap 38-44cm x10 38cm-44cm

Chinstrap
BubbleCPAP
Bubble CPAP System Starter Kits 3.1

Product Code Description

BC461-SK Starter Kit Midline Delivery System


Includes
5 BC161 Infant Delivery System
Nasal Tubing 5 Nasal tubing 70mm
2 Infant Bonnet 17-22cm
3 Infant Bonnet 22-25cm
3 Infant Bonnet 25-29cm
2 Infant Bonnet 29-36cm
2 Nasal Prong 3520
3 Nasal Prong 4030
4 Nasal Prong 4540
4 Nasal Prong 5040
4 Nasal Prong 5050
4 Nasal Prong 5550
4 Nasal Prong 5560
3 Nasal Prong 6060
Infant Bonnet 2 Nasal Prong 6070

BC471-SK* Starter Kit I/F Delivery System


Includes
5 BC171 Infant Delivery System
(with simplified pressure manifold and pressure damping filter)
5 Nasal tubing 70mm
2 Infant Bonnet 17-22cm
3 Infant Bonnet 22-25cm
Nasal Prongs
3 Infant Bonnet 25-29cm
2 Infant Bonnet 29-36cm
2 Nasal Prong 3520
3 Nasal Prong 4030
4 Nasal Prong 4540
4 Nasal Prong 5040
4 Nasal Prong 5050
Damping Filter 4 Nasal Prong 5550
(BC171 only) 4 Nasal Prong 5560
Simplified 3 Nasal Prong 6060
Pressure Manifold 2 Nasal Prong 6070
(BC171 only)
* Flow Driver compatible circuit

10
9
8
7
6
Infant Flow compatible circuit 5

CPAP Generator

Pressure Manifold MR290 Chamber


BubbleCPAP
Patient Interface Starter Kits 3.2

Product Code Description

BC260-SK Nasal Prong Starter Kit


Includes
2 Nasal Prong 3520
3 Nasal Prong 4030
4 Nasal Prong 4540
Nasal Prongs 4 Nasal Prong 5040
4 Nasal Prong 5050
4 Nasal Prong 5550
4 Nasal Prong 5560
3 Nasal Prong 6060
2 Nasal Prong 6070

BC360-SK Infant Bonnet Starter Kit


Includes
2 Infant Bonnet 17-22cm
3 Infant Bonnet 22-25cm
3 Infant Bonnet 25-29cm
2 Infant Bonnet 29-36cm
Infant Bonnet

BC480-SK Starter Kit Midline F&P 50mm


Includes
3 Nasal tubing 50mm
2 Nasal Prong 3520
2 Nasal Prong 4030
2 Nasal Prong 4540
1 Nasal Prong 5040
Nasal Tubing
1 Infant Bonnet 17-22cm
2 Infant Bonnet 22-25cm
1 Infant Bonnet 25-29cm

BC481-SK Starter Kit Midline F&P 70mm


Includes
3 Nasal tubing 70mm
2 Nasal Prong 3520
2 Nasal Prong 4030
2 Nasal Prong 4540
1 Nasal Prong 5040
1 Infant Bonnet 17-22cm
2 Infant Bonnet 22-25cm
1 Infant Bonnet 25-29cm

BC482-SK Starter Kit Midline F&P 100mm


Includes
3 Nasal tubing 100mm
2 Nasal Prong 4540
2 Nasal Prong 5040
2 Nasal Prong 5050
1 Nasal Prong 5550
1 Infant Bonnet 22-25cm
2 Infant Bonnet 25-29cm
1 Infant Bonnet 29-36cm
BubbleCPAP
Universal Patient Interface Starter Kits 3.3

The Universal Patient Interface Starter Kit is designed to allow connection


of other single-heated circuits e.g. Cardinal and Hudson.

Product Code Description

BC490-SK Starter Kit Midline Universal 50mm


Includes
3 Nasal Tubing Universal 50mm
2 Nasal Prong 3520
2 Nasal Prong 4030
2 Nasal Prong 4540
Nasal Tubing
1 Nasal Prong 5040
1 Infant Bonnet 17-22cm
2 Infant Bonnet 22-25cm
1 Infant Bonnet 25-29cm

BC491-SK Starter Kit Midline Universal 70mm


Includes
Nasal Prongs 3 Nasal Tubing Universal 70mm
2 Nasal Prong 3520
2 Nasal Prong 4030
2 Nasal Prong 4540
1 Nasal Prong 5040
1 Infant Bonnet 17-22cm
2 Infant Bonnet 22-25cm
1 Infant Bonnet 25-29cm

BC492-SK Starter Kit Midline Universal 100mm


Includes
3 Nasal Tubing Universal 100mm
2 Nasal Prong 4540
Infant Bonnet
2 Nasal Prong 5040
2 Nasal Prong 5050
1 Nasal Prong 5550
1 Infant Bonnet 22-25cm
2 Infant Bonnet 25-29cm
1 Infant Bonnet 29-36cm
BubbleCPAP
Accessories 4.1

Single-Use

Product Code Description

Gas Supply Line BC400-10 Gas Supply Line 2.1m


(to blender)
BC402-05 Gas Supply Line (to 2 flow meters)

BC420-10 Midline Pressure Luer Adaptor

Gas Supply Line Midline Pressure


(to 2 flow meters) Luer Adaptor
BubbleCPAP
Mounting Options 4.2

Product Code Description

900MR292 Humidifier Stand telescopic

900MR170 Mounting Bracket C-clamp

900MR030 Mounting Bracket Pole 15-50mm

Telescopic
Humidifier Stand C-Clamp Pole Clamp
BubbleCPAP
Devices 4.3

Humidification

Product Code Description

MR850 Humidifier

900MR801 Heater Wire Adaptor MR850

900MR823 Heater Wire/Temp Probe Kit

900MR868 Probe Temp/Flow Ang 1600mm

MR850 Humidifier Temperature/ Heater Wire


Flow Probe

Gas Supply

Product Code Description

Low Flow
Air/O2 Blender
BC501-01 Bird MBlender 2-16LPM DISS

BC502-01 Bird MBlender 2-16LPM NIST

BC506-01 Blender Oxy/Air Bio-Med (DISS) green

BC507-01 Blender Oxy/Air Bio-Med (DISS) white

BC510-01 Bird MBlender Pole Mount Bracket

Blender
Gas Blenders are available on an individual basis.
DISS & NIST
fittings available They come complete with color coded nipples and control knobs.
Flowrate 2-16 L/min (Bird)
0-15 L/min (Bio-Med)
Includes flowmeter nipples (green or white)
Bird MBlender Pole
BIO-MED DEVICES
Mount Bracket Note: Requires 900IW320 Bird Bracket Blender Complete to be fitted to the
CosyCot Infant Warmer. Gas hoses not included. Please refer to Fisher &
O2%
Paykel Healthcare Infant Warming Catalogue.
BubbleCPAP
Sample Setup (pole) 5.1

Bird MBlender
2-16LPM

BC501-01 (DISS) Low Flow


Air/O2 Blender

BC502-01 (NIST)

Note:
Gas Hoses not included.
Please refer to Fisher & Paykel
Infant Warming Catalogue BC510-01
Bird MBlender
Pole Mount Bracket

MR850
Humidifier 900MR030
Mounting Bracket Pole
15-50mm
(for mounting humidifier)

900MR801
Heater Wire Adaptor
MR850

900MR868
Probe Temp/Flow
Ang 1600mm
900MR170
Mounting Bracket C-Clamp
(for mounting CPAP generator)

900MR292
Humidifier Stand Telescopic
BubbleCPAP
Sample Setup (warmer) 5.2

Low Flow
Air/O2 Blender

Bird MBlender BC510-01


2-16LPM Bird MBlender
Pole Mount Bracket
BC501-01 (DISS)
BC502-01 (NIST)
Note:
Gas Hoses not included.
Please refer to Fisher & Paykel
Infant Warming Catalogue

900MR030
Mounting Bracket Pole
MR850 15-50mm
Humidifier (for mounting humidifier)

900MR801
Heater Wire Adaptor
MR850

900MR868
Probe Temp/Flow
Ang 1600mm

900IW302
Mounting Pole long

900MR170
Mounting Bracket C-Clamp
(for mounting CPAP generator)
en
REF 185043599 Rev F 2009-07

MANUFACTURER Fisher & Paykel Healthcare Limited, 15 Maurice Paykel Place, East Tamaki, Auckland 2013; P O Box 14 348, Panmure, Auckland 1741, New
Zealand Tel: +64 9 574 0100 Fax: +64 9 574 0158 Email: info@fphcare.com Web Site: www.fphcare.com AUSTRALIA Tel:+61 3 9879 5022 Fax:+61 3 9879 5232
AUSTRIA Tel: 0800 29 31 23 Fax: 0800 29 31 22 BENELUX Tel:+31 40 216 3555 Fax:+31 40 216 3554 CHINA Tel:+86 20 3205 3486 Fax:+86 20 3205 2132
FRANCE Tel:+33 1 6446 5201 Fax:+33 1 6446 5221 GERMANY Tel:+49 7182 93777 0 Fax:+49 7182 93777 99 INDIA Tel:+91 80 4123 6041 Fax:+91 80 4123 6044
IRISH REPUBLIC Tel: 1800 409 011 ITALY Tel:+39 06 7839 2939 Fax:+39 06 7814 7709 JAPAN Tel:+81 3 3661 7205 Fax:+81 3 3661 7206 NORTHERN IRELAND
Tel: 0800 132 189 SPAIN Tel:+34 902 013 346 Fax:+34 902 013 379 SWEDEN Tel: +46 8 564 76 680 Fax: +46 8 36 63 10 SWITZERLAND Tel: 0800 83 47 63 Fax:
0800 83 47 54 UK (EU Authorised Representative) Fisher & Paykel Healthcare Ltd, Unit 16, Cordwallis Park, Clivemont Road, Maidenhead, Berkshire SL6 7BU, United
Kingdom Tel:+44 1628 626 136 Fax:+44 1628 626 146 USA Tel: +1 800 446 3908 or +1 949 453 4000 Fax:+1 949 453 4001
Bubble CPAP System Set-up Guide
For full User Instructions, please refer to Bubble CPAP System User Instructions
(BC151/BC161 PN 185042847 and BC171 PN 185042848).
BC151 / BC161

BC161 (connects to F&P FlexiTrunk Interface)

BC151 (connects to T-bar Interface)

10
9
8
7
6
5

Checks During Operation

Regularly observe that the water is feeding into the humidification chamber.
Should the water level exceed the maximum level marked on the humidification chamber,
the chamber should be replaced.
Check that all connections are tight before use and after any adjustment.
Ensure air flow is present at all times. If air flow is interrupted, turn off the humidifier.
Regularly observe the circuit for condensate. Drain as required.
Regularly observe the CPAP generator for bubbling. If bubbling is not observed, check
for and minimize air leaks in the system and at the patient. If air leaks have been
minimized, air flow may be increased to achieve continuous bubbling.
Regularly observe the water level in the CPAP generator and overflow container.
Refill the CPAP generator if the water level drops below the minimum water level line.
Check and empty the overflow container once every 8 hours or as needed.
Monitor patient oxygen levels.
Always use pressure monitoring to verify that the patient is receiving the
prescribed CPAP level.

F & P bubb l e cpap s e t - up G ui d e 1


BC151 / BC161 SET-UP

1. Insert Humidification Chamber


Discard the chamber if the seals are not intact
whenreceived.
Slide the chamber on to the humidifier base.
Remove the blue caps.

2. Connect Water Bag


Hang the water bag, unwind the water feed set and
spike the water bag.
Open the vent on the vented spike.
Check the humidification chamber for water flow from
thebag.
If no water is visible in the chamber or water
consumption is low, check that the bag is spiked
properly and that the feed tube is not kinked or
blocked. Try gently squeezing the bag to promote
water flow. Ensure that the height of the water bag
is at least higher than the humidification chamber.
If in doubt replace the chamber.
Discard the chamber if the water level exceeds the
maximum water level line.

3. Fill Bubble CPAP Generator


Using the fill-funnel provided, fill the CPAP generator
with sterile water until water flows into the overflow
container.
Set the CPAP probe to 10 cmH2O, ready for the
leak test.


Ensure that the fill-funnel remains on the Bubble
CPAP generator.
Ensure that the humidifier and Bubble CPAP generator
are mounted below the patient.

Water Water
level level
too low correct

2 F & P bubb l e cpap s e t - up G ui d e


4. Connect Bubble CPAP Pressure Manifold and
Breathing Circuit
Connect the oxygen tubing between the pressure
manifold and the flow source.
Connect the pressure manifold to the chamber
inletport.
Connect the blue inspiratory tube to the remaining
chamber port.
Remove the blue caps from the inspiratory tube
and install the temperature probe ports and heater-
wire adapter.
Ensure that the heater-wire (inside tube) is evenly
distributed along the circuit length and is not
bunched or kinked.
Connect test elbow to expiratory and inspiratorylimbs.

5. Leak Test
Check all connections are tight before use.
Set the CPAP probe to 10 cmH2O and the input flow
rate to 1 L/min.
1 L/min Observe the CPAP generator. Gentle, audible
10
bubbling is acceptable; no bubbling means
9
8
unacceptable leakage.
7 If no bubbling is observed, check the entire system.
6
5 Remove the correct port cap from the pressure
manifold before connecting any monitoring device.

6. Set Flow Rate


Adjust the flow rate to the prescribed input flowrate.
Recommended flow rate: 6 to 8 L/min.
6-8 L/min Allowable flow range: 4 to 15 L/min.
(<15 L/min)

F & P bubb l e cpap s e t - up G ui d e 3


BC151 / BC161 SET-UP

Pressure-Flow Table F&P Bubble CPAP Generator used with


F&P Infant Interface, typical mean pressure values (cmH20) WITH NO LEAK

Probe Setting (cmH2O)

Flow
(L/min)
3 4 5 6 7 8 9 10

4 3.1 4.1 5.1 6.1 7.1 8.1 9.1 10.0

5 3.2 4.2 5.3 6.3 7.3 8.3 9.2 10.1

6 3.4 4.4 5.4 6.4 7.4 8.4 9.3 10.2

7 3.6 4.6 5.6 6.6 7.6 8.5 9.5 10.4

8 3.8 4.8 5.8 6.8 7.8 8.7 9.6 10.5

9 4.0 5.0 6.0 7.0 7.9 8.9 9.8 10.6

10 4.3 5.3 6.2 7.2 8.1 9.0 9.9 10.8

11 4.6 5.6 6.5 7.4 8.4 9.2 10.1 10.9

12 4.9 5.8 6.8 7.7 8.5 9.4 10.2 11.1

13 5.2 6.1 7.0 7.9 8.8 9.6 10.4 11.3

14 5.5 6.4 7.3 8.2 9.0 9.8 10.6 11.4

15 5.8 6.7 7.5 8.4 9.2 10.0 10.8 11.6

Note: Other interfaces or circuits may produce varying pressures.

7. Set CPAP Level


The number on the CPAP probe above the lid
indicates the CPAP pressure in cmH2O. As an example,
10
9 the illustration shows CPAP setting of 6 cmH2O.
8 Set the CPAP probe at the prescribed level
7
6
(3 to 10 cmH2O).

4 F & P bubb l e cpap s e t - up G ui d e


8. Connect Bubble CPAP Circuit to Infant Interface
Remove the flow test elbow and connect the circuits
to the Infant Interface using the instructions provided
with the interface.
BC161 connects to F&P FlexiTrunk Interface.
BC151 connects to short binasal prongs (T-bar).

9. Set the Humidifier


Ensure there is air flow present before turning on
thehumidifier.
Turn on the humidifier.
If using the MR850, ensure it is on Invasive mode
MR850: (37 C).
37 C If using the MR730, set the temperature to
40 C, -3.
Refer to the MR850 or MR730 User Instructions for
further information.

MR730:
40 C, -3

10. Attach Interface to Infant


Connect the interface to the infant using the
instructions provided with the interface.

BC161

BC151

F & P bubb l e cpap s e t - up G ui d e 5


BC171

BC171 (connects to Flow Driver)

10
9
8
7
6
5

Checks During Operation

Regularly observe that the water is feeding into the humidification chamber.
Should the water level exceed the maximum level marked on the humidification chamber,
the chamber should be replaced.
Check all connections are tight before use and after any adjustment.
Ensure air flow is present at all times. If air flow is interrupted, turn off the humidifier.
Regularly observe the circuit for condensate. Drain as required.
Regularly observe the CPAP generator for bubbling. If bubbling is not observed, check
for and minimize air leaks in the system and at the patient. If air leaks have been
minimized, air flow may be increased to achieve continuous bubbling.
Regularly observe the water level in the CPAP generator and overflow container.
Refill the CPAP generator if the water level drops below the minimum water level line.
Check and empty the overflow container once every 8 hours or as needed.
Monitor patient oxygen levels.
Always use pressure monitoring to verify that the patient is receiving the
prescribed CPAP level.

6 F & P bubb l e cpap s e t - up G ui d e


BC171 SET-UP

1. Insert Humidification Chamber


Discard the chamber if the seals are not intact
when received.
Slide the chamber on to the humidifier base.
Remove the blue caps.

2. Connect Water Bag


Hang the water bag, unwind the water feed set and
spike the water bag.
Open the vent on the vented spike.
Check the humidification chamber for water flow from
the bag.
If no water is visible in the chamber or water
consumption is low, check that the bag is spiked
properly and that the feed tube is not kinked or
blocked. Try gently squeezing the bag to promote
water flow. Ensure that the height of the water bag is
at least 50 cm higher than the humidification
chamber. If in doubt replace the chamber.
Discard the chamber if the water level exceeds the
maximum water level line.

3. Fill Bubble CPAP Generator


Using the fill-funnel provided, fill the CPAP generator
with sterile water until water flows into the overflow
container.
Set the CPAP probe to 10 cmH2O, ready for the
leak test.


Ensure that the fill-funnel remains on the
BubbleCPAP generator.
Ensure that the humidifier and Bubble CPAP generator
are mounted below the patient.

Water Water
level level
too low correct

F & P bubb l e cpap s e t - up G ui d e 7


BC171 SET-UP

4. Connect Pressure Manifold, Dry Line and Circuit


Connect the pressure manifold with dry line to the
flow driver and chamber inlet port.
Connect the blue inspiratory limb to the remaining
chamber port.
Remove the blue caps from the inspiratory tube and
install the temperature probe ports and the heater-
wire adapter.
Ensure that the heater wire is evenly distributed
along the circuit length and is not bunched or kinked.
Connect test elbow to expiratory and
inspiratorylimbs.

5. Leak Test
Check all connections are tight before use.
Set the CPAP probe to 10 cmH2O and the input flow
10
rate to 1 L/min.
9 Observe the CPAP generator. Gentle, audible bubbling
1 L/min 8
is acceptable; no bubbling means unacceptable
7
6 leakage.
5 If no bubbling is observed, check the entire system.

6. Set Flow Rate


Adjust the flow rate to the prescribed input flow rate.
Recommended flow rate: 6 to 8 L/min.
Allowable flow range: 4 to 15 L/min.

6-8 L/min
(<15 L/min)
Set CPAP Level
10
9 The number on the CPAP probe above the lid
8
7
indicates the CPAP pressure in cmH2O. As an example,
6
the illustration shows CPAP setting of 6cmH2O.
Set the CPAP probe at the prescribed level
(3 to 10 cmH2O).

8 F & P bubb l e cpap s e t - up G ui d e


Pressure-Flow Table F&P Bubble CPAP Generator used with
F&P Infant Interface, typical mean pressure values (cmH20) WITH NO LEAK

Probe Setting (cmH2O)

Flow
(L/min)
3 4 5 6 7 8 9 10

4 3.1 4.1 5.1 6.1 7.1 8.1 9.1 10.0

5 3.2 4.2 5.3 6.3 7.3 8.3 9.2 10.1

6 3.4 4.4 5.4 6.4 7.4 8.4 9.3 10.2

7 3.6 4.6 5.6 6.6 7.6 8.5 9.5 10.4

8 3.8 4.8 5.8 6.8 7.8 8.7 9.6 10.5

9 4.0 5.0 6.0 7.0 7.9 8.9 9.8 10.6

10 4.3 5.3 6.2 7.2 8.1 9.0 9.9 10.8

11 4.6 5.6 6.5 7.4 8.4 9.2 10.1 10.9

12 4.9 5.8 6.8 7.7 8.5 9.4 10.2 11.1

13 5.2 6.1 7.0 7.9 8.8 9.6 10.4 11.3

14 5.5 6.4 7.3 8.2 9.0 9.8 10.6 11.4

15 5.8 6.7 7.5 8.4 9.2 10.0 10.8 11.6

Note: Other interfaces or circuits may produce varying pressures.

7. Connect Bubble CPAP Circuit to Infant Interface


Remove the flow test elbow and connect the
circuits to the Infant Interface using the instructions
provided with the interface.

F & P bubb l e cpap s e t - up G ui d e 9


BC171 SET-UP

8. Connect Pressure Line to Flow Driver


Connect the pressure line filter end of the pressure
line to the flow driver.
Connect the pressure line elbow connector end of
the pressure line to the nasal tubing pressure port.

9. Set the Humidifier (MR850)


Ensure there is air flow present before turning on
thehumidifier.
Turn on the flow driver.
Turn on the humidifier.
If using the MR850, ensure it is on Invasive mode
(37 C).
Refer to the MR850 User Instructions for further
information.

MR850: 37 C

9A. Set the Humidifier (MR730)


A Ensure there is air flow present before turning on
thehumidifier.
Turn on the flow driver.
Turn on the humidifier.
If using the MR730, set the temperature to
40 C, -3.
Refer to the MR730 User Instructions for further
information.

MR730: 40 C, -3

1 0 F & P bubb l e cpap s e t - up G ui d e


10. Attach Interface to Infant
Connect the interface to the infant using the
instructions provided with the interface.

BC171

F & P bubb l e cpap s e t - up G ui d e 1 1


For more information please contact
your local Fisher & Paykel Healthcare representative

Manufacturer Australia Italy


Fisher & Paykel Healthcare Ltd Fisher & Paykel Healthcare Pty Limited Tel: +39 06 7839 2939
15 Maurice Paykel Place 36-40 New Street, PO Box 167 Fax: +39 06 7814 7709
East Tamaki, Auckland 2013 Ringwood, Melbourne
Spain
PO Box 14 348, Panmure Victoria 3134, Australia Tel: +34 902 013 346
Auckland 1741 Fax: +34 902 013 379
New Zealand Tel: +61 3 9879 5022
Fax: +61 3 9879 5232 Sweden
Tel: +64 9 574 0100 Tel: +46 8 564 76 680
Fax: +64 9 574 0158 Fax: +46 8 36 63 10
Email: info@fphcare.com Austria
Web: www.fphcare.com Tel: 0800 29 31 23 Switzerland
Fax: 0800 29 31 22 Tel: 0800 83 47 63
Fax: 0800 83 47 54
Benelux
Taiwan
Tel: +31 40 216 3555 Tel: +886 2 8751 1739
Fax: +31 40 216 3554 Fax: +886 2 8751 5625
China Turkey
Tel: +86 20 3205 3486 Tel: +90 312 354 34 12
Fax: +86 20 3205 2132 Fax: +90 312 354 31 01
185034215 REVa EN 2011 Fisher & Paykel Healthcare Limited

UK EC REP

France Fisher & Paykel Healthcare Ltd


Tel: +33 1 6446 5201 Unit 16, Cordwallis Park
Fax: +33 1 6446 5221 Clivemont Road, Maidenhead
Berkshire SL6 7BU, UK
Germany
Tel: +49 7181 98599 0 Tel: +44 1628 626 136
Fax: +44 1628 626 146
Fax: +49 7181 98599 66
USA/Canada
India
Tel: +1 800 446 3908
Tel: +91 80 4284 4000
or +1 949 453 4000
Fax: +91 80 4123 6044
Fax: +1 949 453 4001
Irish Republic
Tel: 1800 409 011

www.fphcare.com
Bubble CPAP System

F&P Bubble CPAP SYSTEM


With the goal of improving patient care in over
120 countries, Fisher & Paykel Healthcare
designed the F&P Bubble CPAP System for use
with spontaneously breathing infants requiring
respiratory support.
Incorporating unique features and world-leading
humidification technology, the F&P Bubble CPAP
System has been developed to provide a
respiratory support system that is safe,
effective and easy to use.
Adaptability to current practices has been taken
into consideration with the choice of various
system models.
The F&P Infant Interface complements the system,
providing comfort and optimum
CPAP support to the infant at risk.
F&P Bubble CPAP System | FEATURES

The F&P Bubble CPAP System is the first complete system available in an All-in-One Pack. It includes the Bubble CPAP generator,
heated breathing circuits, humidification chamber and pressure manifold. F&P FlexiTrunk is a lightweight nasal tubing with a unique, flexible
extension for a stable prong and mask fixation while allowing for normal infant movement.
F&P Bubble CPAP System models include:
BC151 connects to other infant CPAP interfaces (e.g. Hudson and Inca prongs)
BC161 connects to the F&P Infant Interface
BC171 connects to the M672P Infant Flow Driver (IFD) or equivalent

BC161 BC171

BC151

10 10
9
9
8
8
7
7
6
6
5
5

10
9
8
7
6
5

Safe and Reliable


Used on over a quarter of a million patients globally
The unique auto-level mechanism of the Bubble CPAP generator provides
consistent and accurate delivery of CPAP
The pressure manifold with pressure relief valve promotes infant safety

Easy to Use
Easy auto-leveling by simply adding water into the Fill Funnel on the Bubble
CPAP generator
Excess condensate automatically drains into the detachable overflow
container (see image on right)
Detachable overflow container can be emptied without interrupting CPAP

Optimum Humidification
The F&P Bubble CPAP System provides respiratory support with
body-temperature pressure-saturated gases to the infant
Optimal humidity promotes mucociliary clearance and reduces the work of
breathing

Simple Yet Versatile


Adjustable CPAP from 3 to 10 cmH2O
The Bubble CPAP System can connect to various interfaces
FOCUS ON FEATURES

Bubble CPAP Circuit


Spiral heater-wire provides even
heat distribution, reducing heat loss
and condensate build-up
Aims to deliver optimal humidity

Bubble CPAP Pressure Manifold


Limits the pressure delivered in
the event of an occlusion
Allows connection to a pressure-
monitoring device and/or an
air/oxygen analyzer

MR290 Humidification Chamber


Auto-fill, easy to use
Maintains constant CPAP
Closed system minimizes the risk
of infection
Bubble CPAP Generator
Adjustable CPAP from 3 to 10 cmH2O
Detachable overflow container allows
uninterrupted CPAP when overflow
container is being emptied
Includes separate Fill Funnel
Mounting bracket is incorporated
on the back of the Bubble CPAP
generator
F&P Bubble CPAP System | SPECIFICATIONS

F&P Bubble CPAP System


Input Flow Range 4 to 15 L/min
Recommended Input Flow 6 to 8 L/min
Set CPAP Pressure Range 3 to 10 cmH2O
Intended Patient Population Premature neonates and infants up to 10 kg
Operating Temperature Range 18 to 26 oC
Storage Temperature Range -10 to 50 oC
Usage Period Single patient use, for a maximum of 7 days
Maximum Pressure Limit 17 cmH2O @ 8 L/min
Oxygen Analyzer Port 22 mm Male or 15 mm Female

Pressure Monitoring Port Female luer


Respiratory Humidifier F&P MR850 or MR730 only (Invasive Mode)
Circuit Length Inspiratory Tube Expiratory Tube
BC161 1.1 m (3.61 ft) 1.2 m (3.94 ft)
BC151 1.4 m (4.59 ft) 1.2 m (3.94 ft)
includes 300 mm (11.8 in) extension
to connect to Hudson Prongs

BC171 1.1 m (3.61 ft) 1.2 m (3.94 ft)

F&P Bubble cpap system models

Product Code & Description

BC151*

BC161*
185034230 EN Rev B 2012 Fisher & Paykel Healthcare Limited

BC171*

* Box of 10

Note: For ordering information, please refer to the Fisher&Paykel Healthcare Bubble CPAP Product Catalogue (Part No. 185043599).
MR850 Humidifier must be purchased separately.

www.fphcare.com

Manufacturer Fisher & Paykel Healthcare Ltd, 15 Maurice Paykel Place, East Tamaki, Auckland 2013, PO Box 14 348, Panmure, Auckland 1741, New Zealand Tel: +64 9 574 0100
Fax: +64 9 574 0158 Email: info@fphcare.co.nz Web: www.fphcare.com Australia Tel: +61 3 9879 5022 Fax: +61 3 9879 5232 Austria Tel: 0800 29 31 23 Fax: 0800 29 31 22
Benelux Tel: +31 40 216 3555 Fax: +31 40 216 3554 China Tel: +86 20 3205 3486 Fax: +86 20 3205 2132 France Tel: +33 1 6446 5201 Fax: +33 1 6446 5221
Germany Tel: +49 7181 98599 0 Fax: +49 7181 98599 66 India Tel: +91 80 4284 4000 Fax: +91 80 4123 6044 Irish Republic Tel: 1800 409 011
Italy Tel: +39 06 7839 2939 Fax: +39 06 7814 7709 Japan Tel: +81 3 3661 7205 Fax: +81 3 3661 7206 Northern Ireland Tel: +0800 132 189 Spain Tel: +34 902 013 346
Fax: +34 902 013 379 Sweden Tel: +46 8 564 76 680 Fax: +46 8 36 63 10 Switzerland Tel: 0800 83 47 63 Fax: 0800 83 47 54 Taiwan Tel: +886 2 8751 1739 Fax: +886 2 8751 5625
Turkey Tel: +90 312 354 34 12 Fax: +90 312 354 31 01 UK (EU Authorised Representative) Fisher & Paykel Healthcare Ltd, Unit 16, Cordwallis Park, Clivemont Road, Maidenhead,
Berkshire SL6 7BU, UK Tel: +44 1628 626 136 Fax: +44 1628 626 146 USA Tel: +1 800 446 3908 or +1 949 453 4000 Fax: +1 949 453 4001

También podría gustarte