Está en la página 1de 4

Spec Sheet Endress+Hauser

Order Code: FMB53-U9X2/101


Quote Ref: 2015984982-40
Date: 08/15/2017

Project
TAG P1PAB10CL001
P1PAB10CL002
P1PAB10CL003
P1PAB20CL001
P1PAB20CL002
P1PAB20CL003
P2PAB10CL001
P2PAB10CL002
P2PAB10CL003
P2PAB20CL001
P2PAB20CL002
P2PAB20CL003
Customer name Access Professional Singapore
Pte Ltd
Customer number 0045802177
Contact role Contact person
Contact name Karl Ngo
Street 29C Benoi Road, Pioneer Lot
ZIP 627777
City/town Singapore
Country Singapore
VAT number 199700515H
Fax number -6568 2199
Phone number -6568 2188
E-mail address sales-
support@accessprofessional.co
m
Manufacturer Contact Data
Enterprise Singapore
Address Endress+Hauser(S.E.A.) Pte Ltd

Address SG Sales Grp 2-LKL


Street 3A International Business Park
ZIP 609935
City/town Singapore
Country Singapore
Product Data
General information
Measuring task Pressure / Level
General explosion-protection information Information about currently
available Ex versions (ATEX,
FM, CSA, TIIS, IECEx, NEPSI,
etc.) can be supplied by your
E+H Sales Center on request.
All explosion protection data are
given in a separate
documentation which is availabe
upon request.
Identification
Nameplate material Stainless steel
Manufacturer name Endress+Hauser
Vendor Endress+Hauser
Product device / transmitter
Product type Pressure sensor with the
CONTITE TM measuring cell
Product name Deltapilot M FMB53
Order code FMB53-U9X2/101
Extended order code FMB53-
AA21JA1HGJ15B3AAAPOZ1
Function and system design
Measuring principle Hydrostatic measurement
Input characteristics
Measurement channels
Measured variable Hydrostatic pressure
Measuring range
Upper range-limit 120.0 kPa
abs/gauge pressure Gauge
Lower range-limit (EU) -1.0 bar
Upper range-limit (EU) 1.2 bar
Measuring span min. (EU) 0.06 bar
Adjusted measuring range
Lower range-limit 410.00000 mbar
Upper range-limit 1,180.00000 mbar
Output characteristics
Analog Output
Output type Analog Output
Signal type 4-20 mA
Lower range value of signal 4.0 mA
Upper range value of signal 20.0 mA
Output configurable
Lower range value of signal configurable 3.8 mA
Upper range value of signal configurable 20.5 mA
Signal on alarm
Signal on alarm available yes
Signal on alarm configurable yes
Lower value 3.6 mA
Upper value 23.0 mA
Communication interface
HART Communication Interface
System Integration
Type of protocol HART
Power supply
Power input circuit
Supply mode 2 wire
Electrical connection Screw terminal
Accuracy
Accuracy 0.2 %
Measured value resolution 1.0 A
Operating conditions
Environment
Ambient temperature C min. -20.0 C
Ambient temperature C max. 70.0 C
Storage temperature C min. -40.0 C
Storage temperature C max. 70.0 C
Climate class 4K4H as per DIN EN 60721-3
Electromagnetic compatibility As per all the relevant
requirements of the EN 61326
series + NAMUR
Recommendation EMC (NE21) /
Maximum deviation: < 0.5 % of
span
Process conditions
Process temperature in C min. -10.0 C
Process temperature in C max. 70.0 C
Mechanical construction
Connection housing / head
Housing type Transmitter housing
Housing type Endress+Hauser F31
Degree of protection, Europe IP66/68
Material housing Die-cast aluminium
Coating Powder protective coating on
polyester base
Colour Blue (housing) / gray (cover)
Colour RAL RAL 5012 (housing) / 7035
(cover)
Additional equipment Glass window
Cable entry signal line
Cable gland M20x1.5
Process connection
Material gasket FKM Viton
Connection cable
Cable type Probe cable, shortable
Cable length in m 15.00 m
Cable cover material PE
Hydrostatic level device
Hydrostatic level measuring device type Cable version with mounting
clamp
Fill fluid Synthetic oil
Material process isolating diaphragm based on AISAlloy C
Operability
Display on device With display
Display languages English
Operating software Hand held DXR375, Field
Communicator 375, Field Xpert
(PDA), WiFi, Fieldgate FXA520,
FieldCare (FDT)
Operating element Button
Local operation yes
Display element
Display type LCD graphic display with
bargraph
Number of lines 4
Certificate and approvals device / transmitter
Approval for general usage CE mark
Sanitary approval FDA
Calibration
Calibration type Customised pressure
Comments

También podría gustarte