300-2
SERVICE
SECTION 5
SERVICE
5.1 GENERAL
This section provides calibration, troubleshooting, and removal and
replacement instructions for the Model C300 Incubator.
5.2 CALIBRATION PROCEDURES
5.2.1 GENERAL
This paragraph provides calibration and test procedures for the Model
300 Incubator.
Unless otherwise indicated, all calibration procedures are performed
under the following conditions:
1, The Controller is unmounted to provide access to the calibration
adjustments. (The PC board need not be removed from the
Controller).
2. The Incubator is connected to a primary power source of the
correct voltage and frequency (see data tag).
3. Ambient temperature for calibration and test is 23°C + 4°C
(73°F = 7°F).
5.2.2 TEST EQUIPMENT REQUIRED
The test equipment listed below is required for calibration and
performing oxygen concentration tests. Equivalent test equipment may
be substituted.
Probe Simulator, Part No. 67 902 75
Digital VOM, Fluke Model 8000A
Oscilloscope - Tektronix 465
Oxygen Analyzer, Sybron/Taylor Model QA580
Flowmeter, Victor - Model 1099-0025
Leakage Tester, Bi-Tek 501
dumper Plug, Part No. 17 731 98
5.2.3 OVERALL CALIBRATION
TEST SET-UP
1. REMOVE SENSOR CONNECTOR and connect the Probe Simulator to
-1,=2, and -3 on the Power Board,
Che 1
(Change 1) st300-2
SERVICE
2. ON THE PROBE SIMULATOR, set the ALARMS switch to NORM and the °C
Swit
3. APPLY MAIN POWER to the Incubator by setting the AC POWER switch
on the Power Chassis to the ON-1 position.
4, APPLY POWER TO THE CONTROLLER by pressing the CONTROLLER ON
‘switchs allow the Incubator to complete the auto-test cycle.
5. ALLOW THE INCUBATOR TO RUN FOR AT LEAST 15 MINUTES before
proceeding with calibration procedures. 7
6. CHECK 150 HZ CLOCK by connecting an oscilloscope to TP10 (Figure
B.1). The clock period should be 430 ms + 30 ms. If necessary,
adjust R92 until period is within specifications. (See Note
below.)
NOTE: In early production C300-2 Incubators, R92 is a fixed
resistor; therefore no adjustment can be made in these units.
BRIDGE (TEMPERATURE) AMPLIFIER CALIBRATION
a
2.
3.
CONNECT DVM between TP2 and TP1 (ground).
ADJUST R58 for a reading of 3.50 Vde + 0.01 Vde at TP2.
DISCONNECT THE DVM.
A/D CONVERTER CALIBRATION
7a
CONNECT JUMPER PLUG (part no. 17 731 98) between TP6 and TP7.
ON THE PROBE SIMULATOR, set the °C switch to 21°C.
ADJUST R76 for a reading of 20.0°C on the temperature display.
ON THE PROBE SIMULATOR, set the °C switch to 41°C.
ADJUST R61 for a reading of 40.0°C on the temperature display.
REPEAT THIS PROCEDURE, as required, to obtain accurate
adjustments.
DO_NOT DISCONNECT the jumper plug from TP6 and TP7.
TEMPERATURE OFFSET ADJUSTMENT
5-2
ON THE PROBE SIMULATOR, set the °C switch to 21°C.
ADJUST R76 for a reading of 21.0°C + 0,1°C on the temperature
‘displa:
DISCONNECT THE JUMPER PLUG from TP6 and TP7.300-2
SERVICE
SINIOd 183.1 GNV SLNAWLSNray b's aHNDIS
usTIOMINOD UV
oy
SOTSIHS Gv we
wowysnont
VAN
altos
5-3300-2
SERVICE
PRESET TEMPERATURE CALIBRATION
NOTE: This adjustment must be completed within 15 seconds after the
‘SET TEMP switch is pressed (SET POINT indicator lit). If the SET
POINT indicator goes out before the adjustment is completed, the
procedure must be repeated.
1. PRESS THE SET TEMP SWITCH; the SET POINT indicator should light.
ADJUST R33 for a reading of 35,0°C on the temperature display.
3. IF THE SET POINT INDICATOR GOES OFF during the adjustment, repeat
teps T ani
MAXIMUM SET POINT TEMPERATURE CALIBRATION
PRESS THE SET TEMP SWITCH; the SET POINT indicator should light.
2. PRESS AND HOLD THE SET POINT RAISE switch ([AJ) until the
‘Temperature display stops changing; do not release the SET POINT
raise switch.
3. SLOWLY ADJUST R31 clockwise until the temperature display
indicates 38.0°C. If the display reading is inadvertently
adjusted to a reading higher than 38.0°C, perform the following
step; otherwise proceed to step 5.
4. ADJUST R31 slightly counterclockwise. Press the SET TEMP switch
‘and then the SET POINT lower switch ( [¥]) until the temperature
display indicates 37.5°C. Repeat steps 1 through 3 of this
procedure until the temperature display indicates 38,0°C.
5. PRESS THE SET TEMP SWITCH; the SET POINT indicator should light.
6. PRESS AND HOLD THE SET pornT Lower swrtcH ([Z)) untit the
Yemperature display stops changing; the display should indicate
21.5°C + 1.0°C.
7. PRESS THE SET TEMP SWITCH; the SET POINT indicator should light.
8. PRESS AND HOLD THE SET POINT RATSE SWITCH ( [AI ) untit the
‘temperature display indicates 30.0°C = 0.5°C; release the switch.
HIGH TEMPERATURE ALARM CALIBRATION
1. ADJUST R103 counterclockwise.
2. CONNECT A JUMPER between TP8 and TP9,
5-43.
4.
5.
300-2
SERVICE
ON THE PROBE SIMULATOR, set the ALARMS switch to HI TEMP.
SLOWLY ADJUST R103 clockwise until the audible alarm "beeps".
DO _NOT DISCONNECT JUMPER from TP8 and TP9.
‘ALARM CHECK
High Temperature
is
WAIT SEVERAL SECONDS until the HI TENP alarm activates. The
audibTe aTarm shouTd sound and the HEATER POWER indicator should
be disabled.
PRESS THE RESET SWITCH. The alarm should not reset.
SET PROBE SIMULATOR ALARMS switch to NORM.
PRESS THE RESET SWITCH. The HIGH TEMP alarm should reset.
CHECK R103 ADJUSTMENT by setting Probe Simulator ALARMS switch
Back to HIGH TEMP position. The audible alarm should beep.
SET PROBE SIMULATOR ALARMS switch to NORM position and disconnect
Jumper from TPB and TPS.
Open Sensor
ey
2.
3.
4
SET PROBE SIMULATOR ALARMS switch to the OPEN position.
THE SENSOR ALARM SHOULD ACTIVATE in about 7 seconds. The audible
‘alarm shouTd sound and the HEATER POWER indicator should be
disabled.
SET THE PROBE SIMULATOR ALARMS switch to NORM position.
PRESS THE RESET SWITCH; the alarm should reset.
Shorted Sensor
b
2.
SET PROBE SIMULATOR ALARMS switch to SHORT position.
THE SENSOR AND HIGH TEMP ALARMS should activate in about 7
Seconds and the audible alarm should be enabled.
SET PROBE SIMULATOR ALARMS SWITCH to NORM position.
PRESS THE RESET SWITCH; the alarm should reset.
5-5300-2
SERVICE =
Power Fi
ure and Set-Point Retention
NOTE: If the Incubator is equipped with optional battery pack,
perform Step 1, otherwise proceed to Step 2.
1. DISCONNECT THE BATTERY TRAY by sliding it out of the Incubator
about two inches.
2. PRESS THE SET TEMP switch and then press the set point raise
TfA]) switch and raise the set point display to 36.0°C.
3. SET THE MASTER AC POWER SWITCH on the Power Chassis to the OFF-O a
position. The POWER FATLURE alarm should be-enabled and a
continuous audible alarm should sound.
4, SET THE MASTER AC POWER SWITCH on the Power Chassis to the ON-1
posTtion. The POWER FATLURE alarm should cease and the auto test
sequence should be initiated. -
5. WAIT FOR THE AUTO TEST SEQUENCE TO END and then check the set
point display; Tt shouTd rea:
6. SET THE MASTER AC POWER SWITCH on the Power Chassis to the OFF-0
posTtion. The POWER FAILURE alarm should be enabled and a
continuous audible alarm should sound.
7. PRESS THE CONTROLLER OFF SWITCH on the Controller, the POWER (
‘alarm should cease,
8. RECONNECT AND SECURE the Battery Tray if Incubator is equipped
with opttonal Battery Pack.
9. SET THE MASTER AC POWER SWITCH on the Power Chassis to the ON-1
posttton, and depress the CONTROLLER ON switch on the Control
Panel to restore operation.
10. DISCONNECT THE PROBE SIMULATOR from 1J1 and reconnect plug 1P1.
11, REASSEMBLE THE INCUBATOR ~
5-6300-2
SERVICE
5.2.4 BATTERY CHARGER (ACCESSORY) CALIBRATION PROCEDURE
If the Incubator is equipped with the Battery/Charger Tray accessory,
calibrate the battery charger as follows:
1, Set the master AC POWER switch on the Power Chassis to the OFF-0
position and depress the CONTROLLER OFF switch on the Controller
front panel.
2. Loosen the Battery/Charger Tray retaining screw completely.
3. Grasp the handle and very slowly withdraw the tray approximately
6 inches (15 cm).
WARNING: The Battery/Charger Tray weights apprxoimately 50 Ibs.
22.7 Kg) with two batteries. The tray must be well supported
during removal and installation. Exercise care to prevent,
pinching of fingers during removal or installation of
Battery/Charger Tray.
4, Position a hand on each side of the tray and as close as possible
to the Incubator.
5. Slowly remove the tray from the Incubator. Place the tray on a
surface that will support its weight.
6. Disconnect the red and black leads from the two batteries.
7. Connect the Battery/Charger Tray to the primary ac power source.
8. Using a digital voltmeter, measure the voltage at 402-11 (see
Figure 7.3). The reading should be 13.8 VOC + 0.1 VDC; adjust
4R10 if required to obtain correct reading.
9. Disconnect all test equipment, reconnect the red and black wires
to the batteries and reinstall the Battery/Charger Tray in the
Incubator.
5.2.5 CHASSIS LEAKAGE CURRENT TEST
TEST SET-UP
1. Connect the Power Chassis to the primary power source through an
ungrounded adapter plug, so that the unit is ungrounded. Set the
AC POWER switch to ON-1 position.
2, The leakage test standards provided in the procedure below assume
leakage through a resistance of 1000 ohms. If the Leakage Tester
being used does not provide this resistance, the test set-up must
be adjusted to provide it.
(Change 1) 5-7300-2
SERVICE
PROCEDURE.
USE THE LEAKAGE TESTER to measure between the chassis of the unit
undér test and @ known ground such as the ground connection of a
wall receptacle. The leakage must not exceed 100 microamps.
2. REVERSE THE PLUG and repeat step 1.
3. PERFORM STEPS 1 AND 2 with the Power Chassis AC POWER switch in
the OFF-0 position. *
5.2.6 OXYGEN CONCENTRATION TESTS
TEST SETUP. =
1. Place a calibrated oxygen analyzer on the mattress in the
Incubator.
2. Connect an oxygen supply, through a flowmeter, to the OXYGEN
INLET connector on the side of the Incubator.
PROCEDURE.
1, TURN THE INCUBATOR ON and adjust oxygen flow rate to 2.0 LPM + -
PM
2. THE HOOD OXYGEN CONCENTRATION SHOULD ATTAIN 32 TO 47% oxygen
within 30 minutes and stabilize.
3. INCREASE THE OXYGEN FLOW RATE to 6.0 LPM +0.5, -0.0 LPM. The
‘ood oxygen concentration should attain 58 to 90% oxygen within
30 minutes and stabilize.
5.2.7 OPERATIONAL CHECKOUT
Perform the Operational Checkout Procedure given in paragraph 2.4.
5.3 TROUBLESHOOTING
5.3.1 GENERAL
Troubleshooting guides for the C300 Incubator are presented under
paragraph 5.3.3 in the form of flowcharts. It is assumed that an ‘
attempt has been made to calibrate the unit and that all cables are in
good condition,
5-8€300-2
SERVICE
5.3.2 TEST EQUIPMENT REQUIRED
The test equipment listed below is required for troubleshooting the
300 Incubator. Equivalent test equipment may be substituted.
Digital VOM, Fluke Model 8000A
Logic Probe - capable of + 12 VOC
Oscilloscope, Tektronix 465
Probe Simulator, Part No. 67 902 75
Jumper Plug, Part No. 17 731 98
5.3.3 TROUBLESHOOTING PROCEDURES
The following flowcharts are provided as an aid in localizing the
cause of equipment malfunctions. The charts are intended for use in
conjunction with the equipment theory of operation (Section 3) and the
schematic diagrams (Section 7). It is assumed that the Operational
Checkout (paragraph 2.4) has been performed, and that the Calibration
Procedures (paragraph 5.2) have been attempted.
5-9300-2
SERVICE
Connect Incubator to
Prinary Pover Source (See
Daca Tag). Tura the Unit
Power
Distribution
Controle vo.
Operating > to Flovehart 5.2.
Auto Test
Test
Sequence
correct,
Go to Flowchart 5.3.
Heater Circuitry
Heater
Bleneat On|
Go to Flowchart 5.7.
Displays
Go to Flowchart 5.4.
Go to Flowehare 5.8.
10 sHEer 2
FLOWCHART 5.1 OVERALL CHECKS
(Sheet 1 of 4)igh Ate
Teap Alarm
300-2
SERVICE
FROM SHEET 1
[Allow Tncubator to
Heat for Approxi~
nately 1 Hour
Press Set Teap Seiteh; Set
Potnt Indicator should 1ight|
land short beep should be
heard. mediately depress
‘and hold both the Setpoint
raise (A) and Setpoint Lover
[Y) Switches and observe
‘Téxperature Display. When
Display indicates 40.0°C,
release the switches,
[Allow Incubator to heat up.
Go to Flowchart 5.6,
Tteater 0
Go to Flowchart 5.8,
[Decrease Teap Set Point to
35°C and allow Incubator
co cool suffictently unt?
HI-TEX lara can be reset.|
|
110 SHEET 3
FLOWCHART 5.1 OVERALL CHECKS
(Sheet 2 of 4)
5-11300-2 :
SERVICE
FROM SHEET 2
Sensor Alam_________4
Disconnect IPI from 11
fon Power Board, Connect|
Probe Simulator to 11,
Set Alarms Switch on
Simator to Open.
‘Sensor xo
(Alarm in? >———™ Go to Plowhart 5.6.
Seconds,
ves
Set Simulator
Alarne Seitch to
Short,
‘irre x0
“and Sensor >» Go to Flowchart 5.6.
are in 7,
ves
Disconnect Probe
Simulator and Re-|
connect 1J1 and
wh,
i
Disconnect 1P5
fron 135 on
Power Board
To Sheet 4
FLOWCHART 5.1 OVERALL CHECKS
(Sheet 3 of 4)
5-12300-2
SERVICE
From Sheet 3
check CRI7, CRIB, O12, 13,
U2, 31-4, UIS-14, U9, 0s:
and assoc{ated conponents.
Tecomect 195 to
, 1s on tover
foard” and eset
8 Alarm
1
Disconnect
AC Power
Cord
Check CR2, CRE, Q3, Ul,
BTL, DS4, ‘and associated
congonents..
¥
Press controller
OFF switch on
controler,
Check UL, 03, CR5, and
associated components.
stop
FLOWCHART 5.1 OVERALL CHECKS
(Sheet 4 of 4)
5-13300-2
SERVICE -
Connect Tncubator to
Prinary Pover Source (See
Data Tag). Tura the Unit -
ons Wale’7 gece. for
Auto-Test to end,
Power Chasais
Check AC Circult Breaker(s).
Check Line Filter 2 FL1 and Power
Switeh 231.
3.” Check Terminal block connections at
2B.
Check Power Transforaer 271.
Check C81 and 1CR2.
Check 10R3 and Associated Componente,
Check 10R2.
Check URS.
Check IK1, IVRI and associated
components.
*ull wave rectified sinewave,
To sueer 2
FLOWCHART 5.2 POWER DISTRIBUTION TROUBLESHOOTING
(Sheet 1 of 2)300-2
SERVICE
FROM SHEET 1
POWER CHASSIS (Cont.).
®
Check 1K operation,
e
Check 1K2 operation.
- Control Board:
- Check interface cable assenbly for
continuity.
: Check on2.
- stor
FLOWCHART 5.2 POWER DISTRIBUTION TROUBLESHOOTING
(Sheet 2 of 2)
t
te300-2
SERVICE
‘Ture unit off,
then on for each
sequence.
to sueeT 2
FLOWCHART 5.3 AUTO-TEST CYCLE TROUBLESHOOTING
45 vae\_ no -
at U-3 Sy cheek 86 and Ul. check oR3 and BTL.
ves
88.8°C\_ 0
TEMP display
for 7 Check DS7-0s9, G7, Q8, 99, U18, 19, U20
30, UBL, and associated components.
ves
NO Check S2, DS3, 65, U9, U10, US, U3,
active 029, 030, v31,"Ql2,"Q13, cRI?, CRIB, 032,
for 7 seconds, ‘and associated components.
Fes
oasbie No
‘beep at end o>—m Check DSL, U3, QlI, 028, U33, and aseoeiated
auto test, coaponents.
YES ~
wEATES
PowER Lighee
on for 7
Check DS13, U26, VR2, U4, U30, 127, CRB,
and associated conponents:
Es. 1%
(Sheet 1 of 2)300-2
SERVICE
a FROM SHEET 1
no
Check 0510, U22, U6, U30, 025, U34, U7,
Racae. Val, and associated’ components.
Display
Indicates:
5.08 attey,
Check UI7, UB, U30, US, U34, UI2, UI3, Ula,
16, 19, "u32’
Check 17-1, U21, US, U7, 0510, and
‘assoctated components.
sTop
FLOWCHART 5.3 AUTO-TEST CYCLE TROUBLESHOOTING
(Sheet 2 of 2)300-2
SERVICE
Turn Incubator on, allow
Incubator tenperature
stabilize af initial SET
POINT 35.0°C.
Press SET TEMP
switch 53
"0
a pat check $3, WH, 6, U2, S10 and assectated
Bs ‘components.
Set. Poin
“emp. Digplayed
(38.08)
0
Go to Flonchart 5.5.
ves
Depress and hold
the Set Point Raise
[Al switch; observe]
display.
Bi splay
reading
Check $4, UIT, UIS, UB, U23-6, U12, Ula,
UIA, RHI, TPS) 32) and associated’ components
To SHEET 2
FLOWCHART 5.4 TEMPERATURE SET POINT CIRCUITRY TROUBLESHOOTING
(Sheet 1 of 2)300-2
SERVICE
FROM SHEET 1
°
Press Se
Switch 53.
Depress and Hold
the Set Point
Lower ( (2) )
Switen; observe
Display.
Check $2, ULL, UIS, UB, U23-6, U12, U13,
UL4, RNL, TP3, 32) and associated ‘components,
ves
STOP
FLOWCHART 5.4 TEMPERATURE SET POINT CIRCUITRY TROUBLESHOOTING
(Sheet 2 of 2)300-2
SERVICE
FLOWCHART 5.5 A/D CONVERTER AND TEMPERATURE DISPLAY TROUBLESHOOTING
5-20
Turn the unit on.|
At the end of
Auto-Test cycle,
press SET TEN
switeh $3,
Bispiay
‘encing
3
Check the following
and associated
PPBETBrakion
2: uta, vie
3, 0572059
2 u20, 07, 0, 99,
u
5, Ul2, U13, Ud
RRL, Ul, U32,
W236
STOPC300-2
SERVICE
Connect Probe Simulator to
Lil"on Power Board. Set alarms
switch on simulator to. NOR.
Turn untt on'ang wait unti?
end of Auto-Test cycle (22 sec) |
HI-TEMP ALARH-
Set Simulator
ALARMS. Switch to
HI-TEHP,
Check DS2, U9, US-3, TPA, U4-11, U28-6,
28-10, git, Us, DSi, and associated
components.
HEATER DISABLE
NO check U26, U4-11, U5-2, and associated
Components,
Check Q6, 1K3 and assoctated components.
(OPEN SENSOR ALAR}
Set Simulator
ALARAS Switch: to
OPEN,
v
To SHEET 2
FLOWCHART 5.6 ALARM CIRCUITRY TROUBLESHOOTING
(Sheet 1 of 3)
5-21€300-2
SERVICE
FROM SHEET 1
Senson
after
T see
0
Check DSS, U20, US-11, U24—14, CRI2, U28-6,
28-10, Qi1, US, 081, and associated
conponents
ves
suorTeD sensor —__——]
LARS
Set, Simulator
ALARNS switch to
SHORT.
free NO. Check a11 conditions for HI-TeNP (sheet 1)
and SENSOR and SENSOR alarm (Sheet 2).
Alarm after
YES
[ALARM RESET
Clear alarm con \
ditions and press
alarm reset Swited)
SL.
Check S1, uaa, ud.
components
|. US-1, and associated
EATER TENP. ALARM
Disconnect 1P5
from 105 on Power|
Board.
To SHEET 3
FLOWCHART 5.6 ALARM CIRCUITRY TROUBLESHOOTING
(Sheet 2 of 3)
5-22300-2
SERVICE
FROM SHEET 2
Check CRI7, CRIB, 12, QL3, U2, UI1-4,
5-14, U9, "083, and associated’ components.
°
4
Reconnect 105 to
105 and reset
o
alarm
PONER FAILURE ALARK >
Disconnect AC
Power Cord.
POWER
FAILURE ata Check CR2, CRE, Q3, UL, BTL, O54, and
associated components.
Press CONTROLLER
OFF switch on
Controller.
‘Check U2, 03, CRE, and associated components
stor
FLOWCHART 5.6 ALARM CIRCUITRY TROUBLESHOOTING
. (Sheet 3 of 3) 5.23300-2
SERVICE
Turn the Incu=
«
1K2 OL 2! check 1K?
3 5, Go to Clacchart 5.6.
Eneraized a
0. TO FLONCHART
Ee seful) wave rectified sinewave
FLOWCHART 5.7 HEATER POWER CIRCUITRY TROUBLESHOOTING300-2
SERVICE
Turn Incubator
on.
POWER CONTROL
Check UI6 (TEMP, ANP.), 2RT1-B;
Check U1G (Diff. Anp) and associated
components.
Board)
ves
‘averore at Check U2y 08, G8, ULL, U25, UIT, and
23 ftsoctated comporents.
conan ee
GND—
x
®av-
aw yin
GND
Check JL, Pl, 1P2, 192, 195, and 15.
WONOTE: These waveforms assume
cold Incubator condition.
Check Heater, 2HR1,
2qh, and 254
0 SHEET 2
FLOWCHART 5.8 POWER CONTROL AND INDICATOR CIRCUITRY TROUBLESHOOTING
(Sheet 1 of 3)
5-25gan.
SERVICE
FRON SHEET 1
HEATER ,
POWER Ge
THDICATOR”
Turn the unit off, connect
probe simulator to 101, Set
alarms switeh te norm and OC
Switeh to 369, Turn unit on
and wait for Setpoint indi~
cator to go out.
{ ‘
Set control panel
Setpoint to 35.65,
Mast for setpoint
Janp to go out.
' -
“ero
ater Pwr
etteps > cnack u27, U6, U26, 030-4, and associated
ue components.
res
Set control panel 2
Setpoint to 364°C,
Wait for setpoint
lamp 0°90 out.
[+ Press Alarm Reset.
Four
fleater Pwr.
Lamps Lit
q
ves
TH SHEET 3 Check DS2 [cheek w7-13, 026,
|ui-11 and associ
FLOWCHART 5.8 POWER CONTROL AND INDICATOR CIRCUITRY TROUBLESHOOTING /
(Sheet 2 of 3) (
26300-2
SERVICE
°
. Fron seer 2
Set, Alarms Switch
on stnutator to
Short Position.
36, U5, Uh, URL, onte, ves,
find associaded eonponents.
7
Turn Untt OFF.
stor
FLOWCHART 5.8 POWER CONTROL AND INDICATOR CIRCUITRY TROUBLESHOOTING
(Sheet 3 of 3)
5-27300-2
SERVICE
Turn the Unft On.
MASTER cLock ——————}
OSCILLATOR
ee
ERMON, MO cneck vater2, w2i-t0, and associates
components .
ves
frurn the unit off
Jand’ then on.
‘AUTD-TEST TIMING}
Tuto ata
Fest Cy
7 secs,
check UI, U2, 030, and associated
aienents .
SET POINT THING >}
Set Pom\__ NO i , and associated
eS heck, wy Us URD-9, UT. and
44 seconds, _ .
ves
0 SHEET 2
FLOWCHART 5.9 CLOCK AND TIMING CIRCUITRY TROUBLESHOOTING
(Sheet 1 of 2)300-2
SERVICE
FROM SHEET 1
ua |
farm unit off. connect
Probe Simulator (set to
NORA") to 1k. Turn
junit on and wait for auto~
test cycle to end, Set
alarms switch to "SHORT".
Check U36-1, U23-9, UIT, and assoctated
components.
ALAR CLOCK.
Check U9-7, UI7-6, and associated components
FLOWCHART 5.9 CLOCK AND TIMING CIRCUITRY TROUBLESHOOTING
(Sheet 2 of 2)
5-29300-2
SERVICE
5.4 REMOVAL AND REPLACEMENT PROCEDURES
8.4.1 GENERAL
This section provides removal and replacement procedures for
components of the Model C300 Incubator. Removal and replacement
procedures for components other than those provided are obvious upon
‘inspection.
5.4.2 CONTROLLER
1. REMOVE THE TWO SCREWS that secure the Controller to the underside
of the Base Assembly,
2, SUPPORT THE CONTROLLER with one hand and remove the screw that
secures the Controller to the center of the front rail.
3. LOWER THE CONTROLLER to the front shelf of the Base Assembly.
4, TO REPLACE THE CONTROLLER, reverse the above procedure.
5.4.3 POWER CHASSIS.
1. REMOVE THE TWO SCREWS that secure the cover to the front of the
‘ower Chassis.
2. DISCONNECT THE CABLES located behind the cover.
3. REMOVE THE SCREW from the bottom center of the Power Chassis and
slide the chass¥s out of the Incubator,
4, TO REPLACE THE POWER CHASSIS, reverse the above procedures.
5.4.4 FAN MOTOR REPLACEMENT KIT, PART NUMBER 67 905 57
IMPORTANT: Fan Motor Replacement Kit, Part Number 67 905 57 is used
For replacing defective fan motors in Model C300 Infant Incubators.
It should be noted that two different fan motors have been used in
this Incubator; the motor will be either cylindrical in shape or
square in shape. The kit can be used to replace either type of motor.
This paragraph contains two separate procedures, one for each type of
motor. Use the appropriate procedure.
5-30