300-2 SERVICE SECTION 5 SERVICE 5.1 GENERAL This section provides calibration, troubleshooting, and removal and replacement instructions for the Model C300 Incubator. 5.2 CALIBRATION PROCEDURES 5.2.1 GENERAL This paragraph provides calibration and test procedures for the Model 300 Incubator. Unless otherwise indicated, all calibration procedures are performed under the following conditions: 1, The Controller is unmounted to provide access to the calibration adjustments. (The PC board need not be removed from the Controller). 2. The Incubator is connected to a primary power source of the correct voltage and frequency (see data tag). 3. Ambient temperature for calibration and test is 23°C + 4°C (73°F = 7°F). 5.2.2 TEST EQUIPMENT REQUIRED The test equipment listed below is required for calibration and performing oxygen concentration tests. Equivalent test equipment may be substituted. Probe Simulator, Part No. 67 902 75 Digital VOM, Fluke Model 8000A Oscilloscope - Tektronix 465 Oxygen Analyzer, Sybron/Taylor Model QA580 Flowmeter, Victor - Model 1099-0025 Leakage Tester, Bi-Tek 501 dumper Plug, Part No. 17 731 98 5.2.3 OVERALL CALIBRATION TEST SET-UP 1. REMOVE SENSOR CONNECTOR and connect the Probe Simulator to -1,=2, and -3 on the Power Board, Che 1 (Change 1) st 300-2 SERVICE 2. ON THE PROBE SIMULATOR, set the ALARMS switch to NORM and the °C Swit 3. APPLY MAIN POWER to the Incubator by setting the AC POWER switch on the Power Chassis to the ON-1 position. 4, APPLY POWER TO THE CONTROLLER by pressing the CONTROLLER ON ‘switchs allow the Incubator to complete the auto-test cycle. 5. ALLOW THE INCUBATOR TO RUN FOR AT LEAST 15 MINUTES before proceeding with calibration procedures. 7 6. CHECK 150 HZ CLOCK by connecting an oscilloscope to TP10 (Figure B.1). The clock period should be 430 ms + 30 ms. If necessary, adjust R92 until period is within specifications. (See Note below.) NOTE: In early production C300-2 Incubators, R92 is a fixed resistor; therefore no adjustment can be made in these units. BRIDGE (TEMPERATURE) AMPLIFIER CALIBRATION a 2. 3. CONNECT DVM between TP2 and TP1 (ground). ADJUST R58 for a reading of 3.50 Vde + 0.01 Vde at TP2. DISCONNECT THE DVM. A/D CONVERTER CALIBRATION 7a CONNECT JUMPER PLUG (part no. 17 731 98) between TP6 and TP7. ON THE PROBE SIMULATOR, set the °C switch to 21°C. ADJUST R76 for a reading of 20.0°C on the temperature display. ON THE PROBE SIMULATOR, set the °C switch to 41°C. ADJUST R61 for a reading of 40.0°C on the temperature display. REPEAT THIS PROCEDURE, as required, to obtain accurate adjustments. DO_NOT DISCONNECT the jumper plug from TP6 and TP7. TEMPERATURE OFFSET ADJUSTMENT 5-2 ON THE PROBE SIMULATOR, set the °C switch to 21°C. ADJUST R76 for a reading of 21.0°C + 0,1°C on the temperature ‘displa: DISCONNECT THE JUMPER PLUG from TP6 and TP7. 300-2 SERVICE SINIOd 183.1 GNV SLNAWLSNray b's aHNDIS usTIOMINOD UV oy SOTSIHS Gv we wowysnont VAN altos 5-3 300-2 SERVICE PRESET TEMPERATURE CALIBRATION NOTE: This adjustment must be completed within 15 seconds after the ‘SET TEMP switch is pressed (SET POINT indicator lit). If the SET POINT indicator goes out before the adjustment is completed, the procedure must be repeated. 1. PRESS THE SET TEMP SWITCH; the SET POINT indicator should light. ADJUST R33 for a reading of 35,0°C on the temperature display. 3. IF THE SET POINT INDICATOR GOES OFF during the adjustment, repeat teps T ani MAXIMUM SET POINT TEMPERATURE CALIBRATION PRESS THE SET TEMP SWITCH; the SET POINT indicator should light. 2. PRESS AND HOLD THE SET POINT RAISE switch ([AJ) until the ‘Temperature display stops changing; do not release the SET POINT raise switch. 3. SLOWLY ADJUST R31 clockwise until the temperature display indicates 38.0°C. If the display reading is inadvertently adjusted to a reading higher than 38.0°C, perform the following step; otherwise proceed to step 5. 4. ADJUST R31 slightly counterclockwise. Press the SET TEMP switch ‘and then the SET POINT lower switch ( [¥]) until the temperature display indicates 37.5°C. Repeat steps 1 through 3 of this procedure until the temperature display indicates 38,0°C. 5. PRESS THE SET TEMP SWITCH; the SET POINT indicator should light. 6. PRESS AND HOLD THE SET pornT Lower swrtcH ([Z)) untit the Yemperature display stops changing; the display should indicate 21.5°C + 1.0°C. 7. PRESS THE SET TEMP SWITCH; the SET POINT indicator should light. 8. PRESS AND HOLD THE SET POINT RATSE SWITCH ( [AI ) untit the ‘temperature display indicates 30.0°C = 0.5°C; release the switch. HIGH TEMPERATURE ALARM CALIBRATION 1. ADJUST R103 counterclockwise. 2. CONNECT A JUMPER between TP8 and TP9, 5-4 3. 4. 5. 300-2 SERVICE ON THE PROBE SIMULATOR, set the ALARMS switch to HI TEMP. SLOWLY ADJUST R103 clockwise until the audible alarm "beeps". DO _NOT DISCONNECT JUMPER from TP8 and TP9. ‘ALARM CHECK High Temperature is WAIT SEVERAL SECONDS until the HI TENP alarm activates. The audibTe aTarm shouTd sound and the HEATER POWER indicator should be disabled. PRESS THE RESET SWITCH. The alarm should not reset. SET PROBE SIMULATOR ALARMS switch to NORM. PRESS THE RESET SWITCH. The HIGH TEMP alarm should reset. CHECK R103 ADJUSTMENT by setting Probe Simulator ALARMS switch Back to HIGH TEMP position. The audible alarm should beep. SET PROBE SIMULATOR ALARMS switch to NORM position and disconnect Jumper from TPB and TPS. Open Sensor ey 2. 3. 4 SET PROBE SIMULATOR ALARMS switch to the OPEN position. THE SENSOR ALARM SHOULD ACTIVATE in about 7 seconds. The audible ‘alarm shouTd sound and the HEATER POWER indicator should be disabled. SET THE PROBE SIMULATOR ALARMS switch to NORM position. PRESS THE RESET SWITCH; the alarm should reset. Shorted Sensor b 2. SET PROBE SIMULATOR ALARMS switch to SHORT position. THE SENSOR AND HIGH TEMP ALARMS should activate in about 7 Seconds and the audible alarm should be enabled. SET PROBE SIMULATOR ALARMS SWITCH to NORM position. PRESS THE RESET SWITCH; the alarm should reset. 5-5 300-2 SERVICE = Power Fi ure and Set-Point Retention NOTE: If the Incubator is equipped with optional battery pack, perform Step 1, otherwise proceed to Step 2. 1. DISCONNECT THE BATTERY TRAY by sliding it out of the Incubator about two inches. 2. PRESS THE SET TEMP switch and then press the set point raise TfA]) switch and raise the set point display to 36.0°C. 3. SET THE MASTER AC POWER SWITCH on the Power Chassis to the OFF-O a position. The POWER FATLURE alarm should be-enabled and a continuous audible alarm should sound. 4, SET THE MASTER AC POWER SWITCH on the Power Chassis to the ON-1 posTtion. The POWER FATLURE alarm should cease and the auto test sequence should be initiated. - 5. WAIT FOR THE AUTO TEST SEQUENCE TO END and then check the set point display; Tt shouTd rea: 6. SET THE MASTER AC POWER SWITCH on the Power Chassis to the OFF-0 posTtion. The POWER FAILURE alarm should be enabled and a continuous audible alarm should sound. 7. PRESS THE CONTROLLER OFF SWITCH on the Controller, the POWER ( ‘alarm should cease, 8. RECONNECT AND SECURE the Battery Tray if Incubator is equipped with opttonal Battery Pack. 9. SET THE MASTER AC POWER SWITCH on the Power Chassis to the ON-1 posttton, and depress the CONTROLLER ON switch on the Control Panel to restore operation. 10. DISCONNECT THE PROBE SIMULATOR from 1J1 and reconnect plug 1P1. 11, REASSEMBLE THE INCUBATOR ~ 5-6 300-2 SERVICE 5.2.4 BATTERY CHARGER (ACCESSORY) CALIBRATION PROCEDURE If the Incubator is equipped with the Battery/Charger Tray accessory, calibrate the battery charger as follows: 1, Set the master AC POWER switch on the Power Chassis to the OFF-0 position and depress the CONTROLLER OFF switch on the Controller front panel. 2. Loosen the Battery/Charger Tray retaining screw completely. 3. Grasp the handle and very slowly withdraw the tray approximately 6 inches (15 cm). WARNING: The Battery/Charger Tray weights apprxoimately 50 Ibs. 22.7 Kg) with two batteries. The tray must be well supported during removal and installation. Exercise care to prevent, pinching of fingers during removal or installation of Battery/Charger Tray. 4, Position a hand on each side of the tray and as close as possible to the Incubator. 5. Slowly remove the tray from the Incubator. Place the tray on a surface that will support its weight. 6. Disconnect the red and black leads from the two batteries. 7. Connect the Battery/Charger Tray to the primary ac power source. 8. Using a digital voltmeter, measure the voltage at 402-11 (see Figure 7.3). The reading should be 13.8 VOC + 0.1 VDC; adjust 4R10 if required to obtain correct reading. 9. Disconnect all test equipment, reconnect the red and black wires to the batteries and reinstall the Battery/Charger Tray in the Incubator. 5.2.5 CHASSIS LEAKAGE CURRENT TEST TEST SET-UP 1. Connect the Power Chassis to the primary power source through an ungrounded adapter plug, so that the unit is ungrounded. Set the AC POWER switch to ON-1 position. 2, The leakage test standards provided in the procedure below assume leakage through a resistance of 1000 ohms. If the Leakage Tester being used does not provide this resistance, the test set-up must be adjusted to provide it. (Change 1) 5-7 300-2 SERVICE PROCEDURE. USE THE LEAKAGE TESTER to measure between the chassis of the unit undér test and @ known ground such as the ground connection of a wall receptacle. The leakage must not exceed 100 microamps. 2. REVERSE THE PLUG and repeat step 1. 3. PERFORM STEPS 1 AND 2 with the Power Chassis AC POWER switch in the OFF-0 position. * 5.2.6 OXYGEN CONCENTRATION TESTS TEST SETUP. = 1. Place a calibrated oxygen analyzer on the mattress in the Incubator. 2. Connect an oxygen supply, through a flowmeter, to the OXYGEN INLET connector on the side of the Incubator. PROCEDURE. 1, TURN THE INCUBATOR ON and adjust oxygen flow rate to 2.0 LPM + - PM 2. THE HOOD OXYGEN CONCENTRATION SHOULD ATTAIN 32 TO 47% oxygen within 30 minutes and stabilize. 3. INCREASE THE OXYGEN FLOW RATE to 6.0 LPM +0.5, -0.0 LPM. The ‘ood oxygen concentration should attain 58 to 90% oxygen within 30 minutes and stabilize. 5.2.7 OPERATIONAL CHECKOUT Perform the Operational Checkout Procedure given in paragraph 2.4. 5.3 TROUBLESHOOTING 5.3.1 GENERAL Troubleshooting guides for the C300 Incubator are presented under paragraph 5.3.3 in the form of flowcharts. It is assumed that an ‘ attempt has been made to calibrate the unit and that all cables are in good condition, 5-8 €300-2 SERVICE 5.3.2 TEST EQUIPMENT REQUIRED The test equipment listed below is required for troubleshooting the 300 Incubator. Equivalent test equipment may be substituted. Digital VOM, Fluke Model 8000A Logic Probe - capable of + 12 VOC Oscilloscope, Tektronix 465 Probe Simulator, Part No. 67 902 75 Jumper Plug, Part No. 17 731 98 5.3.3 TROUBLESHOOTING PROCEDURES The following flowcharts are provided as an aid in localizing the cause of equipment malfunctions. The charts are intended for use in conjunction with the equipment theory of operation (Section 3) and the schematic diagrams (Section 7). It is assumed that the Operational Checkout (paragraph 2.4) has been performed, and that the Calibration Procedures (paragraph 5.2) have been attempted. 5-9 300-2 SERVICE Connect Incubator to Prinary Pover Source (See Daca Tag). Tura the Unit Power Distribution Controle vo. Operating > to Flovehart 5.2. Auto Test Test Sequence correct, Go to Flowchart 5.3. Heater Circuitry Heater Bleneat On| Go to Flowchart 5.7. Displays Go to Flowchart 5.4. Go to Flowehare 5.8. 10 sHEer 2 FLOWCHART 5.1 OVERALL CHECKS (Sheet 1 of 4) igh Ate Teap Alarm 300-2 SERVICE FROM SHEET 1 [Allow Tncubator to Heat for Approxi~ nately 1 Hour Press Set Teap Seiteh; Set Potnt Indicator should 1ight| land short beep should be heard. mediately depress ‘and hold both the Setpoint raise (A) and Setpoint Lover [Y) Switches and observe ‘Téxperature Display. When Display indicates 40.0°C, release the switches, [Allow Incubator to heat up. Go to Flowchart 5.6, Tteater 0 Go to Flowchart 5.8, [Decrease Teap Set Point to 35°C and allow Incubator co cool suffictently unt? HI-TEX lara can be reset.| | 110 SHEET 3 FLOWCHART 5.1 OVERALL CHECKS (Sheet 2 of 4) 5-11 300-2 : SERVICE FROM SHEET 2 Sensor Alam_________4 Disconnect IPI from 11 fon Power Board, Connect| Probe Simulator to 11, Set Alarms Switch on Simator to Open. ‘Sensor xo (Alarm in? >———™ Go to Plowhart 5.6. Seconds, ves Set Simulator Alarne Seitch to Short, ‘irre x0 “and Sensor >» Go to Flowchart 5.6. are in 7, ves Disconnect Probe Simulator and Re-| connect 1J1 and wh, i Disconnect 1P5 fron 135 on Power Board To Sheet 4 FLOWCHART 5.1 OVERALL CHECKS (Sheet 3 of 4) 5-12 300-2 SERVICE From Sheet 3 check CRI7, CRIB, O12, 13, U2, 31-4, UIS-14, U9, 0s: and assoc{ated conponents. Tecomect 195 to , 1s on tover foard” and eset 8 Alarm 1 Disconnect AC Power Cord Check CR2, CRE, Q3, Ul, BTL, DS4, ‘and associated congonents.. ¥ Press controller OFF switch on controler, Check UL, 03, CR5, and associated components. stop FLOWCHART 5.1 OVERALL CHECKS (Sheet 4 of 4) 5-13 300-2 SERVICE - Connect Tncubator to Prinary Pover Source (See Data Tag). Tura the Unit - ons Wale’7 gece. for Auto-Test to end, Power Chasais Check AC Circult Breaker(s). Check Line Filter 2 FL1 and Power Switeh 231. 3.” Check Terminal block connections at 2B. Check Power Transforaer 271. Check C81 and 1CR2. Check 10R3 and Associated Componente, Check 10R2. Check URS. Check IK1, IVRI and associated components. *ull wave rectified sinewave, To sueer 2 FLOWCHART 5.2 POWER DISTRIBUTION TROUBLESHOOTING (Sheet 1 of 2) 300-2 SERVICE FROM SHEET 1 POWER CHASSIS (Cont.). ® Check 1K operation, e Check 1K2 operation. - Control Board: - Check interface cable assenbly for continuity. : Check on2. - stor FLOWCHART 5.2 POWER DISTRIBUTION TROUBLESHOOTING (Sheet 2 of 2) t te 300-2 SERVICE ‘Ture unit off, then on for each sequence. to sueeT 2 FLOWCHART 5.3 AUTO-TEST CYCLE TROUBLESHOOTING 45 vae\_ no - at U-3 Sy cheek 86 and Ul. check oR3 and BTL. ves 88.8°C\_ 0 TEMP display for 7 Check DS7-0s9, G7, Q8, 99, U18, 19, U20 30, UBL, and associated components. ves NO Check S2, DS3, 65, U9, U10, US, U3, active 029, 030, v31,"Ql2,"Q13, cRI?, CRIB, 032, for 7 seconds, ‘and associated components. Fes oasbie No ‘beep at end o>—m Check DSL, U3, QlI, 028, U33, and aseoeiated auto test, coaponents. YES ~ wEATES PowER Lighee on for 7 Check DS13, U26, VR2, U4, U30, 127, CRB, and associated conponents: Es. 1% (Sheet 1 of 2) 300-2 SERVICE a FROM SHEET 1 no Check 0510, U22, U6, U30, 025, U34, U7, Racae. Val, and associated’ components. Display Indicates: 5.08 attey, Check UI7, UB, U30, US, U34, UI2, UI3, Ula, 16, 19, "u32’ Check 17-1, U21, US, U7, 0510, and ‘assoctated components. sTop FLOWCHART 5.3 AUTO-TEST CYCLE TROUBLESHOOTING (Sheet 2 of 2) 300-2 SERVICE Turn Incubator on, allow Incubator tenperature stabilize af initial SET POINT 35.0°C. Press SET TEMP switch 53 "0 a pat check $3, WH, 6, U2, S10 and assectated Bs ‘components. Set. Poin “emp. Digplayed (38.08) 0 Go to Flonchart 5.5. ves Depress and hold the Set Point Raise [Al switch; observe] display. Bi splay reading Check $4, UIT, UIS, UB, U23-6, U12, Ula, UIA, RHI, TPS) 32) and associated’ components To SHEET 2 FLOWCHART 5.4 TEMPERATURE SET POINT CIRCUITRY TROUBLESHOOTING (Sheet 1 of 2) 300-2 SERVICE FROM SHEET 1 ° Press Se Switch 53. Depress and Hold the Set Point Lower ( (2) ) Switen; observe Display. Check $2, ULL, UIS, UB, U23-6, U12, U13, UL4, RNL, TP3, 32) and associated ‘components, ves STOP FLOWCHART 5.4 TEMPERATURE SET POINT CIRCUITRY TROUBLESHOOTING (Sheet 2 of 2) 300-2 SERVICE FLOWCHART 5.5 A/D CONVERTER AND TEMPERATURE DISPLAY TROUBLESHOOTING 5-20 Turn the unit on.| At the end of Auto-Test cycle, press SET TEN switeh $3, Bispiay ‘encing 3 Check the following and associated PPBETBrakion 2: uta, vie 3, 0572059 2 u20, 07, 0, 99, u 5, Ul2, U13, Ud RRL, Ul, U32, W236 STOP C300-2 SERVICE Connect Probe Simulator to Lil"on Power Board. Set alarms switch on simulator to. NOR. Turn untt on'ang wait unti? end of Auto-Test cycle (22 sec) | HI-TEMP ALARH- Set Simulator ALARMS. Switch to HI-TEHP, Check DS2, U9, US-3, TPA, U4-11, U28-6, 28-10, git, Us, DSi, and associated components. HEATER DISABLE NO check U26, U4-11, U5-2, and associated Components, Check Q6, 1K3 and assoctated components. (OPEN SENSOR ALAR} Set Simulator ALARAS Switch: to OPEN, v To SHEET 2 FLOWCHART 5.6 ALARM CIRCUITRY TROUBLESHOOTING (Sheet 1 of 3) 5-21 €300-2 SERVICE FROM SHEET 1 Senson after T see 0 Check DSS, U20, US-11, U24—14, CRI2, U28-6, 28-10, Qi1, US, 081, and associated conponents ves suorTeD sensor —__——] LARS Set, Simulator ALARNS switch to SHORT. free NO. Check a11 conditions for HI-TeNP (sheet 1) and SENSOR and SENSOR alarm (Sheet 2). Alarm after YES [ALARM RESET Clear alarm con \ ditions and press alarm reset Swited) SL. Check S1, uaa, ud. components |. US-1, and associated EATER TENP. ALARM Disconnect 1P5 from 105 on Power| Board. To SHEET 3 FLOWCHART 5.6 ALARM CIRCUITRY TROUBLESHOOTING (Sheet 2 of 3) 5-22 300-2 SERVICE FROM SHEET 2 Check CRI7, CRIB, 12, QL3, U2, UI1-4, 5-14, U9, "083, and associated’ components. ° 4 Reconnect 105 to 105 and reset o alarm PONER FAILURE ALARK > Disconnect AC Power Cord. POWER FAILURE ata Check CR2, CRE, Q3, UL, BTL, O54, and associated components. Press CONTROLLER OFF switch on Controller. ‘Check U2, 03, CRE, and associated components stor FLOWCHART 5.6 ALARM CIRCUITRY TROUBLESHOOTING . (Sheet 3 of 3) 5.23 300-2 SERVICE Turn the Incu= « 1K2 OL 2! check 1K? 3 5, Go to Clacchart 5.6. Eneraized a 0. TO FLONCHART Ee seful) wave rectified sinewave FLOWCHART 5.7 HEATER POWER CIRCUITRY TROUBLESHOOTING 300-2 SERVICE Turn Incubator on. POWER CONTROL Check UI6 (TEMP, ANP.), 2RT1-B; Check U1G (Diff. Anp) and associated components. Board) ves ‘averore at Check U2y 08, G8, ULL, U25, UIT, and 23 ftsoctated comporents. conan ee GND— x ®av- aw yin GND Check JL, Pl, 1P2, 192, 195, and 15. WONOTE: These waveforms assume cold Incubator condition. Check Heater, 2HR1, 2qh, and 254 0 SHEET 2 FLOWCHART 5.8 POWER CONTROL AND INDICATOR CIRCUITRY TROUBLESHOOTING (Sheet 1 of 3) 5-25 gan. SERVICE FRON SHEET 1 HEATER , POWER Ge THDICATOR” Turn the unit off, connect probe simulator to 101, Set alarms switeh te norm and OC Switeh to 369, Turn unit on and wait for Setpoint indi~ cator to go out. { ‘ Set control panel Setpoint to 35.65, Mast for setpoint Janp to go out. ' - “ero ater Pwr etteps > cnack u27, U6, U26, 030-4, and associated ue components. res Set control panel 2 Setpoint to 364°C, Wait for setpoint lamp 0°90 out. [+ Press Alarm Reset. Four fleater Pwr. Lamps Lit q ves TH SHEET 3 Check DS2 [cheek w7-13, 026, |ui-11 and associ FLOWCHART 5.8 POWER CONTROL AND INDICATOR CIRCUITRY TROUBLESHOOTING / (Sheet 2 of 3) ( 26 300-2 SERVICE ° . Fron seer 2 Set, Alarms Switch on stnutator to Short Position. 36, U5, Uh, URL, onte, ves, find associaded eonponents. 7 Turn Untt OFF. stor FLOWCHART 5.8 POWER CONTROL AND INDICATOR CIRCUITRY TROUBLESHOOTING (Sheet 3 of 3) 5-27 300-2 SERVICE Turn the Unft On. MASTER cLock ——————} OSCILLATOR ee ERMON, MO cneck vater2, w2i-t0, and associates components . ves frurn the unit off Jand’ then on. ‘AUTD-TEST TIMING} Tuto ata Fest Cy 7 secs, check UI, U2, 030, and associated aienents . SET POINT THING >} Set Pom\__ NO i , and associated eS heck, wy Us URD-9, UT. and 44 seconds, _ . ves 0 SHEET 2 FLOWCHART 5.9 CLOCK AND TIMING CIRCUITRY TROUBLESHOOTING (Sheet 1 of 2) 300-2 SERVICE FROM SHEET 1 ua | farm unit off. connect Probe Simulator (set to NORA") to 1k. Turn junit on and wait for auto~ test cycle to end, Set alarms switch to "SHORT". Check U36-1, U23-9, UIT, and assoctated components. ALAR CLOCK. Check U9-7, UI7-6, and associated components FLOWCHART 5.9 CLOCK AND TIMING CIRCUITRY TROUBLESHOOTING (Sheet 2 of 2) 5-29 300-2 SERVICE 5.4 REMOVAL AND REPLACEMENT PROCEDURES 8.4.1 GENERAL This section provides removal and replacement procedures for components of the Model C300 Incubator. Removal and replacement procedures for components other than those provided are obvious upon ‘inspection. 5.4.2 CONTROLLER 1. REMOVE THE TWO SCREWS that secure the Controller to the underside of the Base Assembly, 2, SUPPORT THE CONTROLLER with one hand and remove the screw that secures the Controller to the center of the front rail. 3. LOWER THE CONTROLLER to the front shelf of the Base Assembly. 4, TO REPLACE THE CONTROLLER, reverse the above procedure. 5.4.3 POWER CHASSIS. 1. REMOVE THE TWO SCREWS that secure the cover to the front of the ‘ower Chassis. 2. DISCONNECT THE CABLES located behind the cover. 3. REMOVE THE SCREW from the bottom center of the Power Chassis and slide the chass¥s out of the Incubator, 4, TO REPLACE THE POWER CHASSIS, reverse the above procedures. 5.4.4 FAN MOTOR REPLACEMENT KIT, PART NUMBER 67 905 57 IMPORTANT: Fan Motor Replacement Kit, Part Number 67 905 57 is used For replacing defective fan motors in Model C300 Infant Incubators. It should be noted that two different fan motors have been used in this Incubator; the motor will be either cylindrical in shape or square in shape. The kit can be used to replace either type of motor. This paragraph contains two separate procedures, one for each type of motor. Use the appropriate procedure. 5-30

También podría gustarte