Está en la página 1de 5

Tandem Hook Up Kit for Midmark Compressors Applies to Models:

All Midmark / Apollo / Matrx


[30186000] Compressors
Special Tools:
none
Note Intended Use:
Low Voltage Control is required on both compressors. .To provide simultaneous operation of two Midmark compressors. Language of origin: English

Step 1: Disconnect supply power to the compressors.

Step 2: Adjust pressure switches on both compressors.

A) Remove cover on each pressure switch.


B) Turn adjuster clockwise three full turns to bump up the pressure
settings to the highest level so the contact tips never open.
C) Replace covers

Note: Make sure to perform these steps on both compressors so


the system pressure will be regulated by the tandem pressure
switch.

Adjuster

AA206900

2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com TP201 20-42-FO-00013 Rev A1 C2169 10560500 Rev. F
1 Style G
Color Page

Step 3: Attach Air Lines.

A) Connect supply air lines from compressors to check valves on tee assembly.
B) Connect high pressure hose from bottom of tee assembly to operatory.

Note: Verify arrows on check valves are pointing toward tee assembly and shutoff
valves are open on each compressor. Check each compressor for leaks during initial
operational check.

Note
Reducers may be required for air line connections, two are supplied

Air Line to
Check Valve

AA207000

Apollo Air Line


Connection Compressor Shutoff
Air Line to Valve Must be Opened
Operatory Air Line Check Valve

2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com TP201 20-42-FO-00013 Rev A1 C2169 10560500
2 Style G
Color Page

Step 4: Low Voltage Control Wiring. (Midmark& Apollo)

A) Install tandem air switch in control panel.


B) Connect low voltage wires from compressors to pressure switch Midmark & Apollo Control Wiring
and tandem air switch on control panel as shown below. To Control To Pressure
Compressor To Light
Panel Switch
Note: Matrx Compressor Diagram Not Shown .
Master Blue Red White
Slave Blue Red Not Used
Matrx Control Wiring (NOT SHOWN)
Master Red White Blue
Slave Red White Not Used

Pressure Switch

Slave Compressor

Master Compressor
AA210200-1

Control Panel Switch


(Rear View)

2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com TP201 20-42-FO-00013 Rev A1 C2169 10560500
3 Style G
Color Page

Step 5: Testing

A) Connect power supply and start compressors by remote switch.

Note: Compressors should start at 80 PSI and stop at 100 PSI, simultaneously.
Perform adjustment if necessary.

B) If needed turn adjuster nut on the tandem pressure switch.


- Clockwise will increase reading .
- Counter clockwise will decrease reading.

Adjuster AA194100

2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com TP201 20-42-FO-00013 Rev A1 C2169 10560500
4 Style G
6 61945200

77005143 1

5 PCB50615

4 10560700

10378700 2

3 10560400

30186000 Tandem Compressor Hook Up Kit

Item Description Qty.


1 Pressure Switch 1
2 Check Valve, Air 2
3 Wire, 3 Conductor, 18 GA. (50ft) 2
4 Air Low Voltage Control Switch (4 Pole) 1
5 Bushing 1/2x 3/8 2
6 Wire Nut (Blue) 15

Serial Numbers / Dates: all

2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com TP201 20-42-FO-00013 Rev A1 C2169 10560500