Está en la página 1de 6

Metallic Medical Implants:

Electrochemical Characterization of Corrosion Processes


by Patrik Schmutz, Ngoc-Chang Quach-Vu, and Isabel Gerber

T
he use of metallic alloys for a contact, there is a crevice generated at (open circuit potential) (such as, for
whole range of medical implants the contact surface. Electrochemical example, electrochemical impedance
is justified by their superior methods further allow for monitoring spectroscopy, or EIS) can be used to
mechanical properties (hardness, of the release of toxic ions in the follow actively corroding systems like
stiffness, etc.) compared, for example, body, which is another major issue long degradable implants.
to polymers. Other properties like before failure occurs. In the next sections, illustrations
biocompatibility or visibility in X-ray Not all the metallic materials of different types of corrosion pheno-
images also can be mentioned. One of and alloys used for implants have a menon taking place are presented and
their drawbacks is that electrochemical similar risk of corrosive degradation. discussed. For obvious confidentiality
reactions take place on metallic surfaces
in the human body. To replicate
the real environment as closely as
possible, implants should be tested
in vivo in animal experimentation;
but the possibility of monitoring
electrochemical processes is then
very limited and not straightforward.
(Editors Note: See also Hiromotos
article in this issue.) In vitro reactivity
characterizations help to understand
the degradation processes (failure risks)
and the development of new implant
materials. Different macro- and micro-
electrochemical methods allow the
investigation of uniform and localized
corrosion susceptibility and its relation
to material microstructure. A major
difference with classical corrosion
investigations is the complexity of the
physiological media with the presence
of proteins (and cells). The influence
of solution chemistry on degradation
mechanisms, as well as of the specific
temperature and atmosphere (amount
of O2 and CO2), has to be investigated.
Electrochemical methods can also be Fig 1. Main types of metallic materials used for medical implants and their susceptibility
used for implant surface functionalizing to corrosion.
by growing tailored anodic oxide layers
or deposition of coatings. Figure 1 presents some of the main reasons, no detailed indication
There is a whole range of issues (or interesting) categories of materials of products or implant types and
related to corrosion processes that used in implants with their respective geometries can be given and this
needs to be considered and addressed susceptibility and the type of corrosion contribution is focused on a conceptual
experimentally. Degradation of the expected to occur. It is well known discussion of corrosion processes.
implants can be uniform, but for most that Ti and Ti alloys are very corrosion
of the standard metallic materials resistant and therefore the choice of Corrosion Resistant Implants
used (stainless steel, Co based alloys), testing media and conditions is not
localized corrosion related to very critical. On the other extreme, Mg (Ti and Ti Alloys)
microstructural features is observed. alloys are extremely reactive, therefore
The metallic surface is often covered good candidates for degradable Titanium and titanium alloys show
by a native anodically grown oxide implants. Here, an exact understanding a high corrosion resistance due to
layer that guarantees a uniform of the corrosion mechanisms and of their stable passive layer. Therefore,
corrosion resistance, but does not the influence of ions or species present titanium surfaces are mostly mentioned
prevent localized breakdown when an in the physiological media is a major in relation with electrochemical
aggressive environment is present. This challenge. This field is a good example corrosion processes when they react as
is typically the case when chloride of positive use of corrosion processes. a cathode in contact with other metallic
ions are present in physiological media. Concerning the electrochemical materials. Some surface processing,
Crevice corrosion related to the complex methods used for characterization of such as sandblasting, induces rough and
geometries of implants and galvanic the corrosion processes, they can be contaminated surfaces and there might
coupling between dissimilar materials divided in two categories. First the be an increased risk that this surface
used can also occur and can be followed polarization methods used to assess condition results in higher corrosion
in model experimental devices. These the susceptibility to localized corrosion susceptibility. Electrochemical investi-
two types of corrosion often occur for corrosion resistant materials. gations of the corrosion behavior of
simultaneously because when two Second, measurements performed Ti and Ti alloys have almost always
different materials are brought into at the free corrosion potential demonstrated very good passivation
behavior of the surface. In physiological
The Electrochemical Society Interface Summer 2008 35
Schmutz, et al.
(continued from previous page)

media, it can be shown that some


modification of the surface oxide
composition occurs1 or that deposition
of Ca and P, in the case of exposure to
Ringers solution, delays the stabilization
of TiO2 surface oxide.2 When the media
gets acidic (0.5M H2SO4), Ti alloys show
dissolution and are less stable then
pure Ti.3 These conditions are however
extremely aggressive compared to what
is supposed to occur in the human
body and can only be envisaged in the
crevice situation. The different buffering
capacity of the physiological liquids in
the body is still in discussion, and
although the presence of low pH in the
case of infection is always postulated,
no direct evidence is available because
of the difficulty of performing these
measurements in very small amount
of liquids.
However, to conclude that Ti alloys
are totally immune to corrosive attack
would be a mistake. Figure 2 shows an
SEM image of a metallographic section Fig 2. Corrosion fatigue crack propagation in Ti-Al-V implant after human implantation.
for a Ti6Al4V bone replacement pin.
The presence of a crack with secondary slightly more negative potential for
physiological media are considered,
crack ramifications is clearly visible. the Co alloy. The 316L stainless steel
only a few detailed electrochemical
This failure mode is typical of fatigue then shows a breakdown at 0.3 V (SCE)
investigations can be found. For the
crack growth with corrosive dissolution. corresponding to the onset of localized
Co based alloys, there is currently an
The pin broke after 6 months of corrosion in this physiological solution.
increased awareness about potential
implantation and the patient needed it The MP35N alloys demonstrate a
risks related to localized corrosion in
replaced by a new surgical intervention. lower susceptibility to localized
relation with infection and/or toxicity
A combination of cyclic loading and corrosion and the current increase at
of the corrosion products.6,7 Toxicity of
a corrosive environment had to be higher potential correspond to the
elements like Ni, Cr for stainless steel
present for this failure to occur and transpassive dissolution of chromium.
and Co, Cr for the Co alloys is currently
electrochemical characterization of This type of potentiodynamic pola-
being debated. Molybdenum, which is
crack growth is difficult. In the case rization experiment is necessary for
present as an alloying element in both
of Ti and Ti alloys, it can be stated any detailed characterization of the
types of materials, is also included in
that electrochemical methods are localized corrosion susceptibility, but is
this discussion. Tribocorrosion studies
not the best tools to assess corrosion not quite representative of the situation
are gaining in importance in relation
susceptibility unless they are coupled to found for implanted materials. There,
with this toxicity issue because of the
mechanical or tribological solicitations usually crevice conditions with very
influence of friction or fretting on
that are very important but not the small amounts of electrolyte are found
the local depassivation and release of
focus of this article. and the polarization is induced by more
metallic ions.8 When the corrosion
mechanism of Co alloys is discussed, noble materials in the surrounding
Corroding Implants (Stainless it must be kept in mind that quite a area or chemicals acting as oxidizing
Steel and Co-based Alloys) large composition range is considered. agents.
There is the well known Co30Cr6Mo Crevice and galvanic corrosion.
This second category is represented implant, but also alloys like the MP35N Zardiackas, et al. published results on
by implant materials that are suffering (Co35Ni20Cr11Mo1Fe) or other versions galvanic coupling experiments between
from localized corrosion attacks. containing W are used such as L306 Co based and Ti alloys9 as well as
This fact is until now often neglected that are much stiffer. with different stainless steels.10 More
because this corrosion phenomena is Localized corrosion susceptibility. interesting are the galvanic coupling
unknown to surgeons or observed, but When the susceptibility to localized phenomena investigated in crevice
the consequences are accepted. This corrosion needs to be addressed, the conditions, often with Ti implants
statement is worth being made because first electrochemical characterizations being one of the materials.11,12 Only a
the 316L (X17Cr12Ni2Mo) stainless that are performed are potentiodynamic few studies of the corrosion processes
steel is certainly besides the 304 polarization measurements. A standard investigated after in vivo implantation11
(X18Cr10Ni), one of the most studied three electrode cell with a platinum or taking into account fretting between
alloys in terms of the pitting and crevice counter and a calomel (Hg/Hg 2Cl 2 ) dissimilar materials13-15 can be found in
corrosion mechanisms. The influence (replaced now by Ag/AgCl) reference the literature.
of inclusions present in the material, as electrode is usually used. Figure 3 In order to simulate this situation
well as different parameters like applied presents these curves for 316L and an for in vitro testing conditions, a setup
stress or temperature on localized MP35N Co-based alloy. The solution is used where two metallic surfaces are
corrosion initiation susceptibility, used in this case is a Ringers solution brought together without direct contact
has been investigated locally with (9 g/l NaCl: 0.42 g/l KCl; 0.48 g/l CaCl2 ; (Fig. 4a). The current is measured with
the electrochemical microcell and 0.2 g/l NaHCO3) adjusted to pH 5 and the help of an ampere meter (usually
documented by Suter, et al.4,5 But as maintained at a controlled temperature built directly in the potentiostat).
soon as biomedical applications and of 37C. It can be observed that both There are standard tests (for example
alloys are passive at the OCP with a the ASTM G71) that describe the
36 The Electrochemical Society Interface Summer 2008
experiments to be performed but they
are unsatisfactory in the sense that one
of the most important parameters, the
crevice width, is not mentioned. Figure
4b shows an example of typical results
obtained during a galvanic coupling
experiment performed between a 316L
and a MP35N plate separated by a
distance of 1 mm. The current flowing
(blue curve) between the two electrodes
is measured, and the way the electrodes
are connected, the current indicates
that the steel electrode acts as the
cathode and MP35N the anode. The
potential on the 316L electrode is also
recorded during the whole experiments
(red curve). The very interesting fact
is that after approximately 4 hours
of immersion, there is an observed
potential drop that can be associated
with an activation of the steel surface.
This coupling phenomenon is only
observed with crevice widths of 1 mm
or smaller, and support the fact that
standard tests with two electrodes
placed far away from each other will not
Fig. 3. Potentiodynamic polarization measurements: characterization of localized allow to generate the crevice solution
corrosion susceptibility for stainless steel and MP35N Co alloy. condition required for corrosion to
take place. Additional characterization
can be performed simultaneously,
(a) such as pH monitoring or ion release
characterization in the crevice, when
optimized setups are used. Ringers
solution with slightly lower pH to
simulate an infection is used for this
example, but investigations performed
in a whole range of other media are
found in the literature. For this type
of implant materials and corrosion
processes, it is difficult to suggest one
solution as the standard, as they do
not give fundamentally different results
in terms of corrosion processes. The
situation is totally different for the next
type of implant materials.

Degradable Implants
(Mg Alloys)
Permanent implants may induce
long term complications and require
(b) surgery to replace them. An alternative
for specific applications is a degradable
implant made of Mg alloy. Mg is
biocompatible, vital for metabolic
processes, and the alloys show higher
strength than polymers. The positive
use of corrosion processes and a
fundamental understanding of the
mechanisms are here central aspects.
A first requirement is temporary
corrosion protection obtained by
surface oxidation as long as mechanical
strength is needed. Afterward,
uniform corrosion needs to take
place to induce implant dissolution.
Mg alloys corrode fast in neutral
electrolytes and a coating usually
aims at the best possible corrosion
protection. For degradable implants, a
different approach with two challenges

Fig. 4. (a) Crevice corrosion measurement principle and (b) galvanic coupling current
between 316L stainless steel and MP35N cobalt based alloy.

The Electrochemical Society Interface Summer 2008 37


Schmutz, et al. Table I. Ion concentrations (mM) of blood plasma, artificial plasma, and SBF K9.
(continued from previous page)
Blood Plasma Artificial Plasma SBF K9
is necessary: (i) an understanding of
the microscale corrosion mechanisms; Na+ 142.0 144.5 142.0
and (ii) development of a temporary
corrosion protection for at least 3 K +
5.0 5.4 5.0
months (this aspect will not be further Mg 2+
1.5 0.8 1.5
discussed here as it is not the focus of
this article). Another constraint is that Ca 2+
2.5 1.8 2.5
Al-free alloys (the toxicity of Al is still
debated) like Mg-Y-RE (ex. Nd) need to Cl -
103.0 125.3 148.8
be developed for medical applications. (HCO3) -
27.0 26.2 4.2
For this type of implant, the
degradation rates have been first (HPO4) 2-
1.0 3.0 1.0
assessed by in vivo tests16-19 but the
main problem is that without having a SO42- 0.5 0.8 0.5
better understanding of the key factors
controlling the corrosion processes, The different reactions can be followed compounds when exposed to light
life-time prediction is difficult based online as a function of immersion or HEPES can produce toxic oxygen
on the in vivo studies. An important time without perturbing the corrosion metabolites. 26 It is recommended to
scattering in the degradation rate is processes too much. A qualitative avoid HEPES and similar organic buffers
always reported. Here, more than for all comparison of the uniform dissolution in studies of oxidative compounds
other types of implants, there is a need rate for AP and SBF can be obtained by as it interferes with peroxynitrite
for detailed in vitro studies. Currently, considering the impedance modulus and nitric oxide. For a simultaneous
there are only very few studies using amplitude value Z at low frequency electrochemical characterization,
electrochemical methods to address the (10 mHz). This real value, also called solvent conductivity is an additional
role of different ions and buffering polarization resistance, is however still mandatory criterion.
strength on the corrosion rates20 or influenced by the localized corrosion The very basic requirement of a
considering additionally the influence processes that can occur in parallel, so biocompatible material is that the cells
of proteins (albumin).21 Other studies over interpretation of the data should stay alive and are metabolically active
include evaluation of cytotoxicity in the be avoided. However, it can still be seen during long-term cultures. This can
in vitro tests but without performing that uniform dissolution rate differ by a be tested with two rapid and simple
electrochemical characterization. 22 factor of 2030 between the aggressive quantitative assays, namely viability
Testing media.From the in vitro SBF and AP (Fig. 5b) and this is a major based on physical uptake of neutral red
investigation, it can be stated that a concern when prediction of implant (NR) and metabolic activity based on
very critical aspect for biodegradable life is necessary. A critical parameter an MTT (3-(4,5-dimethylthiazol-2-yl-) -
implants is the physiological medium is the surface pH and the buffering 2,5-diphenyltetrazoliumbromide) assay,
to which the surface is exposed. This ability of the different media. Figure 5c which is dependent on the activity
can be blood (simulated in vitro by shows the pH evolution in very small of intracellular enzymes (Fig.6).
Artificial Plasma, or AP) or other body amount of liquid (2 ml on 1 cm2) for Both assays are widely accepted in
fluids (SBF) depending on the implant initially neutral distilled water (H2O), biocompatibility and cytotoxicity
location. Table I presents the ionic AP, and SBF. The pH increase as a result studies for the assessment of viability
contents for these typical media used for of Mg corrosion is the most hindered and growth of cells.27 Another very
in vitro testing. It has to be mentioned for SBF. Only at pH 9 in AP, the Mg common parameter is the quantifica-
that SBF is a very open denomination hydroxide is starting to be stable. In tion of the activity of lactate de-
that allows for example variation of SBF, a different surface oxide is needed hydrogenase in the supernatant of
the concentrations and the buffering in order to guarantee initial implant cell cultures. This enzyme is normally
strength. This absence of a clearly- integrity. The previous example draws localized inside live cells and after cell
defined standard is partly related to the the attention to the fact that there is death the enzyme leaks into the growth
previously mentioned fact that no large clear open question concerning the medium.
difference can be observed for different buffering strength of physiological
SBF when other implant materials are
tested. The main difference between
media especially in crevice or when Future Challenge:
important reaction rates are present.
the two solutions presented here is the Biocompatibility testing procedures. Electrochemical Testing in the
concentration of (HCO3) - ions, 26.2 ISO standard 10993-5 provides detailed Presence of Cells
mM in AP and 4.2 mM in SBF K9. guidelines to perform biocompatibility
It should also be noted, that SBF K9 tests. The cells can be exposed to an One of the major criticisms of
is buffered (with tris-(CH2OH) 3CNH2), extract of the materials at various the actual in vitro electrochemical
whereas AP is a non buffered solution concentrations or the cells can be characterization of medical implants is
although carbonate species show some directly grown on the material the absence of biological species in the
buffering behavior. itself. Considering the exposure simulated physiological media used.
For these investigations, EIS is a of extracts, care has to be taken in (Editors Note: See also Hiromotos
very powerful method to investigate the choice of the eulants such article in this issue.) There is on the
electrochemical processes on samples as organic solvents, e.g. dimethyl other side in biology a huge experience
that do not show high corrosion sulfoxide (DMSO) or ethanol, or the in testing biocompatibity of metallic
resistance. Figure 5a presents EIS buffer system, e.g. phosphate buffer, surfaces with cell cultures. The challenge
characterization of a Mg4Y3RE alloy bicarbonate/CO2 buffer, or organic for corrosion research will be to merge
in AP. The EIS impedance spectra buffers like HEPES. These solutions are these two fields in future investigations,
obtained indicate the presence of two used in studies performed at room this means performing electrochemical
processes: localized attack (fast charge temperature to keep the pH stable tests in media containing living
transfer measured at high frequency) as compared to the bicarbonate/CO2 cells and proteins. There are three
and slower uniform dissolution buffer system. Growth medium buffered different levels of complexity for the
(impedance value at low frequency). with HEPES can generate cytotoxic biocompatibility testing: cell cultures,

38 The Electrochemical Society Interface Summer 2008


animal models, and clinical trials. and can be obtained in a relatively 5. T. Suter and H. Bhni, Electrochim.
This order also reflects the increasing short period of time. There are also Acta, 42, 3275 (1997).
costs from in vitro to in vivo models. disadvantages, which are based on the 6. A. Hodgson, S. Mischler, B. Von
Cell lines, primary cell cultures, or nature of experimentally immortalized Rechenberg, and S. Virtanen,
organ cultures of human or animal or tumor cells as the regulation of Proceedings of the Institute of
origin are widely used and accepted adhesion proliferation or differentiation Mechanical Engineers, Part H,
to investigate biocompatibility of and gene expression may be changed as 221(H3), 291 (2007).
materials. The isolated cells or tissues compared to normal cells. This can be 7. S. Virtanen, Acta Orthopedica, 77, 695
can be kept in commercially available avoided by using primary cell and organ (2006).
and numerous chemically-defined cultures, which are time consuming 8. A. Hodgson, S. Kurz, S. Virtanen, V.
growth media. They are usually and challenging. The intact tissue is Fervel, C. Olsson, and S. Mischler,
supplemented with foetal calf serum used for organ cultures and single cells Electrochim. Acta, 49, 2167 (2004).
to allow cell survival, adhesion, and can be isolated from various tissues of 9. L. Zardiackas, M. Roach, and J.
proliferation. In vitro experiments can different species and age. Organ cultures Disegi, Proceedings from the
be grouped into cell cultures with cell have the advantage that the natural and Materials & Processes for Medical
lines and primary cells isolated from complex organization of a tissue is still Devices Conference, M. Helmus
a tissue (for a review on bone cells see intact, but the time for the cultures is and D. Medlin, Editors, ASM
Ref. 23). Cell lines are immortalized limited dependent of the original size International; Materials Park, OH
cells, which are generally isolated from and growth as the nutrition by diffusion 44073-002, August 25-27, 398- 402
tumors or otherwise experimentally is limited. Culture conditions can have (2004).
transformed cells. The results of the a great influence on proliferation and 10. L. Zardiackas, S. Williamson, M.
experiments are quite reproducible differentiation of osteoblasts.24,25 Roach, and J. Bogan, Stainless Steels
for Medical and Surgical Application,
References G. Winters and M. Nutt, Editors,
ASTM STP 1438, ASTM Inter-
national, West Conshohocken, PA,
1. A. W. E. Hodgson, Y. 107-117 (2003).
Mueller, Y. Forster, and 11. T. Oh Keun, et al., J. Biomed. Mater.
S. Virtanen, Electrochim. Res., Part B, 70, 318 (2004).
Acta, 47, 1913 (2002). 12. S. Al Ali, et al., Biomed. Mater. Eng.,
2. E. Alkhateeb and S. 15, 307 (2005).
Virtanen, J. Biomed. Mater. 13. J. L. Gilbert, C. A. Buckley, and J. J.
Res. Part A, 75A, 934 Jacobs, J. Biomed. Mater. Res., 27, 1533
(2005). (1993).
3. M. Ruzickova, H. Hilde- 14. J. S. Kawalec, S. a. Brown, J. H. Payer,
brand, and S. Virtanen, and K. Merritt, J. Biomed. Mater. Res.,
Zeitschrift fr Phys. Chem., 29, 867 (1995).
(a) 219, 1447 (2005). 15. S. A. Brown, C. A. C. Flemming, J. S.
4. T. Suter and H. Bhni, Kawalec, H. E. Placko, C. Vassaux, K.
in Analytical Methods Merritt, J. H. Payer, and M. J. Kraay,
in Corrosion Science and J. App. Biomaterials, 6, 19 (1995).
Engineering, P. Marcus and 16. L. Xu, G. Yu, E. Zhang, F. Pan, and K.
F. Mansfeld, Editors, p. Yang, J. Biomed. Mat. Res. Part A, 83A
649, CRC Press, New York (3), 703 (2007)
(2005).

(b)

(c)

Fig. 5. Electrochemical Impedance Spectroscopy characterization


of Mg alloy degradation: (a) example of Bode plots; (b) polarization
resistance evolution for SBF and AP; and (c) pH evolution as a
function of time in different media (Ref. 20). Fig. 6. Typical procedure for biocompatibility tests.
The Electrochemical Society Interface Summer 2008 39
Schmutz, et al. Versailles, France. Her current research
(continued from previous page) interests are in materials science and the
degradation of metallic biomaterials.
17. F. Witte, V. Kaese , H. Haferkamp, She may be reached at ngoc-chang.
E. Switzer, A. Meyer-Lindenberg, quach@empa.ch.
C. J. Wirth, and H. Windhagen,
Biomaterials, 26, 3557 (2005). Isabel Gerber is a research assistant in
18. F. Witte, J. Fischer, J. Nellesen, H. the Laboratory of Biological Oriented
A. Crostack, V. Kaese, A. Pisch, F. Materials in the Department of
Beckmann, and H. Windhagen, Materials at the Swiss Federal Institute of
Biomaterials, 27, 1013 (2006). Technology (ETHZ), Zurich, Switzerland.
19. M. P. Steiger, A. M. Pietak, J. Huadmai, She received her PhD in Biology from
and G. Dias, Biomaterials, 27, 1728 the Institute of Biological Sciences at
(2006). the University of Wales Aberystwyth
20. N. Quach-Vu, A. Furrer, P. J. in 1998. Her research interests are
Uggowitzer, and P. Schmutz, Comptes bone biology and cytocompability of
Rendus de Chimie, in press (2008). biomaterials. She may be reached at
21. R. Rettig and S. Virtanen, J. Biomed. isabel.gerber@mat.ethz.ch.
Mater. Res. Part A, 85A, 167 (2008).
22. L. Li, J. Gao, and Y. Wang, Surf.
Coating Tech., 185, 92 (2004).
23. M. Lieberherr, G. Cournot, and S. P.
Robins, British J. Nutr. (Supplement),
89, 59 (2003).
24. I. Gerber, I. ap Gwynn, Eur. Cell.
Mater., 2, 10 (2001).
25. I. Gerber, I. ap Gwynn, Eur. Cell.
Mater., 3, 19 (2002).
26. C. M. Bowman, E. M. Berger, E. N. T he B enefits of M embership C an B e Y ours !
Butler, K. M. Toth, and J. E. Repine, In
Vitro Cell Dev. Biol., 21, 140 (1985).
27. I. Gerber, I. ap Gwynn, M. Alini, and
Join now for exceptional discounts on all ECS publications,
T. Wallimann, Eur. Cell Mater., 10, 8 page charges, meetings, and short courses.
(2005).

About the Authors


Journal of The Electrochemical Professional Development and
Patrik S chmutz is a research group SocietyECS membership includes EducationExchange technical
leader in the Laboratory for Corrosion access to this top-quality, peer- ideas and advances at ECS's two
and Materials Integrity at the Swiss reviewed monthly publication. Each comprehensive meetings in the spring
Federal Laboratories for Materials Testing issue includes more than 70 original and fall of each year, or through the
and Research (EMPA) in Dbendorf, papers selected by a prestigious programs of 23 sections in Brazil,
Switzerland; and he is a Lecturer in the editorial board on topics covering both Canada, Europe, Japan, Korea, and the
Department of Materials at the Swiss
electrochemical and solid-state science United States.
Federal Institute of Technology (ETHZ),
in Zurich, Switzerland. His research
& technology. Papers are published as
available at ecsdl.org/JES. Discounts on Meetings and
interests are in high lateral resolution
electrochemical methods applied
PublicationsKeep aware of pertinent
in the characterization of corrosion Electrochemical and Solid-State scientific and technological advances
processes. This includes the different LettersECS's rapid-publication, through a variety of ECS publications,
aspects of passive film investigation electronic journal. Papers are published including books, meeting abstracts, and
(also ex situ with surface analytical as available at ecsdl.org/ESL. Access monograph volumes.
techniques) and localized corrosion to this peer-reviewed journal, also a
characterization. Investigation of member benefit, covers the leading Honors and Awards Program
degradation mechanisms of metallic edge in research and development in Recognize the accomplishments of your
materials used as biomedical implants all fields of interest to ECS. It is a joint peers through the Honors and Awards
has now become an important part of publication of the ECS and the IEEE Program, which includes over two
its group activities. He may be reached Electron Devices Society. dozen Society, Division, Group, and
at patrik.schmutz@empa.ch. Section awards and the distringuished
Interface, the ECS Members ECS Fellow designation.
Ngoc -C hang Q uach-Vu is a scientific MagazineECS members also receive
collaborator in the Laboratory for Interface, a quarterly publication which Career CenterIncludes an online
Corrosion and Materials Integrity features topical scientific articles, database for posting resumes as well as
at the Swiss Federal Laboratories for news about people and events, and a job bank for prospective employers to
Materials Testing and Research (EMPA) announcements of upcoming meetings. post job openings. There's a discussion
in Dbendorf (Switzerland). She forum and many services for student
received her PhD in electrochemistry members.
and materials sciences from the
Electrochemistry Department of the
Institute of Reactivity, Electrochemistry,

www.electrochem.org
and Microporosity at the University of

40 The Electrochemical Society Interface Summer 2008

También podría gustarte