Isotope-ratio mass spectrometry

Isotope-ratio mass spectrometry

Magnetic sector mass spectrometer used in isotope ratio


analysis, through thermal ionization.
Acronym IRMS
Classification mass spectrometry

Isotope-ratio mass spectrometry (IRMS) is a specialization of mass spectrometry, in which mass


spectrometric methods are used to measure the relative abundance of isotopes in a given sample.[1][2]
This technique has two different applications in the earth and environmental sciences. The analysis
of 'stable isotopes' is normally concerned with measuring isotopic variations arising from mass-
dependent isotopic fractionation in natural systems. On the other hand, radiogenic isotope
analysis[3] involves measuring the abundances of decay-products of natural radioactivity, and is used
in most long-lived radiometric dating methods.

Introduction
The isotope-ratio mass spectrometer (IRMS) allows the precise measurement of mixtures of
naturally occurring isotopes.[4] Most instruments used for precise determination of isotope ratios are
of the magnetic sector type. This type of analyzer is superior to the quadrupole type in this field of
research for two reasons. First, it can be set up for multiple-collector analysis, and second, it gives
high-quality 'peak shapes'. Both of these considerations are important for isotope-ratio analysis at
very high precision and accuracy.[3]

Schematic of an isotope-ratio mass spectrometer for measuring CO 2


The sector-type instrument designed by Alfred Nier was such an advance in mass
spectrometer design that this type of instrument is often called the 'Nier type'. In the most general
terms the instrument operates by ionizing the sample of interest, accelerating it over a potential in
the kilo-volt range, and separating the resulting stream of ions according to their mass-to-charge
ratio (m/z). Beams with lighter ions bend at a smaller radius than beams with heavier ions. The
current of each ion beam is then measured using a 'Faraday cup' or multiplier detector.
Many radiogenic isotope measurements are made by ionization of a solid source, whereas stable
isotope measurements of light elements (e.g. H, C, O) are usually made in an instrument with a gas
source. In a "multicollector" instrument, the ion collector typically has an array of Faraday cups,
which allows the simultaneous detection of multiple isotopes.[5]

Gas source mass spectrometry


Measurement of natural variations in the abundances of stable isotopes of the same element is
normally referred to as stable isotope analysis. This field is of interest because the differences in
mass between different isotopes leads to isotope fractionation, causing measurable effects on the
isotopic composition of samples, characteristic of their biological or physical history.
As a specific example, the hydrogen isotope deuterium (heavy hydrogen) is almost double the mass
of the common hydrogen isotope. Water molecules containing the common hydrogen isotope (and
the common oxygen isotope, mass 16) have a mass of 18. Water incorporating a deuterium atom
has a mass of 19, over 5% heavier. The energy to vaporise the heavy water molecule is higher than
that to vaporize the normal water so isotope fractionation occurs during the process of evaporation.
Thus a sample of sea water will exhibit a quite detectable isotopic-ratio difference when compared to
Antarctic snowfall.
Samples must be introduced to the mass spectrometer as pure gases, achieved through
combustion, gas chromatographic feeds,[6] or chemical trapping. By comparing the detected isotopic
ratios to a measured standard, an accurate determination of the isotopic make up of the sample is
obtained. For example, carbon isotope ratios are measured relative to the international standard for
C. The C standard is produced from a fossil belemnite found in the Peedee Formation, which is a
limestone formed in the Cretaceous period in South Carolina, U.S.A. The fossil is referred to as
VPDB (Vienna Pee Dee Belemnite) and has 13C:12C ratio of 0.0112372. Oxygen isotope ratios are
measured relative the standard, V-SMOW (Vienna Standard Mean Ocean Water).

Isotope-ratio mass spectrometer used to measure stable isotope ratios, with gas bench in foreground

It is critical that the sample be processed before entering the mass spectrometer so that only a
single chemical species enters at a given time. Generally, samples are combusted or pyrolyzed and
the desired gas species (usually hydrogen (H 2), nitrogen (N2), carbon dioxide (CO2), or sulfur dioxide
(SO2)) is purified by means of traps, filters, catalysts and/or chromatography.
The two most common types of IRMS instruments are continuous flow [7] and dual inlet. In dual inlet
IRMS, purified gas obtained from a sample is alternated rapidly with a standard gas (of known
isotopic composition) by means of a system of valves, so that a number of comparison
measurements are made of both gases. In continuous flow IRMS, sample preparation occurs
immediately before introduction to the IRMS, and the purified gas produced from the sample is
measured just once. The standard gas may be measured before and after the sample or after a
series of sample measurements. While continuous-flow IRMS instruments can achieve higher
sample throughput and are more convenient to use than dual inlet instruments, the yielded data is of
approximately 10-fold lower precision.

Static gas mass spectrometry


A static gas mass spectrometer is one in which a gaseous sample for analysis is fed into the source
of the instrument and then left in the source without further supply or pumping throughout the
analysis. This method can be used for 'stable isotope' analysis of light gases (as above), but it is
particularly used in the isotopic analysis of noble gases (rare or inert gases) for radiometric
dating or isotope geochemistry. Important examples are argon-argon dating and helium isotope
analysis.

Thermal ionization mass spectrometry


Several of the isotope systems involved in radiometric dating depend on IRMS using thermal
ionization of a solid sample loaded into the source of the mass spectrometer (hence thermal
ionization mass spectrometry, TIMS). These methods include rubidium-strontium dating, uranium-
lead dating, lead-lead dating and samarium-neodymium dating.
When these isotope ratios are measured by TIMS, mass-dependent fractionation occurs as species
are emitted by the hot filament. Fractionation occurs due to the excitation of the sample and
therefore must be corrected for accurate measurement of the isotope ratio. [8]
There are several advantages of the TIMS method. It has a simple design, is less expensive than
other mass spectrometers, and produces stable ion emissions. It requires a stable power supply,
and is suitable for species with a low ionization potential, such as Strontium (Sr), and Lead (Pb).
The disadvantages of this method stem from the maximum temperature achieved in thermal
ionization. The hot filament reaches a temperature of less than 2500 degrees Celsius, leading to the
inability to create atomic ions of species with a high ionization potential, such as Osmium (Os),
and Tungsten (Hf-W). Although the TIMS method can create molecular ions instead in this case,
species with high ionization potential can be analyzed more effectively with MC-ICP-MS.
Secondary-ion mass spectrometry

Schematic diagram of a SHRIMP instrument illustrating the ion beam path. After Figure 4, Williams, 1998. [9]

An alternative approach used to measure the relative abundance of radiogenic isotopes when
working with a solid surface is secondary-ion mass spectrometry (SIMS). This type of ion-microprobe
analysis normally works by focusing a primary (oxygen) ion beam on a sample in order to generate a
series of secondary positive ions that can be focused and measured based on their mass/charge
ratios.
SIMS is a common method used in U-Pb analysis, as the primary ion beam is used to bombard the
surface of a single zircon grain in order to yield a secondary beam of Pb ions. The Pb ions are
analyzed using a double focusing mass spectrometer that comprises both an electrostatic and
magnetic analyzer. This assembly allows the secondary ions to be focused based on their kinetic
energy and mass-charge ratio in order to be accurately collected using a series of Faraday cups. [10]
A major issue that arises in SIMS analysis is the generation of isobaric interference between
sputtered molecular ions and the ions of interest. This issue occurs with U-Pb dating as Pb ions have
essentially the same mass as HfO 2+.[11] In order to overcome this problem, a sensitive high-resolution
ion microprobe (SHRIMP) can be used. A SHRIMP is a double-focusing mass spectrometer that
allows for a large spatial separation between different ion masses based on its relatively large size.
For U-Pb analysis, the SHRIMP allows for the separation of Pb from other interfering molecular ions,
such as HfO2+.

Multiple collector inductively coupled plasma mass spectrometry


An MC-ICP-MS instrument is a multiple collector mass spectrometer with a plasma source. MC-ICP-
MS was developed to improve the precision achievable by ICP-MS during isotope-ratio
measurements. Conventional ICP-MS analysis uses a quadrupole analyser, which only allows
single-collector analysis. Due to the inherent instability of the plasma, this limits the precision of ICP-
MS with a quadrupole analyzer to around 1%, which is insufficient for most radiogenic isotope
systems.
Isotope-ratio analysis for radiometric dating has normally been determined by TIMS. However, some
systems (e.g. Hf-W and Lu-Hf) are difficult or impossible to analyse by TIMS, due to the high
ionization potential of the elements involved. Therefore, these methods can now be analysed using
MC-ICP-MS.
The Ar-ICP produces an ion-beam with a large inherent kinetic energy distribution, which makes the
design of the mass-spectrometer somewhat more complex than it is the case for conventional TIMS
instruments. First, different from Quadrupole ICP-MS systems, magnetic sector instruments have to
operate with a higher acceleration potential (several 1000V) in order to minimize the energy
distribution of the ion beam. Modern instruments operate at 6-10kV. The radius of deflection of an
ion within a magnetic field depends on the kinetic energy and the mass/charge ratio of the ion
(strictly, the magnet is a momentum analyzer not just a mass analyzer). Because of the large energy
distribution, ions with similar mass/charge ratio can have very different kinetic energies and will thus
experience different deflection for the same magnetic field. In practical terms one would see that ions
with the same mass/charge ratio focus at different points in space. However, in a mass-spectrometer
one wants ions with the same mass/charge ratio to focus at the same point, e.g. where the detector
is located. In order to overcome these limitations, commercial MC-ICP-MS are double-focusing
instruments. In a double-focusing mass-spectrometer ions are focused due to kinetic energy by the
ESA (electro-static-analyzer) and kinetic energy + mass/charge (momentum) in the magnetic field.
Magnet and ESA are carefully chosen to match the energy focusing properties of one another and
are arranged so that the direction of energy focusing is in opposite directions. To simplify, two
components have an energy focus term, when arranged properly, the energy term cancels out and
ions with the same mass/charge ratio focus at the same point in space. It is important to note,
double-focusing does not reduce the kinetic energy distribution and different kinetic energies are not
filtered or homogenized. Double-focusing works for single as well as multi-collector instruments. In
single collector instruments ESA and magnet can be arranged in either forward geometry (first ESA
then magnet) or reversed geometry (magnet first then ESA), as only point-to-point focusing is
required. In multi-collector instruments, only forward geometry (ESA then magnet) is possible due to
the array of detectors and the requirements of a focal plane rather than a focal point.

Accelerator mass spectrometry


Accelerator mass spectrometry

Accelerator mass spectrometer at Lawrence Livermore


National Laboratory
Acronym AMS
Classification Mass spectrometry
Analytes Organic molecules
Biomolecules
Other techniques
Related Particle accelerator

For isotopes occurring at extremely low levels, accelerator mass spectrometry (AMS) can be used.
For example, the decay rate of the radioisotope 14C is widely used to date organic materials, but this
approach is limited to relatively large samples no more than a few thousand years old. AMS
extended the range of 14C dating to about 60,000 years BP, and is about 10 6 times more sensitive
than conventional IRMS.
AMS works by accelerating negative ions through a large (mega-volt) potential, followed by charge
exchange and acceleration back to ground. During charge exchange, interfering species can be
effectively removed. In addition, the high energy of the beam allows the use of energy-loss
detectors, that can distinguish between species with the same mass/charge ratio. Together, these
processes allow the analysis of extreme isotope ratios above 10 12.

Moving wire IRMS


Moving wire IRMS is useful for analyzing Carbon-13 ratios of compounds in a solution, such as after
purification by liquid chromatography. The solution (or outflow from the chromatography) is dried
onto a nickel or stainless steel wire. After the residue is deposited on the wire, it enters a furnace
where the sample is converted to CO 2 and water by combustion. The gas stream finally enters a
capillary, is dried, ionized, and analyzed. [12] This process allows a mixture of compounds to be
purified and analyzed continuously, which can decrease the analysis time by a factor of four.
[12]
Moving wire IRMS is quite sensitive, and samples containing as little as 1 nano-mole of Carbon
can yield precise (within 1) results.[13]