Documentos de Académico
Documentos de Profesional
Documentos de Cultura
Page 1 of 97
All or parts of this publication may be reproduced without further permission, provided the
source is acknowledged.
Although every effort has been made to ensure the accuracy of the material contained in this
publication, complete accuracy cannot be guaranteed. Neither the Environmental Protection Agency
nor the author(s) accept any responsibility whatsoever for loss or damage occasioned or claimed to
have been occasioned, in part or in full, as a consequence of any person acting, or refraining from
acting, as a result of a matter contained in this publication.
Acknowledgments
This document has been prepared on behalf of the Environmental Protection Agency by:
Nicholas Kenny, SiteRIGHT Environmental (Ireland)
Dave Curtis, DRC Consultancy Services (UK)
With the assistance of :
Simon Medhurst, Smedstack Environmental (UK)
The Environmental Protection Agency (EPA) wishes to express its appreciation to the following
organisations for their contributions in various ways towards the preparation of this document:
The Environment Agency of England and Wales
The Source Testing Association.
The National Standards Authority of Ireland. .
The following EPA staff were centrally involved in the development and review of the document:
Mr Peter Webster, Dr Ian Marnane, and Mr Tony Dolan.
and with special thanks to the contribution of the following EPA staff:
Mr Michael Mc Donagh, Mr Martin OReilly, Mr Michael Owens, Ms Eileen Butler, and
Ms Edn Christie.
2
Contents
Preface
Introduction
1.1
1.2
1.3
1.4
10
1.5
12
14
2.1
15
2.2
15
16
3.1
Types of monitoring
16
3.2
Units of measurement
18
3.3
Reference quantities
18
20
4.1
21
4.2
21
4.3
Measurement Uncertainty
23
4.4
Measurement Traceability
24
25
5.1
Total particulate
25
5.2
Combustion gases
27
5.3
Inorganic gases
28
5.4
28
5.5
29
5.6
30
5.7
Formaldehyde
31
5.8
Dioxins
31
5.9
Flow rate
33
5.10
Reference quantities
33
5.11
34
The Equipment
35
6.1
35
6.2
36
6.3
Equipment calibration
37
6.4
Equipment management
37
6.5
Certification of Equipment
38
The Person
39
7.1
39
7.2
Trainee
39
7.3
Technician
39
7.4
Team leader
40
7.5
41
The Organisation
42
8.1
License requirements
42
8.2
42
8.3
43
8.4
Management requirements
43
8.5
Technical requirements
44
8.6
Proficiency testing
46
Standard Methods
47
9.1
Irish standards
47
9.2
Hierarchy of standards
47
9.3
48
9.4
49
9.5
Future standards
49
10
50
10.1
License requirements
50
10.2
50
10.3
51
10.4
51
Appendices
53
54
82
Appendix 3 Calculations
87
89
90
92
93
94
References
97
Preface
A summarised version of this Air Emissions Monitoring Guidance Note #2 (AG2) is to be published by
the end of 2007, and the published document may be ordered from the Agencys publication office,
details of which are available on the Agency website at ww.epa.ie.
The Office of Environmental Enforcement (OEE) is one of the five offices in the Environmental
Protection Agency. The OEEs functions include the regulation of activities licensed under the EPA
and WMA Acts. It is the policy of the OEE to provide information and advice via published guidance to
those it regulates to secure environmental improvements while ensuring value for money.
This Air Emission Monitoring Guidance Note #2 (AG2) is one of a series of guidance notes that the
OEE has planned on the general theme of air pollution monitoring. A forerunner to this document is
Guidance Note No. 1 Air Emissions Sampling Facilities (AG1) which describes the facilities that must
be provided for the safe and effective monitoring of emissions.
The guidance note is intended for use by all Agency staff, (e.g. licensing and enforcement staff), the
licensed operator and test houses that provide an air emissions monitoring service. By raising
awareness among Agency staff and operators of current best practice in stack testing, so the test
houses must ensure that they are providing a comparable standard of service. The Agency advises
licensees to have regard to this guidance when outsourcing their emission monitoring programme
Throughout the guidance note there are examples given of licence conditions which are typical of
those found in Irish IPPC licences. In reality, licence conditions will vary somewhat from one licence to
the next, so reference should be made to the current licence document for the site to determine the
legal obligation for monitoring.
In some existing IPPC licences methods prescribed for monitoring atmospheric emissions may differ
from the recommendations of this guidance. These instances can be dealt with on a case by case
basis and alterations to the monitoring methods may be permitted with the prior approval of the
Agency. New and revised licenses should seek to adopt the recommendations of this guidance.
This is the first air monitoring guidance issued by the Agency and represents a move aimed at
improving the overall quality of the stack emission monitoring. The programme of monitoring will
depend on the nature and complexity of the site operations. The use of best practice in atmospheric
source monitoring is an important strand in operators efforts to protect the environment and licensed
operators should ensure that the practices described in this guidance are applied to their monitoring
programmes as soon as reasonably practicable.
MCERTS
A decision by the Environment Agency (EA), the competent authority for England and Wales, that
MCERTS be a mandatory requirement of permits issued under their new IPPC permitting scheme,
prompted the UK stack testing market to seek and achieve MCERTs accreditation over a number of
years. The EPA (Agency) recognises the many merits of the MCERTS scheme and has encouraged
monitoring companies to continue to seek accreditation to MCERTS through UKAS (UK Accreditation
Service). The Agency has also encouraged the use of MCERTS certified equipment, (or equivalent)
and the personal certification by Irish field technicians through the scheme.
The Agency currently requires MCERTS for their own personnel and contractors who measure
significant air emissions, particularly dioxins, at facilities. For the purposes of consistency, efficiency
and confidence the Agency will extend the MCERTS requirement (or its equivalent) for monitoring
and laboratory personnel (INAB or UKAS accreditation) who are carrying out air monitoring and
6
analyses of licensed incinerator emissions for IPPC and waste facilities, which fall under the
European Directive for the incineration of waste (2000/76/EC), otherwise known as WID plant.
The requirement will also apply to all emission sources where dioxin monitoring is specified, including
those IPPC sites with on site incinerators, unless the operator can demonstrate an alternative which
can deliver equivalent levels of confidence. For these sites, the MCERTS requirement would apply to
a minimum of one full air monitoring campaign per annum. Any other monitoring within this twelvemonth period for parameters other than dioxins could be undertaken by non-certified personnel.
Where operators meet the MCERTS requirements the need for duplicate Agency monitoring may be
removed, and this could then be reflected in the monitoring charges.
This requirement may later be expanded to cater for the following facilities or circumstances:
Large Combustion Plant (2001/80/EC), otherwise known as LCP plant, for some or all of the
parameters that would typically be licensed (elv's) at these facilities following further
assessment and consultation with the relevant sector.
For IPPC and waste sites with atmospheric emissions that are giving rise to concern and
where there is a requirement to ensure that specified monitoring and analyses on certain
priority pollutants is carried out using MCERTS personnel.
On a case-by-case basis having regard to the nature of the emissions and the sensitivity of
the receiving environment, (e.g. using a risk based approach).
The Agency will only consider MCERTS requirements for other sectors following further
assessment and consultation with the relevant sectors.
At the time of development of this Guidance Note, a document is being prepared by CEN/TC 264
working group (WG 19) that provides for the application of ISO 17025 to periodic stack test
measurements. The document is a CEN Technical Specification entitled; CEN/TS (WI264063), Air
Quality Measurement of stationary source emissions Application of EN ISO/IEC 17025:2000 to
periodic stack measurements. The completed document will provide a basis for the accreditation of
test houses across Europe by their respective accreditation bodies. It is intended to revise this
Guidance note, if necessary, to include the requirement of the CEN technical specification.
Introduction
This guidance note (AG2) was commissioned by the Agency to provide information on the subject
of air emission monitoring in the Irish context. The monitoring of air pollution at source is
colloquially known as stack testing and it is this term that will used in this guidance.
Stack testing in Ireland prior to 1994 was confined mainly to those sites that were licensed under
the Air Pollution Act 1987. A small number of test houses provided a service to those facilities that
were required to submit self-monitoring data to the body charged with their regulation (the Local
Authority). The Irish market for commercial stack testing was relatively small in comparison to the
UK which had a history of heavy industry that was regulated originally by Her Majestys
Inspectorate of Pollution (HMIP).
The HMIP published a series of guidance notes on the subject of stack testing, these notes have
now been replaced by a series of Environment Agency (EA) (who regulate England and Wales)
publications. Another rich source of information was the United States Environmental Protection
Agency (US EPA) which published some very prescriptive standard methods for the measurement
of source emissions. These methods remain in use and can be found on the US EPA website.
While standard methods were published in France, Germany and other EU countries, many of
these were not in the English language, therefore it was the American and UK publications that
were most often used in Ireland. This situation has been changing in recent years due to the
development of European Standard methods and their mandatory status in all EU member states.
The number of Irish standard methods for the measurement of air pollutants will continue to
increase over time.
the overriding imperative that ground level pollutant levels beyond the boundary remain below the
appropriate Air Quality Standards (e.g. World Health Organisation Air Quality Guidelines).
The responsibility for monitoring rests, in the first case, with the licensed operator. This is referred
to as self-monitoring. Some of the Irish operators have in-house capability in stack testing,
although the function is more usually contracted-out. The licence document stipulates the
frequency of monitoring and often allows for the scope of the monitoring to be altered with the
approval of the Agency. Where ongoing monitoring indicates compliance with the licence limits
over a significant period the company may apply to the Agency for a reduction in their monitoring
requirements or to amend the parameters that are required to be monitored. The Agency also has
a responsibility to conduct stack tests at licensed sites. Agency data is used to assess compliance,
and can also be used to indicate whether the licensees monitoring data is reliable. All reports,
whether self-monitoring or Agency generated, are placed on the public file. Monitoring data that is
found to exceed the licence limit value may be the subject of enforcement action up to and
including prosecution.
It is important to point out that the scope of this guidance note is confined to manual stack
emission monitoring only (otherwise known as spot-check or non-continuous monitoring). It does
not cover the use of continuously operating Automated Measuring Systems (AMS) although these
systems do depend on the use of periodic stack tests for the purpose of their calibration. At the
time of publication, a separate guidance note (AG3) is been developed for the implementation of
EN14181 in Ireland, (EN14181 is a standard that deals with the Quality Assurance of AMSs).
Agency staff, both licensing personnel that formulate the limit values and monitoring
requirements and enforcement personnel that assess the adequacy and compliance of
the data.
The licensee who must either procure the services of a test house or provide an inhouse testing capability.
The commercial contractor(s) who conduct the field-work and analysis of the samples.
There are approximately 1500 individual emission points that are subject to IPPC control in Ireland.
The single regulatory system that exists in Ireland covers a very broad range of emissions and limit
values. In volumetric flow rate terms, this ranges from laboratory fume hoods as low as 50 Nm3/hr
to power plants at over 1,000,000 Nm3/hr, while emission concentration limits can range from as
little as 0.10 ng/Nm3 for Dioxins and up to 1,700 mg/Nm3 for Sulphur dioxide.
The monitoring frequency at these points can vary, but the normal range of frequency is from
monthly to biennially (every two years). While some IPPC licenses stipulate the monitoring method
to be employed, other licenses require the chosen method to be agreed with the Agency after the
licence has been issued.
Whereas the most frequent use of stack testing is for the purpose of compliance monitoring, there
are many other instances where stack testing may be required at IPPC sites, these include;
The generation of emission data as part of the licence application or licence review
process. The Agency have encountered many instances in which unreliable data has
been provided at the application phase and compliance problems arise when alternative
methods are employed post-licensing. It is particularly important that the licence
Page 9 of 97
application data is based on good monitoring practice as outlined in this note and the
use of appropriate standard methods.
The reporting of emission inventories at national and international level (e.g. European
PER, now referred to as PRTR)
Commissioning of process or abatement plant to confirm that emission levels meet the
manufacturers or regulators specification. Stack tests are an integral part of the
abatement plant test programme that is a condition of many licenses.
Supporting data from air dispersion models that are used to evaluate the impact of
atmospheric emissions.
The generation of data for use in mass balance calculations and the study of fugitive
and non-scheduled releases.
Waste Activities (Waste Management Act iii and PoE Act ii)
Emission Trading (Emissions Trading Regulations iv)
Emission of Solvent VOCs (Solvent Regulations v)
Petroleum Storage (Control of VOC emissions Regulations vi)
This guidance describes current best practice in stack emission monitoring and it establishes
criteria that should be met by those conducting stack tests. In section 9 it provides a listing of
preferred standard methods that will serve to harmonise Irish monitoring practices with other EU
member states.
Chapter 3 deals with the first principles of stack monitoring. It sets out the basic criteria for
sample collection and direct measurement of air pollutants. It provides information on the
units of measurement and standard reference conditions.
Chapter 4 discusses the monitoring plan, the where and when to monitor, the concept of
Emission Indicating Parameters and the uncertainty and traceability of the measurement.
Chapter 5 provides an overview of the methods that are used to measure the pollutants
that are commonly controlled in Irish IPPC licenses.
Chapters 6 through 10 deal, respectively, with the topics of; monitoring equipment,
monitoring personnel, the monitoring organisation, standard methods and the monitoring
report. Each of these five topics are inextricably linked with the quality of the stack emission
data, this is represented in schematic in Figure 1. They are dealt with separately so that the
reader can consult on the topic that is pertinent to their need.
2. Person
3. Organisation
4. Standard
methods
1. Equipment
5. Data Report
Figure 1: Schematic of quality emission data
List of references.
Encourage the personal and professional development of practicing source testers and
students
The STA has been working very closely with the England and Wales Environment Agency (EA)
over the last 5 years in the development of the EAs Monitoring Certification (MCERTS) scheme for
Manual Monitoring. The scheme was launched in February 2002 and is now accepted by the
industry as one of the major contributors in improving quality. The STA web link is www.s-t-a.org
1.5.3 MCERTS
MCERTS is the EA scheme that provides for the certification of equipment, persons and
organisations that are involved in the measurement of emissions.
The current schemes cover;
Emissions monitoring from chimney stacks - using accredited laboratories and certified
staff
In addition the MCERTS schemes also cover Chemical testing of soils and Monitoring of
discharges to waters.
Page 12 of 97
MCERTS was initiated in 1997 and the scheme has grown steadily in scale and importance.
MCERTS is now a mandatory requirement in IPPC permits in England and Wales.
A number of the MCERTS certified test houses based in the UK offer a monitoring service in this
country. Irish field technicians can also seek personal certification under the scheme. Training
programmes in stack testing have grown to meet the demand for the scheme among technicians in
the UK and Ireland. Further information on MCERTS can be found at mcerts.net.
1.5.4 US EPA
The US EPA was one of the earliest sources of references on stack testing methods, particularly
where there were no national standard methods. US EPA Method numbers 1 through 5 were
adopted widely for the measurement of flow, moisture and the total particulate at reference
conditions. Similarly, Method 23 (Dioxin/furan) and Method 29 (metals) also enjoyed widespread
use. With the advent of CEN standards, the use of the US EPA standards within European facilities
has been in decline.
Comparatively speaking, US EPA standard methods are very prescriptive. This makes it difficult for
a test house outside of the USA to comply fully with the method. Those that do comply however
can expect to achieve a high degree of repeatability of measurement. The US EPA web site is very
large and not easy to navigate around; however, the following link accesses the Emission
Measurement Centre directly.
Page 13 of 97
The Agency publication Air Sampling Guidance Note No. 1 (AG1) vii [which will be reviewed
in 2007] provides a more detailed discussion on the subject of Health and Safety. Those
persons with responsibility for the commissioning and the management of stack testing
programmes should visit and conduct a site review of the facility to confirm that it meets
the requirements of AG1 before planning a monitoring programme. A site risk assessment
should be conducted each day before site work commences, by a suitably qualified member
of the monitoring team.
Stack testing is an inherently hazardous occupation but it can be performed safely provided that
rigorous Heath & Safety rules are applied and adopted. Accidents, some causing fatalities, have
happened due to inappropriate Health and Safety and Risk Assessment procedures. Figure 2
shows some of the many hazards that need to be considered when carrying out stack testing. The
risks associated with these hazards can be managed through the application of appropriate control
measures along with proper staff training, the use of suitable PPE and adherence to risk
assessment methodologies.
Figure 2: Hazards associated with stack monitoring
Wind, rain, lightning,
snow and ice
Site traffic
Mechanical operations
Sunburn
Lone working
Chemical operations
Weather,
environment and
welfare
General site
hazards
Confined spaces
Tiredness
Lifting
RISK TO
HEALTH &
SAFETY
Falling
Physical hazards at
the stack
Chemical hazards in
the lab
Exposure to
substances used in
analysis and cleaning
Burns
Electricity
Chemical hazards at
the stack
Exposure to
substances used
in
monitoring tests
Compressed
gases
Radiation
Exposure to
substances from
the flue gas
It is a condition of all IPPC licenses issued by the Agency that safe and permanent access is
provided to all sampling and monitoring points. This condition has in the past and will continue to
be the focus of inspectors site visits and audits and failure to comply will result in enforcement
action.
If persons involved in stack testing of Agency licensed sites have any concerns regarding safety at
a particular site they should raise the matter with the company immediately, and if their concerns
are not adequately addressed within a reasonable time-frame, then they should contact the
Page 14 of 97
Agency for assistance. The Agency will seek to enforce the appropriate condition of the licence to
effect the necessary improvements.
The reader should refer directly to the HS&W at Work Act 2005 for full details. A comprehensive
collection of Acts, Orders, Regulations and Codes of Practice, etc. can be obtained from the Health
and Safety Legislation link on the HSA website.
The Safety Health and Welfare at Work (Work at Height) Regulations 2006 have particular
relevance for those involved in stack testing. Under the regulations an employer should ensure that
work at height is properly planned, appropriately supervised and carried out in a manner that is, so
far as is reasonably practicable, safe and without risk to health.
Page 15 of 97
This section introduces some basic principles of stack emissions monitoring, these principles apply
to all stack testing programmes regardless of the pollutant being measured or the purpose of the
measurement. It sets out the basic types of monitoring activity and the criteria that an organisation
must fulfil if it is to produce credible emission data. The section also discusses the units of
measurement and the concept of reference quantities (i.e. oxygen and moisture).
Page 16 of 97
The range of the analyser is appropriate to the purpose of the measurement. In general,
the lower the range, the more accurate the measurement. This is because accuracy is
usually expressed as a percentage of range. So, for example, it would not be
appropriate to use an analyser whose range is 1 to 100 mg/Nm3 to measure pollutant
levels in and around an emission limit value of 5 mg/Nm3.
xii
Note: The EU Waste Incineration Directive stipulates that analyser ranges should not exceed 1.5x the ELV;
viii
the Large Combustion Plant Directive stipulates 2.5x the ELV.
Calibration before and after measurement using standards that are traceable to certified
reference materials. In general, the standards used should be in line with the expected
measurement concentration [or at the very least the emission limit value (ELV)].
The analyser is free from any bias that can be caused by substances in the waste gas
other than the determinant.
emission to which the detection system is sensitive and the instrument response factor
for the substance is known, (e.g. the use of a portable FID to measure levels of toluene
in an emission).
Mass Concentration: The mass of pollutant per unit volume of waste gas emitted (e.g.
mg/Nm3).
Volume concentration: The volume of pollutant per unit volume of waste gas emitted
(e.g. ppm, ppb)
Volumetric flow rate: The volume of waste gas emitted per unit time (e.g. Nm3/hr)
Mass flow rate: The mass of pollutant emitted per unit time (e.g. kg/hr)
Page 18 of 97
3.3.1 Oxygen
The combustion of a carbon-based fuel consumes oxygen. The 21% oxygen content present in the
combustion air that is fed to a furnace will be depleted to some lower level in the exhaust gas. The
interpretation clause of IPPC licenses typically require emission data to be reported at reference
oxygen conditions that are defined according the fuel type, for example:
3.3.2 Moisture
Correction of emission data for moisture is necessary for certain types of sources. The combustion
of a carbon fuel evolves moisture (H2O) while some other emissions contain moisture by virtue of
the process or the method of abatement. The presence of moisture in a gas stream takes up space
that would otherwise be occupied by pollutant, so the pollutant concentration expressed on a dry
gas basis will always be higher than if it were expressed on a wet gas basis. The reverse is true for
volumetric flow rate or sample volume data. The interpretation clause of IPPC licences typically
requires emissions from combustion plant to be reported on a dry gas basis. The equations that
are used to correct emission data for moisture content are demonstrated in Appendix 3
Calculations. The following are a number of points that are worth noting in relation to moisture in
stack gases.
Water present as a vapour occupies 1000 times the volume of water that is present in liquid
form (droplets).
Droplets in a gas stream can make compliance assessment more complex. The droplets
can themselves contain pollutant species or a derivative of the pollutant, which is not the
subject of an emission limit value. When it is deemed necessary to sample droplets, the
sampling for pollutant determination should be isokinetic. A separate sampling run for
moisture determination only should account for the vapour phase moisture only and should
employ a straight probe (without sample nozzle) and low flowrate sampling (~1 5 l/min).
Some emission limit values are expressed on a wet gas basis. Where the monitoring
technique has dried the gas prior to measurement then the results will need to convert back
to the wet basis using the stack gas moisture result.
Page 19 of 97
The production of monitoring data follows several consecutive steps that all need to be performed
according to either standards or application specific instructions to ensure good quality results.
Those steps are:
Planning
Site measurement
Reporting results
The steps in each process are shown in Figure 3.
Measurements using
manual extractive
sampling to a standard
method.
Measurements using
portable AMS to a
standard method
Analysis of extractive
samples and
quantification of
results
Interpretation of
recorded data
Measurements of
reference conditions
e.g. flow, temperature
and oxygen
Page 21 of 97
The time period within the overall process that the sample/measurement is taken
The averaging time (duration) during which monitoring occurs
The frequency of monitoring
By its nature, self-monitoring bestows the licensee with a great deal of control over the timing of
compliance assessment measurements. The time at which the measurement is conducted should
depend on the process. A process that is known to generate steady state emissions can be
measured at any time. When the process emissions vary with time, then information will be needed
about those factors that affect the emission level so that the timing of the measurement can
properly reflect the average emission (e.g. for use in IPPC licence application) or the maximum
emission (e.g. for use in compliance assessment).
Compliance assessment requires measurement when the process is at a maximum sustainable
level and where emissions are stable (in term of concentration or mass load on the environment) or
as close to that level as is reasonably practical. The scheduling of monitoring campaigns must take
account of this requirement. The monitoring report must detail the process conditions (preferably
through the use of Emission Indicating Parameters along with the emission results.
In relation to the duration of a measurement, the interpretation clause of most IPPC licenses
issued by the Agency states that no 30 minute mean value shall exceed the emission limit value.
There are some exceptions to this rule including certain limits that derive directly from EU
Directives (e.g. dioxin monitoring). What this means for most monitoring programmes is that any
individual sample on which a single measurement value will be made should be collected over a 30
minute period. It may be necessary however to collect some pollutants over a period exceeding 30
minutes which is long enough to collect a measurable amount of pollutant (e.g. particulate, metals
and dioxin/furans). In such cases the monitoring time may require to be defined, particularly where
discharge rates are very low, in order to meet criteria necessary for robust analysis and
assessment. Where such circumstances prevail the above interpretation cannot be applied and a
pragmatic assessment of the likelihood of non-compliance is required
Direct measurements using an analyser making discrete measurements at very short time intervals
(e.g. every five seconds), should ideally cover a period of not less than 30 minutes with the highest
30 minutes average being reported. In the case of volume flow the licence states, no hourly or
daily mean value, calculated on the basis of appropriate spot readings, shall exceed the relevant
limit value. This means that the volumetric flowrate readings taken over a shorter period may be
extrapolated to hourly or daily values provided there are no significant variations in flow that would
render the extrapolation invalid. Where assessment of compliance requires averaging of multiple
monitoring periods (as in the Solvents Directive), then the integrated average value for each
monitoring period (e.g. 3 x 30 minutes) should be averaged to determine the reportable value.
This reportable value should be quoted together with the range of measurements used for its
calculation.
The reading should be the integrated value over the time period, which will be used to assess
compliance, (refer to specific licence and interpretation clause).
The IPPC licence defines the frequency of self-monitoring. When determining the frequency the
Agency seeks to balance the requirement for monitoring with the emission characteristics, risk to
the environment, practicalities of sampling and the costs. On occasion the monitoring frequency for
a pollutant which is particularly difficult or complex to measure/ analyse can be reduced by
maintaining a higher monitoring frequency of a parameter whose concentration is related to the
original but which may be more practicable or cost-effective to measure (e.g. the measurement of
Page 22 of 97
total particulate in place of its active component). Where the emission derives from a variable or
batch type process (e.g. different solvents used in pharmaceutical batch processing) then the
timing of successive monitoring campaigns (e.g. over the year) should seek to cover all emission
scenarios with priority being given to those that present greatest risk to the environment.
The philosophy behind determining timing requirements is well illustrated in Figure 2.5 of the BREF
reference document on the General Principles of Monitoring
It is noted that this is an area of complexity, which may need to be addressed further, but is outside
the scope of this current Guidance note
Page 24 of 97
The most frequently occurring pollutant emission limit values that appear in Irish IPPC licenses are,
in decreasing order, particulates, combustion gases, speciated organics (TA Luft classes), Total
Organic Carbon (as C) and acid gases such as HCl, HF. This section provides a summary
description of the techniques commonly employed for the periodic measurement of these
pollutants. In each case there are numbered bullet points that highlight the important features of
the technique, a consideration of these points may form part of the audit process described in
Appendix 2 Stack emission monitoring - Audit checklist. More detailed descriptions of the
techniques can be found among those standard methods that are listed in Appendix 1 Index of
Preferred Methods.
STACK
the duct (the measurement plane) may vary for example, due to friction at the duct wall. To
achieve a representative sample the probe nozzle must be positioned for an equal time at a
predetermined number of traverse points across the measurement plane, for further detail refer to
the standard method. Other points to note regarding total particulate measurement are:
1. The filter for removal of particulate can be positioned in-stack, Figure 5 or out of stack, as in
Figure 6. When using the latter configuration experience has shown that a significant
percentage of the particulate can drop out on the inner surface of the probe prior to the
filter, it is vital therefore to rinse the probe and collect the washing for analysis.
4
5
2
8
10
Page 26 of 97
4
5
2
7
9
1
8
11
10
2. Isokinetic sampling equipment can be modified so that gaseous pollutants are trapped in
impingers or sorbent traps that are positioned following the filter in the sample train.
3. Isokinetic sampling can only properly be accomplished by personnel with training and
experience.
4. The Irish standard method IS EN 13284 part 1 is suitable for measurement of low level
particulate up to 50mg/m3, the method was validated at around 5mg/m3 There are a
number of factors to consider when measuring very low concentration of particulate
including;
Filter material and filter loses
Filter holder design
Sample flow rate and nozzle size
Monitoring of low level particulates presents unique challenges as discussed in (Appendix 1
Index of Preferred Methods - particulates (low range)
Page 27 of 97
1. Being the products of combustion there is generally the need to measure oxygen (and
sometimes moisture) so as to correct the emission concentrations to reference conditions.
The measurement of reference quantities should be simultaneous.
2. The measurement of NOx (as NO2) is best achieved through the separate measurement of
both Nitric Oxide (NO) and Nitrogen Dioxide (NO2). Certain analysers predict the latter
based on a normal combustion scenario. If predictions are used, it should be stated in the
monitoring report.
3. Combustion sources that cycle on and off can result in frequent surges of Carbon
monoxide. These surges can make it difficult to achieve a representative measurement (for
example a 30 minute average). Start-up peaks in Carbon monoxide must, nonetheless, be
accounted for in the regulatory process.
4. A common reason for spurious data is that the combustion process may have actually
ceased. Oxygen concentration in the region of 20% to 21% is an unequivocal indicator of
such an occurrence however care should be exercised to ensure that high Oxygen
concentrations are not due to leaks within the sampling system itself.
The sum of Cadmium (as Cd) and Thallium (as Tl), and their compounds. Metals include
both gaseous, vapour and particulate phases as well as their compounds (expressed as the
metal or total as specified) [Measurement techniques U.S. EPA Method 29 or as updated
by CEN standard]
Lead [AA/ICP]
The Agency recognises that there are a number of inherent difficulties associated with the
specification, measurement and the interpretation of data on the emission of metallic substances. It
is anticipated that such anomalies will be standardised as IPPC licences are reviewed in line with
this Guidance Note.
Page 28 of 97
Metals should always be sampled isokinetically. Particulate-phase metals are collected on a filter
and vapour-phase metals are scrubbed out in impingers using aggressive absorbing solutions
(normally high concentration acid, see Figure 7).The filter fraction is digested in the laboratory and
analysed along with the impinger fraction using a suitable method such as Atomic Absorption
Spectrometry (AAS) or Inductively Coupled Plasma Spectrometry (ICP). A sample train that
excludes the impinger stage will fail to capture vaporous metal that will pass the filter. It is
important that the choice of standard method for monitoring compliance is consistent with the
parameter that is the subject of the emission limit value.
Temp.
indicator
Heated
Probe
Heated
filter box
STACK
Glass
Nozzle
S-type
Pitot
Impingers
Manometer
Vacuum line
to control
box
Page 29 of 97
Sample collection on to a sorbent tube and laboratory analysis using Gas Chromatography is the
specified monitoring method in many earlier Agency licenses. This sorbent tube method is no
longer best practice, and monitoring should be carried out using an FID method. The licence does
permit a change of methodology with the agreement of the Agency so that the methods
recommended above can be adopted.
Experience has shown that data from the tube / GC analysis method rarely matched data from the
portable FID method. The principal reason for this is that the GC detector responses of any of the
compounds eluted require to be compared and quantified against those of a reference compound
(usually Toluene). Substances that are not present in the calibration mixture cannot therefore be
robustly quantified. In some samples these may represent an appreciable proportion of the total
organic species present. Data is presented as g of Toluene equivalent rather than as C. By
contrast portable FIDs are often calibrated against Propane so it is not surprising that laboratory
and on-site measurement differ.
As with metals a variety of terms have been employed in IPPC licenses for the general limitation of
summed non-specific gaseous organic emissions. These terms include: Total VOCs; Total
Organics; Total Hydrocarbons and others. The licence revision process is likely to see a
standardisation of terminology consistent with this Guidance Note however whatever the origin of
these terms the preferred method of measurement, unless otherwise agreed with the Agency,
should be Propane calibrated portable FID.
A licensed emission point can often contain a range of organic species. In such cases the Agency
may use one of a number of different classification systems for organic substances. The most
common classification system used by the Agency is the German TA Luft regulationsix whereby
different limit values are set according to the class of organic substance rather than the specific
compound. Another classification system derives from the Solvent Regulations S.I. 543 of 2002x
These systems of classification present advantages when a plant is first being licensed because a
single limit value effects control over every substance that is in a particular class (whether they are
being used in the plant or not). Note that TA Luft is not exhaustive and contains a finite list of
organic species and the class into which they fall. Those substances not appearing on the list can
Page 30 of 97
be classified according to the specific set of rules (thus by inference, TA Luft covers every organic
substance).
Despite these advantages, problems can, and do, arise when these sources must be monitored for
compliance. A licence requirement to monitor e.g. for TA Luft class II substances would if
interpreted quite literally mean the determination of every organic that could be so classed. This is
clearly impractical so the following approach should be taken to make the task manageable.
1. Agree with the Agency a list organics that could be expected to be emitted from a particular
stack. If possible subdivide the list into groups of substances that could be expected to be
emitted from each process that is vented to the stack.
2. Where possible select a standard method from those listed in Section 9. Otherwise validate
and submit one or more methods that will achieve the reliable measurement of all possible
emission scenarios. Reliable measurement demands the full quantification of all species
present in a sample. Semi-quantitative analysis will only be permitted where it is
demonstrated to the satisfaction of the Agency that use of (e.g. Toluene) does not
underestimate the levels of any species that are present in the emission
3. The introduction of a new raw material or process that could result in the emission of a new
substance should cause a review of the agreed methods to determine their continued
suitability.
4. Test houses may find it useful to develop and validate a method that covers a suite of
organic substances most commonly found in Irish industry. This approach is used by the
Agencys air emissions laboratory based in Cork.
The above rules ensure that the sampling and analytical methods have full regard for the expected
composition of the emission. The solvent regulation limits may in some cases be more applicable
than TA Luft.
5.7 Formaldehyde
The Agency licenses a number of large timber processing facilities that have Formaldehyde among
their emissions. The Agency has required the use of a method developed and validated for use at
these type of emissions by the US National Council for Air and Stream Improvement (NCASI)xi.
NCASI is an independent research institute that focuses on environmental topics of interest to the
forest products industry.
The method involves impinger train sampling into water and analysis by Acetylacetone colorimetry.
Trapping efficiency is >90% such that 3 traps will effectively quantify all emissions. Bias due to the
presence of other substances such as Methyl Ethyl Ketone (MEK), Acetone and Acetaldehyde
have been evaluated by the Agencys Cork laboratory and shown to be negligible unless they are
present in significant quantities.
Experience has shown the need to avoid (or otherwise recover) any condensation in the sampling
line because Formaldehyde is highly hydrophilic, dissolving readily in water.
5.8 Dioxins
Dioxin is a collective term for the category of 75 Polychlorinated dibenzo-para-dioxins (PCDDs)
and 135 Polychlorinated dibenzofurans (PCDFs). They arise mainly as by-products of incomplete
combustion and from certain chemical processes. An example of the type of equipment used to
sample PCDDs and PCDFs from stack gas is shown in Figure 8.
Page 31 of 97
Page 32 of 97
A key aspect of IS EN 1948 is the use of 13C12 labelled PCDD/PCDF that is spiked onto the
sampling equipment (sampling standards), before extraction (extraction standards) and just before
GC injection (syringe standards). This practice allow the recoveries (or losses) of the congeners to
be determined. Recovery data should always be provided in the monitoring report.
5.10.1 Oxygen
The technologies for oxygen measurement are well developed because the control of oxygen in
the combustion process is critical to the overall efficiency. The following types of portable oxygen
analysers are used for the measurement of stack gases.
5.10.2 Moisture
The measurement of moisture using I.S. EN14789, is achieved by extractive sampling through a
heated line into cooled impingers containing a known volume of deionised water followed by an
impinger filled with pre-weighed silica gel. The liquid volume and silica gel weight gain are
measured post sampling to determine the increase due to moisture in the stack gas.
Page 33 of 97
Page 34 of 97
The Equipment
As mentioned in Section 1.4, there are five principal areas that influence the quality of stack
monitoring data. The first area to be discussed is equipment. Equipment in its broadest sense
covers everything from the contents a field technicians tool box (e.g. screwdriver, spanners, tube
cutter, etc) to a scanning FTIR analyser and stack sampling trains. While all make a contribution to
the monitoring exercise it is obvious that some are more important than others.
If the monitoring is to produce quality data then those items of equipment, (e.g. isokinetic sampling
trains and portable analysers), must be fit for purpose. The same applies to peripheral equipment,
(e.g. temperature and pressure sensing devices), which have a role in the control of the
sampler/analyser or in the measurement of stack conditions.
The licence template states: Monitoring and analysis equipment shall be operated and maintained
as necessary so that monitoring reflects the emission or discharge.
The following sections discuss the types of monitoring equipment, its management to ensure that it
is fit for purpose, and the subject of equipment certification.
Page 35 of 97
Isokinetic
sample trains
Gas sample
trains
Sample probe
(heated )
Determination
of ;
Particulates ,
Metals ,
Dioxins, PCB,
PAHs etc
Determination
of Gaseous
species, SO2,
HCl, HF,
moisture,
speciated
VOC etc
Heated sample
line
Peripheral equipment
Peripheral equipment
Gas conditioner
except for TOC
Analyser
Sample train
Filters
Impingers
Sample train
Impingers
Sorbent tubes
Nozzles
Conditioning
Capture
solutions
Heated probes
Capture
solutions
Sample
recoverey
Laboratory
analysis
Heated probes
Filter / Impinger
housing
Weighing
Sample
recoverey
Impinger/tube
housing
Sample
recoverey
Laboratory
analysis
Laboratory
analysis
Pitot tubes
Pitot tubes
Stack
temperature
Peripheral equipment
Data logging
Calibration
gases
Optional gas
divider
Stack
temperature
Page 36 of 97
Dates, results and copies of reports and certificates of all calibrations, adjustments,
acceptance criteria, and the due date of next calibration;
g) The maintenance plan, where appropriate, and maintenance carried out to date;
h) Any damage, malfunction, modification or repair to the equipment.
Whenever practicable, all equipment under the control of the test house and requiring calibration
should be labelled, coded or otherwise identified to indicate the status of calibration, including the
date when last calibrated and the date or expiration criteria when recalibration is due.
The test house should have procedures for safe handling, transport, storage, use and planned
maintenance of measuring equipment to ensure proper functioning and in order to prevent
contamination or deterioration. Procedures should describe the method for cleaning the equipment
both before use and between sample runs.
Page 37 of 97
When, for whatever reason, equipment goes outside the direct control of the test house, (e.g.
equipment sent for service or rental equipment brought in), the test house should ensure that the
function and calibration status of the equipment are checked and shown to be satisfactory before
the equipment is returned to service.
Page 38 of 97
The Person
With the expansion of the various environmental licensing regimes the number of Irish personnel
involved in compliance monitoring has steadily increased. Entry level staff will typically hold a third
level qualification in an environmental discipline. A number of Irish third level institutions now offer
courses that include modules in environmental monitoring but these do not provide specialist
tuition in stack testing.
This absence of stack emission monitoring education has not been unique to Ireland. Despite the
growth in demand for stack testing that has accompanied the increase in environmental regulation
throughout the EU, there has been a slowness to recognize the need for training in the area. The
training deficit was recognized in the UK with the launch in 2002 of the MCERTS Personnel
Competency Standard for Manual Stack-Emission Monitoring. Apart from the schemes obvious
contribution to the quality of monitoring data, it is important for the opportunity that it offers those
who wish to gain a formal qualification and make a career in stack testing. The number of Irish
people undertaking training in stack emission monitoring through the STA has been steadily
increasing. The Agency recognises that best practice in emission monitoring may depend on
equipment, procedures and quality systems but these are of little value without competent and
reliable personnel. A typical condition in IPPC licenses states:
The licensee shall ensure that personnel performing specifically assigned tasks shall be qualified
on the basis of appropriate education, training and experience, as required and shall be aware of
the requirements of this licence
This section is based on publications by the EA that describe their MCERTS scheme for personnel
certification xv xvi. The organizational structures and personnel competencies set out below are not
mandatory but are included for information purposes.
7.2 Trainee
Trainee denotes those persons at entry level. Trainees should undergo a formal induction
programme that includes an overview of stack-emission monitoring and the preparation of a
personal training plan. It is strongly recommended that the training plan provides for early
attendance at a training course which covers hazard identification and risk assessment relating to
stack-emission monitoring, for example, the Source Testing Association (STA) course Risk
Assessment: Industrial-Emission Monitoring or equivalent.
Thereafter, the trainee may begin supervised on-the-job training. Trainees must not conduct
stack-emission monitoring unless supervised. Trainees should not conduct site reviews or site risk
assessments under any circumstances.
7.3 Technician
Technician level denotes an intermediate competence between that of trainee and team leader. A
technician should have accumulated a degree of experience in stack testing (e.g. at least 10 site
visits involving stack emission monitoring). The technician should be capable of conducting a risk
Page 39 of 97
assessment and site review under the supervision of a team leader. The technician should display
competence in the following areas:
A basic knowledge of the principal atmospheric pollutants, their sources and impact on
the environment
Overview of Irish legislation for the control of air pollutants and guidance for monitoring
Health and safety, including the Irish legal requirements and responsibilities pertaining
to stack testing. A very useful reference source is the STA publication Risk assessment
guide: Industrial emission monitoring
A general knowledge of extractive sampling, the distinction between gas sampling and
particulate sampling and the arrangement of sample train components.
Page 40 of 97
Page 41 of 97
The Organisation
It is the monitoring organisation that must put in place the systems and work practices that ensure
personnel, equipment and standard methodologies combine effectively to produce reliable
emission data and that the data is reported in the prescribed manner. Irish licensed facilities
usually contract their monitoring to an external test house, but some facilities have constituted an
internal monitoring team. Whatever the arrangement, the monitoring organisation should have in
place a Quality Management System (QMS). The QMS should be subject to periodic external
audits by the client, the Agency or an accrediting body as appropriate. This section describes those
elements of a QMS that are particular to stack emissions monitoring and provides a basis for the
auditing process. Elements of the QMS that are dealt with in other sections of this guidance, (i.e.
Equipment, Person, Standard method and Reporting) are not considered in this section.
Such procedures shall be assessed for their suitability for the test matrix and
performance characteristics determined.
Page 42 of 97
Traceable measurement
Participation in and achievement of satisfactory results in proficiency testing and interlaboratory comparison schemes applied scope of accreditation
Page 43 of 97
Records of all original observations should be retained to meet the requirements of the
accreditation body.
Page 44 of 97
8.5.4 Testing
The organisation should hold traceable calibration materials (where available) or all aspects of the
monitoring process. The organisation should participate in proficiency testing or a similar
intercomparison activity where this is available,
8.5.11 Blanks
When monitoring is undertaken requiring the laboratory analysis of samples, sample blanks must
also be analysed and reported. As a minimum, an overall field blank should be produced for each
measurement series and at least once per day. The collection of equipment blanks prior to use of
the sampling equipment (a mandatory requirement of many CEN methods) is also strongly
recommended.
Page 45 of 97
A field blank identifies sample contamination that may arise during handling, storage,
transport and analysis.
An equipment blank acts as an additional field blank but its purpose is principally to
identify any contamination that may have arisen due to the prior use of the sampling
equipment and its impact on sample recovery.
For complex analysis such as the determination of metals it is necessary to incorporate
both blank filters and any impinger solutions to assess their contributions.
Page 46 of 97
Standard Methods
Standard reference methods are essential for the effective measurement and control of air
pollution. Such standards are developed at both European and international level. The robustness
and fitness for purpose of these standards is a function of the accumulated expertise and
experience of the people who work together in committee to produce them.
An IPPC licence issued by the Agency will typically require: Sampling and analysis of all pollutants
as well as reference measurement methods to calibrate automated measurement systems shall be
carried out in accordance with CEN-standards. If CEN standards are not available, ISO, national
or international standards, which will ensure the provision of data of an equivalent scientific quality,
shall apply.
Many licenses will require that monitoring methods are agreed with the Agency subsequent to
licence issue, in which case the licensee should select from the hierarchy of standards the method
most appropriate to their application and seek approval from their licensing inspector. Licenses
may also stipulate the monitoring method in generic terms (e.g. Isokinetic/gravimetric or sorbent
tube/GCMS). Such broad definitions reflect the difficulty of prescribing definitive test methods for
many parameters due to the limited availability of methods and the complexity of their application.
However it is likely that future licence reviews will focus on standards in line with this Guidance
Note. Nonetheless the current licence conditions permit methods and scope of monitoring,
sampling and analyses, to be amended with the agreement of the Agency This allows for some
flexibility in design and selection of suitable monitoring procedures.
Page 48 of 97
Page 49 of 97
10
The reporting of emission monitoring results involves summarising and presenting monitoring data
and related information in an effective way and in a manner that fulfils any legal reporting
requirement. The monitoring report is the culmination of site and laboratory activities and
represents the last step in the data production chain.
The name, function and signature or equivalent identification of the persons authorizing the
report.
An important aspect of compliance monitoring reports is the treatment of measurement uncertainty,
this topic has been discussed in section 4.3 and further guidance can be found in the BREF
reference document on the General Principles of Monitoring xx
10.3 Data Rounding and Treatment of results below the detection limit
The purpose of air emission monitoring is to measure the pollutant load to the atmosphere
however the success of abatement systems means that in many cases the concentration of
pollutant is extremely low. Laboratory procedures are capable of achieving measurements at very
low concentrations ( e.g. sub-parts per million). However, while instrumentation may produce a
response down to its Limit of Detection (LoD), laboratories will generally quote results to a higher
Practical Reporting Limit (PRL) or Limit of Quantitation (LoQ) based on a statistical evaluation of
the variation (standard deviation) of blanks and the measurement of standards at known
concentrations. The PRL is typically 5 -10 times the LoD or may be the lowest calibration standard
used for accredited tests. If the sample response is greater than the LoD but less than the PRL
then values will generally be reported as < n (where n is = PRL).
The use of this qualified (<) laboratory value in calculations will result in an emission measurement
value (as opposed to the lab value) which is similarly qualified and whose magnitude will depend
on several factors as the gas sample volume, impinger volume, etc.
It is acceptable to report less than results provided that the method limit of detection is stated.
The method limit of detection is that which applies to the sampling method in combination with the
analytical method and not simply that latter. The method limit of detection should be no greater
than 10% of the value of the emission limit unless otherwise agreed with the Agency.
In the process of calculating an emission result data rounding of values is commonplace. It is
advisable to calculate all interim values to the accuracy of the most sensitive measurement and to
round data to a reporting value with a significance of 1 decimal place lower than the prescribed
limit e.g. where a limit of 10 mg/Nm3 applies test results should be reported to at least the first
decimal place e.g. 6.6 mg/Nm3. Reporting data which implies an inappropriate degree of sensitivity
e.g. 6.565 mg/Nm3 should be avoided. For procedures capable of producing very accurate
measurements at exceptionally low concentrations e.g. Dioxins, it is recommended that
measurements should be rounded to one decimal place greater than that required to demonstrate
compliance with the licence limit e.g. for a licence limit of 0.1 ng/Nm3, reporting should be to the
second decimal place in this case 0.01 ng/Nm3.
Appendix 8, - Stack emission monitoring report, provides further guidance on the content of
reports.
Page 52 of 97
Appendices
Page 53 of 97
Determinand
Aldehydes
Preferred
Methods
I.S. EN 13649
(sampling)
NIOSH Manual of
analytical
methods
Scope of Method
Comments
Page 54 of 97
Determinand
Preferred
Methods
I.S. EN 13649
(sampling)
NIOSH Manual of
Analytical
Methods
Scope of Method
Comments
Colorimetric
Indicator tubes
Page 55 of 97
Determinand
Ammonia
Preferred
Methods
US EPA method
26 (sampling) as
there is no
current reference
procedure
Colorimetric
Indicator tubes
Scope of Method
Page 56 of 97
Comments
Determinand
Carbon monoxide
Preferred
Methods
I.S. EN 15058
Scope of Method
Page 57 of 97
Comments
Determinand
Dioxin/Furan
Preferred
Methods
I.S. EN 1948 Part
1, 2 and 3
Scope of Method
Page 58 of 97
Comments
Determinand
Formaldehyde (wood
product sources)
Preferred
Methods
NCASI Method
C1/WP-98.01
Scope of Method
Comments
I.S. EN 13284
Part 1
or
The stated scope of the method does not cover flow and
temperature however the measurement of these
parameters are an inherent part of the isokinetic routine
for the collection of particulate samples. The method is it
commonly used for flow and temperature measurement.
ISO9096
US EPA Method
26 & 26A
Page 59 of 97
Determinand
Preferred
Methods
Scope of Method
Comments
Sampling based
on IS EN 1911
Part1. Acid
concentration
(molarity)
determined by
potentiometric
titration
Page 60 of 97
Determinand
Hydrogen fluoride
Preferred
Methods
ISO 15713
Scope of Method
Comments
Page 61 of 97
Determinand
Hydrogen sulphide
Preferred
Methods
US EPA method
11
Scope of Method
Colorimetric
indicator tubes
Mercury
I.S. EN 13211
Comments
Page 62 of 97
Particulate
0 mg/m3 to 20 mg/m3
CxHy
0 mg/m3 to 10 mg/m3
HCI
0 mg/m3 to 50 mg/m3
HF
0 mg/m3 to 10 mg/m3
SO2
CO
NOx
CO2
0 to 15 % (vf)
H2O (g)
2 % to 25 % (vf) (actual)
O2
Temperature
60C to 140C
Determinand
Metals
Preferred
Methods
VDI 3868 part 1
(sampling). VDI
2268 parts 1-4
(analysis)
Scope of Method
Titles of standards
VDI 3868 Part 1 (1994) Determination of total emission of
metals, metalloids, and their compounds - Manual
measurement in flowing, emitted gases - Sampling system
for particulate and filter-passing matter
VDI 2268 Part 1 (1987) Chemical analysis of particulate
matter; determination of Ba, Be, Cd, Co, Cr, Cu, Ni, Pb, Sr,
V, Zn in particulate emissions by atomic spectrometric
methods
VDI 2268 Part 2 (1990) Chemical analysis of particulate
matter; determination of arsenic, antimony and selenium in
dust emissions by atomic absorption spectrometry after
separation of their volatile hydrides
VDI 2268 Part 3 (1988) Chemical analysis of particulate
matter; determination of thallium in particulate emissions by
atomic absorption spectrometry
VDI 2268 Part 4 (1990) Chemical analysis of particulate
matter; determination of arsenic, antimony and selenium in
dust emissions by graphite-furnace atomic absorption
spectrometry
Page 63 of 97
Comments
Determinand
Preferred
Methods
I.S. EN 14385
Scope of Method
Page 64 of 97
Comments
Mass concentration
range
0 to 20 mg/m
0 to 20 mg/m
HCI
0 to 20 mg/m
HF
0 to 2 mg/m
SO2
0 to 100 mg/m
CO
0 to 250 mg/m
NOx as NO2
0 to 500 mg/m
3 to 15 % (dry, actual)
H2O (ga)
10 to 35 % (actual)
O2
3 to 17 % (dry, actual)
Temperature
60 to 200 C
Determinand
Nitrogen oxides
Preferred
Methods
I.S. EN 14792
Scope of Method
Comments
I.S. EN 13725
Page 65 of 97
Determinand
Odour (continued)
Preferred
Methods
Scope of Method
Page 66 of 97
Comments
Determinand
Preferred
Methods
Scope of Method
Page 67 of 97
Comments
Determinand
Preferred
Methods
Scope of Method
Adsorption onto
dual bed ATD
tubes coupled
with adsorption
on charcoal as
backup. Analysis
by ATD / GCMS
/ Solvent
desporption as
appropriate
Page 68 of 97
Comments
Determinand
Preferred
Methods
I.S. EN 13649
Scope of Method
Comments
Page 69 of 97
Determinand
Preferred
Methods
Scope of Method
Comments
US EPA Method
18
Titles of standards
Part 2 - Gaseous emission measurement - Determination of
gaseous organic carbon in waste gases - Adsorption on silica
gel
Part 3 Gaseous emission measurement - Determination of
volatile organic compounds, especially solvents, flame
ionization detector (FID)
Part 6 Gaseous emission measurement - Choice and
application of methods of measuring total gaseous organic
carbon
Page 70 of 97
Determinand
Preferred
Methods
VDI 2457 parts 27
Scope of Method
Titles of standards
Part 2 Gaseous emission measurement - Gas
chromatographic determination of organic compounds Sampling by absorption in a solvent (2-(2methoxyethoxy)ethanol, methyldiglycol) at low temperature
Part 7 Gaseous emission measurement - Gas
chromatographic determination of acetic acid esters with
boiling points up to 90 C
Page 71 of 97
Comments
Determinand
Preferred
Methods
I.S. EN 12619
Scope of Method
Comments
I.S. EN 13526
Page 72 of 97
Determinand
Oxygen
Preferred
Methods
I.S. EN 14789
Scope of Method
Page 73 of 97
Comments
Determinand
PAH's
Preferred
Methods
ISO 11338 part 1
and 2
Scope of Method
Page 74 of 97
Comments
Determinand
Preferred
Methods
US EPA method
201A
Scope of Method
Comments
New standards are being developed by ISO in cooperation with CEN. These standards will be published in
2008 and comprises a standard for below 50 mg/m3
using impactors of different design and performance
characteristic to M201a. The second standard is for
above 50 mg/m3 and uses cyclones similar to those in
M201a.
Page 75 of 97
Determinand
Particulate
Preferred
Methods
ISO 9096
Scope of Method
I.S. EN 13284
Part 1
1)
2)
3)
4)
Page 76 of 97
Comments
Determinand
PCB's
Preferred
Methods
I.S. EN 1948 Part
1, 2 and 3
Scope of Method
Comments
Page 77 of 97
Determinand
Preferred
Methods
I.S. EN 13649
(sampling)
Scope of Method
Comments
NIOSH or OSHA
methods for
analysis
Impinger
sampling as per
IS EN 1911 Part
1 followed by
colorimetry
(Phenol Index),
HPLC, or GC
analysis of
individual
compounds
Page 78 of 97
Determinand
Sulphur dioxide
Preferred
Methods
I.S. EN 14791
Scope of Method
Comments
Validation of an alternative
method
I.S. CEN TS
14793
Page 79 of 97
Determinand
Preferred
Methods
Scope of Method
criteria.
Validation of an alternative
method (continued)
Page 80 of 97
Comments
Determinand
Water vapour
Preferred
Methods
I.S. EN 14790
Scope of Method
Page 81 of 97
Comments
Assessment
Justification
(poor/average/good)
(suggested improvement)
Page 82 of 97
Criteria
Assessment
Justification
(poor/average/good)
(suggested improvement)
Page 83 of 97
Criteria
Assessment
Justification
(poor/average/good)
(suggested improvement)
Page 84 of 97
Criteria
Assessment
Justification
(poor/average/good)
(suggested improvement)
Site review
Risk assessment
Site specific
protocol
Participation in proficiency
testing schemes
Standard methods - 9
Standard method stipulated
in licence is employed or an
Agency agreed method has
been selected from the
hierarchy that is suitable to
the application.
Where there are deviations
from the standard method,
are they documented,
technically justified and
validated.
Are method modifications
and deviations validated in
accordance with CEN/TS
14793:2004
The Monitoring report - 10
The reporting process fulfils
the requirements of the
licence, (e.g. submission
Page 85 of 97
Criteria
Assessment
Justification
(poor/average/good)
(suggested improvement)
Page 86 of 97
Appendix 3 Calculations
Calculations have been provided curtsey of the Source Testing Association www.s-t-a.org
Ft = T/273.15
Fp = 101.325/P
For concentration measurements, P and T will be the pressure and temperature at the point where the
sample volume is metered. Note that the two equations above should only be used to convert mass
concentration (e.g. mg.m-3) and not for volume concentration (e.g. ppm).
When volume flow data is corrected to for temperature and pressure use the reciprocal equations.
Page 87 of 97
Note that when using equation 8 the measured oxygen value should be expressed on a dry gas basis.
In situations where the measurement technique has provided oxygen data on a wet basis then the data
must be converted to a dry basis using equation 7.
Use the reciprocal equations when volume flow data needs to be corrected for moisture and reference
oxygen.
By comparing the velocity of the gas at the nozzle with the velocity of the stack gas at the sampling
plane the isokinetic ratio is determined.
10.
Isokinetic ratio (%) = { velocity at the nozzle x 100 } / velocity of stack gas
It is also possible to check for isokinetic sampling compliance by comparing the required sampling
flow rate to the actual sampling flow rate performed during the monitoring.
11.
Isokinetic ratio (%) = { actual sampling flow rate x 100 } / required sampling flow rate
I.S. EN 13284-1:2001 for the determination of low range concentration of dust states that, if the mean
actual isokinetic ratio during the sampling at the sampling plane differs by more than -5 to +15%, the
measurement is not valid.
-3
Note: the notation mg.m is the same as mg/m , either notation is equally acceptable to the Agency.
Page 88 of 97
Sampling location
Page 89 of 97
The names of sampling team members, their competency and specific responsibility;
For non-continuous processes, the part of the process when sampling will take place;
The process details that need to be collected over the monitoring period;
Any modifications to the technical procedure, with justifications and any resulting
changes to the uncertainties;
The sampling duration and number of samples for each measurement, including
blanks;
For manual methods, the proposed sample flow-rate, volume and minimum sampling
times;
A description of the location of the sampling plane for each release point;
For each sampling plane, a summary of the number, arrangement and orientation of
the sample line(s), and the number of sampling points per line;
Page 90 of 97
For manual methods requiring a separate chemical analysis stage, details of the
analytical method, the laboratory carrying out the analysis, chain-of-custody details,
allowable time for transit to the laboratory, storage conditions and archiving
requirements;
The report format, the person who will be writing the report and the person who will be
checking the report;
The reference number and date of the on-site health and safety risk assessment the
team carried out;
The reference number of the use of hazardous chemicals risk assessments for each
measurement method;
Any site-specific safety requirements, for example, local safety induction course,
intrinsically safe site;
Page 91 of 97
Site-specific protocol;
Site review;
Risk assessment;
Page 92 of 97
The date;
For manual methods, all relevant details of sampling, for example, dry gas meter
readings, solution volumes, pressure and temperature readings;
For manual methods, the sample details, for example, sample bottle or sorbent tube
identification label;
For instrumental methods, the zero and span gas verification results;
Page 93 of 97
Content of reports
The following is an example of the information that should be included in a monitoring report.
A stack-emission monitoring report should include the following information on each page:
The report should contain the following information in the order specified:
Report title;
Page 94 of 97
Name and address of the client organisation (if different from licence holder);
Contents
The content sheet should describe the contents of both parts of the report.
Monitoring Objectives
Monitoring Results
Emission point;
Substances to be monitored;
Emission limit value expressed in the terms and units defined in the licence;
Periodic monitoring result in the same terms as the emission limit value;
Units for the emission limit value, the periodic monitoring result and the
uncertainty;
Date of monitoring
Operating information
Whether the whole of the batch was sampled or the details of the part of the
batch sampled
(if applicable);
Page 95 of 97
The periodic monitoring results and the results obtained for the corresponding
period by the operators CEMS, (this should permit a simple and direct
comparison between the two values).
Monitoring Deviations
The substance(s) monitored, the standard method used and the reference of the
operating procedure used by the monitoring organisation; and
Diagrams showing the dimensions of the stack and the monitoring facilities;
Analysis sheets including the name of the analytical laboratory, the accreditation
for the use of the analytical method and the date of the analysis;
Uncertainty calculations.
Page 96 of 97
References
i
ii
iii
iv SI 437 of 2004 European Communities (Greenhouse Gas Emissions Trading) Regulations 2004
v S.I. No 543 of 2002 EMISSIONS OF VOLATILE ORGANIC COMPOUNDS FROM ORGANIC SOLVENTS
REGULATIONS 2002
vi
SI 374 of 1997 The EPA Act, 1992 (Control of volatile organic compound emissions resulting from the storage of petrol
and its distribution) Regulations, 1997
vii
EPA, Office of Environmental Enforcement, Guidance Note No. 1 Air Emissions Sampling Facilities.
viii
2001/80/EC, Directive on limiting emissions of certain pollutants into the air from large combustion plants.
ix
Technical Instructions on Air Quality Control - TA Luft in accordance with art. 48 of the Federal Immission Control Law
(BImSchG) dated 15 March 1974 (BGBI. I p.721). Federal Ministry for Environment, Bonn 1986, including the
amendment for Classification of Organic Substances according to section 3.1.7 TA.Luft, published in July 1997.
S.I. No 543 of 2002 EMISSIONS OF VOLATILE ORGANIC COMPOUNDS FROM ORGANIC SOLVENTS
REGULATIONS 2002
xi
NCASI Method C1/WP-98.01 "Chilled Impinger Method for use at Wood Products Mills to measure Formaldehyde,
Methanol and Phenol" 1998 [Weblink: http://www.ncasi.org/publications/toc/default.aspx?id=6Methods
Manual]
xii
Directive 2000/76/EC of the European Parliament and of the Council of 4 December 2000 on the incineration of waste
xiii
ISO 17025:2005 General requirements for the competence of testing and calibration laboratories.
xiv
Environment Agency, Manual Stack-Emission Monitoring Performance Standard for Organisations Environment
Agency, September 2003, Version 4. [Weblink: http://publications.environment-agency.gov.uk/pdf/GEHO0903BKAF-ee.pdf]
xv
UK EA MCERTS Personnel Competency Standard for Manual Stack-Emission Monitoring Environment Agency,
Version 4, April 2006. [Weblink: http://publications.environment-agency.gov.uk/pdf/GEHO0705BKAG-e-e.pdf]
xvi
UK EA MCERTS Examination Syllabuses for Manual Stack-Emission Monitoring Environment Agency April 2006
Version 4. [Weblink: http://publications.environment-agency.gov.uk/pdf/GEHO0802BKAI-e-e.pdf]
xvii
Air quality -- Evaluation of the suitability of a measurement procedure by comparison with a required measurement
uncertainty
xviii
CEN/TS 14793:2004, Air Quality Stationary source emission Interlaboratory validation procedure for an alternative
method compared to a reference method
xix
CEN/TS 14793:2004, Air Quality Stationary source emission Interlaboratory validation procedure for an alternative
method compared to a reference method
xx
Integrated Pollution Prevention and Control (IPPC) Reference Document on the General Principles of Monitoring, July
2003.
Page 97 of 97