Installation Instructions
Peripherals
Printer and VCR Configurations

10032747
10032746
08653771
05936518

Siemens AG

Siemens Medical
Ultrasound
lne
n.a.
DivisionSolutions

Print No.: US03-201.812.01.02.02


Replaces: US03-201.812.01.01.02
Part No.: n.a.
sd

2004

The reproduction, transmission or use


of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

English
Doc. Gen. Date: 07.05
n.a.

Revision / Disclaimer

1Revision / Disclaimer

Copyright
Copyright 2004 by Siemens Corporation. All rights reserved. No part of this publication
may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into
any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of
Siemens Corporation. Siemens Corporation reserves the right to change its products and
services at any time. In addition, this manual is subject to change without notice. Siemens
Corporation welcomes customer input on corrections and suggestions for improvements
to this manual.Although Siemens Corporation has attempted to ensure accuracy throughout this manual, Siemens Corporation assumes no liability for any errors or omissions, nor
for any damages resulting from the application or use of this information.
Trademarks
ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE,
Adara, Antares, G20, G40, G50, G60S, Omnia and KinetDx are trademarks of Siemens
Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000,
WS3000, and WebPro are trademarks of Siemens Corporation.
Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet
Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are
registered trademarks of Netscape Communications Corporation. Java and the JavaScript
name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered
trademark of Adobe Systems, Inc.
Other products and brand names are trademarks of their respective owners.
Disclaimer
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
License Agreement
All computer programs copyright 1990-2004 by Siemens Corporation or its suppliers. Programs are licensed under the following agreement:
Siemens or its suppliers retain(s) ownership of and title to any computer program supplied
with the equipment and to the trade secrets embodied in such computer programs. Subject to the Buyers acceptance and fulfillment of the obligations in this paragraph, Siemens
grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any
computer program supplied with the Equipment that is necessary to operate the Equipment solely on the medium in which such program is delivered for the purpose of operating the equipment in accordance with the instructions set forth in the operators manuals
supplied with the Equipment and for no other purpose whatsoever. Buyer may not
reverse-assemble, reverse-compile or otherwise reverse-engineer such computer programs nor may Buyer make a copy of such program or apply any techniques to derive the
trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms

Antares

US03-201.812.01.02.02
07.05

Page 2 of 150
sd

Siemens
Medical Solutions USA, Inc.

Revision / Disclaimer

of this license, the license granted by this paragraph shall terminate. Further, because
unauthorized use of such computer programs will leave Siemens without an adequate
remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to
restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s
suppliers of software is a direct and intended beneficiary of this end-user sublicense and
may enforce it directly against Buyer with respect to software supplied by such supplier,
and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indirect, consequential, incidental or other damages arising out of the sublicense of the computer programs supplied with the equipment.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 3 of 150
sd

Antares

4
1-

Table of Contents
0Table of Contents

1 _______ Overview _______________________________________________________ 6


Report Printers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Mitigating the Risk of Connecting a Report Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Patient Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

2 _______ Hardware Configuration _________________________________________ 14


Single Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CP770 or CP800 Color Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
UP-D23MD Color Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
UP-D55MD Color Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
P91 D or P93 D B/W Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HS-MD3000 VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SVO9500 VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

14
14
17
20
23
26
31

Double Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
B/W Printer and Color Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
B/W Printer and VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Color Printer and VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

34
34
47
62

Peripheral Shelf Mounting Hole Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95


I/O Module Peripheral Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Printer I/O Module Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
VCR I/O Module Cable Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98

3 _______ Printer Setup and Configuration __________________________________ 99


SW v2.1 and V3.0. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CP800 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CP770DW, P91DW, P93DW, UPD23MD, and UPD55MD Installation . . . . . . . . . .
HP1150, HP2500n, HP4050, HP4100, HP4200, and HP6122 Installation . . . . . . .

100
100
107
108

SW v4.0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CP800 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CP770DW, P91DW, P93DW, UPD23MD, and UPD55MD Installation . . . . . . . . . .
HP4050, HP4100 and HP4200 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP1150, HP2500n and HP6122 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

110
110
118
120
121

Reboot System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122


Print/Store Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Restore Factory Settings and Factory Printer Settings . . . . . . . . . . . . . . . . . . . . . .
Configure Routing to Print and Store Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configure Print . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configure Store . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

123
123
125
128
129

Verify Printer Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130

4 _______ VCR Setup and Configuration ___________________________________ 131

Antares

US03-201.812.01.02.02
07.05

Page 4 of 150
sd

Siemens
Medical Solutions USA, Inc.

Table of Contents

HS-MD3000 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132


Front Panel Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
SVO9500 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Front Panel Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Main Menu Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Hidden Menu Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Verify the Operation of the VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142

5________ Printer Troubleshooting _________________________________________ 143


Printer Installation Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Restoring Plug-and-Play Resources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Re-Install A Printer After Troubleshooting or Restoring Plug-and-Play. . . . . . . . . . . . . 147
Report Printers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Re-install Plug-and-Play Printers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 5 of 150
sd

Antares

6
2-

Overview

1Overview

The Peripheral Installation procedure lists the latest peripheral configurations for installation and describes how to install each configuration for the the following Antares systems:

ACUSON Antares ultrasound system


ACUSON Antarespe ultrasound system
SONOLINE Antares ultrasound system
Install an onboard or offboard single configuration or double configuration on the Antares
peripheral shelf. The peripherals consist of color printers, black and white (B/W) printers,
and VCRs.
The following table lists the onboard peripheral image printers and VCR and their software
version installation types.
Tab. 1

ONboard Peripheral Image Printers and VCRs - Software Installation Types:

Peripheral

SONOLINE Antares

ACUSON Antares

ACUSON AntaresPremier Edition

Software Versions
v2.1

v3.0

v4.0

v4.0

v4.0

P91 D

NA

NA

P93 D

CP770DW

CP800DW

X1

NA

NA

UP-D23MD

NA

UP-D55MD

NA

HS-MD3000

X2

SVO9500

NA

NA

1. CP800 is not supported on systems configured with the RM-165 PCB.


2. Valid with combined code installed.

The following table lists the offboard peripheral report printers and their software version
installation types.

Antares

US03-201.812.01.02.02
07.05

Page 6 of 150
sd

Siemens
Medical Solutions USA, Inc.

Overview
Tab. 2

7
Report Printers:

Peripheral

SONOLINE Antares

ACUSON Antares

ACUSON AntaresPremier Edition

Software Versions
v2.1

v3.0

v4.0

v4.0

v4.0

HP1150

HP2500n

HP4050

HP4100

HP4200

HP6122
DeskJet

Before installing report printers, comply with the guidellines in the following sections:

(Report Printers / p. 8)
(Mitigating the Risk of Connecting a Report Printer / p. 10)
(Patient Environment / p. 11)

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 7 of 150
sd

Antares

Overview

Report Printers

WARNING

Accessory equipment connected to the USB port must be certified


according to the respective EN and IEC standards (for example, EN
60950 and IEC 60950 for data processing equipment and EN
60601-1 and IEC 60601-1 for medical equipment). Furthermore, all
configurations shall comply with the system standard EN
60601-1-1 and IEC 60601-1-1.
Anyone who connects additional equipment to the signal
input or signal output port configures a medical system
and is therefore responsible that the system complies
with the requirements of the system standard EN
60601-1-1 and IEC 60601-1-1. Siemens can only guarantee the performance and safety of the devices listed in
the Accessories and Options chapter of the System Reference (shipped with the ultrasound imaging system). If
in doubt, consult the Siemens Uptime Service Center.

WARNING

Equipment connected to the ultrasound system and in the patient


environment must be powered from a medically isolated power
source or must be a medically isolated device.
Equipment powered from a non-isolated source can
result in chassis leakage currents exceeding safe levels.
Chassis leakage current created by an accessory or
device connected to a non-isolated outlet may add to the
chassis leakage current of the ultrasound system.

WARNING

Non-medical grade report printers cannot be used within a patient


environment.
During use of a non-medical grade report printer or when
a non-medical grade report printer is connected to the
ultrasound system, the ultrasound system cannot in any
way be in contact with a patient.

Antares

NOTE

Siemens cannot support or service any external devices connected to the USB port. Siemens cannot assume responsibility for
functionality beyond the scope of this specification.

NOTE

Siemens does not specifically recommend any particular communication or analysis software; please contact a local vendor for assistance with interpretation and analysis of the data transmitted by
the ultrasound imaging system.

US03-201.812.01.02.02
07.05

Page 8 of 150
sd

Siemens
Medical Solutions USA, Inc.

Overview

Relevant standards for some non-medical electrical equipment may have limits for enclosure leakage currents higher than required by medical standards. These higher standards
are acceptable only outside the patient environment. It is essential to reduce enclosure
leakage currents when non-medical electrical equipment is to be used within the patient
environment. Measures for reducing leakage current include use of a medically-approved
isolation transformer.
The ultrasound system has printer drivers compatible with non-medical grade report printers. Siemens does not guarantee the performance and safety of any non-medical grade
report printer. Non-medical grade report printers do not fulfill the following safety requirements:

EN 60601-1-1 and IEC 60601-1-1 (Medical Electrical Equipment, Part 1: General Requirements for Safety).

EN 60601-1-2 and IEC 60601-1-2 (Electromagnetic Compatibility of Medical Devices).


If a non-medical grade report printer is to be used with the ultrasound system, then ensure
mitigation is provided to meet all safety requirements. It is the responsibility of the user to
ensure that the ultrasound system in combination with the non-medical grade report
printer complies with safety requirements.
Always use a medically-approved isolation transformer with a non-medical grade report
printer.
Mitigate risk with regard to the Electromagnetic Compatibility of Medical Devices requirement. A non-medical grade device must be designed, manufactured, and certified to meet
the same EMC (electromagnetic compatibility) requirements as the ultrasound system, or
other means must ensure that the overall EMC requirements are met.
Tab. 3

Manufacturers of Medically Approved Isolation Transformers

Manufacturer

Web Site

Tripp Lite

www.tripplite.com

Toroid Corporation of Maryland

www.toroid.com

Dale Technology

www.daletech.com

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 9 of 150
sd

Antares

10

Overview

Mitigating the Risk of Connecting a Report Printer

To fulfill EN 60601-1-1 and IEC 60601-1-1 (Medical Electrical Equipment, Part 1: General
Requirements for Safety) requirements for non-medical peripheral equipment:
The non-medical peripheral equipment must be approved according to any other EN or
IEC standard (EN XXXXX or IEC XXXXX, e.g., equipment complying with EN 60348 and
IEC 60348, EN 60950 and IEC 60950, etc.)
The connection of non-medical peripheral equipment to the ultrasound system must
adhere to the following conditions:
1. Connect the ultrasound system to an AC power outlet within a medically used room
within the patient environment. A patient environment is defined as an area in which
medical examination, monitoring, or treatment of the patient takes place. The patient
environment is located 1.5 meters (or, in Canada and the U.S.A., 1.8 meters [6 feet])
around the patient location.
2. Connect the peripheral equipment AC power cable to a medically approved isolation
transformer.
3. Connect the medically approved isolation transformer to a main AC outlet either inside
or outside the patient environment. The peripheral device and medically approved isolation transformer combination can be located:
- Within the same room as the ultrasound system, or
- In a non-medically used room.
For additional information and other possible combinations, please refer to the Medical
Electrical Equipment Standard EN 60601-1-1 or IEC 60601-1-1, Annex BBB.7,
Scenario 3c.

NOTE

Antares

The above information is based on current EN 60601-1-1 and IEC


60601-1-1 standards, dated 2000-12. If a country"s regulatory standards for medical equipment do not correspond to EN 60601-1 and
IEC 60601-1, as well as EN 60601-1-1 and IEC 60601-1-1, then the
local requirements may differ.

US03-201.812.01.02.02
07.05

Page 10 of 150
sd

Siemens
Medical Solutions USA, Inc.

Overview

11

Patient Environment

Fig. 1:

Patient Environment (Sheet 1)

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 11 of 150
sd

Antares

12

Overview

Fig. 2:

Antares

Patient Enviroment (Sheet 2)

US03-201.812.01.02.02
07.05

Page 12 of 150
sd

Siemens
Medical Solutions USA, Inc.

Overview

13

Requirements

Time required:

Software setup and configuration: 10 minutes per device


Single hardware configuration: 30 minutes
Double hardware configuration: 60 minutes
Tools:

Pen
Exacto knife or 1/4-inch hollow punch
Phillips head screw driver
Flat head screw driver

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 13 of 150
sd

Antares

14

Hardware Configuration

3-

Single Configuration
2Hardware Configuration

Tab. 4

Single Configurations

Color Printers

B/W Printers

VCRs

(CP770 or CP800 Color


Printer / p. 14)

(P91 D or P93 D B/W


Printer / p. 23)

(HS-MD3000
VCR / p. 26)
(SVO9500 VCR / p. 31)

(UP-D23MD Color
Printer / p. 17)
(UP-D55MD Color
Printer / p. 20)

CP770 or CP800 Color Printer

1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.
4. Remove the feet and the shipment locking screw from the bottom of the printer.
a) Point the bottom side of the printer face up.
b) Remove the four screws that secure the feet to the printer and remove the feet.
c) Remove the shipment locking screw.

Fig. 3:

Antares

CP770 or CP800 Printer - Shipping screw removal

US03-201.812.01.02.02
07.05

Page 14 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

15

5. Secure the mounting bracket to the color printer.


a) Put the mounting bracket on the printer.
b) Align the mounting holes.
c) Secure the bracket to the bottom of the color printer.

Fig. 4:

CP770 or CP800 - Mounting Bracket

WARNING

Use fasteners of the correct length.


Using fasteners that are too long may cause the hardware to contact other components, leading to product
damage or a potential electrical safety hazard. Using fasteners that are too short may not secure the component
to its mount, presenting a performance and safety hazard.

6. Put the color printer assembly on the peripheral shelf and align the mounting holes of
the color printer with the mounting holes in the mat.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 15 of 150
sd

Antares

16

Hardware Configuration
7. Set the color printer dip switches as follows:

Fig. 5:

CP770 Dip Switch Settings to OFF

1-OFF, 2-OFF, 3-OFF, 4-OFF,5-OFF,


6-OFF, 7-OFF, and 8-OFF.

Fig. 6:

CP800 Dip Switch Settings

1-ON, 2-OFF, 3-OFF, 4-OFF,5-OFF,


6-OFF, 7-OFF, and 8-ON.

8. Connect the color printer cables.


a) Connect the power cable to the printer AC power inlet and to the peripheral AC power outlet.
b) Connect the USB cable to the back of the printer.

Fig. 7:

CP770 or CP800 - Cable connections

Pos. 1

USB Cable (CP800 may have a parrallel cable)

Pos. 2

Power cord

c) Route the other end of the USB cable through the peripheral shelf cable passage to
the I/O module.
9. Install the cover.
a) Put the cover over the printer.
b) Align the four captive screws in the cover with the mounting holes.
c) Tighten the four captive screws.

Antares

US03-201.812.01.02.02
07.05

Page 16 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

17

10. See the Mitsubishi User Instruction to perform the following tasks:
- Load the ink cartridge
- Load the paper tray
- Slide the paper tray into the printer
11. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral
Connections / p. 97).
12. To complete the color printer configuration, see (Printer Setup and
Configuration / p. 99).

UP-D23MD Color Printer

1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 17 of 150
sd

Antares

18

Hardware Configuration
4. Put the printer feet into the mounting bracket cut-outs.

Fig. 8:

Antares

Printer Foot Placement Locations

Pos. 1

Rear mounting plate cutout for placement of the printers rear feet

Pos. 2

Front mounting plate cutout for placement of the printers front feet

Fig. 9:

Printer Placement on Mounting Plate

Pos. 1

Cable routing passage through the mat and panel to the I/O module

US03-201.812.01.02.02
07.05

Page 18 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

19

5. Connect the printer cables.

Fig. 10: Cable Connections - AC Power and USB


Pos. 1

Antares OEM AC power outlet

Pos. 2

Printer AC power inlet

Pos. 3

Printer USB connection

a) Connect the AC power cable to the Antares peripheral AC power outlet and to the
printer AC power inlet.
b) Connect the USB cable to the back of the printer.
c) Route the other end of the USB cable through the I/O module cable routing passage.
(See (1/Fig. 9 / p. 18) for the routing passage location.)
NOTE

Do not connect the USB cable to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 19 of 150
sd

Antares

Hardware Configuration

6. Install the printer cover.

Fig. 11: Printer Cover - Installed


Pos. 1

Captive screw

Pos. 2

Captive screw locations at base of cover (not in view)

a) Put the printer cover over the printer.


b) Align the three captive screws with the three fastener holes cut into the mat.
c) Tighten the captive screws
7. See the Sony User Instruction to perform the following tasks:
- Load the ink cartridge
NOTE

Power on the system prior to loading the ink cartridge. This resets
the gears to their default location, permitting the ink cartridge to be
inserted.

- Load the paper tray


- Slide the paper tray into the printer
8. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral
Connections / p. 97).
9. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

UP-D55MD Color Printer

1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.

Antares

US03-201.812.01.02.02
07.05

Page 20 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

21

3. Secure the printer bracket to the bottom of the color printer.

Fig. 12: UP-D55M Color Printer Bracket

NOTE

In (Fig. 12 / p. 21), the flat side of the bracket is oriented on the left
side of the picture.

a) Point the bottom side of the printer face up.


b) Put the bracket on the bottom of the printer.
c) Tighten the bracket captive screw.
4. Align the mounting holes of the mounting bracket to the holes in the peripheral mat.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 21 of 150
sd

Antares

22

Hardware Configuration
5. Connect the printer cables.

Fig. 13: UP-D55M Cable Connections


Pos. 1

Power cord between the OEM AC power outlet and the printer AC power inlet

Pos. 2

USB cable

a) Connect the AC power cable to the Antares peripheral AC power outlet and to the
printer AC power inlet.
b) Connect the USB cable to the back of the printer.
c) Route the other end of the USB cable through the I/O module cable routing passage.
NOTE

Do not connect the USB cable to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.

Antares

US03-201.812.01.02.02
07.05

Page 22 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

23

6. Install the printer cover.

Fig. 14: UP-D55 M Cover - Installed


Pos. 1

Captive screw location

Pos. 2

Captive screw locations at base of cover (not in veiw)

a) Put the printer cover over the printer.


b) Align the three captive screws with the three fastener holes cut into the mat.
c) Tighten the captive screws
7. See the Sony User Instruction to perform the following tasks:
- Load the ink cartridge
- Load the paper tray
- Slide the paper tray into the printer
8. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral
Connections / p. 97).
9. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

P91 D or P93 D B/W Printer

1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable instructions that follow.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 23 of 150
sd

Antares

24

Hardware Configuration
WARNING

Use fasteners of the correct length.


Using fasteners that are too long may cause the hardware to contact other components, leading to product
damage or a potential electrical safety hazard. Using fasteners that are too short may not secure the component
to its mount, presenting a performance and safety hazard.

P91 D Instructions:

P93 D Instructions:

Put the flat side of the bracket face down

Point the bottom side of the P93 D print-

on the peripheral mat.

Align the bracket mounting holes to the


peripheral mat holes.

er up.

Put the P93 D mounting bracket on the


bottom of the P93 D printer.

Put the P91 B/W printer on top of the P91


mounting bracket.

Align the B/W printer feet against the


mounting bracket.

Fig. 16: P93 D Mounting Bracket

Secure the P93 D printer mounting


bracket to the P93 D printer.
Fig. 15: P91 D Positioned on the Bracket

Point the bottom side of the printer facing down.

Put the Printer assembly on the peripheral mat.

Antares

US03-201.812.01.02.02
07.05

Page 24 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

25

4. Set the printer B/W dip switches:


- P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF
- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

Fig. 17: P91 D and P93 D Printer Cables and Dip Switches (P91 D shown)
Pos. 1

Dip Switches

Pos. 2

USB Cable

Pos. 3

AC Power Cord

5. Connect the printer cables.


a) Connect the AC power cable to the Antares peripheral AC power outlet and to the
printer AC power inlet.
b) Connect the USB cable to the back of the printer.
c) Route the other end of the USB cable through the I/O module cable routing passage.
NOTE

Do not connect the USB cable to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 25 of 150
sd

Antares

26

Hardware Configuration
6. Install the printer cover.

Fig. 18: P91 D and P93 D Printer Cover Installed

a) Put the cover over the printer.


b) Tighten the two captive screws.
c) Make sure the back of the printer touches the printer cover stop.
7. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral
Connections / p. 97).
8. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

HS-MD3000 VCR

Fig. 19: HS-MD3000 VRC - Single Configuration

Antares

Pos. 1

VCR bracket

Pos. 2

VCR

Pos. 3

Cover

US03-201.812.01.02.02
07.05

Page 26 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

27

1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Remove the feet from the bottom of the VCR.

Fig. 20: VCR Foot Locations

a) Point the bottom side of the VCR facing up.


b) Remove the four screws that secure the feet to the VCR.
c) Remove the feet.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 27 of 150
sd

Antares

28

Hardware Configuration
4. Secure the mounting bracket to the VCR.

Fig. 21: HS-MD3000 Secured to the Mounting Bracket

WARNING

Use fasteners of the correct length.


Using fasteners that are too long may cause the hardware to contact other components, leading to product
damage or a potential electrical safety hazard. Using fasteners that are too short may not secure the component
to its mount, presenting a performance and safety hazard.

NOTE

In (Fig. 21 / p. 28), the front of the VCR is facing the left side of the
picture.

a) Put the mounting bracket on the VCR.


b) Align the mounting holes.
c) Secure the bracket to the bottom of the VCR.

Antares

US03-201.812.01.02.02
07.05

Page 28 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration
WARNING

29

Use fasteners of the correct length.


Using fasteners that are too long may cause the hardware to contact other components, leading to product
damage or a potential electrical safety hazard. Using fasteners that are too short may not secure the component
to its mount, presenting a performance and safety hazard.

5. Put the VCR assembly on the peripheral shelf by aligning the mounting holes of the
VCR with the mounting holes in the mat.

Fig. 22: HS-MD3000 Positioned on the OEM Bay

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 29 of 150
sd

Antares

30

Hardware Configuration
6. Connect the VCR power cord to the peripheral AC power outlet and to the VCR AC
power inlet.

Fig. 23: HS-MD3000 Power Cord Connection


Pos. 1

OEM power outlets

Pos. 2

VCR power inlet

7. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

Fig. 24: HS-MD3000 VCR Connections from the I/O Module


Pos. 1

Connect P128 to RS-232C

Pos. 2

Connect VIDEO IN

Pos. 3

Connect VIDEO OUT

Pos. 4

Connect AUDIO IN

Pos. 5

Connect AUDIO OUT

Pos. 6

Set the 75 VIDEO and the 75 S-VIDEO switches to ON

8. Route the cables through the I/O panel cable routing passage.

Antares

US03-201.812.01.02.02
07.05

Page 30 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration
NOTE

31

Do not connect the cables to the I/O module at this time.


Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the VCR cover.

9. Install the cover.

Fig. 25: VCR Cover Captive Screw Locations

a) Put the cover over the VCR.


b) Align the three captive screws in the cover with the mounting holes.
c) Tighten the three captive screws.
10. Complete the VCR cable connections to the I/O module, see (I/O Module Peripheral
Connections / p. 97).
11. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

SVO9500 VCR

1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Secure the mounting bracket to the peripheral shelf.
3. Put the VCR assembly on the peripheral shelf by aligning the mounting holes of the
VCR with the mounting holes in the mat.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 31 of 150
sd

Antares

32

Hardware Configuration
4. Set the VCR switches: 1-OFF, 2-OFF, 3-OFF, 4-OFF, 5- ON, 6-ON

Fig. 26: SVO9500 Dip Switch Settings

- Set both 75 ohm switches to ON.


- For PAL model only, set the voltage to 240V.
NOTE

Only the PAL model includes the 240V switch.

5. Connect the VCR power cord to the peripheral AC power outlet and to the VCR AC
power inlet.
6. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

Fig. 27: SVC 9500 VCR Connections from I/O Module


Pos. 1

Audio In (left and right)

Pos. 2

Video In (use either connector)

Pos. 3

Audio Out (left and right)

Pos. 4

Video Out

Pos. 5

Dip Switches

Pos. 6

RS-232C (connect to P128)

Pos. 7

75 (set both switches to ON)

7. Route the cables through the I/O panel cable routing passage.

Antares

US03-201.812.01.02.02
07.05

Page 32 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration
NOTE

33

Do not connect the cables to the I/O module at this time.


Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the VCR cover.

8. Install the cover.


a) Put the cover over the VCR.
b) Align and tighten the two captive screws in the cover with the mounting holes.
9. Complete the VCR cable connections to the I/O module, see (VCR I/O Module Cable
Connections / p. 98).
10. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 33 of 150
sd

Antares

34

Hardware Configuration

Double Configuration
Tab. 5

Double Configuration

B/W Printer/Color Printer

B/W Printer/VCR

Color Printer/VCR

Double Configuration

Double Configuration

Double Configuration

(P91 D or P93 D / CP770 or (P91 D or P93 D /


CP800 / p. 34)
HS-MD3000 / p. 47)

(CP770 or CP800 /
HS-MD3000 / p. 62)

(P91 D or P93 D /
UP-D23MD / p. 37)

(UP-D23MD / HS-MD
3000 / p. 75)

(P91 D or P93 D /
SVO9500 / p. 56)

(P91 D or P93 D /
UP-D55M / p. 42)

(UP-D55MD /
HS-MD3000 / p. 86)
(CP770 or CP800 /
SVO9500 / p. 67)

B/W Printer and Color Printer

P91 D or P93 D / CP770 or CP800


Installation Summary
1. Install the color printer on the peripheral shelf.
2. Install the B/W printer on the color printer.
3. Complete the cable I/O module connections.
4. Complete the printer configuration.

NOTE

When performing this upgrade on a system that already has a B/W


printer or a color printer installed, remove the printer from the peripheral shelf.
Do not disconnect the USB cable from the I/O module. The system
has previously configured the USB port for the printer connected.
Therefore, the same USB port must be used for that printer when
the USB cable is reconnected.
Label the USB cable for the correct reconnection to the printer.

1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.

Antares

US03-201.812.01.02.02
07.05

Page 34 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

35

4. Remove the feet and the shipment locking screw from the bottom of the printer.
a) Point the bottom side of the printer face up.
b) Remove the four screws that secure the feet to the printer and remove the feet.

Fig. 28: CP770 or CP800 Secured to the Mounting Bracket

c) Remove the shipment locking screw.


5. Secure the mounting bracket to the color printer.
a) Put the mounting bracket on the printer.
b) Align the mounting holes.
c) Secure the bracket to the bottom of the color printer.
WARNING

Use fasteners of the correct length.


Using fasteners that are too long may cause the hardware to contact other components, leading to product
damage or a potential electrical safety hazard. Using fasteners that are too short may not secure the component
to its mount, presenting a performance and safety hazard.

6. Put the color printer assembly on the peripheral shelf and align the mounting holes of
the color printer with the mounting holes in the mat.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 35 of 150
sd

Antares

36

Hardware Configuration
7. Set the color printer dip switches as follows:

Fig. 29: CP770 Dip Switch Settings to OFF

Fig. 30: CP800 Dip Switch Settings

1-OFF, 2-OFF, 3-OFF, 4-OFF,5-OFF,


6-OFF, 7-OFF, and 8-OFF.

1-ON, 2-OFF, 3-OFF, 4-OFF,5-OFF,


6-OFF, 7-OFF, and 8-ON.

8. Connect the color printer cables.


a) Connect the color printer AC power cable to the color printer AC power inlet and to
the lower peripheral AC power outlet.
b) Connect the B/W printer AC power cable to the upper peripheral AC power outlet.
c) Connect the color printer USB cable to the back of the color printer.
d) Route the other end of the color printer USB cable through the peripheral shelf cable passage to the I/O module.
e) Route the I/O module end of the B/W printer USB cable through the peripheral shelf
cable passage to the I/O module.
NOTE

Do not connect the USB cables to the I/O module at this time, unless the system previously had one of the printers installed on the
peripheral bay. Then, connect the previously installed printer to
the same port it was connected to before.

9. Install the color printer cover/black and white printer mounting plate assembly.
a) Route the B/W printer USB cable and power cable up through the hole in the B/W
printer mounting plate.
b) Put the printer cover over the printer.
c) Align the four captive screws with the fasener holes cut into the mat.
d) Tighten the captive screws.
10. Set the B/W printer dip switches to:
- P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF
- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF
11. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable instructions that follow.

Antares

US03-201.812.01.02.02
07.05

Page 36 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

37

12. Install the B/W printer.


a) Put the B/W printer on the bracket and mounting bracket.
b) Connect the AC power cable to the back of the B/W printer.
c) Connect the USB cable to the back of the B/W printer.
NOTE

Do not connect the USB cable to the I/O module at this time.

13. Install the B/W printer cover.


a) Put the B/W cover over the B/W printer.
b) Tighten the captive screws (of the B/W printer cover) through the mounting bracket,
to the mounting plate.
c) Make sure the back of the B/W printer touches the B/W printer cover stop.
14. See the Sony User Instruction to perform the following tasks for the color printer:
- Load the ink cartridge
NOTE

Power on the system prior to loading the ink cartridge. This resets
the gears to their default location, permitting the ink cartridge to be
inserted.

- Load the paper tray


- Slide the paper tray into the printer
15. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral
Connections / p. 97).
16. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).
P91 D or P93 D / UP-D23MD
Installation Summary
1. Install the color printer on the peripheral shelf
2. Install the B/W printer on the color printer.
3. Complete the cable I/O module connections.
4. Complete the printer configuration.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 37 of 150
sd

Antares

38

Hardware Configuration
NOTE

When performing this upgrade on a system that already has a B/W


printer or a color printer installed, remove the printer from the peripheral shelf.
Do not disconnect the USB cable from the I/O module. The system
has previously configured the USB port for the printer connected.
Therefore, the same USB port must be used for that printer when
the USB cable is reconnected.
Label the USB cable for the correct reconnection to the printer.

1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.
4. Put the color printer feet into the mounting bracket cut-outs.

Fig. 31: Printer Foot Placement Locations

Antares

Pos. 1

Rear mounting plate cutout for placement of the printers rear feet

Pos. 2

Front mounting plate cutout for placement of the printers front feet

US03-201.812.01.02.02
07.05

Page 38 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

39

5. Connect the color and B/W power cables to the Antares AC outlet.

Fig. 32: UP-D23MD Printer Placement and Cable Connections


Pos. 1

OEM Bay to I/O Module cable routing passage

Pos. 2

AC power outlet UP-D23MD

Pos. 3

AC power outlet P91 D/P93 D

Pos. 4

Power inlet UP-D23MD

Pos. 5

USB cable connection UP-D23MD

a) Connect the AC power cable to the Antares peripheral AC power outlet and to the
color printer AC power inlet.
b) Connect the B/W printer AC power cable to the Antares peripheral AC power outlet
and route the cable through the color printer top cover.
c) Connect the USB cable to the back of the color printer.
d) Route both of the USB cables through the I/O module cable routing passage.
NOTE

Siemens
Medical Solutions USA, Inc.

Do not connect the USB cable to the I/O module at this time, unless
the system previously had this printer installed.

US03-201.812.01.02.02
07.05

Page 39 of 150
sd

Antares

40

Hardware Configuration
6. Install the printer cover.

Fig. 33: UP-D23MD Printer Placement and Cable Connections


Pos. 1

Mounting plate B/W printer

Pos. 2

Cable routing holes

Pos. 3

Captive screw locations

a) Put the printer cover over the printer.


b) Align the three captive screws with the three fastener holes cut into the mat.
c) Tighten the captive screws
NOTE

Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.

7. Set the B/W printer dip switches to:


- P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF
- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF
8. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable instructions that follow.
WARNING

Use fasteners of the correct length.


Using fasteners that are too long may cause the hardware to contact other components, leading to product
damage or a potential electrical safety hazard. Using fasteners that are too short may not secure the component
to its mount, presenting a performance and safety hazard.

Antares

US03-201.812.01.02.02
07.05

Page 40 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

41

P91 D Instructions:

P93 D Instructions:

Put the flat side of the bracket face down

Point the bottom side of the P93 D print-

on the B/W printer mounting plate of the


color printer cover.

Align the bracket mounting holes to the

er up.

Put the P93 D mounting bracket on the


bottom of the P93 D printer.

plate mounting holes.

Put the P91 B/W printer on top of the P91


mounting bracket.

Align the B/W printer feet against the


mounting bracket.

Fig. 35: P93 D Mounting Bracket

Secure the P93 D printer mounting


bracket to the P93 D printer.

Point the bottom side of the printer facFig. 34: P91 D Positioned on the Bracket

ing down.

Put the Printer assembly on the color


printer mounting plate.

9. Install the B/W printer.


a) Put the B/W printer on the bracket and mounting bracket.
b) Connect the AC power cable to the back of the B/W printer.
c) Connect the USB cable to the back of the B/W printer.
NOTE

Do not connect the USB cable to the I/O module at this time.

10. Install the B/W printer cover.


a) Put the B/W cover over the B/W printer.
b) Tighten the captive screws (of the B/W printer cover) through the mounting bracket,
to the mounting plate.
c) Make sure the back of the B/W printer touches the B/W printer cover stop.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 41 of 150
sd

Antares

42

Hardware Configuration
11. See the Sony User Instruction to perform the following tasks for the color printer:
- Load the ink cartridge
NOTE

Power on the system prior to loading the ink cartridge. This resets
the gears to their default location, permitting the ink cartridge to be
inserted.

- Load the paper tray


- Slide the paper tray into the printer
12. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral
Connections / p. 97).
13. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).
P91 D or P93 D / UP-D55M
Installation Summary
1. Install the color printer on the peripheral shelf
2. Install the B/W printer on the color printer.
3. Complete the cable I/O module connections.
4. Complete the printer configuration.

NOTE

When performing this upgrade on a system that already has a B/W


printer or a color printer installed, remove the printer from the peripheral shelf.
Do not disconnect the USB cable from the I/O module. The system
has previously configured the USB port for the printer connected.
Therefore, the same USB port must be used for that printer when
the USB cable is reconnected.
Label the USB cable for the correct reconnection to the printer.

1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.

Antares

US03-201.812.01.02.02
07.05

Page 42 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

43

3. Tighten the printer bracket to the bottom of the color printer.

Fig. 36: UP-D55M Color Printer Bracket

a) Point the bottom side of the printer face up.


b) Put the bracket on the bottom of the printer.
c) Tighten the bracket captive screw.
4. Connect the printer cables.

Fig. 37: UP-D55M Cable Connections


Pos. 1

Power cord between the OEM AC power outlet and the printer AC power inlet

Pos. 2

USB cable

a) Connect the AC power cable to the Antares peripheral AC power outlet and to the
printer AC power inlet.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 43 of 150
sd

Antares

44

Hardware Configuration
b) Connect the USB cable to the back of the printer.
c) Route the other end of the USB cable through the I/O module cable routing passage.
NOTE

Do not connect the USB cable to the I/O module at this time, unless
the system previously had this printer connected.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.

5. Connect the B/W printer AC power cable into the Antares peripheral AC power outlet.
Route the other end of the power cable through the color printer cover.
6. Route the B/W printer USB cable through the I/O panel cable routing passage.
NOTE

Do not connect the USB cable to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.

7. Install the printer cover.

Fig. 38: UP-D55MD Printer Cover with B/W Mounting Plate

a) Put the printer cover over the printer.


b) Align the three captive screws with the three fastener holes cut into the mat.
c) Tighten the captive screws
NOTE

Antares

Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.

US03-201.812.01.02.02
07.05

Page 44 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

45

8. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable instructions that follow.
WARNING

Use fasteners of the correct length.


Using fasteners that are too long may cause the hardware to contact other components, leading to product
damage or a potential electrical safety hazard. Using fasteners that are too short may not secure the component
to its mount, presenting a performance and safety hazard.

P91 D Instructions:

P93 D Instructions:

Put the flat side of the bracket face down

Point the bottom side of the P93 D print-

on the B/W printer mounting plate of the


color printer cover.

Align the bracket mounting holes to the

er up.

Put the P93 D mounting bracket on the


bottom of the P93 D printer.

plate mounting holes.

Put the P91 B/W printer on top of the P91


mounting bracket.

Align the B/W printer feet against the


mounting bracket.

Fig. 40: P93 D Mounting Bracket

Secure the P93 D printer mounting


bracket to the P93 D printer.

Point the bottom side of the printer facFig. 39: P91 D Positioned on the Bracket

ing down.

Put the Printer assembly on the color


printer mounting plate.

9. Set the B/W printer dip switches to:


- P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF
- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 45 of 150
sd

Antares

46

Hardware Configuration
10. Connect the B/W printer cables.

Fig. 41: Back of Black and White Printer


Pos. 1

Dip switch

Pos. 2

USB cable

Pos. 3

Power cord

a) Connect the AC power cable to the back of the P91 D/P93 D printer.
b) Connect the USB cable to the back of the P91 D/P93 D printer.
NOTE

Do not connect the USB cable to the I/O module at this time.

11. Install the B/W printer cover.

Fig. 42: UP-D55M Color Printer and P91 D or P93 D B/W Printer - Double Configuration

a) Put the B/W cover over the P91 D/P93 D printer.


b) Tighten the captive screws (of the B/W printer cover) through the mounting bracket,

Antares

US03-201.812.01.02.02
07.05

Page 46 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

47

to the mounting plate.


c) Make sure the back of the P91 D/P93 D printer touches the B/W printer cover stop.
12. See the Sony User Instruction to perform the following tasks:
- Load the ink cartridge
- Load the paper tray
- Slide the paper tray into the printer
13. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral
Connections / p. 97).
14. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

B/W Printer and VCR

P91 D or P93 D / HS-MD3000


Installation Summary
1. Install the VCR on the peripheral shelf
2. Install the B/W printer on the VCR.
3. Complete the cable I/O module connections.
4. Complete the VCR configuration.

NOTE

When performing this upgrade on a system that already has a B/W


printer installed, remove the B/W printer from the peripheral shelf.
Do not disconnect the USB cable from the I/O module. The system
has previously configured the USB port for the B/W printer. Therefore, the same USB port must be used for the B/W printer when the
USB cable is reconnected.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 47 of 150
sd

Antares

48

Hardware Configuration

Fig. 43: HS-MD3000 VRC/P91 Dor P93 D B/W Printer - Double Configuration
Pos. 1

VCR bracket

Pos. 2

VCR

Pos. 3

VCR Cover with B/W printer plate

Pos. 4

B/W printer with bracket attached to the bottom of the printer (Bracket not shown)

Pos. 5

B/W printer cover

1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.

Antares

US03-201.812.01.02.02
07.05

Page 48 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

49

4. Remove the feet from the bottom of the VCR.

Fig. 44: VCR Foot Locations

a) Point the bottom side of the VCR facing up.


b) Remove the four screws that secure the feet to the VCR.
c) Remove the feet.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 49 of 150
sd

Antares

50

Hardware Configuration
5. Secure the mounting bracket to the VCR.

Fig. 45: HS-MD3000 Secured to the Mounting Bracket

a) Put the mounting bracket on the VCR.


b) Align the mounting holes.
c) Secure the bracket to the bottom of the VCR.
WARNING

Use fasteners of the correct length.


Using fasteners that are too long may cause the hardware to contact other components, leading to product
damage or a potential electrical safety hazard. Using fasteners that are too short may not secure the component
to its mount, presenting a performance and safety hazard.

Antares

US03-201.812.01.02.02
07.05

Page 50 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

51

6. Put the VCR assembly on the peripheral shelf and align the mounting holes of the VCR
bracket with the mounting holes in the mat.

Fig. 46: HS-MD3000 Positioned on the OEM Bay

7. Connect the power cables to the peripheral AC power outlet.

Fig. 47: HS-MD3000 Power Cord Connection


Pos. 1

OEM power outlets

Pos. 2

VCR power inlet

a) Connect the power cable to the VCR AC power inlet and to the lower peripheral AC
power outlet.
b) Connect the printer power cable to the upper peripheral AC outlet.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 51 of 150
sd

Antares

52

Hardware Configuration
8. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

Fig. 48: HS-MD3000 VCR Connections from the I/O Module


Pos. 1

Connect P128 to RS-232C

Pos. 2

Connect VIDEO IN

Pos. 3

Connect VIDEO OUT

Pos. 4

Connect AUDIO IN

Pos. 5

Connect AUDIO OUT

Pos. 6

Set the 75 VIDEO and the 75 S-VIDEO switches to ON

NOTE

Do not connect the cables to the I/O module at this time.


Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the VCR cover.

9. Route the B/W printer power cord and USB cable through the cable channel in the VCR
cover.

Fig. 49: Cable Channels

Antares

Pos. 1

VCR cover to black and white printer

Pos. 2

To I/O module

US03-201.812.01.02.02
07.05

Page 52 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

53

10. Route the other end of the VCR cables through the I/O module cable channel.
11. Route the other end of the B/W printer USB cable through the I/O module cable channel.
12. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable instructions that follow.
WARNING

Use fasteners of the correct length.


Using fasteners that are too long may cause the hardware to contact other components, leading to product
damage or a potential electrical safety hazard. Using fasteners that are too short may not secure the component
to its mount, presenting a performance and safety hazard.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 53 of 150
sd

Antares

54

Hardware Configuration
P91 D Instructions:

P93 D Instructions:

Put the flat side of the bracket face down

Point the bottom side of the P93 D print-

on the B/W printer mounting plate of the


VCR cover.

Align the bracket mounting holes to the

er up.

Put the P93 D mounting bracket on the


bottom of the P93 D printer.

plate mounting holes.

Put the P91 B/W printer on top of the P91


mounting bracket.

Align the B/W printer feet against the


mounting bracket.

Fig. 51: P93 D Mounting Bracket

Secure the P93 D printer mounting


bracket to the P93 D printer.

Point the bottom side of the printer facFig. 50: P91 D Positioned on the Bracket

ing down.

Put the Printer assembly on the VCR


mounting plate.

Antares

US03-201.812.01.02.02
07.05

Page 54 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

55

Fig. 52: VCR Cover with Black and White Mounting Plate
Pos. 1

VCR Cover

Pos. 2

Black and white printer mounting plate

13. Set the B/W printer dip switches:


- P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF
- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

Fig. 53: Back of Black and White Printer


Pos. 1

Dip switch

Pos. 2

USB cable

Pos. 3

Power cord

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 55 of 150
sd

Antares

56

Hardware Configuration
14. Install the B/W printer cover.

Fig. 54: Black and White Printer and VCR Installed

a) Put the printer cover on the B/W printer.


b) Tighten the two printer cover captive screws.
15. Complete the VCR and printer cable connections to the I/O module, see (I/O Module
Peripheral Connections / p. 97).
16. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).
17. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).
P91 D or P93 D / SVO9500
Installation Summary
1. Install the VCR on the peripheral shelf
2. Install the B/W printer on the VCR.
3. Complete the cable I/O module connections.
4. Complete the VCR configuration.

NOTE

When performing this upgrade on a system that already has a B/W


printer installed, remove the B/W printer from the peripheral shelf.
Do not disconnect the USB cable from the I/O module. The system
has previously configured the USB port for the B/W printer. Therefore, the same USB port must be used for the B/W printer when the
USB cable is reconnected.

Antares

US03-201.812.01.02.02
07.05

Page 56 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

57

1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Secure the mounting bracket to the peripheral shelf.
3. Put the VCR assembly on the peripheral shelf by aligning the mounting holes of the
VCR with the mounting holes in the mat.
4. Set the VCR switches:
- Set the dip switches: 1-OFF, 2-OFF, 3-OFF, 4-OFF, 5- ON, 6-ON.

Fig. 55: SVO9500 Dip Switch Settings

- Set both 75 ohm switches to ON.


- For PAL model only, set the voltage to 240V.
NOTE

Only the PAL model includes the 240V switch.

5. Connect the VCR power cord to the peripheral AC power outlet and to the VCR AC
power inlet.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 57 of 150
sd

Antares

58

Hardware Configuration
6. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

Fig. 56: SVC 9500 VCR Connections from I/O Module


Pos. 1

Audio In (left and right)

Pos. 2

Video In (use either connector)

Pos. 3

Audio Out (left and right)

Pos. 4

Video Out

Pos. 5

Dip Switches

Pos. 6

RS-232C (connect to P128)

Pos. 7

75 (set both switches to ON)

7. Route the cables through the I/O panel cable routing passage.
NOTE

Do not connect the cables to the I/O module at this time.


Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the VCR cover.

8. Install the cover.


a) Route the B/W printer power cord and USB cable through the cable channel in the
VCR cover.
b) Route the other end of the VCR cables through the I/O module cable channel.
c) Route the other end of the B/W printer USB cable through the I/O module cable

Antares

US03-201.812.01.02.02
07.05

Page 58 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

59

channel.
d) Put the cover over the VCR.

Fig. 57: Cover Installation - Cables Routed Through the Cover

e) Align and tighten the two captive screws in the cover with the mounting holes.

Fig. 58: Cover Captive Screw

9. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable instructions that follow.
WARNING

Use fasteners of the correct length.


Using fasteners that are too long may cause the hardware to contact other components, leading to product
damage or a potential electrical safety hazard. Using fasteners that are too short may not secure the component
to its mount, presenting a performance and safety hazard.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 59 of 150
sd

Antares

60

Hardware Configuration
P91 D Instructions:

P93 D Instructions:

Put the flat side of the bracket face down

Point the bottom side of the P93 D print-

on the B/W printer mounting plate of the


VCR cover.

Align the bracket mounting holes to the

er up.

Put the P93 D mounting bracket on the


bottom of the P93 D printer.

plate mounting holes.

Put the P91 B/W printer on top of the P91


mounting bracket.

Align the B/W printer feet against the


mounting bracket.

Fig. 60: P93 D Mounting Bracket

Secure the P93 D printer mounting


bracket to the P93 D printer.

Point the bottom side of the printer facFig. 59: P91 D Positioned on the Bracket

ing down.

Put the Printer assembly on the VCR


mounting plate.

Antares

US03-201.812.01.02.02
07.05

Page 60 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

61

10. Set the B/W printer dip switches:


- P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF
- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

Fig. 61: Back of Black and White Printer


Pos. 1

Dip switch

Pos. 2

USB cable

Pos. 3

Power cord

11. Install the B/W printer cover.


a) Put the printer cover on the B/W printer.

Fig. 62: B/W Printer Cover Installation

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 61 of 150
sd

Antares

62

Hardware Configuration
b) Tighten the two printer cover captive screws.

Fig. 63: B/W and VCR Covers Installed

12. Complete the VCR and printer cable connections to the I/O module, see (I/O Module
Peripheral Connections / p. 97).
13. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).
14. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

Color Printer and VCR

CP770 or CP800 / HS-MD3000


Installation Summary
1. Adjust the height of the monitor/control panel.
2. Install the color printer on the peripheral shelf
3. Install the VCR printer on the color printer.
4. Complete the cable I/O module connections.
5. Complete the VCR configuration.

NOTE

When performing this upgrade on a system that already has a color printer installed, remove the color printer from the peripheral
shelf.
Do not disconnect the USB cable from the I/O module. The system
has previously configured the USB port for the color printer.
Therefore, the same USB port must be used for the color printer
when the USB cable is reconnected.

Antares

US03-201.812.01.02.02
07.05

Page 62 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

63

1. Adjust the height of the monitor/control panel.


Press the height adjustment knob to set the monitor/control panel to the highest
position.
2. Remove the upper, right-front panel.
a) Loosen the 1/4-turn captive screw.

Fig. 64: Locking Bolt in the Storage Loction

b) Grip the transducer port insets of the panel and pull to disengage the ball snaps
3. Remove the left-front panel.
a) Loosen the 1/4-turn captive screw.

Fig. 65: Upper, Left-Front Panel Captive Screw

b) Grip the air flow vent insets of the panel and pull the panel to disengage the ball
snaps.
4. Remove the bottom post cover.
a) Loosen the four screws located on either side of the post.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 63 of 150
sd

Antares

64

Hardware Configuration
b) Remove the post cover.
5. Install the locking bolt.
a) Remove the locking bolt from the locking bolt storage location.
b) Tighten the locking bolt to the post to lock the height adjustment on the highest position
6. Attach the bottom post cover.
a) Place the cover with the painted, chamfered side facing up.
b) Tighten the four post cover screw that secure the cover to the system. Locate two
screws on either side of the post.
7. Install the right-front panel.
a) Align the panel ball snap receptacles with the system ball snaps and press the panel to engage the ball snaps.
b) Tighten the 1/4-turn captive screw.
8. Install the left-front panel.
a) Align the panel ball snap receptacles with the system ball snaps and press the panel to engage the ball snaps.
b) Tighten the 1/4-turn captive screw.
9. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
10. Return the mat to the peripheral shelf.
11. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.
12. Remove the feet and the shipment locking screw from the bottom of the printer.
a) Point the bottom side of the printer face up.
b) Remove the four screws that secure the feet to the printer and remove the feet.

Fig. 66: UP-D23M Color Printer Connections

Antares

US03-201.812.01.02.02
07.05

Page 64 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

65

c) Remove the shipment locking screw.


13. Secure the mounting bracket to the color printer.
a) Put the mounting bracket on the printer.
b) Align the mounting holes.
c) Secure the bracket to the bottom of the color printer.
WARNING

Use fasteners of the correct length.


Using fasteners that are too long may cause the hardware to contact other components, leading to product
damage or a potential electrical safety hazard. Using fasteners that are too short may not secure the component
to its mount, presenting a performance and safety hazard.

14. Put the color printer assembly on the peripheral shelf and align the mounting holes of
the color printer with the mounting holes in the mat.
15. Set the color printer dip switches as follows:

Fig. 67: CP770 Dip Switch Settings to OFF

Fig. 68: CP800 Dip Switch Settings

1-OFF, 2-OFF, 3-OFF, 4-OFF,5-OFF,


6-OFF, 7-OFF, and 8-OFF.

1-ON, 2-OFF, 3-OFF, 4-OFF,5-OFF,


6-OFF, 7-OFF, and 8-ON.

16. Connect the color printer cables.


a) Connect the color printer AC power cable to the color printer AC power inlet and to
the lower peripheral AC power outlet.
b) Connect the VCR AC power cable to the upper peripheral AC power outlet.
c) Connect the color printer USB cable to the back of the color printer.
d) Route the other end of the color printer USB cable through the peripheral shelf cable passage to the I/O module.
e) Route the I/O module end of the VCR USB cable through the peripheral shelf cable
passage to the I/O module.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 65 of 150
sd

Antares

66

Hardware Configuration
NOTE

Do not connect the USB cables to the I/O module at this time, unless the system previously had one of the printers installed on the
peripheral bay. Then, connect the previously installed printer to
the same port it was connected to before.

17. Remove the feet and the shipment locking screw from the bottom of the VCR.
a) Point the bottom side of the VCR facing up.
b) Remove the four screws that secure the feet to the VCR.
c) Remove the feet.
18. Secure the mounting bracket to the VCR.
a) Put the mounting bracket on the VCR.
b) Align the mounting holes.
c) Secure the bracket to the bottom of the VCR.
WARNING

Use fasteners of the correct length.


Using fasteners that are too long may cause the hardware to contact other components, leading to product
damage or a potential electrical safety hazard. Using fasteners that are too short may not secure the component
to its mount, presenting a performance and safety hazard.

19. Put the VCR assembly on the color printer.


20. Connect the VCR cables.
a)
b)
c)
d)
e)
f)
g)
h)

Connect the power cable to the VCR and to the Antares peripheral AC power outlet.
Connect the AUDIO IN left and right channel cables to the back of the VCR.
Connect the AUDIO OUT left and right channel cables to the back of the VCR.
Set the VIDEO and S VIDEO switches to the on position.
Connect the VIDEO in/out cables to the back of the VCR.
Connect the S VIDEO in/out cables to the back of the VCR.
Connect the USB cable to the back of the VCR.
Connect the RS232 C cable to the back of the VCR.

21. Install the cover.


a) Put the cover over the VCR.
b) Align the four captive screws in the cover with the mounting holes.
c) Tighten the four captive screws.
22. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).
23. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

Antares

US03-201.812.01.02.02
07.05

Page 66 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

67

24. Complete the VCR and printer cable connections to the I/O module, see (VCR I/O Module Cable Connections / p. 98).
CP770 or CP800 / SVO9500
Installation Summary
1. Adjust the height of the monitor/control panel.
2. Install the color printer on the peripheral shelf
3. Install the VCR printer on the color printer.
4. Complete the cable I/O module connections.
5. Complete the VCR configuration.

NOTE

When performing this upgrade on a system that already has a color printer installed, remove the color printer from the peripheral
shelf.
Do not disconnect the USB cable from the I/O module. The system
has previously configured the USB port for the color printer.
Therefore, the same USB port must be used for the color printer
when the USB cable is reconnected.

1. Adjust the height of the monitor/control panel.


Press the height adjustment knob to set the monitor/control panel to the highest
position.
2. Remove the upper, right-front panel.
a) Loosen the 1/4-turn captive screw.

Fig. 69: Locking Bolt in the Storage Loction

b) Grip the transducer port insets of the panel and pull to disengage the ball snaps

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 67 of 150
sd

Antares

68

Hardware Configuration
3. Remove the left-front panel.
a) Loosen the 1/4-turn captive screw.

Fig. 70: Upper, Left-Front Panel Captive Screw

b) Grip the air flow vent insets of the panel and pull the panel to disengage the ball
snaps.
4. Remove the bottom post cover.
a) Loosen the four screws located on either side of the post.
b) Remove the post cover.
5. Install the locking bolt.
a) Remove the locking bolt from the locking bolt storage location.
b) Tighten the locking bolt to the post to lock the height adjustment on the highest position
6. Attach the bottom post cover.
a) Place the cover with the painted, chamfered side facing up.
b) Tighten the four post cover screw that secure the cover to the system. Locate two
screws on either side of the post.
7. Install the right-front panel.
a) Align the panel ball snap receptacles with the system ball snaps and press the panel to engage the ball snaps.
b) Tighten the 1/4-turn captive screw.
8. Install the left-front panel.
a) Align the panel ball snap receptacles with the system ball snaps and press the panel to engage the ball snaps.
b) Tighten the 1/4-turn captive screw.

Antares

US03-201.812.01.02.02
07.05

Page 68 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

69

9. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
10. Return the mat to the peripheral shelf.
11. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.
12. Put the plastic paper tray on the bracket, aligning the dashed line with the front edge of
the bracket.
13. Position the printer on the peripheral shelf.
a) Put the color printer on the plastic paper tray and bracket.
b) Align the printer so the back panel contacts the two sheet-metal tabs. The front
sheet-metal tabs should sit in the grooves on the bottom of the printer, a few centimeters from the front panel.

Fig. 71: Color Printer Foot Placement on Bracket

14. Set the color printer dip switches as follows:

Fig. 72: CP770 Dip Switch Settings to OFF

Fig. 73: CP800 Dip Switch Settings

1-OFF, 2-OFF, 3-OFF, 4-OFF,5-OFF,


6-OFF, 7-OFF, and 8-OFF.

1-ON, 2-OFF, 3-OFF, 4-OFF,5-OFF,


6-OFF, 7-OFF, and 8-ON.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 69 of 150
sd

Antares

70

Hardware Configuration
15. Connect the color printer cables.

Fig. 74: CP770 and CP800 Cable Connections on Bacl of Printer

a) Connect the color printer AC power cable to the color printer AC power inlet and to
the lower peripheral AC power outlet.
b) Connect the VCR printer AC power cable to the upper peripheral AC power outlet.
c) Connect the parallel end of the parallel-to-USB cable to the back of the color printer.
d) Route the USB end of the parallel-to-USB cable through the peripheral shelf cable
passage to the I/O module.
e) Route the I/O module end of the VCR USB cable through the peripheral shelf cable
passage to the I/O module.
NOTE

Antares

Do not connect the USB cables to the I/O module at this time, unless the system previously had one of the printers installed on the
peripheral bay. Then, connect the previously installed printer to
the same port it was connected to before.

US03-201.812.01.02.02
07.05

Page 70 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

71

16. Put the mounting bracket on the printer.

Fig. 75: SVOC9500 Barcket Placement on the CP770 or CP800

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 71 of 150
sd

Antares

72

Hardware Configuration
17. Position the VCR on the VCR bracket and color printer. Align the rear feet of the VCR
against the tabs of the retaining bracket.

Fig. 76: SVO9500 Positioned on the CP770 or CP800

18. Install the cover.


a) Put the cover over the VCR.

Antares

US03-201.812.01.02.02
07.05

Page 72 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

73

b) Align the four captive screws in the cover with the mounting holes.
c) Do not tighten the four captive screws.

Fig. 77: Install the Cover - SVO9500/CP770 or CP800

19. Connect the VCR cables.

Fig. 78: SVC 9500 VCR Connections from I/O Module


Pos. 1

Audio In (left and right)

Pos. 2

Video In (use either connector)

Pos. 3

Audio Out (left and right)

Pos. 4

Video Out

Pos. 5

Dip Switches

Pos. 6

RS-232C (connect to P128)

Pos. 7

75 (set both switches to ON)

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 73 of 150
sd

Antares

74

Hardware Configuration
a)
b)
c)
d)
e)
f)
g)
h)

Connect the power cable to the VCR and to the Antares peripheral AC power outlet.
Connect the AUDIO IN left and right channel cables to the back of the VCR.
Connect the AUDIO OUT left and right channel cables to the back of the VCR.
Set the VIDEO and S VIDEO switches to the on position.
Connect the VIDEO in/out cables to the back of the VCR.
Connect the S VIDEO in/out cables to the back of the VCR.
Connect the USB cable to the back of the VCR.
Connect the RS232 C cable to the back of the VCR.

20. Tighten the four captive screws to secure the cover.

Fig. 79: Tighten the Cover -Captive Screws - SVO9500/CP770 or CP800

Antares

US03-201.812.01.02.02
07.05

Page 74 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

75

21. Install the back peripheral cover and tighten the four screws.

Fig. 80: Install the Back Cover - SVO9500/CP770 or CP800

22. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).
23. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).
24. Complete the VCR and printer cable connections to the I/O module, see (I/O Module
Peripheral Connections / p. 97).
UP-D23MD / HS-MD 3000
Installation Summary
1. Adjust the height of the monitor/control panel.
Press the height adjustment knob to set the monitor/control panel to the highest
position.
2. Install the VCR on the peripheral shelf
3. Install the color printer on the VCR.
4. Complete the cable I/O module connections.
5. Complete the VCR configuration.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 75 of 150
sd

Antares

76

Hardware Configuration
NOTE

When performing this upgrade on a system that already has a color printer installed, remove the color printer from the peripheral
shelf.
Do not disconnect the USB cable from the I/O module. The system
has previously configured the USB port for the color printer.
Therefore, the same USB port must be used for the color printer
when the USB cable is reconnected.

Fig. 81: HS-MD3000 VRC/UPD-23MD Color Printer - Double Configuration


Pos. 1

VCR bracket

Pos. 2

VCR

Pos. 3

Color printer bracket

Pos. 4

Color printer

Pos. 5

Cover

1. Adjust the height of the monitor/control panel.


Press the height adjustment knob to set the monitor/control panel to the highest
position.

Antares

US03-201.812.01.02.02
07.05

Page 76 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

77

2. Remove the upper, right-front panel.


a) Loosen the 1/4-turn capptive screw.

Fig. 82: Upper, Right-Front Panel Captive Screw

b) Grip the transducer port insets of the panel and pull to disengage the ball snaps
3. Remove the left-front panel.
a) Loosen the 1/4-turn capptive screw.

Fig. 83: Upper, Left-Front Panel Captive Screw

b) Grip the air flow vent insets of the panel and pull the panel to disengage the ball
snaps.
4. Remove the bottom post cover.
a) Loosen the four screws located on either side of the post.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 77 of 150
sd

Antares

78

Hardware Configuration
b) Remove the post cover.

Fig. 84: Bottom Post Cover Screws

5. Install the locking bolt.


a) Remove the locking bolt from the locking bolt storage location.

Fig. 85: Locking Bolt in the Storage Loction

b) Tighten the locking bolt to the post to lock the height adjustment on the highest posi-

Antares

US03-201.812.01.02.02
07.05

Page 78 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

79

tion.

Fig. 86: Locking Height Adjustment

6. Attach the bottom post cover.


a) Place the cover with the painted, chamfered side facing up.
b) Tighten the four post cover screw that secure the cover to the system. Locate two
screws on either side of the post.
7. Install the right-front panel.
a) Align the panel ball snap recepticals with the system ball snaps and press the panel
to engage the ball snaps.
b) Tighten the 1/4-turn captive screw.
8. Install the left-front panel.
a) Align the panel ball snap recepticals with the system ball snaps and press the panel
to engage the ball snaps.
b) Tighten the 1/4-turn captive screw.
9. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
10. Return the mat to the peripheral shelf.
11. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 79 of 150
sd

Antares

80

Hardware Configuration
12. Remove the feet from the bottom of the VCR.

Fig. 87: VCR Foot Locations

a) Point the bottom side of the VCR facing up.


b) Remove the four screws that secure the feet to the VCR.
c) Remove the feet.

Antares

US03-201.812.01.02.02
07.05

Page 80 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

81

13. Secure the mounting bracket to the VCR.

Fig. 88: HS-MD3000 Secured to the Mounting Bracket

a) Put the mounting bracket on the VCR.


b) Align the mounting holes.
c) Secure the bracket to the bottom of the VCR.
WARNING

Use fasteners of the correct length.


Using fasteners that are too long may cause the hardware to contact other components, leading to product
damage or a potential electrical safety hazard. Using fasteners that are too short may not secure the component
to its mount, presenting a performance and safety hazard.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 81 of 150
sd

Antares

82

Hardware Configuration
14. Put the VCR assembly on the peripheral shelf and align the mounting holes of the VCR
bracket with the mounting holes in the mat.

Fig. 89: HS-MD3000 Positioned on the OEM Bay

15. Connect the VCR power cable to the VCR and to the lower peripheral AC power outlet.

Fig. 90: HS-MD3000 Power Cord Connection

Antares

Pos. 1

OEM power outlets

Pos. 2

VCR power inlet

US03-201.812.01.02.02
07.05

Page 82 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

83

16. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

Fig. 91: HS-MD3000 VCR Connections from the I/O Module


Pos. 1

Connect P128 to RS-232C

Pos. 2

Connect VIDEO IN

Pos. 3

Connect VIDEO OUT

Pos. 4

Connect AUDIO IN

Pos. 5

Connect AUDIO OUT

Pos. 6

Set the 75 VIDEO and the 75 S-VIDEO switches to ON

17. Route the other end of the VCR cables through the I/O module cable channel.
NOTE

Do not connect the VCR cables to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the VCR cover.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 83 of 150
sd

Antares

84

Hardware Configuration
18. Secure the color printer to the printer mounting bracket.

Fig. 92: UP-D23M Color Printer Positioned on the MD3000 VCR

a) Point the bottom side of the printer face up.


b) Put the printer mounting bracket on the printer.
c) Tighten the bracket captive screws to secure the printer to the bracket.
19. Put the printer assembly on the VCR.

Fig. 93: UP-D23M Color Printer Positioned on the MD3000 VCR

Antares

US03-201.812.01.02.02
07.05

Page 84 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

85

20. Connect the color printer cables.

Fig. 94: UP-D23M Color Printer Connections

NOTE

(Fig. 94 / p. 85) does not show the VCR with the VCR cables connected. However, the VCR cables were connected in a previous
steps, see (VCR I/O Module Cable Connections / p. 98).

a) Connect the printer power cable into the upper peripheral outlet
b) Connect the printer USB cable to the back of the printer
21. Route the other end of the printer USB cable through the I/O module cable channel.
NOTE

Do not connect the USB cable to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the VCR cover.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 85 of 150
sd

Antares

86

Hardware Configuration
22. Install the cover.

Fig. 95: UP-D23M Color Printer and MD3000 VCR Installed

a) Put the printer cover over the color printer/VCR assembly.


b) Verify the front lip of the cover is against the front of the printer/VCR assembly.
c) Tighten the three printer cover captive screws.
23. See the Sony User Instruction to perform the following tasks:
- Load the ink cartridge
NOTE

Power on the system prior to loading the ink cartridge. This resets
the gears to their default location, permitting the ink cartridge to be
inserted.

- Load the paper tray


- Slide the paper tray into the printer
24. Complete the VCR and printer cable connections to the I/O module, see (I/O Module
Peripheral Connections / p. 97).
25. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).
26. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).
UP-D55MD / HS-MD3000
Installation Summary
1. Adjust the height of the monitor/control panel.
Press the height adjustment knob to set the monitor/control panel to the highest
position.

Antares

US03-201.812.01.02.02
07.05

Page 86 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

87

2. Install the color printer on the peripheral shelf


3. Install the VCR on the color printer.
4. Complete the cable I/O module connections.
5. Complete the VCR configuration.

NOTE

When performing this upgrade on a system that already has a color printer installed, remove the color printer from the peripheral
shelf.
Do not disconnect the USB cable from the I/O module. The system
has previously configured the USB port for the color printer.
Therefore, the same USB port must be used for the color printer
when the USB cable is reconnected.

Fig. 96: UP-D55M Color Printer/HS-MD3000 VRC - Double Configuration


Pos. 1

Color printer bracket

Pos. 2

Color printer

Pos. 3

VCR bracket

Pos. 4

VCR

Pos. 5

Cover

1. Adjust the height of the monitor/control panel.


Press the height adjustment knob to set the monitor/control panel to the highest
position.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 87 of 150
sd

Antares

88

Hardware Configuration
2. Remove the upper, right-front panel.
a) Loosen the 1/4-turn capptive screw.

Fig. 97: Upper, Right-Front Panel Captive Screw

b) Grip the transducer port insets of the panel and pull to disengage the ball snaps
3. Remove the left-front panel.
a) Loosen the 1/4-turn capptive screw.

Fig. 98: Upper, Left-Front Panel Captive Screw

b) Grip the air flow vent insets of the panel and pull the panel to disengage the ball
snaps.
4. Remove the bottom post cover.
a) Loosen the four screws located on either side of the post.

Antares

US03-201.812.01.02.02
07.05

Page 88 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

89

b) Remove the post cover.

Fig. 99: Bottom Post Cover Screws

5. Install the locking bolt.


a) Remove the locking bolt from the locking bolt storage location.

Fig. 100: Locking Bolt in the Storage Loction

b) Tighten the locking bolt to the post to lock the height adjustment on the highest posi-

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 89 of 150
sd

Antares

90

Hardware Configuration
tion.

Fig. 101: Locking Height Adjustment

6. Attach the bottom post cover.


a) Place the cover with the painted, chamfered side facing up.
b) Tighten the four post cover screw that secure the cover to the system. Locate two
screws on either side of the post.
7. Install the right-front panel.
a) Align the panel ball snap recepticals with the system ball snaps and press the panel
to engage the ball snaps.
b) Tighten the 1/4-turn captive screw.
8. Install the left-front panel.
a) Align the panel ball snap recepticals with the system ball snaps and press the panel
to engage the ball snaps.
b) Tighten the 1/4-turn captive screw.
9. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
10. Return the mat to the peripheral shelf.

Antares

US03-201.812.01.02.02
07.05

Page 90 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

91

11. Attach the printer bracket to the bottom of the color printer.

Fig. 102: UP-D55M Color Printer Bracket

a) Point the bottom side of the printer face up.


b) Put the bracket on the bottom of the printer.
c) Tighten the bracket captive screw.
12. Connect the printer cables.

Fig. 103: UP-D55M Cable Connections


Pos. 1

Power cord between the OEM AC power outlet and the printer AC power inlet

Pos. 2

USB cable

a) Connect the AC power cable to the Antares peripheral AC power outlet and to the
printer AC power inlet.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 91 of 150
sd

Antares

92

Hardware Configuration
b) Connect the USB cable to the back of the printer.
c) Route the other end of the USB cable through the I/O module cable routing passage.
NOTE

Do not connect the USB cable to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.

13. Secure the mounting bracket to the VCR.

Fig. 104: MD3000 VCR Bracket for the Mounting on the UP55-MD

a) Put the mounting bracket on the VCR.


b) Align the mounting holes.
c) Tighten the captive screws that secure the bracket to the bottom of the VCR.
14. Put the VCR assembly on the color printer.
15. Connect the AC power cord to the upper peripheral AC power outlet and to the VCR AC
power inlet. (The color printer is connected to the lower power outlet.)

Antares

US03-201.812.01.02.02
07.05

Page 92 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

93

16. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

Fig. 105: HS-MD3000 VCR Connections from the I/O Module


Pos. 1

Connect P128 to RS-232C

Pos. 2

Connect VIDEO IN

Pos. 3

Connect VIDEO OUT

Pos. 4

Connect AUDIO IN

Pos. 5

Connect AUDIO OUT

Pos. 6

Set the 75 VIDEO and the 75 S-VIDEO switches to ON

17. Route the other end of the VCR cables through the I/O module cable channel.
NOTE

Do not connect the cables to the I/O module at this time.


Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the VCR cover.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 93 of 150
sd

Antares

94

Hardware Configuration
18. Install the cover.

Fig. 106: MD3000 VCR/UP-D55M Installed

a) Put the printer cover over the color printer/VCR assembly.


b) Verify the front lip of the cover is flush against the front of the printer/VCR assembly.
c) Tighten the three printer cover captive screws.
19. See the Sony User Instruction to perform the following tasks:
- Load the ink cartridge
- Load the paper tray
- Slide the paper tray into the printer
20. Complete the VCR and printer cable connections to the I/O module, see (I/O Module
Peripheral Connections / p. 97).
21. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).
22. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

Antares

US03-201.812.01.02.02
07.05

Page 94 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

95

Peripheral Shelf Mounting Hole Locations

Usethe following two tables and Printer and VCR Mounting Hole Locations figure to
determine which mounting holes to cut out for the peripheral configuration being installed.
The two tables list which holes to cut from the mat. The hole locations are numbered for
easy identification. The location of the mounting hole cut-outs are shown in the following
figure.
Tab. 6

Single Configurations

Peripheral Mounted on peripheral


Shelf

Mounting Hole Locations

CP770 or CP800 color printer

1-3-10-11

UP-D23MD color printer

1-4-8

UP-D55MD color printer

1-2-9

P91 D or P93 D B/W printer

6-8

HS-MD3000 VCR

1-3-9

SVO9500

2-10

Tab. 7

Double Configurations

Peripheral Mounted on Peripheral


Shelf

Mounting Hole Locations

P91 D or P93 D B/W printer

1-3-10-11

CP770 or CP800 color printer


P91 D or P93 D B/W printer

1-2-9

UP-D23MD color printer


P91 D or P93 D B/W printer

1-2-9

UP-D55MD color printer


P91 D or P93 D B/W printer

1-2-9

HS-MD3000 VCR
CP770 or CP800 color printer

1-3-10-11

HS-MD3000 VCR
UP-D23MD color printer

1-4-8

HS-MD3000 VCR
UP-D55MD color printer

1-2-9

HS-MD3000 VCR

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 95 of 150
sd

Antares

96

Hardware Configuration
Peripheral Mounted on Peripheral
Shelf

Mounting Hole Locations

P91 D or P93 D B/W printer

1-2-9

SVO9500 VCR
CP770 or CP800 color printer

2-10

SVO9500 VCR
UP-D23MD color printer

1-4-8

SVO9500 VCR
UP-D55MD color printer

1-2-9

SVO9500 VCR

Fig. 107: Peripheral Mounting Hole Guide

Antares

US03-201.812.01.02.02
07.05

Page 96 of 150
sd

Siemens
Medical Solutions USA, Inc.

Hardware Configuration

97

I/O Module Peripheral Connections

Complete the cable connections to I/O module from the B/W printer, color printer, or VCR.
The following figure shows the I/O module cable connections for all peripherals.
1. USB-A port (modem), J1
2. Left (Channel 1) audio in from VCR
3. Right (Channel 2) audio in from Right (Channel 2) audio in from VCR
4. S-Video In from VCR
5. Composite Video Out
6. GB/s video and limited COM 2
7. VCR control (RS-232 COM 1)
8. Ethernet (Network)
9. USB-A ports (printers): J5B (color printers)
10. USB-A ports (printers): J5A (black and white)
11. System status LEDs
12. Parallel port (printer)
13. Physio Trigger Out
14. S-Video Out to VCR
15. Right Audio Out to VCR
16. Left Audio Out to VCR

Fig. 108: I/O Module Connections

Printer I/O Module Cable Connections

For B/W printers: connect the USB cable to J5A.


See (Pos. 10/Fig. 108 / p. 97).

For color printers: connect the USB cable to J5B.


See (Pos. 9/Fig. 108 / p. 97).

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 97 of 150
sd

Antares

98

Hardware Configuration

VCR I/O Module Cable Connections

Fig. 109: HS-MD3000 VCR to I/O Module Connections

Fig. 110: SVO9500 VCR to I/O Module Connections

Antares

US03-201.812.01.02.02
07.05

Page 98 of 150
sd

Siemens
Medical Solutions USA, Inc.

Printer Setup and Configuration


4-

99

3Printer Setup and Configuration

This procedure assumes that the printers were installed on the system before installing
software. Software version 4.0 has installation and configuration requirements that differ
from previous software releases. There are two different installation and configuration procedures. Use the following table to identify which procedure to use.
Tab. 8

Software Version 4.0 Printers

Printers

Manufacturer

Printer Type

SW v2.1

SW v3.0

SW v4.0

CP770DW

Mitsubishi

Image

CP800DW

Mitsubishi

Image

HP1150

HP

Report

HP2500n

HP

Report

HP4050

HP

HP4100

1, 2

Report

HP

Report

HP4200

HP

Report

HP6122

HP

Report

P91DW

Mitsubishi

Image

P93DW

Mitsubishi

Image

UPD23MD

Sony

Image

UPD55MD

Sony

Image

1. When installing multiple printers on a system with v4.0, install the CP800 first.
2. CP800 is not supported on systems configured with the RM-165 PCB.

Use the peripheral software setup instructions that match the software version running on
the Antares system.

Software version 2.1 release runs on Windows 2000


Software version 3.0 release runs on Windows 2000
Software version 4.0 release runs on Windows XP OS

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 99 of 150
sd

Antares

100

Printer Setup and Configuration

SW v2.1 and V3.0


NOTE

When installing two printers and one of them is the CP800, install
the CP800 printer first.

CP800 Installation
NOTE

When installing two printers and one of the printers is the CP800,
install the CP800 first.

1. Boot the Antares system to imaging mode.


2. Press the printer Power button. Allow the printer to initialize (approximately ten seconds).
3. Connect the printer USB cable to J5B on the I/O panel. Reference (9/Fig. 108 / p. 97).
NOTE

The system briefly displays a message indicating that the Smart


Cable is installing.

4. Press the Presets button. The system displays the Presets menu.
5. Select Service. The system displays the System Information, Service Options, Remote Access Options, and Video panes.

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 100 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration

101

6. Select Configure Local Printer.

Fig. 111: Presets > Service > Configure Local Printer

The system displays the Add Printer screen.


7. Double-click the Add Printer icon.

Fig. 112: Add Printer Icon (SW v2.1 and v3.0)

The system displays the Add Printer Wizard.


8. Select Next. The Local or Network window displays.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 101 of 150


sd

Antares

102

Printer Setup and Configuration


9. If not already selected, select the Local printer check box.
Verify Automatically detect and install my Plug-and-play printer is not selected.

Fig. 113: Local or Network Printer (SW v2.1 and v3.0)

10. Select Next. The Select a Printer Port screen displays.


11. Select an unused USB virtual printer port from the Use the following Port list.

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 102 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration


NOTE

103

Select a USB port that has not yet been assigned to a printer (usually USB001, but could be USB002 or USB003). Once a USB port
has been assigned to a printer, it cannot be re-assigned to a different printer without first uninstalling and re-installing the USB port
resource.

Fig. 114: Select a Printer Port (SW v2.1and v3.0)

12. Select Next. The Add Printer Wizard screen displays.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 103 of 150


sd

Antares

104

Printer Setup and Configuration


13. If not already selected, select Mitsubishi from the Manufacturers list and Mitsubishi
CP800DW (LPT) from the Printers list.

Fig. 115: Add Printer Wizard (SW v2.1 and v3.0)

14. Select Next.

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 104 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration

105

15. If the Use Existing Driver window displays, select Keep existing driver (recommended).

Fig. 116: Use Existing Driver (SW v2.1 and v3.0)

16. Select Next. The Name Your Printer screen displays.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 105 of 150


sd

Antares

106

Printer Setup and Configuration


17. Verify the name displayed in the Printer Name field is Mitsubishi CP800DW (LPT).

Fig. 117: Name Your Printer (SW v2.1 and v3.0)

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 106 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration

107

18. IF not already selected, select Do not share this printer.

Fig. 118: Printer Sharing (SW v2.1 and v3.0)

19. Select Next. The Print Test Page screen displays.


20. Select Yes to print a test page.
21. Select Next.
22. Select Finish. A dialog box displays to verify the Windows Test Page printed.
Verify the page printed.
23. Select Ok.
NOTE

If the teat page did


Troubleshooting / p. 143).

not

print,

reference

(Printer

24. If there are no more printers to install, continue to (Reboot System / p. 122).

CP770DW, P91DW, P93DW, UPD23MD, and UPD55MD Installation


NOTE

Siemens
Medical Solutions USA, Inc.

When installing two printers and one of them is the CP800, install
the CP800 printer first.

US03-201.812.01.02.02
07.05

Page 107 of 150


sd

Antares

108

Printer Setup and Configuration


1. Boot the Antares system to imaging mode.
2. Press the printer Power button. Allow the printer to initialize (approximately ten seconds).
3. Connect the printer USB cable to the I/O panel.
NOTE

The system briefly displays a message verifying that the printer is


installing.

- For B/W printers: connect the printer USB cable to J5B on the I/O panel. Reference
(9/Fig. 108 / p. 97).
- For color printers: connect the printer USB cable to J5A on the I/O panel. Reference
(10/Fig. 108 / p. 97).
NOTE

The system briefly displays a message indicating that the printer


is installing. If there are problems trying to install the printer, reference (Printer Troubleshooting / p. 143).

4. If there are no more printers to install, continue to (Reboot System / p. 122).

HP1150, HP2500n, HP4050, HP4100, HP4200, and HP6122 Installation

The report printers automatically install and configure during the software installation. The
following steps allow the user to install a printer manually (if problems are encountered
during the software installation) or if the user is adding the printer to an existing, working
system.
1. Power on the Antares system.
2. Power on the printer.
Allow the Antares system to complete the boot sequence to imaging mode.
Wait for the printer to indicate it is ready (indicator light or display message).
3. Connect the printer USB cable to the I/O module.
- For B/W printers: connect the USB cable to J5A. Reference (10/Fig. 108 / p. 97).
- For color printers: connect the USB cable to J5B. Reference (9/Fig. 108 / p. 97).
The system checks for new hardware and displays the following message:
New Hardware has been detected.
Allow the new hardware installation to complete.
4. Register a patient in the Patient Registration screen.
a) Press the Patient key or select the Patient Registration icon. The system displays
the Patient Registration screen.
b) Enter patient information into the screen fields and select Ok.

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 108 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration

109

5. Press the Report key or select the Report icon. The system displays the Patient Report screen.
6. Select Print Report. The system prints the report.
- Compare the printed report to the information on the screen.
- If the report did not print properly, see (Printer Installation Troubleshooting / p. 143)
and if necessary (Restoring Plug-and-Play Resources / p. 145).
7. If there are no more printers to install, continue to (Reboot System / p. 122).

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 109 of 150


sd

Antares

110

Printer Setup and Configuration

SW v4.0

CP800 Installation

NOTE

The CP800 is not supported on systems with the RM-165 hardware.

NOTE

When installing two printers and one of them is the CP800, install
the CP800 printer first.

1. Boot the Antares system to imaging mode.


2. Press the printer Power button. Allow the printer to initialize (approximately ten seconds).
3. Connect the printer USB cable to J5B on the I/O panel. Reference (9/Fig. 108 / p. 97).
NOTE

The system briefly displays a message indicating that the Smart


Cable is installing.

4. The system detects the printer and displays the Found New Hardware Wizard screen.
a) If not already selected, select the Install the software automatically (recommended) check box.
b) Select Next. The system installs the USB software automatically.
c) Verify the installation was successful.
The system displays the following screen when the installation was successful:

Fig. 119: Successful Installation Screen

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 110 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration

111

The system displays the following screen when the installation was not successful:

Fig. 120: Installation Failure Screen

NOTE

If the installation fails, complete the following steps:


1. Select Finish.
2. Disconnect the Smart Cable from J5B on the I/O module.
3. Reconnect the Smart Cable to J5B on the I/O module.
4. Repeat step 4.

d) Select Finish.
The Found New Software Wizard closes.
5. Press the Presets button. The system displays the Presets menu.
6. Select Service. The system displays the System Information, Service Options, Remote Access Options, and Video panes.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 111 of 150


sd

Antares

112

Printer Setup and Configuration


7. Select Configure Local Printer.

Fig. 121: Presets > Service > Configure Local Printer

The system displays the Add Printer screen.


8. Double-click the Add Printer icon.

Fig. 122: Add Printer Icon (SW v4.0)

The system initiates the Add Printer Wizard.


9. Select Next. The Local or Network window displays.

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 112 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration

113

10. If not already selected, select the Local printer attached to this computer check box.
Verify Automatically detect and install my Plug-and-play printer is not selected.

Fig. 123: Local or Network Printer (SW v4.0)

11. Select Next. The Select a Printer Port screen displays.


12. Select an unused USB virtual printer port from the Use the following Port drop-down
menu.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 113 of 150


sd

Antares

114

Printer Setup and Configuration


NOTE

Select a USB port that has not yet been assigned to a printer (usually USB001, but could be USB002 or USB003). Once a USB port
has been assigned to a printer, it cannot be re-assigned to a different printer without first uninstalling and re-installing the USB port
resource.

Fig. 124: Select a Printer Port (SW v4.0)

13. Select Next. The Install Printer Software screen displays.

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 114 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration

115

14. If not already selected, select the MITSUBISHI from the Manufacturer list,
CP800DW (LPT) from the Printers list and select Next. The Use Existing Driver window displays.

Fig. 125: Install Printer Software (SW v4.0)

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 115 of 150


sd

Antares

116

Printer Setup and Configuration


15. If the Use Existing Driver screen displays, select Keep existing driver (recommended).

Fig. 126: Use Existing Driver (SW v4.0)

16. Select Next. The Name Your Printer screen displays.

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 116 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration

117

17. Verify the name displayed in the Printer Name field is Mitsubishi CP800DW (LPT).

Fig. 127: Name Your Printer (SW V4.0)

18. Select No for Do you want to use this printer as the default printer.
19. Select Next. The Print Test Page screen displays.
20. Select Yes for Do you want to print a test page.
21. Select Next. The Completing the Add Printer Wizard screen displays.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 117 of 150


sd

Antares

118

Printer Setup and Configuration


22. Verify the installation was successful.

Fig. 128: Install Printer Software (SW v4.0)

23. Select Finish. A dialog box prompts to verify the Windows Test Page printed.
24. Verify the test page printed.
NOTE

If the test page did not print or If errors occurred during the installation, see (Printer Troubleshooting / p. 143).

25. Select Ok.


26. If there are no more printers to install, continue to the next section, (Reboot
System / p. 122).

CP770DW, P91DW, P93DW, UPD23MD, and UPD55MD Installation


NOTE

When installing two printers and one of them is the CP800, install
the CP800 printer first.

1. Boot the Antares system to imaging mode.


2. Press the printer Power button. Allow the printer to initialize (approximately ten seconds).

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 118 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration

119

3. Connect the printer USB cable to the I/O panel.


NOTE

The system briefly displays a message verifying that the printer is


installing.

- For B/W printers: connect the printer USB cable to J5A on the I/O panel. Reference
(10/Fig. 108 / p. 97).
- For color printers: connect the printer USB cable to J5B on the I/O panel. Reference
(9/Fig. 108 / p. 97).
NOTE

The system may briefly displays a message indicating that the


printer is installing. Select Next and/or Finnish (depending on the
printer).
Some printers install the USB support automatically.
If there are problems trying to install the printer, reference (Printer
Troubleshooting / p. 143).

4. For CP770 and P91DW only:


During the first installation of a USB device using a Smart Cable, the USB installation
fails as shown in (Fig. 120 / p. 111).
If the installation fails, complete the following steps:
a)
b)
c)
d)

Select Finish.
Disconnect the Smart Cable from the J5A (P91DW) or J5B (CP770) I/O module.
Reconnect the Smart Cable to J5A (P91DW) or J5B (CP770) on the I/O module.
Repeat step 3.

If the installation fails again, reference (Printer Troubleshooting / p. 143).


5. The system detects the printer. The system displays the Found New Hardware Wizard for USB printing support.
a) If not already selected, select the Install the software automatically check box.
b) Select Next. The system installs the USB software automatically and then displays
a window indicating the install is complete.
c) Select Finish.
The Found New Hardware Wizard closes.
NOTE

If the first installation attempt of the USB printer will fail. Disconnect the printer(s) USB cable from the I/O panel and start the procedure from step 2.

6. For P91DW only:


Switch off the Bi-directional support.
a) Press Presets.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 119 of 150


sd

Antares

120

Printer Setup and Configuration


b)
c)
d)
e)
f)
g)

Select Service.
Select Configure Local Printer. The system displays the printer screen.
Right-click Mitsubishi P91DW (LPT) and select Properties.
Select the Ports tab.
Make sure Enable bidirectional support is not checked.
Select Ok.

7. If there are no more printers to install, continue to (Reboot System / p. 122).

NOTE

If errors occurred during the installation, see (Printer Installation


Troubleshooting / p. 143).

HP4050, HP4100 and HP4200 Installation

1. With the Antares system powered off, connect the parallel cable to J1 of the I/O module
parallel printer port and the back of the printer. See (Fig. 108 / p. 97) to locate the parallel printer port.
2. Power on the printer.
Allow the Antares system to complete the boot sequence to imaging mode.
Wait for the printer to indicate it is ready (indicator light or display message).
3. Power on the Antares system.
4. The system checks for new hardware and displays the following message:
New Hardware has been detected.
Allow the new hardware installation to complete.
5. Select Next.
Allow the installation to complete and then select Ok.
6. Register a patient in the Patient Registration screen.
a) Press the Patient key or select the Patient Registration icon. The system displays
the Patient Registration screen.
b) Enter patient information into the screen fields and select Ok.
7. Press the Report key or select the Report icon. The system displays the Patient Report screen.
8. Select Print Report. The system prints the report.
- Compare the printed report to the information on the screen.
- If the report did not print properly, see (Printer Installation Troubleshooting / p. 143)
and if necessary (Restoring Plug-and-Play Resources / p. 145).
9. If there are no more printers to install, continue to (Reboot System / p. 122).

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 120 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration

121

HP1150, HP2500n and HP6122 Installation

1. Complete the printer hardware installation.


2. Power on the printer.
Wait for the printer to indicate it is ready (indicator light or display message).
3. Connect the parallel cable to J1 of the I/O module parallel printer port and the back of
the printer. See (Fig. 108 / p. 97) to locate the parallel printer port.
4. Power on the Antares system.
Allow the Antares system to complete the boot sequence to imaging mode.
5. The system checks for new hardware and displays the following message:
New Hardware has been detected.
Allow the new hardware installation to complete.
6. Select Next.
Allow the installation to complete and then select Ok.
7. Register a patient in the Patient Registration screen.
a) Press the Patient key or select the Patient Registration icon. The system displays
the Patient Registration screen.
b) Enter patient information into the screen fields and select Ok.
8. Press the Report key or select the Report icon. The system displays the Patient Report screen.
9. Select Print Report. The system prints the report.
- Compare the printed report to the information on the screen.
- If the report did not print properly, see (Printer Installation Troubleshooting / p. 143)
and if necessary (Restoring Plug-and-Play Resources / p. 145).
10. If there are no more printers to install, continue to (Reboot System / p. 122).

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 121 of 150


sd

Antares

122

Printer Setup and Configuration

Reboot System

Perform a system reboot only after completing the installation of all printers.
1. Reboot the system.
a) Press the Standby button.
b) Select Shut Down from the displayed menu.
c) Wait ten seconds after the system completely shuts down. Then press Standby
again to power on the system.
2. Continue to (Print/Store Setup / p. 123).

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 122 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration

123

Print/Store Setup

After the DICOM store and print devices are configured and the system has been rebooted, it is necessary to configure the Print and Store key assignments. The Print/Store
Setup permits the following configuration of the Antares Print and Store keys:

Automatic transfer of static images, either at the End of exam or During exam.
Automatic transfer of static images, either at the End of exam or During exam.
Automatic print routing, based on the image type.
Storing images as Secondary Capture (global setting).
Burning graphics into pixel data (global setting).

The Print and Store keys on the control panel can be configured to do print only, print and
store, or store only, regardless of the graphic on the button. The graphics located at the
top of the print/store presets screen represent the buttons on the control panel, not the
function of the button.

Fig. 130: Print Key

Fig. 131: Store Key


Fig. 129: Print and Store Keys
Pos. 1

Print Key

Pos. 2

Store Key

Workflow for configuring Print and Store:


1. Restore the factory settings and factory printer settings.
2. Configure routing to Print and Store devices.
3. Configure the Print key.
4. Configure the Store key.
5. Configure Advanced Store Transfer (optional).

Restore Factory Settings and Factory Printer Settings

Perform the Restore Factory Printer Settings function when configuring print settings for
the first time. Perform the Restore Factory Settings function for configuring print and store
settings for the first time.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 123 of 150


sd

Antares

124

Printer Setup and Configuration


1. Restore factory printer settings:
a) Select Presets > Service.
b) Select Restore Factory Printer Settings.

Fig. 132: Presets > Service > Restore Printer Factory Defauts

2. Restore factory settings:


a) Select Print/Store.
b) Select Restore Factory Settings.

Fig. 133: Presets > Print/Store > Restore Factory Settings

Selecting Restore Factory Settings populates the different lists with the installed print
and store devices.

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 124 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration

125

3. Continue to the (US03-103.843.01 / Configure Print) section.

Configure Routing to Print and Store Devices

Configure the routing for different image types to specific printers and storage devices.
1. Display the Print/Store Screen.
a) Select the Presets icon.
b) Select Print/Store from the Presets menu.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 125 of 150


sd

Antares

126

Printer Setup and Configuration


2. Assign the routing for specific ultrasound image types to the Print and/or Store keys.
Use the following table to determine assignments:
Tab. 9

Image Type Routing

BW

Selecting a printer from the drop-down menu sends black and white
images to the printer selected.
Selecting a storage device from the drop-down menu sends black
and white images to the printer selected.

Color Doppler

Selecting a printer from the drop-down menu sends color images to


the printer selected.
Selecting a storage device from the drop-down menu sends color
images to the printer selected.

Tinted

Selecting a printer from the drop-down menu sends tinted images to


the printer selected.
Selecting a storage device from the drop-down menu sends tinted
images to the printer selected.

2D Ref

Selecting a printer from the drop-down menu sends images with


color in the 2D portion of PW Doppler to the printer selected.
Selecting a storage device from the drop-down menu sends images
with color in the 2D portion of PW Doppler to the printer selected.

Misc.

Selecting a printer from the drop-down menu sends secondary captures (screen captures) images to the printer selected.
Selecting a storage device from thre drop-down menu sends secondary captures (screen captures) images to the printer selected.

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 126 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration

127

Fig. 134: Print Routing Setup


Pos. 1

Print key routing configuration for printing or storing specific image types

Pos. 2

Store key routing configuration for storing or printing specific image types

Pos. 3

Store function assigned to the Print key

Pos. 4

Store function assigned to the Store key

3. Assign the store function to the Print or the Store key.


- Select the Print icon Store checkbox (3/Fig. 134 / p. 127).
Select to perform a dual function: print the image to the related image printer
and store the image to the Antares default storage device when the Print key is
pressed.
Uncheck to print the image to the related image printer only.
NOTE

Images are always stored to the local database when either the
Print or Store key is pressed, regardless of the Print/Store configuration.

- Select the Store icon Store checkbox.


Select when printers are added to the store device column, to perform a dual
function: print the image to the related image printer and store the image to the
Antares default storage device when the Store key is pressed.
Uncheck when no printers are assigned to the store device column: to store the
images when the Store key is pressed.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 127 of 150


sd

Antares

128

Printer Setup and Configuration

Configure Print

After configuring a DICOM or local peripheral print device, assign image types to a destination printer, configure transfer types, and set the printed page composition.
1. Display the Print/Store Screen.
a) Select the Presets icon.
b) Select Print/Store from the Presets menu.
2. Select the Configure Printer tab.

Fig. 135: Print/Store > Configure Printers Tab


Pos. 1

Printer column: list of installed printers

Pos. 2

Formatted by selection: System or Printer

3. For each device in the Printer column (1/Fig. 135 / p. 128), make the following selections from the related drop-down menus:
-

Antares

Auto Transfer: defines when to transfer static images.


Layout: defines the number of images per film sheet.
Media Size: defines the size of the media.
Orientation: defines the orientation of the media.

US03-201.812.01.02.02
07.05

Siemens
Page 128 of 150
Medical Solutions USA, Inc.
sd

Printer Setup and Configuration


NOTE

129

Siemens recommends that transfer at the End of Exam be selected for archive servers. Transfer During Exam should only be selected for image review workstations that support In-Progress
review.
Servers configured for Storage Commitment should not be configured for transfer During Exam.
Auto transfer to CD-R is disabled by default.

NOTE

Do not restore factory settings if this screen is already set up for


onboard printers and it is necessary to add a DICOM printer. If this
is the case, skip step 2 and proceed to step 3.

4. Select Printer or System for the Formatted by option (2/Fig. 135 / p. 128).
a) Printer: Assigns the DICOM printers to perform composition and formatting. This is
the default setting. This setting is typically more efficient, allowing smaller blocks of
data to be transferred over the network, and consuming less ultrasound system
CPU resources.
b) System: Assigns the ultrasound system to perform page composition and formatting. This provides uniformity across different printers, and in some cases results in
very large blocks of data being transferred to the printer.
5. After setting the print routing, select Save to save the settings. All changes are discarded unless the Save button is selected.
NOTE

If the Print or Store key is not configured for printing, then all the
drop-down menus under that key icon should be set to empty.

Configure Store

To configure local printers to the Store key, reference the (US03-103.843.01 / Configure
Store) section of System Configuration.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 129 of 150


sd

Antares

130

Printer Setup and Configuration

Verify Printer Operation

Verify the operation of the printers.


The instructions provided in the following steps are a condensed version of the instructions found in the Antares Instructions for Use. See the Antares Instructions for Use for
a complete set of all printer instructions.
1. Print an image during an active study.
a) Press the Print/Store key assigned to the printing function.
The system sends a request to the printer that is configured for the type of image
printed. The image is also stored to the local database.
b) Verify the print function assigned to the Print/Store key was completed.
2. Store an image during an active study:
a) Press the Print/Store key assigned to the storing function.
b) Verify the system stores the image to the local database.

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 130 of 150
Medical Solutions USA, Inc.
sd

VCR Setup and Configuration


5-

131

4VCR Setup and Configuration

Complete the following VCR configurations sections:

VCR Configuration
VCR Controls
Verify the Operation of the VCR

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 131 of 150


sd

Antares

132

VCR Setup and Configuration

HS-MD3000 Configuration

When installing a Mitsubishi HS-MD3000 VCR for use with an Antares, it is necessary to
optimize the presets for the best playback image quality. Select these presets using the
front panel display and controls on the VCR.

Fig. 136: HS-MD3000 VCR Front Panel


Pos. 1

Power

Pos. 2

Monitor - Mix

Pos. 3

Audio - Mix

Pos. 4

Key Lock - All

Pos. 5

Stop

Pos. 6

Play

Pos. 7

Frame Field/Enter

Pos. 8

Menu

Pos. 9

Jog Dial

The configuration instructions describe how to set up the following preset menus on the
HS-MD3000 VCR:

Antares

<Menu 1>
<Menu 2>
<Menu 3>
<Menu 4>
Hidden Menu

US03-201.812.01.02.02
07.05

Siemens
Page 132 of 150
Medical Solutions USA, Inc.
sd

VCR Setup and Configuration

133

Front Panel Controls


Tab. 10

VCR Controls

Control

Function

Menu

To access the VCR preset menus, press the Menu button on the front
panel of the VCR. The first press of the Menu button displays
<MENU 1>. Each press of the Menu button changes the menu
sequentially to the next menu. The sequential order of menus is:
<MENU 1>, <MENU 2>, <MENU 3>, <MENU4>, <ELAPSED TIME>;
and exit the VCR menus.

Stop + Play

To enter the Hidden Menu, simultaneously press and hold STOP


and PLAY for at least ten seconds.

Jog/Adjust

Turn the JOG dial to the left or right to select an item within a menu.
Turn to the right for a higher numbered setting. Turned to the left for a
lower numbered setting.

Frame/ Field Enter

Press the Enter button to edit a menu value. Then turn the Jog dial to
select a different value. Press Enter again to save the change.

Menu 1 Settings
NOTE

This procedure assumes the VCR displays default settings. If the


setting has been changed, the VCR will display the newest settings.

1. Press MENU for approximately two seconds. The VCR displays 101:00.
2. Turn the JOG dial to the left or right to select each value within <Menu 1>.
3. Verify that each menu item value displays correctly. The following table lists the correct
settings:
Tab. 11

<Menu 1>

Menu Item

Antares
Configuration

VCR Front
Panel Display

REC TAPE END

STOP

101:00

TAPE END

STOP

102:00

TAPE IN

STOP

103:00

TAPE LENGTH

PAL = T-120

104:00

NTSC = E-180
VISS

Siemens
Medical Solutions USA, Inc.

REC

US03-201.812.01.02.02
07.05

105:01

Page 133 of 150


sd

Antares

134

VCR Setup and Configuration


Menu Item

Antares
Configuration

VCR Front
Panel Display

TIME CODE REC

OFF

106:00

TIME DATE DISP

OFF

107:00

FOOT SWITCH

L.E.

108:00

DIMMER

OFF

109:00

4. If an item does not match:


a)
b)
c)
d)
e)

Press the ENTER button. The value after the colon (:) will flash.
Turn the JOG dial until the correct value displays.
Press the ENTER button to set.
Turn the JOG dial to the next item as needed to change menu item values.
When the last <Menu 1> item is displayed, press the MENU button. The VCR displays <Menu 2>.

5. Continue to <Menu 2>.


Menu 2 Settings
NOTE

This procedure assumes the VCR displays default settings. If the


setting has been changed, the VCR will display the newest settings.

When <Menu 2> is selected the VCR displays 201:01.


1. Turn the JOG dial to the left or right to select each value within <Menu 2>.
2. Verify that each menu item value displays correctly. The following table lists the correct
settings:
Tab. 12

<Menu 2>

Menu Item

Antares

Antares
Configuration

VCR Front
Panel Display

S-VHS/S-ET (NTSC)

ON/OFF

201:01

SVHS (PAL)

AUTO

Y-NR

ON

202:01

C-NR

ON

203:01

Y-ENHANCER

ON

204:01

TBC

ON

205:01

US03-201.812.01.02.02
07.05

Siemens
Page 134 of 150
Medical Solutions USA, Inc.
sd

VCR Setup and Configuration


Menu Item

Antares

135

Configuration

VCR Front
Panel Display

COLOR FILTER

OFF

206:00

FREEZE MODE

FRAME

207:00

HI-FI AUDIO

ON

208:01

NORMAL AUDIO IN

MIX

209:00

3. If an item does not match:


a)
b)
c)
d)
e)

Press the ENTER button. The value after the colon (:) will flash.
Turn the JOG dial until the correct value displays.
Press the ENTER button to set.
Turn the JOG dial to the next item as needed to change menu item values.
When the last <Menu 2> item is displayed, press the MENU button. The VCR displays <Menu 3>.

4. Continue to <Menu 3>.


Menu 3 Settings
NOTE

This procedure assumes the VCR displays default settings. If the


setting has been changed, the VCR will display the newest settings.

When <Menu 3> is selected the VCR displays 301:00. <Menu 3> is used to set the clock;
it is not necessary to set the clock.Press the MENU button to continue to <Menu 4>.
Menu 4 Settings
NOTE

This procedure assumes the VCR displays default settings. If the


setting has been changed, the VCR will display the newest settings.

When <Menu 4> is selected the VCR displays 401:04.


1. Turn the JOG dial to the left or right to select each value within <Menu 4>.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 135 of 150


sd

Antares

136

VCR Setup and Configuration


2. Verify that each menu item value displays correctly. The following table lists the correct
settings:
Tab. 13

<Menu 4>

Menu Item

Antares
Configuration

VCR Front
Panel Display

TRANSMISSION RATE

19200

401:04

DATA BIT LENGTH

8 BITS

402:01

STOP BIT LENGTH

1 BIT

403:01

PARITY BIT

NONE

404:00

3. If an item does not match:


a)
b)
c)
d)

Press the ENTER button. The value after the colon (:) will flash.
Turn the JOG dial until the correct value displays.
Press the ENTER button to set the change and go to the next menu item.
When the last <Menu 4> item is displayed, press the MENU button twice to exit.

4. Continue to the Hidden Menu.


Hidden Menu Settings
NOTE

This procedure assumes the VCR displays default settings. If the


setting has been changed, the VCR will display the newest settings.

1. Simultaneously press and hold STOP and PLAY for at least ten seconds. The VCR displays PC: 2.
If PC: 2 is not displayed, turn the JOG dial until PC: 2 is displayed.
2. Press the ENTER button to set the change and continue to the next Hidden Menu item.
3. Verify that each menu item value displays correctly. The following table lists the correct
settings:
Tab. 14

Hidden Menu

Menu Item

Antares

Antares
Configuration

VCR Front
Panel Display

PICTURE CONTROL (PC)

PC: 2

SWITCHING POINT (SP)

45

SP:45

POSITION UP/DOWN (Ud)

Ud: 0

POSITION L/R (Lr)

Lr: 0

US03-201.812.01.02.02
07.05

Siemens
Page 136 of 150
Medical Solutions USA, Inc.
sd

VCR Setup and Configuration


Menu Item

Antares

137

Configuration

VCR Front
Panel Display

CHROMA AMPLI-TUDE (CA)

CA: 0

HUE (CP)

CP: 0

COLOR DELAY (CD)

Cd: 1

BRIGHT-NESS CONTROL
(bC)

-1

bC:-1

CONTRAST CONTROL (CC)

-1

CC:-1

BLACK DISPLAY (bd)

OFF

bd:oF

4. If an item does not match:


a)
b)
c)
d)
e)

Press the ENTER button. The value after the colon (:) will flash.
Turn the JOG dial until the correct value displays.
Press the ENTER button to set.
Turn the JOG dial to the next item as needed to change menu item values.
When the last item is displayed, press the Stop button to exit.
NOTE

Siemens
Medical Solutions USA, Inc.

A menu item cannot be exited if the menu item is flashing. Press


the Enter button as needed to stop the menu item from flashing.

US03-201.812.01.02.02
07.05

Page 137 of 150


sd

Antares

138

VCR Setup and Configuration

SVO9500 Configuration

When installing a Sony SVO9500 VCR for use with an Antares, it is necessary to optimize
the presets for the best playback image quality. Select these presets using the front panel
display and controls on the VCR.

Front Panel Settings

Fig. 137: SVO 9500 VCR Front Panel Controls


Pos. 1

HiFi switch

Pos. 3

Program switch

Pos. 4

CTL/Menu switch

Pos. 5

Menu display

Pos. 6

Power button

Pos. 7

Data setting selection button

Pos. 8

Item menu selection button

Pos. 9

Play button

Tab. 15

SVO 9500 Main Menu Controls

Control

Setting

Function

Audio

Hi-Fi

Selects the signal to be output from the Audio


Out CH-1 and CH-2 connectors for hi-fi audio to
be output.

Hi-Fi

ON

Records hi-fi audio.

Program

OFF

The VCR is controled from the Antares system.

CTL/Menu

CTL

Displays the time counter during taping.

Menu

Displays the Hidden menu.

ON

Initiates power to the VCR.

Power

Antares

Audio switch

Pos. 2

US03-201.812.01.02.02
07.05

Siemens
Page 138 of 150
Medical Solutions USA, Inc.
sd

VCR Setup and Configuration

139

Control

Setting

Function

Data

N.A.

When the CTL/Menu control is set to Menu,


press Data to scroll through the settings of each
hidden menu item.

Item

N.A.

When the CTL/Menu control is set to Menu,


press Item to advance to each of the main menu
items.

Play

N.A.

Press Play to play back recorded video.

Main Menu Settings


NOTE

This procedure assumes the VCR displays default settings. If the


setting has been changed, the VCR will display the newest settings.

To display the Main Menu:


1. Set the CTL/Menu switch to Menu.
2. Press the Item button to display each menu and its associated value.
3. Press the Data button to change the item value.
4. Verify the values match the settings in (Tab. 16 / p. 139).
Tab. 16

SVO 9500 Main Menu Items

VCR Front
Antares Option
Panel Display Setting
vIA-

ON

VISS operation mode.

FOt--

LE

Footswitch operation mode.

tPE-

Et

Tape end operation mode. Automatically ejects tape when


the tape reaches the end.

sVH-

At

S-VHS recoreing selection. Automatically detects tape


type.

Crn-

ON

Digital CRN On/Off. Switches on the digital chroma noise


reducer.

Yrn-

ON

Digital YRN On/Off. Switches on the digital luminance


noise reducer.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 139 of 150


sd

Antares

140

VCR Setup and Configuration


VCR Front
Antares Option
Panel Display Setting
Yen-

ON

Digital Y enhancer On/Off. Switches on the digital Y


enhancer.

FrE-

Fr

Freeze moe selection. Displays a frame-freeze picture.

Hidden Menu Settings


NOTE

This procedure assumes the VCR displays default settings. If the


setting has been changed, the VCR will display the newest settings.

To access the Hidden menu:


1. Set the CTL/Menu switch to Menu.
2. Press the Item button until FrE-Fr displays in the menu window.
3. Simultaneously press Item and Play.
4. Press the Item button to display each menu and its associated value.
5. Press the Data button to change the item value.
6. Verify that each menu item value displays correctly. The following table lists the correct
settings:
Tab. 17

SVO 9500 Hidden Menu Items

VCR Front
Antares Option
Panel Display Setting

Antares

SOP-

OFF

Pause status on the 34P 1/F becomes true when the


speed is zero in Search mode, asa well as Pause mode.

34P-

PA

When the host sends a pause command through the


34P I/F, the VCR goes into Freeze instead of Pause
(default is PA). It does not function when the VCR is in
Jog/Shuttle or search mode. It is available in Play mode
only.

FrS-

PA

After the Auto Freeze function completes, the VCR goes


into speed ero of the Shuttle mode, instead of Pause
mode.

1PH-

65

The switching position loch phase can be shifted to either


6.5H, 5.5H, or 4.5H before Vsync. The phase in REC stays
at 6.5H always. 6.5H must be selected when adjusting the
switching postition.

US03-201.812.01.02.02
07.05

Siemens
Page 140 of 150
Medical Solutions USA, Inc.
sd

VCR Setup and Configuration

141

VCR Front
Antares Option
Panel Display Setting
L1n-

ON

Defeats the line resizing function in the Reverse Search


mode. It reduces the image remaining phenomenon. The
image is stretched verticallly as a side effect.

nOn-

OFF

Changes to Non-Interlaced mode om Pause and Field


Freeze. It is valid only if the VM-12 baord is installed, otherwise no image comes out.

FrA-

ON

Switches the framing servo off. Field Memory mode can be


selected. It is valid only if the Non-Interlaced mode is on.

H s-

OFF

Shifts the image to the left approximately 350 ns whenit is


on. The default is off.

v s-

OFF

Shifts the image down four lines during Jog, Shuttle, or


Pause modes. The default is off.

7. If an item does not match:


a)
b)
c)
d)
e)

Press the ENTER button. The value after the colon (:) will flash.
Turn the JOG dial until the correct value displays.
Press the ENTER button to set.
Turn the JOG dial to the next item as needed to change menu item values.
When the last item is displayed, press the Stop button to exit.
NOTE

Siemens
Medical Solutions USA, Inc.

A menu item cannot be exited if the menu item is flashing. Press


the Enter button as needed to stop the menu item from flashing.

US03-201.812.01.02.02
07.05

Page 141 of 150


sd

Antares

142

VCR Setup and Configuration

Verify the Operation of the VCR

1. Power on the Antares system (if not already on).


2. Power on the VCR (if not already on).
3. Insert a VCR tape into the VCR.
NOTE

See the VCR manual for detailed steps.

4. Press the VCR key to toggle between the record and the pause functions.
The VCR key illuminates when the Record function is active.

Fig. 138: VCR - Record/ Pause Key

5. Record the active screen to video tape. Use one of the following methods:
- Use the VCR panel controls to record the active screen.
- Use the VCR group box to control the VCR directly from the Image task card (if this
option is installed on the system).
6. Playback the recording. Use one of the following methods:
- Use the VCR panel controls to record the active screen.
- Use the VCR group box to control the VCR directly from the Image task card (if this
option is installed on the system).
The playback should display the active screen as viewed during the recording
period.

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 142 of 150
Medical Solutions USA, Inc.
sd

Printer Troubleshooting

143

6-

Printer Installation Troubleshooting


5Printer Troubleshooting

If the test page does not print, refer to the following list for possible causes:

The printer is connected to AC power.


The printer is powered on.
The printer cable is securely connected to the system. When using a parallel cable,
tighten the locking screws.

The printer is not out of paper and/or ribbon.


The printer is On Line (On Line light is illuminated).
If the printer is still not functioning, consult the following table:
Tab. 18

Printer Installation Troubleshooting

Symptom

Cause

Solution

Plug-and-play printers
do not install when
connected to the system.

Plug-and-play
resources are used
and must be restored.

1. Follow steps in the Restoring


Plug-and-play Resources.
2. Using the Add/Remove Hardware
Wizard, remove hardware device.
3. Repeat the instructions to install
the printer.

Unable to print.

The printer was


Either nothing happens deleted from the Windows Printer page or is
or an error displays.
not correctly configured in the system
Print/Store routing
page.

Repeat the instructions to install the


printer or check entries in Print/Store
routing page. Printers are not displayed as selections in the Print/
Store routing page until completing
the installation and a system reboot.

Print out does not


match screen.

Printer defaults may


not be correct.

Select Presets > Service > Restore


Printer Factory Defaults. Close Presets screen.

The CP800DW printer


does not print.

SCSI mode may be


enabled. Check the
SCSI light on front
panel.

SW1 on rear panel must be on for


parallel mode.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 143 of 150


sd

Antares

144

Printer Troubleshooting
Symptom

Cause

Solution

Report contains prints


that are shrunken or
expanded.

Printer defaults may be 1. Select Presets > Service > Conincorrect.


figure Local Printer.
2. Right-click on the report printer
and select Properties > Printing
Preferences > Layout > Advanced > Graphic > Print Quality
> 600 DPI
3. Select Ok to close the Advanced
Options screen.
4. Select Ok to close the Printing
Preferences screen.
5. Select Ok to close the Properties
screen.
6. Close the Printer window.
7. Close the Presets screen.

An error message displays on the Printer


screen.

Check printer manual


for list.

The Printer Error indicator lamp illuminates.

Antares

US03-201.812.01.02.02
07.05

Follow instructions on the printer


screen
Identify the error indicator resolution
in the printer manual.

Siemens
Page 144 of 150
Medical Solutions USA, Inc.
sd

Printer Troubleshooting

145

Restoring Plug-and-Play Resources

A driver can be installed for a plug-and-play resource only once. At which time Windows
marks that resource as used.(A used resource displays as a grayed-out device on the
hardware list.) If a printer is not recognized and cannot be installed, the resource may
need to be restored. The device must be deleted and the plug-and-play function re-initiated.

NOTE

A service level 4 password or higher is required to restore


plug-and-play resources.

NOTE

After restoring the plug-and-play resources, the USB cable must


be disconnected and reconnected to initiate the plug-and-play installation.

To restore the Plug-and-play resources:


1. Delete the printer from the Configure Local Printer screen:
a) Disconnect from the I/O panel, all USB cables connected to thermal printers.
It is not necessary to disconnect the LPT cable for a Report Printer if one is installed.
b) Select Presets > Service > Configure Local Printer.
c) Right-click the Printer icon (if present) and select Delete. Repeat for each printer
that is to be deleted.
d) Select Yes.
If printer deleted was the default printer, the system displays a warning message.
e) Select Ok. Close Printer folder.
2. Access NT Command Interpreter:
a)
b)
c)
d)
e)

Select Local Service.


Enter the password in the password field and select Ok.
Select Utilities.
Select Escape to OS from the Source drop-down list.
Select NT Command Interpreter from the Command drop-down list.

3. Access and delete the printer through Explorer:


a) Type explorer in the Parameter field.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 145 of 150


sd

Antares

146

Printer Troubleshooting
b)
c)
d)
e)
f)
g)
h)
i)

Select Go.
Select Next.
Right-click My Computer, then select Properties > Hardware > Device Manager.
Select the View pull-down menu.
Select Show hidden devices and expand Printers.
Select the printer for deletion. Right-click the selected printer and select Uninstall.
Confirm deletion of printer.
Expand Universal Serial Bus controllers, then select the appropriate printing support for the printer to be deleted.
j) Confirm deletion of USB Printing Support.
Repeat steps 3 for each USB Printing Support device entry from the list in turn.
There is a USB Printing Support device for the USB to parallel cable, which is installed as a separate device. All USB Printing Support devices should be removed
to free up resources.

4. Close the Service screen.


Printers may now be installed according to the specific installation procedure in this
document.

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 146 of 150
Medical Solutions USA, Inc.
sd

Printer Troubleshooting

147

Re-Install A Printer After Troubleshooting or Restoring


Plug-and-Play

Use the following table to determine which installation procedure to use.


Tab. 19

Printer Re-installation Instructions

Printers

Manufacturer

Use Instruction

HP1150, HP2500n, HP6122

HP

(Report Printers / p. 147)

Mitsubishi

SW v2.1 or v3.0: (CP800


Installation / p. 100)

HP4050, HP4100, HP4200,


CP800WD

SW v4.0: (CP800 Installation / p. 110)


P91DW, P93DW, CP770DW

Mitsubishi

UPD23MD, UPD55

Sony

(Re-install Plug-and-Play
Printers / p. 148)

Report Printers

The following steps manual printer installation if problems are encountered during the
software install process or if add a printer to an existing, working system.

NOTE

If the printer is unplugged or removed from the system, refer to the


(Restoring Plug-and-Play Resources / p. 145) for the printer uninstall procedure.

HP1150, HP2500n, HP 6122


The Antares system must be powered off before beginning the printer installation.
1. Power on the report printer and permit printer to initialize, illuminating the Ready light.
2. Connect the USB cable between the printer and the Antares:
a) Connect one end of the USB cable into the USB port on the back of the printer.
b) Route the cable through the peripheral cable passage.
c) Connect the other end of the USB cable into a USB port on the I/O panel.
3. The system will detect the printer and display the Found New Hardware Wizard window.
a) If not already selected, select the Install the software automatically (recommended) check box.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 147 of 150


sd

Antares

148

Printer Troubleshooting
b) Select Next. The system installs the USB software automatically.
c) Select Finish.
The Found New Hardware Wizard closes and the system returns to imaging.
4. Press the Patient button. The system displays the Patient Registration screen.
5. Register a patient and select Ok.
6. Press the Report button. The system displays the Patient Report screen.
7. Select Print Report.
8. Compare the printed report to the information on the screen.
HP 4050, HP4100, HP4200
The Antares system must be powered off before beginning the printer installation.
1. Connect the parallel cable between the printer and the Antares:
a) Connect the parallel cable into the parallel printer connector on the I/O panel and
secure with the locking screws.
b) Route the cable through the peripheral cable passage.
c) Connect the other end of the parallel cable into the parallel connector on the back of
the printer and lock it down with the clips.
2. Power on the Report Printer on.
Wait until the printer displays Ready on the front panel.
The system detects the printer and displays the Found New Hardware Wizard
window.
a) If not already selected, select the Install the software automatically (recommended) check box.
b) Select Next. The system installs the USB software automatically.
c) Select Finish.
The Found New Hardware Wizard closes and the system returns to imaging.
3. Press the Patient button. The system displays the Patient Registration screen.
4. Register a patient and select Ok.
5. Press the Report button. The system displays the Patient Report screen.
6. Select Print Report.
7. Compare the printed report to the information on the screen.

Re-install Plug-and-Play Printers

The Plug-and-play procedure assumes the system is booted to imaging. Perform the following steps for each printer to install; one printer at a time.
1. Connect the AC power cable between the printer and the Antares peripheral outlet.

Antares

US03-201.812.01.02.02
07.05

Siemens
Page 148 of 150
Medical Solutions USA, Inc.
sd

Printer Troubleshooting

149

2. Press the printer Power button to power on the printer.


3. Connect the USB cable between the printer and the I/O panel.
a) Connect the square-shaped end of the USB cable to the rear panel of the printer.
b) Route the USB cable through the peripheral cable passage.
c) Connect the rectangular-shaped end of the USB cable to J5B for color printers, or
J5A for black and white printers.
4. The system detects the plug-and-play printers. The system displays the Found New
Hardware Wizard for USB printing support.
a) If not already selected, select the Install the software automatically (recommended) check box.
b) Select Next. The system installs the USB software automatically.
c) Select Finish.
The Found New Hardware Wizard closes.

Siemens
Medical Solutions USA, Inc.

US03-201.812.01.02.02
07.05

Page 149 of 150


sd

Antares

150

Antares

Printer Troubleshooting

US03-201.812.01.02.02
07.05

Siemens
Page 150 of 150
Medical Solutions USA, Inc.
sd