Installation Instructions
Peripherals
Printer and VCR Configurations
10032747
10032746
08653771
05936518
Siemens AG
Siemens Medical
Ultrasound
lne
n.a.
DivisionSolutions
2004
English
Doc. Gen. Date: 07.05
n.a.
Revision / Disclaimer
1Revision / Disclaimer
Copyright
Copyright 2004 by Siemens Corporation. All rights reserved. No part of this publication
may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into
any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of
Siemens Corporation. Siemens Corporation reserves the right to change its products and
services at any time. In addition, this manual is subject to change without notice. Siemens
Corporation welcomes customer input on corrections and suggestions for improvements
to this manual.Although Siemens Corporation has attempted to ensure accuracy throughout this manual, Siemens Corporation assumes no liability for any errors or omissions, nor
for any damages resulting from the application or use of this information.
Trademarks
ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE,
Adara, Antares, G20, G40, G50, G60S, Omnia and KinetDx are trademarks of Siemens
Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000,
WS3000, and WebPro are trademarks of Siemens Corporation.
Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet
Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are
registered trademarks of Netscape Communications Corporation. Java and the JavaScript
name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered
trademark of Adobe Systems, Inc.
Other products and brand names are trademarks of their respective owners.
Disclaimer
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
License Agreement
All computer programs copyright 1990-2004 by Siemens Corporation or its suppliers. Programs are licensed under the following agreement:
Siemens or its suppliers retain(s) ownership of and title to any computer program supplied
with the equipment and to the trade secrets embodied in such computer programs. Subject to the Buyers acceptance and fulfillment of the obligations in this paragraph, Siemens
grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any
computer program supplied with the Equipment that is necessary to operate the Equipment solely on the medium in which such program is delivered for the purpose of operating the equipment in accordance with the instructions set forth in the operators manuals
supplied with the Equipment and for no other purpose whatsoever. Buyer may not
reverse-assemble, reverse-compile or otherwise reverse-engineer such computer programs nor may Buyer make a copy of such program or apply any techniques to derive the
trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms
Antares
US03-201.812.01.02.02
07.05
Page 2 of 150
sd
Siemens
Medical Solutions USA, Inc.
Revision / Disclaimer
of this license, the license granted by this paragraph shall terminate. Further, because
unauthorized use of such computer programs will leave Siemens without an adequate
remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to
restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s
suppliers of software is a direct and intended beneficiary of this end-user sublicense and
may enforce it directly against Buyer with respect to software supplied by such supplier,
and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indirect, consequential, incidental or other damages arising out of the sublicense of the computer programs supplied with the equipment.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 3 of 150
sd
Antares
4
1-
Table of Contents
0Table of Contents
14
14
17
20
23
26
31
Double Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
B/W Printer and Color Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
B/W Printer and VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Color Printer and VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
34
34
47
62
100
100
107
108
SW v4.0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CP800 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CP770DW, P91DW, P93DW, UPD23MD, and UPD55MD Installation . . . . . . . . . .
HP4050, HP4100 and HP4200 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP1150, HP2500n and HP6122 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
110
110
118
120
121
123
123
125
128
129
Antares
US03-201.812.01.02.02
07.05
Page 4 of 150
sd
Siemens
Medical Solutions USA, Inc.
Table of Contents
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 5 of 150
sd
Antares
6
2-
Overview
1Overview
The Peripheral Installation procedure lists the latest peripheral configurations for installation and describes how to install each configuration for the the following Antares systems:
Peripheral
SONOLINE Antares
ACUSON Antares
Software Versions
v2.1
v3.0
v4.0
v4.0
v4.0
P91 D
NA
NA
P93 D
CP770DW
CP800DW
X1
NA
NA
UP-D23MD
NA
UP-D55MD
NA
HS-MD3000
X2
SVO9500
NA
NA
The following table lists the offboard peripheral report printers and their software version
installation types.
Antares
US03-201.812.01.02.02
07.05
Page 6 of 150
sd
Siemens
Medical Solutions USA, Inc.
Overview
Tab. 2
7
Report Printers:
Peripheral
SONOLINE Antares
ACUSON Antares
Software Versions
v2.1
v3.0
v4.0
v4.0
v4.0
HP1150
HP2500n
HP4050
HP4100
HP4200
HP6122
DeskJet
Before installing report printers, comply with the guidellines in the following sections:
(Report Printers / p. 8)
(Mitigating the Risk of Connecting a Report Printer / p. 10)
(Patient Environment / p. 11)
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 7 of 150
sd
Antares
Overview
Report Printers
WARNING
WARNING
WARNING
Antares
NOTE
Siemens cannot support or service any external devices connected to the USB port. Siemens cannot assume responsibility for
functionality beyond the scope of this specification.
NOTE
Siemens does not specifically recommend any particular communication or analysis software; please contact a local vendor for assistance with interpretation and analysis of the data transmitted by
the ultrasound imaging system.
US03-201.812.01.02.02
07.05
Page 8 of 150
sd
Siemens
Medical Solutions USA, Inc.
Overview
Relevant standards for some non-medical electrical equipment may have limits for enclosure leakage currents higher than required by medical standards. These higher standards
are acceptable only outside the patient environment. It is essential to reduce enclosure
leakage currents when non-medical electrical equipment is to be used within the patient
environment. Measures for reducing leakage current include use of a medically-approved
isolation transformer.
The ultrasound system has printer drivers compatible with non-medical grade report printers. Siemens does not guarantee the performance and safety of any non-medical grade
report printer. Non-medical grade report printers do not fulfill the following safety requirements:
EN 60601-1-1 and IEC 60601-1-1 (Medical Electrical Equipment, Part 1: General Requirements for Safety).
Manufacturer
Web Site
Tripp Lite
www.tripplite.com
www.toroid.com
Dale Technology
www.daletech.com
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 9 of 150
sd
Antares
10
Overview
To fulfill EN 60601-1-1 and IEC 60601-1-1 (Medical Electrical Equipment, Part 1: General
Requirements for Safety) requirements for non-medical peripheral equipment:
The non-medical peripheral equipment must be approved according to any other EN or
IEC standard (EN XXXXX or IEC XXXXX, e.g., equipment complying with EN 60348 and
IEC 60348, EN 60950 and IEC 60950, etc.)
The connection of non-medical peripheral equipment to the ultrasound system must
adhere to the following conditions:
1. Connect the ultrasound system to an AC power outlet within a medically used room
within the patient environment. A patient environment is defined as an area in which
medical examination, monitoring, or treatment of the patient takes place. The patient
environment is located 1.5 meters (or, in Canada and the U.S.A., 1.8 meters [6 feet])
around the patient location.
2. Connect the peripheral equipment AC power cable to a medically approved isolation
transformer.
3. Connect the medically approved isolation transformer to a main AC outlet either inside
or outside the patient environment. The peripheral device and medically approved isolation transformer combination can be located:
- Within the same room as the ultrasound system, or
- In a non-medically used room.
For additional information and other possible combinations, please refer to the Medical
Electrical Equipment Standard EN 60601-1-1 or IEC 60601-1-1, Annex BBB.7,
Scenario 3c.
NOTE
Antares
US03-201.812.01.02.02
07.05
Page 10 of 150
sd
Siemens
Medical Solutions USA, Inc.
Overview
11
Patient Environment
Fig. 1:
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 11 of 150
sd
Antares
12
Overview
Fig. 2:
Antares
US03-201.812.01.02.02
07.05
Page 12 of 150
sd
Siemens
Medical Solutions USA, Inc.
Overview
13
Requirements
Time required:
Pen
Exacto knife or 1/4-inch hollow punch
Phillips head screw driver
Flat head screw driver
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 13 of 150
sd
Antares
14
Hardware Configuration
3-
Single Configuration
2Hardware Configuration
Tab. 4
Single Configurations
Color Printers
B/W Printers
VCRs
(HS-MD3000
VCR / p. 26)
(SVO9500 VCR / p. 31)
(UP-D23MD Color
Printer / p. 17)
(UP-D55MD Color
Printer / p. 20)
1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.
4. Remove the feet and the shipment locking screw from the bottom of the printer.
a) Point the bottom side of the printer face up.
b) Remove the four screws that secure the feet to the printer and remove the feet.
c) Remove the shipment locking screw.
Fig. 3:
Antares
US03-201.812.01.02.02
07.05
Page 14 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
15
Fig. 4:
WARNING
6. Put the color printer assembly on the peripheral shelf and align the mounting holes of
the color printer with the mounting holes in the mat.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 15 of 150
sd
Antares
16
Hardware Configuration
7. Set the color printer dip switches as follows:
Fig. 5:
Fig. 6:
Fig. 7:
Pos. 1
Pos. 2
Power cord
c) Route the other end of the USB cable through the peripheral shelf cable passage to
the I/O module.
9. Install the cover.
a) Put the cover over the printer.
b) Align the four captive screws in the cover with the mounting holes.
c) Tighten the four captive screws.
Antares
US03-201.812.01.02.02
07.05
Page 16 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
17
10. See the Mitsubishi User Instruction to perform the following tasks:
- Load the ink cartridge
- Load the paper tray
- Slide the paper tray into the printer
11. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral
Connections / p. 97).
12. To complete the color printer configuration, see (Printer Setup and
Configuration / p. 99).
1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 17 of 150
sd
Antares
18
Hardware Configuration
4. Put the printer feet into the mounting bracket cut-outs.
Fig. 8:
Antares
Pos. 1
Rear mounting plate cutout for placement of the printers rear feet
Pos. 2
Front mounting plate cutout for placement of the printers front feet
Fig. 9:
Pos. 1
Cable routing passage through the mat and panel to the I/O module
US03-201.812.01.02.02
07.05
Page 18 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
19
Pos. 2
Pos. 3
a) Connect the AC power cable to the Antares peripheral AC power outlet and to the
printer AC power inlet.
b) Connect the USB cable to the back of the printer.
c) Route the other end of the USB cable through the I/O module cable routing passage.
(See (1/Fig. 9 / p. 18) for the routing passage location.)
NOTE
Do not connect the USB cable to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 19 of 150
sd
Antares
Hardware Configuration
Captive screw
Pos. 2
Power on the system prior to loading the ink cartridge. This resets
the gears to their default location, permitting the ink cartridge to be
inserted.
1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
Antares
US03-201.812.01.02.02
07.05
Page 20 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
21
NOTE
In (Fig. 12 / p. 21), the flat side of the bracket is oriented on the left
side of the picture.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 21 of 150
sd
Antares
22
Hardware Configuration
5. Connect the printer cables.
Power cord between the OEM AC power outlet and the printer AC power inlet
Pos. 2
USB cable
a) Connect the AC power cable to the Antares peripheral AC power outlet and to the
printer AC power inlet.
b) Connect the USB cable to the back of the printer.
c) Route the other end of the USB cable through the I/O module cable routing passage.
NOTE
Do not connect the USB cable to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.
Antares
US03-201.812.01.02.02
07.05
Page 22 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
23
Pos. 2
1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable instructions that follow.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 23 of 150
sd
Antares
24
Hardware Configuration
WARNING
P91 D Instructions:
P93 D Instructions:
er up.
Antares
US03-201.812.01.02.02
07.05
Page 24 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
25
Fig. 17: P91 D and P93 D Printer Cables and Dip Switches (P91 D shown)
Pos. 1
Dip Switches
Pos. 2
USB Cable
Pos. 3
AC Power Cord
Do not connect the USB cable to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 25 of 150
sd
Antares
26
Hardware Configuration
6. Install the printer cover.
HS-MD3000 VCR
Antares
Pos. 1
VCR bracket
Pos. 2
VCR
Pos. 3
Cover
US03-201.812.01.02.02
07.05
Page 26 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
27
1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Remove the feet from the bottom of the VCR.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 27 of 150
sd
Antares
28
Hardware Configuration
4. Secure the mounting bracket to the VCR.
WARNING
NOTE
In (Fig. 21 / p. 28), the front of the VCR is facing the left side of the
picture.
Antares
US03-201.812.01.02.02
07.05
Page 28 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
WARNING
29
5. Put the VCR assembly on the peripheral shelf by aligning the mounting holes of the
VCR with the mounting holes in the mat.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 29 of 150
sd
Antares
30
Hardware Configuration
6. Connect the VCR power cord to the peripheral AC power outlet and to the VCR AC
power inlet.
Pos. 2
7. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.
Pos. 2
Connect VIDEO IN
Pos. 3
Pos. 4
Connect AUDIO IN
Pos. 5
Pos. 6
8. Route the cables through the I/O panel cable routing passage.
Antares
US03-201.812.01.02.02
07.05
Page 30 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
NOTE
31
SVO9500 VCR
1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Secure the mounting bracket to the peripheral shelf.
3. Put the VCR assembly on the peripheral shelf by aligning the mounting holes of the
VCR with the mounting holes in the mat.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 31 of 150
sd
Antares
32
Hardware Configuration
4. Set the VCR switches: 1-OFF, 2-OFF, 3-OFF, 4-OFF, 5- ON, 6-ON
5. Connect the VCR power cord to the peripheral AC power outlet and to the VCR AC
power inlet.
6. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.
Pos. 2
Pos. 3
Pos. 4
Video Out
Pos. 5
Dip Switches
Pos. 6
Pos. 7
7. Route the cables through the I/O panel cable routing passage.
Antares
US03-201.812.01.02.02
07.05
Page 32 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
NOTE
33
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 33 of 150
sd
Antares
34
Hardware Configuration
Double Configuration
Tab. 5
Double Configuration
B/W Printer/VCR
Color Printer/VCR
Double Configuration
Double Configuration
Double Configuration
(CP770 or CP800 /
HS-MD3000 / p. 62)
(P91 D or P93 D /
UP-D23MD / p. 37)
(UP-D23MD / HS-MD
3000 / p. 75)
(P91 D or P93 D /
SVO9500 / p. 56)
(P91 D or P93 D /
UP-D55M / p. 42)
(UP-D55MD /
HS-MD3000 / p. 86)
(CP770 or CP800 /
SVO9500 / p. 67)
NOTE
1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.
Antares
US03-201.812.01.02.02
07.05
Page 34 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
35
4. Remove the feet and the shipment locking screw from the bottom of the printer.
a) Point the bottom side of the printer face up.
b) Remove the four screws that secure the feet to the printer and remove the feet.
6. Put the color printer assembly on the peripheral shelf and align the mounting holes of
the color printer with the mounting holes in the mat.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 35 of 150
sd
Antares
36
Hardware Configuration
7. Set the color printer dip switches as follows:
Do not connect the USB cables to the I/O module at this time, unless the system previously had one of the printers installed on the
peripheral bay. Then, connect the previously installed printer to
the same port it was connected to before.
9. Install the color printer cover/black and white printer mounting plate assembly.
a) Route the B/W printer USB cable and power cable up through the hole in the B/W
printer mounting plate.
b) Put the printer cover over the printer.
c) Align the four captive screws with the fasener holes cut into the mat.
d) Tighten the captive screws.
10. Set the B/W printer dip switches to:
- P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF
- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF
11. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable instructions that follow.
Antares
US03-201.812.01.02.02
07.05
Page 36 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
37
Do not connect the USB cable to the I/O module at this time.
Power on the system prior to loading the ink cartridge. This resets
the gears to their default location, permitting the ink cartridge to be
inserted.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 37 of 150
sd
Antares
38
Hardware Configuration
NOTE
1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.
4. Put the color printer feet into the mounting bracket cut-outs.
Antares
Pos. 1
Rear mounting plate cutout for placement of the printers rear feet
Pos. 2
Front mounting plate cutout for placement of the printers front feet
US03-201.812.01.02.02
07.05
Page 38 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
39
5. Connect the color and B/W power cables to the Antares AC outlet.
Pos. 2
Pos. 3
Pos. 4
Pos. 5
a) Connect the AC power cable to the Antares peripheral AC power outlet and to the
color printer AC power inlet.
b) Connect the B/W printer AC power cable to the Antares peripheral AC power outlet
and route the cable through the color printer top cover.
c) Connect the USB cable to the back of the color printer.
d) Route both of the USB cables through the I/O module cable routing passage.
NOTE
Siemens
Medical Solutions USA, Inc.
Do not connect the USB cable to the I/O module at this time, unless
the system previously had this printer installed.
US03-201.812.01.02.02
07.05
Page 39 of 150
sd
Antares
40
Hardware Configuration
6. Install the printer cover.
Pos. 2
Pos. 3
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.
Antares
US03-201.812.01.02.02
07.05
Page 40 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
41
P91 D Instructions:
P93 D Instructions:
er up.
Point the bottom side of the printer facFig. 34: P91 D Positioned on the Bracket
ing down.
Do not connect the USB cable to the I/O module at this time.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 41 of 150
sd
Antares
42
Hardware Configuration
11. See the Sony User Instruction to perform the following tasks for the color printer:
- Load the ink cartridge
NOTE
Power on the system prior to loading the ink cartridge. This resets
the gears to their default location, permitting the ink cartridge to be
inserted.
NOTE
1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
Antares
US03-201.812.01.02.02
07.05
Page 42 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
43
Power cord between the OEM AC power outlet and the printer AC power inlet
Pos. 2
USB cable
a) Connect the AC power cable to the Antares peripheral AC power outlet and to the
printer AC power inlet.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 43 of 150
sd
Antares
44
Hardware Configuration
b) Connect the USB cable to the back of the printer.
c) Route the other end of the USB cable through the I/O module cable routing passage.
NOTE
Do not connect the USB cable to the I/O module at this time, unless
the system previously had this printer connected.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.
5. Connect the B/W printer AC power cable into the Antares peripheral AC power outlet.
Route the other end of the power cable through the color printer cover.
6. Route the B/W printer USB cable through the I/O panel cable routing passage.
NOTE
Do not connect the USB cable to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.
Antares
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.
US03-201.812.01.02.02
07.05
Page 44 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
45
8. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable instructions that follow.
WARNING
P91 D Instructions:
P93 D Instructions:
er up.
Point the bottom side of the printer facFig. 39: P91 D Positioned on the Bracket
ing down.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 45 of 150
sd
Antares
46
Hardware Configuration
10. Connect the B/W printer cables.
Dip switch
Pos. 2
USB cable
Pos. 3
Power cord
a) Connect the AC power cable to the back of the P91 D/P93 D printer.
b) Connect the USB cable to the back of the P91 D/P93 D printer.
NOTE
Do not connect the USB cable to the I/O module at this time.
Fig. 42: UP-D55M Color Printer and P91 D or P93 D B/W Printer - Double Configuration
Antares
US03-201.812.01.02.02
07.05
Page 46 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
47
NOTE
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 47 of 150
sd
Antares
48
Hardware Configuration
Fig. 43: HS-MD3000 VRC/P91 Dor P93 D B/W Printer - Double Configuration
Pos. 1
VCR bracket
Pos. 2
VCR
Pos. 3
Pos. 4
B/W printer with bracket attached to the bottom of the printer (Bracket not shown)
Pos. 5
1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Return the mat to the peripheral shelf.
3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.
Antares
US03-201.812.01.02.02
07.05
Page 48 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
49
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 49 of 150
sd
Antares
50
Hardware Configuration
5. Secure the mounting bracket to the VCR.
Antares
US03-201.812.01.02.02
07.05
Page 50 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
51
6. Put the VCR assembly on the peripheral shelf and align the mounting holes of the VCR
bracket with the mounting holes in the mat.
Pos. 2
a) Connect the power cable to the VCR AC power inlet and to the lower peripheral AC
power outlet.
b) Connect the printer power cable to the upper peripheral AC outlet.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 51 of 150
sd
Antares
52
Hardware Configuration
8. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.
Pos. 2
Connect VIDEO IN
Pos. 3
Pos. 4
Connect AUDIO IN
Pos. 5
Pos. 6
NOTE
9. Route the B/W printer power cord and USB cable through the cable channel in the VCR
cover.
Antares
Pos. 1
Pos. 2
To I/O module
US03-201.812.01.02.02
07.05
Page 52 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
53
10. Route the other end of the VCR cables through the I/O module cable channel.
11. Route the other end of the B/W printer USB cable through the I/O module cable channel.
12. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable instructions that follow.
WARNING
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 53 of 150
sd
Antares
54
Hardware Configuration
P91 D Instructions:
P93 D Instructions:
er up.
Point the bottom side of the printer facFig. 50: P91 D Positioned on the Bracket
ing down.
Antares
US03-201.812.01.02.02
07.05
Page 54 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
55
Fig. 52: VCR Cover with Black and White Mounting Plate
Pos. 1
VCR Cover
Pos. 2
Dip switch
Pos. 2
USB cable
Pos. 3
Power cord
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 55 of 150
sd
Antares
56
Hardware Configuration
14. Install the B/W printer cover.
NOTE
Antares
US03-201.812.01.02.02
07.05
Page 56 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
57
1. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
2. Secure the mounting bracket to the peripheral shelf.
3. Put the VCR assembly on the peripheral shelf by aligning the mounting holes of the
VCR with the mounting holes in the mat.
4. Set the VCR switches:
- Set the dip switches: 1-OFF, 2-OFF, 3-OFF, 4-OFF, 5- ON, 6-ON.
5. Connect the VCR power cord to the peripheral AC power outlet and to the VCR AC
power inlet.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 57 of 150
sd
Antares
58
Hardware Configuration
6. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.
Pos. 2
Pos. 3
Pos. 4
Video Out
Pos. 5
Dip Switches
Pos. 6
Pos. 7
7. Route the cables through the I/O panel cable routing passage.
NOTE
Antares
US03-201.812.01.02.02
07.05
Page 58 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
59
channel.
d) Put the cover over the VCR.
e) Align and tighten the two captive screws in the cover with the mounting holes.
9. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable instructions that follow.
WARNING
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 59 of 150
sd
Antares
60
Hardware Configuration
P91 D Instructions:
P93 D Instructions:
er up.
Point the bottom side of the printer facFig. 59: P91 D Positioned on the Bracket
ing down.
Antares
US03-201.812.01.02.02
07.05
Page 60 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
61
Dip switch
Pos. 2
USB cable
Pos. 3
Power cord
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 61 of 150
sd
Antares
62
Hardware Configuration
b) Tighten the two printer cover captive screws.
12. Complete the VCR and printer cable connections to the I/O module, see (I/O Module
Peripheral Connections / p. 97).
13. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).
14. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).
NOTE
When performing this upgrade on a system that already has a color printer installed, remove the color printer from the peripheral
shelf.
Do not disconnect the USB cable from the I/O module. The system
has previously configured the USB port for the color printer.
Therefore, the same USB port must be used for the color printer
when the USB cable is reconnected.
Antares
US03-201.812.01.02.02
07.05
Page 62 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
63
b) Grip the transducer port insets of the panel and pull to disengage the ball snaps
3. Remove the left-front panel.
a) Loosen the 1/4-turn captive screw.
b) Grip the air flow vent insets of the panel and pull the panel to disengage the ball
snaps.
4. Remove the bottom post cover.
a) Loosen the four screws located on either side of the post.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 63 of 150
sd
Antares
64
Hardware Configuration
b) Remove the post cover.
5. Install the locking bolt.
a) Remove the locking bolt from the locking bolt storage location.
b) Tighten the locking bolt to the post to lock the height adjustment on the highest position
6. Attach the bottom post cover.
a) Place the cover with the painted, chamfered side facing up.
b) Tighten the four post cover screw that secure the cover to the system. Locate two
screws on either side of the post.
7. Install the right-front panel.
a) Align the panel ball snap receptacles with the system ball snaps and press the panel to engage the ball snaps.
b) Tighten the 1/4-turn captive screw.
8. Install the left-front panel.
a) Align the panel ball snap receptacles with the system ball snaps and press the panel to engage the ball snaps.
b) Tighten the 1/4-turn captive screw.
9. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
10. Return the mat to the peripheral shelf.
11. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.
12. Remove the feet and the shipment locking screw from the bottom of the printer.
a) Point the bottom side of the printer face up.
b) Remove the four screws that secure the feet to the printer and remove the feet.
Antares
US03-201.812.01.02.02
07.05
Page 64 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
65
14. Put the color printer assembly on the peripheral shelf and align the mounting holes of
the color printer with the mounting holes in the mat.
15. Set the color printer dip switches as follows:
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 65 of 150
sd
Antares
66
Hardware Configuration
NOTE
Do not connect the USB cables to the I/O module at this time, unless the system previously had one of the printers installed on the
peripheral bay. Then, connect the previously installed printer to
the same port it was connected to before.
17. Remove the feet and the shipment locking screw from the bottom of the VCR.
a) Point the bottom side of the VCR facing up.
b) Remove the four screws that secure the feet to the VCR.
c) Remove the feet.
18. Secure the mounting bracket to the VCR.
a) Put the mounting bracket on the VCR.
b) Align the mounting holes.
c) Secure the bracket to the bottom of the VCR.
WARNING
Connect the power cable to the VCR and to the Antares peripheral AC power outlet.
Connect the AUDIO IN left and right channel cables to the back of the VCR.
Connect the AUDIO OUT left and right channel cables to the back of the VCR.
Set the VIDEO and S VIDEO switches to the on position.
Connect the VIDEO in/out cables to the back of the VCR.
Connect the S VIDEO in/out cables to the back of the VCR.
Connect the USB cable to the back of the VCR.
Connect the RS232 C cable to the back of the VCR.
Antares
US03-201.812.01.02.02
07.05
Page 66 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
67
24. Complete the VCR and printer cable connections to the I/O module, see (VCR I/O Module Cable Connections / p. 98).
CP770 or CP800 / SVO9500
Installation Summary
1. Adjust the height of the monitor/control panel.
2. Install the color printer on the peripheral shelf
3. Install the VCR printer on the color printer.
4. Complete the cable I/O module connections.
5. Complete the VCR configuration.
NOTE
When performing this upgrade on a system that already has a color printer installed, remove the color printer from the peripheral
shelf.
Do not disconnect the USB cable from the I/O module. The system
has previously configured the USB port for the color printer.
Therefore, the same USB port must be used for the color printer
when the USB cable is reconnected.
b) Grip the transducer port insets of the panel and pull to disengage the ball snaps
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 67 of 150
sd
Antares
68
Hardware Configuration
3. Remove the left-front panel.
a) Loosen the 1/4-turn captive screw.
b) Grip the air flow vent insets of the panel and pull the panel to disengage the ball
snaps.
4. Remove the bottom post cover.
a) Loosen the four screws located on either side of the post.
b) Remove the post cover.
5. Install the locking bolt.
a) Remove the locking bolt from the locking bolt storage location.
b) Tighten the locking bolt to the post to lock the height adjustment on the highest position
6. Attach the bottom post cover.
a) Place the cover with the painted, chamfered side facing up.
b) Tighten the four post cover screw that secure the cover to the system. Locate two
screws on either side of the post.
7. Install the right-front panel.
a) Align the panel ball snap receptacles with the system ball snaps and press the panel to engage the ball snaps.
b) Tighten the 1/4-turn captive screw.
8. Install the left-front panel.
a) Align the panel ball snap receptacles with the system ball snaps and press the panel to engage the ball snaps.
b) Tighten the 1/4-turn captive screw.
Antares
US03-201.812.01.02.02
07.05
Page 68 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
69
9. Cut the fastener holes into the rubber mat of the peripheral shelf.
See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.
10. Return the mat to the peripheral shelf.
11. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.
12. Put the plastic paper tray on the bracket, aligning the dashed line with the front edge of
the bracket.
13. Position the printer on the peripheral shelf.
a) Put the color printer on the plastic paper tray and bracket.
b) Align the printer so the back panel contacts the two sheet-metal tabs. The front
sheet-metal tabs should sit in the grooves on the bottom of the printer, a few centimeters from the front panel.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 69 of 150
sd
Antares
70
Hardware Configuration
15. Connect the color printer cables.
a) Connect the color printer AC power cable to the color printer AC power inlet and to
the lower peripheral AC power outlet.
b) Connect the VCR printer AC power cable to the upper peripheral AC power outlet.
c) Connect the parallel end of the parallel-to-USB cable to the back of the color printer.
d) Route the USB end of the parallel-to-USB cable through the peripheral shelf cable
passage to the I/O module.
e) Route the I/O module end of the VCR USB cable through the peripheral shelf cable
passage to the I/O module.
NOTE
Antares
Do not connect the USB cables to the I/O module at this time, unless the system previously had one of the printers installed on the
peripheral bay. Then, connect the previously installed printer to
the same port it was connected to before.
US03-201.812.01.02.02
07.05
Page 70 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
71
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 71 of 150
sd
Antares
72
Hardware Configuration
17. Position the VCR on the VCR bracket and color printer. Align the rear feet of the VCR
against the tabs of the retaining bracket.
Antares
US03-201.812.01.02.02
07.05
Page 72 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
73
b) Align the four captive screws in the cover with the mounting holes.
c) Do not tighten the four captive screws.
Pos. 2
Pos. 3
Pos. 4
Video Out
Pos. 5
Dip Switches
Pos. 6
Pos. 7
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 73 of 150
sd
Antares
74
Hardware Configuration
a)
b)
c)
d)
e)
f)
g)
h)
Connect the power cable to the VCR and to the Antares peripheral AC power outlet.
Connect the AUDIO IN left and right channel cables to the back of the VCR.
Connect the AUDIO OUT left and right channel cables to the back of the VCR.
Set the VIDEO and S VIDEO switches to the on position.
Connect the VIDEO in/out cables to the back of the VCR.
Connect the S VIDEO in/out cables to the back of the VCR.
Connect the USB cable to the back of the VCR.
Connect the RS232 C cable to the back of the VCR.
Antares
US03-201.812.01.02.02
07.05
Page 74 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
75
21. Install the back peripheral cover and tighten the four screws.
22. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).
23. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).
24. Complete the VCR and printer cable connections to the I/O module, see (I/O Module
Peripheral Connections / p. 97).
UP-D23MD / HS-MD 3000
Installation Summary
1. Adjust the height of the monitor/control panel.
Press the height adjustment knob to set the monitor/control panel to the highest
position.
2. Install the VCR on the peripheral shelf
3. Install the color printer on the VCR.
4. Complete the cable I/O module connections.
5. Complete the VCR configuration.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 75 of 150
sd
Antares
76
Hardware Configuration
NOTE
When performing this upgrade on a system that already has a color printer installed, remove the color printer from the peripheral
shelf.
Do not disconnect the USB cable from the I/O module. The system
has previously configured the USB port for the color printer.
Therefore, the same USB port must be used for the color printer
when the USB cable is reconnected.
VCR bracket
Pos. 2
VCR
Pos. 3
Pos. 4
Color printer
Pos. 5
Cover
Antares
US03-201.812.01.02.02
07.05
Page 76 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
77
b) Grip the transducer port insets of the panel and pull to disengage the ball snaps
3. Remove the left-front panel.
a) Loosen the 1/4-turn capptive screw.
b) Grip the air flow vent insets of the panel and pull the panel to disengage the ball
snaps.
4. Remove the bottom post cover.
a) Loosen the four screws located on either side of the post.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 77 of 150
sd
Antares
78
Hardware Configuration
b) Remove the post cover.
b) Tighten the locking bolt to the post to lock the height adjustment on the highest posi-
Antares
US03-201.812.01.02.02
07.05
Page 78 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
79
tion.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 79 of 150
sd
Antares
80
Hardware Configuration
12. Remove the feet from the bottom of the VCR.
Antares
US03-201.812.01.02.02
07.05
Page 80 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
81
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 81 of 150
sd
Antares
82
Hardware Configuration
14. Put the VCR assembly on the peripheral shelf and align the mounting holes of the VCR
bracket with the mounting holes in the mat.
15. Connect the VCR power cable to the VCR and to the lower peripheral AC power outlet.
Antares
Pos. 1
Pos. 2
US03-201.812.01.02.02
07.05
Page 82 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
83
16. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.
Pos. 2
Connect VIDEO IN
Pos. 3
Pos. 4
Connect AUDIO IN
Pos. 5
Pos. 6
17. Route the other end of the VCR cables through the I/O module cable channel.
NOTE
Do not connect the VCR cables to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the VCR cover.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 83 of 150
sd
Antares
84
Hardware Configuration
18. Secure the color printer to the printer mounting bracket.
Antares
US03-201.812.01.02.02
07.05
Page 84 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
85
NOTE
(Fig. 94 / p. 85) does not show the VCR with the VCR cables connected. However, the VCR cables were connected in a previous
steps, see (VCR I/O Module Cable Connections / p. 98).
a) Connect the printer power cable into the upper peripheral outlet
b) Connect the printer USB cable to the back of the printer
21. Route the other end of the printer USB cable through the I/O module cable channel.
NOTE
Do not connect the USB cable to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the VCR cover.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 85 of 150
sd
Antares
86
Hardware Configuration
22. Install the cover.
Power on the system prior to loading the ink cartridge. This resets
the gears to their default location, permitting the ink cartridge to be
inserted.
Antares
US03-201.812.01.02.02
07.05
Page 86 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
87
NOTE
When performing this upgrade on a system that already has a color printer installed, remove the color printer from the peripheral
shelf.
Do not disconnect the USB cable from the I/O module. The system
has previously configured the USB port for the color printer.
Therefore, the same USB port must be used for the color printer
when the USB cable is reconnected.
Pos. 2
Color printer
Pos. 3
VCR bracket
Pos. 4
VCR
Pos. 5
Cover
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 87 of 150
sd
Antares
88
Hardware Configuration
2. Remove the upper, right-front panel.
a) Loosen the 1/4-turn capptive screw.
b) Grip the transducer port insets of the panel and pull to disengage the ball snaps
3. Remove the left-front panel.
a) Loosen the 1/4-turn capptive screw.
b) Grip the air flow vent insets of the panel and pull the panel to disengage the ball
snaps.
4. Remove the bottom post cover.
a) Loosen the four screws located on either side of the post.
Antares
US03-201.812.01.02.02
07.05
Page 88 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
89
b) Tighten the locking bolt to the post to lock the height adjustment on the highest posi-
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 89 of 150
sd
Antares
90
Hardware Configuration
tion.
Antares
US03-201.812.01.02.02
07.05
Page 90 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
91
11. Attach the printer bracket to the bottom of the color printer.
Power cord between the OEM AC power outlet and the printer AC power inlet
Pos. 2
USB cable
a) Connect the AC power cable to the Antares peripheral AC power outlet and to the
printer AC power inlet.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 91 of 150
sd
Antares
92
Hardware Configuration
b) Connect the USB cable to the back of the printer.
c) Route the other end of the USB cable through the I/O module cable routing passage.
NOTE
Do not connect the USB cable to the I/O module at this time.
Use cable ties to neatly bundle excess cable length. Excess cable
will be stored under the printer cover.
Fig. 104: MD3000 VCR Bracket for the Mounting on the UP55-MD
Antares
US03-201.812.01.02.02
07.05
Page 92 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
93
16. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.
Pos. 2
Connect VIDEO IN
Pos. 3
Pos. 4
Connect AUDIO IN
Pos. 5
Pos. 6
17. Route the other end of the VCR cables through the I/O module cable channel.
NOTE
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 93 of 150
sd
Antares
94
Hardware Configuration
18. Install the cover.
Antares
US03-201.812.01.02.02
07.05
Page 94 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
95
Usethe following two tables and Printer and VCR Mounting Hole Locations figure to
determine which mounting holes to cut out for the peripheral configuration being installed.
The two tables list which holes to cut from the mat. The hole locations are numbered for
easy identification. The location of the mounting hole cut-outs are shown in the following
figure.
Tab. 6
Single Configurations
1-3-10-11
1-4-8
1-2-9
6-8
HS-MD3000 VCR
1-3-9
SVO9500
2-10
Tab. 7
Double Configurations
1-3-10-11
1-2-9
1-2-9
1-2-9
HS-MD3000 VCR
CP770 or CP800 color printer
1-3-10-11
HS-MD3000 VCR
UP-D23MD color printer
1-4-8
HS-MD3000 VCR
UP-D55MD color printer
1-2-9
HS-MD3000 VCR
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 95 of 150
sd
Antares
96
Hardware Configuration
Peripheral Mounted on Peripheral
Shelf
1-2-9
SVO9500 VCR
CP770 or CP800 color printer
2-10
SVO9500 VCR
UP-D23MD color printer
1-4-8
SVO9500 VCR
UP-D55MD color printer
1-2-9
SVO9500 VCR
Antares
US03-201.812.01.02.02
07.05
Page 96 of 150
sd
Siemens
Medical Solutions USA, Inc.
Hardware Configuration
97
Complete the cable connections to I/O module from the B/W printer, color printer, or VCR.
The following figure shows the I/O module cable connections for all peripherals.
1. USB-A port (modem), J1
2. Left (Channel 1) audio in from VCR
3. Right (Channel 2) audio in from Right (Channel 2) audio in from VCR
4. S-Video In from VCR
5. Composite Video Out
6. GB/s video and limited COM 2
7. VCR control (RS-232 COM 1)
8. Ethernet (Network)
9. USB-A ports (printers): J5B (color printers)
10. USB-A ports (printers): J5A (black and white)
11. System status LEDs
12. Parallel port (printer)
13. Physio Trigger Out
14. S-Video Out to VCR
15. Right Audio Out to VCR
16. Left Audio Out to VCR
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 97 of 150
sd
Antares
98
Hardware Configuration
Antares
US03-201.812.01.02.02
07.05
Page 98 of 150
sd
Siemens
Medical Solutions USA, Inc.
99
This procedure assumes that the printers were installed on the system before installing
software. Software version 4.0 has installation and configuration requirements that differ
from previous software releases. There are two different installation and configuration procedures. Use the following table to identify which procedure to use.
Tab. 8
Printers
Manufacturer
Printer Type
SW v2.1
SW v3.0
SW v4.0
CP770DW
Mitsubishi
Image
CP800DW
Mitsubishi
Image
HP1150
HP
Report
HP2500n
HP
Report
HP4050
HP
HP4100
1, 2
Report
HP
Report
HP4200
HP
Report
HP6122
HP
Report
P91DW
Mitsubishi
Image
P93DW
Mitsubishi
Image
UPD23MD
Sony
Image
UPD55MD
Sony
Image
1. When installing multiple printers on a system with v4.0, install the CP800 first.
2. CP800 is not supported on systems configured with the RM-165 PCB.
Use the peripheral software setup instructions that match the software version running on
the Antares system.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Page 99 of 150
sd
Antares
100
When installing two printers and one of them is the CP800, install
the CP800 printer first.
CP800 Installation
NOTE
When installing two printers and one of the printers is the CP800,
install the CP800 first.
4. Press the Presets button. The system displays the Presets menu.
5. Select Service. The system displays the System Information, Service Options, Remote Access Options, and Video panes.
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 100 of 150
Medical Solutions USA, Inc.
sd
101
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
102
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 102 of 150
Medical Solutions USA, Inc.
sd
103
Select a USB port that has not yet been assigned to a printer (usually USB001, but could be USB002 or USB003). Once a USB port
has been assigned to a printer, it cannot be re-assigned to a different printer without first uninstalling and re-installing the USB port
resource.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
104
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 104 of 150
Medical Solutions USA, Inc.
sd
105
15. If the Use Existing Driver window displays, select Keep existing driver (recommended).
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
106
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 106 of 150
Medical Solutions USA, Inc.
sd
107
not
print,
reference
(Printer
24. If there are no more printers to install, continue to (Reboot System / p. 122).
Siemens
Medical Solutions USA, Inc.
When installing two printers and one of them is the CP800, install
the CP800 printer first.
US03-201.812.01.02.02
07.05
Antares
108
- For B/W printers: connect the printer USB cable to J5B on the I/O panel. Reference
(9/Fig. 108 / p. 97).
- For color printers: connect the printer USB cable to J5A on the I/O panel. Reference
(10/Fig. 108 / p. 97).
NOTE
The report printers automatically install and configure during the software installation. The
following steps allow the user to install a printer manually (if problems are encountered
during the software installation) or if the user is adding the printer to an existing, working
system.
1. Power on the Antares system.
2. Power on the printer.
Allow the Antares system to complete the boot sequence to imaging mode.
Wait for the printer to indicate it is ready (indicator light or display message).
3. Connect the printer USB cable to the I/O module.
- For B/W printers: connect the USB cable to J5A. Reference (10/Fig. 108 / p. 97).
- For color printers: connect the USB cable to J5B. Reference (9/Fig. 108 / p. 97).
The system checks for new hardware and displays the following message:
New Hardware has been detected.
Allow the new hardware installation to complete.
4. Register a patient in the Patient Registration screen.
a) Press the Patient key or select the Patient Registration icon. The system displays
the Patient Registration screen.
b) Enter patient information into the screen fields and select Ok.
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 108 of 150
Medical Solutions USA, Inc.
sd
109
5. Press the Report key or select the Report icon. The system displays the Patient Report screen.
6. Select Print Report. The system prints the report.
- Compare the printed report to the information on the screen.
- If the report did not print properly, see (Printer Installation Troubleshooting / p. 143)
and if necessary (Restoring Plug-and-Play Resources / p. 145).
7. If there are no more printers to install, continue to (Reboot System / p. 122).
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
110
SW v4.0
CP800 Installation
NOTE
NOTE
When installing two printers and one of them is the CP800, install
the CP800 printer first.
4. The system detects the printer and displays the Found New Hardware Wizard screen.
a) If not already selected, select the Install the software automatically (recommended) check box.
b) Select Next. The system installs the USB software automatically.
c) Verify the installation was successful.
The system displays the following screen when the installation was successful:
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 110 of 150
Medical Solutions USA, Inc.
sd
111
The system displays the following screen when the installation was not successful:
NOTE
d) Select Finish.
The Found New Software Wizard closes.
5. Press the Presets button. The system displays the Presets menu.
6. Select Service. The system displays the System Information, Service Options, Remote Access Options, and Video panes.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
112
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 112 of 150
Medical Solutions USA, Inc.
sd
113
10. If not already selected, select the Local printer attached to this computer check box.
Verify Automatically detect and install my Plug-and-play printer is not selected.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
114
Select a USB port that has not yet been assigned to a printer (usually USB001, but could be USB002 or USB003). Once a USB port
has been assigned to a printer, it cannot be re-assigned to a different printer without first uninstalling and re-installing the USB port
resource.
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 114 of 150
Medical Solutions USA, Inc.
sd
115
14. If not already selected, select the MITSUBISHI from the Manufacturer list,
CP800DW (LPT) from the Printers list and select Next. The Use Existing Driver window displays.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
116
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 116 of 150
Medical Solutions USA, Inc.
sd
117
17. Verify the name displayed in the Printer Name field is Mitsubishi CP800DW (LPT).
18. Select No for Do you want to use this printer as the default printer.
19. Select Next. The Print Test Page screen displays.
20. Select Yes for Do you want to print a test page.
21. Select Next. The Completing the Add Printer Wizard screen displays.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
118
23. Select Finish. A dialog box prompts to verify the Windows Test Page printed.
24. Verify the test page printed.
NOTE
If the test page did not print or If errors occurred during the installation, see (Printer Troubleshooting / p. 143).
When installing two printers and one of them is the CP800, install
the CP800 printer first.
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 118 of 150
Medical Solutions USA, Inc.
sd
119
- For B/W printers: connect the printer USB cable to J5A on the I/O panel. Reference
(10/Fig. 108 / p. 97).
- For color printers: connect the printer USB cable to J5B on the I/O panel. Reference
(9/Fig. 108 / p. 97).
NOTE
Select Finish.
Disconnect the Smart Cable from the J5A (P91DW) or J5B (CP770) I/O module.
Reconnect the Smart Cable to J5A (P91DW) or J5B (CP770) on the I/O module.
Repeat step 3.
If the first installation attempt of the USB printer will fail. Disconnect the printer(s) USB cable from the I/O panel and start the procedure from step 2.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
120
Select Service.
Select Configure Local Printer. The system displays the printer screen.
Right-click Mitsubishi P91DW (LPT) and select Properties.
Select the Ports tab.
Make sure Enable bidirectional support is not checked.
Select Ok.
NOTE
1. With the Antares system powered off, connect the parallel cable to J1 of the I/O module
parallel printer port and the back of the printer. See (Fig. 108 / p. 97) to locate the parallel printer port.
2. Power on the printer.
Allow the Antares system to complete the boot sequence to imaging mode.
Wait for the printer to indicate it is ready (indicator light or display message).
3. Power on the Antares system.
4. The system checks for new hardware and displays the following message:
New Hardware has been detected.
Allow the new hardware installation to complete.
5. Select Next.
Allow the installation to complete and then select Ok.
6. Register a patient in the Patient Registration screen.
a) Press the Patient key or select the Patient Registration icon. The system displays
the Patient Registration screen.
b) Enter patient information into the screen fields and select Ok.
7. Press the Report key or select the Report icon. The system displays the Patient Report screen.
8. Select Print Report. The system prints the report.
- Compare the printed report to the information on the screen.
- If the report did not print properly, see (Printer Installation Troubleshooting / p. 143)
and if necessary (Restoring Plug-and-Play Resources / p. 145).
9. If there are no more printers to install, continue to (Reboot System / p. 122).
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 120 of 150
Medical Solutions USA, Inc.
sd
121
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
122
Reboot System
Perform a system reboot only after completing the installation of all printers.
1. Reboot the system.
a) Press the Standby button.
b) Select Shut Down from the displayed menu.
c) Wait ten seconds after the system completely shuts down. Then press Standby
again to power on the system.
2. Continue to (Print/Store Setup / p. 123).
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 122 of 150
Medical Solutions USA, Inc.
sd
123
Print/Store Setup
After the DICOM store and print devices are configured and the system has been rebooted, it is necessary to configure the Print and Store key assignments. The Print/Store
Setup permits the following configuration of the Antares Print and Store keys:
Automatic transfer of static images, either at the End of exam or During exam.
Automatic transfer of static images, either at the End of exam or During exam.
Automatic print routing, based on the image type.
Storing images as Secondary Capture (global setting).
Burning graphics into pixel data (global setting).
The Print and Store keys on the control panel can be configured to do print only, print and
store, or store only, regardless of the graphic on the button. The graphics located at the
top of the print/store presets screen represent the buttons on the control panel, not the
function of the button.
Print Key
Pos. 2
Store Key
Perform the Restore Factory Printer Settings function when configuring print settings for
the first time. Perform the Restore Factory Settings function for configuring print and store
settings for the first time.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
124
Fig. 132: Presets > Service > Restore Printer Factory Defauts
Selecting Restore Factory Settings populates the different lists with the installed print
and store devices.
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 124 of 150
Medical Solutions USA, Inc.
sd
125
Configure the routing for different image types to specific printers and storage devices.
1. Display the Print/Store Screen.
a) Select the Presets icon.
b) Select Print/Store from the Presets menu.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
126
BW
Selecting a printer from the drop-down menu sends black and white
images to the printer selected.
Selecting a storage device from the drop-down menu sends black
and white images to the printer selected.
Color Doppler
Tinted
2D Ref
Misc.
Selecting a printer from the drop-down menu sends secondary captures (screen captures) images to the printer selected.
Selecting a storage device from thre drop-down menu sends secondary captures (screen captures) images to the printer selected.
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 126 of 150
Medical Solutions USA, Inc.
sd
127
Print key routing configuration for printing or storing specific image types
Pos. 2
Store key routing configuration for storing or printing specific image types
Pos. 3
Pos. 4
Images are always stored to the local database when either the
Print or Store key is pressed, regardless of the Print/Store configuration.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
128
Configure Print
After configuring a DICOM or local peripheral print device, assign image types to a destination printer, configure transfer types, and set the printed page composition.
1. Display the Print/Store Screen.
a) Select the Presets icon.
b) Select Print/Store from the Presets menu.
2. Select the Configure Printer tab.
Pos. 2
3. For each device in the Printer column (1/Fig. 135 / p. 128), make the following selections from the related drop-down menus:
-
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 128 of 150
Medical Solutions USA, Inc.
sd
129
Siemens recommends that transfer at the End of Exam be selected for archive servers. Transfer During Exam should only be selected for image review workstations that support In-Progress
review.
Servers configured for Storage Commitment should not be configured for transfer During Exam.
Auto transfer to CD-R is disabled by default.
NOTE
4. Select Printer or System for the Formatted by option (2/Fig. 135 / p. 128).
a) Printer: Assigns the DICOM printers to perform composition and formatting. This is
the default setting. This setting is typically more efficient, allowing smaller blocks of
data to be transferred over the network, and consuming less ultrasound system
CPU resources.
b) System: Assigns the ultrasound system to perform page composition and formatting. This provides uniformity across different printers, and in some cases results in
very large blocks of data being transferred to the printer.
5. After setting the print routing, select Save to save the settings. All changes are discarded unless the Save button is selected.
NOTE
If the Print or Store key is not configured for printing, then all the
drop-down menus under that key icon should be set to empty.
Configure Store
To configure local printers to the Store key, reference the (US03-103.843.01 / Configure
Store) section of System Configuration.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
130
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 130 of 150
Medical Solutions USA, Inc.
sd
131
VCR Configuration
VCR Controls
Verify the Operation of the VCR
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
132
HS-MD3000 Configuration
When installing a Mitsubishi HS-MD3000 VCR for use with an Antares, it is necessary to
optimize the presets for the best playback image quality. Select these presets using the
front panel display and controls on the VCR.
Power
Pos. 2
Monitor - Mix
Pos. 3
Audio - Mix
Pos. 4
Pos. 5
Stop
Pos. 6
Play
Pos. 7
Frame Field/Enter
Pos. 8
Menu
Pos. 9
Jog Dial
The configuration instructions describe how to set up the following preset menus on the
HS-MD3000 VCR:
Antares
<Menu 1>
<Menu 2>
<Menu 3>
<Menu 4>
Hidden Menu
US03-201.812.01.02.02
07.05
Siemens
Page 132 of 150
Medical Solutions USA, Inc.
sd
133
VCR Controls
Control
Function
Menu
To access the VCR preset menus, press the Menu button on the front
panel of the VCR. The first press of the Menu button displays
<MENU 1>. Each press of the Menu button changes the menu
sequentially to the next menu. The sequential order of menus is:
<MENU 1>, <MENU 2>, <MENU 3>, <MENU4>, <ELAPSED TIME>;
and exit the VCR menus.
Stop + Play
Jog/Adjust
Turn the JOG dial to the left or right to select an item within a menu.
Turn to the right for a higher numbered setting. Turned to the left for a
lower numbered setting.
Press the Enter button to edit a menu value. Then turn the Jog dial to
select a different value. Press Enter again to save the change.
Menu 1 Settings
NOTE
1. Press MENU for approximately two seconds. The VCR displays 101:00.
2. Turn the JOG dial to the left or right to select each value within <Menu 1>.
3. Verify that each menu item value displays correctly. The following table lists the correct
settings:
Tab. 11
<Menu 1>
Menu Item
Antares
Configuration
VCR Front
Panel Display
STOP
101:00
TAPE END
STOP
102:00
TAPE IN
STOP
103:00
TAPE LENGTH
PAL = T-120
104:00
NTSC = E-180
VISS
Siemens
Medical Solutions USA, Inc.
REC
US03-201.812.01.02.02
07.05
105:01
Antares
134
Antares
Configuration
VCR Front
Panel Display
OFF
106:00
OFF
107:00
FOOT SWITCH
L.E.
108:00
DIMMER
OFF
109:00
Press the ENTER button. The value after the colon (:) will flash.
Turn the JOG dial until the correct value displays.
Press the ENTER button to set.
Turn the JOG dial to the next item as needed to change menu item values.
When the last <Menu 1> item is displayed, press the MENU button. The VCR displays <Menu 2>.
<Menu 2>
Menu Item
Antares
Antares
Configuration
VCR Front
Panel Display
S-VHS/S-ET (NTSC)
ON/OFF
201:01
SVHS (PAL)
AUTO
Y-NR
ON
202:01
C-NR
ON
203:01
Y-ENHANCER
ON
204:01
TBC
ON
205:01
US03-201.812.01.02.02
07.05
Siemens
Page 134 of 150
Medical Solutions USA, Inc.
sd
Antares
135
Configuration
VCR Front
Panel Display
COLOR FILTER
OFF
206:00
FREEZE MODE
FRAME
207:00
HI-FI AUDIO
ON
208:01
NORMAL AUDIO IN
MIX
209:00
Press the ENTER button. The value after the colon (:) will flash.
Turn the JOG dial until the correct value displays.
Press the ENTER button to set.
Turn the JOG dial to the next item as needed to change menu item values.
When the last <Menu 2> item is displayed, press the MENU button. The VCR displays <Menu 3>.
When <Menu 3> is selected the VCR displays 301:00. <Menu 3> is used to set the clock;
it is not necessary to set the clock.Press the MENU button to continue to <Menu 4>.
Menu 4 Settings
NOTE
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
136
<Menu 4>
Menu Item
Antares
Configuration
VCR Front
Panel Display
TRANSMISSION RATE
19200
401:04
8 BITS
402:01
1 BIT
403:01
PARITY BIT
NONE
404:00
Press the ENTER button. The value after the colon (:) will flash.
Turn the JOG dial until the correct value displays.
Press the ENTER button to set the change and go to the next menu item.
When the last <Menu 4> item is displayed, press the MENU button twice to exit.
1. Simultaneously press and hold STOP and PLAY for at least ten seconds. The VCR displays PC: 2.
If PC: 2 is not displayed, turn the JOG dial until PC: 2 is displayed.
2. Press the ENTER button to set the change and continue to the next Hidden Menu item.
3. Verify that each menu item value displays correctly. The following table lists the correct
settings:
Tab. 14
Hidden Menu
Menu Item
Antares
Antares
Configuration
VCR Front
Panel Display
PC: 2
45
SP:45
Ud: 0
Lr: 0
US03-201.812.01.02.02
07.05
Siemens
Page 136 of 150
Medical Solutions USA, Inc.
sd
Antares
137
Configuration
VCR Front
Panel Display
CA: 0
HUE (CP)
CP: 0
Cd: 1
BRIGHT-NESS CONTROL
(bC)
-1
bC:-1
-1
CC:-1
OFF
bd:oF
Press the ENTER button. The value after the colon (:) will flash.
Turn the JOG dial until the correct value displays.
Press the ENTER button to set.
Turn the JOG dial to the next item as needed to change menu item values.
When the last item is displayed, press the Stop button to exit.
NOTE
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
138
SVO9500 Configuration
When installing a Sony SVO9500 VCR for use with an Antares, it is necessary to optimize
the presets for the best playback image quality. Select these presets using the front panel
display and controls on the VCR.
HiFi switch
Pos. 3
Program switch
Pos. 4
CTL/Menu switch
Pos. 5
Menu display
Pos. 6
Power button
Pos. 7
Pos. 8
Pos. 9
Play button
Tab. 15
Control
Setting
Function
Audio
Hi-Fi
Hi-Fi
ON
Program
OFF
CTL/Menu
CTL
Menu
ON
Power
Antares
Audio switch
Pos. 2
US03-201.812.01.02.02
07.05
Siemens
Page 138 of 150
Medical Solutions USA, Inc.
sd
139
Control
Setting
Function
Data
N.A.
Item
N.A.
Play
N.A.
VCR Front
Antares Option
Panel Display Setting
vIA-
ON
FOt--
LE
tPE-
Et
sVH-
At
Crn-
ON
Yrn-
ON
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
140
ON
FrE-
Fr
VCR Front
Antares Option
Panel Display Setting
Antares
SOP-
OFF
34P-
PA
FrS-
PA
1PH-
65
US03-201.812.01.02.02
07.05
Siemens
Page 140 of 150
Medical Solutions USA, Inc.
sd
141
VCR Front
Antares Option
Panel Display Setting
L1n-
ON
nOn-
OFF
FrA-
ON
H s-
OFF
v s-
OFF
Press the ENTER button. The value after the colon (:) will flash.
Turn the JOG dial until the correct value displays.
Press the ENTER button to set.
Turn the JOG dial to the next item as needed to change menu item values.
When the last item is displayed, press the Stop button to exit.
NOTE
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
142
4. Press the VCR key to toggle between the record and the pause functions.
The VCR key illuminates when the Record function is active.
5. Record the active screen to video tape. Use one of the following methods:
- Use the VCR panel controls to record the active screen.
- Use the VCR group box to control the VCR directly from the Image task card (if this
option is installed on the system).
6. Playback the recording. Use one of the following methods:
- Use the VCR panel controls to record the active screen.
- Use the VCR group box to control the VCR directly from the Image task card (if this
option is installed on the system).
The playback should display the active screen as viewed during the recording
period.
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 142 of 150
Medical Solutions USA, Inc.
sd
Printer Troubleshooting
143
6-
If the test page does not print, refer to the following list for possible causes:
Symptom
Cause
Solution
Plug-and-play printers
do not install when
connected to the system.
Plug-and-play
resources are used
and must be restored.
Unable to print.
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
144
Printer Troubleshooting
Symptom
Cause
Solution
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 144 of 150
Medical Solutions USA, Inc.
sd
Printer Troubleshooting
145
A driver can be installed for a plug-and-play resource only once. At which time Windows
marks that resource as used.(A used resource displays as a grayed-out device on the
hardware list.) If a printer is not recognized and cannot be installed, the resource may
need to be restored. The device must be deleted and the plug-and-play function re-initiated.
NOTE
NOTE
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
146
Printer Troubleshooting
b)
c)
d)
e)
f)
g)
h)
i)
Select Go.
Select Next.
Right-click My Computer, then select Properties > Hardware > Device Manager.
Select the View pull-down menu.
Select Show hidden devices and expand Printers.
Select the printer for deletion. Right-click the selected printer and select Uninstall.
Confirm deletion of printer.
Expand Universal Serial Bus controllers, then select the appropriate printing support for the printer to be deleted.
j) Confirm deletion of USB Printing Support.
Repeat steps 3 for each USB Printing Support device entry from the list in turn.
There is a USB Printing Support device for the USB to parallel cable, which is installed as a separate device. All USB Printing Support devices should be removed
to free up resources.
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 146 of 150
Medical Solutions USA, Inc.
sd
Printer Troubleshooting
147
Printers
Manufacturer
Use Instruction
HP
Mitsubishi
Mitsubishi
UPD23MD, UPD55
Sony
(Re-install Plug-and-Play
Printers / p. 148)
Report Printers
The following steps manual printer installation if problems are encountered during the
software install process or if add a printer to an existing, working system.
NOTE
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
148
Printer Troubleshooting
b) Select Next. The system installs the USB software automatically.
c) Select Finish.
The Found New Hardware Wizard closes and the system returns to imaging.
4. Press the Patient button. The system displays the Patient Registration screen.
5. Register a patient and select Ok.
6. Press the Report button. The system displays the Patient Report screen.
7. Select Print Report.
8. Compare the printed report to the information on the screen.
HP 4050, HP4100, HP4200
The Antares system must be powered off before beginning the printer installation.
1. Connect the parallel cable between the printer and the Antares:
a) Connect the parallel cable into the parallel printer connector on the I/O panel and
secure with the locking screws.
b) Route the cable through the peripheral cable passage.
c) Connect the other end of the parallel cable into the parallel connector on the back of
the printer and lock it down with the clips.
2. Power on the Report Printer on.
Wait until the printer displays Ready on the front panel.
The system detects the printer and displays the Found New Hardware Wizard
window.
a) If not already selected, select the Install the software automatically (recommended) check box.
b) Select Next. The system installs the USB software automatically.
c) Select Finish.
The Found New Hardware Wizard closes and the system returns to imaging.
3. Press the Patient button. The system displays the Patient Registration screen.
4. Register a patient and select Ok.
5. Press the Report button. The system displays the Patient Report screen.
6. Select Print Report.
7. Compare the printed report to the information on the screen.
The Plug-and-play procedure assumes the system is booted to imaging. Perform the following steps for each printer to install; one printer at a time.
1. Connect the AC power cable between the printer and the Antares peripheral outlet.
Antares
US03-201.812.01.02.02
07.05
Siemens
Page 148 of 150
Medical Solutions USA, Inc.
sd
Printer Troubleshooting
149
Siemens
Medical Solutions USA, Inc.
US03-201.812.01.02.02
07.05
Antares
150
Antares
Printer Troubleshooting
US03-201.812.01.02.02
07.05
Siemens
Page 150 of 150
Medical Solutions USA, Inc.
sd