Documentos de Académico
Documentos de Profesional
Documentos de Cultura
DOI 10.1007/s10406-005-0166-x
Springer-Verlag 2005
Dominik Fleischmann
D. Flesichmann (
)
Department of Radiology
Stanford University Medical Center
300 Pasteur Drive, 5-072
Stanford, CA 94305-5105
United States
E-mail: d.fleischmann@stanford.edu
Tel.: 650-7237647
Fax: 650-7257296
Multiple detector-row CT (MDCT) technology continues to rapidly evolve and, from generation to generation, spatial resolution has gradually improved. With
state of the art 16-channel MDCT systems, routine submillimetre resolution data acquisition has become possible, allowing CT angiography to replace diagnostic conventional angiography for many clinical applications. At
the same time, CT acquisition speed has dramatically increased. For example the scan times of an abdominal
CT angiography (CTA) have decreased from 20 s to less
than 5 s when comparing 4-, 8-, 16- and now 64-channel
systems. While fast acquisitions at unprecedented spatial resolution offer new opportunities in CTA, contrast
material (CM) delivery becomes more difficult and less
forgiving at the same time. When one is fortunate
enough to be faced with the problem of designing new
injection protocols for CTA with a newly installed fast
CT scanner, it is necessary to integrate the scanner capabilities with the fundamentals of arterial enhancement, which are not entirely intuitive.
Several traditional concepts of CM injection for
CTA which were conceived empirically for single-de-
E61
tector row CT, are not valid for faster scan times. The
most common misconception regarding arterial enhancement is the assumption that a constant-rate intravenous injection of CM leads to an enhancement
plateau. In addition, the ball-park rule of choosing an
injection duration equal to the scanning time works
well for long scan times (20 s or so), but obviously cannot be used for a 4 s or shorter acquisition time. Finally,
the appealing idea that faster scans will naturally result
in better arterial enhancement is not necessarily true either, the opposite is actually the case; faster scanners
tend to result in lower arterial opacification if one applies traditional injection concepts.
So how can one design CTA injection protocols for a
fast scanner; that is, how can one make sure that adequate arterial opacification is achieved within a much
shorter time window of CT angiographic acquisition?
The key is knowledge of the physiological and pharmacokinetic principles of arterial enhancement, knowledge
of the effects of user-adjustable injection parameters
and the iodine concentration of the contrast medium.
The prerequisites
Early arterial contrast medium dynamics
There are essentially two controllable parameters
which, for a given individual, determine the time-course
of arterial enhancement [1]: these are the iodine administration rate (iodine flux), and the injection duration
(Fig. 1). The iodine administration rate is directly proportional to arterial enhancement and can be controlled
by the injection flow rate and the iodine concentration
of the CM. Thus, an increase in the injection rate and a
higher iodine concentration of the contrast medium directly translate into increased vascular enhancement.
Secondly, arterial enhancement continuously increases
over time with longer injection durations. This relationship is less intuitive but can be illustrated graphically
when considering a long injection (e.g 32 s) as the sum
of 8 smaller injections of 4 s duration (Fig. 1). Note that
the overall enhancement effect can be drawn as the integral (time-sum) of eight separate smaller enhancement curves, which overlap in time. It is obvious from
Fig. 1 Relationship between contrast medium injection and arterial enhancement. The simple
additive model illustrates the effects of the injection flow rate (iodine flux) and injection duration on arterial enhancement. (A) Intravenous
contrast medium injection causes (B) an arterial
enhancement response, which consists of an early
first pass peak and a lower recirculation effect. (C) Doubling the injection flow rate (doubling the iodine administration rate) results in
(D) approximately twice the arterial enhancement. (E) The effect of the injection duration can
be regarded as (F) the sum (time integral) of several enhancement responses. Note that due to the
asymmetric shape of the test enhancement curve
and to recirculation effects, arterial enhancement
following an injection of 128 ml (the time integral of eight consecutive injections of 16 ml) increases continuously over time
E62
E63
not as Injection volume and injection rate. The injection volume is not an important parameter in the setting
of CTA, in contrast to parenchymal imaging, which has
fundamentally different kinetics.
The initial and most important step in the design of
injection protocols for CTA is then to consider the anticipated scan-time.
Injection strategies for slow acquisitions
Slow acquisitions refer to scanning durations in the range
of 20 s or longer. Such scan times are common with 4channel MDCT, notably with large volume coverage
(chest abdomen pelvis CTA), and even with 16 channel MDCT when EKG-gating is used. For slow acquisitions, traditional protocols yield reliable arterial enhancement. For example, one can use a standardized injection (e.g. 1.2-1.5 g of Iodine/s for an average individual
with approximately 75 kg of BW) for the duration of the
scan time. For example, for a 25 s scan, one might inject
at a flow rate of 4 mL/s for 25 s (i.e., a total volume of 100
mL). The scanning delay is set to the patients tCMT, as established by a test-bolus injection or by automated bolus
triggering. Higher (5-6 ml/s) or lower (3 mL/s) flow rates
should be used in larger (>90 kg) and smaller (<60 kg)
patients, respectively. With long scan times, biphasic or
multiphasic injections lead to more uniform arterial enhancement over time [8-10].
Injection Strategies for fast acquisitions
For fast acquisitions, the injection parameters and the
choice of the scanning delay need to be adapted in order to achieve adequate arterial enhancement. Based
on the first two rules (1 and 2) of early contrast medium
dynamics, one can apply two strategies to increase arterial enhancement for fast acquisitions. Alone or in combination, as illustrated in Figure 2: one can (1) increase
the iodine flux (grams of Iodine injected per unit of
time), which will translate into a proportionally stronger
enhancement. The iodine flux can be increased by the
selection of a higher iodine concentration of the contrast medium and/or by increasing the injection flow
rate (Fig. 1). In addition (2) one can increase the scanning delay relative to a patients tCMT. This will also increase arterial enhancement because of the continuing
rise of opacification with longer injection duration
(Fig. 1). It is crucial, however, to also increase the injection duration in this case, otherwise the arterial enhancement would not be maintained long enough. Both
strategies, increased injection rate and increased injection duration with increased scanning delay, are used to
Fig. 2 Strategies to improve arterial enhancement for fast MDCTA. Two strategies to increase arterial enhancement when compared
to a baseline 16 s injection at 4 mL/s (upper
left panel) can be employed; either alone or in
combination. Increasing the injection rate
from 4 to 5 mL/s increases the enhancement
approximately 20% (lower left panel). Increasing the iodine concentration from 300 to
400 mg/mL would achieve an even greater enhancement increase without the need to increase the injection rate. Alternatively, one
can also increase the injection duration and
the scanning delay taking advantage of the
fact that enhancement increases with longer
injection durations (right upper panel). Maximum enhancement can be achieved when
both the injection rate (and or the iodine concentration) and the injection duration are increased (right lower panel) simultaneously
E64
maximum vessel opacification. It is not necessary to operate such powerful scanners at maximum acquisition
speed. There is no advantage to using a 4 s versus a 10 s
acquisition for CT angiographic applications. Maximum
speed may even be detrimental because a very fast acquisition may not allow complete and sufficient opacification of a diseased arterial tree. This has been observed
in mesenteric CTA, in peripheral (lower extremity)
CTA [11], and even in abdominal CTA. Figure 4 shows
the case of a patient with an infrarenal abdominal aortic
aneurysm. Since no diagnostic delay was used, and because the scanner was operated with fast gantry rotation
E65
When scan times are constant, one can also use constant
injection durations for all patients. Flow rates can then
easily be individualised for patient weight (Table 1).
When automated tube current modulation is used, this
strategy results in very simple, robust CTA protocols,
which result in good arterial enhancement and similar
image noise across a wider range of patient physiques.
References
1. Fleischmann D (2003) Use of highconcentration contrast media in
multiple-detector-row CT: principles
and rationale. Eur Radiol 13 [Suppl
5]:M14-20
2. Hittmair K, Fleischmann D (2001)
Accuracy of predicting and controlling
time-dependent aortic enhancement
from a test bolus injection. J Comput
Assist Tomogr 25(2):287-294
3. Sheiman RG, Raptopoulos V, Caruso
P et al (1996) Comparison of tailored
and empiric scan delays for CT
angiography of the abdomen. AJR
Am J Roentgenol 167(3):725-729
4. Bae KT, Heiken JP, Brink JA (1998)
Aortic and hepatic contrast medium
enhancement at CT. Part II. Effect of
reduced cardiac output in a porcine
model. Radiology 207(3):657-662
Copyright of European Radiology Supplement is the property of Springer Science & Business Media B.V. and
its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's
express written permission. However, users may print, download, or email articles for individual use.
Copyright of European Radiology is the property of Springer Science & Business Media B.V. and its content
may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express
written permission. However, users may print, download, or email articles for individual use.