Documentos de Académico
Documentos de Profesional
Documentos de Cultura
performance on Images
Jigar Patel
Dept. of Electrical and Computer Engineering
Florida State University (FSU)
Tallahassee, FL, USA
jp09x@my.fsu.edu
I. INTRODUCTION
Image registration, also known as image fusion,
matching or warping, can be defined as the process of aligning
two or more images together. The goal is to find an optimal
transformation that aligns the structures in the input images.
Image acquisition methods include but are not limited to images
from different times, sensors, viewpoints, or modalities. Once
the images are acquired the images are then aligned/registered
by setting one image as the reference image and setting all
subsequent images as the sensed images. Registration is
especially important in Medical Image Analysis as
doctors/radiologists usually want to analyze a patients scans
from multiple imaging equipment. Imaging in the medical field
is done on patients using a wide variety of equipment such as
computer tomography (CT), X-Ray, positron emission
tomography (PET), magnetic resonance imaging (MRI) to
name a few.
The scope of this paper is to do a simple comparison
of how ICP performs on different images. Points are only one
approach to feature selection, other techniques that are not
discussed are lines, regions, and templates. Points are more
desirable compared to the other techniques. It is not in the scope
of this paper to propose a novel approach or modification to the
existing algorithms.
The paper is organized as follows, Section II
introduces the ICP algorithm in detail followed by section III
which provides pseudo-implementations of the algorithm. In
section IV experiment results are discussed followed by the
conclusion in section V.
Figure 1: (a) Transaxial CT with marker (b) Transaxial MR with markers [1]
Step 5:
Determine the transformation based on the list
created in step 3.2 using least-squares method and
mat set Q to set P and determine the
correspondence again as in step 3.
Step 5 is added into the algorithm to reduce the effect of any
noise that might exist in the image. The value of D in step 3.2
is an arbitrary value which is user defined. This process holds
for both 2-D and 3-D images with a minor change. Instead of
six parameters (a-f) the 3-D version requires computation of 12
parameters (a-l) which are given the equation 2 below.
= + + +
= + + + } . 2
= + + +
Also instead of using three non-collinear point pairs four noncoplanar point pairs are used to calculate the parameters (a-l) of
transformation of equation 2.
IV. EXPERIMENTAL RESULTS
ICP is performed on the following two images shown below
along with the histograms of each image.
Figure 2: Original Images and their histograms: Lena (top) Mammogram (bottom)
Cropped
Image
1
0 0
[0
1 0]
59 89 1
1.011 0.016 0
[ 0.016
1.011 0]
57.528 90.023 1
The matrix above for each image shows the values obtained for the
translation matrix. As seen in row three of each matrix the equation obtained
is fairly close to the actual values from Table 1 where the cropped image
started at point (60,90) and the window size was [128,96] which is the size
of the registered image.
Translation Image
1
0
0
[ 0
1
0]
32 40 1
1
0
0
[ 0
1
0]
32 40 1
V. CONCLUSION
Rotation
Image
0.710
0.714 0
[ 0.714
0.710 0]
132.18 129.26 1
0.886 1.120 0
[1.120 0.886 0]
432.71 313.21 1
The matrix above for each image shows the values obtained for the rotation
matrix. This matrix is harder to compare to the actual rotation value from
Table 1 however visually it is noticed that the Lena image is correctly
registered which the Mammogram image is not registered correctly. This
shows that fewer points leads to errors in registration. The reference and
sensed Mammogram images only had three matched points which is not
sufficient to perform the registration correctly in this case.
[4]
[5]