Documentos de Académico
Documentos de Profesional
Documentos de Cultura
IN THIS ISSUE:
Electrical Units in the New SI:
Saying Goodbye to the 1990 Values
Evaluation of Proficiency Testing Results
with a Drifting Artifact
Calibration of Ultrasonic Flaw Detectors
An Uncertainty Model and Analyzer
for a Space Environmental Test Facility
SMART!
...with AssetSmart
Test Equipment
Management Software.
By PMSC
2800 28th Street, Santa Monica, California 90405 USA 800.755.3968 info@assetsmart.com www.assetsmart.com
CONTENTS
SPECIAL FEATURE
22 Software Analysis and Protection for Smart Metering
Charles B. do Prado, Davidson R. Boccardo, Raphael C. S. Machado,
Luiz F. R. da Costa Carmo, Tiago M. do Nascimento, Lucila M. S. Bento,
Rafael O. Costa, Cristiano G. de Castro, Srgio M. Cmara, Luci Pirmez,
and Renato Oliveira
TECHNICAL PAPERS
30 Electrical Units in the New SI: Saying Goodbye to the 1990 Values
Nick Fletcher, Gert Rietveld, James Olthoff, Ilya Budovsky,
and Martin Milton
36 Realization and Dissemination of the International Temperature Scale of
1990 (ITS-90) above 962C
Andrew D. W. Todd and Donald J. Woods
42 Evaluation of Proficiency Testing Results with a Drifting Artifact
Chen-Yun Hung, Pin-Hao Wang, and Cheng-Yen Fang
48 A 40 GHz Air-Dielectric Cavity Oscillator with Low Phase Modulation Noise
Archita Hati, Craig W. Nelson, Bill Riddle, and David A. Howe
56 A Calibration System for Reference Radiosondes that Meets GRUAN
Uncertainty Requirements
Hannu Sairanen, Martti Heinonen, Richard Hgstrm, Antti Lakka,
and Heikki Kajastie
62 Calibration of Ultrasonic Flaw Detectors
Samuel C. K. Ko, Aaron Y. K. Yan, and Hing-wah Li
70 An Uncertainty Model and Analyzer for a Space Environmental Test Facility
Mihaela Fulop
DEPARTMENTS
3
4
16
80
NCSLI Measure (ISSN #19315775) is a metrology journal published by NCSL International (NCSLI).
The journals primary audience consists of practitioners and researchers in the field of metrology,
including laboratory managers, scientists, engineers, statisticians, and technicians. NCSLI
Measure provides NCSLI members with practical and up-to-date information on calibration
techniques, uncertainty analysis, measurement standards, laboratory accreditation, and quality
processes, as well as metrology review articles. Each issue contains peer reviewed technical
papers, technical notes, national metrology institute news, and other metrology information.
Author instructions are available at www.ncsli.org. If you are interested in purchasing advertising,
please visit www.ncsli.org for more information.
Managing Editor:
Michael Lombardi, National Institute of Standards and Technology (NIST), USA,
lombardi@ncsli.org
Associate Editors:
Jeff Gust, Fluke Corporation, jeff.gust@flukecal.com
Dr. Klaus Jaeger, Jaeger Enterprises, jaegerenterprises@comcast.net
Dr. Leslie R. Pendrill, SP Technical Research Institute of Sweden, leslie.pendrill@sp.se
Dr. James Salsbury, Mitutoyo Corporation, jim.salsbury@mitutoyo.com
Dr. Alan Steele, National Research Council of Canada, alan.steele@nrc-cnrc.gc.ca
Advertising Sales:
Linda Stone, NCSL International, 2995 Wilderness Place, Suite 107, Boulder, CO
80301-5404 USA, lstone@ncsli.org
Michael Lombardi
Managing Editor
lombardi@ncsli.org
HOW TO REACH US
NCSLI Measure, 2995 Wilderness Place, Suite 107, Boulder, CO 80301-5404 USA
www.ncsli.org measure@ncsli.org
NMI NEWS
New NPL Strategic Partnership with
Strathclyde and Surrey
Following a formal competitive process, the universities of Strathclyde
and Surrey have been selected to
develop a strategic partnership with
the government and the National Physical Laboratory (NPL) of the
United Kingdom. This new partnership will help to provide future
leadership of NPL.
The partnership will strengthen both fundamental research and
engagement with business by applying measurement science to
support innovation and growth. The goals are to:
Bring greater expertise and intellectual flexibility to strengthen
the laboratorys science;
Make better use of the existing facilities by strengthening the
laboratorys links with its academic partners, through new and
existing collaborations with academia and industry;
Encourage greater interaction with business, driven by closer
integration of existing innovation infrastructure and commercial
activity;
Make better use of the site at Teddington by granting partners
access to our spare capacity; and
Allow for the formation of a dedicated applied science
postgraduate institute.
The strategic partnership offers exciting prospects to enhance the
reach and impact of NPLs science and commercial activities. NPL will
continue to work with a wide range of academic and industrial partners
both across the UK and internationally. In the new arrangement, the
Department for Business, Innovation and Skills will own the operating
company, NPL Management Ltd. Currently, NPL has been operated
under a government-owned contractor-operated arrangement.
in three-phase high voltage lines, at 110 kV and 150 kV, with currents
up to 5 kA.
After calibrating the individual components in the VSL reference
system, an overall validation was performed at the National Research
Council (NRC) Canada (see photograph). While an agreement of
around 100 parts per million (ppm) was expected, the actual agreement between the VSL and NRC systems was better than 25 ppm at a
power level of 200 MW. Based on these results, a VSL system uncertainty of better than 300 ppm is estimated for actual on-site measurements three times better than the original goal.
The VSL system has been ready for on-site measurements since early
2014. This is just in time for power plant owners and large electricity
consumers in the heavy industry that have already contacted VSL for
on-site verification of their revenue metering systems.
For more information, contact Gert Rietveld: grietveld@vsl.nl
NMI NEWS
Biology/DNA
Chemistry/Instrumental Analysis
Crime Scene/Death Investigation
Information Technology/Multimedia
Physics/Pattern
NMI NEWS
Configuration of the QHE device showing dimensions. The bluegray rectangle in the center is the open face of the Hall bar. The
locations of graphene components are outlined by white lines.
The source and drain are at the left and right ends of the bar,
while electrical contacts are both above and below the bar.
www.ncsli.org
NMI NEWS
vapor deposition on copper or other material; and growth on an insulating silicon carbide substrate, which the PML researchers employ.
The NIST fabrication method involves coating a sheet of graphene
on a section of silicon carbide wafer with about 15 nanometers of
gold before any lithography. Patterns are developed using traditional
photolithography to remove any unwanted gold-coated graphene.
Then, the areas that will be the Hall bar contacts get a thicker coating
of gold, so that they will make good connections for wires used in
electrical measurements. In the last step, the gold layer over the area
of graphene that will serve as the Hall bar is removed with dilute aqua
regia, a mixture of nitric acid, hydrochloric acid, and deionized water,
leaving the graphene almost completely clean.
The aqua regia etching produces helpful p-type doping in the
graphene. Thus molecules from the acids remain on the surface,
reducing the carrier density and improving the mobility of electrons
that remain. Low carrier density is important because the higher the
density of charge-carriers in the Hall bar, the higher the magnetic field
strength required to observe the critical QHE plateaus.
The new devices have carrier densities in the range of 3 1010
per cm2 to 3 1011 per cm2, allowing observation of clearly defined
resistance quantization at magnetic field strengths of less than 4 T. The
p-type molecular doping effect can be reduced by heating in argon
gas, and is restored by dipping in aqua regia.
or more information, contact Rand Elmquist at:
F
randolph.elmquist@nist.gov
C
ROTRONI
TION
APPLICA
NOTE
b Competent?
ty Calibration La ward questions.
Is Your Humidi
the
Question 1 Is
to the ISO
lab accredited
?
17025 standard
17025
to the ISO/IEC
Accreditation
etency and teststandard for comp
is a really good
ing calibration
ditaIn general, accre
place to start.
tion does
not specifically say
how the
calibration
should be
performed,
but rather
it speaks
to a set of
guidance
principles
and results
that the
approved
EC
ld provide. ISO/I
shou
dures
proce
on confirms that
17025 accreditati
suclaboratory has
the calibration
r
ssed five majo
cessfully addre
calibration
areas within the
cies,
e of Competen
process; Scop
,
ed Procedures
Properly Defin
Control of the
Demonstrated
Underated
onstr
Process, Dem
urement
Meas
the
of
standing
onstrated ProfiUncertainty, Dem
urement. The
ciency in the Meas
ditation ensures
ISO 17025 accre
has
ation vendor
that your calibr
and
management
given serious
ERTAINTY
ROTRONIC UNC
eter or
Measured Param
ated
Device Calibr
RTAINTY
HUMIDITY UNCE
ent
urem
Humidity Meas
dity
Relative Humi
BUDGET
Range
Uncertainty (k=2)
MIC
THERMODYNA
% RH
0.1 % RH to 0.8
11.5 % RH
10.5 % RH to
% RH
34 % RH to 36
% RH
79 % RH to 81
UNCERTAINTY
TEMPERATURE
21 C to 27 C
Probe
Thermometer
0.21 % RH
0.22 % RH
0.29 % RH
0.49 % RH
8 Critical Questions
0.065 C
sales@rotronic-usa.com
Discover the
Blue Box
Difference
8000B AutomAted Precision
VoltAge meAsurement system
1200V Range
Bipolar Measurements
Self Calibration
www.mintl.com
NMI NEWS
HIGH ACCURACY
LOW TEMPERATURE COEFFICIENT
STABLE OVER TIME
OPTIONAL TEMPERATURE SENSOR
INCLUDES ACCREDITED CALIBRATION
CALIBRATION SERVICE TO 1000 A
MODEL
ACCURACY
MODEL
ACCURACY
CS-0.1
CS-1
CS-5
CS-10
CS-20
CS-50
< 0.005%
< 0.005%
< 0.01%
< 0.01%
< 0.01%
< 0.01%
CS-100
CS-200
CS-300
CS-500
CS-1000
MCS
< 0.01%
< 0.02%
< 0.05%
< 0.02%
< 0.05%
MULTIPLE
STANDARD
SEE
WWW.OHM-LABS.COM FOR
DETAILS
WWW.OHM-LABS.COM
Vol. 9 No. 3 September 2014
NMI NEWS
10
NMI NEWS
Super Micrometer
Gage Blocks
Dead Weight Tester
Torque Standards
Force Standards
Two-Pressure Humidity
Standard
Height Gages
Thread Set Plugs
Measuring Wires
Plain Plug & Ring Gages
Accredited to ISO/IEC 17025:2005 by NVLAP (Lab Code 200972-0). Over 35 metrology technicians on sta.
www.esscolab.com/capabili es/standards
800.325.2201 Ask about online repor ng via EsscoNet
Vol. 9 No. 3 September 2014
11
NMI NEWS
120,000 lbf
Morehouse Dead
Weight Machine
2,250,000 lbf
Morehouse Universal
Calibrating Machine
Torque Calibrations accurate to 0.002% of applied torque to 2,000 N-m also available
NMI NEWS
Three phases,
one calibrator
Calibrating three-phase
power has never been
this simpleuntil now.
The Fluke Calibration 6003A Three Phase
Electrical Power Calibrator delivers three
independent power phases in a single, costeffective instrument that you can use in the
lab, or wheel on a cart into the factory. Use it
to calibrate power quality analyzers, energy
meters, and energy loggers, as well as power
quality instrumentation and transducers.
www.flukecal.com/6003A_simplify
Fluke Calibration. Precision, performance, confidence.
Electrical
RF Temperature
Pressure
Flow
Software
13
TECHNICAL
EXCHANGE
TRAINING
PROGRAM
An Introduction to Instrument
Control and Calibration
Automation in LabVIEW
Statistical Analysis of
Metrology Data
Process Calibration
Prociency Testing
Dimensional Metrology
Basic Electronics
Temperature Measurement
Introduction to Measurement
Uncertainty
Dilip Shah, E = mc3 Solutions
Raleigh Marriott
Crabtree Valley Hotel
REGISTER TODAY
1-888-236-2427
HALF DAY
ONE DAY
TWO DAY
NCSLI MEMBER
NON-MEMBER
$180
$360
$720
$205
$410
$820
EXHIBITOR REGISTRATION
$800
NCSL International
ncsli.org
15
METROLOGY NEWS
Japan Unveils Plans for CentimeterResolution GPS
METROLOGY NEWS
Mini/Mobile Two-Pressure
Humidity Generator Self contained
humidity calibration standard
NIST traceable certificate of calibration
certified at 0.5%RH uncertainty*.
Humidity Parameter
Volume ratio, V (PPM)
Model 1200
Range
Model 2500
Uncertainty*
0.1 to 3 PPM
3 to 200 PPM
200 to 400000 PPM
-90 to -70 C
4.0% of value
2.0% of value
0.1% of value
0.2 C
-70 to -20 C
-20 to 70 C
0% to 99%
0.1 C
0.05 C
0.3% of reading
Model 3900
www.thunderscientific.com
sales@thunderscientific.com
17
18
METROLOGY NEWS
19
measurement
science
and the
of
life
NCSL
| 2995 J.
Wilderness
Place, Suite 107 | Boulder, CO 80301 | ncsli.org | Phone (303) 440-3339 | Email info@ncsli.org
20 International
| NCSLI Measure
Meas. Sci. www.ncsli.org
2015
......................................................................................
...................................................................................................
Submissions can relate to a variety of measurement and process topics including research and development,
manufacturing and service related fields, new test and measurement techniques, measurement standards and
traceability, statistical process and evaluation, measurement accuracy and uncertainty analysis, laboratory
management and accreditation and new advances in measurement science. Topics can include:
Acceleration
Automation
Chemical
Dimensional
Electrical
Force
Flow
Fundamental Units
Inspection
Humidity
Lab Accreditation
Management Issues
Mass
Optical
Pressure
Quality Topics
RF/Microwave
Standards
Temperature
Time & Frequency
Vacuum
Other Measurement
Topics
EXHIBIT SALES
Vol.www.ncsli.org
9 No. 3 September 2014
exhibits@ncsli.org
SPONSORSHIP PROGRAMS
larcher@ncsli.org
ADVERTISING OPPORTUNITIES
NCSLI
Measure J. Meas. Sci. |
lstone@ncsli.org
21
SPECIAL FEATURE
Abstract: Smart meters are devices equipped with embedded software that are capable of processing complex digital data. These
devices are now a reality in most areas of metrology. They enable a number of new applications, but they also introduce new challenges with regards to their validation. This paper describes initiatives developed by Inmetro, the national metrology institute of
Brazil, and Eletrobrs, a Brazilian power utility company, that are designed to support the validation of smart electricity meters.
1. Introduction
Currently, this scenario has been investigated in the area of cyber security applied to
Smart Grids [4]. Cybersecurity, to which we
refer in this context, refers to the application
of methodologies that make it impossible (or
almost impossible) for an unauthorized person to have access to any service or information inherent in a smart meter, thus protecting
the meter from being tampered with or used
for improper purposes.
Therefore Inmetro, in partnership with
the Eletrobrs Distributor of Rondnia (CERON), has been developing a research project
titled Cyber Security in Smart Metering. In
the present work, we describe the initiatives
to achieve better smart meter validation and
present the following research avenues:
Software analysis, including methodologies to discover vulnerabilities and
how we address the software traceability, i.e., the correspondence of source
code with its compiled version;
Software protection, including methodologies to perform embedded software integrity verification, guaranteeing that the software embedded in a
smart meter corresponds to a version
that was previously validated by the
manufacturer and authorities; software
obfuscation, making the software code
harder to understand to protect against
reverse engineering; and software fingerprinting to discourage code leakage
through the insertion of an identifier
SPECIAL FEATURE
into the software code with the goal of
making it traceable later; and
A measurement confidence chain, including methodologies that allow end
users to be confident that measurements have not been tampered with.
Authors
Charles B. do Prado
cbprado@inmetro.gov.br
Davidson R. Boccardo
drboccardo@inmetro.gov.br
Raphael C. S. Machado
rcmachado@inmetro.gov.br
lfrust@inmetro.gov.br
Tiago M. do Nascimento
1,2
tmnascimento@inmetro.gov.br
Lucila M. S. Bento
1,2
lmbento-eletrobras@inmetro.gov.br
Rafael O. Costa
1,2
rocosta@inmetro.gov.br
Cristiano G. de Castro
cgcastro@inmetro.gov.br
Srgio M. Cmara
1,2
smcamara@inmetro.gov.br
Luci Pirmez
luci@nce.ufrj.br
Renato Oliveira
renato.oliveira@ceron.com.br
1
Inmetro
National Institute of Metrology,
Quality and Technology
Santa Alexandrina St, 416
Rio Comprido - Rio de Janeiro, RJ,
Brazil, 20261-232
Federal University of
Rio de Janeiro (UFRJ)
Av. Pedro Calmon, 550
Cidade Universitria, Rio de Janeiro,
RJ, Brazil, 21941-901
2. Software Analysis
In Brazil, smart meters are regulated by Inmetro, which specifies the set of requirements
and also conducts the type approval and evaluation procedures. The approval process involves the following validations:
Validation of the smart meter architecture and its operations;
Validation of the legally relevant software, i.e., the software that can potentially change measurement information; and
Validation of software protection
mechanisms, which verifies how sensitive information is protected and how
the software integrity is guaranteed.
These aspects are handled in Section 3.
The validation of the smart meter architecture and its operations is based on documentation analysis. The objective is to understand
the smart meter algorithms and protocols,
to identify the communication interfaces, to
check the whole set of commands, to identify
the legally relevant software, and to evaluate
the test cases.
The validation of legally relevant software
is done with by white box analysis, and thus
requires the source code. The legally relevant variables are tracked to verify if they
are correctly manipulated, the common vulnerabilities are scanned based on the CWE
and CERT-C standards, and the consistency
of the commands, protocols and algorithms
with the documentation is reviewed. However, when a source assessment is performed,
the following question arises: How do we ensure that the executable code embedded in an
electricity meter was actually generated from
the source code provided by the manufacturer
and previously evaluated by Inmetro?
This problem is called software traceability.
The goal is to verify if the compiled executable
code corresponds to the source code (which
can be written in any programming language).
For this problem, two immediate approaches
are possible, but may not be practical.
A simple and direct way to test for software traceability is to reproduce the software
development environment and to compile
the approved source code, verifying whether
the generated binary code is as expected. It
can be complex and expensive, however, to
maintain several software development environments. Another way to test for software
traceability is to audit the environment of the
software developer. This step is performed as
NCSLI Measure J. Meas. Sci. |
23
SPECIAL FEATURE
the final procedure of the type approval process. Here, after the evaluation of the source
code by Inmetro, it is compiled in the environment of the software developer and embedded into the smart meter. However, such
an approach is likely to be ineffective when
dealing with a malicious developer.
To identify a malicious developer, the strategy involves verifying whether two program
codes written in different languages (typically
source code and executable code) exhibit the
same behavior. This strategy uses an artificial
neural network, fed with properties collected
from two program codes written in different
languages in order to discover their degree of
similarity. Preliminary results using artificial
neural networks with trivial properties, such
as the number of edges of the control flow
graphs, show that a strong correlation exists
between the source code and executable code
[7-9].
Another avenue of research is based on
black-box analysis and smart meter reverse
engineering and involves the identification
of sections of the firmware that can inappropriately change the behavior of the meter instrument. Here, we assume that is possible to
dump the firmware embedded into the smart
meter and interact with it through JTAG
(Joint Test Action Group) pins. Through the
dump, it is possible to obtain the entire firmware code and all of its execution pathways.
By debugging output with specific input (according to the manufacturers documenta24
SPECIAL FEATURE
ment. These tools allow software to be more efficient and as safe and
secure as possible. However, these same advances have increased the
capacity for reverse engineering with the goal of discovering vulnerabilities. For example, before an attacker exploits vulnerabilities in
a system, they will first have to identify them. Similarly, to change
the code embedded in a device, the attacker first has to analyze how
to tamper with the code without affecting the system functionality or
without triggering an anti-tampering mechanism.
Software protection techniques aim to hinder reverse engineering
and tampering with embedded software, to ensure that it will execute
as expected and to track possible illegal distribution [10]. The protection techniques that can be used for these purposes are: software
obfuscation, tamper proofing, and watermark. Software obfuscation
techniques hinder reverse engineering by syntactic changes that make
it harder to understand the code, but that do not alter its original behavior. Software integrity verification techniques ensure that the software performs as expected, even if tampering attempts are made. Finally, fingerprinting techniques, a special type of watermarking, refers
to the act of embedding a unique identifier in an object with the aim of
making it traceable later.
3.1 Software Obfuscation
Clearly, security should not be based on obscurity. However, there are cases
where secret/private keys are stored in the same memory as the software
binary code. Hence, granting public access to the binary code would also
grant public access to those secret/private keys.
25
SPECIAL FEATURE
SPECIAL FEATURE
to store each possible answer that is, for each possible key in the
malicious software and to simply return the desired answer. Such an
approach is clearly unpractical, as long as any standard MAC has a
key space of at least 2128 keys.
The second strategy that malicious software could use to trick our
verification process would be to keep a copy of the original program and perform all MAC calculations on the copy (instead of on
the tampered program) returning, then, the expected answer. The first
and simplest countermeasure to this strategy is to assure that no extra
space is left for malicious software by keeping a copy of the original program in the memory code. This can be easily done by simply
filling the unused program memory with a random uncompressible
bit sequence. A last attack could be considered by a very motivated
attacker: to compress both original and tampered binary codes and
use on the fly decompression to execute the malicious code and to
compute MAC over the tampered code. Besides the fact that this approach is unpractical when there is about 10 kB of available program
memory, we observe that such software corruption would be detected
from a simple observation of timing characteristics of the device
under verification (the overhead due to the decompression process
would be easily detected).
We developed a software integrity verification tool (SIVT) that
employs the proposed introspection technique for smart meters. The
SIVT provides a single interface for an operator responsible for performing the integrity verification of measuring instruments embedded
with software in the field. This tool also manages the coexistence
of different software versions, each containing their unique identifier.
This is performed by matching the identifier with the correlated set of
MACs retrieved from a web server.
We presented an approach to improve the degree of protection of intellectual property of measurement devices software under evaluation
by regulators. This approach hides the complexity of performing software integrity verification of electricity meters by an operator, making
it easier to disseminate and training of the software integrity verification task. The exposed ideas are simple and easily implementable with
the aim of cheap micro-controllers. This approach was put into practice for the integrity verification of the Brazilian smart energy meters.
3.3 Software Fingerprinting
ble leaks. In this protocol, the regulator assumes the role of a trusted
third party (TTP), responsible for uniquely identifying each software
module that will be outsourced for analysis. Identifying these modules
is done with the use of fingerprints that are embedded in the software
and are not easily removable.
A fingerprint is inserted into a program by an embedder and it can
be recovered by means of a recognizer algorithm. Formally, a fingerprinting scheme is a function : . The input consists
of a software code p P, an information w W, and a secret parameter k K. The output consists of a fingerprinted code with the
following conditions:
q has the same functional semantics of p;
From almost all code q, or close enough to q, it is still possible to recover the information w, knowing k.
It is important to mention that the function f need not be secret, i.e.,
any individual may be able to generate q from p, w, and k. On the other hand, the second condition above tells us that, knowing the secret
k, it is possible to retrieve the information w from the fingerprinted
code even if it was reasonable modified. This means that it is difficult
for a malicious adversary to remove or circumvent the identification
provided by the fingerprint.
Through the use of fingerprints, it is possible to uniquely identify software modules that are distributed to a third party for analysis.
Consider the scenario where a developer has the code p to be analyzed
by the regulator, and the regulator will outsource this analysis. We
propose the following protocol:
1. The regulator generates a cryptographic digest (or hash) of the
whole code to be analyzed together with evaluator identification.
2. The regulator signs the hash. The set (signed hash + evaluator
identification) is the information w to be embedded in the code.
3. The regulator encodes an appropriate graph using the algorithm
presented in [27], and embedding in the original program p (in
a secret region k).
It is important to understand the motivation for each stage.
Stage 1 is important for the uniqueness of the information, w, that
is correlated to the code and the evaluator. Additional information
can be added, for example, the time of the fingerprinting insertion.
Stage 2 ensures that only the regulator will be able to embed valid
information since it is signed, otherwise, any individual would be
able to add a fake identifier to the code. Stage 3 ensures that only
the designated evaluator will have the modified code with a given
fingerprint. In this way, after the code delivery, any leaks from that
modified code can be traced by the fingerprint.
4. Measurement Confidence Chain
27
SPECIAL FEATURE
a)
b)
Figure 4. (a) Trusted module positioned on the meters internal circuit. (b) Consumption authenticator shown on the display.
28
(1)
SPECIAL FEATURE
7. References
29
TECHNICAL PAPERS
Abstract: The proposed redefinition of several International System (SI) base units is a topic that has been on the metrology
agenda for the last decade. Recent progress on several determinations of the fundamental constants means that we now have a
good idea of the defined numerical values that will be given in the new system to the Planck constant, h, and the elementary charge,
e. This is especially relevant to electrical metrology as new numerical values for the von Klitzing and Josephson constants, given
by the relations RK=h/e2 and KJ=2e/h, will replace the existing 1990 conventional values, RK90 and KJ90. The implementation
of the new system cannot be done without introducing small step changes into sizes of the electrical units that are disseminated
using Josephson and quantum Hall intrinsic standards. At the time of writing it looks likely that the relative change from KJ90
to KJ will be of the order 1 107, and that from RK90 to RK will be approximately 2 108. This paper discusses the practical
impact of these changes on electrical metrology and highlights the long term benefits that will come from the updated system.
The CCEM (Consultative Committee for Electricity and Magnetism) of the International Committee for Weights and Measures
is now taking the first steps to ensure a smooth implementation, most probably in 2018.
1. Introduction
Discussions on a possible revision of the International System of Units (SI) [1] have been
ongoing for over a decade. After an initial
concentration on the stability of the kilogram
[2, 3], a consensus has emerged for a major
revision centered around the redefinition of
four of the seven base units, that brings fundamental constants to the fore [4]. Recent
progress towards this goal [5] indicates that
such a redefinition is now a real possibility
for 2018. This paper explores what this upcoming change means for how we realize and
disseminate the SI electrical units.
As we will see, no change to working
practices or traceability routes is required; the
new SI effectively formalizes what is already
standard practice in electrical metrology
laboratories.
In fact, the main change
brought about by the new SI is that we will
no longer need to worry about the distinction
between the representations of the volt and
the ohm maintained in the laboratory and the
presently inaccessible true SI units. The
new SI will make the present representations
of the volt and the ohm equal to the true
SI units. This will change very little for
30
(1)
!"# = ! .
(2)
! = 2 .
(3)
(4)
TECHNICAL PAPERS
1.2 Electrical Traceability Today: the
Conventional 1990 Values
Authors
Nick Fletcher
Bureau International des Poids
et Mesures (BIPM)
Pavillon de Breteuil, 92312
Svres Cedex, France
nick.fletcher@bipm.org
Gert Rietveld
Van Swinden Laboratorium (VSL)
P.O. Box 654
2600 AR Delft, Netherlands
grietveld@vsl.nl
James Olthoff
National Institute of Standards and
Technology (NIST)
100 Bureau Drive
Gaithersburg, MD 20899-8171
james.olthoff@nist.gov
Ilya Budovsky
National Measurement Institute
Australia (NMIA)
Bradfield Road, West Linfield
NSW 2070, Australia
ilya.budovsky@measurement.gov.au
Martin Milton
Bureau International des Poid
et Mesures (BIPM)
Pavillon de Breteuil, 92312
Svres Cedex, France
martin.milton@bipm.org
Since 1990, experimental progress has continued on many determinations of fundamental constants and is regularly reviewed in
adjustments published by CODATA [12]. In
this section, we review what the last 20 years
have brought for our knowledge of the constants RK and KJ.
2.1 An Updated Value for RK
(5)
31
TECHNICAL PAPERS
RK()
u(/)
1990(/)
1990 Value
25 812.807
----
----
1998 CODATA
3.7 10-9
+22 10-9
2002 CODATA
3.3 10-9
+17 10-9
2006 CODATA
0.7 10-9
2010 CODATA
0.32 10
+22 10-9
-9
+17 10-9
! =
!
!!!
. .
(6)
For the purpose of analyzing the contributions to the value of KJ, we can assume the
uncertainty on RK is negligible, and can use
Eq. (6) to convert the reported experimental h
values to individual values of KJ. The results
thus obtained for the most important contributing experiments are shown in Fig. 3. The
values are plotted as relative differences from
KJ90, in parts in 108. The uncertainty bars are
standard uncertainties.
We plot both results that were included in the
latest (2010) CODATA adjustment and those
published since [13, 14]. The convergence of
recent results predicts a value of KJ approximately 10 parts in 108 below the 1990 value.
Although the picture is not yet completely finalized and critical results have only become available within the last six months, it has become
clear that we will have to deal with a small, but
significant, offset from the 1990 value.
3. Implementation of the New SI
Changes for Electricity
Figure 1. Data from Table 1: successive CODATA values of the von Klitzing
constant RK published since the adoption of the 1990 value.
KJ (GHz/V)
u(rel.)
1990(rel.)
1990 Value
43 597.9
----
----
1998 CODATA
43 597.898 (19)
3.9 10-8
-0.4 10-8
2002 CODATA
43 597.879 (41)
8.5 10-8
-4.3 10-8
2006 CODATA
43 597.891 (12)
2.5 10
-8
-1.9 10-8
2010 CODATA
43 597.870 (11)
2.2 10-8
-6.2 10-8
TECHNICAL PAPERS
3.1 Impact for Resistance Measurements
33
TECHNICAL PAPERS
TECHNICAL PAPERS
KJ have to be used. The unavoidable step change introduced on doing
so will be at the limits of visibility for the most demanding users of
resistance and voltage calibrations. The most significant impact will
be in the area of voltage, where a step change of around 0.1V/V is
foreseen, which will be clearly visible in comparisons of Josephson
systems, such as those organized by NCSL International.
This paper is part of the work of a task group of the Consultative
Committee on Electricity and Magnetism (CCEM), created to address
the implementation of the new SI. The details given here will
continue to be updated with new experimental evidence, and the exact
changes to be applied will not be known until just prior to the date
of implementation (most likely 2018). However, the recent progress
means we now have a good picture, and that we can start to prepare
for the necessary changes.
5. References
35
TECHNICAL PAPERS
Abstract: Above 962C, the ITS-90 is realized at the National Research Council (NRC) of Canada by a standard radiation
thermometer with a known spectral responsivity and a silver freezing-point blackbody. Together with Plancks law, the temperature
scale can be extrapolated to temperatures in excess of 2500C, albeit with uncertainties that increase with higher temperatures.
This realization can then be disseminated to other radiation thermometers via a variable-temperature blackbody for use in, for
example, calibration laboratories. In the future, it is expected that new high-temperature fixed points with transition temperatures
exceeding 3000C will allow an interpolated high-temperature realization and lower uncertainties.
1. Introduction
! ! !!" !
where
T90(X)
!!
!!
!!!" !
!
!"# !!! !!
!"
!"#
(1)
Authors
Andrew D. W. Todd
andrew.todd@nrc-cnrc.gc.ca
Donald J. Woods
don.woods@nrc-cnrc.gc.ca
National Research Council
of Canada
1200 Montreal Road
Ottawa, Ontario, Canada K1A 0R6
TECHNICAL PAPERS
L(T90)
L[T90(X)]
c 2
Using Eq. (1), the defined temperature of one of the fixed points
of Ag, Au or Cu (T90(X)), and knowledge of the wavelength, the
unknown temperature (T90) can be determined. Since only one fixed
point is used this realization requires extrapolation and so the farther
the unknown temperature is from the defining fixed point, the higher
the uncertainty in the unknown temperature. In fact, the uncertainty
increases with the square of the temperature.
Since there is no such thing as a perfectly monochromatic radiation
thermometer, some additional knowledge about the operating
wavelength of the radiation thermometer is required. We must also
introduce the emissivities of the T90(X) fixed point and the target at the
temperature being measured. Then, Eq. (1) becomes:
! !!"
= ! !
!" !
where:
r
R()
S(T90)
S[T90(X)]
(,T90)
(,T90(X))
!
! ! !,!!" ! ! !!" ! ! !"
!
!
! ! !,!!" ! ! ! !!" ! ! ! !"
(2)
37
TECHNICAL PAPERS
Uncertainty Component
Ag point
u (k = 1)
u (k = 1) at 1000 C (mK)
u (k = 1) at 2500 C (mK)
51
236
Wavelength calibration
0.1 nm
46 mK
11
546
Reference detector
0.1 nm
11
532
0.0009 nm
Out-of-band transmittance
0.065 nm
355
Interpolation
0.0065 nm
35
Non-linearity
0.0005
Gain Ratio
0.0001
Total (k = 1)
54
874
Total (k = 2)
107
1749
Figure 4. The LP3 and the electrically heated, variable high-temperature blackbody.
38
TECHNICAL PAPERS
thermodynamic temperature of the fixed point used to define the
ITS-90 (e.g. Ag, Au or Cu) will be magnified. If suitable high
temperature fixed-points were available with their thermodynamic
temperature known, then interpolation between the fixed points would
be possible. Uncertainties in this temperature range using a set of
high-temperature fixed points has been shown to have uncertainties
comparable to the ITS-90 but with more robustness and a better link
to thermodynamic temperature. A detailed comparison of extrapolated
versus multiple fixed-point schemes is given in [6].
In order to use a radiation thermometer to realize the ITS-90 as
it is written, the relative spectral responsivity needs to be known
in addition to the measurement of the Ag, Au or Cu freezing point.
Measuring the relative responsivity of a radiation thermometer is
time consuming and out of reach of most users. In a scheme where
three or more fixed points are used, the relative responsivity is not
required to be known and therefore offers some practical advantages
for many users.
In recent years, a great deal of effort has been expended to develop
high-temperature fixed points based on metal-carbon eutectic
materials [7, 8]. These metal-carbon alloys have been shown to exhibit
stable melting transitions and cover temperatures from ~1153 C
(Fe-C) to 3185C (HfC-C). While the determination of the transition
temperatures is still an active area of study at many NMIs, Fig. 5 [8]
shows the approximate melting temperatures of the metal-carbon and
metal carbide-carbon fixed points.
At present, the transition temperatures of these high-temperature
fixed points are not well known, but an international effort to measure
the absolute melting temperatures of the Co-C, Pt-C, and Re-C fixed
points is underway [9, 10]. At NRC we have constructed MC-C fixed
point of Co-C [11], Pt-C and Re-C [12] for radiation thermometry.
We have also measured TiC-C fixed points which have a transition
temperature near 2758C. NRC is also actively working to determine
the thermodynamic temperature of the fixed points of Co-C, Pt-C and
Re-C. Results of preliminary measurements have been presented in
[13] and measurements that will contribute to the final assignment
of the transition temperatures have been completed (but are being
kept unpublished until measurements at all the participating NMIs
have been completed). In this final work our determination of the
thermodynamic temperature of Co-C (~1324 C), Pt-C (~1735 C),
and Re-C (2474C) were made with k = 2 uncertainties of 0.281C,
0.502C, and 0.831C, respectively.
It is anticipated that by using three or more fixed points an
interpolated realization of high temperatures by radiation thermometry
can be achieved with uncertainties comparable to, or better than,
current ITS-90 realizations.
In addition to providing the ability to realize an interpolated hightemperature scale, these fixed points can be used as artifacts to be
circulated among different organizations to determine equivalence
of the locally-realized scales. Using fixed points for this purpose
has a number of advantages over other options for probing a high
temperature scale: they are more robust, less prone to drift, and less
sensitive to emissivity than lamps and are smaller, more robust, and
easier to transport than a radiation thermometer.
5. Summary
Figure 5. Approximate melting temperatures of the metalcarbon and metal carbide-carbon fixed points. The melting
temperature and the uncertainty in the melting temperature
of these fixed points are not well known at this time.
The authors thank Ken Hill for his helpful ideas and perspective.
7. References
39
TECHNICAL PAPERS
[5] http://www.bipm.org/cc/CCT/Allowed/22/CCT03-03.pdf
[6] G. Machin, P. Bloembergen, K. Anhalt, J. Hartmann, M. Sadli,
P. Saunders, E. Woolliams, Y. Yamada, and H. Yoon, Practical
Implementation of the Mise en Practique for the Definition of
the Kelvin Above the Silver Point, Int. J. Thermophys., vol. 31,
no. 8-9, pp. 1779- 1788, 2010.
[7] Y. Yamada, H. Sakate, F. Sakuma, and A. Ono, Radiometric
observation of melting and freezing plateaus for a series of
metal-carbon eutectic points in the range 1330C to 1950C,
Metrologia, vol. 36, no. 3, pp. 207209, 1999.
[8] E. Woolliams, G. Machin, D. Lowe, and R. Winkler, Metal
(carbide)-carbon eutectics for thermometry and radiometry: a
review of the first seven years, Metrologia, vol. 43, no. 6, pp.
R11R25, 2006.
[9] G. Machin, K. Anhalt, P. Bloembergen, M. Sadli, Y. Yamada, and
E. Woolliams, Progress Report for the CCT-WG5 High Temperature Fixed Point Research Plan, in Temperature: Its Measurement and Control in Science and Industry, vol. 8, ed. C.W.
Meyer, AIP Conference Proceeding 1552, pp. 317322, 2013.
[10] G. Machin, J, Engert, R. Gavioso, M. Sadli, and E. Woolliams,
The Euramet Metrology Research Programme Project
Implementing the New Kelvin (InK), Int. J. Thermophys., vol
35, no. 3-4, pp. 405-416, 2014.
[11] A. Todd and D. Woods, Comparison of three Co-C fixed
points constructed using different crucible lining materials,
in Temperature: Its Measurement and Control in Science and
Industry, vol. 8, ed. C.W. Meyer, AIP Conference Proceeding
1552, pp. 369373, 2013.
[12] A. Todd, D. Lowe, W. Dong, and D. Woods, Comparison of
realizations of Re-C fixed points filled and measured at NPL
and NRC, in Temperature: Its Measurement and Control in
Science and Industry, vol. 8, ed. C.W. Meyer, AIP Conference
Proceeding 1552, pp. 797801, 2013.
[13] A. Todd and D. Woods, Thermodynamic temperature
measurements of the melting temperatures of CoC, PtC and
ReC fixed points at NRC, Metrologia, vol. 50, no. 1, pp.
20-26, 2013
40
TECHNICAL PAPERS
Abstract: Proficiency testing (PT) is an evaluation of participants performance against pre-established criteria by means of
interlaboratory comparisons. The normalized error, En, is the most widely used performance statistic for determining the measurement capability of a calibration laboratory. One of the variables in the En equation is Uref, which is the expanded uncertainty
of the reference laboratorys assigned value. To evaluate a participants performance effectively, if any effects of the PT scheme
are significant, the additional uncertainties should be combined with the reference laboratorys reported expanded uncertainty
to estimate Uref. Among such uncertainties, the stability of artifacts is an important uncertainty component in the PT scheme,
especially for a calibration laboratory. Based on practical PT experience, most artifacts can be regarded as sufficiently stable if
the difference between three reference laboratory measurements is small. In such cases, the median of the three measurements
is usually chosen as the assigned value, and its reported expanded uncertainty is used as the Uref value. However, some artifacts,
such as standard resistors, drift over time. This leads to questions about how to accurately determine the assigned values and
expanded uncertainties of these artifacts. This paper presents a PT scheme for standard resistors that demonstrates the evaluation
of PT results with a drifting artifact.
1. Introduction
Authors
Chen-Yun Hung
hungcy@itri.org.tw
Pin-Hao Wang
pin-hao@itri.org.tw
Cheng-Yen Fang
fangcy@itri.org.tw
Center for Measurement Standards
Industrial Technology Research Institute
321 Kuang Fu Rd., Sec. 2
Hsinchu, Taiwan 30011, R.O.C.
TECHNICAL PAPERS
Because standard resistors drift over time,
the evaluation criterion should be selected
carefully. If the standard uncertainty of stability testing is too large, some participants
may receive action and warning signals because the assigned values are inaccurate.
Therefore, based on the guidelines for limiting the uncertainty of the assigned value in
ISO 13528 [2], the criterion for allowing stability uncertainty is
2
2
u sta 0.3 u lab,
min + u ref ,
(1)
where
usta is the standard uncertainty of stability
testing;
ulab, min is the minimum standard uncertainty of
the participants results; and
uref is the standard uncertainty of the reference
laboratorys result.
When Eq. (1) is satisfied, the ratio,
2
2
u lab,
min + u ref
,
will fall
2
2
2
u lab,
min + u ref + u sta
in the range
0.96
2
2
u lab,
min + u ref
2
2
2
u lab,
min + u ref + u sta
Item
1.00 . (2)
Nominal
Value
Manufacturer
Model
Serial No.
Standard
resistor
iET
SRL-1
B2-9425105
Standard
resistor
10 k
iET
SRL-10k
B2-9425117
Nominal Value
Uncertainties
usta
ulab, min
uref
0.00000020
0.000008
0.00000015
10 k
0.0000004 k
0.00005 k
0.0000015 k
43
TECHNICAL PAPERS
0.9991798
Nominal Value (1 )
Nominal Value (1 )
Measured Value
max = 0.9991793
min = 0.9991784
0.9991812
Measured Value ()
0.9991808
0.9991790
0.9991786
0.9991782
0.9991778
0.9991804
0.9991800
0.9991796
0.9991792
0.9991788
0.9991784
2012/8/1
2012/8/3
2012/8/5
2012/8/7
2012/8/9
2012/8/11
2012/8/13
2012/8/15
2012/8/17
2012/8/19
2012/8/21
2012/8/23
2012/8/25
2012/8/27
2012/8/29
2012/8/31
2012/9/2
2012/9/4
2012/9/6
2012/9/8
2012/9/10
2012/9/12
2012/9/14
2012/9/16
2012/9/18
2012/9/20
2012/9/22
2012/9/24
2012/9/26
2012/9/28
2012/9/30
Measured Value ()
0.9991794
2012/10/01
E120519A
10.000753
10.000752
10.000751
2012/8/1
2012/8/3
2012/8/5
2012/8/7
2012/8/9
2012/8/11
2012/8/13
2012/8/15
2012/8/17
2012/8/19
2012/8/21
2012/8/23
2012/8/25
2012/8/27
2012/8/29
2012/8/31
2012/9/2
2012/9/4
2012/9/6
2012/9/8
2012/9/10
2012/9/12
2012/9/14
2012/9/16
2012/9/18
2012/9/20
2012/9/22
2012/9/24
2012/9/26
2012/9/28
2012/9/30
10.000750
10.000749
10.000760
Measured Value
max = 10.000753 k
min = 10.000751 k
10.000754
2013/03/08
E130072A
Date/Report Number
Date
2012/12/07
E120641A
Date
Figure 3. Stability testing results of the 10 k standard resistor.
10.000758
10.000756
10.000754
10.000752
10.000750
10.000748
10.000746
10.000744
2012/10/01
E120520A
2012/12/07
E120642A
2013/03/08
E130073A
Date/Report Number
Figure 6. Calibration results of the 10 k standard resistor.
3.2 Example of Drifting Artifact Standard Resistor
44
TECHNICAL PAPERS
Group
Nominal
Value
1
10 k
II
Uref
Uref
0.9991795
0.0000004
0.9991801
0.0000005
10.000752 k
0.000003 k
10.000752 k
0.000003 k
Nominal Value
Lab Code
A
10 k
0.00
0.09
0.07
0.02
0.01
0.14
0.07
0.04
4. Confidentiality
0.02
0.01
0.00
0.22
3.85#
3.36#
0.08
0.01
0.67
0.07
NA
0.07
0.18
0.01
0.03
0.00
0.00
0.02
0.27
0.26
0.19
0.30
0.17
0.14
Note: NA means the value is not available, and # designates an unsatisfactory result.
(x X)
2
2
U lab
+U ref
(3)
where
x is the participants result;
X is the assigned value;
Ulab is the expanded uncertainty of the
participants result; and
Uref is the expanded uncertainty of the
reference laboratorys assigned value.
The criteria for performance evaluation are
as follows:
|En| 1.0 indicates satisfactory
performance and generates no signal;
5. Conclusions
45
TECHNICAL PAPERS
Measured Value ()
1.002
Nominal Value (1 )
Measured Value
10.006
1.001
1.000
0.999
0.998
0.997
0.996
0.995
Measured Value
10.005
1.003
10.004
10.003
10.002
10.001
10.000
9.999
9.998
9.997
9.996
Ref Lab-I
Ref
Laboratory Code
Laboratory Code
6. References
1.0000
Measured Value ()
0.9998
Nominal Value (1 )
Measured Value
Assigned Value = 0.9991801
0.9996
0.9994
0.9992
0.9990
0.9988
0.9986
0.9984
Ref LabII
Laboratory Code
46
Register Today!
.org
2014
Quality
Summit
WHEN:
September 24 - 25, 2014
Registration opens: 7:30 AM
WHERE:
Suburban Collection Showplace
46100 Grand River Avenue
Novi, MI 48374
REGISTRATION:
Early Registration Deadline:
July 31, 2014
Regular Registration
Deadline: September 10, 2014
Type
Early
Regular
Late
Member
$575
$700
$775
NonMember
$775
$900
$975
TO REGISTER:
Visit www.aiag.org and click
the Events tab or call
248-358-3003.
Sponsors:
2014 AIAG | 26200 Lahser Road, Suite 200 | Southfield, MI 48033 | Tel: 248.358.3003 | Fax: 248.799.7995 | www.aiag.org
47
TECHNICAL PAPERS
Abstract: We describe a 40 GHz cavity stabilized oscillator (CSO) that uses an air-dielectric cavity resonator as a frequency
discriminator to reduce the phase modulation (PM) noise of a commercial 10 GHz dielectric resonator oscillator (DRO) frequency
multiplied by four. Low PM noise and small size were the main design goals. Single sideband (SSB) PM noise equal to
-128 dBc/Hz at a 10 kHz offset from the carrier frequency is achieved for the CSO. In addition, we report on the PM noise of
several Ka-band components.
1. Introduction
Authors
Archita Hati
archita.hati@nist.gov
Craig W. Nelson
craig.nelson@nist.gov
Bill Riddle
bill.riddle@nist.gov
David A. Howe
david.howe@nist.gov
1,2,4
TECHNICAL PAPERS
In Section 2 of this paper, we first provide
the phase-noise performance of an assortment
of active components at 40 GHz, including data that most manufacturers omit from
their specifications. In Section 3, we discuss
the design of our compact dielectric-cavity
resonator. The CSO and its PM noise performance are described in Sections 4 and 5,
respectively. Section 6 presents a summary.
2. PM Noise of Active Components
at 40 GHz
where
kd is the
phase-to-voltage
conversion factor, G is the gain of the IF amplifier and ( t ) is the
is the
PDs phase-to-voltage
where
kd PDs
(0)
difference
between
the
phase
fluctuations
conversion factor, G is the gain of the
theDUT-1 and DUT-2. Taking the Fourier transform of Eq.
(1)
provides
IFIFamplifier
the gain of the
amplifierand
and ( t ) isisthethe difference
between
the phase
fluctuations
DUT-2. Taking
the Fourier
transform
of Eq.of the DUT-1
v n ( t ) transform of
and DUT-2. TakingPSD
the Fourier
2 noise for a sample of commercial amplifiers
3.+ PM
S ( f ) =
S ( f ) ,
= S , DUT ( f ) + SFigure
, NF ( f )frequency
(0)
(carrier
= 40 GHz).
Eq. (1) provides
( k G )2
f ) + 2Swhere,
PM noise, S DUT ( f ) is the actual PM noise of a pair of DUTs, S NF ( f ) is
( f ) ,S ( f ) is the measured
(0)
49
TECHNICAL PAPERS
Amp1
Amp2
Amp2
Amp3
Amp1
Figure 4. Input-referenced
(i.e. 40 GHz) PM noise Amp2
for a pair
Amp2
Amp3
of dividers. Divider-1 is a commercial divider shown in Fig.
1. The regenerative divider is custom-built with two divideby-2 in series.
The input frequency is 40Amp1
GHz and the
Amp3
40
GHz frequency
Amplifiersis 10 GHz.
output
A-1844
JS4-18004000-40-5A
40 GHz A-1844
Amplifiers
Figure 5. Amp3
Output-referenced PM noise for a pair of
multipliers. Multiplier-1 and Multiplier-2 are shown in Fig. 1.
The input frequency is 10 GHz and the output frequency is
40
GHz.
Amp2
Amp3
JS4-18004000-40-5A
JS4-18004000-40-5A
Amp1
mp1
Amp2
Amp2
00-40-5A
Divider-1
40 GHz A-1844
Amplifiers
Amp3
Amp3
Multiplier-1
JS4-18004000-40-5A
Multiplier-2
mplifiers
4
JS4-18004000-40-5A
Divider-1
Multiplier-1
Figure 1. Images of commercial
products used for PM noise measurement atMultiplier-2
40 GHz.
Multiplier-1
ers
Multiplier-2
JS4-18004000-40-5A
kd
volts
radian
(3)
f adifference
DBM
thatestablish
acts
as a the
phase
detector
(PD).
A
shifter
is
used
toIfatthe
between
phase
fluctuations
DUT-1
and
DUT-2.
pathPDlengths
from
the reference
)
between
two
signals
the
inputs
required
for
phase
noise
phase-quadrature
(90
ges
of
commercial
products
used
for
PM
noise
measurement
at
40
GHz.
under
test
(DUTs)
[8].
It
uses
a
reference
oscillator,
a
double
balanced
mixer
(DBM),
and
a
phase
()
of three commernoise, L() shifter.
ble balanced mixer (DBM), and a phase shifter.
between
twoto
signals
inputs
required
for
phase
noise
oscillator
the LO at
andthe
RFPD
ports
are matched,
the
noise
is
cancels,
a outputs
measurement.
The
PD
output
voltage
fluctuations
at
thecommon
baseband
thateach
are proportional
to the of the DUTs are
oscillator
is oscillator
split
into
two tothat
drive
the
inputmode
of
DUT.toThe
products
used
for PM
noise
measurement
at 40 GHz.
t commercial
of each DUT.
The
outputs
ofThe
the
DUTsproduces
aresignal
cially
available
amplifiers.
Note that Amp1
uces
voltage
fluctuations
at the
baseband
that
are
toof
the
large
degree.
The output
of connected
the
after
amplification
isports
analyzed
with that
a fast-Fourier-transform
(FFT)
difference
thePD
phase
fluctuations
and
DUT-2.
Ifacts
theaspath
lengths
from
the reference
to
the
LOproportional
andofRFDUT-1
a DBM
a phase
detector
(PD).
A phase shifter is used to
phasemeasurement
detector
(PD).
Abetween
phase
shifter
ismeasure
used
tonoise
noise
system
(Fig.
2)
is
used
to
of
a
pair
of
devices
0
uations
of DUT-1
and
DUT-2.
If
the
path
lengths
from
the
reference
spectrum
analyzer.
The
voltage
output
of
the
IF
amplifier
is
given
by
to the LO
and
RFphase-quadrature
ports are matched,(90
the) oscillator
that is common
mode
cancels,
to a for
and
Amp3
havephase
approximately
the same gain
establish
between noise
two signals
at the PD
inputs
required
noise
at the
PD oscillator
inputs
required
for
phase
noise
uses
a reference
oscillator,
balanced
mixer
(DBM),
and of
a cancels,
phase
shifter.
matched,
the
oscillator
noise
is
mode
to ais voltage
eare
measurement
system
(Fig. a2)double
isThe
usedmeasurement.
tothat
measure
noise
of after
aPD
pair
devices
large
degree.
output
of common
the
PD
amplification
analyzedfluctuations
with a fast-Fourier-transform
The
output
produces
at the baseband and
that(FFT)
are dB
proportional
to thebut that there is a
P1
compression,
ons
at
the
baseband
that
are
proportional
to
the
split
into
two
to
drive
the
input
of
each
DUT.
The
outputs
of
the
DUTs
are
Figure
6.
cylindrical
aluminum
resonator
40 GHz.
reference
oscillator,
a( tdouble
balanced
mixer
and aof
phase
shifter.
Da after
amplification
is
analyzed
a (DBM),
fast-Fourier-transform
(FFT)cavity
vthat
kasd(a)
Gfrom
Air-dielectric
The
tthe
spectrum
output
the
IFisamplifier
by andatDUT-2.
) analyzer.
(with
)and
,voltage
(0) If the pathvariation
difference
between
the shifter
phase
fluctuations
of DUT-1
lengths from
the reference
n=
RF ports
of
adrive
DBM
acts
phase
detector
(PD).
A phase
used to is given
DUT-2.
the
path
lengths
reference
of almost
20 dB in the PM noise
into
two
toIfIF
the
input
of aeach
DUT.
Thethe
outputs
of the
are
The
diameter
(2a)
length
(d)DUTs
of the
cavity are approximately 2 cm.
0the
put
of
amplifier
is
given
by
ure
(90
)
between
two
signals
at
the
PD
inputs
required
for
phase
noise
oscillator
to
the
LO
and
RF
ports
are
matched,
the
oscillator
noise
that
is
common
mode
cancels,
to avariation in noise
lator
noise
that
is
common
mode
cancels,
to
a
ports of a DBM thatThe
acts SMA
as a phase
detector (PD).
A phase
shifter
is usedcan
to be used as a size reference. (b)
connectors
on
the
circuit
board
performance.
A
similar
t ) is the
0where
kdwith
istwo
theasignals
PDs
phase-to-voltage
conversion
is to
the
gain
the IF amplifier
and (with
utput
produces
voltage
fluctuations
at
thek
baseband
that
are
proportional
the
degree.
The
output
ofGnoise
the
PD
afterofamplification
is| of
analyzed
a fast-Fourier-transform (FFT)
=
tlarge
G|
t40
) between
at vdata
the
PD
inputs
required
forfactor,
phase
(
)
(
)
, GHz
(0)
s90
analyzed
fast-Fourier-transform
(FFT)
n
d
Measured
of
|S
of
a
cavity.
(c)
Measured
data
of
|S
a
40
GHz
11
phase
fluctuations
of
DUT-1
DUT-2.
If21thethat
pathof
lengths
from(0)
the
voltage
fluctuations
atMeasured
thespectrum
baseband
are
proportional
toreference
the
difference
between
theand
phase
fluctuations
the
and
DUT-2.
the Fourier
analyzer.
The
voltage
output
ofTaking
the IF amplifier
is transform
given by of Eq. occurs for the two dividers shown in Fig. 4.
isproduces
cavity.
phase
of
S11DUT-1
for
a 40
GHz
RFgiven
ports by
areofmatched,
the (d)
oscillator
noise
that
is common
mode
cancels,
tocavity.
a
the noise of two dividers: one
fluctuations
DUT-1
and
DUT-2.
If
the
path
lengths
from
the
reference
( t ) iscompared
where
kd is theisPDs
phase-to-voltage
conversion factor,
G is the gain of the IF amplifier and We
the
provides
t of(1)
the
after
amplification
analyzed
with
a fast-Fourier-transform
ports
arePD
matched,
the oscillator
noise
that
is
common
mode and
cancels,
to
t )a(FFT)
(
ge
conversion
factor,
G
is
the
gain
of
the
IF
amplifier
is
the
commercial
(pictured
in Fig. 1) and the other
voltage
output
of
the IF amplifier
is given
by
vn =
(FFT)
DUT-1
difference
the
phase
fluctuations
DUT-2. Taking the Fourier transform of Eq.
( t ) kof
( t ) , andratio,
(0)
d Gthe
(0)
he PD after
amplification
isbetween
analyzed
with
a fast-Fourier-transform
conversion
and PSD is the power spectral a custom regenerative divide-by-four (two
ations
of
the
DUT-1
and
DUT-2.
Taking
the
Fourier
transform
of
Eq.
(1)
provides
PSD byv n ( t )
ge output of the IF amplifier is given
2 measurement method is accurate
This
( t )gain
SIF( famplifier
S ( f ) , factor, G is the(0)
series)
)=
) + S(0), NFdensity.
( f ) + conversion
( t ) is the[9, 10]. In addi,
PDs( fphase-to-voltage
gain of the IFdivide-by-two
amplifier and in
where
k = isSthe, DUT
GGisthe
of the
and
2 (dt ) isthe
k
G
(
)
as
long
as
theofPM
noise
of the
DUT
pairsTaking
is at the
nd( t )DUT-2.
dtransform
,
(0)
difference
between
the
phase
fluctuations
the
DUT-1
and
DUT-2.
Fourier
transform
Eq. of two frequention,
we
compared
theofnoise
PSD
v
t
Taking
the
Fourier
of
Eq.
) and ( t ) is the
se-to-voltage conversion factor, G is the gain of the
n (amplifier
IF
2
S ( f(1)
f ) +10
SdB
f ) +than
Sthe
f ) ,noise floor of the(0) cy multipliers (Fig. 5). The large differences
= provides
= S , DUT (least
hase
higher
PM
)
(
(
NF
2
the
DUT-1
andgain
DUT-2.
Taking
transform
2 Fourier
the
S ( f of
Sthe
( t ) isactual
-voltage
conversion
G is the
of the
IF amplifier
and
theof Eq.
the
measured
PM
noise,
PM noise of a pair of DUTs, S NF ( f ) is
) isfactor,
( f ) is(2)
Sfluctuations
= where,
, DUT( f ) + S , NF ( f ) + S ( f()kd, G ) DUT
(0) measurement system and leakage AM
noise.
fluctuations
of the DUT-1
and DUT-2.
Takingcomprised
the Fourier of
transform
of IF
Eq.amplifier noise, S ( f ) is the measured AM in noise performance between the various
the measurement
system
noise floor
PD and
The
noise
floor
of
the
measurement
system
is devices indicate the importance of selecting
PSD v t actual PM noise of a pair of 2DUTs,
measured
PM
noise,
S ( f ) isisthethe
S DUT ( fn)(is)the
where,
measured
PM
noise,
2 iswhere
S
f
S
f
S
f
S
f )S, NF ( f ) is
=
=
+
+
noise,
the
AM-to-PM
conversion
ratio,
and
PSD
is
the
power
spectral
density.
This
measurement
(
)
(
)
(
)
(
PSD
v
t
f
S
f
+
(
)
(
)
(
)
DUT
NF
,
,
n
(0)
obtained
simply
by
replacing
the
DUTs
with
NF
2 of a pair of DUTs, S NF2( f ) is
oise, S DUT= S( f ) is the
PM
noise
,( f )actual
the correct
PM
f )the
Snoise
f ) floor
+isSlong
+(0)
the
noise
of
acomprised
pairDUT
of((0)
(
(
the
measurement
system
of
PD
and
IF
amplifier
noise,
is
the
measured
AM components when designing a
, DUT
, NFactual
k
G
S
f
(
)
)
,
method
is
accurate
as
as
PM
noise
of
the
pairs
is
at
least
10
dB
higher
than
the
PM
noise
2
d
v n( k( t )G )
coaxial
cables
or
waveguide.
However,
it
is
r comprised
of
PD
IF (amplifier
the measured AM
S ( f ) issystem
low-noise system for this frequency band.
d = S
S noise,
f ) conversion
DUTs,
( f ) +and
) +is 2the
(measurement
, DUT
,
(0) and PSD is the power spectral density. This measurement
noise,
Sis, NF
the fAM-to-PM
ratio,
2
important
toSkeep
the
power
level
at the
LO
andof a pair
G
S
f
f
where,
is
the
measured
PM
noise,
is
the
actual
PM
noise
of DUTs, S
)
(
)
(
)
( f ) is
3
fractional
S
f
actual
PM
noise
of
a
pair
of
DUTs,
is
DUT
(
)
on
ratio,
and
PSD
is
the
power
spectral
density.
This
measurement
floor
comprised
of
PD
amplifier
S DUT
f ) is
noise
NF
sured PM noise,noise
the actualasPM
ofthe
aIF
pair
ofnoise
DUTs,ofSthe
( is
( f ) is pairs is at least 10 dB higher than the PMThe
method
accurate
long
asand
PM
noise 2-sigma NF combined
NF DUT
RF
ports
at the same
level
forIFboth
the noise
M
noise
of
the
DUT
pairs
is
at
least
10
dB
higher
than
the
PM
noise
uncertainties
of
the
PM
noise
measurement
the
measurement
system
noise
floor
comprised
of
PD
and
amplifier
noise,
is
the
measured
AM
S
f
(
)
noise,
measured
noise
floorScomprised
amplifier
IF amplifier
noise,
thethe
measured
AM
Sof
f ) is the
S measured
f and
PM
noise,
PM
noise
ofnoise,
a pair
(DUTs,
( f ) isofSthe
( actual
) isisIF
amplitude
DUT
NF ( f ) is AM
PD
floor3and the DUT measurements.
system discussed
above is approximately
M
conversion
ratio,
PSDIF
isamplifier
the
power
spectral
This measurement
modulation
(AM)
isSthe
is the measured
AM
e floor
comprised
of and
PD
and
noise,
noise,
is
conversion
ratio, and PSD is the power spectral density.
This measurement
( fdensity.
) AM-to-PM
the
power
spectral
density.
This
measurement
3of the DUT pairs is at least 10 dB higher than the PM noise
ng
as is
theatratio,
PM
noise
method
isdensity.
accurate
as measurement
long as the PM noise of the DUT pairs is at least 10 dB higher than the PM noise
nversion
and
the power
spectral
pairs
least
10PSD
dB ishigher
than
the PM
noise This
the PM noise 50
of the |DUTNCSLI
pairs is Measure
at least 10 dB
the PM noise
J. higher
Meas.than
Sci. www.ncsli.org
3
3
TECHNICAL PAPERS
c
2
3.832
5
,
a
d
(4)
where a and d are the radius and length of the cylindrical cavity, respectively. For 2a d 2 cm, the resonant frequency is approximately
40.08 GHz.
The cavity is made of aluminum (Al), and the inside surface of
the cylinder and end caps are electro-silver plated and polished to
NCSLI Measure J. Meas. Sci. |
51
TECHNICAL PAPERS
Offset
frequency
(Hz)
CSO
Microoptoelectronic
oscillator
[22]
Voltage
Phase locked
controlled
dielectric
resonator oscillator oscillator
[24]
[23]
Commercial
frequency
synthesizer
[25]
100
-78
-54
-61
+8
-83
1000
-105
-84
-96
-28
-105
10000
-128
-107
-103
-71
-117
100000
-135
-119
-106
-96
-115
1 1 + 2
1 1 + 2
S21
2 12
. (5)
1 1 + 2
TECHNICAL PAPERS
and spurious close-to-the carrier noise is due to the temperature and
vibration sensitivities of the resonator. The source of flicker frequency
noise (f -3) between 100 Hz to 2 kHz is due to flicker phase noise (f -1)
originating inside the discriminator, likely from the circulator, DBM,
and carrier suppression amplifier [7], and also possibly from AM-toPM conversion in the DBM. Above 2 kHz the noise of the CSO is
consistent with and clearly limited by the 4 multiplier noise, or the
bottom-most noise. The broad structure around 500 kHz offset is due
to the discriminator servo loop.
Note that the multiplier PM noise shown in Fig. 9 is measured with
analog measurement system already discussed in Section 2 (Fig. 5). The
DPNMS is used to measure noise at offset frequencies from 10 Hz to
100 kHz. Above a 100 kHz offset, a photonic delay-line measurement
system (PDLMS) [18, 19] is utilized because it has less instrument noise
above 100 kHz than the DPNMS. It is common practice to use such
hybrid schemes to cover a large range of offsets frequencies. Providing
a complete accounting of the measurement uncertainties is beyond the
scope of this paper, but we can summarize by stating that the measurement
uncertainty of the DPNMS is less than 1 dB and the measurement
uncertainty of the PDLMS is less than 2 dB. Both systems were calibrated
against NISTs PM/AM secondary noise standard [20, 21].
There are two drawbacks of measuring the PM noise at the input
of the multiplier at 10 GHz. First, the measured noise at 10 GHz is
limited by the 4 multiplier noise. Any corrections to the 40 GHz
signal that are lower than the multiplier noise cannot be detected
at the 10 GHz output. Second, any improvements to the 40 GHz
signal at frequencies far from the carrier due to the passive filtration
of the cavity cannot be observed. There are strategies that can be
used to reduce the 4 multiplier noise, hence reduce the DRO noise,
given that the multipliers output is phase stabilized by the overall
CSO scheme. The simplest strategy would be to simply replace the
DRO 4 with a single low-noise 40 GHz voltage-controlled
oscillator (VCO). However, candidates for a sufficiently low-noise
phase-lockable oscillator are not commonly available at a 40 GHz
center frequency. Table 1 shows the performance of the CSO with
respect to few commercially available oscillators in the Ka-band. The
PM noise of all oscillators is normalized to 40 GHz. This comparison
is done only in terms of PM noise; their size, cost and power
consumption are not taken into consideration.
6. Conclusions
7. Acknowledgements
The authors thank Justin Lanfranchi for the construction and noise
measurement of 40 GHz regenerative divide-by-four circuit, and
Stefania Rmisch and Jeff Jargon for useful discussion and suggestions.
We also thank Danielle Lirette and David Smith for carefully reading
and providing comments on this manuscript.
8. References
53
TECHNICAL PAPERS
[16] J. Grove, J. Hein, J. Retta, P. Schweiger, W. Solbrig, and S.
Stein, Direct-digital phase-noise measurement, Proceedings
of 2004 IEEE International Frequency Control Symposium,
Montreal, Canada, pp. 287-291, August 2004.
[17] C. Nelson and D. Howe, A Sub-Sampling Digital PM/AM
Noise Measurement System, NCSLI Measure J. Meas. Sci.,
vol. 7, no. 3, pp. 70-73, 2012.
[18] E. Rubiola, E. Salik, S. Huang, N. Yu, and L. Maleki, Photonicdelay technique for phase-noise measurement of microwave
oscillators, J. Opt. Soc. Am. B, vol. 22, no. 5, pp. 987-997,
2005.
[19] http://www.oewaves.com/phase-noise-measurement
[20] F. Walls, Secondary standard for PM and AM noise at 5, 10 and
100 MHz, IEEE T. Instrum. Meas., vol. 42, no. 2, pp.136-143,
1993.
[21] A. Hati, C. Nelson, N. Ashby, and D. Howe, Calibration
uncertainty for the NIST PM/AM noise standards, NIST
Special Publication 250-90, 33 p., July 2012.
[22] http://www.oewaves.com/products/item/85-micro-optoelectronic-oscillator-oeo.html (http://www.fh-microwave.com/
produits-products/photonics/)
[23] https://miteq.com/docs/MITEQ-PLDRO40000.PDF
[24] https://www.hittite.com/products/view.html/view/HMC-C073
[25] http://cp.literature.agilent.com/litweb/pdf/5989-0698EN.pdf
54
TECHNICAL PAPERS
Abstract: A new International System (SI) traceable calibration set-up for reference radiosondes is presented here with a preliminary uncertainty analysis. By meeting the GRUAN requirements, this development fulfils the needs of the meteorological
community for disseminating SI traceability to upper air humidity measurements with reduced uncertainty. The set-up was designed for calibrations from laboratory temperature down to 183 K and in terms of dew-point temperature from 193K to 283 K.
To enable rapid changes in humidity and to shorten the time needed for a single calibration, the set-up utilizes a hybrid humidity
generator method in which two air flows with known water vapor concentrations are mixed. According to a preliminary uncertainty analysis, the relative expanded uncertainty (k = 2) of the set-up is less than 2% expressed in terms of mixing ratio.
1. Introduction
Authors
Hannu Sairanen
hannu.sairanen@mikes.fi
Martti Heinonen
martti.heinonen@mikes.fi
2. Principle of Operation
Richard Hgstrm
richard.hogstrom@mikes.fi
Antti Lakka
antti.lakka@mikes.fi
! ! , ! = ! ,
(1)
Heikki Kajastie
heikki.kajastie@mikes.fi
Centre for Metrology and Accreditation
MIKES
Tekniikantie 1, FI-02150
Espoo, Finland
TECHNICAL PAPERS
The versatility of the system can be further improved by
introducing accurate mass flow control to the dew-point generators.
This allows us to induce quick changes in the water vapor pressure
(within the limits set by the saturator temperatures of the dewpoint generators) without any changes temperature control settings.
This mixing flow generator principle, however, introduces more
uncertainty sources than a single pressure generator, which can be
seen the equation below derived from Eq. (1) for the water amount
fraction of the air after mixing,
! =
!!
!"! !!
!!
!"! !!
+
!! + !! !!
!! + !! !!
xw
rw
Mw
Mg
1 xw
(2)
(3)
To enable humidity calibrations with high accuracy at 193 K, a dewpoint generator that operates at lower temperatures down to 183 K is
needed. This is achieved by immersing a saturator in a commercially
available liquid bath. By introducing a stable bath temperature with
low temperature gradients, a saturator with high efficiency, and high
quality temperature measurements, we can achieve an uncertainty level that meets the GRUAN requirements. The dew-point generator can
be constructed by applying knowledge presented in [8-11].
4.2 Second Dew-Point Generator with Pressure Reducer
57
TECHNICAL PAPERS
Estimate
Uncertainty of
input quantity
Sensitivity
coefficient
Uncertainty
contribution to
mixing ratio
Dew-point
temperature
183.15 K
0.05 K
7.76 x 10-10
under calibration. After a change of flow mixing the time needed to reach stable humidity
inside the measurement chamber is highly
dependent on the flow profile, the geometry
and the surface quality of the chamber. By
minimizing the volume and by optimizing the
shape of the chamber it is possible to achieve
feasible characteristics as presented by Lakka, et al., in [12].
To prevent diffusion and adsorption/desorption effects between the measurement air
and air around the electronics of the radiosonde under calibration, the chamber is divided into two parts: the measurement chamber
and the chamber for electronics. In Fig. 1 and
Fig. 3 the measurement part of the chamber
is the smaller one below the part reserved for
electronics. In the upper part of the chamber,
tiny leaks of water vapor through radiosonde
wiring may increase the humidity. Also the
stabilization time for the humidity is longer
than in the lower part of the chamber due to
the wiring, electronics and uneven flow. The
air flows from the lower to the upper part of
the chamber through a leaking seal assembled
between the measurement chamber and the
radiosonde and tiny holes drilled right next to
connecting screws (see Fig. 3). The flow path
is designed to prevent the back-diffusion of
water vapor to the lower part.
Water vapor
pressure
9.67 x 10-3 Pa
2.90 x 10-5 Pa
9.99 x 10-11
5. Uncertainty Analysis
103800 Pa
20 Pa
6.42 x 10-12
Enhancement
factor
1.0085
1.01 x 10-3
5.79 x 10-8
3.33 x 10-11
Molar mass
of air
28.96 g mol-1
2.55 x 10-5 g
mol-1
1.55 x 10-13
Molar mass
of water
18.02 g mol-1
2.54 x 10-5 g
mol-1
2.47 x 10-13
Mixing ratio
5.84 x 10-8
Pressure
5.48 x 10-10
Relative standard uncertainty (k = 1)
0.94 %
1.88 %
Table 1. Summary of an uncertainty analysis at the low end of the humidity range,
i.e. when operating with only the low dew-point generator.
4.3 Flow Control
TECHNICAL PAPERS
Estimate
Uncertainty of input
quantity
Sensitivity coefficient
Dew-point temperature
183.15 K
0.05 K
2.58 x 10-10
9.67 x 10 Pa
2.90 x 10 Pa
3.02 x 10 Pa
-1
8.77 x 10-11
6.78 x 10-10
103800 Pa
20 Pa
2.81 x 10
5.63 x 10-12
1.0085
1.01 x 10-3
2.90 x 10-8
2.92 x 10-11
193.15 K
0.05 K
1.31 x 10-9
Dew-point generator 1
-3
Enhancement factor
-5
-7
-13
Pa
-1
Dew-point generator 2
Dew-point temperature
Water vapor pressure
Gas flow rate
Pressure
5.47 x 10-2 Pa
1.64 x 10-4 Pa
4.95 x 10-10
1.09 x 10 g s
1.09 x 10 g s
-1
6.23 x 10 s g
6.78 x 10-10
104000 Pa
20 Pa
3.17 x 10-11
1.0076
1.01 x 10-3
1.64 x 10-7
1.65 x 10-10
-2
Enhancement factor
-1
-4
-1
-6
Other sources
Molar mass of air
28.96 g mol-1
1.71 x 10-13
18.02 g mol
2.54 x 10 g mol
1.08 x 10 mol g
2.74 x 10-13
Mixing ratio
-1
-5
-1
-8
-1
1.94 x 10-7
1.73 x 10-9
Relative standard uncertainty (k = 1)
0.89 %
1.78 %
3.5 %
3.0 %
2.5 %
2.0 %
1.5 %
1.0 %
0.5 %
0.0 %
0.0 %
10.0 %
20.0 %
30.0 %
40.0 %
50.0 %
60.0 %
70.0 %
80.0 %
90.0 %
100.0 %
59
TECHNICAL PAPERS
finalized by carrying out an intercomparison with another system
operating within the same measurement range.
More research is needed for validating the temperature and
pressure characteristics at the level required by GRUAN. Achieving
low pressures at low temperatures with low uncertainty as required
by GRUAN is especially challenging. Additionally, temperature
gradients due to the heat dissipated by a radiosonde in the measurement
chamber may cause problems, particularly at low pressures, and may
introduce additional uncertainties to the measurements.
Our new apparatus allows us to test responses to changing
environments. With the two separate generators it is possibly to
measure response time to slower and faster changes in humidity in a
large relative humidity range. Also the relationship between the air
temperature and the response time can be studied.
7. Acknowledgments
60
TECHNICAL PAPERS
Abstract: A calibration system for ultrasonic flaw detectors has been developed at the Government of the Hong Kong Special Administrative Region Standards and Calibration Laboratory (SCL) in accordance with the international standard EN12668-1:2010.
The calibration covers all the periodic and repair tests (the Group 2 tests) required in the standard for checking the performance of
ultrasonic instruments, including their stability, transmitter pulse parameters, receiver response parameters, and time-base linearity.
During the calibration, the ultrasonic flaw detectors transmitter is connected to a combination of a delay generator and function
generator which simulates a delayed version of the transmitted signal as the reflected signal. The simulated reflected signal is then
fed to the receiver of the ultrasonic flaw detector. The stabilities over time and voltage variation of the received waves are measured
and the receiver frequency response is obtained. Other performance parameters of the receiver such as gain accuracy and linearity
are calibrated by comparing gain steps with step attenuators. Lastly, a burst of pulse waves are generated by the arbitrary waveform
generator to simulate a burst of reflected waves to check the linearity of the time base.
1. Introduction
Authors
Samuel C. K. Ko
samuel.ko@itc.gov.hk
Aaron Y. K. Yan
ykyan@itc.gov.hk
Hing-wah Li
scl_rf@itc.gov.hk
62
TECHNICAL PAPERS
be verified using two groups of tests. Tests
performed at manufacturer on a representative
sample of the ultrasonic instruments produced
are categorized by a Group 1 test. The Group
2 tests shall be performed on every ultrasonic
instrument prior to the shipment of the
ultrasonic instrument by the manufacturer,
every 12 months during the lifetime of the
instrument, and immediately after every
repair of the instrument.
SCL provides the Group 2 tests which
include stability, transmitter pulse parameters,
amplifier frequency response, and linearity
of time base. The tests performed are
summarized in Table 1. The measurement
setup and techniques are detailed in Section
2 of the paper. Section 3 discusses the
measurement uncertainties of the tests.
EN12668-1:2010
Standard
Clause 9.3
Title of Test
Stability after warm-up time
Display jitter
Stability against voltage variation
Clause 9.4
Clause 9.5
Clause 9.6
Clause 8.8.2
2. Measurements
5.
6.
7.
8.
63
TECHNICAL PAPERS
Variable AC/DC
Power Supply
0.50
0.00
UUT
Tx
Rx
-0.50
50 Feedthrough
Termination
Voltage (V)
Fixed Attenuator
RF Step
Attenuator
Transmitter
Pulse
-1.00
-1.50
-2.00
Amplifier
(optional)
Delay/Pulse Generator
or
Arbitrary Waveform Generator*
-2.50
0.0E+00
Function Generator
5.0E-08
1.0E-07
Time (s)
1.5E-07
2.0E-07
O/P
Trigger
O/P
Trigger
Sync
I/P Channel
Digitizing Oscilloscope
Trigger
64
TECHNICAL PAPERS
External step
attenuator
setting (dB)
Target amplitude
on screen (% of
screen height)
Acceptable
amplitude (% of
screen height)
90
88 to 92
80
(Reference line)
64
62 to 66
50
48 to 52
40
38 to 42
12
25
23 to 27
14
20
18 to 22
20
10
8 to 12
26
3 to 7
The gain is set to maximum with the receiver input disconnected and the
noise level on the UUT screen is observed and recorded. Then, the gain
is reduced by 40 dB and the UUT is connected to a signal generator. The
peak to peak amplitude of the signal generator, Vin, is adjusted until the
signal amplitude is the same level as the previously recorded mean noise
level. The equivalent input noise in voltage, Vein, is estimated by Vein =
!"#
Vin / 100 and the equivalent input noise per root bandwidth, nin, is
vii. Decrease the UUT gain by minimum gain steps until the gain is
decreased by a total of 10 dB. Observe the signal amplitude. It
should be monotonically decreasing.
viii. Decrease the external step attenuator by 10 dB and record the
signal amplitude as a percentage of the full screen height.
ix. Repeat the steps until the UUT minimum gain is reached
According to the standard [3], the cumulative error in the fine attenuator/gain shall not exceed 1 dB in any successive 20 dB span, or
across the full range, whichever is smaller. In addition, the cumulative error in the coarse attenuator/gain shall not exceed 2 dB in any
successive 60 dB span, or across the full range, whichever is smaller.
2.8 Linearity of Vertical Display
This test measures the accuracy of the vertical grid line of the UUT
screen. The external attenuator is set to 2 dB in the measurement setup
(Fig. 3). The amplitude of the signal generator and the UUT gain are
adjusted so that the signal is at 80 % of the full screen height (Fig. 4).
Then, the external step attenuator is switched to the values of (1, 4, 6,
8, 12, 14, 20 and 26) dB to verify the vertical grid line accuracy on the
display. At each frequency band and step attenuator setting, the measured signal amplitude shall be within the tolerances given in Table 2.
NCSLI Measure J. Meas. Sci. |
65
TECHNICAL PAPERS
Figure 9. The calibration procedure for the linearity of time base for digital instruments.
2.9 Linearity of Time Base
This test measures the accuracy of the horizontal grid line of the
UUT screen. An arbitrary waveform generator is used instead of
a delay/pulse generator (Fig. 3) such that a burst of 11 pulses is
generated to trigger the signal generator. As a result, 11 regularly
spaced signals should be observed on the UUT screen (Fig. 8). The
standard [3] specifies that the deviation of the reference signals from
their ideal positions shall be within 1 % of the full screen width.
2.10 Linearity of Time Base for Digital Ultrasonic Instruments
is plotted against the delay. A best fit curve to the measured values
is generated by software so that the error for each measurement can
be found. The time base non-linearity shall be within 0.5 % of the
full screen width.
3. Measurement Uncertainties
(1)
where
S is the amplitude or time base stability of the echo signal,
A1 is the amplitude or position of the echo signal after 10 min, 20 min
or 30 min, and
A0 is the initial amplitude or position of the echo signal.
The uncertainties of A1 and A0 are from the UUT discrimination of
the final reading and initial reading, respectively. The standard
SAismaximum
thethe
amplitude
oramplitude
time
stability
of
thethe
echo
signal,
A1 amplitude
is the
or base
position
of the
echo
signal,
and
S is the
or
timeamplitude
base
stability
of
the
echo
signal,
minimum
or
position
of
echo
signal.
0 istraceable
are
tois SCLs
attenuation
measurement.
The relative
standardwith
measurement
Type
Bfrom
uncertainty
with
arelative
rectangular
The
assigned
vol
uncertainty,
u(v),
obtained
the
voltage
measurement
performed
an
are traceable
totheSCLs
attenuation
measurement.
Theecho
measurement
oscilloscope.
is
maximum
amplitude
or
position
of
the
signal,standard
and distribution.
A
where
where
1
the
minimum
amplitude
or
position
of
the
echo
signal.
A1 is Athe
amplitude
or position
of
the
echo
signal,
and
0 ismaximum
uncertainty,
u(v),
is
obtained
from
the
voltage
measurement
performed
withofanthe o
uncertainty
is
assessed
from
the
vertical
deflection
calibration
oscilloscope.
uncertainty,
u(v),
is
obtained
from
the
voltage
measurement
performed
with
an
The
standard
uncertainty
of
v
is
due
to
the
standard
uncertainties
of
a
an
The
standard
uncertainty
of
v
is
due
to
the
standard
uncertainties
of
a
and
m
m signal.
A0 isSimilarly,
theamplitude
minimum
amplitude
or
of
the
echo
position
of position
the
signal.
A0 is the minimum
the or
uncertainties
of Aecho
A8-bit
from
the UUT
discrimination
the finalequ
1 and
0 are in
oscilloscope.
digital
oscilloscope
is
voltage
measurement,
soattenuation
theof
resolution
oscilloscope.
The voltage
measurement
accuracy
is
calculated
as
the
sum
of
resolution
and
is
assigned
TECHNICAL
PAPERS
standard
measurement
uncertainty,
u(a),
is
assessed
from
its
calibrat
standard
measurement
uncertainty,
u(a),
is
assessed
from
its
attenuation
calibratio
S uncertainties
isthe
theinitial
amplitude
or1 time
base
stability
ofstandard
the
SUUT
is echo
thediscrimination
amplitude
signal, or time
basefinal
stability
th
and
A0scale
are
from
ofAs
the
Similarly,
of where
A
readingthe
and
reading,
respectively.
Thethe
measurement
uncertainties
of
Aof
1
scale
equals
eight
vertical
divisions.
aofresult,
the
Theasvoltage
measurement
accuracy
isfull
calculated
asoscilloscope
the
sum
of
resolution
and
the
voltage
measurement
uncertainty
of
the
calibration.
Itisrelative
isassigned
evaluated
as areso
traceable
attenuation
measurement.
The
relative
standard
areare
traceable
toofto
SCLs
attenuation
measurement.
The
standard
m
Similarly,
the
uncertainties
A
and
A
are
from
the
UUT
discrimination
the
final
is
the
maximum
amplitude
or
position
of
the
is
the
echo
maximum
signal,
and
amplitude
or
position
of
A
A
1SCLs
0the
1
1
readingand
and
the
initial
reading,
respectively.
The
standard
measurement
uncertainties
of
A
Similarly,
the
uncertainties
of
A
and
A
are
from
UUT
discrimination
of
the
final
A0 are
equalmeasurement
to 0.5asa1LSD
/ 3.0
The
accuracy
isoscilloscope
calculated
as
the
sum
resolution
andmeasurement
isasassigned
percentage,
and
calculated
as
1The
/ the
256
of
8voltage
It (V
/assigned
div
setting)
/1 measured
as the
voltage
measurement
uncertainty
of
the
calibration.
is
evaluated
a
The
voltage
measurement
accuracy
is
calculated
as
the
sum
of
resolution
and
is
Type
Bvoltage
uncertainty
with
a
rectangular
distribution.
assigned
voltage
uncertainty,
u(v),
is
obtained
from
the
measurement
perfor
uncertainty,
u(v),
is
obtained
from
voltage
measurement
perform
reading
and
the
initial
reading,
respectively.
The
standard
measurement
uncertainties
of
A
A0reading,
isisthe
minimum
amplitude
or
ofAthe
theas
minimum
signal.
a percentage,
andoscilloscope
calculated
1uncertainties
/ 256 x amplitude
8 It
x (V
/Adivor position
0 is echo
measurement uncertainties of A1 and A0 are equal
toand
0.5
LSD
/ 3
reading
and
the
initial
respectively.
The standard
measurement
of evaluated
Ax0asare
equal
to
0.5
expressed
LSD
/ 3. as
the
voltage
uncertainty
of position
the
calibration.
is
as
a1 of
B
uncertainty
with
a /rectangular
distribution.
The assigned
as theType
voltage
measurement
of
oscilloscope
calibration.
It voltage
is of
evaluated
as a1
uncertainty
is measurement
assessed
from
the
vertical
deflection
calibration
the measurement
oscilloscope.
The
oscilloscope.
oscilloscope.
and
Atypical
equal
touncertainty
0.5
LSD
/Rise
3.thevoltage
0 are
setting)
measured
x
100
%.
where LSD is the least significant digit of the
The
3.4
Transmitter
Voltage,
Time,
Duration
and
Reverberation
anddisplay.
A0 are equal
to
0.5
LSD
/
3.
Type
Boscilloscope
uncertainty
a in
rectangular
distribution.
The
assigned
voltage
3.5
Amplifier
Frequency
Response
assessed
from
the
deflection
calibration
the the
oscilloscope.
The
Typeuncertainty
B uncertainty
withSimilarly,
a rectangular
distribution.
assigned
voltage
measurement
digitalis
iswith
8-bit
voltage
measurement,
theof
resolution
equals
1measurement
/ 256
the vertical
uncertainties
of The
Aand
Aso
are
from
the
UUT
uncertainties
discrimination
of of
A1full
and
of thA
1 and
0 Similarly,
3.4 uncertainty
Transmitter
Voltage,
Rise
Time,
Duration
Reverberation
value of LSD equals 1 % of full scale.
is
assessed
from
the
vertical
deflection
calibration
of
the
oscilloscope.
The
digital3.4
oscilloscope
is 8-bit
inequals
measurement,
so
theand
resolution
equals
1 the
/ sum
256
ofisof
full
uncertainty
is assessed
from
the
vertical
deflection
calibration
of
the
oscilloscope.
The
scale
where
full
scale
eight
vertical
divisions.
As
a
result,
the
resolution
expressed
The
voltage
measurement
accuracy
is
calculated
as
sum
resolution
The
voltage
measurement
accuracy
is
calculated
as
the
resolution
ana
Transmitter
Voltage,
Rise
Time,
Duration
Reverberation
reading
and
the
initial
reading,
respectively.
reading
The
standard
and
the
initial
measurement
reading,
uncertaintie
respectively
3.5 model
Amplifier
Frequency
Response
3.4
Transmitter
Voltage,
Rise
Time,
Duration
and
Reverberation
The measurement
for
transmitter
voltage,
rise
time,
and equals
reverberation
isof full
digital
oscilloscope
iscalculated
8-bit
in
voltage
measurement,
so setting)
theduration
resolution
1 / 256
The
measurement
model
for
amplifier
frequency
is
scale
where
full
equals
eight
vertical
divisions.
As
adiv
result,
the
resolution
isof
expressed
digital
oscilloscope
isscale
8-bit
in
voltage
measurement,
so8uncertainty
the
resolution
equals
1 response
/to256
full
as a percentage,
and
as
1
/
256
(V
/
/
measured
voltage
100
%.
where
as
the
voltage
measurement
uncertainty
of
the
oscilloscope
calibration.
It
isev
as
the
voltage
measurement
of
the
oscilloscope
calibration.
It
is
and
A
are
equal
to
0.5
LSD
/
3.
and
A
are
equal
0.5
LSD
/
3.
0 for
0 andresponse
The
measurement
model
forrise
amplifier
frequency
is
where
The measurement
model
transmitter
voltage,
time, duration
reverberation
is is expressed
3.2 Display Jitter
scale
where
full
scale
equals
eight
vertical
As
ameasured
result,
thevoltage
scaleas
where
fullmeasurement
scale
equals
eight
divisions.
arectangular
result,
the
resolution
isresolution
expressed
a percentage,
and
calculated
as
1transmitter
/ 256
8with
with
(Vdivisions.
/ adiv
setting)
/ duration
assigned
100
%.is voltage
TypeBvertical
Bfor
uncertainty
distribution.
The
assigned
voltagem
Type
uncertainty
aAs
rectangular
distribution.
The
The
model
voltage,
rise
time,
and
reverberation
measurement
transmitter
voltage,
rise time,
duration
and reverberation
isvoltage
amodel
v , forand
(4)
vma=percentage,
The measurement model for display jitter is The
as
calculated
as8assessed
1 /(V
256
the
8setting)
frequency
(Vthe
/ div
setting)
/voltage
measured
of
100the
%. oscil
3.5 andAmplifier
Frequency
Response
as a percentage,
calculated
as
1center
/ 256isfrequency
is
div
/3.4
measured
calibration
100
%. (5)
is
the
band,
uncertainty
from
vertical
deflection
assessed
from
the
deflection
calibration
theTime,
oscilloD
3.4 uncertainty
Transmitter
Voltage,
Rise
Time,
Duration
Transmitter
and
Reverberation
Voltage,
Rise
fo /=of
, vertical
(5)of
is the center frequency
of
the
frequency
band,
(4)
vm = a v ,
3.5 vmAmplifier
Frequency
Response
where
is
the oscilloscope
upper
3 dB frequency,
digital
digital
oscilloscope
8-bitininvoltage
voltagemeasurement,
measurement,sosothe
theresolution
resolutionequals
equals1
is is8-bit
=
a
v
,
(4)
is the upper Response
3 dB frequency,
where
vm = a v3.5
,
(4)
= ! ! ,
Amplifier
Frequency
(2) is
= (eight
vertical
).vertical
3.5
Amplifier
Frequency
Response
The measurement
model
for amplifier
response
thewhere
lower
3full
dB
frequency,
and
scale
where
scale
equals
eight
divisions.
As
a and
result,
resolutio
scale
full
scale
equals
divisions.
As
a result,
thethe
resolution
The measurement
forfrequency
transmitter
The is
measurement
rise
time, duration
model
for
transmitter
reverberation
voltis
is the lowermodel
3 dB frequency,
and voltage,
where
The measurement
model
for
amplifier
frequency
response
is
is
the
center
frequency
of
the
frequency
band,
is
the
3
dB
bandwidth.
as
a
percentage,
and
calculated
as
1
/
256
(V
/
div
setting)
/
measured
volt
as
a
percentage,
and
calculated
as
1
/
256
(V
/
div
setting)
/
measured
voltag
where
14
,
model
istransmitter
the for
3 dB
bandwidth.
where
where
!
!
vm ismeasurement
the measured
pulse voltage
or response
reverberation,
The
amplifier
frequency
is
where
The measurement
model
amplifier
frequency
response
is
is
the
upper
3
dB
frequency,
v for
= fao =isv
aand
(5)v ,
(4)
vm = band,
mtransmitter
vm is thea measured
pulse
is the attenuation
in
ratio ofor
thereverberation,
attenuator,
, voltage
the
frequency
offixed
the frequency
,voltage
center
J is the amplitude or time base jitter of the echo signal,
is the
lower
3Amplifier
dB frequency,
The
uncertainties
of and
and
Response
from the 3 dB attenuator accuracy, which
3.5
Frequency
3.5
Amplifier
Frequency
Response
v
is
the
measured
transmitter
pulse
voltage
or
reverberation,
are
m
and
are
from
the
3
dB
attenuator
accuracy,
which
The
uncertainties
of
is measured
thev
attenuation
voltage
ratio
attenuator,
and
is athe
transmitter
voltage
or fixed
reverberation,
= in
(5)
fodelta
is the
theofvoltage
cursors
of the
reading
ispulse
the
upper
3 the
dB
frequency,
oscilloscope.
, of
A1 is the maximum amplitude or position of thevmecho
signal,
and
isfrom
the
3in
dB
bandwidth.
=
(
, ).
its
calibration
(performed
by the
National
Physical Laboratory, NPL, in
a isdelta
the
attenuation
voltage
ratio
of
the
fixed
attenuator,
and
=
(5)
f
where
where
where
oof
from
its
calibration
(performed
by
the
National
Physical
Laboratory,
NPL,
in
v
is
the
reading
the
voltage
cursors
of
the
oscilloscope.
a
is
the
attenuation
in
voltage
ratio
of
the
fixed
attenuator,
and
=
,
(5)
f
and
=
( 3dB
).bandwidth.
= (
).is
The
uncertainties
offrequency
andpulse
of are
fromvmor
the
3band,
dB
attenuator
accuracy,
which
is as
is the
the
frequency
vuncertainty
transmitter
voltage
isreverberation,
the
measured
transmitter
pulse
voltage
where
m is the
measured
14is standard
Thestandard
standardmeasurement
of
vcenter
to the
uncertainties
of a and
v.frequency
Relative
uncertainty,
assessed
from
its
attenuation
calibration
results
that
m is dueu(a),
theis the
center
frequency
of
the
band,
The uncertainties of A1 and A0 are from the UUT discrimination
from
itsuncertainty
calibration
(performed
by
National
Physical
Laboratory,
NPL,
in
the
3.6
Equivalent
Input
Noise
14
The
standard
of
v
is
due
to
the
standard
uncertainties
of
a
and
v.
Relative
is
the
upper
3
dB
frequency,
a
is
the
attenuation
in
voltage
ratio
of
the
fixed
a
is
the
attenuator,
attenuation
and
in
voltage
ratio
of
the
fix
mthe
SCLs
3.6
Equivalent
Input
Noise
14
standard
uncertainty,
u(a),
assessed
its
calibration
results
that
The
standard
uncertainty
of
vuncertainties
is attenuation
due
standard
uncertainties
aattenuator
and
v.
Relative
aremeasurement
traceable
to
The3 of
standard
=
,are
(5) measurement
m
,from
(5)
fto
isfrom
theattenuation
upper
3relative
dB
frequency,
and
the
dB
accuracy,
The
of
o f=ois
measurement.
of the final reading and the initial reading, respectively.
The
standard
Kingdom).
standard
measurement
uncertainty,
u(a),
is
assessed
from
its
attenuation
calibration
results
that
is
the
lower
3
dB
frequency,
and
v
is
the
delta
reading
of
the
voltage
cursors
v
of
is
the
the
oscilloscope.
delta
reading
of
the
voltage
cursors
is
the
center
frequency
of
the
frequency
band,
attenuation
from
are measurement
traceable
to uncertainty,
SCLs
measurement.
The
relative
standard
measurement
standard
is assessed
from
calibration
results
that with an
uncertainty,
obtained
theits
voltage
measurement
performed
attenuation
is
theinput
lower
3 dB
and
which
isisu(a),
assessed
from
its
(performed
by the
National
measurement uncertainties of A1 and A0 are equal to 0.5 x are
LSD
/ 3
. u(v),
The
measurement
model
for
is frequency,
). 3 dB
=
(
calibration
tois
SCLs
measurement.
Thenoise
relative
standardwith
measurement
equivalent
).
is theattenuation
3measurement.
dB
bandwidth.
=
is
the
upper
frequency,
measurement
model
for
equivalent
input
noise
is
uncertainty,
u(v), The
obtained
from
the
voltage
measurement
performed
an
are traceable
totraceable
SCLs
attenuation
The
relative
standard
measurement
oscilloscope.
United
is the 3Kingdom).
dB bandwidth.
Physical
Laboratory
in the
3.6u(v),
Equivalent
Input(NPL)
Noise
uncertainty,
is
obtained
from
the
voltage
measurement
performed
is
due
to
the
standard
uncertainties
is an
due
v. to
R
The
standard
uncertainty
of
v
The
standard
uncertainty
vamand
is
the
lower
3
dB
frequency,
and
oscilloscope.
uncertainty,
u(v), is obtained from the voltagem measurement performed
with an of with
14 14
3.3 Stability Against Voltage Variations
oscilloscope.
and
are
from
the
3
dB
attenuator
accuracy,
whi
The
uncertainties
of
standard
measurement
uncertainty,
u(a),
is
standard
assessed
measurement
from
its
attenuation
uncertainty,
calibration
u(a),
is
resu
a
is
the
3
dB
bandwidth.
V / The
as
oscilloscope. The voltage
measurement
accuracy
the sum of of
resolution
is from
assigned
ninis=calculated
(6)the 3
,
are
uncertainties
and and
3.6
Equivalent
Input
Noise
model
forattenuation
equivalent
noise
is
The measurement model for
voltage variations is The
nin = Vein
,
ein /input
from its
(performed
byare
thetraceable
National
Physical
Laboratory,
NPL,
=stability
! against
are measurement
traceable
tocalibration
SCLs
measurement.
The
to
relative
SCLs
attenuation
standard (6)
measu
meas
! ,
Theasvoltage
measurement
accuracy
is
calculated
as
the
sum
of
resolution
and
is
assigned
the voltage
measurement
uncertainty
of
the
oscilloscope
calibration.
It
is
evaluated
as
a
from
its
calibration
(performed
by
the
National
The measurement modelVfor
equivalent
input noise is
=uncertainties
Vinfrom
/ 10(40/20),
eincalculated
The
voltage
measurement
accuracy
is
as
the
sum
of
resolution
and
is
assigned
Kingdom).
uncertainty,
u(v),
is
obtained
uncertainty,
the
voltage
u(v),
measurement
is
obtained
performed
from
wt
The
of
and
are
from
the
3
dB
attenua
=the
VKingdom).
/ 10(40/20),
as the
voltage
uncertainty
of theVein
oscilloscope
It isisvoltage
evaluated
as a
The
voltage
accuracy
calculated
as
ofcalibration.
resolution
and
assigned
Typemeasurement
Bmeasurement
uncertainty
with
aisrectangular
distribution.
The assigned
measurement
in sum
(3)oscilloscope.
of
=the
(oscilloscope
oscilloscope.
), (performed
= ! ! ,
as
the
voltage
measurement
uncertainty
calibration.
It
is
evaluated
as
a
from
its
calibration
by
the
National
Physical
L
(
), assigned
Buncertainty
uncertainty
a rectangular
The
as theType
voltage
measurement
uncertainty
thendistribution.
calibration.
It voltage
is of
evaluated
as (6)
a (6) The
=
Vein=
/ deflection
is with
assessed
fromofthe
vertical
calibration
the measurement
oscilloscope.
,
inoscilloscope
Type
Boscilloscope
uncertainty
a vertical
rectangular
distribution.
assigned
voltage
Equivalent
Input
Noise
Kingdom).
assessed
from
the
deflection
calibration
theInput
oscilloscope.
The of full
Typeuncertainty
B uncertainty
with a 3.6
rectangular
distribution.
assigned
voltage
measurement
digitalis
iswith
8-bit
in
voltage
soThe
theof
resolution
equals 1measurement
/ 256
nmeasurement,
/ (100
),
3.6
Equivalent
Noise
in = VThe
in
Vvertical
V/in10(40/20),
/ (100
),
ein =nV
in=deflection
where
in
assessed
from
the
calibration
ofthemeasurement
the
Theis ca
The
voltage
measurement
accuracy
The
voltage
asequals
the
sum
resolution
and
a
digitaluncertainty
oscilloscope
is
8-bit
voltage
measurement,
so is
thecalculated
resolution
1 oscilloscope.
/of256
ofisaccuracy
full
uncertainty
is assessed
from
theinequals
vertical
deflection
calibration
of
the
oscilloscope.
The
scale
whereisfull
scale
eight
vertical
divisions.
As
a result,
resolution
expressed
=measurement,
(for
), ofas
S is the amplitude or time base stability of the echo signal,digital oscilloscope
is
8-bit
in
voltage
so
the
resolution
equals
1
/
256
of
full
The
measurement
model
equivalent
input
noise
is
as
the
voltage
measurement
uncertainty
the
the
oscilloscope
voltage
measurement
calibration.
uncertainty
It
is
evaluat
of
3.6
Equivalent
Input
Noise
scale
where
full
scale
equals
eight
vertical
divisions.
As
a
result,
the
resolution
is
expressed
digital
oscilloscope
is
8-bit
in
voltage
measurement,
so
the
resolution
equals
1
/
256
of
full
as a percentage, and calculated as 1 / 256 8 The
(Vmeasurement
/ div setting)model
/ measured
voltage input
100 noise
%. is
for equivalent
where
A1 is the maximum amplitude or position of thescale
echowhere
signal,
andscale
scale
where
full
scale
equals
divisions.
As
result,
the
expressed
Type
Beight
uncertainty
with
distribution.
Type
B
uncertainty
The
assigned
with
aisvoltage
rectangular
measu
di
where
full
equals
vertical
divisions.
a result,
the
resolution
isresolution
expressed
nvertical
(100
as a percentage,
and
calculated
as 1 eight
/ 256
in 8=aVrectangular
(V
/ div
setting)
/ ameasured
voltage
100
%.
),
in / As
A0 is the minimum amplitude or position of theasecho
signal.
as
a percentage,
and
calculated
as8 1The
/(V
256
vertical
(V / div
setting)
/voltage
measured
voltage
100vertical
%.
uncertainty
from
the8setting)
uncertainty
deflection
is
calibration
assessed
of
from
thethe
oscilloscope
measurement
model
for equivalent
input
noise
is
3.5
Frequency
a percentage,
andAmplifier
calculated
as
1is/ assessed
256 Response
/ div
/
measured
100
%.
nin = Vein / ,
(6)
where
oscilloscope
is root
8-bitbandwidth
in voltageinmeasurement,
digital
oscilloscope
so thenresolution
is V8-bit
inequals
voltage
1 / meas
256
in =
ein / ,
3.5
Amplifierdigital
Frequency
Response
n
per
V
/
Hz,
in is the noise
Hz
where
,
is thefull
noise
per
root bandwidth
/divisions.
Hz,
3.5
Amplifier
Frequency
Response
the UUT
discrimination
Similarly, the uncertainties of A1 and A0 are from3.5
Veinvertical
= response
Vinin V/scale
10(40/20),
scaleninwhere
equals
eight
full
As ascale
result,
equals
the resolution
eight vertical
is exd
Amplifier
Frequency
Response
The measurement
model
forscale
amplifier
frequency
is where
Vein = Vin / 10(40/20),
n = V / ,
is the input
signal
amplitude
in
volts
(peak-to-peak),
of the final reading and the initial reading, respectively.
The standardas
),percentage,
1=
(
a Vpercentage,
and
calculated
as
/ 256
as
8a
(Vin/ div ein
setting)
and
calculated
/ measured
asvoltage
1 / 2561
Vinin for
is the
input
signal
amplitude
in
volts
(peak-to-peak),
The measurement
model
amplifier
frequency
response
is
=
( ),
Veinmodel
is the for
equivalent
input
noise inresponse
volts, is Vein = Vin /
The
measurement
amplifier
frequency
measurement uncertainties of A1 and A0 are equalThe
to 0.5
x
LSD
/
.
10(40/20),
3
is the per
equivalent
input
noise
measurement model
amplifier
frequency
response
isVin
nin for
is Vthe
root bandwidth
ein noise
Hz,
ninin=
V/involts,
/ (100 ),
is
the
center
Frequency
, frequency, Response
(5)
),
3.5 fo =
Amplifier
3.5
Amplifier
=
), Vin / (100Response
= ( nFrequency
in
is
thecenter
frequency,
V
is
the
input
signal
amplitude
in
volts
(peak-to-peak),
is
the
upper
3
dB
frequency,
(5)
fo =in ,
3.4 Transmitter Voltage, Rise Time, Duration and Reverberation
the
upper
(
, ).3 dB frequency,
= is=
f
n(5)
lower
in = Vin / (100 ),
Veinis
equivalent
noise
in volts,
the
33input
dB
and
othe
, is
(5) The measurement
fo =
dB frequency,
frequency,
and
measurement
for
amplifier
frequency
response
is model for amplifier frequen
The measurement model for transmitter voltage, rise time, duration
isthe
lowermodel
3 dB
frequency,
and
where
).lower
The
= (
where
is
the
center
frequency,
is
the
3
dB
bandwidth.
14
is
the
3
).
=
(
and reverberation is
= (
).is the 3 dB bandwidth.
is the
upper 3 dB frequency,
14
where
nThe
per
V to
/ Hz,
fonoise
=
root
, bandwidth
fo =(5)
, of Vin and
in
in is the
lower
3
dB
frequency,
andof 14
is theThe
standard
uncertainty
n is in
due
theper
standard
uncertainties
14
due
to
standard
uncertainties
root bandwidth
V of
/ Hz,
vm = a x v ,
(4)
duenoise
to the
the standard
uncertainties
Vin and
Thestandard
standarduncertainty
uncertaintyof
of nnininin isis the
ismeasurement
the
3
dB
bandwidth.
uncertainty,
u(),
is
assessed
from
the
bandwidth
in S
of
and
.uncertainty,
uncertainty,
u(
),=is(assessed
Vin is
the
input
signal
volts
(peak-to-peak),
).u(), isinassessed
measurement
).
=Standard
(
amplitude
measurement
measurement
from
the
bandwidth
measurement
in
Vin is
the
input
signal amplitude
inType
voltsB(peak-to-pea
nin uncertainty,
is the noise
per
root
bandwidth
in V / Hz,
Standard
measurement
u(V
),
is
obtained
through
a
evaluation.
in
Veinthe
is the
equivalent
input
noise in
volts,
from
bandwidth
measurement
in
Section
3.5.
Standard
measureStandard
measurement
uncertainty,
u(V
),
is
obtained
through
a
Type
B
evaluation.
where
in equivalent
input noise inof(Section
volts,
ein istothe
14
The
standard
uncertainty
ofisthe
nobtained
isVmeasurement
due
theamplitude
standard
VItin and 3.4),
. an
St
of
two
components:
voltage
ininput
the
center
frequency,
ment
uncertainty,
u(Vaain),is
through
ausing
Typeuncertainties
Bvolts
evaluation.
signal
inoscilloscope
(peak-to-peak),
is
of two
components:
voltage
measurement
using
oscilloscope
(Section
3.4), an
vm is the measured transmitter pulse voltage or reverberation,
is
the
center
frequency,
measurement
uncertainty,
u(),
is
assessed
from
the
bandwidth
measurement
in
Secti
reading
fluctuation.
It
is
assessed
to
have
a
rectangular
distribution
with
limits
is
the
upper
3
dB
frequency,
VIteinisis the
equivalent
input
in volts,
consists
of two components:
a voltage
measurement
using
oscillo- with limits
reading
fluctuation.
assessed
have
a noise
distribution
a is the attenuation in voltage ratio of the fixed attenuator, and
istothe
upper
3rectangular
dB
frequency,
noise
Standard
u(V
islevel
obtained
through
a assessed
Type B evaluation.
It c
maximum
offrequency,
reading.
in),
measurement
is(Section
thefluctuation
lower
3uncertainty,
dB
and
and
isthe
theUUT
center
frequency,
scope
3.4),
the
UUT
reading
fluctuation.
It
is
of
maximum
fluctuation
the
UUT
noise
level
reading.
v is the delta reading of the voltage cursors of the oscilloscope.
is the lower
dB frequency,(Section
and to 3.4), and the
of two
components:
abandwidth.
voltage
measurement
using3 oscilloscope
is
the
3
dB
is
the
upper
3
dB
frequency,
have a rectangular distribution
with
limits
0.5
x maximum fluctua
Attenuator
is
the 3of
dB
reading
fluctuation.
It of
is Calibrated
assessed
to
have
rectangular
3.7
Accuracy
/ bandwidth.
Gainanddistribution with limits of
is thereading.
lower
3 dB afrequency,
tion
UUT noise
level
3.7 of the
Accuracy
ofCalibrated
Attenuator
/ Gain
The standard uncertainty of vm is due to the standard uncertainties
maximum fluctuation
of the
reading.
The standard
uncertainty
UUT
is thenoise
3 of
dBnlevel
bandwidth.
in is due to the standard uncertainties of Vin and
of a and v. Relative standard measurement uncertainty, u(a), is
The
standard
uncertainty of nin is due isto the stan
Themeasurement
measurementuncertainty,
model for the
accuracy
of the calibrated
u(),
is assessed
from the attenuator/gain
bandwidth measurement
3.7 Accuracy
of Calibrated
Attenuator
Gain
The
measurement
model for the
accuracy
of/the
calibrated
attenuator/gain
is from ti
assessed from its attenuation calibration results that are traceable to
measurement
uncertainty,
u(), is assessed
3.7
Accuracy
of Calibrated
Attenuator
/inGain
Standard
measurement
uncertainty,
u(V
),
is
obtained
through
a
Type
B
evaluatio
The
standard
uncertainty
of
n
is
due
to
the
standard
uncert
in
accuracymeasurement
of the
attenuator/
15 calibrated
uncertainty,
u(Vin), is obtaine
SCLs attenuation measurement. The relative standard measurement The measurement model for theStandard
15 using oscilloscope (Section
of
two
components:
a
voltage
measurement
3.4),
measurement
uncertainty,
u(),
is
assessed
from
the
bandwid
gain is
of two
components:
a voltage measurement
using
uncertainty, u(v), is obtained from the voltage measurement
The measurement
model for
the
accuracy
of the
calibrated
attenuator/gain
is
reading fluctuation.
It is
assessed
to have
a rectangular
with lim
Standard
measurement
uncertainty,
u(Vin), distribution
is obtained
through
a
reading fluctuation. It is assessed to have a recta
performed with an oscilloscope.
maximum fluctuation
of
the
UUT
noise
level
reading.
of Etwo
components:
a
voltage
measurement
using
oscillosco
= 20 x maximum
log10 (R / R15
(7) level reading
fluctuation
of the UUT noise
o),
The voltage measurement accuracy is calculated as the sum
reading fluctuation. It is assessed to have a rectangular dist
of resolution and the voltage measurement uncertainty of the
3.7
Accuracymaximum
of Calibrated
Attenuator
/ Gain
fluctuation
of the UUT
noise level Attenuator
reading.
3.7
Accuracy
of Calibrated
/ Gain
oscilloscope calibration. It is evaluated as a Type B uncertainty where
with a rectangular distribution. The assigned voltage measurement R is the UUT amplitude as a percentage of full screen height at different
The measurement 3.7
model for
the accuracy
of the calibrated
attenuator/gain
is
Accuracy
of Calibrated
Attenuator
/ Gainof the
attenuator/gain settings,
The measurement
model
for the accuracy
calib
uncertainty is assessed from the vertical deflection calibration
of the oscilloscope. The digital oscilloscope is 8-bit in voltage Ro is the reference signal amplitude as a percentage15of full screen height,
15
The measurement model for the accuracy of the calibrated attenu
measurement, so the resolution equals 1 / 256 of full scale where and
15
full scale equals eight vertical divisions. As a result, the resolution E is the measurement error in decibels.
2014
Vol. 9 No. 3 September 2014
67
63
TECHNICAL PAPERS
The standard uncertainty of E is due to the display resolution, dr,
and the 10 dB step attenuator accuracy, a. The standard measurement
uncertainty for display resolution, u(dr), is calculated as 0.5 x LSD
/ 3
where LSD is the resolution or fluctuation of the UUT signal
amplitude reading. The standard measurement uncertainty due to step
attenuator accuracy, u(a), is assessed from the results of attenuation
measurements.
3.8 Linearity of Vertical Display
The measurement model for the linearity of the vertical display is
E = A1 A0 = A1 80 x 10 (2 - a) / 20,
(8)
where
A0 is the target amplitude as a percentage of full screen height,
a is the attenuation of the step attenuator in decibels,
A1 is the UUT amplitude reading as a percentage of full screen height,
and
E is the measured error as a percentage of full screen height.
The standard uncertainty of E is due to the standard uncertainty
of a and A1. The standard measurement uncertainty due to the step
attenuator accuracy, u(a), is assessed from the results of attenuation
measurements. The standard measurement uncertainty, u(A1), is
calculated as 0.5 x LSD / 3
, where the LSD is the resolution of the
UUT.
3.9 Linearity of Time Base
Tests
Clause 9.3 Stability
Stability after warm-up time
1%
Display jitter
1%
1%
3%
3%
Reverberation
3%
Duration (50 ns to 2 s)
1%
2%
3 dB bandwidth
3%
3%
Internal attenuator/gain
0 dB to 110 dB
D = P0 P1,
Clause 8.8.2
The measurement model for the linearity of the time base for digital
ultrasonic instruments is
68
1%
Clause 9.6
where
L is the linearity error as a percentage of full scale,
D1 is the delay time setting of the delay/pulse generator,
D0 is the delay time from a linear best fit curve, and
D is the delay time of full scale.
0.3 dB
(9)
where
D is the deviation from the reference position,
P0 is the reference position, and
Calibration and
Measurement
Capability (CMC)
(10)
1%
1%
TECHNICAL PAPERS
4. Conclusions
69
TECHNICAL PAPERS
Abstract: This paper introduces a measurement uncertainty model and analyzer tool being developed for one of the worlds
largest space environmental test facilities, the Spacecraft Propulsion Research Facility (B2) located at NASA Glenn Research
Centers Plum Brook Station near Sandusky, Ohio. The B2 is the worlds only facility capable of testing full-scale upper-stage
launch vehicles and rocket engines under simulated high-altitude conditions.
Developing an uncertainty tool for the data acquisition of a test facility of this scale presents unique metrology challenges.
Not only must the uncertainty analyzer tool be versatile enough to accommodate a wide range of disciplines and measurement
requirements (such as temperature, pressure, strain, vacuum, and acceleration), but it must provide a user-interactive platform for
evaluating system measurement uncertainty based on customer-chosen measurement scenarios ranging from the most simplistic
tests to the most complex ones. The uncertainty analyzer tool, which was developed in Microsofts Visual Basic for Applications (VBA) in Excel, will serve multiple purposes, including aiding in the optimal selection of measuring and test equipment,
communicating capabilities to customers, and supporting all decisions based on measurements. Although the analysis tool was
developed for the data acquisition system in B2, it can be easily sized to fit other data acquisition systems at the site utilizing
similar measurement methods. This paper outlines the methodology followed, the features of this tool, and how the tool can be
applied to the measurement processes of different facilities.
1. Introduction
NASA Glenn Research Centers Calibration Laboratory was tasked with providing a Measurement Uncertainty Analysis
(MUA) tool for the data acquisition system
at the Spacecraft Propulsion Research Facility (B2) at Glenns Plum Brook Station campus. The MUA tool would need
to accommodate all of the types of measurements required for typical tests in the
B2, including voltage, temperature, strain,
accelerometer, and pressure. In the first
phase, Glenns Calibration Laboratory would
provide the MUA tool for the measurement
disciplines used most frequently for facility
tests: temperature and pressure.
It is essential to understand the measurement path for which the analysis was developed in order to ensure the correctness
of each measurement uncertainty analysis.
Therefore, this paper begins with a short presentation about the measurement path for the
B2 data acquisition system. Because different propellants can be used during tests and
70
because of their known hazards, the measurement path of the B2 data acquisition system
is located in three different places that are all
interconnected during tests.
In the first location, referred to as the test
building, the test article is loaded in a thermal vacuum chamber lined with a liquid
nitrogen cold wall capable of maintaining
195.5 C; the sensors also are located in
the vacuum chamber. The second location,
which is immediately next to the test building, is referred to as the data room. The data
room contains the rest of the measuring and
test equipment. The connections between the
sensors and the rest of the measurement path
are made through cables and chamber feedthroughs. To protect personnel during the
tests, NASA controls all tests from the third
location, which is a control room located
about 790 m from the test site.
The data acquisition system has a matrix
of 18 subsystems, and each subsystem has 32
channels, for a total of 576 available channels. In general, each channel path compris-
Author
Mihaela Fulop
mihaela.fulop-1@nasa.gov
SGT, Inc., Metrology Services
NASA Glenn Research Center
Calibration Laboratory
MS 217, 21000 Brookpark Road
Cleveland, OH 44135
TECHNICAL PAPERS
Error
containment
probability,
%
Uncertainty
type
Distribution
Divisor
Standard
uncertainty,
ui, V
Sensitivity
coefficient,
ci
Product of
standard
uncertainty
and sensitivity
coefficient, V
Degrees
of
freedom
(DOF)
Error
source
Specifications
Error
limits, V
DC (gain,
offset error)
0.10 %
0.0001000
95
Normal
1.96
0.00005
100.000
0.00510
Quantization
error
0.00015 V
0.0001526
95
Normal
1.96
0.00008
1.000
0.00008
Common
voltage
error
100 dB
0.0000005
95
Normal
1.96
0.00000
100.000
0.00003
Crosstalk
110 dB
0.0000002
100
Rectangular
1.73
0.00000
100.000
0.00001
0.0000000
100
Rectangular
1.73
0.00000
100.000
0.00000
uADC
0.00510
Nonlinearity
Effective
DOF
Expanded
UADC (95 %
confidence)
0.01001
the static MUA, or measurement capability). To find the expanded uncertainty for the
measurement process, we combine the static MUA values with the Type A uncertainty
components obtained as described earlier,
following rules in the Guide to the Expression of Uncertainty in Measurement (GUM)
[1] and in the U.S. versionof the guide [2].
2. Project Approach
71
TECHNICAL PAPERS
Summary points
Point
Temperature, C
Uncertainty, 95 %
confidence, C
Temperature of cold
junction, Tref , C
1
2
3
200
201
203
3.881
3.880
3.877
24
24
24
TECHNICAL PAPERS
those on the original spreadsheets. Because
the Link with module accuracy files will
automatically open the accuracy file workbooks
from their network locations, the VBA code for
this command button will also require minimum
modifications to match the network locations of
all the accuracy files at the new facility.
5. Validation of the Measurement
Uncertainty through Repeatable
Measurements
After the uncertainty is determined, it is critical to ensure its validity. By definition, the uncertainty is a range of values that are expected
to contain the true value with a specified level
of confidence. If the uncertainty is estimated
correctly, measurement results obtained at different times should be consistent with the reported uncertainty. This ensures a reliable and
repeatable measurement quality system. The
estimated standard uncertainty will be compared with the standard deviation of a series
of measurements for a test item so that the uncertainty of the measurement can be validated.
Because of the costs involved in this endeavor,
the series of measurement data will be mined
from historical data for similar tests. Statistical process control will also be employed to
make sure that the measuring system maintains in control. Statistical process controls
and validation of the measurement uncertainty
represent the second phase of this project and
are not presented in this paper.
6. Conclusions
100
2
5.5
A. AppendixMeasurement Uncertainty
Analysis for Thermocouple Measurement at the B2 Facility
This appendix describes the measurement uncertainty analysis for a measurement of type-E
thermocouples (TCs) with the data acquisition
system at the Spacecraft Propulsion Research
Facility (B2) at Glenns Plum Brook Station
campus. The analysis in this appendix follows
the system approach in Measurement Uncertainty Analysis Principles and Methods, Annex 3 of a NASA handbook [3]. According
to Annex 3, a system is composed of modules
arranged in series. Because of this series arrangement, the output of any module and its
associated uncertainty represents the input of
the next module.
The TC measurement path determined with
NASAs approach [3] for the data acquisition
system in the B2 is captured in Fig. 3. The
uncertainty analysis is documented for a system input of 200 C, a reference temperature
of 24 C, and the specific conditions listed in
Table 3. It is assumed that the mounting of the
TC does not introduce any errors in the circuit.
Table 4 is a list of the components in the B-2
TC measurement system. The modules for the
B-2 type-E TC measurement system are described in Sections A.1 to A.4.
A.1. Seebeck Temperature Module (M1)
73
TECHNICAL PAPERS
Element component
Manufacturer
Model
Notes
TCPVCE24180, 320 to
250 F
----------------------
OMEGA
-------------------------
25 ft long
TC connector block on
chamber platform
Marlin Manufacturing
Corporation
----------------------
-------------------------
-------------------------
-------------------------
-------------------------
----------------------
----------------------
----------------------
Hy-Cal Engineering
401B
----------------------
-------------------------
-------------------------
40 ft long
Copper terminals
-------------------------
-------------------------
----------------------
Copper wires
-------------------------
S12/S2/C
Precision Filters
28608B; gain = 1
----------------------
Analog-to-digital convertor
DSPCon
9843001004, digitizer
Gain = 100
DSPCon
CCP002 TEFLT
----------------------
Type-E TC
Signal-conditioning/filter
amplifier
E ci (T )n
i 0
Eref ci (Tref )n ,
(A.2)
i 0
E
T
Sref
Eref
.
Tref
(A.3)
where
Figure 4 shows a block diagram of the Seebeck temperature module.
Y1 is the measured voltage, mV;
The error equation for the M1 module can be written as
Y1 cT T cTref Tref cconnectors connectors cextension wires
S is the average Seebeck coefficient, mV/C at T;
extension
(A.4)
wires
Sref is the Seebeck
mV/C
at
T
of
cold
junction;
,
Y1 ccoefficient,
T T
Tref
Tref
connectors connectors
extension wires extension wires
T is the measured temperature, C; and
where ci are the sensitivity coefficients obtained in Eqs. (A.5) to (A.8):
Tref is the temperature of cold junction, C.
74
TECHNICAL PAPERS
Y1
S ,
T
(A.5)
Y1
STref ,
Tref
(A.6)
cT
cTref
Y1
cconnectors
conectors
S,
Y1
S .
extension wires
cextension wires
(A.7)
(A.8)
uY1
2
connectors
transfer function is
Y3 GADC Y2 ADC ,
(A.14)
where
GADC is the ADC gain;
Y2 is the SC output, V;
Y3 is the output (V) corresponding to ADC counts equal to an ADC
input of Y2; and
eADC is the error due to ADC. The components of this error are
captured in the second column of Table 1.
The error equation for the M3 model is
(A.9)
cT T 2 cTref Tref cconnectors connectors 2 cextension
wires extension wires
cY2
2
cextension wires extension wires .
(A.15)
Y3
GADC
Y2
cADC
Y3
1 .
ADC
(A.16)
Y2 G Y1 SC ,
(A.10)
where
G is the gain, V/mV; and
Y2 is the SC output, V.
uY3
cY Y
2
cADC ADC
(A.17)
(A.18)
2
regress
n
multiplied by G to convert it to refer to output (RTO). The eSC
G GADC
G GADC
G GADC
components are captured in the second column of Table 7.
2
n
y3
y3
y3
The error equation Tfor this module
is
a2
... an
regression
computed,C a0 a1
G GADC
G GADC
G GADC
,
Y2 cY1 Y1 cSC SC ,
(A.11)
where y3 is Y3 converted to millivolts and a0, a1a9 are obtained from [4].
The module model error is
and the sensitivity coefficients for Eq. (A.11) are
cY1
Y2
G
Y1
cSC
Y2
G .
SC
uY2
cY Y
1
cSC SC .
(A.13)
(A.12)
(A.19)
Tcomputed
y3
a1
2 a2
G GADC
G
G
ADC
2
8
y3
y3
3 a3
... 9 a9
G GADC
G GADC
Tcomputed
1
cregression
regression
c y3
y3
(A.20)
uTcomputed
Y cY regression cregression
3
(A.21)
75
TECHNICAL PAPERS
Module 1
(TC/alloy wires/connectors/cold junction)
Sensitivity
coefficient,
ci
Product of
standard
uncertainty
and sensitivity
coefficient, V
DOF
Normal
1.96
0.867 C
74.030
V/C
64.2106
Rectangular
1.73
0.000 C
74.030
V/C
0.0000
95
Normal
1.96
0.0000 V
1.0000
0.0000
0.621 C
95
Normal
1.96
0.317 C
60.860
V/C
19.2870
Maximum
gradient
temperature
across
connectors
0.200 C
95
Normal
1.96
0.102 C
74.030
V/C
7.5542
Accuracy
1.700 C
95
Normal
1.96
0.867 C
74.030
V/C
64.2106
Accuracy due
to length of
extension wires
0.000 C
95
Normal
1.96
0.000 C
74.030
V/C
0.0000
inhomogeneity
0.000 C
100
Rectangular
1.73
0.000 C
74.030
V/C
0.0000
M1
module
uncertainty
(95%, 2 )
182.5544
Error source
Error
containment,
%
Type
Distribution
TC accuracy
1.700 C
95
inhomogeneity
0.000 C
100
Error from TC
mounting
0.000
V
Cold
(reference)
junction
Cold junction
expanded
uncertainty
TC
connectors
connectors
TC
T
M1
module
output,
Y1
Divisor
Standard
uncertainty,
ui
Error
limits
Error
contributor
TC
extension
cables
extension
wires
11.987090
mV
TECHNICAL PAPERS
Module 2
Error
contributor
Signal
conditioner
Sensitivity
coefficient,
ci
Product of
standard
uncertainty
and sensitivity
coefficient, V
DOF
Normal
1.96
0.1156 mV
231.3807
Normal
1.96
93.1417 V
186.2834
Normal
1.00
582.2175
Error source
Error
containment,
%
Type
Distribution
Signal
conditioner
expanded
uncertainty
0.2267
mV
95
182.5544
V
95
95
M1 module
Noise
M2
module
output,
Y2
Divisor
Standard
uncertainty,
ui
Error
limits
1
M2 module
uncertainty
(95 %, 2 )
0.02397 V
Table 6. Uncertainty, error components, and output for M2 (DOF is degrees of freedom).
Where the
errors come
from
Components
from RTI
Components
from RTO
Error
containment
probability,
%
Uncertainty
type
Distribution
Divisor
Standard
uncertainty,
ui mV
Sensitivity
coefficient,
ci
Product of
standard
uncertainty
and sensitivity
coefficient, mV
DOF
Error
source
Specifications
Error
limits,
mV
DC accuracy
% of settings
0.20 %
0.0240
95
Normal
1.96
0.0122
1.00
0.0122
Temperature
coefficient
0.008 %/C
0.0053
95
Normal
1.96
0.00267
1.00
0.0027
Noise RTI
2.80 V
0.0196
95
Normal
1.96
0.0100
1.00
0.0100
Offset drift
RTI
1 V/C
0.0055
100
Rectangular
1.73
0.0032
1.00
0.0032
Offset drift
RTO
0 V/C
0.0000
100
Rectangular
1.73
0.0000
0.50
0.0000
Noise RTO
60.00 V
0.4200
95
Normal
1.96
0.2143
0.50
0.1071
All-hostile
crosstalk
90 dB
0.1581
95
Normal
1.96
0.0403
0.50
0.0202
uSC
0.1102
Table 7. Uncertainty contributions of the signal conditioner (RTI, refer to input; RTO, refer to
output; DOF, degrees of freedom; SC, signal conditioner).
Effective
DOF
Expanded
USC (95 %
confidence)
0.2267
wires, such as those used in the B2 data acquisition system, are also
susceptible to noise. The system incorporates a low-pass filter with
low cut-off frequency and use shielded wires to make sure unwanted
noise is eliminated.
A.5.4 Errors Due to Cold Junction Reference
77
TECHNICAL PAPERS
95%. The expanded uncertainty for the cold junction reference with a
confidence level of 95% is 0.621 C.
Eq. (A.3) was used to obtain a sensitivity coefficient of 60.86
mV/C. Therefore, the error limits due to the cold junction reference
(1 ) in units of voltage were computed to be 19.287 mV.
The error limits for the direct-current (DC) accuracy were calculated
from the manufacturer specification of 0.2 % of the input voltage to
the SC. For an input voltage of 11.9870 mV, this error is calculated
to be 0.024 mV. The distribution of this error was considered to be
normally distributed with a confidence level of 95 %.
A.6.2 Temperature Coefficient Error of the Signal Conditioner
crosstalk
(A.22)
Table 8 lists the uncertainty, error components, and output for M3. As
seen in this table, one error contributor to the error in the M3 module
is the uncertainty for M2. The error contribution due to M2 comes
from the uncertainty calculations for this module and is captured in
Table 6. The contributions due to the ADC itself are summarized in
Table1 and explained in the following subsections.
A.7.1 DC Error of the Analog-to-Digital Convertor
e
CMRR in dB 20 log(CMRR) 20 log ADC CMV ,
error
CMV
(A.23)
errorCMV
eCMV GADC
10
CMRR in dB
20
(A.24)
The error limit due to the offset drift RTI was calculated from
manufacturer specifications of 1 V/C. For a change in room
temperature of T = 5.5 C, this was calculated to be 0.0055 mV. This
error is affected by gain and was assumed to be normally distributed
with a confidence level of 95 %.
90
10 20
input voltage
This specification is stated as 110 dB @1 kHz into 50 input impedance. If the maximum difference between channels is assumed to be 50
mV, the crosstalk error Vcrosstalk for a gain of 100 is 0.000912 V. These
limits are assumed to be distributed normally with a confidence of 95 %.
The only DC accuracy specifications provided by the manufacturer are for a
gain setting of 100.
Module 3
TECHNICAL PAPERS
M3
module
output,
Y3
Type
Distribution
Divisor
Sensitivity
coefficient,
ci
Product of
standard
uncertainty
and sensitivity
coefficient, V
DOF
0.0051
100.0
0.0297
M3 output
module
uncertainty
(95 %, 2 )
0.0591
Error
contributor
Error source
ADC
ADC expanded
uncertainty
0.0100
95
Normal
1.96
0.0051
0.0006
95
Normal
1.96
0.0003
M2 module
Error
containment,
%
Standard
uncertainty,
ui, V
Error
limits,
V
2.39742 V
Module 4
Table 8. Uncertainty, error components, and output for M3 (ADC, analog-to-digital convertor; DOF, degrees of freedom).
M4
module
output,
Tcomputed
Divisor
Standard
uncertainty,
ui
Sensitivity
coefficient,
ci
Product of
standard
uncertainty
and sensitivity
coefficient, C
DOF
Error
contributor
Error source
mV
Error
limits
Error
containment,
%
Data
processor
conversion
of M3
uncertainty
0.2953
0.2953
mV
95
Normal
1.96
0.1507 mV
13.5 C/
mV
12.035
0.06 C
95
Normal
1.96
0.03 C
1.00
0.031
M4 module
uncertainty
(95 %, 2 )
3.990
Regression
Type
Distribution
200.001 C
Table 9 lists the uncertainty, error components, and output for M4.
The error contributions in Table 9 are the error due to regression and
errors due to the modified output in the M3 module. The output in M3
was converted to millivolts and was scaled by dividing it by the total
gain. The regression errors were taken from [4] and are assumed to be
normally distributed with a confidence level of 95 %.
A.9. Computing System Output Uncertainty
(A.25)
where
uTcomputed is the standard uncertainty in the system output (1 ); and
t /2,
is the Students t-distribution where = 1 p, p is the
confidence level probability, and is the degrees of freedom.
For this analysis, a confidence level of 95 % (p = 0.95) was used
in accordance with GUM guidelines. In this analysis, the degrees of
freedom are infinite ( = ), resulting in a corresponding Students
t-distribution of t /2, 1.96 . With these assumptions, the confidence
limits were computed to be 200 3.9 C.
79
C NNECT
to the Real Asset World.
ADVERTISERS INDEX
EQUIPMENT MANAGEMENT...
YOUR CUSTOMERS...
YOUR END USERS...
YOUR ERP/FINANCIAL SYSTEM...
WEB BASED ACCESS ANYWHERE IN THE WORLD...
THE CLOUD...
Additel
www.additel.com
Page 10
AIAG
www.aiag.org
Page 47
AssetSmart
www.assetsmart.com
.org
By PMSC
2800 28th Street, Santa Monica, California 90405 USA 310.450.2566 info@assetsmart.com www.assetsmart.com
CENAM
www.cenam.mx/simposio
Page 41
Page 11
Fluke Calibration
www.fluke.com
Page 13
Guildline
www.guildline.com
Page 18
Keysight Technologies
www.keysight.com/find/checkthecal
Back Cover
Measurements International
www.mintl.com
Page 8
Mensor Corporation
www.mensor.com
Page 12
Page 20
Page 21
Page 61
Page 14
NCSLI Membership
www.ncsli.org
Page 55
Ohm-Labs, Inc.
www.ohm-labs.com
Page 9
Rotronic
www.rotronic-usa.com
Page 7
Page 17
Vaisala
www.vaisala.com
Page 6
80
Qua
Certified to ISO 9001:2008 and accredited to ISO/IEC 17025:2005
Calibration
Cert No. 1388.01
Sponsors:
2014 AIAG | 26
CTR9000
CTR5000
Precision Thermometer
CTH7000
Handheld Thermometer