Radar Transmitters

SITRANS LR250 (HART) Quick Start Manual 03/2013

SITRANS

SITRANS LR250 (HART) Quick Start Manual


This manual outlines the essential features and functions of the SITRANS LR250 (HART 1)). We strongly advise you to acquire the detailed version of the manual so you can use your device to its fullest potential. The complete manual can be downloaded from the SITRANS LR250 product page of our web site at: www.siemens.com/LR250. The printed manual is available from your local Siemens Milltronics representative. Questions about the contents of this manual can be directed to: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 Email: techpubs.smpi@siemens.com

English

Copyright Siemens AG 2013. All Rights Reserved


We encourage users to purchase authorized bound manuals, or to view electronic versions as designed and authored by Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments will not be responsible for the contents of partial or whole reproductions of either bound or electronic versions.

Disclaimer of Liability
While we have verified the contents of this manual for agreement with the instrumentation described, variations remain possible. Thus we cannot guarantee full agreement. The contents of this manual are regularly reviewed and corrections are included in subsequent editions. We welcome all suggestions for improvement. Technical data subject to change.

MILLTRONICS is a registered trademark of Siemens Milltronics Process Instruments.

Technical Support
Support is available 24 hours a day. To find your local Siemens Automation Office address, phone number, and fax number, go to: www.siemens.com/automation/partner: Click on the tab Contact, select Service, then click Service again to find your product group (+Automation Technology > +Sensor Systems >+Process Instrumentation > +Level Measurement > +Continuous). Select Radar. Select the country followed by the City/Region. Select Technical Support under Service. For on-line technical support go to: www.siemens.com/automation/support-request Enter the device name (SITRANS LR250) or order number, then click on Search, and select the appropriate product type. Click on Next. Enter a keyword describing your issue. Then either browse the relevant documentation, or click on Next to email a description of your issue to Siemens Technical Support staff. Siemens IA/DT Technical Support Center: phone +49 (0)911 895 7222

1)

HARTis a registered trademark of HART Communication Foundation.

A5E31997170

SITRANS LR250 (HART) QUICK START MANUAL

Page EN-1

Safety Guidelines
English

Warning notices must be observed to ensure personal safety as well as that of others, and to protect the product and the connected equipment. These warning notices are accompanied by a clarification of the level of caution to be observed. WARNING symbol relates to a caution symbol on the product, and means that failure to observe the necessary precautions can result in death, serious injury, and/or considerable material damage. WARNING symbol, used when there is no corresponding caution symbol on the product, means that failure to observe the necessary precautions can result in death, serious injury, and/or considerable material damage. Note: means important information about the product or that part of the operating manual.

FCC Conformity
US Installations only: Federal Communications Commission (FCC) rules
WARNING: Changes or modifications not expressly approved by Siemens Milltronics could void the users authority to operate the equipment. Notes: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference to radio communications, in which case the user will be required to correct the interference at his own expense.

SITRANS LR250
WARNING: SITRANS LR250 is to be used only in the manner outlined in this manual, otherwise protection provided by the equipment may be impaired. Note: This product is intended for use in industrial areas. Operation of this equipment in a residential area may cause interference to several frequency based communications. SITRANS LR250 is a 2-wire 25 GHz pulse radar level transmitter for continuous monitoring of liquids and slurries in storage vessels including high pressure and high temperature, to a range of 20 m (66ft). It is ideal for small vessels and low dielectric media. The device consists of an electronic circuit coupled to a horn antenna and either a threaded or flange type process connection. A threaded PVDF antenna is also available. SITRANS LR250 supports HART communication protocol, and SIMATIC PDM software. Signals are processed using Process Intelligence.

Page EN-2

SITRANS LR250 (HART) QUICK START MANUAL

A5E31997170

Specifications

For a complete listing, see the SITRANS LR250 (HART) Instruction Manual. For Approvals information see Approvals on page 4.

Ambient/Operating Temperature
Notes:
The maximum temperature is dependent on the process connection, antenna materials, and vessel pressure: see Maximum Process Temperature Chart on page 23. For more detailed information see Process Pressure/Temperature derating curves in the full manual. Process temperature and pressure capabilities are dependent upon information on the process connection tag. The reference drawing listed on the ta is available on the product page of our website at www.siemens.com/LR250, under More Info/Installation drawings/ Level Measurement/Installation Drawings/LR250. Additional information on process connections is available on the Installation Drawings page under Process Connection Diagrams. Signal amplitude increases with horn diameter, so use the largest practical size. Optional extensions can be installed below the threads. See Maximum Process Temperature Chart on page 23, for more details.
ambient temperature (surrounding enclosure) 40 C to 80 C (40 F to 176 F) device nameplate

English

process temperature at process connection: standard horn antenna: - with FKM O-ring: 40 oC to +200 oC (40 oF to +392 oF) - with FFKM O-ring: 20 oC to +200 oC (4 oF to +392 oF) 2" NPT / BSPT / G Threaded PVDF antenna: 40 to +80 C (40 to +176 F)

process connection tag

Power
General Purpose Intrinsically Safe Non-Sparking/Energy Limited Non-incendive (FM/CSA US/Canada only) Flameproof Increased Safety Explosion-proof (FM/CSA US/Canada only) Maximum 30 V DC 4 to 20 mA Max. startup current: Nominal 24 V DC at 550 Ohm

Nominal 24 V DC at 250 Ohm

see Startup Behavior on page 24

A5E31997170

SITRANS LR250 (HART) QUICK START MANUAL

Page EN-3

Approvals
Note: The device nameplate lists the approvals that apply to your device. General CSAUS/C, FM, CE, C-TICK, KC Radio Europe (R&TTE), FCC, Industry Canada Hazardous Intrinsically Safe 1) (Europe) ATEX II 1G, EEx ia IIC T4 ATEX II 1D, EEx tD A20 IP67 T90 C (International) IECEx SIR 05.0031X, Ex ia IIC T4, Ex tD A20 IP67 T90 C (US/Canada) FM/CSA Class I, Div. 1, Groups A, B, C, D Class II, Div. 1, Groups E, F, G Class III T4 (Brazil) INMETRO: DNV 12.0087 X Ex ia IIC T4 Ga Ex ia ta IIIC T90 C Da IP65/IP67 -40 C Ta +80 C DNV #OCP 0017 ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005 (China) NEPSI Ex ia IIC T4/ DIP A20 TA, T90 C IP67 Non-sparking/Energy Limited 2) (Europe) ATEX II 3 G Ex nA/nL IIC T4 Gc (China) NEPSI Ex nA II T4/ Ex nL IIC T4

English

Non-incendive 3) (US/Canada) Flameproof 4) (Europe/International)

FM/CSA Class I, Div. 2, Groups A, B, C, D T5 ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex dmbia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C INMETRO: DNV 12.0088 X Ex d ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017

(Brazil)

ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-1:2009, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e

IEC 61241-11:2005

1) 2) 3) 4)

See Intrinsically Safe wiring on page 9. See Non-Sparking/Energy Limited wiring on page 11. See Non-Incendive wiring on page 11. See Flameproof wiring on page 12.

Page EN-4

SITRANS LR250 (HART) QUICK START MANUAL

A5E31997170

Hazardous (continued) Increased Safety 1) Europe/International)

(Brazil)

ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex embia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C INMETRO: DNV 12.0088 X Ex e ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-7:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e

English

IEC 61241-11:2005 Flameproof/Increased Safety (China) NEPSI Ex dmbia IIC T4/ Ex embia IIC T4/ DIP A20 TA, T90 oC IP67 Explosion proof 2) (US/Canada) FM/CSA Class I, Div. 1, Groups A, B, C, D Class II, Div. 1, Groups E, F, G Class III T4 Marine Lloyds Register of Shipping ABS Type Approval

Pressure Application
WARNINGS:

Never attempt to loosen, remove, or disassemble process connection or instrument housing while vessel contents are under pressure. The user is responsible for the selection of bolting and gasket materials which will fall within the limits of the flange and its intended use and which are suitable for the service conditions. Improper installation may result in loss of process pressure.
Notes: The process connection tag shall remain with the process pressure boundary assembly1). In the event the instrument package is replaced, the process connection tag shall be transferred to the replacement unit. SITRANS LR250 units are hydrostatically tested, meeting or exceeding the requirements of the ASME Boiler and Pressure Vessel Code and the European Pressure Equipment Directive.

1) 2)

See Increased safety wiring on page 12. See Explosion-proof wiring (US/Canada only) on page 13.

A5E31997170

SITRANS LR250 (HART) QUICK START MANUAL

Page EN-5

Pressure Equipment Directive, PED, 97/23/EC


Siemens Level Transmitters with flanged, threaded, or sanitary clamp type process mounts have no pressure-bearing housing of their own and, therefore, do not come under the Pressure Equipment Directive as pressure or safety accessories, (see EU Commission Guideline 1/8). 1)

English

Installation
WARNINGS: Installation shall only be performed by qualified personnel and in accordance with local governing regulations. Handle the device using the enclosure not the process connection tag, to avoid damage. Take special care when handling the threaded PVDF antenna. Any damage to the antenna surface, particularly to the tip, could affect performance. Materials of construction are chosen based on their chemical compatibility (or inertness) for general purposes. For exposure to specific environments, check with chemical compatibility charts before installing. Notes:
For European Union and member countries, installation must be according to ETSI EN 302372. Refer to the device nameplate for approval information. The serial numbers stamped in each process connection body provide a unique identification number indicating date of manufacture. Example: MMDDYY XXX (where MM = month, DD = day, YY = year, and XXX= sequential unit produced)

Further markings (space permitting) indicate flange configuration, size, pressure class,
material, and material heat code.

Nozzle design
Threaded PVDF Antenna
min. clearance: 10 mm (0.4")

Stainless Steel Horn Antenna


min. clearance: 10 mm (0.4")

Min. dia. 51 mm (2")

The end of the antenna must protrude a minimum of 10 mm (0.4) to avoid false echoes being reflected from the nozzle. Minimum recommended nozzle diameter for the threaded PVDF antenna is 51 mm (2"). An antenna extension (100 mm/ 3.93") is available for any version except the threaded PVDF antenna.

1)

The process pressure boundary assembly comprises the components that act as a barrier against pressure loss from the process vessel: that is, the combination of process connection body and emitter, but normally excluding the electrical enclosure.

Page EN-6

SITRANS LR250 (HART) QUICK START MANUAL

A5E31997170

Nozzle location
Avoid central locations on tall, narrow vessels, which can generate false echoes.

Environment
Provide an environment suitable to the housing rating and materials of construction. Provide a sunshield if the device will be mounted in direct sunlight. 1)
preferred undesirable

English

Beam angle
Beam angle is the width of the cone where the energy density is half of the peak energy density. The peak energy density is directly in front of and in line with the antenna. There is a signal transmitted outside the beam angle, therefore false targets may be detected.

Emission Cone
Keep emission cone free of interference from ladders, pipes, I-beams or filling streams.

emission cone beam angle: threaded PVDF antenna = 19 1.5" horn = 19 2" horn = 15 = 10 beam 3" horn = 8 angle 4" horn

Access for programming


Provide easy access for viewing the display and programming via the handheld programmer.

Mounting instructions
WARNING: For pressure applications, it will be necessary to use PTFE tape or other appropriate thread sealing compound, and to tighten the process connection beyond hand-tight. (The maximum recommended torque is 40 N-m (30 ft.lbs.) Note: There is no limit to the number of times a device can be rotated without damage.

Threaded Versions
1) 2) Before inserting the device into its mounting connection, check to ensure the threads are matching to avoid damaging them. Simply screw the device into the process connection, and hand tighten, or use a wrench. For pressure applications see Warning note above.

1)

The process pressure boundary assembly comprises the components that act as a barrier against pressure loss from the process vessel: that is, the combination of process connection body and emitter, but normally excluding the electrical enclosure.

A5E31997170

SITRANS LR250 (HART) QUICK START MANUAL

Page EN-7

Flanged Versions
WARNING: The user is responsible for the selection of bolting and gasket materials which will fall within the limits of the flange and its intended use, and which are suitable for the service conditions.

English

Wiring
Power
WARNINGS: The DC input terminals shall be supplied from a source providing electrical isolation between the input and output, in order to meet the applicable safety requirements of IEC 61010-1. All field wiring must have insulation suitable for rated voltages.

Connecting SITRANS LR250


WARNINGS: Check the device nameplate to verify the approval rating. Use appropriate conduit seals to maintain IP or NEMA rating. See Wiring setups for hazardous area installations on page 9. Notes: Use twisted pair cable: AWG 22 to 14 (0.34 mm2 to 2.5 mm2). Separate cables and conduits may be required to conform to standard instrumentation wiring practices or electrical codes.
Use a 2 mm Allen key to loosen the lid-lock set screw. plug (IP 68) optional cable gland1) 2) (or NPT cable entry2)) cable shield (if used)

1) 2) 3) 4) 5)

Strip the cable jacket for approximately 70 mm (2.75") from the end of the cable, and thread the wires through the gland. 1) 2) Connect the wires to the terminal as shown: the polarity is identified on the terminal block. Ground the device according to local regulations. Tighten the gland to form a good seal. Close the lid before programming and device configuration.

1) 2)

May be shipped with the device. If cable is routed through conduit, use only approved suitable-size hubs for waterproof applications.

Page EN-8

SITRANS LR250 (HART) QUICK START MANUAL

A5E31997170

Connecting HART
Typical PLC/mA configuration with HART 1) 2)
active PLC power supply R= 250 HART modem SITRANS LR250 HART communicator

Notes: Depending on the system design, the power supply may be separate from the PLC, or integral to it HART resistance1) must be limited according to the allowable operating area2). Do not connect more than one HART programming device at a time.

English

Wiring setups for hazardous area installations


There are six wiring options for hazardous area installations. In all cases, check the device nameplate and process connection tag to verify the approval rating.

1. Intrinsically Safe wiring


Device nameplate (ATEX/IECEX/INMETRO/C-TICK)

The ATEX certificate listed on the nameplate can be downloaded from the product page of our website at: www.siemens.com/LR250. Go to Support > Approvals / Certificates. The IECEx certificate listed on the nameplate can be viewed on the IECEx website. Go to: http://iecex.iec.ch and click on Ex Equipment Certificates of Conformity then enter the certificate number IECEx SIR 05.0031X.

1) 2)

Total loop resistance = cable resistance plus 250 Ohm (resistor). See either Curve 1: General Purpose, Intrinsically Safe, Non-Sparking/Energy Limited/ Non-incendive on page 23 or Curve 2: Flameproof, Increased Safety, Explosion-proof on page 24.

A5E31997170

SITRANS LR250 (HART) QUICK START MANUAL

Page EN-9

Intrinsically Safe wiring (continued)


Device nameplate (FM/CSA) English
The FM/CSA Intrinsically Safe connection drawing number 23650653 can be downloaded from the product page of our website at: www.siemens.com/LR250. Go to Support > Installation Drawings > Level Measurement > Continuous - Radar. For power demands see Curve 1: General Purpose, Intrinsically Safe, Non-Sparking/ Energy Limited/ Non-incendive on page 23. For wiring requirements: follow local regulations. Approved dust-tight and water-tight conduit seals are required for outdoor NEMA 4X / type 4X / NEMA 6, IP67, IP68 locations. Refer to Instructions specific to hazardous area installations on page 13. Recommended intrinsically safe barriers are listed under Passive Shunt Diode Barriers on page 10 and Active barriers (repeating barriers) on page 11. Note: Selecting a suitable PLC input module, power supply, or barrier requires knowledge about Intrinsic Safety and the application. It is the responsibility of the installer to ensure that the intrinsically safe installation complies with both the apparatus approval requirements and the relevant national code of practice.

Passive Shunt Diode Barriers


Note: A well regulated supply voltage is required. Manufacture
MTL MTL Stahl Stahl

Part Number
787SP+ (Dual Channel) 7787P+ (Dual Channel) 9001/01-280-100-10 (Single Channel) 9002/01-280-110-10 (Dual Channel)

How to select a passive barrier for SITRANS LR250

To make sure that the barrier safety description is suitable for the LR250 Intrinsically Safe (IS) input parameters, carry out the following calculations: Re-e = max. end-to-end resistance of the barrier Rloop = loop resistance (total of cable resistance plus any additional series resistance, for example, PLC inputs and/or displays) Vbarrier = value of any non-linear voltage drops due to the barrier 1) Determine the value for Re-e from the data sheet. 2) Calculate the total value for Rloop. 3) Calculate Rworking = Re-e + Rloop.

Page EN-10

SITRANS LR250 (HART) QUICK START MANUAL

A5E31997170

How to select a passive barrier (continued) 4) Determine the value of Vbarrier from the barrier data sheet (for example, voltage drops due to diodes). 5) Calculate Vworking = Vsupply Vbarrier. 6) Use the values for Vworking and Rworking to confirm that operation is within the shaded area of the graph Curve 1: General Purpose, Intrinsically Safe, Non-Sparking/Energy Limited/ Non-incendive on page 23.

English

Active barriers (repeating barriers)


Manufacturer
MTL MTL Siemens Stahl MTL MTL

Part Number
706 7206 SITRANS I 7NG4124 9001/51-280-110-14 E02009 E02010

2. Non-Sparking/Energy Limited wiring

The ATEX certificate listed on the nameplate can be downloaded from the product page of our website at: www.siemens.com/LR250. Go to: Support > Approvals / Certificates. For power demands see Curve 1: General Purpose, Intrinsically Safe, Non-Sparking/ Energy Limited/ Non-incendive on page 23. For wiring requirements follow local regulations.

3. Non-Incendive wiring

FM/CSA Class 1, Div 2 connection drawing number 23650673 can be downloaded from the product page of our website at: www.siemens.com/LR250. Go to Support > Installation Drawings > Level Measurement > Continuous - Radar. For power demands see Curve 1: General Purpose, Intrinsically Safe, Non-Sparking/Energy Limited/ Non-incendive on page 23.

A5E31997170

SITRANS LR250 (HART) QUICK START MANUAL

Page EN-11

4. Flameproof wiring
English

The ATEX certificate listed on the nameplate can be downloaded from the product page of our website at: www.siemens.com/LR250. Go to: Support > Approvals / Certificates. The IECEx certificate listed on the nameplate can be viewed on the IECEx website. Go to: http://iecex.iec.ch and click on Ex Equipment Certificates of Conformity then enter the certificate number IECEx SIR 08.0107X. For power demands see Curve 2: Flameproof, Increased Safety, Explosion-proof on page 24. For wiring requirements follow local regulations. See also Instructions specific to hazardous area installations on page 13 and the ATEX certificate listed above.

5. Increased safety wiring

The ATEX certificate listed on the nameplate can be downloaded from the product page of our website at: www.siemens.com/LR250. Go to: Support > Approvals / Certificates. The IECEx certificate listed on the nameplate can be viewed on the IECEx website. Go to: http://iecex.iec.ch and click on Ex Equipment Certificates of Conformity then enter the certificate number IECEx SIR 08.0107X. For power demands see Curve 2: Flameproof, Increased Safety, Explosion-proof on page 24. For wiring requirements follow local regulations. See also Instructions specific to hazardous area installations on page 13 and the ATEX certificate listed above.

Page EN-12

SITRANS LR250 (HART) QUICK START MANUAL

A5E31997170

6. Explosion-proof wiring (US/Canada only)


English

FM/CSA Explosion Proof connection drawing number A5E02257843 can be downloaded from the product page of our website at: www.siemens.com/LR250. Go to Support > Installation Drawings > Level Measurement > Continuous - Radar. For power demands see Curve 2: Flameproof, Increased Safety, Explosion-proof on page 24.

Instructions specific to hazardous area installations


(Reference European ATEX Directive 94/9/EC, Annex II, 1/0/6)
The following instructions apply to equipment covered by certificate number SIRA 06ATEX2358X, SIRA 08ATEX1301X, and SIRA 09ATEX4153X. For use and assembly, refer to the main instructions. The equipment is certified for use as Category 1GD equipment per SIRA 06ATEX2358X; Category 1/2 GD, 1D, 2D equipment per SIRA 08ATEX1301X; and Category 3G equipment per SIRA 09ATEX4153X. 3) The equipment may be used with flammable gases and vapors with apparatus group IIC, IIB, and IIA, and temperature classes T1, T2, T3 and T4. 4) The equipment has a degree of ingress protection of IP67 and a temperature class of T90 C and may be used with flammable dusts. 5) The equipment is certified for use in an ambient temperature range of 40 C to +80 C. 6) The equipment has not been assessed as a safety related device (as referred to by Directive 94/9/EC Annex II, clause 1.5): see on page 5. 7) Installation and inspection of this equipment shall be carried out by suitably trained personnel in accordance with the applicable code of practice (EN 60079-14 and EN 6007917 in Europe). 8) The equipment is non-repairable. 9) The certificate numbers have an X suffix, which indicates that special conditions for safe use apply. Those installing or inspecting this equipment must have access to the certificates. 10) If the equipment is likely to come into contact with aggressive substances, then it is the responsibility of the user to take suitable precautions that prevent it from being adversely affected, thus ensuring that the type of protection is not compromised. Aggressive substances: for example, acidic liquids or gases that may attack metals, or solvents that may affect polymeric materials. Suitable precautions: for example, establishing from the materials data sheet that it is resistant to specific chemicals. 1) 2)

A5E31997170

SITRANS LR250 (HART) QUICK START MANUAL

Page EN-13

Programming SITRANS LR250


English

See Quick Start Wizard via the handheld programmer on page 18. See Quick Start Wizard via SIMATIC PDM on page 21.

Activating SITRANS LR250


Power up the device. SITRANS LR250 automatically starts up in Measurement mode.

The LCD Display Measurement mode display ( normal operation) 1) 2)


1 2 3 4
(((

1 toggle indicator1) for linear units or % 2 selected operation: level, space, distance, or volume 3 measured value (level or volume, space, or distance) 4 units 5 bar graph indicates level 6 secondary region indicates on request2) electronics temperature, echo confidence, loop current, or distance 7 text area displays status messages 8 device status indicator (for details please see the full manual)

Fault present indicators S: 0 LOE


When a fault is present the fault code and an error message are displayed in the text area (7), and a service-required icon appears in the device status location (8)

Program mode display


Navigation view current current A visible menu bar indicates the item menu number menu list is too long to display all menu items. current bar item A band halfway down the menu item band bar indicates the current item is halfway down the list. The depth and relative position of the item band on the menu bar indicates the length of the menu list, and approximate position of the current item in the list. A deeper band indicates fewer items. Parameter view
parameter number parameter value/ selection

Edit view

parameter name

1) 2)

Press UP or DOWN arrow to switch. In response to a key press request. For details, see table Function in Measurement mode on page 15.

Page EN-14

SITRANS LR250 (HART) QUICK START MANUAL

A5E31997170

Handheld Programmer (Part No. 7ML1930-1BK)


Ordered separately. 1) 2) 3)

Key

Function in Measurement mode


Updates the loop current1). Updates internal enclosure temperature reading1). Updates echo confidence value1).

English

Updates distance measurement1). Mode opens PROGRAM mode2). RIGHT arrow opens PROGRAM mode3).

UP or DOWN arrow toggles between linear units and %.

Programming via the handheld programmer


Notes: Press Mode to toggle between Measurement and Program Mode. While the device is in PROGRAM mode the output remains fixed and does not respond to changes in the device. SITRANS LR250 automatically returns to Measurement mode after a period of inactivity in PROGRAM mode (between 15 seconds and 10 minutes, depending on the menu level).

Parameter menus

Parameters are identified by name and organized into 1. QUICK START 2. SETUP function groups. 2.1. DEVICE For the complete list of parameters with instructions, see ............. the full manual. 2.7. LINEARIZATION

2.7.1. VOLUME 2.7.1.1. VESSEL SHAPE

1) 2)

3)

New value is displayed in LCD secondary region. Opens the menu level last displayed in this power cycle, unless power has been cycled since exiting PROGRAM mode or more than 10 minutes have elapsed since PROGRAM mode was used. Then top level menu will be displayed. Opens the top level menu

A5E31997170

SITRANS LR250 (HART) QUICK START MANUAL

Page EN-15

1.

Enter PROGRAM mode


Point the programmer at the display (from a maximum distance of 300 mm [1 ft.]).

handheld programmer display

English

RIGHT arrow activates PROGRAM mode and opens menu level 1. Mode opens the menu level last displayed in PROGRAM mode within the last 10 minutes, or menu level 1 if power has been cycled since then.

Max. 300 mm (1 ft)

2.

Navigating: key functions in Navigation mode


, then

Note: For Quick Access to parameters via the handheld programmer, press Home enter the menu number, for example: 2.7.1 (Volume).

Key

Name
UP or DOWN arrow

Menu level Function in Navigation Mode


menu or parameter menu parameter Scroll to previous or next menu or parameter. Go to first parameter in the selected menu, or open next menu. Open Edit mode. Open parent menu. Change to MEASUREMENT mode. Open top level menu: menu 1.

RIGHT arrow LEFT arrow


Mode Home

menu or parameter menu or parameter menu or parameter

3.

Editing in PROGRAM mode


Selecting a listed option a) Navigate to the desired parameter. b) Press RIGHT arrow view. to open parameter
current selection parameter name parameter number

c)

Press RIGHT arrow

again to open Edit

mode. The current selection is highlighted. Scroll to a new selection. d) Press RIGHT arrow to accept it

The LCD returns to parameter view and displays the new selection.

Page EN-16

SITRANS LR250 (HART) QUICK START MANUAL

A5E31997170

Changing a numeric value a) Navigate to the desired parameter. b) c) d) Press RIGHT arrow again to open Edit mode. The current value is highlighted. Key in a new value. Press RIGHT arrow to accept it. The LCD returns to parameter view and displays the new selection.

parameter name

parameter number

English

current value

Key functions in Edit mode Key Name


UP or DOWN arrow

Function in Edit Mode


Selecting options Numeric editing Selecting options Scrolls to item. - Increments or decrements digits. - Toggles plus and minus sign. - Accepts the data (writes the parameter). - Changes from Edit to Navigation mode. - Moves cursor one space to the right - or with cursor on Enter sign, accepts the data and changes from Edit to Navigation mode. Cancels Edit mode without changing the parameter. - Moves cursor to plus/minus sign if this is the first key pressed - or moves cursor one space to the left. Erases the display. Enters a decimal point.

RIGHT arrow

Numeric editing Selecting options

LEFT arrow

Numeric editing Numeric editing Numeric editing

Clear Decimal point Plus or minus sign


to

Numeric editing

Changes the sign of the entered value.

Numeral

Numeric editing

Enters the corresponding character.

A5E31997170

SITRANS LR250 (HART) QUICK START MANUAL

Page EN-17

Quick Start Wizard via the handheld programmer


1. Quick Start
a) Point the programmer at the display from a maximum distance of 300 mm (1 ft), then press RIGHT arrow b) Press RIGHT arrow view. c) Press RIGHT arrow to open Edit mode or DOWN arrow to accept default to activate PROGRAM mode and open menu level 1. twice to navigate to menu item 1.1 and open parameter

English

values and move directly to the next item. d) e) To change a setting, scroll to the desired item or key in a new value. After modifying a value, press RIGHT arrow arrow f) to move to the next item. to accept it and press DOWN

Quick Start settings take effect only after you select Yes to Apply changes in step 1.8.

1.1. Language Selects the language to be used on the LCD and takes effect immediately.
Options
ENGLISH, DEUTSCH, FRANAIS, ESPAOL

1.2. Material Options


LIQUID LIQUID LOW DK (low dielectric liquid)

1.3. Response Rate Sets the reaction speed of the device to measurement changes in the target range.
Response Rate
SLOW MED FAST

Fill Rate/Empty Rate


0.1 m/min (0.32 ft/min) 1.0 m/min (3.28 ft/min) 10.0 m/min (32.8 ft/min)

Use a setting just faster than the maximum filling or emptying rate (whichever is greater).

1.4. Units Sensor measurement units. Options


M, CM, MM, FT, IN

Page EN-18

SITRANS LR250 (HART) QUICK START MANUAL

A5E31997170

1.5. Operating Mode


sensor reference point space sensor reference point

distance High Cal. Point

English

sensor reference point

level Low Cal. Point

Operation
NO SERVICE LEVEL SPACE DISTANCE
1)

Description
Measurement and associated loop current are not updated, and the device defaults to Fail-safe mode1). Distance to material surface referenced from Low Calibration Point. Distance to material surface referenced from High Calibration Point. Distance to material surface referenced from Sensor Reference Point.

For more details on Fail-safe Mode see the full manual.

1.6. Low Calibration Point Distance from Sensor Reference to Low Calibration Point: usually process empty level. See Operating Mode (1.5.) for an illustration. Values
Range: 0.00 to 20.00 m

1.7. High Calibration Point Distance from Sensor Reference to High Calibration Point: usually process full level. See Operating Mode (1.5.) for an illustration. Values
Range: 0.00 to 20.00 m

1.8. Apply? (Apply changes) In order to save the Quick Start settings it is necessary to select Yes to apply changes. Options
Press Mode YES, NO, DONE (Display shows DONE when Quick Start is successfully completed.) to return to Measurement mode. SITRANS LR250 is now ready to operate.

Note: If your application has a tank with obstructions, please see the full manual for details on using Auto False Echo Suppression.

A5E31997170

SITRANS LR250 (HART) QUICK START MANUAL

Page EN-19

SITRANS LR250 Communications: HART


English

You will need the full manual to acquire the list of applicable parameters. We recommend that you use SIMATIC Process Device Manager (PDM) to program your device. Application Guides for setting up HART devices with SIMATIC PDM can be downloaded from the product page of our website at: www.siemens.com/LR250 under Support.

SIMATIC PDM
SIMATIC PDM is a software package used to commission and maintain SITRANS LR250 and other process devices. Please consult the operating instructions or online help for details on using SIMATIC PDM. (You can find more information at www.siemens.com/simatic-pdm.) This aligns SIMATIC PDM with the HART modem for Windows 2000 and Windows XP 1) operating Systems. Notes: SIMATIC PDM operates only in the Windows XP Professional version, not in the Home version. You need administrative rights on your operating system to deactivate buffers. 1) 2) 3) 4) 5) 6) Click Start/Settings/Control Panel to begin configuration. Double click System, select the Hardware tab, and click the Device Manager button. Open Ports folder and double click the COM Port used by the system to open the Communications Port Properties window. Select the Port Settings tab and double click the Advanced button. If the Use FIFO buffers radio box is selected, click to deselect. Click OK to close out. Close all screens and then reboot.

Deactivate Buffers

Update Electronic Device Description (EDD)


Notes:
SIMATIC PDM Rev. 5.2 SP1 is supported only for basic configuration and troubleshooting. For advanced features such as the Quick Start wizard, Rev. 6.0 SP4 or higher is required. In SIMATIC PDM the EDD is located in Device Catalog, under Sensors/Level/Echo/ Siemens Milltronics/SITRANS LR250. Check the product page of our website at: www.siemens.com/LR250 to make sure you have the latest version of SIMATIC PDM, the most recent Service Pack (SP) and the most recent hot fix (HF). Go to Support > Software Downloads

To install a new EDD:


1) 2) 3) Download the EDD from the product page of our website at: www.siemens.com/LR250 and save the files to your computer. Extract the zipped file to an easily accessed location. Launch SIMATIC PDM Manage Device Catalog, browse to the unzipped EDD file and select it.

1)

Windows is a registered trademark of the Microsoft Corporation

Page EN-20

SITRANS LR250 (HART) QUICK START MANUAL

A5E31997170

Configuring a new device


Note: Clicking on Cancel during an upload from device to SIMATIC PDM will result in some
parameters being updated.

English

1) 2)

Check that you have the most recent EDD, and if necessary update it (seeUpdate Electronic Device Description (EDD) on page 20). Launch SIMATIC Manager and create a new project for LR250. Application Guides for setting up HART devices with SIMATIC PDM can be downloaded from the product page of our website at: www.siemens.com/LR250. Open the menu Device Master Reset and click on OK to perform a reset to Factory Defaults. After the reset is complete upload parameters to the PC/PG. Configure the device via the Quick Start wizard.

3) 4) 5)

Quick Start Wizard via SIMATIC PDM


Notes:
The Quick Start wizard settings are inter-related and changes apply only after you click on FINISH AND DOWNLOAD at the end of step 5 to save settings offline and transfer them to the device. Click on BACK to return and revise a setting or Cancel to exit the Quick Start. Launch SIMATIC PDM, open the menu Device Wizard - Quick Start, and follow steps 1 to 5.

Quick Start

Operating via FDT (Field Device Tool)


FDT is a standard used in several software packages designed to commission and maintain field devices. Two commercially available FDTs are PACTware and Fieldcare. To configure a field device via FDT you need the DTM (Device Type Manager) for the device. Siemens instruments use SITRANS DTM and an instrument EDD written for SITRANS DTM. 1) First install SITRANS DTM on your system. You can download it from: http://support.automation.siemens.com. Click on Product Support, and navigate to Product Information/Automation Technology/Sensor systems/Process Instrumentation/Software & Communications. 2) Install the SITRANS LR250 HART EDD for SITRANS DTM. You can download it from the product page of our website at: www.siemens.com/LR250. Go to Support > Software Downloads.

Configuring a new device via FDT


An Application Guide can be downloaded from the product page of our website under Support.

A5E31997170

SITRANS LR250 (HART) QUICK START MANUAL

Page EN-21

Operating via AMS Device Manager


English
AMS Device Manager is a software package designed to commission and maintain field devices. Please consult the operating instructions or online help for details on using AMS Device Manager. You can find more information at: http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD)


SITRANS LR250 requires the EDD for AMS Device Manager version 9.0.

Configuring a new device via AMS Device Manager


1) Check the product page of our website at: www.siemens.com/LR250 to make sure you have the most recent EDD. Go to Support > Software Downloads and if necessary download it. Save the files to your computer, and extract the zipped file to an easily accessed location. Launch AMS Device Manager Add Device Type, browse to the unzipped EDD file and select it.

2)

Launch AMS Device Manager. An Application Guide for setting up HART devices with AMS Device Manager can be downloaded from the product page of our website under Support

Maintenance
SITRANS LR250 requires no maintenance or cleaning under normal operating conditions. If cleaning becomes necessary under severe operating conditions: 1) 2) Note the antenna material and the process medium, and select a cleaning solution that will not react adversely with either. Remove the device from service and wipe the antenna clean using a cloth and suitable cleaning solution.

Unit Repair and Excluded Liability


For detailed information, please see the inside back cover.

Antenna or electronics/enclosure replacement


If the antenna requires replacement due to damage or failure, it may be replaced without the need for re-calibration. Changing to a different antenna type may be performed by a Siemens authorized repair center or personnel. If the electronics or enclosure require replacement due to damage or failure, plesase ensure the correct antenna version is used, otherwise a re-calibration will need to be performed by Siemens authorized personnel.

Page EN-22

SITRANS LR250 (HART) QUICK START MANUAL

A5E31997170

Maximum Process Temperature Chart


Flange Adapter versions of LR250
English
WARNING: Internal temperature must not exceed 80 C (176 F). Notes: The chart is for guidance only and does not represent every possible process connection arrangement. (It will NOT apply if SITRANS LR250 is mounted directly on a metallic vessel surface.) The chart does not take into consideration heating from direct sunshine exposure. Parameter 3.2.1 Current Internal Temperature monitors the internal temperature. Maximum Process Temperatures versus allowable ambient

Ambient Temperature (oC)

process temperature

Process Temperature (oC)

Loop power
Power Supply Requirements
Note: The curves below apply to a standalone device, configured via the Siemens handheld programmer. Curve 1: General Purpose, Intrinsically Safe, Non-Sparking/ Energy Limited/ Nonincendive

Loop Resistance RL

ALLOWABLE OPERATING AREA

Loop Voltage VL

Note: When using HART communications, the minimum voltage with 220 Ohms (RL) is 16.3 VDC.

A5E31997170

SITRANS LR250 (HART) QUICK START MANUAL

Page EN-23

Loop Resistance RL

Curve 2: Flameproof, Increased Safety, Explosion-proof

English

ALLOWABLE OPERATING AREA

Loop Voltage VL

Note: When using HART communications, the minimum voltage with 220 Ohms (RL) is 20.94 VDC.

Startup Behavior
The device draws less than 3.6 mA at startup. Time to first measurement is less than 50 seconds

Page EN-24

SITRANS LR250 (HART) QUICK START MANUAL

A5E31997170

SITRANS LR250 (HART) Manul rychlho uveden do provozu


Tento manul obsahuje zkladn vlastnosti a funkce zazen SITRANS LR250 (HART 1)). Velmi doporuujeme pouvat detailn verzi manulu, abyste mohli vyut vechny monosti zazen. Kompletn manul zazen SITRANS LR250 lze sthnout z naich webovch strnek: www.siemens.com/LR250. Titnou verzi manulu obdrte od svho obchodnho zstupce Siemens Milltronics. Dotazy k obsahu tohoto manulu zaslejte na adresu: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P.O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 Email: techpubs.smpi@siemens.com

esky

Copyright Siemens AG 2013. Vechna prva vyhrazena


Doporuujeme uivatelm nkup autorizovanho titnho manulu nebo staen elektronick verze navren a autorizovan firmou Siemens Milltronics Process Instruments. Firma Siemens Milltronics Process Instruments nen zodpovdn za obsah sten nebo kompletn kopie jak titn nebo elektronick verze manulu.

Prohlen o zodpovdnosti
Akoliv jsme ovili shodu obsahu manulu s popisovanm zazenm, me dochzet ke zmnm. Proto nememe zaruit plnou shodu. Obsah tohoto manulu je pravideln revidovn a opravovn v kadm nsledujcm vydn. Vtme vechny nvrhy k jeho zlepen. Technick daje podlhaj zmn.

MILLTRONICS je registrovan ochrann znmka firmy Siemens Milltronics Process Instruments.

Technick podpora

Sluba technick podpory je dostupn 24 hodin denn. Adresu, telefonn slo a slo faxu nejbli poboky Siemens Automation Office najdete na strnce: www.siemens.com/automation/partner: Kliknte na Contact (Kontakty), vyberte Service (Sluby), opakovanm kliknutm na Service budete moci vybrat skupinu do kter pat vae zazen (+Automation Technology (Automatizan technologie) > +Sensor Systems (Detekn systmy) >+Process Instrumentation (Nstroje pro vrobu) > +Level Measurement (Men hladiny) > +Continuous (Prbn)). Vyberte sloku Radar. Vyberte postupn zem, region a msto. Vyberte Technical Support (Technick podpora) pod zlokou Service. On-line technickou podporu mete kontaktovat pes: www.siemens.com/automation/support-request Zadejte nzev produktu (SITRANS LR250) nebo objednac slo a kliknte na Search (Hledat), nsledovn vyberte vhodn typ zazen. Kliknte na Next. Zadejte klov slovo vztahujc se k vaemu dotazu. Nsledovn proberte nalezenou dokumentaci nebo kliknte na Next pro zasln vaeho dotazu tmu Technick podpory Siemens pomoc elektronick poty. Centrum Technick Podpory Siemens A&D: telefon +49 (0)911 895 7222

1)

HART je registrovan znaka spolenosti HART Communication Foundation.


Str. CS-1

A5E31997170 SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

Bezpenostn pokyny
Varovn pokyny mus bt respektovny, aby byla zajitna bezpenost obsluhy a ostatnch osob stejn jako ochrana vrobku a souvisejcho zazen. Tyto varovn pokyny jsou doprovzeny popisem rovn hrozcho nebezpe. POZOR: vztahuje se k bezpenostnm symbolm na vrobku: Zanedbn nezbytnch bezpenostn opaten me mt za nsledek smrt, vn zrann a znan materiln kody. POZOR: Tento symbol je pouvn v ppad, e neexistuje odpovdajc varovn symbol na vrobku. Zanedbn nezbytnch bezpenostnch opaten me mt za nsledek smrt, vn zrann a znan materiln kody. Poznmka: Oznamuje dleitou informaci o vrobku i jeho sti v manulu.

FCC shoda
esky

Pouze pro instalaci v USA: Pedpisy FCC (Federal Communications Commission)


POZOR: Zmny nebo pravy vslovn neproveden firmou Siemens Milltronics by mohly vst k omezen oprvnn uivatele udrovat zazen v innosti. Poznmky: Toto zazen bylo testovno a spluje limity tdy A digitlnho zazen, podle l. 15 pedpis FCC. Tyto limity jsou navreny tak, aby zajistily rozumnou ochranu proti kodlivmu psoben, je-li zazen komern vyuvno. Toto zazen generuje, vyuv a me vyzaovat vysokofrekvenn vlny a pokud nen instalovno podle pokyn manulu, me ruiv psobit na rdiovou komunikaci. innost zazen v obytnch oblastech by mohla ruiv psobit na rdiovou komunikaci. V takovm ppad mus uivatel odstranit toto ruen na vlastn nklady.

SITRANS LR250
POZOR: SITRANS LR250 me bt uvno pouze podle zsad uvedench v tomto manulu, jinak by mohlo dojt k pokozen bezpenostnch funkc zazen. Poznmka: Toto zajen je koncipovan pro pouit v prmyslovch oblastech. Jeho pouvn v obytnch oblastech me mt za nsledek ruen radiokomunikanch signl. SITRANS LR250 je dvoukanlov impulsov vysla v radarovm psmu 25 GHz, slouc k monitorovn hladiny tekutin i sypkch hmot ve skladovacch silech pi vysokm tlaku a teplot, v rozpt 20 m (66ft). Je ideln pro men skladovac sila a mn dielektrick mdia. Toto zazen se skld z elektronickho obvodu napojenho na trubkovitou antnu pomoc zvitovho i prubovho technologickho spoje. Je t k dispozici roubovac PVDF antna. SITRANS LR250 podporuje komunikan protokol HART a software SIMATIC PDM. Signly jsou zpracovny pomoc softwaru Process Intelligence.

Str. CS-2

SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

A5E31997170

Specifikace

Kompletn vpis najdete v manulu SITRANS LR250 (HART). Informace o homologaci naleznete ne Homologace na str. 4.

Teplota prosted/pracovn
Poznmky:
Maximln teplota je zvisl na typu technologickho spoje, na vrobnm materilu antny a na tlaku uvnit ndoby. Viz Diagram maximln pracovn teploty na str. 23. Pro podrobnj informace viz grafy pouitelnch pracovnch teplot/tlak v kompletnm manulu. Pracovn teplota a povolen rozsah pracovnho tlaku zvisej na dajch uvedench na ttku pracovnch daj. Referenn graf uveden na ttku naleznete na internetovch strnkch produktu www.siemens.com/LR250 pod Support (Podpora) > Installation Drawings (Vkresy) > Level Measurement (Men hladiny) > Continuous Radar (Stabiln radar). Dal daje o technologickm spoji naleznete na strnce Installation Drawings (Instalan grafy) pod Process Connection Diagrams (Grafy technologickho spoje). Amplituda signlu narst s pmrem vyzaovac trubky, proto pouijte nejvt vhodnou velikost. Mete pout voliteln nstavce, kter se instaluj pomoc zvit. Viz Diagram maximln pracovn teploty na str. 23 pro podrobnj daje.
teplota prosted (vn zazen) 40 C a 80 C (40 F a 176 F)

esky

vrobn ttek pstroje

pracovn teplota na technologickm spoji: standardn trubkovit antna: - s tsnicm o-kroukem FKM: 40 oC a +200 oC (-40 oF a +392 oF) - s tsnicm o-kroukem FFKM: 20 oC a +200 oC (-4 oF a +392 oF) 2" NPT / BSPT / G roubovac PVDF antna: -40 a +80 C (-40 a +176 F)

ttek pracovnch daj

Napjen
Veobecn pouit Jiskrov bezpen Nominln 24 V = s odporem Nejiskiv/S omezenm pkonem 550 . Nezpaln zazen (FM/CSA pouze pro USA/Kanada) Ochrana proti ohni Zven bezpenost Ochrana proti vbuchu (FM/CSA pouze pro USA/ Kanada) Maximln 30 V = 4 20 mA Max. proud pi sputn: Nominln 24 V = s odporem 250 .

Viz . Vkon elektrickho obvodu na str. 23

A5E31997170 SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

Str. CS-3

Homologace
Poznmka: Na vrobnm ttku jsou uvedeny homologace vztahujc se k danmu zazen.
Veobecn CSAUS/C, FM, CE, C-TICK, KC Rdiov vlny Evropa (R&TTE), FCC, Industry Canada Vbun prosted Jiskrov bezpen 1) (Evropa) ATEX II 1G, EEx ia IIC T4 ATEX II 1D, EEx tD A20 IP67 T90 C (Mezinrodn) IECEx SIR 05.0031X, Ex ia IIC T4, Ex tD A20 IP67 T90 C (USA/Kanada) FM/CSA Tda I, Div. 1, Skupiny A, B, C, D Tda II, Div. 1, Skupiny E, F, G Tda III T4 (Brazlie) INMETRO: DNV 12.0087 X Ex ia IIC T4 Ga Ex ia ta IIIC T90 C Da IP65/IP67 -40 C Ta +80 C DNV #: OCP 0017 ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-11:2009 ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005 (na) NEPSI Ex ia IIC T4/ DIP A20 TA, T90 C IP67 Nejiskiv/S omezenm pkonem 2) (Evropa) ATEX II 3 G Ex nA/nL IIC T4 Gc (na) NEPSI Ex nA II T4/ Ex nL IIC T4 Nezpaln zazen 3) (USA/Kanada) FM/CSA Tda I, Div. 2, Skupiny A, B, C, D T5 Ochrana proti ohni 4) (Evropa/Mezinrodn) ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex dmbia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C (Brazlie) INMETRO: DNV 12.0088 X Ex d ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #: OCP 0017 ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-1:2009, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005 Zven bezpenost 5) (Evropa/Mezinrodn) ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex embia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C (Brazlie) INMETRO: DNV 12.0088 X Ex e ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #: OCP 0017 ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-7:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

esky
1) 2) 3) 4) 5)

Viz Jiskrov bezpen zapojen na str. 9. Viz Zapojen Nejiskiv/S omezenm pkonem na str. 11. Viz Nezpaln zazen na str. 11. Viz Ohnivzdorn zapojen na str. 12. Viz Zapojen se zvenou bezpenost na str. 12.
SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU A5E31997170

Str. CS-4

Vbun prosted (pokraovn) Ochrana proti ohni/Zven bezpenost (na) NEPSI Ex dmbia IIC T4/ Ex embia IIC T4/ DIP A20 TA, T90 C IP67 Ochrana proti vbuchu 1) (USA/Kanada)FM/CSA Tda I, Div. 1, Skupiny A, B, C, D Tda II, Div. 1, Skupiny E, F, G Tda III T4

Pepravn Nmon registr Lloyds Osvden typu ABS

Pouiteln tlak
POZOR: Nikdy se nesnate odpojovat, odmontovat nebo rozebrat technologick spoj i plt zazen pokud je skladovac silo pod tlakem. Uivatel je odpovdn za volbu vhodnho spojovacho a tsncho materilu pod prubu, v zvislosti na vyuit zazen a na pracovnch podmnkch. Nevhodn instalace me vst ze ztrt pracovnho tlaku. Poznmky:
Visac ttek pracovnch daj mus vdy zstat pipevnn na tlakotsn sestav2). Pokud je mc technologick celek vymnn, mus bt visac ttek pracovnch daj pipevnn na novou sestavu. Jednotky SITRANS LR250 jsou hydrostaticky otestovny, a pinejmenm spluj poadavky zkonu ASME o kotlch a tlakovch ndobch a Evropsk smrnice o tlakovch zazench.

esky

Smrnice o tlakovch zazench STZ 97/23/EC


Monitory hladiny Siemens vybaven zvitovm, prubovm i hygienicky tsnm technologickm spojem nemaj vlastn tlakotsn plt a tud nemohou bt povaovny za tlakov i bezpenostn zazen ve smyslu smrnice o tlakovch zazench (viz Pokyny Komise EU 1/8). 2)

1)

2)

Viz Zapojen chrnn proti explozi (pouze pro USA/Kanada) na str. 13. Sestava technologickch dl pod pracovnm tlakem obsahuje komponenty fungujc jako zbrana proti niku tlaku ze skladovacho sila: toto se tk technologickho spoje a zie, vtinou vak nikoliv elektrick skn.
Str. CS-5

A5E31997170 SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

Instalace
POZOR: Instalace me bt provedena pouze kolenmi pracovnky a v souladu s mstnmi pedpisy. Vdy manipulujte se zazenm pomoc vnjho plt. Manipulace pomoc vynvajcch st me zpsobit jeho pokozen. Bute obzvlt opatrn pi manipulaci se roubovac PVDF antnou. Sebemen pokozen povrchu antny, obzvlst na pice, me mt dsledky na jej vkon. Stavebn materily budou vybran dle jejich chemick kompatibility (inertn materily) pro veobecn pouit. Ped instalac a pouitm ve specifickm prosted, zkontrolujte tabulky chemick kompatibility. Poznmky: V lenskch zemch Evropsk unie mus instalace odpovdat normm ETSI EN 302372. Informace o homologaci najdete na vrobnm ttku. Sriov slo vyznaen na vrobnm ttku udv jednoznan identifikan slo zazen, obsahujc datum vroby. Napklad: MMDDRR XXX (MM = msc, DD = den, RR = rok vroby, XXX = sekvenn vrobn slo zazen) Dal oznaen (dle volnho msta) mohou udvat typ pruby, rozmr, tlakovou tdu, vrobn materil i teplotn kd materilu.

esky

Konstrukce hrdla
roubovac PVDF antna
min. vzdlenost: 10 mm (0,4")

Trubkovit antna z nerezov oceli


min. vzdlenost: 10 mm (0,4")

Min. prmr 51 mm (2")

Konec trubky mus vynvat nejmn 10 mm (0,4) z hrdla, aby se pedelo ruen odrazem od jeho hran. Nejmen doporuen prmr hrdla pro pouit roubovac PVDF antny je 51 mm (2"). Pro vechny typy antn krom roubovac PVDF antny lze objednat 100 mm (3.93") nstavec.

Umstn hrdla
Nepouvejte stedov umstn v ppad vysokho a zkho sila, toto me zpsobit klamn echo. Zajistte dostaten msto pro konstrukci a kryt zazen. Zajistte slunen clonu, je-li zazen umstno na pmm slunci.

Pracovn prosted

doporuen nevhodn

Str. CS-6

SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

A5E31997170

hel zen
hlem zen se rozum hel, ve kterm je intenzita vyzaovanho paprsku polovin oproti maximln intenzit zen. Intenzita vyzaovanho paprsku je maximln v ose antny. Pstroj samozejm vyzauje i vn hlu zen, ale tento signl me vst k chybnmu men. Zajistte, aby kuel zen nebyl ruen ebky, potrubm, I-nosnky nebo proudy materilu.

kuel zen

Kuel zen

hel zen

hel kuelu: roubovac PVDF antna = 19 1.5" trubka = 19 2" trubka = 15 3" trubka = 10 4" trubka = 8

Pstup k programovn
Zajistte snadn pstup pro sledovn displeje a pro programovn pomoc penosnho programtora.

esky

Instrukce k monti
POZOR: Pi prci pod tlakem je nutn pouvat psku PTFE nebo jinou souvislou tsnic hmotu, a uthnout technologick spoj na poadovanou pevnost. Maximln doporuen utahovac moment in 40 Nm (30 ft.lbs.) Poznmka: Se zazenm mete provst libovoln poet otek ani byste riskovali jeho pokozen.

roubovac verze
1) 2) Ped pipevnnm zazen na montn rozhran, zkontrolujte shodu zvit aby se pedelo jejich pokozen. Naroubujte zazen na montn rozhran a run uthnte, ppadn pouijte kl. U zazen pod tlakem viz varovn ve.

Prubov verze
POZOR: Uivatel je odpovdn za volbu vhodnho spojovacho a tsncho materilu pod prubu, v zvislosti na vyuit zazen a na pracovn podmnky.

A5E31997170 SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

Str. CS-7

Elektrick instalace
Napjen
POZOR: Na svorky stejnosmrnho napjen me bt pipojen pouze zdroj vybaven elektrickou izolac mezi vstupem a vstupem, splujc bezpenostn poadavky IEC 61010-1. Veker provozn veden mus mt izolaci vhodnou pro jmenovit napt.

Pipojen zazen SITRANS LR250


POZOR: Ovte si odpovdajc pedpisy na vrobnm ttku zazen a na visacm ttku pracovnch daj. Pouvejte vhodn kabelov prchodky, aby byl dodren pedpis IP nebo NEMA. Viz Schma zapojen v nebezpenm prosted na str. 9. Poznmky: Pouvejte pleten dvojkabel: AWG 22 a 14 (0,34 mm2 a 2,50 mm2). Standardn postupy a pravidla kladen elektrickch veden mohou vyadovat oddlen kabely a veden.
2 mm Allen klem uvolnte aretan roub vka.

esky

ucpvka (IP 68) ochrann vodi (pokud je ptomen) vhodn kabelov prchodka (nebo kabelov hrdlo NPT2))
1) 2)

1) 2) 3) 4) 5)

Obnate plt kabelu cca. 70 mm (2.75") od jeho konce, a prostrte vodie skrz prchodku. 1) 2) Pipojte vodie na svorky dle vyobrazen: polarita je vyznaena na svorkch. Uzemnte zazen v souladu s mstnmi pedpisy. Uthnte prchodku aby bylo zarueno jej utsnn. Ne zanete zazen programovat a konfigurovat uzavete vko.

1) 2)

Lze objednat zrove se zazenm. Jestlie jsou kabely vedeny potrubm, pouvejte pouze vyhovujc chytky, schvlen pro vodotsn pouit.
SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU A5E31997170

Str. CS-8

Pipojen HART
Typick PLC/mA konfigurace se zapojenm HART 1) 2)
aktivn PLC

napjec zdroj

R= 250
HART modem

SITRANS LR250

HART komuniktor

Poznmky: Podle koncepce vaeho systmu me bt zdroj energie oddlen od PLC, nebo jeho soust. Odpor HART1) mus bt dimenzovn dle pracovnch podmnek vaeho zazen2). Nikdy nepipojujte vce ne jedno programovac zazen HART najednou.

esky

Schma zapojen v nebezpenm prosted


Existuje est monost elektrickho zapojen ve vbunm prosted. V kadm ppad je nutn si ovit odpovdajc pedpisy na vrobnm ttku i na visacm ttku pracovnch daj.

1. Jiskrov bezpen zapojen


Vrobn ttek (ATEX/IECEX/INMETRO/C-TICK)

Certifikt ATEX uveden na vrobnm ttku lze sthnout z internetovch strnek produktu: www.siemens.com/LR250.) Vyberte Support (Podpora) > Approvals / Certificates (Schvlen / Certifikty). Certifikt IECEx uveden na vrobnm ttku naleznete na webovch strnkch IECEx na adrese: http://iecex.iec.ch. Zde vyberte Ex Equipment Certificates of Conformity (certifikty o shod zazen) a zadejte slo certifiktu IECEx SIR 05.0031X.

1) 2)

Celkov odpor obvodu = odpor vodi + 250 (rezistor). Viz t Graf 1: Veobecn pouit, Jiskrov bezpen, Nejiskiv/s omezenm pkonem, Nezpaln zazen na str. 23 nebo Graf 2: Ochrana proti ohni, Zven bezpenost, Ochrana proti vbuchu na str. 24.
Str. CS-9

A5E31997170 SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

Jiskrov bezpen zapojen (pokraovn)


Vrobn ttek (FM/CSA)

Vkres jiskrov bezpenho zapojen FM/CSA sla 23650653 lze sthnout z internetovch strnek produktu: www.siemens.com/LR250. Vyberte Support (Podpora) > Installation Drawings (Vkresy) > Level Measurement (Men hladiny) > Continuous Radar (Stabiln radar). Poadavky napjecho zdroje: viz Graf 1: Veobecn pouit, Jiskrov bezpen, Nejiskiv/s omezenm pkonem, Nezpaln zazen na str. 23. Poadavky elektrickho pipojen: respektujte mstn pedpisy. Schvlen prachotsn a vodotsn ochrana veden je vyadovna pro venkovn umstn NEMA 4X / typ 4X / NEMA 6, IP67, IP68. Viz Pokyny specifick pro instalaci v nebezpenm prosted na str. 13. Doporuen jiskrov bezpen ochrany jsou uveden pod Pasivn ochrana pomoc diodovho mstku na str. 10 a Aktivn ochrany (zdvojen ochrana) na str. 11.

esky

Poznmka: Volba vhodnho vstupnho modulu PLC, napjecho zdroje i ochrany vyaduje znalosti a praxi v oboru jiskrov bezpenosti. Montn firma je odpovdn za shodu dan jiskrov bezpen instalace se vemi administrativnmi kolaudanmi poadavky i s platnmi nrodnmi provdcmi pedpisy.

Pasivn ochrana pomoc diodovho mstku


Poznmka: Je teba dokonale regulovanho vstupnho napt. Vrobce MTL MTL Stahl Stahl Katalogov slo 787SP+ (Dva kanly) 7787P+ (Dva kanly) 9001/01-280-100-10 (Jeden kanl) 9002/01-280-110-10 (Dva kanly)

Jak zvolit vhodnou pasivn ochranu pro SITRANS LR250


K oven zda uvazovan bezpenostn ochrana je shodn s jiskrov bezpenmi (IS) parametry pro LR250, provete nsledujc vypoet: Re-e = max. odpor ochrany mezi koncovmi svorky Robvod = odpor obvodu (souet odpor vech vodi a odpor vech vloench komponent) Vochrana = hodnota nelinernho poklesu napt zpsobenho ochranou 1) Podle tabulky dat urete hodnotu Re-e. 2) Spotejte celkovou hodnotu Robvod 3) Vypotejte Rpracovn = Re-e + Robvod.

Str. CS-10

SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

A5E31997170

Jak zvolit vhodnou pasivn ochranu (pokraovn)


4) 5) 6) Podle tabulky dat urete Vochrana (napklad pokles napt v diodch). Vypotejte Vpracovn = Vzdroj - Vochrana. Pomoc hodnot Vpracovn a Rpracovn zkontrolujte zda provozn podmnky jsou uvnit stnovan zny grafu Graf 1: Veobecn pouit, Jiskrov bezpen, Nejiskiv/s omezenm pkonem, Nezpaln zazen na str. 23. Katalogov slo 706 7206 SITRANS I 7NG4124 9001/51-280-110-14 E02009

Aktivn ochrany (zdvojen ochrana)


Vrobce MTL MTL Siemens Stahl MTL MTL

esky

E02010

2. Zapojen Nejiskiv/S omezenm pkonem

Certifikt ATEX uveden na vrobnm ttku lze sthnout z internetovch strnek produktu: www.siemens.com/LR250. Vyberte Support (Podpora) > Approvals / Certificates (Schvlen / Certifikty). Poadavky napjecho zdroje: viz Graf 1: Veobecn pouit, Jiskrov bezpen, Nejiskiv/s omezenm pkonem, Nezpaln zazen na str. 23. Pi zapojen zazen respektujte mstn pedpisy.

3. Nezpaln zazen

Vkres zapojen FM/CSA Tdy 1, Div. 2 sla 23650673 lze sthnout z internetovch strnek produktu: www.siemens.com/LR250. Vyberte Support (Podpora) > Installation Drawings (Vkresy) > Level Measurement (Men hladiny) > Continuous Radar (Stabiln radar). Poadavky napjecho zdroje: viz Graf 1: Veobecn pouit, Jiskrov bezpen, Nejiskiv/s omezenm pkonem, Nezpaln zazen na str. 23.

A5E31997170 SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

Str. CS-11

4. Ohnivzdorn zapojen

Certifikt ATEX uveden na vrobnm ttku lze sthnout z internetovch strnek produktu: www.siemens.com/LR250. Vyberte: Support (Podpora)> Approvals (Schvlen) / Certificates (Certifikty). Certifikt IECEx uveden na vrobnm ttku naleznete na webovch strnkch IECEx na adrese: http://iecex.iec.ch. Zde vyberte Ex Equipment Certificates of Conformity (certifikty o shod zazen) a zadejte slo certifiktu IECEx SIR 08.0107X. Poadavky napjecho zdroje: viz Graf 2: Ochrana proti ohni, Zven bezpenost, Ochrana proti vbuchu na str. 24. Pi zapojen zazen respektujte mstn pedpisy. Viz t Pokyny specifick pro instalaci v nebezpenm prosted na str. 13 a ve uveden certifikt ATEX.

esky

5. Zapojen se zvenou bezpenost

Certifikt ATEX uveden na vrobnm ttku lze sthnout z internetovch strnek produktu: www.siemens.com/LR250. Vyberte: Support (Podpora)> Approvals (Schvlen) / Certificates (Certifikty). Certifikt IECEx uveden na vrobnm ttku naleznete na webovch strnkch IECEx na adrese: http://iecex.iec.ch. Zde vyberte Ex Equipment Certificates of Conformity (certifikty o shod zazen) a zadejte slo certifiktu IECEx SIR 08.0107X. Poadavky napjecho zdroje: viz Graf 2: Ochrana proti ohni, Zven bezpenost, Ochrana proti vbuchu na str. 24. Pi zapojen zazen respektujte mstn pedpisy. Viz t Pokyny specifick pro instalaci v nebezpenm prosted na str. 13 a ve uveden certifikt ATEX.

Str. CS-12

SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

A5E31997170

6. Zapojen chrnn proti explozi (pouze pro USA/Kanada)

Vkres zapojen chrnnho proti explozi FM/CSA, sla A5E02257843 lze sthnout z internetovch strnek produktu: www.siemens.com/LR250. Vyberte Support (Podpora) > Installation Drawings (Vkresy) > Level Measurement (Men hladiny) > Continuous Radar (Stabiln radar). Poadavky napjecho zdroje: viz Graf 2: Ochrana proti ohni, Zven bezpenost, Ochrana proti vbuchu na str. 24.

esky

Pokyny specifick pro instalaci v nebezpenm prosted


(Viz evropskou smrnici ATEX 94/9/EC, Ploha II, 1/0/6)
Nsledujc pokyny se tkaj zazen spadajc pod certifikty SIRA 06ATEX2358X, SIRA 08ATEX1301X a SIRA 09ATEX4153X. Uvn a mont mus odpovdat zkladnm pokynm. Toto zazen je certifikovno v kategorii 1GD certifiktem SIRA 06ATEX2358X, v kategorich 1/2 GD, 1D, 2D certifiktem SIRA 08ATEX1301X a v kategorii 3G certifiktem SIRA 09ATEX4153X. 3) Toto zazen lze pouvat pi styku s holavmi plyny a vpary, spolen s aparty skupiny IIC, IIB a IIA a pi teplotnch tdch T1, T2, T3 a T4. 4) Toto zazen disponuje stupnm ochrany proti vniku vody IP67, pat do teplotn tdy T90 C a me bt pouito ve styku s holavm prachem. 5) Toto zazen je certifikovno pro pouit pi teplot prosted v rozmez40 C a +80 C. 6) Toto zazen nebylo uznno jako bezpenostn zazen (ve smyslu smrnice 94/9/EC, Ploha II, odstavec 1.5). Viz Smrnice o tlakovch zazench STZ 97/23/EC na str. 5. 7) Instalovat a kontrolovat toto zazen mohou pouze patin vykolen osoby v souladu s poadavky platn legislativy (v Evrop EN 60079-14 a EN 60079-17). 8) Toto zazen nen opraviteln svpomoc. 9) sla odpovdajcch certifikt maj pedponu X, co se vztahuje ke specifickm podmnkm bezpenho uvn. Osoby provdjc instalaci nebo kontrolu tchto zazen mus mt pstup k pslunm certifiktm. 10) Jestlie by zazen mohlo pijt do styku s agresivnmi ltkami, je uivatel zodpovdn za vhodn bezpenostn opaten, aby bylo zazen chrnno ped nepznivmi vlivy, a za to, e vlastn zpsob ochrany nebude ohroen. Agresivn ltky jsou napklad kyseliny nebo plyny, kter mohou zpsobit korozi kovu, nebo rozpoutdla, kter poruuj polymery. Vhodn bezpenostn opaten je napklad sledovn pomoc materilovch karet odolnost zazen vi uritm chemiklim. 1) 2)

A5E31997170 SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

Str. CS-13

Programovn SITRANS LR250


Viz Sputn Prvodce rychlm uvedenm do provozu pes penosn programtor na str. 18. Viz Sputn Prvodce rychlm uvedenm do provozu pes SIMATIC PDM na str. 21.

Uveden SITRANS LR250 do provozu


Zapnte zazen. SITRANS LR250 se automaticky spust v mcm mdu.

LCD displej Zobrazen LCD displeje v mcm mdu (normln innost) 1) 2)


1 2 3 4

esky

1 indiktor men1) v linern jednotce nebo v % 2 zvolen funkce: level (rove hladiny), space (prostor), distance (vzdlenost) nebo volume (objem) 3 namen hodnota (rove hladiny, objem, prostor nebo vzdlenost) 4 jednotka 5 sloupcov graf znzorujc hodnotu 6 druh dek indikuje na poadavek2) teplotu elektroniky, spolehlivost odezvy, proud v obvodu nebo vzdlenost 7 textov dek zobrazuje stavov zprvy 8 indikace stavu zazen (podrobnosti naleznete v kompletnm manulu)

Indikace poruchy S: 0 LOE

V ppad poruchy na zazen se na textovm dku zobraz kd chyby a zprva o porue (7), a indikace stavu zazen zobraz ikonku nutn drby (8)

Zobrazen v programovacm mdu Pi navigaci

aktuln aktuln Ptomnost sloupce navigace v menu poloky menu indikuje e seznam poloek sloupec v danm menu je pli dlouh aby navigace v aktuln menu poloka byl cel zobrazen na displeji. pruh poloky Pruh situovn v polovin vky sloupce indikuje e tato poloka je piblin uprosted seznamu poloek danho menu. Relativn vka a poloha pruhu aktuln poloky na sloupci navigace znzoruje dlku seznamu poloek danho menu a piblinou polohu tto poloky v nm. m ir pruh tm je mn poloek v menu.

slo

Pi zmn parametr
slo parametru nzev parametru hodnota/volba parametru

Pi editaci

1)
2)

Pepnte stiskem ipky nahoru nebo dolu. Po stisknut vhodnho tlatka. Vce podrobnost naleznete v tabulce Funkce tlatek v mcm mdu na str. 15.
SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU A5E31997170

Str. CS-14

Penosn programtor (Kat. . 7ML1930-1BK)


Je objednvn zvlt. 1)
2) 3)

Tlatko

Funkce tlatek v mcm mdu Aktualizuje proud v obvodu1). Aktualizuje zmenou teplotu uvnit sila1). Aktualizuje odhad spolehlivosti odezvy1).

Aktualizuje zmenou vzdlenost1). Tlatko Mode otevr programovac md2). Prav ipka otevr programovac md3). ipka nahoru nebo dolu pepn mezi linernm menm a %

esky

Programovn pomoc penosnho programtora


Poznmky: Stisknte Mode (Md) pro pechod mezi Mcm a Programovacm mdem. Pokud je zazen v programovacm mdu, jsou vstupy zablokovan a nereaguj na ppadn zmny mench hodnot. SITRANS LR250 se automaticky vrt do Mcho mdu po uplynut urit lhty neinnosti v Programovacm mdu (mezi 15 vtein a 10 minut, podle aktulnho menu).

Menu parametr
Parametry jsou identifikovny nzvem a organizovny do funknch skupin. Kompletn seznam parametr a instrukc k pouit najdete v kompletnm manulu.

1. QUICK START 2. SETUP 2.1. DEVICE ............. 2.7. LINEARIZATION 2.7.1. VOLUME 2.7.1.1. VESSEL SHAPE

1) 2)

3)

Nov hodnota je zobrazena na druhm dku LCD displeje. Oteve posledn menu programovacho mdu kter bylo zobrazeno v tomto pracovnm cyklu. Pokud v tomto pracovnm cyklu nebyl pouit programovac md nebo uplynulo vce ne 10 minut od jeho poslednho pouit, bude oteveno prvn menu programovacho mdu. Otevr prvn menu programovacho mdu.
Str. CS-15

A5E31997170 SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

1. Vstup do Programovacho mdu


Zamte programtor na displej (z maximln vzdlenosti 30 cm [1 ft.]). Pravou ipkou aktivujete programovac md a zobrazte menu 1. rovn. Tlatkem Mode otevete posledn pouit menu v programovacm mdu bhem uplynulch deseti minut, nebo menu 1. rovn v ppad e bylo zazen restartovno.

penosn programtor LCD displej

Max. 30 cm (1 ft)

2. Navigace: funkce tlatek v naviganm mdu esky


Poznmka: Pro rychl pstup k parametrm pomoc penosnho programtora, stisknte Home (Vchoz) Tlatko Nzev
ipka nahoru nebo dolu

, a zadejte slo poadovanho menu, nap.: 2.7.1 (Hlasitost). rove menu


menu nebo parametry menu

Funkce v naviganm mdu


Rolovn k pedelmu nebo nsledujcmu menu i parametru. Pstup k prvnmu parametru aktulnho menu, nebo postup do nsledujcho menu. Oteve editan md. Oteve pedchoz menu. Pechod do Mcho mdu. Oteve menu prvn rovn: menu 1.

Prav ipka parametry Lev ipka Md Vchoz menu nebo parametry menu nebo parametry menu nebo parametry

3. Editace v programovacm mdu Zmna volby ze seznamu


a) b) Pejdte k danmu parametru. Stisknutm prav ipky nhled do parametru. otevete

nzev parametru

slo parametru

aktuln volba

c)

Druhm stisknutm prav ipky otevete editan md. Aktuln volba je zvraznna. Rolujte k nov volb.

d)

Stisknutm prav ipky

ji akceptujte

LCD displej se vrac do nhledu do parametru a zobrazuje novou volbu.

Str. CS-16

SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

A5E31997170

Zmna seln hodnoty


a) b) Pejdte k danmu parametru. Druhm stisknutm prav ipky otevete Editan md. Aktuln hodnota je zvraznna. Pomoc tlatek zadejte novou hodnotu. Stisknutm prav ipky ji akceptujte. LCD displej se vrac do nhledu do parametru a zobrazuje novou volbu.

nzev parametru

slo parametru

c) d)

aktuln hodnota

Funkce tlatek v Editanm mdu esky


Tlatko Nzev
ipka nahoru nebo dolu

Funkce v editanm mdu


Zmna volby Numerick editace Zmna volby Rolovn k dan poloce. - Zven nebo snen seln hodnoty. - Pepnn znamnek plus a minus. - Potvrzen volby (uloen parametru) - Pechod z Editanho do Naviganho mdu - Posun kurzoru o jedno polko vpravo - Pokud je kurzor na znaku Enter, potvrzen volby a pechod z Editanho do Naviganho mdu Oputn Editanho mdu beze zmny parametr. Pi prvnm stisku posun kurzoru na znamnko plus/ minus Jinak posun kurzoru o jedno polko vlevo. Vymazn daj na displeji

Prav ipka

Numerick editace Zmna volby

Lev ipka

Numerick editace

Vymazn Desetinn teka

Numerick editace

Numerick editace

Vloen desetinn teky.

Znamnko plus nebo Numerick editace minus

Zmna znamnka ped editovanou hodnotou.

Numerick Numerick editace tlatka

Vloen odpovdajcho znaku

na

A5E31997170 SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

Str. CS-17

Sputn Prvodce rychlm uvedenm do provozu pes penosn programtor


1. Rychl uveden do provozu
a) Zamte programtor na displej z maximln vzdlenosti 30 cm (1 ft), nsledn stisknte PRAVOU ipku Menu prvn rovn. b) Stisknte dvakrt PRAVOU ipku do parametr. c) Stisknte PRAVOU ipku dolu pro vstup do Editanho mdu nebo ipku pro pstup k poloce 1.1 v menu a nhled pro oteven programovacho mdu a oteven

pro potvrzen vchozho nastaven a posun k dal poloce.

esky

d)

Jestlie chcete zmnit nastaven, rolujte k poadovan volb nebo zadejte novou hodnotu pomoc tlatek.

e)

Jakmile zmnte hodnotu, stisknte PRAVOU ipku stisknte ipku dolu pro posun k dal poloce.

pro jej potvrzen a

f)

Zmny proveden pomoc Prvodce rychlm uvedenm do provozu se projev a na konci celho postupu po volb Yes (Ano) u Apply changes (Provzt zmny) v kroku (1.8.).

1.1. Language (Jazyk)


Volba jazyka ve kterm se budou zobrazovat zprvy na LCD displeji. Tato volba se projev okamit. Monosti ENGLISH (Anglicky), DEUTSCH (Nmecky), FRANAIS (Francouzsky), ESPAOL (panlsky)

1.2. Material (Materil)


Monosti LIQUID (Tekutina) LIQUID LOW DK (Lehce dielektrick tekutina)

1.3. Response Rate (Rychlost reakce)


Zadejte rychlost reakce zazen pi men zmn ve sledovan hodnot. Rychlost reakce SLOW (Pomal) MED (Stedn) FAST (Rychl) Reakce pi plnn/vyprazdovn 0,1 m/min (0,32 ft/min) 1,0 m/min (3,28 ft/min) 10.0 m/min (32.8 ft/min)

Nastavte vt rychlost reakce ne je maximln rychlost plnn i vyprazdovn ale ne pli.

1.4. Units (Jednotky)


Jednotky men senzorem. Monosti M, CM, MM, FT, IN

Str. CS-18

SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

A5E31997170

1.5. Pracovn md
referenn bod senzoru Horn kalibran bod

referenn bod senzoru prostor

vzdlenost

hladina referenn bod senzoru

doln kalibran bod

esky

Operation Popis (Pracovn md) NO SERVICE (Stop) LEVEL (Hladina) SPACE (Prostor) DISTANCE (Vzdlenost)
1)

Namen hodnota a odpovdajc napt v obvodu nebudou aktualizovny a zazen se pepne do klidovho reimu 1). Vzdlenost povrchu materilu od dolnho kalibranho bodu. Vzdlenost povrchu materilu od hornho kalibranho bodu. Vzdlenost povrchu materilu od referennho bodu senzoru.

Vce podrobnost o klidovm reimu naleznete v kompletnm manulu.

1.6. Low Calibration Point (Doln kalibran bod)


Vzdlenost referennho bodu senzoru od dolnho kalibranho bodu: pouv metodu nulov hladiny. Viz ilustraci Pracovn md (1.5.) Hodnoty Rozsah: 0,00 - 20,00 m

1.7. High Calibration Point (Horn kalibran bod)


Vzdlenost referennho bodu senzoru od hornho kalibranho bodu: pouv metodu pln hladiny. Viz ilustraci Pracovn md (1.5.) . Hodnoty Rozsah: 0,00 - 20,00 m

1.8. Apply? (Provst zmny?)


Aby bylo nastaven Rychlho uveden do provozu uloeno, je nutn zvolit Yes (Ano). Monosti YES (Ano), NO (Ne), DONE (Provedeno) - Npis DONE (Provedeno) na displeji potvrzuje spn proveden zmn pomoc Rychlho uveden do provozu. pro nvrat do mcho mdu. SITRANS LR250 je nyn

Stisknte Mode (Md) pipraveno k innosti.

Poznmka: Pokud ndoba, ve kter probh vae aplikace, obsahuje pekky, pak pouijte kompletn manul pro zjitn podrobnost o pouit funkce automatickho odstrann falench odraz (Auto False Echo Suppression).

A5E31997170 SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

Str. CS-19

SITRANS LR250 komunikan systm: HART


Seznam vech pouitelnch parametr naleznete v kompletnm manulu. K programovn vaeho zazen vm doporuujeme pouvat SIMATIC PDM (Process Device Manager programovac nstroj zazen). Nvod pro nastaven nstroj HART pes SIMATIC PDM je dostupn na naich webovch strnkch: www.siemens.com/LR250 v sekci Support (Podpora).

SIMATIC PDM
SIMATIC PDM je softwarov balek uren k obsluze a drb SITRANS LR250 a jinch podobnch zazen. Prosme respektujte instrukce k pouit SIMATIC PDM obsaen v nvodu nebo na on-line npovd. (Vce informac naleznete na strnkch www.siemens.com/simatic-pdm. Spoj SIMATIC PDM s modemem HART v operanm systmu Windows 2000 nebo Windows XP 1).

Deaktivace vyrovnvacch pamt

esky

Poznmky: SIMATIC PDM spolupracuje pouze s verz operanho systmu Windows XP Professional, nikoliv s verz XP Home. K deaktivaci vyrovnvacch pamt muste mt prva administrtora operanho systmu. 1) 2) 3) 4) 5) 6) Volbou Start/Nastaven/Ovldac panely spustt konfigurovn. Kliknte dvakrt na Systm, zvolte zloku Hardware, potom kliknte na tlatko Sprvce zazen. Otevete poloku Porty, a nsledn otevete okno Vlastnosti komunikanho portu, dvojitm kliknutm na Komunikan port, kter systm pouv. Zvolte zloku Nastaven portu a kliknte dvakrt na tlatko Upesnit. Pokud je zvoleno Pouit vyrovnvac pamti FIFO v zakrtvacm polku, kliknte na nj a tm ji deaktivujete. Volbou OK konfigurovn ukonte. Zavete vechna oteven okna a systm restartujte.

Konfigurtor elektronickho zazen (EDD)


Poznmky: SIMATIC PDM Rev. 5.2 SP1 lze pouvat pouze pro zkladn konfiguraci a odstraovan zvad. Pro pokroil funkce jako nap. Prvodce rychlm startem je zapoteb Rev. 6.0 SP4 nebo vy. Soubor EDD naleznete v SIMATIC PDM ve sloce Device Catalog (Katalog zazen), pod Sensors/Level/Echo/Siemens Milltronics/SITRANS LR250. Zkontrolujte si na naich webovch strnkch: www.siemens.com/LR250 zda vlastnte nejnovj verzi SIMATIC PDM, vetn poslednho Service Pack (SP) a nejnovjho Hot Fix (HF). Vyberte Support (Podpora) > Software Downloads (Soubory ke staen). Sthnte soubory EDD z internetovch strnek produktu: www.siemens.com/LR250 a ulote je do vaeho potae. Zip soubory rozpakujte do snadno dosaiteln sloky. Spustte SIMATIC PDM Manager Device Catalog (katalog konfiguranch nstroj), najdte a vyberte EDD soubor nlec k vaemu zazen.

Instalace novho souboru EDD:


1) 2) 3)

1)

Windows je registrovan znaka Microsoft Corporation


SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU A5E31997170

Str. CS-20

Konfigurace novho zazen


Poznmka: Kliknut na Cancel (Zruit) bhem penosu dat mezi zazenm a SIMATIC PDM nezru ji proveden update nkterch parametr. 1) 2) Zkontrolujte, zda vlastnte nejnovj verzi EDD. Pokud ne, aktualizujte si ji (viz Konfigurtor elektronickho zazen (EDD) na str. 20). Spustte SIMATIC Manager a vytvote nov projekt pro LR250. Nvod pro nastaven nstroj HART pes SIMATIC PDM je dostupn na naich webovch strnkch: www.siemens.com/LR250. Otevete menu Device Master Reset (Hlavn reset zazen) a kliknutm na OK spustte reset do vchozch parametr. Po ukonen Resetu, nahrajte parametry do PC/PG. Nastavte zazen pomoc Prvodce rychlm uvedenm do provozu.

3) 4) 5)

Sputn Prvodce rychlm uvedenm do provozu pes SIMATIC PDM


Poznmky: Jednotliv parametry Prvodce rychlm uvedenm do provozu spolu souvisej a proveden zmny budou aktivovny kliknutm na FINISH AND DOWNLOAD (Ukonen a penos dat) na konci kroku 5, m ulote zadan parametry a penesete je do zazen. Kliknutm na BACK (Zpt) se vrtte k pedel poloce pro jej opravu, a kliknutm na Cancel (Zruit) ukonte Rychl uveden do provozu. Spustte SIMATIC PDM, otevete menu Device Quick Start (Zazen Rychl uveden do provozu) a postupujte podle krok 1 a 5.

esky

Rychl uveden do provozu

Ovldn pes FDT (Field Device Tool Servisn nstroj pro provozn zazen)
FDT je standard pouvan mnoha programovmi balky pro zen a servis provoznch zazen. Dv FDT jsou komern dostupn: PACTware a Fieldcare. Pro konfiguraci provoznho zazen pes FDT potebujete DTM (Device Type Manager) odpovdajc danmu zazen. Siemens instruments pouv SITRANS DTM spolen s nstrojem EDD, vyvynutm pro SITRANS DTM. 1) Za prv nainstalujte SITRANS DTM do vaeho systmu. Lze jej sthnout na adrese: http://support.automation.siemens.com. Vyberte Product Support (Podpora produkt) a zvolte postupn Product Information/Automation Technology/Sensor systems/Process Instrumentation/Software & Communications. Nsledn nainstalujte nstroj SITRANS LR250 HART EDD pro SITRANS DTM. Lze jej sthnout z internetovch strnek produktu: www.siemens.com/LR250. Vyberte Support (Podpora) > Software Downloads (Soubory ke staen).

2)

Konfigurace novho zazen pes FDT


Nvod k pouit lze t sthnout z internetovch strnek produktu pod volbou Support (Podpora).
A5E31997170 SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU Str. CS-21

Ovldn pes AMS Device Manager


AMS Device Manager je programov balek pro zen a servis provoznch zazen. Prosme respektujte instrukce k pouit AMS Device Manager obsaen v nvodu nebo na on-line npovd. Vce informac naleznete na: http://www.emersonprocess.com/AMS/.

Konfigurtor elektronickho zazen (EDD)


Ke konfiguraci SITRANS LR250 potebujete EDD vyvynut pro AMS Device Manager v. 9.0.

Konfigurace novho zazen pes AMS Device Manager


1) Zkontrolujte si na naich webovch strnkch www.siemens.com/LR250 zda vlastnte nejnovj verzi EDD. Vyberte Support (Podpora) > Software Downloads (Soubory ke staen). Pokud je to nutn, sthnte nov soubory. Ulote soubory do vaeho potae a rozpakujte je do snadno dosaiteln sloky. Spustte AMS Device manager Add Device Type (Pidat typ zazen), najdte a vyberte odpovdajc zapakovan EDD soubor.

2)

esky

Spustte AMS Device Manager. Nvod pro nastaven zazen typu HART pes AMS Device Manager lze sthnout z internetovch strnek produktu pod volbou Support (Podpora). SITRANS LR250 nevyaduje dnou drbu ani itn v normlnch pracovnch podmnkch. Pokud je istn nutn z dvodu nronch pracovnch podmnek, postupujte nsledovn: 1) 2) Vnujte pozornost konstruknmu materilu antny a skladovanmu mdiu, a zvolte istic prostedky kter na n nebudou nepzniv psobit. Odstavte zazen z provozu a oistte antnu pomoc tkaniny naputn adekvtn istic tekutinou.

drba

Oprava zazen a vyjmut ze zruky


Viz podrobnosti na vnitn stran krytu.

Vmna antny, elektronick jednotky i plst


Je-li nutn vmna antny z dvodu jejho pokozen nebo poruchy, je mon tuto vmnu provst, ani by to vyadovalo novou kalibraci. Vmna za jin typ antny me bt provedena autorizovanm servisnm centrem nebo zamstnanci spolenosti Siemens. Je-li nutn vmna elektronick jednotky i plst z dvodu pokozen nebo poruchy, ujistte se, zda je pouit opovdajc typ antny. Jinak by bylo nutn provst novou kalibraci autorizovanmi zamstnanci spolenosti Siemens.

Str. CS-22

SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

A5E31997170

Diagram maximln pracovn teploty


SITRANS LR250 s prubovm uchycenm
POZOR: Vnitn teplota nesm pesahovat 80 C (176 F). Poznmky: Tento graf je pouze informativn, jeliko neme zohlednit vechny mon pracovn podmnky a procesy. (Kadopdn jej nelze pout pokud je SITRANS LR250 namontovn pmo na povrch kovov ndoby.) Takt tento graf nezohleduje zahvn v ppad umstn zazen na pmm slunci. Parametr 3.2.1 Current Internal Temperature (Aktuln vnitn teplota) sleduje vnitn teplotu.

SITRANS LR250 s prubovm uchycenm


Teplota prosted (oC)

Maximln provozn teplota / teplota prosted

esky

provozn teplota Provozn teplota (oC)

Vkon elektrickho obvodu


Poadavky na napjec zdroj
Poznmka: Ne uveden grafy plat pro standardn zazen konfigurovan pomoc penosnho programtoru Siemens.
Graf 1: Veobecn pouit, Jiskrov bezpen, Nejiskiv/ s omezenm pkonem, Nezpaln zazen

Odpor obvodu RL

Povolen pracovn zna

Napt v obvodu VL

Poznmka: Pi pouit komunikanho protokolu HART s odporem 220 (RL) je nejmen ppustn napt 16,3 V =.

A5E31997170 SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

Str. CS-23

Graf 2: Ochrana proti ohni, Zven bezpenost, Ochrana proti vbuchu


Odpor obvodu RL

Povolen pracovn zna

Napt v obvodu VL

esky

Poznmka: Pi pouit komunikanho protokolu HART s odporem 220 (RL) je nejmen ppustn napt 20,94 V =.

Chovn zazen pi sputen


Spoteba zazen pi sputn je maximln 3,6 mA. Prvn men probhn do 50 vtein

Str. CS-24

SITRANS LR250 (HART) MANUL RYCHLHO UVEDEN DO PROVOZU

A5E31997170

SITRANS LR250 (HART) Kvikstart manual


Denne manual opridser de vsentligste karakteristika og funktioner af SITRANS LR250 (HART 1)). Vi anbefaler Dem kraftigt at anskaffe den detaljerede version af denne manual, s De kan anvende apparatet fuldt ud. Den komplette manual kan downloades fra produktsiden for SITRANS LR250 p vort website: www.siemens.com/LR250. Den trykte manual kan fs hos Deres lokale Siemens Milltronics reprsentant. Sprgsml vedrrende indholdet af denne manual kan rettes til: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 E-mail: techpubs.smpi@siemens.com

Copyright Siemens AG 2013. Alle rettigheder forbeholdes


Vi opfordrer brugerne til at anskaffe de autoriserede, indbundne manualer eller lse de elektroniske versioner, der er udarbejdet og skrevet af Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments ptager sig intet ansvar for indholdet af delvise eller fuldstndige gengivelser af indbundne eller elektroniske versioner.

Ansvarsfragelse
Sknt vi har kontrolleret, at indholdet af denne manual stemmer overens med de beskrevne instrumenter, kan der stadig forekomme variationer. Vi kan derfor ikke garantere en fuldstndig overensstemmelse. Indholdet af denne manual revideres jvnligt, og eventuelle rettelser inkluderes i de efterflgende udgaver. Vi modtager gerne forslag til forbedringer. Retten til ndringer af de tekniske data forbeholdes.

Dansk

MILLTRONICS er et registreret varemrke, der tilhrer Siemens Milltronics Process Instruments.

Teknisk Support
Support er tilgngelig 24 timer i dgnet. Adresse, telefon- og faxnummer p Siemens Automations lokale kontor kan findes p: www.siemens.com/automation/partner: Klik p fanen Contact (Kontakt), vlg Service og klik derefter en gang til p Service for at finde den relevante produktgruppe (+Automation Technology (Automatiseringsteknik) > +Sensor Systems (Flersystemer) > +Process Instrumentation (Procesinstrumentering) > +Level Measurement (Niveaumling) > +Continuous (Kontinuert)). Vlg Radar. Vlg land og derefter by/egn. Vlg Technical Support (Teknisk support) under Service. For on-line teknisk support, g til: www.siemens.com/automation/support-request Indtast apparatets navn (SITRANS LR250) eller ordrenummeret, klik p Search (Sg) og vlg den relevante produkttype. Klik p Next. Indtast et ngleord, der beskriver problemet. Gennemlb derefter den relevante dokumentation eller klik p Next for at e-maile en beskrivelse af problemet til personalet i Siemens Tekniske Support. Siemens IA/DT Technical Support Center: telefon
1)

+49 (0)911 895 7222

HARTer et registreret varemrke, der tilhrer HART Communication Foundation.

A5E31997170

SITRANS LR250 (HART) KVIKSTART MANUAL

Side DA-1

Sikkerhedsvejledning

De anfrte advarsler skal overholdes for at sikre egen og andres sikkerhed samt for at beskytte produktet og det tilknyttede udstyr. Disse advarsler ledsages af en tydeliggrelse af graden af forsigtighed, der br overholdes. ADVARSEL: vedrrer et advarselssymbol p produktet og betyder, at en manglende overholdelse af de ndvendige forholdsregler kan fre til dd, alvorlig personskade og/eller omfattende materielle skader. ADVARSEL: anvendes, nr der ikke er noget tilsvarende advarselssymbol p produktet, og betyder, at en manglende overholdelse af de ndvendige forholdsregler kan fre til dd, alvorlig personskade og/eller omfattende materielle skader. Bemrkning: betyder vigtige oplysninger om produktet eller denne del af brugsvejledningen.

FCC-overensstemmelse
Kun for installationer i USA: Regler fra Federal Communications Commission (FCC)
ADVARSEL: ndringer eller modifikationer, der ikke er udtrykkeligt godkendt af Siemens Milltronics, kan ophve brugerens ret til at benytte udstyret. Bemrkninger: Dette udstyr er blevet testet og fundet at overholde grnserne for en klasse A digital anordning i henhold til Afsnit 15 i FCC-reglerne. Disse grnser er beregnet til at yde en rimelig beskyttelse mod skadelige interferenser, nr anordningen anvendes i kommercielle omgivelser. Dette udstyr frembringer, bruger og kan udstrle radiofrekvent energi, og kan, sfremt det ikke installeres og bruges i overensstemmelse med instruktionsbogen, forrsage interferens, der kan virke forstyrrende for radiokommunikationen. Brugen af dette udstyr i et beboelsesomrde vil sandsynligvis forrsage skadelig interferens med radiokommunikationer, som brugeren i givet fald vil vre ndsaget til at udbedre for egen regning.

Dansk

SITRANS LR250
ADVARSEL: SITRANS LR250 br kun anvendes som beskrevet i denne manual, da den beskyttelse, udstyret yder, ellers kan forringes. Bemrkning: Dette produkt er beregnet til at anvendes i industriomrder. Brugen at dette udstyr i et beboelsesomrde kan forrsage interferens med diverse frekvensbaserede kommunikationsmidler. SITRANS LR250 er en totrds 25 GHz pulsradarniveautransmitter til kontinuert overvgning af vsker og opslmninger i lagertanke, inklusive ved hjt tryk og hj temperatur, i en afstand af op til 20 m (66 ft). Den er ideelt egnet til sm tanke og medier med lav dielektricitetskonstant. Dette apparat bestr af et elektronisk kredslb, der er koblet til en hornantenne og en procestilslutning af enten gevind- eller flangetypen. Der fs ogs en PVDF-antenne med gevind. SITRANS LR250 understtter HART kommunikationsprotokol og SIMATIC PDM software. Signaler behandles ved hjlp af Process Intelligence.

Side DA-2

SITRANS LR250 (HART) KVIKSTART MANUAL

A5E31997170

Tekniske data

For en fuldstndig liste, jf. SITRANS LR250 (HART) Instruktionsbogen. For oplysninger om godkendelser, se Godkendelser p side 4.

Omgivende temperatur/Driftstemperatur
Bemrk:
Den maksimale temperatur afhnger af procestilslutningen, antennematerialerne og trykket i beholderen: se Diagram over maksimale procestemperaturer p side 23. For yderligere oplysninger, se kurver over procestryk- og temperaturreduktionen i den komplette manual. De tilladelige procestemperaturer og tryk afhnger af oplysningerne p procestilslutningsskiltet. Referencetegningen, der er angivet p skiltet, kan fs p produktsiden p vort website p www.siemens.com/LR250, under Support > Installation Drawings (Installationsdiagrammer) > Level Measurement (Niveaumling) > Continuous Radar (Kontinuert - Radar). Yderligere oplysninger om procestilslutninger kan fs p siden Installation Drawings (Installationsdiagrammer) under Process Connection Diagrams (Procestilslutningsdiagrammer). Signalamplituden stiger med horndiameteren, s brug den strste anvendelige strrelse. Eventuelle forlngelser (ekstraudstyr) kan installeres under gevindet. Se Diagram over maksimale procestemperaturer p side 23 for yderligere detaljer.

omgivelsestemperatur (omkring instrumenthuset) 40 C til 80 C (40 F til 176 C) procestemperaturen ved procestilslutningen: standard hornantenne: - med FKM O-ring: 40 oC til +200 oC (-40 oF til +392 oF) - med FFKM O-ring: 20 oC til +200 oC (-4 oF til +392 oF) PVDF-antenne med 2" NPT / BSPT / G-gevind: 40 til +80 C (40 til +176 F)

apparatets navneplade

Dansk
procestilslutningsskilt

Strmforsyning
Almen brug Egensikker Ikke-gnistdannende/Energibegrnset Ikke-antndingsfarlig (FM/CSA kun USA/Canada) Flammesikker Forhjet sikkerhed Eksplosionssikker (FM/CSA kun USA/Canada) Max. 30 V dc Fra 4 til 20 mA Max. opstartsstrmstyrke: Nominel 24 V dc ved 550 Ohm

Nominel 24 V dc ved 250 Ohm

se Egenskaber under opstart p side 24

A5E31997170

SITRANS LR250 (HART) KVIKSTART MANUAL

Side DA-3

Godkendelser
Bemrk: Apparatets navneplade angiver de godkendelser, der glder for netop Deres apparat. Almen brug Radio Risikoomrder Egensikker 1) (Europa) (International) (USA/Canada) CSAUSA/C, FM, CE, C-TICK, KC Europa (R&TTE), FCC, Industry Canada

(Brasilien)

ATEX II 1G, EEx ia IIC T4 ATEX II 1D, EEx tD A20 IP67 T90 C IECEx SIR 05.0031X, Ex ia IIC T4, Ex tD A20 IP67 T90 C FM/CSA Klasse I, Div. 1, Gr. A, B, C, D Klasse II, Div. 1, Gr. E, F, G Klasse III T4 INMETRO: DNV 12.0087 X Ex ia IIC T4 Ga Ex ia ta IIIC T90 C Da IP65/IP67 -40 C Ta +80 C DNV #OCP 0017

(Kina) NEPSI Ex ia IIC T4/ DIP A20 TA, T90 C IP67 Ikke-gnistdannende/ Energibegrnset 2) (Europa) ATEX II 3 G Ex nA/nL IIC T4 Gc (Kina) NEPSI Ex nA II T4/ Ex nL IIC T4 Ikke-antndingsfarlig 3) (USA/Canada) FM/CSA Klasse I, Div. 2, Gr. A, B, C, D T5 4) Flammesikker (Europa/International) ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex dmbia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C (Brasilien) INMETRO: DNV 12.0088 X Ex d ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017

ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

Dansk

ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-1:2009, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-11-2008, ABNT NBR IEC 60079-31 2011 e IEC 61241-11:2005

1) 2) 3) 4)

Se Egensikker elinstallation p side 9. Se Ikke-gnistdannende/Energibegrnset elinstallation p side 11. Se Ikke-antndingsfarlig elinstallation p side 11. Se Flammesikker elinstallation p side 12.

Side DA-4

SITRANS LR250 (HART) KVIKSTART MANUAL

A5E31997170

Risikoomrder (fortsat) Forhjet sikkerhed 1)


(Europa/International) ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex embia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C INMETRO: DNV 12.0088 X Ex e ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017 ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-7:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

(Brasilien)

Flammesikker/Forhjet sikkerhed (Kina) NEPSI Ex dmbia IIC T4/ Ex embia IIC T4/ DIP A20 TA, T90 oC IP67 Eksplosionssikker 2) (USA/Canada) FM/CSA Klasse I, Div. 1, Gruppe A, B, C, D Klasse II, Div. 1, Gruppe E, F, G Klasse III T4

Maritimt Lloyds Skibsregister ABS Typegodkendelse

Dansk

Anvendelser under tryk 3)


ADVARSLER:

Forsg aldrig at lsne, afmontere eller adskille procestilslutningen eller instrumenthuset, mens indholdet er under tryk. Brugeren er ansvarlig for at vlge sammenboltnings- og pakningsmaterialer, der ligger inden for flangens begrnsninger og dens ptnkte anvendelse, og som passer til anvendelsesforholdene. Ukorrekt installation kan medfre tab af procestryk.
Bemrk: Procestilslutningsskiltet skal forblive sammen med grnsefladen til procestrykket3). Sfremt instrumentpakken udskiftes, skal procestilslutningsskiltet overfres til den nye enhed. SITRANS LR250 enhederne er testet hydrostatisk og opfylder eller overskrider kravene i ASME Boiler and Pressure Vessel Code samt det europiske Trykudstyrsdirektiv.

1) 2) 3)

Se Elinstallation med forhjet sikkerhed p side 12. See Eksplosionssikker elinstallation (kun USA/Canada) p side 13. Grnsefladen til procestrykket omfatter de komponenter, der fungerer som en barriere mod tryktab fra procestanken: det vil sige kombinationen af procestilslutningsdelen og emitteren, men normalt eksklusive den elektriske indkapsling.

A5E31997170

SITRANS LR250 (HART) KVIKSTART MANUAL

Side DA-5

Trykudstyrsdirektivet, 97/23/EF
Siemens niveautransmittere med procesmontering af typen flange, gevind eller sanitr klemme omfatter ikke eget trykbrende hus og er derfor ikke underlagt Trykudstyrsdirektivet som tryk- og sikkerhedshjlpemidler (jf. EU-Kommissionens vejledning 1/8).

Installation
ADVARSLER: Installationen m kun udfres af kvalificeret personale og i overensstemmelse med de lokale bestemmelser. Hndtr altid apparatet ved at tage fat i instrumenthuset, aldrig procestilslutningsskiltet, for at undg beskadigelse. Pas srligt p ved hndtering af PVDF-antennen med gevind. Enhver beskadigelse af antennens overflade, navnlig p spidsen, kan pvirke ydelserne. Konstruktionsmaterialerne er valgt p basis af deres kemiske kompatibilitet (eller inerti) ved almindelig brug. Hvis de skal udsttes for srlige omgivelser, br foreneligheden kontrolleres i tabeller over kemiske kompatibiliteter inden installationen. Bemrkninger:

I den Europiske Union og medlemslandene heraf skal installationen foretages i henhold til ETSI EN 302372. For oplysninger om godkendelser henvises der til apparatets navneskilt. Det serienummer, der er prget i hver procestilslutningsdel, udgr et unikt identifikationsnummer med angivelse af fabrikationsdatoen. Eksempel: MMDDYY XXX (hvor MM = mned, DD = dag, YY = r og XXX= fortlbende nummer p den fremstillede enhed) Yderligere er angivet (hvis pladsen tillader det) flangekonfiguration, strrelse, trykkategori, materiale og materialets varmekode.

Dansk

Mundstykkets udformning
PVDF-antenne med gevind
min. afstand: 10 mm (0,4")

Hornantenne af rustfrit stl


min. afstand: 10 mm (0,4")

Min. diam. 51 mm (2")

Antennens ende skal rage mindst 10 mm (0,4) frem for at undg, at falske ekkoer tilbagekastes fra mundstykket. For PVDF-antennen med gevind er den mindste anbefalede diameter af mundstykket 51 mm (2"). Der fs en antenneforlngelse (100 mm/ 3,93") for alle versioner p nr PVDF-antennen med gevind.

Side DA-6

SITRANS LR250 (HART) KVIKSTART MANUAL

A5E31997170

Mundstykkets placering
P hje, smalle tanke skal centrale placeringer undgs, da de kan forrsage false ekkoer.

Omgivelser
Srg for, at omgivelserne svarer til instrumenthusets klassificering og konstruktionsmaterialer. Hvis apparatet skal monteres i direkte sollys, br det forsynes med en solafskrmning.
foretrukket unskelig

Strlevinkel
Strlevinklen er bredden af keglen, hvor energittheden er halvdelen af den maksimale energitthed. Den maksimale energitthed er lige ud for og i forlngelse af antennen. Der udsendes et signal uden for strlevinklen, hvilket kan give anledning til ekkoer fra falske ml.

emissionskegle strlevinkel: PVDF-antenne m. gevind = 19 1,5" horn = 19 2" horn = 15 = 10 strle 3" horn = 8 vinkel 4" horn

Emissionskegle
Emissionskeglen skal holdes fri for interferens fra stiger, rr, I-profiler eller pfyldningsstrmme.

Dansk

Adgang for programmering


Srg for en lettilgngelig adgang for aflsning af displayet og programmering ved hjlp af den hndholdte programmeringsenhed.

Montageanvisning
ADVARSEL: For anvendelser under tryk er det ndvendigt at bruge PTFE-tape eller et andet passende gevindttningsmiddel og at stramme procestilslutningen mere end med hndkraft. Det maksimale anbefalede stramningsmoment er 40 N.m. (30 ft.lbs.). Bemrk: Der er ingen begrnsning for antallet af gange, et apparat kan drejes uden at tage skade.

Versioner med gevind


1) 2) Inden apparatet indsttes i montagetilslutningen, skal det kontrolleres, at gevindene passer sammen, for at undg at beskadige dem. Apparatet skrues blot p procestilslutningen og strammes med hndkraft eller ved hjlp af en rrtang. For anvendelser under tryk, se Advarsel ovenfor. ADVARSEL: Brugeren er ansvarlig for at vlge sammenboltnings- og pakningsmaterialer, der ligger inden for flangens begrnsninger og dens ptnkte anvendelse, og som passer til anvendelsesforholdene.

Versioner med flange

A5E31997170

SITRANS LR250 (HART) KVIKSTART MANUAL

Side DA-7

Tilslutning
Strmforsyning
ADVARSLER: DC-indgangsklemmerne skal forsynes fra en kilde, der yder elektrisk isolation mellem ind- og udgang, for at opfylde de gldende sikkerhedskrav i henhold til IEC 61010-1. Alle felt-elinstallationer skal vre forsynet med en passende isolering i forhold til de nominelle spndinger.

Tilslutning af SITRANS LR250


ADVARSLER: Tjek godkendelsesniveauet p instrumentets navneplade og procesmrkeskiltet. Brug passende kabelrrsttninger for at bevare IP- eller NEMA-klassificeringen. Se Elektriske opstninger ved installationer i risikoomrder p side 9. Bemrk: Brug parsnoet kabel: trdtykkelse AWG 22 til 14 (0,34 mm2 til 2,5 mm2). Separate kabler og kabelrr kan vre ndvendige for at overholde standardmetoderne for elinstallation eller de elektriske normer.

Dansk

brug en 2 mm unbrakongle til at lsne den stopskrue, der lser dkslet. prop (IP 68) valgfri kabelafslutning1) 2) (eller indfring af NPT kabel2)) kabelafskrmning (eventuelt)

1) 2) 3) 4) 5)

Afisoler kablet p ca. 70 mm (2,75") fra enden af kablet og fr ledningerne gennem pakdsen. 1) 2) Forbind ledningerne med klemmen som vist: polariteten er angivet p klemrkken. Jordforbind apparatet i henhold til de lokale bestemmelser. Stram pakdsen for at opn en god ttning. Luk dkslet, fr apparatet programmeres og konfigureres.

1) 2)

Kan vre leveret sammen med apparatet. Hvis kablerne fres gennem kabelrr, brug kun muffer af en passende strrelse, der er godkendt til vandttte applikationer.

Side DA-8

SITRANS LR250 (HART) KVIKSTART MANUAL

A5E31997170

Tilslutning af HART
Typisk PLC/mA konfiguration med HART 1) 2)
Bemrkninger: Afhngig af systemets design aktiv PLC kan strmforsyningen vre adskilt fra PLC'en eller indbygget R= 250 i den. HART-modstanden2) skal HARTbegrnses alt efter det modem tilladelige driftsomrde1). SITRANS Der m ikke tilsluttes mere end LR250 HARTen HART-programmeringsenhed kommunikationsad gangen. enhed

strmforsyning

Elektriske opstninger ved installationer i risikoomrder


Der findes seks mulige elinstallationer i risikoomrder. Tjek i alle tilflde godkendelsesniveauet p instrumentets navneplade og procestilslutningsskiltet.

1. Egensikker elinstallation
Apparatets navneplade (ATEX/IECEX/INMETRO/C-TICK)

Dansk

ATEX-certifikatet, der er angivet p navnepladen, kan downloades fra produktsiden p vort website p: www.siemens.com/LR250. G til Support > Approvals (Godkendelser) / Certificates (Certifikater). IECEx-certifikatet, der er angivet p navnepladen, kan lses p IECEx-websitet. G til: http://iecex.iec.ch og klik p Ex Equipment Certificates of Conformity og indtast derefter certifikatnummeret IECEx SIR 05.0031X.

1) 2)

Total sljfemodstand = kabelmodstand plus 250 Ohm (resistor). Se enten Kurve 1: Almen brug, Egensikker, Ikke-gnistdannende/Energibegrnset/ Ikkeantndingsfarlig p side 23 eller Kurve 2: Flammesikker, Forhjet sikkerhed, Eksplosionssikker p side 24.

A5E31997170

SITRANS LR250 (HART) KVIKSTART MANUAL

Side DA-9

Egensikker elinstallation (fortsat)


Apparatets navneplade (FM/CSA)

Diagram over FM/CSA egensikker tilslutning nummer 23650653 kan downloades fra produktsiden p vort website: www.siemens.com/LR250. G til Support > Installation Drawings (Installationsdiagrammer) > Level Measurement (Niveaumling) > Continuous - Radar (Kontinuert - Radar). For strmbehov, se Kurve 1: Almen brug, Egensikker, Ikke-gnistdannende/ Energibegrnset/ Ikke-antndingsfarlig p side 23. Mht. kravene til elinstallationen: flg de lokale bestemmelser. Godkendte stv- og vandttte kabelrrsttninger er pkrvet for udendrs NEMA 4X / type 4X / NEMA 6, IP67, IP68 placeringer. Der henvises til Srlige anvisninger vedrrende installation i risikoomrder p side 13. Anbefalede egensikre barrierer er opgivet under Passive barrierer med shunt diode p side 10 og Aktive barrierer (gentagende barrierer) p side 11. Bemrkning: Valget af egnet PLC-inputmodul, strmforsyning eller barriere krver kendskab til egensikkerhed og applikationen. Det phviler installatren at srge for, at den egensikre installation overholder bde apparatets godkendelseskrav og de relevante nationale standarder.

Dansk

Passive barrierer med shunt diode


Bemrkning: Der krves en velreguleret forsyningsspnding. Fabrikant
MTL MTL Stahl Stahl

Delnummer
787SP+ (Dobbelt kanal) 7787P+ (Dobbelt kanal) 9001/01-280-100-10 (Enkelt kanal) 9002/01-280-110-10 (Dobbelt kanal)

For at sikre sig, at barrierens sikkerhedsbeskrivelse passer til inputparametrene for LR250 Egensikker, udfres flgende beregninger: Re-a = barrierens maksimale modstand fra ende til anden Rsljfe = sljfemodstand (summen af kabelmodstand plus eventuelt yderligere seriemodstand, for eksempel PLC-inputs og/eller displays) Vbarriere = vrdi af eventuelle ikke-linere spndingsfald forrsaget af barrieren 1) Bestem vrdien af Re-a ud fra databladet. 2) Beregn den totale vrdi af Rsljfe. 3) Beregn Rarbejde = Re-a + Rsljfe.

Hvordan vlges en passiv barriere for SITRANS LR250?

Side DA-10

SITRANS LR250 (HART) KVIKSTART MANUAL

A5E31997170

Hvordan vlges en passiv barriere (fortsat) 4) Bestem vrdien af Vbarriere ud fra barrierens datablad (for eksempel spndingsfald forrsaget af dioder). 5) Beregn Varbejde = Vforsyning Vbarriere. 6) Brug vrdierne af Varbejde og Rarbejde til at bekrfte, at driften ligger inden for det skraverede omrde p grafen Kurve 1: Almen brug, Egensikker, Ikke-gnistdannende/ Energibegrnset/ Ikke-antndingsfarlig p side 23.

Aktive barrierer (gentagende barrierer)


Fabrikant
MTL MTL Siemens Stahl MTL MTL

Delnummer
706 7206 SITRANS I 7NG4124 9001/51-280-110-14 E02009 E02010

2. Ikke-gnistdannende/Energibegrnset elinstallation

ATEX-certifikatet, der er angivet p navnepladen, kan downloades fra produktsiden p vort website p: www.siemens.com/LR250. G til: Support > Approvals (Godkendelser) / Certificates (Certifikater). For strmbehov, se Kurve 1: Almen brug, Egensikker, Ikke-gnistdannende/ Energibegrnset/ Ikke-antndingsfarlig p side 23. Mht. kravene til elinstallationen, flg de lokale bestemmelser.

Dansk

3. Ikke-antndingsfarlig elinstallation

Diagram over FM/CSA Klasse 1, Div 2 tilslutning nummer 23650673 kan downloades fra produktsiden p vores website p: www.siemens.com/LR250. G til Support > Installation Drawings (Installationsdiagrammer) > Level Measurement (Niveaumling)> Continuous - Radar (Kontinuert - Radar). For strmbehov, se Kurve 1: Almen brug, Egensikker, Ikke-gnistdannende/Energibegrnset/ Ikke-antndingsfarlig p side 23.

A5E31997170

SITRANS LR250 (HART) KVIKSTART MANUAL

Side DA-11

4. Flammesikker elinstallation

ATEX-certifikatet, der er angivet p navnepladen, kan downloades fra produktsiden p vores website p: www.siemens.com/LR250. G til: Support > Approvals / Certificates (Godkendelser / Certifikater). IECEx-certifikatet, der er angivet p navnepladen, kan lses p IECEx-websitet. G til: http://iecex.iec.ch og klik p Ex Equipment Certificates of Conformity og indtast derefter certifikatnummeret IECEx SIR 08.0107X. For strmbehov, se Kurve 2: Flammesikker, Forhjet sikkerhed, Eksplosionssikker p side 24. Mht. kravene til elinstallationen, flg de lokale bestemmelser. Se ogs Srlige anvisninger vedrrende installation i risikoomrder p side 13 og ovenstende ATEX-certifikat.

5. Elinstallation med forhjet sikkerhed

Dansk

ATEX-certifikatet, der er angivet p navnepladen, kan downloades fra produktsiden p vores website p: www.siemens.com/LR250. G til: Support > Approvals / Certificates (Godkendelser / Certifikater). IECEx-certifikatet, der er angivet p navnepladen, kan lses p IECEx-websitet. G til: http://iecex.iec.ch og klik p Ex Equipment Certificates of Conformity og indtast derefter certifikatnummeret IECEx SIR 08.0107X. For strmbehov, se Kurve 2: Flammesikker, Forhjet sikkerhed, Eksplosionssikker p side 24. Mht. kravene til elinstallationen, flg de lokale bestemmelser. Se ogs Srlige anvisninger vedrrende installation i risikoomrder p side 13 og ovenstende ATEX-certifikat.

Side DA-12

SITRANS LR250 (HART) KVIKSTART MANUAL

A5E31997170

6. Eksplosionssikker elinstallation (kun USA/Canada)

Diagram over FM/CSA eksplosionssikker tilslutning nummer A5E02257843 kan downloades fra produktsiden p vores website p: www.siemens.com/LR250. G til Support > Installation Drawings (Installationsdiagrammer) > Level Measurement (Niveaumling)> Continuous - Radar (Kontinuert - Radar). For strmbehov, se Kurve 2: Flammesikker, Forhjet sikkerhed, Eksplosionssikker p side 24.

Srlige anvisninger vedrrende installation i risikoomrder


(Reference: det europiske ATEX Direktiv 94/9/EF, Bilag II, 1/0/6)
Flgende anvisninger glder for udstyr, der er dkket af certifikat nummer SIRA 06ATEX2358X, SIRA 08ATEX1301X og SIRA 09ATEX4153X. Med hensyn til brug og montage henvises der til de generelle anvisninger. Dette udstyr er godkendt til brug som kategori 1GD udstyr for SIRA 06ATEX2358X, som kategori 1/2 GD, 1D, 2D udstyr for SIRA 08ATEX1301X og som kategori 3G udstyr for SIRA 09ATEX4153X. 3) Dette udstyr kan anvendes med brndbare gasser og dampe sammen med apparater af gruppe IIC, IIB og IIA og temperaturklasserne T1, T2, T3 og T4. 4) Dette udstyr har en indtrngningsbeskyttelsesgrad p IP67 og en temperaturklasse p T90 C og kan anvendes med antndeligt stv. 5) Dette udstyr er godkendt til brug i et interval af omgivende temperatur fra -40 C til +80 C. 6) Dette udstyr betragtes ikke som en sikkerhedsanordning (iflge direktiv 94/9/EF Bilag II, paragraf 1.5): se Trykudstyrsdirektivet, 97/23/EF p side 6. 7) Installationen og inspektionen af dette udstyr skal udfres af tilstrkkeligt uddannet personale i henhold til de gldende normer (EN 60079-14 og EN 60079-17 i Europa). 8) Dette udstyr kan ikke repareres. 9) Certifikatnumrene ender p "X", hvilket angiver, at der glder srlige betingelser for en sikker anvendelse. De personer, der skal installere eller inspicere dette udstyr, skal have adgang til certifikaterne. 10) Hvis det er sandsynligt, at udstyret vil komme i kontakt med aggressive stoffer, phviler det brugeren at trffe egnede foranstaltninger for at undg, at det pvirkes i skadelig retning, for sledes at sikre, at denne type beskyttelse ikke bringes i fare. Aggressive stoffer: for eksempel sure vsker eller gasser, der kan angribe metaller, eller oplsningsmidler, der kan pvirke polymere materialer. Egnede sikkerhedsforanstaltninger: for eksempel godtgrelse ud fra materialets datablad af, at det er modstandsdygtigt over for specifikke kemikalier. 1) 2)

Dansk

A5E31997170

SITRANS LR250 (HART) KVIKSTART MANUAL

Side DA-13

Programmering af SITRANS LR250


Aktivering af SITRANS LR250

Se Kvikstartguide via den hndholdte programmeringsenhed p side 18. Se Kvikstartguide via SIMATIC PDM p side 21.

Tnd for apparatet. SITRANS LR250 starter automatisk i Mlingsmode.

LCD-display Mlingsmode (normal drift) 1) 2)


1 2 3 4
(((

1 indikator for skift1) mellem linere enheder og % 2 den valgte drift: level (niveau), space (frirum) eller volume (volumen) 3 den mlte vrdi (niveau eller rumfang, frirum eller afstand) 4 enheder 5 sjlediagram, der angiver niveauet 6 sekundr zone, der p foresprgsel2) angiver temperaturen af elektronikken, ekkokonfidensen, sljfestrmmen eller afstanden 7 tekstzone, der viser statusmeddelelser 5 8 apparatets statusindikator (for detaljer henvises der til den komplette manual)

Indikatorer ved fejl S: 0 LOE


Nr der er en fejl til stede, vises fejlkoden og fejlmeddelelsen i tekstzonen (7), og et ikon for "service pkrvet" tndes i apparatets statuszone (8)

Dansk

Display i Programmode Navigationsskrm

aktuel p aktuelt En synlig menusjle er tegn p, at menu punkt menulisten er for lang til, at alle menuaktuelt punkterne kan vises. sjle punkt bnd Et bnd halvvejs nede af menusjlen angiver, at det aktuelle punkt er halvvejs nede af listen. Bredden og den relative position af bndet p menusjlen angiver menulistens lngde og det aktuelle punkts omtrentlige position i listen. Et bredere bnd er tegn p frre punkter.

nummer

Parameterskrm
parameternavn parameternummer parametervrdi/ valg

Redigeringsskrm

1) 2)

Tryk p OP eller NED pilen for at skifte mellem de to. Som svar p en foresprgsel ved tryk p en tast. For detaljer, se Funktion i Mlingsmode p side 15.

Side DA-14

SITRANS LR250 (HART) KVIKSTART MANUAL

A5E31997170

Hndholdt programmeringsenhed (Del nr. 7ML1930-1BK)


Bestilles separat. 1) 2) 3)

Tast

Funktion i Mlingsmode
Opdaterer sljfestrmmen1). Opdaterer aflsningen af den interne temperatur i huset1). Opdaterer vrdien af ekkokonfidensen1).

Opdaterer afstandsmlingen1). Mode bner PROGRAM mode2). HJRE pil bner PROGRAM mode3).

OP eller NED pilen skifter mellem linere enheder og %

Programmering vha. den hndholdte programmeringsenhed


Bemrkninger: Tryk p Mode for at skifte mellem Mlingsmode og Program mode. Mens apparatet er i PROGRAM mode, fastfryses outputtet og reagerer ikke p ndringer i apparatet. SITRANS LR250 vender automatisk tilbage til Mlingsmode efter en periode uden aktivitet i PROGRAM mode (mellem 15 sekunder og 10 minutter afhngig af menuniveauet).

Dansk

Parametermenuer

Parametrene er identificeret af deres navn og inddelt i 1. QUICK START 2. SETUP funktionsgrupper. 2.1. DEVICE For en fuldstndig liste over parametrene med ............. instruktioner henvises der til den komplette manual. 2.7. LINEARIZATION

2.7.1. VOLUME 2.7.1.1. VESSEL SHAPE

1) 2)

3)

Den nye vrdi vises i displayets sekundre zone. bner det senest viste menuniveau, medmindre strmmen har vret slet fra siden udgangen fra PROGRAM mode, eller der er forlbet mere end 10 minutter siden sidste anvendelse af PROGRAM mode. I s fald vises verste menuniveau. bner verste menuniveau.

A5E31997170

SITRANS LR250 (HART) KVIKSTART MANUAL

Side DA-15

1.

bning af PROGRAM mode.


Ret programmeringsenheden mod displayet (fra en afstand af max. 300 mm [1 ft]). HJRE pil aktiverer PROGRAM mode og bner menuniveau nr. 1. Mode bner det menuniveau, der senest er vist i PROGRAM mode inden for de seneste 10 minutter, eller menuniveau nr. 1, hvis strmmen har vret slukket siden da.

hndholdt programmeringsenhed display

max. 300 mm (1 ft)

2.

Navigation: tasternes funktioner i Navigationsmode


, og indtast menunummeret, for eksempel: 2.7.1 (Volumen).

Bemrkning: For hurtig adgang til parametrene via den hndholdte programmeringsenhed, tryk p Home

Tast

Navn
OP eller NED pil

Menuniveau
menu eller parameter menu parameter

Funktion i Navigationsmode
Rul til forrige eller nste menu eller parameter. G til frste parameter i den valgte menu eller bn nste menu. bn Redigeringsmode. bn forldremenuen. Skifter til MLINGSMODE. bner verste menuniveau: menu nr. 1.

HJRE pil

Dansk

VENSTRE pil
Tilstand Hjem

menu eller parameter menu eller parameter menu eller parameter

3.

Redigering i PROGRAM mode Valg af en valgmulighed i en liste


a) b) Naviger hen til den nskede parameter. Tryk p HJRE pil parameterskrmen. for at bne

parameternavn

parameternummer

aktuelt valg

c)

Tryk en gang til p HJRE pil

for at

bne Redigeringsmode. Det aktuelle valg fremhves. Rul hen til et nyt valg. d) Tryk p HJRE pil for at acceptere

vrdien. Displayet vender tilbage til parameterskrmen og viser det nye valg. Side DA-16 SITRANS LR250 (HART) KVIKSTART MANUAL A5E31997170

ndring af en numerisk vrdi


a) b) Naviger hen til den nskede parameter. Tryk en gang til p HJRE pil for at bne Redigeringsmode. Den aktuelle vrdi fremhves. Indtast en ny vrdi. Tryk p HJRE pil for at acceptere vrdien. Displayet vender tilbage til parameterskrmen og viser det nye valg.

parameternavn

parameternummer

c) d)

aktuel vrdi

Tasternes funktioner i Redigeringsmode Tast Navn


OP eller NED pil

Funktion i Redigeringsmode
Valg af valgmuligheder Talredigering Valg af valgmuligheder Ruller hen til punkt. - Inkrementerer eller dekrementerer tal. - Skifter mellem plus- og minustegn. - Accepterer data (skriver til parameteren). - Skifter fra Redigerings- til Navigationsmode. - Flytter cursoren n plads mod hjre - Eller med cursoren p tegnet Enter, accepterer data og skifter fra Redigerings- til Navigationsmode. Annullerer Redigeringsmode uden at ndre parameteren. - Flytter cursoren til plus-/minustegnet, hvis dette er den frste nedtrykkede tast - Eller flytter cursoren n plads mod venstre Sletter displayet. Indtaster et decimalpunktum.

HJRE pil
Talredigering Valg af valgmuligheder Talredigering Talredigering Talredigering

Dansk

VENSTRE pil

Slet Decimalpunktum Plus- eller minustegn


til

Talredigering

Skifter fortegnet p den indtastede vrdi.

Tal

Talredigering

Indtaster det tilsvarende tegn.

A5E31997170

SITRANS LR250 (HART) KVIKSTART MANUAL

Side DA-17

Kvikstartguide via den hndholdte programmeringsenhed


1. Kvikstart
a) Ret programmeringsenheden mod displayet (fra en afstand af max. 300 mm [1 ft]), og tryk p HJRE pil b) for at aktivere PROGRAM mode og bne menuniveau 1. for at navigere hen til menupunkt 1.1 og bne

Tryk to gange p HJRE pil parameterskrmen.

c)

Tryk p HJRE pil

for at bne Redigeringsmode eller NED pil

for at

acceptere standardvrdierne og g direkte til nste punkt. d) e) For at ndre en indstilling, rul hen til det nskede punkt eller indtast en ny vrdi. Efter at have ndret en vrdi, tryk p HJRE pil NED pil f) for at g til nste punkt. for at acceptere den og tryk p

Quick Start indstillingerne trder frst i kraft, nr der vlges Yes (Ja) til Apply changes (Udfr ndringer) i trin 1.8.

1.1. Sprog Vlger sproget, der anvendes p displayet, med jeblikkelig virkning.
Valgmuligheder
ENGLISH (ENGELSK), DEUTSCH (TYSK), FRANAIS (FRANSK), ESPAOL (SPANSK)

Dansk

1.2. Materiale Valgmuligheder


LIQUID (VSKE) LIQUID LOW DK (vske m. lav dielektricitetskonstant)

1.3. Svarhastighed Indstiller apparatets reaktionshastighed over for ndringer i mlingerne inden for mlomrdet.
Svarhastighed
SLOW (LANGSOM) MED (MIDDEL) FAST (HURTIG)

Fyldnings-/Tmningshastighed
0,1 m/min (0,32 ft/min) 1,0 m/min (3,28 ft/min) 10,0 m/min (32,8 ft/min)

Brug en indstilling, der lige netop er hurtigere end den maksimale fyldnings- eller tmningshastighed (den strste af disse).

1.4. Enheder Flerens mleenheder. Valgmuligheder


M, CM, MM, FT, IN

Side DA-18

SITRANS LR250 (HART) KVIKSTART MANUAL

A5E31997170

1.5. Driftsmode
flerens referencepunkt flerens referencepunkt

afstand frirum Hjt kalib.punkt

flerens referencepunkt

niveau Lavt kalib.punkt

Drift
NO SERVICE (IKKE I DRIFT) LEVEL (NIVEAU) DISTANCE (AFSTAND)
1)

Beskrivelse
Mlingerne og den tilhrende sljfestrm opdateres ikke, og apparatet slr over i fejlsikker drift1). Afstand til materialets overflade fra Lavt kalibreringspunkt. Afstand til materialets overflade i forhold til Flerens referencepunkt.

SPACE (FRIRUM) Afstand til materialets overflade fra Hjt kalibreringspunkt.

For yderligere oplysninger om fejlsikker drift henvises der til den komplette manual.

1.6. Lavt Kalibreringspunkt Afstand fra flerens reference til Lavt Kalibreringspunkt: normalt proces tom-niveau. Se Driftsmode (1.5.) for en illustration. Vrdier
Interval: fra 0,00 til 20,00 m

Dansk

1.7. Hjt Kalibreringspunkt Afstand fra flerens reference til Hjt Kalibreringspunkt: normalt proces fuld-niveau. Se Driftsmode (1.5.) for en illustration. Vrdier
Interval: fra 0,00 til 20,00 m

1.8. Apply? (Udfr ndringer) For at gemme Kvikstartindstillingerne er det ndvendigt at svare Yes for at udfre ndringerne. Valgmuligheder
Tryk p Mode YES (JA), NO (NEJ), DONE (FRDIG) (Displayet viser DONE, nr Kvikstart er afsluttet med succes).

for at vende tilbage til Mlingsmode. SITRANS LR250 er nu klar til drift.

Bemrk: Hvis den betragtede anvendelse vedrrer en tank med forhindringer, se venligst den komplette manual for oplysninger om brugen af Automatisk dmpning af falsk ekko.

A5E31997170

SITRANS LR250 (HART) KVIKSTART MANUAL

Side DA-19

SITRANS LR250 Kommunikation: HART

Det er ndvendigt at rde over den komplette manual for at have adgang til listen over mulige parametre. Vi anbefaler at bruge SIMATIC Process Device Manager (PDM) til at programmere apparatet. Brugsanvisninger for opstning af HART-apparater vha. SIMATIC PDM kan downloades p produktsiden p vort website p: www.siemens.com/LR250 under Support.

SIMATIC PDM
SIMATIC PDM er en softwarepakke, der bruges til at igangstte og vedligeholde SITRANS LR250 og andre procesapparater. Ls venligst betjeningsvejledningen eller online-hjlpen for oplysninger om brugen af SIMATIC PDM. (Flere oplysninger fs p www.siemens.com/simatic-pdm.)

Deaktivering af bufferne

Dette retter SIMATIC PDM ind efter HART-modemet for operativsystemerne Windows 2000 og Windows XP 1). Bemrkninger: SIMATIC PDM fungerer kun i Windows XP Professional version, ikke i Home-versionen. Der krves administrative rettigheder p operativsystemet for at deaktivere bufferne. 1) 2) 3) 4) 5) 6) Klik p Start/Indstillinger/Kontrolpanel for at begynde konfigurationen. Dobbeltklik p System, vlg fanen Hardware, og klik p knappen Enhedshndtering. bn mappen Porte og dobbeltklik p den COM-port, der bruges af systemet, for at bne vinduet Egenskaber for kommunikationsport. Vlg fanen Portindstillinger og dobbeltklik p knappen Avanceret. Hvis boksen Brug FIFO-buffere er valgt, klik for at fravlge den. Klik p OK for at lukke. Luk alle skrme og genstart computeren.

Dansk

Opdatering af Electronic Device Description (EDD - Elektronisk apparatbeskrivelse)


Bemrkninger:
SIMATIC PDM Rev. 5.2 SP1 understttes kun for grundlggende konfiguration og problemlsning. For avancerede funktioner, ssom Kvikstartguiden, krves Rev. 6.0 SP4 eller hjere. I SIMATIC PDM kan EDD findes i Apparatkataloget (Device Catalog) under Sensors/Level/ Echo/Siemens Milltronics/SITRANS LR250. Tjek produktsiden p vort website p: www.siemens.com/LR250 for at sikre Dem, at De har den nyeste version af SIMATIC PDM, den seneste Service Pack (SP) og den seneste hot fix (HF). G til Support > Software Downloads (Download af software). Download om ndvendig EDD-filen, gem filerne p computeren og pak den zippede fil ud til en lettilgngelig placering.

For at installere en ny EDD:


1) 2) 3) Download den ny EDD fra produktsiden p vort website p: www.siemens.com/LR250, og gem filerne p computeren. Pak den ZIP-komprimerede fil ud til en lettilgngelig placering. Start SIMATIC PDM Manager Device Catalog (SIMATIC PDM - Manager Apparatkatalog), find den udpakkede EDD fil og vlg den.
Windows er et registreret varemrke, der tilhrer Microsoft Corporation

1)

Side DA-20

SITRANS LR250 (HART) KVIKSTART MANUAL

A5E31997170

Konfiguration af et nyt apparat


Bemrkning: Hvis der klikkes p Cancel (Annullr) i lbet af en uploading fra apparatet til SIMATIC PDM, vil nogle af parametrene blive opdateret.
1) 2) 3) 4) 5) Tjek, at De rder over den seneste EDD, og opdater den om ndvendigt (seOpdatering af Electronic Device Description (EDD - Elektronisk apparatbeskrivelse) p side 20). Start SIMATIC Manager og opret et nyt projekt for LR250. Brugsanvisninger for opstning af HART-apparater vha. SIMATIC PDM kan downloades p produktsiden p vort website p: www.siemens.com/LR250. bn menuen Device Master Reset og klik p OK for at udfre et reset til fabriksindstillingerne. Nr reset er frdigt, uploades parametrene til PC/PGen. Konfigurer apparatet vha. Kvikstartguiden.

Kvikstartguide via SIMATIC PDM


Kvikstartguidens indstillinger er indbyrdes forbundet og ndres frst, nr der klikkes p FINISH AND DOWNLOAD (AFSLUT OG DOWNLOAD) efter trin 5 for at gemme indstillingerne offline og overfre dem til apparatet. Klik p BACK (Tilbage) for at vende tilbage og ndre en indstilling eller Cancel (Annuler) for at forlade Kvikstart. Start SIMATIC PDM, bn menuen Device Wizard - Quick Start (Apparat - Guide - Kvikstart) og flg trin 1 til 5.

Bemrkninger:

Dansk

Kvikstart

Betjening via FDT (Field Device Tool)

FDT er en standard, som anvendes i adskillige softwarepakker, der er beregnet til idriftsttelse og vedligeholdelse af feltapparater. To kommercielt tilgngelige FDT er PACTware og Fieldcare. For at konfigurere et feltapparat via FDT krves apparatets DTM (Device Type ManagerApparattypemanager). Siemens instruments bruger SITRANS DTM og en instrument-EDD, der er skrevet til SITRANS DTM. 1) Installer frst SITRANS DTM p Deres system. Det kan downloades fra: http://support.automation.siemens.com. Klik p Product Support (Produktsupport) og g til Product Information/Automation Technology/Sensor systems/Process Instrumentation/Software & Communications (Produktinformation/Automatiseringsteknologi/ Flersystemer/Procesinstrumetering/Software og Kommunikation). 2) Installer SITRANS LR250 HART EDD for SITRANS DTM. Den kan downloades fra produktsiden p vort website p: www.siemens.com/LR250. G til Support > Software Downloads (Download af software).

Konfiguration af et nyt apparat via FDT


En brugsanvisning kan downloades fra produktsiden p vort website under Support. A5E31997170 SITRANS LR250 (HART) KVIKSTART MANUAL Side DA-21

Betjening via AMS Device Manager


AMS Device Manager er en softwarepakke, der er beregnet til idriftsttelse og vedligeholdelse af feltapparater. Ls venligst betjeningsvejledningen eller online-hjlpen for oplysninger om brugen af AMS Device Manager. Yderligere information kan findes p: http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD - Elektronisk apparatbeskrivelse)


SITRANS LR250 krver en EDD for AMS Device Manager version 9.0.

Konfiguration af et nyt apparat via AMS Device Manager


1) Tjek produktsiden p vort website p: www.siemens.com/LR250 for at vre sikker p at have den nyeste EDD. G til Support > Software Downloads (Download af software) og download den om ndvendigt. Gem filerne p computeren og pak den zippede fil ud til en lettilgngelig placering. Start AMS Device Manager Add Device Type (Tilfj apparattype), g til den udpakkede EDD-fil og vlg den.

2)

Start AMS Device Manager. En brugervejledning i opstningen af HART-apparater med AMS Device Manager kan downloades fra produktsiden p vort website under Support.

Vedligeholdelse
SITRANS LR250 krver ingen vedligeholdelse eller rengring under normale driftsbetingelser. Sfremt rengring skulle blive ndvendig under hrde driftsbetingelser: 1) Kontrollr antennematerialet og procesmediet og vlg et rengringsmiddel, der ikke reagerer uhensigtsmssigt med det ene eller det andet. Tag apparatet ud af drift og tr antennen ren ved hjlp af en klud og et egnet rengringsmiddel.

Dansk

2)

Reparation af instrumentet og ansvarsbegrnsning


For yderligere oplysninger henvises der til indersiden af bagomslaget.

Udskiftning af antennen eller elektronikken/instrumenthuset


Hvis det bliver ndvendigt at udskifte antennen som flge af beskadigelse eller svigt, kan den udskiftes uden at behve forny kalibreringen. Udskiftning med en anden antennetype kan udfres af et autoriseret Siemens servicecenter eller autoriseret Siemens personale. Hvis elektronikken eller instrumenthuset skal udskiftes som flge af beskadigelse eller svigt, skal den korrekte antenneversion anvendes, ellers vil en ny kalibrering vre ndvendig. Denne skal udfres af autoriseret Siemens personale.

Side DA-22

SITRANS LR250 (HART) KVIKSTART MANUAL

A5E31997170

Diagram over maksimale procestemperaturer


LR250-versioner med flange-adapter
ADVARSEL: Den interne temperatur m ikke overskride 80 C (176 F). Bemrkninger: Diagrammet er kun vejledende og dkker ikke ethvert muligt procestilslutningsarrangement. (Det vil IKKE glde, hvis SITRANS LR250 monteres direkte p overfladen af en metallisk beholder). Diagrammet tager ikke hjde for varme fra direkte sollys. Parameter 3.2.1 Aktuel intern temperatur overvger den interne temperatur.

LR250-versioner med flange-adapter


Omgivelsestemperatur (oC) Maksimale procestemperaturer versus tilladelig omgivelsestemperatur

Procestemperatur Procestemperatur (oC)

Sljfeeffekt
Dansk

Krav til strmforsyningen


Bemrk: Nedenstende kurve glder for en uafhngig anordning, der er konfigureret vha. Siemens hndholdte programmeringsenhed. Kurve 1: Almen brug, Egensikker, Ikkegnistdannende/ Energibegrnset/ Ikke-antndingsfarlig

Sljfemodstand RL

TILLADELIGT DRIFTSOMRDE

Sljfespnding VL

Bemrk: Ved anvendelse af HART-kommunikation er minimumspndingen med 220 Ohm (RL) 16,3 VDC.

A5E31997170

SITRANS LR250 (HART) KVIKSTART MANUAL

Side DA-23

Kurve 2: Flammesikker, Forhjet sikkerhed, Eksplosionssikker Sljfemodstand RL

TILLADELIGT DRIFTSOMRDE

Sljfespnding VL

Bemrk: Ved anvendelse af HART-kommunikation er minimumspndingen med 220 Ohm (RL) 20,94 VDC.

Egenskaber under opstart


Apparatet bruger mindre end 3,6 mA under opstart. Ventetiden indtil frste mling er mindre end 50 sekunder

Dansk
Side DA-24

SITRANS LR250 (HART) KVIKSTART MANUAL

A5E31997170

SITRANS LR250 (HART) Kurzanleitung


Diese Betriebsanleitung ist eine Kurzfassung der wesentlichen Merkmale und Funktionen des SITRANS LR250 (HART 1)). Wir empfehlen, die ausfhrliche Version der Betriebsanleitung zu erwerben, damit Sie Ihr Gert optimal nutzen knnen. Die vollstndige Betriebsanleitung kann von der SITRANS LR250 Produktseite auf unserer Website heruntergeladen werden: www.siemens.com/LR250. Ein gebundenes Exemplar erhalten Sie bei Ihrer zustndigen Siemens Milltronics Niederlassung. Wenn Sie Fragen haben, wenden Sie sich bitte an: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Kanada, K9J 7B1 Email: techpubs.smpi@siemens.com

Copyright Siemens AG 2013. Alle Rechte vorbehalten


Wir fordern Benutzer auf, genehmigte, gebundene Betriebsanleitungen zu erwerben oder die von Siemens Milltronics Process Instruments entworfenen und genehmigten elektronischen Ausfhrungen zu lesen. Siemens Milltronics Process Instruments ist fr den Inhalt auszugsweiser oder vollstndiger Wiedergaben gebundener oder elektronischer Ausfhrungen nicht verantwortlich.

Haftungsausschluss
Wir haben den Inhalt der Druckschrift auf bereinstimmung mit dem beschriebenen Gert geprft. Dennoch knnen Abweichungen nicht ausgeschlossen werden, so dass wir fr die vollstndige bereinstimmung keine Gewhr bernehmen. Die Angaben in dieser Druckschrift werden regelmig berprft, notwendige Korrekturen sind in den nachfolgenden Auflagen enthalten. Fr Verbesserungsvorschlge sind wir dankbar. Technische nderungen vorbehalten.

MILLTRONICS ist ein eingetragenes Warenzeichen der Siemens Milltronics Process Instruments.

Technischer Support
Ein Support steht rund um die Uhr zur Verfgung. Um die Adresse, Telefon- und Faxnummer Ihrer rtlichen Niederlassung von Siemens Automation zu finden, siehe: www.siemens.com/automation/partner: Whlen Sie Deutsch als Sprache, klicken Sie auf Kontakt und whlen Sie Dienstleistungen; klicken Sie erneut auf Dienstleistungen, um Ihre Produktgruppe zu finden (+Automatisierungstechnik > +Sensoriksysteme > +Prozessinstrumentierung > +Fllstandmessung > +Kontinuierliche Verfahren). Whlen Sie Radar aus. Whlen Sie nun das Land aus, gefolgt von der Stadt/Region. Whlen Sie Technical Support unter Service. Fr technischen Support online, siehe: www.siemens.com/automation/support-request Whlen Sie Deutsch als Sprache und geben Sie den Produktnamen (SITRANS LR250) oder die Bestellnummer ein. Klicken Sie auf Suchen und treffen Sie die entsprechende Produktauswahl. Klicken Sie auf Weiter. Geben Sie ein Stichwort zur Beschreibung Ihres Problems ein. Suchen Sie entweder in den zugehrigen Dokumenten oder klicken Sie auf Weiter, um eine Beschreibung Ihres Problems an das Siemens Technical Support Team zu senden. Siemens IA/DT Technical Support Center: Tel. +49 (0)911 895 7222
1)

Deutsch

HARTist ein eingetragenes Warenzeichen von HART Communications Foundation.

A5E31997170

SITRANS LR250 (HART) KURZANLEITUNG

Seite DE-1

Sicherheitstechnische Hinweise

Warnhinweise mssen zu Ihrer persnlichen Sicherheit sowie der Sicherheit Dritter und zur Vermeidung von Sachschden beachtet werden. Zu jedem Warnhinweis wird der jeweilige Gefhrdungsgrad angegeben. Das Symbol WARNUNG bezieht sich auf ein Warnsymbol auf dem Produkt und bedeutet, dass bei Nicht-Einhalt der entsprechenden Vorsichtsmanahmen Tod, schwere Krperverletzung und/oder erheblicher Sachschaden eintreten knnen. Dieses Symbol WARNUNG wird verwendet, wenn sich kein entsprechendes Vorsichtssymbol auf dem Produkt befindet, und bedeutet, dass bei Nicht-Einhalt der entsprechenden Vorsichtsmanahmen Tod, schwere Krperverletzung und/ oder erheblicher Sachschaden eintreten knnen. Hinweis: steht fr eine wichtige Information ber das Produkt selbst oder den Teil der Betriebsanleitung, auf den besonders aufmerksam gemacht werden soll.

FCC-Konformitt
Nur fr Installationen in den USA: Richtlinien der FCC (Federal Communications Commission)
WARNUNG: nderungen oder Modifizierungen, die nicht ausdrcklich von Siemens Milltronics genehmigt wurden, knnen zum Entzug der Betriebsgenehmigung des Benutzers fhren. Hinweise: Dieses Gert wurde getestet und mit den fr ein Digitalgert der Class A geltenden Grenzwerten, gem Teil 15 der FCC-Bestimmungen, fr konform erklrt. Diese Grenzwerte sollen einen angemessenen Schutz vor Strungen gewhrleisten, wenn das Gert in einem gewerblichen Umfeld betrieben wird. Das Gert erzeugt und verwendet Funkfrequenzen und kann sie ausstrahlen. Wenn es nicht gem der Betriebsanleitung installiert und betrieben wird, knnen Funkstrungen auftreten. Der Betrieb des Gertes in einem Wohnbereich kann Strungen verursachen. In diesem Fall ist der Benutzer angehalten, die Strung auf eigene Kosten zu beheben.

SITRANS LR250
Deutsch
WARNUNG: Betriebssicherheit und Schutz des SITRANS LR250 sind nur gewhrleistet, wenn das Gert entsprechend dieser Betriebsanleitung betrieben wird. Hinweis: Dieses Produkt ist fr den Einsatz im Industriebereich vorgesehen. Bei Verwendung in Wohngebieten kann es zu Strungen von verschiedenen Funkanwendungen kommen. SITRANS LR250 ist ein 25 GHz Pulsradar-Fllstandmessumformer in Zweileiter-Technik fr Messbereiche bis 20 m (66 ft). Er eignet sich fr die kontinuierliche berwachung von Flssigkeiten und Schlmmen in Lagerbehltern mit hohen Temperaturen und Druck. Er ist ideal fr kleine Behlter und Medien mit niedriger Dielektrizittszahl. Das Gert besteht aus einer Auswerteelektronik, die mit einer Hornantenne und einem Gewinde- bzw. Flansch-Prozessanschluss verbunden ist. Eine PVDF-Antenne mit Gewindeanschluss steht ebenfalls zur Verfgung. SITRANS LR250 untersttzt das HART-Kommunikationsprotokoll und die SIMATIC PDM-Software. Zur Signalverarbeitung wird die Process Intelligence-Software verwendet.

Seite DE-2

SITRANS LR250 (HART) KURZANLEITUNG

A5E31997170

Technische Daten
Eine vollstndige Liste finden Sie in der ausfhrlichen Betriebsanleitung des SITRANS LR250 (HART). Informationen zu den Zulassungen finden Sie unter Zulassungen auf Seite 4.

Umgebungs-/Betriebstemperatur
Hinweise:
Die maximale Temperatur hngt von Prozessanschluss, Antennenwerkstoff und Behlterdruck ab; siehe Kurve der maximalen Prozesstemperatur auf Seite 23. Nhere Angaben finden Sie in der Betriebsanleitung, unter Prozessdruck-/Temperaturkurven. Prozesstemperatur und Druckwerte sind von den Angaben auf dem TAG-Schild am Prozessanschluss abhngig. Die auf dem TAG-Schild angegebene Bezugszeichnung ist auf unserer Website verfgbar: www.siemens.com/LR250, unter Support/Installationszeichnungen/Fllstandmessung/ Kontinuierliche Verfahren - Radar/LR250. Zustzliche Informationen zu Prozessanschlssen finden Sie auf der Seite Installationszeichnungen unter Prozessanschlusszeichnungen. Die Signalamplitude steigt mit dem Horndurchmesser; verwenden Sie daher den grtmglichen Durchmesser. Optionale Verlngerungen knnen unterhalb der Gewindeanschlsse installiert werden. Weitere Angaben finden Sie unter Kurve der maximalen Prozesstemperatur auf Seite 23.
Umgebungstemperatur (um das Gehuse) 40 C 80 C (40 F 176 F) Typenschild

Prozesstemperatur am Prozessanschluss: Standard Hornantenne: - mit FKM-O-Ring: 40 oC ... +200 oC (40 oF ... +392 oF) - mit FFKM-O-Ring: 20 oC ... +200 oC (4 oF ... +392 oF) 2" NPT / BSPT / G PVDF-Antenne mit Gewinde: 40 ... +80 C (40 ... +176 F)

TAG-Schild am Prozessanschluss

Hilfsenergie
Allgemeine Verwendung Eigensicher Nicht funkend (Non Sparking)/Energiebegrenzt (Energy Limited) Non-incendive (Betriebsmittel fr Zone 2) (nur FM/CSA USA/Kanada) Druckfeste Kapselung (Flameproof) Erhhte Sicherheit Druckfeste Kapselung (Explosion proof) (nur FM/CSA USA/Kanada) Maximal DC 30 V 4 ... 20 mA Max. Anlaufstrom:

Deutsch

Nominal DC 24 V bei max. 550 Ohm

Nominal DC 24 V bei max. 250 Ohm

siehe Anlaufverhalten auf Seite 24

A5E31997170

SITRANS LR250 (HART) KURZANLEITUNG

Seite DE-3

Zulassungen
Hinweis: Das Typenschild des Gertes gibt die fr Ihr Gert gltigen Zulassungen an. Allgemein CSA US/C, FM, CE, C-TICK, KC Funk Europa (R&TTE), FCC, Industry Canada Ex-Bereiche Eigensicher 1) (Europa) ATEX II 1G, EEx ia IIC T4 ATEX II 1D, EEx tD A20 IP67 T90 C (International) IECEx SIR 05.0031X, Ex ia IIC T4, Ex tD A20 IP67 T90 C (USA/Kanada) FM/CSA Class I, Div. 1, Gruppen A, B, C, D Class II, Div. 1, Gruppen E, F, G Class III T4 (Brasilien) INMETRO: DNV 12.0087 X Ex ia IIC T4 Ga Ex ia ta IIIC T90 C Da IP65/IP67 -40 C Ta +80 C DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

(China)

NEPSI Ex ia IIC T4/ DIP A20 TA, T90 C IP67

Nicht funkend (Non Sparking)/Energiebegrenzt (Energy Limited) 2) (Europa) ATEX II 3 G Ex nA/nL IIC T4 Gc (China) NEPSI Ex nA II T4/ Ex nL IIC T4 Betriebsmittel fr Zone 2 (Non-incendive) 3) (USA/Kanada) FM/CSA Class I, Div. 2, Gruppen A, B, C, D T5 Druckfeste Kapselung (Flame proof) 4) (Europa/International) ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex dmbia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C (Brasilien) INMETRO: DNV 12.0088 X Ex d ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017

Deutsch

ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-1:2009, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

1) 2) 3) 4)

Siehe Anschlussmethode Eigensicher auf Seite 9. Siehe Anschlussmethode Nicht funkend (Non Sparking)/Energiebegrenzt (Energy Limited) auf Seite 11. Siehe Anschlussmethode Betriebsmittel fr Zone 2 (Non-incendive) auf Seite 11. Siehe Anschlussmethode Druckfeste Kapselung (Flameproof) auf Seite 12.

Seite DE-4

SITRANS LR250 (HART) KURZANLEITUNG

A5E31997170

Ex-Bereiche (Fortsetzung) Erhhte Sicherheit 1) (Europa/International)

(Brasilien)

ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex embia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C INMETRO: DNV 12.0088 X Ex e ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017

Druckfeste Kapselung (Flame proof)/Erhhte Sicherheit (China) NEPSI Ex dmbia IIC T4/ Ex embia IIC T4/ DIP A20 TA, T90 oC IP67 Druckfeste Kapselung (Explosion Proof) 2) (USA/Kanada) FM/CSA Class I, Div. 1, Gruppen A, B, C, D Class II, Div. 1, Gruppen E, F, G Class III T4 Schiffbau Lloyds Register of Shipping ABS-Schiffbauzulassung

ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-7:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

Einsatzbereich Druck
WARNUNGEN:

Versuchen Sie niemals, die Prozessdichtung oder das Gehuse zu lockern, zu entfernen oder auseinanderzubauen, whrend der Inhalt des Behlters unter Druck steht. Der Benutzer ist fr die Auswahl von Schraub- und Dichtungsmaterial verantwortlich. Dieses muss den fr den Flansch aufgestellten Bedingungen und dessen bestimmter Verwendung entsprechen und fr die Betriebsbedingungen geeignet sein. Eine unsachgeme Installation kann zu Druckverlust im Prozess fhren.
Hinweise: Das TAG-Schild am Prozessanschluss muss mit der Druckbaugruppe1) verbunden bleiben. Falls das Gertegehuse ausgetauscht werden soll, muss das TAG-Schild am Prozessanschluss auf das Austauschgert bertragen werden. Jedes SITRANS LR250 Gert ist hydrostatisch getestet. Es erfllt oder bertrifft die Anforderungen des ASME Boiler and Pressure Vessel Code und der Europischen Druckgerterichtlinie.

Deutsch

1) 2)

Siehe Anschlussmethode Erhhte Sicherheit auf Seite 12. Siehe Anschlussmethode Druckfeste Kapselung (Explosion-proof) (nur USA/Kanada) auf Seite 13.

A5E31997170

SITRANS LR250 (HART) KURZANLEITUNG

Seite DE-5

Druckgerterichtlinie, DGRL, 97/23/EG


Siemens Fllstandmessgerte mit Flansch-, Gewinde- oder Sanitr-Clamp-Prozessanschluss sind mit keinem eigenen, drucktragenden Gehuse ausgestattet und fallen daher nicht als druckhaltendes Ausrstungsteil oder Sicherheitsvorrichtung unter die Druckgerterichtlinie (siehe Richtlinie der EU-Kommission 1/8). 1)

Installation
WARNUNGEN:
Die Installation darf nur durch qualifiziertes Personal und unter Beachtung der rtlichen, gesetzlichen Bestimmungen durchgefhrt werden. Um eine Beschdigung zu vermeiden, halten Sie das Gert nur am Gehuse, und nicht am TAG-Schild des Prozessanschlusses. Die PVDF-Antenne mit Gewindeanschluss erfordert besondere Vorsicht. Jegliche Beschdigung der Antennenoberflche, insbesondere an der Spitze, knnte die Leistung beeintrchtigen. Die Werkstoffe werden entsprechend ihrer chemischen Bestndigkeit (oder Trgheit) fr allgemeine Zwecke gewhlt. Bei Exposition gegenber besonderen Umgebungen ist vor dem Einbau die Tabelle zur chemischen Vertrglichkeit zu prfen.

Hinweise:
Fr Lnder der Europischen Union und Mitgliedslnder muss die Installation in bereinstimmung mit ETSI EN 302372 erfolgen. Das Typenschild des Gerts liefert Angaben zu den Zulassungen. Die Seriennummer ist in jeden Prozessanschluss geprgt und liefert eine eindeutige Kennnummer mit Angabe des Herstellungsdatums. Beispiel: MMTTJJ XXX (mit MM = Monat, TT = Tag, JJ = Jahr und XXX= Laufnummer des Gerts) Weitere Kennzeichnungen (wenn Platz vorhanden) stehen fr Flanschkonfiguration, Gre, Druck-Klasse, Werkstoff und Hitzecode des Werkstoffs.

Design des Montagestutzens


PVDF-Antenne mit Gewinde Edelstahl-Hornantenne
Min. Abstand: 10 mm (0.4") Min. Abstand: 10 mm (0.4")

Deutsch

Min. Durchm. 51 mm (2")

Das Ende der Antenne muss mindestens 10 mm (0,4") in den Tank hineinragen, um Strechos, die vom Stutzen reflektiert werden, zu vermeiden. Empfohlener Mindestdurchmesser des Stutzenz fr die PVDF-Antenne mit Gewinde: 51 mm (2"). Eine Antennenverlngerung (100 mm/3.93") ist fr alle Ausfhrungen auer der PVDFAntenne mit Gewinde verfgbar.

1)

Die Druckbaugruppe besteht aus allen Bauteilen, die vor einem Druckverlust aus dem Prozessbehlter schtzen, d. h. der Prozessanschluss kombiniert mit dem Sender, normalerweise unter Ausschluss des Elektronikgehuses.

Seite DE-6

SITRANS LR250 (HART) KURZANLEITUNG

A5E31997170

Montageort
Vermeiden Sie die mittige Montage auf hohen, schmalen Behltern. Dies kann Strechos erzeugen.

Umgebung
Achten Sie darauf, dass die Umgebungsbedingungen mit den Nennwerten des Gehuses und dem Werkstoff kompatibel sind. Verwenden Sie ein Sonnenschutzdach, wenn das Gert direkter Sonneneinstrahlung ausgesetzt ist.
optimal nicht empfohlen

ffnungswinkel
Der ffnungswinkel entspricht der Kegelweite, wo die Energiedichte halb so gro ist wie ihr hchster Wert. Am grten ist die Energiedichte direkt vor und in einer Linie mit der Antenne. Durch eine Signalbertragung auerhalb des ffnungswinkels knnen Hindernisse erfasst und dadurch Strechos erzeugt werden.

Radarkegel ffnungswinkel: PVDF-Antenne mit Gewinde = 19 1.5" Horn = 19 2" Horn = 15 ffnungs- 3" Horn = 10 winkel 4" Horn = 8

Radarkegel
Der Radarkegel muss ohne Strung durch Leitern, Rohre, Doppel-T-Trger oder Befllstrme die Oberflche des Messstoffs erreichen knnen.

Zugang fr die Programmierung


Sorgen Sie fr einen einfachen Zugang zum Ablesen der Anzeige und zur Programmierung ber das Handprogrammiergert.

Montageanweisungen
WARNUNG: Fr Druckbehlter verwenden Sie PTFE-Band (oder ein anderes, geeignetes Dichtungsmaterial); der Prozessanschluss muss mehr als handfest angezogen werden. (Die maximal empfohlene Drehkraft betrgt 40 N-m (30 ft-lbs).) Hinweis: Das Gert kann beliebig oft gedreht werden, ohne beschdigt zu werden.

Deutsch

Gewindeausfhrungen
1) 2) Vor Einbau des Gertes ist es wichtig, die bereinstimmung der Gewinde zu berprfen, um eine Beschdigung zu vermeiden. Schrauben Sie das Gert in den Prozessanschluss und ziehen Sie es von Hand oder mit einem Schlssel an. Fr den Einsatzbereich Druck, siehe Warnungshinweis oben. WARNUNG: Der Benutzer ist fr die Auswahl von Schraub- und Dichtungsmaterial verantwortlich. Dieses muss den fr den Flansch aufgestellten Bedingungen und dessen bestimmter Verwendung entsprechen und fr die Betriebsbedingungen geeignet sein.

Flanschausfhrungen

A5E31997170

SITRANS LR250 (HART) KURZANLEITUNG

Seite DE-7

Anschluss
Spannungsversorgung
WARNUNGEN: Um die Sicherheitsanforderungen der IEC 61010-1 zu erfllen, sind die Gleichstrom-Eingangsklemmen von einer Spannungsquelle zu versorgen, die ber eine galvanische Trennung zwischen Ein- und Ausgang verfgt. Alle Feldanschlsse mssen entsprechend der angelegten Spannung isoliert sein.

Anschluss des SITRANS LR250


WARNUNGEN: berprfen Sie die Zulassungen auf dem Typenschild des Gertes. Verwenden Sie geeignete Conduit- und Kabelverschraubungen, um die Schutzart IP oder NEMA zu gewhrleisten. Siehe Anschlussmethoden bei Einbau in explosionsgefhrdeten Bereichen auf Seite 9. Hinweise: Verwenden Sie Kabel mit verdrilltem Aderpaar: Leitungsquerschnitt 0,34 mm2 ... 2,5 mm2 (AWG 22 ... 14). Eine getrennte Leitungsverlegung kann erforderlich sein, um Standardanforderungen an den Anschluss oder elektrische Richtlinien zu erfllen.
Die Stellschraube des Deckels mit einem 2 mm Inbusschlssel lockern. Blindverschraubung (IP 68) optionale Kabelverschraubung1) 2) (oder NPT-Kabeleinfhrung2)) Kabelabschirmung (falls verwendet)

Deutsch

1) 2) 3) 4) 5)

Isolieren Sie den Kabelmantel ca. 70 mm (2,75") vom Ende des Kabels ab und fhren Sie die Adern durch die Kabelverschraubung ein. 1) 2) Anschluss der Drhte wie abgebildet: Die Polaritt ist auf den Klemmen gekennzeichnet. Nehmen Sie die Erdung des Gerts entsprechend lokaler Richtlinien vor. Ziehen Sie die Verschraubung so an, dass sie gut abdichtet. Schlieen Sie den Deckel vor der Programmierung und Gertekonfiguration.

1) 2)

Kann im Lieferumfang enthalten sein. Bei einer Kabeleinfhrung ber Schutzrohr verwenden Sie nur zugelassene Rohrverschraubungen geeigneter Gre fr wassergeschtzte Applikationen.

Seite DE-8

SITRANS LR250 (HART) KURZANLEITUNG

A5E31997170

Anschluss HART
Typische SPS/mA-Konfiguration mit HART 1) 2)
aktive SPS Spannungsversorgung R= 250 HARTModem SITRANS LR250 HARTCommunicator

Hinweise: Je nach Systemdesign hat die SPS eine separate oder integrierte Spannungsversorgung. HART-Widerstand1) muss entsprechend dem zulssigen Betriebsbereich begrenzt sein2). Schlieen Sie nicht mehr als jeweils ein HART-Programmiergert an.

Anschlussmethoden bei Einbau in explosionsgefhrdeten Bereichen


Fr Installationen in Ex-Bereichen stehen sechs Anschlussoptionen zur Auswahl. berprfen Sie in allen Fllen die Zulassungen auf dem Typen- und TAG-Schild am Prozessanschluss Ihres Gertes.

1. Anschlussmethode Eigensicher
Typenschild des Gerts (ATEX/IECEX/INMETRO/C-TICK)

Das auf dem Typenschild aufgefhrte ATEX-Zertifikat steht auf der Produktseite unserer Website zum Download zur Verfgung: www.siemens.com/LR250. Whlen Sie Deutsch als Sprache und siehe Support > Approbationen / Prfbescheinigungen / Zertifikate. Das auf dem Typenschild aufgefhrte IECEx-Zertifikat finden Sie auf der IECEx-Website. Siehe: http://iecex.iec.ch, klicken Sie auf Ex Equipment Certificates of Conformity und geben die Zertifikatsnummer IECEx SIR 05.0031X ein.

Deutsch

1) 2)

Gesamtschleifenwiderstand = Leitungswiderstand plus 250 Ohm (Widerstand). Siehe Kurve 1: Allg. Verwendung, Eigensicher, Nicht funkend (Non Sparking)/Energiebegrenzt (Energy Limited)/Betriebsmittel fr Zone 2 (Non-incendive) auf Seite 23 oder Kurve 2: Druckfeste Kapselung (Flameproof), Erhhte Sicherheit, Druckfeste Kapselung (Explosion-proof) auf Seite 24.

A5E31997170

SITRANS LR250 (HART) KURZANLEITUNG

Seite DE-9

Anschlussmethode Eigensicher (Fortsetz.)


Typenschild des Gerts (FM/CSA)

Die Anschlusszeichnung FM/CSA Eigensicher Nr. 23650653 steht auf der Produktseite unserer Website zum Download zur Verfgung: www.siemens.com/LR250. Gehen Sie zu Support > Installationszeichnungen > Fllstandmessung > Kontinuierliche Verfahren - Radar. Angaben zum Strombedarf finden Sie unter Kurve 1: Allg. Verwendung, Eigensicher, Nicht

funkend (Non Sparking)/Energiebegrenzt (Energy Limited)/Betriebsmittel fr Zone 2 (Nonincendive) auf Seite 23.

Anschlussanforderungen: Beachten Sie lokale Vorschriften. Im Auenbereich IP67, IP68, NEMA 4X / Type 4X / NEMA 6 sind zugelassene staub- und wasserdichte Conduit-Verschlsse erforderlich. Siehe Vorschriften bezglich Installationen in Ex-Bereichen auf Seite 13. Empfohlene eigensichere Zenerbarrieren sind aufgefhrt unter Passive Barrieren mit Shunt-Dioden auf Seite 10 und Aktive Barrieren auf Seite 11. Hinweis: Die Auswahl eines geeigneten SPS-Eingangsmoduls (Spannungsversorgung) oder einer Barriere setzt Kenntnisse ber die Eigensicherheit und die Applikation voraus. Die Verantwortung fr die bereinstimmung der eigensicheren Installation mit den Zulassungsanforderungen des Betriebsmittels und den relevanten, nationalen Verfahrensregeln liegt beim Installateur.

Passive Barrieren mit Shunt-Dioden


Hinweis: Eine gut geregelte Versorgungsspannung ist erforderlich. Hersteller
MTL MTL Stahl Stahl

Artikelnummer
787SP+ (zweikanalig) 7787P+ (zweikanalig) 9001/01-280-100-10 (einkanalig) 9002/01-280-110-10 (zweikanalig)

Deutsch

Auswahl einer passiven Barriere fr SITRANS LR250

Um sicherzustellen, dass die Sicherheitsbeschreibung der Barriere fr die Eingangsparameter Eigensicherheit (IS) des LR250 geeignet ist, fhren Sie folgende Rechnung durch: Re-e = max. durchgehender Widerstand der Barriere R Schleife = Schleifenwiderstand (Gesamtleitungswiderstand plus zustzliche Vorwiderstnde, z. B. SPS-Eingnge und/oder -Anzeigen) V Barriere = Wert aller unlinearen Spannungsabflle aufgrund der Barriere 1) Bestimmen Sie den Wert fr Re-e anhand des Datenblattes. 2) Berechnen Sie den Gesamtwert fr R Schleife. 3) Berechnen Sie RBetrieb = Re-e + R Schleife.

Seite DE-10

SITRANS LR250 (HART) KURZANLEITUNG

A5E31997170

Auswahl einer passiven Barriere (Fortsetzung) 4) Bestimmen Sie den Wert fr V Barriere anhand des Datenblattes (z. B. Spannungsabflle durch Dioden). 5) Berechnen Sie V Betrieb = V Versorgung V Barriere. 6) Unter Verwendung der Werte von V Betrieb und R Betrieb ist zu besttigen, dass der Betrieb innerhalb des schattierten Bereichs der Kurve 1: Allg. Verwendung, Eigensicher,

Nicht funkend (Non Sparking)/Energiebegrenzt (Energy Limited)/Betriebsmittel fr Zone 2 (Non-incendive) auf Seite 23 liegt.
Artikelnummer
706 7206 SITRANS I 7NG4124 9001/51-280-110-14 E02009 E02010

Aktive Barrieren
Hersteller
MTL MTL Siemens Stahl MTL MTL

2. Anschlussmethode Nicht funkend (Non Sparking)/ Energiebegrenzt (Energy Limited)

Das auf dem Typenschild aufgefhrte ATEX-Zertifikat steht auf der Produktseite unserer Website zum Download zur Verfgung: www.siemens.com/LR250. Siehe: Support > Approbationen / Prfbescheinigungen / Zertifikate. Angaben zum Strombedarf finden Sie unter Kurve 1: Allg. Verwendung, Eigensicher, Nicht Fr Anschlussanforderungen beachten Sie lokale Richtlinien.

funkend (Non Sparking)/Energiebegrenzt (Energy Limited)/Betriebsmittel fr Zone 2 (Nonincendive) auf Seite 23.

Deutsch

3. Anschlussmethode Betriebsmittel fr Zone 2 (Non-incendive)

Die Anschlusszeichnung FM/CSA Class 1, Div 2, Nr. 23650673 steht auf der Produktseite unserer Website zum Download zur Verfgung: www.siemens.com/LR250. Gehen Sie zu Support > Installationszeichnungen > Fllstandmessung > Kontinuierliche Verfahren - Radar. Angaben zum Strombedarf finden Sie unter Kurve 1: Allg. Verwendung, Eigensicher, Nicht

funkend (Non Sparking)/Energiebegrenzt (Energy Limited)/Betriebsmittel fr Zone 2 (Nonincendive) auf Seite 23.
SITRANS LR250 (HART) KURZANLEITUNG

A5E31997170

Seite DE-11

4. Anschlussmethode Druckfeste Kapselung (Flameproof)

Das auf dem Typenschild aufgefhrte ATEX-Zertifikat steht auf der Produktseite unserer Website zum Download zur Verfgung: www.siemens.com/LR250. Siehe: Support > Approbationen / Prfbescheinigungen / Zertifikate. Das auf dem Typenschild aufgefhrte IECEx-Zertifikat finden Sie auf der IECEx-Website. Siehe: http://iecex.iec.ch, klicken Sie auf Ex Equipment Certificates of Conformity und geben die Zertifikatsnummer IECEx SIR 08.0107X ein. Angaben zum Strombedarf finden Sie unter Kurve 2: Druckfeste Kapselung (Flameproof), Erhhte Sicherheit, Druckfeste Kapselung (Explosion-proof) auf Seite 24. Fr Anschlussanforderungen beachten Sie lokale Richtlinien. Siehe auch Vorschriften bezglich Installationen in Ex-Bereichen auf Seite 13 und das oben aufgefhrte ATEX-Zertifikat.

5. Anschlussmethode Erhhte Sicherheit

Das auf dem Typenschild aufgefhrte ATEX-Zertifikat steht auf der Produktseite unserer Website zum Download zur Verfgung: www.siemens.com/LR250. Siehe: Support > Approbationen / Prfbescheinigungen / Zertifikate. Das auf dem Typenschild aufgefhrte IECEx-Zertifikat finden Sie auf der IECEx-Website. Siehe: http://iecex.iec.ch, klicken Sie auf Ex Equipment Certificates of Conformity und geben die Zertifikatsnummer IECEx SIR 08.0107X ein.

Deutsch

Angaben zum Strombedarf finden Sie unter Kurve 2: Druckfeste Kapselung (Flameproof), Erhhte Sicherheit, Druckfeste Kapselung (Explosion-proof) auf Seite 24. Fr Anschlussanforderungen beachten Sie lokale Richtlinien. Siehe auch Vorschriften bezglich Installationen in Ex-Bereichen auf Seite 13 und das oben aufgefhrte ATEX-Zertifikat.

Seite DE-12

SITRANS LR250 (HART) KURZANLEITUNG

A5E31997170

6. Anschlussmethode Druckfeste Kapselung (Explosion-proof) (nur USA/Kanada)

Die Anschlusszeichnung FM/CSA Explosion Proof Nr. A5E02257843 steht auf der Produktseite unserer Website zum Download zur Verfgung: www.siemens.com/LR250. Gehen Sie zu Support > Installationszeichnungen > Fllstandmessung > Kontinuierliche Verfahren - Radar. Angaben zum Strombedarf finden Sie unter Kurve 2: Druckfeste Kapselung (Flameproof), Erhhte Sicherheit, Druckfeste Kapselung (Explosion-proof) auf Seite 24.

Vorschriften bezglich Installationen in Ex-Bereichen


(Europische ATEX-Richtlinie 94/9/EG, Anhang II, 1/0/6)
Folgende Vorschriften finden Anwendung auf die Gerte, die Gegenstand des Zertifikats Nr. SIRA 06ATEX2358X, SIRA 08ATEX1301X und SIRA 09ATEX4153X sind. 1) 2) Angaben zu Verwendung und Zusammenbau finden Sie im Hauptteil der Vorschriften. Das Gert ist gem SIRA 06ATEX2358X fr den Einsatz als Betriebsmittel der Kategorie 1GD zertifiziert, gem SIRA 08ATEX1301X als Kategorie 1/2 GD, 1D, 2D und gem SIRA 09ATEX4153X als Kategorie 3G. 3) Das Gert kann mit brennbaren Gasen und Dmpfen mit Betriebsmitteln der Gruppen IIA, IIB und IIC und Temperaturklasse T1, T2, T3 und T4 eingesetzt werden. 4) Das Gert besitzt die Schutzart IP67 und eine Temperaturklasse von T90 C. Es kann mit brennbaren Stuben eingesetzt werden. 5) Das Gert ist fr einen Einsatz bei Umgebungstemperaturen von 40C bis +80 C zugelassen. 6) Das Gert wird nicht als Sicherheitseinrichtung (im Sinne der Richtlinie 94/9/EG Anhang II, Klausel 1,5) eingestuft: siehe auf Seite 5. 7) Installation und Prfung dieses Gerts drfen nur durch entsprechend geschultes Personal in bereinstimmung mit den geltenden Verfahrensregeln (EN 60079-14 und EN 6007917 in Europa) durchgefhrt werden. 8) Das Gert ist nicht reparierbar. 9) Die Zertifizierungsnummer besitzt eine Endung X zur Angabe, dass fr die sichere Verwendung Sonderbedingungen gelten. Personen, die diese Betriebsmittel installieren oder warten, mssen Zugriff auf die Bescheinigungen haben. 10) Es sind geeignete Manahmen zu treffen, um eine Beschdigung des Gerts im Falle eines Kontakts mit aggressiven Stoffen zu verhindern und die Schutzart zu gewhrleisten. Aggressive Stoffe: z. B. surehaltige Flssigkeiten oder Gase, die Metalle angreifen knnen, oder Lsungen, die polymerische Stoffe angreifen. Geeignete Manahmen: z. B. Aufstellung der Bestndigkeit gegen bestimmte Chemikalien anhand des Datenblatts des Materials.

Deutsch

A5E31997170

SITRANS LR250 (HART) KURZANLEITUNG

Seite DE-13

Programmierung des SITRANS LR250


Einschalten des SITRANS LR250

Siehe Schnellstartassistent mit dem Handprogrammiergert auf Seite 18. Siehe Schnellstartassistent ber SIMATIC PDM auf Seite 21.

Gert einschalten. SITRANS LR250 startet automatisch im Messmodus.

Die Anzeige (LCD) Anzeige im Messmodus (Normalbetrieb) 1) 2)


1)
(((

1 Symbol Umschalten fr lineare Einheiten oder % 2 Gewhlte Betriebsart: Fllstand, Leerraum, Abstand oder Volumen 3 Messwert (Fllstand oder Volumen, Leerraum oder Abstand) 4 Einheiten 5 Balkenanzeige zur Angabe des Fllstands 6 Im Zusatz-Anzeigefeld erscheint auf Anfrage2) die Elektroniktemperatur, Echogte, der Schleifenstrom oder Abstand 7 Textbereich zur Anzeige von Statusmeldungen 8 Symbol Gertezustand (nhere Angaben finden Sie in der ausfhrlichen Betriebsanleitung)

Anliegen von Fehlern S: 0 LOE


Wenn ein Fehler anliegt, erscheint der Fehlercode und eine Fehlermeldung im Textbereich (7). Im Bereich Gertestatus (8) erscheint das Symbol Service erforderlich.

Anzeige im Programmiermodus
Navigationsansicht aktuelle Eine sichtbare Menleiste zeigt an, aktuelles EintragsMen nummer dass die Menliste zu lang ist, um Menalle Eintrge anzuzeigen. aktueller leiste Eintrag Befindet sich der Balken halb Eintragsbalken unten in der Menleiste, bedeutet dies, dass der aktuelle Eintrag etwa in der Mitte der Liste ist. Die Hhe und relative Position des Eintragsbalkens auf der Menleiste gibt die Lnge der Menliste, sowie die ungefhre Position des aktuellen Eintrags in der Liste an. Je hher der Balken, desto weniger Eintrge. Parameteransicht
Parameternummer Parameterwert/ -auswahl

Deutsch

Bearbeitungsansicht

Parametername

1) 2)

Drcken Sie den Pfeil nach OBEN oder UNTEN zum Umschalten. Als Antwort auf eine Anfrage per Tastendruck. Nhere Angaben finden Sie unter Tastenfunktion im Messmodus auf Seite 15.

Seite DE-14

SITRANS LR250 (HART) KURZANLEITUNG

A5E31997170

Handprogrammiergert (Artikelnr. 7ML1930-1BK)


Nicht im Lieferumfang enthalten. Muss separat bestellt werden! 1) 2) 3)

Taste Tastenfunktion im Messmodus


Aktualisiert den Schleifenstrom1). Aktualisiert die Anzeige der Temperatur im Gehuse1). Aktualisiert den Echogtewert1).
C

Aktualisiert die Abstandsmessung1). Modus ffnet den PROGRAMMIER-Modus2). RECHTS-Pfeil ffnet den PROGRAMMIER-Modus3). Pfeil nach OBEN oder UNTEN schaltet zwischen linearen Einheiten und % um.

Programmierung ber Handprogrammiergert


Hinweise: Drcken Sie Modus zum Umschalten zwischen Mess- und Programmiermodus. Whrend sich das Gert im PROGRAMMIER-Modus befindet, bleibt der Ausgang unvernderlich und reagiert nicht auf nderungen des Gertes. SITRANS LR250 kehrt nach einer gewissen Ruhezeit im PROGRAMMIER-Modus (zwischen 15 Sekunden und 10 Minuten, je nach Men-Ebene) automatisch in den Messmodus zurck.

Parametermens
Die Parameter sind durch Namen gekennzeichnet und 1. QUICK START 2. SETUP in Funktionsgruppen gegliedert. 2.1. GERT ............. Eine Liste aller Parameter mit Anweisungen finden Sie 2.7. KENNLINIE in der ausfhrlichen Betriebsanleitung.

Deutsch

2.7.1. VOLUMEN 2.7.1.1. BEHLTERFORM

1) 2)

3)

Anzeige des neuen Wertes im Zusatzfeld des LCD. ffnet die zuletzt angezeigte Men-Ebene, seitdem das Gert zuletzt eingeschaltet wurde, es sei denn das Gert wurde seit Verlassen des PROGRAMMIER-Modus ausgeschaltet oder mehr als 10 Minuten sind vergangen, seit der PROGRAMMIER-Modus zuletzt aktiviert wurde. In diesem Fall erscheint die oberste Men-Ebene. ffnet die oberste Men-Ebene.

A5E31997170

SITRANS LR250 (HART) KURZANLEITUNG

Seite DE-15

1.

Aufruf PROGRAMMIER-Modus
Richten Sie das Programmiergert auf die Anzeige (aus maximal 300 mm [1 ft] Abstand). Der RECHTS-Pfeil aktiviert den PROGRAMMIER-Modus und ffnet Men-Ebene 1. Modus ffnet die im PROGRAMMIER-Modus innerhalb der letzten 10 Minuten zuletzt angezeigte Men-Ebene oder Men-Ebene 1, wenn das Gert seither aus- und wieder eingeschaltet wurde.

Handprogrammiergert Anzeige

Max. 300 mm (1 ft)

2.

Navigation: Tastenfunktion im Navigationsmodus

Hinweis: Fr einen Schnellzugriff auf Parameter ber das Handprogrammiergert drcken Sie die Taste Home , gefolgt von der Mennummer, z. B.: 2.7.1 (Volumen).

Taste

Name

MenEbene

Funktion im Navigationsmodus
Auf das vorige oder nchste Men bzw. den vorigen oder nchsten Parameter blttern. Aufruf des ersten Parameters im gewhlten Men, oder nchstes Men ffnen. Aufruf des Bearbeitungs-Modus. Aufruf bergeordnetes Men. bergang auf MESS-Modus. ffnen der obersten Men-Ebene: Men 1.

Pfeil nach OBEN Men oder oder UNTEN Parameter

RECHTS-Pfeil LINKS-Pfeil
Modus Home

Men Parameter Men oder Parameter Men oder Parameter Men oder Parameter

3. Deutsch

Bearbeiten im PROGRAMMIER-Modus
Option aus einer Liste whlen a) Rufen Sie den gewnschten Parameter auf. b) Drcken Sie den RECHTS-Pfeil Parameteransicht zu ffnen. , um die
aktuelle Auswahl Parametername Parameternummer

c)

Drcken Sie erneut den RECHTS-Pfeil erneut, um den Bearbeitungs-Modus zu starten. Die aktuelle Auswahl ist hervorgehoben. Blttern Sie auf eine neue Auswahl.

d)

Drcken Sie den RECHTS-Pfeil

zur

Besttigung. Das LCD kehrt auf die Parameteransicht zurck und zeigt die neue Auswahl an. Seite DE-16 SITRANS LR250 (HART) KURZANLEITUNG A5E31997170

Einen numerischen Wert ndern a) Rufen Sie den gewnschten Parameter auf. b) Drcken Sie erneut den RECHTS-Pfeil erneut, um den Bearbeitungs-Modus zu starten. Der aktuelle Wert wird hervorgehoben. Geben Sie einen neuen Wert ein. Drcken Sie den RECHTS-Pfeil zur Besttigung. Das LCD kehrt auf die Parameteransicht zurck und zeigt die neue Auswahl an.

Parametername

Parameternummer

c) d)

aktueller Wert

Tastenfunktionen im Bearbeitungsmodus Taste Name Funktion im Bearbeitungsmodus


Blttert auf den Eintrag. - Erhht oder vermindert Zahlen. - Schaltet das Vorzeichen um (plus und minus). - Besttigt Daten (schreibt den Parameter). - Modus wechselt von Bearbeiten auf Navigation. - Bewegt den Cursor eine Stelle nach rechts - oder, wenn sich der Cursor auf dem Enter-Zeichen befindet: besttigt Daten und Modus wechselt von Bearbeiten auf Navigation. Abbruch der Bearbeitung, ohne den Parameter gendert zu haben. - Bewegt den Cursor auf das Plus/Minus-Zeichen, wenn dies die erste Taste ist, die gedrckt wird - oder bewegt den Cursor eine Stelle nach links. Lscht die Anzeige.

Pfeil nach Optionsauswahl OBEN oder Numerische UNTEN Bearbeitung Optionsauswahl

RECHTSPfeil

Numerische Bearbeitung Optionsauswahl

LINKS Pfeil:

Numerische Bearbeitung Numerische Bearbeitung Numerische Bearbeitung Numerische Bearbeitung Numerische Bearbeitung

Lschen Dezimalstelle Plus oder MinusZeichen


...

Deutsch

Eingabe eines Dezimalpunkts.

ndert das Vorzeichen des eingegebenen Wertes.

Ziffern

Eingabe des entsprechenden Zeichens.

A5E31997170

SITRANS LR250 (HART) KURZANLEITUNG

Seite DE-17

Schnellstartassistent mit dem Handprogrammiergert


1. Quick Start
a) Richten Sie das Programmiergert auf die Anzeige (aus maximal 300 mm [1 ft] Abstand); drcken Sie dann den RECHTS-Pfeil MIERUNG und ffnen Sie Men-Ebene 1. b) Drcken Sie den RECHTS-Pfeil Parameteransicht zu ffnen. c) Drcken Sie den RECHTS-Pfeil den Pfeil nach UNTEN Eintrag zu gelangen. d) Zur nderung einer Einstellung blttern Sie auf den gewnschten Eintrag oder geben Sie einen neuen Wert ein. e) Nachdem der Wert gendert wurde, drcken Sie den RECHTS-Pfeil gung und den Pfeil nach UNTEN f) zur Bestti, um den Bearbeitungsmodus zu ffnen, oder zweimal, um Meneintrag 1.1 aufzurufen und die zur Aktivierung der PROGRAM-

, um die Vorgabewerte zu besttigen und zum nchsten

, um zum nchsten Eintrag zu gelangen.

Schnellstarteinstellungen treten erst in Kraft, wenn Ja in nderungen durchfhren in Schritt 1.8. gewhlt wurde.

1.1. Language Whlt die auf der Anzeige zu verwendende Sprache und tritt sofort in Kraft.
Optionen
ENGLISH, DEUTSCH, FRANAIS, ESPAOL

1.2. Material Optionen


FLSSIGKEIT FLSSIGK. NIED. (Flssigkeit mit niedriger Dielektrizittszahl)

Deutsch

1.3. Reaktionszeit Einstellung der Geschwindigkeit, mit der das Gert auf nderungen im Zielbereich reagiert.
Reaktionszeit
LANGSAM MITTEL SCHNELL

Befllgeschwindigkeit/Entleergeschwindigkeit
0,1 m/min (0.32 ft/min) 1,0 m/min (3.28 ft/min) 10,0 m/min (32.8 ft/min)

Die Einstellung sollte die max. Geschwindigkeit beim Befllen oder Entleeren (es gilt der grere Wert) gerade etwas bersteigen.

1.4. Einheiten Maeinheiten des Sensors. Optionen


M, CM, MM, FT, IN

Seite DE-18

SITRANS LR250 (HART) KURZANLEITUNG

A5E31997170

1.5. Betriebsart
SensorBezugspunkt Leerraum SensorBezugspunkt

Abstand Ob. Kal. Punkt

SensorBezugspunkt

Fllstand Unt. Kal. Punkt

Betriebsart

Beschreibung

Keine Aktualisierung der Messung und des zugehrigen Schleifenstroms; das Gert KEIN BETRIEB nimmt automatisch den fehlersicheren Modus (Fail-safe) an1). FLLSTAND LEERRAUM ABSTAND
1)

Abstand zur Materialoberflche bezogen auf den Unteren Kalibrierungspunkt. Abstand zur Materialoberflche bezogen auf den Oberen Kalibrierungspunkt. Abstand zur Materialoberflche bezogen auf den Sensor-Bezugspunkt.

Genauere Angaben zur Fehlersicherheit (Fail-safe) finden Sie in der ausfhrlichen Betriebsanleitung.

1.6. Unterer Kalibrierungspunkt Abstand vom Sensor-Bezugspunkt zum Unteren Kalibrierungspunkt: entspricht in der Regel dem Nullpunkt des Prozesses. Eine Darstellung finden Sie unter Betriebsart (1.5.). Werte
Bereich: 0,00 ... 20,00 m

1.7. Oberer Kalibrierungspunkt Abstand vom Sensor-Bezugspunkt zum Oberen Kalibrierungspunkt: entspricht in der Regel dem Vollpunkt des Prozesses. Eine Darstellung finden Sie unter Betriebsart (1.5.). Werte
Bereich: 0,00 ... 20,00 m

1.8. Durchfhren? (nderungen durchfhren) Um die Schnellstart-Einstellungen zu speichern, whlen Sie Ja in nderungen durchfhren. Optionen
Drcken Sie Modus betriebsbereit. JA, NEIN, ABGESCHLOSSEN (Auf der Anzeige erscheint ABGESCHL, wenn der Schnellstart erfolgreich durchgefhrt wurde.) , um in den Messmodus zurckzukehren. SITRANS LR250 ist nun

Deutsch

Hinweis: Wenn es sich um einen Tank mit Einbauten handelt, lesen Sie bitte in der Betriebsanleitung die Angaben zur Autom. Strechoausblendung nach.

A5E31997170

SITRANS LR250 (HART) KURZANLEITUNG

Seite DE-19

SITRANS LR250 Kommunikation: HART

Eine Liste aller zutreffenden Parameter finden Sie in der ausfhrlichen Betriebsanleitung. Zur Programmierung Ihres Gerts empfehlen wir SIMATIC Process Device Manager (PDM). Anwendungsanleitungen zur Einstellung von HART-Gerten mit SIMATIC PDM knnen von der Produktseite unserer Website unter folgender Adresse heruntergeladen werden: www.siemens.com/LR250 unter Support.

SIMATIC PDM
SIMATIC PDM ist ein Softwarepaket fr die Inbetriebnahme und Wartung des SITRANS LR250 und anderer Prozessgerte. Detailgenaue Angaben zur Verwendung von SIMATIC PDM sind in der Betriebsanleitung oder Online-Hilfe enthalten. (Weitere Informationen finden Sie unter www.siemens.com/simatic-pdm.)

Deaktivieren von Puffern

Hiermit wird SIMATIC PDM mit dem HART-Modem fr die Betriebssysteme Windows 2000 und Windows XP 1) abgestimmt. Hinweise: SIMATIC PDM arbeitet nur in der Version Windows XP Professional, nicht in der Home Edition. Sie bentigen administrative Rechte ber Ihr Betriebssystem, um die Puffer zu deaktivieren. 1) 2) 3) 4) 5) 6) Klicken Sie zur Konfiguration auf Start/Einstellungen/Systemsteuerung. Doppelklicken Sie auf System, whlen Sie das Register Hardware und klicken Sie darin auf Device Manager. ffnen Sie den Ordner Ports (Anschlsse) und doppelklicken Sie auf den vom System verwendeten COM Port (COM-Anschluss), um das Fenster Communications Port Properties zu ffnen. Whlen Sie Port Settings (Anschlusseinstellungen) und doppelklicken Sie auf die Schaltflche Advanced (Erweitert). Entfernen Sie eine ggf. vorhandene Markierung im Kontrollkstchen zu Use FIFO buffers (FIFO-Puffer verwenden). Klicken Sie zur Besttigung auf OK. Schlieen Sie alle geffneten Fenster und fhren Sie einen Neustart des Computers durch.

Aktualisieren der Electronic Device Description (EDD)


Hinweise:
SIMATIC PDM Rev. 5.2 SP1 wird nur fr eine grundlegende Konfiguration und Fehlersuche untersttzt. Erweiterte Funktionen, wie der Schnellstartassistent, erfordern die Revision 6.0 SP4 oder hher. In SIMATIC PDM befindet sich die EDD im Gertekatalog, unter Sensors/Level/Echo/ Siemens Milltronics/SITRANS LR250. Gehen Sie zur Produktseite unserer Website: www.siemens.com/LR250, um sicherzustellen, dass Sie die neueste Version von SIMATIC PDM, das aktuellste Servicepaket (SP) und den aktuellsten Hotfix (HF) haben. Whlen Sie Deutsch als Sprache und siehe Support > Software Downloads.

Deutsch

Zum Installieren einer neuen EDD:


1) 2) 3) Download der EDD von der Produktseite unserer Website: www.siemens.com/LR250 und Speichern der Dateien auf Ihrem Computer. Entpacken Sie die komprimierte Datei an eine leicht erreichbare Stelle. Starten Sie SIMATIC PDM Manager Device Catalog, blttern Sie bis zur entzippten EDDDatei und whlen Sie diese.
Windows ist ein eingetragenes Warenzeichen der Microsoft Corporation

1)

Seite DE-20

SITRANS LR250 (HART) KURZANLEITUNG

A5E31997170

Konfiguration eines neuen Gertes


Hinweis: Das Anklicken von Abbrechen whrend eines Uploads vom Gert in SIMATIC PDM hat die Aktualisierung einiger Parameter zur Folge.
1) 2) berprfen Sie, ob Sie die neueste EDD besitzen und aktualisieren Sie sie bei Bedarf (siehe Aktualisieren der Electronic Device Description (EDD) auf Seite 20). Starten Sie den SIMATIC Manager und legen Sie ein neues Projekt fr LR250 an. Anwendungsanleitungen zur Einstellung von HART-Gerten mit SIMATIC PDM knnen von der Produktseite unserer Website unter folgender Adresse heruntergeladen werden: www.siemens.com/LR250. ffnen Sie das Men Gert Gert zurcksetzen und klicken Sie auf OK, um ein Rcksetzen auf Werkseinstellungen durchzufhren. Nach Beenden des Rcksetzens laden Sie die Parameter in PC/PG. Konfigurieren Sie das Gert mit dem Schnellstartassistenten.

3) 4) 5)

Schnellstartassistent ber SIMATIC PDM


Hinweise:
Die Einstellungen des Schnellstartassistenten sind zusammenhngend und nderungen werden erst wirksam, wenn Sie am Ende von Schritt 5 auf FERTIGSTELLEN UND BERTRAGEN drcken; die Einstellungen werden damit offline gespeichert und an das Gert bertragen. Klicken Sie auf ZURCK, um zurckzukehren und die Einstellung zu berprfen oder auf Abbrechen, um den Schnellstart zu verlassen. Starten Sie SIMATIC PDM, ffnen Sie das Men Gert Assistent - Schnellstart und folgen Sie den Schritten 1 bis 5.

Schnellstart

Bedienung ber FDT (Field Device Tool)


FDT ist ein in mehreren Softwarepaketen integrierter Standard fr die Inbetriebnahme und Wartung von Feldgerten. PACTware und Fieldcare sind zwei handelsbliche FDTs. Um ein Feldgert ber FDT zu projektieren, ist der DTM (Device Type Manager) fr dieses Gert erforderlich. In Siemens Messgerten wird SITRANS DTM, sowie eine fr SITRANS DTM geschriebene EDD eingesetzt. 1) Installieren Sie zunchst SITRANS DTM auf Ihrem System. Der Download ist unter folgender Adresse mglich: http://support.automation.siemens.com. Whlen Sie Deutsch als Sprache und klicken Sie auf Produkt Support; weiter navigieren zu Produktinformationen/Automatisierungstechnik/ Sensoriksysteme/Prozessinstrumentierung/Software & Communications. 2) Installieren Sie die SITRANS LR250 HART EDD fr SITRANS DTM. Diese steht auf der Produktseite unserer Website zum Download zur Verfgung: www.siemens.com/LR250. Whlen Sie Deutsch als Sprache und siehe Support > Software Downloads. A5E31997170 SITRANS LR250 (HART) KURZANLEITUNG Seite DE-21

Deutsch

Konfiguration eines neuen Gerts ber FDT


Applikationsbeispiele stehen auf der Produktseite unserer Website unter Support zum Download zur Verfgung.

Bedienung ber AMS Device Manager


AMS Device Manager ist ein Softwarepaket fr die Inbetriebnahme und Wartung von Feldgerten. Detailgenaue Angaben zur Verwendung von AMS Device Manager sind in der Betriebsanleitung oder Online-Hilfe enthalten. Weitere Informationen finden Sie unter: http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD)


SITRANS LR250 erfordert die EDD fr den AMS Device Manager Version 9.0.

Konfiguration eines neuen Gerts ber AMS Device Manager


1) Gehen Sie zur Produktseite unserer Website: www.siemens.com/LR250 und berprfen Sie, ob Sie die neueste EDD besitzen. Siehe Support > Software Downloads und laden Sie sie bei Bedarf herunter. Speichern Sie die Dateien auf Ihrem Computer und extrahieren Sie die komprimierte Datei an eine leicht erreichbare Stelle. Starten Sie AMS Device Manger Add Device Type (Gertetyp hinzufgen), blttern Sie bis zur entzippten EDD-Datei und whlen Sie diese.

2)

Starten Sie den AMS Device Manager. Applikationsbeispiele zur Einstellung von HART-Gerten mit AMS Device Manager knnen von der Produktseite unserer Website unter Support heruntergeladen werden.

Wartung
Unter normalen Betriebsbedingungen erfordert SITRANS LR250 keine Wartung oder Reinigung. Bei sehr rauen Betriebsbedingungen empfehlen wir jedoch eine Reinigung: 1) 2) Bercksichtigen Sie das Antennenmaterial und das Prozessmedium und whlen Sie ein geeignetes Reinigungsmittel. Bauen Sie das Gert aus und wischen Sie die Antenne mit einem Tuch und einem geeigneten Reinigungsmittel ab.

Gertereparatur und Haftungsausschluss


Nhere Angaben finden Sie auf der hinteren Umschlaginnenseite.

Deutsch

Ersatz von Antenne oder Elektronik/Gehuse


Wenn die Antenne in Folge einer Beschdigung oder Strung ersetzt werden muss, ist keine Neukalibrierung erforderlich. Der Austausch der Antenne kann von einem genehmigten Siemens Reparaturzentrum oder von befugtem Siemens Personal durchgefhrt werden. Wenn die Elektronik oder das Gehuse in Folge einer Beschdigung oder Strung ersetzt werden muss, stellen Sie bitte sicher, dass die richtige Antennenausfhrung verwendet wird. Andernfalls ist eine Neukalibrierung durch befugtes Siemens Personal vorzunehmen.

Seite DE-22

SITRANS LR250 (HART) KURZANLEITUNG

A5E31997170

Kurve der maximalen Prozesstemperatur


Flanschadapter-Ausfhrung des LR250
WARNUNG: Die Innentemperatur darf 80 C (176 F) nicht berschreiten! Hinweise: Das Schaubild dient nur der Veranschaulichung. Es werden nicht alle mglichen Prozessanschlsse dargestellt. (Es trifft beispielsweise NICHT zu, wenn das SITRANS LR250 direkt an einer Behlteroberflche aus Metall montiert ist.) Eine Erwrmung durch direkte Sonneneinstrahlung wird im Schaubild nicht bercksichtigt. Parameter 3.2.1 Aktuelle Innentemperatur berwacht die Innentemperatur. Umgebungstemperatur (oC) Max. Prozesstemperatur im Vgl. zur zulssigen Umgebungstemperatur

Prozesstemperatur

Prozesstemperatur (oC)

Schleifenstrom
Anforderungen an die Spannungsversorgung
Hinweis: Die unten gezeigten Kurven beziehen sich auf ein Stand-Alone-Gert, das ber das Siemens Handprogrammiergert konfiguriert wurde. Kurve 1: Allg. Verwendung, Eigensicher, Nicht funkend (Non Sparking)/ Energiebegrenzt (Energy Limited)/ Betriebsmittel fr Zone 2 (Non-incendive)

Schleifenwiderstand RL

ZULSSIGER BETRIEBSBEREICH

Deutsch

Schleifenspannung VL

Hinweis: Bei einer Kommunikation ber HART betrgt die Mindestspannung bei 220 Ohm (RL) DC 16,3 V.

A5E31997170

SITRANS LR250 (HART) KURZANLEITUNG

Seite DE-23

Schleifenwiderstand RL

Kurve 2: Druckfeste Kapselung (Flameproof), Erhhte Sicherheit, Druckfeste Kapselung (Explosion-proof)

ZULSSIGER BETRIEBSBEREICH

Schleifenspannung VL

Hinweis: Bei einer Kommunikation ber HART betrgt die Mindestspannung bei 220 Ohm (RL) DC 20,94 V.

Anlaufverhalten
Die Stromaufnahme bei Inbetriebnahme betrgt weniger als 3,6 mA. Bis zur ersten Messung vergehen weniger als 50 Sekunden.

Deutsch
Seite DE-24

SITRANS LR250 (HART) KURZANLEITUNG

A5E31997170

SITRANS LR250 (HART)


SITRANS LR250 (HART 1)). , . SITRANS LR250 Web : www.siemens.com/LR250. Siemens Milltronics. : Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P.O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 Email: techpubs.smpi@siemens.com
: Siemens AG 2013. Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments , . , . , . . . .

MILLTRONICS Siemens Milltronics Process Instruments.


24 . , Siemens, : www.siemens.com/automation/partner: Contact (), Service (), Service () (+Automation Technology ( ) > +Sensor Systems ( ) >+Process Instrumentation ( ) > +Level Measurement ( ) > +Continuous ()). Radar (). /. Technical Support ( ) Service (). on-line : www.siemens.com/automation/support-request (SITRANS LR250) , Search () . Next. . , Next Siemens. Siemens IA/DT: +49 (0)911 895 7222

1)

HART HART Communication Foundation.

A5E31997170

SITRANS LR250 (HART)

L-1

. . , , / . , , / . : .

FCC
...: (FCC)
: Siemens Milltronics . : A, 15 FCC. , . , , , . , .

SITRANS LR250
: SITRANS LR250 , . : . . SITRANS LR250 2 25 GHz , 20 m (66ft). . . PVDF. SITRANS LR250 HART SIMATIC PDM . Process Intelligence.

EL-2

SITRANS LR250 (HART) -

A5E31997170


SITRANS LR250 (HART). , 4.

/ :
, : 23. , / , . . Web www.siemens.com/LR250, Support () > Installation Drawings
( ) > Level Measurement ( ) > Continuous - Radar ( - ).

Installation Drawings ( ), Process Connection Diagrams ( ). , . . 23 .


( ) 40 C 80 C (40 F 176 F)

: : - FKM: 40 oC +200 oC (40 oF +392 oF) - FFKM: 20 oC +200 oC (-4 oF +392 oF) PVDF 2" NPT / BSPT / G: -40 +80 C (-40 +176 F)

/ (FM/CSA, .../ ) (FM/CSA, .../ ) : 24 V DC 550 Ohm : 24 V DC 250 Ohm

30 V DC 4 20 mA . :

24

A5E31997170

SITRANS LR250 (HART)

L-3


: . CSAUS/C, FM, CE, C-TICK, KC (R&TTE), FCC, Industry Canada
1)

() () (.../)

()

ATEX II 1G, EEx ia IIC T4 ATEX II 1D, EEx tD A20 IP67 T90 C IECEx SIR 05.0031X, Ex ia IIC T4, Ex tD A20 IP67 T90 C FM/CSA I, . 1, A, B, C, D II, . 1, E, F, G III T4 INMETRO: DNV 12.0087 X Ex ia IIC T4 Ga Ex ia ta IIIC T90 C Da IP65/IP67 -40 C Ta +80 C DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-11:2009 ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

()

NEPSI Ex ia IIC T4/ DIP A20 TA, T90 C IP67

/ 2) () ATEX II 3 G Ex nA/nL IIC T4 Gc () NEPSI Ex nA II T4/ Ex nL IIC T4 3) (.../) FM/CSA I, . 2, A, B, C, D T5 4) (/) ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex dmbia IIC T4 Ga/Gb Ex tD A20 IP67 T90 oC () INMETRO: DNV 12.0088 X Ex d ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-1:2009, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

1)

2) 3) 4)

9. / 11. 11. 12. SITRANS LR250 (HART) - A5E31997170

EL-4

() 1) (/)

()

ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex embia IIC T4 Ga/Gb Ex tD A20 IP67 T90 oC INMETRO: DNV 12.0088 X Ex d ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-7:2008,
ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010,
ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e

IEC 61241-11:2005 () / () NEPSI Ex dmbia IIC T4/ Ex embia IIC T4/ DIP A20 TA, T90 oC IP67 FM/CSA I, . 1, A, B, C, D II, . 1, E, F, G III T4

2) (/)

Lloyds ABS


:
, , . , , . . : 2). , . SITRANS LR250 ASME .

, PED, 97/23/EK
Siemens , , , ( 1/8). 3)

1) 2)

12. (.../ ) 13. SITRANS LR250 (HART) L-5

A5E31997170


: . , . PVDF. , , . ( ) . , , , . : , ETSI EN 302372. , , . . : MM XXX ( MM = , = , = XXX= ) ( ) , , , .


PVDF . : 10 mm (0,4") . : 10 mm (0,4")

. . 51 mm (2")

10 mm (0,4), . PVDF 51 mm (2"). (100 mm/ 3,93") , PVDF.

3)

: . , . SITRANS LR250 (HART) - A5E31997170

EL-6


, , .

. , . . . , .

: PVDF = 19 1,5" = 19 2" = 15 3" = 10 4" = 8


, , I .


: PTFE , . 40 N-m (30 ft-lbs.). : .


1) 2) , , . . , .


: , , .

A5E31997170

SITRANS LR250 (HART)

L-7

: DC , IEC 61010-1. , .

SITRANS LR250
: , . IP NEMA. 9. : : AWG 22 14 (0,34 mm2 2,5 mm2). .
Allen 2 mm .

(IP 68) ( )

1) 2) ( NPT2))

1) 2) 3) 4) 5)

70 mm (2,75") . 1) 2) , : . . . .

1) 2)

. , .
SITRANS LR250 (HART) - A5E31997170

EL-8

HART
PLC/mA HART 1) 2) : , PLC HART1)1) 2). HART .

PLC R= 250 HART SITRANS LR250 HART communicator


. , , .

1.
(ATEX/IECEX/INMETRO/C-TICK)

ATEX Web : www.siemens.com/LR250. Support () > Approvals / Certificates ( / ). IECEx IECEx. : http://iecex.iec.ch Ex Equipment Certificates of Conformity ( ) IECEx SIR 05.0031X.

1) 2)

= 250 Ohm (). 1: , , / / 23 2: , , 24. SITRANS LR250 (HART) L-9

A5E31997170

()
(FM/CSA)

FM/CSA, 23650653, Web : www.siemens.com/LR250. Support () > Installation Drawings ( ) > Level Measurement ( ) > Continuous - Radar ( - ).

1: , , / / 23. : . NEMA 4X / 4X / NEMA 6, IP67, IP68. 13. ( ) 10 ( ) 11.

: PLC, . .

( )
: .
MTL MTL Stahl Stahl


787SP+ ( ) 7787P+ ( ) 9001/01-280-100-10 ( ) 9002/01-280-110-10 ( )

SITRANS LR250
(IS) LR250, : R. = . R. = ( , , PLC / ) V. = 1) R. . 2) R. 3) R = R. + R.

EL-10

SITRANS LR250 (HART) -

A5E31997170

()
4) 5) 6) V. ( , ). V = V. V. V R 1: , , / / 23.
706 7206 SITRANS I 7NG4124 9001/51-280-110-14 E02009 E02010

( )

MTL MTL Siemens Stahl MTL MTL

2. /

ATEX Web : www.siemens.com/LR250. : Support () > Approvals / Certificates ( /

). 1: , , / / 23. .

3.

FM/CSA 1, . 2, 23650673, Web : www.siemens.com/LR250. Support () >


Installation Drawings ( ) > Level Measurement ( ) > Continuous - Radar ( - ).

1: , , / / 23.

A5E31997170

SITRANS LR250 (HART)

L-11

4.

ATEX Web : www.siemens.com/LR250. : Support () > Approvals / Certificates ( / ). IECEx IECEx. : http://iecex.iec.ch Ex Equipment Certificates of Conformity ( ) IECEx SIR 08.0107X.

2: , , 24. . 13 ATEX .

5.

ATEX Web : www.siemens.com/LR250. : Support () > Approvals / Certificates ( / ).


IECEx IECEx. : http://iecex.iec.ch Ex Equipment Certificates of Conformity ( ) IECEx SIR 08.0107X.

2: , , 24. . 13 ATEX .

EL-12

SITRANS LR250 (HART) -

A5E31997170

6. (.../ )

FM/CSA, A5E02257843, Web : www.siemens.com/LR250.


Support () > Installation Drawings ( ) > Level Measurement ( ) > Continuous - Radar ( - ).

2: , , 24.


( ATEX 94/9/E, II, 1/0/6)
SIRA 06ATEX2358X, SIRA 08ATEX1301X SIRA 09ATEX4153X. . 1GD SIRA 06ATEX2358X, 1/2 GD, 1D, 2D SIRA 08ATEX1301X 3G SIRA 09ATEX4153X. 3) IIC, IIB IIA T1, T2, T3 T4. 4) IP67 T90 C, . 5) -40 C +80 C. 6) ( 94/9/E, II, 1.5): , PED, 97/23/EK 5. 7) , (EN 60079-14 EN 60079-17 ). 8) . 9) X, . . 10) , , . : , . 1) 2)

: , .

A5E31997170

SITRANS LR250 (HART)

L-13

SITRANS LR250
18. SIMATIC PDM 21.

SITRANS LR250
. SITRANS LR250 Measurement ().

LCD
Measurement () ( ) 1) 2)((( 1 1) %
1 2 3 4

2 : level (), space ( ), distance () volume () 3 ( , ) 4 5 6 , 2), , , 7 8 ( )

S: 0 LOE

, (7), (8)

Program Navigation ()

, . , . . .

Parameter ()
/

Edit ()

1)
2)

. . Measurement () 15. SITRANS LR250 (HART) - A5E31997170

EL-14

(. 7ML1930-1BK)
. 1)
2) 3)

- Measurement ()
1). 1). 1). 1). Mode PROGRAM2).

PROGRAM3).

%.


: Mode ( ) Measurement Program. PROGRAM, . SITRANS LR250 Measurement PROGRAM ( 15 10 , ).

. .

1. QUICK START 2. SETUP 2.1. DEVICE ............. 2.7. LINEARIZATION 2.7.1. VOLUME 2.7.1.1. VESSEL SHAPE

1) 2)

3)

LCD. , PROGRAM 10 PROGRAM. . SITRANS LR250 (HART) L-15

A5E31997170

1.

PROGRAM
( 300 mm [1 ft.]). PROGRAM 1 . Mode PROGRAM 10 1 , .

. 300 mm (1 ft)

2.

: Navigation ()
,

: , Home : 2.7.1 ().


Home Navigation () . . Edit (). . MEASUREMENT (). : 1.

3.

PROGRAM
a) b) . .

c)

Edit (). . d) . LCD .


EL-16 SITRANS LR250 (HART) - A5E31997170


a) b) . Edit (). . . . LCD .

c) d)

Edit ()
Edit ()
. - . - . - ( ). - . - - Enter, . . - / , - . .

A5E31997170

SITRANS LR250 (HART)

L-17


1.
a) 300 mm (1 ft), b) c) d) e) f) PROGRAM 1 . 1.1 .

. , . , .

Yes () Apply changes ( ) (1.8.).

1.1.
LCD . ENGLISH (), DEUTSCH (), FRANCAIS (), ESPANOL ()

1.2.
LIQUID () LIQUID LOW DK ( )

1.3.
. SLOW () MED () FAST () / 0,1 m/min (0,32 ft/min) 1,0 m/min (3,28 ft/min) 10,0 m/min (32,8 ft/min)

( ).

1.4.
. M, CM, MM, FT, IN

EL-18

SITRANS LR250 (HART) -

A5E31997170

1.5.
.

NO SERVICE ( ) LEVEL () SPACE ( ) DISTANCE ()


1)

, 1). . . .

1.6.
: . (1.5.) . : 0,00 20,00 m

1.7.
: . (1.5.) .. : 0,00 20,00 m

1.8. Apply? ( )
Yes () . YES (), NO (), DONE () ( DONE .)

Mode Measurement (). SITRANS LR250 . : , .

A5E31997170

SITRANS LR250 (HART)

L-19

SITRANS LR250: HART


. SIMATIC Process Device Manager (PDM) . HART SIMATIC PDM Web : www.siemens.com/LR250, Support ().

SIMATIC PDM
SIMATIC PDM SITRANS LR250 . SIMATIC PDM, . ( www.siemens.com/simatic-pdm.) SIMATIC PDM HART Windows 2000 Windows XP 1). : SIMATIC PDM Windows XP Professional, Home. buffer. 1) 2) 3) // . , . Ports Communications Port : Communications Port. . buffer FIFO , . OK . .

buffer

4) 5) 6)

(EDD)
: 5.2 SP1 SIMATIC PDM . , , 6.0 SP4 . SIMATIC PDM EDD Device Catalog ( ), Sensors/Level/Echo/Siemens Milltronics/SITRANS LR250. Web : www.siemens.com/LR250, SIMATIC PDM, Service Pack (SP) hot fix (HF). Support () > Software Downloads ( ). EDD Web : www.siemens.com/LR250 . . SIMATIC PDM Manager Device Catalog ( ), EDD .
Windows Microsoft Corporation SITRANS LR250 (HART) - A5E31997170

EDD:
1) 2) 3)

1)

EL-20


: Cancel () SIMATIC PDM, . 1) 2) EDD , , ( (EDD) 20). SIMATIC Manager LR250. HART SIMATIC PDM web : www.siemens.com/LR250. Device Master Reset ( ) OK . , PC/PG. .

3) 4) 5)

SIMATIC PDM
: FINISH AND DOWNLOAD ( ) 5, offline . BACK () Cancel () . SIMATIC PDM, Device Wizard - Quick Start ( ) 1 5.

FDT (Field Device Tool)


FDT . FDT PACTware Fieldcare. FDT DTM (Device Type Manager) . Siemens SITRANS DTM EDD SITRANS DTM. 1) SITRANS DTM . : http://support.automation.siemens.com. Product Support ( ) Product Information/ Automation Technology/Sensor systems/Process Instrumentation/Software & Communications. HART EDD SITRANS LR250 SITRANS DTM. Web : www.siemens.com/LR250. Support () > Software Downloads ( ).

2)

FDT
Web , Support ().
A5E31997170 SITRANS LR250 (HART) L-21

AMS Device Manager


AMS Device Manager . AMS Device Manager, . : http://www.emersonprocess.com/AMS/.

(EDD)
SITRANS LR250, EDD AMS Device Manager, 9.0.

AMS Device Manager


1) Web : www.siemens.com/LR250 EDD. Support () > Software Downloads ( ) , . . AMS Device Manager Add Device Type, EDD .

2)

AMS Device Manager. HART AMS Device Manager Web , Support ().

SITRANS LR250 . : 1) 2) , . .

/
, . Siemens. , , Siemens.

EL-22

SITRANS LR250 (HART) -

A5E31997170


LR250

H: 80 C (176 F). : . ( SITRANS LR250 .) , 3.2.1 .


(oC)

(oC)



: , Siemens.
1: , , / /

R.

V.

: HART, 220 Ohm (RL) 16,3 VDC.

A5E31997170

SITRANS LR250 (HART)

L-23

2: , , R.

V.

: HART, 220 Ohm (.) 20,94 VDC.


3,6 mA . 50

EL-24

SITRANS LR250 (HART) -

A5E31997170

SITRANS LR250 (HART) Puesta en marcha rpida


Espaol
Este manual contiene indicaciones importantes y condiciones para la utilizacin del SITRANS LR250 (HART 1)). Le recomendamos encarecidamente que obtenga la versin completa del manual de instrucciones para beneficiarse de todas las funciones del instrumento. Puede descargar la versin completa del manual de la pgina SITRANS LR250 de nuestro sitio web: www.siemens.com/LR250. Los manuales tambin estn disponibles en versin impresa contacte su representante Siemens Milltronics. Para ms informacin acerca de este documento contacte: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canad, K9J 7B1 E-mail: techpubs.smpi@siemens.com

Copyright Siemens AG 2013. Todos los derechos reservados


Recomendamos a nuestros usuarios obtengan copias impresas de la documentacin o consulten las versiones digitales diseadas y comprobadas por Siemens Milltronics Process Instruments. En ningn caso ser Siemens Milltronics Process Instruments responsable de reproducciones totales o parciales de la documentacin, ya sea de versiones impresas o electrnicas.

Exencin de responsabilidad
Hacemos todo lo necesario para garantizar la conformidad del contenido de este manual con el instrumento proporcionado. Sin embargo, estas informaciones quedan sujetas a cambios y no asumimos responsabilidad alguna en caso de modificaciones. Examinamos y corregimos el contenido de este manual regularmente y nos esforzamos en proporcionar publicaciones cada vez ms completas. No dude en contactarnos si tiene preguntas o comentarios. Las especificaciones estn sujetas a cambios.

MILLTRONICS es una marca registrada de Siemens Milltronics Process Instruments.

Soporte Tcnico
El soporte est disponible las 24 horas al da. Para encontrar los interlocutores de Siemens Automation en todo el mundo (oficinas, infolines y fax), consulte: www.siemens.com/automation/partner: Seleccione el idioma, haga clic en el botn Contacto, seleccione Servicios y luego haga clic otra vez en Servicios (+Automation Technology > +Sensor Systems >+Instrumentacin de procesos > +Medida de nivel > +Continuous). Seleccione Radar. Seleccione un pas y una ciudad/regin. Seleccione Technical Support bajo Servicio. Para soporte online, consulte: www.siemens.com/automation/support-request Introduzca el nombre de aparato (SITRANS LR250) o el nmero de pedido, haga clic en Search (Buscar), y seleccione el tipo de producto correspondiente. Haga clic en Seguir. Bsqueda por palabra clave: introduzca la palabra. Acceda a la documentacin relacionada con el producto o haga clic en Next (Seguir) para contactar con los servicios de soporte tcnico Siemens. Servicio de soporte tcnico Siemens IA/DT: Tel: +49 (0) 911 895 7222
1)

HARTes un marca registrada de la HART Communication Foundation.

A5E31997170

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

Pgina ES-1

Notas de seguridad
Espaol

Es imprescindible respetar las indicaciones de seguridad para una utilizacin sin peligro alguno para el usuario, el personal, el instrumento y los equipos conectados a ste. Por motivos de claridad expositiva en los textos de indicacin y de precaucin se destaca el nivel de precaucin necesario para cada intervencin. El smbolo de ADVERTENCIA se refiere a un marcado colocado en el producto. Significa que al no observar las precauciones de seguridad se puede provocar la muerte, lesiones corporales graves y/o daos materiales considerables. El smbolo de ADVERTENCIA se refiere a un marcado colocado en el producto. Significa que al no observar las precauciones de seguridad se puede provocar la muerte, lesiones corporales graves y/o daos materiales considerables. Nota:es una informacin importante acerca del instrumento o de la parte respectiva del manual, a la cual se debe atender especialmente.

Conformidad FCC
Slo para EE.UU.: Reglas FCC (Federal Communications Commission)
ADVERTENCIA: los cambios o modificaciones efectuados sin autorizacin previa de Siemens Milltronics pueden provocar la anulacin de la autorizacin del usuario a utilizar este aparato. Notas: El equipo se ha comprobado para garantizar su conformidad con los lmites aplicables a los aparatos digitales tipo Clase A, de acuerdo con la Seccin 15 de la reglamentacin FCC. Estos lmites estn diseados para proporcionar una proteccin razonable contra interferencias perjudiciales cuando el instrumento se usa en zonas residenciales. Este equipo genera, utiliza y puede radiar energa de frecuencia de radio. Al no observar las indicaciones para la instalacin proporcionadas en este manual el equipo puede provocar interferencias daosas en la comunicacin radio. El funcionamiento de este equipo en entornos residenciales puede provocar interferencias radio perjudiciales. El usuario es el nico responsable de corregir las interferencias.

SITRANS LR250
ADVERTENCIA: El SITRANS LR250 debe funcionar nicamente de la manera como se especifica en ste manual. La realizacin tcnica de estas condiciones es la condicin para una utilizacin sin peligro alguno. Nota: este aparato se ha diseado para el uso en mbito industrial. El uso de este instrumento en instalaciones residenciales puede causar interferencias a varias comunicaciones por radio. El radar pulsado a 2 hilos SITRANS LR250 mide el nivel de lquidos y lodos de forma continua y sin contacto en tanques de almacenamiento con presin y temperaturas extremas, en rangos hasta 20 m (66 ft). Es la solucin ideal para depsitos estrechos con productos con bajo dielctrico. El instrumento consta de una electrnica con una antena de bocina, con conexin al proceso bridada o roscada. Est disponible una antena roscada de PVDF. El SITRANS LR250 soporta el protocolo de comunicacin HART, y el software SIMATIC PDM. El procesamiento de seales se obtiene con los algoritmos exclusivos Process Intelligence.

Pgina ES-2

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

A5E31997170

Datos tcnicos

Para un listado completo, ver la versin completa del manual de instrucciones del SITRANS LR250 (HART). Para ms detalles acerca de las Homologaciones verHomologaciones pgina 4.

Temperatura ambiente y de operacin


Notas:
La mxima temperatura depende de la conexin al proceso, de los materiales de la antena y de la presin en el depsito: ver Tabla de temperaturas mximas de proceso pgina 23. Para informaciones ms detalladas ver curvas de reduccin de Presin de proceso/Temperatura en las instrucciones de servicio. Los lmites de temperatura y presin en el proceso pueden variar. Para ello vase la placa de caractersticas en el instrumento. El diagrama de referencia mencionado en la placa puede descargarse en: www.siemens.com/LR250, bajo Support/Installation drawings/ Level Measurement/Continuous - Radar/LR250. Para informaciones ms detalladas sobre las conexiones al proceso ver los diagramas en la pgina Installation Drawings, captulo Process Connection Diagrams. La amplitud de la seal aumenta conforme aumenta el dimetro de bocina. Utilizar el tamao de bocina ms grande posible. Existe la posibilidad de instalar extensiones opcionales por debajo de las conexiones roscadas. Para ms detalles ver Tabla de temperaturas mximas de proceso pgina 23.
temperatura ambiente (medida a proximidad de la caja) 40 C ... 80 C (40 F ... 176 F) placa de identificacin del equipo

Espaol

temperatura de proceso (medida en la conexin al proceso): antena de bocina estndar: - con junta trica de FKM: 40 oC ... +200 oC (-40 oF ... +392 oF) - con junta trica de FFKM: 20 oC ... +200 oC (4 oF ... +392 oF) antena roscada de PVDF, 2" NPT / BSPT / G: 40 ... +80 C (40 ... +176 F)

etiqueta en la conexin al proceso

Alimentacin
Uso general Seguridad intrnseca Sin chispa/Energa limitada No incendiario (FM/CSA EE.UU./Canad nicamente) A prueba de llamas Seguridad aumentada A prueba de explosiones (FM/CSA EE.UU./Canad nicamente) Mximo 30 V DC 4 20 mA Corriente mxima de arranque: Nominal 24 V DC a 550 ohmios

Nominal 24 V DC a 250 ohmios

ver Condiciones de arranque pgina 24

A5E31997170

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

Pgina ES-3

Homologaciones
Nota: Las homologaciones del instrumento estn indicadas en la placa de caractersticas. General CSAUS/C, FM, CE, C-TICK, KC RadiointerferenciaEuropa (R&TTE), FCC, Industry Canada Zonas con riesgo de explosin Seguridad intrnseca 1) (Europa) ATEX II 1G, EEx ia IIC T4 ATEX II 1D, EEx tD A20 IP67 T90 C (Internacional) IECEx SIR 05.0031X, Ex ia IIC T4, Ex tD A20 IP67 T90 C (EE.UU./Canad) FM/CSA: Clase I, Div. 1, Grupos A, B, C, D Clase II, Div. 1, Grupos E, F, G Clase III T4 (Brasil) INMETRO: DNV 12.0087 X Ex ia IIC T4 Ga Ex ia ta IIIC T90 C Da IP65/IP67 -40 C Ta +80 C DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-11:2009 ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

Espaol

(China)

NEPSI Ex ia IIC T4/ DIP A20 TA, T90 C IP67

Sin chispa/Energa limitada 2) (Europa) ATEX II 3 G Ex nA/nL IIC T4 Gc (China) NEPSI Ex nA II T4/ Ex nL IIC T4 No incendiario 3) (EE.UU./Canad) FM/CSA Clase I, Div. 2, Grupos A, B, C, D T5

A prueba de llamas 4) (Europa/Internacional) ATEX II 1/2 GD,1D, 2D IECEx SIR 08.0107X Ex dmbia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C (Brasil) INMETRO: DNV 12.0088 X Ex d ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017

ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-1:2009, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

1) 2) 3) 4)

Vase Conexiones elctricas de seguridad intrnseca pgina 9. Vase Cableado sin chispa/Energa limitada pgina 11. Vase Cableado no incendiario pgina 11. Vase Conexiones elctricas a prueba de llamas pgina 12.

Pgina ES-4

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

A5E31997170

Seguridad aumentada 1) (Europa/Internacional) ATEX II 1/2 GD,1D, 2D IECEx SIR 08.0107X Ex embia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C (Brasil) INMETRO: DNV 12.0088 X Ex e ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-7:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

Espaol

Zonas peligrosas Antideflagrante/Seguridad aumentada (China) NEPSI Ex dmbia IIC T4/ Ex embia IIC T4/ DIP A20 TA, T90 oC IP67 Proteccin contra explosiones 2) (EE.UU./Canad) FM/CSA Clase I, Div. 1, Grupos A, B, C, D Clase II, Div. 1, Grupos E, F, G Clase III T4 Instalaciones martimas Lloyds Register of Shipping Aprobacin tipo ABS (American Bureau of Shipping)

Aplicacin bajo presin


ADVERTENCIAS:

Es imprescindible no destornillar, retirar o desmontar la conexin al proceso o la caja de la electrnica mientras el contenido del depsito est bajo presin. El usuario deber utilizar tornillos y selladores conformes a los lmites de la brida y a su utilizacin, y que se adapten a las condiciones de servicio. La instalacin incorrecta puede provocar una cada de presin del proceso.
Notas: Es recomendable no separar la placa de caractersticas del conjunto protegido de la presin1). Tras sustituir el instrumento por otro equipo habr que retirar y volver a colocar la placa de caractersticas en la unidad de sustitucin. Los aparatos SITRANS LR250 se someten a controles de calidad que incluyen la realizacin de pruebas hidrostticas. Los resultados cumplen con (o exceden) los requisitos de ASME para calderas y recipientes a presin y la Directiva europea de equipos a presin.

1) 2)

Vase Conexiones elctricas de seguridad aumentada pgina 12. Vase Conexiones elctricas a prueba de explosiones (EE.UU./Canad nic.) pgina 13.

A5E31997170

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

Pgina ES-5

Directiva de equipos a presin, PED 97/23/CE


Los transmisores de nivel de Siemens con conexin al proceso bridada, roscada o sanitaria tipo clamp no vienen provistos de carcasas resistentes a la presin y por consiguiente estn excluidos del mbito de aplicacin de la Directiva de equipos a presin (aplicable a accesorios a presin y de seguridad). Para ms detalles ver la directriz de la Comisin Europea 1/8. 1)

Espaol

Instalacin
ADVERTENCIAS: El instrumento deber ser instalado nicamente por personal tcnico cualificado y siguiendo las normativas locales del pas. Manipular la caja del aparato y no tocar la placa de caractersticas para evitar daarla. La antena roscada de PVDF debe ser manipulada con mucha precaucin. Cualquier dao a la superficie de la antena, y en particular a la punta de la antena puede afectar el rendimiento. Los materiales de construccin son seleccionados en base a su compatibilidad qumica (o inertidad) para usos generales. Antes de instalar el aparato en ambientes especficos consulte las tablas de compatibilidad qumica. Notas:
En la Unin Europea y en los estados miembros, la instalacin del instrumento debe realizarse de acuerdo con la norma ETSI EN 302372. Para ms detalles acerca de las aprobaciones, vase por favor la placa indicadora en el producto. El nmero de serie marcado en cada conexin al proceso proporciona un nmero de identificacin nico en el que se indica la fecha de fabricacin. Ejemplo: MMDDAA XXX (MM = mes, DD = da, YY = ao y XXX= unidad secuencial fabricada) El instrumento puede incluir referencias suplementarias para indicar la configuracin, el tamao, la presin el material y las condiciones trmicas aplicables a la brida.

Diseo de la boquilla
Antena PVDF roscada
espacio libre mn.: 10 mm (0.4")

Antena de bocina de acero inoxidable


espacio libre mn.: 10 mm (0.4")

Dimetro mn. 51 mm

El extremo de la bocina debe sobresalir unos 10 mm (0.4) para evitar interferencias con la boquilla o tubuladura (y ecos perturbadores). El dimetro mnimo recomendado de la boquilla, o tubuladura, para antenas de PVDF es 51 mm (2"). La extensin (100 mm/3.93'') se adapta a todas las antenas excepto a la antena roscada de PVDF.

1)

El instrumento incluye piezas no sujetas a la presin en el proceso. Este conjunto de piezas ofrece una barrera efectiva contra prdidas de presin del tanque. El conjunto combina generalmente la conexin al proceso y el emisor pero no incluye la caja de la electrnica.

Pgina ES-6

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

A5E31997170

Ubicacin de la tubuladura
Evite montar el aparato en la parte central de depsitos altos y estrechos que pueden generar ecos perturbadores.

Espaol

Condiciones ambientales
Montar el aparato tomando en cuenta las caractersticas de la caja y los materiales de construccin. Evitar que el aparato est expuesto al sol. De lo contrario, instalar una pantalla protectora.
preferible desaconsejada

ngulo de dispersin del haz


El ngulo de dispersin se define como el ngulo en que la densidad de energa de las ondas de radar es la mitad del valor de la densidad de energa mxima. La densidad de energa mxima se encuentra alineada, frente a la antena. Las microondas que se emiten fuera del haz de seal pueden reflejarse en elementos estructurales y provocan interferencias.

cono de emisin ngulo de haz: antena roscada de PVDF = 19 bocina de 1.5" = 19 bocina de 2" = 15 ngulo bocina de 3" = 10 de haz bocina de 4" = 8

Cono de emisin
Mantener el cono de emisin lejos de fuentes de interferencia: peldaos, tubos, vigas en I, o corrientes de llenado.

Acceso para la programacin


El aparato debe situarse en un lugar visible, de fcil acceso (programacin con el calibrador).

Instrucciones de montaje
ADVERTENCIA: En aplicaciones a presin garantizar la unin hermtica de las roscas con una cinta de PTFE u otro sellador y apretar adecuadamente la conexin al proceso (el apriete manual es insuficiente). (El par de apriete mximo recomendado es de 40 N-m (30 ft-lbs.) Nota: El dispositivo soporta un nmero ilimitado de rotaciones.

Versiones roscadas
1) 2) Se recomienda comprobar que las roscas sean idnticas antes de colocar el dispositivo en la conexin de montaje (para evitar daarlas). Atornillar el aparato en la conexin de proceso roscada. Apriete a mano o con una llave. En el caso de presin en el depsito prestar atencin a la Advertencia arriba indicada.

A5E31997170

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

Pgina ES-7

Versiones bridadas
ADVERTENCIA: El usuario se compromete a utilizar tornillos y selladores conformes a los lmites de la brida y a su utilizacin, y que se adapten a las condiciones de servicio.

Espaol

Conexin elctrica
Alimentacin
ADVERTENCIAS: Para garantizar la conformidad con los requisitos de seguridad (IEC 61010-1), los bornes de conexin (CC) deben recibir la alimentacin elctrica de una fuente que provea aislamiento elctrico (entrada/salida). Aislar todos los cableados tomando en cuenta las tensiones utilizadas.

Conexin SITRANS LR250


ADVERTENCIAS: Consulte la etiqueta de identificacin para informacin sobre las homologaciones. Para mantener el grado de proteccin IP/NEMA se deben utilizar prensaestopas y sellos adecuados. Ver Cableado en zonas peligrosas pgina 9. Notas: Utilizar un cable de par trenzado: AWG 22 - 14 (0,34 mm2 - 2,5 mm2). Para la instalacin lectrica deben observarse las normas y disposiciones pertinentes. Pueden ser necesarios cables y conductos separados.
Utilizar una llave Allen 2 mm para aflojar el tornillo de presin en la tapa. tapn (IP 68) prensaestopas opcional1) 2) (o entrada de cables NPT2)) blindaje del cable (si necesario)

1) 2) 3) 4)

Pelar el extremo de la cubierta del cable unos 70 mm (2.75"), y pasar los cables a travs del prensaestopas. 1) 2) Los cables van conectados a los bornes como se indica: polaridad indicada en los bornes de conexin. El aparato va conectado a tierra observando las disposiciones pertinentes. Apretar el prensaestopas para conseguir un sellado ptimo.

1) 2)

Se puede suministrar junto con el dispositivo. Para tender cables en conductos utilice siempre prensaestopas certificados adecuados para garantizar la estanqueidad.

Pgina ES-8

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

A5E31997170

5)

Antes de la programacin o configuracin del dispositivo, comprobar que la tapa est cerrada.

Conexin HART
Espaol Configuracin tpica PLC/mA con HART 1) 2)
PLC activo alimentacin elctrica R= 250 Mdem HART SITRANS LR250 Comunicador HART

Notas: La alimentacin elctrica puede ser de dos tipos, segn el diseo del aparato: integrada o separada del PLC. Limitar la resistencia HART1) en conformidad con el rea de operacin admisible2). Slo se debe conectar una unidad de programacin HART a la vez.

Cableado en zonas peligrosas


Existen seis tipos de cableado para cubrir las necesidades de las zonas Ex. En cualquier caso, es recomendable verificar las aprobaciones indicadas en la placa indicadora y en la placa de caractersticas del instrumento.

1. Conexiones elctricas de seguridad intrnseca


Placa indicadora del dispositivo (ATEX/IECEX/INMETRO/C-TICK)

El certificado ATEX indicado en la placa de identificacin del equipo puede descargarse en la pgina web: www.siemens.com/LR250. Ver Support > Approvals / Certificates. El certificado IECEx referenciado en la placa de identificacin puede visualizarse en el sitio web IECEx. Ver: http://iecex.iec.ch, seleccione Ex Equipment Certificates of Conformity. Introduzca el nmero de certificado IECEx SIR 05.0031X.

1) 2)

Resistencia total del bucle = resistencia del cable + 250 ohmios (resistor). Ver Curva 1: Uso general, seguridad intrnseca, antichispa/energa limitada/no incendiario pgina 23 oCurva 2: A prueba de llamas, seguridad aumentada, a prueba de explosin pgina 24.

A5E31997170

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

Pgina ES-9

Conexiones elctricas de seguridad intrnseca (...)


Placa indicadora del dispositivo (FM/CSA) Espaol
El diagrama de conexiones seguridad intrnseca FM/CSA nmero 23650653 puede descargarse en la pgina web: www.siemens.com/LR250. Vase Support > Installation Drawings > Level Measurement > Continuous - Radar. Requisitos de alimentacin: verCurva 1: Uso general, seguridad intrnseca, antichispa/ energa limitada/no incendiario pgina 23. Requisitos relativos al cableado: observar la normativa local. Utilizar conductos aptos, resistentes a la intemperie y estancos al polvo y al agua para aplicaciones NEMA 4X / Tipo 4X / NEMA 6, IP67, IP68. Ver Instrucciones especficas para instalaciones en zonas peligrosas pgina 13. Barreras de seguridad intrnseca recomendadas, ver Barreras pasivas, diodo limitador pgina 10 y Barreras activas (barreras de repeticin) pgina 11. Nota: La seleccin de una tarjeta de entrada PLC, una alimentacin lectrica o una barrera apropiada requiere conocer la seguridad intrnseca y la aplicacin. El instalador es responsable de garantizar que la instalacin intrnsecamente segura sea conforme con los requisitos de certificacin del aparato y los cdigos de prctica aplicables.

Barreras pasivas, diodo limitador


Nota: Requiere tensin de alimentacin regulada. Fabricante
MTL MTL Stahl Stahl

Nmero de la pieza
787SP+ (Dos canales) 7787P+ (Dos canales) 9001/01-280-100-10 (Un canal) 9002/01-280-110-10 (Dos canales)

Los clculos siguientes permiten determinar la compatibilidad de la barrera con los parmetros de entrada intrinsecamente segura del LR250. Re-e = mx. resistencia de la barrera (extremo-a-extremo) Rloop = resistencia del bucle (resistencia del cable + resistencia adicional p.ej. entradas PLC o indicadores) Vbarrier = valor de cadas de tensin no lineales provocadas por la barrera 1) Determinar el valor Re-e en base a la hoja de datos. 2) Calcular el valor total Rloop. 3) Calcular Roperacin = Re-e + Rlazo.

Seleccin de una barrera pasiva para el SITRANS LR250

Pgina ES-10

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

A5E31997170

Seleccin de una barrera pasiva (...) 4) Determinar el valor Vbarrier en base a la hoja de datos de la barrera (p.ej. cadas de tensin provocadas por diodos). 5) Calcular Voperacin = Valimentacin Vbarrera. 6) Los valores Vworking y Rworking permiten confirmar si el instrumento funciona respetando los lmites indicados en el grfico (en gris) Curva 1: Uso general, seguridad intrnseca, antichispa/energa limitada/no incendiario pgina 23.

Espaol

Barreras activas (barreras de repeticin)


Fabricante
MTL MTL Siemens Stahl MTL MTL

Nmero de la pieza
706 7206 SITRANS I 7NG4124 9001/51-280-110-14 E02009 E02010

2. Cableado sin chispa/Energa limitada

El certificado ATEX indicado en la placa de identificacin del equipo puede descargarse en la pgina web: www.siemens.com/LR250. Ver: Support > Approvals / Certificates. Requisitos de alimentacin: verCurva 1: Uso general, seguridad intrnseca, antichispa/ energa limitada/no incendiario pgina 23. Realizar las conexiones elctricas observando la normativa local.

3. Cableado no incendiario

El diagrama de conexiones seguridad intrnseca FM/CSA Clase 1, Div 2, nmero 23650673 puede descargarse en la pgina web: www.siemens.com/ LR250. Vase Support > Installation Drawings > Level Measurement > Continuous - Radar. Requisitos de alimentacin: verCurva 1: Uso general, seguridad intrnseca, antichispa/ energa limitada/no incendiario pgina 23.

A5E31997170

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

Pgina ES-11

4. Conexiones elctricas a prueba de llamas


Espaol

El certificado ATEX indicado en la placa de identificacin del equipo puede descargarse en la pgina web: www.siemens.com/LR250. Ver: Support > Approvals / Certificates. El certificado IECEx referenciado en la placa de identificacin puede visualizarse en el sitio web IECEx. Ver: http://iecex.iec.ch, seleccione Ex Equipment Certificates of Conformity e introduzca el nmero de certificado IECEx SIR 08.0107X. Requisitos de alimentacin: verCurva 2: A prueba de llamas, seguridad aumentada, a prueba de explosin pgina 24. Realizar las conexiones elctricas observando la normativa local. Ver tambin Instrucciones especficas para instalaciones en zonas peligrosas pgina 13 y el certificado ATEX arriba mencionado.

5. Conexiones elctricas de seguridad aumentada

El certificado ATEX indicado en la placa de identificacin del equipo puede descargarse en la pgina web: www.siemens.com/LR250. Ver: Support > Approvals / Certificates. El certificado IECEx referenciado en la placa de identificacin puede visualizarse en el sitio web IECEx. Ver: http://iecex.iec.ch, seleccione Ex Equipment Certificates of Conformity e introduzca el nmero de certificado IECEx SIR 08.0107X. Requisitos de alimentacin: verCurva 2: A prueba de llamas, seguridad aumentada, a prueba de explosin pgina 24. Realizar las conexiones elctricas observando la normativa local. Ver tambin Instrucciones especficas para instalaciones en zonas peligrosas pgina 13 y el certificado ATEX arriba mencionado.

Pgina ES-12

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

A5E31997170

6. Conexiones elctricas a prueba de explosiones (EE.UU./Canad nic.)

Espaol

El diagrama de conexiones FM/CSA a prueba de explosin nmeroA5E02257843 puede descargarse en la pgina web: www.siemens.com/ LR250. Vase Support > Installation Drawings > Level Measurement > Continuous - Radar. Requisitos de alimentacin: verCurva 2: A prueba de llamas, seguridad aumentada, a prueba de explosin pgina 24.

Instrucciones especficas para instalaciones en zonas peligrosas


(Directiva europea ATEX 94/9/CE, Anexo II, 1/0/6)
Las instrucciones se refieren a dispositivos con certificacin de examen SIRA 06ATEX2358X, SIRA 08ATEX1301X y SIRA 09ATEX4153X. Para ms detalles acerca de la utilizacin y el montaje, ver el manual de instrucciones. El dispositivo cumple con las exigencias de la categora 1GD con certificacin SIRA 06ATEX2358X; categora 1/2 GD, 1D, 2D con certificacin SIRA 08ATEX1301X; y categora 3G con SIRA 09ATEX4153X. 3) El aparato puede ser utilizado en zonas con gases y vapores inflamables, con aparatos del Grupo IIC, IIB y IIA, clases de temperatura T1, T2, T3 y T4. 4) El aparato posee un ndice de proteccin IP67, clase de temperatura T90 C es eficaz en atmsferas con polvos inflamables. 5) El aparato puede soportar temperaturas ambientes de 40C a +80 C. 6) El aparato no se ha analizado como sistema de proteccin (como se indica en la Directiva 94/9/CE Anexo II, clusula 1.5): ver pgina 5. 7) La instalacin e inspeccin de este instrumento deben ser efectuadas por personal calificado en conformidad con los cdigos de prctica aplicables (EN 60079-14 y EN 60079-17 en Europa). 8) El aparato no se puede reparar. 9) El sufijo 'X' en el nmero de certificado indica la aplicacin condiciones de seguridad especficas. El personal responsable de la instalacin y la verificacin del instrumento debe tener acceso al contenido de los certificados. 10) El usuario es responsable de tomar las precauciones necesarias para evitar el dao del aparato y garantizar el nivel de proteccin obtenido, si existe la posibilidad de que est en contacto con productos agresivos. Productos agresivos: p.ej. : lquidos cidos o gases que pueden daar los metales, o disolventes que pueden daar los polmeros. Precauciones adecuadas: (p.ej.) confirmacin de la resistencia a productos qumicos especficos en base a la hoja de datos. 1) 2)

A5E31997170

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

Pgina ES-13

Programacin del SITRANS LR250


Espaol

Ver Asistente de arranque rpido (acceso con el programador) pgina 18. Vase Utilizacin del asistente de arranque rpido con SIMATIC PDM pgina 21.

Activacin del SITRANS LR250


Conectar la alimentacin elctrica. El SITRANS LR250 arranca automticamente en el modo de Medida.

Display LCD Indicacin de datos en el modo de medicin ( funcionamiento normal) 1) 2)


1 2 3 4
(((

1 indicador de visualizacin alternada1), unidades lineales % 2 funcionamiento seleccionado: nivel, espacio, distancia o volumen 3 valor medido (nivel o volumen, espacio, distancia) 4 unidades 5 grfico de barras, indicacin del nivel 6 zona secundaria indica (ajustable)2) temperatura de la electrnica, fiabilidad del eco, corriente de bucle, o distancia 7 zona de texto para visualizar mensajes de estado 8 indicador estado del aparato (para ms detalles ver el manual de instrucciones)

Indicadores de fallo S: 0 LOE


En caso de fallo, el cdigo de fallo correspondiente y un mensaje de error aparecen en la zona de texto (7) del indicador, y el icono mantenimiento necesario aparece en la zona estado del aparato (8)

Visualizacin en modo Program


Navegacin nmero men La aparicin de la barra de mens de tem actual actual indica que la lista de mens barra de excede la capacidad mxima del tem mens actual indicador. banda de tem Una banda intermedia aparece cuando el tem est situado a media altura de la lista. El ancho y la posicin de la banda tem en la barra de mens indican la longitud de la lista de mens y la posicin del tem seleccionado. Si la barra es ms ancha indica que la lista de tems es ms corta. Visualizacin de parmetros
nmero de parmetro valor/seleccin de parmetro

Edicin

nombre de parmetro

1) 2)

Pulsar la flecha UP (ARRIBA) o DOWN (ABAJO) para commutar entre valores. Al pulsar una tecla. Para ms detalles ver Funcin de las teclas en modo de medida pgina 15.

Pgina ES-14

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

A5E31997170

Programador porttil (ref. 7ML1930-1BK)


Vendido por separado. 1) 2) 3)

Espaol

Tecla Funcin de las teclas en modo de medida


Actualiza la corriente de bucle.1) Actualiza el valor correspondiente a la temperatura interna de la caja1). Actualizacin del valor de fiabilidad del eco1).
C

Actualiza la medida de distancia1). Mode permite acceder al modo PROGRAM2). Flecha DERECHA permite acceder al modo PROGRAM3). Flecha ARRIBA o ABAJO alterna la visualizacin unidades lineales / %.

Programacin con el programador porttil


Notas: Pulsar Mode para alternar entre el modo de medida (Measurement) y de programacin (Program). En modo PROGRAM, la salida se mantiene constante y no refleja las eventuales variaciones en el aparato. SITRANS LR250 vuelve automticamente al modo de medida despus un periodo de inactividad en modo PROGRAM (entre 15 segundos y 10 minutos en funcin del nivel de men).

Mens de parmetros

Los parmetros se identifican por su nombre y se dividen en grupos de funciones. Para ms detalles acerca de todos los parmetros consulte la versin completa del manual.

1. QUICK START 2. SETUP 2.1. DEVICE ............. 2.7. LINEARIZATION 2.7.1. VOLUME 2.7.1.1. VESSEL SHAPE

1) 2)

3)

Nuevo valor indicado en la zona secundaria del display LCD. Abre el ltimo nivel de men indicado en el actual ciclo de accionamiento, excepto si se ha rearrancado el equipo despus de salir del modo PROGRAM, o si han transcurrido ms de 10 minutos desde la ltima accin en modo PROGRAM. Aparece el men, nivel superior. Abre el men, nivel superior.

A5E31997170

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

Pgina ES-15

1.

Activar el modo PROGRAM


Orientar el programador hacia el indicador (mxima distancia 300 mm [1 ft.])

unidad de programacin porttil indicador

Espaol

La flecha DERECHA activa el modo PROGRAM y permite acceder al men, nivel 1. Pulsar Modo para abrir el nivel de men visualizado en modo PROGRAM durante los ltimos 10 minutos, o el men nivel 1 si se ha rearrancado el equipo.
Mx. 300 mm (1 ft)

2.

Navegacin: asignacin de las teclas en modo Navegacin

Nota: Para acceder directamente a los parmetros con el programador porttil, pulsar Home , e introducir el nmero de men, por ejemplo: 2.7.1 (Volumen).

Tecla

Nombre

Nivel de men

Funcin en modo navegacin


Visualizacin del men o parmetro precedente/ siguiente. Acceder al primer parmetro en el men seleccionado, o abrir el men siguiente. Acceder al modo Edit. Acceder al men padre. Acceder al modo MEASUREMENT. Abrir el men nivel principal: men 1.

Flecha ARRIBA o men o ABAJO parmetro

Flecha DERECHA Flecha IZQUIERDA


Modo Home

men parmetro men o parmetro men o parmetro men o parmetro

3.

Edicin en modo PROGRAM


Seleccionar una opcin de la lista a) Acceder al parmetro deseado. b) Pulsar la flecha DERECHA visualizar el parmetro. para
seleccin actual nombre de parmetro nmero de parmetro

c)

Pulsar nuevamente la flecha DERECHA para acceder al modo Edicin . Seleccin del valor actual. Visualizar otro valor.

d)

Pulsar la flecha DERECHA para confirmar. Se visualiza nuevamente el parmetro y el nuevo valor en el indicador de cristal lquido. SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA A5E31997170

Pgina ES-16

Modificar un valor numrico a) Acceder al parmetro deseado. b) Pulsar nuevamente la flecha DERECHA para acceder al modo Edicin. Seleccin del valor actual. Introducir otro nuevo. Pulsar la flecha DERECHA para confirmar. Se visualiza nuevamente el parmetro y el nuevo valor en el indicador de cristal lquido.

nombre de parmetro

nmero de parmetro

Espaol

c) d)

valor actual

Utilizacin de las teclas en modo Edicin Tecla Nombre


Flecha ARRIBA o ABAJO

Principales funciones del modo Edicin


Seleccin de opciones Edicin numrica Seleccin de opciones Selecciona el tem. - Incrementa o decrementa dgitos. - Alterna entre el signo positivo y negativo. - Confirmacin de datos (acceso de escritura/ parmetro). - Visualizacin alternada modo Edicin - Navegacin. - Mueve el cursor un espacio hacia la derecha - o con el cursor colocado en el signo Enter, acepta los datos y permite alternar entre el modo Edicin y el modo Navegacin. Anula el modo Edicin sin modificar el parmetro. - Mueve el cursor hacia el signo positivo/negativo si es la primera tecla pulsada - o mueve el cursor un espacio hacia la izquierda. Cancela el contenido del indicador. Introduce un punto decimal.

Flecha DERECHA
Edicin numrica Seleccin de opciones Edicin numrica Edicin numrica Edicin numrica

Flecha IZQUIERDA

Cancelar Punto decimal Signo + -

Edicin numrica

Modifica el signo del valor introducido.

Numrico

Edicin numrica

Introduce el carcter correspondiente.

A5E31997170

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

Pgina ES-17

Asistente de arranque rpido (acceso con el programador)


1. Arranque rpido Espaol
a) Apuntar la unidad porttil de programacin hacia el indicador (mx. distancia 300 mm (1 ft), y pulsar la flecha DERECHA acceder al men, nivel 1. b) Pulsar dos veces la flecha DERECHA parmetros. c) Pulsar la flecha DERECHA para acceder al modo Edicin la flecha ABAJO para acceder al tem men 1.1 y visualizar para activar el modo PROGRAM y

para aceptar valores predefinidos y acceder directamente al tem siguiente. d) e) Los valores se modifican accediendo a un tem o introduciendo un nuevo valor. Despus de modificar un valor pulsar la flecha DERECHA flecha ABAJO f) para acceder al tem siguiente. para confirmar y la

Los ajustes efectuados durante el Arranque rpido slo se implementan seleccionando Yes (S) en respuesta a Apply changes (Aplicar cambios) , en la etapa 1.8.

1.1. Idioma Define el idioma utilizado en la pantalla LCD (aplicable inmediatamente).


Opciones
ENGLISH, DEUTSCH, FRANAIS, ESPAOL

1.2. Material Opciones


LIQUID LIQUID LOW DK (lquido con bajo dielctrico)

1.3. Velocidad de reaccin Tiempo de reaccin del instrumento a cambios de nivel en el rango de medida.
Velocidad de reaccin
LENTO MEDIO RPIDO

Velocidad de llenado/vaciado
0,1 m/min (0.32 ft/min) 1,0 m/min (3.28 ft/min) 10,0 m/min (32.8 ft/min)

Seleccionar un valor ligeramente superior a la mxima velocidad de llenado/vaciado (valor ms alto).

1.4. Unidades Unidades de medida del instrumento. Opciones


M, CM, MM, FT, IN

Pgina ES-18

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

A5E31997170

1.5. Modo de funcionamiento


punto de referencia sensor espacio punto de referencia sensor

distancia punto de calibracin alto

Espaol

punto de referencia sensor

nivel punto de calibracin bajo

Funcionamiento
NO SERVICE NIVEL ESPACIO DISTANCIA
1)

Descripcin
La medida y la corriente de bucle asociada no se actualizan. El aparato conmuta al modo autoproteccin1). Distancia a la superficie del material referenciada al punto inferior de calibracin. Distancia a la superficie del material referenciada al punto superior de calibracin. Distancia a la superficie del material referenciada al punto de referencia del sensor.

Para ms detalles acerca del modo Fail-safe consulte las instrucciones de uso.

1.6. Punto inferior de calibracin Distancia del punto de referencia sensor al punto inferior de calibracin: generalmente el nivel proceso vaco. Ver Modo de funcionamiento (1.5.) para ms detalles. Valores
Rango: 0,00 20,00 m

1.7. Punto superior de calibracin Distancia del punto de referencia sensor al punto superior de calibracin: generalmente el nivel proceso lleno. Ver Modo de funcionamiento (1.5.) para ms detalles. Valores
Rango: 0,00 20,00 m

1.8. Apply? (Aplicar modificaciones ?) Los ajustes realizados durante el Arranque Rpido slo se implementan seleccionando Yes (S). Opciones
Pulsar Mode YES (SI), NO, DONE (FIN) (en el indicador aparece DONE al final de la Puesta en marcha rpida.) para volver al modo de medida. El SITRANS LR250 est listo para funcionar.

Nota: En caso de obstrucciones en el depsito, consultar la versin completa de las instrucciones de servicio para informaciones ms detalladas acerca de la autosupresin de falsos ecos.

A5E31997170

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

Pgina ES-19

Comunicacin SITRANS LR250: HART


Espaol

Para informaciones ms detalladas acerca de los parmetros, ver la versin completa de las instrucciones de servicio. Recomendamos parametrizar el aparato utilizando el software SIMATIC Process Device Manager (PDM). Las guas de aplicacin para la configuracin de aparatos HART con SIMATIC PDM estn disponibles en la pgina Producto de nuestro sitio web: www.siemens.com/LR250 bajo Support.

SIMATIC PDM
SIMATIC PDM es una herramienta universal destinada a poner en servicio y mantener los aparatos SITRANS LR250 y otros instrumentos de proceso. Para ms detalles, consulte las instrucciones de servicio del SIMATIC PDM o la ayuda en lnea. (Para ms detalles ver www.siemens.com/simatic-pdm.)

Desactivar memoria intermedia

Permite alinear SIMATIC PDM con un mdem HART para los sistemas operativos Windows 2000 y Windows XP 1). Notas: SIMATIC PDM solo es compatible con Windows XP Professional, no con la versin Home. La desactivacin de la memoria intermedia (buffer) solo puede ser realizada por usuarios con derechos administrativos. 1) 2) 3) 4) 5) 6) Hacer clic en Start/Settings/Control Panel para empezar la configuracin. Hacer doble clic en System, seleccionar la fichaHardware, y hacer clic en el botn Device Manager. Abrir Ports y hacer doble clic en el puerto COM utilizado por el sistema para acceder a la ventana Communications Port Properties. Seleccionar la ficha Port Settings y hacer doble clic en el botn Advanced. Deseleccionar la casillaUse FIFO buffers si es necesario. Hacer clic en OK para cerrar. Cerrar todas las pantallas y reinicializar el instrumento.

Actualizar Electronic Device Description (EDD).


Notas:
SIMATIC PDM Rev. 5.2 SP1 slo se utiliza para las funciones bsicas de configuracin y diagnstico. Las funciones avanzadas como el Asistente de arranque rpido requieren la versin 6.0 SP4 (o ulterior). El fichero EDD est disponible en el Catlogo de aparatos (Device Catalog) SIMATIC PDM, en Sensors/Level/Echo/Siemens Milltronics/SITRANS LR250. Compruebe si hay actualizaciones ms recientes en la pgina Producto de nuestra web: www.siemens.com/LR250, para obtener la ltima versin de SIMATIC PDM, el SP (Service Pack) y H (hot fix) ms recientes. Vase Support > Software Downloads.

Para instalar un nuevo EDD:


1) 2) 3) Descargar el EDD en la pgina producto de nuestra web: www.siemens.com/LR250 y guardar los ficheros en el PC. Extraer el contenido del archivo zip al emplazamiento seleccionado. Ejecutar SIMATIC PDM Manager Device Catalog, seleccionar el archivo EDD no comprimido.
Windows es una marca registrada de Microsoft Corporation.

1)

Pgina ES-20

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

A5E31997170

Configuracin de equipos nuevos


Nota: Haciendo clic en Cancel/Anular mientras se cargan los datos equipo-SIMATIC PDM
pueden actualizarse automticamente algunos parmetros. 1) 2)

Espaol

Comprobar que se dispone del fichero EDD ms reciente, y actualizar si es necesario (ver Actualizar Electronic Device Description (EDD). pgina 20). Ejecutar SIMATIC Manager y crear un proyecto nuevo para el LR250. Las guas de aplicacin para la configuracin de aparatos HART con SIMATIC PDM estn disponibles en la pgina Producto de nuestro sitio web: www.siemens.com/LR250. Abrir el men Dispositivo Reset Maestro y hacer clic en OK para reinicializar a los valores de fbrica. Despus de la reinicializacin transmitir los parmetros al PC/PG. Configurar el equipo con el Asistente de Arranque rpido.

3) 4) 5)

Utilizacin del asistente de arranque rpido con SIMATIC PDM


Notas:
Los ajustes del Asistente de Arranque Rpido estn interrelacionados. Para aplicar modificaciones, hacer clic en FINISH (TERMINAR) y DOWNLOAD (DESCARGAR) al final de la etapa 5 para almacenar ajustes y transferirlos al aparato. Para corregir un valor, hacer clic en BACK (PRECEDENTE). Seleccionar Cancel (Anular) para salir de la secuencia de arranque rpido. Ejecutar el programa SIMATIC PDM, abrir el men Device Wizard - Quick Start, y seguir los pasos 1 - 5.

Arranque rpido

Utilizacin y manejo con el interfaz FDT (Field Device Tool)


El estndar FDT es compatible con numerosos paquetes de software utilizados para la puesta en servicio y el mantenimiento de dispositivos de campo. PACTware y Fieldcare son ejemplos de soluciones basadas en la tecnologa FDT. La configuracin de un dispositivo de campo con FDT requiere la tecnologa DTM (Device Type Manager). Los instrumentos de Siemens utilizan la herramienta de configuracin SIMATIC DTM con la tecnologa EDD para SITRANS DTM. 1) El primer paso consiste en instalar SITRANS DTM. Est disponible en: http://support.automation.siemens.com. Hacer clic en Product Support y acceder a Product Information/Automation Technology/Sensor systems/Process Instrumentation/ Software & Communications. 2) Instalar el EDD SITRANS LR250 HART para SITRANS DTM. La Descripcin de Dispositivo Electrnica (EDD) est disponible en: www.siemens.com/LR250. Vase Support > Software Downloads. A5E31997170 SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA Pgina ES-21

Configuracin de equipos nuevos con FDT


La gua de aplicacin est disponible en la pgina web, bajo Support.

Espaol

Interfaz con AMS Device Manager


El software AMS Device Manager est diseado para la puesta en servicio y el mantenimiento de dispositivos de campo. Para ms detalles acerca de AMS Device Manager, consulte las instrucciones de servicio o la ayuda en lnea. Vase: http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD)


Para su funcionamiento SITRANS LR250 requiere el EDD para la versin 9.0 de AMS Device Manager.

Configuracin de equipos con AMS Device Manager


1) Compruebe si hay actualizaciones ms recientes en la pgina Producto de nuestra web: www.siemens.com/LR250 para descargar el ltimo EDD. Ver Support > Software Downloads y descargar el EDD en caso necesario. Guardar los datos en el ordenador y extraer el archivo zip, dejndolo fcilmente accesible. Ejecutar AMS Device Manager Add Device Type, y seleccionar el EDD no comprimido.

2)

Ejecutar AMS Device Manager. El usuario dispone de una Gua de aplicacin para configurar equipos HART con AMS Device Manager, descargable de nuestro sitio web bajo Support.

Mantenimiento
Al funcionar en condiciones normales el SITRANS LR250 no necesita mantenimiento o limpieza. Sin embargo, en condiciones de operacin difciles se recomienda: 1) 2) Utilizar un producto de limpieza compatible con el material de construccin de la antena, y que no contamine el medio a medir. Retirar el equipo del depsito y aplicar el producto de limpieza con un pao en la superficie sucia de la antena.

Reparaciones y lmite de responsabilidad


Para ms detalles vase la portada interior.

Sustitucin de la antena o electrnica/caja


La antena puede retirarse para ser sustituida. No es necesario volver a calibrar el instrumento. La antena se puede sustituir con otro tipo de antena contactando con un centro de reparacin Siemens autorizado. Al sustituir una electrnica o una caja que presenta daos o averas, por una unidad igual es imprescindible comprobar la compatibilidad de la antena utilizada. De lo contrario, ser necesario volver a calibrar el instrumento contactando con un centro Siemens autorizado.

Pgina ES-22

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

A5E31997170

Tabla de temperaturas mximas de proceso


LR250 con adaptador de brida
Espaol
ADVERTENCIA: la temperatura interna no debe sobrepasar 80 C (176 F). Notas: Este grfico es de carcter informativo y no representa los distintos tipos de conexin al proceso. (NO se aplica p.ej. colocando el SITRANS LR250 directamente en una superficie metlica del tanque/depsito.) La exposicin directa al sol no se ha tomado en cuenta en el grfico. El Parmetro 3.2.1, Temperatura interna actual controla la temperatura interna. Mx. temperaturas de proceso/temperatura ambiente admisible

Temperatura ambiente (oC)

temperatura de proceso

Temperatura de proceso (oC)

Alimentacin de bucle
Requisitos aplicables a la alimentacin elctrica
Nota: Las curvas a continuacin se refieren a instrumentos autnomos, configurados con el programador manual comercializado por Siemens. Curva 1: Uso general, seguridad intrnseca, antichispa/energa limitada/no incendiario

Resistencia del bucle RL

ZONA DE FUNCIONAMIENTO AUTORIZADA

Tensin del bucle VL

Nota: La comunicacin HART requiere una tensin mnima 16,3 VDC (220 Ohms (RL)).

A5E31997170

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

Pgina ES-23

Resistencia del bucle RL

Espaol

Curva 2: A prueba de llamas, seguridad aumentada, a prueba de explosin

ZONA DE FUNCIONAMIENTO AUTORIZADA

Tensin del bucle VL

Nota: La comunicacin HART requiere una tensin mnima 20.94 VDC (220 Ohms (RL)).

Condiciones de arranque
Consumo del equipo es mnimo durante el arranque (inferior a 3,6 mA). Un lapso de tiempo inferior a 50 segundos precede la primera medida.

Pgina ES-24

SITRANS LR250 (HART) PUESTA EN MARCHA RPIDA

A5E31997170

SITRANS LR250 (HART) Manuel de mise en service rapide


Ce document dcrit les caractristiques et les fonctions essentielles du SITRANS LR250 (HART 1)). Il est fortement recommand de se reporter la version intgrale des instructions de service pour accder lensemble des fonctions. La version complte du manuel est disponible sur la page produit SITRANS LR250, sur le site : www.siemens.com/LR250. Les versions imprimes de la documentation sont fournies par les reprsentants Siemens Milltronics. Toute question sur le contenu de ce document peut tre adresse : Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 veuillez nous crire : techpubs.smpi@siemens.com

Franais

Copyright Siemens AG 2013. Tous droits rservs


Nous encourageons les utilisateurs se procurer les exemplaires imprims de ces manuels ou les versions lectroniques prpares et valides par Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments ne pourra tre tenu responsable du contenu de toute reproduction totale ou partielle des versions imprimes ou lectroniques.

Clause de non-responsabilit
Le contenu de ce document a t vrifi pour garantir la conformit avec les caractristiques du systme. Des divergences tant possibles, nous ne pouvons en aucun cas garantir la conformit totale. Ce document est rvis et actualis rgulirement pour inclure les nouvelles caractristiques. Nhsitez pas nous faire part de vos commentaires. Sous rserve de modifications techniques.

MILLTRONICS est une marque dpose de Siemens Milltronics Process Instruments.

Assistance technique
L'assistance technique est disponible 24h/24. Vous trouverez les coordonnes de l'interlocuteur Siemens Automatisation le plus proche l'adresse : www.siemens.com/automation/partner : Slectionner d'abord la langue d'affichage (franais). Cliquer sur l'onglet Contact, slectionner Prestations, puis cliquer sur Prestations de service et chercher la branche souhaite (+Automation Technology > +Sensor Systems > Instrumentation de processus > +Mesure de niveau > +Continuous). Slectionner Radar. Slectionner un pays, puis une ville/rgion. Slectionner Technical Support sous Service. Vous trouverez l'assistance technique en ligne l'adresse : www.siemens.com/automation/ support-request Indiquez le nom du produit (SITRANS LR250) ou le numro de commande et cliquez sur Rechercher, puis slectionnez le type de produit. Cliquer sur Suite. Entrer un mot cl pour la requte. Accder la documentation souhaite en ligne ou cliquez sur Suivant pour envoyer un courriel aux spcialistes de l'Assistance Technique Siemens. Assistance Technique Siemens IA/DT : Tl. +49 (0)911 895 7222
1)

HARTest une marque dpose de HART Communication Foundation.

A5E31997170

SITRANS LR250 (HART) MISE EN SCE RAPIDE

Page FR-1

Consignes de scurit

ll est important de respecter les consignes fournies dans ce document afin de garantir la scurit de l'utilisateur ou de tiers et la protection de l'instrument ou de tout quipement connect ce dernier. Chaque avertissement s'associe une explication dtaille du niveau de prcaution recommand pour chaque opration. Le symbole d'AVERTISSEMENT fait rfrence une mention sur le produit. Il signifie que la mort, des blessures corporelles graves et/ou des dommages matriels consquents peuvent se produire si les dispositions de scurit correspondantes ne sont pas respectes. Le symbole d'AVERTISSEMENT est utilis lorsqu'il n'y a pas de mention correspondante sur le produit. Il signifie que la mort, des blessures corporelles graves et/ou des dommages matriels consquents peuvent se produire si les dispositions de scurit correspondantes ne sont pas respectes.

Franais

N.B. : information importante concernant linstrument ou une section particulire de la notice dutilisation.

Conformit FCC
U.S.A. uniquement : Rgles FCC (Federal Communications Commission)
AVERTISSEMENT : tout(e) changement ou modification effectu(e) sans laccord pralable de Siemens Milltronics peut remettre en cause les droits dutilisation de l'quipement. N.B. : Cet instrument est conforme aux limites imposes (alina 15 de la rglementation FCC, Classe A). Cette rglementation vise assurer une protection suffisante contre les interfrences nocives lorsque l'instrument est utilis dans un environnement commercial. Cet instrument gnre, utilise et peut mettre de lnergie frquence radio. Pour cette raison, il doit tre install et utilis suivant les instructions fournies, pour viter toute interfrence nocive aux communications radio. En cas d'utilisation en zone rsidentielle cet instrument peut provoquer des interfrences radio nocives. Il appartiendra lutilisateur de prendre en charge les consquences de ces dysfonctionnements.

SITRANS LR250
AVERTISSEMENT : pour garantir la scurit, le SITRANS LR250 doit tre utilis suivant les consignes fournies dans ce manuel d'utilisation. N.B. : cet instrument est conu pour une utilisation en milieu industriel. En cas d'utilisation en zone rsidentielle, cet instrument peut provoquer diffrentes interfrences radio. Le SITRANS LR250 est un radar 25 GHz impulsions, en technologie 2 fils. Conu pour la mesure en continu de liquides et de boues liquides dans les cuves de stockage prsentant pression et tempratures leves, il convient parfaitement aux plages de mesure jusqu 20 m (66 ft). Cet instrument est idal pour les petites cuves ou les produits faible constante dilectrique. Cet instrument se compose d'une lectronique associe une antenne conique et un raccord process (filetage ou bride). Une antenne en PVDF avec raccord filet est disponible galement. Le SITRANS LR 250 est compatible avec le protocole de communication HART et le logiciel SIMATIC PDM. Le logiciel Process Intelligence assure le traitement des signaux.

Page FR-2

SITRANS LR250 (HART) MISE EN SERVICE RAPIDE

A5E31997170

Caractristiques techniques

Pour une liste exhaustive se reporter au manuel dutilisation SITRANS LR250 (HART). Pour plus de dtails sur les Homologations se reporter Homologations page 4.

Temprature ambiante/de fonctionnement


N.B. :
La temprature maximale varie en fonction du raccord process, du type dantenne et de la pression dans la cuve/ Cf. Temprature de process maximum page 23. Pour plus de dtails se reporter aux Courbes de limitation de la Pression/Temprature de process, fournies dans les instructions de service. La temprature et la pression de process varient en fonction des informations reportes sur la plaque d'identification de l'instrument. Le schma de rfrence mentionn est disponible sur la page produit, sous www.siemens.com/LR250, Support/Installation drawings/Level Measurement/Continuous - Radar/LR250. Pour plus de dtails sur les raccords process se reporter Installation Drawings, Continuous - Radar. Lamplitude du signal est proportionnelle au diamtre du cne. Il est donc prfrable dutiliser une antenne qui soit le plus large possible. Les antennes peuvent tre quipes de rallonges en option. Pour plus de dtails se reporter Temprature de process maximum page 23.
temprature ambiante (autour du botier) 40 C 80 C (40 F 176 F) plaque signaltique

Franais

temprature de process (raccord process) : antenne conique standard : - avec joint torique FKM : 40 oC ... +200 oC (40 oF ... +392 oF) - avec joint torique FFKM : 20 oC ... +200 oC (4 oF ... +392 oF) antenne PVDT filete 2" NPT / BSPT / G : 40 ... +80 C (40 ... +176 F)

plaque fixe au raccord process

Alimentation
Usage gnral Scurit intrinsque Anti-tincelles/ nergie limite Non-Ex (FM/CSA USA/Canada uniquement) Enveloppe antidflagrante Scurit augmente Antidflagrant (FM/CSA USA/Canada uniquement) : 30 V CC max. 4 20 mA Courant de dmarrage max.: 24 V CC nominale, 550 Ohm

24 V CC nominale, 250 Ohm

cf. Comportement au dmarrage page 24

A5E31997170

SITRANS LR250 (HART) MISE EN SCE RAPIDE

Page FR-3

Homologations
N.B. : les homologations applicables sont reportes sur la plaque de l'appareil. Usage gnralCSAUS/C, FM, CE, C-TICK, KC Radio Europe (R&TTE), FCC, Industry Canada Zone dangereuse Scurit intrinsque 1) (Europe) ATEX II 1G, EEx ia IIC T4 ATEX II 1D, EEx tD A20 IP67 T90 C (International) IECEx SIR 05.0031X, Ex ia IIC T4, Ex tD A20 IP67 T90 C (USA/Canada) FM/CSA Classe I, Div. 1, Gr. A, B, C, D Classe II, Div 1, Gr. E, F & G Classe III T4 (Brsil) INMETRO: DNV 12.0087 X Ex ia IIC T4 Ga Ex ia ta IIIC T90 C Da IP65/IP67 -40 C Ta +80 C DNV #OCP 0017

Franais

ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

(Chine)NEPSI Ex ia IIC T4/ DIP A20 TA, T90 C IP67 Anti-tincelles/ nergie limite 2) (Europe) ATEX II 3 G Ex nA/nL IIC T4 Gc (Chine) NEPSI Ex nA II T4/ Ex nL IIC T4 NI (non-incendive) 3) (USA/Canada) FM/CSA Classe I, Div. 2, Groupes A, B, C, D T5

Enveloppe antidflagrante 4) (Europe/International) ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex dmbia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C Brsil) INMETRO: DNV 12.0088 X Ex d ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-1:2009, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

1) 2) 3) 4)

Se reporter Cblage scurit intrinsque page 9. Se reporter Cblage Anti-tincelles/ nergie limite page 11. Se reporter Cblage Non-incendiaire page 11. Se reporter Scurit antidflagrante page 12.

Page FR-4

SITRANS LR250 (HART) MISE EN SERVICE RAPIDE

A5E31997170

Homologations (suite) Scurit augmente 1) (Europe/International)

(Brsil)

ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex embia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C INMETRO: DNV 12.0088 X Ex e ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-7:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

Franais

Antidflagrant/Scurit augmente (Chine) NEPSI Ex dmbia IIC T4/ Ex embia IIC T4/ DIP A20 TA, T90 oC IP67 Antidflagrant 2) (USA/Canada) FM/CSA Classe I, Div. 1, Gr. A, B, C, D Classe II, Div 1, Gr. E, F & G Classe III T4

Applications maritimes Lloyd's Register of Shipping Certification ABS (American Bureau of Shipping)

Applications avec pression


AVERTISSEMENTS :

Ne jamais dvisser, retirer ou dmonter le raccord process ou le botier de l'instrument lorsque lintrieur de la cuve est sous pression. Il incombe l'utilisateur de choisir le matriel de fixation (boulonnerie, tanchit) adapt aux spcifications de la bride et aux conditions de service attendues dans l'application. L'installation incorrecte peut entraner une perte de pression dans le process.
N.B. : Ne pas dissocier la plaque d'identification de lensemble non expos la pression de process1). En cas de substitution de lensemble (instrument), veiller attacher la plaque d'identification sur l'unit de remplacement. Les appareils SITRANS LR250 sont soumis des tests hydrostatiques, et prsentent des rsultats conformes ou suprieurs aux exigences tablies par le Code sur les chaudires et appareils pression de l'ASME, et par la Directive europenne relative aux quipements sous pression.

1) 2)

Se reporter Scurit augmente page 12. Se reporter Cblage antidflagrant (USA/Canada uniquement) page 13.

A5E31997170

SITRANS LR250 (HART) MISE EN SCE RAPIDE

Page FR-5

Directive Europenne 97/23/CE concernant les quipements sous pression


Les transmetteurs de niveau Siemens quips de raccords process type bride, filet ou clamp sanitaire ne sont pas quips d'un botier rsistant la pression et ne rentrent pas dans le cadre de la Directive Equipements sous Pression en tant qu'accessoires de pression ou de scurit (voir Directive de la Commission UE 1/8). 1)

Installation
AVERTISSEMENTS : L'installation doit tre effectue par un personnel qualifi, en accord avec les dispositions rglementaires locales. Pour viter toute dtrioration, il est prfrable de manipuler l'instrument en le tenant par le botier, sans toucher la plaque d'identification. Redoubler de prcaution lors de toute manipulation de l'antenne PVDF filete. Tout endommagement de la surface de l'antenne, et notamment de la pointe, peut affecter la performance. Cet instrument est conu avec des matriaux choisis en fonction de leur compatibilit chimique, pour une exploitation gnrale. Se reporter aux tableaux de compatibilit chimique avant toute utilisation dans un environnement spcifique. N.B. :
L'installation dans les pays de l'Union Europenne doit tre ralise en conformit avec la norme ETSI EN 302372. Les agrments sont indiqus sur l'tiquette de l'appareil. Les numros de srie imprims sur les raccords process correspondent un seul numro didentification qui indique la date de fabrication. Ex. : MMJJAA XXX (MM = mois, JJ = jour, AA = anne, et XXX= unit sq. fabrique) L'appareil peut comporter des marquages supplmentaires : type de bride, dimensions, pression applicable, matriau et conditions thermiques.

Franais

Type de piquage
Antenne et raccord filet en PVDF
dgagement min. : 10 mm (0.4")

Antenne conique en acier inoxydable


dgagement min. : 10 mm (0.4")

Diamtre min. 51 mm

Pour viter toute interfrence (chos parasites), veillez ce lextrmit de l'antenne dpasse le piquage de 10 mm (0.4) minimum. Le diamtre minimum recommand du piquage utilis pour le montage d'une antenne avec raccord filet PVDF est 51 mm (2"). A l'exception de l'antenne avec raccord filet en PVDF, toutes les antennes peuvent tre quipes d'une rallonge de 100 mm (3.93").

1)

L'instrument comporte un ensemble de pices non exposes la pression du process. Cet ensemble sert de barrire contre une ventuelle perte de pression dans la cuve. Il associe le corps du raccord process et lmetteur mais exclut gnralement le botier de llectronique.

Page FR-6

SITRANS LR250 (HART) MISE EN SERVICE RAPIDE

A5E31997170

Positionnement du piquage
L'installation de l'instrument au centre du toit d'un cuve haute et troite doit tre vit car il peut donner lieu des chos parasites.

Conditions environnementales
Assurer des conditions de service compatibles avec lindice de protection du botier et les matriaux de construction. Prvoir une protection spciale lorsque l'instrument est expos au soleil.
conseill dconseill

Angle du faisceau
L'angle du faisceau d'mission correspond la largeur du cne, o la densit d'nergie est rduite de moiti par rapport la densit d'nergie maximale. La densit d'nergie est maximale devant l'antenne, dans le prolongement de celle-ci. La transmission d'un signal extrieur l'angle du faisceau fait obstacle et entrane des chos parasites.

Franais

cne dmission angle du faisceau : antenne et raccord filet en PVDF= 19 cne 1.5" = 19 cne 2" = 15 angle cne 3" = 10 cne 4" = 8 du faisceau

Cne d'mission
Le faisceau d'mission doit tre dgag de toute source d'interfrence : chelles, tuyaux, poutrelles, ou points de remplissage.

Accs pour la programmation


Veiller ce que l'instrument soit facilement accessible pour la visualisation et la programmation.

Instructions pour le montage


AVERTISSEMENT : en cas de pression dans lapplication, prvoir un ruban PTFE ou un produit dtanchit quivalent, associ un vissage plus serr du raccord process. (Couple de serrage max. : 40 N-m (30 ft.lbs.) N.B. : L'appareil peut supporter un nombre illimit de rotations, sans risque d'endommagement.

Versions filetes
1) 2) Avant de visser l'appareil sur le raccord de montage, vrifier que les filetages soient identiques pour viter de les endommager. Visser l'instrument sur le raccord process. Serrer manuellement ou avec une cl adapte. En prsence de pression dans l'application, se reporter l'Avertissement cidessus.

A5E31997170

SITRANS LR250 (HART) MISE EN SCE RAPIDE

Page FR-7

Versions bride
AVERTISSEMENT : le boulonnage et les joints choisis par l'utilisateur doivent permettre de respecter les consignes et les limites d'utilisation de la bride, et s'adapter aux conditions de fonctionnement.

Cblage
Alimentation
AVERTISSEMENTS : Les bornes d'entre CC doivent tre alimentes par une source mme de fournir l'isolation lectrique entre/sortie requise pour la conformit avec les rgles de scurit de la norme CEI 61010-1.

Franais

Isoler tous les cblages en tenant compte des tensions nominales.

Raccordement du SITRANS LR250


AVERTISSEMENTS : Vrifier les homologations indiques sur la plaque signaltique de l'instrument. Pour garantir la protection IP/NEMA veiller l'tanchit des conduits. Cf. Instructions relatives au cblage pour l'installation en zone dangereuse page 9. N.B. : Utiliser un cble paire torsade : AWG 22 - 14 (0,34 mm2 - 2,5 mm2). Des cbles et des conduits spars peuvent tre ncessaires pour garantir la conformit avec les consignes de cblage ou les normes lectriques.
Desserrer les vis de pression du capot l'aide d'une cl Allen 2 mm bouchon (IP 68) presse-toupe optionnel1) 2) (ou entre de cble NPT2)) blindage cble (si utilis)

1) 2) 3) 4) 5)

Retirer environ 70 mm (2.75") de gaine l'extrmit du cble. Faire passer les cbles travers le presse-toupe. 1) 2) Connecter les cbles au bornier tel quillustr : la polarit est indique sur le bornier. Effectuer la mise la terre suivant les prescriptions locales. Serrer le presse-toupe pour garantir ltanchit. Fermer le couvercle avant toute programmation et configuration de l'instrument.

1) 2)

Peut tre livr avec l'instrument. Si le cble est install sous conduit, utiliser des bouchons tanches adapts.

Page FR-8

SITRANS LR250 (HART) MISE EN SERVICE RAPIDE

A5E31997170

Connexion HART
Configuration API/mA avec HART (standard) 1) 2)
API actif alimentation lectrique R= 250 Modem HART SITRANS LR250 Communicateur HART

N.B. : Deux types d'alimentation sont possibles, suivant la configuration : intgre, ou spare de l'API. La rsistance HART1) doit tre limite conformment la zone de fonctionnement autorise2). Connecter un seul programmateur HART la fois.

Franais

Instructions relatives au cblage pour l'installation en zone dangereuse


Il existe six possibilits de cblage pour l'installation en zone dangereuse. Dans tous les cas il est impratif de vrifier les homologations indiques sur la plaque signaltique et l'tiquette de l'appareil.

1. Cblage scurit intrinsque


Plaque signaltique de linstrument (ATEX/IECEX/INMETRO/C-TICK)

Les certificats ATEX indiqus sur la plaque signaltique sont disponibles sur notre site web : www.siemens.com/LR250. Consulter Support > Approvals / Certificates. Le certificat IECEx indiqu sur la plaque signaltique est accessible partir du site internet IECEx. Consulter : http://iecex.iec.ch, cliquer sur Ex Equipment Certificates of Conformity, puis entrer le numro du certificat IECEx SIR 05.0031X.

1) 2)

Rsistance totale de la boucle = rsistance du cble + 250 Ohm (rsistance). Se reporter Courbe 1 : Usage gnral, scurit intrinsque, anti-tincelles/ nergie limite, protection non-incendiaire page 23 ou Courbe 2 : Antidflagrant, scurit augmente, protection contre les explosions page 24.

A5E31997170

SITRANS LR250 (HART) MISE EN SCE RAPIDE

Page FR-9

Cblage scurit intrinsque (suite)


Plaque signaltique de linstrument (FM/CSA)

Le schma de raccordement FM/CSA Scurit Intrinsque numro 23650653 est disponible sur la page produit de notre site : www.siemens.com/LR250. Consulter Support > Installation Drawings > Level Measurement > Continuous - Radar. Pour plus de dtails sur la consommation de courant, voir Courbe 1 : Usage gnral, scurit intrinsque, anti-tincelles/ nergie limite, protection non-incendiaire page 23. Exigences relatives au cblage : respecter les prescriptions locales. Les conduits utiliss doivent tre tanches la poussire et leau pour les applications en extrieur (NEMA 4X / type 4X / NEMA 6, IP67, IP68). Se reporter Instructions spcifiques pour linstallation en zone dangereuse page 13. Les barrires de scurit intrinsque recommandes sont indiques sous Barrires passives diodes shunt page 10 et Barrires actives (barrires de rptition) page 11. N.B. : la slection d'un module d'entre API, d'une alimentation lectrique ou d'une barrire appropri(e) ncessite de connatre les principes de scurit intrinsque et l'application envisage. L'installateur doit s'assurer que la scurit intrinsque de l'installation soit conforme aux exigences d'agrment de l'instrument ainsi qu'aux rgles et normes d'utilisation applicables.

Franais

Barrires passives diodes shunt


N.B. : requiert une tension dalimentation bien rgule. Fabricant
MTL MTL Stahl Stahl

Numro de pice
787SP+ (Double) 7787P+ (Double) 9001/01-280-100-10 (Simple) 9002/01-280-110-10 (Double)

Pour vrifier que la barrire de scurit soit conforme aux paramtres de scurit intrinsque du LR250, effectuer les calculs suivants : Re-e = max. rsistance de la barrire de bout en bout Rboucle = rsistance de la boucle (rsistance totale du cble et toute rsistance srie supplmentaire, entres API et/ou affichages par ex.) Vbarrire = valeur d'une chute de tension non linaire associe la barrire 1) Dfinir la valeur Re-e (cf. fiche technique). 2) Calculer la valeur totale de Rboucle. 3) Calculer Rfonct = Re-e + Rboucle.

Procdure pour vrifier une barrire passive associe au SITRANS LR250

Page FR-10

SITRANS LR250 (HART) MISE EN SERVICE RAPIDE

A5E31997170

Slection d'une barrire passive (suite) 4) Dfinir la valeur Vbarrire (cf. fiche technique) ; par ex. chute de tension provoque par les diodes. 5) Calculer Vfonct = Valim Vbarrire. 6) Utiliser les valeurs Vfonct et Rfonct pour confirmer que l'appareil respecte les seuils de fonctionnement (zone grise dans le graphique) Courbe 1 : Usage gnral, scurit intrinsque, anti-tincelles/ nergie limite, protection non-incendiaire page 23.

Barrires actives (barrires de rptition)


Fabricant
MTL MTL Siemens Stahl MTL MTL

Numro de pice
706 7206 SITRANS I 7NG4124 9001/51-280-110-14 E02009 E02010

Franais

2. Cblage Anti-tincelles/ nergie limite

Les certificats ATEX indiqus sur la plaque signaltique sont disponibles sur notre site web : www.siemens.com/LR250. Consulter : Support > Approvals / Certificates. Pour plus de dtails sur la consommation de courant, voir Courbe 1 : Usage gnral, scurit intrinsque, anti-tincelles/ nergie limite, protection non-incendiaire page 23. Effectuer le cblage selon les dispositions rglementaires locales.

3. Cblage Non-incendiaire

Le schma de raccordement FM/CSA Classe 1, Div 2 numro 23650673 est disponible sur la page produit de notre site : www.siemens.com/LR250. Consulter Support > Installation Drawings > Level Measurement > Continuous - Radar. Pour plus de dtails sur la consommation de courant, voir Courbe 1 : Usage gnral, scurit intrinsque, anti-tincelles/ nergie limite, protection non-incendiaire page 23.

A5E31997170

SITRANS LR250 (HART) MISE EN SCE RAPIDE

Page FR-11

4. Scurit antidflagrante

Les certificats ATEX indiqus sur la plaque signaltique sont disponibles sur notre site web : www.siemens.com/LR250. Consulter : Support > Approvals / Certificates. Le certificat IECEx indiqu sur la plaque signaltique est accessible partir du site internet IECEx. Consulter : http://iecex.iec.ch, cliquer sur Ex Equipment Certificates of Conformity, puis entrer le numro du certificat IECEx SIR 08.0107X. Pour plus de dtails sur la consommation de courant, voir Courbe 2 : Antidflagrant, scurit augmente, protection contre les explosions page 24. Effectuer le cblage selon les dispositions rglementaires locales. Consulter Instructions spcifiques pour linstallation en zone dangereuse page 13 ainsi que le certificat ATEX indiqu ci-dessus.

Franais

5. Scurit augmente

Les certificats ATEX indiqus sur la plaque signaltique sont disponibles sur notre site web : www.siemens.com/LR250. Consulter : Support > Approvals / Certificates. Le certificat IECEx indiqu sur la plaque signaltique est accessible partir du site internet IECEx. Consulter : http://iecex.iec.ch, cliquer sur Ex Equipment Certificates of Conformity, puis entrer le numro du certificat IECEx SIR 08.0107X. Pour plus de dtails sur la consommation de courant, voir Courbe 2 : Antidflagrant, scurit augmente, protection contre les explosions page 24. Effectuer le cblage selon les dispositions rglementaires locales. Consulter Instructions spcifiques pour linstallation en zone dangereuse page 13 ainsi que le certificat ATEX indiqu ci-dessus.

Page FR-12

SITRANS LR250 (HART) MISE EN SERVICE RAPIDE

A5E31997170

6. Cblage antidflagrant (USA/Canada uniquement)

Le schma de raccordement FM/CSA Antidflagrant numro A5E02257843 est disponible sur la page produit de notre site : www.siemens.com/LR250. Consulter Support > Installation Drawings > Level Measurement > Continuous - Radar. Pour plus de dtails sur la consommation de courant, voir Courbe 2 : Antidflagrant, scurit augmente, protection contre les explosions page 24.

Franais

Instructions spcifiques pour linstallation en zone dangereuse


(rf. Directive Europenne ATEX 94/9/CE, Annexe II, 1/0/6)
Les consignes ci-dessous s'appliquent l'appareil objet des certificats n SIRA 06ATEX2358X, SIRA 08ATEX1301X et SIRA 09ATEX4153X. L'utilisation et l'assemblage sont dcrits dans les instructions de service. L'appareil est certifi en tant qu'instrument de la catgorie 1GD suivant SIRA 06ATEX2358X; catgorie 1/2 GD, 1D, 2D suivant SIRA 08ATEX1301X; et catgorie 3G suivant SIRA 09ATEX4153X. 3) L'appareil peut tre utilis en prsence de gaz et de vapeurs inflammables, avec des instruments de groupe IIC, IIB et IIA, classification de temprature T1, T2, T3 et T4. 4) L'appareil prsente une tanchit IP67, classe de temprature T90 C et supporte les atmosphres avec des poussires inflammables. 5) L'appareil est certifi pour utilisation dans une plage de temprature ambiante de 40 C ... +80 C. 6) Conformment la Directive 94/9/CE Annexe II, clause 1.5, cet appareil n'a pas t valu en tant que dispositif de scurit : Cf. page 5. 7) Linstallation et la vrification de cet appareil doivent tre effectues par un personnel qualifi, en accord avec le code de bonne pratique applicable (EN 60079-14 et EN 6007917 pour lEurope). 8) L'appareil ne peut pas tre rpar. 9) Les numros de certificat comportent le suffixe 'X', indiquant l'application de conditions spciales pour garantir la scurit. Toute personne habilite installer ou vrifier cet instrument doit avoir accs aux certificats. 10) Lorsque l'appareil peut entrer en contact avec des substances aggressives, il incombe l'utilisateur de prendre les mesures adquates pour empcher la dtrioration de l'appareil et garantir lindice de protection. Substances agressives : liquides ou gaz acides susceptibles d'attaquer les mtaux, ou solvants agressifs pour les matriaux polymriss. Prcautions adquates : (exemple) dterminer, l'aide de la fiche technique, si le produit rsiste certaines substances chimiques. 1) 2)

A5E31997170

SITRANS LR250 (HART) MISE EN SCE RAPIDE

Page FR-13

Programmation du SITRANS LR250


Activation du SITRANS LR250

Cf. Accs l'assistant de mise en service via le programmateur portatif page 18. Voir Assistant de mise en service via SIMATIC PDM page 21.

Mettre l'instrument sous tension. Le SITRANS LR250 dmarre automatiquement en mode de Mesure.

Affichage LCD Affichage en mode de mesure ( fonctionnement normal) 1) 2)


1 2 3 4
(((

Franais

1 indicateur d'affichage altern1), units linaires ou % 2 mode de fonct. slectionn : niveau, espace, distance ou volume 3 valeur mesure (niveau ou volume, espace ou distance) 4 units 5 reprsentation du niveau (bargraphe) 6 zone secondaire pour affichage (sur demande2)) de la temprature de l'lectronique, de la fiabilit de l'cho, du courant de boucle ou de la distance 7 zone ddie laffichage de messages de texte 5 8 indicateur de l'tat de l'appareil (pour plus de dtails se reporter aux instructions de service).

Indicateurs de prsence de dfauts S : 0 LOE affichs dans la zone de texte (7) ; l'icne associe la maintenance est
affiche dans la zone tat de l'appareil (8). En prsence d'un dfaut, le code d'erreur et un message d'erreur sont

Affichage en mode Program


Visualisation durant la navigation numro menu L'apparition de la barre de menu d'item actuel actuel indique que le menu dpasse la barre de capacit d'affichage de l'cran. item menu actuel Une bande situe au milieu de la bande barre de menu indique que l'item d'item actuel se trouve au milieu de la liste. La largeur et la position de la bande item dans la barre de menu reflte la longueur du menu et la position approximative de l'item actuel sur la liste. Plus la bande est large, plus le nombre d'items est rduit. Visualisation de paramtres
numro de paramtre valeur/slection de paramtre

Visualisation en mode dition

nom de paramtre

1) 2)

Appuyer sur la flche HAUT ou BAS pour modifier l'affichage. En rponse la demande d'appui d'une touche. Pour plus de dtails, voir le tableau sous Fonction en mode de Mesure page 15.

Page FR-14

SITRANS LR250 (HART) MISE EN SERVICE RAPIDE

A5E31997170

Programmateur portatif (rf. produit 7ML1930-1BK)


Command sparment. 1) 2) 3)

Touche Fonction en mode de Mesure


Actualise le courant de boucle1). Actualise l'affichage de la temprature interne du botier1). Actualise l'affichage de la fiabilit de l'cho1).
C

Actualise l'affichage de la distance1).

Franais

Mode permet d'accder au mode PROGRAM2). La flche DROITE permet d'accder au mode PROGRAM3). Les flches HAUT ou BAS permettent un affichage altern : units linaires et %.

Programmation avec le programmateur portatif


N.B. : Appuyer sur Mode pour un affichage altern mode de Mesure/Programmation. Lorsque l'appareil se trouve en mode PROGRAM, la sortie reste sur une valeur fixe et ne correspond pas aux variations de l'appareil. La non-utilisation du SITRANS LR250 en mode PROGRAM durant une priode allant de 15 secondes 10 minutes (suivant le niveau de menu) entrane le retour automatique au mode de Mesure.

Menus des paramtres

Les paramtres sont identifis par leur nom et rpartis 1. QUICK START 2. CONFIGURATION en groupes de fonction. 2.1. APPAREIL Pour plus de dtails sur les paramtres et les instructions ............. correspondantes consulter les instructions de service. 2.7. LINEARISATION

2.7.1. VOLUME 2.7.1.1. FORME DE CUVE

1) 2)

3)

Indique la nouvelle valeur dans la zone secondaire de l'afficheur LCD. Accde au dernier niveau de menu affich durant le cycle de fonctionnement en cours, except si l'instrument a t remis sous tension en quittant le mode PROGRAM, ou si plus de 10 minutes se sont coules depuis l'utilisation du mode PROGRAM. On obtient l'indication du menu niveau suprieur. Ouverture du menu, niveau suprieur.

A5E31997170

SITRANS LR250 (HART) MISE EN SCE RAPIDE

Page FR-15

1.

Accder au mode PROGRAM


Diriger le programmateur vers l'affichage (distance maximum : 300 mm [1 ft.]). La flche DROITE lance le mode PROGRAM et permet d'accder au niveau menu 1. Mode permet daccder au dernier niveau de menu affich en mode PROGRAM durant les 10 dernires minutes, ou au niveau de menu 1 lorsque linstrument a t arrt et remis sous tension.

programmateur portatif affichage

Max. 300 mm (1 ft)

2. Franais

Navigation : utilisation des touches en mode Navigation

N.B. :Pour accder aux paramtres directement partir du programmateur portatif appuyer sur la touche Home et entrer le numro de menu, par exemple : 2.7.1 (Volume).

Touche

Nom

Niveau menu

Fonction en mode Navigation


Accder au menu ou au paramtre suivant. Accder au premier paramtre du menu slectionn ou ouvrir le menu suivant. Accder au mode Edition. Accder au menu parent. Accder au mode de MESURE. Ouvrir le menu, niveau suprieur : menu 1.

Flche HAUT ou menu ou BAS paramtre

Flche DROITE Flche GAUCHE


Mode Home

menu paramtre menu ou paramtre menu ou paramtre menu ou paramtre

3.

Edition en mode PROGRAM


Slection d'une option indique a) Accder au paramtre souhait. b) Appuyer sur la flche DROITE visualiser le paramtre. pour
slection actuelle nom de paramtre numro de paramtre

c)

Appuyer de nouveau sur la flche DROITE pour accder au mode Edition. La valeur slectionne est accentue. Accder la nouvelle valeur slectionne.

d)

Appuyer sur la flche DROITE

pour

valider. L'affichage LCD revient aux paramtres et indique la nouvelle valeur slectionne. Page FR-16 SITRANS LR250 (HART) MISE EN SERVICE RAPIDE A5E31997170

Modification d'une valeur numrique a) Accder au paramtre souhait. b) Appuyer de nouveau sur la flche DROITE pour accder au mode Edition. La valeur actuelle est accentue. Entrer une nouvelle valeur. Appuyer sur la flche DROITE pour valider. L'affichage LCD revient aux paramtres et indique la nouvelle valeur slectionne.

nom de paramtre

numro de paramtre

c) d)

valeur courante

Franais

Utilisation des touches en mode Edition Touche Nom


Flche HAUT ou BAS

Fonction en mode Edition


Slection d'options Edition numrique Slection d'options Accs l'item. - Augmente ou diminue les chiffres. - Alterne les signes plus et moins. - Accepte les donnes (modifie le paramtre). - Commute du mode Edition au mode Navigation. - Dplace le curseur d'un espace vers la droite - ou, si le curseur est plac sur l'icne Enter, confirme les donnes et commute du mode Edition au mode Navigation. Annule le mode Edition sans modifier le paramtre. - Dplace le curseur sur le signe plus/moins lorsqu'il s'agit de la premire touche utilise - ou dplace le curseur un espace vers la gauche. Efface les valeurs affiches. Introduit une virgule dcimale.

Flche DROITE

Edition numrique

Slection d'options

Flche GAUCHE

Edition numrique Edition numrique Edition numrique

Effacer Virgule dcimale

Signe + ou - Edition numrique

Modifie le signe de la valeur introduite.

Numrique Edition numrique

Introduit le caractre correspondant.

A5E31997170

SITRANS LR250 (HART) MISE EN SCE RAPIDE

Page FR-17

Accs l'assistant de mise en service via le programmateur portatif


1. Mise en service rapide
a) Diriger le programmateur vers l'affichage ( une distance max. de 300 mm [1 ft.]), et appuyer sur la flche DROITE niveau menu 1. b) Appuyer deux fois sur la flche DROITE visualiser les paramtres. c) Appuyer sur la flche DROITE BAS pour accder au mode Edition ou sur la flche pour accder l'item menu 1.1 et pour accder au mode PROGRAM et ouvrir le

pour accepter les valeurs par dfaut et accder l'item suivant.

Franais

d) e)

Pour modifier une valeur, accder l'item souhait ou entrer une nouvelle valeur. Pour confirmer une valeur modifie, appuyer sur la flche DROITE la flche BAS pour accder l'item suivant. . Appuyer sur

f)

Pour confirmer les rglages effectus durant la Mise en service rapide, slectionner Yes en rponse Apply changes (Appliquer modifications) dans ltape 1.8.

1.1. Langue Slection de la langue d'interface (cran LCD), avec prise d'effet immdiate.
Options
ENGLISH, DEUTSCH, FRANAIS, ESPAOL

1.2. Matriau Options


LIQUIDE LIQUID LOW DK (liquide faible constante dilectrique)

1.3. Temps de rponse Dfinit la vitesse de raction de l'instrument aux variations de mesure dans la plage spcifie.
Temps de rponse
SLOW (LENTE) MED (MOY) FAST (RAPIDE)

Vitesse de remplissage/vidange
0,1 m/min (0.32 ft/min) 1,0 m/min (3.28 ft/min) 10,0 m/min (32.8 ft/min)

Slectionner une vitesse lgrement suprieure la vitesse de remplissage ou de vidange maximale (la plus leve des deux).

1.4. Units Units de mesure capteur. Options


M, CM, MM, FT, IN

Page FR-18

SITRANS LR250 (HART) MISE EN SERVICE RAPIDE

A5E31997170

1.5. Mode de fonctionnement


point de rfrence du capteur espace point de rfrence du capteur

distance point d'talonnage max.

point de rfrence du capteur

niveau point d'talonnage min.

Franais

Fonctionnement
PAS DE SERVICE NIVEAU ESPACE DISTANCE
1)

Description
Les mesures et le courant de boucle correspondant ne sont pas actualiss et l'appareil commute automatiquement en mode scurit-dfaut1). Distance entre le point d'talonnage min. et la surface du matriau. Distance entre le point d'talonnage maximum et la surface du matriau. Distance entre le point de rfrence capteur et la surface du matriau.

Pour plus de dtails sur le mode scurit-dfaut consulter la version complte du manuel.

1.6. Point d'talonnage min. Distance entre le point de rfrence du capteur et le point d'talonnage min. : gnralement le niveau process vide. Cf. Mode de fonctionnement (1.5.) pour plus de dtails. Valeurs
Plage : 0,00 20,00 m

1.7. Point d'talonnage max. Distance entre le point de rfrence du capteur et le point d'talonnage max. : gnralement le niveau process plein. Cf. Mode de fonctionnement (1.5.) pour plus de dtails. Valeurs
Plage : 0,00 20,00 m

1.8. Appliquer? (valider les modifications) Pour sauvegarder les modifications effectues durant la Mise en service rapide, slectionner Yes (Oui). Options
Appuyer sur Mode fonctionner. YES, NO, DONE (L'affichage indique DONE (TERMINE) la fin de la Mise en service.) pour revenir au mode de Mesure. Le SITRANS LR250 est prt

N.B. : La fonction de suppression automatique des chos parasites est efficace en prsence d'obstacles dans le rservoir. Pour plus de dtails, consulter les instructions de service.

A5E31997170

SITRANS LR250 (HART) MISE EN SCE RAPIDE

Page FR-19

Communication SITRANS LR250 : HART

Vous trouverez des explications dtailles sur les paramtres applicables dans la version complte des instructions de service. Nous vous conseillons de programmer l'instrument avec SIMATIC Process Device Manager (PDM). Les guides d'application concernant le paramtrage d'appareils HART avec SIMATIC PDM sont disponibles sur la page produit de notre site internet : www.siemens.com/ LR250 sous Support.

SIMATIC PDM
Le logiciel SIMATIC PDM est utilis pour la configuration et la maintenance d'appareils tels que le SITRANS LR250. Pour plus de dtails sur l'utilisation de SIMATIC PDM, consulter les instructions de service ou l'aide en ligne. (Plus de dtails sur www.siemens.com/simatic-pdm.)

Dsactivation de mmoires tampon


Franais

Permet d'aligner SIMATIC PDM avec le modem HART pour les systmes dexploitation Windows 2000 et Windows XP 1). N.B. : SIMATIC PDM fonctionne uniquement avec la version Windows XP Professionnel, pas avec l'Edition Familiale. Pour dsactiver les mmoires tampon, vous devez disposer de droits d'administrateur sur le systme d'exploitation. 1) 2) 3) 4) 5) 6) Cliquer sur Dmarrer/Paramtres/Panneau de configuration pour lancer la configuration. Double-cliquer sur Systme, slectionner l'onglet Matriel, puis cliquer sur le bouton Device Manager. Ouvrir le dossier Ports et double-cliquer le port COM utilis par le systme pour ouvrir la fentre Proprits du port de communication. Slectionner l'onglet Paramtres Port et double-cliquer sur le bouton Avanc. Si la case Use FIFO buffers (utiliser tampons FIFO) est coche, cliquer pour dslectionner. Cliquer sur OK pour conclure. Fermer tous les crans, puis redmarrer.

Actualisation de l'Electronic Device Description (EDD)


Accs via SIMATIC PDM Rv. 5.2 SP1 pour la configuration et le diagnostic de base uniquement. Les fonctions avances telles que l'Assistant de mise en service ncessitent la rvision 6.0 SP4 (minimum). En SIMATIC PDM, cette description (EDD) est votre disposition dans la rubrique Device Catalog, sous Sensors/Level/Echo/Siemens Milltronics/SITRANS LR250. Consulter la page produit sur notre site web : www.siemens.com/LR250, pour obtenir la dernire version de SIMATIC PDM, le dernier SP (Service Pack) et le dernier HF (hot fix). Consulter Support > Software Downloads

N.B. :

Pour installer une nouvelle EDD :


1) 2) 3) Tlcharger l'EDD sur la page produit de notre site : www.siemens.com/LR250 et enregistrer les fichiers sur votre ordinateur. Extraire le fichier zipp vers un emplacement accessible. Lancer SIMATIC PDM Manager Device Catalog, accder au fichier EDD dzipp et le slectionner.
Windows est une marque dpose de Microsoft Corporation.

1)

Page FR-20

SITRANS LR250 (HART) MISE EN SERVICE RAPIDE

A5E31997170

Configuration d'un nouvel instrument


N.B. : La slection de Cancel (Annuler) durant l'actualisation (appareil vers SIMATIC PDM)
entrane l'actualisation de certains paramtres. 1) 2) 3) 4) 5) Il est impratif d'utiliser le tout dernier fichier EDD. Vous pouvez, si ncessaire, l'actualiser (cf. Actualisation de l'Electronic Device Description (EDD) page 20). Lancer SIMATIC Manager et crer un nouveau projet pour le LR250. Les guides d'application concernant le paramtrage d'appareils HART avec SIMATIC PDM sont disponibles sur la page produit de notre site internet : www.siemens.com/LR250. Ouvrir le menu Appareil Remise zro et cliquer sur OK pour restaurer les valeurs usine. Une fois la rinitialisation effectue, exporter les paramtres vers le PC/PG. Effectuer le rglage de l'instrument suivant l'Assistant de mise en service simplifie.

Assistant de mise en service via SIMATIC PDM


N.B. :
Les rglages de mise en service rapide sont interlis. Toute modification doit tre valide en cliquant sur TERMINER ET TELECHARGER la fin de l'tape 5 pour sauvegarder les rglages off-line et les transfrer vers l'instrument de mesure. Slectionner BACK/RETOUR pour revenir une valeur et la modifier ou Cancel/Annuler pour quitter la Mise en service. Lancer SIMATIC PDM, ouvrir le menu Appareil Assistant - Mise en service rapide, et suivre les tapes 1 5.

Franais

Mise en service rapide

Fonctionnement via FDT (Field Device Tool)


Le standard de fichiers FDT est utilis dans de nombreux logiciels conus pour la mise en service et la maintenance des appareils de terrain. Parmi ces logiciels, on trouve PACTware et Fieldcare. La configuration d'un appareil de terrain via FDT ncessite le programme DTM (Device Type Manager). Les instruments de mesure Siemens utilisent SITRANS DTM ainsi qu'un fichier de description lectronique EDD pour SITRANS DTM. 1) La premire tape consiste installer SITRANS DTM. Vous pouvez le tlcharger sur : http://support.automation.siemens.com. Cliquer sur Product Support et accder Product Information/Automation Technology/Sensor systems/Process Instrumentation/Software & Communications. 2) Installer le fichier EDD du SITRANS LR250 HART pour SITRANS DTM. Vous pouvez le tlcharger sur la page produit correspondante, l'adresse : www.siemens.com/LR250. Consulter Support > Software Downloads.

Configuration d'un autre appareil via FDT


Un guide d'application est disponible sur la page produit de notre site internet, sous Support. A5E31997170 SITRANS LR250 (HART) MISE EN SCE RAPIDE Page FR-21

Utilisation avec AMS Device Manager


Le logiciel AMS Device Manager permet de configurer, calibrer et dpanner les appareils de terrain. Pour plus de dtails sur l'utilisation de AMS Device Manager, consulter les instructions de service ou l'aide en ligne. Pour plus d'informations, consulter : http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD)


SITRANS LR250 requiert le fichier EDD pour la version 9.0 d'AMS Device Manager.

Configuration d'un nouvel appareil avec AMS Device Manager


1) Consulter la page produit sur notre site web : www.siemens.com/LR250 pour obtenir le dernier fichier EDD. Consulter Support > Software Downloads et tlcharger le fichier, si ncessaire. Sauvegarder les fichiers sur l'ordinateur. Extraire le fichier zipp vers un emplacement accessible. Lancer AMS Device Manager Add Device Type, accder au fichier dzipp EDD et le slectionner.

Franais

2)

Lancer AMS Device Manager. Un Guide d'application ddi la configuration des appareils HART avec le logiciel AMS Device Manager est disponible sur la page produit de notre site internet, sous Support.

Maintenance
Utilis en conditions de fonctionnement normales, le SITRANS LR250 ne requiert pas de maintenance ou de nettoyage. Si des conditions extrmes sont prsentes et le nettoyage savre ncessaire : 1) 2) Choisir un produit de nettoyage adapt aux matriaux de construction de lantenne et au produit prsent dans le process contrl. Retirer l'instrument de l'application et nettoyer l'antenne avec un tissu et un produit adapt.

Rparation de l'instrument et limite de responsabilit


Pour plus de dtails, veuillez vous reporter la dernire page.

Remplacement de l'antenne ou de l'lectronique/botier


En cas d'endommagement ou de dysfonctionnement, l'antenne peut tre remplace sans ncessiter un rtalonnage. L'installation d'une antenne diffrente peut tre ralise par un centre de rparation agre Siemens. En cas de remplacement dune llectronique ou dun botier endommag(e) ou dfectueux(euse), il faut vrifier la compatibilit avec l'antenne utilise, pour viter le rtalonnage de l'instrument (effectu par le personnel autoris de Siemens).

Page FR-22

SITRANS LR250 (HART) MISE EN SERVICE RAPIDE

A5E31997170

Temprature de process maximum


Versions LR250 quipes d'adaptateur bride
AVERTISSEMENT : La temprature interne ne doit pas dpasser 80 C (176 F). N.B. : Ce graphique est fourni titre indicatif uniquement et ne tient pas compte des diffrentes possibilits de raccord process. (Il ne s'applique PAS lorsque le SITRANS LR250 est fix directement sur une surface mtallique du rservoir.) Ce graphique ne tient pas compte de la chaleur due lexposition de l'appareil au soleil. Paramtre 3.2.1 Temprature interne actuelle permet de contrler la temprature interne. Tempratures max. process/temprature ambiante admissible

Temprature ambiante (oC)

Franais

temprature de process

temprature de process (oC)

Boucle de courant
Tension d'alimentation ncessaire
N.B. : ces courbes s'appliquent un instrument utilis en mode autonome, configur avec le programmateur portatif de Siemens. Rsistance de la boucle RL Courbe 1 : Usage gnral, scurit intrinsque, antitincelles/ nergie limite, protection non-incendiaire

ZONE DE FONCTIONNEMENT AUTORISEE

Tension de la boucle VL

N.B. : la tension minimum requise pour la communication HART avec 220 Ohms (RL) est 16,3 VCC.

A5E31997170

SITRANS LR250 (HART) MISE EN SCE RAPIDE

Page FR-23

Rsistance de la boucle RL

Courbe 2 : Antidflagrant, scurit augmente, protection contre les explosions

ZONE DE FONCTIONNEMENT AUTORISEE

Tension de la boucle VL

Franais

N.B. : La tension minimum requise pour la communication HART avec 220 Ohms (RL) est 20.94 VCC.

Comportement au dmarrage
L'appareil consomme moins de 3,6 mA la mise sous tension. Dure jusqu' la premire mesure infrieure 50 s

Page FR-24

SITRANS LR250 (HART) MISE EN SERVICE RAPIDE

A5E31997170

Manuale per l'avvio rapido del SITRANS LR250 (HART)


Questo manuale descrive le caratteristiche e le funzioni pi importanti del SITRANS LR250 (HART 1)). Si consiglia all'operatore di leggere il manuale d'uso integrale per ottenere i massimi risultati. Il manuale disponibile sulla pagina prodotto SITRANS LR250 del nostro sito web: www.siemens.com/LR250. Per la versione stampata, rivolgersi al proprio rappresentante di Siemens Milltronics. Per ulteriori informazioni su questo manuale, rivolgersi a: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 Email: techpubs.smpi@siemens.com

Copyright Siemens AG 2013. Tutti i diritti riservati


Si consiglia agli utenti di acquisire manuali stampati autorizzati oppure di consultare le versioni elettroniche progettate e realizzate da Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments non potr essere ritenuta responsabile per la riproduzione parziale o totale del contenuto delle versioni stampate o elettroniche.

Clausola di esclusione della responsabilit


La conformit tra lo stato tecnico dell'apparecchio e il contenuto di questo manuale stata verificata; tuttavia, potrebbero essere riscontrate alcune variazioni. SMPI non garantisce pertanto la completa conformit del manuale con l'apparecchio descritto. Si avvisa inoltre che tutti i manuali vengono regolarmente controllati ed aggiornati e che le eventuali correzioni vengono incluse nelle versioni successive. Si invitano gli utenti a trasmettere i propri suggerimenti e commenti a SMPI. SMPI si riserva il diritto di apportare modifiche tecniche.

Italiano

MILLTRONICS un marchio registrato di Siemens Milltronics Process Instruments.

Supporto tecnico
Il servizio disponible 24 ore su 24. Per trovare l'ufficio di rappresentanza Siemens Automation pi vicino, consultare il sito web: www.siemens.com/automation/partner: (Selezionare la lingua desiderata). Cliccare sul tab Contatti, scegliere Servizi, e Service per scegliere il prodotto (+Automation Technology > +Sensor Systems >+Strumentazione di processi > +Misura di livello > +Continuous). Selezionare Radar. Selezionare un paese e una citt/Regione. Selezionare Technical Support nell'elenco Servizi. Per accedere al supporto tecnico on-line, visitare: www.siemens.com/automation/supportrequest Introdurre il nome dell'apparecchio (SITRANS LR250) o il numero dell'ordine di acquisto, cliccare su Cerca, e selezionare il tipo di prodotto. Cliccare su Next. Introdurre un keyword per la richiesta. Visualizzare la documentazione o cliccare suNext (Avanti) per inviare una descrizione della richiesta via e-mail al nostro team di Supporto Tecnico. Centro di Supporto Tecnico Siemens IA/DT: tel. +49 (0)911 895 7222
1)

HART un marchio registrato di HART Communication Foundation.

A5E31997170

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

Pagina IT-1

Indicazioni di sicurezza

L'apparecchio deve essere utilizzato osservando le avvertenze. Un utilizzo non corretto potrebbe causare danni anche gravi alle persone, sia al prodotto e agli apparecchi ad esso collegati. Le avvertenze contengono una spiegazione dettagliata del livello di sicurezza da osservare. Il simbolo AVVERTENZA sul prodotto indica all'utente che la mancata osservazione delle precauzioni necessarie pu causare morte o gravi lesioni personali nonch seri danni materiali. Il simbolo AVVERTENZA usato quando sul prodotto non riportato un simbolo di avvertenza. Indica all'utente che la mancata osservazione delle precauzioni necessarie pu causare morte o gravi lesioni personali nonch seri danni materiali. Avvertenza:le note contengono importanti informazioni sul prodotto o sulla sezione del manuale d'istruzioni a cui viene fatto riferimento.

Conformit FCC
Installazioni negli USA: Regole della FCC (Federal Communications Commission)
AVVERTENZA: Ogni cambiamento o modifica non espressamente approvato da Siemens Milltronics potrebbe revocare il diritto all'utilizzo dell'apparecchio. Note: Il presente apparecchio stato collaudato e risultato conforme ai vincoli relativi ai dispositivi digitali di Classe A, inclusi nella Parte 15 della normativa FCC. Tali vincoli sono stati stabiliti ai fini di garantire un'adatta protezione da interferenze dannose durante l'utilizzo dell'apparecchio in ambienti commerciali. Il presente apparecchio genera, utilizza e pu emettere energia in radio frequenza. Se installata e utilizzata in modo improprio, vale a dire in difformit con il relativo manuale d'istruzioni, il presente apparecchio pu causare interferenze dannose alle radiocomunicazioni. L'impiego del presente apparecchio all'interno di una zona residenziale potrebbe causare interferenze dannose. In tal caso, l'utente dovr risolvere il problema ed eliminare tali interferenze a proprie spese.

Italiano

SITRANS LR250
AVVERTENZA: Per garantire la protezione il SITRANS LR250 deve essere utilizzato osservando le istruzioni e avvertenze contenute in questo manuale. Avvertenza: Questo prodotto stato progettato per l'uso in ambienti industriali. Luso di questo apparecchio in zone residenziali pu causare disturbi alle comunicazioni radiofoniche. SITRANS LR250 un misuratore di livello a circuito alimentato a due fili per campi di misura fino a 20 m (66 ft). Utilizza una tecnologia radar ad impulsi 25 Ghz per misurare liquidi e fanghi nei serbatoi di stoccaggio con condizioni di lavoro pericolose quali pressione e alte temperature. E' ideale per serbatoi di dimensioni ridotte contenenti prodotti a basso dielettrico. I componenti dell'apparecchio sono: la custodia stagna con l'elettronica collegata ad un'antenna a cono, e l'attacco di processo filettato o a flangia. Antenna con attacco filettato in PVDF. Il SITRANS LR250 compatibile con il protocollo di comunicazione HART e SIMATIC PDM . L'elaborazione dei segnali effettuata grazie al Process Intelligence.

Pagina IT-2

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

A5E31997170

Specifiche tecniche

Il dati tecnici completi sono descritti nel manuale di istruzioni SITRANS LR250. Le Omologazioni sono riportate nella sezione Certificazioni pagina 4.

Temperatura ambientale/di funzionamento


Note:
La temperatura massima di funzionamento dipende dalla connessione al processo, i materiali utilizzati per l'antenna e la pressione (serbatoio). Vedi Grafico della temperatura massima di lavoro pagina 23. Per ulteriori dettagli vedere le curve di limitazione della Pressione/Temperatura di processo riportate nelle istruzioni d'uso. Le caratteristiche relative alle condizioni ambientali quali temperatura di processo e pressione sono specificate sull'apposita etichetta dell'apparecchio. L'illustrazione di riferimento citata sull'etichetta puo' essere scaricata direttamente sul nostro sito internet www.siemens.com/LR250, rubrica More Info/Installation drawings/Level Measurement/ Installation Drawings/LR250. Ulteriori dati riguardanti le connessioni al processo sono disponibili sulla pagina Installation Drawings, rubrica Process Connection Diagrams. L'ampiezza del segnale aumenta proporzionalmente al diametro del cono. E' quindi consigliabile scegliere un'antenna pi larga. Eventuali prolunghe possono essere installate sotto le filettature. Per ulteriori dettagli vedere Grafico della temperatura massima di lavoro pagina 23.
temperatura ambientale (intorno alla custodia) 40 C ... 80 C (40 F ... 176 F) targhetta identificativa

Italiano

temperatura di lavoro (all'attacco di processo): antenna a cono standard: - con anello di tenuta FKM: 40 oC ... +200 oC (-40 oF ... +392 oF) - con anello di tenuta FFKM: 20 oC ... +200 oC (4 oF ... +392 oF) antenna con attacco filettato in PVDF, 2" NPT / BSPT / G: 40 ... +80 C (40 ... +176 F)

etichetta dell'apparecchio

Alimentazione
Uso generale Sicurezza intrinseca Antiscintilla/a energia limitata Non incendivo (FM/CSA solo USA/Canada) Antifiamma Sicurezza aumentata A prova di esplosione (FM/CSA solo USA/Canada) Massima 30 V DC 4 ... 20 mA Corrente di avvio max.:: Nominale 24 V DC a 550 Ohm

Nominale 24 V DC a 250 Ohm

vedi Messa in servizio pagina 24

A5E31997170

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

Pagina IT-3

Certificazioni
Avvertenza: Verificare le omologazioni consultando l'etichetta dell'apparecchio. Generali CSAUS/C, FM, CE, C-TICK, KC Radio Europa (R&TTE), FCC, Industry Canada Aree a rischio esplosione Sicurezza intrinseca 1) (Europa) ATEX II 1G, EEx ia IIC T4 ATEX II 1D, EEx tD A20 IP67 T90 C (Internazionale) IECEx SIR 05.0031X, Ex ia IIC T4, Ex tD A20 IP67 T90 C (US/Canada) FM/CSA Classe I, Div. 1, Gruppi A, B, C, D Classe II, Div. 1, Gruppi E, F, G Classe III T4 (Brasile) INMETRO: DNV 12.0087 X Ex ia IIC T4 Ga Ex ia ta IIIC T90 C Da IP65/IP67 -40 C Ta +80 C DNV #OCP 0017

(Cina)NEPSI Ex ia IIC T4/ DIP A20 TA, T90 C IP67

ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

Italiano

Antiscintilla/a energia limitata 2) (Europa) ATEX II 3 G Ex nA/nL IIC T4 Gc (Cina) NEPSI Ex nA II T4/ Ex nL IIC T4 Non incendivo 3) (USA/Canada) FM/CSA Classe I, Div. 2, Gruppi A, B, C, D T5

Antifiamma 4) (Europa/Internazionale) ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex dmbia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C (Brasile) INMETRO: DNV 12.0088 X Ex d ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-1:2009, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

1) 2) 3) 4)

Vedi Cablaggio intrinsecamente sicuro pagina 9. Vedi Cablaggio antiscintilla/a energia limitata pagina 11. Vedi Cablaggio NI (non incendivo) pagina 11. Vedi Cablaggio antifiamma pagina 12.

Pagina IT-4

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

A5E31997170

Aree a rischio di esplosione (segue) Sicurezza aumentata 1) (Europa/Internazionale) ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex embia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C (Brasile) INMETRO: DNV 12.0088 X Ex e ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-7:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

Antifiamma/Sicurezza aumentata (Cina) NEPSI Ex dmbia IIC T4/ Ex embia IIC T4/ DIP A20 TA, T90 oC IP67 Antideflagrante 2) (USA/Canada) FM/CSA Classe I, Div. 1, Gruppi A, B, C, D Classe II, Div. 1, Gruppi E, F, G Classe III T4

Marittima Lloyds Register of Shipping Certificato tipo ABS (American Bureau of Shipping).

Italiano

Applicazioni in pressione
ATTENZIONE:

Non svitare, smontare o rimuovere la connessione di processo o la custodia dell'apparecchio quando il contenuto del serbatoio sotto pressione. L'utente responsabile della selezione dei materiali quali bulloni o guarnizioni che consentano di rispettare i limiti e l'uso appropriato della flangia e che siano adeguati alle condizioni di servizio. L'installazione errata pu provocare un calo della pressione di lavoro.
Note: E' importante lasciare l'etichetta di identificazione collegata all'insieme che delimita la soglia della pressione di lavoro1). In caso di sostituzione dell'apparecchio, assicurarsi che venga spostata l'apposita etichetta di identificazione. Gli apparecchi SITRANS LR560 sono sottoposti a prove di pressione. Rispettano rigorosamente ed eccedono gli standard stabiliti dall'ASME (Boiler and Pressure Vessel Code) e dalla direttiva Europea PED sulle apparecchiature a pressione.

1) 2)

Vedi Cablaggio di sicurezza aumentata pagina 12. Vedi Cablaggio antideflagrante (solo USA/Canada) pagina 13.

A5E31997170

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

Pagina IT-5

Direttiva Europea sulle attrezzature in pressione, PED, 97/23/EC


I trasmettitori di livello Siemens dotati di attacco al processo flangiato, filettato o tipo sanitario (clamp) non sono dotati di custodia resistente alla pressione e sono quindi esclusi dalla direttiva europea sulle attrezzature in pressione o di sicurezza (vedi regolamento della commissione Europea 1/8). 1)

Installazione
ATTENZIONE: L'installazione deve essere eseguita esclusivamente da personale qualificato e in conformit con le normative locali correnti. Manipolare l'apparecchio con cura tenendo bene la custodia per evitare di danneggiare l'antenna o l'etichetta. Manipolare l'antenna con attacco filettato in PVDF con estrema cura. Qualsiasi danno alla superficie dell'antenna, in particolare alla punta, puo' impedire il funzionamento corretto dell'apparecchio. I materiali di composizione vengono scelti in base alla compatibilit chimica (o inerzia) per uso generico. Per l'esposizione in ambienti specifici, verificare il grafico della compatibilit chimica prima di procedere all'installazione. Note:
L'installazione nell'Unione Europea e nei paesi membri dovr essere effettuata in conformit con la norma ETSI EN 302372. Per ulteriori informazioni sulle omologazioni vedi la targhetta di identificazione. I numeri di serie stampati all'interno dell'attacco di processo costituiscono un'identificativo unico che indica la data di fabbricazione. Esempio: MMDDYY XXX (MM = mese, DD = giorno, YY = anno, e XXX= unit sequenziale prodotta) Altre marcature (in base allo spazio disponibile) indicano: configurazione flangia, dimensioni, classe di pressione, materiale, e resistenza del materiale al calore.

Italiano

Tipo di tronchetto
Antenna con attacco filettato in PVDF
spazio libero minimo: 10 mm (0.4")

Antenna a cono di acciaio inossidabile


spazio libero minimo: 10 mm (0.4")

Diametro min. 51 mm

Per impedire la riflessione di falsi echi sul tronchetto importante che l'estremit dell'antenna a cono sporga di almeno 10 mm (0.4''). Diametro minimo consigliato del tronchetto per l'antenna con attacco filettato in PVDF: 51 mm (2"). Tutte le antenne, tranne la versione con attacco filettato in PVDF, possono essere dotate di un'estensione da 100 mm (3.93")

1)

L'apparecchio dotato di un'insieme di componenti che delimitano la soglia della pressione di lavoro ed agiscono come barriera contro le perdite di pressione dal serbatoio di processo. L'insieme composto dall'attacco al processo e l'emettitore), generalmente senza la custodia con elettronica.

Pagina IT-6

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

A5E31997170

Montaggio tronchetto
In caso di serbatoi alti e stretti, si raccomanda di non montare lapparecchio al di sopra della parte centrale del silo per prevenire interferenze (echi di disturbo).

Ambiente

Installare l'apparecchio in un ambiente idoneo errato alle caratteristiche e ai materiali di costruzione della custodia. E' opportuno prevedere un rivestimento di protezione solare se l'apparecchio deve essere esposto al sole.

consigliato

Angolo del fascio


Per angolo del fascio si intende il punto del cono in cui la densit dell'energia si riduce a met della massima densit di energia. La massima densit di energia situata davanti all'antenna. Con la dispersione di un segnale all'esterno del fascio possono essere rilevati target errati.

cono di trasmissione angolo del fascio: antenna con attacco filettato in PVDF = 19 angolo cono 1.5" = 19 cono 2" = 15 del fascio cono 3" = 10 cono 4" = 8

Cono di emissione
Scale, tubi, assali a doppio T o flussi di carico provocano riflessioni o echi falsi e non devono trovarsi all'interno del cono di emissione.

Italiano

Accesso per la programmazione


Installare l'apparecchio in modo da ottenere un accesso facilitato al display, per la visualizzazione e il controllo nonch per la programmazione tramite il programmatore portatile.

Istruzioni generali di montaggio


AVVERTENZA: Per applicazioni in pressione, necessario utilizzare un nastro PTFE (o un altro composto appropriato per sigillare le filettature) e stringere ulteriormente l'attacco di collegamento all'apparecchio rispetto alla chiusura manuale. (La torsione massima di 40 N-m (30 ft-lbs). Avvertenza: La custodia dell'apparecchio pu essere ruotata innumerevoli volte.

Versione filettata
1) 2) Prima di inserire l'apparecchio nell'attacco di collegamento, verificare la corrispondenza delle filettature onde evitare di danneggiarle. E' sufficente avvitare l'apparecchio sull'attacco di processo e stringere manualmente o con un'apposita chiave a forchetta. Per applicazioni in pressione vedi Avvertenza sopra indicata.

A5E31997170

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

Pagina IT-7

Versioni flangiate
AVVERTENZA: L'utente responsabile della selezione dei materiali quali bulloni o guarnizioni che consentano di rispettare i limiti e l'uso appropriato della flangia e che siano adeguati alle condizioni di servizio.

Collegamento elettrico
Alimentazione
ATTENZIONE: I terminali di ingresso CC devono essere alimentati da una fonte dotata di isolamento elettrico tra l'ingresso e l'uscita, per garantire la conformit ai requisiti di sicurezza applicabili della direttiva IEC 61010-1. Tutti i collegamenti elettrici di campo devono disporre di adeguato isolamento per le tensioni di esercizio.

Collegamenti SITRANS LR250


ATTENZIONE: Per informazioni sulle omologazioni consultare la targhetta identificativa. Utilizzare guarnizioni adatte a tenuta stagna per garantire il grado di protezione IP o NEMA. Vedi Cablaggio per aree a rischio esplosione pagina 9. Note: Utilizzare un cavo bifilare ritorto: calibro da AWG 22 a 14 (0.34 mm2 a 2.5 mm2). Potr essere necessario separare i cavi e conduit in conformit con le procedure standard di cablaggio della strumentazione o con codici locali per materiali elettrici.
Svitare la vite di pressione del coperchio con una chiave a brugola da 2 mm. tappo (IP 68) pressacavo opzionale1) 2) (o ingresso cavo NPT2)) schermatura del cavo (se utilizzata)

Italiano

1) 2) 3) 4)

Rimuovere il rivestimento del cavo per circa 70 mm (2.75") partendo dall'estremit, quindi passare il fili attraverso il serracavo. 1) 2) Collegare i fili con gli appositi morsetti seguendo il diagramma sottoriportato: la polarit viene indicata sul blocco terminale. Effettuare il collegamento a terra dell'apparecchio attenendosi alle prescrizioni vigenti. Avvitare il pressacavo per garantire la tenuta stagna.

1) 2)

Pu essere fornito con l'apparecchio. Se il cavo viene inserito in un tubo protettivo o conduit, utilizzare un pressatubo di dimensioni adatte per garantire la tenuta stagna.

Pagina IT-8

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

A5E31997170

5)

Chiudere il coperchio prima di configurare l'apparecchio.

Collegamenti HART
Impostazione tipica PLC/mA con HART 1) 2)
PLC attivo alimentazione R= 250 modem HART SITRANS LR250 comunicatore HART

Note: A seconda dell'apparecchio, l'alimentazione puo' essere separata o legata al PLC Limitare la resistenza HART1) a seconda della zona di impiego autorizzata2). Non collegare pi di un programmatore HART contemporaneamente.

Cablaggio per aree a rischio esplosione


Per l'installazione in zone potenzialmente esplosive esistono sei opzioni. L'importante in ogni caso di controllare le certificazioni riportate sulla targhetta dell'apparecchio e sull'apposita etichetta.

1. Cablaggio intrinsecamente sicuro


Targhetta apparecchio (ATEX/IECEX/INMETRO/C-TICK) Italiano

Il certificato ATEX scaricabile dalla pagina prodotti del nostro sito: www.siemens.com/LR250. Rubrica Support > Approvals / Certificates. Il certificato IECEx riportato sulla targhetta dell'apparecchio disponibile sul sito internet IECEx. Vedere: http://iecex.iec.ch e cliccare su Ex Equipment Certificates of Conformity, poi digitare il numero di certificato IECEx SIR 05.0031X.

1) 2)

Resistenza totale del loop = resistenza del cavo + 250 Ohm (resistore). Vedi Curva 1: Uso generale, intrinsecamente sicuro, antiscintilla/a energia limitata, non incendivo pagina 23 oCurva 2: Antifiamma, sicurezza aumentata, a prova di esplosione pagina 24.

A5E31997170

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

Pagina IT-9

Cablaggio intrinsecamente sicuro (segue)


Targhetta apparecchio (FM/CSA)

Il diagramma di collegamento FM/CSA sicurezza intrinseca numero 23650653 scaricabile dalla pagina prodotto del nostro sito: www.siemens.com/ LR250. Vedi Support > Installation Drawings > Level Measurement > Continuous - Radar. Per informazioni sull'alimentazione necessaria vediCurva 1: Uso generale, intrinsecamente sicuro, antiscintilla/a energia limitata, non incendivo pagina 23. Per informazioni sul cablaggio: attenersi alle prescrizioni vigenti. Le applicazioni esterne tipo NEMA 4X / tipo 4X / NEMA 6, IP67, IP68 richiedono limpiego di pressacavi e guarnizioni a tenuta stagna alla polvere e allacqua. Consultare Istruzioni specifiche per aree potenzialmente esplosive pagina 13. I tipi di barriere di sicurezza consigliati sono riportati nelle sezioni Barriere a diodo derivatore passivo pagina 10 e Barriere attive (barriere di ripetizione) pagina 11. Avvertenza: Per scegliere un modulo di ingresso, un'alimentazione o una barriera adatta per il PLC necessario conoscere la sicurezza intrinseca e l'applicazione. L'installatore deve garantire che l'installazione intrinsecamente sicura sia conforme ai requisiti stabiliti dalla certificazione dell'apparecchio ed ai codici di procedura o le istruzioni per l'installazione.

Italiano

Barriere a diodo derivatore passivo


Avvertenza: L'apparecchio richiede una tensione di alimentazione regolata. Fabbricante
MTL MTL Stahl Stahl

Numero pezzo
787SP+ (due canali) 7787P+ (due canali) 9001/01-280-100-10 (un canale) 9002/01-280-110-10 (due canali)

Verificare, eseguendo il seguente calcolo, che la barriera di sicurezza conforme ai parametri di entrata intrinsecamente sicura dell'apparecchio LR250: Re-e = massima resistenza della barriera, tra i due estremi Rloop = resistenza del loop (resistenza totale del cavo + resistenza in serie, ad esempio entrate PLC e display) Vbarrier = cadute di tensione dovute alla barriera 1) Definire il valore Re-e basandosi sulla scheda tecnica. 2) Calcolare il valore totale Rloop. 3) Calcolare Rworking = Re-e + Rloop.

Come scegliere una barriera passiva per il SITRANS LR250

Pagina IT-10

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

A5E31997170

Come scegliere una barriera passiva (segue) 4) Definire il valore Vbarrier in base alla scheda tecnica della barriera (esempio: cadute di tensione dovute ai diodi). 5) Calcolare Vworking = Vsupply Vbarrier. 6) Utilizzare i valori Vworking e Rworking per confermare che l'apparecchio funziona nei limiti definiti (zona grigia del grafico)Curva 1: Uso generale, intrinsecamente sicuro, antiscintilla/a energia limitata, non incendivo pagina 23.

Barriere attive (barriere di ripetizione)


Marca, fabbricante Numero pezzo
MTL MTL Siemens Stahl MTL MTL 706 7206 SITRANS I 7NG4124 9001/51-280-110-14 E02009 E02010

2. Cablaggio antiscintilla/a energia limitata

Il certificato ATEX scaricabile dalla pagina prodotti del nostro sito: www.siemens.com/LR250. Vedere: Support > Approvals / Certificates. Per informazioni sull'alimentazione necessaria vediCurva 1: Uso generale, intrinsecamente sicuro, antiscintilla/a energia limitata, non incendivo pagina 23. Per il cablaggio attenersi alle prescrizioni vigenti.

Italiano

3. Cablaggio NI (non incendivo)

Il diagramma di collegamento FM/CSA Classe 1, Div 2 numero 23650673 scaricabile dalla pagina prodotto nel nostro sito: www.siemens.com/ LR250. Vedi Support > Installation Drawings > Level Measurement > Continuous - Radar. Per informazioni sull'alimentazione necessaria vediCurva 1: Uso generale, intrinsecamente sicuro, antiscintilla/a energia limitata, non incendivo pagina 23.

A5E31997170

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

Pagina IT-11

4. Cablaggio antifiamma

Il certificato ATEX scaricabile dalla pagina prodotti del nostro sito: www.siemens.com/LR250. Vedi: Support > Approvals / Certificates. Il certificato IECEx riportato sulla targhetta dell'apparecchio disponibile sul sito internet IECEx. Vedi: http://iecex.iec.ch e cliccare su Ex Equipment Certificates of Conformity, poi digitare il numero di certificato IECEx SIR 08.0107X. Per informazioni sull'alimentazione necessaria vediCurva 2: Antifiamma, sicurezza aumentata, a prova di esplosione pagina 24. Per il cablaggio attenersi alle prescrizioni vigenti. Vedi anche Istruzioni specifiche per aree potenzialmente esplosive pagina 13 e la certificazione ATEX sopraindicata.

5. Cablaggio di sicurezza aumentata

Il certificato ATEX scaricabile dalla pagina prodotti del nostro sito: www.siemens.com/LR250. Vedi: Support > Approvals / Certificates. Il certificato IECEx riportato sulla targhetta dell'apparecchio disponibile sul sito internet IECEx. Vedi: http://iecex.iec.ch e cliccare su Ex Equipment Certificates of Conformity, poi digitare il numero di certificato IECEx SIR 08.0107X. Per informazioni sull'alimentazione necessaria vediCurva 2: Antifiamma, sicurezza aumentata, a prova di esplosione pagina 24. Per il cablaggio attenersi alle prescrizioni vigenti. Vedi anche Istruzioni specifiche per aree potenzialmente esplosive pagina 13 e la certificazione ATEX sopraindicata.

Italiano

Pagina IT-12

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

A5E31997170

6. Cablaggio antideflagrante (solo USA/Canada)

Il diagramma di collegamento a prova di esplosione FM/CSA, numero A5E02257843 scaricabile dalla pagina prodotto del nostro sito: www.siemens.com/LR250. Vedi Support > Installation Drawings > Level Measurement > Continuous - Radar. Per informazioni sull'alimentazione necessaria vediCurva 2: Antifiamma, sicurezza aumentata, a prova di esplosione pagina 24.

Istruzioni specifiche per aree potenzialmente esplosive


(Direttiva Europea ATEX 94/9/CE, Allegato II, 1/0/6)
Le istruzioni riportate sotto riguardano specificamente gli apparecchi contrassegnati dai numeri di certificato SIRA 06ATEX2358X, SIRA 08ATEX1301X e SIRA 09ATEX4153X. Per ulteriori informazioni sull'uso e l'assemblaggio, consultare il manuale d'uso integrale. L'apparecchiatura certificata per l'uso come dispositivo di categoria 1GD secondo SIRA 06ATEX2358X; Categoria 1/2 GD, 1D, 2D secondo SIRA 08ATEX1301X; e Categoria 3G secondo SIRA 09ATEX4153X. 3) L'apparecchiatura pu essere utilizzata in atmosfere potenzialmente esplosive composte da gas o vapori pericolosi, con dispositivi del Gruppo IIC, IIB e IIA, classi di temperatura T1, T2, T3 e T4. 4) L'apparecchio caratterizzato da un grado di protezione IP67, classe di temperatura T90 C. Puo' essere utilizzato in atmosfere che contengono polveri infiammabili. 5) Questo apparecchio certificado resiste a temperature ambientali tra 40 C e +80 C. 6) Questo apparecchio non stato definito dispositivo di sicurezza nell'ambito della direttiva 94/9/CE Allegato II, clausola 1,5. Vedi pagina 5. 7) L'installazione e l'ispezione di questo apparecchio deve essere eseguita da personale qualificato in conformit con le normative locali correnti (EN 60079-14 e EN 60079-17 in Europa). 8) L'apparecchiatura non pu essere riparata. 9) Il suffisso 'X' del numero di certificato si riferisce a condizioni speciali per l'uso sicuro. E' importante garantire l'accesso ai certificati per il personale responsabile dell'installazione e dell'ispezione dell'apparecchio. 10) Se esiste la possibilit che l'apparecchio entri in contatto con sostanze aggressive, l'utente tenuto a prendere le debite precauzioni in modo da evitare che esso venga danneggiato, assicurandosi pertanto che non venga compromesso il tipo di protezione. Sostanze aggressive: come ad esempio, liquidi o gas acidi in grado di attaccare e corrodere metalli, o solventi in grado di danneggiare materiali polimerici. Precauzioni: come ad esempio, consultazione delle specifiche dei materiali relative alla resistenza a sostanze chimiche specifiche. 1) 2)

Italiano

A5E31997170

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

Pagina IT-13

Programmazione del SITRANS LR250


Attivazione del SITRANS LR250

Vedi Accesso all'assistente di avvio rapido con il programmatore portatile pagina 18. Vedi Accesso all'assistente di avvio rapido tramite SIMATIC PDM pagina 21.

Collegare l'alimentazione. SITRANS LR250 si avvia automaticamente nel modo di Misura.

Indicatore LCD Modo di misura ( funzionamento normale) 1) 2)


1 2 3 4
(((

1 indicatore di commutazione1) unit lineari/ % 2 modo di funz.mento impostato: livello, spazio, distanza o volume 3 valore misurato (livello o volume, spazio o distanza) 4 unit 5 diagramma a barre indica il livello 6 altra zona del display, indica (su richiesta)2) la temp. dell'elettronica, l'attendibilit dell'eco, la corrente del loop o la distanza 7 area di visualizzazione testo indica le risposte (status message) 5 8 indicatore di stato dell'apparecchio (descritto in dettaglio nel manuale d'uso integrale)

Indicatori guasti S: 0 LOE


La presenza di un guasto provoca l'indicazione del codice e di un messaggio di errore nella zona di testo (7), e l'apparizione dell'icona manutenzione nella zona stato dell'apparecchio (8)

Italiano

Visualizzazione in modo Program


Navigazione numero men Se la barra di men visibile, la item attuale attuale lista di men troppo lunga e non barra di si visualizzano tutti gli item. item men attuale Se la banda situata nel centro banda di item della barra di men, l'item si trova a met della lista. Lampiezza e la relativa posizione della barra di scorrimento allinterno della barra men indicano la lunghezza della lista di men, e la posizione approssimativa dell'item nella lista. Con una barra di scorrimento pi profonda l'utente dispone di meno items tra cui scegliere. Visualizzazione parametri
numero di parametro valore/ selezione del parametro

Modo editazione

nome del parametro

1) 2)

Premere la freccia UP (ALTO) o DOWN (BASSO) per alternare il display. Se viene premuto in tasto. Per maggiori dettagli vedi la tabella Principali funzioni, modo di misura pagina 15.

Pagina IT-14

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

A5E31997170

Programmatore portatile (n. 7ML1930-1BK)


Da ordinare separatamente. 1) 2) 3)

Tasto Principali funzioni, modo di misura


Aggiorna la corrente del loop1). Aggiorna il valore di temperatura (interno custodia)1). Aggiorna il valore di attendibilit dell'eco1).
C

Aggiorna la misura di distanza1). Con Mode si accede al modo PROGRAM2). Con la freccia DESTRA si accede al modo PROGRAM3). Premere i tasti freccia SU o GIU per commutare tra unit e %.

Italiano

Uso del programmatore portatile


Note: Premere Mode per commutare tra il modo di Misura e il modo Program. L'uso dell'apparecchio in modo PROGRAM blocca l'uscita, che non reagisce in tempo reale a eventuali cambi nell'apparecchio. SITRANS LR250 torna automaticamente al modo di misura dopo un periodo di inattivit nel modo PROGRAM (da 15 secondi a 10 minuti a seconda del livello di men).

Men dei parametri

I parametri sono elencati per nome e trattati in gruppi 1. QUICK START 2. SETUP di funzioni. 2.1. DEVICE Per una lista completa di parametri con le relative istruzioni ............. vedi il manuale d'uso integrale. 2.7. LINEARIZATION

2.7.1. VOLUME 2.7.1.1. VESSEL SHAPE

1) 2)

3)

Il valore nuovo indicato nella zona secondaria dell'indicatore LCD. Apre l'ultimo livello di men visualizzato durante l'attuale sequenza di funzionamento, a condizione che l'alimentazione elettrica non sia stata riattivata dopo la chiusura del modo PROGRAM, e che non siano trascorsi pi di 10 minuti dall'accesso al modo PROGRAM. Viene visualizzato il men superiore. Si accede al men superiore.

A5E31997170

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

Pagina IT-15

1.

Attivare il modo PROGRAM


Puntare il programmatore verso il display (distanza max. 300 mm [1 ft.]). FRECCIA destra attiva il modo PROGRAM e accede al men livello 1. Modo accede all'ultimo men visualizzato in modo PROGRAM durante gli ultimi 10 minuti o il men livello 1 se l'alimentazione stata riattivata.

programmatore portatile display

Max. 300 mm (1 ft)

2.

Navigazione: principali funzioni, modo navigazione

Avvertenza: L'acceso rapido ai parametri tramite il programmatore portatile si ottiene premendo Home , e immettendo il numero di men, ad esempio: 2.7.1 (Volume).

Tasto

Nome
Freccia SU o GIU'

Livello men
men o parametro men parametro men o parametro men o parametro men o parametro

Funzione in modo navigazione


Accedere al men o parametro precedente/ successivo. Accedere al primo parametro del men selezionato o aprire il men successivo. Accedere al modo Editazione. Accedere al men genitore. Passare al modo di MISURA. Accedere al men livello superiore: men 1.

Freccia DESTRA

Italiano

Freccia SINISTRA
Modo Home

3.

Editazione in modo PROGRAM


Selezione di un'opzione a) Accedere al parametro desiderato. b) Premere la freccia DESTRA visualizzare il parametro. per
selezione attuale nome del parametro numero di parametro

c)

Premere nuovamente la freccia DESTRA per accedere al modo Editazione . La selezione attuale evidenziata. Accedere a una selezione nuova.

d)

Premere la freccia DESTRA

per

confermare. La nuova selezione viene visualizzata sull'indicatore LCD. Pagina IT-16 SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO A5E31997170

Modificazione di un valore numerico a) Accedere al parametro desiderato. b) Premere nuovamente la freccia DESTRA per accedere al modo Editazione . Il valore attuale evidenziato. Digitare un'altro valore. Premere la freccia DESTRA per confermare. La nuova selezione viene visualizzata sull'indicatore LCD.

nome del parametro

numero di parametro

c) d)

valore attuale

Principali funzioni, modo editazione Tasto Nome


Freccia SU o GIU'

Funzione in modo Editazione


Selezione di opzioni Editazione numerica Selezione di opzioni Accede all'item. - Aumenta o riduce le cifre. - Passa dal pi al meno. - Conferma i dati (accede al parametro in scrittura). - Passa dal modo Editazione al modo Navigazione. - Sposta il cursore un posto verso destra - o con il cursore posizionato su Enter, conferma i dati e passa dal modo Editazione al modo Navigazione. Annulla il modo Editazione senza modificare il parametro. - Sposta il cursore sul pi/meno se si tratta del primo tasto premuto - o sposta il cursore un posto verso sinistra. Cancella il display. Immette una virgola decimale.

Freccia DESTRA

Editazione numerica Selezione di opzioni

Italiano

SINISTRA (freccia)

Editazione numerica Editazione numerica Editazione numerica

Annulla Virgola decimale +o.

Editazione numerica

Cambia il segno del valore impostato.

Numerico

Editazione numerica

Immette il carattere corrispondente.

A5E31997170

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

Pagina IT-17

Accesso all'assistente di avvio rapido con il programmatore portatile


1. Avvio rapido
a) Premere il programmatore verso il display (a una distanza massima di 300 mm (1 ft), e premere la freccia DESTRA men livello 1. b) Premere la freccia DESTRA visualizzare i parametri. c) Premere la freccia DESTRA successivo. d) Per modificare un valore impostato accedere all'item desiderato o digitare un valore nuovo. e) Per confermare un valore, premere la freccia DESTRA accedere all'item successivo. f) Per memorizzare i valori impostati durante l'avvio rapido necessario selezionare Yes (Si) in risposta alla domanda Apply changes? (Confermare modifiche?) nella tappa 1.8. e la freccia GIU' per per accedere al modo Editazione o la freccia GIU' due volte per accedere all'item men 1.1 e per attivare il modo PROGRAM e accedere al

per confermare i valori preimpostati ed accedere direttamente all'item

Italiano

1.1. Lingua Selezionare la lingua utilizzata sul display LCD dell'apparecchio (impostazione immediata).
Opzioni
ENGLISH, DEUTSCH, FRANAIS, ESPAOL

1.2. Materiale Opzioni


LIQUID LIQUID LOW DK (liquido a basso dielettrico)

1.3. Velocit di risposta Regola la risposta di misurazione dell'apparecchio all'interno della portata.
Velocit di risposta
SLOW (Lenta) MED (Media) FAST (Veloce)

Velocit di riempimento/svuotamento
0.1 m/min (0.32 ft/min) 1.0 m/min (3.28 ft/min) 10.0 m/min (32.8 ft/min)

Impostare un valore leggermente superiore alla massima velocit di riempimento o svuotamento (a seconda del valore pi alto).

1.4. Unit Unit di misura sensore. Opzioni


M, CM, MM, FT, IN

Pagina IT-18

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

A5E31997170

1.5. Modo operativo


piano di riferimento sensore superiore spazio piano di riferimento sensore

distanza punto di calibrazione alto

piano di riferimento sensore inferiore

livello punto di calibrazione basso

Funzionamento
NO SERVICE LEVEL (Livello) SPACE (Spazio) DISTANCE (Distanza)
1)

Descrizione
La misura e relativa corrente di loop non vengono aggiornate, e l'apparecchio viene impostato in modo Fail-safe1). Distanza alla superficie del prodotto, misurata dal piano di riferimento sensore (inferiore). Distanza alla superficie del prodotto, misurata dal piano di riferimento sensore (superiore). Distanza alla superficie del prodotto, misurata dal piano di riferimento sensore.

Italiano

Consultare il manuale d'uso integrale per ulteriori informazioni sul modo Fail-safe.

1.6. Punto di calibrazione basso Distanza tra il piano di riferimento sensore e il punto di calibrazione basso: generalmente livello processo vuoto. Per un'illustrazione vedi Modo operativo (1.5.). Valori
Campo di misura: 0.00 ... 20.00 m

1.7. Punto di calibrazione alto Distanza tra il piano di riferimento sensore e il punto di calibrazione alto: generalmente il livello processo pieno. Per un'illustrazione vedi Modo operativo (1.5.) . Valori
Campo di misura: 0.00 ... 20.00 m

1.8. Apply? (confermare modifiche) Per memorizzare i valori impostati (avvio rapido) necessario selezionare Yes (Si). Opzioni
Premere Mode per l'uso. Il display visualizza YES, NO, DONE. (Visualizza DONE (fine) una volta concluso l'Avvio rapido.) per accedere nuovamente al modo di misura. Il SITRANS LR250 pronto

Nota: In presenza di ostacoli fissi all'interno del serbatoio utile la funzione di elimazione automatica falsi echi (AFES). Consultare il manuale d'uso integrale.

A5E31997170

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

Pagina IT-19

Comunicazione SITRANS LR250: HART

E' necessario consultare il manuale d'uso integrale per ottenere la lista di tutti i parametri. Si consiglia di programmare l'apparecchio con SIMATIC Process Device Manager (PDM). Le note esplicative per l'impostazione di apparecchiature HART con SIMATIC PDM sono scaricabili dalla pagina Prodotto del nostro sito: www.siemens.com/LR250 rubrica Support.

SIMATIC PDM
SIMATIC PDM un pacchetto software per la messa in servizio e la manutenzione del SITRANS LR250 e di altre apparecchiature da campo. Per ulteriori informazioni su SIMATIC PDM si prega di consultare le istruzioni d'uso o l'aiuto in linea. (Per ulteriori informazioni sul software potete consultare www.siemens.com/simatic-pdm.) Consente di usare SIMATIC PDM con il modem HART per sistemi operativi Windows 2000 e Windows XP 1). Note: SIMATIC PDM compatibile solo con Windows XP Professional (non Home). Per disattivare la memoria tampone (buffer) necessario disporre di privilegi amministrativi. 1) 2) 3) 4) 5) 6) Cliccare su Start/Settings/Control Panel per configurare l'apparecchio. Doppio-cliccare su System, selezionare Hardware, e cliccare su Device Manager. Aprire la cartella Ports e doppio-cliccare sulla porta di comunicazione utilizzata dall'apparecchio per consultare le propriet (Communications Port Properties.) Selezionare Port Settings e doppio-cliccare su Advanced. Cliccare sulla casella Use FIFO buffers se stata selezionata per disattivare l'uso dei buffer. Cliccare su OK. Chiudere tutte le finestre e riavviare.

Disattivare buffer

Italiano

Aggiornare il file EDD (Electronic Device Description)


La versione 5.2 SP1 di SIMATIC PDM utilizzabile soltanto per le impostazioni di base e la ricerca guasti. Le funzioni avanzate quali l'Assistente di avvio rapido richiedono la versione 6.0 SP4 (o una versione ulteriore).. Il file EDD pu essere scaricato dalla rubrica Device Catalog (SIMATIC PDM), pagina Sensors/Level/Echo/Siemens Milltronics/SITRANS LR250. Consultare la pagina Prodotto sul nostro sito internet: www.siemens.com/LR250, per ottenere l'ultima versione di SIMATIC PDM, del Service Pack (SP) e del hotfix (HF). Vedi Support > Software Downloads

Note:

Per installare un nuovo file EDD:


1) 2)
1)

Il file EDD scaricabile dalla pagina Prodotto del nostro sito internet: www.siemens.com/ LR250 e memorizzare i file sul computer. Estrarre il file zip e salvarlo in modo che sia facilmente accessibile.
Windows un marchio registrato di Microsoft Corporation.

Pagina IT-20

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

A5E31997170

3)

Avviare SIMATIC PDM Manager Device Catalog, accedere al file EDD dezippato e selezionarlo.

Impostazione di un apparecchio
Nota: Cliccando su Cancel (Annulla) durante il telecaricamento dall'apparecchiatura a
SIMATIC PDM si ottiene l'aggiornamento di alcuni parametri. 1) 2) E' importante disporre dell'ultima versione del file EDD. Il file aggiornabile (vediAggiornare il file EDD (Electronic Device Description) pagina 20). Avviare SIMATIC Manager e creare un progetto nuovo per il LR250. Le note esplicative per l'impostazione di apparecchiature HART con SIMATIC PDM sono scaricabili dalla pagina Prodotto del nostro sito: www.siemens.com/LR250. Accedere al men Device Master Reset , selezionare OK per ripristinare i valori di fabbrica. Dopo la sequenza di ripristino trasmettere i parametri al PC/PG. Calibrare l'apparecchio con l'assistente di avvio rapido.

3) 4) 5)

Accesso all'assistente di avvio rapido tramite SIMATIC PDM


Le impostazioni di Avvio Rapido sono correlate tra di loro e le eventuali modifiche vengono applicate solo dopo avere cliccato FINISH e DOWNLOAD dopo la tappa 5, per memorizzare i valori impostati offline e trasferirli all'apparecchio. Cliccare su BACK (Precedente) per correggere un'impostazione o Cancel (Annulla) per uscire dal programma di Avvio rapido. Attivare SIMATIC PDM, accedere al men Device Wizard - Quick Start, e seguire le tappe 1 5.

Note:

Italiano

Avvio rapido

Uso con la tecnologia FDT (Field Device Tool)


FDT uno standard di interfaccia progettato per la messa in servizio e la manutenzione di apparecchiature di campo. PACTware e Fieldcare sono esempi di software FDT. Per configurare un'apparecchiatura con FDT necessario un DTM (Device Type Manager). Gli strumenti Siemens impiegano SIMATIC PDM e una descrizione dell'apparecchiatura (EDD) per SIMATIC DTM. 1) Per prima cosa necessario installare SITRANS DTM. Il programma scaricabile da: http://support.automation.siemens.com. Cliccare su Product Support e accedere a Product Information/Automation Technology/Sensor systems/Process Instrumentation/Software & Communications.

A5E31997170

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

Pagina IT-21

2) Installare la descrizione EDD SITRANS LR250 HART per SITRANS DTM. La descrizione scaricabile dalla pagina prodotto del nostro sito internet: www.siemens.com/LR250. Vedi Support > Software Downloads.

Configurazione di un apparecchiatura con FDT


L'utente ha la possibilit di scaricare una guida di applicazione dal nostro sito internet, sezione Support.

Uso con AMS Device Manager


Il software AMS Device Manager consente di configurare, tarare e rilevare guasti di apparecchiature di campo. Per ulteriori informazioni su AMS Device Manager si prega di consultare le istruzioni operative o l'aiuto in linea. Per ulteriori informazioni sul software: http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD)


SITRANS LR250 richiede l'utilizzo della descrizione EDD per la versione 9.0 di AMS Device Manager.

Configurazione di un'apparecchio con AMS Device Manager


1) Consultare la pagina Prodotto sul nostro sito internet: www.siemens.com/LR250 per ottenere l'ultima descrizione EDD. Vedi Support > Software Downloads e scaricare se necessario. Memorizzare i file sul vostro computer, ed estrarre il file zip assicurandosi che sia facilmente accessibile. Avviare AMS Device Manager - Add Device Type, accedere al file EDD dezippato e selezionarlo.

2)

Italiano

Avviare AMS Device Manager. L'utente dispone di una Guida di applicazione per apparecchiature HART con AMS Device Manager, scaricabile dal nostro sito internet. Vedi Support.

Manutenzione
SITRANS LR250 non richiede interventi di manutenzione o pulizia in condizioni operative normali. In condizioni operative particolari pu essere necessario effettuare una pulizia dell'apparecchio. 1) 2) procedere alla pulizia tenendo conto dei materiali utilizzati per l'antenna e del materiale all'interno del serbatoio. Rimuovere l'apparecchio e pulire l'antenna con un panno e una soluzione detergente appropriata.

Riparazione dell'apparecchio ed esclusione di responsabilit


Per ulteriori informazioni consultare la copertina finale di questo manuale.

Sostituzione dell'antenna o dell'elettronica/custodia


L'antenna dell'apparecchio puo' essere sostituita se danneggiata o difettosa. Non necessaria la ricalibrazione. E' possibile cambiare il tipo di antenna contattando un centro di assistenza Siemens. Anche l'elettronica o la custodia dell'apparecchio possono essere sostituite se danneggiate o difettose. In questo caso necessario verificare la compatibilit dell'antenna per non dover procedere alla ricalibrazione, realizzabile solo da personale autorizzato da Siemens.

Pagina IT-22

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

A5E31997170

Grafico della temperatura massima di lavoro


LR250 - versioni tipo adattatore per flangia
AVVERTENZA: La temperatura interna non deve eccedere 80 C (176 F). Note: Il grafico fornito a scopo informativo ed illustra solo alcuni degli attacchi di processo. (Il grafico NON valido se l'apparecchio SITRANS LR250 installato direttamente su una superficie metallica.) In tale grafico non si tiene conto del calore legato all'esposizione diretta al sole. Parametro 3.2.1 Temperatura interna attuale, controlla la temperatura interna. Massima temperatura processo vs. temperatura ambientale permessa

Temp. ambientale (oC)

temperatura di lavoro

Temperatura di lavoro (oC)

Alimentazione del loop


Italiano

Alimentazione elettrica
Avvertenza: I seguenti grafici si riferiscono ad apparecchi singoli, configurati con il programmatore portatile Siemens. Curva 1: Uso generale, intrinsecamente sicuro, antiscintilla/a energia limitata, non incendivo

Resistenza loop RL

ZONA DI IMPIEGO AUTORIZZATA

Tensione loop VL

Avvertenza: La comunicazione HART richiede una tensione minima di 16.3 VDC (con una resistenza di linea da 220 Ohm).

A5E31997170

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

Pagina IT-23

Curva 2: Antifiamma, sicurezza aumentata, a prova di esplosione Resistenza loop RL

ZONA DI IMPIEGO AUTORIZZATA

Tensione loop VL

Avvertenza: La comunicazione HART richiede una tensione minima di 20.94 VDC (con una resistenza di linea da 220 Ohm).

Messa in servizio
L'apparecchio consuma meno di 3,6 mA durante la messa in servizio. La prima misura si ottiene dopo un intervallo inferiore a 50 secondi.

Italiano
Pagina IT-24

SITRANS LR250 (HART) MANUALE DI AVVIO RAPIDO

A5E31997170

SITRANS LR250 (HART) Quick Start Handleiding


Deze handleiding beschrijft de essentile kenmerken en functies van de SITRANS LR250 (HART 1)). Wij raden u sterk aan de uitgebreide versie van de handleiding aan te schaffen zodat u uw instrument zo optimaal mogelijk kunt gebruiken. U kunt de volledige handleiding downloaden op de SITRANS LR250 productpagina van onze website: www.siemens.com/LR250. De gedrukte handleiding is verkrijgbaar via uw lokale Siemens Milltronics vertegenwoordiging. Vragen omtrent de inhoud van deze handleiding kunnen worden gericht aan: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 Email: techpubs.smpi@siemens.com

Auteursrecht Siemens AG 2013. Alle rechten voorbehouden


Wij raden gebruikers aan geautoriseerde, ingebonden gebruikershandleidingen te kopen, of om de elektronische versies te raadplegen, zoals ontworpen en goedgekeurd door Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments is niet aansprakelijk voor de inhoud van gedeeltelijk, of geheel gekopieerde versies, gebonden of elektronisch.

Disclaimer
Hoewel we de inhoud van deze gebruikershandleiding hebben geverifieerd aan de omschreven instrumentatie, kunnen desondanks afwijkingen voorkomen. Wij kunnen derhalve niet een volledige overeenstemming garanderen. De inhoud van deze handleiding wordt regelmatig herzien, en correcties worden opgenomen in volgende uitgaven. Wij houden ons aanbevolen voor suggesties ter verbetering. Technische gegevens kunnen worden gewijzigd.

MILLTRONICS is een geregistreerd handelsmerk van Siemens Milltronics Process Instruments.

Technische ondersteuning

Ondersteuning is 24-uur per dag beschikbaar. Ga na de hierna vermelde website om het adres, telefoonnummer en faxnummer van uw lokale Siemens Automation kantoor te vinden: www.siemens.com/automation/partner: Klik op het tabblad Contact, selecteer Service, klik vervolgens weer op Service om uw productgroep te vinden (+Automation Technology > +Sensor Systems > +Process Instrumentation > +Level Measurement > +Continuous). Kies Radar. Selecteer het land, gevolgd door plaats/regio. Selecteer Technical Support onder Service. Ga voor on-line technische ondersteuning naar: www.siemens.com/automation/support-request Voer de naam van het apparaat in (SITRANS LR250) of het bestelnummer, klik vervolgens op Search, en selecteer het juiste producttype. Klik op Next. Voer een sleutelwoord in dat uw vraag omschrijft. Blader vervolgens door de relevante documentatie, of klik op Next om een omschrijving van uw vraag op te sturen naar het technische ondersteuningspersoneel van Siemens. Siemens IA/DT Technical Support Center: telefoon +49 (0)911 895 7222
1)

Nederlands

HARTis een geregistreerd handelsmerk van de HART Communications Foundation.

A5E31997170

SITRANS LR250 (HART) QUICK START Handleiding

Pagina NL-1

Veiligheidsrichtlijnen

Waarschuwingsmeldingen moeten worden aangehouden om de eigen veiligheid en die van anderen te waarborgen en om het product en de aangesloten apparatuur te beschermen. Deze waarschuwingsmeldingen gaan vergezeld met een verduidelijking van de mate van voorzichtigheid die moet worden aangehouden. Het WAARSCHUWING symbool is gerelateerd aan een voorzichtig-symbool op het product en betekent dat het niet aanhouden van de noodzakelijke voorzorgsmaatregelen kan leiden tot ernstig of dodelijk letsel en/of aanzienlijke materile schade. WAARSCHUWING symbool gebruikt wanneer er geen corresponderend voorzichtig symbool op het product is, betekent dat het niet aanhouden van de noodzakelijke voorzorgsmaatregelen kan leiden tot ernstig of dodelijk letsel en/of aanzienlijke materile schade. Opmerking: geeft belangrijke informatie aan over het product of dat deel van de gebruikershandleiding.

FCC conformiteit
Alleen voor US installaties: Federal Communications Commission regelgeving
Waarschuwing: wijzigingen of modificaties die niet uitdrukkelijk zijn goedgekeurd door Siemens Milltronics kunnen de toestemming om de apparatuur te bedienen ongeldig maken. Opmerkingen: Deze apparatuur is getest, en voldoet aan de grenswaarden voor een Class A digital device, zoals vastgelegd in Part 15 van de FCC regelgeving. Deze grenswaarden zijn vastgelegd voor het bieden van een redelijke bescherming tegen schadelijke interferentie wanneer de apparatuur wordt gebruikt in een commercile omgeving. Deze apparatuur genereert, gebruikt en kan radiofrequentie-energie uitzenden en kan, indien niet gebruikt en genstalleerd conform de gebruikershandleiding schadelijke interferentie veroorzaken voor radiocommunicatie. Gebruik van deze apparatuur in woonomgevingen zal hoogstwaarschijnlijk leiden tot schadelijke interferentie voor radiocommunicatie. In dat geval moet de gebruiker de interferentie voor eigen rekening corrigeren.

SITRANS LR250
Nederlands
Waarschuwing: de SITRANS LR250 mag uitsluitend worden gebruikt op de manier zoals beschreven in deze handleiding, anders kan de bescherming die deze apparatuur biedt in gevaar komen. Opmerking: dit product is bedoeld voor gebruik in industrile omgevingen. Gebruik van deze apparatuur in woonomgevingen kan leiden tot interferentie van communicatie op verschillende frequenties. SITRANS LR250 is een 2-draads, 25 GHz pulsradar niveautransmitter voor continue bewaking van vloeistoffen en slurries in opslagtanks, waaronder hoge druk en hoge temperatuur configuraties, tot een bereik van 20 m (66ft). Het instrument is ideaal voor kleine tanks en media met een lage dilektrische constante. Het apparaat bestaat uit een elektronica circuit gekoppeld aan een hoornantenne en een schroefdraad- of flens procesaansluiting. Een PVDF antenne met schroefdraadaansluiting is eveneens leverbaar. SITRANS LR250 ondersteunt het HART communicatieprotocol en SIMATIC PDM software. Signalen worden verwerkt m.b.v. Process Intelligence.

Pagina NL-2

SITRANS LR250 (HART) QUICK START Handleiding

A5E31997170

Specificaties

Raadpleeg voor een compleet overzicht de SITRANS LR250 (HART) gebruikershandleiding. Voor informatie over goedkeuringen, zie Goedkeuringen op pagina 4.

Omgevings-/bedrijfstemperatuur
Opmerkingen:
De maximale temperatuur is afhankelijk van de procesaansluiting, antennematerialen en tankdruk. Zie Maximum procestemperatuurgrafiek op pagina 23. Zie voor meer gedetailleerde informatie de Proces Druk/Temperatuur derating curves in de uitgebreide handleiding. Procestemperatuur en drukspecificaties zijn afhankelijk van informatie op het typeplaatje op de procesaansluiting. De referentietekening die is vermeld op de tag is beschikbaar op de productpagina van onze website op www.siemens.com/LR250, onder Support > Installation Drawings > Level Measurement > Continuous - Radar. Aanvullende informatie over procesaansluitingen is beschikbaar op de installatietekeningen onder Process Connection Diagrams. De signaalamplitude neemt toe met de hoorndiameter, dus gebruik de grootste maat die praktisch haalbaar is. Optionele verlengingen kunnen worden genstalleerd onder de schroefdraden. Zie Maximum procestemperatuurgrafiek op pagina 23, voor meer informatie.
Omgevingstemperatuur (rondom behuizing) 40 C tot 80 C (40 F tot 176 F) procestemperatuur bij procesaansluiting: standaard hoornantenne: - met FKM O-ring: 40 oC tot +200 oC (40 oF tot +392 oF) - met FFKM O-ring: 20 oC tot +200 oC (4 oF tot +392 oF) 2" NPT / BSPT / G-schroefdraad PVDF antenne: 40 tot +80 C (40 tot +176 F) instrument typeplaatje

proces identificatieplaatje

Voeding
General purpose Intrinsiekveilig Niet vonkend/beperkte energie Niet brandveroorzakend (alleen FM/CSA US/Canada) Vlambestendig Verhoogde veiligheid Explosieveilig (alleen FM/CSA US/Canada) maximaal 30 V DC 4 tot 20 mA Max. opstartstroom: Nominaal 24 V DC bij 550 Ohm Nominaal 24 V DC bij 250 Ohm

Nederlands

Zie Gedrag bij opstarten op pagina 24

A5E31997170

SITRANS LR250 (HART) QUICK START Handleiding

Pagina NL-3

Goedkeuringen
Opmerking: Het typeplaatje van het instrument geeft een opsomming van alle goedkeuringen voor uw instrument. General CSAUS/C, FM, CE, C-TICK, KC Radio Europa (R&TTE), FCC, Industry Canada Explosiegevaarlijk Intrinsiekveilig 1) (Europa) ATEX II 1G, EEx ia IIC T4 ATEX II 1D, EEx tD A20 IP67 T90 C (Internationaal) IECEx SIR 05.0031X, Ex ia IIC T4 Ex tD A20 IP67 T90 C (US/Canada) FM/CSA Class I, Div. 1, Groups A, B, C, D Class II, Div. 1, Groups E, F, G Class III T4 (Brazili) INMETRO: DNV 12.0087 X Ex ia IIC T4 Ga Ex ia ta IIIC T90 C Da IP65/IP67 -40 C Ta +80 C DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

(China) NEPSI Ex ia IIC T4/ DIP A20 TA, T90 C IP67 Niet vonkend/Energiebeperkt 2) (Europa) ATEX II 3 G Ex nA/nL IIC T4 Gc (China) NEPSI Ex nA II T4/ Ex nL IIC T4 Niet brandveroorzakend 3) (US/Canada) Vlambestendig 4) (Europa/Internationaal) FM/CSA Class I, Div. 2, Groups A, B, C, D T5 ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex dmbia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C INMETRO: DNV 12.0088 X Ex d ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-1:2009, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005
1) 2) 3) 4)

Nederlands

(Brazili)

Zie Intrinsiekveilige bekabeling op pagina 9. Zie Niet vonkend / energiebeperkte bekabeling op pagina 11. Zie Niet brandveroozakende bekabeling op pagina 11. Zie Vlambestendige bekabeling op pagina 12.

Pagina NL-4

SITRANS LR250 (HART) QUICK START Handleiding

A5E31997170

Gevaarlijk (vervolgd) Verhoogde veiligheid 1) (Europa/Internationaal)

(Brazili)

ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex embia IIC T4 Ga/Gb Ex tD A20 IP67 T90 C INMETRO: DNV 12.0088 X Ex e ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-7:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

Vlambestendig/Verhoogde veiligheid (China) NEPSI Ex dmbia IIC T4/ Ex embia IIC T4/ DIP A20 TA, T90 oC IP67 2) Explosieveilig (US/Canada) FM/CSA Class I, Div. 1, Groups A, B, C, D Class II, Div. 1, Groups E, F, G Class III T4 Marine Lloyds Register of Shipping ABS typegoedkeuring

Druktoepassing
WAARSCHUWINGEN:

probeer nooit de procesaansluiting los te draaien, te verwijderen of te demonteren terwijl de tank onder druk staat. De gebruiker is verantwoordelijk voor het kiezen van bevestingingsmateriaal en pakkingen die binnen de flensmaat vallen, voldoen aan het bedoelde gebruik en geschiktheid voor de bedrijfsomstandigheden. Onjuiste installatie kan leiden tot verlies aan procesdruk.
Opmerkingen: Het procesaansluitingtypeplaatje moet bij de procesdrukgrenseenheid blijven2). In geval dat deze eenheid wordt vervangen, moet het procesaansluiting typeplaatje worden overgebracht op de vervangende eenheid. SITRANS LR250 eenheden zijn hydrostatisch getest, waarbij (ruim) wordt voldaan aan de vereisten van de ASME "Boiler and Pressure Vessel" norm en de Europese richtlijn voor drukapparatuur 97/23/EG.

Nederlands

1) 2)

Zie Verhoogde veiligheid bekabeling op pagina 12. Zie Explosieveilige bekabeling (alleen US/Canada) op pagina 13.

A5E31997170

SITRANS LR250 (HART) QUICK START Handleiding

Pagina NL-5

Richtlijn drukapparatuur, PED, 97/23/EC 1)


Siemens niveautransmitters met flens-, schroefdraad of sanitaire procesaansluitingen hebben zelf geen drukhoudende behuizing en vallen dus niet onder de richtlijn drukapparatuur als zijnde druk- of veiligheidstoebehoren, (raadpleeg de EU commissie richtlijn 1/8).

Installatie
WAARSCHUWINGEN: De installatie mag uitsluitend worden uitgevoerd door gekwalificeerd personeel en in overeenstemming met lokale regelgeving. Hanteer het instrument bij de behuizing, niet aan het procesaansluiting typeplaatje, om beschadiging te voorkomen. Wees extra voorzichtig bij het omgaan met de PVDF antenne met schroefdraad. Elke beschadiging aan het antenne-oppervlak, met name aan de punt, kan een negatieve invloed hebben op de prestaties. Constructiematerialen zijn gekozen op basis van hun chemische compatibiliteit (of inertheid) voor algemene doeleinden. Controleer voor blootstelling aan specifieke omgevingen, de chemische compatibiliteitstabellen voorafgaande aan de installatie. Opmerkingen:

Voor de EU en lidstaten moet de installatie conform ETSI EN 302372 zijn. Raadpleeg het typeplaatje van het instrument voor goedkeuringsinformatie. De serienummers die zijn ingeslagen op het huis van elke procesaansluiting vormen een uniek identificatienummer en geven de productiedatum aan. Voorbeeld: MMDDYY XXX (waarbij MM = maand, DD = dag, YY = jaar, en XXX= sequentieel geproduceerde unit) Andere markeringen (voor zover de ruimte het toelaat) geven flensconfiguratie aan, grootte, drukklasse, materiaal, en materiaalhittecode.

Hoornontwerp
PVDF antenne met schroefdraad
min. afstand: 10 mm (0.4")

RVS hoornantenne
min. afstand: 10 mm (0.4")

Nederlands

Min. dia. 51 mm (2")

Het uiteinde van de antenne moet minimaal 10 mm (0.4) uitsteken om stoorecho's te voorkomen die door de hoorn worden gereflecteerd. De minimum aanbevolen hoorndiameter voor de PVDF antenne met schroefdraad is 51 mm (2"). Een antenneverlenging (100 mm/ 3.93") is beschikbaar voor elke versie, behalve voor de PVDF antenne met schroefdraad.

1)

De proces drukgrenseenheid bestaat uit de componenten die fungeren als een barrire tegen drukverlies uit het procesvat: d.w.z. de combinatie van procesaansluiting huis en zender, maar normaal gesproken zonder de elektrische behuizing.

Pagina NL-6

SITRANS LR250 (HART) QUICK START Handleiding

A5E31997170

Locatie van de sok


Voorkom centrale locaties op hoge, smalle tanks, hetgeen stoorecho's kan opleveren.

Omgeving
Zorg voor een omgeving die geschikt is voor de beschermingsklasse van de behuizing en de gebruikte constructiematerialen. Zorg voor een zonnedak wanneer het instrument in direct zonlicht wordt opgesteld.
voorkeur niet wenselijk

Bundelhoek
Bundelhoek is de breedte van de conus waar de energiedichtheid half die van de piekenergiedichtheid is. De piekenergiedichtheid bevindt zicht direct voor en in lijn met de antenne. Er wordt een signaal uitgezonden buiten de bundelhoek, daarom kunnen valse doelen worden gedetecteerd.

zendconus bundelhoek: PVDF antenne met schroefdraad = 19 1.5" hoorn = 19 2" hoorn = 15 bundel- 3" hoorn = 10 4" hoorn = 8 hoek

Zendconus
Houd de zendconus interferentievrij door deze uit de buurt te houden van ladders, leidingen, I-balken of vulstromen.

Toegang voor programmering


Zorg voor gemakkelijke toegang om het display te bekijken en voor programmering via de handprogrammer.

Montage-instructies
Waarschuwing: voor druktoepassingen is het nodig om PTFE-tape (of een ander schroefdraadafdichtmiddel) te gebruiken en vervolgens de procesaansluiting meer dan handvast aandraaien. Het maximum aandraaimoment is 40 N-m (30 ft.lbs.). Opmerking: er is geen grens aan het aantal keer dat een apparaat kan worden geroteerd zonder deze te beschadigen.

Nederlands

Uitvoeringen met schroefdraad


1) 2) Controleer vr het aanbrengen van het instrument in de montage-aansluiting of de schroefdraadaansluitingen van hetzelfde type zijn zodat ze niet beschadigd raken. Schroef het instrument gewoon in de procesaansluiting en draai deze vervolgens handvast, of gebruik een sleutel. Zie voor druktoepassingen de waarschuwingsopmerkingen hierboven.

Uitvoeringen met flens


Waarschuwing: de gebruiker is verantwoordelijk voor het kiezen van bevestigingsmateriaal en pakkingen die binnen de flensmaat vallen, voldoen aan het bedoelde gebruik en geschiktheid voor de bedrijfsomstandigheden. A5E31997170 SITRANS LR250 (HART) QUICK START Handleiding Pagina NL-7

Bedrading
Voeding
WAARSCHUWINGEN: De DC-ingangsklemmen moeten worden gevoed uit een bron die galvanisch is gescheiden tussen in- en uitgang, teneinde te voldoen aan de geldende veiligheidsvereisten van IEC 61010-1. Alle veldbekabeling moet zijn voorzien van een isolatie die geschikt is voor nominale spanningen.

Aansluiten van de SITRANS LR250


WAARSCHUWINGEN: Controleer het typeplaatje van het instrument en het procestypeplaatje, om de toelatingen te verifiren. Gebruik de juiste doorvoerafdichtingen om de IP- of NEMA-beschermingsklasse te handhaven. Zie Bedrading configuraties voor installaties in explosiegevaarlijke gebieden op pagina 9. Opmerkingen: Gebruik twisted pair kabel: AWG 22 tot 14 (0,34 mm2 tot 2,5 mm2). Er kunnen aparte kabels en doorvoeren nodig zijn om te voldoen aan de standaard instrumentatie-bekabelingsnormen of elektrische regelgeving.
Gebruik een 2 mm inbussleutel voor het losdraaien van de borgschroef van het deksel.

plug (IP 68)

kabelafscherming (indien gebruikt)

Nederlands

optionele kabelwartel1) 2) (of NPT kabeldoorvoer2))

1) 2) 3) 4) 5)

Strip de kabelmantel over een lengte van ca. 70 mm (2.75") vanaf het uiteinde van de kabel en voer de draden door de wartel. 1) 2) Sluit de draden aan op de klemmen zoals weergegeven: de polariteit wordt aangegeven op het klemmenblok. Aard het instrument conform de lokale regelgeving. Draai de wartel aan om een goede afdichting te vormen. Sluit het deksel voorafgaande aan programmering en configuratie van het apparaat.

1) 2)

Kan met het apparaat zijn meegeleverd. Wanneer de kabel door een doorvoer wordt gevoerd, gebruik dan uitsluitend wartels van de goede maat die zijn goedgekeurd voor waterdichte toepassingen.

Pagina NL-8

SITRANS LR250 (HART) QUICK START Handleiding

A5E31997170

Aansluiten van HART


Typische PLC/mA configuratie met HART 1) 2)
actieve PLC Voeding R= 250 HART modem SITRANS LR250 HART communicator

Opmerkingen: Afhankelijk van het systeemontwerp, kan de voeding zijn gescheiden van de PLC of er in zijn gentegreerd. HART weerstand1) moet zijn beperkt, naar gelang het toegestane bedrijfsgebied2). Sluit niet meer dan n HART programmeerapparaat tegelijkertijd aan.

Bedrading configuraties voor installaties in explosiegevaarlijke gebieden


Er zijn zes bekabelingsopties voor installaties in gevaarlijke zones. Controleer in alle gevallen de typeplaat van de procesaansluiting, om de toelatingen te verifiren.

1. Intrinsiekveilige bekabeling
Apparaat typeplaatje (ATEX/IECEx/INMETRO/C-TICK)

Het ATEX certificaat dat is vermeld op de naamplaat kan worden gedownload vanaf de productpagina van onze website: www.siemens.com/LR250. Ga naar Support > Approvals / Certificates. Het IECEx certificaat vermeld op de naamplaat kan worden bekeken op de IECEx website. Ga naar: http://iecex.iec.ch en klik opEx Equipment Certificates of Conformity en voer vervolgens het certificaatnummer IECEx SIR 05.0031X in.

Nederlands

1) 2)

Totale lusweerstand = kabelweerstand 250 Ohm (weerstand). Zie of Curve 1: General Purpose, intrinsiekveilig, niet vonkend / beperkte energie / niet brandveroozakend op pagina 23 of Curve 2: Vlambestendig, verhoogde veiligheid, explosieveilig op pagina 24.

A5E31997170

SITRANS LR250 (HART) QUICK START Handleiding

Pagina NL-9

Intrinsiekveilige bekabeling (vervolgd)


Apparaat typeplaatje (FM/CSA)

Het FM/CSA intrinsiekveilig aansluitschema nummer 23650653 kan worden gedownload vanaf de productpagina van onze website: www.siemens.com/LR250. Ga naar Support > Installation Drawings > Level Measurement > Continuous - Radar. Voor voedingsvereisten zieCurve 1: General Purpose, intrinsiekveilig, niet vonkend / beperkte energie / niet brandveroozakend op pagina 23. Voor bedradingsvereisten: Volg de lokale regelgeving. Goedgekeurde stofdichte en waterdichten doorvoerafdichtingen zijn vereist voor buiten NEMA 4X / type 4X / NEMA 6, IP67, IP68 locaties. Raadpleeg Instructies speciaal voor installatie in explosiegevaarlijke zones: op pagina 13. Aanbevolen intrinsiekveilige barrires worden vermeld onder Passieve shunt-diode barrires op pagina 10 en Actieve barrires (herhalende barrires) op pagina 11. Opmerking:voor de keuze van een geschikte PLC ingangsmodule, voeding of barrire is kennis vereist van intrinsieke veiligheid en de toepassing. Het is de verantwoordelijkheid van de installateur om er voor te zorgen dat de intrinsiekveilige installatie voldoet aan zowel de toelatingsvereisten van het apparaat als de geldende landelijke normen.

Passieve shunt-diode barrires


Opmerking: een goed geregelde voedingsspanning is vereist. Fabrikaat
MTL MTL Stahl Stahl

Onderdeelnummer
787SP+ (Dubbel kanaal) 7787P+ (Dubbel kanaal) 9001/01-280-100-10 (Enkel kanaal) 9002/01-280-110-10 (Dubbel kanaal)

Nederlands

Voer, om te waarborgen dat de veiligheidsomschrijving van de barrire geschikt is voor de LR250 intrinsiekveilige ingangsparameters, de volgende berekeningen uit: Re-e = max. end-to-end weerstand van de barrire Rloop = lusweerstand (totale kabelweerstand plus eventuele aanvullende weerstand, zoals bijvoorbeeld PLC ingangen en/of displays). Vbarrier = waarde van enige a-lineaire spanning daalt als gevolg van de barrire 1) Bepaal de waarde voor Re-e uit het datablad. 2) Bereken de totale waarde voor Rloop. 3) Bereken Rworking = Re-e + Rloop.

Selecteren van een passieve barrire voor de SITRANS LR250

Pagina NL-10

SITRANS LR250 (HART) QUICK START Handleiding

A5E31997170

Selecteren van een passieve barrire (vervolgd)


4) 5) 6) Bepaal de waarde van Vbarrier uit het barrire datablad (bijvoorbeeld spanningsval als gevolg van diodes). Bereken Vworking = Vsupply Vbarrier. Gebruik de waarden voor Vworking en Rworking om te bevestigen dat de werking binnen het gearceerde gebied van figuur Curve 1: General Purpose, intrinsiekveilig, niet vonkend / beperkte energie / niet brandveroozakend op pagina 23 valt. Onderdeelnummer
706 7206 SITRANS I 7NG4124 9001/51-280-110-14 E02009 E02010

Actieve barrires (herhalende barrires)


Producent
MTL MTL Siemens Stahl MTL MTL

2. Niet vonkend / energiebeperkte bekabeling

Het ATEX certificaat dat is vermeld op de naamplaat kan worden gedownload vanaf de productpagina van onze website: www.siemens.com/LR250. Ga naar Support > Approvals / Certificates. Zie voor voedingsvereisten Curve 1: General Purpose, intrinsiekveilig, niet vonkend / beperkte energie / niet brandveroozakend op pagina 23. Hanteer de lokale regelgeving voor de bekabelingsvereisten.

3. Niet brandveroozakende bekabeling


Nederlands

FM/CSA Class 1, Div 2 aansluitschema nummer 23650673 kan worden gedownload vanaf de productpagina van onze website: www.siemens.com/LR250. Ga naar Support > Installation Drawings > Level Measurement > Continuous - Radar. Voor voedingsvereisten zieCurve 1: General Purpose, intrinsiekveilig, niet vonkend / beperkte energie / niet brandveroozakend op pagina 23.

A5E31997170

SITRANS LR250 (HART) QUICK START Handleiding

Pagina NL-11

4. Vlambestendige bekabeling

Het ATEX certificaat dat is vermeld op de naamplaat kan worden gedownload vanaf de productpagina van onze website: www.siemens.com/LR250. Ga naar: Support > Approvals / Certificates. Het IECEx certificaat vermeld op de naamplaat kan worden bekeken op de IECEx website. Ga naar: http://iecex.iec.ch en klik opEx Equipment Certificates of Conformity en voer vervolgens het certificaatnummer IECEx SIR 08.0107X in. Voor voedingsvereisten zieCurve 2: Vlambestendig, verhoogde veiligheid, explosieveilig op pagina 24. Hanteer de lokale regelgeving voor de bekabelingsvereisten. Zie ook Instructies speciaal voor installatie in explosiegevaarlijke zones: op pagina 13 en het hierboven vermelde ATEX certificaat.

5. Verhoogde veiligheid bekabeling

Het ATEX certificaat dat is vermeld op de naamplaat kan worden gedownload vanaf de productpagina van onze website: www.siemens.com/LR250. Ga naar: Support > Approvals / Certificates. Het IECEx certificaat vermeld op de naamplaat kan worden bekeken op de IECEx website. Ga naar: http://iecex.iec.ch en klik opEx Equipment Certificates of Conformity en voer vervolgens het certificaatnummer IECEx SIR 08.0107X in. Voor voedingsvereisten zieCurve 2: Vlambestendig, verhoogde veiligheid, explosieveilig op pagina 24. Hanteer de lokale regelgeving voor de bekabelingsvereisten. Zie ook Instructies speciaal voor installatie in explosiegevaarlijke zones: op pagina 13 en het hierboven vermelde ATEX certificaat.

Nederlands

Pagina NL-12

SITRANS LR250 (HART) QUICK START Handleiding

A5E31997170

6. Explosieveilige bekabeling (alleen US/Canada)

FM/CSA Explosieveilig aansluitschema nummer A5E02257843 kan worden gedownload vanaf de productpagina van onze website: www.siemens.com/LR250. Ga naar Support > Installation Drawings > Level Measurement > Continuous - Radar. Voor voedingsvereisten zieCurve 2: Vlambestendig, verhoogde veiligheid, explosieveilig op pagina 24.

Instructies speciaal voor installatie in explosiegevaarlijke zones:


(Referentie Europese ATEX richtlijn 94/9/EC, Annex II, 1/0/6)
De volgende instructies gelden voor apparatuur die wordt gedekt door certificaatnummer SIRA 06ATEX2358X, SIRA 08ATEX1301X en SIRA 09ATEX4153X. Raadpleeg de volledige handleiding voor gebruik en montage. De apparatuur is gecertificeerd voor gebruik als Category 1GD apparatuur volgens SIRA 06ATEX2358X; Category 1/2 GD, 1D, 2D apparatuur volgens SIRA 08ATEX1301X; en Category 3G apparatuur volgens SIRA 09ATEX4153X. 3) De apparatuur mag worden gebruikt met ontbrandbare gassen en dampen met apparaatgroepen IIC, IIB, en IIA, en temperatuurklassen T1, T2, T3 en T4. 4) De apparatuur heeft een beschermingsklasse IP67 en een temperatuurklasse van T90C en mag worden gebruikt bij brandbare stofproducten. 5) De apparatuur is gecertificeerd voor gebruik in een omgevingstemperatuurbereik van 40C tot +80C. 6) De apparatuur is niet getest als veiligheidstoestel (zoals omschreven in richtlijn 94/9/EC Annex II, bepaling 1.5). Zie Richtlijn drukapparatuur, PED, 97/23/EC op pagina 6. 7) Installatie en inspectie van deze apparatuur moet worden uitgevoerd door voldoende opgeleid personeel, conform de geldende normen (EN 60079-14 en EN 60079-17 in Europa). 8) De apparatuur kan niet worden gerepareerd. 9) De certificaatnummers hebben een X suffix, hetgeen aangeeft dat speciale condities van toepassing zijn voor veilig gebruik. Diegene die deze apparatuur installeren of inspecteren moeten toegang hebben tot de certificaten. 10) Wanneer de apparatuur in contact kan komen met agressieve stoffen, dan is het de verantwoordelijkheid van de gebruiker om de nodige maatregelen te nemen om te voorkomen dat het instrument negatief wordt benvloed, en zo te waarborgen dat de typeclassificatie niet in gevaar komt. - Agressieve stoffen: bijvoorbeeld zure vloeistoffen of gassen die metalen aantasten, of oplosmiddelen die invloed kunnen hebben op polymeren. - Geschikte voorzorgsmaatregelen: bijvoorbeeld, vaststellen aan de hand van het datablad van het materiaal dat het bestand is tegen specifieke chemische stoffen. 1) 2)

Nederlands

A5E31997170

SITRANS LR250 (HART) QUICK START Handleiding

Pagina NL-13

Programmeren van de SITRANS LR250


Activeren van de SITRANS LR250

Zie Quick Start Wizard via de handprogrammer op pagina 18. Zie Quick Start Wizard via SIMATIC PDM op pagina 21.

Inschakelen van het instrument. SITRANS LR250 start automatisch op in de meetmodus.

Het LCD Meetmodus display (normaal bedrijf) 1) 2)


1 2 3 4

1 toggle indicator1) voor lineaire eenheden of % 2 geselecteerde bedrijf: Niveau, ruimte, afstand of volume 3 meetwaarde (niveau of volume, ruimte of afstand) 4 eenheden 5 bargraph geeft niveau aan 6 secondaire gebied op verzoek2) temperatuur van de elektronica, echo confidence, lusstroom of afstand 7 tekstgebied toont statusmeldingen 8 device status indicator (zie voor details de uitgebreide handleiding)

Fout aanwezigheidsindicatoren S: 0 LOE


Wanneer een fout aanwezig is wordt een foutmelding weergegeven in het tekstgebied (7) en een service-vereist pictogram verschijnt in de apparaatstatuslocatie (8)

PROGRAM modus display Navigatieweergave


huidige nummer Een zichtbare menubalk geeft aan menu dat de menulijst te lang is om alle menubalk huidige items in n keer weer te geven. item item Een band halverwege de menubalk band geeft aan dat het huidige item halverwege de lijst is. De diepte en relatieve positie van de item-band op de menubalk geeft de lengte aan van de menulijst en bij benadering de positie van het huidige item in de lijst. Een diepere band geeft minder items aan. huidige item-

Nederlands

Parameterweergave
parameternaam parameter nummer parameterwaarde/selectie

Bewerken weergave

1) 2)

Druk op de OMHOOG of OMLAAG pijl om te schakelen. Als response op een ingedrukte toets. Raadpleeg voor details tabel Functie in de meetmodus op pagina 15.

Pagina NL-14

SITRANS LR250 (HART) QUICK START Handleiding

A5E31997170

Handprogrammer (Onderdeelnr. 7ML1930-1BK)


Afzonderlijke besteld. 1) 2) 3)

Toets Functie in de meetmodus


Update de lusstroom1). Update de temperatuuraanwijzing van de temperatuur in de behuizing1). Update de echobetrouwbaarheidswaarde1).
C

Update de afstandsmeting1). Modus opent PROGRAM modus2). Pijl naar RECHTS opent PROGRAM modus3). OMHOOG of OMLAAG pijl schakelt tussen lineaire eenheden en %.

Programmering via de handprogrammer


Opmerkingen: Druk op Modus om te schakelen tussen de Meet- en Programmeer modus. Wanneer het apparaat zich in de PROGRAM modus bevindt, blijft de uitgang vast en reageert deze niet op veranderingen in het apparaat. De SITRANS LR250 keert automatisch terug naar de Meetmodus na een periode van inactiviteit in de PROGRAM modus (tussen 15 seconden en 10 minuten, afhankelijk van het menuniveau).

Parametermenu's

Nederlands

Parameters worden gedentificeerd op basis van de 1. QUICK START 2. SETUP naam en in groepen gerangschikt. 2.1. DEVICE Raadpleeg voor een volledig overzicht van parameters met ............. instructies de uitgebreide handleiding. 2.7. LINEARIZATION

2.7.1. Volume 2.7.1.1. vorm van de tank

1) 2)

3)

Nieuwe waarde wordt op LCD weergegeven in secundair gebied. Opent het menuniveau dat het laatst is weergegeven tijdens deze inschakelcyclus, tenzij de voeding is geschakeld sinds de PROGRAMMEER modus of dat meer dan 10 minuten zijn verstreken sinds de PROGRAMMEER modus is gebruikt. Vervolgens wordt het menu van het hoogste niveau weergegeven. Opent het top niveau menu

A5E31997170

SITRANS LR250 (HART) QUICK START Handleiding

Pagina NL-15

1.

Naar de PROGRAM modus


Richt de handprogrammer op het display (maximale afstand 300 mm [1 ft.]). Pijl naar RECHTS activeert de PROGRAM modus en opent menuniveau 1. Modus opent het menuniveau dat het laatst is weergegeven in de PROGRAM modus binnen de laatste 10 minuten, of menuniveau 1 wanneer de voeding sindsdien is geschakeld.

handprogrammer display

max. 300 mm (1 ft)

2.

Navigatie: toetsfuncties in Navigatie modus


, voer vervolgens het menunummer in, bijvoorbeeld: 2.7.1 (Volume).

Opmerking: druk voor een snelle toegang tot de parameters via handprogrammer op Home

Toets

Naam
Pijl omhoog of omlaag

Menuniveau
menu of parameter menu parameter menu of parameter menu of parameter menu of parameter

Functie in navigatiemodus
Blader naar vorig of volgende menu-item of parameter. Ga naar eerste parameter in het geselecteerde menu, of open volgende menu. Open Edit modus. Open bovenliggende menu. Schakel naar MEET modus. Open het menu van het hoogste niveau: menu 1.

Pijl naar RECHTS Pijl naar LINKS


Modus Home

3. Nederlands

Bewerken in PROGRAM modus Selecteren van een vermelde optie


a) b) Navigeer naar de gewenste parameter. Druk op de pijl naar RECHTS parameterweergave te openen. om

parameternaam parameternummer

huidige selectie

c)

Druk weer op pijl naar RECHTS om de Edit modus te openen. De huidige selectie is gemarkeerd. Blader naar een nieuwe selectie. Druk op de pijl naar RECHTS om deze te accepteren. Het display keert terug naar de weergave van de parameters en toont de nieuwe selectie. SITRANS LR250 (HART) QUICK START Handleiding A5E31997170

d)

Pagina NL-16

Wijzigen van een numerieke waarde


a) b) Navigeer naar de gewenste parameter. Druk opnieuw op de pijl naar RECHTS om de Edit modus te openen. De huidige waarde wordt gemarkeerd. Voer een nieuwe waarde in. Druk op de pijl naar RECHTS om deze te accepteren Het display keert terug naar de weergave van de parameters en toont de nieuwe selectie.

parameternaam

parameternummer

c) d)

huidige waarde

toetsfuncties in Edit modus Toets Naam


Pijl omhoog of omlaag

Functie en Edit modus


Selecteren van opties Bladert naar item. Numerieke bewerking Selecteren van opties - Verhoogt of verlaagt posities - Schakelt tussen plus- en minteken - Accepteert de data (schrijft de parameter). - Wijzigt van Edit naar Navigatie modus

Pijl naar RECHTS

- Verplaatst de cursor een spatie naar rechts Numerieke bewerking - of met de cursor op het Enter teken, accepteert de data en schakelt van de Edit naar de Navigatie modus Selecteren van opties Verlaat de Edit modus zonder de parameter te wijzigen.

Pijl naar LINKS

Numerieke bewerking

- Verplaatst de cursor naar plus/min teken wanneer dit de eerste toets is die wordt ingedrukt - of verplaatst de cursor n spatie naar links.

Clear Decimale punt Plus- of minteken


in

Numerieke bewerking Wist het display.

Nederlands

Numerieke bewerking Voert een decimale punt in.

Numerieke bewerking Verandert het teken van de ingevoerde waarde.

Numeral

Numerieke bewerking Voert het corresponderende karakter in.

A5E31997170

SITRANS LR250 (HART) QUICK START Handleiding

Pagina NL-17

Quick Start Wizard via de handprogrammer


1. Quick Start
a) Richt de handprogrammer op het display vanaf een maximale afstand van 30 cm (1 ft) en druk dan op de pijl naar RECHTS activeren en menuniveau 1 te openen. b) Druk tweemaal op de pijl naar RECHTS parameterweergave te openen. c) Druk op de pijl naar RECHTS volgende item. d) Blader naar het gewenste item of voer een nieuwe waarde in om een instelling te wijzigen. e) Druk, na het wijzigen van een waarde, op de pijl naar RECHTS accepteren en druk op de pijl OMLAAG item. f) Quick Start instellingen worden pas actief nadat u Yes kiest om de wijzigingen toe te passen in in stap (1.8.). om deze te om de Edit modus te openen of op de pijl OMLAAG om naar menu-item 1.1 te gaan en de om de PROGRAMMEER modus te

om de standaard waarden te accepteren en direct verder te gaan naar het

om verder te gaan naar het volgende

1.1. LANGUAGE Selecteert de taal die gebruikt moet worden op het LCD en wordt direct doorgevoerd.
Opties
ENGLISH, DEUTSCH, FRANAIS, ESPAOL

1.2. Product Opties Nederlands


LIQUID LIQUID LOW DK (dilektrische vloeistof laag)

1.3. Responsiesnelheid Stelt de reactiesnelheid in van het apparaat op wijzigingen in de meetwaarden in het doelbereik.
Responsiesnelheid
Slow Med Fast

Vulsnelheid/leegsnelheid
0,1 m / min (0.32 ft / min) 1,0 m / min (3.28 ft / min) 10,0 m / min (32.8 ft / min)

Gebruik een instelling die net iets sneller is dan de maximale vul- of leegsnelheid (welke maar groter is).

1.4. Eenheden Sensor meeteenheden. Opties


M, CM, MM, FT, IN

Pagina NL-18

SITRANS LR250 (HART) QUICK START Handleiding

A5E31997170

1.5. Bedrijfsmodus
sensor referentiepunt ruimte sensor referentiepunt high cal. point

afstand

sensor referentiepunt

niveau low cal.point

Bediening Omschrijving
NO SERVICE Niveau Ruimte Afstand
1)

Meting en de bijbehorende stroomkring worden niet bijgewerkt en het apparaat schakelt standaard in de Fail-safe modus1). Afstand tot het materiaaloppervlak ten opzichte van het "Laag" kalibratiepunt. Afstand tot het materiaaloppervlak ten opzichte van het "Hoog" kalibratiepunt. Afstand tot materiaaloppervlak ten opzichte van het sensorreferentiepunt.

Zie voor meer details over de Fail-safe modus de uitgebreide handleiding.

1.6. Low Calibration Point Afstand van de sensorreferentie tot Low Calibration Point: doorgaans het proces leeg niveau. Zie Bedrijfsmodus (1.5.) voor een illustratie. Waarden
Bereik: 0,00 tot 20,00 m

1.7. High Calibration Point Afstand van de sensorreferentie tot High Calibration Point: doorgaans proces vol niveau. Zie Bedrijfsmodus (1.5.) voor een illustratie. Waarden
Bereik: 0,00 tot 20,00 m

1.8. Apply? (Pas wijzigingen toe) Om de Quick Start instellingen te bewaren, is het noodzakelijk om Yes te kiezen om de wijzigingen toe te passen. Opties
Druk op Mode voor bedrijf. YES, NO, DONE (Display toont DONE wanneer Quick Start succesvol is afgerond.) om terug te keren naar de meetmodus. De SITRANS LR250 is nu gereed

Nederlands

Opmerking: wanneer uw toepassing een tank met obstructies bevat, raadpleeg dan de uitgebreide handleiding voor het gebruik van automatische onderdrukking van stoorecho's.

A5E31997170

SITRANS LR250 (HART) QUICK START Handleiding

Pagina NL-19

SITRANS LR250 communicatie: HART

U heeft de uitgebreide handleiding nodig voor de lijst van parameters die van toepassing zijn. Wij raden het gebruik aan van de Simatic Process Device Manager (PDM) voor het programmeren van uw apparaat. Applicatierichtlijnen voor het configureren van HART-apparaten met SIMATIC PDM kunnen worden gedownload van de productpagina van onze website: www.siemens.com/LR250 onder Support.

SIMATIC PDM
SIMATIC PDM is een softwarepakket dat word gebruikt voor het inbedrijfnemen en onderhouden van de SITRANS LR250 en andere procesapparaten. Raadpleeg de bedieningsinstructies of online help voor details omtrent het gebruik van SIMATIC PDM. (U kunt meer info vinden op www.siemens.com/simatic-pdm.

Deactiveer buffers

Hiermee wordt de SIMATIC PDM op n lijn gebracht met het HART modem voor Windows 2000 en Windows XP 1) besturingssystemen. Opmerkingen: SIMATIC PDM werkt uitsluitend onder de Windows XP Professional versie, niet onder de Home versie. U heeft admininstrator bevoegdheden nodig op uw besturingssysteem om buffers te deactiveren. 1) 2) 3) 4) 5) 6) Klik op Start/Settings/Control Panel om te beginnen met de configuratie. Dubbelklik op System, selecteer het Hardware tabblad, en klik op de Device Manager knop. Open de Ports map en dubbelklik op de COM Port die wordt gebruikt door het systeem om het Communications Port Properties venster te openen. Selecteer het Port Settings tabblad en dubbelklik op de Advanced knop. Wanneer de Use FIFO buffers radioknop is geselecteerd, klik hier dan op om deze te deselecteren. Klik op OK om af te sluiten. Sluit alle schermen en reboot vervolgens.

Updaten van de Electronic Device Description (EDD)


Opmerkingen: Nederlands
SIMATIC PDM Versie 5.2 SP1 wordt alleen ondersteund voor basisconfiguratie en probleemoplossen. Voor geavanceerde functies zoals de Quick Start wizard, is Rev. 6.0 SP4 of hoger vereist. In SIMATIC PDM kunt u de EDD vinden in de Device Catalog, onder Sensors/Level/Echo/ Siemens Milltronics/SITRANS LR250. Zie de productpagina van onze website: www.siemens.com/LR250, om er zeker van te zijn dat u de meest recente versie heeft van SIMATIC PDM, het meest recente Service Pack (SP) en de meest recente hot fix (HF). Ga naar Support > Software Downloads.

Installeren van een nieuwe EDD:


1) 2) 3) download de EDD van de productpagina van onze website: www.siemens.com/LR250 en sla de bestanden op uw computer op. Pak het zip-bestand uit in een gemakkeijk toegankelijke locatie. Start SIMATIC PDM Manager Device Catalog, blader naar het uitgepakte EDD bestand en selecteer het.
Windows is een geregistreerd handelsmerk van Microsoft.

1)

Pagina NL-20

SITRANS LR250 (HART) QUICK START Handleiding

A5E31997170

Configureren van een nieuw apparaat


Opmerking: door op Cancel te drukken tijdens het uploaden van een apparaat naar SIMATIC PDM worden enkele parameters bijgewerkt. 1) 2) Verifieer of o beschikt over de meeste recente EDD, en update deze indien nodig (zieUpdaten van de Electronic Device Description (EDD) op pagina 20). Start SIMATIC Manager en creer een nieuw project voor LR250. Applicatiehandleidingen voor het instellen van HART apparaten met SIMATIC PDM kunnen worden gedownload vanaf de productpagina van onze website: www.siemens.com/LR250. Open het menu Device Master Reset en klik op OK om een reset naar de fabrieksinstellingen uit te voeren. Upload nadat de reset is uitgevoerd de parameters naar de PC/PG. Configureer het apparaat via de Quick Start wizard.

3) 4) 5)

Quick Start Wizard via SIMATIC PDM


de Quick Start wizard instellingen zijn onderling gerelateerd en het kan zijn dat wijzigingen pas worden doorgevoerd nadat u op FINISH AND DOWNLOAD heeft geklikt aan het einde van stap 5 voor het offline opslaan van de instellingen en het overdragen hiervan naar het apparaat. Klik op BACK om terug te gaan en een instelling te wijzigen of kies Cancel om de Quick Start te verlaten. Start SIMATIC PDM, open het menu Device Wizard - Quick Start en volg de stappen 1 t/m 5.

Opmerkingen:

Quick Start

Bediening via FDT (Field Device Tool)


FDT is een standaard die wordt gebruikt in verschillende softwarepakketten die zijn bedoeld voor inbedrijfstelling en onderhoud van veldapparaten. Twee op de markt verkrijgbare FDT's zijn PACTware en Fieldcare. Voor de configuratie van een veldapparaat via FDT heeft u de DTM (Device Type Manager) voor uw apparaat nodig. Siemens instrumenten maken gebruik van SITRANS DTM en een instrument-EDD geschreven voor SITRANS DTM. 1) Installeer eerst SITRANS DTM op uw systeem. U kunt dit downloaden via: http://support.automation.siemens.com. Klik op Product Support en ga naar Product Information/Automation Technology/Sensor systems/Process Instrumentation/Software & Communications. 2) Installeer de SITRANS LR250 HART EDD voor SITRANS DTM. U kunt deze downloaden vanaf de productpagina van onze website: www.siemens.com/LR250. Ga naar Support > Software Downloads.

Nederlands

Configureren van een nieuw apparaat via FDT


Een Applicatiegids kan worden gedownload vanaf de productpagina van onze website onder Support. A5E31997170 SITRANS LR250 (HART) QUICK START Handleiding Pagina NL-21

Bediening via AMS Device Manager


AMS Device Manager is een softwarepakket dat is ontworpen voor inbedrijfstelling en onderhoud van veldapparaten. Raadpleeg de bedieningsinstructies of online help voor meer informatie over het gebruik van de AMS Device Manager. U kunt meer informatie vinden op: http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD)


SITRANS LR250 vereist de EDD voor AMS Device Manager versie 9.0.

Een nieuw apparaat configureren via de AMS Device Manager


1) Zie de productpagina van onze website: www.siemens.com/LR250 om er zeker van te zijn dat u de meest recente EDD heeft. Ga naar Support > Software Downloads en download deze indien nodig. Sla de bestanden op uw computer op en pak het zip-bestand uit in een gemakkelijk toegankelijke map. Start AMS Device Manager voeg Device Type toe, blader naar het uitgepakte EDD bestand en selecteer dit.

2)

Start de AMS Device Manager. Een Applicatiegids voor het instellen van HART-apparaten met AMS Device Manager kunt u downloaden vanaf de productpagina van onze website onder Support. Onderhoud De SITRANS LR250 vereist geen onderhoud of reiniging onder normale bedrijfsomstandigheden. Wanneer reiniging nodig wordt onder extreme omstandigheden: 1) 2) houd rekening met het materiaal van de antenne en het procesmedium en kies een reinigingsmiddel dat op beide geen negatieve invloed heeft. Neem het instrument uit bedrijf en veeg de antenne schoon met een doek en een oplossing van een geschikt reinigingsmiddel.

Reparatie van de eenheid en uitsluiting van aansprakelijkheid


Raadpleeg voor gedetailleerde informatie de binnenkant van de achterflap.

Vervangen van antenne of elektronica/behuizing


Wanneer de antenne vervangen moet worden vanwege beschadiging of een storing, dan kan deze worden vervangen, zonder dat herkalibratie nodig is. Omschakeling naar een ander antennetype kan worden uitgevoerd door een door Siemens geautoriseerd reparatiecentrum of -personeel. Wanneer de elektronica of behuizing moet worden vervangen vanwege beschadiging of storing, zorg er dan voor dat de juiste antenneversie wordt gebruikt, anders moet er opnieuw een kalibratie worden uitgevoerd door personeel dat is geautoriseerd door Siemens.

Nederlands

Pagina NL-22

SITRANS LR250 (HART) QUICK START Handleiding

A5E31997170

Maximum procestemperatuurgrafiek
Flensadapteruitvoeringen van LR250
WAARSCHUWING: de interne temperatuur mag niet hoger zijn dan 80 C (176 F). Opmerkingen: De grafiek dient uitsluitend als richtlijn en toont niet elke mogelijke procesaansluitingsconfiguratie. (Deze geldt NIET wanneer de SITRANS LR250 direct op een metalen tankoppervlak wordt gemonteerd.) De grafiek houdt geen rekening met opwarming als gevolg van blootstelling aan direct zonlicht. Parameter 3.2.1 Huidige interne temperatuur bewaakt de interne temperatuur. Maximale procestemperaturen versus toegestane omgevingstemp.

Omgevingstemperatuur (oC)

Procestemperatuur

Procestemperatuur (oC)

Lusvermogen
voedingsvereisten
Opmerking: De onderstaande grafieken gelden voor een standalone apparaat, geconfigureerd via de Siemens handprogrammeer eenheid. Curve 1: General Purpose, intrinsiekveilig, niet vonkend / beperkte energie / niet brandveroozakend

Lusweerstand RL

TOEGESTAAN BEDRIJFSGEBIED

Nederlands

Lusspanning VL

Opmerking: Bij gebruik van HART communicatie, is de minimale spanning bij 220 Ohm (RL) 16,3 VDC.

A5E31997170

SITRANS LR250 (HART) QUICK START Handleiding

Pagina NL-23

Curve 2: Vlambestendig, verhoogde veiligheid, explosieveilig Lusweerstand RL

TOEGESTAAN BEDRIJFSGEBIED

Lusspanning VL

Opmerking: Bij gebruik van HART communicatie, is de minimale spanning bij 220 Ohm (RL) 20,94 VDC.

Gedrag bij opstarten


Het apparaat trekt minder dan 3,6 mA tijdens opstarten. Tijd tot eerste meting is minder dan 50 seconden.

Nederlands
Pagina NL-24

SITRANS LR250 (HART) QUICK START Handleiding

A5E31997170

SITRANS LR250 (HART) Manual de Partida Rpida


Este manual descreve as caractersticas e funes essenciais do SITRANS LR250 (HART 1)). Aconselhamos categoricamente que voc adquira a verso detalhada do manual para poder utilizar o mximo potencial de seu dispositivo. O manual completo pode ser baixado na pgina do produto SITRANS LR250, em nosso website: www.siemens.com/LR250. O manual impresso est disponvel com seu representante local Siemens Milltronics. Quaisquer dvidas acerca do contedo deste manual devem ser dirigidas a: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontrio, Canad, K9J 7B1 E-mail: techpubs.smpi@siemens.com

Copyright Siemens AG 2013. Todos os direitos reservados


Aconselhamos os usurios a adquirirem manuais autorizados relacionados ou a verem verses eletrnicas conforme projetadas e autenticadas pela Siemens Milltronics Process Instruments. A Siemens Milltronics Process Instruments no se considera responsvel pelo contedo de reprodues parciais ou totais de verses relacionadas ou eletrnicas.

Excluso de Responsabilidade
Apesar de termos verificado o contedo deste manual quanto conformidade com os instrumentos descritos, este pode ter algumas variaes. Dessa forma, no podemos garantir conformidade integral. O contedo deste manual revisto com regularidade, e as correes so includas em edies posteriores. Todas as sugestes de aprimoramento so bem-vindas. Os dados tcnicos esto sujeitos a alteraes.

MILLTRONICS uma marca comercial registrada da Siemens Milltronics Process Instruments.

O suporte est disponvel 24 horas por dia. Para saber o endereo, nmero de telefone e nmero de fax do Escritrio de Automao local da Siemens, acesse: www.siemens.com/automation/partner: Clique na guia Contact (Contato), selecione Service (Servio), em seguida, clique em Service novamente para localizar o grupo de produtos (+Automation Technology (Tecnologia de Automao) > +Sensor Systems (Sistemas de Sensor) > +Process Instrumentation (Instrumentao de Processo) > +Level Measurement (Medio de Nvel) > +Continuous (Contnuo). Selecione Radar. Selecione o pas seguido por regio/cidade. Selecione Technical Support (Suporte Tcnico) em Service. Para suporte tcnico on-line, acesse: www.siemens.com/automation/support-request Digite o nome do dispositivo (SITRANS LR250) ou nmero do pedido; em seguida, clique em Search (Pesquisar) e selecione o tipo de produto apropriado. Clique em Next. Digite uma palavra-chave que descreva seu assunto. Em seguida, navegue pela documentao relevante ou clique em Next para enviar por e-mail a descrio de seu assunto para a equipe de Suporte Tcnico da Siemens. Centro de Suporte Tcnico da Siemens IA/DT: Telefone
1)

Suporte Tcnico

Portugus

+49 (0)911 895 7222

HART uma marca comercial registrada da HART Communication Foundation.

A5E31997170

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

Pgina PT-1

Diretrizes de Segurana

As mensagens de advertncia devem ser observadas para garantir sua segurana pessoal bem como a de terceiros e para proteger o produto e o equipamento a ele associado. Essas mensagens de advertncia so acompanhadas de um esclarecimento do nvel de cautela a ser observado. O smbolo de ADVERTNCIA est relacionado com um smbolo de cuidado no produto e significa que o no cumprimento das precaues necessrias pode resultar em morte, ferimentos graves e/ou danos materiais considerveis. O smbolo de ADVERTNCIA usado quando no h um smbolo de cuidado no produto significa que o no cumprimento das precaues necessrias pode resultar em morte, ferimentos graves e/ou danos materiais considerveis. Observao: significa informao importante acerca do produto ou dessa parte do manual de operaes.

Conformidade com a FCC


Apenas para instalaes nos EUA: Normas da Federal Communications Commission (FCC)
ADVERTNCIA: As alteraes ou modificaes no aprovadas expressamente pela Siemens Milltronics podem anular a autoridade do usurio de utilizar o equipamento. Observaes: Este equipamento foi testado e considerado em conformidade com os limites para um dispositivo digital da Classe A, de acordo com a Parte 15 das Regras da FCC. Estes limites foram concebidos para proporcionar uma proteo razovel contra interferncia nociva quando o equipamento utilizado em um ambiente comercial. Este equipamento gera, utiliza e pode irradiar energia de radiofrequncia e, se no for instalado e utilizado de acordo com o manual de instrues, pode causar interferncias nocivas nas comunicaes via rdio. provvel que o funcionamento deste equipamento em uma rea residencial cause interferncias nocivas nas comunicaes via rdio; nesse caso, o usurio precisar corrigir a interferncia sua prpria custa.

SITRANS LR250
ADVERTNCIA: O SITRANS LR250 deve ser usado apenas da forma descrita neste manual, caso contrrio, a proteo fornecida pelo equipamento pode ser prejudicada. Observao: Este produto foi projetado para uso em reas industriais. A operao deste equipamento em uma rea residencial pode causar interferncia em vrias comunicaes com base em frequncia. O SITRANS LR250 um transmissor de nvel radar de pulsos de 2 fios e 25 GHz para monitoramento contnuo de lquidos e fluidos em tanques de armazenamento, incluindo alta presso e alta temperatura, at um intervalo de 20 m (66 ps). ideal para tanques pequenos e meios dieltricos baixos. O dispositivo consiste em um circuito eletrnico acoplado a uma antena em forma de corneta e/ ou uma conexo de processo do tipo rosqueada ou flangeada. Tambm est disponvel uma antena PVDF rosqueada. O SITRANS LR250 oferece suporte ao protocolo de comunicao HART e ao software SIMATIC PDM. Os sinais so processados utilizando-se Process Intelligence.

Portugus

Pgina PT-2

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

A5E31997170

Especificaes

Para obter uma listagem completa, consulte o Manual de Instrues do SITRANS LR250 (HART). Para informaes sobre as Aprovaes, ver Aprovaes, na pgina 4.

Temperaturas Ambiente/Operacional
Observaes:
A temperatura mxima depende da conexo do processo, materiais da antena e presso do tanque: consulte Grfico de Temperatura Mxima do Processo, na pgina 23. Para mais informaes detalhadas, consulte as curvas de classificao de Presso/ Temperatura do processo no manual completo. A temperatura de processo e as capacidades de presso dependem das informaes na etiqueta do dispositivo de processo. O desenho de referncia listado na etiqueta est disponvel na pgina do produto em nosso site, www.siemens.com/LR250, em Support (Suporte) > Installation Drawings (Desenhos da Instalao) > Level Measurement (Medio de Nvel) > Continuous (Contnuo) - Radar. Informaes adicionais nas conexes do processo esto disponveis na pgina Desenhos de Instalao em Process Connection Diagrams (Diagramas de conexo do processo). A amplitude do sinal aumenta com o dimetro da antena, portanto, use o maior tamanho prtico. Podem ser instaladas extenses adicionais sob as roscas. Consulte Grfico de Temperatura Mxima do Processo, na pgina 23, para obter mais detalhes.
temperatura ambiente (ao redor do invlucro) 40 C a 80 C (40 F a 176 F) temperatura de processo durante a conexo de processo: antena em forma de corneta padro: - com junta circular FKM: 40 oC a +200 oC (-40 oF a +392 oF) - com junta circular FFKM: 20 oC a +200 oC (-4 oF a +392 oF) antena PVDF rosqueada 2" NPT / BSPT / G: 40 a +80 C (40 a +176 F) placa de identificao do dispositivo

etiqueta de conexo do processo

Alimentao
Finalidades Gerais Segurana Intrnseca Sem formao de fascas/Energia limitada No Acendvel (apenas para EUA/Canad FM/CSA) prova de exploso Segurana aumentada prova de exploso (apenas para EUA/Canad FM/CSA) Mximo 30 V CC 4 a 20 mA Corrente inicial mx.: 24 V CC Nominal a 550 Ohm 24 V CC Nominal a 250 Ohm

consultar Comportamento de inicializao, na pgina 24

Portugus

A5E31997170

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

Pgina PT-3

Aprovaes
Observao: A placa de identificao do dispositivo relaciona as aprovaes que se aplicam ao seu dispositivo. Geral CSAUS/C, FM, CE, C-TICK, KC Rdio Europa (R&TTE), FCC, Industry Canada Atmosferas Explosivas Segurana Intrnseca 1) (Europa) ATEX II 1G, EEx ia IIC T4 ATEX II 1D, EEx tD A20 IP67 T90 C (Internacional) IECEx SIR 05.0031X, Ex ia IIC T4 Ex tD A20 IP67 T90 C (EUA/Canad) FM/CSA Classe I, Div. 1, Grupos A, B, C, D Classe II, Div. 1, Grupos E, F, G Classe III T4 (Brasil) INMETRO: DNV 12.0087 X Ex ia IIC T4 Ga Ex ia ta IIIC T90 C Da IP65/IP67 -40 C Ta +80 C DNV #OCP 0017

ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

(China)

Sem formao de fascas/Energia limitada 2) (Europa) ATEX II 3 G Ex nA/nL IIC T4 Gc (China) NEPSI Ex nA II T4/ Ex nL IIC T4 No Acendvel 3) (EUA/Canad)

NEPSI Ex ia IIC T4/ DIP A20 TA, T90 C IP67

prova de exploso 4) ( Europa/Internacional) ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex dmbia IIC T4 Ga/Gb Ex tD A20 IP67 T90 oC (Brasil) INMETRO: DNV 12.0088 X Ex d ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017

FM/CSA Classe I, Div. 2, Grupos A, B, C, D T5

ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-1:2009, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

Portugus

1) 2) 3) 4)

Consulte Instalao eltrica com Segurana Intrnseca, na pgina 9. Consulte Ligao Eltrica sem formao de fascas/com energia limitada, na pgina 11. Consulte Ligao Eltrica no-acendvel, na pgina 11. Consult Instalao eltrica prova de exploso, na pgina 12.

Pgina PT-4

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

A5E31997170

Atmosferas Explosivas (continuao) Segurana aumentada 1) (Europa/Internacional) ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex embia IIC T4 Ga/Gb Ex tD A20 IP67 T90 oC (Brasil) INMETRO: DNV 12.0088 X Ex e ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-7:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

prova de exploso/Segurana aumentada (China) NEPSI Ex dmbia IIC T4/ Ex embia IIC T4/ DIP A20 TA, T90 oC IP67 2) prova de exploso (EUA/Canad) FM/CSA Classe I, Div. 1, Grupos A, B, C, D Classe II, Div. 1, Grupos E, F, G Classe III T4 Naval Registro de Embarque do Lloyds Aprovao Tipo ABS

Aplicao de Presso
ADVERTNCIAS:

Jamais tente soltar, retirar ou desmontar a conexo do processo ou o invlucro de instrumentos enquanto o contedo do tanque estiver sob presso. O usurio responsvel pela seleo dos materiais de aparafusamento e juntas que estejam dentro dos limites do flange e da sua utilizao pretendida e que sejam adequados s condies de servio. Uma instalao incorreta pode resultar na perda de presso do processo.
Observaes: A etiqueta de conexo do processo deve permanecer junto ao conjunto do limite de presso do processo1). Caso a embalagem do dispositivo seja substituda, a etiqueta do dispositivo de processo dever ser transferida para a unidade de substituio. As unidades SITRANS LR250 so hidrostaticamente testadas, atendendo ou excedendo as exigncias do Cdigo de Caldeira e Vaso de Presso da ASME e da Diretiva Europeia de Equipamentos de Presso.

Portugus

1) 2)

Consulte Instalao eltrica de segurana aumentada, na pgina 12. Consulte Instalao eltrica prova de exploso (apenas EUA/Canad), na pgina 13.

A5E31997170

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

Pgina PT-5

Diretiva de Equipamentos sob Presso, PED, 97/23/EC 1)


Os transmissores de nvel da Siemens com montagens de processo do tipo flangeado, roscado ou sanitrio, no possuem um local de presso de suporte prprio e, por conseguinte, no so abrangidos pela Diretiva de Equipamentos sob Presso como acessrios de presso ou segurana (consulte a Diretiva da Comisso Europeia 1/8).

Instalao
ADVERTNCIAS A instalao s dever ser efetuada por pessoal qualificado e em conformidade com os regulamentos locais vigentes. Manuseie o dispositivo utilizando a caixa e no a etiqueta de conexo do processo para evitar danos. Tenha cuidado principalmente quando manusear a antena PVDF rosqueada. Quaisquer danos superfcie da antena, principalmente ponta, podem afetar o desempenho. Os materiais de construo so escolhidos com base em suas capacidades qumicas (ou inatividade) para finalidades gerais. Para exposio a ambientes especficos, verificar as tabelas de compatibilidade qumica, antes da instalao. Observaes:

Na Unio Europeia e pases membros, a instalao dever estar de acordo com a norma ETSI EN 302372. Consulte a placa de identificao do dispositivo para obter informaes sobre aprovao. Os nmeros de srie estampados na estrutura de cada conexo de processo fornecem um nmero de identificao nico que indica a data de fabricao. Exemplo: MMDDYY XXX (onde MM = ms, DD = dia, YY = ano, e XXX= unidade sequencial produzida) Outras marcaes (se o espao permitir) indicam a configurao, tamanho, classe de presso, material e cdigo trmico do material da flange.

Design do bocal
Antena PVDF rosqueada
Folga mn.: 10 mm (0,4)

Antena em forma de corneta de ao inoxidvel


Folga mn.: 10 mm (0,4)

Dim. mn. 51 mm (2")

A extremidade da antena deve ser projetada com, no mnimo, 10 mm (0,4) a fim de evitar que falsos ecos ressoem a partir do bocal. O dimetro mnimo recomendado para o bocal da antena PVDF rosqueada de 51 mm (2"). Uma extenso da antena (100 mm/ 3,93") est disponvel para qualquer verso, exceto para a antena PVDF rosqueada.

Portugus

1)

A unidade do limite de presso do processo engloba os componentes que atuam como uma barreira contra a perda de presso a partir do vaso do processo: ou seja, a combinao do corpo de conexo do processo e o emissor, mas normalmente se exclui a caixa eltrica.

Pgina PT-6

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

A5E31997170

Localizao do bocal
Evite localizaes centrais em tanques altos e estreitos, as quais podem gerar falsos ecos.

Meio Ambiente
Fornecer um ambiente adequado classificao do invlucro e dos materiais de construo. Fornecer proteo solar se o dispositivo for montado em luz solar direta.
prefervel no desejvel

ngulo do feixe
O ngulo do feixe a largura do cone, onde a densidade de energia metade da densidade de energia de pico. A densidade de energia de pico direcionada diretamente em frente da antena e alinhada com ela. H um sinal transmitido fora do ngulo de feixe; portanto, alvos falsos podem ser detectados.

cone de emisso
ngulo do feixe: antena PVDF rosqueada = 19 haste de 1,5" = 19 haste de 2" = 15 ngulo haste de 3" = 10 do haste de 4" = 8 feixe

Cone de emisso
Mantenha o cone de emisso sem interferncias de escadas, tubos, vigas ou trajetos de enchimento.

Acesso para programao


Fornecer acesso fcil para visualizar o display e programar por meio do programador manual.

Instrues de montagem
ADVERTNCIA: Para aplicaes de presso, ser necessrio utilizar fita de PTFE ou outro composto vedante de roscas apropriado e apertar a conexo de processo com mais presso do que o aperto manual. O torque mximo recomendado de 40 N-m (30 ps-lbs.). Observao: No h limite ao nmero de vezes que o dispositivo pode ser girado sem danos.

Verses rosqueadas
1) 2) Antes de inserir o dispositivo em sua conexo de montagem, verifique se as roscas so compatveis para evitar danific-las. Simplesmente parafuse o dispositivo na conexo de processo e aperte manualmente ou use uma chave. Para aplicaes de presso, consulte a Advertncia acima. ADVERTNCIA: O usurio responsvel pela seleo dos materiais de parafusamento e juntas que estejam dentro dos limites do flange e da sua utilizao pretendida e que sejam adequados s condies de servio.

Verses flangeadas Portugus

A5E31997170

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

Pgina PT-7

Ligao Eltrica
Alimentao
ADVERTNCIAS Os terminais de entrada CC devem ser alimentados por uma fonte que proporcione isolamento eltrico entre a entrada e a sada para que sejam respeitadas as exigncias de segurana da norma IEC 61010-1 Toda a ligao eltrica de campo deve ter um isolamento adequado s tenses especificadas.blindagem do

Conexo do SITRANS LR250


ADVERTNCIAS: Verifique a identificao do dispositivo e o rtulo de identificao do processo para verificar a classificao de aprovao. Utilize vedaes de condute adequadas para manter a classificao de IP ou NEMA. Consulte Configuraes da conexo eltrica para instalaes em atmosferas explosivas, na pgina 9. Observaes: Utilize cabo de par tranado: de AWG 22 a 14 (de 0,34 mm2 a 2,5 mm2). Podem ser necessrios cabos e condutes separados de acordo com as prticas de instalao eltrica de instrumentao padro ou cdigos de instalao eltrica.
Utilize uma chave Allen de 2 mm para desapertar o parafuso de fechamento da tampa. plugue (IP 68) prensa-cabo opcional1) 2) (ou entrada do cabo NPT2)) blindagem do cabo (se for utilizado)

1) 2) 3) 4) 5)

Retire a cobertura do cabo aproximadamente 70 mm (2,75") a partir do seu extremo e passe os fios pelo prensa-cabo. 1) 2) Ligue os fios aos terminais, conforme exibido: a polaridade est indicada no bloco de terminais. Ligue o instrumento ao aterramento de acordo com os regulamentos locais. Aperte o prensa-cabo para fornecer vedao adequada. Aperte a tampa antes da configurao do dispositivo e programao.

Portugus

1) 2)

Pode ser entregue com o dispositivo. Se o cabo passar atravs do condute, utilizar apenas encaixes de tamanho apropriado prprios para aplicaes prova de tempo.

Pgina PT-8

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

A5E31997170

Conexo do HART
Configurao tpica do PLC/mA com HART 1) 2)
PLC ativo fonte de alimentao R= 250 modem do HART SITRANS LR250 comunicador do HART

Observaes: Dependendo do design do sistema, a fonte de alimentao pode ser separada do PLC ou estar ligada a ela. A resistncia HART1) deve ser limitada de acordo com a rea de funcionamento permissvel2). No conectar mais do que um dispositivo de programao HART ao mesmo tempo.

Configuraes da conexo eltrica para instalaes em atmosferas explosivas


H seis opes de instalao eltrica em atmosferas explosivas. Em todos os casos, consulte a placa de identificao do dispositivo e o rtulo do dispositivo do processo para verificar a classificao nominal.

1. Instalao eltrica com Segurana Intrnseca


Placa de identificao do dispositivo (ATEX/IECEX/INMETRO/C-TICK)

O certificado ATEX, listado na placa de identificao, pode ser baixado da pgina do produto em nosso website: www.siemens.com/LR250. Acesse Support (Suporte) > Approvals (Aprovaes)/ Certificates (Certificados). O certificado IECEx, listado na placa de identificao, pode ser consultado no website da IECEx. Acesse: http://iecex.iec.ch, clique em Ex Equipment Certificates of Conformity e, a seguir, insira o nmero de certificado IECEx SIR 05.0031X.

Portugus

1) 2)

Resistncia total do circuito = resistncia do cabo mais 250 Ohm (resistncia). Consulte Curva 1: Propsito Geral, Segurana Intrnseca, Sem Formao de Fascas/Energia Limitada, No-acendvel, na pgina 23 ou Curva 2: prova de fogo, Segurana aumentada, prova de Exploso, na pgina 24.

A5E31997170

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

Pgina PT-9

Instalao eltrica com segurana intrnseca (continua)


Placa de identificao do dispositivo (FM/CSA)

O nmero de conexo de Segurana Intrnseca nmero 23650653 pode ser baixado da pgina do produto em nosso website: www.siemens.com/LR250. Acesse Support (Suporte) > Installation Drawings (Desenhos da Instalao) > Level Measurement (Medio de Nvel) > Continuous (Contnuo) - Radar. Para exigncias de alimentao, consulte Curva 1: Propsito Geral, Segurana Intrnseca, Sem Formao de Fascas/Energia Limitada, No-acendvel, na pgina 23. Para requisitos de instalao eltrica: seguir os regulamentos locais. As vedaes de condute impermevel a p e gua so obrigatrias para locais externos de NEMA 4X / tipo 4X / NEMA 6, IP67, IP68. Consulte Instrues especficas para instalaes em atmosferas explosivas, na pgina 13. As barreiras de segurana intrnseca recomendadas encontram-se listadas em Barreiras Passivas de Diodo em Shunt, na pgina 10 e Barreiras ativas (barreiras repetidoras), na pgina 11. Observao: A seleo de um mdulo de entrada do PLC, fonte de alimentao ou barreira adequada requer conhecimentos sobre Segurana Intrnseca e sobre a aplicao. de responsabilidade do instalador assegurar que a instalao intrinsecamente segura esteja em conformidade, tanto com os requisitos de aprovao do aparelho como com o cdigo correspondente de prtica nacional.

Barreiras Passivas de Diodo em Shunt


Observao: necessria uma fonte de alimentao bem regulada. Fabricao
MTL MTL Stahl Stahl

Cdigo de Pea
787SP+ (Canal duplo) 7787P+ (Canal duplo) 9001/01-280-100-10 (Canal nico) 9002/01-280-110-10 (Canal duplo)

Como selecionar uma barreira passiva para o SITRANS LR 250

Portugus

Para garantir que a descrio de segurana da barreira seja adequada aos parmetros de entrada de Segurana Intrnseca (IS) do LR250, realize os clculos a seguir: Re-e = mx. resistncia de ponta a ponta da barreira Rloop = Resistncia do circuito (total da resistncia do cabo mais quaisquer sries de resistncia adicionais, por exemplo entradas e/ou displays do PLC) Vbarreira = valor de qualquer quedas de tenso no linear devido barreira 1) Determinar o valor de Re-e da planilha de dados. 2) Calcular o valor total de Rloop. 3) Calcular a R trabalho = Re-e + Rloop.

Pgina PT-10

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

A5E31997170

Como selecionar uma barreira passiva (continuao) 4) Determinar o valor da Vbarreira na planilha de dados da barreira (por exemplo, a voltagem diminui devido aos diodos). 5) Calcular a V trabalho = Valimentao Vbarreira. 6) Utilizar os valores para a V trabalho e R trabalho para confirmar que a operao est dentro da rea escura do grfico Curva 1: Propsito Geral, Segurana Intrnseca, Sem Formao de Fascas/Energia Limitada, No-acendvel, na pgina 23.

Barreiras ativas (barreiras repetidoras)


Fabricante
MTL MTL Siemens Stahl MTL MTL

Cdigo de Pea
706 7206 SITRANS I 7NG4124 9001/51-280-110-14 E02009 E02010

2. Ligao Eltrica sem formao de fascas/com energia limitada

O certificado ATEX, listado na placa de identificao, pode ser baixado da pgina do produto em nosso website: www.siemens.com/LR250. Acesse Support (Suporte) > Approvals (Aprovaes)/ Certificates (Certificados). Para requisitos de energia, consulte Curva 1: Propsito Geral, Segurana Intrnseca, Sem Formao de Fascas/Energia Limitada, No-acendvel, na pgina 23. Para as exigncias de instalao eltrica, observe os regulamentos locais.

3. Ligao Eltrica no-acendvel

O nmero de conexo da FM/CSA Classe 1, Div 2: 23650673 pode ser baixado da pgina do produto em nosso website: www.siemens.com/LR250. Acesse Support (Suporte) > Installation Drawings (Desenhos da Instalao) > Level Measurement (Medio de Nvel) > Continuous (Contnuo) - Radar. Para exigncias de alimentao, consulte Curva 1: Propsito Geral, Segurana Intrnseca, Sem Formao de Fascas/Energia Limitada, No-acendvel, na pgina 23.

Portugus

A5E31997170

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

Pgina PT-11

4. Instalao eltrica prova de exploso

O certificado ATEX, listado na placa de identificao, pode ser baixado da pgina do produto em nosso website: www.siemens.com/LR250. Acesse: Support (Suporte) > Approvals (Aprovaes) / Certificates (Certificados). O certificado IECEx, listado na placa de identificao, pode ser consultado no website da IECEx. Acesse: http://iecex.iec.ch, clique em Ex Equipment Certificates of Conformity e, a seguir, insira o nmero de certificado IECEx SIR 08.0107X. Para exigncias de alimentao, consulte Curva 2: prova de fogo, Segurana aumentada, prova de Exploso, na pgina 24. Para as exigncias de instalao eltrica, observe os regulamentos locais. Veja tambm Instrues especficas para instalaes em atmosferas explosivas, na pgina 13 e o certificado ATEX relacionado acima.

5. Instalao eltrica de segurana aumentada

O certificado ATEX, listado na placa de identificao, pode ser baixado da pgina do produto em nosso website: www.siemens.com/LR250. Acesse: Support (Suporte) > Approvals (Aprovaes) / Certificates (Certificados). O certificado IECEx, listado na placa de identificao, pode ser consultado no website da IECEx. Acesse: http://iecex.iec.ch, clique em Ex Equipment Certificates of Conformity e, a seguir, insira o nmero de certificado IECEx SIR 08.0107X. Para exigncias de alimentao, consulte Curva 2: prova de fogo, Segurana aumentada, prova de Exploso, na pgina 24. Para as exigncias de instalao eltrica, observe os regulamentos locais. Veja tambm Instrues especficas para instalaes em atmosferas explosivas, na pgina 13 e o certificado ATEX relacionado acima.

Portugus
Pgina PT-12

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

A5E31997170

6. Instalao eltrica prova de exploso (apenas EUA/Canad)

O nmero de conexo Prova de Exploso da FM/CSA A5E02257843 pode ser baixado da pgina do produto em nosso website: www.siemens.com/LR250. Acesse Support (Suporte) > Installation Drawings (Desenhos da Instalao) > Level Measurement (Medio de Nvel) > Continuous (Contnuo) - Radar. Para exigncias de alimentao, consulte Curva 2: prova de fogo, Segurana aumentada, prova de Exploso, na pgina 24.

Instrues especficas para instalaes em atmosferas explosivas


(Referncia Europeia ATEX Diretiva 94/9/CE, Anexo II, 1/0/6)
As seguintes instrues aplicam-se aos equipamentos abrangidos pelo nmero de certificado SIRA 06ATEX2358X, SIRA 08ATEX1301X e SIRA 09ATEX4153X. 1) 2) Para utilizao e montagem, consulte as instrues principais. O equipamento certificado para uso como equipamento de categoria 1GD segundo SIRA 06ATEX2358X; equipamento de categoria 1/2 GD, 1D, 2D segundo SIRA 08ATEX1301X e equipamento de categoria 3G segundo SIRA 09ATEX4153X. 3) O equipamento pode ser utilizado com gases e vapores inflamveis com os grupos de aparelhos IIC, IIB e IIA, e classes de temperatura T1, T2, T3 e T4. 4) O equipamento tem um grau de proteo de IP67 e classe de temperatura de T90 C e pode ser utilizado com poeiras combustveis. 5) O equipamento est certificado para ser utilizado dentro de um intervalo de temperatura ambiente de -40 C a +80 C. 6) O equipamento no foi avaliado como dispositivo relacionado com segurana (conforme mencionado na Diretiva 94/9/CE, Anexo II, clusula 1.5). Consulte Diretiva de Equipamentos sob Presso, PED, 97/23/EC, na pgina 6. 7) A instalao e a inspeo deste equipamento sero efetuadas por pessoal com a formao adequada, de acordo com o cdigo de prtica aplicvel (EN 60079-14 e EN 60079-17 na Europa). 8) O equipamento no pode ser reparado. 9) Os nmeros do certificado tm um sufixo X, que indica que se aplicam condies para o uso seguro. Os que forem instalar ou inspecionar este equipamento devem ter acesso aos certificados. 10) Em caso de o equipamento entrar em contato com substncias agressivas, da responsabilidade do usurio tomar as precaues adequadas para que o equipamento no seja adversamente afetado, garantindo, portanto, que a proteo adotada no seja comprometida. Substncias agressivas: incluem, por exemplo, lquidos ou gases cidos que possam atacar os metais ou solventes e afetar os materiais polimricos. Precaues adequadas: por exemplo, estabelecer que a planilha de dados do material seja resistente a produtos qumicos especficos. A5E31997170 SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA Pgina PT-13

Portugus

Programao do SITRANS LR250


Inicializao do SITRANS LR250

Consulte Assistente de Partida Rpida pelo programador porttil, na pgina 18. Consultar Assistente de Partida Rpida por meio de SIMATIC PDM, na pgina 21.

Ligue o instrumento. O SITRANS LR250 iniciado automaticamente no modo Medio.

O Display de LCD Display no modo Medio (funcionamento normal) 1) 2)


(

1 indicador de alternncia1) para unidades lineares ou %


1 2 3 4

2 operao selecionada: level (nvel), space (espao), distance (distncia) ou volume (volume) 3 valor medido (nvel ou volume, espao ou distncia) 4 unidades 5 o grfico de barras indica o nvel 6 a regio secundria indica a temperatura da eletrnica2) mediante solicitao, confiabilidade de eco, corrente de circuito ou distncia 7 a rea de texto exibe mensagens de status 8 indicador de status do dispositivo (para obter mais detalhes, consulte o manual)

Indicadores de falha atual S: 0 LOE


Em caso de falha atual, o cdigo de falha e uma mensagem de erro so exibidos na rea de texto (7) e um cone de servio necessrio aparece no local de status do dispositivo (8).

Display do modo programa Visualizao de Navegao


menu de item Uma barra de menu visvel indica atual atual que a lista de menus est longa barra de item atual demais para exibir todos os itens. menus faixa de Uma faixa no meio da barra de item menus indica que o item atual est no meio da lista. A posio de profundidade e relativa da faixa do item na barra de menus indica a extenso da lista de menus e a posio aproximada do item atual na lista. Uma faixa mais profunda indica menos itens. nmero

Visualizao Parmetros
nome do parmetro nmero do parmetro valor/seleo do parmetro

Visualizao Editar

Portugus

1) 2)

Pressione a seta UP (para cima) ou DOWN (para baixo) para alternar. Em resposta a uma solicitao de tecla pressionada. Para obter mais detalhes, consulte a tabela Funo no modo Medio, na pgina 15.

Pgina PT-14

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

A5E31997170

Programador Manual (No. da Pea 7ML1930-1BK)


Solicitado separadamente. 1) 2) 3)

Tecla Funo no modo Medio


Atualiza a corrente do circuito1). Atualiza a leitura da temperatura interna da caixa1). Atualiza o valor de confiabilidade do eco1).
C

Atualiza a medio da distncia1). Modo: abre o modo PROGRAM2). Seta DIREITA: abre o modo PROGRAM3). Seta PARA CIMA ou PARA BAIXO: alterna entre as unidades lineares e a porcentagem.

Programao pelo programador manual


Observaes: Pressione Modo para alternar entre o modo Medio e o modo Program. Enquanto o dispositivo se encontra no modo PROGRAM, a sada permanece fixa e no responde s alteraes no dispositivo. O SITRANS LR250 retorna automaticamente ao modo Medio aps um perodo de inatividade no modo PROGRAM (entre 15 segundos e 10 minutos, dependendo do nvel de menu).

Menus de parmetros

Os parmetros so identificados por nome e 1. QUICK START 2. SETUP organizados em grupos de funo. 2.1. DEVICE Para obter a lista completa de parmetros com instrues, ............. consulte o manual completo. 2.7. LINEARIZATION

2.7.1. VOLUME 2.7.1.1. VESSEL SHAPE

Portugus

1) 2)

3)

O novo valor exibido na regio secundria do LCD. Abre o ltimo nvel de menu exibido neste ciclo de alimentao, a no ser que a alimentao tenha sido cclica desde a sada do modo PROGRAM ou por mais de 10 minutos, ou tenha sido programvel desde a utilizao do modo PROGRAM. Em seguida, o menu de nvel mais elevado ser exibido. Abre o menu de nvel elevado.

A5E31997170

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

Pgina PT-15

1.

Selecione o modo PROGRAM


Direcione o programador ao display (de uma distncia mxima de 300 mm [1 p]). A seta DIREITA o nvel do menu 1. ativa o modo PROGRAM e abre

programador porttil display

O modo abre o ltimo nvel de menu exibido no modo PROGRAM nos ltimos 10 minutos ou o menu nvel 1 se a alimentao houver passado por uma reinicializao desde ento.

Mx. 300 mm (1 p)

2.

Navegao: funes das teclas em modo Navegao

Observao: Para um Acesso Rpido aos parmetros pelo programador manual, pressione Home e, a seguir, insira o nmero de menu, por exemplo: 2.7.1 (Volume).

Tecla

Nome
Seta PARA CIMA ou PARA BAIXO

Nvel de menu
menu ou parmetro menu parmetro

Funo no Modo Navegao


Desliza a barra de rolagem para o menu ou parmetro anterior ou seguinte. Acessa o primeiro parmetro no menu selecionado ou abra o menu seguinte. Abre o modo Editar. Abre o menu original. Altera para o modo MEDIO. Abre o menu de nvel elevado: menu 1.

Seta DIREITA Seta ESQUERDA


Modo Pgina inicial

menu ou parmetro menu ou parmetro menu ou parmetro

3.

Edio no modo PROGRAM Seleo de uma opo relacionada


a) b) Navegue at o parmetro desejado. Pressione a seta DIREITA visualizao de parmetros. para abrir a

nome do parmetro

nmero do parmetro

seleo atual

c)

Portugus

Pressione a seta DIREITA mais uma vez para abrir o modo Edio. A seleo atual destacada. Deslize com a barra de rolagem para uma nova seleo. Pressione a seta DIREITA para confirmar. O display LCD retorna para a visualizao de parmetros e exibe a nova seleo. SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA A5E31997170

d)

Pgina PT-16

Alterao de um valor numrico


a) b) Navegue at o parmetro desejado. Pressione a seta DIREITA uma vez mais para abrir o modo Editar. O valor atual destacado. Digite um novo valor. Pressione a seta DIREITA para confirmar. O display LCD retorna para a visualizao de parmetros e exibe a nova seleo.

nome do parmetro

nmero do parmetro

c) d)

valor atual

Funes das teclas no modo Editar Tecla Nome Funo no Modo Edio
Desliza com a barra de rolagem at o item. - Acrescenta ou subtrai dgitos - Alterna entre os sinais de mais e de menos - Confirma dados (grava o parmetro). - Altera do modo Editar para o modo Navegao - Move o cursor um espao para a direita - ou com o cursor no sinal Enter, confirma os dados e altera do modo Editar para o modo Navegao. Cancela o modo Editar sem alterar o parmetro - Move o cursor para o sinal de mais/menos se esta for a primeira tecla pressionada - ou move o cursor um espao para a esquerda. Apaga a tela. Adiciona um ponto decimal.

Seta PARA Seleo de opes CIMA ou PARA BAIXO Edio numrica Seleo de opes

Seta DIREITA

Edio numrica Seleo de opes

Seta ESQUERDA

Edio numrica Edio numrica Edio numrica

Limpar Ponto decimal Sinal de mais ou de menos


a

Edio numrica

Muda o sinal do valor inserido.

Numeral

Edio numrica

Insere o caractere correspondente.

o Portugus

A5E31997170

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

Pgina PT-17

Assistente de Partida Rpida pelo programador porttil


1. Partida Rpida
a) Aponte o programador para o display (a uma distncia mxima de 300 mm (1 p); depois pressione a seta DIREITA no nvel 1. b) Pressione a seta DIREITA duas vezes para navegar at o item de menu 1.1 e para ativar o modo PROGRAM e abra o menu

abrir a visualizao de parmetros. c) Pressione a seta DIREITA BAIXO d) para abrir o modo Edio ou a seta PARA

para confirmar os valores padro e ir diretamente para o item seguinte.

Para alterar uma configurao, deslize com a barra de rolagem at o item desejado ou digite um novo valor.

e)

Aps modificar um valor, pressione a seta DIREITA seta PARA BAIXO para ir at o item seguinte.

para aceit-lo e pressione a

f)

As configuraes de Partida Rpida so aplicadas somente depois de selecionar Sim para Aplicar alteraes na etapa (1.8.).

1.1. Idioma Seleciona o idioma a ser utilizado no LCD e funciona de imediato.


Opes
ENGLISH (INGLS), DEUTSCH (ALEMO), FRANAIS (FRANCS), ESPAOL (ESPANHOL)

1.2. Material Opes


LIQUID (LQUIDO) LIQUID LOW DK (lquido dieltrico baixo)

1.3. Tempo de Resposta Define a velocidade de reao do dispositivo s alteraes de medio no intervalo de alcance.
Tempo de Resposta
SLOW (LENTA) MED (MD.) FAST (RPIDA)

Taxa de Enchimento/Taxa de Esvaziamento


0,1 m/min (0,32 ps/min) 1,0 m/min (3,28 ps/min) 10,0 m/min (32,8 ps/min)

Use uma configurao mais rpida do que a taxa mxima de enchimento ou esvaziamento (a que for maior).

1.4. Unidades Unidades de medida de sensor. Portugus Opes


M, CM, MM, FT, IN

Pgina PT-18

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

A5E31997170

1.5. Modo de Funcionamento


ponto de referncia do sensor espao ponto de referncia do sensor

distncia ponto de calibrao alto

ponto de referncia do sensor

nvel ponto de calibrao baixo

Operao

Descrio

NO SERVICE A medida e a corrente do circuito associada no esto atualizadas, e o dispositivo (NO FUNCIONA) passa para o modo Falha de Segurana1). LEVEL (NVEL) SPACE (ESPAO) DISTANCE (DISTNCIA)
1)

Distncia at superfcie do material referenciada ao Ponto de Calibrao Baixo. Distncia at superfcie do material referenciada ao Ponto de Calibrao Alto. Distncia at superfcie do material referenciada ao Ponto de Referncia do Sensor.

Para obter mais detalhes sobre o modo Falha de Segurana, consulte o manual.

1.6. Ponto de Calibrao Baixo Distncia da Referncia do Sensor em relao ao Ponto de Calibrao Baixo: normalmente nvel de processo vazio. Consultar Modo de Funcionamento (1.5.) para uma ilustrao. Valores
Faixa: 0,00 a 20,00 m

1.7. Ponto de Calibrao Alto Distncia da Referncia do Sensor em relao ao Ponto de Calibrao Alto: normalmente nvel de processo cheio. Consultar Modo de Funcionamento (1.5.) para uma ilustrao. Valores
Faixa: 0,00 a 20,00 m

1.8. Apply? (Aplicar?) (Aplicar Alteraes) Para poder salvar as configuraes de Partida Rpida, necessrio selecionar Sim para aplicar as alteraes. Opes
Pressione Modo para funcionar. YES (Sim), NO (No), DONE (Concludo) (O display exibe DONE quando a Partida Rpida concluda com sucesso). para retornar ao modo Medio. Agora o SITRANS LR250 est pronto

Portugus

Observao: Se sua aplicao tiver um tanque com obstrues, consulte o manual completo para detalhes sobre como usar Autossupresso de Eco Falso.

A5E31997170

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

Pgina PT-19

Comunicaes com o SITRANS LR250: HART

Voc precisar do manual completo para obter a lista dos parmetros aplicveis. Recomendamos que utilize o SIMATIC Process Device Manager (PDM) para programar o seu dispositivo. As Guias de Aplicao para a configurao de dispositivos HART com SIMATIC PDM podem ser baixadas da pgina do produto em nosso site: www.siemens.com/LR250 em Support (Suporte).

SIMATIC PDM

O SIMATIC PDM um pacote de softwares utilizado para designar e preservar o SITRANS LR250 e outros dispositivos do processo. Consultar as instrues de funcionamento ou a ajuda on-line quanto a detalhes sobre a utilizao do SIMATIC PDM. (Obtenha mais informaes no seguinte website: www.siemens.com/simatic-pdm.)

Desativar Buffers

Isso alinha o SIMATIC PDM com o modem HART para os sistemas operacionais Windows 2000 e Windows XP 1). Observaes: O SIMATIC PDM funciona apenas com a verso Windows XP Professional e no funciona com a verso Home. Voc precisa de direitos administrativos em seu sistema operacional para desativar buffers. 1) 2) 3) 4) 5) 6) Clique em Start (Iniciar)/Settings (Configuraes)/Control Panel (Painel de controle) para iniciar a configurao. D um clique duplo em System (Sistema), selecione a guia Hardware e clique no boto Device Manager (Gerenciador de dispositivos). Abra a pasta Ports (Portas) e d um clique duplo na Porta COM usada pelo sistema para abrir a janela Communications Port Properties (Propriedades da porta de comunicaes). Selecione a guia Port Settings (Configuraes da porta) e d um duplo clique no boto Advanced (Avanado). Se a caixa de seleo Use FIFO buffers (Usar FIFO buffers) estiver selecionada, clique para desmarcar. Clique em OK para fechar. Feche todas as telas e reinicie.

Atualizar Descrio do Dispositivo Eletrnico (EDD)


Observaes:
O SIMATIC PDM Rev. 5.2 SP1 serve apenas para configuraes bsicas e resoluo de problemas. Para caratersticas mais avanadas como o assistente de Partida Rpida, necessrio Rev. 6.0 SP4 ou mais elevado. Em SIMATIC PDM, o EDD encontra-se localizado no Catlogo do Dispositivo, em Sensors/Level/Echo/Siemens Milltronics/SITRANS LR250. Verifique a pgina do produto em nosso website: www.siemens.com/LR250, para garantir que voc tem a verso mais atual do SIMATIC PDM, o Service Pack (SP) mais atual e as correes (HF) mais atuais. Acesse Support (Suporte) > Software Downloads (Downloads de Software).

Para instalar um novo EDD:


1)

Portugus

2) 3)
1)

Baixe o EDD da pgina do produto em nosso website: www.siemens.com/LR250 e salve os arquivos em seu computador. Extraia o arquivo compactado para um local de fcil acesso. Inicie o SIMATIC PDM Manager Device Catalog (Catlogo do Dispositivo de Administrao), acesse o arquivo EDD descompactado e selecione-o.
Windows marca registrada da Microsoft Corporation

Pgina PT-20

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

A5E31997170

Configurao de um novo dispositivo


Observao: Clicar em Cancelar durante um upload a partir do dispositivo para o SIMATIC PDM resultar na atualizao de alguns parmetros.
1) 2) 3) 4) 5)

Atualizar Descrio do Dispositivo Eletrnico (EDD), na pgina 20).

Verifique se voc tem o EDD mais recente e, se necessrio, atualize-o (consulte

Inicie o Gestor SIMATIC e crie um novo projeto para LR250. As Guias de Aplicao para a configurao de dispositivos HART com SIMATIC PDM podem ser baixadas da pgina do produto em nosso site: www.siemens.com/LR250. Abra o menu Dispositivo Master Reset e clique em OK para reiniciar segundo os Padres de Fbrica. Aps reiniciar, faa a atualizao dos parmetros para o PC/PG. Configure o dispositivo por meio do assistente de Partida Rpida.

Assistente de Partida Rpida por meio de SIMATIC PDM


As configuraes do assistente de Partida Rpida esto interligadas e as alteraes so aplicadas somente depois de clicar em FINISH (CONCLUDO) E DOWNLOAD no fim da etapa 5 para salvar as configuraes e transferi-las para o dispositivo. Clique em RETORNAR para voltar e analisar uma configurao ou Cancelar para sair da Partida Rpida. Inicie o SIMATIC PDM, abra o menu Device (Dispositivo), Wizard - Quick Start (Assistente Partida Rpida) e siga as etapas 1 a 5.

Observaes:

Partida Rpida

Operao via FDT (Field Device Tool (Ferramenta de Dispositivo de Campo))


O FDT um padro utilizado em vrios pacotes de software projetados para vincular e manter dispositivos de campo. Dois FDTs disponveis no mercado so o PACTware e o Fieldcare. Para configurar um dispositivo de campo via FDT, voc precisar do DTM (Device Type Manager (Gerenciador de Tipo de Dispositivo)) para o dispositivo. Instrumentos Siemens de uso SITRANS DTM e um instrumento de escrita para EDD SITRANS DTM. 1) Primeiro instale o SITRANS DTM em seu sistema. Voc pode baix-lo do website: http://support.automation.siemens.com. Clique em Product Support (Suporte ao Produto) e desa at Product Information/Automation Technology/Sensor systems/Process Instrumentation/Software & Communications (Informaes do Produto/Tecnologia de Automao/Sistemas de Sensores/Instrumentao do Processo/Software e Comunicaes) 2) Instale o SITRANS LR250 HART EDD para o SITRANS DTM. Pode ser baixado da pgina do produto em nosso website: www.siemens.com/LR250. Acesse Support (Suporte) > Software Downloads (Downloads de Software).

Portugus

Configurao de um novo dispositivo via FDT

Um Guia de Aplicao pode ser baixado da pgina do produto do nosso website em Support (Suporte). A5E31997170 SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA Pgina PT-21

Operao por meio do AMS Device Manager


O Gerenciador de Dispositivos AMS um pacote de software projetado para vincular e manter dispositivos de campo. Consulte as instrues de funcionamento ou a ajuda on-line quanto a detalhes sobre a utilizao do AMS Device Manager. possvel encontrar mais informaes no site: http://www.emersonprocess.com/AMS/.

Descrio do Dispositivo Eletrnico (EDD)


O SITRANS LR250 exige o EDD para a verso 9.0 do AMS Device Manager.

Configurar um novo dispositivo via AMS Device Manager


1) Verifique a pgina do produto em nosso website: www.siemens.com/LR250 para ter certeza de que est utilizando o EDD mais recente. Acesse Support (Suporte) > Software Downloads (Downloads de Software) e, se necessrio, faa o download. Salve os arquivos em seu computador e extraia os arquivos comprimidos em um local de fcil acesso. Inicie o AMS Device Manager Add Device Type, acesse o arquivo EDD descompactado e selecione-o.

2)

Inicie o AMS Device Manager. Um Guia de Aplicao para configurao dos dispositivos HART com o AMS Device Manager pode ser baixado da pgina do produto de nosso site em Support (Suporte).

Manuteno
O SITRANS LR250 no necessita de manuteno ou limpeza em condies de funcionamento normais. Se for necessrio fazer uma limpeza em condies de funcionamento difceis: 1) 2) Anote o material da antena e do meio de processo e selecione uma soluo de limpeza que no reaja de forma adversa com ambos. Retire o dispositivo de operao e limpe a antena, utilizando um pano e uma soluo de limpeza adequada.

Reparo da unidade e Excluses de Responsabilidade


Para obter informaes detalhadas, consulte a contracapa interna.

Substituio de antena ou eletrnicos/invlucro


Se a antena exigir substituio devido a danos ou falhas, ela poder ser substituda sem necessidade de recalibrao. A mudana para um tipo de antena diferente pode ser feita por uma pessoa ou centro autorizado de reparos da Siemens. Se os eletrnicos ou invlucros precisarem de substituio devido a danos ou falhas, verifique a utilizao da verso correta da antena, caso contrrio ser necessria uma recalibrao feita pela equipe autorizada da Siemens.

Portugus
Pgina PT-22

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

A5E31997170

Grfico de Temperatura Mxima do Processo


Verses do Adaptador de Flange do LR250
ADVERTNCIA: a temperatura interna no deve exceder 80 C (176 F). Observaes: A tabela deve ser utilizada somente como orientao e no representa todas as possveis disposies de conexo do processo. (No se aplica se o SITRANS LR250 estiver montada diretamente em uma superfcie de recipiente metlico.) O grfico no leva em conta o aquecimento resultante da exposio direta luz solar Parmetro 3.2.1 A Temperatura Interna Corrente monitora a temperatura interna.

Temperatura Ambiente (oC)

Temperaturas Mximas do Processo vs. ambiente permissvel

temperatura do processo

Temperatura do Processo (oC)

Alimentao do circuito
Exigncias de Fonte de Alimentao
Observao: As curvas abaixo aplicam-se a um dispositivo independente, configurado atravs do programador porttil da Siemens. Curva 1: Propsito Geral, Segurana Intrnseca, Sem Formao de Fascas/ Energia Limitada, No-acendvel

Resistncia do Circuito RL

REA OPERACIONAL AUTORIZADA

Tenso de Circuito VL

Observao: Ao usar comunicaes HART, a tenso mnima com 220 Ohms (RL) de 16,3 VCC.

Portugus

A5E31997170

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

Pgina PT-23

Resistncia do Circuito RL

Curva 2: prova de fogo, Segurana aumentada, prova de Exploso

REA OPERACIONAL AUTORIZADA

Tenso de Circuito VL

Observao: Ao usar comunicaes HART, a tenso mnima com 220 Ohms (RL) de 20,94 VCC.

Comportamento de inicializao
O dispositivo consome menos de 3,6 mA na inicializao. O tempo para primeira medio de menos de 50 segundos.

Portugus
Pgina PT-24

SITRANS LR250 (HART) MANUAL DE PARTIDA RPIDA

A5E31997170

SITRANS LR250 (HART) Pikakyttopas


Tm opas kertoo SITRANS LR250 (HART) -laitteen ominaisuuksista ja toiminnoista. On erittin suositeltavaa tutustua mys oppaan laajaan versioon, jossa laitteen ominaisuudet on selvitetty yksityiskohtaisesti. Tydellinen opas voidaan ladata SITRANS LR250 -tuotesivulta verkkosivuiltamme osoitteessa www.siemens.com/LR250. Painetun oppaan saat paikalliselta Siemens Milltronics -edustajaltasi. Tmn oppaan sislt koskevia asioita voi tiedustella osoitteesta: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 Shkposti: techpubs.smpi@siemens.com

Suomi

Copyright Siemens AG 2013. Kaikki oikeudet pidtetn


Kehotamme kyttji hankkimaan oppaan hyvksytyn painetun laitoksen tai tutustumaan Siemens Milltronics Process Instruments Inc:n laatimaan ja hyvksymn shkiseen versioon. Siemens Milltronics Process Instruments ei vastaa osittain tai kokonaan kopioitujen painettujen tai shkisten versioiden sisllst.

Vastuuvapauslauseke
Tmn oppaan sislt on tarkistettu vastaamaan kuvattua laitteistoa, mutta muutokset ovat mahdollisia. Tmn vuoksi valmistaja ei voi taata tydellist yhtpitvyytt. Tmn oppaan sislt tarkistetaan snnllisesti ja korjaukset sisllytetn seuraaviin painoksiin. Otamme parannusehdotuksia mielellmme vastaan. Teknisiin tietoihin saattaa tulla muutoksia.

MILLTRONICS on Siemens Milltronics Process Instruments Inc:n rekisterity tavaramerkki.

Tekninen tuki
Tukea on saatavissa 24 tuntia vuorokaudessa. Lydt lhimmn paikallisen Siemens Automation -toimipaikan osoitteen sek puhelin- ja faksinumeron osoitteesta www.siemens.com/automation/partner: Napsauta vlilehte Contact (Yhteystiedot) ja valitse Service (Huolto) ja napsauta sen jlkeen Service uudelleen ja etsi tuoteryhm (+Automation Technology (Automaatioteknologia) > +Sensor Systems (Anturijrjestelmt) >+Process Instrumentation (Prosessi-instrumentointi) > +Level Measurement (Tasonmittaus) > +Continuous (Jatkuva)). Valitse Radar (Tutka). Valitse maa ja sen jlkeen kaupunki/alue (City/Region). Valitse Technical Support (Tekninen tuki) kohdasta Service. Jos haluat teknist online-tukea, vieraile sivustossa www.siemens.com/automation/supportrequest Syt laitteen nimi (SITRANS LR250) tai tilausnumero ja napsauta sen jlkeen Search (Etsi). Valitse sopiva tuotetyyppi. Napsauta Next. Syt avainsana, joka kuvaa ongelmaasi. Siirry sen jlkeen sopivaan asiakirjaan tai napsauta Next, jolloin voit lhett ongelmasi kuvauksen Siemensin tekniselle tukihenkilstlle. Siemens IA/DT Technical Support Center: puhelin +49 (0)911 895 7222

A5E31997170

SITRANS LR250 (HART) PIKAKYTTOPAS

Sivu SU-1

Annettuja varoituksia on noudatettava kyttjn ja muihin henkilihin kohdistuvien henkilvahinkojen sek tuotteen ja siihen kytkettyjen varusteiden laitevahinkojen vlttmiseksi. Varoitusten yhteydess ilmoitetaan mys vaaratasosta.

Turvaohjeet

Suomi

VAROITUS liittyy tuotteen varoitussymboliin. Varoituksen huomiotta jttminen saattaa aiheuttaa hengenvaaran, vakavia ruumiinvammoja ja merkittvi aineellisia vahinkoja. VAROITUSTA kytetn, kun tuotteessa ei ole vastaavaa varoitussymbolia. Varoituksen huomiotta jttminen saattaa aiheuttaa hengenvaaran, vakavia ruumiinvammoja ja merkittvi aineellisia vahinkoja. Huomautus: Huomautuksissa on trkeit tuotteeseen tai kyttoppaaseen liittyvi tietoja.

FCC-sntjen mukaisuus
Ainoastaan asennuksissa Amerikan Yhdysvaltoihin: Amerikan Yhdysvaltojen liittovaltion televalvontaviranomaisen (Federal Communications Commission, FCC) snnt
VAROITUS: Laitteeseen tehdyt muutokset ja muokkaukset, joita Siemens Milltronics ei ole erikseen hyvksynyt, saattavat vaikuttaa laitteen kyttoikeuksiin. Huomautukset: Tm laite on FCC-sntjen kohdan 15 mukaisten luokan A digitaalilaitteiden vaatimusten mukainen. Nm vaatimukset on mritetty niin, ett ne suojaavat laitteen haitallisilta hiriilt kytettess laitetta kaupallisessa ympristss. Tm laite steilee ja kytt radiotaajuusenergiaa. Tm saattaa hirit radiolhetysten vastaanottoa, jos laitetta ei asenneta ja kytet kyttoppaan ohjeiden mukaan. Laitteen kyttminen asuinalueilla saattaa aiheuttaa hiriit radioliikenteeseen. Kyttj vastaa itse tllaisten hiriiden korjaamisesta.

SITRANS LR250
VAROITUS: SITRANS LR250:t saa kytt ainoastaan tss kyttoppaassa kuvatulla tavalla, koska muutoin laitteen suojaustaso voi heikenty. Huomautus: Tm tuote on suunniteltu kytettvksi teollisuusalueilla. Laitteen kyttminen asuinalueilla saattaa aiheuttaa hiriit erilaiseen radiotaajuuksia kyttvn viestintn. SITRANS LR250 on kaksijohtiminen 25 GHz:n pulssitekniikkaa kyttv laite, joka on tarkoitettu jatkuvaan neste- ja lietetason seurantaan varastosiliiss, mys suuressa paineessa ja korkeassa lmptilassa enintn 20 m:n (66 jalan) mittausalueella. Laite soveltuu optimaalisesti pieniin siliihin ja aineille, joiden eristevakio on pieni. Laite koostuu elektroniikkapiirist, joka on kytketty torviantenniin ja joko kierre- tai laippatyyppiseen prosessiliitntn. Saatavana on mys kierreliitntinen PVDF-antenni. SITRANS LR250 tukee HART-viestintprotokollaa ja SIMATIC PDM -ohjelmistoa. Signaalien ksittelyyn kytetn Process Intelligence -tekniikkaa.

Sivu SU-2

SITRANS LR250 (HART) PIKAKYTTOPAS

A5E31997170

Tekniset tiedot

Tydellinen erittely lytyy SITRANS LR250 (HART) -laitteen kyttoppaasta. Hyvksynttiedot on esitetty kohdassa Hyvksynnt sivulla 4.

Ympristn lmptila/kyttlmptila
Huomautukset:
Maksimilmptila riippuu prosessiliitnnst, antennin materiaaleista ja astian paineesta: katso Prosessin maksimilmptilakuvaaja sivulla 23. Prosessipaineen ja -lmptilan yksityiskohtaiset suhteutuskyrt ovat oppaan laajassa versiossa. Prosessilmptila ja paineominaisuudet riippuvat prosessiliitnnn merkintkilven tiedoista. Kilvess mainittu kytkentpiirustus voidaan ladata sivustomme LR250 tuotesivulta osoitteesta www.siemens.com/LR250, kohdasta Support > Installation Drawings > Level Measurement > Continuous - Radar (Tuki / Asennuspiirustukset / Tason mittaus / Jatkuva - Tutka). Listietoja prosessiliitnnist voi lukea asennuspiirustussivulta kohdasta Process Connection Diagrams. Signaalin amplitudi kasvaa torven lpimitan myt, joten kyt suurinta kytnnllist kokoa. Kierteiden alapuolelle voidaan asentaa valinnaisia jatkeita. Katso listietoja Prosessin maksimilmptilakuvaaja sivulla 23.
ympristn lmptila: (kotelon ymprist) 40 C...80 C (40 F...176 F) prosessin lmptila prosessiliitnnss: vakiotorviantenni: - FKM O-renkaalla: 40 oC +200 oC (-40 oF +392 oF) - FFKM O-renkaalla: 20 oC +200 oC (4 oF +392 oF) 2" NPT-/BSPT-/G-kierreliitntinen PVDF-antenni: 40+80 C (40+176 F) laitteen nimikilpi

Suomi

prosessiliitnnn kilpi

Virta
Yleiskytt Luonnostaan vaaraton Kipinimtn / Energialtaan rajoitettu Sytyttmtn (FM/CSA vain USA/Kanada) Tulenkestv Varmennettu rakenne Rjhdyssuojattu (FM/CSA vain USA/Kanada) Enintn 30 V (tasavirta) 420 mA Suurin kynnistysvirta: Nimellisjnnite 24 V (tasavirta) ja 550 Ohm. Nimellisjnnite 24 V (tasavirta) ja 250 Ohm.

katso Kynnistyskyttytyminen sivulla 24

A5E31997170

SITRANS LR250 (HART) PIKAKYTTOPAS

Sivu SU-3

Hyvksynnt
Huomautus: Laitteen nimikilvess on lueteltu kyseiselle laitteelle mynnetyt hyvksynnt. Yleinen CSAUS/C, FM, CE, C-TICK, KC Radio Eurooppa (R&TTE), FCC, Industry Canada Rjhdysvaaralliset tilat Luonnostaan vaaraton 1) (Eurooppa) ATEX II 1G, EEx ia IIC T4 ATEX II 1D, EEx tD A20 IP67 T90 C (Kansainvlinen) IECEx SIR 05.0031X, Ex ia IIC T4, Ex tD A20 IP67 T90 C (USA/Kanada) FM/CSA Luokka I, jako 1, ryhmt A, B, C, D Luokka II, jako 1, ryhmt E, F, G Luokka III T4 (Brasilia) INMETRO: DNV 12.0087 Ex ia IIC T4 Ga Ex ia ta IIIC T90 C Da IP65/IP67 -40 C Ta +80 C DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

Suomi

(Kiina) NEPSI Ex ia IIC T4/ DIP A20 TA, T90 C IP67 Kipinimtn / Energialtaan rajoitettu 2) (Eurooppa) ATEX II 3 G Ex nA/nL IIC T4 Gc (Kiina) NEPSI Ex nA II T4/ Ex nL IIC T4 Sytyttmtn 3) (USA/Kanada) Tulenkestv 4) (Eurooppa/kansainvlinen) FM/CSA Luokka I, div. 2, Ryhmt A, B, C, D T5 ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex dmbia IIC T4 Ga/Gb Ex tD A20 IP67 T90 oC INMETRO: DNV 12.0088 X Ex d ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017

(Brasilia)

ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-1:2009, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

1) 2) 3) 4)

Katso Luonnostaan vaaraton johdotus sivulla 9. Katso Kipinimtn / Energialtaan rajoitettu johdotus sivulla 11. Katso Sytyttmtn johdotus sivulla 11. Katso Tulenkestv johdotus sivulla 12.

Sivu SU-4

SITRANS LR250 (HART) PIKAKYTTOPAS

A5E31997170

Rjhdysvaaralliset tilat (jatkoa) Varmennettu rakenne 1) (Eurooppa/kansainvlinen)

(Brasilia)

ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex embia IIC T4 Ga/Gb Ex tD A20 IP67 T90 oC INMETRO DNV 12.0088 X Ex e ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-7:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

Suomi

Rjhdyssuojattu 2) (USA/Kanada) FM/CSA Luokka I, jako 1, ryhmt A, B, C, D Luokka II, jako 1, ryhmt E, F, G Luokka III T4 Meriymprist Lloyds Register of Shipping ABS-tyyppihyvksynt

Tulenkestv / Varmennettu rakenne (Kiina) NEPSI Ex dmbia IIC T4/ Ex embia IIC T4/ DIP A20 TA, T90 oC IP67

Painesovellukset
VAROITUKSET:

l yrit irrottaa, poistaa tai purkaa prosessiliitnt tai mittarikoteloa silin ollessa paineistettuna. Kyttj vastaa laipan kiinnitys- ja tiivistemateriaaleista, kytttavasta ja huoltokelpoisuudesta. Virheellinen asennus saattaa aiheuttaa prosessipaineen hvimisen.
Huomautukset: Prosessiliitnnn kilpe ei saa poistaa prosessipaineen rajapinta-asennelmasta2). Jos laitekokonaisuus vaihdetaan uuteen, prosessiliitnnn kilpi on siirrettv korvaavaan yksikkn. SITRANS LR250 -yksikt on testattu hydrostaattisesti, ja ne tyttvt tai ylittvt ASME Boiler and Pressure Vessel Code -snnstn ja Euroopan yhteisn painelaitedirektiivin vaatimukset.

1) 2)

Katso Varmennetun rakenteen johdotus sivulla 12. Katso Rjhdyssuojattu johdotus (vain USA/Kanada) sivulla 13.

A5E31997170

SITRANS LR250 (HART) PIKAKYTTOPAS

Sivu SU-5

Painelaitedirektiivi 97/23/EY 1)
Siemens-tasoantureissa, jotka on varustettu laippa- tai kierre-tyyppisell tai hygieenisell clamp-tyyppisell prosessiliitnnll, ei ole omaa paineenpitv koteloa, joten ne eivt kuulu painelaitedirektiivin soveltamisalaan paineenalaisina lislaitteina tai varolaitteina (katso Euroopan komission opas 1/8).

Suomi

Asennus
VAROITUKSET: Asennuksen saa suorittaa ainoastaan asiantunteva henkilst ja asennuksessa on noudatettava paikallisia mryksi. Kun ksittelet laitetta, pid kiinni kotelosta lk laitteen prosessiliitnnn kilvest. Nin vltt vauriot. Ole erityisen varovainen ksitellesssi kierreliitntist PVDF-antennia. Antennin pinnan, erityisesti sen krjen, vaurioituminen voi vaikuttaa sen suorituskykyyn. Rakenneaineet valitaan yleisiin kytttarkoituksiin niiden kemiallisen yhteensopivuuden (eli reagoimattomuuden) perusteella. Erityisympristille altistuvissa sovelluksissa kemiallinen yhteensopivuus on tarkistettava taulukoista ennen asennusta. Huomautukset:

Euroopan unionissa ja sen jsenvaltioissa asennuksen on oltava standardin ETSI EN 302372 mukainen. Katso hyvksynnt laitteen nimikilvest. Kunkin prosessiliitnnn runkoon leimatut sarjanumerot toimivat ainutkertaisina tunnistusnumeroina, jotka osoittavat valmistuspivn. Esimerkki: KKPPVV XXX (KK = kuukausi, PP = piv, VV = vuosi ja XXX= valmistetun laitteen jrjestysnumero) Lismerkinnt osoittavat (jos tilaa j) laipan rakenteen, koon, paineluokan, materiaalin ja materiaalin lmpkoodin.

Steilyaukon suunnittelu
Kierreliitntinen PVDF-antenni
vhimmisvlys: 10 mm (0,4")

Torviantenni ruostumatonta terst


vhimmisvlys: 10 mm (0,4")

Pienin lpim. 51 mm (2")

Antennin pn on ulotuttava vhintn 10 mm (0,4") phn steilyaukosta, jotta vltetn vrt kaiut. Kierreliitntisen PVDF-antennin pienin suositeltu steilyaukon lpimitta on 51 mm (2"). Antennijatke (100 mm / 3,93") on saatavana kaikkiin muihin versioihin paitsi kierreliitntiseen PVDF-antenniin.

1)

Prosessipaineen rajapinta-asennelmaan sisltyy osia, jotka estvt prosessiastian paineen purkautumista: tll tarkoitetaan prosessiliitnnn runkoa ja lhetint mutta yleens ei shkiset osat sisltv koteloa.

Sivu SU-6

SITRANS LR250 (HART) PIKAKYTTOPAS

A5E31997170

Steilyaukon sijoitus
Vlt sijoittamasta keskelle pitkiss kapeissa siliiss, jotka voivat aiheuttaa vri kaikuja.

Suomi

Ymprist
Asennuspaikan on oltava laitteen kotelon luokituksen ja rakenneaineiden mukainen. Asenna aurinkosuoja, jos laite asennetaan suoraan auringonvalolle alttiiksi.
suositeltava vltettv

Steen kulma
Steen kulma on keilan leveys, jonka kohdalla energiatiheys on puolet huippuenergiatiheydest. Huippuenergiatiheys sijoittuu suoraan antennin eteen sen keskikohdalle. Signaali vlittyy mys steen kulman ulkopuolelle, joten se voi havaita mys vri kohteita.

steilykeila steen kulma: kierreliitntinen PVDFantenni = 19 1,5":n torvi = 19 2":n torvi = 15 steen 3":n torvi = 10 kulma 4":n torvi = 8

Steilykeila
Varmista, ettei steilykeilan alueella ole hiritsevi tikkaita, putkia, I-tankoja eik tyttvirtauksia.

Ohjelmoimaan psy
Sopivassa asennuspaikassa laitteen nkee helposti ja sen voi ohjelmoida helposti ksiohjelmointilaitteella.

Asennusohjeet
VAROITUS: Painesovelluksissa on kytettv PTFE-nauhaa tai muuta hyvksytty kierretiivistett ja prosessiliitnt on kiristettv koneellisesti. Enimmisvntvoima on 40 Nm (30 ft.lbs.). Huomautus: Laitetta voidaan kiert rajattomasti sen vahingoittumatta.

Kierreversiot
1) 2) Ennen kuin tynnt laitteen asennusliitntns, varmista, ett kierteet ovat samaa tyyppi, jotta ne eivt vahingoitu. Kiinnit laite prosessiliitntn kiertmll paikoilleen ja kirist ksin tai avaimella. Painesovelluksissa on luettava edell oleva varoitus. VAROITUS: Kyttj vastaa laipan kiinnitys- ja tiivistemateriaaleista, kytttavasta ja huoltokelpoisuudesta.

Laipalliset versiot

A5E31997170

SITRANS LR250 (HART) PIKAKYTTOPAS

Sivu SU-7

Johdotus
Virta
VAROITUKSET: Tasavirtaliittimiin on sytettv virtaa virtalhteest, jossa tulo- ja lhtpuoli on eristetty shkisesti toisistaan, jotta sovellettavat IEC 61010-1 turvallisuusvaatimukset tyttyvt. Kaikissa kenttjohtimissa on oltava ilmoitetun jnnitteen mukainen eristys.

Suomi

SITRANS LR250:n kytkeminen


VAROITUKSET: Tarkista laitteen nimikilvest ja prosessikilvest hyvksyntluokitus. Kyt asianmukaisia johdintiivisteit, jotta laitteen IP- tai NEMA-luokitus silyy. Katso Kytkentmritykset asennettaessa rjhdysvaarallisiin tiloihin sivulla 9. Huomautukset: Kyt kierrettyj parikaapeleita: lankamitaltaan 2214 (0,34 mm22,5 mm2). Standardien tai shkasetusten mukaiseen johdotukseen saatetaan tarvita erilliset kaapelit ja johtimet.
Irrota kansilukituksen ruuvit 2 mm:n kuusiokoloavaimella. tulppa (IP 68) valinnainen kaapeliholkki1) 2) (tai NPT-kaapelin sisntulo2)) kaapelisuojaus (jos kytss)

1) 2) 3) 4) 5)

Pura kaapelin vaippaa noin 70 mm:n (2,75") matkalta toisesta pst ja pujota johdot holkin lpi. 1) 2) Kytke johdot liittimiin yll kuvatulla tavalla: napaisuus on merkitty riviliittimeen. Maadoita laite paikallisten vaatimusten mukaisesti. Kirist holkki, niin ett johdot lukittuvat hyvin paikoilleen. Sulje kansi ennen ohjelmointia ja laitteen konfigurointia.

1) 2)

Voidaan toimittaa laitteen mukana. Jos vedt kaapelin putken kautta, kyt vain hyvksyttyj ja sopivankokoisia vedenkestvi keskiit.

Sivu SU-8

SITRANS LR250 (HART) PIKAKYTTOPAS

A5E31997170

HART-kytkent
Tyypillinen ohjelmoitavan logiikkaohjaimen mA-kokoonpano ja HART-laite 1) 2)
aktiivinen PLC virtalhde R= 250 HART-modeemi HART communicator

SITRANS LR250

Huomautukset: Jrjestelmsuunnittelusta riippuu, onko virtalhde erillinen PLC-ohjaimesta vai sen yhdysrakenteinen osa HART-vastus1) on rajoitettava sallitulle kyttalueelle2). Laitteeseen saa kytke vain yhden HART-ohjelmointilaitteen kerrallaan.

Suomi

Kytkentmritykset asennettaessa rjhdysvaarallisiin tiloihin


Rjhdysvaarallisten tilojen asennuksia varten on kaksi johdotusvaihtoehtoa. Tarkista aina laitteen nimikilvest ja prosessiliitnnn kilvest hyvksyntluokitus.

1. Luonnostaan vaaraton johdotus


Laitteen nimikilpi (ATEX/IECEX/INMETRO/C-TICK)

Tuotekilvess mainittu ATEX-sertifikaatti voidaan ladata tuotesivulta sivustostamme osoitteesta www.siemens.com/LR250. Siirry kohtaan Support > Approvals / Certificates (Tuki / Hyvksynnt / Sertifikaatit). Tuotekilvess mainittu IECEx-sertifikaatti on tarkistettavissa IECEx:n verkkosivustossa. Mene osoitteeseen http://iecex.iec.ch ja napsauttamalla kohtaa Ex Equipment Certificates of Conformity. Syt sen jlkeen sertifikaatin numero IECEx SIR 05.0031X.

1) 2)

Tykalun silmukkavastus = johdinresistanssi plus 250 Ohmia (vastus). Katso joko Kyr 1: Yleiskytt, luonnostaan turvallinen, kipinimtn / energialtaan rajoitettu / sytyttmtn sivulla 23 tai Kyr 2: Tulenkestv, varmennettu rakenne, rjhdyssuojattu sivulla 24.

A5E31997170

SITRANS LR250 (HART) PIKAKYTTOPAS

Sivu SU-9

Luonnostaan vaaraton johdotus (jatkoa)


Laitteen nimikilpi (FM/CSA) Suomi
FM/CSA: luonnostaan turvallisen kytkentpiirustus numero 23650653 voidaan ladata tuotesivulta sivustostamme osoitteesta www.siemens.com/LR250. Siirry kohtaan Support > Installation Drawings > Level Measurement > Continuous - Radar (Tuki / Asennuspiirustukset / Tason mittaus / Jatkuva Tutka). Tehontarve: katso Kyr 1: Yleiskytt, luonnostaan turvallinen, kipinimtn / energialtaan rajoitettu / sytyttmtn sivulla 23. Johdotusvaatimukset: noudata paikallisia mryksi. NEMA 4X / tyyppi 4X / NEMA 6, IP67, IP68 -tyypin ulkosovelluksissa on kytettv hyvksyttyj ply- ja vesitiiviit johdintiivisteit. Katso Rjhdysvaarallisten tilojen asennusta koskevat ohjeet sivulla 13. Suositeltavien luonnostaan turvallisten suojarajoittimien luettelo Passiiviset ohitusdiodisuojarajoittimet sivulla 10 ja Aktiiviset suojarajoittimet (toistuvatoimiset suojarajoittimet) sivulla 11. Huomautus: Sopivan PLC-syttmoduulin, teholhteen tai suojarajoittimen valinta edellytt tietoja luontaisesta turvallisuudesta ja sovelluksesta. Asentajan vastuulla on varmistaa, ett luonnostaan turvallinen asennus on sek laitteen hyvksyntvaatimusten ett asiaa koskevien kansallisten kytnnesntjen mukainen.

Passiiviset ohitusdiodisuojarajoittimet
Huomautus: Tarvitaan tarkasti sdettv virtalhde. Valmistaja
MTL MTL Stahl Stahl

Osanro
787SP+ (kaksikanavainen) 7787P+ (kaksikanavainen) 9001/01-280-100-10 (yksikanavainen) 9002/01-280-110-10 (kaksikanavainen)

Varmista, ett suojarajoittimen kuvaus vastaa LR250:n luonnostaan turvallisen asennuksen (IS) tuloparametreja, tekemll seuraavat laskelmat: Re-e = suojan enimmisvastus pst phn Rloop = silmukkavastus (johdinresistanssi + mahdollinen muu sarjaresistanssi, esim. PLC-sytt ja/tai nytt) Vbarrier = suojien aiheuttamien eplineaaristen jnnitteenalenemien arvo 1) Mrit Re-e-arvo teknisist tiedoista. 2) Laske Rloop-kokonaisarvo. 3) Laske Rworking = Re-e + Rloop.

Passiivisen suojarajoittimen valinta SITRANS LR250 -laitteeseen

Sivu SU-10

SITRANS LR250 (HART) PIKAKYTTOPAS

A5E31997170

Passiiviset ohitusdiodisuojarajoittimet (jatkoa) 4) Mrit suojarajoittimen Vbarrier-arvo teknisist tiedoista (esimerkiksi diodien aiheuttamat jnnitteenalenemat). 5) Laske Vworking = Vsupply Vbarrier. 6) Vahvista Vworking- ja Rworking-arvoja kytten, ett laskutoimituksen arvo sijoittuu varjostetulle alueelle kuvaajassa Kyr 1: Yleiskytt, luonnostaan turvallinen, kipinimtn / energialtaan rajoitettu / sytyttmtn sivulla 23.

Suomi

Aktiiviset suojarajoittimet (toistuvatoimiset suojarajoittimet)


Valmistaja
MTL MTL Siemens Stahl MTL MTL

Osanro
706 7206 SITRANS I 7NG4124 9001/51-280-110-14 E02009 E02010

2. Kipinimtn / Energialtaan rajoitettu johdotus

Tuotekilvess mainittu ATEX-sertifikaatti voidaan ladata tuotesivulta sivustostamme osoitteesta www.siemens.com/LR250. Siirry kohtaan Support > Approvals / Certificates (Tuki / Hyvksynnt / Sertifikaatit). Tehontarve: katso Kyr 1: Yleiskytt, luonnostaan turvallinen, kipinimtn / energialtaan rajoitettu / sytyttmtn sivulla 23. Noudata johdotuksessa aina paikallisia mryksi.

3. Sytyttmtn johdotus

FM/CSA Luokka 1, div. 2 johdotuspiirustus numero 23650673 voidaan ladata tuotesivulta sivustostamme osoitteesta www.siemens.com/LR250. Siirry kohtaan Support > Installation Drawings > Level Measurement > Continuous - Radar (Tuki / Asennuspiirustukset / Tason mittaus / Jatkuva - Tutka). Tehontarve: katso Kyr 1: Yleiskytt, luonnostaan turvallinen, kipinimtn / energialtaan rajoitettu / sytyttmtn sivulla 23.

A5E31997170

SITRANS LR250 (HART) PIKAKYTTOPAS

Sivu SU-11

4. Tulenkestv johdotus

Suomi

Tuotekilvess mainittu ATEX-sertifikaatti voidaan ladata tuotesivulta sivustostamme osoitteesta www.siemens.com/LR250. Mene osoitteeseen Support > Approvals / Certificates (Tuki / Hyvksynnt / Sertifikaatit). Tuotekilvess mainittu IECEx-sertifikaatti on tarkistettavissa IECEx:n verkkosivustossa. Mene osoitteeseen http://iecex.iec.ch ja napsauttamalla kohtaa Ex Equipment Certificates of Conformity. Syt sen jlkeen sertifikaatin numero IECEx SIR 08.0107X. Tehontarve: katso Kyr 1: Yleiskytt, luonnostaan turvallinen, kipinimtn / energialtaan rajoitettu / sytyttmtn sivulla 23. Noudata johdotuksessa aina paikallisia mryksi. Katso mys Rjhdysvaarallisten tilojen asennusta koskevat ohjeet sivulla 13 ja edell mainittu ATEX-sertifikaatti.

5. Varmennetun rakenteen johdotus

Tuotekilvess mainittu ATEX-sertifikaatti voidaan ladata tuotesivulta sivustostamme osoitteesta www.siemens.com/LR250. Mene osoitteeseen Support > Approvals / Certificates (Tuki / Hyvksynnt / Sertifikaatit). Tuotekilvess mainittu IECEx-sertifikaatti on tarkistettavissa IECEx:n verkkosivustossa. Mene osoitteeseen http://iecex.iec.ch ja napsauttamalla kohtaa Ex Equipment Certificates of Conformity. Syt sen jlkeen sertifikaatin numero IECEx SIR 08.0107X. Tehontarve: katso Kyr 2: Tulenkestv, varmennettu rakenne, rjhdyssuojattu sivulla 24. Noudata johdotuksessa aina paikallisia mryksi. Katso mys Rjhdysvaarallisten tilojen asennusta koskevat ohjeet sivulla 13 ja edell mainittu ATEX-sertifikaatti.

Sivu SU-12

SITRANS LR250 (HART) PIKAKYTTOPAS

A5E31997170

6. Rjhdyssuojattu johdotus (vain USA/Kanada)

Suomi

FM/CSA: Rjhdyssuojattu johdotuspiirustus numero A5E02257843 voidaan ladata tuotesivulta sivustostamme osoitteesta www.siemens.com/LR250. Siirry kohtaan Support > Installation Drawings > Level Measurement > Continuous - Radar (Tuki / Asennuspiirustukset / Tason mittaus / Jatkuva - Tutka). Tehontarve: katso Kyr 2: Tulenkestv, varmennettu rakenne, rjhdyssuojattu sivulla 24.

Rjhdysvaarallisten tilojen asennusta koskevat ohjeet


(ATEX-direktiivi 94/9/EY, liite II, kohta 1/0/6)
Seuraavat ohjeet koskevat sertifikaattien SIRA 06ATEX2358X, SIRA 08ATEX1301X ja SIRA 09ATEX4153X mukaisia laitteita. 1) 2) Kytt- ja kokoamisohjeet ovat pohjeissa. Laite on sertifioitu luokan 1GD -laitteeksi sertifikaatissa SIRA 06ATEX2358X; luokan 1/2 GD, 1D, 2D laitteeksi sertifikaatissa SIRA 08ATEX1301X ja luokan 3G laitteeksi sertifikaatissa SIRA 09ATEX4153X. 3) Laitetta saadaan kytt palavien kaasujen ja hyryjen yhteydess laiteluokissa IIC, IIB ja IIA sek lmptilaluokissa T1, T2, T3 ja T4. 4) Laitteen kotelointiluokka on IP67 ja lmptilaluokka T90 C. Laitetta voidaan kytt ympristiss, joissa on leimahtavia plyj. 5) Laite on sertifioitu kytettvksi 40 C+80 C:n ympristn lmptilassa. 6) Laitetta ei ole luokiteltu direktiivin 94/9/EY liitteen II kohdan 1.5 mukaiseksi turvallisuuteen mytvaikuttavaksi laitteeksi: katso Painelaitedirektiivi 97/23/EY sivulla 6. 7) Tmn laitteen saa asentaa ja tarkistaa vain koulutettu henkil sovellettavan standardin (EN 60079-14 ja EN 60079-17 Euroopassa) mukaisesti. 8) Laitetta ei voi korjata. 9) X-pte sertifiointinumerossa tarkoittaa, ett laitteen kyttn sovelletaan turvallista kytt koskevia erikoisehtoja. Laitetta asentavien tai tarkastavien henkiliden on voitava tarkistaa tarkastustodistukset. 10) Jos laite joutuu todennkisesti kosketuksiin voimakkaasti reagoivien aineiden kanssa, kyttjn on varmistettava, ett kyseiset aineet eivt pse vaikuttamaan laitteeseen ja heikentmn sen suojausta. Voimakkaasti reagoivat aineet: esimerkiksi metalleja syvyttvt happoliuokset tai kaasut ja polymeeriin vaikuttavat liuottimet. Varotoimet: sen tarkistaminen materiaalin teknisist tiedoista, ett materiaali kest tiettyj kemikaaleja.

A5E31997170

SITRANS LR250 (HART) PIKAKYTTOPAS

Sivu SU-13

SITRANS LR250: n ohjelmointi


Suomi

Katso Ohjattu pikakytttoiminto ksiohjelmointilaitteen avulla sivulla 18. Katso Ohjattu pikakytttoiminto SIMATIC PDM:n avulla sivulla 21.

SITRANS LR250: n aktivointi


Kytke laitteeseen virta. SITRANS LR250 kynnistyy automaattisesti mittaustilaan.

Nestekidenytt Mittaustilan nytt (normaali toimintatila) 1) 2)


1 2 3 4
(((

1 kytkin:1) ilmaisimen lineaariset yksikt tai % 2 valittu toiminta: level (pinnan korkeus), space (vli), distance (etisyys) tai volume (tilavuus) 3 mittausarvo (pinnan korkeus tai tilavuus, vli tai etisyys) 4 yksikk 5 pinnankorkeuden palkkikuvaaja 6 toissijainen alue nytt pyydettess2) elektroniikan lmptilan, kaiun luotettavuusarvon, silmukkavirran tai etisyyden 7 tekstialue nytt tilaviestit 5 8 laitteen tilailmaisin (katso tiedot oppaan laajasta versiosta)

Vikaosoittimet S: 0 LOE
Jos laitteessa on vika, tekstialue (7) nytt virhekoodin ja virheviestin ja huoltotarpeen ilmaisin nkyy laitteen tilailmaisimen kohdalla (8)

Program-ohjelmointitilan nytt Navigaationytt

valikon kohdan Nkyviin tuleva valikkopalkki nimi numero tarkoittaa, ett valikkoluettelo on valittu niin pitk, etteivt kaikki kohdat ny. valikkopalkki kohta kohdan Kun neli on puolivliss neli valikkopalkkia, se tarkoittaa, ett nykyinen kohta on puolivliss luetteloa. Valikon kohtaa ilmaisevan nelin pituus ja sijainti valikkopalkissa osoittaa valikkoluettelon pituuden ja nykyisen kohdan suhteellisen sijainnin luettelossa. Pitempi neli tarkoittaa, ett kohtia on vhemmn.

valikon

Parametrinytt
parametrin nimi parametrin numero parametrin arvo/valinta

Muokkausnytt

1) 2)

Vaihda painamalla nuolta YLS tai ALAS. Nppimen painalluksen jlkeen. Katso listietoja taulukosta Toiminto mittaustilassa sivulla 15.

Sivu SU-14

SITRANS LR250 (HART) PIKAKYTTOPAS

A5E31997170

Ksiohjelmointilaite (osanro 7ML1930-1BK)


Tilattava erikseen. 1) 2) 3)

Painike Toiminto mittaustilassa


Pivitt silmukkavirran1). Pivitt kotelon sisisen lmptilalukeman1). Pivitt kaiun luotettavuusarvon1).

Suomi

C
Pivitt etisyysmittauksen1). Mode siirt laitteen PROGRAM-ohjelmointitilaan2). Nuoli OIKEALLE siirt PROGRAM-ohjelmointitilaan3).

Nuoli YLS tai ALAS vaihtaa lineaariset yksikt %:eihin ja pinvastoin

Ohjelmointi ksiohjelmointilaitteesta
Huomautukset: Voit vaihtaa mittaustilan ohjelmointitilaan ja pinvastoin painamalla Mode (Tila). Kun laite on PROGRAM-ohjelmointitilassa, sen nytt pysyy muuttumattomana eik vastaa laitteessa tapahtuvia muutoksia. SITRANS LR250 palaa automaattisesti mittaustilaan, kun sit ei ole kytetty vhn aikaan PROGRAM-ohjelmointitilassa (aika vaihtelee valikon tason mukaan 15 sekunnista 10 minuuttiin).

Parametrivalikot

Parametreilla on nimi, ja ne on jrjestetty toimintoryhmittin. Katso tydellinen parametriluettelo ja ohjeet kyttoppaan laajasta versiosta.

1. QUICK START 2. SETUP 2.1. DEVICE ............. 2.7. LINEARIZATION 2.7.1. VOLUME 2.7.1.1. VESSEL SHAPE

1) 2)

3)

Uusi arvo nkyy nestekidenytn toissijaisella alueella. Nyttn tulee viimeksi tll kyttjaksolla nytetty valikkotaso, jos virtaa ei ole sammutettu PROGRAM-ohjelmointitilasta poistumisen jlkeen tai jos PROGRAM-ohjelmointitilan kytst on kulunut yli 10 minuuttia. Nyttn tulee yltason valikko. Avaa yltason valikon.

A5E31997170

SITRANS LR250 (HART) PIKAKYTTOPAS

Sivu SU-15

1.

Siirry PROGRAM-ohjelmointitilaan
Suuntaa ohjelmointilaite nytt kohti (enintn 300 mm:n [1 jalan] pst).

ksiohjelmointilaite nytt

Suomi

Nuoli OIKEALLE aktivoi PROGRAMohjelmointitilan ja avaa valikkotason 1. Mode avaa viimeksi PROGRAMohjelmointitilassa enintn 10 min aiemmin avatun valikkotason tai tason 1, jos virta on katkaistu.

Enint. 300 mm (1 jalan)

2.

Navigointi: painikkeiden toiminnot navigaatiotilassa

Huomautus: Pset nopeasti ksiksi parametreihin ksiohjelmointilaitteella painalla Home -painiketta ja antamalla sen jlkeen valikon numeron, esimerkiksi: 2.7.1 (Volume tilavuus).

Painike

Nimi
Nuoli YLS tai ALAS

Valikon taso
valikko tai parametri valikko parametri

Toiminta navigaatiotilassa
Siirry vierittmll edelliseen tai seuraavan valikkoon tai parametriin. Siirry valitun valikon ensimmiseen parametriin tai avaa seuraava valikko. Avaa muokkaustila. Avaa ylvalikon. Siirt MITTAUSTILAAN. Avaa yltason valikon: valikko 1.

Nuoli OIKEALLE

Nuoli valikko tai VASEMMALLE parametri


Tila Home-painike valikko tai parametri valikko tai parametri

3.

Muokkaus PROGRAM-ohjelmointitilassa Luetellun vaihtoehdon valinta


a) b) Siirry haluttuun parametriin. Paina nuoli OIKEALLE parametrinyttn. , niin siirryt

parametrin nimi

parametrin numero

nykyinen valinta

c)

Paina nuoli OIKEALLE uudestaan, niin siirryt muokkaustilaan. Valittuna oleva vaihtoehto nkyy korostettuna. Vierit uuden valinnan kohdalle.

Sivu SU-16

SITRANS LR250 (HART) PIKAKYTTOPAS

A5E31997170

d)

Hyvksy se painamalla nuoli OIKEALLE . Nestekidenytt palaa parametrinyttn ja nytt uuden valinnan.
parametrin nimi

Numeroarvon muuttaminen
a) b) Siirry haluttuun parametriin. Paina nuoli OIKEALLE uudestaan, niin siirryt muokkaustilaan. Nykyinen arvo nkyy korostettuna. Nppile uusi arvo. Hyvksy se painamalla nuoli OIKEALLE . Nestekidenytt palaa parametrinyttn ja nytt uuden valinnan.

parametrin numero

Suomi

c) d)

nykyinen arvo

Painikkeiden toiminnot muokkaustilassa


Painike Nimi Toiminta muokkaustilassa
Vaihtoehtojen valinta Nuoli YLS tai ALAS Numeron muokkaus Vaihtoehtojen valinta Nuoli OIKEALLE Numeron muokkaus Vaihtoehtojen valinta VASEMMALLE -nuoli Numeron muokkaus Numeron muokkaus Vieritt kohtaan. - Suurentaa tai pienent lukua. - Vaihtaa plus- tai miinusmerkin. - Hyvksyy tiedot (tallentaa parametrin). - Siirt muokkaustilasta navigaatiotilaan. - Siirt kohdistinta yhden merkin oikealle - tai kohdistimen ollessa Enter-merkiss hyvksyy tiedot ja siirt muokkaustilasta navigaatiotilaan. Peruuttaa muokkaustilan muuttamatta parametria. - Siirt kohdistimen plus-/miinusmerkkiin, jos painetaan ensimmisen - tai siirt kohdistimen yhden merkin vasemmalle. Tyhjent nytn.

Poisto

Desimaalipiste

Numeron muokkaus

Lis desimaalipisteen.

Plus- tai miinusmerkki

Numeron muokkaus

Muuttaa sytetyn arvon etumerkin.

Numerot

Numeron muokkaus

Lis kyseisen merkin.

A5E31997170

SITRANS LR250 (HART) PIKAKYTTOPAS

Sivu SU-17

Ohjattu pikakytttoiminto ksiohjelmointilaitteen avulla


1. Pikakytttoiminto
a) Suuntaa ohjelmointilaite nytt kohti enintn 300 mm:n (1 jalan) etisyydelt ja paina nuoli OIKEALLE valikkotaso 1. b) Siirry valikon kohtaan 1.1 ja avaa parametrinytt painamalla nuoli OIKEALLE kahdesti. c) Paina nuoli OIKEALLE ALAS d) , jolloin muokkaustila tulee nyttn, tai paina nuoli , jolloin PROGRAM-tila aktivoituu ja nyttn aukeaa

Suomi

, jos haluat hyvksy oletusarvot ja siirty suoraan seuraavan kohtaan.

Asetuksia muutetaan vierittmll haluttuun kohtaan tai syttmll uusi arvo painikkeilla.

e)

Kun olet muuttanut arvoa, hyvksy se painamalla nuoli OIKEALLE seuraavaan kohtaan painamalla nuoli ALAS .

ja siirry

f)

Pikakytttoiminnon asetukset tulevat voimaan vasta, kun valitset Yes (Kyll) Apply changes (Toteuta muutokset) kohdassa (1.8.).

1.1. Kieli Valitsee nestekidenytn kielen. Asetus tulee voimaan vlittmsti.


Vaihtoehdot
ENGLISH (ENGLANTI), DEUTSCH (SAKSA), FRANAIS (RANSKA), ESPAOL (ESPANJA)

1.2. Materiaali Vaihtoehdot


LIQUID (NESTE) LIQUID LOW DK (neste, pieni eristevakio)

1.3. Vastenopeus Mritt laitteen reagointinopeuden mittaustulosten muutoksiin valitulla mittausalueella.


Vastenopeus
SLOW (HIDAS) MED (KESKI) FAST (NOPEA)

Tytt-/tyhjennysnopeus
0,1 m / min (9,75 cm / min) 1,0 m / min (99,97 cm / min) 10,0 m / min (999,74 cm / min)

Kyt asetusta, joka on hieman nopeampi kuin tytn tai tyhjentymisen enimmisnopeus (sen mukaan, kumpi on suurempi).

1.4. Yksikt Mittapn mittausyksikt. Vaihtoehdot


M, CM, MM, FT, IN

Sivu SU-18

SITRANS LR250 (HART) PIKAKYTTOPAS

A5E31997170

1.5. Kytttila
mittapn viitepiste

mittapn viitepiste vli

etisyys korkea kalibrointipiste

Suomi

mittapn viitepiste

taso matala kalibrointipiste

Kytt
NO SERVICE (EI TOIMINTAA) LEVEL (TASO) SPACE (VLI) DISTANCE (ETISYYS)
1)

Kuvaus
Mittaustietoja ja niihin liittyv silmukkavirtaa ei pivitet, laite palautuu vikaturvatilaan1). Etisyys materiaalin pintaan suhteessa matalaan kalibrointipisteeseen. Etisyys materiaalin pintaan suhteessa korkeaan kalibrointipisteeseen. Etisyys materiaalin pintaan suhteessa mittapn viitepisteeseen.

Listietoja vikaturvatilasta on oppaan laajassa versiossa.

1.6. Matala kalibrointipiste Etisyys mittapn viitepisteest matalaan kalibrointipisteeseen: yleens prosessin perustaso. Katso kuva kohdasta Kytttila (1.5.). Arvot
Alue: 0,0020,00 m

1.7. Korkea kalibrointipiste Etisyys mittapn viitepisteest korkeaan kalibrointipisteeseen: yleens prosessin ylin taso. Katso kuva kohdasta Kytttila (1.5.) . Arvot
Alue: 0,0020,00 m

1.8. Apply? (Toteuta muutokset) Pikakyttasetukset tallentuvat valitsemalla thn kohtaan Yes (Kyll). Vaihtoehdot
YES (KYLL), NO (EI), DONE (VALMIS) (Nyttn tulee DONE, kun Quick Start -pikakynnistysasetukset asetettu suoritettu onnistuneesti). -painiketta. SITRANS LR250 on nyt toimintavalmis.

Palaa mittaustilaan painamalla Mode

Huomautus: Jos sovelluksessasi on siliss esteit, katso listietoja automaattisesta hirikaiun vaimennuksesta.

A5E31997170

SITRANS LR250 (HART) PIKAKYTTOPAS

Sivu SU-19

SITRANS LR250 - viestint: HART


Suomi

Luettelo kaikista kytettviss olevista parametreista lytyy oppaan laajasta versiosta. Suosittelemme kyttmn laitteen ohjelmoinnin yhteydess SIMATIC Process Device Manager (PDM) -ohjelmistoa. Sovellusoppaat SIMATIC PDM:n avulla mriteltville HART-laitteille voidaan ladata tuotesivulta sivustostamme osoitteesta www.siemens.com/LR250 kohdasta Support (Tuki).

SIMATIC PDM
SIMATIC PDM on ohjelmistopaketti, jolla otetaan kyttn ja kunnossapidetn SITRANS LR250:t ja muita prosessilaitteita. Katso yksityiskohtaiset tiedot SIMATIC PDM:n kytst kyttohjeista tai online-ohjeista. (Voit lukea listietoja osoitteesta www.siemens.com/simatic-pdm.)

Poista puskurit kytst

Tm mukauttaa SIMATIC PDM:n ja HART-modeemin Windows 2000 ja Windows XP 1) kyttjrjestelmiin. Huomautukset: SIMATIC PDM toimii ainoastaan Windows XP Professional -versiossa, ei Homeversiossa. Puskureiden poistamiseen kytst tarvitaan jrjestelmkyttjn oikeudet. 1) 2) 3) 4) 5) 6) Aloita konfigurointi napsauttamalla Kynnist/Asetukset/Ohjauspaneeli. Kaksoisnapsauta Jrjestelm ja valitse Laitteisto-vlilehti ja napsauta Laitehallintapainiketta. Avaa Portit-kansio ja kaksoisnapsauta jrjestelmn kyttm COM-porttia, jolloin Tietoliikenneportin ominaisuudet -ikkuna aukeaa. Valitse Portin asetukset -vlilehti ja kaksoisnapsauta Lisasetukset-painiketta. Jos valintaruutu Kyt FIFO-puskureita on valittuna, poista rasti napsauttamalla sit. Sulje ikkuna napsauttamalla OK. Sulje kaikki ikkunat ja kynnist laite uudelleen.

Pivit shkinen laitemritys (EDD)


Huomautukset:
SIMATIC PDM Rev. 5.2 SP1 on tuettu ainoastaan peruskokoonpanon mrityksen ja vianmrityksen osalta. Ohjatun pikakytttoiminnon ja muiden kehittyneiden toimintojen kyttn tarvitaan versio 6.0 SP4 tai uudempi. SIMATIC PDM:ss laitemritys lytyy laiteluettelosta (Device Catalog) kohdasta Sensors/Level/Echo/Siemens Milltronics/SITRANS LR 250. Tarkista tuotesivulta verkkosivustostamme osoitteessa www.siemens.com/LR250, onko Sinulla varmasti SIMATIC PDM:n uusin versio ja uusimmat Service Pack (SP) - ja hot fix (HF) -tiedostot. Siirry kohtaan Support > Software Downloads (Tuki / Ohjelmisto / Lataukset).

Uuden laitemrityksen asentaminen:


1) 2) 3) Lataa laitemritys tuotesivulta sivustostamme osoitteesta www.siemens.com/LR250 ja tallenna tiedostot tietokoneellesi. Pura pakattu tiedosto helppopsyiseen kansioon. Kynnist SIMATIC PDM Manager Device Catalog, selaa puretun EDD-tiedoston kohdalle ja valitse se.
Windows on Microsoft Corporationin rekisterity tavaramerkki.

1)

Sivu SU-20

SITRANS LR250 (HART) PIKAKYTTOPAS

A5E31997170

Uuden laitteen kokoonpanoasetukset


Huomautus: Vaikka napsautat Cancel (Peruuta), kun lataat tietoja laitteesta SIMATIC PDM
-ohjelmistoon, osa parametreista pivittyy. 1) 2) 3) 4) 5) Tarkista, ett sinulla on uusin laitemritys, ja pivit se tarvittaessa (katsoPivit shkinen laitemritys (EDD) sivulla 20). Kynnist SIMATIC Manager ja luo uusi projekti laitteelle LR250. Sovellusoppaat SIMATIC PDM:n avulla mriteltville HART-laitteille voidaan ladata tuotesivulta sivustostamme osoitteesta www.siemens.com/LR250. Avaa valikko Device Master Reset (Laite Isntlaitteen palautus) ja napsauta OK. Laite nollautuu tehdasasetuksiin. Kun nollaus on tehty, lataa parametrit PC/PG-laitteeseen. Mrit laite ohjatun pikakytttoiminnon (Quick Start Wizard) avulla.

Suomi

Ohjattu pikakytttoiminto SIMATIC PDM:n avulla


Ohjatun pikakytttoiminnon asetukset ovat sidoksissa toisiinsa, ja muutokset tulevat voimaan vasta, kun napsautat pikakytn vaiheen 5 lopuksi FINISH AND DOWNLOAD , jolloin asetukset tallentuvat pysyvsti ilman verkkoyhteytt ja siirtyvt laitteeseen. Napsauta BACK (TAKAISIN), jos haluat palata taaksepin tarkistamaan jonkin asetuksen, tai Cancel (Peruuta), jos haluat poistua Quick Start -pikakynnistystoiminnosta. Kynnist SIMATIC PDM, avaa valikko Device Quick Start (Laite Pikakytttoiminto) ja noudata ohjeita vaiheissa 15.

Huomautukset:

Pikakytttoiminto

Kytt FDT-kenttvyltykalun (Field Device Tool) avulla

FDT on useissa eri ohjelmistopaketeissa kytetty vakiomuotoinen tykalu, jolla otetaan kyttn ja yllpidetn kenttlaitteita. Markkinoilla on saatavissa kaksi kenttvyltykalua: PACTware ja Fieldcare. Jotta kenttlaite voidaan konfiguroida kenttvyltykalun avulla, tarvitaan laitteen DTMohjelma (Device Type Manager). Siemens-mittauslaitteissa kytetn SITRANS DTM ohjelmistoa ja SITRANS DTM -ohjelmistoa varten kirjoitettua mittauslaitteen EDDlaitemrityst. 1) Asenna ensin SITRANS DTM jrjestelmsi. Sen voi ladata osoitteesta http://support.automation.siemens.com. Napsauta Product Support (Tuotetuki) ja siirry kohtaan Product Information/Automation Technology/Sensor systems/Process Instrumentation/Software & Communications (Tuotetiedot/Automaatioteknologia/ Anturijrjestelmt/Prosessi-instrumentointi/Ohjelmistot ja tiedonsiirto). 2) Asenna SITRANS LR250 HART -laitemritys (EDD), joka on tarkoitettu SITRANS DTM ohjelmistolle. Voit ladata sen tuotesivulta sivustostamme osoitteesta www.siemens.com/ LR250. Siirry kohtaan Support > Software Downloads (Tuki / Ohjelmistolataukset).

Uuden laitteen konfigurointi FDT:n avulla


Sovellusopas voidaan ladata tuotesivulta sivustostamme kohdasta Support (Tuki). A5E31997170 SITRANS LR250 (HART) PIKAKYTTOPAS Sivu SU-21

Kytt AMS-laitehallintaohjelmiston avulla


Suomi
AMS Device Manager on kenttlaitteiden kyttnottoon ja yllpitmiseen tarkoitettu ohjelmistopaketti. Katso yksityiskohtaiset tiedot AMS Device Manager -ohjelmiston kytst kyttohjeista tai online-ohjeista. Voit lukea listietoja osoitteesta http://www.emersonprocess.com/AMS/.

Shkinen laitemritys (EDD)


SITRANS LR250 edellytt AMS Device Manager -ohjelmistoversiolle 9.0 suunnitellun EDDlaitemrityksen.

Uuden laitteen konfigurointi AMS Device Managerin avulla


1) Tarkista tuotesivulta verkkosivustostamme osoitteessa www.siemens.com/LR250, ett kytsssi on uusin EDD-tiedosto. Siirry kohtaan Support > Software Downloads (Tuki / Ohjelmistolataukset) ja lataa se tarvittaessa. Tallenna tiedostot tietokoneellesi ja pura pakattu tiedosto helppopsyiseen kansioon. Kynnist AMS Device Manager Add Device Type (Lis laitetyyppi), selaa puretun EDD-tiedoston kohdalle ja valitse se.

2)

Kynnist AMS Device Manager. Sovellusopas HART-laitteiden asetusten mrittmiseksi AMS Device Managerilla on ladattavissa Internet-sivustomme tuotesivulta kohdasta Support (Tuki).

Huolto
SITRANS LR250 ei normaaleissa kyttolosuhteissa vaadi huoltoa tai puhdistamista. Toimi seuraavasti, jos laitteen kytt riolosuhteissa edellytt sen puhdistamista: 1) 2) Valitse puhdistusaine, joka ei vahingoita antennia eik reagoi prosessinesteen kanssa. Poista laite kytst ja pyyhi antenni puhtaaksi kankaalla ja sopivalla puhdistusaineella.

Laitteen korjaaminen ja vastuuvapauslauseke


Katso listietoja takakannen sissivulta.

Antennin tai elektroniikan/kotelon vaihto


Jos antenni on vaihdettava vahingoittumisen tai vikaantumisen vuoksi, se voidaan vaihtaa uuteen tarvitsematta kalibroida laitetta uudelleen. Antennintyypin vaihto edellytt laitteen toimittamista Siemensin valtuuttamaan korjauspisteeseen tai valtuuttamalle korjaajalle. Jos elektroniikkaa tai kotelo on vaihdettava vahingoittumisen tai vikaantumisen vuoksi, varmista, ett kytt oikeaa antenniversiota. Muussa tapauksessa laite on toimitettava Siemensin valtuuttamalle korjaajalle kalibroitavaksi uudelleen.

Sivu SU-22

SITRANS LR250 (HART) PIKAKYTTOPAS

A5E31997170

Prosessin maksimilmptilakuvaaja
LR250:n laippasovitinversiot
Suomi
VAROITUS: Sislmptila ei saa ylitt 80 C-astetta (176 F)! Huomautukset: Kuvaaja on ainoastaan ohjeellinen, eik siin ole huomioitu kaikkia mahdollisia prosessiliitntmahdollisuuksia. (Se EI KOSKE SITRANS LR250:n asennuksia suoraan metallisen astian pintaan.) Kuvaajassa ei huomioida myskn suoran auringonpaisteen aiheuttamaa lmpvaikutusta. Parametri 3.2.1 Current Internal Temperature seuraa laitteen sisist lmptilaa.

Ympristn lmptila (oC)

Suurimmat prosessilmptilat suhteessa sallittuihin ympristn lmptiloihin

prosessin lmptila

Prosessilmptila (oC)

Silmukan teho
Virtalhdevaatimukset
Huomautus: Alla olevat kyrt koskevat erillislaitetta, joka on konfiguroitu Siemensksiohjelmointilaitteen avulla.

Silmukkavastus RL

Kyr 1: Yleiskytt, luonnostaan turvallinen, kipinimtn / energialtaan rajoitettu / sytyttmtn

SALLITTU KYTTALUE

Silmukkajnnite VL Huomautus: Kun kytetn HART-tiedonsiirtoa, vhimmisjnnite 220 Ohmin (RL) vastuksella on 16,3 V DC (tasavirta).

A5E31997170

SITRANS LR250 (HART) PIKAKYTTOPAS

Sivu SU-23

Silmukkavastus RL

Kyr 2: Tulenkestv, varmennettu rakenne, rjhdyssuojattu

Suomi

SALLITTU KYTTALUE

Silmukkajnnite VL

Huomautus: Kun kytetn HART-tiedonsiirtoa, vhimmisjnnite 220 Ohmin (RL) vastuksella on 20,94 V DC (tasavirta).

Kynnistyskyttytyminen
Laite vie kynnistyessn virtaa alle 3,6 mA. Ensimmisen mittauksen odotusaika on alle 50 sekuntia.

Sivu SU-24

SITRANS LR250 (HART) PIKAKYTTOPAS

A5E31997170

SITRANS LR250 (HART) Snabbstartsmanual


Denna manual ger en versikt ver de viktigaste egenskaperna och funktionerna fr SITRANS LR250 (HART 1)). Vi rder er uttryckligen att skaffa er den detaljerade versionen av manualen s att ni kan f ut det mesta av ert instrument. Den kompletta manualen kan laddas ner frn SITRANS LR250 produktsida p vr hemsida p: www.siemens.com/LR250. Den tryckta manualen kan anskaffas frn er lokala representant fr Siemens Milltronics. Frgor angende innehllet i denna manual kan sndas till: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Kanada, K9J 7B1 E-post: techpubs.smpi@siemens.com

Svenska

Copyright Siemens AG 2013. Med ensamrtt


Vi rder anvndare att skaffa sig auktoriserade inbundna handbcker, eller att konsultera av Siemens Milltronics Process Instruments framtagna och utgivna elektroniska versioner. Siemens Milltronics Process Instruments ansvarar inte fr innehllet i ofullstndiga eller kompletta kopior av inbundna eller elektroniska versioner.

Ansvarsbegrnsning
Trots att vi kontrollerat att innehllet i denna handbok verensstmmer med beskriven instrumentering, kan avvikelser frekomma. Drfr kan vi inte garantera full verensstmmelse. Innehllet i denna handbok revideras regelbundet och ndringar tas med i fljande versioner. Vi vlkomnar alla frslag till frbttringar. Tekniska data kan komma att ndras utan fregende varsel.

MILLTRONICS r ett registrerat varumrke fr Siemens Milltronics Process Instruments.

Teknisk support
Support ges dygnet om. Fr att hitta ditt lokala Siemens Automation-kontors adress, telefonnummer och faxnummer, g till: www.siemens.com/automation/partner: Klicka p fliken Contact (Kontakt), vlj Service, och klicka sedan p Service igen fr att komma till produktgruppen (+Automation Technology (Automationsteknik) > +Sensor Systems (Sensorsystem) > +Process Instrumentation (Processinstrumentering) > +Level Measurement (Nivmtning) > +Continuous (Kontinuerlig)). Vlj Radar. Vlj land och sedan Stad/Region. Vlj Technical Support (Teknisk support) under Service. Fr on-line teknisk support, g till: www.siemens.com/automation/support-request Fr in apparatnamnet (SITRANS LR250) eller ordernumret, och klicka sedan p Search (Sk), och vlj motsvarande produkttyp. Klicka p Next. Fr in ett nyckelord som beskriver din frga. Blddra igenom motsvarande dokumentation, eller klicka p Next fr att skicka en e-post med din frga till Siemens Tekniska Support-stab. Siemens IA(DT Technical Support Center (Siemens A&D Teknisk support-central): tfn. +49 (0)911 895 7222

1)

HARTr ett registrerat varumrke fr HART Communications Foundation.

A5E31997170

SITRANS LR250 (HART) SNABBSTARTSMANUAL

Sida SV-1

Riktlinjer fr skerhet

Varningsnotiser mste efterlevas fr att din personliga och andras skerhet skall kunna skerstllas samt skydda produkten och ansluten utrustning. Dessa varningsnotiser tfljs av anvisningar om den skerhetsniv som skall observeras. Denna VARNING-symbol hnvisar till en varningssymbol p produkten. Den informerar om att underltenhet att vidta erforderliga frsiktighetstgrder kan orsaka ddsfall, svr kroppsskada och/eller stor materiell skada. Denna VARNING-symbol som anvnds nr det inte finns ngon motsvarande varningssymbol p produkten, informerar om att underltenhet att vidta erforderliga frsiktighetstgrder kan orsaka ddsfall, svr kroppsskada och/ eller stor materiell skada. Anmrkning: innebr viktig information om produkten eller denna del av anvndarmanualen.

Svenska

FCC-verensstmmelse
Endast fr anlggningar i USA: Regler uppstllda av Federal Communications Commission (FCC)
VARNING: ndringar eller modifikationer som inte uttryckligen godknts av Siemens Milltronics kan upphva anvndarens rtt att bruka utrustningen. Anmrkningar: Denna utrustning har testats och befunnits klara de grnser som uppsatts fr digital utrustning av Klass A, i enlighet med Del 15 av FCC-reglerna. Dessa grnser har satts s att skligt skydd mot skadlig strning uppns nr utrustningen anvnds i en industriell milj. Denna utrustning genererar, anvnder och kan utstrla radiofrekvensenergi samt kan, om den inte installeras och anvnds i enlighet med anvndarmanualen, frorsaka besvrande strningar av radiokommunikationer. Om denna utrustning anvnds i bostadsomrden uppstr frmodligen besvrande strningar p radiokommunikationerna, och anvndaren kan d lggas att avhjlpa strningarna p egen bekostnad.

SITRANS LR250
VARNING: SITRANS LR250 skall endast anvndas p det stt som anvisas i denna manual; i annat fall kan det skydd, som utrustningen erbjuder, visa sig otillrckligt. Anmrkning: Denna produkt r avsedd att anvndas i industriomrden. Om denna utrustning anvnds i bostadsomrden kan den frorsaka strningar p mnga frekvensbaserade kommunikationer. SITRANS LR250 r en 2-trdars 25 GHz puls radarnivtransmitter fr kontinuerlig vervakning av vtskor och slurryn i frvaringstankar vid hgt tryck och hg temperatur, upp till 20 m (66ft). Den r idealisk fr sm tankar och lgdielektriska media. Anordningen bestr av en kapslad elektronisk krets kopplad till en hornantenn och antingen en gngad eller flnsad processanslutning. En gngad PVDF-antenn finns ocks tillgnglig. SITRANS LR250 stder HART:s kommunikationsprotokoll, och programvaran SIMATIC PDM. Signaler bearbetas med hjlp av Process Intelligence.

Sida SV-2

SITRANS LR250 (HART) SNABBSTARTSMANUAL

A5E31997170

Specifikationer

Fr fullstndig lista, var god se SITRANS LR250 (HART) Anvndarmanual. Fr information om godknnanden, se Godknnanden p sida 4.

Omgivnings- och drifttemperatur


Anmrkningar:
Maximal temperatur beror p processanslutningen, antennmaterialen och behllartrycket: se Tabell ver maximala processtemperaturer p sida 23. Fr nrmare information, se Processtryck/temperaturkurvor i den kompletta manualen. Processtemperatur och tryckkapacitet str angivna p anslutningsbrickan. Referensritningen angiven p brickan finns tillgnglig p produktsidan p vr webbplats www.siemens.com/LR250, under Support > Installation Drawings (Installationsritningar) > Level Measurement (Nivmtning) > Continuous (Kontinuerligt) - Radar. Ytterligare information om processanslutningar finns p Installation Drawings page (Installationsritningssidan) under Process Connection Diagrams (Processkopplingsscheman). Signalamplituden kar med horndiametern; anvnd drfr den strsta, praktiskt anvndbara storleken. Extra frlngningar kan installeras under gngorna. Se Tabell ver maximala processtemperaturer p sida 23, fr nrmare detaljer.
omgivande temperatur (inkapsling) 40 C till 80 C (40 F till 176 F) Processtemperatur vid processanslutning: standardhornantenn: - med FKM O-ring: 40 oC till +200 oC (-40 oF till +392 oF) - med FFKM O-ring: 20 oC till +200 oC (-4 oF till +392 oF) 2" NPT / BSPT / G Gngad PVDF-antenn: 40 till +80 C (40 till +176 F) apparatens mrkskylt

Svenska

processanslutningsbricka

Effektfrbrukning
Allmnt syfte Egensker Gnistfri/Energibegrnsad Icke-tndande (FM/CSA endast USA/Kanada) Flamsker kad skerhet Explosionssker (FM/CSA endast USA/Kanada) Maximalt 30 V DC 4 till 20 mA Max. startstrm: Mrkspnning 24 V DC vid 550 Ohm Mrkspnning 24 V DC vid 250 Ohm

se Startbeteende p sida 24

A5E31997170

SITRANS LR250 (HART) SNABBSTARTSMANUAL

Sida SV-3

Godknnanden
Anmrkning: Apparatens namnskylt listar de godknnanden som gller fr er apparat. Allmnt CSAUS/C, FM, CE, C-TICK, KC Radio: Europa (R&TTE), FCC, Industry Canada Riskfylld Egenskere 1) (Europa) ATEX II 1G, EEx ia IIC T4 ATEX II 1D, EEx tD A20 IP67 T90 C (Internationell) IECEx SIR 05.0031X, Ex ia IIC T4, Ex tD A20 IP67 T90 C (USA/Kanada) FM/CSA Klass I, Div. 1, Gr. A, B, C, D Klass II, Div. 1, Grupperna E, F, G Klass III T4 (Brasilien) INMETRO: DNV 12.0087 Ex ia IIC T4 Ga Ex ia ta IIIC T90 C Da IP65/IP67 -40 C Ta +80 C DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

Svenska

(Kina) Gnistfri/Begrnsad energi 2) (Europa) (Kina) Icke-tndande 3) (USA/Kanada) Flamsker 4) (Europa/Internationell)

NEPSI Ex ia IIC T4/ DIP A20 TA, T90 C IP67 ATEX II 3 G Ex nA/nL IIC T4 Gc NEPSI Ex nA II T4/ Ex nL IIC T4 FM/CSA Klass I, Div. 2, Gr. A, B, C, D T5 ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex dmbia IIC T4 Ga/Gb Ex tD A20 IP67 T90 oC INMETRO: DNV 12.0088 X Ex d ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017

(Brasilien)

ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-1:2009, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

1) 2) 3) 4)

Se Egensker kabeldragning p sida 9. Se Gnistfritt/Energibegrnsat kablage p sida 11. Se Icke-antndligt kablage p sida 11. Se Flamsker kabeldragning p sida 12.

Sida SV-4

SITRANS LR250 (HART) SNABBSTARTSMANUAL

A5E31997170

Farliga (forts.) Frhjd skerhet 1) (Europa/Internationell)

(Brasilien)

ATEX II 1/2 GD, 1D, 2D IECEx SIR 08.0107X Ex embia IIC T4 Ga/Gb Ex tD A20 IP67 T90 oC INMETRO: DNV 12.0088 X Ex e ia mb IIC T4 Ga/Gb Ex ia ta IIIC T90 C Da IP67 -40 C Ta +80 C Um = 250 V DNV #OCP 0017
ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-7:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 e IEC 61241-11:2005

Svenska

Flamsker/Frhjd skerhet (Kina) NEPSI Ex dmbia IIC T4/ Ex embia IIC T4/ DIP A20 TA, T90 oC IP67 Explosionssker 2) (USA/Kanada) FM/CSA Klass I, Div. 1, Grupperna A, B, C, D Klass II, Div. 1, Grupperna E, F, G Klass III T4

Marin Lloyd's Register ABS Typgodknnande

Anvndning under tryck


VARNINGAR:

Frsk inte att lossa, ta bort eller ta isr processanslutningen eller instrumenthljet medan innehllet i behllaren str under tryck. Anvndaren r ansvarig fr val av material i skruvfrband och packningar som skall hlla sig inom flnsens grnsvrden och passa fr dess avsedda anvndning och rdande driftfrhllanden. Felaktig installation kan orsaka bortfall av processtryck.
Anmrkningar: Processanslutningsbrickan skall sitta kvar p tryckgrnsenheten2). Om apparatpaketet byts, skall processanslutningsbrickan flyttas ver till den nya enheten. SITRANS LR250-enheter r hydrostatiskt testade, och uppfyller eller verskrider alla krav i pann- och tryckkrlslagen ASME Boiler and Pressure Vessel Code och det Europeiska direktivet fr tryckkrl.

1) 2)

Se Frhjd skerhetskabeldragning p sida 12. Se Icke-tndande kabeldragning (endast USA/Kanada) p sida 13.

A5E31997170

SITRANS LR250 (HART) SNABBSTARTSMANUAL

Sida SV-5

Tryckutrustningsdirektiv, PED, 97/23/EC 1)


Siemens nivtransmittrar med flnsade eller gngade infstningar eller infstningar av sanitr klmtyp, har inga egna tryckhllande hljen och klassas drfr inte av Tryckutrustningsdirektivet som tryck- eller skerhetstillbehr, (se EU-kommissionen riktlinje 1/8).

Installation
Varningar: Installation fr endast utfras av kompetent personal och enligt gllande lokala bestmmelser. Greppa apparaten i hljet, inte processanslutningsbrickan, fr att undvika skador. Var extra frsiktig vid hantering av den gngade PVDF-antennen. Varje skada p antennytan, i synnerhet p spetsen, kan pverka prestandan. Konstruktionsmaterial vljs p basis av dess kemiska kompatibilitet (eller lga reaktionsbengenhet) fr allmnna ndaml. Fr exponering mot specifika miljer, kontrollera mot kemiska kompatibilitetstabeller fre installation. Anmrkningar:

Svenska

Fr lnder som r medlemmar i den Europeiska unionen mste installation ske enligt ETSI EN 302372. Se godknnandeinformation p apparatens mrkskylt. Tillverkningsnumret som str stmplat i varje processanslutningskropp r ett unikt identifikationsnummer som anger tillverkningsdatumet. Exempel: MMDDYY XXX (dr MM = mnad, DD = dag, YY = r, och XXX= producerad sekvensenhet) Ytterligare mrkningar (nr utrymme finns) anger flnskonfiguration, storlek, tryckklass, material, och materialvrmekod.

Munstycksutformning
Gngad PVDF-antenn
min. frigng: 10 mm (0,4")

Rostfri hornantenn
min. frigng: 10 mm (0,4")

Min. dia. 51 mm (2")

nden p antennen mste sticka ut minst 10 mm (0,4") fr att undvika att falska ekon reflekteras frn munstycket. Minsta rekommenderade munstycksdiameter fr den gngade PVDF-antennen r 51 mm (2"). En antennfrlngning (100 mm/ 3,93") finns tillgnglig fr varje version med undantag av den gngade PVDF-antennen.

1)

Tryckgrnsenheten innehller komponenter som fungerar som en barrir mot tryckfrluster frn processkrlet: d.v.s. kombinationen av processanslutningskropp och sndare, men normalt exklusive elektrisk inkapsling.

Sida SV-6

SITRANS LR250 (HART) SNABBSTARTSMANUAL

A5E31997170

Munstycksplacering
Undvik centrala placeringar p hga, smala tankar som kan generera falska ekon.

Milj
Ombesrj en milj som lmpar sig fr hljets skyddsklass och konstruktionsmaterialen. Stt in en solskrm om apparaten skall monteras i direkt solljus.
fredragen olmplig

Strlningsvinkel
Strlvinkel r bredden p konan dr energittheten r hlften av toppenergittheten. Toppenergidensiteten finns direkt framfr och i linje med antennen. Det avges en signal utanfr strlvinkeln, vilket innebr att falska ml kan detekteras.

Svenska

emissionskona strlningsvinkel: gngad PVDF antenn = 19 1,5" horn = 19 2" horn = 15 3" horn = 10 4" horn = 8

Emissionskona
Hll emissionskonan fri frn strningar frn stegar, rr, I-balkar eller fyllningsstrmmar.

strlningsvinkel

Tilltrde fr programmering
Se till att displayen r ltt att se och programmera med den handhllna programmeringsenheten.

Monteringsanvisningar
VARNING: Fr trycktillmpningar mste man anvnda PTFE-tejp eller annat lmpligt gngttningsmedel, och dra t processanslutningen s mycket att det inte gr att vrida ur anslutningen fr hand. Det maximalt rekommenderade vridmomentet r 40 N-m. (30 ft-pund). Anmrkning: Det finns ingen grns fr hur mnga gnger ett instrument kan roteras utan att skadas.

Gngade versioner
1) 2) Innan du stter in apparaten i dess fattning, kontrollera att gngorna passar ihop s att de inte skadas. Skruva bara i enheten i processanslutningen, och dra t fr hand, eller anvnd en skruvnyckel. Fr trycktillmpningar se Varning ovan. VARNING: Anvndaren r ansvarig fr val av material i skruvfrband och packningar som skall hlla sig inom flnsens grnsvrden och passa fr dess avsedda anvndning och rdande driftfrhllanden.

Flnsade versioner

A5E31997170

SITRANS LR250 (HART) SNABBSTARTSMANUAL

Sida SV-7

Kabeldragning
Effekt
Varningar: DC-ingngarna skall komma frn en klla som ger elektrisk isolering mellan ingng och utgng, fr att uppfylla tillmpliga skerhetskrav i IEC 61010-1. Allt fltkablage mste ha isolering lmpad fr mrkspnningarna.

Anslutning av SITRANS LR250


VARNINGAR: Kontrollera p apparatens mrkskylt och processbricka, att vrdena r riktiga. Anvnd lmpliga ledningsttningar fr att upprtthlla IP- eller NEMA-klassning. Se Kabeldragningar fr installationer i farligt omrde p sida 9 Anmrkningar: Anvnd partvinnad kabel: AWG 22 till 14 (0,34 mm2 till 2,5 mm2). Separata kablar och ledningar kan krvas fr att uppfylla standardiserad kabeldragningspraxis, eller elbestmmelser.
anvnd en 2 mm insexnyckel fr att lossa locklsets stllskruv. plugg (IP 68) alternativt kabelpackbox1) 2) (eller NPT-kabelingng2)) kabelskrmning (vid frekomst)

Svenska

1) 2) 3) 4) 5)

Strippa kabelisoleringen ver ungefr 70 mm (2.75") frn nden p kabeln och tr ledningarna genom packboxen. 1) 2) Anslut ledningarna till plinten s som visas: polariteten finns angiven p plinten. Jorda apparaten i enlighet med lokala bestmmelser. Dra t packboxen s att den ttar ordentligt. Stng locket innan enheten programmeras och konfigureras.

1) 2)

Fr transporteras med anordningen. Om kabel dras genom skyddsrr, anvnd endast godknda lmpligt dimensionerade nav fr vattentta tillmpningar.

Sida SV-8

SITRANS LR250 (HART) SNABBSTARTSMANUAL

A5E31997170

Anslutning av HART
Typisk PLC/mA-konfiguration med HART 1) 2)
aktiv PLC strmfrsrjning R= 250 HARTmodem SITRANS LR250 HARTkommunikator

Anmrkningar: Beroende p systemets utformning, kan strmfrsrjningen vara skild frn PLC:n, eller integrerad i den. HART-motstnd1) mste vara begrnsat till det tilltna driftomrdet2). Anslut inte mer n en HARTprogrammeringsenhet t gngen.

Svenska

Kabeldragningar fr installationer i farligt omrde


Det finns sex kabeldragningsalternativ fr installationer i farligt omrde. Kontrollera alltid p apparatens mrkskylt och processanslutningsbricka, att vrdena r riktiga.

1. Egensker kabeldragning
Mrkskylt (ATEX/IECEX/INMETRO/C-TICK)

ATEX-intyg listat p mrkskylten kan laddas ner frn produktsidan p vr webbplats: www.siemens.com/LR250. G till Support > Approvals / Certificates (Godknnanden / Intyg). IECEx-intyg listat p mrkskylten kan ses p IECEx-webbsidan. G till: http://iecex.iec.ch och klicka p Ex Equipment Certificates of Conformity/ verensstmmelseintyg fr utrustning och fr sedan in intygsnumret IECEx SIR 05.0031X.

1) 2)

Total kretsresistans = kabelresistans plus 250 Ohm (resistor). Se antingen Kurva 1: Allmnt syfte, Egensker, Gnistfri/Energibegrnsad/ Icke antndlig p sida 23 eller Kurva 2: Flamsker, Frhjd skerhet, Explosionssker p sida 24.

A5E31997170

SITRANS LR250 (HART) SNABBSTARTSMANUAL

Sida SV-9

Egensker kabeldragning (forts.)


Mrkskylt (FM/CSA)

FM/CSA Egensker anslutningsritning nummer 23650653 kan laddas ner frn produktsidan p vr webbsida: www.siemens.com/LR250. G till Support > Installation Drawings (Installationsritningar) > Level Measurement (Nivmtning) > Continuous (Kontinuerligt) - Radar. Fr effektbehov se Kurva 1: Allmnt syfte, Egensker, Gnistfri/Energibegrnsad/ Icke antndlig p sida 23. Fr kabeldragningskrav: flj lokala bestmmelser. Godknda dammtta och vattentta kanalttningar krvs fr utomhus NEMA 4X / typ 4X / NEMA 6, IP67, IP68 platser. Se Instruktioner gllande installationer i riskzoner p sida 13. Rekommenderade egenskra skerhetsbarrirer listas under Diodbarrirer med passiv shunt p sida 10 och Aktiva barrirer (upprepande barrirer) p sida 11. Anmrkning: Fr att kunna vlja en PLC-inmatningsmodul, strmfrsrjning, eller barrir krvs kunskap om egenskerhet och tillmpningen. Det ligger installatren att skerstlla att den egenskra installationen uppfyller bde apparatens godknnandekrav och gllande nationell praxis.

Svenska

Diodbarrirer med passiv shunt


Anmrkning: Det krvs en vlreglerad matarspnning. Fabrikat
MTL MTL Stahl Stahl

Artikelnummer
787SP+ (Dubbelkanal) 7787P+ (Dubbelkanal) 9001/01-280-100-10 (Enkelkanal) 9002/01-280-110-10 (Dubbelkanal)

Fr att skerstlla att barrirens skerhetsbeskrivning lmpar sig fr ingngsparametrarna i LR250 Egensker, skall fljande berkningar utfras: Re-e = max. nde-till-nde-motstnd i barriren Rkrets = kretsresistans (totalt kabelresistans plus varje extra serieresistans, t.ex. PLC-ingngar och/eller displayer) Vbarrir = vrde p varje icke-linjrt spnningsfall p grund av barriren 1) Bestm vrdet fr Re-e frn databladet. 2) Berkna det totala vrdet fr Rkrets. 3) Berkna Rarbetande = Re-e + Rkrets.

Hur man vljer en passiv barrir fr SITRANS LR250

Sida SV-10

SITRANS LR250 (HART) SNABBSTARTSMANUAL

A5E31997170

Hur man vljer en passiv barrir (forts.)


4) 5) 6) Bestm vrdet p Vbarrir frn barrirdatabladet (t.ex. spnningsfall p grund av dioder). Berkna Varbetande = Vfrsrjning Vbarrir. Anvnd vrdena fr Varbetande Rarbetande fr att bekrfta att operationen ligger inom det skuggade omrdet i diagrammet Kurva 1: Allmnt syfte, Egensker, Gnistfri/ Energibegrnsad/ Icke antndlig p sida 23 Artikelnummer
706 7206 SITRANS I 7NG4124 9001/51-280-110-14 E02009 E02010

Aktiva barrirer (upprepande barrirer)


Tillverkare
MTL MTL Siemens Stahl MTL MTL

Svenska

2. Gnistfritt/Energibegrnsat kablage

ATEX-intyget listat p mrkskylten kan laddas ner frn produktsidan p vr webbsida p: www.siemens.com/LR250. G till Support > Approvals / Certificates (Godknnanden / Intyg). Fr effektbehov, se Kurva 1: Allmnt syfte, Egensker, Gnistfri/Energibegrnsad/ Icke antndlig p sida 23. Fr kabeldragningskrav, flj lokala bestmmelser.

3. Icke-antndligt kablage

FM/CSA Klass 1, Div 2 anslutningsritning nummer 23650673 kan laddas ner frn produktsidan p vr webbsida p: www.siemens.com/LR250. G till Support > Installation Drawings (Installationsritningar) > Level Measurement (Nivmtning) > Continuous (Kontinuerligt) - Radar. Fr effektbehov se Kurva 1: Allmnt syfte, Egensker, Gnistfri/Energibegrnsad/ Icke antndlig p sida 23.

A5E31997170

SITRANS LR250 (HART) SNABBSTARTSMANUAL

Sida SV-11

4. Flamsker kabeldragning

ATEX-intyget listat p mrkskylten kan laddas ner frn produktsidan p vr webbsida p: www.siemens.com/LR250. G till: Support > Approvals / Certificates (Godknnanden / Intyg). IECEx-intyg listat p mrkskylten kan se p IECEx-webbsidan. G till: http://iecex.iec.ch och klicka p Ex Equipment Certificates of Conformity/ verensstmmelseintyg fr utrustning och fr sedan in intygsnumret IECEx SIR 08.0107X.

Svenska

Fr effektbehov se Kurva 2: Flamsker, Frhjd skerhet, Explosionssker p sida 24. Fr kabeldragningskrav, flj lokala bestmmelser. Se ven Instruktioner gllande installationer i riskzoner p sida 13 och ATEX-intyg listat ovan.

5. Frhjd skerhetskabeldragning

ATEX-intyget listat p mrkskylten kan laddas ner frn produktsidan p vr webbsida p: www.siemens.com/LR250. G till: Support > Approvals / Certificates (Godknnanden / Intyg). IECEx-intyg listat p mrkskylten kan se p IECEx-webbsidan. G till: http://iecex.iec.ch och klicka p Ex Equipment Certificates of Conformity/ verensstmmelseintyg fr utrustning och fr sedan in intygsnumret IECEx SIR 08.0107X. Fr effektbehov se Kurva 2: Flamsker, Frhjd skerhet, Explosionssker p sida 24. Fr kabeldragningskrav, flj lokala bestmmelser. Se ven Instruktioner gllande installationer i riskzoner p sida 13 och ATEX-intyg listat ovan.

Sida SV-12

SITRANS LR250 (HART) SNABBSTARTSMANUAL

A5E31997170

6. Icke-tndande kabeldragning (endast USA/Kanada)

FM/CSA Explosionssker anslutningsritning nummer A5E02257843 kan laddas ner frn produktsidan p vr webbsida: www.siemens.com/LR250. G till Support > Installation Drawings (Installationsritningar) > Level Measurement (Nivmtning) > Continuous (Kontinuerligt) - Radar. Fr effektbehov se Kurva 2: Flamsker, Frhjd skerhet, Explosionssker p sida 24.

Svenska

Instruktioner gllande installationer i riskzoner


(Referens Europeiska ATEX-direktivet 94/9/EC, Bilaga II, 1/0/6)
Fljande instruktioner gller fr utrustning som tcks av intygsnummer SIRA 06ATEX2358X och SIRA 08ATEX1301X, och SIRA 09ATEX4153X. Fr anvndning och montering, se huvudinstruktionerna. Utrustningen r certifierad fr anvndning som Kategori 1GD utrustning enligt SIRA 06ATEX2358X; Kategori 1/2 GD, 1D, 2D utrustning enligt SIRA 08ATEX1301X; och Kategori 3G utrustning enligt SIRA 09ATEX4153X. 3) Utrustningen kan anvndas med antndbara gaser och ngor med apparatgrupp IIC, IIB och IIA och temperaturklasser T1, T2, T3 och T4. 4) Utrustningen har ett intrngningsskydd motsvarande IP67 och en temperaturklass p T90 C och kan anvndas med antndbart damm. 5) Utrustningen r certifierad fr anvndning i omgivningstemperaturer frn 40 C till +80 C. 6) Utrustningen har inte bedmts som en skerhetsrelaterad utrustning (enligt meningen i direktiv 94/9/EC Bilaga II, klausul 1.5). se Tryckutrustningsdirektiv, PED, 97/23/EC p sida 6. 7) Installation och inspektion av denna utrustning skall utfras av lmpligt utbildad personal i enlighet med tillmpliga delar av arbetspraxis (EN 60079-14 och EN 60079-17 i Europa). 8) Utrustningen kan inte repareras. 9) Intygsnumren har ett X-suffix, som anger att speciella villkor fr sker anvndning gller. De som installerar eller inspekterar denna utrustning mste ha tillgng till intygen. 10) Om utrustningen riskerar komma i kontakt med frtande mnen ligger det anvndaren att vidta lmpliga tgrder fr att frhindra att utrustningen pverkas negativt och p s vis skerstlla att skyddsgraden rcker till. Frtande mnen: t.ex. sura vtskor eller gaser som kan angripa metaller, eller lsningsmedel som kan pverka polymermaterial. Lmpliga frsiktighetstgrder: t.ex. att med ledning av materialets datablad kontrollera att det r bestndigt mot vissa kemikalier. 1) 2)

A5E31997170

SITRANS LR250 (HART) SNABBSTARTSMANUAL

Sida SV-13

Programmering SITRANS LR250


Aktivering av SITRANS LR250

Se Vgledning fr snabbstart via den handhllna programmeringsenheten p sida 18 Se Vgledning fr snabbstart via SIMATIC PDM p sida 21.

Inkoppling av apparaten p elntet. SITRANS LR250 startar automatiskt i Mtningslget.

LCD-displayen Mtningsvisning (normal drift) 1) 2)


1 2 3 4

Svenska

1 vippindikator1) fr linjra enheter eller % 2 vald funktion: level (niv), space (rymd), eller distance (avstnd) 3 uppmtt vrde (niv eller volym, rymd eller avstnd) 4 enheter 5 stapeldiagram indikerar niv 6 sekundrt omrde indikerar p begran2) elektroniktemperatur, ekotillfrlitlighet, kretsstrm eller avstnd 7 textyta visar statusmeddelanden 8 apparatstatusindikator (fr detaljer, se den kompletta manualen)

Frekommande-fel-indikatorer S: 0 LOE
Nr ett fel freligger, visas ett felmeddelande i textzonen (7), och en service-krvs-ikon visas p apparatstatusplatsen (8)

Visning Programlge Menyruta Navigation


aktuell positionsnummer En synlig menystapel indikerar att meny menylistan r fr lng fr att kunna aktuell position menystapel visa alla positioner. positionsband Ett band halvvgs ner i menystapeln indikerar att den aktuella positionen r halvvgs ner i listan. Positionsbandets djup och relativa lge i menystapeln indikerar menylistans lngd, och den aktuella positionens lge i listan. Ett djupare band indikerar frre positioner. aktuellt

Menyruta Parametrar
parameternummer parameternamn parametervrde/val

Menyruta Redigering

1) 2)

Tryck in UP(P) eller DOWN/NED pil fr att vxla. Som svar p tangenttryckning. Fr detaljer, se tabell Funktion i mtlge p sida 15.

Sida SV-14

SITRANS LR250 (HART) SNABBSTARTSMANUAL

A5E31997170

Handhllen programmeringsenhet (Art.nr 7ML1930-1BK)


Bestlls separat. 1) 2) 3)
Tangent Funktion i mtlge Uppdaterar kretsstrmmen1). Uppdaterar inre kapslings temperaturavlsning1). Uppdaterar eko konfidensvrde1).

C
Uppdaterar distansmtning1).

Svenska

Lge ppnar PROGRAM-lge2). HGER pil ppnar PROGRAM-lge3).

UPP- eller NER-pil vxlar mellan linjra enheter och %.

Programmering via den handhllna programmeringsenheten


Anmrkningar: Tryck in Mode (Lge) fr att vxla mellan Mtnings- och Programlge. Medan apparaten str i PROGRAM-lge ndras inte utgngen och svarar inte p ndringar i apparaten. SITRANS LR250 tergr automatiskt till Mtningslge efter en inaktivitetsperiod i PROGRAM-lge (mellan 15 sekunder och 10 minuter, beroende p menyniv).

Parametermenyer

Parametrar identifieras med namn och organiseras i funktionsgrupper. Fr den kompletta parameterlistan med instruktioner, se den kompletta manualen.

1. QUICK START 2. SETUP 2.1. DEVICE ............. 2.7. LINEARIZATION 2.7.1. VOLUME 2.7.1.1. VESSEL SHAPE

1) 2)

3)

Nytt vrde visas i LCD:ns sekundra omrde. ppnar den senast visade menynivn i denna effektcykel, om inte strm har cirkulerat efter utgng ur PROGRAM-lge eller mer n 10 minuter har gtt sedan PROGRAM-lge anvndes. Sedan visas toppnivmenyn. ppnar toppnivmenyn

A5E31997170

SITRANS LR250 (HART) SNABBSTARTSMANUAL

Sida SV-15

1.

G in i PROGRAM-lge
Rikta programmeringsenheten mot displayen (frn ett maximalt avstnd p 300 mm [1 fot]). HGER pil aktiverar PROGRAM-lge och ppnar menyniv 1. Lge ppnar den menyniv som senast visades i PROGRAM-lget inom de senaste 10 minuterna, eller menyniv 1 om strm har cirkulerat sedan dess.

handhllen programmeringsenhet display

max. 300 mm (1 fot)

2. Svenska

Navigering: knappfunktioner i arbetslge Navigation

Anmrkning: Fr Snabb tkomst av parametrarna via den handhllna programmeringsenheten, tryck in Home och fr sedan in menynumret, t.ex.: 2.7.1 (Volym).

Tangent

Namn
UPP eller NER-pil HGER pil VNSTER pil
Lge Hem

Menyniv Funktion i Navigeringslge


meny eller parameter meny parameter meny eller parameter meny eller parameter meny eller parameter Rulla till fregende eller nsta meny eller parameter. G till frsta parametern i den valda menyn, eller ppna nsta meny. ppna lge Edit (Redigering). ppna modermeny. Byt till MTNINGSLGE. ppna den hgsta menynivn: meny 1.

3.

Redigering i PROGRAM-lge Val av ett listat alternativ


a) b) Navigera till den nskade parametern. Tryck p HGER pil parametermeny. fr att ppna

parameternamn

parameternummer

aktuellt val

c)

Tryck p HGER pil igen fr att ppna Redigeringslge . Det aktuella valet r upplyst. Rulla till ett nytt val. Tryck in HGER pil fr att acceptera det. LCD:n tergr till menyruta parametrar och visar det nya valet.

d)

Sida SV-16

SITRANS LR250 (HART) SNABBSTARTSMANUAL

A5E31997170

ndring av ett numeriskt vrde


a) b) Navigera till den nskade parametern. Tryck p HGER pil igen fr att ppna Redigeringslge. Det aktuella vrdet r upplyst. Skriv in ett nytt vrde. Tryck in HGER pil fr att acceptera det. LCD:n tergr till menyruta parametrar och visar det nya valet.

parameternamn

parameternummer

c) d)

aktuellt vrde

Svenska

Tangentfunktioner i Redigeringslge
Tangent Namn UPP eller NER-pil Funktion i redigeringslge Val av alternativ Numerisk redigering Val av alternativ HGER pil Rullar till position. - Inkrementerar eller dekrementerar siffror. - Vxlar plus- och minustecken. - Accepterar data (skriver parameter). - ndrar frn Redigerings- till Navigerings-lge

- Flyttar markr ett steg t hger Numerisk redigering - eller med markr p Enter-tecknet, accepterar data och vxlar frn Redigerings- till Navigerings-lge Val av alternativ Upphver Redigerings-lge utan att ndra parametern - Flyttar markren till plus/minus-tecken om det r den frst intryckta tangenten - eller flyttar markren ett steg t vnster.

VNSTER pil

Numerisk redigering

Tm

Numerisk redigering Raderar visningen.

Decimalpunkt Numerisk redigering Fr in en decimalpunkt. Plus eller Numerisk redigering Byter tecknet p det infrda vrdet. minus tecken

till

Numerisk

Numerisk redigering Fr in motsvarande tecken.

A5E31997170

SITRANS LR250 (HART) SNABBSTARTSMANUAL

Sida SV-17

Vgledning fr snabbstart via den handhllna programmeringsenheten


1. Snabbstart
a) Rikta programmeringsenheten mot displayen frn ett maximalt avstnd p 300 mm [1 fot.], och tryck sedan p HGER-pil ppna menyniv 1. b) Tryck in HGER-pil parameterruta. c) Tryck in HGER pil fr att ppna Redigeringslge eller NER-pil fr att tv gnger fr att navigera till menyposition 1.1 och ppna fr att aktivera PROGRAM-lge och

Svenska

acceptera standardvrden och g direkt till nsta position. d) Fr att ndra en instllning, rulla till den nskade positionen eller knappa in ett nytt vrde. e) Efter ndring av ett vrde, tryck in HGER pil NER-pil f) fr att g till nsta position. fr att acceptera det och tryck in

Snabbstartsinstllningar blir aktiva frst efter att du valt Yes (Ja) p Apply changes

(Tillmpa ndringar) i steg (1.8.).

1.1. Sprk Vlj det sprk som skall anvndas p LCD:n och som trder i kraft omedelbart.
Alternativ
ENGLISH (ENGELSKA) , DEUTSCH (TYSKA), FRANAIS (FRANSKA), ESPAOL

(SPANSKA)

1.2. Material Alternativ


LIQUID (VTSKA) LIQUID LOW DK (lg dielektrikumvtska)

1.3. Svarshastighet Stller in enhetens reaktionshastighet p mtvrdesndringar i mlomrdet.


Svarshastighet
SLOW (LNGSAM) MED (MEDEL) FAST (SNABB)

Fyllnings-/tmningshastighet
0,1 m/min (0,32 fot/min) 1,0 m / min (3,28 fot/min) 10,0 m / min (32,8 fot/min)

Anvnd en instllning just ver den maximala fyllnings- eller tmningshastigheten (den som r strst av dem).

1.4. Enheter Sensormtenheter. Alternativ


M, CM, MM, FT, IN

Sida SV-18

SITRANS LR250 (HART) SNABBSTARTSMANUAL

A5E31997170

1.5. Funktionsstt
sensornollpunkt vre kal.punkt

avstnd sensornollpunkt volym

sensornollpunkt

niv undre kal. punkt

Svenska

Funktionsstt
NO SERVICE (INGEN DRIFT) LEVEL (NIV) SPACE (VOLYM) DISTANCE (AVSTND)
1)

Beskrivning
Mtning och tillhrande kretsstrm uppdateras inte, och apparaten begrnsar sig till Felsker funktion1). Avstnd frn Undre kalibreringspunkt till materialyta Avstnd frn vre kalibreringspunkt till materialyta Avstnd till materialyta rknat frn Sensorns nollpunkt.

Fr nrmare detaljer om Felsker funktion, se den kompletta manualen.

1.6. Undre kalibreringspunkt Avstnd frn Sensornollpunkt till Undre kalibreringspunkt: vanligen process tom niv. Se Funktionsstt (1.5.) fr en illustration. Vrden
Omrde: 0,00 till 20,00 m

1.7. vre kalibreringspunkt Avstnd frn Sensornollpunkt till vre kalibreringspunkt: vanligen process full niv. Se Funktionsstt (1.5.) fr en illustration. Vrden
Omrde: 0,00 till 20,00 m

1.8. Apply? (Tillmpa ndringar) Fr att spara Snabbstartsinstllningarna mste man vlja Ja p tillmpa ndringar. Alternativ
Tryck p Mode YES (JA), NO (NEJ), DONE (GJORT) (Displayen visar DONE (GJORT) nr Snabbstarten r framgngsrikt genomfrd. fr att terg till Mtlge. SITRANS LR250 r nu klar fr anvndning.

Anmrkning: Om er tillmpning har en tank med hinder, var vnlig studera detaljer i den kompletta manualen betrffande anvndning av Auto undvikande av falska ekon.

A5E31997170

SITRANS LR250 (HART) SNABBSTARTSMANUAL

Sida SV-19

SITRANS LR250 Communications: HART


Du behver den detaljerade manualen fr att f en lista p tillmpliga parametrar. Vi rekommenderar att du anvnder processinstrumenthanteraren SIMATIC Process Device Manager (PDM) fr att programmera instrumentet. Applikationsvgledningar fr installation av HART-apparater med SIMATIC PDM kan laddas ner frn produktsidan p vr webbplats: www.siemens.com/LR250 under Support.

SIMATIC PDM
SIMATIC PDM r ett programvarupaket som anvnds fr att kra igng och upprtthlla funktionen hos SITRANS LR250 och andra processapparater. Var god ls anvndarmanualen eller online-hjlp fr detaljer i anvndandet av SIMATIC PDM. (Du finner mer information p www.siemens.com/simatic-pdm.)

Avaktivera buffertar
Svenska

Detta anpassar SIMATIC PDM till HART-modemet fr Windows 2000 och Windows XP 1) operativsystem. Anmrkningar: SIMATIC PDM fungerar endast med Windows XP Professional-versionen, inte med Home-versionen. Du behver administrativa rttigheter fr ert operativsystem fr att avaktivera buffertar. 1) 2) 3) 4) 5) 6) Klicka p Start-/Instllnings-/Styrpanel fr att brja konfigurera. Dubbelklicka System, vlj Maskinvara tabulera och klicka p Device Manager -knappen. ppna mappen Portar och dubbelklicka p den COM-port som anvnds av systemet fr att ppna fnstret Kommunikationsport Egenskaper. Vlj Portinstllningar tabulera och dubbelklicka p Avancerat-knappen. Om Anvnd FIFO-buffertar radiorutan r vald, klicka fr att vlja bort den. Klicka p OK fr att g ur. Stng alla skrmbilder och boota om.

Uppdatera Electronic Device Description (EDD) (Beskrivning Elektroniska Instrument)


Anmrkningar:
SIMATIC PDM V 5.2 SP1 understds endast fr baskonfiguration och felavhjlpning. Fr avancerade funktioner ssom Snabbstartsvgledning, krvs version 6.0 SP4 eller hgre. I SIMATIC PDM terfinns EDD i Instrumentkatalogen Device Catalog, under Sensors/ Level/Echo/Siemens Milltronics/SITRANS LR250. Kontrollera produktsidan p vr webbplats: www.siemens.com/LR250, fr att skerstlla att du har den senaste versionen av SIMATIC PDM, senaste Service Pack (SP) och senaste hot fix (HF). G till Support > Software (Programvara) Downloads (Nedladdningar).

Fr att installera en ny EDD:


1) 2) 3) Ladda ner EDD:n frn produktsidan p vr webbsida p: www.siemens.com/LR250 och spara filerna p din dator. Packa upp den zippade filen p en ltt tkomlig plats. Lansera SIMATIC PDM Manager Device Catalog (Instrumentkatalog), blddra till den uppackade EDD-filen och vlj den.

1)

Windows r ett registrerat varumrke fr Microsoft Corporation

Sida SV-20

SITRANS LR250 (HART) SNABBSTARTSMANUAL

A5E31997170

Konfigurering av en ny apparat
Anmrkning: Klickning p Cancel (Upphv) under en uppladdning frn instrumentet till SIMATIC PDM resulterar i att vissa parametrar uppdateras.
1) 2) 3) 4) 5) Kontrollera att du har den senaste EDD:n, och uppdatera den vid behov (seUppdatera Electronic Device Description (EDD) (Beskrivning Elektroniska Instrument) p sida 20). Starta SIMATIC Manager och skapa ett nytt projekt fr LR250. Applikationsvgledningar fr installation av HART-apparater med SIMATIC PDM kan laddas ner frn produktsidan p vr webbplats: www.siemens.com/LR250. ppna menyn Device Master Reset (Instrument Masterterstllning) och klicka p OK fr att utfra en terstllning till Fabriksstandardvrden. Efter att terstllningen fullbordats, ladda upp parametrar till PC:n/PG:n. Konfigurera apparaten med hjlp av Snabbstartsvgledningen.

Vgledning fr snabbstart via SIMATIC PDM


Snabbstartsvgledningens instllningar r internt relaterade och ndringar trder i kraft frst efter att du klickat p FINISH AND DOWNLOAD (AVSLUTA OCH NEDLADDNING) i slutet av steg 5 fr att spara instllningar offline och sedan verfra dem till apparaten. Klicka p BACK (BAKT) fr att g tillbaka och gra om en instllning eller Cancel (Upphv) fr att g ur Snabbstarten. Starta SIMATIC PDM, ppna menyn Device Wizard - Quick Start (Apparat Snabbstart), och flj stegen 1 till 5.

Svenska

Anmrkningar:

Snabbstart

Funktion via FDT (Field Device Tool)


FDT r en standard som anvnds i mnga programvarupaket avsedda fr att kra igng och istndhlla fltapparatur. Tv kommersiellt tillgngliga FDT:n r PACTware och Fieldcare. Fr att konfigurera en fltapparat via FDT behver du DTM (Device Type Manager) fr enheten. Siemens-instrument anvnder SITRANS DTM och ett EDD-instrument skrivet fr SITRANS DTM. 1) Installera SITRANS DTM i ert system. Du kan ladda ner det frn: http://support.automation.siemens.com. Klicka p Product Support (Produktsupport) och navigera till Product Information/Automation Technology/Sensor systems/Process Instrumentation/Software & Communications (Produktinformation/ Automatiseringsteknik/Sensorsystem/Processinstrumentering/Programvara & Kommunikationer). 2) Installera SITRANS LR250 HART EDD fr SITRANS DTM. Du kan ladda ner den frn produktsidan p vr hemsida p: www.siemens.com/LR250. G till Support > Software Downloads (Nedladdningar av programvara).

Konfigurering av en ny apparat via FDT


En Applikationsvgledning kan laddas ner frn produktsidan p vr webbsida under Support. A5E31997170 SITRANS LR250 (HART) SNABBSTARTSMANUAL Sida SV-21

Funktion via AMS Device Manager


AMS Device Manager r ett programvarupaket utformat fr igngkrning och istndhllning av fltapparater. Var god ls anvndarmanualen eller online-hjlp fr detaljer rrande anvndningen av AMS Device Manager. Mer information ges p: http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD) (Beskrivning Elektroniska Instrument)


SITRANS LR250 krver EDD fr AMS Device Manager version 9.0.

Konfiguration av en ny apparat via AMS Device Manager


1) Kontrollera produktsidan p vr webbplats: www.siemens.com/LR250 fr att skerstlla att du har den senaste EDD:n. G till Support > Software Downloads (Nedladdning av programvara) och ladda ner den om ndvndigt. Spara filerna p din dator, och extrahera den zippade filen till en ltttkomlig plats. Hmta in AMS Device Manager Add Device Type (AMS Apparatanvisning Lgg till apparattyp), blddra till den uppackade EDD-filen och vlj den.

Svenska

2)

Hmta apparatanvisningen AMS Device Manager. En Applikationsvgledning fr installation av HART-apparater med AMS Device Manager kan laddas ner frn produktsidan p vr webbsida under Support.

Underhll
SITRANS LR250 krver vare sig underhll eller rengring under normala driftfrhllanden. Om rengring blir ndvndig under svra driftfrhllanden: 1) 2) Notera antennmaterial och processmedium och vlj en rengringslsning som inte r skadlig fr ngondera. Ta ur apparaten och rengr antennen med en duk och lmplig rengringslsning.

Reparation och ansvarsbegrnsning


Fr detaljerad information, v.g. se omslagets tredje sida.

Byte av antenn eller elektronik/hlje


Om antennen behver bytas p grund av skada eller bortfall, kan den bytas utan krav p omkalibrering. Byte till en annan antenntyp kan utfras av en av Siemens auktoriserad reparationsverkstad eller fackman. Om elektronik eller hlje behver bytas till fljd av skada eller bortfall, var vnlig se till att rtt antennversion anvnds, annars mste omkalibrering utfras av person auktoriserad av Siemens.

Sida SV-22

SITRANS LR250 (HART) SNABBSTARTSMANUAL

A5E31997170

Tabell ver maximala processtemperaturer


Flnsadapterversioner av LR250
VARNING: Inre temperatur fr inte verstiga 80 C (176 F). Anmrkningar: Tabellen r endast avsedd fr vgledning och terger inte varje mjlig typ av processanslutning. (Den kommer t.ex. INTE att glla om ni monterar SITRANS LR250 direkt p en plttanksyta. Tabellen beaktar inte uppvrmning frn direkt solljusexponering. Parameter 3.2.1 Aktuell inre temperatur vervakar den inre temperaturen.

Omgivande temperatur (oC)

Maximala processtemperaturer mot tillten omgivande

Svenska

processtemperatur Processtemperatur (oC)

Kretseffekt
Effektmatningsbehov
Anmrkning: Kurvorna nedan gller fr en fristende apparat, konfigurerad med Siemens handhllna programmeringsenhet. Kurva 1: Allmnt syfte, Egensker, Gnistfri/ Energibegrnsad/ Icke antndlig

Kretsresistans RL

TILLTET DRIFTOMRDE

Kretsresistans VL

Anmrkning: Vid anvndning av HART-kommunikationer, r minimispnningen med 220 Ohm (RL) 16,3 VDC.

A5E31997170

SITRANS LR250 (HART) SNABBSTARTSMANUAL

Sida SV-23

Kurva 2: Flamsker, Frhjd skerhet, Explosionssker Kretsresistans RL

TILLTET DRIFTOMRDE

Kretsresistans VL

Svenska

Anmrkning: Vid anvndning av HART-kommunikationer, r minimispnningen med 220 Ohm (RL) 20,94 VDC.

Startbeteende
Anordningen drar mindre n 3,6 mA vid starten. Tid till den frsta mtningen r mindre n 50 sekunder.

Sida SV-24

SITRANS LR250 (HART) SNABBSTARTSMANUAL

A5E31997170

Unit Repair and Excluded Liability

All changes and repairs must be done by qualified personnel, and applicable safety regulations must be followed. Please note the following: The user is responsible for all changes and repairs made to the device. All new components must be provided by Siemens. Restrict repair to faulty components only. Do not re-use faulty components

Reparation af enheden og ansvarsbegrnsning:


Alle ndringer og reparationer skal udfres af kvalificeret personale, og de gldende sikkerhedsbestemmelser skal overholdes. Bemrk venligst flgende: Brugeren er ansvarlig for alle de p apparatet udfrte ndringer og reparationer. Alle nye komponenter skal vre leveret af Siemens. Reparr kun defekte komponenter. Defekte komponenter m ikke genbruges

Gertereparatur und Haftungsausschluss:


Alle nderungen und Reparaturen mssen von qualifiziertem Personal unter Beachtung der jeweiligen Sicherheitsbestimmungen vorgenommen werden. Bitte beachten Sie: Der Benutzer ist fr alle nderungen und Reparaturen am Gert verantwortlich. Alle neuen Bestandteile sind von Siemens bereit zu stellen. Reparieren Sie lediglich defekte Bestandteile. Defekte Bestandteile drfen nicht wiederverwendet werden.

:
, . : . Siemens. . .

Reparacin del dispositivo y lmite de responsabilidad:


Las modificaciones y reparaciones debern ser efectuadas por personal calificado de acuerdo con las normas de seguridad aplicables. Notas importantes: El usuario es el nico responsable de las modificaciones y reparaciones del dispositivo. Recomendamos utilizar slo recambios originales Siemens. Reparar slo los componentes defectuosos. Los componentes defectuosos no se deben reutilizar.

Rparation de lunit et limite de responsabilit :


Les modifications et rparations doivent tre effectues par un personnel qualifi en accord avec les consignes de scurit applicables. Remarques importantes : Lutilisateur est seul responsable des modifications et rparations effectues sur lunit. Utiliser seulement des composants fournis par Siemens. Rparer uniquement les composants dfectueux. Les composants dfectueux ne doivent pas tre rutiliss.

Riparazioni dellapparecchiatura e limiti di responsabilit:


Le modifiche e le riparazioni devono essere effettuate solo da personale qualificato, rispettando le normative sulla sicurezza. Note importanti: Lutente responsabile delle eventuali modifiche e riparazioni effettuate sullapparecchiatura. Utilizzare solo pezzi di ricambio originali forniti da Siemens. Riparare solo i componenti difettosi. E importante non riutilizzare i componenti difettosi.

Reparatie van apparatuur en uitsluiting van aansprakelijkheid:


Alle modificaties en reparaties moeten worden uitgevoerd door gekwalificeerd personeel en de geldende veiligheidsvoorschriften moeten worden aangehouden. Let op: De gebruiker is verantwoordelijk voor alle modificaties en reparaties die worden uitgevoerd aan het apparaat. Alle nieuwe onderdelen moeten zijn geleverd door Siemens. Beperk de reparatie uitsluitend tot de defecte componenten. Defecte componenten niet opnieuw gebruiken.

Reparao da Unidade e Responsabilidade Excluda


Todas as alteraes e reparaes devem ser realizadas por pessoal qualificado e devem ser seguidas as regras de segurana aplicveis. Por favor, note o seguinte: O usurio responsvel por todas as alteraes e reparos efetuados no dispositivo. Todos os novos componentes devem ser fornecidos pela Siemens. Reparo restrito a apenas a componentes danificados. No reutilize componentes danificados.

Yksikn korjaaminen ja vastuuvapaus:


Muutos- ja korjaustyt saa suorittaa ainoastaan ptev henkilkunta, ja voimassa olevia turvallisuusmryksi on noudatettava. Pyydmme ottamaan huomioon seuraavat seikat: Kyttj on vastuussa kaikista laitteeseen tehdyist muutoksista ja korjauksista. Kaikki uudet osat on hankittava Siemens:lt. Korjaukset on kohdistettava ainoastaan viallisiin osiin. Viallisia osia ei saa kytt uudelleen.

Reparation och ansvarsfrihet:


Alla ndringar och reparationer mste utfras av kompetent personal och under iakttagande av gllande skerhetsbestmmelser. Observera att: Anvndaren ansvarar fr alla ndringar och reparationer som grs p enheten. Alla nya delar mste komma frn Siemens. Reparara endast med fel behftade delar. Delar behftade med fel fr ej teranvndas.

www.siemens.com/processautomation

For more information


www.siemens.com/level www.siemens.com/weighing