Supplement to the Quick Start for category 3 apparatus (Ex zone 2)

1.1 General
This supplement is valid for the following flowmeters, provided with these markings: Flowmeter OPTIFLUX 1300 C OPTIFLUX 2300 C OPTIFLUX 4300 C OPTIFLUX 5300 C OPTIFLUX 6300 C OPTIFLUX 1000 F OPTIFLUX 2000 F OPTIFLUX 4000 F OPTIFLUX 5000 F OPTIFLUX 6000 F IFC 300 F marking II 3G Ex nA [nL] IIC T4...T3

II 3G Ex nA nL IIC T4...T3

II 3G Ex nA [nL] IIC T4

1.2 Installation Instructions for safe use


Installation and connection of the flowmeter must be realised conform standard EN IEC 60079-14 ("Electrical Installations in hazardous areas, other than mines") or equivalent national standard.

Electrical data of terminals / connections: Terminals Mains: L, N

I/O: A, A-, A+, B, B-, C, C-, D, DOnly for IFC 300 F or OPTIFLUX x000 F (x = 1, 2, 4, 5 or 6) Field current: 7, 8, 9 Only for OPTIFLUX x000 F (x = 1, 2, 4, 5 or 6) Electrode circuit: 1, 2, 3, 4 Only for IFC 300 F Electrode circuit : 1, 2, 20, 3, 30, 4, 40

Type of Ex protection / electrical data Ex nA Un = 100230 VAC or 1224 VDC or 24 VAC/DC (order specific) Pn: 22 VA or 12 W. Ex nA Un < 32 V, In < 100 mA Ex nA Un < 40 V, In = 125 mA Ex nL Ui = 20 V, Ii = 175 mA, Ci = 0 nF, Li = 0 mH Ex [nL] Uo = 14 V, Io = 70 mA, Po = 300 mW, Co = 430 nF, Lo = 2 mH

Notes:
7.30978.22.00 Supplement to the Quick Start 1/2

The Ex [nL] electrode circuit of the OPTIFLUX x300 C (compact flowmeter) is a internal circuit. Hence no Ui, Ii, Pi etc parameters are applicable for this circuit. Connections to mains (L, N or L+, L-), signal I/O (A, A-, A+, B, B-, C, C-, D and D-) and field current (7, 8, 9) can only be made when the flowmeter is not energized or when there is no explosive gas atmosphere surrounding the flowmeter. Used cable glands must have an IP degree to EN 60529 of at least IP65. The glands must be suitable for the types of cables used. For example the sealing ring must be suitable for the outer diameter of the cables. The temperature class T4 or T3 of OPTIFLUX x300 C and OPTIFLUX x000 F, depends on ambient temperature Ta and maximum process temperature Tp, according to following table: Ta < 40 C 130 140 40 Ta< 50 C 130 -50 Ta < 60 C 60 --

maximum Tp in C T4 T3

1.3 Serviceable parts


With exception of the mains fuse link, the IFC 300 electronic unit contains no serviceable parts. In case of replacement of the main fuse link, 5x20 mm fuse links to standard IEC 60 127-2, breaking capacity High, rated voltage 250 V, must be used: Mains 100230 VAC 1224 VDC 24 VAC/DC Mains fuse link 0,8 A Time-lag (T800H250V) 2,0 A Time-lag (T2H250V) 2,0 Time-lag (T2H250V)