MEC-1000OEM Portable Patient Monitor

MEC-1000 Portable Patient Monitor

10.4 colorful TFT display, with maximum 4 waveforms. Arrhythmia analysis, S-T segment and pacemaker detection Three lead ECG, and ECG waveforms cascade Suitable for adult, pediatric and neonatal patient All-in-one designed module with standard parameters: ECG, Resp, Spo2, 2-Temps, NIBP, Pulse Rate Built-in recorder as option with numerous print formats selectable TECHNICAL SPECIFICATIONS
Dimension and Weight
Dimension: Weight: 318mm (W) x 260mm (H) x 122mm (D) 3.5kg

Operation Environment
Power requirements: Temperature: Humidity: AC100~250V, 50/60 Hz 0~40 C 30~85% Class I equipment

Classification
Anti-electroshock type
Mindray Co., Ltd.

MEC-1000OEM Portable Patient Monitor

EMC type Anti-electroshock degree Harmful liquid proof degree Working system

Class A ECG(RESP), SpO2, NIBP, TEMP CF Ordinary equipment (sealed equipment without liquid proof) Continuous running equipment

Application Scope
Neonatal, pediatric, and adult patients

Performance Specifications
Display: Traces: Sweep Speed: Indicator: 10.4 TFT display Resolution: 640 x 480 4 waveforms 6.25, 12.5, 25 and 50mm/s individually selectable Alarm indicator Power indicator Battery charge indicator Parameter module interface Power input socket Built-in and Rechargeable, lasting for maximum 1.5 hours 1~72 hours User-adjustable High & Low limits , audible and visual 3-level alarm Optional, Built-in, thermal array Plethysmogram waveforms: 2 channels Record mode: manual, on alarm, time- defined, etc Recording width: 48mm Recording speed: 25mm/s, 50mm/s

Interface: Battery: Trend time: Alarm: Recorder:

ECG
3-lead Input: Lead selection: ECG waveform: Gain selection: Sweep speed: Heart rate Measurement Range: (R, L, F or RA, LA, LL) I; II; III 1 channel x0.25, x0.5; x1; x2; auto 12.5mm/s, 25mm/s, 50mm/s Adult: 15~300bpm Child/Neonate: 15~350bpm Accuracy: 1% or 1bpm whichever is greater Resolution: 1bpm Sensitivity > 200 uV(VP-P) Differential Input Impedance > 5 M CMRR Monitor > 105dB Operation > 105 dB Diagnosis > 85 dB Electrode offset potential 300mV Leakage Current <10 uA Baseline Recovery < 3 s After Defi. ECG Signal Range 8 mV(Vp-p) Bandwidth Surgery 1 ~ 15 Hz Monitor 0.5 ~ 35 Hz
Mindray Co., Ltd.

MEC-1000OEM Portable Patient Monitor

Diagnostic 0.05 ~ 100 Hz Calibration Signal 1 mV p-p, 5% Accuracy ST Segment Monitoring Range Measure and Alarm -2.0 ~ +2.0 mV Alarm range: -2.0mV~2.0mV ARR Detecting Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC Alarm Available Pace Detection: Yes Protection: Withstand 4000VAC/50Hz voltage in isolation. Against electrosurgical interference and defibrillation Alarm : YES, audible and visual alarm, alarm events recallable

Respiration
Method: Differential Input Impedance Measuring Impedance Range: Base line Impedance Range: Bandwidth Operation modes: Measurement range: Resolution: Accuracy: Alarm range: Apnea alarm: RA-LL impedance >2.5 M 0.3~5.0 100 2.5 K 0.3 ~ 2.5 Hz Auto / Manual Adult: 0-120rpm; Child/Neonate: 0-150rpm 1rpm 2rpm Adult:0~120rpm Child/Neonate:0~150rpm YES

NIBP
Method Oscillometric Mode Manual, Auto, STAT Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 (Min) Measuring Period in STAT Mode 5 Min Pulse Rate Range 40 ~ 240 bpm Measuring and alarm Range Adult Mode SYS 40 ~ 270 mmHg DIA 10 ~ 215 mmHg MEAN 20 ~ 235 mmHg Pediatric Mode SYS 40~ 220 mmHg DIA 10 ~ 150 mmHg MEAN 20 ~ 165 mmHg Neonatal Mode SYS 40 ~ 135 mmHg DIA 10 ~ 100 mmHg MEAN 20 ~ 110 mmHg
Mindray Co., Ltd.

MEC-1000OEM Portable Patient Monitor

Alarm Type Resolution

Pressure

SYS, DIA, MEAN 1mmHg

Accuracy( Pressure ) Maximum Mean error Maximum Standard deviation Over-pressure protection: Adult:: Child: Neonate: 5mmHg 8mmHg 2973mmHg 2403mmHg 1473mmHg

SpO2
Measurement range: Resolution: Accuracy: Alarm range: Pulse Rate: 0~100% 1% 2% (70~100%), 0~69% unspecified 0~100% Range: 0-254bpm Resolution: 1bpm Accuracy: 1bpm Alarm range: 0-254bp

Temperature
Measurement range: Resolution: Accuracy: Channel: Alarm range: 0~50 oC 0.1 oC 0.1 oC One channel 0~50 oC

Recorder
Record Width Paper Speed Trace Recording types: 48 mm 25/50 mm/S 2 Continuous real-time recording 8 second real-time recording Auto 8 second recording Parameter alarm recording Waveform freeze recording Trend graph/table recording ARR events review recording Alarm event review recording NIBP review recording Drug Calculation and titration table recording Monitor information recording

Recall and storage


Trend Recall
Mindray Co., Ltd.

MEC-1000OEM Portable Patient Monitor

Short 1 hrs, 1 Second Resolution Long 72 hrs, 1 Min. Resolution Alarm Event Recall 60 alarm events of all parameters and 8/16/32seconds of corresponding waveform. NIBP Measurement Recall 400 NIBP measurement data. Power-off data storage Patient information, alarm limitation, clock

Specifications subject to change without prior notice