Está en la página 1de 4

Measurement of Power Devices

Using External DC Power Supply

Application Note 4156-5


HP 4155B / 4156B
Semiconductor
ParameterAnalyzer

Expansion of measurement range to cover several


amperes easily achieved using external DC power
supply

1
Introduction Shield Box

If used in combination with an INTLK


INTLK
external DC power supply, the SMU 1
HP 4155B/4156B semiconductor Vg
parameter analyzer can easily Cv+ SMU 2
evaluate the DC characteristics of D Vvp
+
a power device that requires a A1
VMU 1
current of several amperes. This A3
V Vim
application note describes how to G
- VMU 2
connect the highly cost-effective S A6
HP E3615A DC power supply to To Common HP 4155B
the HP 4155B and to use the HP E3615A
HP 4155B's automatic analysis HP 16495H/J
Connector Plate
functions to automatically
calculate threshold voltage (Vth). Figure 1. Diagram of Connections for Power Transistor Measurement

Power transistor CHA NNEL S : CHA NNEL D E F I NI T I ON 9 4 OCT 0 5 0 5 : 1 5 PM


V

measurement * M E A S U R E ME N T MO D E
S WE E P I

Curve tracers are widely used for * CHANNE L S


power transistor measurement, ME A S U R E STBY SERI ES
UNI T V N A ME I N A ME MO D E F CT N RES I S T A NCE V PUL SE
but have few analysis functions S MU 1 : M P Vg I g VPUL SE VAR2 0 ohm

and have difficulty in extracting S MU 2 : M P Vv p I v p V VAR1 0 ohm


S MU 3 : M P
transistor parameters. They also S MU 4 : M P
I PUL SE

have the disadvantage of being V SU1


V SU2
- - - - - - -
- - - - - - -
difficult to automate in order to V MU 1 Vi m - - - - - - - DV OL T - - - - - - - - - C O MM O N
V MU 2 - - - - - - - DV OL T - - - - - - - - -
improve productivity.

The HP 4155B/4156B offers a


wide range of analysis functions
including log sweep measurement DEL ET E
R OW
functions, tangent line, regression
line, and marker Min/Max. It also Se l e c t Mo d e wi t h s of t k ey or r ot ar y k nob. B

has built-in HP Instrument BASIC, CHANNEL


DEF
USER
F CT N
US ER
VAR
NEXT
P A GE
which allows the control and Figure 2. CHANNEL DEFINITION Page
automation of all measurements
without any external controller, can be monitored remotely via its SMU2 is connected to the control
and offers major improvements in sense resistor. This paper terminal Cv+ to control output
productivity from device describes how this function can be voltage, and the voltage
development and design to used to increase the output power measurement units (VMU1 and
incoming inspection. The and expand the current VMU2) are connected to sense
dedicated expander (HP 41501B) measurement range of the resistor terminals A1 and A3 to
gives a measurement range of up HP 4155B. Our example uses the monitor output current. To give
to 1A/200V, and using an external HP E3615A, which provides a the HP E3615A a circuit common
DC power supply that can be maximum of 60W and 3A/20V, but potential, terminal A6 is connected
remotely controlled allows further the HP E3614A (6A/8V) or the HP to the common terminal which is
expansion of the measurement E3617A (1A/60V) are also suitable. the outer shield of VMU2. These
range. connections are made using the
Connections HP 16495H/J option 002 connector
The HP E3615A can be remotely plate*1.
controlled by applying a voltage to The HP 4155B and HP E3615A are
its control terminal, and current connected as shown in Figure 1.

2
ME A S U R E : S WE E P SE T UP 9 3 SEP0 2 0 3 : 3 8 AM

CHANNEL S : USER F U NCT I ON D E F I NI T I ON 9 3 SEP0 2 0 3 : 3 7 AM * VARI ABL E V AR1 VAR2


UNI T S MU 2 : H R S MU 1 : H R
N A ME Vv p Vg
S WE E P MO D E D OU B L E S I NGL E
L I N / L OG L OG 10 L I NE A R
ST A RT 1 0 . 0 mV 5. 400 V

* USER F U N C T I ON S T OP 18. 000 V 5. 800 V


STEP - - - - - - - - - 2 0 0 . 0 mV * S MU PUL SE
N A ME UNI T DE F I NI T I ON
NO OF ST EP 33 3 UNI T S MU 1 : H R
Vd V Vv p * ( 2/ 3)
C OMP L I A N C E 1 0 0 . 0 0 mA 1 0 0 . 0 0 mA N A ME Vg
I d A V i m/ . 2
P O WE R C O MP OF F OF F P E R I OD 2 0 0 . 0 ms
WI D T H 2 0 . 0 ms
* T I MI N G BASE 0. 0000 V
H OL D T I ME 3 0 0 ms
* S WE E P CONT I NU E AT A NY St a t u s
* C ON S T A N T
UNI T
N A ME
MO D E
S OU R C E - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
C OMP L I A N C E - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

Figure 3. USER FUNCTION DEFINITION Page Figure 4. SWEEP SETUP Page


GRAPH/ L I ST : GRAPHI CS SHORT 9 4 J UN0 7 0 5 : 0 9 AM GR A P H / L I S T : GR A P H I C S ME D I U M 9 4 J UN0 7 0 5 : 1 1 AM

MA RK E R ( 9. 4940000000 V 1. 4080000000 A ) MA R K E R ( 6 . 0 0 4 V 1 . 5 6 1 0 30 00 0 0 A 1 . 3 6 9 3 4 21 05 3 )
Vt h 4. 8 6 4 01 46 0 56 V
( A) ( A) ( )

1. 50 3 . 0 0 1 . 5 0
o
*

I d I d GM

200. m 2 0 0 . m o 1 00 . m
/ di v / d i v / d i v

0. 00 0 . 0 0 0 . 0 0

6. 60m Vd ( V) 1. 00 / di v 12. 0 3 . 0 0 Vg ( V) 50 0. m / d i v 7 . 6 0

Figure 5. Example of Drain Characteristic Measurements Figure 6. Drain Current versus Gate Voltage
Characteristics to Obtain Vth

HP 4155B settings and factor depending on the DC power set so that pulse of SMU1 has a
results of measurements supply (2/3 for the HP E3615A). duty cycle of 10 %. The
The sense resistance is 0.2 W for measurement results of drain
Figures 2 to 4 show the HP 4155B the HP E3615A, so Id = Vim/0.2. characteristics are shown in
setup screens for the drain So, you set the user functions as Figure 5.
characteristic measurements. On follows:
the CHANNEL DEFINITION page
Automatic extraction of
Vd = Vvp * (2/3) transistor parameters
(Figure 2), SMU1, which is
connected to the gate, is set to Id = Vim / 0.2
pulse output. Driving the gate The HP 4155B/4156B can
using a pulse voltage reduces the For other DC power supplies, see automatically extract parameters
average output current, so thermal *2 at end of this note. after measurements*3. Figure 6
drift is reduced. VMU1 and VMU2 shows the automatic extraction of
The SWEEP SETUP page (Figure threshold voltage from the "drain
are set to differential mode and 4) is used to set the sweep
the potential difference (Vim) is current versus the gate voltage"
conditions for SMU1 and SMU2, characteristics graph. Before
measured. and the pulse output for SMU1. In measurement, you set up the
The USER FUNCTION this example, to reduce the ANALYSIS SETUP page to extract
DEFINITION page (Figure 3) is number of measurements in the desired parameters. After
used to calculate the Vd and Id high voltage range and prevent measurement, HP 4155B/4156B
variables. Vd is the SMU2 output heating, SMU2 is set to log sweep. automatically extracts the
voltage (Vvp) multiplied by a Pulse width and pulse period are parameters.

3
In this example, the HP 4155B The HP 4155B/4156B's automatic For more information about Hewlett-
Packard test and measurement products,
finds maximum Gm, then draws analysis functions and built-in HP applications, services, and for a current
tangent to Id curve of this point. Instrument BASIC environment sales office listing, visit our web site:
The x-intercept of this tangent line also allow HP 4155B/4156B to http://www.hp.com/go/tmdir
You can also contact one of the following
is Vth, which is automatically become an automatic test station. centers and ask for a test and measure-
displayed. Measurement ment sales representative.
productivity, repeatability, and Notes:
*1 Please observe following safety United States:
speed are all significantly Hewlett-Packard Company
improved. precautions when using the Test and Measurement Call Center
HP 4155B. P.O. Box 4026
Englewood, CO 80155-4026
Tips on maximizing Make sure that the wiring used
(tel) 1 800 452 4844
measurement accuracy with the connector plate and
shield box has adequate Canada:
Hewlett-Packard Canada Ltd.
To obtain the highest possible capacity. The maximum output 5150 Spectrum Way
accuracy of measurement, you voltage and current of each Mississauga, Ontario
external DC power supply are L4W 5G1
should observe the following (tel) (905) 206 4725
points: as follows:
Europe:
Hewlett-Packard Company
Instead of calculating Vd by Modle No. Max. output Max. output
European Marketing Centre
user functions, directly connect voltage current
P.O. Box 999
one SMU to drain and one SMU HP E3614A 8V 6A 1180 AZ Amstelveen
The Netherlands
to source, and subtract voltage HP E3615A 20 V 3A (tel) (31 20) 547 9900
difference between the drain HP E3616A 35 V 1.7 A
and the source. This will Japan:
HP E3617A 60 V 1A Hewlett-Packard Japan Ltd.
eliminate the error caused by Measurement Assistance Center
remote programming and the 9-1, Takakura-Cho, Hachioji-Shi,
To protect yourself against Tokyo 192, Japan
voltage loss caused by the
electrical shocks when (tel) (81) 426 56 7832
cables. (fax) (81) 426 56 7840
replacing the DUT, be sure to fit
To minimize voltage loss in an INTLK switch and install the Latin America:
measurement system, including Hewlett-Packard Company
cables that carry a high current, Latin American Region Headquarters
use cables that are as short and external DC power supply in a 5200 Blue Lagoon Drive
thick as possible. shield box. 9th Floor
Miami, Florida 33126
*2 For other external DC power U.S.A.
After a sweep, when the drain (tel) (305) 267 4245/4220
voltage returns abruptly to zero, supplies, use the following user (fax) (305) 267 4288
the HP E3615A may generate a functions:
Australia/New Zealand:
spike. To eliminate this spike, HP E3614A: Hewlett-Packard Australia Ltd.
use the double-staircase sweep Vd = Vvp * (4/9), Id = Vim / 0.1 31-41 Joseph Street
mode. HP E3616A: Blackburn, Victoria 3130
Australia
Vd = Vvp * (7/9), Id = Vim / 0.6 (tel) 1 800 629 485 (Australia)
As much as possible, you should HP E3617A: (tel) 0 800 738 378 (New Zealand)
use a heat sink. Vd = Vvp * (6/7), Id = Vim / 0.89 (fax) (61 3) 9210 5489
Asia Pacific:
Conclusion *3 For a detailed explanation of Hewlett-Packard Asia Pacific Ltd
automatic analytical functions, 17-21/F Shell Tower, Times Square,
1 Matheson Street, Causeway Bay,
This application note described see "Application Note 4156-2, Hong Kong
one example of how the HP 4155B Automated Extraction of (tel) (852) 2599 7777
Semiconductor Parameters (fax) (852) 2506 9285
can be used with the highly
cost-effective HP E3615A DC Using the HP 4155B/4156B" (HP © Hewlett-Packard Company 1998
publication number Data subject to change
power supply to extract power Printed in Japan 1/1998
transistor parameters, which are 5963-1249E). 5963-2365E
difficult to extract using the
conventional curve tracer method.