Documentos de Académico
Documentos de Profesional
Documentos de Cultura
Procedure
Type
SAFETY
Door
interlock
(beam
off)
Door
closing
safety
Audiovisual
monitors
Beam
on
indicator
Stereotactic
interlocks
Radiation
area
monitor
Laser
guard-interlock
test
Manufacturers
test
procedures
Functional
Functional
Functional
Functional
Functional
Functional
MECHANICAL
Functional
Daily
Annually
Functional
Monthly
Annually
General:
2
mm
IMRT:
1.5
mm
SBRT:
1
mm
2
mm
2
mm
SBRT:
1mm
1 mm or 1% on a side
1mm
1 degree
2 mm
2 mm
1 mm
1
mm
2
mm/1
degree
SBRT:
1
mm/0.5
degree
2
mm
1
mm
Laser localization
2 mm or 1% on a side
Procedure
Type
MECHANICAL
Annually
Functional
General:
+/-
2
mm
IMRT:
+/-
1mm
SBRT:
<
+/-
1
mm
Functional
DOSIMETRY
Daily
Monthly
Annually
3%
3%
2%
2%
2%
2%
at
operating
dose
rate
(IMRT,
SBRT)
1%
1%
2%
or
2
mm
1%
+/-
1%
1%
Procedure
Type
DOSIMETRY
Annually
+/-
1%
MU
set
vs.
delivered:
1.0MU
or
2%,
whichever
is
greater
Gantry
arc
set
vs.
delivered:
1
degree
or
2%,
whichever
is
greater
+/-
1%
(absolute)
2%
for
field
size
<4x4cm2,
1%
>
4x4cm2
+/-
2%
from
baseline
MULTILEAF
COLLIMATOR
Qualitative
test
(matched
segments,
picket
fence)
Setting
vs.
radiation
field
for
two
patterns
(non-IMRT)
Travel
speed
(IMRT)
Leaf
position
accuracy
(IMRT)
Weekly
Monthly
Annually
2 mm
+
1.0
mm
<
1.0
mm
radius
+
2.0
mm
Functional
(Universal)
CAX
45o
or
60o
WF
(within
2%)
Virtual:
5%
from
unity,
otherwise
2%
Check
of
off-center
ratios
at
80%
field
width
@
10
cm
to
be
within
2%
Respiratory
Gating
Quality
Assurance1
Frequency
&
Tolerance
Daily
Monthly
Procedure
Type
Beam
output
constancy
Phase,
amplitude
beam
control
In-room
respiratory
monitoring
system
Gating
interlock
Beam
energy
constancy
Temporal
accuracy
of
phase/amplitude
gate
on
Calibration
of
surrogate
for
respiratory
phase/amplitude
Interlock
testing
Annually
2%
Functional
Functional
Functional
2%
100
ms
of
expected
100 ms of expected
Functional
Procedure
Type
PLANAR
kV
AND
MV
(EPID)
Collision
interlocks
Positioning/repositioning
Imaging
and
treatment
coordinate
coincidence
Scaling
Annually
Spatial
resolution
Contrast
Uniformity
and
noise
MV:
Full
range
of
travel
SDD
Imaging
dose
Collision interlocks
Functional
<
2
mm
SBRT:
<
1
mm
<
1
mm
Geometric distortion
Spatial resolution
<
2
mm
SBRT:
<
1
mm
Baseline
Contrast
Baseline
HU constancy
Baseline
Baseline
Baseline
<
5
mm
Baseline
Baseline
Procedure
Type
SAFETY
Collision
and
other
interlocks
Warning
lights
Replace/refresh
+
1
mm
+
1
mm
SYSTEM OPERATION/ACCURACY
kV/MV/laser
alignment
Couch
shifts:
accuracy
of
motions
Functional
Functional
Annually
AP/mediolateral/craniocaudal
orientations
are
maintained
(upon
upgrade
from
CT
to
IGRT
system)
Accurate
IMAGE QUALITY
Scale/distance/orientation accuracy*
Baseline
Baseline
<
2
mm
(or
<
5
lp/cm)
Baseline
Baseline
DOSE
Imaging dose
Baseline
Baseline
SYSTEM OPERATION
Uniformity/noise*
High
contrast
spatial
resolution*
Low
contrast
detectability*
CT
number
accuracy
and
stability*
(if
used
for
dose
calculation)
*These
tests
may
be
performed
on
a
semiannual
basis
after
stability
has
been
demonstrated,
6-12
months
after
commissioning
Initially
TOLERANCE
NCRP
recommendations
or
applicable
regulatory
limits
+
20%
of
manufacturer
specifications
FREQUENCY
TOLERANCE
Daily
+ 2 mm
+ 2 mm
Monthly
Annually
TOLERANCE
Annually
Annually
+
1o
or
+
1
mm
from
nominal
position
Annually
Manufacturer specifications
Manufacturer
specifications
or
Report
No.
39
recommendations
FREQUENCY
TOLERANCE
0
+
5
HU
for
water
Manufacturer
specifications
+
1
mm
Within
0
+
5
HU
Consistent
with
commissioning
results
and
test
phantom
manufacturer
specifications
Manufacturer
specifications
Manufacturer
specifications
Scan
localization
Radiation
profile
width
Daily
Monthly
Annually
Daily
Daily
Monthly
Monthly
Annually
Annually
or
after
scanner
calibration
Spatial
resolution
Contrast
resolution
Annually
Annually
+ 1 mm of nominal value
Procedure
Type
Localizing
lasers
Optical
distance
indicator
(ODI)
Field
size
indicator
Gantry/collimator
angle
indicators
Crosshair
centering
Focal
spot-axis
indicator
Flouroscopic
image
quality
Emergency/collision
avoidance
Light/radiation
field
coincidence
2
mm
2
mm
MECHANICAL
2
mm
1o
2
mm
diameter
2
mm
Baseline
Functional
2
mm
or
1%
Baseline
Daily
Annually
Monthly
Annually
2
mm
diameter
2
mm
diameter
2
mm
diameter
2
mm
diameter
2
mm
2
mm
Daily
Monthly
Annually
Baseline
Baseline
Baseline
Baseline
Instrument/Device
Ionization
chamber
and
electrometer
Beam
scanning
systems*
*tests
should
be
performed
to
evaluate
reliability
and
constancy
of
response
over
time
interval
required
to
make
a
full
set
of
measurements
Ancillary
equipment**
Phantom
positioning
devices
(rulers,
chamber
positioning
devices,
solid
phantom
materials)
Thermometers
Barometers
Electronic
devices
Relative
dose
measuring
equipment
Diodes,
TLDs,
film-
DO
NOT
require
calibration
Survey
Meters
Frequency
Regularly
Aneroid/Electronic:
weekly
or
monthly
Mercury:
before
first
use
Weekly
or
monthly
Regularly
Before
first
use
Annually
Before
each
use
Before
each
use
Procedure
Individual
field
and
plan
verification
Dose
to
a
test
point
in
each
IMRT
field
Static
field
v.
sliding
window
field
dose
distribution
as
a
function
of
gantry
and
collimator
angles
All
commissioning
procedures:
-Stability
of
leaf
speed,
acceleration
and
deceleration
-MLC
transmission
-Leaf
positional
accuracy
-Standard
plan
verification
Tolerance
Annually
MLC
Quality
Assurance4
Frequency
Procedure
Check
of
MLC-generated
field
v.
simulator
film
(or
DRR)
before
each
field
treated
Double
check
of
MLC
field
by
therapists
for
each
fraction
On-line
imaging
verification
for
patient
on
each
infraction
Port
film
approval
before
second
fraction
Setting
v.
light
field
v.
radiation
field
for
two
designated
patterns
Testing
of
network
system
Check
of
interlocks
Setting
v.
light
field
v.
radiation
field
for
patterns
over
range
of
gantry
and
collimator
angles
Water
scan
of
set
patterns
Film
scans
to
evaluate
interleaf
leakage
and
abutted
leaf
transmission
Review
of
procedures
and
in-service
with
therapists
Tolerance
Patient specific
2 mm
Patient specific
Expected field
Patient specific
Physician discretion
Patient
specific
Quarterly
Physician
discretion
1
mm
Quarterly
Quarterly
Annually
Annually
Annually
Annually
Treatment
Planning
Computer:
Commissioning~
Ongoing~
(to
address
major
sources
of
uncertainty)
Understand
treatment
planning
software
algorithms
and
their
dependence
on
measured
data
Collection
of
appropriate
data
required
by
TPS
and
entry
of
this
data
Collection
of
additional
data
fully
characterizing
beams
for
which
treatment
planning
will
be
done
and
representing
a
range
of
clinical
situations
Calculation
of
dose
distributions,
point
doses,
and
MU
or
treatment
time
settings,
along
with
comparison
of
these
calculations
with
measured
data
Inaccuracies
in
measured
beam
and
patient-
specific
data
Inaccuracies
of
data
entry
Inaccuracies
of
data
output
Algorithm
reproducibility
and
accuracy
of
calculations
Check
contours
weekly,
agree
within
2
mm
Check
no
less
than
annually,
agree
within
2
mm
or
2%
per
ICRU
guidelines
IMRT:
Commissioning~
Verify
the
dose
predicted
by
the
planning
system
is
accurate
to
within
acceptable
limits
Test
for
single,
simple
geometry
and
for
cumulative
dose
from
multiple
segments.
Ensure
leaf
and
jaw
transmission
are
calculated
correctly
in
the
TPS,
especially
for
small
MU
fields
Patient-specific
Dose
Verification~
*Per
ASTRO/ACR,
dose
delivery
MUST
be
documented
for
every
IMRT
plan
by
comparing
measured
dose
in
a
phantom
to
planned
dose
from
the
TPS
Can
be
accomplished
by
measuring:
Point
or
planar
dose
for
a
single
or
composite
field
SRS:
Brachytherapy:
Instrument/Device
Applicators:
Radioactive
Sources:
Daily QA:
Monthly QA:
Procedure
Radiation
safety
surveys
Frequency
Prior
to
initial
use
After
each
use
Routine
Prior
to
initial
use,
at
least
semi-annually
(some
cesium
sources
only
tested
every
3
years)
Objective
References
1. Klein
EE,
Hanley
J,
Bayouth
J,
et
al.
TG-142:
Quality
assurance
of
medical
accelerators.
Med
Phys.
2009;36(9):4197-4212.
http://dx.doi.org/10.1118/1.3190392
2. Bissonnette
J-P,
Balter
PA,
Dong
L,
et
al.
TG-179:
Quality
assurance
for
image-guided
radiation
therapy
utilizing
CT-based
technologies.
Med
Phys.
2012;39(4):1946-1963.
http://dx.doi.org/10.1118/1.3690466
3. Mutic
S,
Palta
JR,
Butker
EK,
et
al.
TG-66:
Quality
assurance
for
computed-tomography
simulators
and
the
computed-
tomography-simulation
process.
Med
Phys.
2003;30(10):2762-2792.
http://dx.doi.org/10.1118/1.1609271
4. Khan
FM.
Quality
assurance.
In:
The
physics
of
radiation
therapy.
4th
ed.
Baltimore,
MD:
Lippincott
Williams
&
Wilkins;
2010:375-404.
5. Lenards
N.
Quality
assurance:
QA
of
instruments.
[SoftChalk].
LaCrosse,
WI:
UW-L
Medical
Dosimetry
Program,
2015.
6. Khan
FM.
Measurement
of
ionizing
radiation.
In:
The
physics
of
radiation
therapy.
4th
ed.
Baltimore,
MD:
Lippincott
Williams
&
Wilkins;
2010:70-86.
7. Khan
FM.
Intensity-modulated
radiation
therapy.
In:
The
physics
of
radiation
therapy.
4th
ed.
Baltimore,
MD:
Lippincott
Williams
&
Wilkins;
2010:430-452.
8. Vann
AM,
Lenards
N.
Quality
assurance:
QA
of
TPS,
IMRT,
SRS.
[SoftChalk].
LaCrosse,
WI:
UW-L
Medical
Dosimetry
Program,
2015.
9. Lenards
N.
Quality
assurance:
QA
of
brachytherapy
equipment
&
sources.
[SoftChalk].
LaCrosse,
WI:
UW-L
Medical
Dosimetry
Program,
2015.