CALPELAB Co. Ltda.

FAX: (02) 322 6796

Certificate of Calibration
Company:

Tripoint S.A.

Certificate No .

015-15

Gag e:
Model:

Console Console WellTech-IV

OIT:

013-15

S/N
WT4C0511869

Calibration Date:

15-ene-15
15-jul-15

Serie:

Next Calibration:

Tolerance : +/-10% Desviation of Nominal

~ildi~i'.!l'';,p4~11,111911;;.~M.!Ji'"'''
_

2.5

2,5

2.50

==r=

120,00

2.50

122,00

=r=

:J

1,67%

Comments:
Procedures Used: Mag. coil measurements are made using OHMS Law (l=E/R) Method C.D. measurements are made using non-contact
induction method gauss measurements are determined by standardized magnetic probe.
ASTM E1444-0S.

We certify the equipment usedfor this calibration is traceable to NIST through one or more ofthefollowing numbers:
Reference Standard Serial : 502
Gaussmeter
Traceable to NIST certification no: MR-03A-I

Gage Status:

Last / Next Cal Dates: 0 1/12/2014 ---> 01 /1 2/2015

PASS

Geovanny Columba
ASNT TC-IA NIVEL II

This certificate is not valid unless all 1 page(s) are present:

CALPELAB Cia.Ltda. certifies that the instrument listed above meets the specifications of the manufacturer at the completion of its calibration.
Standards used are traceable to the National lnstitute of Standards and Technology (NJST), or have been derived from accepted values, natural physical
constants, or through the use of the ratio method of self-calibration techniques. Methods used are in accordance with ANSJ/NCSL Z540-I-I 994 . This
certificate is not to be reproduced other than in full , except with prior written approval of CALPELAB Cia.Ltda.
Laboratory Enviromental temperature: 20 C +/- 1 C, Relative Humidity: 50 % +/- 1O%

Page 1 of 1