Calibration Report: FB103 - A3G2 - CORVINA3


Meter ID: FIT11407
Meter Description: separador de prueba tanque 2
Operator: LOI
Differential Pressure Calibration (AIN A1, EU)
Tag: Diff Pres Range: 0.0 - 250.0 End-Element Diameter: 1.75
21/05/2014 10:41:53 Set Low Reading
As Found:
-3.423
21/05/2014 10:43:43 Set High Reading
As Found:
246.592

Pipe Diameter: 3.0


As Left:
0.000

Deviation:

-3.423

-1.3583 %

As Left:

250.000

Deviation:

-3.408

-1.3524 %

21/05/2014 10:43:59 Set Mid Point 1

As Found:

246.558

As Left:

250.000

Deviation:

-3.442

-1.3659 %

21/05/2014 10:44:09 Set Mid Point 2

As Found:

246.558

As Left:

250.000

Deviation:

-3.442

-1.3659 %

21/05/2014 10:44:21 Set Mid Point 3

As Found:

246.524

As Left:

250.000

Deviation:

-3.476

-1.3794 %

21/05/2014 10:45:22 Verify Calibration

Actual:

-0.068

Expected:

0.000

Deviation:

-0.068

-0.0272 %

21/05/2014 10:46:11 Verify Calibration

Actual:

62.329

Expected:

62.500

Deviation:

-0.171

-0.0684 %

21/05/2014 10:47:00 Verify Calibration

Actual:

124.726

Expected:

125.000

Deviation:

-0.274

-0.1096 %

21/05/2014 10:47:39 Verify Calibration

Actual:

187.295

Expected:

187.500

Deviation:

-0.205

-0.0820 %

21/05/2014 10:48:23 Verify Calibration

Actual:

249.897

Expected:

250.000

Deviation:

-0.103

-0.0412 %

21/05/2014 10:49:15 Set Zero Shift

As Found:

-0.103

As Left:

0.000

Deviation:

-0.103

-0.0412 %

21/05/2014 10:49:21 Set Zero Shift

As Found:

0.000

As Left:

0.000

Deviation:

0.000

0.0000 %