Está en la página 1de 438

ABL800 FLEX operators manual

ABL800 FLEX operators manual

Note to users of the ABL800 FLEX analyzers


Introduction This note to users gathers changes from previous note to users in one document and outlines some new changes to the operators manual of your ABL800 FLEX analyzer (from software version 6.10). Please remove the existing note to users from the binder of your manual and place this note to users in the binder instead.

Instructions to user

Brief overview of the change

Changes/Description Interference new interference results for ClO4 Limitations of use and known interfering substances: CAUTION - Known interfering substances Substance ClO4 (drugs) Interference For ClO4 , interference on cCa2+ (1.25 mmol/L level), cCl (110 mmol/L level), and cK+ (4 mmol/L level) has been detected: cCa2+ (1.25 mmol/L level): 0.27* cCl (110 mmol/L level): 4-30 cK+ (4 mmol/L level): 0.3. * Depending on the pH level USB connector Instead of the CD-ROM drive, a USB connector may be available.

USB connector or CD-ROM drive

The USB connector can be used for storing data on an USB flash drive (memory stick) or for connecting USB devices. Can also be used for installation of software.
Radiometer, the Radiometer logo, ABL, AQT, TCM, RADIANCE, PICO and CLINITUBES are trademarks of Radiometer Medical ApS. 2011 Radiometer Medical ApS. All rights reserved. 995-922. 201104A.

Measured parameters

New table for pleural liquid: Parameter Unit Measuring Range Test range

pH

pH scale

6.300-8.000

7.0-7.5*

* If the measured values obtained lie outside the test range, Radiometer advises you to repeat the measurement by means of another method. Environmental requirements Environmental ratings Indoor use statement added according to CSA regulations: Location Indoor use only

Pollution degree 2 statement added according to CSA regulations: Installation category II. Pollution degree 2.

Clot detection

A new feature is included in software version 6.06 of the following products: ABL7xx Series (XPE) analyzers ABL800 FLEX analyzers ABL800 BASIC analyzer. The new feature enables the analyzer to detect clots caught in the measurement chamber under the pH electrode. This document describes it.
When to activate the clot-detection feature

It is especially beneficial to activate the clot detection feature in analyzers frequently used to analyze samples known to be prone to clotting, for example, samples drawn from umbilical cords. Once activated clot detection is active during all sample measurements. NOTICE: It is important to note that activating the feature will delay measurement results by approximately one minute, even though it only increases the measurement cycle time by five seconds. It also increases consumption of Rinse solution, but this is minimal.

To activate the clot-detection feature Contact a Radiometer representative and request that clot detection be enabled via the service program. Clot-detection process During calibration the measuring chamber is rinsed and the pH of the Rinse solution is measured. The value is stored in the analyzer. During a sample measurement the measuring chamber is also rinsed and the pH of the rinse solution is measured and compared with the pH value stored during the last calibration. If the difference between these values (the pH drift) exceeds a pre-determined maximum value, it could indicate the presence of a clot beneath the pH electrode. When enabled, the clot-detection feature can generate three new messages that are listed below. Analyzer messages NOTICE: Operator actions are listed in order of priority. Perform the first action in the list and the actions indicated in the "Removal condition(s) list. If the message persists, perform the next listed action and the actions indicated in the "Removal condition(s) list, and so on. No. 1025 Message Clot suspected beneath pH electrode* Interpretation Operator action(s)

The difference Remove the pH electrode and between the pH of check the measuring chamber for the rinse solution, clots: clean the measuring measured during the chamber with a cotton stick sample measurement, moistened with distilled water. and that stored after Make sure no cotton fibers are the last calibration left in the measuring chamber exceeds the Clean pH electrode maximum allowable Remembrane Ref electrode drift. Replace pH electrode This indicates the Perform a 1- or 2-point calibration presence of a clot in the measuring chamber beneath the Removal condition(s): pH electrode. Acknowledge execution of the "Clot removal procedure" Successful 1- or 2-point calibration

* NOTICE: This is a system message.

1026

Clot (pH) detection not possible

The pH of the rinse solution - stored during the latest calibration - was invalidated when either the Rinse solution was replaced, or when maintenance was performed on an electrode. A new pH value needs to be established by performing a calibration.

Perform a 1- or 2-point calibration. Removal condition(s): Successful 1- or 2-point calibration.

1027

Clot suspected beneath pH electrode

The difference Remove the pH electrode and between the pH of check the measuring chamber for the rinse solution, clots: clean the measuring measured during the chamber with a cotton stick sample measurement, moistened with distilled water. and that stored after Make sure no cotton fibers are the last calibration left in the measuring chamber exceeds the Clean pH electrode maximum allowable Remembrane Ref electrode drift. Replace pH electrode This indicates the Perform a 1- or 2-point calibration presence of a clot in the measuring chamber beneath the Removal condition(s): pH electrode. Acknowledge execution of the "Clot removal procedure" Successful 1- or 2-point calibration Action Press Analyzer status > Electrodes and others. Press Replace. Press Clots removed.

Acknowledging execution of a "Clot-removal procedure"

Step 1. 2. 3.

Technical documentation

Data in this document will be added to the manual next time it is updated.

Radiometer Medical ApS kandevej 21 2700 Brnshj Denmark www.radiometer.com

Table of contents

ABL800 FLEX

1.

Introduction

2.

What is what

3.

Installation and setup

4.

Sample measurements

5.

Quality control

6.

Calibration

7.

Replacements

Operators manual

8.

Disk functions

9.

Data management

10. Analyzer shutdown

11. Troubleshooting

12. Sampling

13. Specifications

14. Ordering information

15. Radiometer settings

Index - Date of issue

System performance The procedures described in this manual must be observed in order to ensure proper system performance, and to avoid hazards. Radiometer cannot provide or verify system performance characteristics if the system is not installed, used and maintained in accordance with Radiometer procedures or if accessories not meeting the specifications provided by Radiometer are used. Radiometer warrants that the data media on which the software included in the system is furnished is free from defects in material and workmanship under normal use for three (3) months from the date of delivery as evidenced by a copy of invoice or receipt. Third-party software and trademarks The ABL800 FLEX analyzers comprise the Microsoft WindowsXP Embedded, VxWorks and Sybase SQL Anywhere software. By using the system, you accept the terms of the Software License Agreement(s) of the provider(s) of the above software as shown in the End User License Agreement(s) in the analyzer start up picture and to the terms of the Microsoft WindowsXP Embedded End-User Agreement included in this manual. If you cannot accept the terms of the Software License Agreement(s), you should not use the system, but immediately contact your provider for a return of the system and a refund of the purchase price. Microsoft and Windows are trademarks of Microsoft Corporation. VxWorks is a registered trademark of WindRiver Systems Incorporated. Sybase SQL Anywhere is a registered trademark of Sybase Incorporated. Warranties and disclaimer Radiometer makes no warranties, express or implied, other than expressly stated. Any warranties expressly stated in this document are conditional upon the system being installed, used and maintained in accordance with Radiometer procedures, including that only accessories meeting the specifications provided by Radiometer are used. Radiometer disclaims any liability for system performance if the system is not installed, used and maintained in accordance with Radiometer procedures or if accessories not meeting the specifications provided by Radiometer are used. Further, Radiometer disclaims any liability for loss of data and direct, consequential or other damages, including loss of profit or loss of business, whether such claim for damages is based upon contract, negligence or tort (including strict liability), and even if Radiometer has knowledge of the possibility of the potential damage or loss.

Confidentiality The contents of this document shall not be reproduced or communicated to any third party without the prior written consent of Radiometer.

Changes This document is subject to change without notice and you are urged to contact Radiometer to verify whether the document has been changed. While every effort is made to ensure the correctness of the information provided in this document as changed from time to time, Radiometer disclaims any liability for errors and omissions.

Radiometer, the Radiometer logo, ABL, AQT, TCM, RADIANCE, PICO and CLINITUBES are trademarks of Radiometer Medical ApS. 2011 Radiometer Medical ApS. All rights reserved.

End-User License Agreement for the ABL800 FLEX Analyzers


T

You have acquired a device ("DEVICE") THAT INCLUDES SOFTWARE LICENSED BY Radiometer Medical ApS from Microsoft Licensing Inc. or its affiliates ("MS"). Those installed software products of MS origin, as well as associated media, printed materials and "online" or electronic documentation ("SOFTWARE") are protected by international intellectual property laws and treaties. The SOFTWARE is licensed, not sold. All rights reserved. IF YOU DO NOT AGREE TO THIS END-USER LICENSE AGREEMENT ("EULA"), DO NOT USE THE DEVICE OR COPY THE SOFTWARE . INSTEAD, PROMPTLY CONTACT THE SUPPLIER OF THE INSTRUMENT FOR INSTRUCTIONS ON RETURN OF THE UNUSED DEVICE(S) FOR A REFUND. ANY USE OF THE SOFTWARE, INCLUDING BUT NOT LIMITED TO USE ON THE DEVICE WILL CONSTITUTE YOUR AGREEMENT TO THIS EULA (OR RATIFICATION OF ANY PREVIOUS CONSENT). GRANT ON SOFTWARE LICENSE. This EULA grants you the following license: You may use the SOFTWARE only on the DEVICE. NOT FAULT TOLERANT. THE SOFTWARE IS NOT FAULT TOLERANT. RADIOMETER MEDICAL ApS HAS INDEPENDENTLY DETERMINED HOW TO USE THE SOFTWARE IN THE DEVICE, AND MS HAS RELIED UPON RADIOMETER MEDICAL ApS TO CONDUCT SUFFICIENT TESTING TO DETERMINE THAT THE SOFTWARE IS SUITABLE FOR SUCH USE. NO WARRANTIES FOR THE SOFTWARE. THE SOFTWARE is provided "AS IS" and with all faults. THE ENTIRE RISK AS TO SATISFACTORY QUALITY, PERFORMANCE, ACCURACY, AND EFFORT (INCLUDING LACK OF NEGLIGENCE) IS WITH YOU. ALSO, THERE IS NO WARRANTY AGAINST INTERFERENCE WITH YOUR ENJOYMENT OF THE SOFTWARE OR AGAINST INFRINGEMENT. IF YOU HAVE RECEIVED ANY WARRANTIES REGARDING THE DEVICE OR THE SOFTWARE, THOSE WARRANTIES DO NOT ORGINATE FROM, AND ARE NOT BINDING ON, MS. Note on Java Support. The SOFTWARE may contain support for programs written in Java. Java technology is not fault tolerant and is not designed, manufactured or intended for use or resale as online control equipment in hazardous environments requiring fail-safe performance, such as in the operation of nuclear facilities, aircraft, navigation or communication systems, air traffic control, direct life support machines or weapons systems, in which the failure of Java could lead directly to death, personal injury, or severe physical or enviromental damage. Sun Microsystems, Inc. has contractually obligated MS to make this disclaimer. No Liability for Certain Damages. EXCEPT AS PROHIBITED BY LAW, MS SHALL HAVE NO LIABILITY FOR ANY INDIRECT, SPECIAL, CONSEQUENTIAL OR INCIDENTAL DAMAGES ARISING FROM OR IN CONNECTION WITH THE USE OR PERFORMANCE OF THE SOFTWARE. THIS LIMITATION SHALL APPLY EVEN IF ANY REMEDY FAILS OF ITS ESSENTIAL PURPOSE. IN NO EVENT SHALL MS BE LIABLE FOR ANY AMOUNT IN EXCESS OF US. TWO HUNDRED FIFTY DOLLARS (US$250.00). Limitations on Reverse Engineering, Decompilation and Disassembly. You may not reverse engineering, decompile, or disassemble the SOFTWARE, except and only to the extent that such activity is expressly permitted by applicable law notwithstanding this limitation. SOFTWARE TRANSFER ALLOWED BUT WITH RESTRICTIONS. You may permanently transfer rights under this EULA only as a part of a permanent sale or transfer of the DEVICE, and only if the recipient agrees to this EULA. If the SOFTWARE is an upgrade, any transfer must also include all prior versions of the SOFTWARE. EXPORT RESTRICTIONS. You acknowledge that SOFTWARE is of US-origin. You agree to comply with all applicable international and national laws that apply to the SOFTWARE, including the US Export Administration Regulations, as well as end-user, end-use and country destination restrictions issued by US and other governments. For additional information on exporting the SOFTWARE, see http://www.microsoft.com/exporting/.

Contents
This manual contains the following topics.

1. Introduction ...................................................................................................... 1-1


Overview........................................................................................................................1-1 Names and intended use.................................................................................................1-2 Limitations of use and known interfering substances ....................................................1-4 Warning/Caution and Notices........................................................................................1-7 Symbols overview..........................................................................................................1-8

2. What is what ..................................................................................................... 2-1


Overview........................................................................................................................2-1 Analyzer - front..............................................................................................................2-2 Analyzer - rear ...............................................................................................................2-4 Measuring section ..........................................................................................................2-5 Inlet module ...................................................................................................................2-7 FLEXQ module..............................................................................................................2-8 Thermal printer ............................................................................................................2-10 Communication ports...................................................................................................2-11 Barcode reader .............................................................................................................2-13 AutoCheck module ......................................................................................................2-14 Screen elements ...........................................................................................................2-15 Menu structure .............................................................................................................2-24 Analyzer status.............................................................................................................2-28 Online aid facilities ......................................................................................................2-36 Sample counter ............................................................................................................2-39

3. Installation and setup ....................................................................................... 3-1


Overview........................................................................................................................3-1 Installation .....................................................................................................................3-2 Setup menu structure .....................................................................................................3-3 Analyzer security ...........................................................................................................3-6 Analysis setup ..............................................................................................................3-13 Patient reports ..............................................................................................................3-26 Calibration setup ..........................................................................................................3-32 Quality control setup....................................................................................................3-35 Replacement setup .......................................................................................................3-48 Parameters and input setup ..........................................................................................3-55 Analyzer settings..........................................................................................................3-62

Contents

ABL800 FLEX Operator's Manual

Communications ..........................................................................................................3-68 Printers .........................................................................................................................3-77 Disk Functions setup....................................................................................................3-80 Corrective actions ........................................................................................................3-83 Miscellaneous setup .....................................................................................................3-86

4. Sample measurements...................................................................................... 4-1


Overview........................................................................................................................4-1 General information .......................................................................................................4-2 Immediately before analysis ..........................................................................................4-9 Measurements with FLEXQ ........................................................................................4-10 Introducing a blood sample..........................................................................................4-12 Introducing a pleura sample.........................................................................................4-15 Introducing an expired air sample................................................................................4-16 Patient identification ....................................................................................................4-17 Patient result ................................................................................................................4-21

_
Calculation of FShunt and ctO2(a - V )........................................................................4-25 Patient result messages ................................................................................................4-26

5. Quality control .................................................................................................. 5-1


Overview........................................................................................................................5-1 General information .......................................................................................................5-2 Preparing a control solution ...........................................................................................5-4 Manual quality control measurement.............................................................................5-6 AutoCheck measurement ...............................................................................................5-7 Quality control identification .........................................................................................5-8 Quality control result ...................................................................................................5-10 Quality control result messages ...................................................................................5-15

6. Calibration ........................................................................................................ 6-1


Overview........................................................................................................................6-1 General information .......................................................................................................6-2 Unscheduled calibrations ...............................................................................................6-4 Interrupted, pending or expired calibrations..................................................................6-5 tHb calibration ...............................................................................................................6-6 Calibration result............................................................................................................6-8 Calibration result messages..........................................................................................6-10

7. Replacements .................................................................................................... 7-1


Overview........................................................................................................................7-1 General information .......................................................................................................7-2 Replacing membranes or electrodes ..............................................................................7-6 Replacing pump tubes....................................................................................................7-9

ABL800 FLEX Operator's Manual

Contents

Replacing inlet gasket unit and inlet probe..................................................................7-12 Replacing waste container, fan filter, printer paper .....................................................7-14 Replacing solutions and gases .....................................................................................7-16 Refilling the AutoCheck carousel................................................................................7-19 Automatic auxiliary programs .....................................................................................7-20 Decontamination and Protein Removal programs .......................................................7-21 Cleaning the analyzer...................................................................................................7-23 List of references .........................................................................................................7-25

8. Disk Functions .................................................................................................. 8-1


Overview........................................................................................................................8-1 General information .......................................................................................................8-2 Creating a WDC report ..................................................................................................8-4 Backing up all data ........................................................................................................8-6 Restoring all data ...........................................................................................................8-8 Exporting data logs ........................................................................................................8-9 Importing/exporting archives.......................................................................................8-11 Saving setup.................................................................................................................8-13 Loading/restoring setup ...............................................................................................8-14

9. Data management............................................................................................. 9-1


Overview........................................................................................................................9-1 General information .......................................................................................................9-2 Patient Results Log ........................................................................................................9-4 Patient Profiles Log .......................................................................................................9-7 Quality Control Log.....................................................................................................9-11 Calibration Log ............................................................................................................9-17 Activity Log.................................................................................................................9-20 Replacement Log .........................................................................................................9-23 Archived data logs .......................................................................................................9-24 RADIANCE browser (optional) ..................................................................................9-26

10. Analyzer shutdown......................................................................................... 10-1


Overview......................................................................................................................10-1 General information .....................................................................................................10-2 Standby mode ..............................................................................................................10-3 Full waste container .....................................................................................................10-5 Temporary Shutdown ..................................................................................................10-6 Long Term Shutdown ..................................................................................................10-8

Contents

ABL800 FLEX Operator's Manual

11. Troubleshooting.............................................................................................. 11-1


Overview......................................................................................................................11-1 General information .....................................................................................................11-2 Forced Hold causes......................................................................................................11-5 Analyzer messages.......................................................................................................11-7 Fluid transport troubleshooting procedure.................................................................11-74 Inlet probe troubleshooting procedure .......................................................................11-75 Inlet troubleshooting procedure .................................................................................11-76 Leak troubleshooting procedure ................................................................................11-77 Electrode troubleshooting procedures........................................................................11-78 Pump troubleshooting procedure ...............................................................................11-80 Fluid transport system description .............................................................................11-81

12. Sampling.......................................................................................................... 12-1


Overview......................................................................................................................12-1 Causes of errors in preanalytical phase........................................................................12-2 Preparation prior to arterial/venous sampling..............................................................12-5 Preparation prior to capillary sampling........................................................................12-7 Sampling devices .........................................................................................................12-8 Storage and preparation prior to analysis.....................................................................12-9 Sampling procedures..................................................................................................12-12 References..................................................................................................................12-15

13. Specifications .................................................................................................. 13-1


Overview......................................................................................................................13-1 Measured parameters ...................................................................................................13-2 Input parameters ..........................................................................................................13-5 Derived parameters ......................................................................................................13-6 Sample handling ..........................................................................................................13-8 Calibration and maintenance programs......................................................................13-11 Analyzer requirements ...............................................................................................13-13 Analyzer specifications..............................................................................................13-14 Approvals and patents................................................................................................13-16

14. Ordering information..................................................................................... 14-1


Overview......................................................................................................................14-1 Analyzer accessories....................................................................................................14-2 Quality control .............................................................................................................14-5 Sampling devices .........................................................................................................14-7

15. Radiometer settings ........................................................................................ 15-1


Overview......................................................................................................................15-1 Radiometer default settings..........................................................................................15-2

ABL800 FLEX Operator's Manual

Contents

Contents of setup settings ..........................................................................................15-15 Calibration verification ..............................................................................................15-18 Interfacing facilities ...................................................................................................15-20

Index Date of issue

Contents

ABL800 FLEX Operator's Manual

1. Introduction Overview
Introduction The chapter briefly describes the intended use of the analyzer, lists all measured parameters and the substances known to interfere with the measurements, and explains the different notices that appear in the manual. Throughout this manual, "ABL800 FLEX analyzer" is used for all ABL8xx FLEX analyzers, i.e.: ABL837/835/830/827/825/820/817/815/810/805 and ABL810 BG only. The abbreviation "ABL8x7 FLEX analyzer" is used for the ABL837/27/17 FLEX analyzers throughout this manual. Contents This chapter contains the following topics. Names and intended use................................................................................... Limitations of use and known interfering substances ...................................... Warning/Caution and Notices .......................................................................... Symbols overview........................................................................................... 1-2 1-4 1-7 1-8

1. Introduction

ABL800 FLEX Operator's Manual

Names and intended use


Names Proprietary name: Common name: ABL800 FLEX blood gas, oximetry, electrolyte and metabolite analyzer. Blood gas, oximetry, electrolyte and metabolite measuring system.

Intended use

The ABL800 FLEX analyzers are intended for: In Vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) in vitro testing of samples of expired air for the parameters pO2 and pCO2 in vitro testing of pleura samples for the pH parameter. The following parameters can be measured on blood: Parameter group pH/blood gas: pH (acidity) pCO2 (carbon dioxide tension) pO2 (oxygen tension) Oximetry: ctHb (total hemoglobin concentration) sO2 (oxygen saturation) FO2Hb (fraction of oxyhemoglobin in total hemoglobin) FCOHb (fraction of carboxyhemoglobin in total hemoglobin) FHHb (fraction of deoxyhemoglobin in total hemoglobin) FMetHb (fraction of methemoglobin in total hemoglobin) FHbF (fraction of fetal hemoglobin) Electrolytes: cK+ (potassium ion concentration) cNa+ (sodium ion concentration) cCa2+ (calcium ion concentration) cCl (chloride ion concentration) Metabolites: cGlu (D-glucose concentration) cLac (L(+)-lactate concentration) ctBil (concentration of total bilirubin, measurement in plasma is possible, see Limitations of use later in this chapter) cCrea (concentration of creatinine, measurement on plasma and serum possible, see Limitations of use later in this chapter)
Continued on next page

Parameters

1-2

ABL800 FLEX Operator's Manual

1. Introduction

Names and intended use, Continued


Intended use (continued) The following parameters can be measured on pleura samples: Parameter group Measured value pH (acidity) Parameters

The following parameters can be measured on expired air samples: Parameter group Measured values pO2 (oxygen tension) Derived parameters are listed in chapter 13: Specifications and described in detail in the Reference Manual, chapter 6. The analyzer should be used by personnel who have received special education and training with regard to procedures utilizing in vitro diagnostic medical devices. Parameters pCO2 (carbon dioxide tension)

Requirement to the operator

Measurements Animal blood has not been tested on the ABL800 FLEX analyzer. Some on animal blood components in animal blood differ from those in human blood, and variations in the composition of blood from different animal species may also exist. FLEXMODE This mode allows you to analyze a blood sample of 35 L and higher up to the maximum volume accepted by your analyzer. Depending on the available sample volume, the FLEXMODE provides the highest number of parameters: from all available to as many as reliably possible. This mode is not available in the ABL8x7 FLEX analyzers. Measurements Pleura pH can be measured on pleural fluids. Corrections are present in the device. on pleural fluids Other fluids mode All parameters available on your ABL800 FLEX analyzers can be measured on fluids other than heparinized human whole blood NOTICE: Before using this mode you must establish "user-defined corrections" for each parameter used, on the fluid in question. The corrections assume a linear correlation between the measured value and the reference instrument. The data used for establishing "user-defined corrections" have to cover the desired measuring range. If no user-defined corrections are entered, you will measure in this mode as if on heparinized human whole blood. FLEXQ module The FLEXQ module can accommodate up to three samplers simultaneously. It reads a sampler's barcode, mixes the sample and transports the sampler to the inlet for aspiration and analysis without any further assistance from the operator. Results can be delivered via FLEXLINK (for information refer to the RADIANCE Installation and Setup Manual). NOTICE: The model ABL810 can also be ordered without oximetry parameters: as ABL810 BG only. 1-3

1. Introduction

ABL800 FLEX Operator's Manual

Limitations of use and known interfering substances


Limitations of use The following limitations should be taken into consideration: The ABL800 FLEX analyzers are designed for measurements of adult and fetal hemoglobin with normal spectrum characteristics. Some spectra deviate from the normal characteristics, e.g. for certain hemoglobinopathies and the ABL800 FLEX analyzers do not compensate for this.

CAUTION - Fulfillment of user-specific analytical needs The user should review the analyzer performance data to assure that the performance fulfills the user-specific analytical needs. WARNING Clinical decisions The validity of the test results from this instrument must be carefully examined by a clinician and related to the patients clinical condition, before any clinical decisions are taken on the basis of the test results. CAUTION - Risk of erroneous results Always meticulously follow the sampling procedures described in chapter 12: Sampling. Failure to follow these procedures may introduce clots or air bubbles in the sample and yield erroneous results. NOTICE: Bilirubin measurements on plasma and creatinine measurements on plasma and serum need to be measured in Other fluids mode, as the other modes are intended for measurement on human whole blood only. Corrections for Creatinine can be found in the Reference Manual. FHbF measurement: The uncertainty in FHbF measurements exceeds the level required to measure normal HbF levels in the adult range (FHbF reference range is 0-1 %). The following substances are known to affect or interfere with measurements on the ABL800 FLEX analyzers. CAUTION Known interfering substances Substance Halothane (anesthetic) Lipid therapy (treatment) Interference May give unreliable pO2 results. In OXI measurements. After measurement on blood from a patient who has received lipid therapy it may be necessary to clean the analyzer using the Cleaning program. Methylene Blue, HiCN (medication)
Anions: Br , I , S 2 and ClO 4 (drugs)

In OXI measurements.

Erroneously high cCl- results.

Continued on next page

1-4

ABL800 FLEX Operator's Manual

1. Introduction

Limitations of use and known interfering substances,


Continued
Substance CAUTION Known interfering substances (continued) Interference

Anticoagulants (sampling)

Anticoagulants that contain sodium salts will give erroneously high cNa+ results. Sodium fluoride with or without EDTA and oxalate (di Na) influence cGlu results. Sodium fluoride gives erroneously high cNa+ and low cCa2+, cGlu and cLac results. Tri sodium citrate influences cNa+, cK+ and cGlu results. Thus Radiometer recommends the exclusive use of heparin as anticoagulant. Solutions containing organic preservatives may damage the ionselective membranes of the cK+ and cGlu electrodes when introduced into the analyzer. Do not use EDTA, as it leads to erroneous pH, pCO2, cNa+, cK+ and cCa2+ results. Use of EDTA will also affect subsequent measurements on the Ca electrode and it will reduce the lifetime of this electrode. Erroneously high cGlu and cLac measurements.

Thiocyanic acid (degradation product from treatment with nitroprussid. Also produced in thiosulphate treatment of cyanide poisoning) Glycolic acid (ethyleneglycol degradation product) Insufficiently stabilized blood. Caustic fluids (e.g. strong acids or bases, detergents, etc.). Fluids that precipitate. Fluids that affect the sensor enzymes. Fluids that form complexes with the analyzer solutions (calcium). High viscosity fluids. Hydrophobic fluids. Reactive fluids.

Erroneously high cLac measurements. Other fluids mode allows you to measure on fluids other than heparinized human blood. Be aware that some substances, such as listed in the left column, measured in the Other fluids mode may damage the instrument or the electrodes. This can affect the subsequent measurement on human blood or quality control solutions.

Continued on next page

1-5

1. Introduction

ABL800 FLEX Operator's Manual

Limitations of use and known interfering substances,


Continued

Substance CAUTION Known interfering substances (continued)

Interference

Carboxymethyl cellulose (CMC) Some auto-venting arterial blood samplers contain carboxymethyl cellulose (CMC) in the porous vent. CMC can dissolve into the sample and give erroneously low cCa2+ results. Therefore we recommend Radiometer accessories together with our analyzers, e.g. the safePICO arterial blood sampler which is specifically designed to minimize sample contamination with CMC. Galactose, glucosamine, maltose, Erroneously high cGlu measurements. mannose, xylose

For detailed information see Interference Tests in chapter 5 of the ABL800 FLEX Reference Manual.

1-6

ABL800 FLEX Operator's Manual

1. Introduction

Warning/Caution and Notices


Definitions

Throughout the manual, the various procedures may contain operational cautions and warnings, which are important and should be read carefully before performing the related procedures. The manual also contains a number of NOTICES. The following table indicates the type of information given in Warnings, cautions and notes.
Symbol WARNING Explanation

A warning alerts the reader about a situation, which, if not avoided, could result in death or serious injury. It may also describe potential serious adverse reactions and safety hazards. The designation of a hazard alert as a "warning" is reserved for the most significant problems. The term WARNING is generally used as signal word for this type of hazard alert.
CAUTION

The term precaution is used for the statement of a hazard alert that warns the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or the patient or damage to the equipment or other property. It may also be used to alert against unsafe practices. This includes the special care necessary for the safe and effective use of the device and the care necessary to avoid damage to a device that may occur as a result of use or misuse. The word CAUTION is generally used as signal word for a precaution statement.
NOTICE

Notices give practical information.

1-7

1. Introduction

ABL800 FLEX Operator's Manual

Symbols overview
List of symbols

The symbols below are used by Radiometer.


Symbol Explanation

Biohazard

Expiry date See instructions for use For In Vitro Diagnostic Use Lot number Sensitive to light. Store in a dark place. Code number For single use only Sterilized by ethylene oxide Sterilized by irradiation Storage temperature from 2 C to 32 C (36 F to 90 F)
Waste of Electrical and Electronic Equipment (WEEE)

The symbol indicates that: Radiometer Medical ApS and its distributors within the European Union (EU) and associated states have taken the necessary steps to comply with the directive, 2002/96/EC on waste electrical and electronic equipment (WEEE).

The instrument, when reaching its end of life, must be collected and recycled separately from other waste according to national requirements. Please contact your local Radiometer distributor for instructions.

Environmental implications: WEEE contains materials that are potentially hazardous to the environment and to human health.

1-8

2. What is what Overview


Introduction The ABL800 FLEX analyzer is a complete unit comprised of several different modules each performing a specific function and controlled by comprehensive software. The modules are collected into well-defined sections according to their related function. This chapter describes the basic parts of the ABL800 FLEX analyzer and its software. Contents This chapter contains the following topics. Analyzer - front................................................................................................ Analyzer - rear ................................................................................................. Measuring section ............................................................................................ Inlet module ..................................................................................................... FLEXQ module................................................................................................ 2-2 2-4 2-5 2-7 2-8

Thermal printer ................................................................................................ 2-10 Communication ports ....................................................................................... 2-11 Barcode reader ................................................................................................. 2-13 AutoCheck module........................................................................................... 2-14 Screen elements................................................................................................ 2-15 Menu structure ................................................................................................. 2-24 Analyzer status ................................................................................................. 2-28 Online aid facilities .......................................................................................... 2-36 Sample counter................................................................................................. 2-39

2. What Is what

ABL800 FLEX Operator's Manual

Analyzer - front
Parts and functions The components of the analyzer front (with covers) is shown below.

Touch screen

Cover with window

Capillary inlet Thermal printer Syringe inlet Left cover Right cover Waste container Cleaning Solution

Rinse Solution FLEXQ module Barcode reader

AutoCheck module

Cal 1 solution

Cal 2 solution

Part Color touch screen Thermal printer Left cover Waste container Rinse Solution FLEXQ module Barcode reader AutoCheck module Cal 1 Solution Cal 2 Solution Cleaning Solution Right cover Syringe inlet flap Capillary inlet flap Cover with window

Function 10.4" LCD for operation and management of the analyzer. For automatic printout of data. To access the waste container/Rinse Solution. For waste collection. A sensor detects when container is full, and a message is displayed on the screen. For rinsing the liquid transport system after various analyzer activities. See FLEXQ module in this chapter. See Barcode reader in this chapter. See AutoCheck module in this chapter. For performing 1- and 2-point calibrations. For performing 2-point calibrations. For cleaning the liquid transport system of lipid deposits. To access solutions and pumps see overleaf. Lift to introduce syringe samples and quality control solutions. Lift to introduce capillary samples. See Measuring Section in this chapter.
Continued on next page

2-2

ABL800 FLEX Operator's Manual

2. What Is what

Analyzer - front, Continued


Parts and functions (continued) The components of the analyzer front (without covers) is shown below.

pH/BG module El/Met module Met II module CD-ROM drive Oximetry module Inlet module Solution pump Waste pump

Magnets

Part Inlet module CD-ROM drive

Function Accepts samples from a syringe/test tube or a capillary see Inlet Module in this chapter. For storing data on a CD-RW disk or retrieving data from a CD (e.g. installation software). CAUTION Installation of correct software Install only software which is strictly intended for use with the analyzer. Installation of other types may affect analyzer performance.

pH/Blood Gas (BG) module Electrolyte/Metabolite (El/Met) module Met II module

Measures pH, pO2, pCO2 and cCl. See measuring component detail. Measures cCa2+, cK+, cNa+, cGlu and cLac. See measuring section detail. Measures creatinine. (Only available for the ABL8x7 FLEX analyzers) Measures ctHb, sO2, FHHb, FO2Hb, FCOHb, FMetHb, FHbF and ctBil. See measuring component detail. Transports solutions through the liquid transport system. Transports liquid to the waste container. Hold the covers in place.

Oximetry (Oxi) module Solution pump Waste pump Magnets

2-3

2. What Is what

ABL800 FLEX Operator's Manual

Analyzer - rear
Parts and functions The parts and components of the analyzer's rear panel are shown below.

Gas 1 cylinder Gas 2 cylinder

Communication ports

Fan

Power switch

Power socket Gas cylinder sockets

Part Gas 1 cylinder Gas 2 cylinder Fan Gas cylinder socket Communication ports Power switch Power socket

Function Contains a gas mixture of 5.61 % CO2, 19.76 % O2; balance 74.64 % N2. Contains a gas mixture of 11.22 % CO2, < 0.04 % O2; balance > 88.74 % N2. For cooling internal components. For mounting the gas cylinders. See section Communication Ports in this chapter. For turning the analyzer on (position I) and off (position O). For connecting the power cord. In USA: If source is 120 V, use Radiometer cord 615-403 with parallel blade attachment plug. If source is 240 V, use Radiometer cord 615-405 with tandem blade attachment plug.

2-4

ABL800 FLEX Operator's Manual

2. What Is what

Measuring section
Parts and functions The modules and components of the measuring section are shown below.

Membrane valves

Electrode module pumps El/Met module Latch Met II module

Oximetry module pH/BG module

Sample path

Part Membrane valves Electrode module pumps Latch

Function Control the flow of gas and solutions into the measuring modules. Transport solutions and gas mixtures through the measuring modules. Establishes electric contact between the electrode in the measuring chamber and the electrode amplifier. To access an electrode, push inwards on the upper portion of the latch, and lift upwards.

Sample path pH/BG module El/Met module

Transports samples, solutions and gas mixtures. Contains the measuring chamber with the pH, reference, pCO2, pO2 and cCl electrodes. Contains the measuring chamber with the cCa2+, cK+, cNa+, cGlu, and cLac electrodes (a filter is installed for the cGlu and cLac electrodes). Contains the measuring chamber with the cCrea A and cCrea B electrodes (Only available for the ABL8x7 FLEX analyzers) Contains the hemolyzer with glass cuvette, spectrophotometer for measuring ctHb, sO2, FHHb, FO2Hb, FCOHb, FMetHb, FHbF, ctBil on 128 wavelengths, and the lamp unit.
Continued on next page

Met II module

Oximetry module

2-5

2. What Is what

ABL800 FLEX Operator's Manual

Measuring section, Continued


Electrode locations The electrodes are located in the electrode modules as follows:
-

cCl electrode pCO2 electrode pO2 electrode Reference electrode

cLac electrode cGlu electrode


+

cNa electrode cK electrode


2+ +

pH electrode

cCa

electrode

cCrea A electrode cCrea B electrode

The table below describes each electrode. Module pH/BG Electrode cCl pCO2 pO2 Reference pH El/Met cLac cGlu cNa cK Met II
+ 2+ +

Type E744 E788 E799 E1001 E777 E7077 E7066 E755 E722 E733 E8088 E8089

cCa

cCrea A cCrea B

NOTICE: An orange filter is placed in the measuring chambers for the cGlu, cLac, cCrea A and B electrodes.

2-6

ABL800 FLEX Operator's Manual

2. What Is what

Inlet module
Parts and functions The inlet module accepts samples from a syringe/test tube or a capillary (the syringe and capillary inlet flaps are removed on the picture). Syringe and capillary inlet flaps are interlocked so that only one may be opened at a time.

Probe attachment

Preheater

Inlet probe Capillary inlet flap post Syringe inlet flap post Capillary inlet Inlet gasket unit Syringe inlet

Part Probe attachment Preheater Inlet probe Capillary inlet flap post Syringe inlet flap post Capillary inlet Syringe inlet Inlet gasket unit

Function Secures inlet probe. Open the clip to remove probe. Heats all samples and gases to 37 C. Automatically moves into the syringe to aspirate the required sample. Holds the inlet flap for capillary samples. Holds the inlet flap for syringe samples. Receives sample from a capillary. Receives sample from a syringe and test tube. Provides transfer of samples from a sampling device to the inlet probe; can be removed.

2-7

2. What Is what

ABL800 FLEX Operator's Manual

FLEXQ module
Parts and functions FLEXQ module for automatic transport of samplers to the inlet is shown below.

FLEXQ position before or after analysis:

Barcode reader Indicator diodes

Samplers

Sampler tray FLEXQ position during analysis:

Continued on next page

2-8

ABL800 FLEX Operator's Manual

2. What Is what

FLEXQ module, Continued


Parts and functions (continued) The components are as follows: Part FLEXQ sampler tray Function FLEXQ sampler tray moves to the position for measurement by means of a step motor (not shown). The samplers used in the sampler tray should be safePICO sampler with the safeTIPCAP. Indicator diodes Change the color from green to yellow when the optical switch detects the sampler placed in the slot. Laser barcode reader reads the sampler barcode. A short beep indicates that the barcode has been read. You can scan the Sampler ID on the analyzers scanner and then place it in the sample tray or you can place the sampler in the sampler tray and let the FLEXQ scanner read the sampler barcode. The sampler tray cover has three slots for the samplers; the slots have optical switches for detecting the sampler. The cover can be removed (as shown) for cleaning.

Barcode reader

Sampler tray cover

Mixer tray

The mixer rotates and moves the ball in the sampler in order to mix the blood sample. The sample is adequately mixed after 7 seconds. The mixer tray moves into position under the sampler immediately before the measurement.

2-9

2. What Is what

ABL800 FLEX Operator's Manual

Thermal printer
Parts and functions The components of the thermal printer (with the cover open) are shown below.
Release lever Drive roller Paper guide

Paper roll holder

Cover

Part Release lever

Function Push the lever fully back to lift the drive roller from the printer head and paper guide in order to adjust paper alignment. Remember to return the lever to its original position (as shown) in order to print.

Drive roller Paper guide Paper roll holder Cover

Feeds the thermal paper through the printer. Located behind the drive roller, it is between this and the drive roller that paper is fed. Holds the paper roll in place. Open to replace the paper roll. Contains instructions on how to make a replacement. Close the cover when a new paper roll has been mounted in the printer.

2-10

ABL800 FLEX Operator's Manual

2. What Is what

Communication ports
Ports and functions The following communication ports are available:

Mouse Keyboard

USB ports (two)

Display unit

Printer port

Network connector

Ext. VGA monitor

Fuse for printer unit

Serial port (COM2) Main fuse

The functions of the communication ports are as follows: Port Display unit port Mouse port Keyboard port Serial port (COM2) Network Printer port Function For connection to the analyzers display PS/2 connector used for the connection of a standard mouse (user supplied). PS/2 connector used for connection of a keyboard (user supplied). 9-pin serial port used for HIS/LIS communication or connection of an external barcode reader. RJ45 ethernet interface connection to network. Parallel port for connection to a local printer
Continued on next page

2-11

2. What Is what

ABL800 FLEX Operator's Manual

Communication ports, Continued


Ports and functions (continued) Port External VGA monitor port USB (2 ports) Function 15-pin connector for external monitor (optional; is enabled by Radiometer service representative). For connecting USB devices, e.g. a removable drive. USB keyboard supported. Other USB devices need an approved XP driver (user-supplied mouse or modem, etc.). The compartment contains 1 protective fuse: 5 20 mm, 4 Amp, Slow blow (T4AL) (order code number 450-035). The compartment contains 2 protective fuses: 5 20 mm, 4 Amp, High break (T4AH). Type Shurter No. 0001.2510 (order code number 450-144). WARNING Risk of fire Fire hazard. Replace fuse as marked.

Fuse for printer unit Main fuse

2-12

ABL800 FLEX Operator's Manual

2. What Is what

Barcode reader
Barcoded items The following items have barcodes that can be read into the analyzer: Item All solution containers QC ampoules tHb calibration ampoules Electrodes Electrode membranes Gas cylinders Pump tubing Fan filter Inlet gasket Location of barcode Label on the container Insert in the box containing the ampoules Insert in the box containing the ampoules Label on the box containing the electrode Label on the box containing the membrane unit Label on the cylinder Label on the tubing packaging Label on the box containing the filter Label on the box containing the inlet gasket

In addition to the items above, you can enable every text box on the Patient Profile, Patient ID, Patient Result, QC ID and Recording Fluid Replacement screens where it is possible to enter a barcode refer to Miscellaneous Setup in chapter 3: Installation and setup. Reading in a barcode Hold the barcode of the item you wish to read in parallel to the sensor part of the barcode reader. A short beep from the barcode reader indicates that the information has been read in successfully.

Sensor part

2-13

2. What Is what

ABL800 FLEX Operator's Manual

AutoCheck module
Parts and functions The AutoCheck module is shown below.

AutoCheck carousel

To close

To open

Retractable cover

Part Retractable cover

Function Gives access to the carousel. You can either slide the carousel cover open or press Menu - Analyzer Status - AutoCheck - More - Open Module. Make sure that no objects prevent the cover from opening freely.

AutoCheck carousel

Contains the AutoCheck ampoules packed according to the optimal packing list see Refilling the AutoCheck carousel in chapter 7: Replacements.

2-14

ABL800 FLEX Operator's Manual

2. What Is what

Screen elements
Ready screen Operation and management of the ABL800 FLEX analyzers is performed via a touch screen. Almost all commands to the analyzer come from the touch of a button or area on the screen. The touch screen is divided into three sections:
Top section

Center section

Bottom section

For the analyzers without the FLEXQ module, the Ready screen looks as follows:

The Ready screen appears automatically if the touch screen remains idle for more than 3 minutes. The screen intensity reduces by 50 % when the analyzer is not used. Top section The status bar varies slightly, according to which screen you are in and the status of the analyzer. Example:

The status describes the current task of the analyzer (e.g. calibration, measurement, etc.) or its status (e.g. ready for use, on hold, etc.).
Continued on next page

2-15

2. What Is what

ABL800 FLEX Operator's Manual

Screen elements, Continued


Top section (continued) The time bar is present only when the analyzer is performing a task; it follows the progress of the current task. For example:

The Stop button in progress.

is visible only when it is allowed to interrupt an activity

The parameter bar lists all measured parameters available and activated on your analyzer. It allows you to judge the parameter status at a glance before you have to perform a measurement. Green parameter: Parameter status is okay; no problem is detected on the given measuring channel. Yellow parameter: Error associated with the given parameter during last calibration or quality control measurement. The parameter is unreliable and will have ? (if requested in Corrective Actions) in front of the result. Serious error associated with the given measuring channel. Parameter cannot be used at all and will be displayed as ..; or a parameter was repressed in the Parameter Setup program due to errors during last calibration or quality control measurement.

Red parameter:

NOTICE: A parameter disabled in the Parameter Setup program will be removed from the parameter bar see Analysis Setup Programs (Disabled versus Deselected Parameters) in chapter 3.
Continued on next page

2-16

ABL800 FLEX Operator's Manual

2. What Is what

Screen elements, Continued


Center section The center section is the main information and interaction area of which there are many different types. The center section of the Ready screen is described part by part, starting from the top (for the analyzers without FLEXQ module the center part contains the analyzer type and if entered in Miscellaneous Setup (see chapter 3) analyzer message). This part of the Ready screen contains the following elements:

Analyzer name The messages to the operator regarding the FLEXQ activities Urgent Manual Sample request button. The button allows you to book an inlet for an urgent measurement while the analyzer performs a task. The button is grayed out when the analyzer is ready for any measurement. This part of the Ready screen contains the following elements:

"Slot #": condition of each sampler tray slot (1, 2 and 3) is indicated with the arrows that reflect the indicator diodes in the sampler tray: Green Blinking green Place the sampler into the slot Yellow or remove it from the slot. Wait for the sampler to be identified. Blinking yellow Do not remove the sampler from the slot. Remove the sampler from the slot.

"Sampler ID": appears when the sampler is placed in the sampler tray slot and is registered. The name of this column can be changed in the Sample Logistics Setup (see Analysis Setup in chapter 3 of this manual). "Time to Result": shows time left till a result will be ready; the clock indicates that the sample age has exceeded settings selected in Sample Logistics setup. "Status": gives information about a sampler in the slot: Slot empty Ready for processing Processing Completed A slot is ready to accept a sampler. Sampler has been registered in the slot. Patient and sampler data are being processed. Measurement is completed and the result can be viewed.
Continued on next page

2-17

2. What Is what

ABL800 FLEX Operator's Manual

Screen elements, Continued


Center section (continued) Patient data can be edited.

Result can be viewed.

This part of the Ready screen is reserved for the analyzer messages and is entered in the Miscellaneous Setup (see chapter 3 in this manual):

Other screen types and their elements are described on the following pages. Icon and screen name identify each screen. Example:
Icon Screen name

The icon and title of a screen are the same as the button that gives access to that screen.
Continued on next page

2-18

ABL800 FLEX Operator's Manual

2. What Is what

Screen elements, Continued


Center section (continued) Text boxes show a text in the form of lists, items, tables, etc. Various types of text boxes and navigation in them are given on the next page. The first line is highlighted. To highlight another line, touch it on the screen or use up and down arrow scroll keys displayed on the screen beside a relevant text box. A scroll bar scrolls a text box horizontally when the text box extends beyond the area available on the screen. A single-arrow scroll button highlights one item at a time upward or downward. A double-arrow scroll button highlights an item at the top or the bottom of the screen. The first line is highlighted. To highlight another line, use the up and down arrow buttons or touch this line in the box. Enter the data using the screen keypad or keyboard.

Scroll bar

Scroll keys

In this text box you can select one of the predefined options. Use the up and down arrow buttons to select an item. NOTICE: If the text box already contains an entry, it will be overwritten and cannot be retrieved.

Text boxes are designated in quotation marks, e.g. "Operator". The same applies to the other elements on the center section of the screen: names of the columns (e.g. "Status"), input fields (e.g. "Draw time"), etc.
Continued on next page

2-19

2. What Is what

ABL800 FLEX Operator's Manual

Screen elements, Continued


Center section (continued) Check buttons allow you to enable/disable or select/deselect an item on the screen. For example: A function is selected (e.g. acoustic signal if the inlet remains open) or activated (e.g. communication with the RADIANCE system) A function is deselected or deactivated. Parameter is selected.

Parameter is deselected.

Screen keypad allows the entry of numerical data on the screen.

Numerical keys from 0 to 9.

Decimal point.

The buttons listed below have the following functions: Press to confirm a numerical entry and to highlight the next line in the text box.

Press to delete a character (from right to left) or to type in the new data (the box is cleared as soon as the first character is typed). Press to display the full alphanumerical keyboard.

Continued on next page

2-20

ABL800 FLEX Operator's Manual

2. What Is what

Screen elements, Continued


Center section (continued) Screen keyboard is used for entering both alphabetical and numerical data see the description on the next page.
Activated text box

The keyboard functions as a normal keyboard with alphabetical and numerical characters. The typed text appears in the activated text box. Enter Press to confirm the entry and to return to the previous screen. Backspace Press to delete characters from right to left. Press to cancel any changes made in the activated text box and to return to the previous screen Press to delete an entry in the activated text box.

Shift Press to type a capital letter or symbol. Press to lock the keyboard in order to type capital letters and symbols on the numerical keys. When activated, indicator is green. Up/down/left/right arrows move the cursor in the activated text box in order to edit a text.
Continued on next page

2-21

2. What Is what

ABL800 FLEX Operator's Manual

Screen elements, Continued


Buttons Each button has an icon and a name placed on it. When pressed, it opens a screen or a menu.

The buttons are designated in bold italics in this manual, e.g. Menu, Utilities, etc. The buttons displayed in full color can be activated. A button in a weaker color, i.e. grayed-out, is temporarily inactive. The buttons at the top and the bottom of the screen can be selected in the Setup program Access Profiles (in the Analyzer Security menu) together with the access profiles for each operator for detailed information see the description in chapter 3: Installation and setup. Note the functions of the following buttons: Returns you to the previous screen in the same program, e.g. in the Patient Results log, it will return you from the Patient Identification screen to the Patient Result screen. Returns you to the Ready screen.

Information bar The information bar is placed in the lower right corner of the screen. and icons

Continued on next page

2-22

ABL800 FLEX Operator's Manual

2. What Is what

Screen elements, Continued


Information bar The following icons are available on the information bar: and icons (continued) Icon Function Shows the current time in the selected format.

Shows that an analyzer is connected to the RADIANCE system. Shows that an analyzer is connected to the QA Portal.

Shows the time of the next scheduled calibration. The time changes when a calibration has been completed. Shows the time of the next scheduled QC measurement. The time changes when a QC measurement has been completed. Shows that the remote operator is connected. Click to make more icons visible.

Click to reduce the number of visible icons to a clock.

2-23

2. What Is what

ABL800 FLEX Operator's Manual

Menu structure
When the analyzer is taken into use, only the following limited menu is available: Menu at analyzer startup Press Menu.

Log on Analyzer Status

Last patient result Help

Tutorials

Calibration Progs:
1 Point Calibration 2 Point Calibration

Start Progs

Data Logs

Auxiliary Progs:
Rinse Cleaning Protein Removal Decontamination Utilities

Patient Results Log Patient Profile Log QC Log Calibration Log Activity Log

AutoCheck Programs Run High Crea Check


(only the ABL8x7 analyzers)

Standby Temporary Shutdown

Allowed actions The following actions are allowed at analyzer startup: at analyzer startup To perform a measurement To perform a replacement To call a calibration To edit data in the data logs The menu above and the scope of actions are suitable for those users whose activities include performing measurements and occasional replacements.
Continued on next page

To start an AutoCheck measurement

2-24

ABL800 FLEX Operator's Manual

2. What Is what

Menu structure, Continued


Entering standard password To access the complete menu, do the following: Step 1. 2. Action Press Menu on the Ready screen. Then click on Logon. Type in the standard password: 123456.
Standard password:

123456.

See chapter 3: Installation and setup, Analyzer Security, in this manual for further information about the logon possibilities. Confirm with Enter. 3. Press Menu to access the complete menu see the next section.

The access possibilities for each user of the analyzer and their passwords are entered in Setup program Access Profiles see chapter 3: Installation and setup in this manual.
Continued on next page

2-25

2. What Is what

ABL800 FLEX Operator's Manual

Menu structure, Continued


Complete menu Press Menu.

My results* Log on Analyzer Status

Last patient result Help Tutorials

Calibration Progs:
1 Point Calibration 2 Point Calibration tHb Calibration

Start Progs

Data Logs Patient Results Log Patient Profile Log QC Log Calibration Log Activity Log Utilities Archived Data Logs

Auxiliary Progs:
Rinse Cleaning Protein Removal Decontamination Liquid Sensor Adjust Pump Calibration Tubing Refill

Disk Functions:
WDC Report Back up All Data Restore All Data Export Data Logs Import/Export Archives Save Setup Load Setup Restore Default Setup Eject CD

AutoCheck Programs Run High Crea Check


(only the ABL8x7 analyzers)

Standby Sample Counter Temporary Shutdown Long-term Shutdown

Patient Report Archive Quality Control Archive Calibration Archive Activity Archive

RADIANCE Browser (optional)

Setup:
Analysis Setup QC Setup Calibration Setup Replacement Setup General Setup Analyzer Security Print Setup

* Is selected in the Setup program Access Profiles (in the Analyzer Security menu) see chapter 3. Structure of the Setup programs see chapter 3: Installation and Setup.
Continued on next page

2-26

ABL800 FLEX Operator's Manual

2. What Is what

Menu structure, Continued


Access to analyzer menu The access to various parts/functions of the analyzer menu is determined by the rights defined in the Setup program Access Profiles (in the Analyzer Security menu). Example: You as a user can enter the following: Rinse, Patient Report and QC logs, and Standby programs. Then your menu after you have logged on will look like this:

Log on

Last patient result

Analyzer Status

Help

Start Progs

Data Logs

Rinse

Utilities

Patient Report Log QC log

Standby

For details see the chapter 3: Installation and setup and chapter 15: Radiometer Settings.

2-27

2. What Is what

ABL800 FLEX Operator's Manual

Analyzer status
Analyzer status at a glance The working condition of the analyzer is continuously monitored during its operation. To evaluate the analyzer status at a glance before a measurement, use the following facilities: parameter bar color of the traffic light on the Analyzer Status button.
Analyzer Status button

Red = cannot be used; Yellow = unreliable; Green = OK

Analyzer Status To enter the Analyzer Status screen, press Menu - Analyzer Status. elements

Traffic light color of the Analyzer Status button is determined by the traffic light colors of the following status elements: Status element Calibration Status GREEN OK. The colors indicate

YELLOW error(s) in the last calibration and/or cal schedule reminders. RED error-prone parameters are repressed.
Continued on next page

2-28

ABL800 FLEX Operator's Manual

2. What Is what

Analyzer status, Continued


Analyzer Status elements (continued) Button Quality Control Status GREEN OK. The colors indicate

YELLOW error(s) in the last QC measurement and/or QC schedule reminders. RED error-prone parameters are repressed. no replacements due at the present time.

Reagents Status

GREEN

YELLOW a replacement is due, Calibration/Cleaning solutions expire soon or Calibration/Cleaning solutions have expired. GREEN no replacements due at the present time.

Electrode and others System Messages

YELLOW a replacement is due. GREEN YELLOW RED no (critical) messages. non-critical messages. critical messages. The analyzer cannot calibrate or measure. no replacements due at the present time. a replacement is due.

AutoCheck Status

GREEN YELLOW

Calibrations status

The Calibrations status gives an overview of the status of the most recently performed calibration of each type and relevant messages.

The following information is available on the screen: Calibration Type Lists each calibration type along with its status: OK ? Calibration was accepted. Error(s) detected during calibration. Pending or overdue calibration. The last calibration was accepted.
Continued on next page

2-29

2. What Is what

ABL800 FLEX Operator's Manual

Analyzer status, Continued


Calibration status (continued) Calibration Type ?+ Last Time Next Time Lists each calibration type along with its status: Pending or overdue calibration. The last calibration was not accepted.

The date and time the last calibration of the specified type was performed. The date and time the next calibration of the specified type is due according to the Calibration Schedule - see Calibration Schedule, chapter 3. The time interval between calibrations as set up in the Calibration Schedule.

Interval

The following buttons are available: During a calibration:

Result Electrode Upd. After a calibration:

Press to display the result of the highlighted calibration. Press to display the electrode updatings for the ongoing calibration.

Result Start Calibration

Press to display the result of the highlighted calibration. Is activated for the overdue calibrations or the calibrations with "?".
Continued on next page

2-30

ABL800 FLEX Operator's Manual

2. What Is what

Analyzer status, Continued


Quality Control The Quality Control status provides the following information: status status of the last measurement on each control solution type messages referring to quality control measurements.

For the ABL8x7 FLEX analyzers, please note that Slot 11 will be automatically installed when you scan the barcode for the Cleaning Met II Solution during replacement.

Element Solution

Function Lists the quality control slot and its solution type along with a status symbol. Cleaning Met II Solution (for ABL8x7 FLEX analyzers only) is used for quality control of high creatinine levels. OK ? The last measurement was accepted. One or more of the following occurred: error in the last calibration. analyzer error during last QC measurement. a parameter measurement is outside the defined ranges or a Westgard Rule has been violated. The next measurement is overdue, and the previous measurement, if any, was accepted. The last quality control measurement had errors present, and the next measurement is overdue.

Lot Last Time Next Time

The solution lot number for the slot. Lists the time the last measurement was performed. Lists the next scheduled time to perform the measurement on the slot. See Quality Control Schedule, chapter 3.
Continued on next page

2-31

2. What Is what

ABL800 FLEX Operator's Manual

Analyzer status, Continued


Quality Control Buttons: status Run AC (continued) ampoule

Press to start a measurement on the highlighted AutoCheck solution. For ABL8x7 FLEX analyzers only: When Slot 11 is highlighted, this button changes to Run High Crea Check to start a measurement on the Cleaning Met II Solution.

Result

Press to display the last measurement result on the control solution.

Reagents status

The Reagents status shows the following: status of the solutions and gas cylinders When Calibration/Cleaning solutions are expired, the container in question is marked with a clock above it and the contents are colored orange. messages referring to solutions and gas cylinders.

Buttons: Replace Adjust Level Troubleshoot Close Press to enter the screen for replacements of fluids see chapter 7: Replacements. Press to enter the Solution Level Adjustment program to make adjustments see chapter 3 in this manual. Press to view the error interpretation and operator actions see the description in chapter 11: Troubleshooting. Press to exit to the Ready screen.

NOTICE: The Calibration and Cleaning Met II Solution containers are filled so that they provide up to 14-days of use in the ABL8x7 FLEX analyzer. This will be registered on the screen by percentages that are lower than 100 % (i.e. 87%, 75%, 50% for the Calibration Solution 1, Calibration Solution 2 and Cleaning Met II Solution, respectively). This has been made so that the solution warning function will not be impaired and you can select the warning values in the usual manner.
Continued on next page

2-32

ABL800 FLEX Operator's Manual

2. What Is what

Analyzer status, Continued


Electrodes and Other status The Electrodes and Other status shows the following: replacement schedule for items replaced in the Hold mode messages referring to replacements or user activities.

Buttons: Replace Highlight the item and press this button to enter the screen for replacements of electrodes/membranes see Replacing membrane or electrode in chapter 7. Activate User Activities part of the screen to make this button visible. When the button is visible, press it to enter the User Activity program to edit the user activities list see chapter 3 in this manual. Press to exit to the Ready screen.

Log Activity

Close

System messages See chapter 11: Troubleshooting. AutoCheck status AutoCheck Status is available only if the AutoCheck module has been installed. It shows the following: status of the carousel and the measurements scheduled in the QC Schedule program (see section Quality control setup in chapter 3). Messages referring to AutoCheck.
Continued on next page

2-33

2. What Is what

ABL800 FLEX Operator's Manual

Analyzer status, Continued


AutoCheck status (continued)

Element Next Scheduled Next Replacement Available

Function To show the next AutoCheck solution/lot combination according to the scheduled setup. To show the date and time for refilling the carousel according to the scheduled setup. To show the presently available AutoCheck control solutions: type, lot and quantity of ampoules in the carousel. Use the up/down arrow buttons to highlight the desired solution.
Continued on next page

2-34

ABL800 FLEX Operator's Manual

2. What Is what

Analyzer status, Continued


AutoCheck status (continued) The carousel shows the number of ampoules available in it. The black spots indicate used ampoules.

Button Packing List Reset

unrolls to access the following buttons: Displays the Optimal Packing List screen. Cancels any pending quality control measurements. Interrupts ampoule conditioning in the carousel. Resets wet section programs.

Open Module AutoCheck Programs

Opens the cover of the AutoCheck module. Displays the AutoCheck Programs screen to start an AutoCheck measurement.

2-35

2. What Is what

ABL800 FLEX Operator's Manual

Online aid facilities


Online facilities The online aid facilities include the following: Online Help Tutorials Online troubleshooting (see chapter 11 in this manual). Help This function assists you in using the analyzer. Press the button in the Main menu (if not moved to the Ready screen see Access Profiles (in the Analyzer Security menu) in chapter 3) or on the Ready screen. The help text is available for practically every screen your analyzer shows during its operation. The Contents screen displays all help topics. Each topic is hyperlinked to either the next topic level or to the appropriate help text screen. Use the scroll bar or touch the text to search through the topics and access the relevant help information. Press the Previous Topic button to display the help text of a previous screen. The Index screen is an alphabetic list of subjects in the help program. Each letter at the top of the screen is a hyperlink to that section in the index. Touch the letter to display the list of entries under that letter. Each index entry is displayed in bold, purple text (a hyperlink). Simply touch the text to display the relevant help text.
Continued on next page

2-36

ABL800 FLEX Operator's Manual

2. What Is what

Online aid facilities, Continued


Tutorials Tutorials are short video sequences of commonly used procedures. Press the button in the main menu or button next to the Troubleshooting button (when performing troubleshooting) to enter the Tutorial Playlist screen.

The Preview shows the highlighted sequence animation. To start a tutorial on full screen, highlight the topic of interest from the menu, using the arrows, and press the Play button. Topic Blood Samples Quality Check Replacement Contents Videos of a capillary, syringe or expired air sample introduction by highlighting it. Videos of manual or AutoCheck quality control. Video sequences on various replacements.

The buttons on the Tutorial screen during a sequence allow to do the following:

Continued on next page

2-37

2. What Is what

ABL800 FLEX Operator's Manual

Online aid facilities, Continued


Tutorials (continued) Button Function Play. Pause. Stop. Show previous picture show next picture. Switch on the sound (voice presentation of a sequence) with volume regulation. Switch off the sound. Show the list of video sequences.

The "-" button puts Tutorials in the background in order to perform an action and then to return to Tutorials again. The "x" button exits Tutorials. When a button is grayed out, it is not functional.

2-38

ABL800 FLEX Operator's Manual

2. What Is what

Sample counter
Purpose The sample counter allows you to keep track of the measurements, calibrations and quality control. The screen elements are described below:

Description

Element Parameter Count

Function Lists the parameters and how many times each of them was measured by the analyzer. Normally the count is the same as the total number of measurements provided the parameters were not excluded from the measurement(s). Shows the number of sample measurements, calibrations and QC measurements made since the sample counter was last reset (User column). The following is registered: Activity Total Number of... Completed sample/quality control measurements/ calibrations only. Interrupted or aborted measurements are excluded. Aborted sample/quality control measurements/ calibrations due to sample errors, wet section errors, etc. interrupted measurements excluded. All completed sample/quality control measurements/calibrations performed by all operators since the sample counter was last reset.

Counters

Aborted

User

User counters last reset

Gives the date when the counters in the User column were last reset to zero.
Continued on next page

2-39

2. What Is what

ABL800 FLEX Operator's Manual

Sample counter, Continued


Description (continued) Element Reset Counters Function

For the analyzers with no logon protection of the Setup programs, press this button to reset the counters in the User column. Print Close Press to start the printout of information in Counters and in Parameters. Press to return to the Ready screen.

2-40

3. Overview
Introduction

Installation and setup

After the analyzer has been installed, you can define the settings on your ABL800 FLEX analyzer according to your own needs and requirements. This chapter contains the following topics. Installation........................................................................................................ Setup menu structure........................................................................................ Analyzer security ............................................................................................. 3-2 3-3 3-6

Contents

Analysis setup .................................................................................................. 3-13 Patient reports .................................................................................................. 3-26 Calibration setup .............................................................................................. 3-32 Quality control setup ........................................................................................ 3-35 Replacement setup ........................................................................................... 3-48 Parameters and input setup............................................................................... 3-55 Analyzer settings.............................................................................................. 3-62 Communications .............................................................................................. 3-68 Printers ............................................................................................................. 3-77 Disk Functions setup ........................................................................................ 3-80 Corrective actions ............................................................................................ 3-83 Miscellaneous setup ......................................................................................... 3-86

3. Installation and setup

ABL800 FLEX Operator's Manual

Installation
Information The ABL800 FLEX analyzers are installed and fully prepared for use by Radiometer representatives in your country. The reason for it is that due to the analyzer's modular design each user can select only those modules that cover his/her specific needs. These modules are installed and checked at the user's place. Installation of consumables and accessories are exactly the same as the procedures described in chapter 7: Replacements.

3-2

ABL800 FLEX Operator's Manual

3. Installation and setup

Setup menu structure


Accessing Setup To access the Setup programs, press the buttons: programs

Setup menu structure

Setup
Analyzer Security:
General Security Operators and Passwords Access Profiles

Replacement Setup:
Replacement Schedule User Activities Solution Level Forecast Solution Warning Expiration Warning AutoCheck Warning

QC Setup:
QC Solutions QC Schedule QC Ranges QC Input Setup QC Statistics Westgard Rules Rilibk Ranges

Calibration Setup:
Cal Schedule Drift Tolerances

Analysis Setup:
Capillary Modes Syringe Modes Patient Reports Reportable Ranges Reference Ranges Sample Pre-registration Sample Logistics Setup

Print Setup

General Setup

See the next page

Continued on next page

3-3

3. Installation and setup

ABL800 FLEX Operator's Manual

Setup menu structure, Continued


Setup menu structure (continued)

General Setup

Printers:
Printer Setup Automatic Printing

Communications:
RADIANCE Connection LIS/HIS Connection QA Portal Automatic Data Transmission Automatic Data Request Patient Lookup Setup Remote Support

Analyzer Settings:
Analyzer ID Time/Date Acoustic Signal Environment Language

Parameters and Input:


Parameters Units User-defined Data Items User-defined Notes

Disk Functions Setup:


Automatic Backup Automatic Archiving

Miscellaneous Setup

Corrective Actions

Note that the exit from a setup program returns you to the Ready screen. You will find it more convenient to place some buttons, e.g., Utilities, on the Ready screen to shortcut the entry to a desired setup program - refer to Access Profiles program (in the Analyzer Security menu) in chapter 3.
Continued on next page

3-4

ABL800 FLEX Operator's Manual

3. Installation and setup

Setup menu structure, Continued

Entering password

To enter the Setup programs, do the following. Step 1. 2. Action Press Menu on the Ready screen. Then click on Logon. Depending on the authentication setting chosen in the General Security screen, enter or scan your user name and password, or your logon-barcode.

Confirm with Enter.

Print Setup

This program allows you to print parts of the setup. To select the parts of the setup for printing, do the following: Step 1. 2. Action Press Menu Utilities Setup Print Setup. All setups are selected as a default. Press the relevant check buttons to deselect the setup that you do not wish to be printed out.

3.

Press Print to start printout of the selected setups or press Close to return to the Ready screen.

3-5

3. Installation and setup

ABL800 FLEX Operator's Manual

Analyzer security
Programs Analyzer Security programs are described in this section.

General Security

This program allows you to hand over the control of the operators and passwords to the RADIANCE system and to allow an anonymous use of the analyzer. To enter this program, press Menu Utilities Setup Analyzer Security General Security.

To hand over the control of the operators and passwords to the RADIANCE system activate the check button in the "Centralized User Management" box. With this option enabled you can only view the operators, not add, edit or remove any of them. All users defined on the ABL are deleted and the list of users in the RADIANCE system is copied to the ABL. NOTICE: This option can only be enabled if RADIANCE communication is enabled in the RADIANCE Connection Setup program.

Continued on next page

3-6

ABL800 FLEX Operator's Manual

3. Installation and setup

Analyzer security, Continued


General Security (continued) To define, how the user should log on, use the up/down arrow buttons in the "Authenticate" box to select the desired logon option. The following options are available: UserId/Password as primary This option allows you to enter or scan a User name and password in the Logon screen. By pressing the Log On BC button a Logon-barcode can be scanned. UserId/Password only This option allows you to enter or scan the user name and password in the Logon screen. Logon-barcode as primary This option allows you to enter or scan a Logon-barcode in the Logon screen. By pressing the Extented Log On button a user name and password can be scanned. Logon-barcode only This option allows you to enter or scan a Logon-barcode in the Logon screen. To allow an anonymous use of the analyzer, i.e. use without logon, use the up/down arrow buttons to select "Yes" in the "Anonymous use" box (default) and select the desired access profile of the anonymous operator, in the "Access profile for anonymous operator" box. The "Access profile for anonymous operator" box only appears, when "Yes" is chosen in the "Anonymous use" box. See Access Profiles further in this chapter for information on how to define the access profiles. To set the time interval to elapse, before an operator is automatically logged off, press the Logoff Time button. Select the logoff time in minutes (from 0 to 60) and seconds (from 0 to 50 in 10-second intervals). The default logoff time is three minutes. Press Back to return to the General Security screen.
Continued on next page

3-7

3. Installation and setup

ABL800 FLEX Operator's Manual

Analyzer security, Continued


Operators and Passwords This program allows you to add, edit or remove operators and to assign an access profile to each operator. NOTICE: If the Centralized User Management option is enabled in the General Security screen, you cannot add, remove or edit the operators, but only view the access profiles of the individual operators. To enter the Operators and Passwords program, press Menu Utilities Setup Analyzer Security Operators and Passwords.

Shows the list of available operator profiles Shows the list of actual operators

When the analyzer is taken into use, the following default operators are available: Operator Manager Has access to All menu items and programs (not service programs). It is recommended to remove this operator with the standard password: 123456, and enter the actual users with their profiles and passwords. All menu items and programs (user and service) on the analyzer. Note that "Radiometer" cannot be removed from the operator list. Only if the Remote Support option key is installed.
Continued on next page

Radiometer

Remote operator

3-8

ABL800 FLEX Operator's Manual

3. Installation and setup

Analyzer security, Continued


Operators and Passwords (continued) To add an operator to the list, do the following: Step 1. Action Press the Add Operator button to display the Add New Operator screen.

2. 3.

Type the name of the operator or operator category in the "Operator ID" box, using the screen keyboard. Enter or scan the password: in the "Password" box. The password must be at least four characters long, and not more than 32.

4. 5.

Re-enter or re-scan the password in the "Confirm" box. Enter or scan the logon barcode in the "Logon barcode:" box The logon barcode must be at least four characters long. The logon barcode and the password can be identical.

6. 7.

Re-enter or re-scan the logon barcode in the "Confirm" box. Press Back. If the password is not accepted, the Add New Operator screen remains open and a message, telling you what was wrong, appears. If the password is accepted, the Operators and Passwords screen is displayed.

8.

In the Operators and Passwords screen, select the desired access profile of the new operator.
Continued on next page

3-9

3. Installation and setup

ABL800 FLEX Operator's Manual

Analyzer security, Continued


Operators and Passwords (continued) To remove an operator from the list, use the up/down arrow buttons in the "Operator" box to highlight the operator and press Remove Operator.

NOTICE: If the Centralized User Management option is enabled in the General Security screen you cannot add, remove or edit the operators, but only view the access profiles of the individual operators. This program allows you to define the permitted actions, the available menu items and button shortcuts of an access profile. To enter this program, press Menu Utilities Setup Analyzer Security Access Profiles.

Access Profiles

To define the permitted actions of an access profile, select the desired profile in the "Profile names" box and activate the desired check buttons in the "Permitted actions" box. To deactivate an action, press the check buttons once again.
Continued on next page

3-10

ABL800 FLEX Operator's Manual

3. Installation and setup

Analyzer security, Continued


Access Profiles (continued) To define the available menu items and button shortcuts of an access profile do the following: Step 1. Action In the Access Profiles screen highlight the desired access profile in the "Profile names" box and press Menus and Buttons. Note that this button is grayed-out for the service technician profile. 2. Select the desired menu items in the "Menu Items in Quick Menu" box.
Selected profile

is named on the screen.

A grayed-out item in the Menu & Button Configuration screen indicates that only some subitems were selected in this group. Clear checked items indicate that all subitems have been selected. The buttons allows you to do the following: Highlight menu items

Open/close submenus

Select/deselect a menu item.

Continued on next page

3-11

3. Installation and setup

ABL800 FLEX Operator's Manual

Analyzer security, Continued


Access Profiles (continued) Step 3. Action To create a button shortcut for a specific item, highlight the desired item in the "Menu Items in Quick Menu" box and then, in the "Button configuration" box, press the button position that you wish to give the selected item.

4. 5. 6.

Select other five buttons in the same manner, if desired. To deselect a button, press it once again. Press Back to return to the Access Profiles screen.

Enabling the My Results option will give the operator an easy access to all Patient Results made by that operator, by displaying the Patient Result Log, filtered on the operator name.

3-12

ABL800 FLEX Operator's Manual

3. Installation and setup

Analysis setup
Programs Analysis Setup programs are described in this section.

Activate a button to enter the program.

Syringe Modes and Capillary Modes Setup

The Syringe Modes Setup and Capillary Modes Setup are shown below:
To select a measuring mode To enable a mode

The activated button will be selected as default mode when opening the inlet

Continued on next page

3-13

3. Installation and setup

ABL800 FLEX Operator's Manual

Analysis setup, Continued


Selecting a measuring program To select a measuring program, do the following: Step 1. Action Blood, pleura and expired air samples: Press the desired mode button, enable it and select the desired measuring program with the arrow buttons. Other fluids: Activate the desired mode button and enable the Other fluids check button. 2. 3. 4. Press Parameters to select a parameter profile for a measuring program see Selecting parameter profile below. Edit, if desired, the left part of the button name for the mode by pressing Edit Name. Enter corrections for the Other fluids mode by pressing Corrections see Entering corrections for Other Fluids Modes further in this section. Select the report layout for a measuring program by pressing Layout see Selecting a report layout further in this section. You can use one of the syringe measuring modes for test tubes. Use Capillary 35 L OXI for bilirubin analysis (ctHb = 0) on plasma. Deselect all other parameters for this mode in the Parameter Profile. Alternatively, you can use the Other Fluids mode for the bilirubin measurement on plasma (this is not relevant for the ABL8x7 FLEX analyzers). In the Syringe Modes Setup and Capillary Modes Setup screens you can define up to 12 measuring programs. To get access to the last 6 mode buttons press Next Page.
Continued on next page

5.

NOTICES:

3-14

ABL800 FLEX Operator's Manual

3. Installation and setup

Analysis setup, Continued


Selecting parameter profile To select parameter profile for each measuring program, do the following: Step 1. 2. Action Select a measuring program and press Parameters on the Syringe Modes Setup or Capillary Modes Setup screen. Select parameters for a given measuring mode by pressing a parameter check button (see Screen elements in chapter 2). The list of available parameters depends on your analyzer version.

3.

Select the dynamic parameters option in the "Use dynamic parameters" box as follows: Parameters can be selected before a measurement. Parameters can be selected after a measurement.

4.

Press the Back button to return to the Syringe Modes Setup or Capillary Modes Setup screen.

Disabled versus deselected parameter

A parameter is disabled, i.e. excluded from the Parameter Profile screen and the parameter bar, in General Setup Parameters and Input Parameters. A parameter deselected for the given syringe or capillary measuring program will be measured, but excluded from the displayed and printed patient report. You can further select or deselect a parameter before or after a measurement see chapter 4: Sample measurements.
Continued on next page

3-15

3. Installation and setup

ABL800 FLEX Operator's Manual

Analysis setup, Continued


Selecting HbF correction To give the user the possibility of selecting or deselecting HbF correction during a measurement, do the following: Step 1. Action Select HbF correction for all levels or for HbF levels higher than 20 % in Miscellaneous Setup (Setup General Setup Miscellaneous Setup) see the description further in this chapter. Return to the Analysis Setup, choose the desired syringe or capillary program and press the Parameters button to enter the Parameter Profile screen. Activate the check button in the "Use HbF correction" box (if "Disabled for all levels" has been selected in Miscellaneous Setup, the text box and the check button in it will not be visible on this screen).

2.

3.

4.

Press the dynamic parameters check button in the Use dynamic parameters box in order to be able to exclude the HbF correction and other parameters during a sample measurement.

HbF correction was disabled in Miscellaneous Setup

Continued on next page

3-16

ABL800 FLEX Operator's Manual

3. Installation and setup

Analysis setup, Continued


Entering corrections for Other Fluids mode The Other Fluids mode allows measurement on samples other than heparinized whole human blood and, therefore, requires entering corrections (offset and slope). Step 1. 2. Action Select the Other Fluids mode and press the Corrections button on one of the Modes Setup screens (Capillary or Syringe). Highlight the parameter with the arrow buttons and press Edit.

3.

Key in the slope and offset values. Confirm each entry with Enter.

Please note that only values within the range programmed into the software will be accepted. 4. 5. Press Back to confirm the entries and return to the previous screen. To delete a correction, highlight it and then press the Remove button.
Continued on next page

3-17

3. Installation and setup

ABL800 FLEX Operator's Manual

Analysis setup, Continued


Selecting a report layout To select a report layout for a measuring program, do the following: Step 1. 2. Action Select a measuring program. Press the Layout button on one of the Mode Setup screens.

3.

Select the layout for the measuring program from the list made in the Patient Report Setup see Patient report further in this chapter. The selected layout will be the default layout for the given measuring program. Press Back to confirm the settings.
Continued on next page

4.

3-18

ABL800 FLEX Operator's Manual

3. Installation and setup

Analysis setup, Continued


Reference ranges and critical limits This program allows you to enter your own reference ranges and critical limits for all measured and calculated parameters. For each parameter, you can choose whether or not to differentiate between the categories of sample type, sex and age group.

Press the check button to activate a function; press the check button again to deactivate it. To view the reference ranges and critical limits, highlight a parameter in the Viewing the reference ranges "Parameter list" box using the arrow buttons. and critical limits

View the settings for a highlighted parameter

Selecting sample To select a sample type, highlight a parameter in the "Parameter" box, using the arrow buttons. type Press the check button in the "Sample type" box and select a sample type, using the arrow buttons in the box. To deactivate sample type differentiation, press the check button again. Selecting sex To select sex, highlight a parameter. Press the check button in the "Sex" box and select sex type, using the arrow buttons in the box. To deactivate sex differentiation, press the check button again.
Continued on next page

3-19

3. Installation and setup

ABL800 FLEX Operator's Manual

Analysis setup, Continued


Setting age group limits To set the age group limits, do the following: Step 1. Action Press the Age groups button to display the following screen:

0 days-1 month, 1-5 months, 5-8 months, >8 months"

2.

Use the following: left/right arrows to choose the age group limit you want to alter (indicated by a blue circle containing a white cross). up/down arrows to scroll through the list of possible age limits. As the list is scrolled the text on the age group bar changes accordingly.

3. 4. 5.

Repeat step 2 for each limit to be changed. Press Back when completed to return to the Reference Ranges and Critical Limits screen. Activate the Age group check button. To deactivate age group differentiation, press the check button again.

Setting reference and critical limits

To set the reference and critical limits for each parameter do the following: Step 1. Action Highlight a parameter by using the up/down arrow buttons.

2.

Choose the sample type, sex, age group and press the Edit button.
Continued on next page

3-20

ABL800 FLEX Operator's Manual

3. Installation and setup

Analysis setup, Continued


Setting reference and critical limits (continued) Step 3. Action To remove all current ranges, press the Clear limits button.

4.

To change the value in a range limit, touch and highlight the range. Then enter the limit, using the screen keyboard and confirm with Enter. Repeat for other limits in a similar way. Press Back when completed.

5. 6. Reportable ranges

To change or enter a reportable range (must be selected smaller than or equal to the measuring range), do the following: Step 1. Action Scroll to the desired parameter, using the up/down arrows or scroll bar.

Measured parameters show their measuring ranges. Derived parameters show .....

2. 3. 4.

Key in the desired lower limit and confirm with Enter on the keypad. Key in the upper limit and confirm with Enter on the keypad. To change the reportable range to the default (primary) setting, highlight the desired parameter and press Set Default.
Continued on next page

3-21

3. Installation and setup

ABL800 FLEX Operator's Manual

Analysis setup, Continued


Reportable ranges (continued) Step 5. Action To change all parameters to the default values, press the 1-n Set All Default button.

Press Continue to change all parameters reportable ranges to the default ones. Press Cancel to keep the user-defined reportable ranges and return to the previous screen. 6. Press Close to exit the program and confirm the selected settings.

See also Calibration verification in chapter 15 in this manual. Sample Preregistration setup This program allows you to select interpretation of the barcode and the patient data that can be confirmed before a sample measurement.

To select the settings, do the following: Step 1. Action Use the up/down arrow buttons to select the interpretation of the barcode setting in the "Interpret barcode input as" box. Choose one of the following: Patient ID Accession Number Sampler ID.
Continued on next page

3-22

ABL800 FLEX Operator's Manual

3. Installation and setup

Analysis setup, Continued


Sample Preregistration setup (continued) Step Action

Note that choosing the Accession Number or Sampler ID will gray out 1 (cont.) its check button (Sampler ID on the screen above). Choosing the Accession Number or Sampler ID ensures that this data will be included in the "Data confirmation" box. 2. 3. Select the barcode entry. Activate the relevant check buttons in the "Included fields" box: Accession Number, Sampler ID, Patient First Name, Patient Last Name, Birth Date, Patient Sex. Activate the check button in the "Data confirmation" box so that the patient data can be confirmed before the sample measurement. Press Close to confirm the settings and return to the Ready screen.

4. 5.

Sample Logistics This program allows you to select the following: the batch mode (automatic sample handling), Patient Identification shown on the Ready screen and to define Setup maximum allowed sample age for each measured parameter.

Batch mode (automatic sample handling)

The batch mode allows you to place your sampler in an empty slot in the FLEXQ sampler tray and leave the analyzer. The sampler will be registered, sample analyzed and the results sent to the required person. To achieve this the following is required: Analyzer has live connection to the RADIANCE system Barcode interpretation is set to Sampler ID in the Sample Pre-registration "When entering Sampler ID" in "Request patient demographics" box is selected and "From connection" must be set to RADIANCE in the Automatic Data Request setup All sample/patient information is present on the RADIANCE system
Continued on next page

3-23

3. Installation and setup

ABL800 FLEX Operator's Manual

Analysis setup, Continued


Batch mode (automatic sample handling) (continued) If the above requirements are fulfilled, activate the check button in the "FLEXQ" box. This box is not available for the analyzers without the FLEXQ module. If a live RADIANCE connection is detected, the analyzer will automatically set the connect option when activating the Enable batch mode button.

To select the information in the second column on the Ready screen, use the Patient identification on up/down arrow buttons in the "FLEXQ" box. the Ready screen

The choices are: Patient Last Name, Accession No. or Sampler ID. Sample age Sample age is the time within which the sample should be analyzed. It can be defined for each measured parameter individually or for all parameters. Patient Report Layout (described further in this chapter) offers the following time registration items in the Patient ID: "Draw time", i.e. when the sample was obtained from the patient; is entered by the operator on the Patient Identification screen on the analyzer or obtained from the RADIANCE system. "Sample registration", i.e. when the sample was registered by the analyzer; is automatically registrered by the analyzer. "Time before FLEXQ", i.e. sample transport time to the analyzer; is obtained if the "Draw time" has been entered; is automatically registrered by the analyzer. "Time in the FLEXQ", i.e. waiting time; is automatically registrered by the analyzer. Max. sample age (is entered on the RADIANCE system). This setting, if used, will overrule the settings selected on the analyzer. The sample age is calculated, using the following: Manual measurement FLEXQ measurement "Sample age " = "Sample registration" "Draw time". Sample age = 0 if "Draw time" is not entered. "Sample age" = "Time before FLEXQ" + "Time in FLEXQ" where: "Time before FLEXQ" = "Sample registration" "Draw time"; Time in FLEXQ" = "Time the sample aspiration begins" "Sample registration"
Continued on next page

3-24

ABL800 FLEX Operator's Manual

3. Installation and setup

Analysis setup, Continued


Sample age (continued) To define maximum allowed sample age, do the following: Step 1. 2. Action Activate the check button in the "Calculation of sample age" box. To select the same age for all parameters, press Set all. To select the sample age for each parameter individually, highlight a parameter in the "Sample aging time per parameter" box, using the up/down arrow buttons. Select sample age for the highlighted parameter, using up/down arrow buttons.

3.

3-25

3. Installation and setup

ABL800 FLEX Operator's Manual

Patient reports
Purpose This program allows you to create a number of new layouts for patient reports or to modify the existing ones.

The screen accomodates two main functions: creating a layout and editing a layout. Creating a layout To create a new layout, do the following: Step 1. Action Press New to make a new blank layout with the name "New" displayed on the right side of the screen in the "Name" box. If you wish to make a copy of an existing layout and use it as a basis for a new layout, select the desired layout and press Copy. "Copy" will be added to the name in the "Name" box (i.e. "-R- default Copy").

Press the keyboard icon beside the "Name" box Type in a new name for your layout. Confirm with Enter to return to the Patient Report Setup screen. The newly named layout will be added to the list of layouts. 2. Edit your layout as described in Editing a layout further in this section.
Continued on next page

3-26

ABL800 FLEX Operator's Manual

3. Installation and setup

Patient reports, Continued


Creating a layout (continued) To make the highlighted layout a default for your analyzer press Make Default. The default layout will be marked with () in the list of layouts. To make a test printout of the highlighted layout (patient ID items and selected parameter groups with the parameters/units for each parameter group) press Preview. This test print will be labeled "Preview". To reset any user defined layout to the R- Default layout, select the desired layout and press R Default. To delete a highlighted layout, press Delete. Note that the Radiometer default layout cannot be deleted. The button is disabled if only the Radiometer default layout is available.

Editing a layout To edit a layout, do the following: Step 1. 2. Action Highlight a layout in the list by touching it on the screen. Press the Edit Patient ID Layout button to edit the patient ID items see description on the next page.

3. 4.

Press the Edit Patient Results Layout button to edit the parameter groups see description further in this section. If an automatic printout of the Acid-Base chart for this layout is desired, activate the Print Acid-Base Chart button.
Continued on next page

3-27

3. Installation and setup

ABL800 FLEX Operator's Manual

Patient reports, Continued


Editing patient ID layout You can customize the patient identification (Patient ID) information as follows: select the items to be included in the Patient ID. set Patient ID entries as mandatory. define default values for Patient ID.
Selected patient report layout

To edit a patient ID for a selected patient report layout, do the following: Step 1. Action To add a highlighted (by touching it on the screen) item in the "Available items" box, to the list of selected items, press the button. To remove a highlighted (by touching it on the screen) item from the "Selected items" box, press the button. To make a highlighted item in the list of selected items mandatory, press the Set as Mandatory button. The item will be indicated by a on the Patient ID screen and must be entered during a measurement before a patient result can be viewed. To remove the mandatory mark, highlight the item in the "Selected items" box and press the Set as Mandatory button again. NOTICE: If you are using mandatory input fields, you are not allowed to change the report layout during measurement, unless the mandatory fields have been filled out. 4. Press Back to exit the screen.
Continued on next page

2. 3.

3-28

ABL800 FLEX Operator's Manual

3. Installation and setup

Patient reports, Continued


Entering default To enter default values for patient identification items, do the following: values Step Action 1. Use the arrow buttons on the right to highlight the desired item in the "Selected items" box.

Changes to

List 2. Set the default: If the item has a value, press Keyboard, enter the value and press Enter on the keyboard to confirm. If the item has a list of options, press List, highlight the option using the arrow buttons and press Enter to confirm. 3. 4. Set or change other default values in the similar manner. Press Back to exit and confirm changes.

NOTICES: It is not possible to set default values for all items. The values can be changed on a result-by-result basis on the Patient Identification screen. An item placed in the Selected items list does not appear in the Available items list. To use the Patient Lookup function, Department (Pat.) should be selected for the Patient Identification screen. To use the Request function, Accession number, Sampler ID and/or Patient ID should be selected for the Patient Identification screen.
Continued on next page

3-29

3. Installation and setup

ABL800 FLEX Operator's Manual

Patient reports, Continued


Editing patient result layout You can do the following: select parameter groups and the parameters in each group, and make report layout by using layout commands. To edit a patient result layout, do the following: Step 1. Action Select the patient report layout by highlighting it on the screen and

press Edit Patient Results Layout. 2. Use the arrow buttons to highlight a parameter group and press

Include Exclude

to include the group in the patient result. It will be shown in the "Selected items" box. 3. 4. 5. Select parameters for this parameter group by highlighting them one by one and pressing (Include). Select another parameter group and parameters for this group in the same manner. Use layout commands <New group> (items following this command are placed on the top of the next half of the screen), <New Line> (a line is inserted between items) or <New Page> (items following this command appear on next screen page) as desired and press .
Continued on next page

3-30

ABL800 FLEX Operator's Manual

3. Installation and setup

Patient reports, Continued


Editing patient result layout (continued) Step 6. Action To exclude an item from the selected parameter list, highlight it and

press 7. If you wish to display parameters with their reference ranges and have them checked against reference ranges/critical limits, do the following: Highlight a parameter. Press Show Ranges to indicate it by [xxx-xxx]. Repeat for other parameters in the same manner. If the [xxx-xxx] is not included for a parameter, checking against reference ranges/critical limits is disabled for this parameter. However, checking against the reportable ranges is always performed. NOTICE: Refer to the Reference Manual, chapter 6: Parameters, for information on parameters and their groups.

3-31

3. Installation and setup

ABL800 FLEX Operator's Manual

Calibration setup
Programs Calibration Setup programs are described in this section.

Activate a button to enter the program. Calibration drift This program allows you to change the default drift tolerances for the pH, pCO2, tolerances pO2, electrolyte and metabolite electrodes see Radiometer default settings in chapter 15 in this manual. NOTICE: Radiometer recommends the use of the default drift tolerances. Too narrow drift tolerances will cause electrode drift errors even for normal electrode fluctuations. If the drift tolerances are made wider, no warning will be given if the electrode should become unstable. Significant measurement errors could result. To define your own drift tolerances, do the following: Step 1. Action Highlight the desired parameter by touching it on the screen. The list of available parameters depends on your analyzer version.

2. 3. 4.

Type in your own value and confirm with Enter. Change other drift tolerances in the same manner. Press Close to confirm the changes.
Continued on next page

3-32

ABL800 FLEX Operator's Manual

3. Installation and setup

Calibration setup, Continued


Calibration schedule This program allows you to schedule automatic calibrations, manual tHb calibration and automatic cleaning (the first automatic cleaning also includes quality control of the Crea A and B electrodes for high creatinine values on the ABL8x7 FLEX analyzers), and select a calibration to be performed after a sample measurement.
24-hour scale shows the time for each scheduled automatic calibration or cleaning.

To edit the schedule for automatic calibrations/cleaning and tHb calibration, do the following: Step 1. Action Highlight the desired calibration or cleaning on the screen.

List of activities: calibrations and cleaning.

2.

Press the Edit button.

Activity after each measurement

Use the arrow buttons to select start time for calibrations and cleaning and the intervals between each activity.
Continued on next page

3-33

3. Installation and setup

ABL800 FLEX Operator's Manual

Calibration setup, Continued


Calibration schedule (continued) Step 3. Action Select activity after each measurement by activating the check button for the desired calibration.

The following options are available. Available calibration schedule options Option 1 Point calibration 2 Point calibration 1 Point pH/BG calibration (for the USA only) tHb calibration Start time

Interval

30 min, 1 hour, 2, or 4 hours. 1 hour, 2, 4 or 8 hours. 30 min, 1 hour, 2 hours. Never, 7 days, 1 month, 2, 3, 4 or 6 months 00:00, 00:15, 00:30, 00:45.. 23:45. or 12 midnight, 12:15 am, 12:30 am, 12:45 am.12 midday, 12:15 pm..11:45 pm.

Activity after measurement Cleaning intervals

none, 1 Point pH/BG calibration (for the USA only), 1 Point calibration, 2 Point calibration. 8, 24 hours.

3-34

ABL800 FLEX Operator's Manual

3. Installation and setup

Quality control setup


Programs The quality control setup programs are described in this section.

Activate a button to enter the program. QC solutions This program allows you to assign or change a quality control solution to a specific slot. NOTICE: Changing a control solution assigned to a slot will delete all current quality control statistics obtained on that slot. If you want a copy of the statistics for the last QC month, create a WDC Report see chapter 8: Disk Functions.

NOTICE: (for the ABL8x7 FLEX analyzers only) Slot 11 is assigned to the Cleaning Met II Solution used for high Crea check.
Continued on next page

3-35

3. Installation and setup

ABL800 FLEX Operator's Manual

Quality control setup, Continued


QC solutions (continued) To assign a control solution to a slot, do the following: Step 1. 2. Action Highlight a slot, using the arrow buttons. For control solutions from Radiometer, scan the bar code or press the keypad button 3. to enter the barcode.

For non-Radiometer control solutions, press Add Non-R-.

The warning

Press Accept to continue with the installation of the control solution. Press Cancel to cancel installation of the control solution. 4. To delete a control solution, highlight the desired slot and press Delete. The warning as in step 3 above appears. Press Accept to continue with the removal of the control solution. Press Cancel to cancel. 5. To define whether an AutoCheck quality control measurement is to be run manually or automatically, press the Manual/Auto button. Useful if the AutoCheck module is out of order.

The following quality control solutions can be used with the ABL800 FLEX analyzers: Quality control system QUALICHECK5+; AutoCheck5+ QUALICHECK3+; QUALICHECK5+; AutoCheck3+; AutoCheck5+ QUALICHECK5+; AutoCheck5+ AutoCheck6+, S8377 Cleaning Met II Solution Analyzer ABL805 ABL830/20/10 ABL835/25/15 ABL837/27/17

For the solution types for each quality control system, please refer to chapter 14: Ordering information in this manual. You can also use a suitable non-Radiometer quality control system for your analyzer.
Continued on next page

3-36

ABL800 FLEX Operator's Manual

3. Installation and setup

Quality control setup, Continued


QC ranges This program allows you to do the following: globally update all control ranges of a slot to a calculated lot-to-date range. individually edit parameter control ranges by entering your own ranges or updating to a calculated lot-to-date range. define a minimum allowed control range by entering a Fixed SD (standard deviation). See the Quality Control Systems Reference Manual for detailed information on statistics and its parameters. To edit data for the control solution, do the following: Step 1. Action Use Next Slot to display the desired slot.
Selected control solution

Press Edit. 2. Select the parameter to be edited, using Next Param. or Prev Param.

Parameter

3.

Press Update to change the range to the one shown in the "Lot-to-Date range (2 SD)" box - if available (i.e. the range calculated over the course of the lot, represented mathematically by: mean value 2 SD. This is the range within which 95 % of the measurements taken are found).
Continued on next page

3-37

3. Installation and setup

ABL800 FLEX Operator's Manual

Quality control setup, Continued


QC ranges (continued) Step 4. 5. Action Highlight the limit by touching it on the screen and enter your own value(s), using the keypad. To activate or deactivate the Fixed SD (i.e. a minimum allowed control range is defined by setting a Fixed SD), press the check button.

To change the SD value, touch the SD field to highlight it and enter the value, using the keypad. Confirm the entry with Enter. 6. 7. Repeat the procedure for other parameters in the same manner. To update the control ranges for all parameters of a displayed level to the corresponding lot-to-date (2 SD) range shown, display the desired slot, using Next Slot.

Press Update All. 8. Press Continue to update the control ranges for all parameters under the specified solution slot.

Press Cancel to cancel updating.


Continued on next page

3-38

ABL800 FLEX Operator's Manual

3. Installation and setup

Quality control setup, Continued


QC ranges (continued) NOTICE: Once the Fixed SD is activated, you cannot update the control ranges to the limits that are narrower than those determined by the Fixed SD, for both singleparameter and multiple-parameter updates. This program allows you to select the following for the Quality Control Identification screen during a measurement: mandatory temperature entry by the operator the default temperature always displayed (unless altered by the operator). To select the settings, do the following: Step 1. Action Activate the Mandatory temperature check button.

QC input setup

A will appear next to an empty temperature box on the Quality Control Identification screen during each quality control measurement and a temperature must be entered, otherwise the result cannot be retrieved. 2. Highlight the default temperature in the "Default temperature" box and enter a default temperature on the keypad; confirm with Enter. The value in C or F is automatically entered on the Quality Control Identification screen during measurement. The temperature can be altered for a particular measurement but will return to the default setting for future measurements.
Continued on next page

3-39

3. Installation and setup

ABL800 FLEX Operator's Manual

Quality control setup, Continued


QC schedule This program allows you to schedule quality control measurements for your analyzer.

The screen shows the timetable for all days of the week and time during a day. Step 1. 2. Action To select time during a day, use the and buttons. To display other weekdays, use the and buttons.

Symbols for manual and automatic quality control are as follows: measurement(s) on the control solution manually performed. measurement(s) on the control solution automatically performed. To add a new quality control solution to a schedule, do the following: Step 1. 2. Action Select the desired time. Press the Add button to display the screen below.

Slots/quality control solutions

Touch the QC slot box to activate it.


Continued on next page

3-40

ABL800 FLEX Operator's Manual

3. Installation and setup

Quality control setup, Continued


QC schedule (continued) Step 3. Action Select the slot/quality control solution (entered in the Quality Control Solutions program), using the up/down arrow buttons in the box. Confirm with Select. 4. Highlight the "Weekdays" box using the Field down button. Activate the relevant check buttons to select the days of the week this measurement should be performed. 5. Highlight the "Start time" box. Key in the time to perform a measurement and confirm with Enter on the keypad. NOTICE: The symbols for the automatic or manual quality control (given on the previous page) will automatically appear in the schedule. 6. 7. Highlight the "Repeat" box. Select the interval the measurement should be repeated, using the up/down arrow buttons in the box. Press Back to return to the Quality Control Schedule Setup screen. If more than one quality control solution was scheduled, the schedule screen will look like this:

To edit the schedule, press Edit and follow the procedure above.
Continued on next page

3-41

3. Installation and setup

ABL800 FLEX Operator's Manual

Quality control setup, Continued


QC schedule (continued) To remove a quality control measurement from the schedule do the following: Step 1. Action Highlight the desired quality control measurement and press Remove.

2.

Press the desired button to remove the quality control measurement from the schedule: For this day, For all days or All entries for QC slot (number). Press Back to return to the QC Schedule screen.

3. QC Statistics

This program allows you to select the following: the statistics factor the initiation of warning messages at the end of the statistical month. To select the settings, do the following: Step 1. Action Statistics factor expands the control range to the statistics range (it is a range within which quality control results must fall in order to be included in the quality control statistics).

Key in the desired statistics factor (from 1.0 to 9.9) on the keypad and confirm with Enter. The default value is 1.5.
Continued on next page

3-42

ABL800 FLEX Operator's Manual

3. Installation and setup

Quality control setup, Continued


QC Statistics (continued) Step 2. Action To set up the analyzer to send reminders to print or export the monthly statistics, activate the appropriate check button (or both) for the message(s) to appear in System Messages upon occurrence. Press Close to exit the program.

3.

Westgard Rules The Westgard Rules are a set of statistical rules that, when applied to the quality control results, can increase the probability of detecting an error in the sampling procedure or in the analyzer itself, or help detecting a shift or trend in your quality control results by comparing current measurement values of a control solution with previous values. This program allows you to select Westgard Rules for all slots or for specific parameters. To activate or deactivate the assigned Westgard Rules, do the following. Step 1. Action Press On/Off button to activate the assigned Westgard Rules for the slot.
Slot, solution, type, lot

Press On/Off again to deactivate the assigned Westgard Rules.

Buttons are grayed out

Continued on next page

3-43

3. Installation and setup

ABL800 FLEX Operator's Manual

Quality control setup, Continued


Westgard Rules (continued) Step 2. 3. 4. Action Press Next Slot and activate/deactivate the Westgard Rules for the other slots in the same manner. To activate the Westgard rules for specific parameters in a slot, display the desired slots by using Next Slot. Press the Edit button. The screen shows all the Westgard Rules and their current settings for the parameter.
Parameter

5. 6.

Display the desired parameter using the Next Param or Prev Param buttons. Activate the desired Westgard Rule(s) by pressing the corresponding check button. Once edited, all future quality control measurements on the parameter for the given slot will be evaluated according to the selected Westgard Rule(s).

7. 8.

Select Westgard Rules for other parameters or levels in the same manner. To select/deselect all six Westgard Rules for every parameter of a displayed control solution slot, select the desired slot by pressing the Next Slot button.

Continued on next page

3-44

ABL800 FLEX Operator's Manual

3. Installation and setup

Quality control setup, Continued


Westgard Rules (continued) Step 9. 10. Action Press Select All or Deselect All. Verify information on the screen.

Press Continue. Changes are made and shown in the Westgard Rules Setup. Press Cancel. No changes are made. 11. Select or deselect the Westgard Rules for other slots in the same manner.

NOTICES: All future measurements taken on the slot will be evaluated with respect to the Westgard Rules if the Select All function was activated Settings will be removed under Deselect All. To have the possibility to return to the settings as they were, use the On/Off button. When a quality control measurement violates an applied Westgard Rule, a W is added to the parameter in the result. For a detailed explanation of how to interpret and evaluate results with respect to Westgard Rules, see chapter 7 of the Quality Control Reference Manual. Reference: Westgard JO, Barry PLL. Cost effective quality control: managing the quality and productivity of analytical processes. Washington: AACC Press, 1992.

Rilibk Ranges

The Rilibk Ranges program allows you to define a set of rules to control the maximum deviation of any parameter from the assigned target value. The target values are given on the QC insert. It is possible to define more than one rule for the individual parameters.
Continued on next page

3-45

3. Installation and setup

ABL800 FLEX Operator's Manual

Quality control setup, Continued


Rilibk Ranges (continued) To activate or deactivate the Rilibk Rules, do the following. Step 1. Action Press the On/Off button to activate/deactiviate the assigned Rilibk rules.

To add a new Rilibk rule, do the following: Step 1. Action Press the Add button to display the screen below:

2. 3. 4. 5. 6.

Select the desired parameter from the parameter list shown in the right side of the screen. Press the Left arrow button until the first "Lower Limit" box is highlighted and enter the desired lower limit. Highlight the next box and select "<" or "<=". Highlight the first "Upper Limit" box and select "<" or "<=". Highlight the next "Upper Limit" box and enter the desired upper limit.
Continued on next page

3-46

ABL800 FLEX Operator's Manual

3. Installation and setup

Quality control setup, Continued


Rilibk Ranges (continued) Step 7. 8. 9. Action To select the desired +/- range press the desired radio button. Enter the desired +/- range in the "Ranges" box. Press Back to return to the Rilibk Ranges screen. The added Rilibk rule is now shown in the screen.

To edit a Rilibk rule, do the following: Step 1. Action Select the desired rule in the Rilibk Ranges screen and press Edit to display the screen below:

2. 3.

Use the Right or Left arrows to jump between the input boxes and edit the desired values. Press Back to return to the Rilibk Ranges screen.

To remove a Rilibk rule, do the following: Step 1. Action Highlight the desired rule in the Rilibk Ranges screen and press Delete.

NOTICE: When a quality control measurement violates an applied Rilibk Rule, a red R is shown in front of the parameter name in the result.

3-47

3. Installation and setup

ABL800 FLEX Operator's Manual

Replacement setup
Programs Replacement Setup includes the programs described in this section.

Activate a button to enter the program. Replacement schedule This program allows you to schedule routine replacements along with the current scheduled date and interval for replacement. The settings selected here are then used in the Replacements on the Analyzer Status screen. To edit the schedule of replacement actions, do the following: Step 1. Action Highlight the replacement action to be scheduled and press Edit.

Continued on next page

3-48

ABL800 FLEX Operator's Manual

3. Installation and setup

Replacement setup, Continued


Replacement schedule (continued) Step 2. Action Change the interval for the selected replacement action (displayed in the "Action" box) with the up/down arrow buttons (see recommended replacement intervals below).

3. 4. 5. Recommended replacement intervals

NOTICE: The replacement schedule reminder "Lock analyzer when 10 % overdue" (selected in Corrective Actions see further in this chapter) will work on the basis of the setting selected in the "Action" box. Touch to highlight the "Next date" box and change the date using screen keypad. Confirm with Enter. Press Back to return to the Replacement Setup screen. Repeat steps 1-4 for each replacement action to be scheduled.

The following replacement intervals are recommended by Radiometer. NOTICE: The replacement intervals below are guidelines only. Under no circumstances do they guarantee the lifetime of the replacement items. For analyzers with high throughput the replacement intervals should be adjusted accordingly in the Replacement schedule. For an average of 40 samples per day: Action Replacing Crea A and Crea B electrode membranes Replacing metabolite membranes Replacing blood gas and electrolyte electrode membranes Replacing pump tubing Replacing inlet gasket Replacing fan filter Replacing electrodes Replacing reference electrode membranes: 40 samples per day: All analyzers, except ABL8x7 FLEX analyzers 40 samples per day: >40 samples per day: ABL8x7 FLEX analyzers only ABL8x7 FLEX analyzers only Interval 14 days 1 month 3 months 6 months 6 months as required as required

1 month 14 days 1 week

The status of solutions, waste container and gas cylinders are not scheduled and can be viewed in the Replacements on the Analyzer Status screen.
Continued on next page

3-49

3. Installation and setup

ABL800 FLEX Operator's Manual

Replacement setup, Continued


User Activities This program allows you to formulate and schedule your own activities (e.g. cleaning analyzer, replacing printer paper, etc.) along with the current scheduled date and interval. The settings selected here are then used in the Replacement Status. To add the user activity, do the following:

Adding a user activity

Step 1.

Action Press Add to display the Edit User Activities Schedule screen.

2.

To add text, press the keyboard icon and type the activity. Confirm with Enter on the keyboard.

Choose the Interval with the up and down arrow keys in the "Interval" box: 1-2-3-4-5-6-7-10-14-20-30 days or 2-3-4-6-8-12-18-24 months. 3. 4. 5. Type the "Next date" using screen keypad. Confirm with Enter. Press Back to return to the User Activities screen. Repeat steps 1-4 for each activity to be scheduled.
Continued on next page

3-50

ABL800 FLEX Operator's Manual

3. Installation and setup

Replacement setup, Continued


Editing a user activity To edit a user activity, do the following: Step 1. 2. Action Highlight the desired user activity on the User Activities screen. Press Edit. Press the keyboard icon to edit the text. Confirm the text with ENTER.

Change, if desired, the interval or next date (confirm the date with Enter). 3. 4. Deleting a user activity Press Back to return to the User Activities screen. Edit other user activities in the same manner.

To delete a user activity from the list, do the following: Step 1. 2. Action Press Delete on the User Activities screen. Press Continue to delete the activity.

Press Cancel to return to the User Activities screen.


Continued on next page

3-51

3. Installation and setup

ABL800 FLEX Operator's Manual

Replacement setup, Continued


Solution warning This program allows you to select a solution level value in % below which a warning will be given. To select the solution level value for the warning to be activated (default value is 25 %), do the following: Step 1. Action Highlight the desired solution bottle: touch the bottle on the screen or use Next Bottle or Prev. Bottle.

2. 3. 4.

Select the solution warning level (in %), using Increase or Decrease. Select warning values for solution levels in other bottles in the same manner. Press Close to confirm the settings.

Reagents Expiration Warning

This program allows you to set the time for the analyzer to give a warning before the calibration and cleaning solutions with additives will expire. To set the time for the warning, do the following: Step 1. Action Select time, using the up/down arrow buttons in the "Give warning before reagents expire" box. The actual bottles of the analyzer depend on the analyzer type.

2.

Press Close to exit the program and confirm the settings.


Continued on next page

3-52

ABL800 FLEX Operator's Manual

3. Installation and setup

Replacement setup, Continued


Solution Level Forecast This program allows you to change the solution level forecast so that the analyzer will predict the solution levels in the bottles more precisely. It will improve both the graphical presentation on the Replacement Status screen and the accuracy of the solution low level warning. To change solution level forecast, do the following: Step 1. Action Select the solution bottle, using the Next bottle or Prev. bottle button (the Rinse solution has been selected on the screen).

2.

Select the true level (in %) with the up and down arrow buttons on the screen. The bottom of each solution bottle has markings for solution levels.
Rinse Solution Other Solutions

25 % min.

25 % min.

You can also remove the solution bottle to check the level, and the analyzer will not enter Hold mode. However, the Refill will be performed as soon as the bottle has been mounted back again. The selected setting will be displayed on the Analyzer Status screen. The screen below is an example.

Continued on next page

3-53

3. Installation and setup

ABL800 FLEX Operator's Manual

Replacement setup, Continued


Solution Level Forecast (continued) Step 3. 4. 5. Action Select solution level forecasts for other solutions in the same manner. To reset all solution level forecasts to the default values: = -1.00 and = 100.0, press the Set Default button. Press Close to confirm the settings.

AutoCheck Replacement Warning

This program allows you to select the time for a reminder to appear in the System Messages in order to fill up the carousel. To select time for the reminder, do the following: Step 1. Action Use the up/down arrow buttons to select one of the settings: No, 1 hour, 2 hours, 4, 8, 12, 24, 48 hours.

2.

Press Close to exit the program.

3-54

ABL800 FLEX Operator's Manual

3. Installation and setup

Parameters and input setup


Programs Parameters and Input Setup includes the programs described in this section.

Activate a button to enter the program. Parameters This program allows you to do the following: disable or enable a parameter repress parameters if any problems are detected lock a parameter make user-defined corrections for each measured parameter make out-of-range suppression of oximetry parameters and bilirubin. To disable/enable a parameter, do the following: Step 1. Action Highlight a parameter on the screen, using the scroll facilities.

2.

Press Enable/Disable to include or exclude the parameter from a parameter profile and the parameter bar. Note that pH, pCO2 and pO2 cannot be excluded.
Continued on next page

3-55

3. Installation and setup

ABL800 FLEX Operator's Manual

Parameters and input setup, Continued


Parameters (continued) Parameters can be locked/unlocked from the RADIANCE system. To lock or unlock a parameter from the analyzer, do the following: Step 1. Action Make sure the analyzer is not connected to the RADIANCE system. Highlight a parameter on the screen, using the scroll facilities.

2.

Press Lock/Unlock. (This button is grayed out if the analyzer is connected to the RADIANCE system.) The locked parameter will show red on the parameter bar and will change the overall analyzer status traffic light on the Analyzer Status button to red. The parameter value will be absent from the printout. However, the locked parameter will be calibrated.

3.

To unlock a parameter, highlight it and press Lock/Unlock. The traffic light on the Analyzer Status button will change from red to a color corresponding to the analyzers overall status.

To edit the parameter setup, do the following. Step 1. 2. Action Highlight the desired parameter and press Edit. Activate (or deactivate) the following check buttons to select (or deselect) the following functions:

Highlighted parameter

Continued on next page

3-56

ABL800 FLEX Operator's Manual

3. Installation and setup

Parameters and input setup, Continued


Parameters (continued) Step 2. (cont.) Action Repression (Repress value in patient result, if the parameter is marked with a question mark, either because there is a calibration problem or QC problem). Out of range suppression for oximetry parameters or ctBil only. When activated, this function is applied to the oximetry/ctBil results (including those obtained in the past) as follows: Oximetry parameters and ctBil values inside the measuring range but lower than 0 or higher than "100 %" will be shown as 0 or "100 %", respectively. ctBil values higher than "1000 mol/L" but inside the measuring range will be shown as "1000 mol/L".

3.

Use the keypad to enter the following for the highlighted parameter: Correction offset. Confirm the entry with Enter. Correction slope. Confirm the entry with Enter.

4.

Press Back to return to the previous screen and repeat steps 1-3 for another parameter, if desired.

NOTICE: User-defined corrections for blood measurements will influence the measurement results from blood and QC analyses and change the specific performance characteristics unless Apply parameter corrections to QC was disabled in Miscellaneous Setup.

Units

This program allows you to select the unit for each parameter or group of parameters. To select the unit for a group or for a single parameter, do the following: Step 1. Action Highlight the group (i.e. parameters which use the same unit as "pressures" = (pCO2, pO2, Baro, etc) or a parameter, using the arrow buttons. Select the desired unit, using the arrow buttons.
Continued on next page

2.

3-57

3. Installation and setup

ABL800 FLEX Operator's Manual

Parameters and input setup, Continued


Units (continued) Step 3. Action Confirm the selected unit with Select.

4. 5.

Change units for other parameters in a similar manner. Press Close to return to the Ready screen.

User-defined Patient Data Items

This program allows you to include other patient data in the Patient ID layout than those already available there. To add a new item to the list, do the following: Step 1. Action Press Add.

2.

Follow the procedure for editing an item from step 2 see the next page.
Continued on next page

3-58

ABL800 FLEX Operator's Manual

3. Installation and setup

Parameters and input setup, Continued


User-defined Patient Data Items (continued) To edit an item in the list, do the following: Step 1. 2. Action Highlight an item in the list and press Edit. To change the name, press the keyboard icon on the keypad and type in the new name of up to 20 characters. Confirm with Enter on the keyboard. Select the data type with the up/down arrow buttons and press Select.

3.

For "Text" entry proceed to step 8. For "Numerical" entry proceed to step 4. 4. Highlight "Unit". The keypad is now displayed. Press the keyboard icon on the keypad and type in the new name of up to 20 characters. Confirm with Enter. "Decimals" is now highlighted and the box with "None", "1", "2","3" is displayed. Choose the number of decimals with the up/down arrow keys and press Select to confirm. Highlight "Max. value". Type in the value and confirm with Enter on the keypad. "Min. value" is now highlighted. Type in the value and confirm with Enter on the keypad.
Continued on next page

5.

6. 7.

3-59

3. Installation and setup

ABL800 FLEX Operator's Manual

Parameters and input setup, Continued


User-defined Patient Data Items (continued) Step 8. Action "Use selection list" function:

Press the check button. To make a list, press Add, type in the item on the displayed keyboard (up to 20 characters) and confirm with Enter. Add as many items as you wish in the same manner. The check button can be activated only if the selection list contains two or more items. 9. Press Back to return to the User-defined Patient Data Items screen. The new entry will be included in the list.

To include the new items in a Patient ID Layout, press Menu - Utilities - Setup Analysis Setup - Patient Reports - Patient Report ID and follow the procedure described under Including an Item into Patient ID in Patient reports in this chapter.
Continued on next page

3-60

ABL800 FLEX Operator's Manual

3. Installation and setup

Parameters and input setup, Continued


User-defined Notes This program allows you to formulate and store notes. To make a Note text, do the following. Step 1. Action Activate one of the check buttons (e.g. Patient Result, Replacements, etc.). See also the remaining options by pressing Next Page.

List of Notes for a selected option.

2.

Press Add New and type the text for the Note, using the screen keyboard. Confirm with Enter to save the text and return to the previous screen. Your Note is now included in the list. To edit a Note, highlight it in the Notes box, using the up and down arrow buttons, and press Edit. The keyboard will be displayed with the Note text on it. Edit the text and confirm with Enter.

3.

4. 5. 6.

To delete a Note, highlight it in the Notes box, using the up and down arrow buttons, and press Remove. Make notes for other options in the same manner. Press Back to exit to the Ready screen.

If a list of Notes has been made for a given option, it will be marked with a pencil icon on the relevant screen(s).

3-61

3. Installation and setup

ABL800 FLEX Operator's Manual

Analyzer settings
Programs The Analyzer Settings includes the programs described in this section.

Activate a button to enter the program.

Analyzer ID

This program allows you to change the analyzers identification. To enter analyzer ID do the following: Step 1. Action Touch and highlight the "Analyzer name" box.

2.

Type in an identification name and/or number for the analyzer (up to 32 characters), using the screen keypad or keyboard. Confirm with Enter.

NOTICE: Installation number cannot be changed. Quote this number in any technical inquiries you have to Radiometer.
Continued on next page

3-62

ABL800 FLEX Operator's Manual

3. Installation and setup

Analyzer settings, Continued


Environment Setup This program allows you to enter altitude above sea level, to adjust the built-in barometer in accordance with the reference barometer in your laboratory and to enter the ambient temperature required for the Crea reagents.

To enter the altitude at which your analyzer is placed (this is generally used for analyzers at particularly high altitudes), do the following: Step 1. 2. Action Touch the "Altitude above sea level" box to activate. Type in the height above sea level in meters, using the keypad. Confirm with Enter.

At high altitudes, the difference between pCO2 values in blood and in the QC solutions is very small due to the high buffer capacity of blood and the QC solutions, and is considered to be negligible. However, the aqueous QC solutions have a different capacity than blood to take up oxygen, and correcting for different pO2 QC values at high altitudes is imperative as deviations from the respective values at sea level are significant. To correct for the different pO2 values, the QC control ranges are modified according to the altitude. For detailed information, please refer to the Quality Control Systems Reference Manual, chapter 4. NOTICE: To make altitude corrections operative for the installed QC solutions, re-enter the QC solutions in their slots and re-scan their barcodes. To adjust the internal barometer, do the following: Step 1. 2. Action Touch the "Adjust to:" in the "Barometer values" box to activate. Key in the desired pressure value on the keypad and confirm with Enter. The value will be shown in the "Measured adjusted" box. Maximum accepted correction is 19 mmHg (i.e. the difference between the "Measured unadjusted" and "Measured adjusted" settings).
Continued on next page

3-63

3. Installation and setup

ABL800 FLEX Operator's Manual

Analyzer settings, Continued


Environment Setup (continued) Step Action

Barometer pressure limits are 450-800 mmHg, or 60.0-106.7 kPa, or 2 (cont). 450-800 torr. The units are selected in the Setup program: Units. Check the barometer of the ABL800 FLEX analyzer against a known calibrated barometer at least once a month. See ranges in chapter 13 in this manual. To enter the ambient temperature, do the following: Step 1. 2. Action Touch the "Ambient temperature" box to activate. Key in the new temperature, using the keypad. Confirm with Enter.

The concentrations of creatinine and creatine in calibration solutions depend on the ambient temperature. This dependency is well-known and taken into account as follows: the analyzer calculates the correct concentrations, using the equilibration constants, time and the ambient temperature entered by user in the Environment Setup (Menu > Utilities > Setup > General Setup > Analyzer settings). For ambient temperatures between 15 C and 28 C or temperatures fluctuating through the entire operating range 15-32 C of the analyzer, enter 23.5 C. For ambient temperatures constantly above 28 C and up to 32 C enter 30 C. Using these settings will provide analyzer performance according to the performance characteristics specified in the ABL800 Reference manual, chapter 5. In the temperature-controlled environment with constant temperature enter the average ambient temperature in the Environment Setup in order to improve the creatinine performance. If an error in the keyed-in ambient temperature is about 5 %, the bias in the reported cCrea results will be < 0.5 % unless creatine concentration in a sample is cCrea concentration.
Continued on next page

3-64

ABL800 FLEX Operator's Manual

3. Installation and setup

Analyzer settings, Continued


Time/Date setup This program allows you to change the current time and date setting. To reset the time and date, do the following: Step 1. Action Highlight the "Time" box by touching it on the screen.

2.

Key in the time on the screen keypad. Confirm with Enter. A separator is automatically added between hours, minutes and seconds. Repeat steps 1-2 to reset the Date. To revert to the previous settings, press Current.

3. 4. Acoustic signal setup

This program allows you to set up a short beep to sound after certain events. To activate the acoustic signal, do the following: Step 1. Action Activate the desired check button(s).

2.

Select volume for the acoustic signal or activate the "Mute all acoustic signals" check button.
Continued on next page

3-65

3. Installation and setup

ABL800 FLEX Operator's Manual

Analyzer settings, Continued


Acoustic signal setup (continued) The following events are available: Event Explanation

Value exceeds critical One of the measured values exceeds the specified range critical limits for that parameter. Close inlet Result is ready Inlet is open too long Beep before AutoCheck opens Language The inlet should be closed. A sample has been analyzed and the results are ready for viewing. The inlet should be closed. A beep sounds to indicate that an AutoCheck measurement is about to take place.

This program allows you to select a language of your choice from the list of languages available on your analyzer. To change to another language do the following: Step 1. Action Select the desired language with the arrow buttons.

Press Set Language. 2. Press Continue to continue or press Cancel to continue operating the analyzer with the language unchanged.

Continued on next page

3-66

ABL800 FLEX Operator's Manual

3. Installation and setup

Analyzer settings, Continued


Language (continued) Step 3. Action If you pressed Continue: Press Continue once again to restart the analyzer right away and to use the new selected language right away or Press Cancel to continue operating the analyzer with the current language until the next restart, where the language then will be changed.

3-67

3. Installation and setup

ABL800 FLEX Operator's Manual

Communications
Programs The Communications Setup programs are described in this section.

Activate a button to enter the program. Refer to Chapter 2, Communication Ports for the identification and location of a serial RS232 interface connection (COM2) and a network (TCP/IP) RJ45 ethernet connection. This program allows you to connect the analyzer to the RADIANCE system. RADIANCE Communication NOTICE: Connecting the analyzer to the RADIANCE system should be Setup performed by the RADIANCE administrator in your institution. To enable communication with the RADIANCE system, do the following: Step 1. Action Touch and highlight the "Server address" box.

Type in the TCP/IP address of your RADIANCE PC, using the screen keypad or keyboard. 2. 3. Touch and highlight the "Port" box. Type in port number, using the keyboard. Touch and highlight the "Password" box. Type in your RADIANCE password, using the keyboard.
Continued on next page

3-68

ABL800 FLEX Operator's Manual

3. Installation and setup

Communications, Continued
RADIANCE Communication Setup (continued) Step 4. Action Press the check button in the "Radiance Communication" box to activate connection.

The Icon in the "Connection status" box indicates the state of the RADIANCE connection. "Connected" indicates an established connection to the RADIANCE system.

The 5.

icon will indicate the established connection as well.

The "Output queue" box shows the number of data queued up for transfer. It will be sent to the RADIANCE system. Otherwise clear the queue by activating the recycle bin icon. Press Close to exit the screen.

6.

LIS/HIS Connection Setup

This program allows you to select the communication protocol for a connected device. To select the communication protocol, do the following: Step 1. Action Press Add on the screen.

The "Output queue" box shows the number of data queued up for transfer. It will be sent to LIS/HIS. Otherwise clear the queue by activating the recycle bin icon.
Continued on next page

3-69

3. Installation and setup

ABL800 FLEX Operator's Manual

Communications, Continued
LIS/HIS Connection Setup (continued) Step 2. Action Press Keyboard, type the name of the connection instead of the default one, and press Enter.

Press Back to return to LIS/HIS Connection Setup screen. 3. Select the high-level protocol according to the requirements of the connected device.

Select between: ASTM, ASTM6xx, HL7 version 2.2, HL7 version 2.5 or POCT1-A DML, using the up/down arrows in the box. 4. Select the low-level protocol as follows: Use Serial or Serial (RAW) for the serial connection. Use Network (TCP/IP) for the network connection. Use Network (TCP/IP)ASTM" for additional serial connection (not all combinations of high-level protocols and low-level protocols are possible).
Continued on next page

3-70

ABL800 FLEX Operator's Manual

3. Installation and setup

Communications, Continued
LIS/HIS Connection setup (continued) Step 5. Action a. Connection specifications for serial low-level protocol: Press Edit to display the Connection Specifications screen below.

b. Connection specifications for network low-level protocol: Press Edit to display the Connection Specifications screen below (HL 7 version 2.5 is used as an example).

c. Connection specifications for POCT1-A DML low-level protocol:

Continued on next page

3-71

3. Installation and setup

ABL800 FLEX Operator's Manual

Communications, Continued
LIS/HIS Connection Setup (continued) Step 6. Action a. Connection specifications for serial low-level protocol: Use the up and down arrows in each box to select baud rate, stop bits, data bits and parity. Port number should be COM2. Box Baud rate Data bits Stop bits Parity Option 1200, 2400, 4800, 9600, 14400, 19200, 38400, 115200 5, 6, 7, 8 1, 1.5, 2 Even, odd, none.

b. Connection specifications for network low-level protocol: Touch the screen to highlight the following boxes one after another: Server address Port number Reconnect Interval (time interval in seconds to attempt re-connection if connection failed) Retransmit max Transmit timeout. Use the keypad/keyboard to enter the relevant information. c. Connection specifications for POCT1-A DML low-level protocol: Touch the screen to highlight the following boxes one after another: Server address Port number Reconnect Interval. Use the keypad/keyboard to enter the relevant information.

QA Portal

This program allows you to connect the analyzer to the QA portal. If the QC Portal communication is enabled, the analyzer will automatically send QC results and Cal Verification measurements to the QA portal. To enable communication with the QA portal, do the following: Step 1. Action Touch and highlight the "Server address" box.

Type in the TCP/IP address of your QA portal, using the screen keypad or keyboard.
Continued on next page

3-72

ABL800 FLEX Operator's Manual

3. Installation and setup

Communications, Continued
QA Portal (continued) Step 2. 3. Action Touch and highlight the "Port" box. Type in port number, using the keyboard. Press the check button in the "Enable communication with QA Portal" box to activate connection.

The Icon in the "Connection status" box indicates the state of the QA portal connection. "Enabled" indicates an established connection to the QA portal.

The icon in the information bar will indicate the established connection as well. 4. Press Close to exit the screen.

3-73

3. Installation and setup

ABL800 FLEX Operator's Manual

Communications, Continued
Automatic data request This program allows you to select the conditions for requesting patient demographics automatically from the connected the RADIANCE system or from LIS/HIS computer system when entering Patient ID, Accession Number or Sampler ID. FLEXQ: Manual measurement: The request will be made during scanning the sampler. The request will be made when the corresponding data is entered on the Patient Identification screen during or after analysis.

To select the settings, do the following: Step 1. Action Select a connected device in the "From connection" box, using the up/down arrows buttons.

2.

Activate the relevant check button(s) to request patient demographics from the connected RADIANCE system or LIS/HIS when: entering Patient ID entering Accession Number entering Sampler ID.

3.

Press Close when completed.

NOTICE: If the requested patient data (e.g. Patient Last Name) was received after exiting the Patient Identification screen, the patient report will be stored without the requested data in the Patient Report log. The requested patient data will be stored as a patient profile in the analyzer's database without, however, being attached to any patient report.
Continued on next page

3-74

ABL800 FLEX Operator's Manual

3. Installation and setup

Communications, Continued
Automatic Data This program allows you to set up automatic sending of data to a connected LIS/HIS computer system or to the RADIANCE system. Transmission Setup To select the settings, do the following: Step 1. Action Highlight a desired connection device on the screen, using the up/down arrows (note that the "FLEXQ Sample Status" is not available if connection to the RADIANCE system has been selected).

2.

Activate the relevant check button(s) to select the data to be sent to the highlighted connection.

NOTICE: If the requested patient data (e.g. Patient Last Name) was received after exiting the Patient Identification screen, the patient report will be transmitted without the requested data. To prevent this, select one of the patient data items transferred from LIS/HIS as mandatory. Patient Lookup Setup This program allows you to select the data source in order to obtain the patient information on the Patient Identification screen. To select the settings, do the following: Step 1. Action Select a data source from the established connections (local database, RADIANCE or LIS/HIS).

Continued on next page

3-75

3. Installation and setup

ABL800 FLEX Operator's Manual

Communications, Continued
Patient Lookup Setup (continued) Step 2. Action Select the number of days you want each patient to be kept in the list, using the up and down arrow buttons in the box.

Remote Support This program allows a remote operator, e.g. a service technician, to log onto your analyzer and monitor its functions. To select access, do the following: Step 1. Action Press the "Enable remote access" check button to allow a remote operator to log onto your analyzer.

When connected (logged on), the "Remote operator" appears on the analyzer status bar. 2. Deactivate the check button to prevent a remote operator logging onto your analyzer.

NOTICE: To select the access profile for the remote operator see description in General Security in Analyzer Security earlier in this chapter.

3-76

ABL800 FLEX Operator's Manual

3. Installation and setup

Printers
Programs The Printers programs are described in this section. Activate a button to enter the program.

Automatic Printing

This program allows you to select automatic printout of patient, QC and calibration results, activity log messages, and registration receipts. A registration receipt contains patient data and the barcode with the analyzer ID and the sampler number. Scanning the barcode will display the corresponding result. To select options available on the screen, do the following: Step 1. Action Activate the desired check buttons for the automatic printout of the patient results, QC results, calibration results, activity log messages and registration receipts for FLEXQ.

The last line will not be visible on the analyzers without the FLEXQ module. 2. Select automatic printout of several copies of patient results (from 1 to 5), using the up/down arrow buttons in the "Patient results print options" box.
Continued on next page

3-77

3. Installation and setup

ABL800 FLEX Operator's Manual

Printers, Continued
Automatic Printing (continued) Step 3. Action Press "User", "Manager" or "Service" in the "Message level" box to select the level for the messages in the Activity log.

Printer Setup

This program allows you to set up other printers than the analyzer's printer for making printouts. To select a printer for the analyzer, do the following: Step 1. Action Highlight a printer from the list by using the arrow buttons.

2.

Press Select/Deselect to select the highlighted printer for printing. You can install any number of printers, but only up to 10 printers can be selected. To display the list of printers every time the Print button is pressed, activate the check button in the "Manual printing" box. If not activated, all the selected printers will make a printout every time the Print button is pressed.

3.

4.

To change the highlighted printer's name, press Edit Name to display the keyboard. Type a name and confirm the entry with Enter.
Continued on next page

3-78

ABL800 FLEX Operator's Manual

3. Installation and setup

Printers, Continued
Printer Setup (continued) Step 5. Action To install a new printer, press the Install Printer button.

This function can be used by Radiometer service engineer or a person with network knowledge.

3-79

3. Installation and setup

ABL800 FLEX Operator's Manual

Disk Functions setup


Programs The Disk Functions Setup programs are described in this section.

The programs are Automatic Backup and Automatic Archiving. Automatic Backup Setup This program allows you to select automatic backup of all data and system files. Do the following: Step 1. Action To select automatic backup of all data and system files, activate the check button.

2. 3.

Select time for auto backup by highlighting the "Time" box and typing the time, using the screen keypad. Confirm the entry with Enter. Enter the interval between subsequent backups in the "Interval" box and type the number of days, using the screen keypad. Confirm with Enter.
Continued on next page

3-80

ABL800 FLEX Operator's Manual

3. Installation and setup

Disk Functions setup, Continued


Automatic Backup Setup (continued) Step 4. 5. Action Press the drive icon in the Destination box to select destination. Highlight the drive or folder and press the Expand/Collapse button to open a folder in a directory or within a folder.

Note that automatic backup can be selected for the internal hard disk or the network. When completed, the correct destination should appear in the upper part of the box.

This program allows you to select automatic archiving of the data logs by Automatic Archiving Setup activating the relevant check button. Do the following: Step 1. Action To select automatic archiving on the analyzer's hard disk, activate the check button.

2.

To select another destination, press the drive icon to select a directory as described in step 5 for Automatic Backup Setup above.
Continued on next page

3-81

3. Installation and setup

ABL800 FLEX Operator's Manual

Disk Functions setup, Continued


The oldest records (500 patient reports, quality control or calibration results, or Automatic Archiving Setup 2000 entries in the Activity log) will be automatically removed from a data log and placed into the relevant archive. The archives can be stored on the analyzer's hard (continued) disk and viewed in "Archived Data logs" or at a remote location. For detailed information on archiving the old data, please refer to chapter 8: Disk functions in this manual.

3-82

ABL800 FLEX Operator's Manual

3. Installation and setup

Corrective actions
Purpose This program allows you to do the following: to select corrective actions for the events listed in this program, to select the traffic light indication, if available, for an event, to select analyzer action for the subsequent measurements. To specify the corrective action and traffic light signal (if available) for each condition listed in the "Condition" box, do the following: Step 1. Action Highlight the desired condition using the up/down arrows in the box.

2. 3.

Select an action for this condition, using the arrow buttons in the "Corrective action(s)" box see the table below. Select the desired traffic light signal (yellow or green, if available) for the specified event by pressing the traffic light in the "Traffic light signal" box see the table below. Select corrective actions/traffic light signal for the other conditions in a similar way.

4.

Conditions and correction actions

Conditions and corresponding corrective action choices are as follows: Condition Calibration error(s) present Corrective action ? on specific parameters Perform AutoCheck measurement Do not run scheduled AutoCheck Calibration schedule reminder(s) QC error(s) present Message on next Patient Result ? on specific parameters Green or yellow
Continued on next page

Traffic light

3-83

3. Installation and setup

ABL800 FLEX Operator's Manual

Corrective actions, Continued


Conditions and corrective actions (continued) Condition QC schedule reminder(s) Corrective action Message on next Patient Result Lock analyzer immediately when overdue Message on next Patient Result Lock analyzer when 10 % overdue System message(s) present User activity reminder(s) AutoCheck QC error(s) present Re-run same level once Re-run same level twice Message on next Patient Result Traffic light Green or yellow Green or yellow Green or yellow Green or yellow Green or yellow

Replacement schedule reminder(s)

AutoCheck scheduled Use ampoule last in schedule level missing Explanation of corrective actions The explanation of the corrective actions is given below. Corrective action ? on specific parameters Message on next Patient Result Lock analyzer immediately when overdue Explanation

The affected parameter(s) will be marked with ? in subsequent patient results. To mark the subsequent patient results on the Message screen If a quality control measurement is not run 1 minute after the scheduled time for it, the analyzer will be locked. As a QC takes more than 1 minute to complete, the analyzer will enter the "Locked" state during measurement. It will, however, automatically leave the "Locked" mode again, if the quality control measurement results are valid. If a scheduled replacement procedure is more than 10 % overdue compared with the scheduled time for it, the analyzer will be locked. If an error has been registered during a scheduled AutoCheck measurement, it will be re-run provided the control solution is available in the carousel.
Continued on next page

Lock analyzer when 10 % overdue Re-run same level once

3-84

ABL800 FLEX Operator's Manual

3. Installation and setup

Corrective actions, Continued


Explanation of corrective actions (continued) Corrective action Re-run same level twice Use ampoule last in schedule Explanation If an error has been registered during a scheduled AutoCheck measurement, it will be repeated twice provided the control solution is available in the carousel. If a control solution scheduled for AutoCheck is missing in the carousel, the measurement will be on the control solution type scheduled before this one.

NOTICES:

The critical system messages will always result in a red traffic light signal. If the analyzer locks due to QC or replacement schedule reminders, a critical system message will be generated and the traffic light signal will be red The specified traffic light signal and the messages will continue to appear until the condition no longer exists.

3-85

3. Installation and setup

ABL800 FLEX Operator's Manual

Miscellaneous setup
Purpose This program allows you to select the following options (use the arrow buttons to display the rest of the options):

List of options

The options are as follows: Option Analyzer locked Function Suspends all measurements on the analyzer; the other functions such as calibrations and service programs are still enabled. The analyzer can be locked via this program or via a lock command from an externally connected system, e.g. LIS or the RADIANCE system. After the syringe inlet has been opened, two Start buttons will be displayed: one for the samplers without a safeTIPCAP and the other for samplers with a safeTIPCAP. If the check button is disabled, only the Start button for samplers without a safeTIPCAP will be available. Enable estimated derived parameters Fixed pO2/pCO2 decimals Enable general barcode support Enables estimation of the derived parameters based on default values even though the measured parameters have been deselected or are not available. If enabled, these parameters will be reported with a fixed number of decimals. Enables every text box on the Patient Profile, Patient Identification, Patient Result, QC ID and Recording Fluid Replacement screens where it is possible to enter a barcode.

Use safeTIPCAP

Enable patient result Enables the additional buttons on the Patient Result approval screen used for approval of the result. For detailed information please refer to chapter 4: Sample measurements. Apply parameter corrections to QC If enabled, the user-defined corrections (slope and offset) will be applied to the quality control results.
Continued on next page

3-86

ABL800 FLEX Operator's Manual

3. Installation and setup

Miscellaneous setup, Continued


List of options (continued) Option Log all measurement activities Function If enabled, "Ready", "Rinse", "Aspirating" and "Measurement" will be registered in the Activity log. Otherwise these activities will not be registered in the Activity log. This option aims to avoid too many entries in the Activity log. C will be automatically changed to F if the entered temperature is over the value of 45. The screen saver will show up if the analyzer was idle for 10 minutes.

Auto temp. unit conversion Show screen saver

Show parameter bar If disabled, the parameter bar will not be shown on the Ready screen. To activate/deactivate an option, do the following: Step 1. 2. 3. Selecting HbF correction option Action Scroll the list of options with the up/down arrow keys. Highlight the option and press the check button ( ) beside it. To deactivate the option, press the check button again. Press Close to confirm the settings and return to the Ready screen.

This option disables HbF correction for all levels, or enables it for all levels or for HbF levels higher than 20 % (HbF correction will be automatically disabled if FCOHb < 0.5 % in order to decrease a false HbF detection on adult blood samples and to ensure non-biased oximetry results). The guidelines for selecting/deselecting HbF correction are as follows: For neonatal samples: Use "Enabled for all levels". It is important to enable HbF correction in order to obtain correct results for ctBil, sO2, FO2Hb, FMetHb, FCOHb and FHHb. For adult samples: Use "Disabled" or "Enabled for levels > 20 %".

NOTICE: When an adult sample is measured with HbF correction "Enabled for all levels" or "Enabled for levels > 20 %", it will slightly affect measurement of sO2, FO2Hb, FMetHb, FCOHb and FHHb, and will cause a marginal number of adult samples reported with HbF present. If this is not acceptable, the HbF correction for adult samples can be turned off before a measurement see Selecting HbF Correction in the Analysis Setup in this chapter. To select the desired option, use the arrow buttons in the box.
Continued on next page

3-87

3. Installation and setup

ABL800 FLEX Operator's Manual

Miscellaneous setup, Continued


Analyzer message A message, sent from the RADIANCE system to the connected analyzer and displayed on the Ready screen, can be changed or deleted in this program as follows: Step 1. Action Press the keyboard icon to display the keyboard.

Type the message and confirm with Enter. To delete the current message, press Delete on the keyboard, or delete a message and type a new one if desired. NOTICE: The analyzer message can be up to 40 characters long. 2. Confirm the change with Enter on the keyboard to return to the Miscellaneous Setup screen.

Screen saver

To set the time for the screen saver to appear, do the following: Step 1. Action Check that the screen saver check button is activated.

2.

In the "Screen saver" box select the desired time with the arrow buttons.

3-88

4. Overview
Introduction

Sample measurements

This chapter describes how to perform measurements on patient samples and how to interpret the results. This chapter contains the following topics. General information ......................................................................................... Immediately before analysis ............................................................................ 4-2 4-9

Contents

Measurements with FLEXQ ............................................................................ 4-10 Introducing a blood sample.............................................................................. 4-12 Introducing an expired air sample.................................................................... 4-15 Patient identification ........................................................................................ 4-17 Patient result..................................................................................................... 4-21
)................................................................ 4-25 Calculation of FShunt and ctO2(a-v

Patient result messages..................................................................................... 4-26

4. Sample measurements

ABL800 FLEX Operator's Manual

General information
Measuring modes with FLEXQ With the FLEXQ module installed, it is possible to make the measuring process automatic (i.e. just place up to three samples in the sampler tray and go), or to place a single sampler in the sampler tray and perform a single measurement. The following rules apply to the batch mode: The analyzer is connected to the RADIANCE system All information is entered on the RADIANCE system via FLEXLINK Automatic data request is based on Sampler ID Measuring mode will always be the largest syringe mode, and it will be marked "Flex" Default layout is always used. The following rules apply to a single measurement: Parameter profile received from the RADIANCE system Measuring modes will be those syringe modes that can provide the requested parameter profile; these modes will be marked "Flex" Default layout is always used. Parameter profile not received from the RADIANCE system The measuring modes are those that can be selected on the analyzer (see Available modes and parameters below) The layout will be the one assigned to the selected measuring mode.

Available modes The modes and the measured parameters available on the ABL800 FLEX and parameters analyzers are listed below. Analyzer 805 Measuring modes Syringe S165 L Syringe S95 L Capillary C95 L Syringe S85 L pH+BG Capillary C55 L pH+BG Capillary C35 L MET Capillary FLEXMODE Syringe Pleura pH* Other fluids (any mode) Expired air
* See page 4-5

Measured parameters pH, pCO2, pO2, cK+, cNa+, cCa2+, cCl, cGlu, cLac pH, pCO2, pO2, cK+, cNa+, cCa2+, cCl, cGlu, cLac pH, pCO2, pO2 cGlu, cLac See below pH Depends on the selected mode pCO2, pO2
Continued on next page

4-2

ABL800 FLEX Operator's Manual

4. Sample measurements

General information, Continued


Available modes and parameters (continued) Analyzer 805 (cont.) 810 BG only Measuring modes Ampoule QC Syringe S85 L Capillary C55 L Capillary FLEXMODE Syringe Pleura pH* Other fluids (any mode) Expired air Ampoule QC 810/20/30 Syringe S85 L Capillary C55 L Measured parameters All available pH, pCO2, pO2 See below pH Depends on the selected mode pCO2, pO2 All available pH, pCO2, pO2 and 810: ctHb, sO2 820: ctHb, sO2, FO2Hb, FCOHb, FMetHb 830: ctHb, sO2, FO2Hb, FCOHb, FMetHb, ctBil, FHbF Capillary C35 L OXI 810: ctHb, sO2 820: ctHb, sO2, FO2Hb, FCOHb, FMetHb 830: ctHb, sO2, FO2Hb, FCOHb, FMetHb, ctBil, FHbF See below pH Depends on the selected mode pCO2, pO2 All available

Capillary FLEXMODE Syringe Pleura pH* Other fluids (any mode) Expired air Ampoule QC * See page 4-5

Continued on next page

4-3

4. Sample measurements

ABL800 FLEX Operator's Manual

General information, Continued


Available modes and parameters (continued) Analyzer 815/25/35 Measuring modes Syringe S195/95 L Capillary C95 L Measured parameters pH, pCO2, pO2, cK+, cNa+, cCa2+, cCl, cGlu, cLac and 815: ctHb, sO2 825: ctHb, sO2, FO2Hb, FCOHb, FMetHb 835: ctHb, sO2, FO2Hb, FCOHb, FMetHb, ctBil, FHbF Syringe S85 L pH+BG + OXI Capillary C55 L pH+BG+OXI pH, pCO2, pO2 and 815: ctHb, sO2 825: ctHb, sO2, FO2Hb, FCOHb, FMetHb 835: ctHb, sO2, FO2Hb, FCOHb, FMetHb, ctBil, FHbF Capillary C35 L MET Capillary C35 L OXI cGlu, cLac 815: ctHb, sO2 825: ctHb, sO2, FO2Hb, FCOHb, FMetHb 835: ctHb, sO2, FO2Hb, FCOHb, FMetHb, ctBil, FHbF Capillary FLEXMODE Syringe Pleura pH* Other fluids (any mode) Expired air Ampoule QC * See page 4-5
Continued on next page

See below pH Depends on the selected mode pCO2, pO2 All available

4-4

ABL800 FLEX Operator's Manual

4. Sample measurements

General information, Continued


Available modes and parameters (continued) Analyzer 837/27/17 Measuring modes Syringe S250 L Capillary C125 L Measured parameters pH, pCO2, pO2, cK+, cNa+, cCa2+, cCl, cGlu, cLac, cCrea and 817: ctHb, sO2 827: ctHb, sO2, FO2Hb, FCOHb, FMetHb 837: ctHb, sO2, FO2Hb, FCOHb, FMetHb, ctBil, FHbF pH, pCO2, pO2 and 817: ctHb, sO2 827: ctHb, sO2, FO2Hb, FCOHb, FMetHb 837: ctHb, sO2, FO2Hb, FCOHb, FMetHb, ctBil, FHbF Capillary C55 L pH + BG + OXI pH, pCO2, pO2 and 817: ctHb, sO2 827: ctHb, sO2, FO2Hb, FCOHb, FMetHb 837: ctHb, sO2, FO2Hb, FCOHb, FMetHb, ctBil, FHbF Capillary C35 L MET Capillary C35 L OXI cGlu, cLac 817: ctHb, sO2 827: ctHb, sO2, FO2Hb, FCOHb, FMetHb 837: ctHb, sO2, FO2Hb, FCOHb, FMetHb, ctBil, FHbF Syringe Pleura pH* Other fluids (any mode) Expired air Ampoule QC pH only Depends on the selected mode pCO2, pO2 All available

Syringe S85 L pH + BG + OXI

* NOTICE: The safeTIPCAP cannot be used with pleura fluids, and as a consequence of this the Pleura pH measuring mode is not available when placing a sampler in the FLEXQ module. The safeTipCap start button is also not available as an option when starting a manual pleura pH measurement NOTICE: In order to perform measurements in the 35 L mode in accordance with performance specifications it is highly recommended to use 35 L capillaries with mixing wire and a clot catcher.
Continued on next page

4-5

4. Sample measurements

ABL800 FLEX Operator's Manual

General information, Continued


FLEXMODE This mode allows you to analyze a blood sample of 35 L and higher up to the maximum volume accepted by your analyzer. Depending on the available sample volume, the FLEXMODE provides the highest number of parameters: from all available to as many as reliably possible. The following parameters can be obtained depending on the available sample volume (for the ABL835): Volume 35-40 L* 40-55 L* 50-70 L* 65-100 L* > 90 L* 195 L pH + OXI pH + OXI pH + OXI + BG pH + OXI + BG + MET pH + OXI + BG + MET + EL pH + OXI + BG + MET + EL + ctBil Parameter profiles Message No. ** 869 870 871, 872 873 874 No message

* All volumes are approximate volumes at ctHb = 15 g/dL. ** The messages will accompany the obtained result see Patient result messages at the end of this chapter. For interpretation of the messages see Troubleshooting messages in chapter 11 in this manual. 1. CAUTION notices CAUTION Presence of glucolic acid Never use analyzer for lactate measurements if there is any suspicion of presence of glycolic acid in blood as e.g. in case of ethylene glycol poisoning or xylitol infusion. Glycolic acid interferes with the lactate sensor resulting in erroneously high lactate readings 2. CAUTION Risk of incorrect result Visually inspect the measuring chambers for blockages when lifting the inlet flap prior to performing a measurement. If there is a blockage, close the flap and the analyzer will perform a rinse. If repeated visual inspection shows blockages in the measuring chamber, clean the electrodes and the measuring chamber according to step 7 in chapter 11: Troubleshooting, section "Fluid transport troubleshooting procedure" in this manual. 3. CAUTION - Risk of incorrect result Always remove air bubbles from a sample before introducing it into the analyzer. Air bubbles may result in erroneous pO2 values. 4. CAUTION - Risk of incorrect result During measurement, use the sample path view to observe the sample in the measuring chambers. If any bubbles are seen, disregard the result and repeat the measurement.
Continued on next page

4-6

ABL800 FLEX Operator's Manual

4. Sample measurements

General information, Continued


5. CAUTION notices (continued) 6. CAUTION - Risk of incorrect result When measurements are flagged always conduct the operator actions required by the troubleshooting program to prevent possible clots from remaining in the sample path. Fluid path clots may cause erroneous measurement results. CAUTION - Risk of erroneous results Always meticulously follow the sampling procedures described in chapter 12: Sampling. Failure to follow these procedures may introduce clots or air bubbles in the sample and yield erroneous results. When the FLEXQ is in use, the CAUTION notices 2 and 4 above do not apply.

Interrupting a measurement

To interrupt a measurement, press

(Stop) button.

Before using the To use the FLEXQ efficiently, the setup should be made thoroughly. FLEXQ Program Analysis Setup Syringe Modes Recommendations Make the report layout for each measuring program you are going to use, including the Patient ID items and parameter profile Select the default layout for each selected measuring program. This will define the number of measuring programs to select from before each measurement. Analysis Setup Sample Preregistration Analysis Setup Sample Logistics Select interpretation of the barcode entry.

Select batch mode for automatic measuring process or deselect it if single measurements are to be performed Select sample age for each measured parameter
Continued on next page

4-7

4. Sample measurements

ABL800 FLEX Operator's Manual

General information, Continued


Before using the FLEXQ (continued) Program Connections Recommendations Select connection to the RADIANCE system, LIS/HIS Select conditions for automatic patient data request Select data to be transmitted to the RADIANCE system, LIS/HIS Select patient data source (Patient Lookup) Printers Automatic Printing Select printing of the registration receipt for FLEXQ measurements

To make the measuring process automatic, the following is required: Batch mode selected in the Sample Logistics Setup Barcode interpretation (patient ID, Accession number, Sampler ID) selected in Sample Pre-registration and corresponds to that on the RADIANCE system Analyzer connected to the RADIANCE system Patient information entered on the RADIANCE system via FLEXLINK. Sampling See chapter 12 in this manual.

4-8

ABL800 FLEX Operator's Manual

4. Sample measurements

Immediately before analysis


Purpose It is necessary to mix a blood sample before introducing it into the system to ensure its homogeneity. If a sample is transferred from a syringe to an analyzer without being properly mixed, either the plasma phase or packed red blood cells may be analyzed, rendering the oximetry results meaningless. CAUTION - Risk of incorrect result Always remove air bubbles from a sample before introducing it into the analyzer. Air bubbles may result in erroneous pO2 values. CAUTION Risk of erroneous results Always meticulously follow the sampling procedures described in chapter 12: Sampling. Failure to follow these procedures may introduce clots or air bubbles in the sample and yield erroneous results. FLEXQ module Place a safePICO sampler in the FLEXQ sampler tray. The sample will be automatically mixed before the measurement. Mixing a sample For the analyzers without FLEXQ module, mix a blood sample in a sampling device as follows: Sampling device Description Mix a capillary sample with a magnet. Then move the mixing wire to the end of the capillary opposite to that from which the blood is to be aspirated. Remove both capillary caps. safePICO sampler or another syringe: Invert the syringe repeatedly and roll it between the palms of your hands.

Invert the test tube repeatedly. Then remove the cap.

CAUTION - Risk of incorrect result Use capillary tubes with a volume relevant for the selected measuring mode. Too small a volume will give the Insufficient sample error. Too large a volume will render the internal analyzer corrections invalid and the results inaccurate. Guidelines for handling capillary blood samples See chapter 12 in this manual.

4-9

4. Sample measurements

ABL800 FLEX Operator's Manual

Measurements with FLEXQ


To make a single measurement, do the following: Single measurement Step Action from the FLEXQ sampler Place the sampler in a slot on the tray. A short beep indicates that the 1. tray barcode has been accepted. 2. The available measuring modes are shown; the selected mode is highlighted. Change a measuring mode if desired. Press Proceed. 3. Enter, if required, additional patient data or mandatory data.

Press Accept to accept all the data (patient data, measuring mode, selected parameters, layout, etc.). 4. 5. 6. Remember to take the registration receipt printed out on the analyzers printer. Repeat steps 1-4 for another sampler. The measurement starts. The patient name and the time left till the result is ready are displayed on the Ready screen. If further editing of the patient data has to be made, press the button on the Ready screen to return to the Sample Registration screen. 7. When the result is ready, it will be automatically transmitted to its destination (the RADIANCE system or LIS/HIS).

Batch measurement

To make a batch measurement, do the following: Step 1. 2. 3. 4. Action Make sure that a slot indicator is green and ready to accept a sampler. Place the sampler in an empty FLEXQ sampler tray slot. A short beep indicates that the barcode has been accepted. Make sure that your sampler is ready for processing. If not, check analyzer connection to FLEXLINK. Proceed with the next sampler in the same manner as described above.
Continued on next page

4-10

ABL800 FLEX Operator's Manual

4. Sample measurements

Measurements with FLEXQ, Continued


An urgent measurement To bypass a queued measurement while the analyzer processes the samplers in the sampler tray, do the following: Step 1. Action Check that Urgent Manual Sample is active on the Ready screen and press it.

2.

If the urgent measurement violates parameter lifetime of one or more queued samples, accept violation and book the inlet for manual measurement by pressing Yes or abort booking of the inlet by pressing No. When allowed on the screen, open the inlet flap, insert the sampling device in the inlet and press Start see Introducing a blood sample in this chapter.

3.

Note that if the Urgent Manual Sample button is grayed out, the inlets are free and the sample can be analyzed immediately. Editing sample registration To change or add information to sample registration, do the following: Step 1. Action Locate the required sample in the "FLEXQ Status" on the Ready screen and press the Registration screen. 2. 3. button to proceed to the Sample

Change the patient data on the Sample Registration screen. Press Accept to confirm the patient data.

4-11

4. Sample measurements

ABL800 FLEX Operator's Manual

Introducing a blood sample


Syringe samples To analyze a blood or other fluid sample without the use of the FLEXQ sampler tray, do the following. Step 1. Action Check that the analyzer is in the Ready mode and scan the sample barcode.

Accept or cancel the displayed patient data (selected in Sample Preregistration setup program see chapter 3 in this manual). If no sample pre-registration was selected, proceed to step 2 below. 2. Lift the syringe inlet flap. The following screen appears:

If sample pre-registration was selected (see Analysis Setup in chapter 3), press Accept to accept displayed patient data or Cancel to cancel the patient data. 3. CAUTION - Risk of incorrect result Always remove air bubbles from a sample before introducing it into the analyzer. Air bubbles may result in erroneous pO2 values. Place the syringe tip firmly in the inlet as shown. SafePICO sampler: Do not remove the safeTIPCAP. 4. Select the desired mode by pressing the appropriate button.
Continued on next page

4-12

ABL800 FLEX Operator's Manual

4. Sample measurements

Introducing a blood sample, Continued


Syringe samples (continued) Step 5. Action Press Start with the safeTIPCAP for safePICO samplers or press Start without safeTIPCAP for other samplers (two Start buttons will be available only if "Use safeTIPCAP" check button is enabled in the Miscellaneous Setup. Check the selection of parameters depending on their status (green, yellow, red) and your needs (if the "Dynamic Parameters" check button has been activated for the given measuring mode). Deselect any parameters or the HbF correction for this mode by pressing the parameters or the "HbF correction" check button. 7. 8. Press Aspirate to start the measurement. When prompted by the analyzer, remove the sampling device and close the inlet flap. safePICO sampler: remove the sampler by holding the safeTIPCAP. Enter the information on the Patient Identification screen.

6.

9. Capillary samples

To analyze a blood or other fluid sample from a capillary do the following: Step 1. Action Check that the analyzer is in the Ready mode and scan the sample barcode. Accept or cancel the displayed patient data (selected in Sample Preregistration setup program see chapter 3 in this manual). If no sample pre-registration was selected, proceed to step 2 below. 2. Lift the capillary inlet flap. CAUTION - Risk of incorrect result Always remove air bubbles from a sample before introducing it into the analyzer. Air bubbles may result in erroneous pO2 values. 3. Remove the capillary end caps and insert the capillary into the inlet as shown.

4. 5. 6-9.

Select the desired mode by pressing the appropriate button. Press the Start button. See steps 6-9 for the syringe mode.

NOTICE: If the capillary breaks in the inlet, do not attempt to remove it, but replace the inlet gasket as described in chapter 7: Replacements.
Continued on next page

4-13

4. Sample measurements

ABL800 FLEX Operator's Manual

Introducing a blood sample, Continued


Test tube samples To analyze a blood or other fluid sample, do the following. Step 1. Action Check that the analyzer is in the Ready mode and scan the sample barcode. Accept or cancel the displayed patient data (selected in Sample Preregistration setup program see chapter 3 in this manual). If no sample pre-registration was selected, proceed to step 2 below. 2. Lift the syringe inlet flap. CAUTION - Risk of incorrect result Always remove air bubbles from a sample before introducing it into the analyzer. Air bubbles may result in erroneous pO2 values. 3. Hold the test tube at the syrin ge inlet. Select the desired mode by pressing the appropriate button. Press Start. The probe will move out of the syringe inlet into the test tube. 6-9. See steps 6-9 for the syringe mode.

4. 5.

CAUTION Risk of erroneous results Always meticulously follow the sampling procedures described in chapter 12: Sampling. Failure to follow these procedures may introduce clots or air bubbles in the sample and yield erroneous results. NOTICE: After measurement on blood from a patient who has received protamine sulphate it may be necessary to clean the analyzer using the Cleaning program to avoid clots.

4-14

ABL800 FLEX Operator's Manual

4. Sample measurements

Introducing a pleura sample


Procedure To analyze a pleura sample, do the following: Step 1. Action Open the syringe inlet flap and select the Pleura pH mode. NOTICE: The Pleura pH measuring program has to be selected in the Analysis Setup for syringe samples in order to obtain the button on the screen. 2. Place the syringe tip in the inlet. NOTICE: Radiometer recommends always to use heparinized samplers for pleural liquids to avoid introducing clots in the analyzer.

3. 4. 5.

Press Start. When prompted by the analyzer status, remove the syringe and close the inlet flap. Enter the patient information on the Patient Identification screen.

NOTICE: If pleura samples with identified bacterial infections are introduced into the analyzer, it is recommended to run a Decontamination.

4-15

4. Sample measurements

ABL800 FLEX Operator's Manual

Introducing an expired air sample


Procedure To analyze an expired air sample, do the following: Step 1. Action Open the syringe inlet flap and select the Expired Air mode.

NOTICE: The Expired Air measuring program has to be selected in the Analysis Setup for syringe samples in order to obtain the button on the screen. 2. Place the syringe tip in the inlet. NOTICE: The Expired Air mode is exclusively used for samples of expired air. Injection of blood or other liquid in this mode may cause damage to the analyzer. 3. 4. Press Start. Inject the sample slowly, following the volume bar on the screen. Make sure to continue injecting the expired air sample as long as the volume is higher than 0 mL. When prompted by the analyzer status, remove the syringe and close the inlet flap. Enter the patient information on the Patient Identification screen.

5. 6.

4-16

ABL800 FLEX Operator's Manual

4. Sample measurements

Patient identification
Entering information with barcode reader Unless the Batch mode is used, the Patient Identification screen is displayed while the sample is being aspirated. To enter information with a barcode reader (the "Enable General Barcode Support" must be activated in Miscellaneous Setup) in the item boxes, which have barcode support, do the following: Step 1. Action Scan the barcode with the barcode reader. Primarily, Operator, Accession No., Patient ID and Sampler ID are scanned. If the operator is scanned from the ID-card, and the operator is not included in the list of the analyzer's registered users, the barcode will be read directly from the ID-card. Entering information manually To enter patient information manually, do the following: Step 1. Action If desired, select another report layout by selecting the "Report layout" item/box on the screen and choosing the desired layout in the appearing list Highlight the desired text boxes and use the keyboard to type the desired data and press Enter. Sample demographics on the basis of an entered/scanned Sampler ID, i.e. all information about the sample (measuring mode, parameters, all input values and patient demographics). Data based on the Sampler ID is transferred from the RADIANCE system. Patient demographics on the basis of an entered/scanned Accession No., Patient ID or Department (Pat.)., i.e. all information about Patient ID, Patient Last Name, Patient First Name, Sex, Date of birth, Department (Pat.), Patient Note, Birth weight and Gestational Age. Data based on Accession No., Patient ID or Department (Pat.). can be transferred from all connections

2.

Mandatory text boxes (with a next to it) must be filled in order to view the measurement results.
Continued on next page

4-17

4. Sample measurements

ABL800 FLEX Operator's Manual

Patient identification, Continued


Entering information manually (continued) NOTICES: If the requested patient data (e.g. Patient Last Name) was received after the Patient Identification screen was exited, the patient report will be stored without the requested data in the Patient Report log. The requested patient data will be stored as a patient profile in the analyzer's database without, however, being attached to any patient report. If the requested patient data (e.g. Patient Last Name) was received after the Patient Identification screen was exited, the patient report will be transmitted without the requested data. To prevent this, select one of the patient ID items transferred from LIS/HIS as mandatory.

Patient Lookup

The Patient Lookup function allows you to transfer the patient information from a department's specific list to the Patient Identification screen. Department (Pat.) must be included in the report layout and filled in on the Patient Identification screen. To use the Patient Lookup function, do the following: Step 1. Action Highlight and fill in the Department (Pat.) in the Patient Identification screen. Press Patient Lookup. 2. 3. Select your patient from the list by highlighting the line on the screen. Press the Update button to update the Patient List. Press Select to transfer the specific patient information to the Patient Identification screen and return to the previous screen. Press Back to return to the previous screen without updating the patient information.
Continued on next page

4-18

ABL800 FLEX Operator's Manual

4. Sample measurements

Patient identification, Continued


Selecting a report layout

To select a report layout, do the following:


Step 1. 2. Action Highlight "Report Layout" with the up/down arrows on the Patient Identification screen. If "Report Layout" is included in the Patient ID, select a layout from the list displayed on the right side of the screen, using the up/down arrows. (The list of report layouts has been made in the Patient Report Setup - see chapter 3: Installation and setup in this manual.) If "Report Layout" is not included in the Patient ID, the text box is separated from the rest of the items on the Patient Identification screen. You can select another report layout by highlighting the line on the screen and selecting the desired layout. The patient report will be saved in this layout. NOTICE: If you are using mandatory input fields you are not allowed to change the report layout during measurement unless the mandatory fields have been filled. It is possible to see updatings of the pH, pCO2, pO2, cK+, cNa+, cCa2+, cCl, cGlu, cLac and cCrea electrodes (the number of electrodes depends on the type of your analyzer) during a measurement. To display the Electrode Updatings screen, press the Electrode Upd. button on the Patient Identification screen. Press Back to return to the on the Patient Identification screen.
Continued on next page

Electrode updatings

4-19

4. Sample measurements

ABL800 FLEX Operator's Manual

Patient identification, Continued


To change the parameter profile after a given measurement, do the following: Selecting parameters after a measurement Step Action 1. 2. 3. 4. Press Parameters on the Patient Identification screen. To exclude a parameter, deactivate the relevant button. To include a parameter, activate the relevant button. Press Back to return to the Patient Identification screen and press Result to display the changed result. Press Print to print out the changed patient report.

NOTICE: The Selected Parameters screen displays all the parameters selected in the Parameter Profile for a given measuring mode, depending on the analyzer version.

4-20

ABL800 FLEX Operator's Manual

4. Sample measurements

Patient result
Retrieving patient result The FLEXQ module is installed: The measurement result will not be displayed automatically. You can do one of the following: The FLEXQ module is not installed: The Patient Result screen will be displayed automatically when the measurement has been completed. However, if the patient identification information took longer time to enter than the measurement, you can do one of the following: Press the Result button on the Ready screen Press Menu Latest Result Press Menu My Results Press Menu Data Logs Patient Results Log, highlight the desired result and press Result.

Press the Ready screen

button on the

Scan the registration receipt Press Menu Latest Result Press Menu My Results Press Menu Data Logs Patient Results Log, highlight the desired result and press Result.

Press Print to print out the patient report. Parameter status The absence of any markings next to a parameter indicates that the parameter was measured without any fault. The following markings may appear next to a parameter: Marking ? Explanation Error in the last quality control measurement or calibration, or due to sample problems Parameter has violated the sample age. Parameter value is outside the reference range (the range within which the parameter value is considered normal, for the specific type of sample being measured), but inside the critical limits.
Continued on next page

4-21

4. Sample measurements

ABL800 FLEX Operator's Manual

Patient result, Continued


Parameter status (continued) Marking Explanation Parameter value is outside the upper or lower critical limit (the limits outside of which a parameter value is dangerously high or low, for the specific type of sample being measured), but inside the reportable range. Parameter value is outside the reportable range (userdefined and equal to or narrower than the measuring range). These markings will be printed together with the result and a verbal explanation. ".." instead of the value next to the value A parameter cannot be calculated or exceeds the numerical limit of the analyzer. Values with user-defined correction factors.

Critical limit

Critical limit

(no marking) Low parameter value High parameter value

Reference Range

Reportable Range

Patient result approval

Result approval is possible if the "Enable Result Approval" check button has been activated () in the Miscellaneous Setup program. To use this function, do the following: Step 1. 2. Action Press Approval on the Patient Result screen. The following buttons become available when you press the button: Approve Press Accept to accept the result and to send it to the connected RADIANCE system or LIS/HIS. Press the keyboard icon to write a note. Press Cancel to return to the previous screen.
Continued on next page

4-22

ABL800 FLEX Operator's Manual

4. Sample measurements

Patient result, Continued


Patient result approval (continued) Step Action Reject Press Accept to reject the result and to send it to the connected RADIANCE system or LIS/HIS. Press the keyboard icon to write a note. Press Cancel to return to the previous screen. Rerun Press Accept to mark that the measurement should be repeated, and to send it to the connected RADIANCE system or LIS/HIS. Press the keyboard icon to write a note. Press Cancel to return to the previous screen. Note Press the keyboard icon to type a note prior to the approval or to edit a highlighted note. To delete a highlighted note, press the keyboard icon and delete the note.

Once the result has been approved/rejected/rerun, the ID data can only be viewed, not changed (grayed-out input boxes). The Parameters, Patient Lookup and Request buttons will not be available either. In the Patient Report log you can filter patient reports according to the approval status. The Approval button changes to Send (if an external connection is used) after the approval of the patient result has been made. If result approval is used, data will not be transmitted to a LIS/HIS connection before the approval has been made. Acid-Base Chart Press Acid-Base Chart to view the acid-base chart for the selected patient result (the button will not be available if pH and pCO2 were deselected in the Parameter setup. The sample type should be either Arterial or Capillary).

The chart illustrates the patient result according to the Siggard-Andersen nomogram. The clinical conditions corresponding to each of the 8 zones listed to the right of the screen.
Continued on next page

4-23

4. Sample measurements

ABL800 FLEX Operator's Manual

Patient result, Continued


Acid-Base Chart The point position is determined by the pH and pCO2 values from the patient result. If the pH and pCO2 values are outside the defined limits of the acid-base (continued) chart, the message Measurement out of range will appear on the chart in red and the Print button will disappear. Available buttons: Print = to print out the displayed acid-base chart and the corresponding patient result, if available - see above. Back = to return to the Patient Result screen. NOTICE: The information provided by the acid-base chart should only be used as a guideline for interpreting the patients condition. The patient results must always be examined carefully by a clinician before a proper diagnosis is made. Recalling patient ID To recall the Patient Identification screen, press the ID button. For detailed information, refer to the topic Patient Identification in this chapter. Any changes made to an existing result will be registered in the Audit Trail. The Messages button is substituted with the Log button to indicate that changes were made. To see the audit trail, do the following: Step 1. 2. Action Press Log on the Patient Result screen to display the following buttons: Audit Trail and Messages. Press Audit Trail. The audit trail contains information about the operator who made the change, time of change and new/old values. The changes made at the same time are indicated with in Time and User columns. 3. Use the up/down arrow buttons to scroll the list of changes. Press Back to return to the Patient Result screen.

Patient result audit trail

4-24

ABL800 FLEX Operator's Manual

4. Sample measurements

Calculation of FShunt and ctO2(a - V )


Calculation of FShunt and
ctO2(a-v ) ), it is necessary to To obtain the calculated values of FShunt and ctO2(a-v analyze a mixed-venous and an arterial (or capillary) blood sample from the patient.

Do the following: Step 1. Action Make, if required, a new patient report layout (see Patient Reports in chapter 3) with the following parameters included:
FShunt and/or ctO2(a-v ) into the Patient Result ), sO2(v ), FO2(I), RQ and T (patient temperature) into the pO2(v

Patient ID 2. 3. 4.
) and sO2(v ). Analyze the mixed-venous sample and record the pO2(v

Analyze the arterial sample. On the Patient Identification screen, select the sample type as Arterial or Capillary and key in the following values:
pO (v 2 ) and sO2(v) from the mixed-venous sample (step 2 above)

FO2(I) if it differs from the default value of 0.21 (for Fshunt) RQ if it differs from the default value of 0.86 (for Fshunt) T if it differs from the default value of 37 C (for Fshunt). p50(st) from the mixed-venous sample if sO2(a) > 97 %. NOTICE: The FShunt value will be available as an estimated value even without the data obtained from the mixed-venous sample.

4-25

4. Sample measurements

ABL800 FLEX Operator's Manual

Patient result messages


Screen messages The Patient Result Messages screen is displayed after the Messages button has been pressed. The screen gives the erroneous parameter(s) and the message(s) with number(s) referring to the message described in Troubleshooting Messages, chapter 11. Messages can be seen on the following three information levels: Level User Manager Service Explanation Messages for the user familiar with the basic daily operation of the analyzer and primarily responsible for performing measurements. Messages for the user with deeper knowledge of the analyzer functions and responsible for the analyzer's proper operation. Messages for the service technician with thorough knowledge of the operation and construction of the analyzer.

Buttons and functions

The following buttons are available: Button Result Function Press to return to the Patient Result screen.

Troubleshoot Press to display the error description and operator actions see Troubleshooting Messages, chapter 11 for detailed information. Print Back Note Press to print the screen message(s). Press to return to the previous screen. Press to display the keyboard, type the note and confirm with Enter. If Notes for patient result were entered in the User-defined Notes program (see chapter 3: Installation and Setup), select a Note from the list with the up/down arrow buttons. To edit a note use Edit Note and type the note on the keyboard. Remember to confirm it with Enter on the keyboard. To delete a note use Delete Note and delete it.

4-26

5. Overview
Introduction

Quality control

Quality control is used to evaluate the performance of the analyzer to ensure the reliability, accuracy and precision of patient sample results. This chapter describes how to perform a quality control measurement and the subsequent presentation of results.

Contents

This chapter contains the following topics. General information ......................................................................................... Preparing a control solution ............................................................................. Manual quality control measurement............................................................... AutoCheck measurement ................................................................................. Quality control identification ........................................................................... 5-2 5-4 5-6 5-7 5-8

Quality control result........................................................................................ 5-10 Quality control result messages ....................................................................... 5-15

5. Quality control

ABL800 FLEX Operator's Manual

General information
Purpose The purpose of quality control is to validate the analyzers performance by evaluating accuracy and precision. Performing quality control routinely will help to ensure that results from actual patient samples are accurate. Each quality control system includes four types of control solutions in order to provide control of analyzer performance when measuring parameters at low, normal and high levels refer to the Quality control setup in chapter 3. It is possible to use non-Radiometer solutions although the accuracy and validity of the results cannot be guaranteed. Inserts Control solutions are packed in ampoule boxes. Each ampoule box contains an insert with the Radiometer-determined assigned values and control limits valid for that particular lot of control solutions. The solution type, lot number, assigned value and expiration date is barcoded for easy entry. Quality control solutions are solutions with predetermined values that cover the clinically relevant ranges for the measured parameters, the objective being to simulate a patient sample. Each quality control system from Radiometer includes four levels of quality control solutions in order to cover the whole clinically significant range: low, normal and high. The number of levels should cover the analytical measuring range. Special country, state and local regulations must be complied with. Additional quality control should be run after any troubleshooting or preventive maintenance which might alter performance and whenever the technician has questions about the performance of the analyzer. An example of daily routine with the 4-level quality control system is given below: Shifts per day 3 Quality control routine One level tested at the beginning of each shift. The 4th level is tested in the 8-hour shift with the highest volume of patient samples. Two levels at the beginning of each shift. All levels are tested at the beginning of each shift.
Continued on next page

Solutions

Quality control frequency

2 1

5-2

ABL800 FLEX Operator's Manual

5. Quality control

General information, Continued


Glossary of terms The following terms are used in the quality control: Term Accepted result Control range Explanation A measurement value which falls within the statistics range. The range within which a measurement should fall. Typically the control range is set to be the mean 2 SD (see Quality control setup in chapter 3). This range can be set using the lot-to-date range (2 SD) calculated by the analyzer, or it may be user-defined. Upper and lower control limits, established by Radiometer, printed in the insert provided with each box of control solutions. A range calculated by the analyzer from measurements taken on a lot of a particular control solution. It is represented by mean 2 SD. The range for each parameter the analyzer is capable of measuring. Refer to Measured parameters in chapter 13 for the analyzer measuring range for each parameter. The factor by which the control range is expanded (multiplied by) to determine the statistics range. The recommended statistics factor is 1.5. The range within which a measurement must fall in order to be included in the quality control statistics. It is determined by multiplying the control range limits by the entered statistics factor. It is typical that the statistics range is set to the mean 3 SD.

Insert limits

Lot-to-date range

Measuring range

Statistics factor

Statistics range

For detailed information on terminology and principles for the quality control systems, please refer to the Quality Control Systems Reference Manual. NOTICE: In cases where the first blood sample measurement after a measurement on a QUALICHEK3+/5+ or AutoCheck3+/5+/6+ control solution is to be performed with HbF correction (either set up for all levels or for FHbF> 20 %), a rinse must be called prior to the blood measurement to ensure accurate oximetry results.

5-3

5. Quality control

ABL800 FLEX Operator's Manual

Preparing a control solution


About control solutions Brief information is given below: Item Quantity Conditioning before measurement Open ampoule stability Description Each QC ampoule of QUALICHECK+ contains 2 mL of QC solution. Each QC ampoule of AutoCheck contains 0.7 mL of QC solution.

Store the ampoules for five hours at 18-32 C. Control solutions are sensitive to light, so always keep the ampoule box closed. In order to ensure the reliability of the measurement, each QC ampoule must be used immediately after opening, for one measurement on one analyzer only.

Required materials

The following materials are required to make a manual measurement on a control solution: Quality control system QUALICHECK3+ QUALICHECK5+ Materials required Four levels of control solutions H700 Ampoule adapter Ampoule opener Rubber gloves AutoCheck3+ AutoCheck5+ AutoCheck6+ Four levels of control solutions H700 Ampoule adapter H705 Ampoule holder Rubber gloves

Before measure- Do the following: ment on QUALICHECK+ Step Action control solution 1.

Holding the ampoule between two fingers as shown, shake it vigorously for at least 15 seconds.

2.

Tap the top of the ampoule until all of the solution collects at the bottom.
Continued on next page

5-4

ABL800 FLEX Operator's Manual

5. Quality control

Preparing a control solution, Continued


Before measurement on
QUALICHECK+

Step 3.

Action
2

control solution (continued)


1

Place the ampoule in the ampoule opener and break off the ampoule neck.

4.

Place the ampoule fully into the H700 adapter.

Before manual Do the following: measurement on AutoCheck Step Action control solution Place the ampoule in the H705 Ampoule holder. 1. 2. 3. Holding the ampoule in its holder between two fingers, shake it vigorously for at least 15 seconds. Place the H700 Ampoule adapter over the ampoule and press it down to open the ampoule.

5-5

5. Quality control

ABL800 FLEX Operator's Manual

Manual quality control measurement


Procedure To make a manual measurement on a QUALICHECK+ or AutoCheck control solution, do the following: Step 1. 2. 3. Action Open the syringe inlet flap. Place the adapter tip up into the syringe inlet. Press Ampoule-QC to select the measuring program.

4.

Press Start. The sample is aspirated and the Quality Control Identification screen appears. Refer to the next section, for information on entering data.

5. 6.

When prompted by the analyzer, remove the adapter and close the syringe inlet flap. Dispose of the used ampoules as infectious waste*.

* Reference: Clinical laboratory waste management. CLSI document GP5-A2.

High Crea Check

S8377 Cleaning Met II Solution is used to check the analyzer performance at a high creatinine level. The check up is made automatically during the system alignment procedure. To call an unscheduled High Crea Check, do one of the following: Press Menu > Start Programs > Run High Crea Check On the Analyzer Status screen highlight the line for Slot 11 and press Run High Crea Check to start a measurement.

5-6

ABL800 FLEX Operator's Manual

5. Quality control

AutoCheck measurement
Scheduled AutoCheck measurement A scheduled AutoCheck measurement will start on time provided that no calibrations are pending. A pending calibration will be performed before the scheduled AutoCheck measurement. On the screen that appears when an AutoCheck measurement is about to run (scheduled either from QC Schedule or from the RADIANCE system), select one of the three possibilities: No action taken: the AutoCheck measurement starts after 20 seconds (see counter on the screen) Proceed pressed: the AutoCheck measurement starts immediately. Postpone pressed: the AutoCheck measurement will be delayed by five minutes.

Unscheduled AutoCheck measurement

To start an unscheduled AutoCheck measurement, do the following: Step 1. 2. 3. Action Press Analyzer Status Quality Control. Highlight the solution to run an unscheduled measurement on. Press Run AC Ampoule to start the measurement.

The Quality Control Result is shown after the measurement is completed. NOTICE: It is only possible to choose ampoules that are packed in the carousel. If an ampoule is not available, the Run AC Ampoule button is inactive.

5-7

5. Quality control

ABL800 FLEX Operator's Manual

Quality control identification


Entering information with barcode reader To enter the information with the barcode reader, do the following: Step 1. Action Scan "Operator" from your ID-card.

If the name is included in the list of the analyzer's registered users, the barcode will identify the person and fill in the name automatically.

Entering information manually

To enter information manually, do the following: Step 1. Action Highlight the item to be edited, using the arrow buttons. Solution Are automatically detected and entered if only one quality and Lot control system is used. Otherwise highlight the desired solution from the screen and confirm with Enter. If the same type of QUALICHECK+/AutoCheck+ solution is used, then only the solution type will be identified. Choose the specific slot from the screen. 2. For manual QC measurements only: Enter the temperature (default: 25 C) on the keypad and confirm with Enter. Mandatory entry is indicated by a .
Continued on next page

5-8

ABL800 FLEX Operator's Manual

5. Quality control

Quality control identification, Continued


Entering information manually (continued) Step 3. 4. Action Enter "Department" on the keypad and confirm with Enter. Enter "Operator" on the keypad or keyboard and confirm with Enter (filled in automatically when operator is logged on).

"Scheduled: Yes" indicates that a QC is performed according to a planned QC in the Quality Control Schedule. "Scheduled: No" indicates that a QC is performed on behalf of the user. To display the electrode updatings during measurement, press the Electrode Upd. button.

Electrode Updatings

5-9

5. Quality control

ABL800 FLEX Operator's Manual

Quality control result


Quality Control In most cases the Quality Control Result screen will be displayed automatically when the measurement has been completed. However, if the quality control Result screen identification took longer time to enter than the measurement, you can: Press Result from on the Quality Control Identification screen; from the Menu > Analyzer status > Quality Control; from the Quality Control log.

The measured values are compared with the defined control range, measuring range and statistics range, then given a status mark accordingly. NOTICE: The user-defined corrections (slope and offset) do not influence the quality control results, unless the "Apply parameter corrections to QC" function has been activated in the Miscellaneous Setup. Parameter status The absence of any markings next to a parameter indicates that the parameter was measured without any fault. The following markings may appear next to a parameter: Marking ? W R Explanation Error in the previous calibration, or analyzer malfunction. A violated Westgard Rule. A violated Rilibk Rule. Parameter value outside the control range, but inside the statistics range. Only the values within the statistics range are considered accepted and are included in the quality control statistics. Parameter value outside the statistics range and not included into statistics Parameter value outside the measuring range. Measurement is not included into statistics.
Continued on next page

5-10

ABL800 FLEX Operator's Manual

5. Quality control

Quality control result, Continued


Parameter status (continued) Marking * .. Explanation Parameter values with user-defined corrections see Parameters and input, chapter 3 for details. Parameter value could not be calculated, most likely due to a system error or malfunction. These values for the most part will be accompanied with a ?. To obtain a possible explanation, press the Message button.

Unknown solutions

Quality control on a solution identified as unknown is not compared with any previous measurements or statistics, and therefore does not receive a status marking. The parameter results for unknown levels cannot be plotted and are not included in any statistical data unless later changed to a defined slot. For control solutions from Radiometer temperature corrections are made automatically by using the measured or typed-in temperature. For non-Radiometer control solutions, temperature corrections must be made manually. Refer to manufacturers literature for procedure.

Temperature corrections

Reference

For a detailed explanation of the evaluation of results, refer to chapter 6 of the Quality Control Systems Reference Manual. To change data that can be edited (non-grey lettered) on the screen, press QC ID. For detailed information, refer to previous section Quality Control Identification in this chapter. NOTICES (for manual QC measurements only): Changing the temperature will initiate recalculation of the last result. It will be substituted by the recalculated result, and the statistics will also be recalculated. In case the temperature is mandatory, the result cannot be viewed until it has been entered. Exiting the QC ID screen without entering a temperature, the previous temperature will be recalled.
Continued on next page

Recalling quality control identification

5-11

5. Quality control

ABL800 FLEX Operator's Manual

Quality control result, Continued


High Crea Check (for the ABL8x7 FLEX analyzers only) The high Crea check result of the ABL8x7 FLEX analyzers can be retrieved as follows: Press Result on the Quality Control Identification screen from the Menu > Analyzer status > Quality Control from the Quality Control log

The initial cCrea concentration in the Cleaning Met II Solution is approximately 1500 mol/L when the container is installed in the analyzer. The exact concentration is entered from the barcode when the solution is installed in the analyzer (the value shown as cCrea on the High cCrea Check screen above) . With time, creatinine in the solution transforms into creatine, the speed of this transformation depends on the ambient temperature. cCrea
1500 mol/L

Creadif

Time

Expected cCrea concentration in the Cleaning Met II Solution The upper/lower drift limits for cCrea is 225 mol/L, i.e. if the initial cCrea is 1500 mol/L, the corresponding control ranges will bwe 1275-1725 mol/L. Expected cCrea concentration in the Cleaning Met II Solution at the time of the High Crea check has been performed Measured cCrea concentration in the Cleaning Met II Solution at the time of the High Crea check has been performed
Continued on next page

5-12

ABL800 FLEX Operator's Manual

5. Quality control

Quality control result, Continued


High Crea Check (for the ABL8x7 FLEX analyzers only) (continued) Creadiff. is the difference between the expected creatinine value in the Cleaning Met II Solution and the measured value. It is shown on the plot see page 5-14. To evaluate the Creadiff. accurately, it is required to enter the expected ambient temperature for the 14-day period in the Environment Setup see Analyzer settings in chapter 3.

The ambient temperature on the High Crea Check ID screen is grayed out, i.e. it can only be viewed. The ambient temperature setting can be changed in the Environment Setup.

Quality control plot

The Quality Control Plot screen appears once the Plot button is pressed.

Parameter

The following diagram shows the details of a typical plot.


Slot number Solution type Control range upper limit Date Time Measurement Measured value Lot change Control range lower limit

Out-of-range symbol

Continued on next page

5-13

5. Quality control

ABL800 FLEX Operator's Manual

Quality control result, Continued


Quality control plot (continued) Plot element Shaded block Function Date, time and measured value for the highlighted measurement. Use the arrow buttons to scroll the plot and view other measurements. Show the upper and the lower limits of the control range for the highlighted measurement. Show the number of measurements for the selected parameter. To view a measurement result, highlight a dot on the screen and press Result. Measurement outside the control range and inside the statistical range. Measurement outside both the control and statistical ranges. Use the following buttons to do the following: Next Param. Prev. Param. Next Slot Result Print Back Display the quality control plots for the requested parameter. Display plots of the next control solution slot. View quality control result for the highlighted measurement of the highlighted plot. Print out the plot. Return to the previous screen.

Control range limits Dots

Out-of-range symbols

(For the ABL8x7 FLEX analyzers only): The plot for the High Crea Check of the ABL8x7 FLEX analyzers presents Creadiff.

The screen is similar to the quality control solution plot, but the Next Param. and Prev Param. buttons are absent.

5-14

ABL800 FLEX Operator's Manual

5. Quality control

Quality control result messages


Quality control result messages To see the quality control result and check the messages, do the following: Step 1. 2. 3. Action Press Menu - Analyzer Status - Quality Control. Highlight the desired quality control measurement and press Result. Press Messages to see the messages in connection with the given quality control result. The erring parameter(s) and the message(s) with number(s) referring to the message are described in Troubleshooting Messages, chapter 11. 4. Troubleshoot, if required, by highlighting the message (the first message is already highlighted) and pressing Troubleshoot. Follow instructions on the screen to remedy the error. Use Tutorials for assistance. When it is done, press Troubleshooting completed. 5. Press Back to return to the list of messages screen. Press Print to print out the list of messages. Press Note to display the keyboard, type the note and confirm with enter. If Notes for quality control result were entered in User-defined Notes program (see chapter 3: Installation and Setup), select a Note from the list with the up/down arrow buttons. To edit a note, use Edit Note. To delete a note, use Delete Note. 6. Press Back to close the Note popup window. Press Back to return to the Quality Control Result screen.

AutoCheck status

See Refilling the AutoCheck carousel in chapter 7: Replacements.

5-15

5. Quality control

ABL800 FLEX Operator's Manual

5-16

ABL800 FLEX Operator's Manual

6. Calibration

6. Overview
Introduction

Calibration

This chapter describes the available calibration programs and the calibration results. A short glossary of terms used in the chapter is also given.

Contents

This chapter contains the following topics. General information ......................................................................................... Unscheduled calibrations ................................................................................. Interrupted, pending or expired calibrations .................................................... tHb calibration.................................................................................................. Calibration result.............................................................................................. 6-2 6-4 6-5 6-6 6-8

Calibration result messages.............................................................................. 6-10

6-1

6. Calibration

ABL800 FLEX Operator's Manual

General information
Purpose The calibration process determines and checks the accuracy with which the analyzer measures its parameters. The process is therefore important in ensuring the reliability of results. Calibrations are performed using solutions and gases of known concentration, for each of the measured parameters. Calibration programs The following calibrations performed on the analyzer are briefly described below. (For detailed explanation please refer to the Reference Manual, chapters 1,2,3.) Calibration 1 point Description Measures each parameter on one solution and/or gas of known composition, giving one value per parameter. Relates the measured values to the theoretical values of a solution and/or gas of the same composition. Gives the Drift 1 and status values of the electrode. 2 point Measures each parameter on two different solutions and/or gases both of known composition, giving two values per parameter. Relates the measured values to the theoretical values of solutions and/or gases of the same composition. Gives the Drift 1, Drift 2 values and the status or zero point, sensitivity of the electrodes (two sensitivity values are given for the Crea B electrode and one for the Crea A electrode of the ABL8x7 FLEX analyzers). 1 point pH/BG Measures pCO2 and pO2 on one gas mixture of known (the USA only) composition and pH on a known solution, giving one value per parameter Relates the measured values to the theoretical values of a mixture/solution of the same composition. Gives the Drift 1 of the electrodes. tHb Calibrates the spectrophotometer. See tHb calibration in this chapter.

Calibrations performed automatically at specific intervals selected in the Calibration Schedule setup (see Calibration Setup in chapter 3) are called scheduled calibrations. All calibration results are automatically recorded in the Calibration Log from which they can be viewed and printed. Refer to chapter 9: Data Management for details. The calibration process and a detailed explanation of the terms are described in the Reference Manual.
Continued on next page

6-2

ABL800 FLEX Operator's Manual

6. Calibration

General information, Continued


NOTICE: Additional 1-point calibrations are performed on the analyzers with the metabolite electrodes: after 11 measurements made one after another, no more than 10 minutes apart and 30 minutes after the last measurement was made. The additional 1-point calibration does not change the intervals selected for the scheduled calibrations. In cases when a scheduled 2-point calibration is imminent, the additional 1-point calibration will be substituted with the 2-point calibration. NOTICE: (for the USA only) If an unacceptable pH drift is detected during a 1 point pH/BG calibration, the ongoing calibration is aborted and a normal 1-point calibration is performed.

6-3

6. Calibration

ABL800 FLEX Operator's Manual

Unscheduled calibrations
Calling an unscheduled calibration Calibrations called by the operator at any time are called unscheduled calibrations. To call an unscheduled calibration, do the following: Step 1. 2. Action Check that the analyzer is in the Ready mode. Press Menu Start Programs Calibration Programs.

Call a calibration. If a button is grayed out, a calibration cannot be called (as 1-point calibration on the screen). 3. To see the electrode updatings during a calibration, go to the Analyzer Status screen and press the Electrode Upd. button.

The 24-hour calibration period is only necessary after the replacement of Calibrations during 24 hours electrodes and/or electrode membranes. after restart The calibration intervals for the first 24 hours after restart are as follows: Time after restart First 4 hours Calibration intervals 1-point calibration 2-point calibration Remaining 20 hours 1-point calibration 2-point calibration Every 30 minutes Every hour Every hour Every 4 hours

NOTICE: If the restart is caused by cGlu/cLac electrode/membrane maintenance 1-point calibrations will be performed every 30 minutes for the first 4 hours and every hour for the following 20 hours. 2-point calibrations will follow the normal schedule. NOTICE: If the calibration schedule is set up to perform calibrations at intervals less frequent than those programmed for the 24-hour calibration period, the calibration schedule is overridden.

6-4

ABL800 FLEX Operator's Manual

6. Calibration

Interrupted, pending or expired calibrations


Interrupted calibrations A calibration is interrupted when the operator presses the Stop button during the calibration process. The analyzer performs a rinse before returning to the Ready mode. If the interrupted calibration was a scheduled calibration, it will be automatically performed after the analyzer has been in the Ready mode for three minutes without any activity, i.e. measurements. The scheduled calibration becomes pending if interrupted. Pending calibrations A calibration is pending when a scheduled calibration cannot be performed at its scheduled time because the analyzer is occupied with another function, e.g. measurement. A pending calibration is indicated by a clock symbol next to it on the Analyzer Status Calibrations screen as a warning. A pending calibration will be performed three minutes after the analyzer has returned to the Ready mode without any activity. If more than one type of calibration is pending, the highest-priority calibration is performed. The calibrations are given in descending order of priority, the 2-point calibration having the highest priority. If a calibration is pending, a scheduled AutoCheck measurement will be postponed until after the calibration has been performed. Expired calibrations A full calibration means that all parameters are calibrated. Normally, a full calibration is a 2-point calibration, but in the case of glucose and lactate electrodes it is a 1-point calibration. A calibration is expired and considered to be invalid when: an electrode or electrode membrane is replaced. the time interval since the last accepted full calibration exceeds the time interval between two full calibrations plus a 1-point calibration. Under these conditions a Cal expired (parameter) message appears in the Activity and System Messages logs, and the replaced item is also recorded in the Activity log. An expired calibration can be remedied by performing a successful full calibration after a replacement. Upgraded calibrations Whenever a 2-point calibration contains an error, the next calibration will be upgraded to a 2-point calibration. If a 2-point calibration contains errors on Glu, Lac and oximetry parameters only, the errors can be remedied by a 1-point calibration. Thus the next calibration will not be updated to a 2-point calibration. A 1-point pH/BG calibration (the USA only) will, however, be updated to a 1-point calibration.

6-5

6. Calibration

ABL800 FLEX Operator's Manual

tHb calibration
Purpose The calibration is used to calibrate the analyzers spectrophotometer and is performed once every three months on S7770 tHb Calibrating Solution. The calibration can be included in the Calibration Schedule. It is recommended to perform the Protein Removal Program before a tHb calibration in order to remove protein deposits see Decontamination and Protein Removal program, chapter 7. Check that the analyzer is in the Ready mode. Prepare an ampoule of S7770 Calibrating Solution, an H700 Ampoule adapter and an ampoule opener. Performing a tHb calibration To perform a tHb calibration, do the following. Step 1. 2. Action Perform Cal 1 or Cal 2, if not done after the Protein Removal. Press the Menu - Start Programs - Calibration Programs - tHb Calibration. Enter the barcode from the tHb Calibration Solution insert, using the barcode reader or the keyboard. To cancel the program, press Close. 3. When the barcode has been accepted, a screen with requirements appears. Requirements 1-3 should be fulfilled before you press Start. 4. Tap the top of the tHb Calibration Solution ampoule to collect the liquid at the bottom and break off the ampoule neck, using the ampoule opener. Put the ampoule in the H700 Ampoule adapter. Open the syringe inlet flap and place the adapter tip into the inlet. Press Start to aspirate the calibrating solution. When prompted by the analyzer, remove the adapter and close the syringe inlet flap. Do not discard the S7770 ampoule, as it will be used for the tHb Calibration verification. After the measurement a rinse is performed and then the analyzer returns to the Ready mode. 9. If the calibration results are accepted, perform the tHb calibration check immediately see below. Otherwise remedy the error and perform a new tHb calibration.
Continued on next page

Preparation

5. 6. 7. 8.

6-6

ABL800 FLEX Operator's Manual

6. Calibration

tHb calibration, Continued


tHb calibration check To verify the validity of the tHb calibration, do the following. Step 1. 2. 3. 4a. 4b. Action Open the syringe inlet flap and place the adapter with the S7770 solution in the inlet. Select the mode: Syringe S195 L or Syringe S85 L or Syringe S250 L depending on the ABL800 FLEX version. Press the Start button. Remove the ampoule when prompted and close the inlet. If the calibration check value* is within the insert limits, the tHb calibration is accepted. If the calibration check value* is outside the insert limits, repeat the tHb calibration and then the calibration check immediately after it. If the second tHb calibration and the calibration check fail again, contact a service technician. * Can be ctHb, or ctHb and ctBil, or ctBil. NOTICE: If the tHb calibration check has not been performed immediately after the tHb calibration, repeat the calibration and the calibration check immediately after it.

6-7

6. Calibration

ABL800 FLEX Operator's Manual

Calibration result
Viewing a calibration result The result can be viewed as soon as the calibration is completed. Results from the most recent calibrations can be viewed via the Analyzer Status Calibration screen or via the Calibration Log. To recall the result, highlight the desired calibration and press the Result button to display the Calibration Result screen.

Press button: To display the remainder of the calibration result. Messages Print (Send) To display interpretation of any detected errors. To print out the result. To send the result to a connected LIS/HIS system.

"Scheduled: Yes" indicates that a QC is performed according to a planned QC in the Quality Control Schedule. "Scheduled: No" indicates that a QC is performed on behalf of the user. 1- or 2-point calibration results The calibration data is grouped together by parameter: Bold characters Indicate data updated during this calibration.

Grey characters Indicate data taken from the previous calibration and not included in this calibration. "? " Indicates an error or the value outside the recommended range, such as: Drift value Status Sensitivity = outside the drift tolerances see Calibration Setup in chapter 3. = = outside the default limits. outside the default limits. (Two sensitivity values are given for the Crea B electrode and one for the Crea A electrode of the ABL8x7 FLEX analyzers.)

".."

Indicates the value could not be calculated, most likely due to a system error or malfunction. These values for the most part will be accompanied by a ?. Press the Messages button for an explanation of the error.
Continued on next page

6-8

ABL800 FLEX Operator's Manual

6. Calibration

Calibration result, Continued


tHb calibration result An example of a successfully performed tHb calibration is shown below:

FCUV is a factor which expresses the ratio of the effective light path of the analyzer cuvette to that of a reference cuvette determined by Radiometer. The value should be between 0.80-1.20 and have no errors, i.e. no ? in it.

6-9

6. Calibration

ABL800 FLEX Operator's Manual

Calibration result messages


Access to the screen Press the Messages button on the Calibration Result screen to display the Calibration Result Messages screen. The following buttons are available: Result Troubleshoot Note To return to the Calibration Result screen. To display the interpretation and operator actions see Troubleshooting Messages, chapter 11. A Note can be selected from a list of Notes (made in the User-defined Notes program), typed by pressing the keyboard icon, edited by pressing the Edit Note button, or deleted by pressing the Delete Note button. To print out the messages. To return to the previous screen.

Print Back

Message levels

The message levels can be considered as a type of filter in which the messages displayed are based on the following: Level User Manager Service Explanation Messages for the user familiar with the basic daily operation of the analyzer and primarily responsible for performing measurements. Messages for the user with deeper understanding the analyzer functions and responsible for the analyzer's proper operation. Messages for the service technician with thorough knowledge of the operation and construction of the analyzer.

6-10

7. Replacements Overview
Introduction This chapter describes the replacement procedures that ensure proper functioning of your analyzer. Remember to perform Decontamination before replacing the parts that were in direct contact with blood. Contents This chapter contains the following topics. General information ......................................................................................... Replacing membranes or electrodes................................................................. Replacing pump tubes ...................................................................................... 7-2 7-6 7-9

Replacing inlet gasket and inlet probe ............................................................. 7-12 Replacing waste container, fan filter, printer paper ......................................... 7-14 Replacing solutions and gases.......................................................................... 7-16 Refilling the AutoCheck carousel .................................................................... 7-19 Automatic auxiliary programs.......................................................................... 7-20 Decontamination and Protein Removal programs ........................................... 7-21 Cleaning the analyzer....................................................................................... 7-23 List of references.............................................................................................. 7-25

7. Replacements

ABL800 FLEX Operator's Manual

General information
Viewing replacement requirements To view the replacements that are due, do the following: Step 1. Action Check the following on the Analyzer Status screen:

Replacement schedule for the electrode membranes and the electrodes Messages referring to the membranes/electrodes or user activities Traffic light color (GREEN = No replacements are due at the present time; YELLOW = A replacement is due) 2. 3. 4. Highlight the item to enter the screen for replacements of electrodes/membranes see the next page. Press Log Activity to enter the User Activity program to view user activities list see chapter 3 in this manual. Press Close to exit to the Ready screen.

Entering the Hold mode

The Hold mode ensures that all wet section activities are suspended and that the proper restart sequence will follow the exit from the Hold mode. To enter the Hold mode, do one of the following: Go to Menu Analyzer Status Electrodes and Other. Press Replace. Remove the cover to the measuring modules. Open the window to the electrode module and wait for draining to complete. Remove both inlet flaps. Remove a solution container.
Continued on next page

7-2

ABL800 FLEX Operator's Manual

7. Replacements

General information, Continued


To record replacements in the Hold mode, do the following: Recording replacements in Hold mode Step Action 1. 2. 3. Enter the Hold mode as described earlier. Remove the component to be replaced according to the recommended procedure. Preferably scan the barcode of the new item before mounting it in the analyzer. Otherwise highlight the item in the list and press Replaced in order to have it appear in Recorded replacements box on the right of the screen. Repeat step 2 for another replacement in the similar manner.

4.

To remove an item from the list of recorded replacements, highlight it (with the up/down arrow buttons) and press Undo. 5. If desired, enter the operator initials and a note for the Activity Log by pressing the keyboard icon in the Operator: box (or the keyboard) in the Note: box. Select a Note from the list with the up/down arrow buttons. Edit, if desired, a highlighted Note by pressing Edit and displaying the keyboard. Remember to confirm a change with Enter. Delete, if desired, a highlighted Note by pressing Delete Note.

6. 7.

When all replacement actions are completed, mount all components and the covers, and close the inlets. Press Restart to begin the proper restart sequence. For the ABL8x7 FLEX analyzers only: In case the Crea A and Crea B electrode membranes or electrodes have been replaced, pressing Restart will display the following screen (see the next page):
Continued on next page

7-3

7. Replacements

ABL800 FLEX Operator's Manual

General information, Continued


Recording replacements in Hold mode (continued) Step 7. (continued) Action

(In case only the Crea electrodes were remembraned or replaced, the Confirm Membrane Replacement screen will be used.) Press Accept to start the restarting sequence or press Back to return to the previous screen. After the required action is completed, press Restart and then Accept. Entering Fluid Replacements mode The Fluid Replacements mode ensures that the proper restart sequence will follow the exit from this mode. To enter the Fluid Replacements mode, do one of the following: Press Analyzer Status Reagents Replace. Remove a solution container To record replacements in the Fluid Replacements mode, do the following: Recording replacements in the Fluid Step Action Replacements Press Analyzer Status Reagents Replace. 1. mode

2.

Remove the component to be replaced according to recommended procedure.


Continued on next page

7-4

ABL800 FLEX Operator's Manual

7. Replacements

General information, Continued


Recording replacements in the Fluid Replacements mode (continued) Step 3. Action Scan the barcode on a new component, or press the keyboard icon in the "For each item" box, type in the barcode and confirm with Enter. The replaced items appear in the Recorded replacements box as soon as the code has been scanned or typed. 4. 5. 6. Install a new component according to recommended procedure. Repeat steps 1-4 for another replacement in a similar manner. To remove an item from the recorded replacements list, highlight it with the up/down arrow buttons in the box and press Undo.

If desired, enter the operator initials and a note for the Activity Log. Restart after replacements Press Restart and then Accept on the Confirm Replacement screen.

If the barcode for the new Rinse, Calibration or Cleaning solution has not been entered, the analyzer will not restart and the pop-up window appears with the message to enter the solution barcode. Press Back to return to the Recording Fluid Replacements screen and to scan the barcode. Press Restart and then Accept. The analyzer performs a restart sequence, depending on the replaced item. After electrode and electrode membrane replacements, calibrations are performed more frequently during the first 24 hours see Unscheduled Calibrations in chapter 6 of this manual.

7-5

7. Replacements

ABL800 FLEX Operator's Manual

Replacing membranes or electrodes


Removing an electrode from measuring chamber To remove an electrode from the measuring chamber, do the following: Step 1. 2. Action Enter the Hold mode and record the replacement action. Press the electrode connector latch, and lift the electrode connector open and lift the electrode out of the measuring chamber.

Remembraning an electrode

To remembrane an electrode, do the following: Step 1. Action All electrodes: Remove the electrode jacket by pressing the tabs on the sides and pulling. Further for the reference electrode: Remove the electrode jacket by pulling it off. If the Oring remains on the electrode, remove it. If the electrode is stuck in its jacket due to salt crystals around the top of the jacket, soak it in water until the crystals dissolve. Further for the pO2 electrode: Brush the electrode tip with the supplied brush.

2.

All electrodes: Rinse the electrode with tap water and shake to remove excess drops of water. Do not dry the electrode. Further for cK+, cCl, cCa2+, cNa+ and reference electrodes: Remove any salt deposits, using tap water.

3.

All electrodes: Remove the protective foil of a sealed electrode jacket in the membrane box. Further for cGlu, cLac or cCrea A + cCrea B electrodes: Open a capsule of the electrolyte solution supplied and empty it into the electrode jacket. NOTICE: Always remembrane both cCrea A and cCrea B electrodes at the same time.
Continued on next page

7-6

ABL800 FLEX Operator's Manual

7. Replacements

Replacing membranes or electrodes, Continued


Remembraning an electrode (continued) Step 4. Action All electrodes: Press the electrode firmly into the jacket until a click is heard. Further for the reference electrode: Check that the old O-ring has been removed and press the electrode through the protective film covering the electrode jacket. CAUTION Safety precautions Solution is irritating to eyes, respiratory system and skin. In case of contact with eyes, rinse immediately with plenty of water and seek medical advice. 5. All electrodes: Remove the membraned electrode from the membrane box. Further for the reference electrode: Rinse the electrode with tap water and dry with a lint-free tissue. Further for pCO2 and pO2 electrodes: Remove any air bubbles trapped between the electrode tip and the membrane by pressing the tabs on the electrode jacket and moving the electrode up and down slightly in the electrolyte until the bubble disappears. 6. All electrodes: Dry the electrode contact.

7.

8.

All electrodes: Place the electrode into the measuring chamber and close the electrode connector, pushing on its top until it clicks shut. If no more replacement actions are to be made, mount all covers and parts, and press Restart.

Replacing an electrode

To replace an electrode, do the following: Step 1. 2. Action All electrodes: Remove the old electrode from the measuring chamber. For the new pH, pCO2 and pO2 electrodes only: 2a. Rinse the electrode under tap water. 2b. Shake it to remove excess drops of water. Do not dry the electrode. 4. All electrodes, except the pH electrode: Membrane the electrode at once.
Continued on next page

7-7

7. Replacements

ABL800 FLEX Operator's Manual

Replacing membranes or electrodes, Continued


Replacing an electrode (continued) Step 5. Action All electrodes: Dry the electrode contact and place the new electrode in the measuring chamber. 6. If no more replacement actions are to be made, mount all covers and parts, and press Restart.

7-8

ABL800 FLEX Operator's Manual

7. Replacements

Replacing pump tubes


Replacing the electrode pump tubes The following diagram identifies the parts of the electrode module pumps to aid you in performing the replacement procedure.
pH/BG module El/Met module Met II module

Connector

Connector Pump rollers

Fastener in fastening slot

Fastener in fastening slot

Connector

To replace an electrode module pump tube, do the following: Step 1. 2. 3. 4. 5. 6. 7. Action Enter the Hold mode and record the replacement action. Disconnect the ends of the tube from connectors. Gripping one end of the tube, free the fastener from the fastening slot by pulling upward. Pull upward to free the remainder of the tube from around the pump rotor and the other fastening slot. Place a new pump tube around the pump rotor. Secure the upper fastener of the end of the tube in the fastening slot. Gripping the other fastener, wrap the tube around the pump rotor, laying the tube along the top of the rollers. Secure the fastener in the other fastening slot. Press the Rotate Pumps button to turn the pump and seat the tube. This can also be done manually by turning the rotor a few rotations in the both directions. NOTICE: If seated correctly, the tube should begin to be drawn downwards onto the rollers. The tubing is correctly drawn onto the rollers by the pump rotation during restart. 9. 10. Attach the ends of the tube to the connectors, pushing the tube onto the connector as far as possible. If no more actions are to be performed, remount all covers and restart the analyzer by pressing Restart.

8.

Continued on next page

7-9

7. Replacements

ABL800 FLEX Operator's Manual

Replacing pump tubes, Continued


Replacing the solution pump tube The solution pump tube:
Fastening slots

Pump clamp Connector (pump tube - short end) Pump tube long end

To replace the solution pump tube, do the following: Step 1. 2. 3. 4. 5. Action Enter the Hold mode and record the replacement action. Remove the pump clamp and disconnect the ends of the old tube from the connectors. Gripping the end of the tube, free the right fastener from the fastening slot by first pulling outward and then lifting upward. Pull upward to free the remainder of the old tube from around the pump rotor and out of the other fastening slot. Secure the fastener of the left short end of the new tube in the fastening slot. The short end of the tube is to be on the left, the long end on the right. Gripping the longer fastener, wrap the new tube around the pump rotor, laying the tube along the top of the rollers. Secure the fastener in the other fastening slot. Press the Rotate Pumps button to turn the pump and seat the tube. This can also be done manually by turning the rotor a few rotations in both directions. NOTICE: If seated correctly, the tube should begin to be drawn downwards onto the rollers by the pump rotation during restart. 8. 9. Attach the ends of the tube to the connectors, pushing the tube onto the connector as far as possible. Slide a new pump clamp (supplied with the pump tubes) over the right-side connector so that it is seated over the enlarged portion of the tube end, then clamp it shut. If no more actions are to be performed, remount all covers and restart the analyzer by pressing Restart.
Continued on next page

6.

7.

10.

7-10

ABL800 FLEX Operator's Manual

7. Replacements

Replacing pump tubes, Continued


Replacing the waste pump tubes When replacing the waste pump tubes, use the yellow tubes with black ends. These have a greater capacity than the other pump tubes, and leakage of fluid from the inlet due to a smaller capacity tube will be avoided.
Connectors - right side

Connectors left side Upper and lower tubes Fastener and fastening slot Upper and lower tubes Fastener and fastening slot

To replace the waste pump tubes, do the following: Step 1. 2. Action Enter the Hold mode and record the replacement action. Gripping the left end of the upper tube, free the fastener from the fastening slot by first pulling outward, then lifting upward. Pull upward to free the remainder of the tube from around the pump rotor, sliding the right fastener out of the fastening slot. 3. 4. Remove the lower tube in the same manner. Take the short end of one of the new yellow tubes and place the fastener into the lower right-side fastening slot. Gripping the other fastener, wrap the tube around the pump rotor, laying the tube along the top of the rollers. Secure the fastener in the lower left-side fastening slot. 5. Press the Rotate Pumps button to turn the pumps and seat the tube. This can also be done manually by turning the rotor a few rotations in the counterclockwise direction. If seated correctly, the tube should begin to be drawn downwards onto the rollers by the pump rotation during restart. 6. Connect the right end of the new tube to one of the right-side connectors and the left end of the tube to one of the left-side connectors. Ensure that tube ends are pressed completely onto the connectors. Repeat steps 4-6 with the other yellow tube except securing the tube fasteners in the upper fastening slots. If no more replacement actions are to be performed, remount all covers and restart the analyzer by pressing Restart.

7. 8.

7-11

7. Replacements

ABL800 FLEX Operator's Manual

Replacing inlet gasket unit and inlet probe


To replace the inlet gasket, do the following: Replacing the inlet gasket unit
Inlet gasket unit

Mount

Step 1. 2. 3.

Action Remove the capillary and then the syringe inlet flaps and record the action in the Hold mode. Remove the old inlet gasket unit by grabbing it on either side and lifting it upward. Slide the new inlet gasket unit onto the mount and press down the top of the unit to secure it in place. Mount the syringe inlet flap and then the capillary inlet flap. Press Restart.

4. Replacing the inlet probe

To replace the inlet probe, do the following:

Attachment and clip

Tube Inlet probe

Step 1. 2. 3. 4.

Action Remove the inlet flaps to enter the Hold mode. Remove the gasket unit see Replacing the inlet gasket unit above. Lift the clip to open it. Lift the inlet probe out of the attachment.
Continued on next page

7-12

ABL800 FLEX Operator's Manual

7. Replacements

Replacing inlet gasket unit and inlet probe, Continued


Replacing the inlet probe (continued) Step 5. 6. Action Disconnect the tube from the inlet probe and the nipple on the inlet module. Connect the new tube to the nipple on the inlet module and to the new probe. Then insert the inlet probe with the tube into the groove of the attachment and seat it properly. Close the clip. 7. Mount the inlet gasket unit and inlet flaps, then press Restart.

7-13

7. Replacements

ABL800 FLEX Operator's Manual

Replacing waste container, fan filter, printer paper


Replacing the waste container To replace the waste container, do the following: Step 1. 2. Action Remove the existing container by gripping its sides and pulling it off the connector. Place a new waste container onto the connector with the waste label facing upwards. Use an empty rinse solution container as waste container. Prior to use, remove the rinse label to reveal the underlying waste label. Close the filled waste container with a cap immediately.

3.

To monitor the volume of fluid in the waste container: look at the waste bottle in the Analyzer Status Reagents. check the waste bottle under the analyzers cover with the naked eye. If the waste container is critically full, the analyzer will enter the Forced Standby mode. If required by local authorities, a concentrated disinfectant can be added to the waste container before use or disposal. The final concentration of disinfectant in a filled container should be 20 % household bleach (1 % sodium hypochlorite), 10 % formalin (3.7 % formaldehyde) or 20 % glutaraldehyde[1]. NOTICE: Always handle a used waste container with care. Keep the waste container standing at all times in order to prevent any leaks from it. Replacing the fan filter If the filter is noticeably dirty, replace it to avoid unnecessary heating of the analyzer. To replace the fan filter (at the rear of the analyzer), do the following: Step 1. 2. Action Remove the existing filter from the fan compartment. Place the new filter in position on top of the fan.

To have this replacement recorded, replace the fan filter in the Hold mode.

Replacing the printer paper

NOTICE: Avoid contact of paper with direct sunlight, water, high temperature and humidity, alcoholic and organic solvents, PVC-containing materials, freshly developed diazo copy sheets, extensive pressure and scratching. For storage, use folders and cases made of polyethylene, polypropylene, polyester, etc.
Continued on next page

7-14

ABL800 FLEX Operator's Manual

7. Replacements

Replacing waste container, fan filter, printer paper, Continued


Replacing the printer paper (continued) To prepare the new roll, tear off and discard approximately the first 25 cm of paper and cut the leading edge of the paper straight.
Release lever

Printer cover (opened)

To replace the printer paper, do the following: Step 1. 2. 3. 4. Action Move the release lever fully back to the nearly horizontal position. Remove any leftover paper from the printer. Move the release lever forward. Make sure that the paper edge is cleanly cut. Place the new roll into position in the printer so that the paper unreels from underneath the roll. The thermal side of the paper is on the outside of the roll. Aligning the leading edge of paper straight, feed it behind the drive roller. The paper will feed through the printer automatically once the internal sensor detects the edge of the paper. Check that alignment is satisfactory. Realign as required. Close the printer cover, making sure the paper feeds out of the printer.

5.

6.

To realign the paper or to remove a paper jam, do the following: Step 1. 2. Action Place the release lever in the down position. Grip the paper edges close to the printer and gently pull the paper through the printer until alignment is satisfactory. Draw enough paper through the printer so that all damaged paper is removed. Place the release lever in the upright position. Close the printer cover making sure the paper feeds out of the printer through the provided slot.

3. 4.

7-15

7. Replacements

ABL800 FLEX Operator's Manual

Replacing solutions and gases


Preparing the Calibration and Cleaning Solution

CAUTION Risk of allergic reaction Contains streptokinase. May cause allergic reaction.

Prepare S1827 Calibration Solution 1, S1837 Calibration Solution 2 or S8375 Cleaning Solution, S8377 Cleaning Met II Solution as follows: Step 1. Action Remove the foil from the DosiCapZip and unscrew it.

2.

Turn the DosiCapZip upside down and screw it onto the container again. NOTICE: If the contents of the DosiCapZip or the container have been spilt by accident, both the container and the DosiCapZip should be discarded to prevent incorrect concentrations in the solution.

3.
x 20

Invert the container at least 20 times to dissolve the additive. Place the container horizontally so that the solution may enter the DosiCapZip and leave it for 3 minutes.

4.

5.
x 20

Invert the container again at least 20 times to fully dissolve the additive.

6. 7. 8. 9. 10.

Unscrew the lid from the new solution container. Remove the used solution container by holding it on the sides and pulling. Scan the barcode of the new solution, using the barcode reader. Place the new solution container in position on the analyzer and push it firmly onto the connector as far as possible. Press Restart and Accept to restart the analyzer.
Continued on next page

7-16

ABL800 FLEX Operator's Manual

7. Replacements

Replacing solutions and gases, Continued


Replacing solutions To replace a solution container, do the following: Step 1. Action Remove the solution container by gripping its sides and pulling it off the connector. The Recording Fluid Replacements screen is displayed. 2. 3. Remove the cap from the new bottle. Scan the barcode on the new solution container Always scan the barcode for each new solution container to provide precise concentration of the installed solution. 4. Place the new bottle in position label upward, and push it firmly and completely onto the connector.

5. 6.

Repeat steps 1-4 for the other solutions if required. For ABL8x7 FLEX analyzers only: Press Accept to confirm that the additive has been added and to start the proper restarting sequence or press Back to prepare the Calibration or Cleaning Solution properly.

7. Replacing gas cylinders

Press Restart to start the proper restarting sequence.

To replace a gas cylinder, do the following: Step 1. 2. Action Press Analyzer Status Reagents Replace. The Recording Fluid Replacements screen is displayed. Remove the gas cylinder by turning it counterclockwise until it becomes free. CAUTION Removal of safety valve Before discarding an empty cylinder remove the safety valve using the Valve Key (code number 922-509). 3. Remove the cap covering the valve on the new gas cylinder. Check that the valve is clear of any debris.
Continued on next page

7-17

7. Replacements

ABL800 FLEX Operator's Manual

Replacing solutions and gases, Continued


Replacing gas cylinders (continued) Step 4. 5. Action Scan the barcode on a new gas cylinder. Place the gas cylinder valve into position on the socket of the regulator. To ease alignment, use the analyzer enclosure as a guide.

6.

The replaced items appear in the Recorded replacements box as soon as the code has been scanned or typed. To remove an item from the recorded replacements list, highlight it with the up/down arrow buttons in the box and press Undo.

7.

Press Restart to start the proper restarting sequence.

CAUTION Risk of personal injury Pressurized container. Non-flammable compressed gas. Do not breathe gas. Gas mixtures containing less than 19.5 % oxygen may cause suffocation. Protect from sunlight and do not expose to temperatures exceeding 50 C. Store and use with adequate ventilation. Keep away from oil and grease. Do not refill.

7-18

ABL800 FLEX Operator's Manual

7. Replacements

Refilling the AutoCheck carousel


Refilling AutoCheck carousel To refill the AutoCheck carousel, do the following: Step 1. 2. 3. Action Open the retractable cover or go to the Analyzer Status screen and press AutoCheck - More - Open Module. Remove the carousel. Dispose of the used ampoules as infectious waste*. Fill the carousel with the control solutions according to the Optimal Packing List screen (the packing list is based on the selected quality control schedule).

4.

Place the packed carousel in the module and check that it sits properly in the module. NOTICE: Do not refill the carousel when it sits in the AutoCheck module as the analyzer will not register the action and will not scan the carousel contents.

5.

Close the cover either from the screen or on the module.

* Reference: Clinical laboratory waste management. CLSI document GP5-A2.

Messages after refilling AutoCheck carousel

After a packed carousel has been placed in the AutoCheck module and scanned, the following messages can be observed: Message Explanation

"AutoCheck Packing list Appears for 10 seconds in the status field if the confirmed" carousel has been packed according to the list. This message can, however, be overwritten in case of an activity that updates the status field. "Carousel packing not optimal" Appears at the bottom of the status field if the carousel has not been packed according to the list.

If, however, you have exited the screen and, for example, started a sample measurement, no message will be displayed in the status field.

7-19

7. Replacements

ABL800 FLEX Operator's Manual

Automatic auxiliary programs


Purpose The following programs might be requested: Program... Rinse Cleaning Is used... To remove traces of a sample, calibrating solution or a quality control solution from the liquid transport system. To remove lipid deposits from the liquid transport system and the electrodes by using the Cleaning Solution. The program is followed by a rinse. To make the cleaning start automatically see section Calibration schedule in Calibration setup in chapter 3 for details. Tubing refill To fill the liquid transport system with solution (is performed automatically during restart after replacement of any solution container). To adjust the liquid sensors in the Wet Section. To calibrate the rotation of the analyzer's pumps (is performed automatically during restart after replacement of the pump tubes).

Liquid sensor adjustment Pump calibration

Starting a program

To start one of the above programs, do the following: Step 1. Action Go to the Auxiliary programs menu: Menu - Start Programs Auxiliary Programs.

2. Interrupting a program

Press the button to start a program.

The following programs can be interrupted by pressing the Stop button: Cleaning Tubing refill Liquid sensor adjustment Pump calibration.

The Rinse program, once started, cannot be interrupted.

7-20

ABL800 FLEX Operator's Manual

7. Replacements

Decontamination and Protein Removal programs


Decontamination program The Decontamination program disinfects the liquid transport system and items in direct contact with blood such as electrodes, pump tubing, etc. CAUTION Replacements before decontaminations Do not perform the Decontamination Program without first replacing the cCl, cGlu, cLac, cCrea A and cCrea B electrodes with dummy electrodes, unless the membranes are to be replaced at this time. The Hypochlorite Solution used in decontamination will damage the membranes. It is recommended to perform Decontamination program once a month. To perform Decontamination, do the following: Step 1. 2. 3. Action Replace the cCl, cGlu, cLac, cCrea A and cCrea B electrodes with dummy electrodes, code 945-626. Fill a syringe with 0.5 mL of S5362 Hypochlorite Solution. Press Menu Start Programs Auxiliary Programs Decontamination. (To cancel the program at this time, press Close.) Open the syringe inlet flap and place the syringe tip in the inlet. Press Start to start aspiration of the solution and the program to begin. Remove the syringe and close the inlet flap as prompted on the screen. When the program is completed, the analyzer returns to the Ready screen. Replace the dummy electrodes with the cGlu, cLac, cCl, cCrea A and cCrea B electrodes. Perform a 2-point calibration.

4. 5. 6. 7. 8. 9.

Protein Removal The Protein Removal program is a shorter version of the Decontamination program and is used when required to clean the pH/BG and oximetry modules, pump program tubings and the inlets. Replacement of the cGlu, cLac, cCrea and cCl electrodes with dummy electrodes is not required. The Protein Removal program must be performed once a week in order to maintain the performance of the reference electrode membrane. To perform Protein Removal, do the following: Step 1. 2. Action Fill a syringe with 0.5 mL of S5362 Hypochlorite Solution. Press Menu Start Programs Auxiliary Programs Protein Removal. (To cancel the program at this time, press Close.)
Continued on next page

7-21

7. Replacements

ABL800 FLEX Operator's Manual

Decontamination and Protein Removal programs, Continued


Protein Removal program (continued) Step 3. 4. 5. 6. Action Open the syringe inlet flap and place the syringe tip in the inlet. Press Start to start aspiration of the solution and the program to begin. Remove the syringe and close the inlet flap as prompted on the screen. When the program is completed, the analyzer returns to the Ready screen.

NOTICE: The program can be terminated at any time by pressing the STOP button.

7-22

ABL800 FLEX Operator's Manual

7. Replacements

Cleaning the analyzer


Introduction The analyzer surfaces should always be kept clean of blood and/or other liquids. Immediately clean all surfaces if they become contaminated with blood or other liquids. CAUTION Safety precautions The use of rubber gloves is recommended when cleaning the analyzer

Cleaning the measuring chambers

To clean the measuring chamber, do the following. Step 1. 2. 3. 4. 5. 6. Action Lift the window to the measuring modules to enter the Hold mode automatically. Remove the electrode from the measuring chamber. Clean the measuring chamber using a cotton swab moistened in distilled water. Check that no cotton fibers are left in the chamber. Clean the electrode contact with a dry tissue if dirty or wet. Remount the electrode in the measuring chamber. Remount cover and press Restart to restart the analyzer.

Cleaning the inlet flaps and inlet gasket

To clean the syringe or capillary inlet, do the following. Step 1. 2. 3. 4. 5. 6. Action Remove the capillary inlet flap by sliding it off to the right. Remove the syringe inlet flap by sliding it to the right. The analyzer is placed in the Hold mode after draining. Remove the inlet gasket by pulling it vertically upwards. Clean the flaps and inlet area as required. Soak the inlet gasket in DeconexTM or a similar detergent. (Deconex is used in soaking baths and ultrasonic cleaning systems for cleaning laboratory utensils and precision components to remove moderate to most resilient contamination of organic nature.) Slide the inlet gasket onto the inlet mount, the syringe inlet flap onto the mounting post, followed by the capillary inlet flap. Check that the inlet probe is in the correct position. Ensure that both flaps are closed and press Restart to restart the analyzer.
Continued on next page

7.

8.

7-23

7. Replacements

ABL800 FLEX Operator's Manual

Cleaning the analyzer, Continued


Cleaning the analyzer exterior When cleaning the analyzer covers and outer case, use soapy water or a mild detergent. Do not use abrasive cleansers or pads, ethanol-based substances or aggressive detergents for cleaning. Use a dry or lightly dampened soft lint-free cloth to clean the analyzers touch screen. Wipe the screen gently to remove fingerprints and/or dirt. To avoid streaking, an approved screen cleaner is recommended.

Cleaning the analyzer touch screen

Use a tissue lightly dampened in mild soapy water. Cleaning the FLEXQ sampler If you wish to clean the sampler tray cover separately, do the following: tray cover Step 1. Action Carefully lift the empty sampler tray cover as shown.

2. 3. Disinfection of outer surfaces

Use a tissue lightly dampened in mild soapy water to clean it. Remount the sampler tray cover

Disinfection of outer surfaces is performed when appropriate. Disinfection frequency depends on local requirements and the use of the analyzer. Prior to disinfection always ensure that that analyzer surfaces are clean and without residues from blood and/or liquids. CAUTION Safety precautions Follow legal and local rules for safe work practices with chemicals. Use the following disinfectants: 70 % isopropyl alcohol 70 % ethanol 4 % Diversol BX

Wipe the outer surfaces of the analyzer and the touch screen, using disinfectant on a paper towel or tissue.

7-24

ABL800 FLEX Operator's Manual

7. Replacements

List of references
1. Protection of laboratory workers from occupationally acquired infections; Approved Guideline Second edition. CLSI (former NCCLS) document M29-A2. Wayne, Pa: CLSI (former NCCLS), 2001.

7-25

7. Replacements

ABL800 FLEX Operator's Manual

7-26

8. Overview
Introduction

Disk Functions

This chapter describes the disk functions available on your analyzer. The Disk Functions programs are a data management tool for storing and retrieving data logs, analyzer setup configurations, and other data and system files. This chapter contains the following topics. General information ......................................................................................... Creating a WDC report .................................................................................... Backing up all data........................................................................................... Restoring all data ............................................................................................. Exporting data logs .......................................................................................... 8-2 8-4 8-6 8-8 8-9

Contents

Importing/exporting archives ........................................................................... 8-11 Saving setup ..................................................................................................... 8-13 Loading/restoring setup.................................................................................... 8-14

8. Disk functions

ABL800 FLEX Operator's Manual

General information
Disk Functions programs Press the following buttons to access the Disk Functions programs: Menu Utilities - Disk Functions.

The following programs are available by pressing a corresponding button: Button WDC Report Backup All Data Restore All Data Export Data Logs Function To make a Worldwide DATACHECK report. To make a backup of all data. Data is stored as a backup on a designated location. To restore a backup of all data files to the analyzer hard drive from a designated location. To export selected records from selected data logs.

Import/Export To export or delete archived data logs. Archives To import externally archived data logs. Save Setup Load Setup Restore Default Setup Eject CD To save the current setup of your analyzer. To load a previously saved setup. To restore all or only some Radiometer default settings. To eject a CD from the CD-drive. Similar buttons will be on the Source and Destination screens.
Continued on next page

8-2

ABL800 FLEX Operator's Manual

8. Disk functions

General information, Continued

Definitions

Setup data refers to information or files that configure the analyzer to operate according to settings entered through the Setup programs. All data refers to data contained in the analyzer's internal database, including but not limited to data logs, setup and system files.

Data storage options Disk handling rules

Information is stored on or retrieved from an internal hard disk, a network, CDRW, CD-R/RW or a removable drive (USB mass storage device). The CD-RW, CD-R/RW and a removable drive (USB mass storage device) should be handled according to the instructions printed on the packaging.

8-3

8. Disk functions

ABL800 FLEX Operator's Manual

Creating a WDC report


Purpose This function allows you to make a Worldwide DATACHECK (WDC) file for reporting monthly quality control data. For information regarding Worldwide DATACHECK reporting, refer to the Worldwide DATACHECK Manual. Storing a WDC report To make a WDC report, do the following: Step 1. Action Select the desired month in the "Select period" box, using the up/down arrows.

2.

Select destination by pressing the destination icon

Highlight the desired destination (removable or CD-RW drive, or another directory) by touching it on the screen. Use the Expand/Collapse button to choose a source directory/drive. Press the Back button to return to the previous screen.
Continued on next page

8-4

ABL800 FLEX Operator's Manual

8. Disk functions

Creating a WDC report, Continued


Storing a WDC report (continued) Step 3. Action In the File name box (the WDC Report screen), press the keyboard icon to type the file name. You can change the four characters WDC_. Confirm the entry on the keyboard and return to the WDC Report screen. 4. Send the file to the selected destination by pressing Export data. Or insert a disk and press Export data in the "Export data" box. Wait until the WDC Report screen appears and remove the disk with the WDC report.

Error messages

The following error messages may appear in the analyzer status: "Could not create output file" if the wrong disk type (e.g. filled or read-only) is inserted. "No statistical data found. WDC data not generated" if no data is available for the selected month.

8-5

8. Disk functions

ABL800 FLEX Operator's Manual

Backing up all data


Purpose This function is intended as a protection or security against the loss of data or system files that includes, but is not limited to, the following: Patient report data Patient profile data Setup data Quality control data (i.e. results, statistics, plots) Calibration results and setup (i.e. schedule) Activity data (i.e. replacement actions, system messages)

Manually performed backup: data can be stored on the internal hard disk, the network, CD-RW or removable drive. Automatic backup (can be selected see Disk Functions Setup in chapter 3): data can be stored on the internal hard disk or the network. In case of data loss or similar problem, the loss can be minimized by using the backup file and the Restore All Data function. NOTICE: It is the user's responsibility to ensure that all valuable data is regularly backed up. During the warranty period of the ABL800 FLEX analyzers Radiometer accepts warranty responsibility for the original storage hardware and installed software only. To make a backup of all data, do the following: Step 1. Action Press Change Destination to choose the destination.

Creating a backup of all data

2.

Select the drive/directory by touching it on the screen. If a CD-RW is used, place it in the CD-drive. If a removable drive is used, connect it to the USB port. Use the Expand/Collapse button to choose a destination directory, then press Back to return to the previous screen.
Continued on next page

8-6

ABL800 FLEX Operator's Manual

8. Disk functions

Backing up all data, Continued


Creating a backup of all data (continued) Step 3. 4. Action On the Backup All Data screen press Start to continue. (Or press Back to cancel and return to the Ready screen.) The backup process begins. Network drive or internal hard disk drive: backup continues without any further action from the operator. CD- or removable drive: wait for the data to be prepared (see the timer in the current task field located beside the status indicator in the upper left corner of the screen) and press Start.

5.

If the analyzer status shows a Backup done message, the process is complete. Press Close to return to the Ready screen.

8-7

8. Disk functions

ABL800 FLEX Operator's Manual

Restoring all data


Purpose You can restore all data in case of loss or damage, provided the backup of all your data is available. NOTICES: When restoring all data, any existing data will be overwritten by the data obtained from the backup files. When restoring all data is complete, the analyzer will automatically and without warning shut down and restart. Restoring all data To restore a backup of data files, do the following: Step 1. Action Press Change Source to choose the source drive/directory.

2.

Select the drive/directory by touching it on the screen. If a CD-RW is used, place it in the CD-drive. If a removable drive is used, connect it to the USB port. Use Expand/Collapse to choose a source directory, then press Back to return to the previous screen.

3. 4.

On the Restore All Data screen press Start to continue. (Or press Close to cancel and return to the Ready screen.) The restore process begins. Network drive or internal hard disk drive: restoring does not require any further action. CD or removable drive: press Start.

5.

Complete restoring all data. When restoring is complete, the analyzer shuts down and then restarts automatically, configured to the information obtained from the backup file.

8-8

ABL800 FLEX Operator's Manual

8. Disk functions

Exporting data logs


Purpose You can export data from the data logs to a CD-RW, removable disk, network. The exported files are made in a form of a compressed "comma separated value (CSV)" file which can be read using a number of standard database and spreadsheet programs, e.g. Microsoft Excel, Access, Lotus 123, etc. NOTICE: It is important that the disk to which data is being exported is not ejected from the disk drive during the process. To export data logs, do the following: Step 1. Action Activate the check buttons beside the data logs that are to be exported.

Exporting data logs

2.

Activate the calendar icon.

Type the From: date; confirm with Enter. Type the To: date; confirm with Enter. Press Back to return to the Export Data Logs screen.
Continued on next page

8-9

8. Disk functions

ABL800 FLEX Operator's Manual

Exporting data logs, Continued


Exporting data logs (continued) Step 3. Action On the Export Data Logs screen press the drive icon. Activate the desired drive by touching it on the screen. Press Expand/Collapse to select the destination directory. Press Back to return to the Export Data Logs screen. 5. 6. On the Export Data Logs screen press Start to begin the export of data to the selected destination. If the dates are different for each exported data log, repeat steps 2-5 for each data log.

8-10

ABL800 FLEX Operator's Manual

8. Disk functions

Importing/exporting archives
Purpose This function allows you to do the following: 1. To export archived data logs onto any media. 2. To delete archived data logs. 3. To import externally archived data logs into the analyzer's archive directory from any location.

For exporting archives

For importing archives

Exporting an archive

To export an archive, do the following: Step 1. Action Select the desired archive type by activating one of the four archive type buttons.
Select source or destination

2.

Highlight the desired archive with the up/down arrow buttons.


Continued on next page

8-11

8. Disk functions

ABL800 FLEX Operator's Manual

Importing/exporting archives, Continued


Exporting an archive (continued) Step 3. Action To export the highlighted archive, select the location by pressing the icon. Touch and highlight the desired location on the Source/Destination screen. To open a folder contained in a drive or within a folder, highlight the drive or folder, then press Expand/Collapse. Press Back to return to the Import/Export Archives screen. 4. On the Import/Export Archives press the Export button. Press Refresh to update the contents of a drive or directory.

Importing an archive

To import an archive, follow the procedure for exporting an archive, using the right-hand section of the screen and the Import button.

Deleting an archive

To delete an archive from a directory, use the Delete button.

8-12

ABL800 FLEX Operator's Manual

8. Disk functions

Saving setup
Purpose You can copy your analyzer's current setup configuration onto a CD-RW, removable drive, hard disk or network. It can be reloaded in the event the current setup is lost or damaged or the same setup configuration should be loaded on other analyzers without performing all the Setup programs. To store the current analyzer setup, do the following: Step 1. Action Press Edit Location to select destination.

Storing the setup

2.

Select required location by touching and highlighting it on the screen. If a CD-RW is used, place a CD-RW in the CD-drive. If a removable drive is used, connect it to the USB port. Press Expand/Collapse to select the destination directory. Press Back to return to the Save Setup screen. On the Save Setup screen press Start. When saving is complete, the Ready screen is displayed. Remove CD-RW (use the Eject CD button) or removable drive.

3. 4.

8-13

8. Disk functions

ABL800 FLEX Operator's Manual

Loading/restoring setup
Purpose You can re-install a saved setup quickly and easily without performing the Setup programs. If desired, only part of the setup can be loaded, e.g. operators.

Loading a setup To load a setup, do the following: Step 1. Action Press Select All to include all items from the list on the screen. Or press Deselect All to exclude all items from the list.

2.

To select single items, do the following: Highlight the desired item, using the up and down arrow buttons. Press the check button () to include an item.

3. 4.

To open or close a group of items, highlight the group title (e.g. General) and press the button. Press the Change Source button to select the source. If a CD-RW is used, place it in the CD-drive. If a removable disk is used, connect it to the USB port. Select required source by touching and highlighting it on the screen. Press the Expand/Collapse button to access the required folder. The chosen source appears in the Choose a directory box.

5. 6.

Press the Back button to return to the Load Setup screen. Press Continue. The analyzer will shut down and then restart with reloaded setup configuration. Pressing Cancel will terminate loading the setup.

Contents of Setup settings

See chapter 15: Radiometer settings in this manual.

8-14

9. Overview
Introduction

Data management

This chapter describes the functions of the Patient Results log, Patient Profile log, Quality Control log, Calibration log and Activity log. It also describes the archiving of data logs and the optional feature: RADIANCE browser. This chapter contains the following topics. General information ......................................................................................... Patient Results Log .......................................................................................... Patient Profiles Log.......................................................................................... 9-2 9-4 9-7

Contents

Quality Control Log ......................................................................................... 9-11 Calibration Log ................................................................................................ 9-17 Activity Log ..................................................................................................... 9-20 Replacement Log ............................................................................................. 9-23 Archived Data Logs ......................................................................................... 9-24 RADIANCE Browser (optional)...................................................................... 9-26

9. Data management

ABL800 FLEX Operator's Manual

General information
Access to data logs To access Data Logs, press Menu - Data Logs.

The following data logs are available: Log Patient Results Log Description Contains up to 2000 patient reports from which you can view patient report, edit Patient ID, print and/or send patient report, search for a patient report, and make a parameter trend for the selected patient reports. Provides a list of patient profiles containing general information about patients, from which you can find and view patient profiles, add, edit or remove a patient profile.

Patient Profiles Log

Quality Control Contains up to 1500 quality control results from which you Log can view a quality control result, view statistics and plots of results, print and/or send result, search for a particular result, and make a parameter trend for the selected quality control results. Calibration Log Contains up to 1000 calibration results from which you can find and view result, make a parameter trend for the selected calibrations, print and/or send result. Activity Log Contains up to 5000 records of all measurements, replacements, system messages and operator messages. From the log, you can search an activity, add an operator message, print and/or send records.
Continued on next page

9-2

ABL800 FLEX Operator's Manual

9. Data management

General information, Continued


Archived data logs To keep the database clean and responsive, data can be archived into respective archived data logs. If a need to examine the old archived data should arise, the archived data can be reloaded.

Press Menu - Data Logs - Archived Data Logs to access the archived Patient Results Log, Quality Control Log, Calibration Log and Activity Log.

9-3

9. Data management

ABL800 FLEX Operator's Manual

Patient Results Log


Purpose The Patient Results Log is a historical file of patient reports automatically saved in the Patient Results Log after a measurement.

Each patient report contained in the log is listed as an individual record. The reports are listed chronologically, the latest report being at the top of the screen. Each report is identified as follows: Item Sample # Time Sample type Status Description Indicates the number of the sample. Date and time sample measurement was performed. Shows the type of blood sample specified in Patient ID. Indicates the status of the sample measurement:
OK

a successful measurement an error detected or a parameter exceeded the reportable range a measurement stopped by the operator a measurement stopped by the analyzer most likely due to insufficient sample.

?
Interrupted Aborted

Patient ID Last Name

Patient identification number, up to 20 characters. Patients last name, up to 50 characters.

NOTICE: The columns displayed in the log and their positions can be changed. Contact a qualified service technician.
Continued on next page

9-4

ABL800 FLEX Operator's Manual

9. Data management

Patient Results Log, Continued


Button functions The following functions are available by activating a button: Button Result Filter Close Trend and Print Function Displays the highlighted patient report see chapter 4: Sample measurements for the detailed information. Sets conditions for finding a patient report or for trend function see below. Returns to the Ready screen. Are inactive until the filter criteria are chosen and applied.

Filter function

To set the filter criteria in order to find a patient report or to select a number of patient reports, do the following: Step 1. Action Set the start date by using the 1-day, 7-day, 14-day or 30-day icons in the Criteria box. Or type in the date, using the keypad and confirming the entry with Enter.

2. 3.

Change if necessary the end date by highlighting it and typing date other than the current date. Confirm with Enter. Use the arrow buttons to highlight the desired search criteria. Press More to display more filter criteria. Note that the right side of the screen changes to show choices for the highlighted search criterion.

4.

Select the desired choice and confirm each entry with Enter. Or use the keyboard to enter the desired data and confirm each entry with Enter. When all the criteria have been selected, press Apply to start the search.
Continued on next page

5.

9-5

9. Data management

ABL800 FLEX Operator's Manual

Patient Results Log, Continued


Trend function To make a parameter trend for the selected patient reports, do the following: Step 1. Action Set up search criteria as described under Filter function and apply them. The screen of the selected patient reports will be displayed.

The Trend and Print buttons are now active. 2. 3. 4. Press Trend to select the parameter(s). Press View Trend to display the trend of the selected parameter(s) for the selected patient reports. To view the selected filter criteria, press View Filter. The criteria can only be viewed, not changed. To apply new criteria, repeat the Filter function procedure described in this topic and the procedure for the Trend function. 5. Press Back to return to the previous screen.

9-6

ABL800 FLEX Operator's Manual

9. Data management

Patient Profiles Log


About a patient profile A patient profile is a record containing general information about a particular patient. It is created automatically whenever a new patient ID number is entered during measurement and recorded in the Patient Profiles log. Alternatively, a patient profile can be added to the log. The given patient data recorded in a patient profile will be automatically used as soon as patient ID number is entered on the Patient Identification screen during a measurement. NOTICE: If the analyzer is set up to request data from a LIS/HIS or the RADIANCE system, this patient information will always override that from the Patient Profiles log.

Patient Profiles log

The patient profiles are listed in the order they were created.

The following data can be entered in a patient profile (use the left/right scroll bar to see the items): Item Patient ID Last Name First Name Sex Date of birth Department (Pat.) Patient note Birth weight Gestational age Description Lists the patient ID number Provides the patients surname. Provides the patients first name. Identifies the patients sex. Identifies the patients date of birth. Identifies the department in which the patient is located. Lists a note regarding the patient. Gives the weight at birth. Gives the duration of pregnancy.

The functions available on the Patient Profiles screen are described in this section.
Continued on next page

9-7

9. Data management

ABL800 FLEX Operator's Manual

Patient Profiles Log, Continued


Editing a patient To edit a patient profile, do the following: profile Step Action 1. Highlight the patient profile and press Edit.

2.

Use a barcode reader to scan the patient data supported by the barcode (the "Enable General Barcode Support" should be activated in the Miscellaneous Setup) see the procedure in Entering information with barcode reader in chapter 4, section Patient identification. Or request (if connected) the data from LIS/HIS by verifying the patient ID and pressing the Request button.

3.

To edit data on the screen, highlight an item, using the arrow buttons, and type in the data, using the screen keyboard or keypad. Confirm each entry with Enter.

4.

When completed, press Back to return to the previous screen.


Continued on next page

9-8

ABL800 FLEX Operator's Manual

9. Data management

Patient Profiles Log, Continued


Adding a new patient profile To add a new patient profile, do the following: Step 1. Action Press Add New to display the screen below.

2.

Use a barcode reader to scan the patient data supported by the barcode (the "Enable General Barcode Support" should be activated in the Miscellaneous Setup) see the procedure in Entering information with barcode reader in chapter 4, section Patient identification. Or request (if connected) the data from LIS/HIS by verifying the patient ID and pressing the Request button.

3.

To edit data on the screen, highlight an item, using the arrow buttons and type in the data, using the screen keyboard or keypad. Confirm each entry with Enter. When completed, press Back to return to the previous screen.

4.

Deleting a patient profile

To delete a patient profile from the log, do the following: Step 1. 2. Action Highlight the desired patient profile. Press Remove.
Continued on next page

9-9

9. Data management

ABL800 FLEX Operator's Manual

Patient Profiles Log, Continued


Finding a patient profile To find a patient profile, do the following: Step 1. Action Press Find on the screen.

2.

Using the arrow buttons, highlight the search criterion and enter the data using the screen keypad or keyboard. Or use a barcode reader to scan the Patient ID barcode. Press Find. If a patient profile is found, it will be highlighted on the Patient Profile Log screen. If a patient profile is not found, "No item found that matches search criteria" message appears in analyzer status. Repeat step 2 or press Back to cancel.

3.

9-10

ABL800 FLEX Operator's Manual

9. Data management

Quality Control Log


Purpose The Quality Control Log is a historical file of all quality control results automatically saved in the Quality Control Log after a measurement. Each quality control result is listed chronologically as an individual record.

Each result is identified as follows (use the left/right scroll bar to see the items): Item Time QC# Slot Solution Lot Status Description Date and time the measurement was performed. The ID number of a quality control measurement. Control solution slot as defined in setup. Control solution type as defined in setup. Lot number of the control solution. Indicates the status of the quality control measurement:
OK

a successful measurement an error detected or a measured value exceeded a control, statistical or measuring range. a measurement stopped by the operator a measurement stopped by the analyzer most likely due to insufficient sample.

Interrupted Aborted

NOTICE: The columns displayed in the log and their positions can be changed. Contact a qualified service technician. The functions of the buttons available on the Quality Control Log screen are described in this section.
Continued on next page

9-11

9. Data management

ABL800 FLEX Operator's Manual

Quality Control Log, Continued


Viewing a result To view a result, do the following: Step 1. 2. Action Highlight the desired quality control result using scroll facilities. Press the Result button. For details see Quality Control Result in chapter 5.

Filter/trend function

This function allows you to find a quality control result or to select a number of results in order to make a trend. To set the filter criteria, do the following: Step 1. 2. Action Press the Filter button on the Quality Control Log screen. Set the start date by using the 1-day, 7-day, 14-day or 30-day icons in the Criteria box.

Or type in the date using the keypad and confirming the entry with Enter. Change if necessary the end date by highlighting it and typing a date other than the current date. Confirm with Enter. 3. 4. Use the up and down arrow buttons to highlight the desired search criteria. Select the desired choice and confirm each entry with Enter. Or use the keyboard to enter the desired data and confirm each entry with Enter.
Continued on next page

9-12

ABL800 FLEX Operator's Manual

9. Data management

Quality Control Log, Continued


Filter/trend function (continued) Step 5. Action When all the criteria have been selected, press Apply to make the search. Press Trend to select the parameters. 6. Select the desired parameter(s) by activating the check buttons (e.g. pH, pCO2, pO2). Press View Trend. 7. The trend of the selected parameter(s) is displayed. To view the filter criteria, press the View Filter (the criteria cannot be changed, only viewed). To print the displayed trend, press Print. To return to the screen in step 5, press Back.

Viewing quality Only those measurement values that fall within the statistical range are included in control statistics the statistical data. Therefore parameters outside the measuring range or with a "?" are not included.
Parameter

The screen shows statistics for each control solution (identified by its slot, type and lot number) for a single parameter. Press Next param.. or Prev. param.. to view other parameters. Press WDC Report to create a WDC Report see Creating a WDC Report in chapter 8.
Continued on next page

9-13

9. Data management

ABL800 FLEX Operator's Manual

Quality Control Log, Continued


Viewing quality Month-to-date statistics are obtained from all measurements taken on the current control statistics lot of a control solution over the course of the month see QC Statistics in Quality control setup in chapter 3 for information. (continued) Lot-to-date statistics are obtained from all measurements taken over the entire course of the current lot of the control solution. The Month-to-date and Lot-to-date statistics give the following information: Item N Mean 2 SD CV Description Number of measurements included in statistics for the slot. Calculated mean of the measurement values*. Calculated 2 standard deviation of measurement values*. Coefficient of variation*.

Control Range Current control range for the control solution level. * See Quality Control Systems Reference Manual for detailed information. To print QC statistics, do the following: Step 1. 2. Action Press Print on the Quality Control Statistics screen. To print QC statistics for an entire lot, activate Print lot to date button and press Print.

Printing QC statistics

3. 4.

To print monthly QC statistics, activate Print for period. Select the month, using up and down buttons. Then press Print. Press Back to return to the Quality Control Statistics screen.
Continued on next page

9-14

ABL800 FLEX Operator's Manual

9. Data management

Quality Control Log, Continued


Quality control plot To view the Quality Control Plot screen press Plot on the Quality Control Log screen.

Parameter

The following diagram shows the details of a typical plot.


Slot number Solution type Control range upper limit Date Time Measurement Measured value Lot change Control range lower limit

Out-of-range symbol

Plot element Shaded block

Function Date, time and measured value for the highlighted measurement. Use the arrow buttons to scroll the plot and view other measurements. Show the upper and the lower limits of the control range for the highlighted measurement. Show the number of measurements for the selected parameter. To view a measurement result, highlight a dot on the screen and press Result. Measurement outside the control range and inside the statistical range. Measurement outside both the control and statistical ranges.
Continued on next page

Control range limits Dots

Out-of-range symbols

9-15

9. Data management

ABL800 FLEX Operator's Manual

Quality Control Log, Continued


Quality control plot (continued) Use the following buttons to do the following: Next Param. Prev. Param. Next Slot Result Print Back Display the quality control plots for the requested parameter. Display plots of the next control solution slot. View quality control result for the highlighted measurement of the highlighted plot. Print out the plot. Return to the previous screen.

(For the ABL8x7 FLEX analyzers only): The plot for the High Crea Check of the ABL8x7 FLEX analyzers presents Creadiff.

The screen is similar to the quality control solution plot, but the Next Param. and Prev Param. buttons are absent.

9-16

ABL800 FLEX Operator's Manual

9. Data management

Calibration Log
Purpose The Calibration Log is a historical file of calibrations automatically saved in the Calibration Log after a calibration.

Each calibration report is listed as an individual record. The reports are listed chronologically, the latest report being at the top of the screen. Each report is identified as follows (use the left/right scroll bar to see the items): Item Time Cal # Type Status Description Date and time the calibration was performed. The ID number of a performed calibration. Indicates the calibration type: 1-point, 2-point or tHb. Indicates the calibration status:
OK

a successful calibration an error detected or a parameter exceeded the measuring range a calibration stopped by the operator a calibration stopped by the analyzer most likely due to missing calibration solution.

?
Interrupted Aborted

NOTICE: The columns displayed in the log and their positions can be changed. Contact a qualified service technician.
Continued on next page

9-17

9. Data management

ABL800 FLEX Operator's Manual

Calibration Log, Continued


Viewing a calibration result To view a calibration result, do the following: Step 1. 2. Action Highlight the desired calibration. Press Result see chapter 6 for details about calibration results.

Filter function

This function allows you to find a calibration or to select a number of calibrations. To set the filter criteria, do the following: Step 1. 2. Action Press the Filter button on the Calibration Log screen. Set the start date by using the 1-day, 7-day, 14-day or 30-day icons.

Or type in the date using the keypad and confirm with Enter. Change if necessary the end date by highlighting it and typing the date other than the current date. Confirm with Enter or Select. 3. Select the desired calibration type and status, using the relevant arrow buttons. Confirm each entry with Enter or Select. 4. When the criteria have been selected, press Apply to start the filter.
Continued on next page

9-18

ABL800 FLEX Operator's Manual

9. Data management

Calibration Log, Continued


Trend function This function allows you to make a parameter trend for the selected calibrations. To make a trend, do the following: Step 1. Action Set up the search criteria as described above and apply them.

Press Trend to select a parameter. 2. 3. Select a parameter (only one for calibrations) and press View Trend. To view the filter criteria, press the View Filter (the criteria cannot be changed, only viewed). To print the displayed trend, press Print. To return to the screen in step 5, press Back 4. To apply the new criteria, repeat the Filter function procedure described in this topic and the procedure for the Trend function.

9-19

9. Data management

ABL800 FLEX Operator's Manual

Activity Log
Purpose The Activity Log provides a historical record of all replacements, system messages and errors during operation. It also lists any operator messages and allows the entry of additional notes.
Level

The records are listed chronologically and can be displayed at user, manager or service level. The record is identified as follows (use the left/right scroll bar to see the items): Item Time Message Description Shows the date and time when the activity occurred. Gives a message number and an explanation of the activity. Operator note text appears as written.

NOTICE: The columns displayed in the log and their positions can be changed. Contact a qualified service technician. Adding a message To add a message to the log, do the following: Step 1. 2. Action Press Add Message on the Activity Log screen. If Notes were entered in Setup program User-defined Notes (see Parameters and Input setup in chapter 3: Installation and Setup), then you can edit a highlighted message by pressing Edit Msg. to display the keyboard and make the changes.

To delete a message, highlight it and press Delete Msg.


Continued on next page

9-20

ABL800 FLEX Operator's Manual

9. Data management

Activity Log, Continued


Adding a message (continued) Step Action

If Notes were not entered in Setup program User-defined Notes 2 (cont.).

To key in a note, press the keyboard touch-key to display the keyboard, type a message and press Enter to confirm and return to the Activity Log screen
3.

Press Back to return to the Activity Log screen

Filter function

This function allows you to find an activity or to select a number of activities. To set the filter criteria, do the following: Step 1. 2. Action Press the Filter button on the Activity Log screen. Set the start date by using the 1-day, 7-day, 14-day or 30-day icons.

Or type in the date using the keypad and confirm with Enter. Change if necessary the end date by highlighting it and typing the date other than the current date. Confirm with Enter.
Continued on next page

9-21

9. Data management

ABL800 FLEX Operator's Manual

Activity Log, Continued


Filter function (continued) Step 3. Action Use the arrow buttons to highlight the desired search criterion or criteria. NOTICE: If you filter the Activity Log for "Replacement" only and set the message level to Manager or Service, the "Activity Log" changes to "Replacement Log". 4. 5. When all the criteria have been selected, press Apply to start. The Activity Log displays the list of selected activities.

By selecting the message level, you can see messages on the user, manager or service levels. The Add Message button is available on all levels. 6. Press Filter to set other search criteria if desired. Press Close to return to the Ready screen.

9-22

ABL800 FLEX Operator's Manual

9. Data management

Replacement Log
Replacement Log The Replacement Log provides a historical record of replacements only. To get a list of all the replacements, do the following: Step 1. 2. 3. 4. Action Enter the Activity Log screen. Select the Manager or Service message level. Press the Filter button. Set the start date by using the 1-day, 7-day, 14-day or 30-day icons.

Or type in the date using the keypad and confirm with Enter. 5. 6. Highlight the criteria Replacement and press Apply. The Replacement Log appears with all the replacements and an Operator and Note field.

7.

Press Close to return to the Ready screen.

9-23

9. Data management

ABL800 FLEX Operator's Manual

Archived data logs


Purpose You can view the old patient reports, quality control results, calibration results and activities which have been archived automatically when maximum log size (2000 patient reports, 1500 quality control results, 1000 calibration results, and 5000 activity log entries) has been reached. The archives are located on the analyzer's hard disk and can be exported to a CDRW, removable disk, network, etc. see Importing/exporting archives in chapter 8: Disk functions. The archives are archived after date: 500 reports from each log, and 2000 entries from the Activity log. Press Menu - Data Logs - Archived Data Logs and select the log archives.

Press the button on the screen to select the log archives. As the functions are similar for all archives, the Calibration Archive is used as an example. Selecting archive stored on the analyzer To select an archive saved on the analyzer, do the following: Step 1. Action Highlight the desired archive with the up/down arrow buttons (if a few archives are present).

Information about analyzer type, installation number and the date this archive was generated is on the screen.
Continued on next page

9-24

ABL800 FLEX Operator's Manual

9. Data management

Archived data logs, Continued


Selecting archive stored on the analyzer (continued) Step 2. 3. 4. 5. Action Press Select Archive to obtain it for viewing. Press, if desired, Filter to enter the search criteria and apply the filter. Highlight the desired report by touching it on the screen and press Result to display the result. Press Close to exit to the Ready screen.

Moving an archive

To move an archive to another media (e.g. computer) or from another media to the analyzer, do the following: Step 1. 2. Action Press Import Archive to move it. The Import/Export Archives screen will be displayed. Proceed as described in Importing/exporting archives, chapter 8: Disk functions.

To convert an archive into .csv format, do the following: Converting an archive into .csv format Step Action 1. Highlight an archive and press Export Archive.

2. 3.

Press Start to start conversion. Press Back to cancel and return to the previous screen. Export the converted archive to a disk as described in Importing/exporting archives, chapter 8: Disk functions.

9-25

9. Data management

ABL800 FLEX Operator's Manual

RADIANCE browser (optional)


Purpose This option allows you to get in touch with the RADIANCE system in order to find information on patients, quality control, calibrations, from all analyzers connected to the RADIANCE network. To get access to the RADIANCE system, do the following: Step 1. 2. Action Make sure that a connection to the RADIANCE server is established in the RADIANCE Communication Setup program. Press Menu - Utilities - RADIANCE Browser.

Access to the RADIANCE system

The analyzer will exchange information with the RADIANCE system about your access profile, and the contents of the RADIANCE main page will reflect the scope of your privileges.

3.

Refer to the RADIANCE Operator's Manual on how to browse on the RADIANCE system.

9-26

10. Overview
Introduction

Analyzer shutdown

This chapter describes the Standby mode, Temporary Shutdown, Long Term Shutdown programs and the Forced Standby mode. This chapter contains the following topics. General information ......................................................................................... 10-2 Standby mode................................................................................................... 10-3 Full waste container ......................................................................................... 10-5 Temporary Sutdown......................................................................................... 10-6 Long Term Sutdown ........................................................................................ 10-8

Contents

10. Analyzer shutdown

ABL800 FLEX Operator's Manual

General information
Shutdown programs The following programs are available: Standby Temporary Shutdown Long Term Shutdown. To access the programs, press Menu - Utilities.

Then press the desired button to enter the program. When the waste container becomes critically full, the analyzer automatically enters Forced Standby mode.

10-2

ABL800 FLEX Operator's Manual

10. Analyzer shutdown

Standby mode
Purpose The Standby mode is used for the following: to keep the analyzer ready for measurements (the electrodes conditioned and solutions thermostatted) during short-term time periods such as weekends or holidays. to prevent spillage from the waste container when it gets critically full. By placing the analyzer on standby the consumption of solutions is reduced, as the scheduled calibrations are suspended. Entering Standby To put the analyzer into Standby mode, do the following: Step 1. Action Enter date and time for the analyzer to automatically exit the Standby mode, using the arrow buttons.

2.

Press Enter Standby to enter the Standby mode. The button changes to Exit Standby.
Continued on next page

10-3

10. Analyzer shutdown

ABL800 FLEX Operator's Manual

Standby mode, Continued


Exiting Standby The analyzer automatically exits the Standby mode on the selected day and time.

Press Exit Standby to exit standby before the scheduled date/time. If the Exit Standby button is not shown, you have to logon to the analyzer as a user with permission to exit standby. The analyzer will return to the Ready mode upon exit from Standby, unless: all parameters are marked "Cal Expired" analyzer is looked due to "QC Expired". Analyzer is locked due to "Replacement overdue".

Pending calibrations and/or QC measurements will be performed when possible.

10-4

ABL800 FLEX Operator's Manual

10. Analyzer shutdown

Full waste container


Forced Standby When the waste container gets critically full, the analyzer enters the Forced Standby mode.

To return the analyzer to the Ready mode, do the following: Step 1. 2. Action Replace the waste container. Press the Exit Standby button. For users with limited access (selected in the Access Profiles program see chapter 3), this button may not be visible. Press Menu Log on and type in your password. If your access profile allows the exit from the Standby mode, the Exit Standby button will be visible.

10-5

10. Analyzer shutdown

ABL800 FLEX Operator's Manual

Temporary Shutdown
Purpose This program prepares the analyzer software for switching off power, and is used for temporary (short-term) shutdown of the analyzer. This program differs from the Long Term Shutdown used for a long-term storage of an analyzer that requires removal of system components. Executing Temporary Shutdown program To switch the analyzer off for a short time, do the following: Step 1. Action Enter the program.

2. 3.

Press Confirm Shutdown to continue. Press Close to cancel. Follow the instructions on the screen to complete the temporary shutdown.

Restarting the analyzer

To restart the analyzer after Temporary Shutdown, do the following: Step 1. 2. Action Verify that the system requirements are met and all components are installed. Place the power switch in the ON or I position. The startup procedure begins. When completed, the analyzers Ready screen appears Depending on the length of the shutdown, the analyzer performs either a cold start or a warm start.
Continued on next page

10-6

ABL800 FLEX Operator's Manual

10. Analyzer shutdown

Temporary Shutdown, Continued


Restarting the analyzer (continued) Start Cold Is performed if... The thermostatting temperature is outside predefined limits. It lasts approximately 25 minutes and includes: Warm Loading of software Inlet calibration (not shown in status) Leak test Initializing Liquid sensor adjustment Pump Calibration Rinse Startup (conditioning of the electrodes) 2-point calibration.

The thermostatting temperature is within the predefined limits. It lasts approximately 5 minutes and includes: Loading of software Inlet calibration (not shown in status) Leak test Rinse

Over the first 24 hours after the cold start, the analyzer will automatically perform a number of calibrations at specific intervals as follows: The first four hours: remaining 20 hours: 1-point calibration every 30 minutes 2-point calibration every hour 1-point calibration every hour 2-point calibration every four hours. During the startup, availabe analyzer functions are indicated by the appearance of the buttons. A greyed-out button indicates that the function is not available. NOTICE: Message 962 "Ambient temperature not specified" will be present after the ABL837/27/17 cold startup in the System messages on the Analyzer Status screen. To remedy it, enter the ambient temperature in the Environment Setup (press Menu Utilities Setup General Setup Analyzer Settings Environment).

10-7

10. Analyzer shutdown

ABL800 FLEX Operator's Manual

Long Term Shutdown


Purpose This program gives the procedure for emptying the analyzer of solutions before transport or storage for an extended period of time. The Shutdown procedure takes approximately 15 minutes. Before shutdown You will need the following: Four ~50 cm long tubes with an internal diameter capable of fitting onto the solution connectors. One large glass beaker filled with distilled water The shutdown procedure Enter the Long Term Shutdown program and follow the prompts on the screen.

Screen 1

Prompts and operator actions Perform Decontamination (use the button on the screen to enter the program).

Press Confirm Shutdown to continue, or Close to cancel the procedure. After this point there is no opportunity to cancel the shutdown. 2 3 4 Remove the Rinse, Cal 1, Cal 2 and Cleaning Solution containers. Press Continue. Wait for the fluid transport path to drain. Place tubes on the Rinse, Cal 1, Cal 2 and Cleaning solution connectors and the free ends in a beaker filled with distilled water. Press Continue. Wait while the fluid transport path is rinsed with distilled water. Remove beaker with distilled water and disconnect the tubes. Press Continue. Wait for the fluid transport path to drain.
Continued on next page

5 6 7

10-8

ABL800 FLEX Operator's Manual

10. Analyzer shutdown

Long Term Shutdown, Continued


The shutdown procedure (continued) Screen 8 Prompts and operator actions Read the steps below prior to performing any actions. Remove Gas 1 and Gas 2 cylinders. Remove all electrodes and place them in their respective protective jackets. Remove waste container. Place covers on all solution connectors and gas inlets. Press Finish to power down. When the message "You can now turn off your PC" appears, turn off the power (the power switch at the back of the analyzer). Remove pump tubings. Remove paper from printer, and network and power cables from the back of the unit. Shutdown is now complete. Storing the analyzer Transporting the analyzer Radiometer recommends to place the analyzer on a trolley in its normal position, and protect it from dust using a plastic cover. To transport the analyzer from one location to another, put it back in its original packaging and seal. To transport the analyzer without its original packaging, follow the recommendations below: A minimum of two persons is recommended. Remove all solutions and gas containers; disconnect power and peripheral devices prior to lifting. The analyzer can be lifted and carried from any position along the base apart from the gas regulator. As a recommendation, see the diagram below outlining the analyzer base. Hand positions are indicated by the arrows.
Back

Front

NOTICE: Do not lift analyzer by the gas regulator. Damage to the analyzer can occur.

10-9

10. Analyzer shutdown

ABL800 FLEX Operator's Manual

10-10

11. Overview
Introduction Contents

Troubleshooting

This chapter describes Hold causes and standard troubleshooting procedures. This chapter contains the following topics. General information ......................................................................................... 11-2 Forced Hold causes .......................................................................................... 11-5 Analyzer messages ........................................................................................... 11-7 Fluid transport troubleshooting procedure ....................................................... 11-74 Inlet probe troubleshooting procedure ............................................................. 11-75 Inlet troubleshooting procedure ....................................................................... 11-76 Leak troubleshooting procedure....................................................................... 11-77 Electrode troubleshooting procedures.............................................................. 11-78 Pump troubleshooting procedure ..................................................................... 11-80 Fluid transport system description ................................................................... 11-81

11. Troubleshooting

ABL800 FLEX Operator's Manual

General information
Analyzer action Depending on the severity of the error the analyzer will do one of the following: in case of error continue its activity, but mark relevant parameter results with a ? interrupt and abort its activity enter the Hold mode deny measurements and calibrations. FLEXQ sampler messages If a sampler was accidentally removed from the slot, the following screen appears:

If the batch mode was selected, but the analyzer is not connected to the RADIANCE system, the message Remove sampler will appear on the screen If two samplers were placed in the sampler tray one immediately after the other so that the FLEXQ had no time to register the first sampler, the message Remove sampler will appear on the screen. To locate and remedy messages/errors, do the following: Operator actions in case of error Step Action 1. 2. Check which of the status button indicators is red or yellow on the Analyzer Status screen. Press the relevant button (example: System Messages was pressed).

Highlight the error (the first line is automatically highlighted).


Continued on next page

11-2

ABL800 FLEX Operator's Manual

11. Troubleshooting

General information, Continued


Operator actions in case of error (continued) Step 3. Action Press Troubleshoot for error description and operator action.

Is placed in the top right corner of the screen so that you can continue troubleshooting. Use Tutorials to assist in the troubleshooting procedure. Online Help at the bottom of the screen. Returns to the previous help topic.

Press to return to the Analyzer Status screen when all messages have been remedied. Press to return to the Analyzer Status screen. 4. 5. 6. Remedy the error as described in the Actions on the screen. Remedy other error(s) if present. Press Troubleshooting completed after all errors have been remedied. (The Help button will then reappear in the top right corner of the screen).

NOTICE: For description of analyzer messages and operator actions see Analyzer messages.
Continued on next page

11-3

11. Troubleshooting

ABL800 FLEX Operator's Manual

General information, Continued


Logging system messages All system messages are automatically recorded in the Activity Log at the time of their occurrence. The items in the Activity Log are listed in chronological order and form a permanent record for the user of all system messages that have taken place. Using the filter function (see description in Activity Log, chapter 9 in this manual), you can select the desired type of messages and view them at the user, manager and service levels.

11-4

ABL800 FLEX Operator's Manual

11. Troubleshooting

Forced Hold causes


Purpose Forced Hold mode suspends all the wet section activities in case certain analyzer conditions occur. Until the condition has been corrected the analyzer cannot be restarted.

The causes for the Forced Hold mode are as follows: Condition Forced Standby Rinse error Screen message Replace waste container Rinse Error Corrective action Replace the waste container. Press the Exit Standby button. Replace the Rinse Solution bottle see the procedure in chapter 7: Replacements. Check Analyzer Status and remedy the condition/error. Verify the cover is properly mounted and the window shut prior to restart. Verify the inlet flaps are properly mounted prior to restart. Contact service representative. Perform the procedure described on the next page. Perform the procedure described on the next page.

Operator intervention required

Intervention Required

Cover to the measuring Cover removed or modules removed or window opened the window opened. Both inlet flaps removed. Inlet flaps removed

The inlet cannot obtain Inlet positioning proper position. error Inlet calibration failed Leakage test failed Inlet calibration error Leak detected

Inlet calibration This condition occurs when the calibration of the inlet fails, most likely caused by a problem with the inlet components. error To correct the error, do the following: Step 1. 2. 3. Action Remove the analyzer cover and the inlet flaps. Check that the inlet probe and gasket are mounted properly. Remove the inlet gasket and check for wear or damage. Replace as required.
Continued on next page

11-5

11. Troubleshooting

ABL800 FLEX Operator's Manual

Forced Hold causes, Continued


Inlet calibration error (continued) Step 4. 5. Action Check the condition of the inlet probe. If it appears bent or damaged, replace it. Mount all components, then restart the analyzer by pressing Restart. If the error has been corrected, the analyzer performs an inlet calibration during restart and enters the Ready mode. If the Hold mode remains, contact a service representative.

Leak Detected error

This condition occurs when a leakage test has failed, indicating a leak in the fluid transport system. To correct the error, do the following: Step 1. 2. 3. 4. 5. Action Remove the analyzer covers and inspect the fluid transport system for leaks. Check for poor tube connections or worn tubes. Repair as required. Check the inlet gasket; change if required. Ensure that electrodes are properly mounted in the measuring chamber and the electrode connector is latched shut. Mount all items, restart the analyzer by pressing Restart. If the error has been corrected, the analyzer performs a restart sequence (includes a leakage test) and enters the Ready mode. If the Hold mode remains, contact a service representative.

11-6

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages
Messages at manager level The following messages will be seen at the manager levels. The messages are listed in numerical order. Operator actions are listed in order of priority. Perform the first action in the list; if unsuccessful, try the next action, etc. No. 1 Message Inconsistent software versions. Please contact service. Interpretation Inconsistent software versions for different modules. May appear after replacing a complete module or as a result of an incomplete software upgrade. Operator action Contact Radiometer service representative. Removal condition: Successful software consistency check. No action required.

83

Value above Parameter value is reference above the user-defined range reference range. This is only a message, not an error.

84

Value below Parameter value is reference below the user-defined range reference range. This is only a message, not an error.

No action required.

85

Value is Parameter value is below the below the user-defined critical limit critical limit. This is only a message, not an error.

No action required.

86

Value is Parameter value is above the above the user-defined critical limit critical limit. This is only a message, not an error.

No action required.

Continued on next page

11-7

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 89 Message Measured QC value above the control range Interpretation Measured parameter value is above the control range. Operator action - Verify procedure and repeat measurement. - Refer to Quality Control Systems Reference Manual. Crea: - Mount a new Cleaning Met II Solution. - Remembrane both Crea A and Crea B electrodes 90 Measured QC value below control range Measured parameter value is below the control range. - Verify procedure and repeat measurement. - Refer to Quality Control Systems Reference Manual. Crea: - Mount a new Cleaning Met II Solution. - Remembrane both Crea A and Crea B electrodes 93 Value above the reportable range Value below the reportable range Parameter value is above the reportable range. See chapter 13: Specifications in the Operator's Manual Parameter value below the reportable range. See chapter 13: Specifications in the Operator's Manual. - Check for and remedy other errors related to the result, system messages, or calibration status. - Perform QC. If the QC result is accepted, the blood sample may be suspected. - Repeat measurement on a new blood sample. - Perform the Electrode Troubleshooting procedure. Check the communication parameters specified in Communications setup.

94

117

LIS/HIS: Invalid connection configuration

The comunication configuration or the protocol definition was invalid.

Continued on next page

11-8

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 128 Message LIS/HIS: Failed to open connection Interpretation The communication hardware was busy or the remote system did not respond. Operator action - Check that the remote system is running, correctly configured and responding. - Check communication parameters, e.g. baud rate, parity, IP address, etc., as defined in Communication setup. - Reboot the analyzer. If the problem persists, check the communication hardware. The remote system may lack buffer capacity.

129

LIS/HIS: Failed to close connection

Messages were queued when the communication channel was closed. Results and other messages sent by the analyzer to a remote system may be lost. A communication error occured while sending a message. The message was not sent.

131

LIS/HIS: Failed to send packet

132

LIS/HIS: Failed to receive packet

133

LIS/HIS: Connection lost

- Check that the remote system is running and responding. - Check communication hardware including cables. - Repeat sending. An error occured while - Check that protocol types are correctly configured on receiving a message. both the analyzer and the The analyzer was not remote system. able to recognize the - Contact Radiometer received massage. service representative. - Check that the remote A previously system is running and established LIS/HIS responding. connection has been - Check cables. lost.
Continued on next page

11-9

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 134 Message LIS/HIS: Connection established LIS/HIS: High level protocol could not generate high level packet LIS/HIS: General communication error LIS/HIS: High level protocol received packet in wrong format User msg: Interpretation The connection was successfully established. An error occurred while formatting a message. Operator action No action required. The message is informative. - Check protocol configurations. - Contact Radiometer service representative.

165

166

An internal error occurred in the LIS/HIS communication module. An error occurred while parsing (interpreting) a message.

Contact Radiometer service representative if the problem persists.

167

- Check protocol configurations. - Contact Radiometer service representative.

200

This is only a message. No action required. An operator has entered a note in the log. Measured parameter value is outside the mean 2 SD range. Measured parameter value is outside the mean 3 SD range. Two consecutive measurements are outside the mean 2 SD range on the same side of the mean. This may indicate a shift.

201

Westgard Rule (1.2s) violation Westgard Rule (1.3s) violation Westgard Rule (2.2s) violation

- Verify procedure and repeat measurement. - Check Replacement Status for pending replacements (including elctrodes). - Refer to the Quality Control Systems Reference Manual for detailed evaluation procedure.

202

203

Continued on next page

11-10

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 204 Message Westgard Rule (R.4s) violation Interpretation The difference between two consecutive measurements exceeds 4 SD. This may indicate an inconsistency in your procedure or an unstable analyzer. Operator action - Verify procedure and repeat measurement. - Check Replacement Status for pending electrode replacements. - Refer to the Quality Control Systems Reference Manual for detailed evaluation procedure. - Check for excesive electrode sensor calibration drift. - Check Replacement Status for pending electrode replacements. - Refer to Quality Control Systems Reference Manual for evaluation procedure. - Check the electrode drift during last calibration. - Check Replacement Status for pending electrode replacements. - Refer to Quality Control Systems Reference Manual for evaluation procedure. Check the Calibration Status and perform any pending calibrations. Check the Quality Control Status and perform the pending quality control.

205

Westgard Rule (4.1s) violation

Four consecutive measurements are outside the mean 1 SD range on the same side of the mean. A trend or shift is indicated. Patient results should be considered unreliable until the problem is remedied. Ten consecutive measurements are on the same side of the mean. A trend or shift is indicated. Patient results should be considered unreliable until the problem is remedied. One or more scheduled calibrations are overdue. One or more scheduled QC measurements are overdue.

206

Westgard Rule (10.x) violation

207

Calibration schedule reminder(s) present Quality control schedule reminder(s) present Replacement schedule reminder(s) present

208

209

One or more scheduled replacements are overdue.

Check the Replacement Status and perform any pending replacement actions.

11-11

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 210 Message Calibration error(s) present Interpretation An error registered on one or more parameters during the last calibration. Operator action Check the Calibration Status for errors in latest calibration results for the given parameter. View calibration error messages and take required corrective action. Check the Quality Control Status for errors. View QC error messages and take required corrective action. Check the System Messages Status for errors. Take required corrective action. - Check Automatic Backup Setup. - Check network and servers used for the backup. - Contact your IT engineer. 214 Automatic backup succeeded The scheduled No action required. automatic backup was completed successfully.

211

Quality control error(s) present System message(s) present Automatic backup failed

One or more errors registered during last QC measurment in one of the installed QC levels. One or more systems errors are present.

212

213

An error occurred during the scheduled data backup.

216

Printer error The printer is out of - Check printer paper. paper, paper is jammed, Clear any jam or mount a or some other printer roll. problem has occurred. - Power down and restart the analyzer. - Contact Radiometer Service Representative.

217

Replacement:

The message is used in the Activity Log to indicate a performed replacement.

No action required.

Continued on next page

11-12

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 218 Message Inlet positioning error Interpretation Inlet probe was not positioned properly within required time frame. Causes the activity to abort. 219 Conditioning error in El/Met electrode module Rinse Solution was not detected in El/Met electrode module. The glucose and lactate results will be reported as "....." in the next measurement. - Check Rinse Solution for adequate volume, proper installation or obstructions in container opening. Replace as necessary. - Perform the Fluid transport troubleshooting procedure. Removal condition: Successful Rinse. 232 Oximetry calibration error Water spectrum intensity was too low or too high during absorbance calculation. OXI measurements cannot be performed. - Perform a 2-point calibration. - Contact Radiometer Service Representative if the problem persists. Removal condition: Successful 2-point calibration. 234-238, Oximetry 245-252, hardware 254-258, error 269, 314316 259, 270, 354-357 The Oxi module has a hardware error. - Perform a 2-point calibration. - Contact Radiometer Service Representative if the problem persists. - Ensure the ambient temperature is between 15 and 32 degrees Celsius. - If the system has just performed a cold start, wait for error to disappear. - Replace the fan filter if dirty. - Shield analyzer from direct sunlight and other heat sources. - Contact Radiometer service representative.
Continued on next page

Operator action Perform the Inlet Troubleshooting Procedure Removal condition: Successful inlet calibration.

Temperature Analyzer temperature error is outside 37.0 0.2 degrees Celsius.

11-13

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 290 Message Warning: SHb detected SHb too high Interpretation FSHb detected in the range of 1-10 %. Detected FSHb is greater than 10 %. Measurement accuracy is affected. Turbidity is greater than 5 %: too high for reliable measurements. Operator action No action required. For information only. Repeat the measurement.

291

292

Turbidity too high

- Check pump tubes and inlet tube for leaks. Repair as necessary. - Hyperlipemic sample; decrease the lipemic content by, e.g. centrifuge or extraction. - Perform the measurement on a blood sample from a healthy donor. - Contact Radiometer service representative.

293

Warning: HbF detected and compensated for Empty Gas 1 cylinder Empty Gas 2 cylinder

Detected FHbF is greater than 20 % of ctHb. Oxi parameters are automatically corrected. Gas 1 cylinder is detected as empty. Gas 2 cylinder is detected as empty.

No action required. For information only.

326 327

- Enter Replacement mode via Replacement Status. - Scan the barcode of a new Gas 1 or 2 cylinder. - Replace Gas 1 cylinder. (See chapter 7: Replacements for detailed information.) Removal condition: Gas pressure in Gas 1 or 2 cylinder.
Continued on next page

11-14

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 328 Message No leading air segment in inlet's liquid sensor within time frame Interpretation Air segment in front of sample is too small or missing during sample aspiration, possibly due to obstructions in inlet or problems with liquid sensors. Measurement is aborted. The quality control measurement was performed on an expired control solution. No sample detected in the pH/BG module. Measurement is aborted. Operator action - Check inlet for obstructions and clean as required. - Perform liquid sensor adjustment. - Perform the Fluid Transport Troubleshooting procedure. Discontinue the use of the lot and set up a valid lot for the control solution.

329

QC expiration date exceeded No sample detected during sample aspiration

331

- Ensure that adequate sample volume is used. - Check the sample for clots. - Perform the Fluid Transport Troubleshooting procedure.

332, 333

pH/BG module not filled

Sample did not fill the pH/BG module properly during aspiration. Measurement is aborted.

- Ensure that adequate sample volume is used. - Check the sample for clots. - Perform the Fluid Transport Troubleshooting procedure. - Check rinse solution for sufficient volume. Replace as necessary. - Repeat pump calibration. - Perform the Pump Troubleshooting procedure. Removal condition: Successful pump calibration.
Continued on next page

339-340. 342-343, 345-346

Pump calibration error

The pump calibration failed for one or more pumps. The previous pump calibration is used. A successful pump calibration should be attempted as soon as possible.

11-15

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 369 Message Interpretation Operator action - Check the solution containers. Replace as required. - Perform the Fluid transport troubleshooting procedure. Removal condition: Successful calibration. 374 Inhomogeneous sample at pH/BG module pH/BG lower liquid sensor detected air bubbles in the sample; they might affect the measurement. All parameters are marked with "?". 375 Calibration The status value is status out of outside the range for limits the given parameter: pH: 6.7-8.1 pCO2: 6.2-260 mmHg or 0.83-34.66 kPa cK+: 0.5-12 mmol/L cNa : 10-250 mmol/L cCa2+: 0.1-20 mmol/L cCl: 30-900 mmol/L
+

El/Met not El/Met module did not filled within fill properly with time frame calibration solution. Calibration is aborted.

- Ensure the sample is free of air bubbles and that sufficient volume is used. - Check the condition of the inlet gasket. Replace as necessary. - Check for and remedy any System Messages. - Perform any pending electrode replacements. - Check that electrodes are properly installed. - Verify that proper solutions and gases are being used. - Perform the Electrode Troubleshooting procedure. Removal condition: Successful 1- or 2-point calibration.
Continued on next page

11-16

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 376 Message Calibration Drift 1 out of range Calibration Drift 2 out of range Interpretation The Drift 1 value exceeds the userdefined tolerance. The Drift 2 value exceeds the userdefined tolerance. Operator action - Check for and remedy any System Messages. - Perform any pending electrode replacements. - Check that electrodes are properly installed. - Verify that proper solutions (all additives are added) and gases are used. - Perform the Electrode Troubleshooting procedure. Removal condition: Successful 1- or 2-point calibration. 378 Calibration The sensitivity value is sensitivity out of range for the out of range given parameter: pH: 92-103 % pCO2: 85-100 % - Check for and remedy any System Messages. - Check that electrodes are properly installed.

377

- Verify that proper solutions (all additives are pO2: 5-40 pA/mmHg or added) and gases are used. 37.5-300 pA/kPa - Perform the Electrode cK+: 92-105 % Troubleshooting procedure. cNa+: 90-105 % Removal condition: cCa++: 90-105 % Successful 1- or 2-point calibration. cCl : 85-100 % cGlu: 100-1800 pA/mM cLac: 150-2000 pA/mM cCrea: 5-15 pA/M (1) cCrea: 65-85 % (2) cCrea: 50-200 % or 3.5-20 pA/M (3) Affected parameters and measurements are marked with "?".
Continued on next page

11-17

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 379 Message Calibration unstable. (response fault) Interpretation An electrode response fault occurred during calibration. Affected parameters and measurements (only creatinine) are marked with "?". Operator action - Check for and remedy any System Messages. - Perform any pending replacements including electrodes. - Check that electrodes are properly installed. - Verify that proper solutions and gases are used. - Perform the Electrode Troubleshooting procedure. Removal condition: Successful 1- or 2-point calibration. 386 Barometer The barometer reading out of range is out of the range: 450800 mmHg or 60.0106.7 kPa. Measurement results are marked with "?". - Check the measured pressure in Barometer Setup program and compare it to that of an external barometer. Adjust the analyzer pressure accordingly. - Contact Radiometer service representative. 408 Zero current Zero current was too - Check for and remedy error high during calibration. System Messages. Affected parameters and measurements are marked with "?". - Perform any pending replacements, including electrodes. - Perform the Electrode troubleshooting procedure. - Contact Radiometer service representative. 443 Ca(7.4) not usable cCa++ at a pH of 7.4 is No action required. not usable as the actual pH is out of the range 7.2-7.6.
Continued on next page

11-18

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 452 Message Interpretation Operator action Check the patient record for medication containing possible interfering substances. - Repeat the measurement ensuring that the sample is free of air bubbles and that proper sample volume is used. - Replace Pump Tubing in pH/BG and El/Met modules and solution pump. - Contact Radiometer Service Representative. 468 Inhomogene ous sample in Met II module Lower liquid sensor at Met II module detected air bubbles. The affected parameters are marked with "?". - Repeat the measurement ensuring that the sample is free of air bubbles and that proper sample volume is used. - Replace Pump Tubing in pH/BG, El/Met and Met II modules and solution pump. - Contact Radiometer Service Representative. 474 The sample did not fill No sample the El/Met module in El/Met upper liquid properly. sensor within time limit - Check the sample for clots. - Ensure sample volume matches the selected sample mode. - Perform the Fluid Transport Troubleshooting procedure.
Continued on next page

Interference Interference was detected during during measurement. measurement Inhomogeneous sample at El/Met module Lower liquid sensor at El/Met module detected air bubbles. The affected parameters are marked with "?".

467

11-19

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 475 Message No sample in Oxi liquid sensor within time limit Interpretation The sample did not fill the oximetry module properly. Operator action - Check the sample for clots. - Ensure sample volume matches the selected sample mode. - Perform the Fluid Transport Troubleshooting procedure. 476 Measurement unstable The electrode response fault occurred during measurement. - Check for and remedy other errors in System Message or Calibration Status. - Repeat the measurement. - Perform any pending replacements, including electrodes. - Ensure the electrode is properly installed. - Ensure the analyzer is properly grounded. - Perform QC. If the QC result value is within the control range, the blood sample may be suspected. Continue troubleshooting. - Perform the Electrode Troubleshooting procedure. 478 pCO2 drifting The pCO2 drift value exceeds the userdefined limits + 0.3 kPa during flush. The pO2 drift value exceeds the user defined limits + 0.3 kPa during flush. - Call 1-point calibration. - Check for and remedy any specific errors in the last calibration result, then call another 1-point calibration. - Contact Radiometer service representative.
Continued on next page

480

pO2 drifting

11-20

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 481-482 Message Interpretation Operator action

Conditioning Indicates a fluid error transport problem of pH/BG module during Flush. Today is last day in current statistical month remember to print QC statistics After the current day, Print the QC statistics if a copy is required. quality control statistics obtained over the month will be deleted and new statistics started.

484

487

A new statistical A new month has begun. statistical month has begun remember to export WDC data Warning: Bilirubin detected and compensated for

Make a WDC report disk. Removal condition: A WDC report disk has been made.

493

No action required. Detected bilirubin concentration is greater than 50 umol/L. The corresponding plasma bilirubin concentration can be calculated as follows: ctBil(blood) = (1-Hct) x ctBil(plasma).

494

No action required. Bilirubin too Detected bilirubin high concentration, ctBil(blood), is greater than 2000umol/L. The corresponding plasma bilirubin concentration can be calculated as follows: ctBil(blood) = (1-Hct) x ctBil(plasma).
Continued on next page

11-21

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 499 Message Sample too early in Oxi liquid sensor. Interpretation The Oxi module detected liquid prematurely during sample aspiration. The activity is aborted. Operator action - Check the sample for clots. - Repeat measurement. - Replace inlet gasket. - Perform the Fluid transport troubleshooting procedure. 500 Rinse detected by pH/BG module when not expected Rinse not aspirated into the pH/BG module within time limit Unable to fill inlet with continuous rinse segment during Pump Calibration No sample in El/Met lower liquid sensor within time limit The pH/BG module detected liquid unexpectedly. Perform the Fluid transport troubleshooting procedure.

501

The pH/BG module was not filled properly with Rinse Solution.

- Check the Rinse Solution for sufficient volume and proper mounting. Replace as necessary. - Perform the Fluid transport troubleshooting procedure.

503

Three attempts to fill the inlet probe with a continuous rinse segment have failed; the Pump Calibration program aborts.

- Check Rinse Solution for sufficient volume. Replace as necessary. - Perform the Pump troubleshooting procedure.

504

The sample did not fill - Check the sample for El/Met module clots. properly. The activity is - Ensure that sample aborted. volume matches selected sample mode. - Perform the Fluid transport troubleshooting procedure.
Continued on next page

11-22

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 505 Message Met II upper liquid sensor did not detect Cleaning Solution Interpretation Operator action

506

The Met II module was not filled properly during Cleaning - Check the Cleaning program. The activity solution for sufficient is aborted. volume. Replace as required. The El/Met module El/Met upper liquid was not filled properly - Perform Fluid transport troubleshooting procedure. during Cleaning sensor did program. The activity not detect Removal condition. is aborted. Cleaning Successful Cleaning. Solution The pH/BG module was not filled properly during Cleaning program. Cleaning is aborted. Rinse Solution was not detected during Rinse program. Measurements and calibrations cannot be performed until the condition is removed. - Check the rinse solution for sufficient volume. Replace as required. - If the message "Inlet flow impeded" (765) has been reported, replace or clean inlet probe. - Perform Fluid transport troubleshooting procedure. Removal condition: Successful Rinse.

507

pH/BG upper liquid sensor did not detect Cleaning solution Rinse error

508

509

pH/BG upper liquid sensor did not receive Rinse from above El/Met upper liquid sensor did not receive Rinse from above

The pH/BG module was not filled properly during Rinse program; automatic Refill is attempted. The El/Met module was not filled properly during Rinse program; automatic Refill is attempted.

- Check the Rinse Solution for sufficient volume. Replace as required. - Perform Fluid transport troubleshooting procedure. Removal condition: Successful Rinse.

510

Continued on next page

11-23

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 512 Message Temperature error Interpretation The temperature was outside the required range during measurement or calibration. All results are marked with "?". Operator action Ensure ambient temperature is between 15 and 32 C. - If the analyzer has recently performed a Cold start, wait for temperature error to disappear. - Shield analyzer from direct sunlight or heat sources. - Contact Radiometer service representative. 513 No gas Measurement or calibration was performed with Gas error present. Activity is aborted. - Check preasure in gas cylinders. Replace as required. - Verify that cylinders are properly installed. - Enter the Replacement screen (via Replacement Status) and re-scan the barcode labels of the gas cylinders. 521 Inhomogeneous sample Air bubbles were detected in the sample. Results may have "?". - Ensure sampler and sample volume match the selected sample mode. - Repeat measurement. - Check pump tubes and inlet for leaks. Replace tubing and inlet gasket as required. 529, 531- Liquid 532, 534- sensor 537 calibration error One or more of the liquid sensors failed to calibrate. - Repeat liquid sensor calibration. - Perform Fluid transport troubleshooting procedure. - Contact Radiometer Service Representative.
Continued on next page

11-24

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 538 Message pH/BG pump error Interpretation The pH/BG module pump failed to calibrate. Previous calibration is used until a successful pump calibration is performed. The El/Met module pump failed to calibrate. Previous calibration is used until a successful pump calibration is performed. Operator action - Perform Pump troubleshooting procedure. - Contact Radiometer service representative.

539

El/Met pump error

- Perform Pump troubleshooting procedure. - Contact Radiometer service representative.

540

Met II pump The Met II module error pump failed to calibrate. Previous calibration is used until a successful pump calibration is performed. Solution pump error The solution pump failed to calibrate. Previous calibration is used until a successful pump calibration is performed. No Cleaning Solution was detected by Oxi module during Startup.

- Perform Pump troubleshooting procedure. - Contact Radiometer service representative.

541

- Perform Pump Troubleshooting procedure. - Contact Radiometer service representative.

564

No Cleaning Solution detected by Oxi liquid sensor during initializing. No Cal 2 Solution detected by Oxi liquid sensor during initializing

- Check the Cleaning solution for sufficient volume. Replace as necessary. - Perform Fluid transport troubleshooting procedure.

567

No Cal 2 Solution was - Check the Cal 2 solution detected by Oxi module for sufficient volume. during Startup. Replace as necessary. - Perform Fluid transport troubleshooting procedure.

Continued on next page

11-25

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 569 Message No Cal 1 Solution detected by Oxi liquid sensor during initializing No Rinse Solution detected by OXI liquid sensor during initializing Oximetry hardware error Interpretation No Cal 1 Solution was detected by the Oxi module during Startup. Operator action Check the Cal 1 solution for sufficient volume. Replace as necessary. - Perform Fluid transport troubleshooting procedure.

571

No Rinse Solution was detected by the OXI module during Startup.

- Check the Rinse solution for sufficient volume. Replace as necessary. - Perform Fluid transport troubleshooting procedure.

573

The OXI module has a hardware error.

- Perform a 2-point calibration. - Contact Radiometer Service Representative if the problem persists.

574

Oximetry calibration error

The measuring system - Perform a 2-point has no calibration data. calibration. It is not possible to - Contact Radiometer make a measurement. Service Representative. Removal condition: Successful 2-point calibration.

577-578

Option key error

An error occurred in the analyzer protection system. The analyzer cannot be used.

- Shut down the analyzer using the Temporary Shutdown function, then restart it. - Contact Radiometer Service Representative. Removal condition: Key installed correctly and accepted by analyzer.

579

Oximetry hardware problem. Not possible to measure.

Parameter cannot be realiably measured because of a hardware problem.

Perform a 2-point calibration. - Contact Radiometer service representative. Removal condition: No errors detected by oximetry hardware.
Continued on next page

11-26

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued)

No. 580

Message Oximetry hardware problem. Possible to measure.

Interpretation A hardware problem exists in the OXI module. Parameter can be measured.

Operator action - Perform a 2-point calibration. - Contact Radiometer service representative. Removal condition: No errors detected by oximetry hardware.

581

Oximetry measuring error

Spectrum deviates from expected blood or QC spectrum. Measurement may be unrealiable.

- Check patient record for medication containing possible interfering substances. - Call a 2-point calibration. - Contact Radiometer Service Representative.

582

tHb calibration outside limits Measured value outside reportable range for the parameter

tHb calibration failed.

Repeat tHb calibration. Removal condition: Successful tHb calibration.

583

The parameter value is outside the defined reportable range or the analyzer measurement range.

- Check for and remedy other errors in the result, and any errors in System Messages or Calibration Status. - Perform QC. If the QC result value lies within the measuring range, the blood sample may be suspected. - Perform the Electrode Troubleshooting procedure.

588

Measured QC value lower than statistical range

The parameter value is below the lower limit of the user-defined statistical range. Measurement is not included in statistics.

- Verify procedure and repeat the measurement. - Refer to the Quality Control Systems Reference Manual for details on the evaluation of the results.
Continued on next page

11-27

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 589 Message Measured QC value higher than statistical range Interpretation The parameter value is above the upper limit of the user-defined statistical range. Measurement not included into statistics. Sample volume is too small for the selected measuring mode. Affected parameters will be marked with "?". Operator action - Verify procedure and repeat the measurement. - Refer to the Quality Control Systems Reference Manual for details on the evaluation of the results. - Make sure sample volume matches the selected measuring mode. - Repeat the measurement, ensuring sufficient sample volume. - Contact Radiometer service representative. 597 Liquid detected in pH/BG module during Gas Calibration Demo option will expire soon Unexpected liquid is detected by the pH/BG module during calibration. Calibration is aborted. One or more currently installed demo options will expire and be removed within the next 7 days. - Repeat calibration. - Perform Fluid transport troubleshooting procedure.

593

Insufficient sample

600

To have the option permanently, contact a Radiometer sales or service representative. Removal condition: Options expired or permanent options installed. Contact Radiometer service representative.

604

Parameter Parameter was not not installed installed or is corrupted. Parameter cannot be measured.

Continued on next page

11-28

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 605 606 608 609 610 611 612 Message Cal expired Cal expired (pH) Cal expired (pCO2) Cal expired (pO2) Cal expired (K) Cal expired (Na) Cal expired (Ca) Interpretation Too long time passed since the last successful calibration of the parameter. Calibrations of the parameter are considered invalid and values reported as ".....". Operator action Perform 1- or 2-point calibration. Removal condition: Successful 1- or 2-point calibration.

613 614 615 616 618

Cal expired (Cl) Cal expired (Glu) Cal expired (Lac) Cal expired (OXI) Waste detector error The waste detector may Contact Radiometer service be faulty or damaged. representative. Removal condition: Functioning waste detector installed. The waste container is missing or improperly mounted. Mount the waste container properly. Removal condition: Waste container properly mounted.
Continued on next page

619

Waste container missing

11-29

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 620 Message Waste container full Interpretation Operator action

Replace waste container. Waste container is filled to near-maximum Removal condition: Empty capacity. Unless Waste container installed. replaced soon, the analyzer will enter Standby until the container is replaced. The waste container is filled to maximum capacity. The analyzer enters Standby mode; no measurements or calibrations can be performed. One or more of the solution barcodes was entered incorrectly during replacement. The analyzer cannot monitor solution consumption. Replace waste container. Removal condition: Empty Waste container installed.

621

Waste container critically full

622

Expected liquid level below 0

Check the solution containers and replace as necessary. Remember to scan barcodes. Removal condition: Solution containers correctly installed and solution barcodes correctly recorded by the analyzer. Check the solution levels and replace containers as necessary. Removal condition: Sufficient level of all solutions.

623

Solution empty

Wet section activities are suspended due to one or more empty solution containers.

624

Solution missing

One or more solution Check that all solution containers is missing or containers are mounted improperly mounted. properly. Removal condition: All solution containers are mounted correctly.
Continued on next page

11-30

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 628 Message No Cleaning Solution detected by Oxi liquid sensor during Refill No Cal 2 Solution detected by Oxi liquid sensor during Refill No Cal 1 Solution detected by Oxi liquid sensor during Refill No Rinse Solution detected by Oxi liquid sensor during Refill Interpretation OXI module was not filled correctly with Cleaning Solution. Operator action - Check Cleaning solution level in the container. Replace as necessary. - Perform Fluid transport troubleshooting procedure.

630

OXI module was not - Check Cal 2 Solution filled correctly with Cal level in the container. 2 Solution. Replace as necessary. - Perform Fluid transport troubleshooting procedure.

632

OXI module was not filled with Cal 1 Solution.

- Check Cal 1 Solution level in the container. Replace as necessary. - Perform Fluid transport troubleshooting procedure.

634

OXI module was not filled with Rinse Solution.

- Check Rinse Solution level in the container. Replace as necessary. - Perform Fluid transport troubleshooting procedure.

636

Gas 1 Low pressure level is cylinder low detected in Gas 1 pressure cylinder.

Replace the Gas 1 cylinder within a few days. Removal condition: Sufficient pressure in Gas 1 cylinder. Replace the Gas 2 cylinder within a few days. Removal condition: Sufficient pressure in Gas 2 cylinder.
Continued on next page

637

Gas 2 Low pressure level is cylinder low detected in Gas 2 pressure cylinder.

11-31

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 641 Message Interpretation Operator action No action required. For information only. - Check pump tubing in El/Met module and solution pump. Replace as necessary. - Contact Radiometer service representative.

ABL/DMS The analyzer was PC restarted restarted from power off. Missing sample at El/Met module in C95 uL mode The El/Met module was not filled properly with sample. Measurement is aborted.

644

648

Calibration The last calibration was - Check solution levels and failed or not aborted or not replace containers as accepted accepted. necessary. - Check for and remedy System Messages. - Repeat the calibration. Removal condition: Successful calibration. Oxi liquid sensor did not receive Rinse from the Fluidic module Rinse solution was not detected by the OXI module. - Check Rinse Solution for sufficient volume and proper mounting. Replace as necessary. - Perform Fluid transport troubleshooting procedure. Removal condition: Successful Rinse.

652

653

pH/BG module was not Sample in filled correctly during pH/BG upper liquid sample aspiration. sensor detected prematurely during aspiration

- Ensure sample is free of clots. - Repeat measurement. - Perform Fluid transport troubleshooting procedure.

Continued on next page

11-32

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 654 Message Interpretation Operator action - Ensure sample is free of clots. - Repeat measurement. - Perform Fluid transport troubleshooting procedure.

El/Met module was not Sample in filled correctly during El/Met upper liquid sample aspiration. sensor detected prematurely during aspiration pH/BG module was not Sample in filled correctly during pH/BG upper liquid sample aspiration. sensor detected prematurely during aspiration OXI module was not Sample in OXI liquid filled correctly during sample aspriration. sensor detected prematurely during aspiration Inlet flaps removed The analyzer entered HOLD because both inlet flaps were removed.

655

-Ensure sample is free of clots. - Repeat measurement. - Perform Fluid transport troubleshooting procedure.

656

- Ensure sample is free of clots. - Repeat measurement. - Perform the Fluid Transport Troubleshooting procedure. Ensure the inlet flaps are properly mounted prior to restart. Removal condition: Inlet flaps properly mounted. Position the cover and close the window prior to restart. Removal condition: Cover in position and window closed. - Remove the inlet gasket. Inspect the gasket and inlet probe. Replace as necessary. - Press Restart. Removal condition: Successful Inlet calibration
Continued on next page

658

659

Cover removed or window opened

The analyzer entered HOLD because the cover was removed or the window to the electrode modules opened. An error occurred during Inlet calibration. Analyzer entered HOLD mode.

660

Inlet calibration error

11-33

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 661 Message Leak detected Interpretation A leak was detected. The analyzer entered HOLD mode. Operator action - Remove the analyzer covers and inspect the fluid transport system for leaks. - Check the inlet gasket, pump tubes and ensure electrodes are mounted correctly. Repair as necessary - Restart the analyzer. During restart the analyzer performs a leak detection. Removal condition: Leak test completed without any leaks detected. 662 Barometer Measured barometer out of range value is outside the measuring range: 60106.7 kPa. Leakage A leak was detected in the fluid transport system. Contact Radiometer service representative.

663

Remove the analyzer covers and inspect the fluid transport system for leaks. - Check the inlet gasket, pump tubes and ensure electrodes are mounted correctly. Repair as necessary. - Restart the analyzer. During restart the analyzer performs a leak detection. Removal condition: Leak test completed without any leaks detected.

664

Sample problem

A problem was detected during sample transport through the analyzer.

- Check the sample for clots. - Ensure sample volume matches selected sample mode. - Perform the Fluid Transport Troubleshooting procedure.
Continued on next page

11-34

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 671 Message Analyzing Unit service setup nondefault Interpretation An option in the service setup was changed from default setting. Operator action Contact Radiometer service representative. Removal condition: Options restored to default values. - Check whether Conditioning error (525) was reported during the last analyzer activity. If not, repeat measurement/ calibration. - Check Rinse Solution level in the container. - Perform Rinse. Removal condition: Rinse without Conditioning error. 678 Heater error One or more measuring - Ensure the ambient modules are overtemperature is between 15 heated. and 32 degrees Celsius. - Shut down the analyzer using the Temporary Shutdown function, then restart the analyzer. - Replace fan filter if dirty. - Shield analyzer from direct sunlight and other heat sources. - Contact Radiometer Service Representative. Removal condition: Analyzer temperature within required range. 679 Barometer error Measured parameter may be unreliable due to barometer error. Contact Radiometer service representative.

677

Unreliable

A measurement or calibration was attempted during "Conditioning error in El/Met electrode module".

Continued on next page

11-35

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 680 Message pH/BG module not active Interpretation The pH/BG module is not responding due to internal communication problem. Operator action - Shut down the analyzer using the Temporary Shutdown function, then restart it. - Contact Radiometer Service Representative. Removal condition: pH/BG module ready. 681 El/Met module not active The El/Met module is not responding due to internal communication problem, or software configuration does not match the analyzer type. - Shut down the analyzer using the Temporary Shutdown function, then restart it. - Contact Radiometer Service Representative. Removal condition: El/Met module ready, or software configured without El/Met module support. - Shut down the analyzer using the Temporary Shutdown function, then restart it. - Contact Radiometer Service Representative. Removal condition: OXI module ready, or software configured without OXI module support. 683 Inlet module not active The inlet module is not responding due to internal communication problem. - Shut down power using the Temporary Shutdown function, then restart the analyzer. - Contact Radiometer Service Representative. Removal condition: Inlet module ready.
Continued on next page

682

Oxi module not active

The Oxi module is not responding due to internal communication problem, or software configuration does not match the analyzer type.

11-36

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 688 Message ctHb/ceHb too low for OXI calculation Interpretation Operator action

ctHb < 1 mmol/L or Informative message. No ceHb < 0.75 mmol/L. If analyzer error detected. ctHb is too low FHHb, FO2Hb, FCOHb and FmetHb are not calculated. If ceHb = cHHb + cO2Hb is too low sO2 is not calculated. The gas pressure offset voltage measured by the transducer is out of range: 0.07-0.35 V. - Restart analyzer. - Contact Radiometer service representative. Removal condition: Gas pressure transducer offset within the required range. Contact your RADIANCE /IT engineer. - Check RADIANCE communication setup including TCP/IP address, port No. and password. - Check that RADIANCE is responding. - Check network connections. Removal condition: RADIANCE connection established or disabled.

689

Gas pressure offset voltage out of range.

692

ABL not The analyzer is not connected to connected to RADIANCE RADIANCE.

693

ABL not connected to RADIANCE - incorrect password.

The analyzer was refused connection to RADIANCE due to incorrect password.

Enter the correct password in the analyzer's RADIANCE Communication setup. Removal condition: RADIANCE connection established or disabled. No action required.

694

ABL The analyzer is connected to connected to RADIANCE RADIANCE.

Continued on next page

11-37

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 695 Message Interpretation Operator action No action required.

ABL discon- The analyzer was nected from disconnected from RADIANCE RADIANCE. ABL<> RADIANCE communication error Communication error between the analyzer and RADIANCE.

696

Contact Radiometer service representative.

697

Requested A scheduled QC AutoCheck ampoule was absent in QC ampoule the carousel. not present in carousel Performing AutoCheck on last ampoule in schedule instead of non-present AutoCheck QC measurement started due to calibration error Scheduled AutoCheck not run due to errors in last calibration QC expired If the scheduled QC ampoule was absent in the carousel, the analyzer performs QC measurement on the ampoule scheduled previously.

Refill AutoCheck carousel according to the packing list.

698

Refill AutoCheck carousel according to the packing list.

699

The analyzer was set up Check Calibration Status to perform AutoCheck and remedy any reported measurements in case of calibration errors. 1- or 2-point calibration errors. Last calibration contained an error, and the analyzer was set up to suspend AutoCheck measurements in case of calibration errors. QC measurement is overdue (corrective action "Lock analyzer" has been selected in Setup program Corrective Actions). Check Calibration Status and remedy calibration errors.

700

703

Perform a quality control measurement. Removal condition: No QC measurements are pending.

704

AutoCheck QC measurement is repeated

No action required. The scheduled QC measurement was not accepted; the measurement was repeated as requested in Setup program Corrective Actions.
Continued on next page

11-38

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 705 Message AutoCheck QC measurement is repeated twice Interpretation The scheduled QC measurement was not accepted; the measurement was repeated twice as requested in Setup program: Corrective Actions. Replacement is overdue by 10% (corrective action "Lock analyzer" was selected in Setup program: Corrective Actions). When the analyzer is locked, scheduled calibrations are performed, but no patient samples or QC measurements are allowed. Scheduled AutoCheck measurement was requested, but the carousel was empty. Operator action No action required.

707

Replacement overdue by 10 %. Analyzer locked

- Check Replacement Status and replace as required. - Unlock analyzer in Miscellaneous Setup program. Removal condition: No replacement pending.

708

Corrective action not possible due to empty AutoCheck carousel El/Met module not expected

Refill AutoCheck carousel.

709

Software configuration does not match analyzer type.

Contact Radiometer service representative. Removal condition: Matching software and hardware configuration. Contact Radiometer service representative. Removal condition: Matching software and hardware configuration. Informative message. No analyzer error was detected.

710

OXI module Software configuration not does not match expected. analyzer type.

711

Warning: Blood sample has a DysHb high high level of DysHb. FCOHb + FMetHb > 5%.

Continued on next page

11-39

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 712 Message FHbF measurement not possible Interpretation Composition of blood sample makes FHbF measurement too inaccurate, but OXI parameters are corrected for HbF. See explanation in the Reference Manual. Blood sample ctHb is so high that nearly no plasma is left to measure plasma bilirubin on. ctHb > 15.5mmol/L. 714, 740-743 AutoCheck mechanical error A mechanical error occurred in one of the AutoCheck submodules. - Check the AutoCheck module cover. Close if opened. - If error persists, contact Radiometer service representative. - Shut down the analyzer using the Temporary Shutdown function, then restart the analyzer. - Contact Radiometer Service Representative. 716 AutoCheck cover missing The cover was removed Mount the AutoCheck from the AutoCheck cover. module. Removal condition: AutoCheck cover is mounted correctly. The AutoCheck Repack AutoCheck carousel was not carousel according to the packed according to the packing list. schedule. Removal condition: AutoCheck carousel packed according to Packing list.
Continued on next page

Operator action Informative message. No analyzer error was detected.

713

ctBil measurement not possible

Informative message. No analyzer error was detected.

715

AutoCheck Communication error communica- between AutoCheck tion error submodules.

717

Carousel packing not optimal

11-40

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 718 Message AutoCheck carousel will need refilling soon Insufficient sample in pH/BG module Interpretation AutoCheck ampoule stock is low. Operator action Refill AutoCheck carousel when convenient. Removal condition: AutoCheck refilled. Air bubbles were detected in the sample with possible implications on measurement accuracy. All parameters are marked with "?". - Ensure that sample is free of air bubbles. - Ensure that sample volume matches the selected sample mode. - Check inlet tube. Replace if necessary. - Repeat the measurement. 720 Insufficient sample at El/Met module Air bubbles were detected in the sample with possible implications on measurement accuracy. All affected parameters are marked with "?". - Ensure that sample is free of air bubbles. - Ensure that sample volume matches the selected sample mode. - Check inlet tube. Replace if necessary.

719

721

Insufficient sample in Met II module

Air bubbles were detected in the sample - Repeat measurement. with possible implications on measurement accuracy. All affected parameters are marked with "?".

722

Sample error

Insufficient or - Check and remedy other inhomogeneous sample fluid transport errors. during measurement. - Repeat the measurement. pH/BG module was not - Check Cleaning solution level. Replace container as filled correctly with necessary. Cleaning Solution during Startup. - Perform Fluid transport troubleshooting procedure. pH/BG module was not - Check Cal 2 solution filled correctly with Cal level. Replace container as 2 Solution during necessary. Startup. - Perform Fluid transport troubleshooting procedure.
Continued on next page

723

No Cleaning Solution detected by inlet during Startup No Cal 2 Solution detected by inlet during Startup

724

11-41

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 725 Message No Cal 1 Solution detected by inlet during Startup No Rinse Solution detected by inlet during Startup No Cleaning Solution detected by inlet during Refill No Cal 2 Solution detected by inlet during Refill No Cal 1 Solution detected by inlet during Refill No Rinse Solution detected by inlet during Refill Met II upper liquid sensor did not receive Rinse from above Interpretation Operator action

pH/BG module was not - Check Cal 1 solution filled correctly with Cal level. Replace container as necessary. 1 Solution during Startup. - Perform Fluid transport troubleshooting procedure. pH/BG module was not - Check Rinse solution filled correctly with level. Replace container as Rinse Solution during necessary. Startup. - Perform Fluid transport troubleshooting procedure. pH/BG module was not - Check Cleaning solution level. Replace container as filled correctly with necessary. Cleaning Solution during Refill. - Perform Fluid transport troubleshooting procedure. pH/BG module was not - Check Cal 2 solution filled correctly with Cal level. Replace container as 2 Solution during necessary. Refill. - Perform Fluid transport troubleshooting procedure. pH/BG module was not - Check Cal 1 solution filled correctly with Cal level. Replace container as necessary. 1 Solution during Refill. - Perform Fluid transport troubleshooting procedure. pH/BG module was not - Check Rinse solution filled correctly with level. Replace container as Rinse Solution during necessary. Refill. - Perform Fluid transport troubleshooting procedure. Met II module was not filled correctly during Rinse program; automatic Refill is attempted. - Check the Rinse Solution for sufficient volume. Replace as required. - Perform Fluid transport troubleshooting procedure.
Continued on next page

726

727

728

729

730

731

11-42

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 734 Message DMS not connected to Analyzing Unit Interpretation Data management system establishes connection to the Analyzing Unit, or the connection is lost. Operator action - Wait a few minutes for the connection to establish. - Restart the analyzer. - If the error persists, contact Radiometer service representative. Read in the barcodes for all ampoules present in AutoCheck carousel use inserts. Removal condition: Barcodes read in for all ampoules present in the AutoCheck carousel. - Repeat measurement. - Contact Radiometer service representative if the problem persists. - Restart analyzer. - If the error persists, contact Radiometer service representative. Move archive files to another storage device. Removal condition: More than 100 Mb of free hard disk space. 746 Rinse Rinse Solution level is Solution below the defined low warning warning level. Replace container as required. Removal condition: Solution level above warning level.
Continued on next page

738

Ampoule(s) in AutoCheck carousel not installed

AutoCheck carousel contains an ampoule which was not selected in Control Solutions Setup program.

739

AutoCheck ampoule Solution's composition solutions do not match labeling. not identifiable AutoCheck Laser Setup Error AutoCheck barcode reader has failed.

744

745

Low hard disk space

Free hard disk space is below 100 Mb.

11-43

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 747 Message Interpretation Operator action Replace container as required. Removal condition: Solution level above warning level. Replace container as required. Removal condition: Solution level above warning level.

Cal 1 Cal 1 Solution level is solution low below the defined warning warning level.

748

Cal 2 Cal 2 Solution level is Solution below the defined low warning warning level.

749

Cleaning Cleaning Solution level Replace container as Solution is below the defined required. low warning warning level. Removal condition: Solution level above warning level. Solution The estimated solution No action required. projection level was adjusted by adjusted the operator. Message 751 is only found in the Activity Log to inform the user about activities that have taken place. The message is blank (empty) in the database, and when an activity occurs the actual status information is appended to it resulting in the logged 751-message. If the setting "Log All Measuring Activities" is enabled in Miscellaneous Setup, all wet section activities will be logged in the Activity Log as 751-messages. Service Electrode Updatings enabled Logging of sensor signals was enabled. This may be due to ongoing technical service activities on the analyzer. The condition does not affect analyzer functions. Contact Radiometer service representative if logging was not enabled on purpose. Removal condition: Electrode signal logging disabled by Radiometer service representative.
Continued on next page

750

751

753

11-44

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 763 Message AutoCheck temperature out of required range Interpretation The QC temperature was outside the range valid for the temperature correction algorithms. Operator action - Ensure the ambient temperature is between 15 and 32 degrees Celsius. - If the system has just performed a cold start, wait for error to disappear. - Replace fan filter if dirty. - Shield analyzer from direct sunlight and other heat sources. - Contact Radiometer Service Representative. Removal condition: AutoCheck temperature within the required range. 764 Liquid detected in pH/BG module during Expired Air sample Inlet flow impeded Liquid sample detected Do not use the Expired Air during aspiration in the mode for liquid samples. Expired Air mode.

765

Inlet flow is impeded - Perform Inlet by some foreign matter. troubleshooting procedure. - Perform the Protein Removal program. Removal condition: Inlet has no obstructions.
Continued on next page

11-45

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 766 Message ABL not connected to RADIANCE no RADIANCE connection license ABL not connected to RADIANCE ABL StatLink version too high ABL not connected to RADIANCE ABL StatLink version too low Failed to restore Custom Setup Interpretation The analyzer has been refused connection to RADIANCE because there is no connection licence available on RADIANCE. The analyzer has been refused connection to RADIANCE because the ABL StatLink version is higher than the RADIANCE StatLink version. The analyzer has been refused connection to RADIANCE because the ABL StatLink version is lower than the RADIANCE StatLink version. The setup could not be restored. Operator action Contact RADIANCE /IT engineer or Radiometer service representative. Removal condition: Connection to RADIANCE established. Contact RADIANCE /IT engineer or Radiometer service representative. Removal condition: RADIANCE connection established. Contact RADIANCE /IT engineer or Radiometer service representative. Removal condition: RADIANCE connection established. - Download the setup data from another floppy disk, hard disk or network. - Contact Radiometer service representative if the error persists. No action required.

767

768

770

771

Succeeded to Restoring of setup is completed. restore Custom Setup User Activity: Remote operator logged on with user:

772 773

User activity logged by No action required. operator. A remote operator has logged on the analyzer via NetOp No action required.

Continued on next page

11-46

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 774 Message Remote operator logged off with user Failed to restore Default Setup Succeeded to restore Default Setup Inlet gasket should be replaced soon Cleaning failed Interpretation An operator, remotely logged on via NetOp, has logged off or has been logged off by a local operator. Restoring analyzer setup to default values has failed. Restoring setup to default values is completed. Operator action No action required.

775

Contact Radiometer service representative.

776

No action required.

778

Inlet gasket is nearly Replace inlet gasket. worn out and should be replaced soon. The last Cleaning was aborted or interrupted. - Check solution level and replace container as necessary. - Check for and remedy System Messages. - Repeat Cleaning. Removal condition: Successful Cleaning.

779

780

RADIANCE communication enabled RADIANCE communication disabled

RADIANCE communication has been enabled as part of the RADIANCE Connection Setup. RADIANCE communication has been diabled as part of the RADIANCE Connection Setup.

Informative message. No action required.

781

Informative message. No action required.

Continued on next page

11-47

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 782 Message RADIANCE output queue cleared Automatic backup started Automatic archiving started Automatic archiving completed Export of data logs started Hypochlorite detected remembrane Glu, Lac, Cl and both Crea electrodes Interpretation Operator action

The output queue was Informative message. No cleared in RADIANCE action required. Connection Setup.

783

Automatic backup (selected in Disk Functions Setup) has started. Automatic archiving (selected in Disk Functions Setup) has started.

Informative message. No action required.

785

Informative message. No action required.

786

Informative message. No Automatic archiving action required. (selected in Disk Functions Setup) completed successfully. Export of data logs was Informative message. No started by the user. action required. Measurement on a substance resembling hypochlorite has been performed. Any installed Glu, Lac, Cl and Crea electrodes have been damaged and must be remembraned before blood measurements can be performed. As a precaution all parameters are marked with Calibration Expired.
Continued on next page

787

788

- Remembrane any installed Glu, Lac, Cl and Crea (A and B) electrodes. - Perform a 2-point calibration. Removal condition: Successful 2-point calibration.

11-48

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 790 Message Adaptive measuring mode applied Interpretation Due to insufficient or inhomogeneous sample, the analyzer has automatically chosen an adaptive sample handling in order to report as many of the desired parameters as possible. Operator action If missing parameters are needed, repeat the measurement with sufficient sample free of clots and air bubbles.

791

Parameter Due to insufficient or could not be inhomogenous sample, measured the parameter in question could not be measured. Liquid sensor calibration error Inlet lower liquid sensor failed to calibrate.

If missing parameters are needed, repeat the measurement with sufficient sample free of clots and air bubbles. - Repeat liquid sensor calibration. - Perform the Fluid Transport Troubleshooting procedure. - Contact Radiometer service representative.

792

793

pH/BG pump asymmetry error

The pH/BG module pump failed to calibrate pump asymmetry. The same calibration values will be used for both directions until a successful calibration of pump asymmetry is performed. The El/Met module pump failed to calibrate pump asymmetry. The same calibration values will be used for both directions until a successful calibration of pump asymmetry is performed.

- Perform the Pump troubleshooting procedure. - Contact Radiometer service representative.

794

El/Met pump asymmetry error

- Perform the Pump troubleshooting procedure. - Contact Radiometer service representative.

Continued on next page

11-49

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 795 Message Interpretation Operator action - Perform the Pump troubleshooting procedure. - Contact Radiometer service representative.

Met II pump The Met II module asymmetry pump failed to calibrate error pump asymmetry. The same calibration values will be used for both directions until a successful calibration of pump asymmetry has been performed. Solution pump asymmetry error The Solution module pump failed to calibrate pump asymmetry. The same calibration values will be used for both directions until a successful calibration of pump asymmetry is performed.

796

- Perform the Pump Troubleshooting procedure. - Contact Radiometer service representative.

797

Waste pump The Waste pump in the asymmetry Solution module failed error to calibrate pump asymmetry. The same calibration values will be used for both directions until a successful calibration of pump asymmetry is performed. User logged User logged on on successfully. User logged User logged off. off Logon attempt failed Hard disk S.M.A.R.T. threshold(s) exceeded

- Perform the Pump Troubleshooting procedure. - Contact Radiometer service representative.

798 799 800

No action required. Informative message. Informative message. No action required.

User tried to log on but Provide a valid password to did not provide a valid log on. password. The analyzer's hard disk drive has given a warning. This may mean imminent hard disk drive failure. Contact Radiometer service representative.

801

Continued on next page

11-50

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 805 Message Adaptive measuring mode applied (decay to pH/BG/Met/ Oxi) Adaptive measuring mode applied (decay to all micro mode) Unauthorized software Calibration drift not available Interpretation Due to insufficient or inhomogeneous sample, the analyzer has automatically chosen an adaptive sample handling in order to report as many of the desired parameters as possible. Operator action - If missing parameters are needed, repeat the measurement with sufficient sample free from clots and air bubbles.

806

807

Software option key is not installed. Analyzer cannot be used. A calibration drift value could not be calculated, since this is the first calibration result after calibration drift monitoring was started. This message may appear on the first calibration result after the analyzer has been turned on or after the parameter in question has been enabled. A drift value will be calculated for the next calibration result.

Contact Radiometer service representative. No action required. The message is informative.

808

809

Service Electrode Updatings disabled

The message is used in the Activity Log to indicate that Service Electrode Updatings have been disabled.

No action required. The message is for information only.

Continued on next page

11-51

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 810-827 829, 830 Message Parameter locked (pH, pCO2, pO2, cK+, etc.) Interpretation Parameter has been locked by a RADIANCE operator, as reflected in the Activity Log. When a parameter is locked, presumably due to problems with QC, the parameter is repressed in patient results. 831-848, 850, 851 Parameter unlocked (pH, pCO2, pO2, cK+, etc.) Radiance: LIS/HIS: Missing Communication Option key No leading air segment in inlet's lower liquid sensor within time frame The message is used in the Activity Log to indicate that a previously locked parameter has been unlocked. Message from RADIANCE Operator action Await corrective actions initiated by the RADIANCE operator. Removal condition: Determined by the RADIANCE operator.

Informative message. No action required.

852 853

Informative message. No action required.

No valid option key for Contact Radiometer service representative. LIS/HIS communication is installed on the analyzer. Air segment in front of sample is too small or missing during sample aspiration, possibly due to obstructions in inlet or problems with liquid sensor. Measurement is aborted. - Check inlet for obstructions and clean as required. - Perform liquid sensor adjustment. - Perform the Fluid Transport Troubleshooting procedure. Informative message. No analyzer error was detected.
Continued on next page

854

855

Base Excess Base Excess exceeds out of range +/- 30 mmol/L range

11-52

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 862 Message pH-only sample handling adapted to 35uL range pH-only sample handling adapted to 55uL range pH/BG sample handling adapted to 35uL range (decay to pH only) Interpretation To counteract insufficient or inhomogeneous sample, the analyzer has automatically adjusted the sample handling to the sample volume. Operator action None.

864

865

Due to insufficient or inhomogeneous sample, the analyzer has automatically chosen an adaptive measuring mode in order to report as many of the desired parameters as possible. To counteract insufficient or inhomogeneous sample, the analyzer has automatically adjusted the sample handling to the sample volume.

If missing parameters are needed, repeat the measurement with sufficient sample free from clots and air bubbles.

867

pH/BG sample handling adapted to 55uL range pH/BG/Oxi sample handling adapted to 55uL range

None.

868

Continued on next page

11-53

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 869 Message Adaptive sample handling applied (decay to pH/Oxi without Bil) Adaptive sample handling applied (decay to pH/Oxi) Adaptive sample handling applied (decay to pH/BG/Oxi) Adaptive measuring mode applied (pH/BG/Oxi [2]) Adaptive measuring mode applied (pH/BG/ Met/Oxi) Adaptive measuring mode applied (decay to all micro mode) Sample aged The specified limit for sample age has been exceeded. Draw and analyze new sample. Interpretation Operator action

870

871

872

Due to insufficient or inhomogeneous sample, the analyzer has automatically chosen an adaptive measuring mode in order to report as many of the desired parameters as possible.

If missing parameters are needed, repeat the measurement with sufficient sample free from clots and air bubbles.

873

874

875

Continued on next page

11-54

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 876 Message FLEXQ positioning warning Interpretation A problem was detected during positioning of the sampler tray Operator action The message is informative and intended for service use only Removal condition: No FLEXQ positioning problems observed for 24 hours. - Remove all the samplers from the sampler tray (the sampler registration is lost). - If the sampler tray is blocking the syringe inlet or AutoCheck module, try to push it gently to its home position. - If the syringe inlet is not blocked, scan the sampler barcode, using the analyzers barcode reader and perform a manual measurement. - To get the FLEXQ operational again, restart the analyzer, using Temporary Shutdown and power off/on. - If the error persists, contact Radiometer service representative. - Removal condition: Successful FLEXQ sampler tray calibration.

877

FLEXQ positioning error

A sampler in the sampler tray could not be aligned with the syringe inlet; aspiration of the sample was not possible.

Continued on next page

11-55

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 878 Message No Rinse solution during calibration No Cal1 solution during calibration No Cal2 solution during calibration LIS/HIS: Missing CIC Communication Option Key Interpretation The analyzer was not able to retrieve Rinse solution during calibration. The analyzer was unable to retrieve Cal1 solution during calibration. The analyzer was unable to retrieve Cal2 solution during calibration. Operator action Check level in Rinse Solution bottle. Replace bottle as necessary. Check level in Cal1 solution bottle. Replace bottle as necessary. Check level in Cal2 solution bottle. Replace bottle as necessary.

879

880

881

Contact Radiometer service No valid Option Key for CIC communication representative. is installed on the analyzer. - Lift the sampler tray cover and check for obstructions in the mixer tray area. - Remove any foreign objects. - Remount the sampler tray cover. Removal condition: Alignment operation performed without slip.

882

Mixer tray A small positioning slip detected problem of the mixer tray has been detected and corrected.

883

Mixer error

The sample may not have been adequately mixed.

- Lift the sampler tray cover and check for obstructions in the mixer tray area. - Remove any foreign objects. - Remount the sampler tray cover.
Continued on next page

11-56

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 884 Message Sampler tray adjustment needed Interpretation It was not possible to position one or more samplers directly below the syringe inlet. The affected sampler tray positions in the sampler tray must be adjusted. 885 Cyclic QC schedule reset from RADIANCE Adaptive measuring mode disabled Operator action Contact your service engineer to perform the Sampler Tray Adjustment procedure. Removal condition: All sampler tray positions are adjusted.

The cyclic QC schedule No action required. Informative message only. has been reset and all related reminders have been removed as a result of a RADIANCE command. Due to insufficient or inhomogeneous sample, the analyzer would normally automatically have chosen an adaptive measuring mode but this option has been disabled. Measurement is aborted. Repeat the measurement with sufficient sample free from clots and air bubbles.

887

888-889, 891

Pump calibration error

The pump calibration failed for one or more pumps. The previous pump calibration is used. A successful pump calibration should be attempted as soon as possible.

- Repeat Pump calibration. - Perform the Pump Troubleshooting procedure. Removal condition: Successful pump calibration.

Continued on next page

11-57

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 890 Message Error measuring Inlet's liquid sensor to Oxi's liquid sensor volume Interpretation During pump calibration, measurement of internal volume part failed. Causes the Fluid Transport Control program to abort. During pump calibration measurement of internal volume part failed. Causes the Fluid Transport Control program to abort. Operator action

- Check Rinse Solution for sufficient volume. Replace as necessary. - Perform Pump Calibration. - Perform the Fluid Transport Troubleshooting procedure. Removal condition: Successful pump calibration.

892

Inlet not filled properly during pump calibration using reagent as backup

893

Met II module was not Sample in filled correctly during Met II's upper liquid sample aspiration. sensor detected prematurely during aspiration

- Ensure sample is free of clots. - Repeat measurement. - Perform the Fluid Transport Troubleshooting procedure.

Continued on next page

11-58

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 894 Message Adaptive measuring mode applied (pH/BG/ Met/Oxi [2]) Interpretation Operator action

895

Due to insufficient or inhomogeneous sample, the analyzer has automatically Adaptive chosen an adaptive measuring measuring mode in mode order to report as many applied of the desired (pH/BG/ Met/El/Oxi) parameters as possible. Adaptive measuring mode applied (pH/BG/ Met/El/Oxi [2]) Adaptive measuring mode applied (pH/BG/ Met/El/Oxi [3]) Adaptive measuring mode applied (pH/BG/ Met/El/Crea /Oxi) Adaptive measuring mode applied (pH/BG/ Met/El/ Crea/Oxi [2])

If missing parameters are needed, repeat the measurement with sufficient sample free from clots and air bubbles.

896

897

898

899

Continued on next page

11-59

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 900 Message Adaptive measuring mode applied (pH/BG/ Met/El/ Crea/Oxi [3]) Adaptive measuring mode applied (pH/BG/ Met/El/ Crea/Oxi [4]) Interpretation Due to insufficient or inhomogeneous sample, the analyzer has automatically chosen an adaptive measuring mode in order to report as many of the desired parameters as possible. Due to insufficient or inhomogeneous sample, the analyzer has automatically chosen an adaptive measuring mode in order to report as many of the desired parameters as possible. Operator action If missing parameters are needed, repeat the measurement with sufficient sample free from clots and air bubbles.

901

If missing parameters are needed, repeat the measurement with sufficient sample free from clots and air bubbles.

909

Met II not Met II module did not filled within fill properly with time limit calibration solution. Causes the calibration program to abort.

- Check the solution containers for adequate volume. Replace as required. - Perform the Fluid Transport Troubleshooting procedure. Removal condition: Successful calibration.

919

Inlet not in CONE3 position before timeout prior to wash with cleaning

Inlet was not in proper position within required time frame. The activity is aborted.

- Remove the inlet gasket and inspect the condition of inlet probe and gasket. Replace as necessary. Install all parts. - Restart the analyzer. - Contact Radiometer Service Representative. Removal condition: Successful inlet calibration.
Continued on next page

11-60

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 920-922 Message Interpretation Operator action - Check the solution containers for adequate volume. Replace as required. - Perform the Fluid Transport Troubleshooting procedure. Removal condition: Successful Rinse 923 Inlet not refilled properly during bidirectional pump calibration The inlet module was not filled properly during bi-directional pump calibration. - Check Rinse Solution for sufficient volume. Replace as necessary. - Perform Pump Calibration. - Perform the Fluid Transport Troubleshooting procedure. Removal condition: Successful pump calibration. 924 Inlet's liquid sensor to Oxi's liquid sensor volume could not be confirmed 1 During pump calibration, measurement of internal volume part failed. Causes the Fluid Transport Control program to abort. - Check Rinse Solution for sufficient volume. Replace as necessary. - Perform Pump Calibration. - Perform the Fluid Transport Troubleshooting procedure. Removal condition: Successful pump calibration. 925 Inlet's liquid sensor to Oxi's liquid sensor volume could not be confirmed 2 During pump calibration, measurement of internal volume part failed. Causes the Fluid Transport Control program to abort. - Check Rinse Solution for sufficient volume. Replace as necessary. - Perform Pump Calibration. - Perform the Fluid Transport Troubleshooting procedure Removal condition:Successful pump calibration.
Continued on next page

Module not Module did not fill filled within properly with cleaning time limit in solution. wash with cleaning

11-61

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 926 Message Inlet's liquid sensor to Oxi's liquid sensor volume could not be confirmed 3 Interpretation During pump calibration, measurement of internal volume part failed. Operator action - Check Rinse Solution for sufficient volume. Replace as necessary. - Perform Pump Calibration. - Perform the Fluid Transport Troubleshooting procedure. Removal condition: Successful pump calibration.

927

Causes the Fluid Transport Control Inlet's liquid program to abort. sensor to Oxi's liquid sensor volume could not be confirmed 4 Unable to fill Inlet with continuous rinse during pump calibration TIPCAP conflict detected 1 TIPCAP conflict detected 2 The detection of the TIPCAP failed. The activity is aborted.

928

929

930

- Remove the inlet flaps and the inlet gasket. Inspect the condition of the inlet probe and gasket. Look for obstructions that may hinder probe movement. Replace as necessary. Install all parts. - Perform the Inlet Probe Troubleshooting Procedure. - Restart the analyzer. - Contact Radiometer Service Representative. Removal condition: Successful inlet calibration.
Continued on next page

11-62

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 931 Message No sample in Met II's lower liquid sensor within time limit No sample in Met II's upper liquid sensor within time limit Inlet not in CONE3 position before timeout prior to wash with MetCal1 Inlet probe was not in proper position within the required time limit. Interpretation The sample did not fill Met II module properly. The activity is aborted. Operator action - Check the sample for clots. - Ensure that sample volume matches selected sample mode. - Perform the Fluid Transport Troubleshooting procedure.

932

934

- Remove the inlet gasket and inspect the condition of inlet probe and gasket. Replace as necessary. Install all parts. - Restart analyzer. - Contact Radiometer Service Representative. Removal condition: Successful inlet calibration.

935

Cal1 Solution Crea will expire soon

The additive used in the Calibration Solution 1 to calibrate Crea will expire soon. This is a warning - the solution lifetime is 14 days.

Replace the solution before the lifetime is actually exceeded.

936

Cal2 Solution Crea will expire soon

The additive used in the Calibration Solution 2 to calibrate Crea will expire soon. This is a warning - the solution lifetime is 14 days.

Replace the solution before the lifetime is actually exceeded.

Continued on next page

11-63

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 937 Message Cleaning Solution Crea will expire soon Interpretation The additive used in the Cleaning Solution to measure the High Crea Check will expire soon. This is a warning - the solution lifetime is 14 days. 938 Cal1 Solution Crea is expired The additive used in the Calibration Solution 1 to calibrate Crea is expired. The solution is more than 14 days old. It is no longer possible to calibrate Crea. 939 Cal2 Solution Crea is expired The additive used in the Calibration Solution 2 to calibrate Crea is expired. The solution is more than 14 days old. It is no longer possible to calibrate Crea. 940 Cleaning Solution Crea is expired The additive used in the Cleaning Solution to measure the High Crea Check is expired. The solution is more than 14 days old. It is no longer possible to perform the High Crea Check.
Continued on next page

Operator action Replace the solution before the lifetime is actually exceeded.

Replace the Calibration Solution 1 bottle remember to add the additive. Removal condition: Fresh bottle installed.

Replace the Calibration Solution 2 bottle remember to add the additive. Removal condition: Fresh bottle installed.

Replace the Cleaning Solution bottle - remember to add the additive. Removal condition: Fresh bottle installed.

11-64

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 941 942 Message Cal expired (Crea) Cal expired (Creatine) Interpretation Too long time passed since the last successful calibration of the parameter. Parameter measurement values are reported as ".....". Software configuration does not match analyzer type. Operator action Perform a 2-point calibration. Removal condition: Successful 2-point calibration.

944

Met II module not expected

945

Met II module not active

Contact Radiometer Service Representative. Removal condition: Matching software and hardware configuration. - Shut down the analyzer The Met II module is using the Temporary not responding due to internal communication Shutdown function, then restart it. problems, or software configuration does not - Contact Radiometer match the analyzer Service Representative. type. Removal condition: Met II module ready, or software configured without Met II module support. Rinse Solution was not detected in Met II electrode module. The Crea results will be reported as "....." in the next measurement. - Check Rinse Solution for adequate volume, proper installation or obstructions in bottle opening. - Replace as necessary. - Perform the Fluid Transport Troubleshooting procedure. Removal condition: Succesful Rinse.
Continued on next page

946

Conditioning error in Met II electrode module

11-65

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 951 Message Creatine is below the reportable range Interpretation Creatine is below the reportable range. See chapter 14: Specifications in the Operator's Manual. Operator action - Check for and remedy other errors related to the result, system messages or calibration status. - Perform QC. If the QC result is accepted, the blood sample may be suspected. - Perform measurement on new blood sample. - Perform the Electrode Troubleshooting procedure. 952 Creatine is above the reportable range Creatine is above the reportable range. See chapter 14: Specifications in the Operator's Manual. - Check for and remedy other errors related to the result, system messages or calibration status. - Perform QC. If the QC result is accepted, the blood sample may be suspected. - Perform measurement on new blood sample. - Perform the Electrode Troubleshooting procedure. 953 Measured QC value above the control range (Crea A) Measured QC value below the control range (Crea A) Creatine value is above - Verify procedure and the control range. repeat measurement. Error on the Creatine measurement is possible. - Refer to Quality Control Systems Reference Manual.

954

Creatine value is below - Verify procedure and the control range. repeat measurement. Error on the Creatine measurement is possible. - Refer to the Quality Control Systems Reference Manual.
Continued on next page

11-66

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 955 Message Measured QC value lower than statistical range (Crea A) Measured QC value higher than statistical range (Crea A) Solution lifetime exceeded Interpretation The parameter value is below the lower limit of the user-defined statistical range. Measurement is not included in the statistics. The parameter value is above the upper limit of the user-defined statistical range. Measurement is not included in the statistics. A solution is more than 14 days old and cannot be used for Crea calibration or High Crea Check. Calibration Solutions: - Crea is no longer reported during measurements and is marked red on the parameter bar. Cleaning Solution: - It is no longer possible to perform a High Crea Check.
Continued on next page

Operator action - Verify procedure and repeat the measurement. - Refer to the Quality Control Systems Reference Manual for details on the evaluation of the results. - Verify procedure and repeat the measurement. - Refer to the Quality Control Systems Reference Manual for details on the evaluation of the results. Replace required solution bottle - remember to add the additive. Removal condition: Fresh bottle installed.

956

958

11-67

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 959 Message Crea Calibration Check failed. Possible mix-up. Interpretation During a 2-point calibration, the Crea A electrode detects too high a signal in Cal 1. This situation indicates a possible mix-up of either the electrodes or the Calibration Solutions. Operator action - Check that the Crea A and Crea B electrodes are placed in the correct measuring chambers. - Check that the Calibration Solutions are placed in the correct positions on the analyzer. - Replace Calibration Solutions and perform a 2point calibration. - Remembrane both Crea A and Crea B electrodes. Removal condition: Successful 2-point calibration. 960 Measurement unstable (Crea A) An electrode response fault occurred during measurement on the Crea A electrode. - Check for and remedy other errors in System Message or Calibration Status. - Repeat the measurement. - Perform any pending replacements, including electrodes. - Ensure the electrode is properly installed. - Ensure the analyzer is properly grounded. - Perform QC. If the QC result value is within the control range, the blood sample may be suspected. Continue troubleshooting. - Perform the Electrode Troubleshooting procedure.
Continued on next page

11-68

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 961 Message Total concentration outside measuring range Interpretation The total concentration of Crea + Creatine is above the measuring range and the analyzer is not able to measure Crea within the specifications. See chapter 14: Specifications in the Operator's Manual. Operator action - Check for and remedy other errors related to the result, system messages or calibration status. - Perform a High Crea Check. If the High Crea Check result is accepted, the blood sample may be suspected. - Perform measurement on new blood sample. - Perform the Electrode Troubleshooting procedure. 962 Ambient temperature not defined Ambient temperature has not been defined by the user in Environment Setup. Enter ambient temperature in Environment Setup - see Operator's Manual chapter 3.

Until the ambient Removal condition: temperature has been Ambient temperature defined, a default value entered. of 25 C is used during High Crea Check and the internal Crea calibration calculations. Please maintain an ambient temperature between 18 C and 29 C 963, 964 Leak currents detected Leak currents are detected during calibration and may distort the measuring results. The affected parameters are marked with "?". The Cleaning additive is too old and the Cleaning Solution cannot be used anymore. Contact Radiometer service representative.

965

Cleaning Additive is expired

Replace Cleaning Solution bottle remember to add the additive.

Continued on next page

11-69

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


Messages at manager level (continued) No. 966 Message Cleaning Additive will expire soon Creatine high. Unable to perform Crea measurement Interpretation The Cleaning additive will expire soon. Operator action Replace the solution before the lifetime is actually exceeded. - Check for and remedy other errors related to the result, system messages or calibration status. - Perform a High Crea Check. If the High Crea Check result is accepted, the blood sample may be suspected. - Perform measurement on a new blood sample. - Perform the Electrode Troubleshooting procedure Contact Radiometer Service Representative. Removal condition: Matching software and hardware configuration.

967

The Creatine concentration is too high compared to the Crea concentration and the analyzer is not able to measure Crea within the specifications. See chapter 14: Specifications in the Operator's Manual

969

FLEXQ module not expected

Software configuration does not match analyzer type.

970

Rilibk violation: Value above upper limit Rilibk violation: Value belov lower limit
Continued on next page

971

11-70

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 975 Message Interpretation Operator action Check level in Cleaning solution bottle. Replace bottle as necessary.

No Cleaning The analyzer was solution unable to retrieve Cleaning solution during calibration or High Crea Check Error reading QA Portal configuration file The file QAPortalSetup.xml was not found or was corrupted.

977

Check if the file QAPortalSetup.xml is missing in the Abl700 directory (root) and that the file is not empty or corrupted.

978

pH of Rinse Solution outside range of pH/BG Calibration 1 point pH/BG Calibration upgraded due to large pH drift Discrepancy The pO2 and sO2 between pO2 deviates so much from the standard Oxygen and sO2 Dissociation Curve that an error is assumed. Causes may be: If air buble is suspected: - Make sure, that all air is expelled from the sample prior to mixing. - This includes 'tapping' the sampler to loosen any air adhering to the walls af - The pO2 electrode may have measured the the sampler. - Repeat sampling and pO2 of a bubble of measurement. atmospheric air. - The patient may be treated with Oxygen enriched air yet still have a low oxygenation. - Your ABL is working at high altitude (low atmospheric pressure).
Continued on next page

979

1010

If the problem occurs repeatedly: - Contact your service technician.

11-71

11. Troubleshooting

ABL800 FLEX Operator's Manual

Analyzer messages, Continued


T T

Messages at manager level (continued)


T

No. 1012

Message Error writing to QA Portal configuration file Volume Calibration Failed (Wet Section internal Volume Measurement) Volume measurement failed: pH/BG module to Oxi's LS Volume measurement failed: El/Met module to Oxi's LS Volume measurement failed: Met II module to Oxi's LS Volume measurement below limit: pH/BG module to Oxi's LS

Interpretation

Operator action

1013

1014

1015

1016

1017

Continued on next page

11-72

ABL800 FLEX Operator's Manual

11. Troubleshooting

Analyzer messages, Continued


Messages at manager level (continued) No. 1018 Message Volume measurement below limit: El/Met module to Oxi's LS Volume measurement below limit: Met II module to Oxi's LS Volume measurement above limit: pH/BG module to Oxi's LS Volume measurement above limit: El/Met module to Oxi's LS Volume measurement above limit: Met II module to Oxi's LS Software Upgrade Initiated Upgraded From Interpretation Operator action

1019

1020

1021

1022

1023

1024

11-73

11. Troubleshooting

ABL800 FLEX Operator's Manual

Fluid transport troubleshooting procedure


Using the troubleshooting procedure NOTICES: The actions are listed in order of priority. It is recommended to perform a step in the procedure below and verify whether or not the error has been remedied. If the error persists, proceed to the next step of the procedure. For specific instructions on the replacement of components, refer to the relevant sections of chapter 7. Record all replacement actions. Use the table of solutions below to identify and correct errors in the fluid transport system. Step 1. Action Remove the analyzer cover to the solutions and call the Rinse program. Observe the flow of solution through the fluid transport system as this may help to identify the problem see Fluid transport system description for details. And check the fluid transport system for clots. 2. Remove the cover to the measuring modules and check the following: The electrodes are properly mounted in the measuring chamber and the electrode connector is latched shut. The pump tubings and connectors have no leaks or damage. The pump rotor seatings. The inlet module has no leaks or damaged parts. 3. 4. 5. 6. 7. 8. 9. Make any necessary repairs. Mount all the parts and press Restart. Call the Liquid Sensor Adjustment program (in the Auxiliary Programs). Call the Pump Calibration program (in the Auxiliary Programs). Clean the electrodes and the measuring chambers, using distilled water and a cotton swab. Check the inlet probe and gasket unit for obstructions see the Inlet probe troubleshooting procedure further in this chapter. If the error persists, contact a service representative.

Fluid transport troubleshooting procedure

11-74

ABL800 FLEX Operator's Manual

11. Troubleshooting

Inlet probe troubleshooting procedure


Causes of error The following conditions are the causes of error: Condition Inlet positioning error Explanation This condition occurs when the inlet probe cannot obtain the proper position within a defined time limit. This condition occurs when the calibration of the inlet fails, most likely caused by a problem with the inlet components.

Inlet calibration error

Procedure

To correct the error, do the following: Step 1. 2. 3. Action Remove the analyzer cover and the inlet flaps. Check that the inlet probe and gasket are mounted properly. Remove the inlet gasket and check for wear or damage. Replace as required. If the inlet probe has not retracted, move it in by pressing the probe attachment back. Note that you will feel resistance while moving the probe in. 4. 5. Check the condition of the inlet probe. If it appears bent or damaged, replace it. Mount all parts and press Restart. The inlet calibration is performed automatically during restart. The error has been corrected when the analyzer enters the Ready mode. If the error persists, the analyzer remains in the Hold mode. Contact a service representative.

11-75

11. Troubleshooting

ABL800 FLEX Operator's Manual

Inlet troubleshooting procedure


Indication of obstructions in the inlet The following conditions may indicate an obstruction somewhere in the inlet system. No sample flow through the fluid transport system. Solution dripping from the inlet. Excessive suction noise coming from inlet when waste pump is activated. Fluid flowing upwards in the electrode measuring modules during the wrong stage of a rinse. Erratic flow of fluid through all three modules and/or excessive air bubbles flowing through the system. Lack of fluid movement through the small canals off the waste pump tubes, just above the pump rotor. If one or more of the above conditions are present, perform the following Inlet troubleshooting procedure. To troubleshoot possible obstructions in the inlet, do the following: Step 1. Action Remove the inlet gasket unit and inspect its condition. If it appears worn or damaged, replace it. Rinse the gasket and its openings by filling a syringe with distilled water and injecting the water through both the syringe and capillary openings. 2. 3. Remove the inlet probe and tube. Fill a syringe with distilled water. Take the end of the inlet tube, press it firmly against the tip of the syringe and inject the distilled water. The water should stream freely out of the opening at the end of the probe. If not, replace the inlet probe and tube. 4. Install all parts and press Restart.

Procedure

For detailed instructions on removal and/or replacement of parts, refer to chapter 7: Replacements. NOTICE: The spillage of liquid from the inlet may be caused by damaged waste pump tubes, improper mounting of the inlet gasket unit, clots, etc, so all these factors should be inspected, then corrected or replaced.

11-76

ABL800 FLEX Operator's Manual

11. Troubleshooting

Leak troubleshooting procedure


Leak error This condition occurs when a leak detection program has failed, possibly indicating a leak in the fluid transport system. To correct the error, do the following: Step 1. 2. 3. Action Remove the analyzer covers and inspect the fluid transport system for poor tube connections or worn tubes. Repair as required. Ensure that electrodes are properly mounted in the measuring chamber and the electrode connector is latched shut. Mount all parts and press Restart. The leak detection is performed automatically during the restart. The error is corrected when the analyzer enters the Ready mode. If the error persists, the analyzer remains in the Hold mode. Contact a service representative.

Procedure

11-77

11. Troubleshooting

ABL800 FLEX Operator's Manual

Electrode troubleshooting procedures


Using the procedures pH, cK+, cNa+, cCa2+, and cCl troubleshooting procedure The actions are listed in order of priority. It is recommended to perform each action and verify whether or not the error has been remedied. If not, proceed to the next step. To troubleshoot pH, cK+, cNa+, cCa2+, and cCl electrodes, do the following: Step 1. Action The pH electrode only: Clean the electrode, using a cotton swab moistened with Cleaning Solution. 2. Check the measuring chamber for clots: clean the measuring chamber with a cotton stick moistened with distilled water. Make sure that no cotton fibers are left in the measuring chamber. Replace the electrode membrane (not the pH electrode). Call a 2-point calibration. If the error persists, replace the electrode.

3. 4.

If an error has occurred on two or more parameters, troubleshoot the reference electrode. Reference electrode troubleshooting procedure To troubleshoot the reference electrode, do the following: Step 1. Action Remove the electrode and check the measuring chamber for clots: clean the measuring chamber with a cotton stick moistened with distilled water. Make sure that no cotton fibers are left in the measuring chamber. Replace the electrode membrane. Call a 2-point calibration. If the error persists, replace the electrode.

2. 3. pCO2 and pO2 troubleshooting procedure

To troubleshoot the pCO2 and pO2 electrodes, do the following: Step 1. Action Remove the electrode and check for air bubble(s) trapped between the electrode tip and the membrane. If any air bubbles exist, remove them by pressing the tabs on the electrode jacket and moving the electrode slightly up and down until they dissipate. 2. Replace the electrode membrane. The pO2 electrode: brush the electrode tip with the supplied brush before remembraning the electrode. 3. Call a 2-point calibration. If the error persists, replace the electrode.
Continued on next page

11-78

ABL800 FLEX Operator's Manual

11. Troubleshooting

Electrode troubleshooting procedures, Continued


cGlu and cLac troubleshooting procedure To troubleshoot the cGlu and cLac electrodes, do the following:

Step 1.

Action Remove the electrode and check the measuring chamber for clots; clean the measuring chamber with a cotton stick moistened with distilled water. Make sure that no cotton fibers are left in the measuring chamber. Replace the electrode membrane. Call a 1-point calibration. If the error persists, replace the electrode.

2. 3.

cCrea electrodes To troubleshoot the cCrea-A and cCrea-B electrodes, do the following: troubleshooting procedure Step Action 1. Remove the electrodes and check the measuring chamber for clots; clean the measuring chamber with a cotton stick moistened with distilled water. Make sure that no cotton fibers are left in the measuring chamber. Replace the membranes of both electrodes. Call a 2-point calibration. If the error persists, replace the electrodes.

2. 3.

NOTICE:

Refer to chapter 7 for specific instructions on replacing electrodes and membranes. Remember to record all replacements.

11-79

11. Troubleshooting

ABL800 FLEX Operator's Manual

Pump troubleshooting procedure


Procedure To troubleshoot pumps, do the following: Step 1. 2. 3. 4. Action Check the Rinse Solution container for sufficient volume and proper mounting. Replace as required. Inspect the pump tube(s) for leaks, wear and proper seating. Replace as required. Call the Liquid Sensor Adjustment program. Call the Pump Calibration program. This program is performed automatically after any replacements made in step 2 above. 5. If the error persists, contact a service representative.

11-80

ABL800 FLEX Operator's Manual

11. Troubleshooting

Fluid transport system description


Diagram of fluid The description of the fluid transport system and a detailed explanation of the rinse transport system process aim to help troubleshooting. The fluid transport system of the analyzer is shown below. The available moduls depends on the analyzer version.
pH/BG module Gas2 Gas1 Solution valve Membrane valve Liquid sensor Rinse solution Gas/air Capillary opening Sample path Syringe opening Cuvette Inlet probe Pump El/Met module Met II module Oximetry module

Waste Rinse

Cal1

Cal2

Cleaning

Diagram of the fluidic module

A detailed illustration of the fluidic module is given below.

V8

V5 V4 P1 V6 V7 P2

V3

V2 V1 Cal 2 Cleaning

Optional

Cal 1

Continued on next page

11-81

11. Troubleshooting

ABL800 FLEX Operator's Manual

Fluid transport system description, Continued


Description of the fluidic module The function of the fluidic module is to transport solutions from the containers to the measuring modules, and to transport samples from the inlet and measuring modules to the waste container. The right fluidic module consists of a transparent fluid unit including all flow lines and four solution connectors. Eight membrane valves and two pumps select the flow path and control the flow. The functions of the principal parts are given below. Part V1 V2 V3 V4 V5 V6 V7 V8 P1 Function Pulled to enable supply of cleaning solution to the wet section. Pulled to enable supply of Cal 2 solution to the wet section. Pulled to enable supply of Cal 1 solution to the wet section. Not used. Not used. Pulled to enable supply of rinse solution to the wet section. Pulled to enable supply of air to the wet section. Pulled to enable flow to the waste container (used when the sample is transported into the measuring modules). Waste pump. Includes two tubes with larger inner diameter than the other pump tubes. It aspirates from the inlet and transports solutions and sample to the waste container. The pump rotation is always counterclockwise. Solution pump. Includes one tube. Transports solutions from the containers to the measuring modules, and solutions from the measuring modules to the waste container (when sample is transported into the measuring modules). Rotates in both directions.

P2

Rinse process

The basic stages of the rinse process are described below. Stage 1 Description The measuring modules are drained of fluid flowing through the inlet and out to the waste container. The pH/BG, El/Met and Met II measuring chambers drain from the top down. Fluid from the Oxi module (cuvette) flows from right to left through the sample path. The system is washed with alternating segments of air and rinse solution. The flow should be similar to that of Stage 1.
Continued on next page

11-82

ABL800 FLEX Operator's Manual

11. Troubleshooting

Fluid transport system description, Continued


Rinse process (continued) Stage 3 4 Description The measuring modules are filled with rinse solution flowing through the measuring chambers from the bottom up. The pH/BG module is filled with gas containing CO2. The gas is filled from the top, expelling the rinse solution from the measuring chambers. The outer portion of the inlet system is filled with air drawn from the exterior (not shown).

Evaluating the rinse

Carefully observe the flow of fluid through the system during the rinse. Use the table below to help identify possible sources of error. Possible cause Obstruction in inlet Indication Fluid flowing upwards through pH/BG, El/Met and Met II modules during Stage 1. Recommended action Perform inlet troubleshooting procedure.

Erratic flow of fluid through all three modules that contains excessive air bubbles. Fluid does not move through the small canals off the waste pump tubes, just above the pump rotor. Excessive suction noise coming from inlet. Improper electrode installation Fluid and air bubbles rising up into a measuring chamber. Leakage from electrode compartment. Obstruction in the measuring chamber of a measuring module Fluid does not flow freely through all measuring chambers of a particular measuring module. Clean the measuring chambers of the measuring module. Ensure the electrodes are properly installed.

11-83

11. Troubleshooting

ABL800 FLEX Operator's Manual

11-84

12. Sampling Overview


Introduction This chapter provides guidelines and recommendations for blood, pleura and expired air sample collection prior to analysis, i.e. the preanalytical phase, in order to avoid errors and obtain the most accurate results. For more comprehensive information regarding blood sampling, see the references at the end of this chapter. Contents This chapter contains the following topics. Causes of errors in preanalytical phase ............................................................ 12-2 Preparation prior to arterial/venous sampling .................................................. 12-5 Preparation prior to capillary sampling............................................................ 12-7 Sampling devices ............................................................................................. 12-8 Storage and preparation prior to analysis......................................................... 12-9 Sampling procedures........................................................................................ 12-12 References ........................................................................................................ 12-15

12. Sampling

ABL800 FLEX Operator's Manual

Causes of errors in preanalytical phase


Blood sampling should be performed by authorized personnel only. Always handle blood and the collection devices with care. Avoid direct contact with the sample by using certified rubber gloves. Sterile techniques are required at all times to avoid WARNING Risk of infection infection. Introduction The major source of errors in blood sample analysis is the preanalytical phase, i.e. sample collection and handling prior to transfer into the analyzer. In order to avoid errors, it is necessary to observe certain guidelines during: preparation prior to sampling sampling transport storage preparation prior to transfer of the sample into the analyzer Causes of error The possible causes of errors during the preanalytical phase and preventative measures are given below. Cause of error Air bubbles in sample Clot in sample** Major affected parameters pO2* pH Preventative measures Remove air bubbles immediately. Sedimented samples ctHb, pO2, pCO2 Hemolysis cK+, cCa2+, cCl, cNa+ Use correct pre-analytical handling Use Cleaning additive Capillaries: Use the mixing wire Use clot catcher.

Mix sample prior to transfer into analyzer. Sampling device construction, narrow needle diameter causes hemolysis choose a proper device for sampling. Avoid squeezing the tissue during sample collection into a capillary Avoid mixing too vigorously See section Storage and preparation prior to analysis further in this chapter
Continued on next page

12-2

ABL800 FLEX Operator's Manual

12. Sampling

Causes of errors in preanalytical phase, Continued


Causes of error (continued) Cause of error Major affected parameters Preventative measures

Dilution from liquid Electrolytes and Use preheparinized devices with heparin metabolites, cHCO3, dry heparin. pCO2, ctHb, Heparin interference Arterial blood mixed with venous blood Electrolytes, especially cCa2+ pH, pO2, sO2 Use electrolyte-balanced heparin. Use proper sampling technique.

Instability of patient pH and blood gases

Sample at least 20 minutes after ventilatory adjustment Inform patient of procedure Store glass capillary samples for maximum 30 minutes in cold (0- 4 C) condition. For patients receiving phototherapy, lights should be extinguished prior to drawing the sample. Discard sufficient volume from the catheter before taking a blood sample. Stop infusion for a period or use another sampling site.

Leakage from red blood cells Exposure to artificial light or sunlight

cK+

ctBil

Electrolytes and Inappropriate metabolites amount of flush solution is discarded from the catheter Infusion solution given in the same arm Electrolytes and metabolites

* CAUTION Risk of incorrect results Always remove air bubbles from a sample before introducing it into the analyzer. Air bubbles may result in erroneous pO2 values. ** CAUTION Risk of clots When measurements are flagged always conduct the operator actions required by the troubleshooting program to prevent possible clots from remaining in the sample path. Fluid path clots may cause erroneous measurement results.

See also Recommended literature at the end of this chapter.


Continued on next page

12-3

12. Sampling

ABL800 FLEX Operator's Manual

Causes of errors in preanalytical phase, Continued


Selection of anticoagulant To avoid introduction of clotted blood samples into the analyzer which may give inaccurate measurement results and cause analyzer damage, Radiometer recommends the following: Exclusive use of preheparinized sampling devices Exclusive use of dry heparin, preferably of sodium or lithium Do not use liquid heparin as this causes dilution of the sample [1] Use heparin in sufficient concentration. The recommended concentration depends on the sampling device and specific blood sample please refer to documentation for specific sampling devices.

Use electrolyte-balanced heparin to minimize bias from heparin on electrolyte values CAUTION Risk of erroneously high results Do not add sodium fluoride to the blood samples, as it gives erroneously high cNa+ and low cCa2+, cGlucose and cLactate results. Sodium fluoride also damages the Glucose and Lactate electrodes.

Preparing the patient

The respiratory condition of the patient should be stable. The patient should preferably be in a steady state of ventilation before and during the collection of blood sample The patient should be informed about the procedure in order to avoid unnecessary anxiety. If the patient hyperventilates due to anxiety, for example, pH and blood gases may be affected.

12-4

ABL800 FLEX Operator's Manual

12. Sampling

Preparation prior to arterial/venous sampling


Sample types The recommendations and considerations for the sample type are as follows: Sample type Arterial Recommendations and considerations The most preferred and recommended type of blood for pH/blood gas analysis. Gives the best information about oxygen uptake in the lungs and oxygen transport Information is basically the same regardless of the sampling site Venous Not generally recommended for blood gas analysis. Affected by peripheral circulatory efficiency and cellular metabolic needs Should not be used to evaluate oxygen status Can be used to reflect acid-base status (pH, pCO2, cBase(Ecf)), ctHb, FCOHb, FMetHb, FHbF and ctBil The recommendations and considerations for selecting a sampling site are as Arterial and venous sampling follows: sites Sample Site Advantages/disadvantages Arterial Radial artery Good collateral circulation Easy access Prior to sampling from the radial artery, perform a Modified Allens Test to confirm adequate collateral circulation. Brachial artery Femoral artery Difficult to puncture due to location Greater risk of damage to surrounding structures Large and easy to puncture Poor collateral circulation Should be avoided on newborns and elderly Arterial line Remove the flush solutions used in A-lines completely from the system to avoid dilution of the blood sample. It is recommended to withdraw a volume equal to three to six times the dead space of the catheter. Brachial vein Convenient site. For sampling on venous blood, use the standard procedure of your institution.
Continued on next page

Venous

12-5

12. Sampling

ABL800 FLEX Operator's Manual

Preparation prior to arterial/venous sampling, Continued


Arterial sampling guidelines The following are guidelines for sampling by arterial puncture: Sampling method Puncture Guidelines Be careful not to mix venous blood with the arterial sample. This can occur if a vein is punctured before locating the artery. Arterial blood (due to the higher pressure) will generally fill faster than venous blood, and is lighter in color. It is important to expel any air bubbles immediately without prior agitation. An air bubble with a relative volume of 0.5 % or more of the blood volume in the syringe is a potential source of significant error [2, 3]. The effect of such an air bubble increases with the storage time and agitation of the sample. Mix sample thoroughly with heparin immediately after sampling by inverting the sampler several times and then rolling the sample between your palms. A-line If aspirating the sample, do so slowly in order to avoid hemolyzation of the sample Expel any air bubbles immediately (see above) Mix the sample thoroughly with heparin immediately after sampling by inverting the sampler several times and then rolling the sample between your palms.

Disposal of used sampling devices

WARNING Risk of infection Discard the waste as infectious waste [4].

12-6

ABL800 FLEX Operator's Manual

12. Sampling

Preparation prior to capillary sampling


Capillary sampling sites The following are recommended capillary sampling sites: Earlobe Finger tip Big toe* * The heel and big toe should not be used on patients older than six months. NOTICE: In general, results obtained from capillary samples, particularly pO2 values, should be interpreted with caution. Capillary sampling guidelines The guidelines for capillary sampling are as follows: Arterialize the puncture site by warming it up to approximately 42 C for 5 to 10 minutes prior to actual sampling. Failure to do so will make the blood sample only representative of the local tissue, and not the general status of the patient. Make a puncture using a lancet or the like, so that blood flows freely. Do not squeeze the area. Remove the first drop of blood, since it may be diluted with tissue fluid. Take the sample from the center of the second blood drop. To prevent air from entering the capillary, take the sample from the center of the drop. Squeezing or milking the puncture site is to be avoided as this will result in an admixture of blood and tissue fluid, resulting in faulty measurements. It can also cause hemolysis of the blood sample (or part of the sample), resulting in cK+ readings being too high. Mix the sample with heparin immediately after collection to prevent blood from clotting. (Clots can clog the analyzer and cause unnecessary downtime.) Use a mixing wire and magnet when mixing a capillary sample. Fetal scalp Heel*

Disposal of used sampling devices

WARNING Risk of infection Discard the waste as infectious waste [4].

12-7

12. Sampling

ABL800 FLEX Operator's Manual

Sampling devices
Introduction Radiometer has an assortment of specialized blood gas sampling devices. The PICO product line is used for puncture and A-line sampling, and CLINITUBES product line are used for capillary sampling. For ordering information on sampling devices, see chapter 14: Ordering Information or contact your local distributor of Radiometer products. Arterial blood samplers The following arterial blood samplers available from Radiometer are recommended: PICO70: Self-filling arterial blood sampler for arterial puncture or A-line sampling PICO50 (2 mL): Arterial blood sampler for the aspiration of arterial blood from A-lines SafePICO: Self-filling arterial blood sampler for arterial puncture or A-line sampling (contains a ball for automatic sample mixing in the FLEXQ sampler tray)

CLINITUBES

The following CLINITUBES for capillary blood sampling available from Radiometer are recommended: D956G-70-210 D956G-70-100 D957P-70-100 D941G-240-55 (do not use for measuring electrolytes) D956G-70-35 (G = glass; P = plastic; the last figure designates volume in L) CAUTION - Risk of erroneous results Do not use 85 L CLINITUBES sampling device with high heparin concentration in FLEXMODE.

12-8

ABL800 FLEX Operator's Manual

12. Sampling

Storage and preparation prior to analysis


Storage recommendation Radiometer recommends the following for samples obtained in plastic and glass syringes: Material Plastic syringe Storage recommendation Storage should be avoided whenever possible or, at least, kept to a minimum If it is not possible to analyze the sample immediately, analyze it within 30 minutes after collection [5, 6, 7, 8, 9, 10, 11] Recommended sample storage temperature: room temperature [5, 6, 8] Samples with expected high pO2 values or for special studies like shunt studies should be analyzed immediately or within 5 minutes. The use of glass syringes can also be considered [5, 7, 8] Glass syringe Storage should be avoided whenever possible or, at least, kept to a minimum If it is not possible to analyze the sample immediately, analyze it within 30 minutes after collection when stored at room temperature [5, 12] Alternatively, store the sample in ice-water (0-4 C ). The storage time should not exceed 1 hour [13] Samples with expected high pO2 values or for special studies like shunt studies should be analyzed immediately or within 5 minutes [13]

Radiometer recommends the following for samples obtained in plastic and glass capillaries: Material Plastic capillary Storage recommendation Analyze the sample within 10 minutes. Note that for samples with pO2 > 80 mmHg (10.7 kPa) a positive bias from 1 to 9 % is observed. The bias depends on the pO2, pH and time. Analyze the sample within 10 minutes. If storage is unavoidable, store the sample horizontally at 0-4 oC for maximum 30 minutes.
Continued on next page

Glass capillary

12-9

12. Sampling

ABL800 FLEX Operator's Manual

Storage and preparation prior to analysis, Continued


Pay attention to For some samples the recommendation above does not apply and individual the following: guidelines should be used or developed. Examples of these samples are samples with an increased leukocyte or platelet count, fetal scalp samples, samples with atypical metabolism, fast-clotting samples, etc. [9, 14]. Typical metabolic activity in blood samples causes an increase in the lactate concentration and a decrease in the glucose concentration. For samples within the typical reference range this corresponds to an average change in the lactate concentration of 0.25 mmol/L and of 0.2 mmol/L for glucose over 30 minutes at room temperature [15]. Arterial puncture and A-line samples: Preparation prior to analysis If there is an air bubble in a sample that has been stored, pH and blood gas values should not be reported Thoroughly mix the sample by inverting the sampler several times and rolling it between the palms of your hands.

Capillary samples: Thoroughly mix the sample with heparin using a mixing wire and magnet. Capillary blood sample measurements In order to perform in accordance with the analyzer performance specifications, Radiometer highly recommends to follow the recommendations below. Capillary blood sample When performing capillary blood sample measurements: measurements, use the mixing wire in order to ensure appropriate mixing of the sample and thereby prevent clots from being introduced into the analyzer. The sample should be mixed immediately after collection, moving the mixing wire with a magnet 20 times along the full length of the capillary. This procedure should be repeated immediately before analyzing the sample. When performing capillary blood sample measurements and clots are suspected, use a clot catcher in order to prevent blockage of the measuring chamber. Do not remove the mixing wire before the sample is aspirated for measurement. Slide the mixing wire to the end of the capillary opposite to that from which the blood is to be aspirated. Remove both capillary end caps.
Continued on next page

12-10

ABL800 FLEX Operator's Manual

12. Sampling

Storage and preparation prior to analysis, Continued


Capillary blood sample measurements (continued) Capillary blood sample When performing capillary blood sample measurements on fetal measurements on fetal scalp blood, insert the scalp blood: mixing wire into the capillary tube after the sample has been collected. When performing measurements on fetal scalp blood, introduce the capillary sample into the analyzer from the cap end free from Vaseline smear. In case both capillary ends are smeared, dry one of the ends with a lint-free tissue before measurement. Due to their inherent characteristics fetal scalp blood samples are generally regarded as potentially difficult to work with. An important prerequisite for successfully analyzing fetal scalp blood samples is the sample quality and that the preanalytical precautions listed above are followed very closely.

12-11

12. Sampling

ABL800 FLEX Operator's Manual

Sampling procedures
NOTICE: This section describes non-Radiometer sampling devices. For information on use and handling Radiometer sampling devices, please refer to the inserts supplied with the sampling devices.

Sampling with a Prepare the following together with a standard syringe (dead space 2-6 %): standard syringe Anticoagulant: lithium or sodium heparin (1000 IU per mL sample) Two needles, 20-25 gauge Tourniquet for venous samples Skin antiseptic, e.g. 70 % alcohol Sterile gauze sponge or cotton wool Rubber stopper or syringe end cap (stainless steel) Adhesive tape and pressure dressing

Step 1. 2. 3. 4. 5. 6.

Action Mount a needle on the syringe and draw the heparin into the syringe. Change the needle and flush the barrel walls, then expel air and heparin through the needle. Extend the patients arm. Apply the tourniquet for a venous puncture. Locate the blood vessel with the finger tip and clean the skin area with antiseptic and a sterile gauze sponge or cotton wool. Insert the needle in the vessel with the bevel upward. Withdraw the syringe plunger gently to collect blood. Release the tourniquet (if used), remove the needle and immediately apply dry gauze or cotton wool over the puncture site. Maintain pressure on an arterial puncture for at least 5 minutes. Apply a pressure dressing over the puncture site and elevate the site for at least 2-3 minutes, if possible. Expel any air bubbles from the syringe and remove the needle according to your standard procedure. Mix the sample by inverting the syringe several times and then rolling it between your palms. Mix the sample immediately before analysis. Analyze the sample immediately.
Continued on next page

7. 8. 9. 10

12-12

ABL800 FLEX Operator's Manual

12. Sampling

Sampling procedures, Continued


Sampling with a Prepare the following items: test tube A test tube and cap Anticoagulant: lithium or sodium heparin (1000 IU per mL sample) Pipette or syringe and a needle for heparin dosage One hypodermic needle, 20-25 gauge Tourniquet Skin antiseptic, e.g. 70 % alcohol Sterile gauze sponge or cotton wool Adhesive tape and pressure dressing Step 1. 2. 3. 4. 5. 6. 7. 8. Action Transfer the heparin to the test tube using a syringe or a pipette: 50 L heparin per 5 mL blood. Extend the patients arm. Apply the tourniquet for a venous puncture. Locate the vein with the fingertip. Clean the skin area with antiseptic and a sterile gauze sponge or cotton wool. Collect the sample according to the procedure in your institution. Apply dry gauze or cotton wool over the puncture site. Close the test tube with a cap, then mix the contents by inverting it or rolling it between your palms about 20 times. Apply a pressure dressing over the puncture site and elevate the site for at least 2-3 minutes, if possible. Mix the sample immediately before analysis. Analyze the sample immediately.

Pleural liquid sampling

To collect a pleural liquid sample, do the following: Step 1. Action Collect the sample directly from the patient or from a drain. NOTICE: Radiometer recommends always to use heparinized samplers for pleural liquids to avoid introducing clots in the analyzer.
Continued on next page

12-13

12. Sampling

ABL800 FLEX Operator's Manual

Sampling procedures, Continued


Expired air sampling To collect an expired air sample, do the following: Step 1. 2. 3. Action Collect the patient's expired air in the Douglas bag. Connect a syringe (minimum 20 mL, with a rubber cap) to the bag. Fill and empty the syringe twice, then fill it with the sample. Close the syringe with the rubber cap.

12-14

ABL800 FLEX Operator's Manual

12. Sampling

References
List of references This section lists the references applicable to blood sampling and blood gas analysis associated with this chapter. 1. 2. Avoid Errors in arterial blood sampling. Copenhagen: Radiometer Medical A/S, 1992. Code 919-467. Biswas CK, Ramos JM, Agroyannis B, Kerr DNS. Blood gas analysis: effect of air bubbles in the syringe and delay estimation. B Med J 1982; 284: 923-27. Mueller RG, Lang GE, Beam JM. Bubbles in samples for blood gas determinations a potential source of error. Am J Clin Pathol 1976; 65: 242-49. Clinical laboratory waste management. CLSI document GP5-A2. Blonshine S, Alberti R, Olesinski RL. Procedures for the collection of arterial blood specimens; approved standard Third Edition. NCCLS document H11-A3 (ISBN 1-56238-374-4). Wayne, Pa,: NCCLS, 1999: 12, 8. Mahoney JJ, Van Kessel A. Arterial blood gas analysis. Respiratory Care 1997: 249-79. Smeenk F, Janssen J, Arends B, Harff G, Bosch J, Schnberger J, Postmus P. Effects of four different methods of sampling arterial blood and storage time on gas tensions and shunt calculation in the 100% oxygen test. Eur Respir J 1996; 10: 910-13. Mahoney JJ, Harvey JA, Wong RJ, Kessel VLA. Changes in oxygen measurements when whole blood is stored in iced plastic or glass syringes. Clin Chem 1991; 37: 1244-48. Woolley A and Hickling K. Errors in measuring blood gases in the intensive care unit: Effect of delay in estimation. Journal of Critical Care 2003; 18: 31-37.

3.

4. 5.

6. 7.

8.

9.

10. Blonshine S. To ice or not to ice. AARC Times 2000: 37-39. 11. Liss P, Payne P. Stability of blood gases in ice and at room temperature. Chest 1993; 103: 1120-21. 12. Burnett RW, Chatburn RL, Christiansen TF et al. Blood gas preanalytical considerations: specimen collection, calibration and controls; Approved guideline. NCCLS document C27-A (ISBN 1-56238-190-3). Villanova Pa,: 1993; 13, 6. 13. Burnett RW, Covington AK, Fogh-Andersen N et al. Approved IFCC recommendations on whole blood sampling, transport and storage for simultaneous determination of pH, blood gases and electrolytes. Eur J Clin Chem Clin Biochem 1995; 33: 247-53.
Continued on next page

12-15

12. Sampling

ABL800 FLEX Operator's Manual

References, Continued
List of references (continued) 14. Nickelsen CN. Fetal capillary blood pH. http://www.bloodgas.org /2002. 15. Wennecke G. Useful tips to avoid preanalytical errors in blood gas testing: metabolites. http://www.bloodgas.org / 2004.

Recommended literature

1. 2.

Capillary sampling. In: The whole-blood sampling handbook. Copenhagen: Radiometer Medical ApS, 2003. Code 989-466. Christiansen TF. Heparin and blood sampling for pH, blood gases and direct potentiometric electrolyte analysis. Radiometer Publication AS96. Copenhagen: Radiometer Medical A/S, 1986. The blood gas handbook. Copenhagen: Radiometer Medical ApS, 2004. Code 989-318 Bower LK, Barnhart SL, Betit P, Hendon B, Masi-Lynch J, Wilson BG. AARC Clinical practice guideline. Capillary blood gas sampling for neonatal and pediatric patients. Respir Care 1994; 39, 12: 1180-83. Moran RF, Bergkuist C, Graham GA et al. Considerations in the simultaneous measurement of blood gases, electrolytes and related analytes in whole blood. Proposed Guideline. CLSI (former NCCLS) Publication C32-P. Villanova Pa.: NCCLS, 1993; 13, 17. Christiansen TF. Measurement of potassium in whole blood. Radiometer Publication AS92. Copenhagen: Radiometer Medical A/S, 1983. Code 918-231. Wandrup JH. Clinical and analytical requirements and needs of glucose measurements on whole blood. Radiometer Publication AS122. Copenhagen: Radiometer Medical A/S, 1996. Code 918-549. Wandrup JH, Tvede K, Grinsted J, Jordening H. STAT measurements of L-lactate in whole blood and cerebrospinal fluid assessed. Clin Chem 1989; 35, 8: 1740-43. Wu E, Barazanji K, Johnson R. Sources of error in A - aDo2 calculated from blood stored in plastic and glass syringes. J Appl Phycial 1997; 82: 196-202.

3. 4.

5.

6.

7.

8.

9.

12-16

13. Overview
Introduction Contents

Specifications

This chapter lists specifications for the ABL800 FLEX analyzers. This chapter contains the following topics. Measured parameters ....................................................................................... 13-2 Input parameters............................................................................................... 13-5 Derived parameters .......................................................................................... 13-6 Sample handling............................................................................................... 13-8 Calibration and maintenance programs............................................................ 13-11 Analyzer requirements ..................................................................................... 13-13 Analyzer specifications .................................................................................... 13-14 Approvals and patents ...................................................................................... 13-16

13. Specifications

ABL800 FLEX Operator's Manual

Measured parameters
Definitions The measuring range for a parameter is the range within which the analyzer is physically capable of measuring. The measuring range corresponds to the "range of indication" as defined in the "International vocabulary of basic and general terms in metrology" (VIM). The test range for a parameter is the range within which accuracy and precision of a measured parameter has been specified and intended to lie within specified limits. The test range corresponds to the "measuring range" as defined in the "International vocabulary of basic and general terms in metrology" (VIM). Blood The following measured parameters are available, independent of analyzer configuration. Parameter pH cH+ pCO2 Unit pH scale nmol/L mmHg; Torr kPa pO2 mmHg; Torr kPa ctHb g/dL g/L mmol/L sO2 % fraction FO2Hb % fraction FCOHb % fraction FMetHb % fraction FHHb % fraction For (*) see explanation on the next page.
Continued on next page

Measuring range 6.300-8.000 10.0-501 5.0-250 0.67-33.3 0.0-800 0.00-107 0.00-27.7 0.0-277 0.00-17.2 0.0-100.0* 0.00-1.000* 0.0-100.0* 0.00-1.000* 0.0-100.0* 0.00-1.000* 0.0 100.0* 0.00-1.000* 0.0 100.0* 0.00-1.000*

Test range 6.85-7.55 28-141 17-160 2.27-21.3 20-580 2.67-77.3 2.5-23 25-230 1.55-14.2 0-100 0-1 0-100 0-1 0-20 0.0-0.2 0-20 0.0-0.2 0-100 0-1

13-2

ABL800 FLEX Operator's Manual

13. Specifications

Measured parameters, Continued


Blood (continued) Parameter FHbF Unit % fraction cK+ cNa+ cCa2+ mmol/L; meq/L mmol/L; meq/L mmol/L meq/L mg/dL cCl cGlu mmol/L; meq/L mmol/L mg/dL cLac mmol/L; meq/L mg/dL cCrea mol/L mg/dL ctBil mol/L mg/dL mg/L Measuring range 0.0-100.0* 0.00-1.000* 0.5-25.0 7-350 0.20-9.99 0.40-19.98 0.80-40.04 7-350 0.0-60 0-1081 0.0-30 0-270 10-1800 0.11-20.4 1-1000* 0.0-58.5* 0-585* Test range 0-80 0.0-0.8 2-8 120-180 0.51-2.2 1.0-4.4 2.0-8.8 95-150 0.5-15 9.0-270 0.5-15 4.5-135 50-1500 0.57-17.0 0-400 0.0-23.4 0-234

* The values are for the analyzer with the activated "Out of range suppression" function. In case this function has not been activated, the measuring ranges will be as follows: Parameter sO2, FO2Hb, FCOHb, FMetHb FHbF Unit % fraction % fraction ctBil mol/L mg/dL mg/L Measuring range 2.0 % to 102.0 % 0.020 to 1.020 15 % to 102 % 0.15 to 1.02 20 to 1000 1.2 to 58.5 12 to 585

Continued on next page

13-3

13. Specifications

ABL800 FLEX Operator's Manual

Measured parameters, Continued


Pleural liquid Parameter pH Unit pH scale Measuring Range 6.300-8.000 Test range 7.0-7.5*

* If the measured values obtained lie outside the test range, Radiometer advises you to repeat the measurement by means of another method. Expired air Parameter pCO2 Unit mmHg kPa pO2 mmHg kPa Baro Parameter p(amb) Unit mmHg; Torr kPa Measuring Range 5.0-250 0.67-33.3 0.0-800 0.00-107 Measuring Range 450-800 60.0 -106.7 Test range 15-150 2.00-20.00 15-530 2-70 Test range 450-800 60.0-106.7

Other fluids

All parameters available on the analyzer.

13-4

ABL800 FLEX Operator's Manual

13. Specifications

Input parameters
List of parameters All the input parameters available on the ABL800 FLEX analyzers, independent of configuration, are listed below. Symbol T Parameter Patient temperature Unit
o

Input range Defaults 15.0-45.0 59.0-113.0 18.0-32.0 64.4-89.6 0.0-100.0 0.000-1.000 0.0-33.0 0.0-330 0.0-20.5 0.00-2.00 37 98.6 25 77 21.0 0.210 0.86

C F

QUALICHECK+/AutoCheck+ ampoule temperature FO2(I) Fraction of oxygen in dry inspired air Total hemoglobin concentration (if not measured)

C F

% fraction g/dL g/L mmol/L

ctHb

RQ

Respiratory quotient ratio of carbon dioxide production and oxygen consumption

) Oxygen tension in mixed pO2(v venous blood ) sO2(v Qt VO2 Oxygen saturation in mixed venous blood Cardiac output Total oxygen consumption

mmHg kPa % fraction L/min mL/min mmol/min

0.0-xxx.x* 0.00-xx.xx* 0.0-100.0 0.000-1.000 0.0-xxx.x* 0-xxxx* 0.0-xxx.x* 0.0-xxx.x* 0.01-100.00 0.001-13.332 0.0 -100.0 0.000-1.000 0.0 -100.0 0.000-1.000

26.84 3.578 -

VCO

Volume of carbon monoxide

mL mmHg kPa % fraction % fraction

p50(st) Half saturation tension of oxygen in blood at standard conditions FCOHb The fraction of COHb measured before the CO-injection (1) FCOHb The fraction of COHb measured after the CO-injection (2)

* numerical format

13-5

13. Specifications

ABL800 FLEX Operator's Manual

Derived parameters
List of parameters All derived parameters of the ABL800 FLEX analyzers are listed below. Symbol pH(T) cH+(T) pCO2(T) cHCO3(P) cBase(B) cBase(B,ox) cBase(Ecf) cBase(Ecf,ox) cHCO3(P,st) ctCO2(P) ctCO2(B) VCO2/V(dry air) pH(st) Hct pO2(T) pO2(A) pO2(A,T) pO2(a)/FO2(I) pO2(a,T)/FO2(I) p50 p50(T) p50(st) pO2(A-a) pO2(A-a,T) pO2(a/A) pO2(a/A,T) ctHb T FO2(I) + RQ FO2(I) + RQ + T FO2(I) FO2(I) + T T FO2(I) + RQ FO2(I) + RQ + T FO2(I) + RQ FO2(I) + RQ + T
Continued on next page

Input parameter required (if not determined by analyzer) T T T ctHb ctHb ctHb -

13-6

ABL800 FLEX Operator's Manual

13. Specifications

Derived parameters, Continued


List of parameters (continued) Symbol pO2(x) pO2(x,T) ctO2(B) ) ctO (a-v
2

Input parameter required (if not determined by analyzer) ctHb, p50(st) ctHb, p50(st), T ctHb ctHb ctHb ctHb, p50(st) Q t VO2 Q t ctHb + FO2(I) + RQ ctHb + FO2(I) + RQ + T FO2(I) + RQ FO2(I) + RQ + T ctHb, p50(st)

BO2 ctO2(x) D O2 Qt VO2 FShunt FShunt(T) RI RI(T) Qx VO2/V(dry air) sO2 FO2Hb FHHb V(B) Anion Gap(K+) Anion Gap cCa2+(pH=7.4) mOsm GFR, if non AA GFR, if AA

ctHb, VCO, FCOHb(1), FCOHb(2)

Crea, Sex, Age Crea, Sex, Age

NOTICE: All parameters are calculated in SI units. If other units are selected, the results are obtained by converting the SI units into those selected. For detailed information please refer to the Reference Manual, chapter 6.

13-7

13. Specifications

ABL800 FLEX Operator's Manual

Sample handling
ABL837/27/17 Mode Syringe S250 Other fluids S250 Ampoule - QC Syringe S85 pH + BG + OXI Capillary C125 Capillary C55 Capillary C35 Met Capillary C35 OXI Syringe Pleura pH Expired air Sample volume 250 L Measuring time (sec)* 100 Cycle time (sec) 170 Samples per hour 21

85 L 125 L 55 L 35 L 35 L 85 L 15 mL

80 150 100 80 80 80 80

170 225 80 145 145 170 170

21 16 20 24 24 21 21

ABL835/25/15

Mode Syringe S195 Other fluids S195 Ampoule - QC Capillary FLEXMODE Other fluids FLEXMODE (no message) Capillary FLEXMODE Other fluids FLEXMODE (message 869 or 870) Capillary FLEXMODE Other fluids FLEXMODE (message 871 or 872) Capillary FLEXMODE Other fluids FLEXMODE (message 873) Capillary FLEXMODE Other fluids FLEXMODE (message 874) Syringe S95 Capillary C95

Sample volume 195 L

Measuring time (sec)* 80

Cycle time (sec) 150

Samples per hour 24

195 L

80

150

24

30-40 L or 40-55 L 50-70 L

100

170

21

100

170

21

65-100 L > 90 L

135

200

18

135

200

18

95 L

135

200

18

* From the moment the inlet flap is lifted until the results are displayed.
Continued on next page

13-8

ABL800 FLEX Operator's Manual

13. Specifications

Sample handling, Continued


ABL835/25/15 (continued) Mode Syringe S85 Capillary C55 Capillary C35 Met Capillary C35 Oxi Syringe Pleura pH Expired air Sample volume 85 L 55 L 35 L 35 L 85 L 15 mL Measuring time (sec)* 80 100 80 80 80 80 Cycle time (sec) 170 170 145 145 170 170 Samples per hour 21 21 25 25 21 21

ABL830/20/10/ 10 BG only

Mode Syringe S85 Other fluids S85 Ampoule - QC Capillary FLEXMODE Other fluids FLEXMODE (no message)

Sample volume 85 L

Measuring time (sec)* 80

Cycle time (sec) 170

Samples per hour 25

85 L

80

145

25

Capillary FLEXMODE 30-40 L Other fluids FLEXMODE or 40-55 L (message 869 or 870) Capillary FLEXMODE 50-70 L Other fluids FLEXMODE (message 871 or 872) Capillary C55 Capillary C35 Oxi (not active on the ABL810 BG only) Syringe Pleura pH Expired air 55 L 35 L

100

170

21

100

170

21

100 80

170 145

21 25

85 15 mL

80 80

170 170

21 21

* From the moment the inlet flap is lifted until the results are displayed.
Continued on next page

13-9

13. Specifications

ABL800 FLEX Operator's Manual

Sample handling, Continued


ABL805 Mode Syringe S165 Other fluids S165 Ampoule - QC Capillary FLEXMODE Other fluids FLEXMODE (no message) Capillary FLEXMODE 30-40 L Other fluids FLEXMODE or 40-55 L (message 869 or 870) Capillary FLEXMODE 50-70 L Other fluids FLEXMODE (message 871 or 872) Capillary FLEXMODE Other fluids FLEXMODE (message 873) Capillary FLEXMODE Other fluids FLEXMODE (message 874) Syringe S95 Capillary C95 Syringe S85 Capillary C55 Capillary C35 Met Syringe Pleura pH Expired air 95 L 85 L 55 L 35 L 85 L 15 mL 100 80 80 80 80 65 170 170 145 145 170 170 19 21 25 25 21 21 > 90 L 100 170 21 65-100 L 80 170 21 80 145 25 Sample volume 165 L Measuring time (sec)* 80 Cycle time (sec) 170 Samples per hour 21

165 L

80

150

24

80

145

25

* From the moment the inlet flap is lifted until the results are displayed.

13-10

ABL800 FLEX Operator's Manual

13. Specifications

Calibration and maintenance programs


Calibrations Calibr Adjustable Duration, min ation type frequency ABL837/27/17 ABL835/ 25/15 ABL830/20/10 (hours) Cal 1 Cal 2 1 Point pH/BG (for the USA only) tHb , 1, 2, 4 1, 2, 4, 8 , 1, 2 5 10 2 4 9 2 3 8 2

ABL805 4 9 2

Manual

High Crea Check Auxiliary programs

Duration: 3 min.

Program

Started

Duration, min ABL837/27/ ABL835/25 ABL830/20 ABL805 17 /15 /10

Cleaning

Every 8 or 24 hours, or called by operator After changing all solution bottles or called by operator

Tubing Refill

Decontami- Called by nation operator Protein Remover Called by operator

4 3 3

2 2 3

2 2 3

2 2 3

Liquid Called by Sensor operator Adjustment

Fortsttes p nste side

13-11

13. Specifications

ABL800 FLEX Operator's Manual

Calibration and maintenance programs, Continued


Auxiliary programs (continued) Program Started Duration, min ABL837/27/ ABL835/25 ABL830/20 ABL805 17 /15 /10 Pump Called by Calibration operator Rinse After each measurement, calibration or called by operator Automatically Automatically 4 1 4 80 sec 3 80 sec 4 80 sec

Flush Draining

50 sec 21 sec

50 sec 21 sec

50 sec 21 sec

System alignment

ABL830: approximately 21 min. ABL835/05: approximately 24 min. ABL837/27/17: Approximately 26 min.

13-12

ABL800 FLEX Operator's Manual

13. Specifications

Analyzer requirements
Analyzer requirements Power Rated voltage: 100-240 V; 50/60 Hz; 270 VA. Maximum voltage fluctuations: 10 %. Leakage current measured at 230 V and 50 Hz: Normal condition (ground connected): < 0.005 mA Single fault condition (ground disconnected): Fuses < 0.5 mA

Printer unit has 1 protective fuse: 5 20 mm, 4 Amp, Slow blow (T4AL). Main fuse includes 2 protective fuses: 5 20 mm, 4 Amp, High break (T4AH). Type Shurter No. 0001.2510. WARNING Risk of fire Fire hazard. Replace fuse as marked.

Start up Gas pressure

Cold start: typically 25 minutes Warm start: 5 minutes Operating pressure: 1.0-38 bar Maximum pressure: 38 bar

Environmental requirements

Location Relative humidity Operating temperature

Indoor use only 20 to 80 % 15 C to 32 C

Storage temperature 20 C to 60 C Altitude correction Baro Up to 3000 meters above sea level at standard barometric pressure of 760 mmHg. 450-800 mmHg 60.0-106.7 kPa 0.600-1.067 bar 450-800 Torr

Environmental ratings

Installation category II. Pollution degree 2.

13-13

13. Specifications

ABL800 FLEX Operator's Manual

Analyzer specifications
Specifications Thermostatting Spectrophometer Hemolyzer External serial ports USB ports Ethernet Solid state, 37.0 0.15 C Wavelength range: 478-672 nm Hemolyzation: at approximately 30 kHz cuvette light path: approximately 0.01 cm 1 x RS-232 (9-pin) connector. Baud rate: 1200, 2400, 4800, 9600, 14400, 19200, 38400, 115200 Pin header for two USB (Universal Serial Bus) ports 1 x RJ45 connector, 32-bit PCI bus interface, Novell NE 2000 compatible, onboard 10-Base T, fully compliant with IEEE 802.3 10 Mbps CSMA/CD standards. Bi-directional parallel (ECP/EPP/SPP) port (25-pin) connector Keyboard/PS/2 mouse interface Connector for VGA screen Built-in thermal; paper 11 cm wide Built-in. Reading distance: contact 70 mm. Barcode length: max. 65 mm Bar width: 127 m (5 mil) Number of characters: <62 Code types: 128, 39, I 2 of 5 Class 1 laser product Codabar barcode support Height: 548 mm(21.95 in.) with the vertical screen Width: 700 mm (27.6 in.) Depth: 476 mm(18.5 in.) ABL805: ABL830/20/10: ABL835/25/15: ABL837/27/17 32.9 32.9 33.9 36.2
Continued on next page

External printer port Keyboard/Mouse External VGA Screen Printer Barcode reader

Dimensions (basic module, without accessories) Weight (kg) (with AutoCheck)

13-14

ABL800 FLEX Operator's Manual

13. Specifications

Analyzer specifications, Continued


AutoCheck module specifications Number of ampoules in the carousel Positioning of ampoules in the carousel Lot change Ampoule volume, liquid Expiration of ampoules 0 to 20 Random 2 Lots of the same level at the same time possible 0.7 mL 24 months at 25 C (including 15 days at up to 32 C) < 60 sec with filled carousel Yes Remote monitoring and start of measurement via the RADIANCE system.

Conditioning time (from room temperature) < 15 min. with filled carousel Scanning time Manual quality control measurement possible Remote control

FLEXQ module The specifications for the FLEXQ module are as follows: specifications Number of samplers in the sampler tray up to 3 Sampler type Positioning of samplers in the sampler tray Sample mixing time Scanning time per sample Warning signal for missing sampler ID Barcode scanner reads safePICO with safeTIPCAP Random 7 sec Typically <1 sec < 5 sec Code 128 between 5-10 mil.

13-15

13. Specifications

ABL800 FLEX Operator's Manual

Approvals and patents


Approvals The equipment complies with the following standards: UL 61010-1: Electrical Equipment for Measurement, Control, and Laboratory Use; Part 1: General Requirements IEC 61010-2-81: Electrical Equipment for Laboratory Use; Particular requirements for automatic and semiautomatic laboratory equipment for analysis and other purposes IEC 61010-2-101: Electrical Equipment for Measurement, Control, and Laboratory Use; Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment The equipment complies with the requirements in 98/79/EC on in vitro diagnostic medical devices The equipment complies with the emission requirements for Class A equipment in EN 61326-1: Electrical equipment for measurement, control and laboratory use EMC requirements - Part 1: General requirements.

CE-mark

EMC emission

EMC immunity The equipment complies with the immunity requirements in EN 61326-2-6: Electrical equipment for measurement, control and laboratory use EMC requirements - Part 2-6 Particular requirements - In vitro diagnostic (IVD) medical equipment Patents One or more of the following patents and patent applications may apply: US Patent Nos.: US6051389, US6099804, US6551480, US6689318, US6880384, US6980285 European Patent Nos.: EP889951, EP944731, EP1084398 Japanese Patent Nos.: JP2972351, JP3285879, JP3369547, JP4229912 German Patent Nos.: DE69729185, DE69735127, DE69938550 US Patent Application Nos.: US2005-0130285, US2006-0275857, US2006-0275859, US2006-0275860 European Patent Application Nos.: EP1086366, EP1273920, EP1583952, EP1692522, EP1885871, EP1889075 Japanese Patent Application Nos.: JP2007-512519, JP2008-541103, JP2008-541104, JP2008-545952 Chinese Patent Application Nos.: CN101175999, CN101184851, CN101223284

13-16

14. Overview
Introduction Contents

Ordering information

This chapter lists accessories used for the analyzer. This chapter contains the following topics. Analyzer accessories ........................................................................................ 14-2
X X

Quality control ................................................................................................. 14-5


X X

Sampling devices ............................................................................................. 14-7


X X

14. Ordering information

ABL800 FLEX Operator's Manual

Analyzer accessories
Introduction The following accessories are available on order for the ABL800 FLEX analyzer depending on analyzer configuration. Item Reference Electrode pH Electrode pO2 Electrode
T T B B

Electrodes

Code No. 945-603 945-614 945-613 945-612 945-615 945-618 945-617 945-616 945-620 945-619 945-662 945-663 945-626

Type E1001 E777 E799 E788 E722 E755 E744 E733 E7066 E7077 E8088 E8089 -

pCO2 Electrode
T T B B

K Electrode Na Electrode Cl Electrode Ca Electrode Glucose Electrode Lactate Electrode Crea A electrode Crea B electrode Dummy Electrode

Membrane boxes
T

Item Reference Membrane Box (4 units) pO2 Membrane Box (4 units)


T B B

Code No. 942-058 942-064 942-063 942-059 942-062 942-061 942-060 942-065 942-066 942-073

Type D711 D799 D788 D722 D755 D744 D733 D7066 D7077 D8088 and D8089

pCO2 Membrane Box (4 units)


T T B B

K Membrane Box (4 units) Na Membrane Box (4 units) Cl Membrane Box (4 units) Ca Membrane Box (4 units) Glucose Membrane Box (4 units) Lactate Membrane Box (4 units) Membrane Box for Crea A and Crea B electrodes (2 2 units)

Continued on next page

14-2

ABL800 FLEX Operator's Manual

14. Ordering information

Analyzer accessories, Continued


T T

Solutions

ABL835/830/825/820/815/810/810 BG only/805: Item Waste Container 600 mL ctHb Calibration Solution (4 ampoules)
T T

Code No. 905-802 944-021 944-126 944-128 944-129 944-132

Type D513 S7770 S8375 S1820 S1830 S4980

Cleaning Solution 175 mL Calibration Solution 1 200 mL Calibration Solution 2 200 mL Rinse Solution 600 mL

ABL837/27/17: Item Waste Container 600 mL ctHb Calibration Solution (4 ampoules)


T T

Code No. 905-802 944-021 944-136 944-133 944-134 944-159

Type D513 S7770 S8377 S1827 S1837 S4987

Cleaning Met II Solution, 100 mL 6 Calibration Solution 1, 175 mL 6 Calibration Solution 2, 150 mL 6 Rinse Solution, 600 mL

Gas accessories

Item Gas Cal 1 (10 bar) Onboard gas bottle Gas Cal 2 (10 bar) Onboard gas bottle Gas Cal 1 (34 bar) Onboard gas bottle for non-EU countries Gas Cal 2 (34 bar) Onboard gas bottle for non-EU countries Gas Cal 1 (34 bar), Onboard gas bottle for the USA and Canada Gas Cal 2 (10 bar), Onboard gas bottle for the USA and Canada Gas Cal 1 (26 bar), Onboard gas bottle for Japan Gas Cal 2 (10 bar), Onboard gas bottle for Japan Valve Key

Code No. 962-169 962-170 962-183 962-184 962-174 962-175 962-176 962-177 922-509

Type -

Continued on next page

14-3

14. Ordering information

ABL800 FLEX Operator's Manual

Analyzer accessories, Continued


T T

Power cords

Item Line Cord 120 V, USA and Japan Line Cord 230 V, UK Line Cord 230 V, ITA Line Cord 230 V, DK Line Cord 230 V, ISR Line Cord 230 V, CHE Line Cord 230 V, other 230 V countries Line Cord 230 V, AUS and NZA Line Cord 230 V, ZAF and IND

Code No. 615-403 615-312 615-313 615-314 615-315 615-316 615-303 615-317 615-318

Type -

Other accessories

Item Thermal Paper (8 rolls) Hypochlorite Solution Inlet Gasket Inlet Probe Inlet Probe Tube Inlet Syringe Handle Inlet Capillary Handle Pump Tube for electrode modules Waste Pump Tube, yellow (2 pcs.) Pump Tube for solutions Fan Filter Clot Catcher for the ABL800 FLEX analyzer, 250 units

Code No. 984-070 943-906 902-668 902-797 841-780 902-669 902-670 842-328 842-326 842-327 924-073 906-020

Type S5362 -

Documentation

Item Communication Protocol Specifications, English ABL800 FLEX Operators manual, English ABL800 FLEX Reference manual, English ABL800 FLEX Reference manual on CD-ROM, English

Code No. 989-329 994-908 989-963 990-449

Type -

14-4

ABL800 FLEX Operator's Manual

14. Ordering information

Quality control
Quality control accessories Item Ampoule Opener QC Adapter Code No. 920-712 924-036 Type H700

Automatic quality control accessories

Item Carousel Ampoule Holder

Code No. 887-865 924-185

Type H705

QUALICHECK3+ The QUALICHECK3+ quality control system contains the following solutions:
T

Type S7330 S7340 S7350 S7360 Level 1 RED

Description

Code 944-049 944-050 944-051 944-052

Level 2 YELLOW Level 3 BLUE Level 4 GREEN

QUALICHECK5+ The QUALICHECK5+ quality control system contains the following solutions:
T

Type S7730 S7740 S7750 S7760 Level 1 RED

Description

Code 944-017 944-018 944-019 944-020

Level 2 YELLOW Level 3 BLUE Level 4 GREEN

AutoCheck3+

The AutoCheck3+ quality control system contains the following solutions:


T

Type S7335 S7345 S7355 S7365 Level 1 RED

Description

Code 944-082 944-083 944-084 944-085


Continued on next page

Level 2 YELLOW Level 3 BLUE Level 4 GREEN

14-5

14. Ordering information

ABL800 FLEX Operator's Manual

Quality control, Continued


T T

AutoCheck5+

The AutoCheck5+ quality control system contains the following solutions:


T

Type S7735 S7745 S7755 S7765 Level 1 RED

Description

Code 944-074 944-075 944-076 944-077

Level 2 YELLOW Level 3 BLUE Level 4 GREEN

AutoCheck6+

The AutoCheck6+ quality control system contains the following solutions:


T

Type S7835 S7845 S7855 S7865 Level 1 RED

Description

Code 944-094 944-095 944-096 944-097

Level 2 YELLOW Level 3 BLUE Level 4 GREEN

Range+ QUALICHECK

The Range+ QUALICHECK quality control system contains the following items: Type S7930 S7940 S7950 H700 Description Level 1, box of 30 ampoules Level 2, box of 30 ampoules Level 3, box of 30 ampoules ABL700 QC Adaptor Ampoule Opener Code 944-151 944-152 944-153 924-036 920-712

Quality Control Documentation

Type

Description Quality Control Systems Reference Manual QUALICHECK Short-form Instructions Logbook Control Binder, QUALICHECK+ Logbook Control Charts, Blood Gas Logbook Control Charts, Electrolytes and Metabolites Control Charts Oximetry

Code 989-046 983-313 984-129 989-417 989-418 989-419

14-6

ABL800 FLEX Operator's Manual

14. Ordering information

Sampling devices
Arterial blood gas samplers The following sampling devices are available: Item PICO50, 2 mL aspirator (100 units) PICO70 w/o needle (100 units) PICO70 w/o needle (w/o needle cube) (100 units) PICO70 22G 1 (100 units) PICO70 22G 1 1/4 (100 units) PICO70 23G 5/8 (100 units) PICO70 23G 1 (100 units) PICO70 23G 1 1/4 (100 units) PICO70 23G 5/8 (w/o needle cube) (100 units) PICO70 25G 5/8 (100 units) PICO70 22G 1 (w/o needle cube) (100 units) safePICO70 22G 1" with needle shield device (100 units)
T T

Code no. 956-552 956-518 956-519 956-522 956-525 956-529 956-533 956-534 956-546 956-547 956-563 956-608 956-609 956-624

Type PICO50 PICO70 PICO70 PICO70 PICO70 PICO70 PICO70 PICO70 PICO70 PICO70 PICO70 PICO70 PICO70 PICO70

safePICO70 23G 5/8" with needle shield device (100 units)


T T

safePICO70 22G 1" with needle shield device (100 units)


T T

Capillary samplers, glass

Item CLINITUBES with balanced heparin, 125 L including mixing wire and capillary caps (5 50 units) CLINITUBES with balanced heparin, 100 L including mixing wire and capillary caps (575 units) CLINITUBES with sodium heparin, 55 L including mixing wire and capillary caps (575 units)

Code no. 942-880

Type D957G-70125x5 D957G-70100x5 D941G-24055x5

942-878

942-885

Continued on next page

14-7

14. Ordering information

ABL800 FLEX Operator's Manual

Sampling devices, Continued


T T

Capillary samplers, plastic

Item CLINITUBES with balanced heparin, 125 L including mixing wire and capillary caps (500 units) CLINITUBES with balanced heparin, 100 L including mixing wire and capillary caps (1 250 units) CLINITUBES with balanced heparin, 70 L including mixing wire and capillary caps (1250 units)

Code no. 942-893 942-892

Type D957P-70125x1 D957P-70100x1 D957P-7070x1

942-898

Sampling devices for FLEXQ

The following sampling devices are available on order for the ABL800 FLEX analyzers with FLEXQ module: Item * safePICO Self-fill without needle with safeTIPCAP, 0.7-1.5 mL (100 units)
T T T T

Code No. 956-610 956-611 956-612 956-613 956-614

safePICO Self-fill with safeTIPCAP, without needle shield device, 22G x 32 mm, 0.7-1.5 mL (100 units)
T T

safePICO Self-fill with safeTIPCAP, without needle shield device, 23G x 16 mm, 0.7-1.5 mL (100 units)
T T T T

safePICO Self-fill with safeTIPCAP, without needle shield device, 22G x 25 mm, 0.7-1.5 mL (100 units)
T T T T

* safePICO Self-fill with needle shield device and safeTIPCAP, 22G x 32 mm, 0.7-1.5 mL (100 units)
T T T T

* safePICO Self-fill with needle shield device and safeTIPCAP, 23G x 16 mm, 0.7-1.5 mL (100 units)
T T T T

956-615 956-616 956-620 956-622 956-623

* safePICO Self-fill with needle shield device and safeTIPCAP, 23G x 25 mm, 0.7-1.5 mL (100 units)
T T T T

* safePICO Self-fill with needle shield device and safeTIPCAP, 22G x 25 mm, 0.7-1.5 mL (100 units)
T T T T

* safePICO Aspirating without needle with safeTIPCAP, 0.7-1.7 mL (100 units)


T T T T

* safePICO Self-fill with needle shield device and safeTIPCAP, 25G x 16 mm, 0.7-1.5 mL (100 units)
T T T T

* Without a needle cube.

14-8

15. Overview
Introduction

Radiometer settings

This chapter describes Radiometer settings, contents of various Ini files and calibration verification procedure. This chapter contains the following topics. Radiometer default settings.............................................................................. 15-2 Contents of setup settings................................................................................. 15-15 Calibration verification .................................................................................... 15-18 Interfacing facilities ......................................................................................... 15-20

Contents

15. Radiometer settings

ABL800 FLEX Operator's Manual

Radiometer default settings


Radiometer default setup screen Press Menu - Utilities - Disk Functions - Restore Default Setup:

You can select the parts of the Setup to be set back to Radiometer defaults.

Analyzer Security

Default settings for General Security are as follows: Item Enable Centralized User Management Authenticate operator by Allow anonymous use Logoff Time Profile for anonymous user Default setting Off Logon-barcode as primary Yes 3 min User
Continued on next page

15-2

ABL800 FLEX Operator's Manual

15. Radiometer settings

Radiometer default settings, Continued


Default settings for Access Profiles are as follows: A User Supervisor Manager Service techn. Guest Custom 1 Custom 2 Custom 3 Remote operator X X X X X X X X X B X X X X C X X X X X X X D E (X) X X X (X) (X) (X) (X) X X X X X F X X X X X X X X G H I X X X X X X X J

A = Perform measurement B = Perform calibration C = Perform replacements D = Perform Disk Functions E = View Data Logs

F = Edit data in logs G = Enter Setup Programs H = Enter Service Programs I = Start AutoCheck J = Approve results

Columns D, E, G and H are controlled via the Menu and Button Configuration screen settings, not via the check buttons on the Access Profiles screen. (X) means restricted access to data logs: User can view the logs, but there is no access to the archived data logs. Guest and Custom can view Patient Results Log and Quality Control Log.
Continued on next page

15-3

15. Radiometer settings

ABL800 FLEX Operator's Manual

Radiometer default settings, Continued


Analysis Setup defaults The following default measuring modes are available: Analyzer ABL837/27 /17 Modes Syringe Default button Syringe S250 L; Syringe S85 L; Syringe Pleura pH Ampoule QC; Expired air Capillary C125 L; Capillary C55 L; Capillary C35 L MET; Capillary C35 L OXI Syringe S195 L; Syringe S95 L; Syringe Pleura pH Ampoule-QC

Capillary

ABL835/25 /15

Syringe:

Capillary: Capillary FLEXMODE; Capillary C95 L; Capillary C55 L pH+BG+OXI; Capillary C35 L MET Capillary C35 L OXI ABL830/20 /10/10 BG only ABL805 Syringe: Syringe S85 L; Syringe Pleura pH Capillary: Ampoule-QC Capillary FLEXMODE; Capillary C55 L Capillary C35 L OXI Syringe: Syringe S165 L; Syringe S95 L; Syringe Pleura pH Ampoule-QC

Capillary: Capillary FLEXMODE; Capillary C95 L; Capillary C55 L pH+BG; Capillary C35 L MET ABL800* N/A N/A

* Not available at the time of the release. All user-defined modes and Other Fluids mode are deleted. All modes are set to Radiometer default layout. Reference ranges and critical limits are deleted. Parameter profile: All parameters are selected. Dynamic Parameters: Off. Sample Pre-registration settings: - Interpret barcode input as: Sampler ID - Confirm pre-registered data : On. - Included fields: All fields On. Sample Logistics Setup settings: - Batch mode: Off - Sample age: On (30 min for all parameters) - Primary patient ID field in "FLEXQ" box: Patient Last Name.
Continued on next page

15-4

ABL800 FLEX Operator's Manual

15. Radiometer settings

Radiometer default settings, Continued


Analysis Setup defaults (continued) Available layouts: -R- Default (Default) GFR not activated by default. Patient ID layout settings included in the -R- Default Layout: - Patient ID - Patient Last Name - Patient First Name - Sample type - Temp. C Patient Result settings included in the -R- Default Layout (bold text = a new title; [xxxxxx] = the reference range for a parameter): Blood Gas Values pH [xxxxxx] pCO2 pO2 [xxxxxx] [xxxxxx]

< New Line > Oximetry Values ctHb [xxxxxx] sO2 FO2Hb FCOHb FHHb FMetHb FHbF [xxxxxx] [xxxxxx] [xxxxxx] [xxxxxx] [xxxxxx] [xxxxxx]

< New Line > Electrolyte Values cK+ [xxxxxx] cNa+ cCa
2+ -

[xxxxxx] [xxxxxx] [xxxxxx]

cCl

< New Line > Metabolite Values cGlu [xxxxxx] cLac cCrea ctBil < New Page >
Continued on next page

[xxxxxx] [xxxxxx] [xxxxxx]

15-5

15. Radiometer settings

ABL800 FLEX Operator's Manual

Radiometer default settings, Continued


Analysis Setup defaults (continued) Temperature Corrected Values pH(T) pCO2(T) pO2(T) < New Group > Oxygen Status ctO2 p50 < New Line > Acid Base Status cBase(Ecf)
cHCO 3 ( P,st )

Calibration Setup defaults

Default settings for Drift tolerances are as follows:


Parameter Drift 1 Drift 2 Units

pH pCO2 pO2 ctHb cK


+ +

0.020 2.5 6 10 0.2 3 0.05 2 0.5 0.2 15

0.020 5 6 N/A 1.5 1 0.2 3 N/A N/A 15 mmHg mmHg pA mmol/L mmol/L mmol/L mmol/L mmol/L mmol/L
mol/L

cNa

cCa2+ cClcGlu cLac cCrea

Continued on next page

15-6

ABL800 FLEX Operator's Manual

15. Radiometer settings

Radiometer default settings, Continued


Calibration Setup defaults (continued)

Default settings for Calibration Schedule are as follows:


Activity Default setting

2 Point Cal 1 Point Cal 1 Point pH/BG calibration (for the USA only) tHb Cal Cleaning Start time for 1- and 2-point calibration and cleaning Activity after measurement

8 hours 4 hours 2 hours 3 months 8 hours 00:00 None

Quality Control Default settings for Quality Control Statistics are as follows: Setup defaults Item Default setting

Statistics factor Remind to print statistics each month Remind to export WDC data each month

1.5 No No

Default settings for Quality Control Mandatory Input are as follows:


Item Default Setting

Mandatory temperature Default temperature

No 25 oC

QC Schedule settings are deleted. AutoCheck Setup: replacement warning before carousel empty: No. Westgard Rules: all rules are Off and the use of Westgard Rules is disabled. Rilibk Ranges: the use of Rilibk rules is disabled.
Continued on next page

15-7

15. Radiometer settings

ABL800 FLEX Operator's Manual

Radiometer default settings, Continued


Replacement Setup defaults

The default settings for Replacements are as follows:


Item Default setting

pCO2 Electrode Membrane pO2 Electrode Membrane K Electrode Membrane Na Electrode Membrane Ca Electrode Membrane Cl Electrode Membrane Glucose Electrode Membrane Lactate Electrode Membrane Crea A and Crea B Electrode Membranes Reference Electrode Membrane K Electrode Na Electrode Ca Electrode Cl Electrode Glucose Electrode Lactate Electrode Crea A Electrode Crea B Electrode Reference Electrode pH Electrode pCO2 Electrode pO2 Electrode Inlet Gasket Rubber Tube for Inlet Pump Tube, Electrode Modules Pump Tube, Waste Pump Tube, Solutions Fan Filter

Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never Never
Continued on next page

15-8

ABL800 FLEX Operator's Manual

15. Radiometer settings

Radiometer default settings, Continued


Replacement Setup defaults (continued) Solution warning settings are 25 %. Reagents Expiration warning setting: 24 hours. User Activities settings are blank. AutoCheck Warning: No.

General Setup defaults

The default settings for Parameter Setup are as follows:


Parameter Enabled/ Locked Repression Offset Slope Units Out of Range Suppression

pH pCO2 pO2 ctHb FHHb FO2Hb FHbF sO2 FCOHb FMetHb ctBil cK
+

Yes/No Yes/No Yes/No Yes/No Yes/No Yes/No Yes/No Yes/No Yes/No Yes/No Yes/No Yes/No Yes/No Yes/No Yes/No Yes/No Yes/No Yes/No

No No No No No No No No No No No No No No No No No No

0.000 0.0 0.0 N/A N/A N/A 0 0.0 0.0 0.0 0 0.0 0 0.00 0 0.0 0.0 0.0

1.000 1.000 1.000 1.000 N/A N/A 1.000 1.000 N/A N/A 1.000 1.000 1.000 1.000 1.000 1.000 1.000 1.000 mmHg mmHg g/dL % % % % % %
mol/L

No No Yes No No No Yes

mmol/L mmol/L mmol/L mmol/L mmol/L mmol/L


mol/L
Continued on next page

cNa+ cCa2+ cCl cGlu CLac cCrea

15-9

15. Radiometer settings

ABL800 FLEX Operator's Manual

Radiometer default settings, Continued


General Setup defaults (continued)

Default settings for Units are as follows:


Parameter Unit

Pressures cCrea ctBil ctHb Oximetry fractions Saturation Gas fractions FO2(I) Sensitivity pH/cK+/cNa+/cCa2+/cCl Hct pO2(a/A) FShunt RI cK+/cNa+/cCa2+/cCl cGlu/cLac Temperatures ctO2/ctCO2 DO2 V O2 Age Weight Height Altitude Birth weight pO2 sensitivity cCrea sensitivity 1 cCrea sensitivity 2 and 3

mmHg
mol/L mol/L

g/dL % % % % % % % % % mmol/L mmol/L


o

Vol % mL/min mL/min years kg m m g PA/mmHg pA/M %

Continued on next page

15-10

ABL800 FLEX Operator's Manual

15. Radiometer settings

Radiometer default settings, Continued


General Setup defaults (continued)

Default settings for User-defined Patient Data Items are as follows:


Name Type Unit Decimals

Spontaneous RR Set RR Vt Ve Peak Flow Liter Flow Ti PEEP Pressure Support CPAP CMV SIMV Flow-by HFV I:E Ratio Wave ICD9 Code Oxygen Device 1 Oxygen Device 2 Diagnostic Code

Numerical Numerical Numerical Numerical Numerical Numerical Numerical Numerical Numerical Numerical Numerical Numerical Numerical Numerical Numerical Numerical Numerical Numerical Numerical Numerical

b/min b/min L L L/min L/min seconds cmH2O cmH2O cmH2O Rate Rate L/min Rate

1 1 2 2 1 1 1 1 1 1 1 1 1 1 2 None None None None None

Default settings for User-defined Notes: No notes defined. Default settings for Language: English. Default settings for Acoustic Signals are as follows:
Event Default setting

Value exceeds critical limits Close inlet Result is ready Inlet is open too long Beep before AutoCheck opens

No Yes Yes Yes Yes


Continued on next page

15-11

15. Radiometer settings

ABL800 FLEX Operator's Manual

Radiometer default settings, Continued


General Setup defaults (continued)
Default settings for Corrective Actions are as follows: Event Calibration error(s) present Calibration schedule reminder(s) QC error(s) present QC schedule reminders Replacement schedule reminders System message(s) present User Activity Reminder(s) AutoCheck QC error(s) present AutoCheck scheduled level missing Default setting ? on specific parameters No setting No setting No setting No setting No setting No setting No setting No setting Traffic light Yellow Yellow Green Yellow Yellow Yellow Yellow N/A Green

Default settings for Miscellaneous Setup are as follows: Event Analyzer locked Use safeTIPCAP Enable Estimated Derived Parameters Fixed pO2/pCO2 decimals Enable General Barcode Support Enable Patient Result Approval Apply Parameter Corrections to QC Log All Measurement Activities Show Parameter Bar Auto temp unit conversion HbF correction Analyzer Message Screen saver Default setting Not set Yes No No Yes No Yes No Yes No "Enabled for levels > 20 %" (Blank) Yes, Minutes to wait when idle: 5

Default settings for Automatic Printing are as follows: Item Patient results QC results Calibration results Activity log messages FLEXQ registration receipt Message level Number of copies Default setting On Off Off Off Off User 1
Continued on next page

15-12

ABL800 FLEX Operator's Manual

15. Radiometer settings

Radiometer default settings, Continued


General Setup defaults (continued)

Default settings for Printer Setup are as follows:


Item Default setting

Installed printers

Internal Printer (added printers are not deleted)

Select printer dialogue

Off

Default settings for Automatic Backup are as follows:


Item Default setting

Auto Backup

Off

Default settings for Automatic Archiving are as follows:


Item Default setting

Patient report log Calibration log Quality control log Activity log Store archives on the analyzer

On On On On On

Default settings for Communication Setup are as follows:


Item Default setting

RADIANCE Connection LIS/HIS Connection QA Portal Automatic Data Request Automatic Data Transmission Patient Lookup Setup Remote Support

Off None Off "When entering sampler ID" on Patient results, Calibration results, QC results, Activity log messages Local database Enable remote access
Continued on next page

15-13

15. Radiometer settings

ABL800 FLEX Operator's Manual

Radiometer default settings, Continued


Setups without Radiometer settings

The following setups have no Radiometer settings:


Environment setup time and date setup analyzer identification setup

15-14

ABL800 FLEX Operator's Manual

15. Radiometer settings

Contents of setup settings


Groups of Setup The setup is divided in the following groups of settings: settings Parameters General Schedules, etc.

You can restore the Radiometer Default Setup or a setup you have customized (Customer Setup) and saved. Selecting or deselecting items in the setup see Loading/restoring Setup in chapter 8. Each setup group of settings is described in this section.
Parameters

The following settings (i.e. screens and their data) will be restored in the Parameters group:
Item Setup (screens) Syringe Mode Capillary mode Parameter Setup (Offset and slope only)

Sample Modes

Parameter Units Corrections Critical Ranges

Units Setup Parameter Setup (Repression and Out-of-range suppression only)


Reference ranges, Critical limits Age groups
Continued on next page

15-15

15. Radiometer settings

ABL800 FLEX Operator's Manual

Contents of setup settings, Continued


General

The following settings (i.e. screens and their data) will be restored in the General group:
Item Setup (screens) Corrective Actions Acoustic Signals Low Level Warning Selected Language Printer path Reagents Expiration Warning RADIANCE Connection LIS/HIS Connection QA Portal Automatic Data Transmission Automatic Data Request Patient Lookup Setup Operators and Passwords (Logon Protection Level and Logoff Time only) Miscellaneous Setup (All, except Analyzer Locked) Automatic Printing Automatic Archiving Automatic Backup Save Setup (Destination) Load Setup (Source) Backup All Data (Destination) Export Data Logs (Destination) Function: External keyboard enabling Function: Enable remote access when operator is logged on QC Statistics Setup QC Input Setup AutoCheck Setup Westgard Rules (enable Westgard Rules) Rilibk Ranges (enable Rilibk rules) Printer Setup (Show list of printers)
Continued on next page

Analyzer Setup

Ini File

Ini Settings and Communications

15-16

ABL800 FLEX Operator's Manual

15. Radiometer settings

Contents of setup settings, Continued


General (continued) Item Setup (screens) Patient Report Setup Patient ID Layout Patient Result Layout User-defined Data Items The width of the following column setups: Patient Results Log; Patient Lookup; Patient Profiles Log; QC Log; Calibration Log; System Messages; Replacement Schedule

Layouts

Operators

General Security Operators and Passwords Access Profiles User-defined Notes

Pre-def. Notes
Schedules, etc.

The following ini files (i.e. screens and their data) will be restored in the Schedules, etc. group:
Item Setup (screens)

QC Schedule

QC Schedule (QC schedule is restored for the slots with the control solutions installed in them. The schedule follows the slots, not the QC levels). Rilibk rules are restored
Calibration Schedule (minus tHb Cal and the start time) Calibration Drift Tolerances

Rilibk Ranges Wetsection Setup

Replacement Schedule User Activities Schedule

Replacement Schedule
User Activities Edit User Activities

15-17

15. Radiometer settings

ABL800 FLEX Operator's Manual

Calibration verification
Purpose

Regulations in some countries require verification of the reportable ranges selected for the measured parameters, and a special feature described below is provided to assist in the collection of this data. Once the reportable ranges for the measured parameters are established, the limits may be entered in the Reportable Ranges program described in Analysis Setup, chapter 3.

Preparatory steps

The procedure given below is only a suggestion to easily identify your specific layout and measuring mode.
Step 1. 2. Action

Make a new report layout called, e.g. "Cal. Verification", based on the Radiometer default layout see Patient Report Setup in chapter 3. Press Patient ID layout and deselect all items except for the Patient ID and Sample type. Set default sample type to Cal. Verification.

3. 4.

Press Edit Patient Results Layout and deselect all parameters except for the measured parameters. Deselect range checks for all parameters. In the Analysis Setup - Syringe Mode Setup dedicate one unoccupied button to this specific mode (call it, e.g. Cal. Verification) using Edit name.

Select desired parameters (press Parameters) and layout (press Layout) that you have made in step 1 above.
Continued on next page

15-18

ABL800 FLEX Operator's Manual

15. Radiometer settings

Calibration verification, Continued


Preparatory steps (continued) Step 5. 6. Action

Condition calibration verification kit see Preparing a control solution in chapter 5 of this manual. Place the ampoule in the ampoule opener and break off the ampoule neck. Place the ampoule fully into the H700 Adapter.

Verification procedure

To perform a verification measurement, do the following:


Step 1. 2. 3. 4. 5. 6. Action

Check that the analyzer is in the Ready mode. Open the syringe inlet flap and place the adapter tip into the syringe inlet. Select the Cal. verification mode on the screen and press Start. Press Aspirate to start sample aspiration. When prompted by the analyzer, remove the sampling device and close the inlet flap. Enter the information on the Patient Identification screen.

Use a dedicated identifier for each solution (e.g. Level 1) as Patient ID.
7.

After all measurements on the calibration verification solutions have been completed, review the data by filtering it in the Patient Results log or transfer the data to a data analysis program see Exporting Data Logs in chapter 8: Disk Functions. After the new reportable ranges have been determined, enter any changes in Setup program: Reportable Ranges.

8.

Once the first measurement in this mode has been performed, the mode is available for the next seven days. During this time you can verify your reportable ranges according to the procedure used in your institution. After seven days the mode will not be available for the next 14 days.

15-19

15. Radiometer settings

ABL800 FLEX Operator's Manual

Interfacing facilities
Connecting a mouse

A mouse connected to the analyzer may be used to activate all the analyzers screen functions instead of touching the screen. This mouse facility is first and foremost intended for service technicians and not for the daily use of the analyzer. A standard PS/2 port mouse is the sole item that is required for connection to the analyzer. To connect the mouse to the analyzer, do the following:
Step 1. 2. 3. Action

Switch off the analyzer. Connect the mouse to the mouse port at the rear of the analyzer. Switch on the analyzer.

Connecting an alphanumeric keyboard

An external alphanumeric keyboard may be used instead of the on-screen keyboard to enter data. However, to select individual buttons on the analyzers screen, a mouse must be used or the operator must touch the screen. An IBM enhanced personal computer keyboard is the sole item that is required for connection to the analyzer.
NOTICE: The keyboard layout must correspond to the language version used by the analyzer.

The transmission format for an alphanumeric keyboard is as follows:


bit 11 5V 0V 8 data bits (ASCII) Stop bit Parity bit (odd) Start bit bit 1

Pin designation on the cable connector are as follows: Pin 1 Clock in/out Pin 2 Data in/out Pin 3 Not connected Pin 4 Ground Pin 5 +5 V Pin 6 Not connected
Continued on next page

15-20

ABL800 FLEX Operator's Manual

15. Radiometer settings

Interfacing facilities, Continued


Connecting an alphanumeric keyboard (continued)

To connect the keyboard to the analyzer, do the following:


Step 1. 2. 3. Action

Switch off the analyzer. Connect the keyboard to the keyboard port at the rear of the analyzer. Switch on the analyzer.

Connecting to a network

Many hospitals utilize a computer-controlled information system such as the Hospital Information System (HIS) or the Laboratory Information System (LIS). Connecting the analyzer to such an information system via a network enables the user to exercise greater control over the amount of patient data circulating within the hospital. The types of information that can be communicated via a network between the central computer controlling the information system and the analyzer are:
Patient results Quality control results Calibration data System messages

Recommendations:
1. 2.

Use a shielded data cable with an RJ45 connector to connect the analyzer to a network. The analyzer is first connected to the computer controlling the information system via one of the following two interfaces:
A serial line (RS232 Interface) An Ethernet interface (TCP/IP)

3.

Once the analyzer has been physically connected to the network, one of the protocols stated below is used for communication with the central computer.
ASTM HL7 POCT1-A

For further information refer to the Communication Protocol Specifications for Radiometer Products (code 989-329). Radiometer recommends that a qualified service technician carries out connection of the analyzer to a network.
External barcode reader

An external barcode reader can be connected and used side by side with the builtin barcode reader contact your Radiometer service representative.

15-21

15. Radiometer settings

ABL800 FLEX Operator's Manual

15-22

ABL800 FLEX Operators Manual

Index

Index

1 1- or 2-point calibration results.......................................................................................... 6-8 A Access to analyzer menu.................................................................................................. 2-27 Access to Data Logs .......................................................................................................... 9-2 Accessing Setup programs................................................................................................. 3-3 Acid-Base Chart............................................................................................................... 4-23 Acoustic signal setup ....................................................................................................... 3-65 Activity Log..................................................................................................................... 9-20 Adding a Message ....................................................................................................... 9-20 Filter function .............................................................................................................. 9-21 Adding a new patient profile.............................................................................................. 9-9 Adding a user activity ...................................................................................................... 3-50 Adjusting low level warning............................................................................................ 3-52 Allowed actions at analyzer startup ................................................................................. 2-24 Analysis Setup ................................................................................................................. 3-13 Analysis Setup Defaults................................................................................................... 15-4 Analyzer - Front................................................................................................................. 2-2 Analyzer - Rear.................................................................................................................. 2-4 Analyzer ID ..................................................................................................................... 3-62 Analyzer Message............................................................................................................ 3-88 Analyzer messages........................................................................................................... 11-7 Analyzer Requirements.................................................................................................. 13-13 Analyzer security ............................................................................................................... 3-6 Analyzer Settings Setup................................................................................................... 3-62 Analyzer Specifications ................................................................................................. 13-14 Analyzer Status ................................................................................................................ 2-28 Analyzer status at a glance............................................................................................... 2-28 Approvals and Patents ................................................................................................... 13-16 Archived Data Logs ..................................................................................................9-3, 9-24 Arterial blood samplers.................................................................................................... 12-8 AutoCheck Measurement................................................................................................... 5-7 AutoCheck Module.......................................................................................................... 2-14 AutoCheck Module Specifications ................................................................................ 13-15 AutoCheck Status ...................................................................................................2-33, 5-15 AutoCheck Warning program.......................................................................................... 3-54 Automatic Archiving Setup ............................................................................................. 3-81 Automatic Auxiliary Programs ........................................................................................ 7-20 Automatic Backup Setup ................................................................................................. 3-80 Automatic Data Request .................................................................................................. 3-74 Automatic Data Transmission Setup................................................................................ 3-75 Automatic Printing........................................................................................................... 3-77 Auxiliary programs Interrupting a program................................................................................................. 7-20 Starting a program ....................................................................................................... 7-20 Available Calibration Schedule Options.......................................................................... 3-34 Available modes and parameters ....................................................................................... 4-2 B Backing up All Data .......................................................................................................... 8-6 Barcode Reader................................................................................................................ 2-13 Batch measurement.......................................................................................................... 4-10

Index

ABL800 FLEX Operators Manual

Batch mode (automatic sample handling)........................................................................ 3-23 Before measurement on AutoCheck control solution ........................................................ 5-5 Before measurement on QUALICHECK+ control solution .............................................. 5-4 Before using the FLEXQ ................................................................................................... 4-7 C Calculation of FShunt and ctO2(a-v) ............................................................................... 4-25 Calibration Filter function .............................................................................................................. 9-18 Trend function ............................................................................................................. 9-19 Calibration Drift Tolerances ............................................................................................ 3-32 Calibration Log ................................................................................................................ 9-17 Calibration programs ......................................................................................................... 6-2 Calibration Result .............................................................................................................. 6-8 Calibration Result Messages............................................................................................ 6-10 Calibration Schedule........................................................................................................ 3-33 Calibration Setup ............................................................................................................. 3-32 Calibration Setup Defaults............................................................................................... 15-6 Calibration Status............................................................................................................. 2-29 Calibrations during 24 hours after Restart ......................................................................... 6-4 Calling an unscheduled calibration .................................................................................... 6-4 Capillary samples............................................................................................................. 4-13 Causes of errors in preanalytical phase............................................................................ 12-2 CE-mark......................................................................................................................... 13-16 Cleaning the analyzer exterior .................................................................................................... 7-24 the analyzer touch screen............................................................................................. 7-24 the inlet flaps and inlet gasket ..................................................................................... 7-23 the measuring chambers .............................................................................................. 7-23 Cleaning the Analyzer ..................................................................................................... 7-23 Cleaning the FLEXQ sampler tray cover......................................................................... 7-24 CLINITUBES .................................................................................................................. 12-8 Communication Ports ...................................................................................................... 2-11 Communications Setup .................................................................................................... 3-68 Complete menu ................................................................................................................ 2-26 Conditions and correction actions.................................................................................... 3-83 Connecting a mouse....................................................................................................... 15-20 Connecting an alphanumeric keyboard.......................................................................... 15-20 Connecting to a network ................................................................................................ 15-21 Converting an Archive into .csv Format .......................................................................... 9-25 Corrective Actions program............................................................................................. 3-83 Creating a backup of all data ............................................................................................. 8-6 Creating a layout.............................................................................................................. 3-26 Creating a WDC Report..................................................................................................... 8-4 D Data storage options........................................................................................................... 8-3 Decontamination program ............................................................................................... 7-21 Definition NOTE ............................................................................................................................ 1-7 WARNING/CAUTION ................................................................................................. 1-7 Deleting a Patient Profile ................................................................................................... 9-9 Deleting a user activity .................................................................................................... 3-51 Deleting an archive .......................................................................................................... 8-12 Description of the Fluidic Module ................................................................................. 11-82 Diagram of fluid transport system ................................................................................. 11-81 Diagram of the Fluidic Module...................................................................................... 11-81 Disabled Versus Deselected Parameter............................................................................ 3-15 Disk Functions menu ......................................................................................................... 8-3 Disk Functions programs ................................................................................................... 8-2

ABL800 FLEX Operators Manual

Index

Disk Functions Setup ....................................................................................................... 3-80 Disk handling rules ............................................................................................................ 8-3 E Editing a layout................................................................................................................ 3-27 Editing a Patient Profile ..................................................................................................... 9-8 Editing a user activity ...................................................................................................... 3-51 Editing patient ID layout.................................................................................................. 3-28 Editing patient result layout ............................................................................................. 3-30 Editing sample registration .............................................................................................. 4-11 Electrode Troubleshooting Procedures .......................................................................... 11-78 Electrode Updatings..................................................................................................4-19, 5-9 Electrodes and Other Status ............................................................................................. 2-33 Entering Corrections for Other Fluids Mode ................................................................... 3-17 Entering default values .................................................................................................... 3-29 Entering Information Manually ....................................................................................... 4-17 Entering Information with Bar Code Reader ................................................................... 4-17 Entering information with Barcode Reader ....................................................................... 5-8 Entering password.............................................................................................................. 3-5 Entering standard password ............................................................................................. 2-25 Entering Standby.............................................................................................................. 10-3 Entering the Fluid Replacements mode ............................................................................. 7-4 Entering the Hold Mode..................................................................................................... 7-2 Environment Setup .......................................................................................................... 3-63 Evaluating the Rinse ...................................................................................................... 11-83 Executing the Temporary Shutdown program ................................................................. 10-6 Exiting Standby................................................................................................................ 10-4 Expired air sampling...................................................................................................... 12-14 Expired calibrations ........................................................................................................... 6-5 Explanation of Corrective Actions................................................................................... 3-84 Exporting an Archive....................................................................................................... 8-11 Exporting Data Logs.......................................................................................................... 8-9 External barcode reader ................................................................................................. 15-21 F Finding a Patient Profile .................................................................................................. 9-10 FLEXMODE...............................................................................................................1-3, 4-6 FLEXQ module ................................................................................................................. 2-8 FLEXQ module specifications....................................................................................... 13-15 Fluid Transport System Description .............................................................................. 11-81 Fluid Transport Troubleshooting Procedure .................................................................. 11-74 Forced Hold Causes ......................................................................................................... 11-5 Forced Standby ................................................................................................................ 10-5 Full Waste Container ....................................................................................................... 10-5 G General Security ................................................................................................................ 3-6 General Setup Defaults .................................................................................................... 15-9 Glucose and Lactate troubleshooting trocedure............................................................. 11-79 Guidelines for handling capillary blood samples............................................................... 4-9 H Help ................................................................................................................................. 2-36 High Crea check........................................................................................................5-12, 5-6 I Importing an Archive....................................................................................................... 8-12 Importing/Exporting Archives ......................................................................................... 8-11 Indication of obstructions in the inlet ............................................................................ 11-76 Inlet calibration Error....................................................................................................... 11-5

Index

ABL800 FLEX Operators Manual

Inlet Module....................................................................................................................... 2-7 Inlet Probe Troubleshooting Procedure ......................................................................... 11-75 Inlet Troubleshooting Procedure.................................................................................... 11-76 Installation ......................................................................................................................... 3-2 Intended use ....................................................................................................................... 1-2 Interfacing Facilities ...................................................................................................... 15-20 Interrupted calibrations ...................................................................................................... 6-5 Interrupting a measurement ............................................................................................... 4-7 Introducing a blood sample.............................................................................................. 4-12 Introducing a pleura sample............................................................................................. 4-15 K Known interfering substances............................................................................................ 1-4 L Leak Detected Error......................................................................................................... 11-6 Leak Troubleshooting Procedure ................................................................................... 11-77 Limitations of use .............................................................................................................. 1-4 LIS/HIS Connection Setup .............................................................................................. 3-69 Loading/Restoring Setup ................................................................................................. 8-14 Location ......................................................................................................................... 13-13 Long-term Shutdown ....................................................................................................... 10-8 M Manual Quality Control Measurement .............................................................................. 5-6 Measurement on the S8377 Solution ................................................................................. 5-6 Measurements on animal blood ......................................................................................... 1-3 Measurements with FLEXQ ............................................................................................ 4-10 Measuring modes with FLEXQ ......................................................................................... 4-2 Measuring Section ............................................................................................................. 2-5 Menu at analyzer startup.................................................................................................. 2-24 Menu Structure ................................................................................................................ 2-24 Messages after refilling AutoCheck carousel .................................................................. 7-19 Miscellaneous Setup ........................................................................................................ 3-86 List of options.............................................................................................................. 3-86 Mixing a blood sample in a capillary .................................................................................................................. 4-9 in a syringe .................................................................................................................... 4-9 in a test tube................................................................................................................... 4-9 Moving Archive............................................................................................................... 9-25 N Names ................................................................................................................................ 1-2 O Online Aid Facilities ........................................................................................................ 2-36 Operators and Passwords ................................................................................................... 3-8 Other fluids ...................................................................................................................... 3-14 Other fluids mode .............................................................................................................. 1-3 P Parameter Setup ............................................................................................................... 3-55 Parameter status ............................................................................................................... 4-21 Parameters and Input Setup ............................................................................................. 3-55 Password ............................................................................................................................ 3-8 Patents............................................................................................................................ 13-16 Patient filter criteria ........................................................................................................... 9-5 Patient identification on the Ready screen ....................................................................... 3-24 Patient Lookup................................................................................................................. 4-18 Patient Lookup Setup....................................................................................................... 3-75

ABL800 FLEX Operators Manual

Index

Patient Profile Log ............................................................................................................. 9-7 Patient Report Layout ...................................................................................................... 3-30 Patient Reports Setup....................................................................................................... 3-26 Patient Result ................................................................................................................... 4-21 Patient result approval ..................................................................................................... 4-22 Patient result audit trail .................................................................................................... 4-24 Patient Result Messages................................................................................................... 4-26 Patient Results Log ............................................................................................................ 9-4 Patient Trend Function....................................................................................................... 9-6 pCO2 and pO2 troubleshooting procedure...................................................................... 11-78 Pending calibrations........................................................................................................... 6-5 Performing a tHb Calibration............................................................................................. 6-6 pH, K, Na, Ca, and Cl troubleshooting procedure ......................................................... 11-78 Pleural liquid sampling .................................................................................................. 12-13 Preparation prior to arterial/venous sampling .................................................................. 12-5 Preparation prior to capillary sampling............................................................................ 12-7 Preparing a Control Solution ............................................................................................. 5-4 Preparing the Calibration and Cleaning Solution............................................................. 7-16 Print Setup ......................................................................................................................... 3-5 Printer Setup .................................................................................................................... 3-78 Printing Monthly QC Statistics ........................................................................................ 9-14 Printing QC statistics ....................................................................................................... 9-14 Protein Removal Program................................................................................................ 7-21 Pump Troubleshooting Procedure.................................................................................. 11-80 Q QA Portal ......................................................................................................................... 3-72 QC Input Setup program.................................................................................................. 3-39 QC Ranges program......................................................................................................... 3-37 QC Shedule program ....................................................................................................... 3-40 QC Solutions program ..................................................................................................... 3-35 QC Statistics program...................................................................................................... 3-42 Quality Control Entering Information Manually ..................................................................................... 5-8 filter criteria................................................................................................................. 9-12 General information....................................................................................................... 5-2 Glossary of terms........................................................................................................... 5-3 Quality Control Frequency ................................................................................................ 5-2 Quality Control Log......................................................................................................... 9-11 Quality control plot.......................................................................................................... 9-15 Quality Control Plot................................................................................................5-13, 9-15 Quality Control Result status markings ............................................................................................................ 5-10 Quality Control Result Messages..................................................................................... 5-15 Quality Control Setup ...................................................................................................... 3-35 Quality Control Setup Defaults........................................................................................ 15-7 Quality Control Status ..................................................................................................... 2-31 R RADIANCE Browser (optional)...................................................................................... 9-26 RADIANCE Communication Setup ................................................................................ 3-68 Reading In A Barcode...................................................................................................... 2-13 Ready screen.................................................................................................................... 2-15 Reagents Status ................................................................................................................ 2-32 Recalling patient ID ......................................................................................................... 4-24 Recalling Quality Control Identification.......................................................................... 5-11 Recommended Replacement Intervals............................................................................. 3-49 Recording Replacements in Hold Mode ............................................................................ 7-3 Recording replacements in the Fluid Replacements mode ................................................ 7-4 Reference electrode troubleshooting procedure............................................................. 11-78

Index

ABL800 FLEX Operators Manual

Refilling the AutoCheck Carousel ................................................................................... 7-19 Remebraning an Electrode................................................................................................. 7-6 Remote Support ............................................................................................................... 3-76 Replacement Log ....................................................................................................9-22, 9-23 Replacement Setup .......................................................................................................... 3-48 Replacement Setup Defaults ............................................................................................ 15-8 Replacing an Electrode ...................................................................................................... 7-7 Replacing gas cylinders ................................................................................................... 7-17 Replacing Pump Tubes ...................................................................................................... 7-9 Replacing solutions.......................................................................................................... 7-17 Replacing the electrode pump tubes .................................................................................. 7-9 Replacing the Fan Filter................................................................................................... 7-14 Replacing the inlet gasket unit......................................................................................... 7-12 Replacing the Inlet Probe................................................................................................. 7-12 Replacing the Printer Paper ............................................................................................. 7-14 Replacing the Solutions Pump Tube................................................................................ 7-10 Replacing the Waste Container........................................................................................ 7-14 Replacing the Waste Pump Tubes ................................................................................... 7-11 Requirement to the operator............................................................................................... 1-3 Restart after replacements.................................................................................................. 7-5 Restarting the Analyzer ................................................................................................... 10-6 Restoring All Data ............................................................................................................. 8-8 Retrieving patient result................................................................................................... 4-21 Rinse Process ................................................................................................................. 11-82 S Sample age....................................................................................................................... 3-24 Sample Counter................................................................................................................ 2-39 Sample Logistics Setup.................................................................................................... 3-23 Sample Pre-registration.................................................................................................... 3-22 Samples Preparation Prior to Analysis..................................................................................... 12-10 Sampling devices ....................................................................................................12-8, 14-7 Sampling devices for FLEXQ.......................................................................................... 14-8 Sampling with a standard syringe .................................................................................. 12-12 Sampling with a test tube............................................................................................... 12-13 Saving Setup .................................................................................................................... 8-13 Scheduled AutoCheck measurement ................................................................................. 5-7 Screen elements buttons ......................................................................................................................... 2-22 center section............................................................................................................... 2-17 check buttons............................................................................................................... 2-20 information bar and icons ............................................................................................ 2-22 parameter bar............................................................................................................... 2-16 screen keyboard ........................................................................................................... 2-21 screen keypad .............................................................................................................. 2-20 status bar...................................................................................................................... 2-15 text boxes..................................................................................................................... 2-19 time bar........................................................................................................................ 2-16 top section.................................................................................................................... 2-15 Screen saver ..................................................................................................................... 3-88 Selecting Sex .................................................................................................................. 3-19 Selecting a Report Layout.......................................................................................3-18, 4-19 Selecting access to menus................................................................................................ 3-10 Selecting Archive Stored on the Analyzer ....................................................................... 9-24 Selecting HbF Correction ................................................................................................ 3-16 Selecting HbF Correction option ..................................................................................... 3-87 Selecting Options............................................................................................................. 3-87 Selecting Parameter Profile.............................................................................................. 3-15 Selecting parameters after a measurement....................................................................... 4-20

ABL800 FLEX Operators Manual

Index

Selecting Sample Type .................................................................................................... 3-19 Selection of Anticoagulant.............................................................................................. 12-4 Set Language ................................................................................................................... 3-66 Setting Reference and Critical Limits .............................................................................. 3-20 Setup Menu Structure ........................................................................................................ 3-3 Shutdown programs ......................................................................................................... 10-2 Single measurement from the FLEXQ sampler tray........................................................ 4-10 Solution Level Forecast program..................................................................................... 3-53 Solution Warning program .............................................................................................. 3-52 Specifications Derived Parameters ..................................................................................................... 13-6 Input Parameters.......................................................................................................... 13-5 Measured Parameters................................................................................................... 13-2 Standby Mode.................................................................................................................. 10-3 Storage and preparation prior to analysis......................................................................... 12-9 Storing a WDC report ........................................................................................................ 8-4 Storing the analyzer ......................................................................................................... 10-9 Storing the setup .............................................................................................................. 8-13 Syringe Mode Setup......................................................................................................... 3-13 Syringe samples ............................................................................................................... 4-12 T Temperature Corrections (QC) ........................................................................................ 5-11 Temporary Shutdown ...................................................................................................... 10-6 Test tube samples............................................................................................................. 4-14 tHb Calibration check.............................................................................................................................. 6-7 tHb calibration result ......................................................................................................... 6-9 The shutdown procedure.................................................................................................. 10-8 Thermal Printer ................................................................................................................ 2-10 Time/Date Setup .............................................................................................................. 3-65 To adjust the internal barometer ...................................................................................... 3-63 To enter the altitude ......................................................................................................... 3-63 To enter the ambient temperature .................................................................................... 3-64 Transporting the analyzer ................................................................................................ 10-9 Tutorials........................................................................................................................... 2-37 U Units Setup....................................................................................................................... 3-57 Unknown quality control solutions .................................................................................. 5-11 Unscheduled AutoCheck Measurement............................................................................. 5-7 Unscheduled Calibrations .................................................................................................. 6-4 Updated calibrations .......................................................................................................... 6-5 User Activities ................................................................................................................. 3-50 User-defined Data Items .................................................................................................. 3-58 User-defined Notes .......................................................................................................... 3-61 V Viewing a calibration result ......................................................................................6-8, 9-18 Viewing quality control statistics..................................................................................... 9-13 Viewing replacement requirements ................................................................................... 7-2 Viewing the Reference Ranges and Critical Limits ......................................................... 3-19 W Westgard Rules program.................................................................................................. 3-43

Index

ABL800 FLEX Operators Manual

ABL800 FLEX Operator's manual

Date of issue

Date of issue

Radiometer representative:

Manufacturer:

If you have any questions or need assistance, please contact your local Radiometer representative.

Radiometer Medical ApS kandevej 21 2700 Brnshj Denmark www.radiometer.com

ABL800 FLEX operators manual - from software version 6.10

Publication: 201104 Edition: C Code number: 994-908

Date of issue

ABL800 FLEX Operator's manual

También podría gustarte